Guth Laboratories 2100 Operation Manual

Guth Laboratories, Inc.
MODEL 2100 SIMULATOR
OPERATION MANUAL
590 North 67th Street Harrisburg, PA 17111-4511
ELEPHONE: 717-564-5470 / 800-233-2338
T FAX: 717-564-2555 WEB SITE AT: www.guthlabs.com PRICE $5.00
© 1999 Guth Laboratories, Inc./all rights reserved
RECORDS (SERIAL NUMBERS, CALIBRATION AND NOTES)
NIST TRACEABLE THERMOMETER SERIAL NUMBER___________________
CERTIFIED BY________________________________DATE_____________
TEST TEMPERATURE_______qC NIST TRUE TEMPERATURE_______qC
MODEL 2100 SERIAL NUMBER_________________________________
CALIBRATION CHECK DATE
CHECKED BY
NOTE:
Only a Guth Laboratories, Inc. certified technician shall perform Field
calibration.
INTRODUCTION
The Guth Model 2100 Alcohol Breath Simulator is a State-of-the-Art, electronically temperature controlled, water-alcohol instrument for the purpose of providing a precise calibration standard. Temperature of the water-alcohol solution placed in the simulator is maintained at 34oC + .05oC or better.
The Guth Model 2100 Simulator is approved by the National Highway Traffic Safety Administration (NHTSA), published on their Conforming Products List (CPL), and is listed with Underwriters Laboratories, Inc. (UL), as meeting United States Safety Standards and Canadian Safety Standards (CUL).
CERTIFIED* SIMULATOR SOLUTION STANDARD
PLEASE NOTE: It is imperative your Simulator Solution be of the highest quality in
order to acquire the scientific and legal standards of acceptance
.
To establish a precise and accurate calibration standard when using a Guth simulator, GUTH LABORATORIES, INC. recommends using Guth Certified Simulator Solution. Guth Laboratories, Inc., a pioneer and leader in the science of alcohol breath testing, has provided Certified Simulator Solution to state, municipal, and local law enforcement agencies for many years.
* * * * * * * * * * * *
Guth Premixed Certified Simulator Solution is the only solution available which is prepared by a FDA / DEA Registered Laboratory.
* * * * * * * * * * * *
Guth Certified Simulator Solutions are prepared in standard concentrations of .02%, .04%, .05%, .08%, .10%, .15% and .20%. Other special concentrations are available. Please contact Guth Laboratories, Inc. for availability and pricing:
Toll free: (800) 233-2338 Fax: (717) 564-2555
or visit our web site at: www.guthlabs.com
STORAGE: Store Simulator Solution in a cool location. Do not place solution in a freezer
DO NOT REFRIGERATE SIMULATOR.
*Guth Certified Simulator Solutions are tested by recognized laboratories.
1
EXCLUSIVE FEATURES OF THE MODEL 2100 SIMULATOR
x MICROPROCESSOR CONTROLLED: Mercury column is eliminated.
x TEMPERATURE SENSOR: The Model 2100
incorporates a high precision temperature sensor.
x Maintains a precise temperature of 34
o
C + .05oC or better.
x Will not heat if temperature sensor is defective (shorted or open).
x GUTH LABORATORY GLASS CONTAINER: Graduated 500ml +
5%.
x NON-MAGNETIC METAL TOP HOUSING ASSEMBLY: Provides uniform headspace
heating.
x POWER SWITCH: Safety On/Off - Double pole, double throw (DPDT).
x ELECTRICALLY ISOLATED: Internal power transformer and optically isolated control
components prevent AC leakage.
x FUSE: External fuse provides protection in the event of an electrical malfunction.
x HIGH INTENSITY LED DISPLAY: Provides maximum visibility of temperature and heater
activity.
x RELIABLE AGITATOR MOTOR.
x AGITATOR PADDLE: Designed by Guth Laboratories, Inc. Offers maximum circulation
and uniformity of solution.
x HEATER ELEMENT: Electronically protected from overheating when removed from
solution or when temperature sensor is defective.
x RFI PROTECTED: The heating element and motor will turn off when the simulator is
subjected to a RF field of 10 volts/meter or greater.
x INTERNAL DIAGNOSTICS: Simulator computer performs internal diagnostics.
x MALFUNCTION INDICATION: Simulator provides an audible and visual indication if a
malfunction occurs.
x POWER : 120 volts AC 60 Hz.
x
x RS-232 COMMUNICATIONS: Provides the ability to monitor the temperature and status of
the simulator.
xOOPPTTIIOONNAALL::SSEE
R
RIIAALLIIZZEEDDNNIISSTTRREEFFEERREENNCCEEMMEERRCCUURRYY--IINN--GGLLAASSSS
TTHHEERRMMOOMMEETTEERRFFOORRCCAALLIIBBRRAATTIIOONNCCHHEECCKKSS.
.
2
2
REMEDY: Turn off the simulator and turn on. If the simulator does not heat, the heating element may be defective. Requires service*
*REQUIRES SERVICE - Simulator must be repaired by a Guth certified technician or returned to Guth Laboratories, Inc. (Seeinstructions on page 9.)
The temperature of the solution is belo
w
33.8oC after it has initially obtaine
d
34.0oC.
MO DE L 2100
ON
OFF
RESET
GUTH LABORATORIES, IN C.
.5 .5
G U TH MO D EL 2100 SIMULATOR
C
3
MODEL 210 0 REAR V I EW
!
CAU TIO N
REMOVE SOLUTION BEFORE
MOVING INSTRUMENT
REPL ACE FUS E WI TH BU SS TYPE G M W
1amp @ 125 vac Fast Acting
4
Loading...
+ 11 hidden pages