Gunson G4125 Handbook

DDiiggiittaall EExxhhaauusstt CCOO AAnnaallyysseerr wwiitthh PPuullssee PPuummpp
GASTESTER
Handbook G4125
www.gunson.co.uk
33
Digital Exhaust CO Analyser with Pulse Pump
INDEX Page
1. Contents 4
2. Assembly 4
3. Descriptions and controls 5
4. Preparations before use 5
5. Instructions for use 6
6. Further information 7
7. Common problems 8
8. Multiple carburettors 9
9. Carburettor Adjustments – General Information 10
10. Single Fixed Venture Type 12
11. Single Variable Venture Type 12
12. Idle Adjustment 13
13. Technical Specification 18
14. General Warning 18
15. Warranty 19
GGuunnssoonn GGaasstteesstteerr
44
Gastester is an exhaust gas analyzer that works on the “Hot Wire” or “Thermal Conductivity principle. According to this principle, the thermal conductivity of exhaust gas varies in proportion to the amount of carbon monoxide present.
CO Function
Calibrated range: 0-10% CO
(indicates uncalibrated to 20% C)
Accuracy: +/- 0.5% CO Typical
(Throughout the indicated range 0.5% CO to 6.5% CO)
1. Contents
1. Gastester instrument including collector box and pulse pump/water trap assembly
2. Metal exhaust probe with retaining springs
3. Flexible pipes (3)
4. Instruction handbook
2. Assembly
Warming the various pipes will assist assembly. Only the connection to the vented collector box inlet needs to be pressed fully home.
Press the shortest pipe onto the lower boss of the Pulse Pump/Water Trap as shown in the diagram Figure 1. This is the automatic water drain.
Connect the other plastic pipe between the Pulse Pump outlet and the collector box inlet. (This connection is intentionally vented at the inlet boss). The connection to the collector box inlet needs to be pressed fully home.
GGuunnssoonn GGaasstteesstteerr
FFiigguurree 11
55
3. Descriptions & Controls
DISPLAY This is calibrated in volume percent carbon monoxide (CO%)
CALIBRATION CONTROL This is used to set the reading of the display to show 2.0% at
the start of the tests, before the probe is inserted into the exhaust pipe. (NB: The calibrate position represents what the instrument should register when the probe is in the air. It is coincidental that air should measure the same as exhaust gas with 2% CO). When the probe is subsequently inserted into the exhaust pipe, the display of the instrument may increase or fall from the calibrate condition.
TWO CORE CABLE WITH CLIPS (for connecting to the car battery)
EXHAUST PROBE. The metal pipe is for insertion into the exhaust pipe and is retained
in position by the metal springs which press against the inside of the exhaust pipe.
NB In use the pipe from the exhaust probe should slope down continuously to the Pulse Pump/Water Trap so that water runs down and may be automatically expelled from the drain pipe.
4. Preparations before use
Any service maintenance such as air filter renewal, tappet adjustment, carburetor maintenance, ignition maintenance including ignition timing, should be carried out before setting the carburetor or fuel injection mixture adjustment. Petrol mixture setting should be the FINAL ADJUSTMENT in any engine tuning work.
Carbon monoxide is an EXTREMELY POISONOUS gas and any work on the car with the engine running should therefore be carried out in the open air. Take care not to breathe in gases when using the gas tester.
Study the workshop manual for the particular car (or consult the information given later in this booklet) to identify the correct adjustment screws, that control the mixture strength and the idle speed. Make a note of the initial position of the adjustments before commencing work.
The car should be thoroughly warm before the tests begin. It is not enough to leave the car to warm while parked. The car should be taken for a drive and the tests not commenced until water temperature, exhaust system and engine are running at normal temperature.
Have to hand the correct tools for making the necessary adjustments. If the manual recommends a CO level at a particular engine speed, then a tachometer would be an advantage.
GGuunnssoonn GGaasstteesstteerr
FFiigguurree 22
5. Instructions for use
Ensure that the car is thoroughly warm as mentioned above, that it is parked in a convenient position in the open air, with the handbrake applied.
NB: THE INSTRUMENT SHOULD NOT BE PLACED DIRECTLY IN THE EXHAUST STREAM. CLEAN AIR IS REQUIRED IN THE REFERENCE CELL FOR ACCURATE READINGS.
1. Place the Gastester Professional on a convenient flat surface close to the vehicle’s exhaust pipe outlet. Switch off the engine temporarily while making connections. Connect the RED (+) and BLACK (-) clips to the vehicle’s 12 V battery (Do not use
this Gastester with 6v or 24v batteries)
2. The probe should be fitted to the instrument but do not insert the exhaust probe into the exhaust pipe at this stage
3. Restart the engine and allow a minimum period of eight minutes with the instrument probe in the air. Set the calibration control to achieve 2% on the CO range and observe the display for a further 2 minutes minimum to ensure that the reading has stabilized.
4. IMPORTANT if it has not stabilized allow a further 2 minutes before attempting to measure exhaust CO level.
5. During the warm up period other ranges will give accurate readings and it is therefore useful to check and adjust engine idle RPM.
6. Switch to the CO range with the probe in air.
Do not insert the exhaust probe into the exhaust pipe at this stage
7. Use the rotary calibration control to carefully reset to the CALIBRATE reading of
2.0% CO. The instrument displays two decimal places on the CO range as a result of
the high resolution display. The last decimal place is insignificant and when setting to 2% in air the user should not be concerned if 2.00% exactly can not be obtained.
Having set the CALIBRATE condition do not move the instrument, or move to a different location during subsequent CO tests.
8. When the probe is subsequently inserted into the exhaust pipe, the display of the instrument may increase or fall from the CALIBRATE condition, depending on whether the exhaust has more than, or less than, 2% CO.
9. Ensure that the engine is set to the idle RPM stated by the manufacturer. Now insert the probe into the vehicle’s exhaust pipe to a minimum of 3/4 of its length, i.e. 8” or 20cm. In order for the automatic water drain to function, the probe pipe should fall continuously from the exhaust end to the inlet end to allow water droplets to run down. Otherwise the water will collect at the lowest point and will have to be drained manually.
10. Wait for a period of 15 seconds for the meter to respond and a further 1 minute to stabilize (the reading may overshoot before returning to a steady value, particularly during the initial measurement).
11. Make a note of the reading and observe the display for a further one or two minutes to ensure that the reading is indeed steady and within tolerance for the vehicle concerned. If the reading is not between the manufacturer’s recommended maximum and minimum, or is not below that specified as a legal requirement, then adjustment of the carburettor or fuel injection system will be required.
66
GGuunnssoonn GGaasstteesstteerr
Loading...
+ 14 hidden pages