Guralp Systems CMG-5TD Installation

Installation
1
CMG-5TD
Accelerometer
Check you have all components and cables.
CMG-5TD strong motion broadband accelerometer
GPS Receiver
Blue/grey sensor data and power cable
Brown GPS sensor cable
You will need a 12–24 Volt DC power supply and a Windows or Linux PC with an RS232 (COM) port.
Keep the packaging, in case you need to transport the sensor at a later date.
Prepare the installation site. The instrument should be installed on a solid surface, such as a
2
concrete floor.
To create a suitable surface in a temporary site:
1. Dig a pit, or machine-auger a suitable hole for the sensor.
2. Prepare a quick-setting cement/sand mixture, and pour it into the hole.
3. Agitate (“puddle”) the mixture until the surface is level.
4. Leave the mixture for between 2 and 12 hours, depending on the type of cement used.
Scribe an accurate North/South line on the surface.
3
Place the sensor in position and rotate it until it’s studs and orientation lines align with the scribed line.
If you are using a base-plate and ring:
1. Fit the longer three screws half way into alternate holes in the round base-plate and add a lock-nut to each (left-hand top picture).
2. Drill through the centre of the North/South line and use the anchor bolt to fasten the round base-plate to the surface with the screw-heads uppermost. Rotate to the correct orientation before tightening.
4. Screw the indented mounting ring onto the bottom of the sensor (right-hand top picture).
5. Fix the sensor onto the base-plate using the shorter set of three screws through the mounting ring.
6. Use the bubble level to check the sensor attitude and adjust the mounting feet as required.
Installation
4
CMG-5TD
RS232
PC
+
Power supply
Connect the various components together:
GPS
Connect the brown cable from the GPS receiver to the GPS connector on the sensor;
Connect the 9-pin ‘D’ connector on the blue cable to your PC’s serial (COM) connector;
Connect the free, stripped end of the grey cable to the power supply. Connect the black wire to the negative (–) terminal and the red wire to the positive (+) terminal;
The blue and grey cables are joined together at a 10­pin mil-spec socket. Attach this to the DATA connector on the sensor. Do this step last.
5
6
Start the PC and run Güralp Systems’ Scream! software.
If you have not run Scream! before, the Setup window will open. Otherwise, choose File Setup... from the main menu and view the Com Ports tab.
Set the Baud Rate to 38,400 and click OK. Data streams should start appearing in the main window. Highlight the Z, N and E streams by control-clicking them and press Enter to open a waveview window.
Remove the screw-on cap protecting the offset adjusters. With a small screwdriver, adjust each offset until the corresponding output in the waveview window is close to zero. Zoom in as necessary.
Replace the pressure cap. The sensor is now ready for use.
Please refer to the full manual for detailed instructions on usage, calibration and troubleshooting.