Guido Rayos Nestophot Service manual

GUIDO RAYOS X, S.A. 1
TABLE OF CONTENTS
PAG
I Preliminary 7
II Operation 8 III Lamps 9 IV Installation instructions 10
V Maintenan ce and cleaning 12
Electric wiring 13 Spare parts list 14 Warranty 16 Certifications 16
MAN-002 Jan.08 Ed. 2 / Rev. 1
NESTOPHOT
GUIDO RAYOS X, S.A. 2
NESTOPHOT
MAN-002 Jan.08 Ed. 2 / Rev. 1
GUIDO RAYOS X, S.A. 3
DISCLAIMER
The safety devices and other controls provided in this equipment will perform reliably when operated, maintained, and repaired in accordance with the instructions of this manual.
Safety devices must be checked periodically and reset, repaired, or replaced as necessary to ensure that they will operate reliably. Equipment and parts that are broken, missing, badly worn, distorted or contaminated should be replaced with appropriate GUIDO RAYOS X parts. The equipment or its components should not be modified without the approval of the manufacturer.
The Manufacturer disclaims all responsability for any malfunction of this equipment resulting from faulty operation, maintenance or repair, or if any of its components are damaged or modified by anyone other than the manufacturer.
MAN-002 Jan.08 Ed. 2 / Rev. 1
NESTOPHOT
SYMBOLS CHART
SYMBOL DESCRIPTION
In compliance with the requirement of the Med i ca l
GUIDO RAYOS X, S.A. 4
Devices Directive 93/42/EEC. Certified by notified body no. 0120
Alterning current
Atention, read the attached documents
Fuse
Off
On
Patient eyes protection
Earth
Class I, Type B Equipment
Electrical waste. Selective collection
NESTOPHOT
MAN-002 Jan.08 Ed. 2 / Rev. 1
GUIDO RAYOS X, S.A. 5
SPECIFICATIONS
Description Phototherapy
Model NESTOPHOT
Power External AC 230 V 50/60 Hz 1,5 A 100 W
Electric isulation Class : I Type : B
Incoming water protection Ordinary (IPX0)
Operation mode Continuos
Fuses 2 slow fuses 2 A 5x20 mm
Measurement Mounting: height 1,40 m foot : 59x47 y 27x47 Lamp unit: 60x30 cm
Wavelenght range 460 nm (blue specter)
Thermal power 4 lamps 18 w each
Radiation 8,3 ìW/cm 19,7 ìW/cm
2
/nm at 80 cm from light source
2
/nm at 50 cm from light source
Storage and handling Temperature -10ºC to +40ºC conditions Humidity : 10% to 90%
Noise level Less than 40 dBa
MAN-002 Jan.08 Ed. 2 / Rev. 1
NESTOPHOT
GUIDO RAYOS X, S.A. 6
WARNINGS
Generals
It has not been detected anomalies in the operation of the equipment by the effect of external electromagnetic interferences, such as mobile phones, laptops, etc.
It has not been detected any alteration in the operation of the equipment by variation
of the environmental conditions
It has not been detected injurious effects of the equipment on the patient nor on the
operator. Despite of it, the use of protective goggles or screens is recommended.
It has not been detected any perturbation in the water balance of the patient by effect
of the lamps of cold light.
Monitor bilirrubin values according to the indications of the medical facultative.
Do not use NESTOPHOT equipment in combination with flammable products, such as
cleaning or antiseptic agents
Do not store medicaments nor infusion liquids into radiation area
Do not use NESTOPHOT equipment in the presence of gases that can help the com-
bustion
Do not use two or more phototherapy units together nor adjacent
Carry the equipment holding the vertical bar. Do not push the frontal nor the back of
the lamps module to transport.
In chapters
Bilirubin photoisomer can cause detrimental effect in the patient
I
NESTOPHOT uses cold light lamps, so its use do not increase patient temperature
This equipment can be used in combination with a therm otherapy device, without af-
fecting the temperatur e of the patient. So, the use of skin temperature sensor is not indispensable.
- Cover infant's eyes to protect from high light levels.
II
- The fluorescence lamps must be collected according the current legislation.
V
- The substitution of the lamps must be made only by authorized qualified personnel.
- The use of lamps others than recommended (DULUX ® L lamps) can affect both the
security and the effectiveness of the equipment.
NESTOPHOT
MAN-002 Jan.08 Ed. 2 / Rev. 1
GUIDO RAYOS X, S.A. 7
I. PRELIMINARY
Intended use
Bilirrubin is produced primarily in the decomposition of red blood cells; it becomes water-solubles in the liver through combination with glucuronic acid and thus be separated (through the bile).
If an icterus occurs on the first day of the newborn infant, treatm ent to reduce the bilirubin is nec­essary. The phototherapy method has predominated as a reliable method. Through a photo­chemical reaction, light changes insoluble bilirubin into the water-soluble form, permitting removal through the bile and urine without coupling to the glucuronic acids in the liver. For an effective treatment at least 4 ìW/cm
NOTE : bilirubin photoisomer can cause detrimental effect in the patient
2
/nm of radiant flux is required.
NESTOPHOT phototherapy unit em its 8 ìW/cm
2
/nm exactly in 460 nm range to re duce the treat­ment duration and, consequently, increase lamps life. This turns it to an extremely suitable instru­ment for bilirubin treatment
NOTE: NESTOPHOT uses cold light lamps, so its use do not increase patient tem­perature
NESTOPHOT equipment is provided with an electric clock for total operation time and an electric clock for treatment time with zero reset. NESTOPHOT has four castors to its better mobility.
NOTE : this equipment can be used in combination with a thermotherapy device, with­out affecting the temperature of the patient. So, the use of sk in temperature sensor is not indispensable.
MAN-002 Jan.08 Ed. 2 / Rev. 1
NESTOPHOT