Gundlach PrimoX, PrimoX 11.12.00, PrimoX 11.12.10 Series Manual

Manual for PrimoX microscopes
11.12.00 PrimoX, binocular
11.12.10 PrimoX, trinocular
Specifications:
Total Magnification: 40x - 1000x.
plan 10x N.A. 0.25 plan 20x N.A. 0.40 plan 40x N.A. 0.65 plan 100x N.A. 1.25 plan
Mechanical stage: Size:
140 mm x 210 mm. X-Y range: 75 mm x 50 mm. Graduation: 1 mm Focusing: Graduation for fine
focus is 0.002 mm. Abbé-
condenser: N.A. = 1.25.With
built-in diaphragm
and filter holder.
Extra swing-out lens
that is used with the
4x objective.
Construction of the microscope:
Objectives:
All objectives are produced accor­ding to DIN-standard. 40x and 100x objectives have a spring mount to avoid breakage of specimens and objectives. The five-hole objective revolver makes it easy to change between the different objectives.
Focusing knobs and stage stop adjustment:
The coaxial coarse and fine adjust­ment knobs make it easy to attain a sharp image. The biggest knobs are used for the coarse adjustment and the smallest knobs for the fine adju­stment.
Between the left large focusing knob and the column is there also a ring. With this ring it is possible to adjust the friction when turning it. In this way it can be adapted to diffe­rent users’ wishes.
On the right side (between the right large focusing knobs and the colu­mn) is a stop ring with a small peg. This is called the stage stop. With this ring it is possible to settle the upper limit of the stage movement to protect the specimen from touching and/or damaging the ob ­jectives. Furthermore, the stop ring will make you able to focus more quickly. The stage stop ring is loo­sened by turning the ring towards you/down, when you are sitting in front of the microscope.
Mechanical stage:
The specimen is placed on the mechanical stage by pulling back the movable handle of the speci­men holder. Gently release the handle and the specimen is placed securely. The mechanical stage can then be moved in the X and Y-axis with the coaxial mounted knobs under the right side of the mechani­cal stage.
PrimoX microscopes, binocular and trinocular VE111200_10
Vareno. 111200_10 Rev. 19.06.2007
Gundlach A/S Tel. +45 8694 1388 gundlach@gundlach.eu Silkeborgvej 765 · DK-8220 Brabrand Fax +45 8694 2486 www.gundlach.eu
Condenser and iris diaphragm:
The condenser can be focused by turning the condenser knob (left side) up or down, until the best picture is being obtained. At high magnifications the position should normally be at the top.
The condenser has a built-in iris with an iris handle for adjustment. Closing the iris diaphragm decrea­ses the light but increases the con­trast and depth of sharpness of the image. Adjustment of the iris diap­hragm might be necessary when switching between objectives.
The swing-out lens is used when the 4x objective is used. When using the other objectives this lens is simply swung out.
It is important that the condenser is aligned with the light path to obtain the best resolution and contrast.
Centring of the condenser:
– Use the 4x objective and focus
on a specimen.
– Place the condenser in the top
position.
– Reduce the diameter of the light
cone by turning the field dia ­phragm on top of the light hous­ing. The light is now visible in the eyepieces as a small round dot in the field of view.
– Use the two clamp screws that
are mounted on the condenser holder to centre the light cone in the field of view. The two silver coloured clamp screws are mounted in a 45 degrees angle compared with the other adjust­ment knobs (for example the focussing knobs).
– Turn on the field diaphragm on
top of the lamp housing until the light cone follows the outer edge of the field of view. It is opened more and more until the light is adequate.
– The condenser is now centred.
Field diaphragm:
The field diaphragm is placed on top of the lamp housing, and can be adjusted by turning it. At low magnification it is normally open and it is closed more and more when the magnification increases. The field diaphragm should never be more open than the diaphragm in the condenser.
Filters:
A blue filter can be placed in the fil­ter holder just beneath the iris diap­hragm. This can be an advantage for example at greater magnificati­ons, where it will give a more detai­led image.
Bulb:
The bulb is placed in the lamp hou­sing in the bottom of the main microscope body. It is a 2-pin halo­gen bulb 6 V / 20 Watt (Osram HLX
64250). The bulb must have the correct Voltage and Wattage; othe­rwise it might damage the micros­cope’s electronic.
Fuse:
The fuse is mounted on the back side of the main microscope body. It can be found under the plastic cover labelled “FUSE”.
Connecting the microscope to mains
The microscope is developed for the use with 220-240 V / 50-60 Hz.
General use of the microscope:
– Take out the main body and put
it on a plan and steady place.
– Hold on to the optical head, and
loosen the clamp screw on the right side of the head. Turn the optical head so that it faces the correct way, and fasten it with the clamp screw.
– Put the eyepieces into eyepiece
tubes.
– Put the mains cord in a socket
and turn on the microscope.
– Adjust the light intensity at mid
level with the brightness control placed on the lover left side of the microscope body. The inten­sity of the light can be adjusted to the desired level with this con­trol slider.
– Place a specimen on its place on
the mechanical stage.
– Move the mechanical stage up
and down with the focusing knobs to obtain a clear and sharp image. The mechanical stage can be moved on the Y­axis and X-axis with coaxial mounted control knobs.
– The part of the specimen that
you want to observe is placed in the field of vision with these two control knobs.
– It is necessary to adjust the eye-
piece tubes to fit the distance between the eyes. This is done by pulling or pushing the right and left black eyepiece tubes. The adjustment is finished, when the image is comfortable and shows a completely round image. Notice the number on the graduated scale on the front of the optical head, and use this number to make a quick adjust­ment next time the microscope is used.
– The microscope is then adjusted
to your sight to achieve a good and sharp image. Look through the right eyepiece with your right eye. Use the fine focusing knob to get a sharp image of the spe­cimen. Then look with your left eye through the left eyepiece and adjust the sharpness with the dioptre ring on the eyepiece tube. The dioptre ring is the
PrimoX microscopes, binocular and trinocular VE111200_10
Vareno. 111200_10 Rev. 19.06.2007
Gundlach A/S Tel. +45 8694 1388 gundlach@gundlach.eu Silkeborgvej 765 · DK-8220 Brabrand Fax +45 8694 2486 www.gundlach.eu
SIDE 2 / 5
Loading...
+ 3 hidden pages