Guralp Systems CMG-3T User Manual

CMG-3T
Triaxial Broadband Seismometer
Ope rat or's g uide
Part No. MAN-030-0001
Designed and manufactured by Güralp Systems Limited 3 Midas House, Calleva Park Aldermaston RG7 8EA England
Proprietary Notice: The information in this manual is proprietary to Güralp Systems Limited and may not be copied or distributed outside the approved recipient's organisation without the approval of Güralp Systems Limited. Güralp Systems Limited shall not be liable for technical or editorial errors or omissions made herein, nor for incidental or consequential damages resulting from the furnishing, performance, or usage of this material.
Issue N 2009-07-06
CMG-3T
Table of Contents
1 Introduction...............................................................................................................3
1.1 Options................................................................................................................4
2 Installing the 3T........................................................................................................7
2.1 First encounters..................................................................................................7
2.2 Installation notes..............................................................................................10
2.3 Installing in vaults............................................................................................11
2.4 Installing in pits................................................................................................14
2.5 Installing in post-holes....................................................................................18
3 Calibrating the 3T...................................................................................................20
3.1 The calibration pack.........................................................................................20
3.2 Calibration methods.........................................................................................23
3.3 Calibration with Scream! ................................................................................23
3.4 Calibration with a hand-held control unit.......................................................28
3.5 The coil constant..............................................................................................28
4 Accessories...............................................................................................................30
4.1 The breakout box..............................................................................................30
4.2 The hand-held control unit..............................................................................31
4.3 Integrated State-of-Health Controller...............................................................33
5 Inside the 3T............................................................................................................35
5.1 The sensors.......................................................................................................35
5.2 The control system...........................................................................................36
5.3 RS232 control interface....................................................................................41
5.4 The feedback system........................................................................................42
6 Connector pin-outs..................................................................................................47
6.1 Sensor and control unit pin output.................................................................47
6.2 Sensor output (“D”-type connector option).....................................................48
6.3 Sensor output (waterproof connector option).................................................49
6.4 Breakout box POWER connector......................................................................50
7 Specifications..........................................................................................................51
8 Revision history.......................................................................................................53
2 Issue N
Operator's guide
1 Introduction
The CMG-3T is a three-axis seismometer consisting of three sensors in a sealed case, which can measure the north/south, east/west and vertical components of ground motion simultaneously. Each sensor is sensitive to ground vibrations in the frequency range 0.003 – 50 Hz, a broadband frequency response made possible by advanced force­balance feedback electronics. Because of this wide response range, the 3T can replace many of the instruments conventionally used in a seismic observatory; it also produces true pulse-shape records suitable for modern earthquake mechanism analysis.
The 3T is designed for mounting on a hard, near-horizontal surface well coupled to the bedrock. After levelling and orienting the case, you can perform accurate adjustments internally by sending the instrument control signals. These electronics allow it to compensate for a tilt of up to 3 ° from horizontal.
Once levelled and centred, the 3T will begin operating automatically. It outputs analogue voltages representing ground velocity on balanced differential lines. These voltages can be recorded using a separate logging device or digitizer. For testing and installation purposes, a hand-held control unit is supplied which can monitor the instrument's output.
July 2009 3
CMG-3T
The seismometer unit is self-contained apart from its 12 Volt power supply. Centring and mass locking can be carried out by sending control signals to the instrument, either through the hand-held control unit or through an attached Güralp digitizer.
Each seismometer is delivered with a detailed calibration sheet showing its serial number, measured frequency response in both the long period and the short period sections of the seismic spectrum, sensor DC calibration levels, and the transfer function in poles/zeros notation.
1.1 Options
Form factors
The CMG-3T can be supplied in several forms, besides its standard configuration:
The CMG-3T Compact is internally identical to the standard 3T,
but has a different arrangement of components allowing it to fit in a smaller casing.
The 3T can also be supplied in a slimline form factor, with
vertically-stacked sensors, suitable for installation in post-holes (see Section 2.5, page 18.).
The 3T is also available as a fully-fledged borehole instrument,
the CMG-3TB.
Any of these can be supplied with integral digitizers and data modules, allowing the 3T to form a complete, integrated seismic installation. For example, the 3TD is a full-height 3T with an integrated CMG-DM24 digitiser. As an additional option, the 3TD can be ordered with an Integrated State-of-Health Controller, which provides the mass-control functions (normally accessed via the break­out box) directly from controls mounted on the instrument. See section 4.3 on page 33 for more details.
Output types
The standard 3T has a single 26-pin mil-spec waterproof connector for signals, control and power:
4 Issue N
Operator's guide
As an option, the 3T can be supplied with a 26-way “D”-type connector with standard pin assignments (see section 6.2, page 48.)
Other output connector types may also be supplied, to your own pin­out specification.
Sensor control interface
3T instruments can be fitted with an optional RS232 control interface. The control interface can be used to lock, unlock, and centre the sensor masses, and query mass positions. See section 5.3, page 41, for more details.
July 2009 5
CMG-3T
Sensor response
The 3T can be supplied with a response which is flat to velocity from 100 Hz to any of 0.1 Hz (10 seconds), 0.033 Hz (30 seconds), 0.016 Hz (60 seconds), 0.01 Hz (100 seconds), 0.0083 Hz (120 seconds) or 0.0027 Hz (360 seconds). Alternatively, a hybrid response function may be provided. See Chapter 5 for more details.
If you do not require high-frequency data, a low-pass filter may be installed at any frequency (below 100 Hz) that you specify.
6 Issue N
Operator's guide
2 Installing the 3T
2.1 First encounters
Unpacking
The 3T seismometer is delivered in a single transportation case. The packaging is specifically designed for the 3T and should be reused whenever you need to transport the sensor. Please note any damage to the packaging when you receive the equipment, and unpack on a safe, clean surface. The package should contain:
the seismometer;
a cable to join the sensor to the breakout box;
the breakout box (which provides separate connections for the
signal, control and power lines);
a Hand-held Control Unit (HCU) for monitoring sensor outputs
and calibration, if ordered;
a 10-pin connector for your power lead (see below); and
a calibration and installation sheet.
Assuming all the parts are present, stand the seismometer in the centre of a bench and identify its external features:
a handle with North indication,
a multi-pole socket for input and output,
other optional connectors as ordered;
a bubble level,
an air vent port,
three adjustable feet, and
two accurate orientation pins (one brass and one steel).
July 2009 7
CMG-3T
Serial number
The sensor's serial number can be found on the label stuck to the top lid of the sensor. It is also stamped onto the side of the sensor base, next to the N/S indicator, and into the lid itself. You should quote this serial number if you need assistance from Güralp Systems.
Handling notes
The 3T is a sensitive instrument, and is easily damaged if mishandled. If you are at all unsure about the handling or installation of the device, you should contact Güralp Systems for assistance.
Do not bump or jolt any part of the sensor when handling or
unpacking.
Do not kink or walk on the data cable (especially on rough
surfaces such as gravel), nor allow it to bear the weight of the sensor.
Do not connect the instrument to power sources except where
instructed.
Do not ground any of the signal lines from the sensor.
Do not move the instrument whilst the masses are unlocked.
You should report any sign of loose components, or any sound of parts moving inside the instrument, to Güralp Systems.
Connections
The instrument has a single connector, which can be joined using the cable provided to a digitizer or breakout box. Individually shielded twisted-pair cabling must be used for the sensor outputs, control lines and power supply. If you need to make up a suitable cable, you should confirm the cable type with Güralp Systems.
Using a digitizer
The 3T can be connected directly to any Güralp Systems digitizer using the signal cable provided. This is the simplest way to use a 3T instrument. All the instrument's functions are available through the digitizer, including centring, locking and unlocking.
We recommend that you keep the digitizer near the instrument if at all possible, to minimize the length of analogue cable required. Once digitized, the signal is robust to degradation by noise or attenuation. Keeping the digitizer in the quiet, stable conditions of a seismic installation also provides it with an optimum environment for the on-
8 Issue N
Operator's guide
board ADCs.
Breakout box and hand-held control unit
The 3T can be supplied with an optional robust breakout box, which provides mass control functions in installations which do not use compatible digitizers. A hand-held control unit is also available which can control calibration lines and monitor sensor outputs in addition to mass control. See chapter 4, page 30, for more details.
Power supply
The sensor requires a 12 Volt power supply, connected via the socket and breakout box or digitizer. You will need to make up a suitable cable to connect a 12 V power source to the 10-pin connector on the breakout box (spare 10-pin mil-spec connectors are provided for this purpose.) Using a 12 Volt, 25 Ampere-hour, sealed, heavy-duty, lead­acid battery, you should expect the instrument to operate for around a week without recharging.
If you prefer, you can power the 3T directly from the connector on the top panel (see Chapter 6, page 47.)
A power management module can be installed as an option, which allows the 3T to operate from a 10 – 15 V supply range. This module also cuts the input power to the sensor electronics if it drops below
10.5 V, to minimize discharge from battery-operated installations. Units with serial numbers beginning T33 are provided with a wide input range DC-DC converter, and can be powered from a 10 – 36 V supply.
The 3T draws a nominal current of 75 milliamps from a 12 Volt supply when in use. During locking and unlocking of the sensor masses, this current rises briefly to 600 milliamps. It is recommended that you carry a spare 12 Volt battery when visiting an installation for maintenance, in case the sensor needs to be moved and the on-site batteries no longer have sufficient charge to perform the locking procedure.
Signal output
The sensors output voltages representing ground velocity on floating differential lines. The breakout box provides a RECORDER connector for attaching to a recording system or digitizer. You can use any multi­channel recording system, provided that it has high-impedance floating differential inputs.
If you are using a Güralp Systems digitizer, you can connect the instrument directly to the digitizer without using the breakout box;
July 2009 9
CMG-3T
power will be supplied through the digitizer, which can also activate the sensor control lines.
The breakout box also provides a CONTROL output, which can be connected to the Hand-held Control Unit. This device lets you monitor output signals from the instrument, and perform on-site calibration. For more information, see Section 6, page 47.
2.2 Installation notes
The goal of any seismic installation is to ensure that wave-trains arriving at the instrument accurately reflect the internal motion of subsurface rock formations. To achieve this, the seismometer and its emplacement need to be considered as a mechanical system, which will have its own vibrational modes and resonances. These frequencies should be raised as high as possible so that they do not interfere with true ground motion: ideally, beyond the range of the instrument.
In particular, the sensor needs to be protected against environmental factors such as
fluctuations in temperature,
turbulent air flow around walls or trees, or around sharp
corners or edges in the immediate vicinity of the sensor;
vibration caused by equipment in or near the installation,
particularly computer equipment; and
vibration caused by heavy machinery (even at a distance), or by
overhead power lines.
In seismic vaults, instruments are often installed on piers. It is important to ensure that the interface between the pier and the floor does not introduce noise, and that the pier itself does not have resonant frequencies within the passband. Ideally, a seismic pier will be significantly wider than it is high (to minimize flexing) and will form a single piece with the floor, e.g. by moulding a poured concrete floor within a wooden frame.
Many situations do not allow for the construction of a seismic vault. For example, you may need to deploy quickly to monitor the activity of a volcano showing signs of rejuvenation, or to study the aftershocks of a major earthquake. In other cases the site itself may be too remote to ship in construction equipment.
10 Issue N
Operator's guide
Temporary installations can be protected against spurious vibrations by
selecting a suitable site,
placing the instrument in a protective enclosure (an open-sided
box of 5 cm expanded polystyrene slabs, placed over the instrument and taped down to exclude draughts, makes an excellent thermal shield),
standing the sensor on bedrock where possible, or at least deep
in well-compacted subsoil;
clearing the floor of the hole of all loose material; and
using as little extra mass as possible in preparing the chamber.
After installation, the instrument case and mounting surface will slowly return to the local temperature and settle in their positions. This will take around four hours from the time installation is completed. If you require long-period recording, you should re-zero the instrument after this time.
2.3 Installing in vaults
You can install a 3T in an existing seismic vault with the following procedure:
1. Unpack the sensors from their container, saving the shipping boxes for later transportation.
2. Prepare the mounting surface, which should be smooth and free of cracks. Remove any loose particles or dust, and any pieces of loose surfacing. This ensures good contact between the instrument's feet and the surface.
3. If it is not already present, inscribe an accurate North-South line on the mounting surface.
4. Place the sensor over the scribed line, so that the brass and steel pointers are aligned with the marked directions, with the brass pointer facing North. This can be done by rotating the base of the sensor whilst observing it from above. The brass pointer can be found next to one of the feet.
July 2009 11
CMG-3T
If you cannot easily see the pointers, you should align the sensor using the north arrow on the handle. However, the alignment of the handle with the sensors inside is less accurate than the metal pointers, so they should be used wherever possible.
5. The top panel of the 3T includes a spirit level.
Level the sensor using each of the adjustable feet of the instrument in turn, until the bubble in the spirit level lies entirely within the inner circle. (The instrument can operate with up to 2° of tilt, but with reduced performance.)
The feet are mounted on screw threads. To adjust the height of a foot, turn the brass locking nut anticlockwise to loosen it, and rotate the foot so that it screws either in or out. When you are happy with the height, tighten the brass locking nut clockwise to secure the foot. When locked, the nut should be at the bottom of its travel for optimal noise performance.
6. Connect the sensor to a breakout box, or a Güralp digitizer if
12 Issue N
Operator's guide
you are using one.
7. Connect a 12 Volt power supply, either directly or through the breakout box or digitizer.
8. Unlock the sensor. If you have a breakout box or hand-held control unit, you can do this by holding the ENABLE and
UNLOCK buttons on the unit down together for 7 seconds. The BUSY LED will start flashing, and then go out.
Alternatively, if you are using a DM24 digitizer and Scream!,
right-click on the digitizer's entry in Scream! and select
Control.... Click on the Mass control tab, followed by Unlock. (If
the Mass control tab is unavailable, check the sensor type in the
Sensor type tab, apply, and open a new Control window.)
Alternatively, if you are using a DM24 digitizer and a DCM,
navigate to the Actions Digitizer Setup Port x page for the digitizer and click on the Unlock instrument button at the bottom of the page.
After unlocking the masses, you should be careful not to tilt the instrument or you may damage it.
9. Check the mass position outputs using a digital multimeter, digitizer or hand-held control unit.
Re-centre the masses if required. If you have a breakout box or hand-held control unit, you can do this by holding the ENABLE and CENTRE buttons on the unit down together for 7 seconds. The BUSY LED will start flashing, and then go out
Alternatively, if you are using a DM24 digitizer and Scream!, right-click on the digitizer's entry in Scream! and select
Control.... Click on the Mass control tab, followed by Centre.
Alternatively, if you are using a DM24 digitizer and a DCM, navigate to the Actions Digitizer Setup Port x page for the digitizer and click on the Centre instrument button at the bottom of the page.
10.Cover the instrument with thermal insulation, for example, a 5 cm expanded polystyrene box. This will shield it from thermal fluctuations and convection currents in the vault. It also helps to stratify the air in the seismometer package. Position the thermal insulation carefully so that it does not touch the sensor package.
July 2009 13
CMG-3T
11.Ensure that the sensor cable is loose and that it exits the seismometer enclosure at the base of the instrument. This will prevent vibrations from being inadvertently transmitted along the cable.
2.4 Installing in pits
For outdoor installations, high-quality results can be obtained by constructing a seismic pit.
Depending on the time and resources available, this type of installation can suit all kinds of deployment, from rapid temporary installations to medium-term telemetered stations.
Ideally, the sensor should rest directly on the bedrock for maximum coupling to surface movements. However, if bedrock cannot be reached, good results can be obtained by placing the sensor on a granite pier on a bed of dry sand.
1. Prepare a hole of 60 – 90 cm depth to compacted subsoil, or
down to the bedrock if possible.
14 Issue N
Operator's guide
2. On granite or other hard bedrock, use an angle grinder to plane off the bedrock at the pit bottom so that it is flat and level. Stand the instrument directly on the bedrock, and go to step 7.
3. On soft bedrock or subsoil, you should install a pier as depicted below.
4. Pour a layer of loose, fine sand into the pit to cover the base. The type of sand used for children's sand-pits is ideal, since the grains are clean, dry and within a small size range. On top of the sand, place a smooth, flat granite plinth around 20 cm across, and shift it to compact the sand and provide a near­level surface.
Placing a granite plinth on a sand layer increases the contact between the ground and the plinth and improves the performance of the instrument. There is also no need to mix concrete or to wait for it to set, as in step 5.
5. Alternatively, if time allows and granite is not available,
July 2009 15
CMG-3T
prepare a concrete mix with sand and fine grit, and pour it into the hole. Agitate (“puddle”) it whilst still liquid, to allow it to flow out and form a level surface, then leave to set. Follow on from step 7.
Puddled concrete produces a fine-textured, level floor for situating the seismometer. However, once set hard, the concrete does not have the best possible coupling to the subsoil or bedrock, which has some leeway to shift or settle beneath it.
6. Alternatively, for the most rapid installation, place loose soil over the bottom of the pit, and compact it with a flat stone. Place the seismometer on top of this stone. This method emulates that in step 3, but can be performed on-site with no additional equipment.
7. Set up the instrument as for a vault installation (Section 2.3, page 11, steps 4 to 9.)
8. The instrument must now be shielded from air currents and temperature fluctuations. This is best done by covering it with a thermal shield.
An open-sided box of 5 cm expanded polystyrene slabs is recommended. If using a seismic plinth on sand (from steps 3– 4 or 5), ensure that the box is firmly placed in the sand, without touching the plinth at any point. In other installations, tape the box down to the surface to exclude draughts.
9. Alternatively, if a box is not available, cover the instrument with fine sand up to the top.
The sand insulates the instrument and protects it from thermal fluctuations, as well as minimizing unwanted vibration.
10. Ensure that the sensor cable is loose and that it exits the seismometer enclosure at the base of the instrument. This will prevent vibrations from being inadvertently transmitted along the cable.
11. Cover the pit with a wooden lid, and back-fill with fresh turf.
Other installation methods
The recommended installation methods have been extensively tested in a wide range of situations. However, past practice in seismometer installation has varied widely.
16 Issue N
Operator's guide
Some installations introduce a layer of ceramic tiles between a rock or concrete plinth and the seismometer (left):
However, noise tests show that this method of installation is significantly inferior to the same concrete plinth with the tiles removed (right). Horizontal sensors show shifting due to moisture trapped between the concrete and tiling, whilst the vertical sensors show "pings" as the tile settles.
Other installations have been attempted with the instrument encased in plaster of Paris, or some other hard-setting compound (left):
Again, this method produces inferior bonding to the instrument, and moisture becomes trapped between the hard surfaces. We recommend the use of fine dry sand (right) contained in a box if necessary, which can also insulate the instrument against convection currents and temperature changes. Sand has the further advantage of being very easy to install, requiring no preparation.
Finally, many pit installations have a large space around the seismometer, covered with a wooden roof. Large air-filled cavities are susceptible to currents which produce lower-frequency vibrations, and sharp edges and corners can give rise to turbulence. We recommend that a wooden box is placed around the sensor to protect it from these currents. Once in the box, the emplacement may be backfilled with fresh turf to insulate it from vibrations at the surface, or simply roofed as before.
July 2009 17
CMG-3T
By following these guidelines, you will ensure that your seismic installation is ready to produce the highest quality data.
2.5 Installing in post-holes
The 3T is suitable for installation in post-holes. In soft subsoil, a hole two to four metres deep and 20cm wide can be conveniently excavated using a tractor-mounted or hand-operated post-hole auger. To minimize surface effects, you should ensure that the hole is one metre deeper than the length of the instrument or, preferably, somewhat more.
Since the hole has no lining, it may occasionally flood. However, most soil types are sufficiently permeable to allow water to soak away, leaving the packing material moist.
A slim-profile 3T instrument is available with vertically stacked sensors and built-in inclinometer, specifically designed for post-hole installations. Digitizers or other recording equipment can be placed either within the hole or in a separate enclosure; alternatively, the Güralp DM24 is available in modular form, allowing you to attach the digitizer directly to the instrument.
To install a 3T in a post-hole:
1. Clean the post-hole, making sure there is no loose material around the mouth of the hole or on its base.
2. Prepare the instrument package, making sure the inclinometer is visible, and attach it to a winch or hoist by clamping a light steel cable to the centre of the handle so that the package hangs vertically. Connect the signal cable to the instrument.
3. Add packing material to the hole to about 15cm depth. Fine
18 Issue N
Operator's guide
crushed rock, with a high proportion of rock flour and fine particles, makes excellent packing material. Alternatively, a mixture of 3 mm grade angular coarse grit with around 30% medium grit gives good results. Moisten the packing material in the hole and ram firm.
4. Lower the instrument to the bottom of the hole, but without slackening the lifting cable.
5. Fill more packing material around the instrument for about 30 cm, moisten, and ram firm.
6. Use the inclinometer to check that the instrument remains within its tilt tolerance (± 2°).
7. Continue filling, moistening and packing until the instrument is buried, checking that the tilt remains within tolerance.
8. Release the strain on the lifting cable, and allow the packing material to settle for 24 hours.
9. If all is well after the settling period, release the lifting tackle, coil a tail of the lifting wire into the top of the hole and backfill almost to the surface.
10. Ensure that the signal cable is slack, and fix it to a support at the top of the hole.
11. Ram a split wooden bung into the top of the hole, and cover with sandbags.
12. Attach the signal cable to your recording equipment or breakout box. Power the sensor, and unlock it. Carry out preliminary tests using a hand-held control unit, if required.
July 2009 19
CMG-3T
3 Calibrating the 3T
3.1 The calibration pack
All Güralp sensors are fully calibrated before they leave the factory. Both absolute and relative calibration calculations are carried out. The results are given in the calibration pack supplied with each instrument:
Works Order : The Güralp factory order number including the instrument, used internally to file details of the sensor's manufacture.
Serial Number : The serial number of the instrument
Date : The date the instrument was tested at the factory.
Tested By : The name of the testing engineer.
There follows a table showing important calibration information for each component of the instrument, VERTICAL, NORTH/SOUTH, and
EAST/WEST. Each row details:
Velocity Output (Differential) : The sensitivity of each component to
velocity at 1 Hz, in volts per m/s. Because the 3T uses balanced differential outputs, the signal strength as measured between the +ve and –ve lines will be twice the true sensitivity of the instrument. To remind you of this, the sensitivities are given as 2 × (single-ended sensitivity) in each case.
Mass Position Output : The sensitivity of the mass position outputs to acceleration, in volts per m/s². These outputs are single-ended and referenced to signal ground.
Feedback Coil Constant : A constant describing the characteristics of the feedback system. You will need this constant, given in amperes per m/s², if you want to perform your own calibration calculations (see below.)
Power Consumption : The average power consumption of the sensor during testing, given in amperes and assuming a 12 V supply.
Calibration Resistor : The value of the resistor in the calibration circuit. You will need this value if you want to perform your own calibration calculations (see below.)
20 Issue N
Operator's guide
Poles and zeroes
Most users of seismometers find it convenient to consider the sensor as a “black box”, which produces an output signal V from a measured input x. So long as the relationship between V and x is known, the details of the internal mechanics and electronics can be disregarded. This relationship, given in terms of the Laplace variable s, takes the form
( V / x ) (s) = G × A × H (s)
In this equation
G is the acceleration output sensitivity (gain constant) of the
instrument. This relates the actual output to the desired input over the flat portion of the frequency response.
A is a constant which is evaluated so that A × H (s) is
dimensionless and has a value of 1 over the flat portion of the frequency response. In practice, it is possible to design a system transfer function with a very wide-range flat frequency response.
The normalising constant A is calculated at a normalising frequency value fm = 1 Hz, with s = j fm, where j = √–1.
H (s) is the transfer function of the sensor, which can be
expressed in factored form:
In this equation Zn are the roots of the numerator polynomial, giving the zeros of the transfer function, and Pm are the roots of the denominator polynomial giving the poles of the transfer
function.
In the calibration pack, G is the sensitivity given for each component on the first page, whilst the roots Zn and Pm, together with the
normalising factor A, are given in the Poles and Zeros table. The poles and zeros given are measured directly at Güralp Systems' factory using a spectrum analyser. Transfer functions for the vertical and horizontal sensors may be provided separately.
July 2009 21
CMG-3T
Frequency response curves
The frequency response of each component of the 3T is described in the normalised amplitude and phase plots provided. The response is measured at low and high frequencies in two separate experiments. Each plot marks the low-frequency and high-frequency cut-off values (also known as –3 dB or half-power points).
If you want to repeat the calibration to obtain more precise values at a frequency of interest, or to check that a sensor is still functioning correctly, you can inject calibration signals into the system using a Güralp digitizer or your own signal generator, and record the instrument's response.
Obtaining copies of the calibration pack
Our servers keep copies of all calibration data that we send out. In the event that the calibration information becomes separated from the instrument, you can obtain all the information using our free e-mail service. Simply e-mail caldoc@guralp.com with the serial number of the instrument in the subject line, e.g.
From: your@email.net To: caldoc@guralp.com Subject: T3A15
The server will reply with the calibration documentation in Word format. The body of your e-mail will be ignored. If you want data for several sensors, put their serial numbers together on the subject line, separated by commas.
22 Issue N
Operator's guide
3.2 Calibration methods
Velocity sensors such as the 3T are not sensitive to constant DC levels, either as a result of their design or because of an interposed high-pass filter. Instead, three common calibration techniques are used.
Injecting a step current allows the system response to be
determined in the time domain. The amplitude and phase response can then be calculated using a Fourier transform. Because the input signal has predominantly low-frequency components, this method generally gives poor results. However, it is simple enough to be performed daily.
Injecting a sinusoidal current of known amplitude and
frequency allows the system response to be determined at a spot frequency. However, before the calibration measurement can be made the system must be allowed to reach a steady state; for low frequencies, this may take a long time. In addition, several measurements must be made to determine the response over the full frequency spectrum.
Injecting white noise into the calibration coil gives the response
of the whole system, which can be measured using a spectrum analyser.
You can perform calibration either using a Güralp DM24 digitizer, which can generate step and sinusoidal calibration signals, or by feeding your own signals into the instrument through a hand-held control unit.
Before you can calibrate the instrument, its calibration relays need to be activated by pulling low the CAL ENABLE line on the instrument's connector for the component you wish to calibrate. Once enabled, a calibration signal provided across the CAL SIGNAL and SIGNAL GROUND lines will be routed through the feedback system. You can then measure the signal's equivalent velocity on the sensor's output lines. Güralp Hand-held Control Units provide a switch for activating the CAL ENABLE line.
3.3 Calibration with Scream!
Güralp digitizers provide calibration signal generators to help you set up your sensors. Calibration is most easily done through a PC running Güralp's Scream! software.
Depending on the digitizer type, sine-wave, step and broadband noise signal generators may be available. In this section, broadband noise
July 2009 23
CMG-3T
calibration will be used to determine the complete sensor response in one action. Please refer to the digitizer's manual for information on other calibration methods.
1. In Scream!'s main window, right-click on the digitiser's icon and
select Control.... Open the Calibration pane.
2. Select the calibration channel corresponding to the instrument, and choose Broadband Noise. Select the component you wish to calibrate, together with a suitable duration and amplitude, and click Inject now. A new data stream, ending Cn (n = 0 – 7) or MB, should appear in Scream!'s main window containing the returned calibration signal.
3. Open a Waveview window on the calibration signal and the returned streams by selecting them and double-clicking. The streams should display the calibration signal combined with the sensors' own measurements. If you cannot see the calibration signal, zoom into the Waveview using the scaling icons at the
24 Issue N
Operator's guide
top left of the window or the cursor keys.
Drag the calibration stream Cn up the Waveview window, so that it is at the top.
4. If the returning signal is saturated, retry using a calibration signal with lower amplitude. Repeat until the entire curve is visible in the Waveview window.
5. If you need to scale one, but not another, of the traces, right-
click on the trace and select Scale.... You can then type in a
suitable scale factor for that trace.
6. Pause the Waveview window by clicking on the icon.
7. Hold down SHIFT and drag across the window to select the calibration signal and the returning component(s). Release the mouse button, keeping SHIFT held down. A menu will pop up. Choose Broadband Noise Calibration. This option runs a script called bbnoisecal which automatically performs appropriate averaging to reduce the effects of aliasing and cultural noise.
8. The script will ask you to fill in sensor calibration parameters for each component you have selected.
July 2009 25
CMG-3T
Most data can be found on the calibration sheet for your sensor. Under Instrument response, you should fill in the sensor response code for your sensor, according to the table below. Instrument Type should be set to the model number of the sensor.
If the file calvals.txt exists in the same directory as Scream!'s executable (scream.exe), Scream! will look there for suitable calibration values. A sample calvals.txt is supplied with Scream!, which you can edit to your requirements. Each stream has its own section in the file, headed by the line [instrument-id]. The instrument-id is the string which identifies the digitizer in the left-hand pane, e.g. GURALP-DEMO. It is always 6 characters (the system identifier) followed by a dash, then 4 characters (the serial number.) For example:
[instrument-id] Serial-Nos=T3X99 VPC=3.153,3.147,3.159 G=1010,1007,1002 COILCONST=0.02575,0.01778,0.01774 CALVPC=3.161 CALRES=51000 TYPE=sensor-type RESPONSE=response-code
9. Click OK. The script will return with a graph showing the responsivity of the sensor in terms of amplitude and phase plots for each component (if appropriate.)
The accuracy of the results depends on the amount of data you have selected, and its sample rate. To obtain good-quality results at low frequency, it will save computation time to use data collected at a lower sample rate; although the same
26 Issue N
Operator's guide
information is present in higher-rate streams, they also include a large amount of high-frequency data which may not be relevant to your purposes.
The bbnoisecal script automatically performs appropriate averaging to reduce the effects of aliasing and cultural noise.
Sensor response codes
Sensor Sensor type code
Units
(V/A)
CMG-5T or 5TD, DC – 100 Hz response
CMG-5_100HZ A
CMG-40T-1 or 6T-1, 1s – 50 Hz response
CMG-40_1HZ_50HZ V
CMG-40T-1 or 6T-1,
1s – 100 Hz response
CMG-40_1S_100HZ V
CMG-40T-1 or 6T-1,
2s – 100 Hz response
CMG-40_2S_100HZ V
CMG-40T-1 or 6T-1,
10s – 100 Hz response
CMG-40_10S_100HZ V
CMG-40, 20s – 50 Hz response
CMG-40_20S_50HZ V
CMG-40, 30s – 50 Hz response
CMG-40_30S_50HZ V
CMG-3T or 3ESP, 30s – 50 Hz response
CMG-3_30S_50HZ V
CMG-40, 60s – 50 Hz response
CMG-40_60S_50HZ V
CMG-3T or 3ESP, 60s – 50 Hz response
CMG-3_60S_50HZ V
CMG-3T or 3ESP, 100s – 50 Hz response
CMG-3_100S_50HZ V
CMG-3T or 3ESP, 120s – 50 Hz response
CMG-3_120S_50HZ V
July 2009 27
CMG-3T
Sensor Sensor type code
Units
(V/A)
CMG-3T, 360s – 50 Hz response
CMG-3_360S_50HZ V
CMG-3TB or 3V / 3ESP borehole,
30s – 50 Hz response
CMG-3B_30S_50HZ V
CMG-3TB or 3V / 3ESP borehole,
100s – 50 Hz response
CMG-3B_100S_50HZ V
CMG-3TB or 3V / 3ESP borehole,
120s – 50 Hz response
CMG-3B_120S_50HZ V
3.4 Calibration with a hand-held control unit
If you prefer, you can inject your own calibration signals into the system through a hand-held control unit. The unit includes a switch which activates the calibration relay in the seismometer and 4 mm banana sockets for an external signal source. As above, the equivalent input velocity for a sinusoidal calibration signal is given by
v = V / 2 π f R K
where V is the peak-to-peak voltage of the calibration signal, f is the signal frequency, R is the value of the calibration resistor and K is the feedback coil constant. R and K are both given on the calibration sheet supplied with the 3T.
The calibration resistor is placed in series with the transducer. Depending on the calibration signal source and the sensitivity of your recording equipment, you may need to increase R by adding further resistors to the circuit.
3.5 The coil constant
The feedback coil constant K is measured at the time of manufacture and printed on the calibration sheet. Using this value will give good results at the time of installation. However, it may change over time.
The coil constant can be determined by tilting the instrument and measuring its response to gravity. To do this, you will need apparatus for measuring tilt angles accurately.
1. Measure the acceleration due to gravity, g, at your location.
2. Tilt the instrument slightly, and measure its attitude and the gain of the mass position output for the component you wish to calibrate.
28 Issue N
Operator's guide
3. Repeat this measurement for several tilt angles.
4. For the vertical sensor, the input acceleration is given by
a = g sin φ, whilst for the horizontal sensor, it is a = g ( 1 – cos φ ).
Calculate the input acceleration for each of the tilt angles used, and plot a graph of mass position output against input acceleration.
5. The gradient of the line obtained gives the sensitivity of the coil (in V/m/s², if g was measured in m/s² and the mass position in V.)
6. The coil constant K is equal to this sensitivity divided by the value of the displacement feedback resistor, as given on the calibration sheet.
July 2009 29
CMG-3T
4 Accessories
4.1 The breakout box
This unit separates the lines in the signal cable, so you can connect a power supply, a recording system, and the hand-held control unit:
You can also use the breakout box to centre, lock and unlock the sensor masses. You will need to provide power through the breakout box's POWER connector to do this (see below.)
To unlock the sensor masses, hold down the ENABLE and
UNLOCK buttons simultaneously for 7 seconds. The BUSY LED
will light. All three masses are unlocked, each in turn. The sensor then automatically moves on to centre the masses, during which time the BUSY LED will flash. When the BUSY LED goes out, the instrument is ready for use.
You should not attempt to move the instrument without re­locking the masses.
To lock the sensor masses, hold down the ENABLE and LOCK
buttons simultaneously for 7 seconds. When the BUSY LED goes
30 Issue N
Operator's guide
out, the instrument is ready for transportation.
To re-centre the sensor masses, hold down the ENABLE and
CENTRE buttons simultaneously for 7 seconds. When the BUSY
LED stops flashing, the centring process has finished. You may need to initiate several rounds of centring before the instrument is ready; when no more centring is required, pressing the ENABLE and CENTRE buttons has no effect.
For more details on the control system, see Section 5.2, page 36.
The standard breakout box is rain resistant but not waterproof. If you intend to use a breakout box in your installation, you should site it away from potential flooding. If this is not possible, a larger unit is optionally available which can be immersed in water. (The 3T itself is, however, completely waterproof.)
4.2 The hand-held control unit
This portable control unit provides easy access to the seismometer's control commands, as well as displaying the output velocity and mass position (i.e. acceleration) on an analogue meter. It takes input from the 26-pin connector at the bottom, and repeats it at the connector on the side for connection to further equipment.
The hand-held control unit can be sited up to fifty metres from the breakout box.
The meter
The meter at the top of the unit allows you to monitor the voltage outputs of the instrument. You can use the knob below to select, for each of the three components, either the mass position output or the
July 2009 31
CMG-3T
velocity output. There is also a RANGE switch allowing you to alter the sensitivity of the meter.
Calibration
The hand-held control unit can be used to calibrate the 3T. To activate the calibration relays, turn the knob to the component you wish to calibrate, and introduce a calibration signal on the CAL SIG banana sockets.
Control commands
You can use the hand-held control unit to centre, lock and unlock the sensor masses.
To unlock the sensor masses, press the ENABLE switch down,
and the LOCK/UNLOCK switch up simultaneously. The BUSY LED will light. All three masses are unlocked, each in turn. The sensor then automatically moves on to centre the masses, during which time the BUSY LED will flash. When the BUSY LED goes out, the instrument is ready for use.
You should not attempt to move the instrument without re­locking the masses.
To lock the sensor masses, press the ENABLE and
LOCK/UNLOCK switches down simultaneously. When the BUSY
LED goes out, the instrument is ready for transportation.
To re-centre the sensor masses, press the ENABLE and CENTRE
switches down simultaneously. When the BUSY LED stops flashing, the centring process has finished. You may need to initiate several rounds of centring before the instrument is ready. When no more centring is required, pressing the ENABLE and CENTRE buttons has no effect.
Note: The ENABLE, LOCK, CENTRE and UNLOCK switches require only a single quick press to initiate the processes. Do not hold them down.
For more details on the control system, see Section 5.2, page 36.
Outputs
The remaining banana sockets provide easy access to the output voltages of the instrument. For each component (vertical, N/S and E/W):
32 Issue N
Operator's guide
the left-hand two sockets expose the balanced differential
outputs representing ground velocity, and
the right-hand socket exposes the mass position (acceleration)
output.
Ground references for each of these voltages are provided at the bottom of the unit. Ensure that you do not connect either side of a differential output to ground.
4.3 Integrated State-of-Health Controller
The CMG-3TD variant (a CMG-3T with an integrated DM24 digitiser) is available with an optional, in-built state-of-health controller.
This option provides a simple means to centre, lock and unlock the masses without the need for a hand-held control unit.
The lid of the device is fitted with four momentary, colour-coded push-buttons, labelled “ENABLE” (black), “CENTRE” (blue), “UNLOCK” (green) and “LOCK” (red) - together with a red LED, labelled “BUSY”.
These buttons are used in a similar manner to those on the break-out box with the exception that it is not necessary to hold the buttons down for seven seconds - one second is adequate.
July 2009 33
CMG-3T
To unlock the sensor masses, hold down the ENABLE and
UNLOCK buttons simultaneously for one second. The BUSY LED
will light. All three masses are unlocked, each in turn. The sensor then automatically moves on to centre the masses, during which time the BUSY LED will flash. When the BUSY LED goes out, the instrument is ready for use.
You should not attempt to move the instrument without re­locking the masses.
To lock the sensor masses, hold down the ENABLE and LOCK
buttons simultaneously for one second. When the BUSY LED goes out, the instrument is ready for transportation.
To re-centre the sensor masses, hold down the ENABLE and
CENTRE buttons simultaneously for one second. When the BUSY LED stops flashing, the centring process has finished. You
may need to initiate several rounds of centring before the instrument is ready; when no more centring is required, pressing the ENABLE and CENTRE buttons has no effect.
For more details on the mass control system, see Section 5.2, page 36.
34 Issue N
Operator's guide
5 Inside the 3T
5.1 The sensors
The horizontal and vertical sensors are similar in design. The inertial mass in both cases consists of a transducer coil and a leaf-spring suspended boom which swings on a frictionless hinge. A triangular spring supports the weight of the mass; in the vertical sensor this spring is pre-stressed with a natural period of around 0.5 seconds, whilst the horizontal sensor has an unstressed flat spring with a natural period of around 1 second. CMG-3T sensors have no spurious resonances below 140 Hz, and weigh around 180 g. The small boom size and stiff springs allow three independent instruments to be mounted within the casing, together with all the associated feedback electronics.
The 3T functions by monitoring the position of each mass with a capacitative position sensor. The three sensors are identical. Signals from the sensors are fed into an electronic processing unit, which is mounted in a screened compartment above the mechanical components (see below for details on the feedback circuitry.)
When the instrument is being transported, the masses are locked securely in their frames so as to relieve strain on the support hinges. This locking is performed by a small motor-driven clamp in response to a signal from the surface controller unit.
Before using the instrument, the boom of the vertical sensor must be levelled and the bases of the horizontal sensors tilted, so that the masses are centred in their equilibrium positions. These adjustments are made by small DC motors controlled remotely.
July 2009 35
CMG-3T
The signal voltages output by the 3T are proportional to ground velocity and are transmitted from the instrument on balanced differential lines. In addition, mass position signals are sent as single­ended circuits referred to analogue ground on the output plug. The 3T also receives control signals, which are used to clamp and un-clamp the masses and to run the motors which level and centre the instrument once in position. Finally, a line is provided for you to apply a calibrating voltage to the force transducers, thereby measuring the deflection sensitivity.
5.2 The control system
The internal operations of the 3T are supervised by a control microprocessor, which drives the mass clamping and centring
36 Issue N
Operator's guide
adjustment motors. It responds to commands sent on three input lines by grounding for 0.2 – 7 seconds.
The signals you can send to the microprocessor are termed LOCK, UNLOCK, and CENTRE. Each command acts on the vertical, N/S and E/ W masses in turn. The microprocessor prevents the system from attempting incompatible actions (e.g. centring when the masses are clamped.)
While a command is taking place, if you are using a Hand-held Control Unit, its BUSY LED will flash; you can use this for diagnostic purposes. See the description of each command for full details.
When no command is active, i.e. all three lines are high, the control microprocessor goes into a power-saving mode. In routine operation, the lines are controlled from the breakout box, Hand-held Control Unit or digitizer. If you send control signals to the 3T manually, you must ensure that the lines are pulled high after sending the signal, or the equipment may be damaged. A “biased-OFF” type switch can be used for this purpose.
LOCK
This command locks the masses and clamps the horizontal sensors by tilting them up to their end stops.
If LOCK is activated when the masses are already locked, the processor will unlock them and attempt to lock again. This is useful if you suspect that the locking procedure has failed.
In detail, the process acts as shown in the following graph. The top three streams are the mass position outputs of each component (Z, N/S and E/W, respectively), whilst the bottom one represents the state of the BUSY LED (up = on).
July 2009 37
CMG-3T
The five-stage process comprises locking the Z, N/S and E/W masses, followed by longer periods tilting the N/S and E/W sensor base to their end stops. During the latter two periods, the position of the N/S and E/ W masses will “flip” to one side.
The BUSY LED is lit during each stage, but goes out briefly between stages, allowing you to follow the progress of the lock.
UNLOCK
This command unlocks the sensor masses and prepares the instrument to begin operating.
If UNLOCK is activated when the masses are already unlocked, the processor will lock them and attempt to unlock again. This is useful if you suspect that the locking procedure has failed.
38 Issue N
Operator's guide
Again, you can use the BUSY LED to monitor the progress of unlocking.
1. The instrument checks to see whether the Z, N/S and E/W masses are locked.
2. The instrument checks the N/S and E/W sensor bases.
3. The Z component is unlocked.
4. The N/S and E/W components are unlocked. These are quicker than the Z component.
5. (see graph below) The N/S sensor base is unlocked, followed by the E/W base. These processes take longer still.
After unlocking, the instrument automatically performs a round
of centring (see below).
CENTRE
This command re-centres the masses. If the masses are clamped, or if the sensor mass positions do not exceed ± 1.2 V, the CENTRE command does nothing. Otherwise, it attempts to zero the output of the vertical, N/S and E/W sensors in sequence by exerting small extra forces on the boom. For the vertical sensor, a motor-driven adjuster presses a small spring lever against the boom until the mass position
July 2009 39
CMG-3T
sensor indicates an offset close to zero. In the case of the horizontal sensors, the sensor frame is tilted on its base plate. Again, the controller monitors the mass position sensor and stops the centring process once it reaches its lowest offset.
This graph shows the entire process of unlocking and centring:
1 – 5. The unlocking process as described above.
6. The BUSY LED pulses to indicate that it is centring the Z component. The mass position output does not change for a while, as it is beyond the range of the output. However, after a few pulses, the position of the Z component comes within range and is centred. The pulses become more brief as this goes on, until a pulse is missed (signifying that no corrective impulse is needed.)
7. The N/S and E/W components follow in the same way, until all three masses are centred and the process completes. The first round of centring has to move the N/S and E/W components all the way from their end stops, whilst the Z component is often closer to the proper position. Because of this, the first Z centring operation takes much less time than the others, and you may not notice it.
After successful centring, the mass position outputs should be in the range 0.1 – 0.8 V. If the centring process leaves the mass position outputs above ±1.1 V, you should start another centring cycle by
40 Issue N
Operator's guide
activating the CENTRE command again. You will probably need to initiate the centring process several times before the masses are adequately centred.
5.3 RS232 control interface
As an option, the 3T can be supplied with an RS232 control interface, which uses three additional pins on the sensor output connector.
The control electronics save power by switching off the sensor feedback electronics whilst the masses are locked. When you unlock the masses, either using the standard logic lines (see above) or over the RS232 interface, the control electronics automatically “wake up” the rest of the sensor.
To connect to the RS232 control interface, attach the sensor to a PC's serial port using the cable supplied, and use a terminal program such as minicom (on Linux) or HyperTerminal (on Windows) to connect to the port. Set the baud rate to 4800, with 8 data bits, no parity bit, 0 stop bits and no flow control.
Now power up the sensor. You should see the message PWR OK indicating that it is ready to receive commands. Each command is a single character. When you enter a command character, the controller will echo it back to your terminal so you can see it.
After around 10 s of inactivity, the control electronics will go into power-saving mode. You will need to wake up the controller by sending any character before you can issue commands. This character will not be interpreted as a command, and the controller will not echo it back to your terminal.
H Help
Sending H causes the controller to reply with a short list of available commands.
[U]nlock [C]entre [L]ock [S]tatus [Q]uit [H]elp
U Unlock
Sending U will start an unlock sequence, exactly as if you had activated the Unlock line on the output port. The sensor will automatically perform one round of centring after it has been unlocked, and the electronics will be fully activated.
If the unlocking process finishes normally, the controller will reply
July 2009 41
CMG-3T
with OK. If there was a problem, such as one or more components failing to unlock or centre, the controller will report NG (“no good”.)
L Lock
Sending L will start a lock sequence, as if you had activated the Lock line on the output port (including unlocking the sensor before locking, if necessary.) The sensor electronics will be placed in power saving mode as soon as the masses are successfully locked.
If the locking process finishes normally, the controller will reply with OK. If there was a problem, the controller will report NG.
C Centre
Sending C will start a centring sequence, as if you had activated the Centre line on the output port. When the process finishes, the controller will reply with OK, or NG if there was a problem. You may need to issue the command several times before the masses are adequately centred.
If you try to centre the masses when they are locked, the controller will reply with OK immediately but take no action.
S Status
Sending S causes the controller to reply with the current instantaneous mass positions, e.g.:
V: +5 N: +8 E: -10
The mass positions are measured by the controller's on-board ADC, which has a nominal range of ±127 counts. For accurate mass position information, you should use the analogue lines provided elsewhere on the output connector.
Q Quit
Sending Q ends your command session, and puts the controller into power-saving mode. To issue further commands, you will need to wake up the controller by sending any character. This character will not be treated as a command, and will not be echoed back to your terminal.
5.4 The feedback system
The output from a modern broadband seismometer does not depend on the natural characteristics of the instrument. Instead, the period
42 Issue N
Operator's guide
and damping of the sensor is completely determined by a feedback loop which applies a force to the sensor mass opposing any motion. The force required to restrain the movement of the mass can then be used to measure the inertial force which it exerts as a result of ground motion.
All CMG-3 series units are based on these general principles. The capacitative position sensor for each mass produces a voltage proportional to the displacement of the mass from its equilibrium position. After amplification, this voltage generates a current in the force transducer coil which tends to force the mass back toward equilibrium. The feedback loop has a sufficiently high gain to cancel the motion of the mass. Since the mass is not moving, the forces acting on it must be balanced; the feedback voltage then directly measures the force, and hence the acceleration, which is being applied to the mass. The feedback loop introduces a phase shift, which must be carefully controlled if the instrument is to remain stable over its entire frequency range. This is achieved using compensation components in the forward and feedback paths.
Force feedback seismometers of this type rely on the assumption that the force transducer produces a field of constant strength. The magnetic circuit and magnet/pole assembly in the 3T are designed so that the field strength from the feedback transducer is constant over large deflections and current levels. Tests have shown that the mechanical suspension system and electronics of a 3T instrument are linear to better than 107 dB (source: measurements made at ASL during evaluation for the USGS National Network.)
In a feedback seismometer with a displacement transducer, it is essential to monitor the acceleration output. This provides the position of the displacement transducer and therefore also the mass position, as the displacement transducer is attached to the sensor inertial mass. The sensor should always be operated with the displacement transducer centred or nulled, so that the response to input acceleration is linear.
There are two types of feedback system which can be used in a 3T instrument, known as hybrid and conventional-response feedback.
These are both described below.
Hybrid feedback
The hybrid feedback method of operation is illustrated by the following schematic diagram:
July 2009 43
CMG-3T
The hybrid feedback circuit contains a single capacitor in parallel with a resistor, resulting in a single dominant pole. Below this frequency, which can be varied from the standard 30 seconds, the response of the seismometer is flat to ground acceleration. Above it, the response is flat to velocity. Hybrid-feedback systems provide a stable response, particularly for portable systems, with a high saturation level at high frequencies and a high dynamic range at long periods.
An active low-pass filter provides a high-frequency cut-off point at a frequency you specify. Without the filter, the velocity response is flat up to 100 Hz. Outside the feedback loop there is an active high-pass filter with a corner frequency of 0.01 Hz (100 s) or 0.005 Hz (200 s), which serves to remove any DC offsets.
Conventional-response feedback
44 Issue N
Operator's guide
The conventional-response feedback system has an additional parallel feedback circuit, consisting of a non-inverting integrator in series with a resistor. This arrangement results in two poles at specified frequencies. The velocity response of a conventional-response system is defined by a transfer function identical to that of a conventional long-period sensor with a damping constant of 0.707 (1/√2) ζ
The seismometer can be supplied with an equivalent resonant frequency of 0.033 Hz (30 seconds), 0.01 Hz (100 seconds) or 0.0083 Hz (120 seconds) as required. An active low-pass filter provides a high­frequency cut-off point at a frequency you specify.
Comparisons
The figures below plot the comparative response of a conventional velocity output broadband sensor and a hybrid output broadband sensor. The first graph shows the response in terms of output against input acceleration in units of V/m/s², whilst the second is plotted in terms of output against input velocity, in V/m/s.
July 2009 45
CMG-3T
46 Issue N
Operator's guide
6 Connector pin-outs
6.1 Sensor and control unit pin output
Models with a 26-pin mil-spec plug (02E-16-26P) have the following pin assignments. The RECORDER and HCU connectors on the breakout box are connected directly to the SENSOR connector, and therefore behave identically.
Pin Function
A Velocity +ve, vertical channel
B Velocity –ve, vertical channel
C Velocity +ve, N/S channel
D Velocity –ve, N/S channel
E Velocity +ve, E/W channel
F Velocity –ve, E/W channel
G Mass position, vertical channel
H RS232 ground (RS232 control interface option)
J Mass position, N/S channel
K BUSY LED
L Mass position, E/W channel
M Power –ve (not used in sensors with internal DC-DC converters)
N Signal ground
P Calibration signal (all channels)
R Calibration enable, vertical channel
S Calibration enable, N/S channel
T Calibration enable, E/W channel
U Centre
W Unlock
X Lock
Y Logic ground
Z RS232 transmit from instrument (RS232 control interface option)
a RS232 receive at instrument (RS232 control interface option)
b Power ground
July 2009 47
CMG-3T
c Power +ve
In sensors with the RS232 control interface option, the RS232 ground line (pin H) is isolated from the other ground lines (pins Z and a) using 10 nF decoupling capacitors.
6.2 Sensor output (“D”-type connector option)
As an option, sensors can be equipped with a 26-pin high-density “D”­type plug in place of the mil-spec connector.
Pin Function
1 Mass position, N/S component
2 Lock
3 BUSY LED
4 Logic ground
5 Mass position, E/W component
7 RS232 transmit from instrument (RS232 control interface option)
8 Signal ground
9 RS232 receive at instrument (RS232 control interface option)
10 Calibration signal
11 Calibration enable, N/S component
12 Calibration enable, vertical component
13 0 V power
14 Velocity –ve, vertical component
15 Calibration enable, E/W component
16 Velocity +ve, vertical component
17 + 10 – 36 V power
18 Velocity –ve, N/S component
19 Centre
20 Velocity +ve, N/S component
21 Factory use only, do not connect
22 Velocity –ve, E/W component
23 Velocity +ve, E/W component
24 Unlock
25 Mass position, vertical component
48 Issue N
Operator's guide
26 RS232 ground (RS232 control interface option)
In sensors with the RS232 control interface option, the RS232 ground line (pin 26) is isolated from the other ground lines (pins 4 and 8) using 10 nF decoupling capacitors.
6.3 Sensor output (waterproof connector option)
Waterproofed models have a 32-pin waterproof connector in place of the 26-pin output connector. These instruments are supplied with an adapter cable for connecting to a standard digitizer input.
The pin connections are:
Pin Function
1 Mass position, N/S channel
2 Lock
3 BUSY LED
4 Logic ground
5 Mass position, E/W channel
6 Power –ve (not used in sensors with internal DC-DC converters)
7 RS232 transmit from instrument (RS232 control interface option)
8 Signal ground
July 2009 49
CMG-3T
9 RS232 receive at instrument (RS232 control interface option)
10 Calibration signal (all channels)
11 Calibration enable, N/S channel
12 Calibration enable, vertical channel
13 Power ground
14 Velocity +ve, vertical channel
15 Calibration enable, E/W channel
16 Velocity –ve, vertical channel
17 Power +ve
18 Velocity +ve, N/S channel
19 Centre
20 Velocity –ve, N/S channel
22 Velocity +ve, E/W channel
23 Velocity –ve, E/W channel
24 Unlock
25 Mass position, vertical channel
26 RS232 ground (RS232 control interface option)
In sensors with the RS232 control interface option, the RS232 ground line (pin 26) is isolated from the other ground lines (pins 4 and 8) using 10 nF decoupling capacitors.
6.4 Breakout box POWER connector
This is a standard 10-pin mil-spec plug (02E-12-10P).
Pin Function
A 0 V
B +12 V DC supply
H –12 V DC supply
50 Issue N
Operator's guide
7 Specifications
Hybrid sensors Velocity output bandwidth 0.1 – 50 Hz
High pass filter output flat to acceleration
0.01 Hz – spec*
High pass filter output flat to velocity
spec – 50 Hz*
Mass position output DC – 0.1 Hz
Velocity sensitivity 1400 V/m/s
Acceleration sensitivity 2000 V/m/s²
Velocity sensors Velocity output bandwidth spec – 50 Hz*
Mass position output DC – spec Hz*
Velocity sensitivity 2 × 750 V/m/s
Mass position sensitivity 1000 V/m/s²
Controls
Mass locking and unlocking
remotely operated
Mass centring
automatic, microprocessor controlled
Mechanics and electronics
Sensors
3 orthogonal sensors, each 0.180 kg
Lowest spurious resonance above 140 Hz
Total weight 13.5 kg
Sensor transducer type
capacitive displacement
Feedback transducer type magnet/coil
Connector pressure tight
Temperature range with masses locked
–35 to +75 °C
Operational temperature range
–20 to +65 °C**
Power Supply 10 – 36 V
Current at 12 V DC 75 mA†
Current at 12 V DC during calibration
100 mA†
Current at 12 V DC during 330 mA†
July 2009 51
CMG-3T
centring (average)
Current at 12 V DC during locking and unlocking
490 mA†
*spec refers to the quoted frequency response value, e.g., for a “30 s” sensor, the value of spec would be 30 s = 0.033 Hz.
**Temperatures below –20 °C may be accommodated with additional care. Please consult Güralp Systems for advice.
†Because centring, locking, and unlocking consume varying amounts of power, it is recommended that you use a power supply capable of delivering 1 A at 12 V.
52 Issue N
Operator's guide
8 Revision history
2009-07-06 N Added integrated SOH controller
2006-10-19 M Added waterproof connector pin-out
2006-08-14 L Added RS232 command interface
2006-06-06 K Moved BB to “Accessories”
2006-02-20 J Corrections; added revision history
2004-10-29 H New document
July 2009 53
Loading...