Guralp Systems CMG-40T User Manual

CMG-40T
Triaxial Broadband Seismometer
Operator's guide
Part No. MAN-040-0001
Designed and manufactured by Güralp Systems Limited 3 Midas House, Calleva Park Aldermaston RG7 8EA England
Proprietary Notice: The information in this manual is proprietary to Güralp Systems Limited and may not be copied or distributed outside the approved recipient's organisation without the approval of Güralp Systems Limited. Güralp Systems Limited shall not be liable for technical or editorial errors or omissions made herein, nor for incidental or consequential damages resulting from the furnishing, performance, or usage of this material.
Issue A 2006-01-12
CMG-40T
Table of Contents
1 Introduction............................................................................................................................4
1.1 Response options..............................................................................................................5
2 First encounters......................................................................................................................6
2.1 Handling notes..................................................................................................................6
2.2 Connections......................................................................................................................6
The breakout box...............................................................................................................6
The handheld control unit..................................................................................................7
Signal meter..................................................................................................................8
Calibration.....................................................................................................................9
Control commands........................................................................................................9
Banana plugs.................................................................................................................9
2.3 Zeroing the instrument.....................................................................................................9
Adjusting the mass position offsets manually...................................................................9
Adjusting the mass position offsets with a Handheld Control Unit................................11
Zeroing a 40TD digital instrument..................................................................................11
2.4 Installation notes.............................................................................................................14
3 Installing the 40T.................................................................................................................16
3.1 Installing in vaults..........................................................................................................16
Choosing a location.........................................................................................................16
Temperature stability.......................................................................................................17
Other considerations........................................................................................................18
3.2 Installing in pits..............................................................................................................19
Other installation methods...............................................................................................21
4 Calibrating the 40T..............................................................................................................23
4.1 The calibration pack.......................................................................................................23
Poles and zeroes...............................................................................................................24
Frequency response curves..............................................................................................25
Obtaining copies of the calibration pack.........................................................................25
4.2 Calibration methods.......................................................................................................26
4.3 Calibration with Scream! ...............................................................................................26
Sensor response codes.....................................................................................................30
4.4 Calibration with a handheld control unit........................................................................30
4.5 The coil constant............................................................................................................31
Appendix A Connector pinouts.............................................................................................32
2 Issue A
Operator's guide
Appendix A.1 Output port and breakout box RECORDER connector................................32
Appendix A.2 Output port and breakout box RECORDER connector (high gain option)..33
Appendix A.3 Breakout box power connector.....................................................................34
Appendix B Specifications......................................................................................................35
January 2006 3
CMG-40T

1 Introduction

The CMG-40T is an ultra-lightweight seismometer consisting of three sensors in a sealed case, which can measure the north/south, east/west and vertical components of ground motion simultaneously.
The 40T has a rugged, waterproof stainless steel design for ease of installation. The lightweight sensor elements are designed so that no mechanical clamping is required. Because of this, the 40T is ready to record ground movements as soon as you provide it with power. In addition, the sensor does not have to be levelled or centred as long as the base is within 3 ° of horizontal. For the best results, however, you should install where possible on a hard, near-horizontal surface well coupled to the bedrock.
Each seismometer is delivered with a detailed calibration sheet showing its serial number, measured frequency response in both the long period and the short period sections of the seismic spectrum, sensor DC calibration levels, and the transfer function in poles/zeros notation.
4 Issue A
Operator's guide

1.1 Response options

The 40T can be supplied with a response which is flat to velocity from 50 Hz to any of 0.1 Hz (10 s), 0.050 Hz (20 s) or 0.033 Hz (30 s).
If you do not require high-frequency data, a low-pass filter may be installed at a frequency (below 50 Hz) that you specify.
Standard 40T instruments output signals representing ground velocity on three pairs of balanced differential lines. An option is available which provides a second, parallel set of outputs at higher gain. The high-gain outputs have a sensitivity nominally 10 times higher than the standard (low-gain) outputs.
January 2006 5
CMG-40T

2 First encounters

2.1 Handling notes

Although the 40T has a rugged design, it is still a sensitive instrument, and can be damaged if mishandled. If you are at all unsure about the handling or installation of the device, you should contact Güralp Systems for assistance.
Avoid bumping or jolting the sensor when handling or unpacking.
Do not kink or walk on the data cable (especially on rough surfaces such
as gravel), nor allow it to bear the weight of the sensor.
Do not connect the instrument to power sources except where instructed.
Do not ground any of the signal lines from the sensor.
All parts of the 40T are waterproof.

2.2 Connections

The instrument has an integrated cable ending in a 26-pin mil-spec socket which carries both power and output signals. This is suitable for connecting directly to a Güralp digitizer.
The breakout box, if ordered, provides individual signal and power connectors, or you can make up your own cable if you prefer.

The breakout box

If you are using a Güralp breakout box, it should be attached to the sensor through its SENSOR connector. Connectors are also provided at the CONTROL and RECORDER outputs, for attaching to a handheld control unit or a Güralp digitizer. If you have ordered a 40T with optional high gain outputs, you will need to make up a suitable cable to expose these outputs.
6 Issue A
Operator's guide
The breakout box also provides a standard Güralp power connector on a 10-pin mil-spec plug. The 40T draws a nominal current of 48 mA from a 12 V supply when in use; thus, using a 12 V, 25 Ah sealed heavy-duty lead-acid battery, you should expect the instrument to operate for around a week without recharging.
The CENTRE button switches the instrument into ACC/VEL mode whilst it is pressed. This mode allows you to monitor the mass positions whilst you adjust the offsets manually. If you prefer, you can use the equivalent switch on a Handheld Control Unit (see below.)

The handheld control unit

This portable control unit provides easy access to the seismometer's control commands, as well as displaying the output velocity and mass position (i.e. acceleration) on an analogue meter.
January 2006 7
CMG-40T
Signal meter
The upper section of the HCU contains a simple voltmeter for monitoring various signals from the instrument.
To monitor the low-gain outputs, switch the dial to V, N/S or E/W LOW
VEL according to the component you want to monitor.
To monitor the high-gain outputs (on a 40T with that option), switch the
dial to V, N/S or E/W HIGH VEL.
To monitor the mass position outputs, switch the dial to V, N/S or E/W
MASS POS. Whilst you are adjusting mass position offsets, you should
also switch the instrument out of broadband mode by switching the rightmost CENTRING SELECT switch to 1 SEC VEL, or by holding down the CENTRE button on a breakout box.
You can set the range of the meter with the RANGE switch. When
switched to 10 V, the meter ranges from –10 to + 10 V (as marked.) When switched to 1 V, the range is –1 to +1 V.
8 Issue A
Operator's guide
Calibration
You can calibrate a 40T sensor through the HCU by connecting a signal generator across the yellow and green CALIBRATION SIGNAL inputs and setting the adjacent switch to ON. The sensor's response can now be monitored or recorded, and calibration calculations carried out. See Chapter 4, “Calibrating
the 40T” for full details.
Control commands
If you have ordered a 40T with the remote null facility, you can zero its mass position offsets from the HCU.
1. Select the component you want to centre from the CENTRING SELECT dial.
2. Switch the signal meter dial to one of the MASS POS settings.
3. Switch the rightmost switch to 1 SEC VEL to enable the centring lines.
4. Press the +/– switch towards – to centre a mass from a positive value, or towards + to centre it from a negative value.
Banana plugs
The remainder of the HCU provides useful connections for each of the signal lines from the instrument, for attaching to your own equipment as necessary.

2.3 Zeroing the instrument

Before installing the 40T, you should check that the mass positions are not significantly offset from zero. The mass position offsets can be affected by any tilt to the instrument, as well as handling during transportation. The normal range of the mass positions is ±10 V; you should zero the instrument if any mass reads more than around ±3.5 V when the sensor is stationary.
The velocity outputs of the 40T are set at the factory to a nominal value below ±3 mV. Once the instrument is installed and has reached thermal equilibrium with its environment, these outputs should be similar to the factory-set value.

Adjusting the mass position offsets manually

The 40T has three potentiometers (“pots”) accessible within its casing, which should be used to remove any DC offsets electronically:
1. Bring the instrument into 1 second response mode by applying a voltage across the Acc/Vel and Signal Ground pins of the input. If you are using
January 2006 9
CMG-40T
a DM24, you can do this by sending a CENTRE command. If you are using a Handheld Control Unit, you should select 1 SEC VEL from the Velocity Select switch.
2. Measure the vertical mass position output with a 10 V voltmeter (see
Appendix A, “Connector pinouts”) or by selecting MASS POS, V from a
Handheld Control Unit's Display Select knob. If using a HCU, also check that the Centring Select knob is set to OFF.
3. If the vertical component needs adjusting, remove the cap on the lid which protects the Verticalpot with a flat-bladed screwdriver (provided).
4. Insert the screwdriver through the opening, and engage the pot. An LED flashlight may be useful for locating the head.
5. Turn the pot either way until the offset readout is as close to 0 V as possible.
10 Issue A
Operator's guide
6. Repeat steps 2 – 5 for the north/south and east/west components.

Adjusting the mass position offsets with a Handheld Control Unit

Some 40T units are equipped with a remote mass centring option, which allows you to adjust the internal potentiometers by applying voltages across control lines to the sensor:
1. Bring the instrument into 1 second response mode by selecting 1 SEC VEL from the Velocity Select switch.
2. Measure the vertical mass position output by selecting MASS POS, V from the Handheld Control Unit's Display Select knob.
3. Set the Centring Select knob to V.
4. Press the spring-loaded switch towards + or – to bring the mass position offset from negative or positive values towards zero.
5. Repeat steps 3 and 4 for the N/S and E/W components.
6. Return the instrument to broadband mode by selecting BB VEL from the Velocity Select switch.

Zeroing a 40TD digital instrument

The offset potentiometers in a 40TD are in the same place as on the 40T. To access them, you will need to remove the digitizer module, which lies on top of the sensor itself. You can monitor the mass position outputs of the sensor using a Handheld Control Unit and an adapter cable, available from Güralp Systems.
To change the offsets of a 40TD without digital centring:
1. Check the bubble level on the lid of the instrument, to ensure it is not tilted. If necessary, re-level the instrument by adjusting its feet.
2. Unscrew the vent cap on the lid to allow the air pressure to equalise.
January 2006 11
Loading...
+ 24 hidden pages