GW Instek AFG-100, AFG-200 User Manual

Arbitrary Function Generator Module
AFG-125/225/125P/225P
USER MANUAL
GW INSTEK PART NO. 82DS-23042EA1
ISO-9001 CERTIFIED MANUFACTURER
August 2014
This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Good Will company.
The information in this manual was correct at the time of printing. However, Good Will continues to improve products and reserves the rights to change specification, equipment, and maintenance procedures at any time without notice.
Good Will Instrument Co., Ltd. No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan
TABLE OF CONTENTS
Table of Contents
SAFETY INSTRUCTIONS ................................................... 4
GETTING STARTED ........................................................... 8
AFG-125/225/125P/225P Overview ........ 9
Panel Overview .................................... 12
DS2-FH1 Housing Installation ............. 14
Module Installation/Removal ............... 15
USB Configuration ............................... 17
AFG App Installation ........................... 18
Uninstalling the AFG App .................... 19
QUICK REFERENCE ......................................................... 20
Options Menu Tree .............................. 21
OPERATION .................................................................... 25
Basic Operation ................................... 26
Modulation .......................................... 40
Sweep .................................................. 50
Burst .................................................... 53
ARB ...................................................... 56
Power Supply Function ........................ 70
APPENDEX ...................................................................... 71
Dimensions ......................................... 71
Preset ARB Waveform .......................... 72
Declaration of Conformity .................... 80
INDEX............................................................................. 81
3
AFG-125/225 Series Options Manual
WARNING
Warning: Identifies conditions or practices that could result in injury or loss of life.
CAUTION
Caution: Identifies conditions or practices that could result in damage to the function generator or to other objects or property.
DANGER High Voltage
Attention: Refer to the Manual
Protective Conductor Terminal
Earth (Ground) Terminal DANGER Hot Surface
SAFETY INSTRUCTIONS
This chapter contains important safety instructions that should be followed when operating and storing the function generator. Read the following before any operation to ensure your safety and to keep the function generator in the best condition.
Safety Symbols
These safety symbols may appear in this manual or on the instrument.
4
SAFETY INSTRUCTIONS
Double Insulated
Do not dispose electronic equipment as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased.
General Guideline
CAUTION
Do not place heavy objects on the instrument. Do not place flammable objects on the
instrument.
Avoid severe impact or rough handling that
may damage the function generator.
Avoid discharges of static electricity on or near
the function generator.
Use only mating connectors, not bare wires, for
the terminals.
The instrument should only be disassembled by
a qualified technician.
Power Supply
WARNING
DC Input voltage: 5V/2A. Do not exceed an input voltage of 5V±5%.
Fuse
WARNING
Fuse type: F3.15A/125V. Only qualified technicians should replace the
fuse.
To ensure fire protection, replace the fuse only
with the specified type and rating.
Disconnect the power and all test leads before
replacing the fuse.
Make sure the cause of fuse blowout is fixed
before replacing the fuse.
Safety Guidelines
5
AFG-125/225 Series Options Manual
Cleaning the function generator
Disconnect the power cord before cleaning the
function generator.
Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid into the function generator.
Do not use chemicals containing harsh products
such as benzene, toluene, xylene, and acetone.
Operation Environment
Location: Indoor, no direct sunlight, dust free,
almost non-conductive pollution (Note below) and avoid strong magnetic fields.
Relative Humidity: < 80% Altitude: < 2000m Temperature: 0°C to 40°C
(Pollution Degree) EN 61010-1:2010 specifies pollution degrees and their requirements as follows. The function generator falls under degree 2.
Pollution refers to “addition of foreign matter, solid, liquid, or
gaseous (ionized gases), that may produce a reduction of dielectric strength or surface resistivity”.
Pollution degree 1: No pollution or only dry, non-conductive
pollution occurs. The pollution has no influence.
Pollution degree 2: Normally only non-conductive pollution
occurs. Occasionally, however, a temporary conductivity caused by condensation must be expected.
Pollution degree 3: Conductive pollution occurs, or dry, non-
conductive pollution occurs which becomes conductive due to condensation which is expected. In such conditions, equipment is normally protected against exposure to direct sunlight, precipitation, and full wind pressure, but neither temperature nor humidity is controlled.
Storage environment
Location: Indoor Relative Humidity: < 70% Temperature: -10°C ~ 70°C
6
SAFETY INSTRUCTIONS
Disposal
Do not dispose this instrument as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased. Please make sure discarded electrical waste is properly recycled to reduce environmental impact.
7
AFG-125/225 Series Options Manual
AFG-125/225/125P/225P Overview .................................... 9
Model Lineup ......................................................................... 9
Main Features ....................................................................... 10
Accessories ............................................................................ 11
Optional Accessories ........................................................... 11
Panel Overview ............................................................... 12
Front Panel ........................................................................... 12
Rear Panel ............................................................................. 13
DS2-FH1 Housing Installation ......................................... 14
Module Installation/Removal .......................................... 15
USB Configuration .......................................................... 17
AFG App Installation ....................................................... 18
Uninstalling the AFG App ................................................ 19
GETTING STARTED
This chapter gives a brief overview of how to install the AFG-125/225/125P/225P module onto the GDS-2000A.
8
GETTING STARTED
Model
AFG-125
AFG-125P
AFG-225
AFG-225P
Frequency Range
1uHz-25MHz
Output Channels
1 1 2
2
Power Output
None
Yes
None
Yes
AFG-125/225/125P/225P Overview
The AFG-125, AFG-225, AFG-125P and AFG-225P are arbitrary function generator modules for use with the GDS-2000A series DSOs. The options require the DS2-FH1 module extension bay to secure the module to the DSO. This user manual will explain how to become familiar with the arbitrary function generator modules. The Getting Started chapter will introduce the modules with an overview of the features and installation of the APP and accompanying PC software. The Operation chapter will go over the operation details.
Note1: The AFG-125/225/125P/225P are only supported with GDS-2000A series DSOs with firmware version V1.19 or above installed.
Note2: Throughout this manual AFG-125 will refer to both the AFG-125 and the AFG-125P, unless stated otherwise. Similarly, AFG-225 will refer to both the AFG-225 and AFG-225P, unless stated otherwise.
Model Lineup
9
Main Features
Performance
DDS signal generator 1μHz resolution over the full range 20ppm frequency stability Arbitrary Waveform Capability
120 MSa/s sample rate
60 MSa/s repetition rate
4 k-point waveform length
4k waveform memory, 10 groups
User-defined output
DWR (Direct waveform reconstruction) capability
PC waveform editing
Features
Sine, Square, Ramp, Pulse & Noise as standard
waveforms
Internal LIN/LOG sweeps with marker output AM, FM, FSK, SUM modulation Triggered burst function Save/recall 10 setup memories Output overload protection
Interface
USB interface as standard AWES (arbitrary waveform editing software) PC
software
Power Supply (AFG-125P/ 225P only
2.5V/3.3V/5V supply output 0.6A current output
AFG-125/225 Series Options Manual
10
GETTING STARTED
Part Number
Description
DS2-FH1
Module extension bay
GTL-254
USB A – USB A/B converter
Quick start guide
CD User manual
Part Number
Description
GPA-501
Power adapter
GTL-246
USB Type A – Type B cable
GTL-201A
Ground lead
Accessories
Optional Accessories
11
AFG-125/225 Series Options Manual
GND CH1 Output
Power Supply Status LEDs
CH2 Output
SYNC Output
Positive OutputNegative Output
C

CH CH


GD
VV V
FIED

WP
Power Supply Status LEDs
VV V
FIED
These LEDs indicate the immediate status of the power supply function on the AFG-125/225:
2.5V
2.5V output is on
3.5V
3.5V output is on
5V
5V output is on
OVER LOAD
Overload condition
Negative Output
8
GD
Negative output port
GND
Ground port
Positive Output
Positive output port
SYNC
C
Sync output. A TTL signal is output as the sweep marker or sync output signal. See page 50 for sweep details or page 34 for sync details.
CH1
CH

CH1 (Signal 1) output.
Panel Overview
Front Panel
12
CH2

CH
CH2 (Signal 2) output (AFG-225 only).
 MX
 V
INPUT
Vent USB Device Port Input Power
Vent
Cooling vent.
USB Device Port
Interfaces with the GDS-2000A and can also provide power.
Input Power
MX
V
INPUT
Input power source: DC 5V; 2A max.
Rear Panel
GETTING STARTED
13
AFG-125/225 Series Options Manual
Background
The DS2-FH1 consists of 2 housings that are attached to the feet on the underside of the case.
Note
Make sure the power is turned off before installing the AFG-125/225 module.
Steps
1. Slip the housing over the front of the feet on the
GDS-2000A.
2. Make sure that the rear tab clips securely over
the fan vent grid on the rear panel, as shown below.
1
2
3. The AFG-125/225 module is now ready to be
installed.
DS2-FH1 Housing Installation
14
GETTING STARTED
Background
The AFG-125/225 modules are installed into the area that is left between both of the DS2-FH1 housings.
Note
Make sure the power is turned off before installing the AFG-125/225 module.
Installation Steps
1. Slide the module into the slot that was created
between the DS2-FH1 housings. The front of the module should be facing forwards.
2. Make sure the module is secure. The module
will click into place when it is inserted properly.
3. Make sure the GDS-2000A is turned off before
proceeding.
4. Connect one end of the GTL-254 USB cable to
the rear panel USB Device port and to the USB Host port. Connect the other end to the Device port on the AFG-125/225, as shown below.
5. Turn the power back on. The AFG-125/225 will
now be accessible in the Option menu.
Module Installation/Removal
15
AFG-125/225 Series Options Manual
Removal
1. At the rear of the housings are two tabs. Pull
both tabs outwards.
2. The module can now be slid out from the
housing (back to front).
16
GETTING STARTED
Background
The USB Device port needs to be configured to provide power for the AFG-125/225 if an external power supply is not used.
Steps
1. As shown previously in the “AFG-125/225
Module Installation/Removal” section, connect
the GTL-254 USB cable.
2. Press the Utility key then I/O from
the bottom menu.
Utility
3. From the side menu press USB
Device Port and select USB Power.
The port will now supply power to the AFG-125/225.
Caution
The USB Device Port should be reconfigured to “Computer” or “Printer” when the AFG-125/225 is not used. Failure to do so may damage the PC or printer when connected in the “USB Power” mode.
USB Configuration
17
AFG-125/225 Series Options Manual
Background
Install the AFG app as you would any application for the GDS-2000A series.
The AFG app file (AFG.gz) can be found on the User Manual CD under the APP directory.
For the latest files and information for applications, see the GW Instek website: www.gwinstek.com or contact your nearest distributor.
Steps
1. Make sure the DS2-FH1 and AFG-125/225 are
installed and turned on.
Panel Operation
2. Insert a USB flash drive that has
the AFG app file (AFG.gz) copied onto it into the front panel USB port.
Demo
3. Press the Utility key then the File
Utilities soft-key.
Utility
4. Navigate to the desired file in the
USB file path.
When the AFG.gz file has been found, press the Select key twice to start the installation.
VARIABLE

AFG App Installation
18
GETTING STARTED
5. The installation will complete in a few seconds.
When finished a pop-up message will appear asking you to restart the GDS-2000A.
6. Restart the GDS-2000A.
Background
The AFG app can be uninstalled from the TEST menu, like the other optional apps.
Panel Operation
1. Press the Test key.
T
1. Press the APP button from the
bottom menu.
2. Use the Variable knob to select the
AFG app.
VRIBL
Up Down
2. Press Uninstall on the side menu
twice. Once to select and once to confirm uninstallation.
3. The uninstallation process is complete when a
message showing “Please turn off the oscilloscope and turn on again” appears.
Uninstalling the AFG App
19
AFG-125/225 Series Options Manual
Options Menu Tree ......................................................... 21
Option Key ........................................................................... 21
Arbitrary Function Generator ............................................ 22
Arbitrary Function Generator – Signal 1/2 Setup ........... 23
Arbitrary Function Generator – MOD part II ................ 24
Power Supply ........................................................................ 24
QUICK REFERENCE
This chapter describes the menu tree for the AFG­125/225.
20
QUICK REFERENCE
Accesses the functions in the Option menu.
Function
Gen 1
Function
Gen 2
I/O
L
z
O
Logic Analyzer Function
Gen 1
Goes to the Logic Analyzer menu Goes to the Function Generator menu
for the GEN 1 output
Function
Gen 2
Goes to the Function Generator menu for the GEN 2 output
I/O
Goes to the options I/O menu
DVM
AFG
AFG
Goes to the options AFG menu
*Note: Any option that is not installed/turned on will be grayed-out.
Options Menu Tree
Option Key
21
AFG-125/225 Series Options Manual
Set up the AFG-225/125 arbitrary function generator.
Note: The Power Supply function is only available for the AFG-125P and AFG-225P
The Dual Chan and Sync Setup functions are only available for the AFG-225 and AFG-225P.
O
Signal 1 Setup
 UTIL
Signal 1 Signal 2
Select
ON
O
On Off
Output
50
High Z
Load
-180.0º~180.0º
Phase
_P
Signal 1 Setup
Signal 2 Setup
Goes to the Signal 1/2 Setup menu
Signal 1 Setup
P

On Off
Ampl Cpl

OFF
ON
Inverted
Tracking
_P
Bk
OFF
Offset
Ratio
Freq Cpl Type
-999.99Hz ~
24.999MHz
Offset
0.001 ~
1000.000
Ratio

On Off
Sync
Signal 1 Signal 2
Source
Carrier Marker
Mode
Normal
Inverted
Polarity
P
Power Supply
Goes to the Power Supply menu
Arbitrary Function Generator
22
QUICK REFERENCE
Setup the signal type for each channel. From the AFG Arbitrary Function Generator menu on page 22.
Waveform Mode
MO 
ARB
Sine
Square
Pulse
Rm
N
On Off
ARB
L

Om
M1
On Off
MOD
M
M
K
M1
On Off
Sweep
B
On Off
Burst
N
User
1μHz ~ 60MHz
Frequency
1mVpp to
2.5Vpp
(into 50Ω)
2mVpp to 5Vpp
(open-circuit)
Amplitude
User
2μHz ~ 120MHz
Rate
Linear
Log
Type
M
M
Mk
Go Back
1~65535 cycles
Cycles
Infinite
-360º ~ 360º
Phase
1ms ~ 500
seconds
Period
TRI
INT
Manual Trigger
0~655350 nS
Delay
Bk
PM
SUM
1μHz~ 25MHz 1μHz~ 15MHz
(Square)
1μHz~ 1MHz
(Ramp)
Start
1μHz~ 25MHz 1μHz~ 15MHz
(Square)
1μHz~ 1MHz
(Ramp)
Stop
M1
0.001 ~ 500.000 Sec
SWP Time
1μHz ~ 25MHz 1μHz ~ 15MHz
(Square)
1μHz ~ 1MHz
(Ramp)
Span
450Hz ~ 25MHz 450Hz ~ 15MHz
(Square)
450Hz ~ 1MHz
(Ramp)
Center
±1.25Vpk ac+dc
(into 50Ω)
±2.5Vpk ac+dc
(Open circuit)
Offset
On Off
Marker
Bk
1μHz ~ 25 MHz
1μHz ~ 1 MHz
(Ramp)
Frequency
AM
FM
Goes to the Signal 1 Setup menu -MOD part II
FSK
PM
SUM
Arbitrary Function Generator – Signal 1/2 Setup
23
AFG-125/225 Series Options Manual
Configures AM, FM, FSK, PM or SUM modulation. From the Arbitrary Function Generator – Signal 1/2 Setup – MOD menu on page 23.
UM
Bk
0.0% ~ 100.0%
SUM Ampl
0.002Hz ~ 20kHz
SUM Freq
Sine
Square
Triangle UpRamp DnRamp
Shape
M
Bk
0.0% ~ 120.0%
Depth
0.002Hz ~ 20kHz
AM Freq
Sine
Square
Triangle UpRamp DnRamp
Shape
M
Bk
0.002Hz ~ 20kHz
FM Freq
Sine
Square Triangle UpRamp DnRamp
Shape
K
Bk
0.002Hz ~ 100kHz
FSK Rate
Bk
0.0º ~ 360º
Phase Dev
0.002Hz ~ 20kHz
PM Freq
Sine
Square Triangle UpRamp DnRamp
Shape
PM
DC~25MHz DC~15MHz
(square)
DC~1MHz
(Ramp)
Freq Dev
1 μHz~25MHz
(sine)
1 μHz~15MHz
(Square, Pulse)
1 μHz~1MHz
(Ramp)
Hop Freq
Power supply menu for the AFG-125P and AFG-225P.
P
On Off
Power
2.5V
3.3V 5V
Voltage
Arbitrary Function Generator – MOD part II
Power Supply
24
OPERATION
Basic Operation .............................................................. 26
Entering the AFG Menu ..................................................... 26
State Display ......................................................................... 27
Selecting a Basic Waveform ................................................ 28
Turning the Output On ...................................................... 29
Dual Channel Tracking ....................................................... 31
Recalling the Preset Settings ............................................... 33
Sync Setup ............................................................................. 34
Modulation ..................................................................... 40
AM Modulation .................................................................... 40
FM Modulation .................................................................... 42
FSK Modulation ................................................................... 44
PM Modulation .................................................................... 46
SUM Modulation .................................................................. 48
Sweep ............................................................................. 50
Sweep ..................................................................................... 50
Burst ............................................................................... 53
Burst Waveform ................................................................... 53
ARB ................................................................................. 56
Loading a Preset ARB Waveform ...................................... 56
Creating/Editing a Preset ARB Waveform ...................... 59
Setting ARB Output Length ............................................... 67
Setting ARB Frequency, Rate, Amplitude & Offset ........ 68
OPERATION
25
AFG-125/225 Series Options Manual
Background
Like the other options for the GDS-2000A, the AFG option can be accessed via the front panel option key.
Panel Operation
1. Press the Option key.
O
2. Press AFG from the bottom menu
to enter the first level of the AFG menu. (The model type will be shown below the icon)
AFG Menu
Basic Operation
Entering the AFG Menu
26
State Display
Background
The state display function provides a visual display of the status of each channel and the function that is used.
The state display is on by default.
The state display is also used as a visual guide when editing/creating an ARB waveform.
State Display
Signal1
Status
Signal2 Status
Signal1
Function
Settings
Signal2 Function Settings
Signal1 Function
Visual Display
Signal2 Function Visual Display
Panel Operation
1. From the first level AFG menu,
press State Disp to toggle the display pop-up on or off.
OPERATION
27
AFG-125/225 Series Options Manual
Background
The AFG-125/225 can output 5 different types of basic waveforms: Sine, Square, Pulse, Ramp, Noise.
The basic waveforms can be output as is or they can be used as the baseband waveforms for the modulation functions or as the primary waveforms for the sweep and burst functions.
Connection
1. The AFG-125/225 arbitrary function has 2
output channels, CH1 and CH2. Each output can be selected individually.
2. Connect a BNC cable
to the appropriate output (CH1 or CH2).
Panel Operation
1. From the first level AFG menu,
select Signal 1 Setup or Signal 2 Setup from the bottom menu to select the output channel.
2. Press Waveform Mode from the
bottom menu and then select a waveform from the side menu.
Waveforms
Sine, Square, Pulse Ramp, Noise
3. See page 29 to set the output
settings and turn the output on.
Page 29
Selecting a Basic Waveform
28
OPERATION
Background
The AFG-125/225 can output 5 different types of basic waveforms: Sine, Square, Pulse, Ramp, Noise.
The basic waveforms can be output as is or they can be used as the baseband waveforms for the modulation functions or as the primary waveforms for the sweep and burst functions.
Connection
1. The AFG-125/225 arbitrary function has 2
output channels, CH1 and CH2. Each output can be selected individually.
2. Connect a BNC cable
to the appropriate output (CH1 or CH2).
Steps
1. Press the Output Setup key to set
the amplitude.
2. From the side menu choose the
output parameters
Choose the signal that will be output:
Signal 1: Signal 2:
CH1 CH2
Turns the output of the selected signal on/off.
Sets the output impedance to 50Ω or High Z.
Turning the Output On
29
AFG-125/225 Series Options Manual
Sets the output phase relative to 0º.
Synchronizes the phase of both channels on dual channel models.
The signal will be output as soon as the output is turned on.
30
Background
As only the AFG-225 has dual channels, it has a number of tracking functions not available on the AFG-125. The tracking functions are listed below.
Dual Channel Functions
Function
Description
Frequency Coupling
Frequency coupling will couple the channel outputs by frequency. Frequency coupling can be achieved by frequency offset or by a frequency ratio.
Amplitude Coupling
Amplitude coupling will couple the both channel outputs by amplitude. Any changes in amplitude in one channel is reflected in the other channel.
Tracking
There are two tracking modes, ON and Inverted. When turned on, tracking performs frequency and amplitude tracking so the channels behave as a single channel. When set to inverted, the one channel will output the inverse of the other.
S_Phase
S_Phase will synchronize the phase of both channels.
Steps
1. From the first level AFG menu,
press UTIL and then press Dual Chan on the side menu.
OPERATION
Dual Channel Tracking
31
AFG-125/225 Series Options Manual
2. From the side menu choose the
relevant tracking function(s):
Turns frequency coupling on and sets the type of frequency coupling:
Offset: Sets frequency coupling as an offset.
Ratio: Sets frequency coupling as a ratio.
Turns amplitude coupling on or off. Ampl Cpl:
OFF, ON
Turns tracking on or off, or turns tracking on and inverts the output of the second output.
Tracking:
OFF, ON, INVERTED
Press S_Phase to synchronize the phase of the both channels.
32
OPERATION
Background
Recalling the preset settings will effectively reset the settings to the factory defaults.
Preset Settings
Waveform Mode
Sine
Frequency
1kHz
Amplitude
1.000Vpp
Offset
0.0Vdc
Phase
0.0º
Load
50
Output
Off
ARB
Off
MOD
Off
Sweep
Off
Burst
Off
Frequency Coupling*
Off
Amplitude Coupling*
Off
Tracking*
Off
*Only applicable to the AFG-225.
Steps
1. From the first level of the AFG
menu, press UTIL.
2. Press Preset. The preset settings
will be recalled straight away.
Recalling the Preset Settings
33
Sync Setup
Background
The sync output signal is output from the SYNC port on the front panel. The sync output signal is based on either the channel 1 or channel 2 output signals. Each periodic type of waveform output function has an associated sync output signal.
The characteristics of the sync output depend on the selected signal source.
Sync Signal Output
Sine, Square, Pulse, Ramp Source
For these types of waveforms, the sync output is a square wave pulse. The pulse is high for the positive transition of the waveform and low for the negative transition. The signal is a TTL level signal.
Sine output
SYNC output
0V
0V
AFG-125/225 Series Options Manual
34
OPERATION
MOD waveform Source
The sync output is a positive pulse for the positive transition of the modulating waveform and low for the negative transition. The signal is a TTL level signal.
FM output
SYNC output
0V
0V
Arbitrary Waveform Source
The ARB function can output either carrier or marker signals.
Carrier: The sync output is a single TTL positive pulse for the positive transition of the ARB waveform and low for the negative transition.
Marker: A single TTL level positive pulse is output at the start of each period of the ARB waveform.
35
AFG-125/225 Series Options Manual
Carrier
ARB output
SYNC output
0V
0V
Marker
ARB output
SYNC output
0V
0V
Sweep Source
The sweep function can output either carrier or marker signals.
Carrier: The sync signal output is a positive TTL level pulse for the positive transition of the sweep waveform and low for the negative transition.
Marker: The marker signal output is a TTL level positive
pulse from the “marker”
frequency to the stop frequency and a low level signal from the start frequency to the “marker” frequency. See page 50 to set the marker settings.
36
OPERATION
Carrier
Sweep output
SYNC output
0V
0V
Marker
Sweep output
SYNC output
0V
0V
Marker
frequency
Burst Source
Like the sweep function, the burst function can output Carrier or Marker signals from the SYNC output.
Carrier: The sync signal output is a positive TTL level pulse for the positive transition (0V and above) of the burst waveform and low for the negative transition.
Marker: The marker signal output is a TTL level positive pulse for the duration of the burst period and low for the remainder of waveform period
37
AFG-125/225 Series Options Manual
Carrier
FSK output
SYNC output
0V
0V
Marker
FSK output
SYNC output
0V
0V
Steps
1. From the first level of the AFG
menu, press UTIL and then press Sync Setup on the side menu.
2. From the side menu choose the
sync setting:
The Sync function will turn the sync output on for all waveforms except for noise.
(Note: If Marker is turned on in the Sweep function, the Sync On setting will be overridden.
Selects the source signal on which to base the sync signal.
Source:
Signal 1, Signal 2
When Sync is on, You can force the sync signal to follow the carrier or the marker* mode.
Mode:
Carrier, Marker
38
OPERATION
The polarity can be set to Normal or Inverted.
Polarity:
Normal, Inverted
Note
The marker mode can only be used for the ARB and mod waveforms.
For the Sweep function, the marker output must be turned on in the Sweep function menu. It cannot be activated in the Sync Setup menu. See page 51 to turn the marker option on.
39
Modulation
Background
The AM Modulation function can set the carrier frequency (AM Freq), amplitude modulation depth (Depth) and carrier waveform (Shape).
Steps
1. Set the carrier waveform and
frequency. You can set the carrier waveform and frequency in the Selecting a Basic Waveform section.
Page 26
2. From the first level of the AFG
menu, press the Signal 1 Setup or Signal 2 Setup key to select the signal source.
3. From the bottom menu press
MOD.
4. From the side menu turn MOD On
and then select the AM modulation.
Note
Only one modulation mode can be active at any one time. The function generator also will not allow sweep, burst or ARB mode to be used with a modulation function. Activating a modulation mode will turn the previous modulation mode off.
5. Select the AM parameters from the side menu:
AM Modulation
AFG-125/225 Series Options Manual
40
OPERATION
Sets the amplitude modulation depth: 0~ 120%
Sets the AM frequency: 2 mHz~20kHz.
Sets the carrier waveform shape: Sine, Square, Triangle, Upramp, Dnramp.
6. See page 29 to set the output
settings and turn the output on.
Page 29
Example
Amplitude Modulation
41
FM Modulation
Background
The FM Modulation function can set the carrier frequency (FM Freq), frequency deviation (FM DEV) and carrier waveform (Shape).
Steps
1. Set the carrier waveform and
frequency. You can set the carrier waveform and frequency in the Selecting a Basic Waveform section.
Page 26
2. From the first level of the AFG
menu, press the Signal 1 Setup or Signal 2 Setup key to select the signal source.
3. From the bottom menu press
MOD.
4. From the side menu turn MOD On
and then select the FM modulation.
Note
Only one modulation mode can be active at any one time. The function generator also will not allow sweep, burst or ARB mode to be used with a modulation function. Activating a modulation mode will turn the previous modulation mode off.
5. Select the FM parameters from the side menu:
AFG-125/225 Series Options Manual
42
OPERATION
Sets the frequency deviation: DC~25MHz DC~15MHz(square) DC~1MHz (Ramp)
Sets the FM frequency: 2mHz~20kHz.
Sets the carrier waveform shape: Sine, Square, Triangle, Upramp, Dnramp.
6. See page 29 to set the output
settings and turn the output on.
Page 29
Example
Frequency Modulation
43
AFG-125/225 Series Options Manual
Background
The FSK Modulation function can set the hop frequency (FSK Hop) and the frequency-shift keying rate (FSK Rate). FSK Modulation essentially jumps between the carrier frequency and the hop frequency at a rate determined by the FSK rate frequency.
Steps
1. Set the carrier waveform and
frequency. You can set the carrier waveform and frequency in the Selecting a Basic Waveform section.
Page 26
2. From the first level of the AFG
menu, press the Signal 1 Setup or Signal 2 Setup key to select the signal source.
3. From the bottom menu press
MOD.
4. From the side menu turn MOD On
and then select the FSK modulation.
Note
Only one modulation mode can be active at any one time. The function generator also will not allow sweep, burst or ARB mode to be used with a modulation function. Activating a modulation mode will turn the previous modulation mode off.
5. Select the FSK parameters from the side menu:
FSK Modulation
44
OPERATION
Sets the hop frequency: 1μHz~25MHz (Sine) 1μHz~15MHz (Square, Pulse) 1μHz~1MHz (Ramp)
Sets the FSK rate: 2mHz~100kHz.
6. See page 29 to set the output
settings and turn the output on.
Page 29
Example
Frequency-Shift Keying Modulation
45
PM Modulation
Background
The PM Modulation function can set the phase modulation frequency (PM Freq), phase deviation (Phase Dev) and the PM shape (Shape).
The phase deviation of the carrier waveform deviates from a reference phase value in proportion to changes in the modulating waveform.
Steps
1. Set the carrier waveform and
frequency. You can set the carrier waveform and frequency in the Selecting a Basic Waveform section.
Page 26
2. From the first level of the AFG
menu, press the Signal 1 Setup or Signal 2 Setup key to select the signal source.
3. From the bottom menu press
MOD.
4. From the side menu turn MOD On,
press more 1 of 2 and then select the PM modulation.
Note
Only one modulation mode can be active at any one time. The function generator also will not allow sweep, burst or ARB mode to be used with a modulation function. Activating a modulation mode will turn the previous modulation mode off.
AFG-125/225 Series Options Manual
46
OPERATION
5. Select the PM parameters from the side menu:
Sets the phase deviation: 0º ~ 360º
Sets the phase modulation frequency: 2mHz ~ 20kHz
Sets the PM wave shape: Sine, Square, Triangle, UpRamp, DnRamp.
6. See page 29 to set the output
settings and turn the output on.
Page 29
Example
Pulse Modulation
47
AFG-125/225 Series Options Manual
Background
The SUM Modulation function can set the SUM amplitude (SUM Ampl) the carrier frequency (SUM Freq) and carrier waveform (Shape).
In general SUM modulation adds a modulating signal to a carrier wave. Typically, sum modulation is used to add noise to a carrier wave. The modulating signal is added as a percentage of the carrier amplitude.
Steps
1. Set the carrier waveform and
frequency. You can set the carrier waveform and frequency in the Selecting a Basic Waveform section.
Page 26
2. From the first level of the AFG
menu, press the Signal 1 Setup or Signal 2 Setup key to select the signal source.
3. From the bottom menu press
MOD.
4. From the side menu turn MOD On,
press more 1 of 2 and then select the SUM modulation.
Note
Only one modulation mode can be active at any one time. The function generator also will not allow sweep, burst or ARB mode to be used with a modulation function. Activating a modulation mode will turn the previous modulation mode off.
SUM Modulation
48
OPERATION
5. Select the SUM parameters from the side menu:
Sets the SUM amplitude (relative to the carrier): 0.0% ~ 100.0%
Sets SUM modulation frequency: 2mHz ~ 20kHz
Sets the SUM wave shape.
6. See page 29 to set the output
settings and turn the output on.
Page 29
Example
SUM Modulation
49
Sweep
Background
The AFG-125/225 can perform a sweep for sine, square or ramp waveforms. In Sweep mode, the function generator will sweep from a start frequency to a stop frequency. The sweep can be performed in a linear or logarithmic fashion. The AFG-125/225 can also output a TTL signal from the SYNC output when the marker function is turned on.
Steps
1. Set the carrier waveform and
frequency. You can set the carrier waveform and frequency in the Selecting a Basic Waveform section.
Page 26
2. From the first level of the AFG
menu, press the Signal 1 Setup or Signal 2 Setup key to select the signal source.
3. From the bottom menu press
Sweep.
4. From the side menu turn Sweep
On.
Note
Only one modulation mode can be active at any one time. The function generator also will not allow sweep, burst or ARB mode to be used with a modulation function. Activating a modulation mode will turn the previous modulation mode off.
Sweep
AFG-125/225 Series Options Manual
50
OPERATION
5. Select the sweep parameters from the side
menu:
Selects the type of sweep: logarithmic or linear.
Sets the start frequency*:
1μHz ~ 25MHz 1μHz ~ 15MHz (Square) 1μHz ~ 1MHz (Ramp)
Sets the Stop frequency*:
1μHz ~ 25MHz 1μHz ~ 15MHz (Square) 1μHz ~ 1MHz (Ramp)
Press to access the remaining sweep settings, shown below:
Sets the sweep time: 1ms ~ 500s.
Sets the Span*:
1μHz ~ 25MHz 1μHz ~ 15MHz (Square) 1μHz ~ 1MHz (Ramp)
Sets the center frequency*.
Note
*Setting the Start and Stop frequencies will override the Span and Center settings, and conversely setting the Span and Center settings will override the Start and Stop settings.
Marker
6. To set a marker signal, press
Marker and turn Marker On. The marker will output a TTL-level square wave signal at the designated frequency from the SYNC output port.
51
AFG-125/225 Series Options Manual
7. Press Frequency and set a marker
frequency. The marker frequency cannot be lower than the center frequency.
1μHz ~ 25MHz 1μHz ~ 1MHz (Ramp)
8. See page 29 to set the output
settings and turn the output on.
Page 29
Example
Sweep Waveform (CH1) with marker output (CH2).
52
OPERATION
Background
The function generator can create a waveform burst with a designated number of cycles. Burst mode supports sine, square and ramp waveforms.
Steps
1. Set the carrier waveform and burst
frequency. You can set the carrier waveform and burst frequency in the Selecting a Basic Waveform section.
Page 26
2. From the first level of the AFG
menu, press the Signal 1 Setup or Signal 2 Setup key to select the signal source.
3. From the bottom menu press Burst.
4. From the side menu turn Burst On.
Note
Only one modulation mode can be active at any one time. The function generator also will not allow sweep, burst or ARB mode to be used with a modulation function. Activating a modulation mode will turn the previous modulation mode off.
5. Press N Cycle from the side menu.
Burst
Burst Waveform
53
AFG-125/225 Series Options Manual
6. Select the N Cycle parameters from the side
menu:
Sets the number of burst cycles* : 1~ 65,535, infinite.
Sets the starting phase of the burst waveform.
Sets the period* length in seconds for the burst waveform.
Note
*The Period setting will affect the number of cycles that can be chosen for any given frequency. The relationship is shown in the following equation:
Burst Cycle < (Burst Period x Wave Frequency)
Trigger Settings
7. Press Trigger set from the side
menu.
8. Select the Trigger parameters from the side
menu:
Sets the trigger as internal. An internal trigger is generated at the end of each period.
Sets the trigger to Manual mode. This Manual Trigger key must be pressed each time to output a burst waveform.
Sets a delay time between the trigger and the burst waveform output: 0 ~ 655350nS.
9. See page 29 to set the output
settings and to turn the output on.
Page 29
54
OPERATION
Example
Burst Waveform
55
AFG-125/225 Series Options Manual
Background
The AFG-225 can load a number of different waveforms as an arbitrary waveform file: CH1~CH4, Ref1~Ref4, Wave1~Wave20, a previously saved waveform (LSF and fast CSV format*), a previously created ARB waveform or one of the 66 preset ARB waveforms**. The preset waveforms are categorized into Common, Math, Window and Engineer waveforms, as shown in the table below.
* See the DSO user manual for save file formats. ** See the appendix on page 72 for a full list and description of the preset ARB waveforms.
Preset waveforms
Common:
Absatan, Abssin, Abssinehalf, Ampalt, Attalt, Direven, Diriodd, Gauspuls, Havercos, Haversin, Negramp, N_pulse, Rectpuls, Roundhaf, Sawtooth, Sinetra, Sinever, Stair_dw, Stair_ud, Stair_up, stepresp, trapezia, tripuls
Math:
Arccos, Arccot, Arccsc, Arcsec, Arcsin, Arcsinh, Arctan, Arctanh, Cosh, Cot, Csc, Dlorentz, Expofall, Exporise, Gauss, Ln, Lorentz, Sec, Sech, Sinec, Sinh, Sqrt, Tan, Tanh, Xsquare
Window:
Barthwin, Bartlett, Blackman, Bohmwin, Chebyshe, Flattwin, Hamming, Hann, Hanning, Kaiser, Triang, Tukeywin
Engineer:
ARB
Loading a Preset ARB Waveform
56
OPERATION
Airy, Bessel, Betainc, Gamm, Legendre, Neumann
Steps
1. From the first level of the AFG
menu, press the Signal 1 Setup or Signal 2 Setup key to select the signal source.
2. From the bottom menu press ARB.
3. From the side menu turn ARB On.
Note
The function generator also will not allow ARB to be turned on when sweep, burst or a modulation function is active. Turn off any other active modes before using ARB mode.
4. Press Load from the side menu.
5. You can load an ARB waveform from one of the
source channels, from internal memory or from a number of predefined waveforms:
Press From to select an input channel or internally saved waveform.
From:
CH1 ~ CH4 Ref1 ~ Ref4 Math Wave1 ~ Wave20
Press From File to load a previously saved waveform or a predefined waveform.
57
AFG-125/225 Series Options Manual
From File:
LSF, Fast CSV, User Preset, Common, Math, Window, Engineer
If a previously saved waveform or predefined waveform type was chosen above, use the File Utilities to select the desired ARB waveform.
Press Recall Now to load the ARB waveform.
A message will appear on the screen when the ARB waveform is successfully recalled.
Example
Triangle Pulse waveform
58
OPERATION
Background
The AFG-125/225 has a number of methods to create and edit arbitrary waveforms. The ARB waveforms can have a maximum of 4096 (0~4095) points in length and 1023 (0±511) points in amplitude.
Editing Methods
Point/Line: Creates a vertical line of a user-defined length and amplitude.
Diagonal: Creates a diagonal line of a user-defined length and amplitude.
Scale: Scales the full-scale of the ARB waveform.
Copy/Paste: This method will copy and paste a user-defined section of the ARB waveform.
Clear: This will clear a user-defined section of the ARB waveform.
Note: By default, any section of the ARB waveform that is not edited will be set to 0Vpp. By default the editing will automatically match the amplitude of each point.
Steps
1. From the first level of the AFG
menu, press the Signal 1 Setup or Signal 2 Setup key to select the signal source.
Creating/Editing a Preset ARB Waveform
59
AFG-125/225 Series Options Manual
2. From the bottom menu press ARB.
3. From the side menu turn ARB On.
Note
The function generator also will not allow ARB to be turned on when sweep, burst or a modulation function is active. Turn off any other active modes before using ARB mode.
4. Press Edit from the side menu.
Point/Line Editing
5. Press Edit Method and choose Point
Line.
6. Press Action to begin editing the
waveform using the Point/Line method.
7. Use the side menu to create a point or vertical
line:
Press Address to set the starting address of the line or point.
The maximum address depends on the Length setting below:
Max address = 4096 – Length
Range:
0 ~ 4095
Press Length to set the length of the line. The maximum length depends on the address set above:
Max length = 4096 – Address
Length
1 ~ 4096
60
OPERATION
Press Data to set the amplitude of the point/line.
Data
-511 ~ 0 ~ +511
When the point/line has been edited, press Preview to view the edited line.
Press Undo to cancel the point/line edit.
Press Done to confirm and save the point/line edit.
Example
Figure showing a horizontal line (black highlight)
Diagonal Editing
1. Press Edit Method and choose
Diagonal.
2. Press Action to begin editing the
waveform using the Diagonal method.
3. Use the side menu to create a diagonal line:
61
AFG-125/225 Series Options Manual
Press Addr1/Data1 to set the first data point for the diagonal line. Press once to set the Addr1, press again to set Data1 (amplitude).
Addr1 range:
0 ~ 4095
Data1 range:
±511
Press Addr2/Data2 to set the second data point for the diagonal line. Press once to set the Addr2, press again to set Data2 (amplitude).
Addr2 range:
0 ~ 4095
Data2 range:
±511
When the line has been edited, press Preview to view the edited line.
Press Undo to cancel the line edit.
Press Done to confirm and save the line edit.
Example
Figure showing a diagonal line (black highlight).
62
OPERATION
Scale Editing
1. Press Edit Method and choose Scale.
2. Press Action to begin editing the
waveform using the Scale method.
3. Use the side menu to set the scale:
The scale function sets the vertical scaling factor* for the whole ARB waveform.
Scale
0.1 ~ 10X
Example
Figure of 2.0X scaling. Notice that the waveform is clipped at the bottom.
Note
* If any data points that exceed a magnitude of ±511 will be clipped.
Copy / Paste
1. Press Edit Method and choose
Copy /Paste.
63
AFG-125/225 Series Options Manual
2. Press Action to enter the
Copy/Paste menu.
3. Use the side menu to copy a section of the ARB
waveform:
Press Start* to set the start address of the section you want to copy.
Start range:
0 ~ 4095
Press Length* to set the length of the section you wish to copy.
Length range
1~ 4096
Press Paste To to choose the position to paste the copied section of waveform**.
Paste range
0~ 4095
When the copy and paste areas have been selected, press Preview to view the result.
Press Undo to cancel the copy/paste operation.
Press Done to confirm and save the copy/paste operation.
64
OPERATION
Example
Figure showing the copied section (black) and the pasted section (red bars).
Note
* The maximum start address depends on the Length setting: Max Start = 4096 – Length
The maximum length depends on the start address set above: Max Length = 4096 – Start
** You cannot paste over an area that was originally copied. For example, if you copy from points 100 ~ 200, you cannot paste that to points 150 ~ 250. A setting conflict error will appear on the screen when the “Copy” and “Paste To” area overlap.
Clear
1. Press Edit Method and choose
Clear.
2. Press Action to enter the Clear
menu.
3. Use the side menu to select a section of the ARB
waveform to clear or to clear the whole ARB waveform:
65
AFG-125/225 Series Options Manual
Sets the start address of the section of the ARB waveform to be cleared.
Start range
0 ~ 4095
Sets the length of the section of the ARB waveform to be cleared.
Length range
1 ~ 4096
Press Done to confirm and then clear the selected section of the ARB waveform.
Press Done to clear the whole ARB waveform instantly.
Example
Figure showing the cleared section of the waveform.
Saving the ARB Waveform
After you have created/edited your waveform, you can save it for use at a later time. See page 56 to load a User Preset ARB waveform.
1. Press Save Now to save the ARB
waveform to the current directory.
66
OPERATION
Background
The Output Confirm menu allows the user to specify a section of the ARB waveform to be output.
The length of the outputted section will directly affect the possible frequency or rate of the ARB waveform. See page 68 for details.
Steps
1. From the ARB menu press Output
Confirm.
2. Select the section of the ARB waveform that
you wish to have output from the side menu:
Press Start* to set the start address for the output section.
Press Length* to set the total length of the output section.
Example
Output section is shown in black.
Setting ARB Output Length
67
AFG-125/225 Series Options Manual
Note
* The maximum length depends on the start address setting.
The maximum length = 4096 – Start address.
3. Press Confirm to set the chosen
output section.
4. See page 29 to set the output
settings and turn the output on.
Page 29
Background
The rate, frequency, amplitude and (amplitude)offset output parameters can also be set.
Steps
1. From the ARB menu press more 1
of 2.
2. Select the parameters from the side menu:
Sets the frequency* of the ARB waveform.
Sets the amplitude** of the waveform: 1mVpp to 2.5Vpp (into 50Ω) 2mVpp to 5Vpp (open-circuit)
Sets the Offset** of the ARB waveform: ±1.25Vpk ac +dc (into 50Ω) ±2.5Vpk ac +dc (Open circuit)
Setting ARB Frequency, Rate, Amplitude & Offset
68
OPERATION
Sets the rate of the ARB waveform*. This will set the number of times the ARB waveform will be output each second.
Note
*The frequency setting is inversely proportional to the Rate setting. The frequency and rate settings are also directly affected by the length of the output ARB waveform. See page 67 for setting the output length of the ARB waveform.
**The amplitude and offset are linked. Together the amplitude and offset cannot exceed 2.5Vpp (into 50Ω).
3. See page 29 to set the output
settings and turn the output on.
Page 29
69
AFG-125/225 Series Options Manual
Background
The AFG-125P and AFG-225P have an additional power supply function. The power supply has three fixed output levels: 2.5V, 3.3V, 5V.
Steps
1. Press Power Supply from the first
level of the AFG menu.
2. Press Power to turn the power
supply output on.
3. Press Voltage to select one of the
fixed output levels. Status LEDs on the front panel will light up appropriately.
Voltage
2.5V, 3.3V, 5V
Note
The current power supply has a current output of 6A and a voltage output of (2.5V, 3.3V or 5V)±5%.
If the power supply is overload, the OVER LOAD LED on the front panel will light up.
Power Supply Function
Using the Power Supply Function (AFG-125P & AFG-225P only)
70
APPENDEX
106.9
30.7
223.6
3.5
APPENDEX
Dimensions
71
AFG-125/225 Series Options Manual
Function
Description
Function
Description
ABSATAN
y=|atan(x)| The absolute of atan(x)
RECTPULS
Sampled aperiodic rectangle
ABSSIN
y=|sin(x)| The absolute of sin(x)
ROUNDHAF
y=sqrt(1-x^2) Half round function
ABSSINEH
y=sin(x),0<x<pi y=0,pi<x<2pi Half- wave function
SAWTOOT
Sawtooth or triangle wave
AMPALT
y=e(x).sin(x) Oscillation rise
SINETRA
Piecewise function
Preset ARB Waveform
Common
72
APPENDEX
ATTALT
y=e(-x).sin(x) Oscillation fall
SINEVER
Piecewise function
DIRIEVEN
Even f(x)=-1^(x*(n-1)/2*pi) x=0,±2*pi,±4*pi,……
STAIR_DW
Step down
DIRIODD
Odd f(x)=sin(nx/2)/n*sin(x/2) x=±pi,±3pi……
STAIR_UD
Step up and step down
GAUSPULS
f(x)=a*e^(-(x-b)^2)/c^2) Gaussian-modulated sinusoidal pulse
STAIR_UP
Step up
HAVERCOS
y=(1-sin(x))/2 Havercosine function
STEPRESP
Heaviside step function
73
AFG-125/225 Series Options Manual
HAVERSIN
=(1-cos(x))/2 Haversine function
TRAPEZIA
Piecewise function
NEGRAMP
y=-x Line segment
TRIPULS
Sampled aperiodic triangle
N_PULSE
Negative pulse
Function
Description
Function
Description
ARCCOS
Basic trigonometric function
EXPORISE
Exponential rise
Math
74
APPENDEX
ARCCOT
Basic trigonometric function
GAUSS
A waveform representing a gaussian bell curve
ARCCSC
Basic trigonometric function
LN
Logarithm function
ARCSEC
Basic trigonometric function
LORENTZ
The derivative of the lorentz function y=1/(k*x^2+1)
ARCSIN
Basic trigonometric function
SEC
Basic trigonometric function
ARCSINH
Basic trigonometric function
SECH
Basic trigonometric function
75
AFG-125/225 Series Options Manual
ARCTAN
Basic trigonometric function
SINEC
y=sin(x)/x
ARCTANH
Basic trigonometric function
SINH
Basic trigonometric function
COSH
Basic trigonometric function
SQRT
y=sqrt(x)
COT
Basic trigonometric function
TAN
Basic trigonometric function
CSC
Basic trigonometric function
TANH
Basic trigonometric function
76
DLORENTZ
The derivative of the lorentz function. y=-2x/(k*x^2+1)
XSQUARE
Parabola
EXPOFALL
Exponential decay
Window
Function
Description
Function
Description
BARTHWIN
Modified Bartlett-Hann window
HAMMING
The Hamming window function
BARTLETT
The Bartlett window is very similar to a triangular window as returned by the TRIANG function.
HANN
The Hann window function
APPENDEX
77
AFG-125/225 Series Options Manual
BLACKMAN
The Blackman window function
HANNING
The Hanning window function
BOHMWIN
The Bohman window function
KAISER
The Kaiser window function
CHEBYSHE
The Chebyshev window function
TRIANG
The Triang window function
FLATTWIN
The Flattopwin window function
TUKEYWIN
The Tukey window function
Function
Description
Function
Description
AIRY
The Airy function
GAMM
The Gamma function
Engineer
78
APPENDEX
BESSEL
The Bessel function
LEGENDRE
Associated Legendre functions
BETA
The Beta function
NEUMANN
The Neumann function
79
AFG-125/225 Series Options Manual
Type of Product: Arbitrary Function Generator (With DC Power Supply)
Model Number: AFG-125, AFG-225, AFG-125P & AFG-225P
EMC
EN 61326-1:
Electrical equipment for measurement, control and laboratory use –– EMC requirements (2013)
Conducted and Radiated Emissions EN 55011: 2009+A1: 2010
Electrostatic Discharge EN 61000-4-2: 2009
Current Harmonic EN 61000-3-2: 2006+A1: 2009+A2: 2009
Radiated Immunity EN 61000-4-3: 2006+A1: 2008+A2 :2010
Voltage Fluctuation EN 61000-3-3: 2008
Electrical Fast Transients EN 61000-4-4: 2012
-------------------------
Surge Immunity EN 61000-4-5: 2006
-------------------------
Conducted Susceptibility EN 61000-4-6: 2009
-------------------------
Power Frequency Magnetic Field EN 61000-4-8: 2010
-------------------------
Voltage Dips/Interrupts EN 61000-4-11: 2004
Declaration of Conformity
We
GOOD WILL INSTRUMENT CO., LTD.
No. 7-1, Jhongsing Rd, Tucheng Dist., New Taipei City 236, Taiwan
GOOD WILL INSTRUMENT (SUZHOU) CO., LTD.
No. 69 Lushan Road, Suzhou New District Jiangsu, China.
declare that the below mentioned products
are herewith confirmed to comply with the requirements set out in the Council Directive on the Approximation of the Law of Member States relating to Electromagnetic Compatibility (2004/108/EC). For the evaluation regarding the Electromagnetic Compatibility, the following standards were applied:
80
INDEX
INDEX
Accessories ................................. 11
AFG menu .................................. 26
App installation ......................... 18
App uninstallation ..................... 19
Appendix .................................... 71
ARB
Amplitude ..................................... 68
Clear................................................ 65
copy/paste ..................................... 63
diagonal line .................................. 61
Editing ............................................ 59
Frequency ...................................... 68
line .................................................. 60
Load waveforms ........................... 56
Offset .............................................. 68
Output ............................................ 67
point ................................................ 60
Preset waveform list .................... 72
Preset waveforms ......................... 56
Rate ................................................. 68
Save ................................................. 66
scale ................................................ 63
Baseband waveforms ................ 28
Basic waveforms ........................ 28
Burst ............................................ 53
Caution symbol ............................ 4
Cleaning the instrument ............. 6
Coupling ..................................... 31
Declaration of conformity ......... 80
Default settings .......................... 33
Disposal
symbol .............................................. 5
Disposal instructions ................... 7
DS2-FH1 installation ................. 14
Dual tracking .............................. 31
EN61010
pollution degree.............................. 6
Entering the ARB function
menu ........................................ 26
Environment
safety instructions .......................... 6
Front panel diagram .................. 12
Function generator
operation ........................................ 26
Fuse
type ................................................... 5
Ground
symbol .............................................. 4
List of features ............................ 10
Main features .............................. 10
Menu tree .................................... 21
Modulation
AM .................................................. 40
FM ................................................... 42
FSK .................................................. 44
PM ................................................... 46
SUM ................................................ 48
Module installation .................... 15
Module removal ......................... 15
Operation .................................... 25
Output ................................... 29, 34
Overview ....................................... 9
Power on/off
safety instruction ............................ 5
Power supply operation ............ 70
Preset settings ............................. 33
Rear panel diagram ................... 13
Service operation
about disassembly .......................... 5
State display ................................ 27
Sweep .......................................... 50
Sync Output ................................ 34
Tracking ...................................... 31
Turn the output on or off .......... 29
81
AFG-125/225 Series Options Manual
USB configuration ..................... 17 Warning symbol .......................... 4
82
Loading...