SPX GSE
1525 Fairway Circle
Allen Park, MI 48101
USA
Information in this Technical Manual is subject to change without notice due to correction or
enhancement. The information described in this manual is solely the property of GSE. No part of
this manual may be reproduced or transmitted in any form or by any means, electronic or
mechanical, including photocopying and recording and sold for any monetary figure without the
express written permission of GSE.
GSE Locations
SPX GSE
1525 Fairway Circle
Allen Park, MI 48101
U.S.A.
Phone: (800) 755-7875
Changing Targets from the Weigh Mode (Percentage Check-Weighing) ................................ 75
Example (ChecP) ...................................................................................................................... 76
FILL ............................................................................................................................................... 77
Thank you for selecting the GSE Model 350/355 Indicator. The Model 350 and Model 355 continue the GSE
tradition of Excellence in Weighing and Counting Technology. A properly installed and maintained Model
350 or Model 355 will provide many years of reliable, accurate performance.
The chapters of this manual focus on various aspects of the Model 350/355:
Chapter 1: Introduction Features and Specifications
Chapter 2: Indicator Installation Mounting and wiring connections
Chapter 3: Option Installation Installing and setting up options
Chapter 4: Configuration Using the Model 350/355
Chapter 5: Calibration Calibrate the indicator to a scale
Chapter 6: Legal for Trade NTEP and OIML requirements
Chapter 7: Troubleshooting Troubleshooting help and error messages
Standard Features
The Model 350 and Model 355 include built-in functions that you can enable through the Indicator Setup.
Refer to Chapter 4: Configuration for information on the setup and operation of the following standard
functions:
• Check-weighing
• Fill, two-speed, single ingredient
• Batch, three ingredients, single-speed
• Loss-in weight, two-speed emptying
• Independent setpoints
• Parts counting
• Remote key operation
• Selectable, built-in data transmission formats
• Custom data transmissions
• Real time clock with battery backed time and date.
• Numeric keypad (Model 355 only)
• Easily update firmware via the RS-232 port
Specifications
PERFORMANCE
Full Scale (F.S.) Selectable 0 to 999,999
Resolution 20-bit A/D converter, 100,000d displayed,
A/D Conversion 60 Hz
Zero Track 0 – 100% of F.S.
Operating Temperature -10°C to +40°C
Units of Measure lb, kg, oz, g, lb oz
1,000,000d internal
1
Chapter 1
ELECTRICAL
Power Requirement
350 Zinc Die Cast Enclosure
Input (J4): 12 – 26VAC / 12 – 36 VDC
One of four wall mount transformers supplied:
Output: 20VAC, 800mA
IEC 320: (table top transformer with IEC 320 receptacle)
Input: 230VAC, 28W, 50 Hz
Output: 20VAC, 800mA
350/355 Stainless Steel Enclosure
Input (J10): 85 – 265VAC, 0.5A; 50/60 Hz (internal power supply version)
Input (J3): 10 – 36VDC, minimum 0.8A w/no options or 1.25A w/options installed (internal power supply
version)
Excitation Voltage 10 VDC
Excitation Current 180 mA max. / (6) 350Ω bridge
F.S. Signal Input 0.1 mV/V min – 20 mV/V max
Signal Connection 4 lead or 6 lead with sense
PROCESS CONTROL
Outputs See OPTIONS
Remote Input 1 momentary contact closure (100ms minimum)
TARE, PRINT, ZERO
COMMUNICATIONS
Serial RS232 bi-directional serial port
Data Output 16 selectable fixed-format transmissions or
2 custom formats (programmable via RS232)
Protocol Selectable
Baud Rate 150 – 115200 bps
DISPLAY
LED 6-digit weight display, 0.8” (22mm) height
11 LED annunciators for operational status
LCD 6-digit weight display, 1.0” (25.4mm) height
12 LCD annunciators for operational status
Built in LCD status bargraph
KEYPAD
350 Five key, durable elastomeric (rubber)
355 22 key, full numeric, durable elastomeric (rubber)
ENCLOSURE
Zinc Die Cast Black powder coat paint, self-standing on flat surface
Wall/Ceiling Mount Optional stainless steel swivel bracket
Analog Output Module 0–10 VDC/5mA, 0–20mA/10V, 4–20mA/10V, electrically isolated, 16 bit
Relay Output Module Three (3) solid-state 24–280 VAC, 1A with up to 10 updates per second
Wall Mount Kit Swivel bracket/stand for zinc die cast enclosure
Panel Mount Kit Mounts zinc die cast enclosure to user panel
Cutout: 7.33” – 7.45”w x 5.25” – 5.37”h x 2.25”d
186.2 – 189.2mm x 133.4 – 136.4mm x 57.2mm
Two Option Mount Kit Mounts up to two option boards inside the stainless enclosure
Battery Power Supply Kits Two versions: one mounts inside the stainless steel enclosure, the other
Splash Shield Durable adhesive plastic that adheres to the front surface of the stainless
20 mA Current Loop Enables the communication port to be a digital 20 mA current loop port.
Fiber Optic Transceiver
resolution with up to 10 updates per second response time, mounts internal
to enclosure
response time, mounts internal to enclosure or three (3) 3– 60 VDC, 2A
mounts inside the die cast enclosure
steel enclosure model
Model 350/355 stainless steel, installs in the safe area and connects to a
Model 350/355. Allows setpoints and/or analog output options.
Communicates with hazardous area indicator via fiber-optic cable.
Display
The Model 350 and Model 355 indicators are available with either a six digit, 7-segment green LED display,
a six digit, 7-segment black LCD display or a 7-segment backlit LCD display. The Model 350 and Model 355
will display alphanumeric data, but due to the nature of 7-segment LEDs/LCD and the limitation of six digits,
some information is abbreviated.
All segments and annunciators are illuminated for a brief display test upon power up. The current gross
weight is then displayed in default units.
LEDDISPLAY
The LED display is a six digit, 7-segment bright green LED screen with 12 annunciators to show
weight and status information. The SP1, SP2, and SP3 annunciators are red, green, and yellow. Also
there is an annunciator for a third unit under kg. Place the third unit label above the third annunciator
(the third unit is available on both the LED and LCD displays). See page 53 for third unit setup.
Figure 1: Model 350/355 LED Display
3
Chapter 1
LCDDISPLAY
The LCD display is a six digit, 7-segment black LCD screen with 12 annunciators and a bargraph to show
the operational status.
Figure 2: Model 350/355 LCD Display
ANNUNCIATORS
Annunciators provide mode and status information. When illuminated, they indicate the following conditions:
Displayed weight is at center-of-zero (± ¼ display graduation).
Scale is in motion. Motion inhibited transmits and motion inhibited setpoint activation will
be delayed until motion ceases.
Displayed value represents the current gross weight.
Displayed value represents the current net weight.
Displayed value represents the current piece quantity (Count)
The displayed value is represented in pounds.
The displayed value is represented in kilograms.
The displayed value is represented in either ounces, pound ounces or grams.
Illuminates when the battery reaches a low tolerance. LCD only
350 Keypad
The Model 350 offers a sealed 5-button elastomer keypad is used for operator input. Each key is assigned
two distinct functions. Various key combinations are also used. Each key has secondary functions; allowing
an operator to enter target values, perform piece samples, access setup parameters, etc.
A
Figure 3: Model 350 Keypad
4
The Model 350 keypad performs different functions in the Weigh Mode, the Setup Mode, and the Calibration
Mode. Secondary functions for each key allow you to perform additional tasks.
Key Press Weigh Mode Count Mode Setup Mode
Performs a gross zero function
and/or clears an entry in progress.
Hold this key on power-up to
turn on the display regardless
of P420.
Performs a quantity zero
function and/or clears an entry
in progress.
Exits the Setup Mode and/or
answers “NO” to query prompts
and/or clears an entry in progress.
2
1
+
1 +
+ 1
Performs a print function and/or
‘scrolls’ through digits during
setpoint entry.
Toggles between ‘lb’ and ‘kg’
and/or advances cursor to next
entry position.
Performs an auto-tare function (if
enabled) and/or accepts an entry
in progress.
Toggles between display modes
and/or restores power to the
indicator (if auto-shutoff
enabled).
Access Setup Mode. Hold these
keys on power-up to bypass the
remote display mode.
No function. No function.
Absolute clear – clears an entry in
progress and/or clears the value of
a parameter.
Performs a print function
and/or ‘scrolls’ through digits
during setpoint entry.
Toggles through standard
sample sizes and/or begins a
new sample entry.
Performs an auto-tare function
and requests a piece sample
and/or accepts an entry in
progress.
Toggles between display modes
and/or restores power to the
indicator (if auto-shutoff
enabled).
Access Setup Mode. No function.
No function. Clears any entry in progress.
‘Scrolls’ through digits during
data entry.
Advances cursor to next entry
position and/or cycles prompts.
Accepts an entry in progress
and/or ‘scrolls’ through
parameter sub-set selections
and/or answers ‘YES’ to query
prompts.
Advances to the next setup
parameter.
Return to the previous setup
parameter.
+
+
2 + 1
+ 2
Backspace – erases the right-most
digit during data entry.
Extended gross. Extended gross. No function.
Turn off indicator by holding key
for approximately 1 second.
Reverse character scroll during
data entry.
Backspace – erases the rightmost digit during sample entry.
Turn off indicator by holding
key for approximately 1
second.
Reverse character scroll during
sample entry.
Backspace – erases right-most
digit during data entry.
Turn off indicator by holding key
for approximately 1 second.
Reverse character scroll during
data entry.
5
Chapter 1
355 Keypad
The Model 355 keypad performs different functions in the Weigh Mode, the Setup Mode, and the Calibration
Mode. The number keys make entering a tare value or average piece weight easier.
TARGET
ZERO
PRINT
SELECT
TARE
START
STOP
UNITS
SAMPLE
ENTER
Figure 4: Model 355 Keypad
Key Press Weigh Mode Count Mode Setup Mode
"
Performs a gross zero function
and/or clears an entry in progress.
Performs a quantity zero function
and/or clears an entry in progress.
YES
123
4596
78
CLR
0
NO
Exits the Setup Mode and/or
answers “NO” to query prompts
and/or clears an entry in
progress.
2
#
!
(
Performs a print function and/or
‘scrolls’ through digits during
setpoint entry.
Toggles between ‘lb’ and ‘kg’
and/or advances cursor to next
entry position.
Performs an auto-tare function (if
enabled) and/or accepts an entry
in progress. If the tare weight is
known, key in the value and press
#.
Toggles between display modes
and/or restores power to the
indicator (if auto-shutoff
enabled).
Performs a sample. If a number is
keyed in before hand, it will be
used as the sample size. Accepts
an entry.
Clears an entry in progress. Hold
this key on power-up to turn on
the display regardless of P420.
Performs a print function and/or
‘scrolls’ through digits during
setpoint entry.
Toggles through standard sample
sizes and/or begins a new sample
entry.
Performs an auto-tare function
and/or accepts an entry in
progress.
Toggles between display modes
and/or restores power to the
indicator (if auto-shutoff
enabled).
Performs a sample. If a number is
keyed in before hand, it will be
used as the sample size.
Performs a quantity zero function
and/or clears an entry in progress.
‘Scrolls’ through digits during
data entry.
Advances cursor to next entry
position and/or cycles prompts.
Accepts an entry in progress
and/or ‘scrolls’ through
parameter sub-set selections
and/or answers ‘YES’ to query
prompts.
Advances to the next setup
parameter.
Accepts an entry.
Exits the Setup Mode and/or
answers “NO” to query prompts
and/or clears an entry in
progress.
If setpoints are enabled, causes a
process to start or resume. See
&
parameter 5003 on page 58 for
details on enabling the START
function.
6
If setpoints are enabled, causes a
process to start or resume. See
parameter 5003 on page 58 for
details on enabling the START
function.
No function
Key Press Weigh Mode Count Mode Setup Mode
%
If setpoints are enabled, causes a
pause. Press % again to abort
the process. See parameter 5002
on page 58 for details on enabling
the STOP function.
If setpoints are enabled, causes a
pause. Press
the process. See parameter 5002
on page 58 for details on enabling
the STOP function.
% again to abort
No function
$
+
Wake up the indicator if in sleep
mode. Turn on the indicator if
power is off. Access the target
entry mode. See parameter 5000
on page 58 for details on enabling
the TARGET function.
Access Local Setup Mode. Hold
these keys on power-up to
bypass the remote display
mode.
Wake up the indicator if in sleep
mode. Turn on the indicator if
power is off. Access the target
entry mode. See parameter 5000
on page 58 for details on enabling
the TARGET function.
Access Local Setup Mode. No function
When the Model 355 keypad is installed with a new main board, the keypad must be initialized.
Hold down the
“EntHld”. To reestablish the keypad as a Model 350, hold down the
while power is applied by pressing the
( key while power is applied. If the keypad was enabled, the display will show
1 key of the 350 keypad
key.
WEIGH MODE FUNCTIONS
The Model 350 and Model 355 keypads have five primary Weigh Mode functions:
No function
"
2
#
!
Performs a gross zero and selects the gross mode
Initiates data transmission out the communication port
Toggles the units of measure between lb, kg, g, lb oz, oz
Tares any displayed weight and selects the net mode
Toggles the display between GROSS, NET, QUANTITY and setpoint TARGETS (if enabled)
7
Chapter 1
ENTERING A TARE VALUE (MODEL 350)
If a tare value is known, it is possible to enter that value into the tare register. Follow the steps below. P167
must be enabled to use this feature.
1. From the gross or net mode press the
2. Use the
3. Press the
4. Repeat steps 2 and 3 until the desired number is showing on the display.
5. Press the
key to scroll in the first number.
2 key to move to the next digit.
1 key to accept the entered tare value.
key until tare is displayed.
ENTERING A TARE VALUE (MODEL 355)
If a tare value is known, it is possible to enter that value into the tare register. Follow the step below. P167
must be enabled to use this feature. Key in the known tare value and press
the net mode.
If the indicator is being installed in a legal for trade application, refer to Chapter 6 on page 123 for
details on the tare feature.
#. The display will access
8
350/355 Technical Reference
(
)
(
)
CHAPTER 2:INDICATOR INSTALLATION
This chapter contains information necessary for proper installation of the Model 350 and Model 355. Please
review these instructions before installing the indicator.
High voltages may exist within the enclosure. To prevent the risk of electrical shock, ALWAYS unplug the
Model 350/355 when opening the enclosure. Installation and servicing of the Model 350/355 should be
performed only by authorized and qualified service personnel.
For information on installing options, see Chapter 3: Option Installation. For NTEP and OIML details, see Chapter 6: Legal-for-Trade.
IMPORTANT! The 350 Series indicators do not include an on/off switch and therefore must be
installed near a power outlet socket that is easily accessible and in keeping with UL/CSA Safety
Standards.
INFORMATION IMPORTANT! Prendre note que les contrôleurs de serie 350 ne sont pas munis
d'interrupteurs "Marche / Arrêt". Par conséquent, il devront être installés près d'une source
d'alimentation secteur accessible pour demeurer sous les exigences des normes de sécurité
UL/CSA.
Mounting
The standard Model 350 zinc die cast enclosure is a NEMA1 (IP 20) equivalent. The Model 350/355
stainless steel enclosure meets NEMA 4X type specifications.
When choosing a mounting location for the Model 350 die cast, ensure that the unit is not installed in a
washdown area or conductive dust environment.
DESKTOP MOUNTING
The stainless steel enclosure is designed for desktop mounting. The 350 die cast may be installed as a
desktop unit with the optional mounting bracket 24350B-301C0 (refer to Figure 5 for installation details).
When set on a flat surface, the front face is angled for easy viewing. All wiring enters from the rear and can
be secured with the included screw mounted cable ties.
PANEL MOUNTING (DIE CAST ONLY)
The optional panel mount kit allows for the zinc die cast enclosure to be installed in a user panel. The panel
mount kit is only available in die cast not in stainless steel. Allow for 2.00” (57.2mm) depth behind the panel
surface. See Panel Mount Kit on page 18 for more details.
PANEL
5.25” – 5.37”
133.4 – 136.4mm
7.33” – 7.45”
186.2 – 189.2mm
CUTOUT
9
Chapter 2
PERMANENT MOUNTING
The optional mounting bracket allows the zinc die cast enclosure to be securely fastened to another surface.
The bracket is attached to the indicator with two thumbscrews and can be swiveled to an optimal viewing
angle.
Figure 5: Model 350 Zinc Die Cast with Optional Mounting Bracket
OUTLINE DRAWINGS
Figure 6: Model 350 Zinc Die Cast Dimensions
10
350/355 Technical Reference
Figure 7: Model 350 Stainless Steel Dimensions
Figure 8: Model 355 Stainless Steel Dimensions
11
Chapter 2
Wiring
Wiring a load cell or communication port is simple on both the die cast and stainless steel models. Refer to
the information in this section for more details.
STAINLESS STEEL MODEL 350/355
The stainless steel models must be wired through the strain reliefs located on the bottom or rear of the
enclosure and connected to the main board. Refer to Figure 9 and Figure 10.
Figure 9 : Model 355 SS Rear Figure 10: Model 350 SS Rear
GSE INC. a division of SPX Corp.
www.gse-inc.com MADE IN USA
DIE CAST MODEL 350
A description of all wiring terminals is included on the bottom label of the zinc die cast enclosure as shown in
Figure 11.
Figure 11: Model 350 Zinc Die Cast Enclosure Wiring Label
LOAD CELL CONNECTIONS
A high quality braided shield cable with 16 to 24 AWG stranded wire is recommended for load cell or
summing box connections. Secure the cable by the cable tie on the die cast model, or route it through the
strain relief supplied on the back of the internal power supply models (see Figure 12). Either four or six
conductor cables can be used.
12
350/355 Technical Reference
Do not tin the ends of the load cell wire! A terminal connection free from the effects of vibration
and oxidation can be assured only if the load cell terminals securely grip a bare, stranded wire.
When using four conductor cables, (+ Excitation) must be connected together with (+ Sense), and (Excitation) must be connected together with (- Sense) on the Model 350 die cast. Sense jumpers are
standard on the Model 350/355 Stainless Steel (E3 and E4). Utilizing the (+) and (-) Sense leads of six
conductor cables provides compensation for variations in the excitation voltage due to resistance changes in
the cable.
Figure 12: Model 350 Cable Tie on Die Cast & Load Cell Connection (J8)
Figure 13: Model 350/355 Stainless Steel Load Cell Connector (J10)
SERIAL PORT CONNECTIONS
Before connecting to the serial port, consideration should be given to the communication protocol and to any
remote key requirements. For information on remote key operation, refer to Remote Key Operation on page
97. Use Table 1: Serial Port Connections to determine proper wiring to the communication port.
Communication connections are made through the DB9 male connector on the rear of the enclosure or a
pigtailed cable through a strain relief (internal power supply model on J6 comm connector). The cable
should be 20 to 28 AWG with a braided or foil shield for either model type. For maximum noise immunity on
the die cast model, use a mating DB9 connector with a metal hood and a braided shield cable. Ensure that
the braid makes good connection with the hood. The maximum recommended cable length is 50 feet (15
13
Chapter 2
meters). However, much longer connections are possible if using a properly shielded, low-capacitance
cable.
+5V +5 VDC
GND Digital Ground GND
GND Digital Ground GND
RTS Request-to-Send RTS
CTS Clear-to-Send
I/O Do Not Connect
Label
(Die cast encl. only)
VCC
CTS
REMOTE KEY CONNECTION
A remote key may be connected to the Model 350/355 communication port to provide remote activation of
print, tare, or zero functions.
The connection for the remote key input is between pin 1 and pin 6 of the DB9 communication port
connector (see Figure 15 for die cast, Figure 14 for stainless). A two-conductor shielded cable between 28
and 20 AWG is recommended. The input requires a contact closure from a push-button switch, a ‘dry’ relay
contact, a photo-eye, and a proximity sensor or other such device. A closure initiates the operation specified
at P800.
The open circuit voltage across the remote key pins is +5 VDC. A closed switch will conduct about 0.25 mA.
Therefore, a low-voltage switch with gold-plated contacts is recommended. A Mercury-wetted switch will
also work well. A minimum contact duration of 100 ms is required. Once invoked, the selected remote key
operation will not repeat until the contact is released and closed again.
Do not apply an external voltage to remote key terminals! Only a contact closure is
required to activate the remote key input.
Figure 15: Remote Key Connection
(Model 350 die cast)
14
Figure 14: Remote Key Connection
(Model 350 stainless and Model 355)
350/355 Technical Reference
REMOTE DISPLAY CONNECTIONS
It is possible to connect one Model 350/355 indicator to another Model 350/355 and have the remote
indicator display a copy of the master indicator or customize the display of the remote (slave) indicator. The
keypad on both the master and slave indicators will operate. Refer to the drawing below to connect the
master indicator to the remote (slave) indicator. Refer to page 100 to configure the master and remote
(slave) displays. There are several different configurations for Master to remote (slave) indicator connection
and setup.
Figure 16: Master Indicator to Remote (slave) Indicator Connections
POWER CONNECTION
There are four ways to power the Model 350/355.
• 120 VAC wall mount transformer
• 12-36 VDC external source
• 12-26VAC external source
• Internal battery option
120VACWALL MOUNT TRANSFORMER (DIE CAST 350)
An external wall mount transformer is supplied with the die cast Model 350 for connection to 120VAC, 60Hz
power and is plugged into J4 of the 350 die cast main board.
To connect the transformer:
The Model 350/355 does not include an on/off switch. In keeping with UL/CSA Safety Standards it
must be installed near an easily accessible power outlet. Note that the [ON] key does not
connect/disconnect the line voltage. It ‘awakens’ the Model 350/355 from a ‘sleep’ mode.
Do not cut the ground prong off the wall mount transformer!
1. Insert the polarized plug into the rectangular power connector (J4).
2. Press the plug firmly into the hole to ensure it is seated properly.
3. Plug the wall mount transformer into a nearby wall outlet. Use only a three-wire grounded outlet.
15
Chapter 2
EXTERNAL AC OR DCPOWER SUPPLY (DIE CAST 350)
The Model 350 die cast may be powered by an external 12-36 VDC or 12-26 VAC power source. The power
supply should have a minimum current rating of 800mA with no options installed, a minimum current rating
of 1.25A with options installed. Recommended plugs are AMP MTA .156” or MOLEX .156”.
External DC power supply cable GSE P/N 22-30-35459 plugs into the Model 350 die cast unit rear
connector with bare wires on the other end.
An internal option battery module is available, GSE part number 24350B-120B1. Refer to page 35 for
details on installation.
Pin 1 Pin 2 Pin 3 Range
DC
AC L GND N
+
N/C
12-36 VDC
-
12-26 VAC
50/60 HZ
Figure 17: Model 350 Die Cast Power Connections
INTERNAL AC OR DCPOWER SUPPLY (STAINLESS STEEL 350/355)
Figure 18: Model
350/355 Internal Power
Supply Model AC
Connections (J14)
The Model 350/355 internal power supply model may be powered by an external
VAC power source connected to (J14) on the main board. Input power can be 85
– 265VAC, 0.5A; 50/60 Hz. See the internal AC power connector as shown in
Figure 18. The ground wire is connected to a stud on the enclosure.
The Model 350/355 internal power supply model may be powered by an external
10-36 VDC power source connected to (J12) on the main board (also see Battery Power Supply on page 35). The power supply should have a minimum current
rating of 800mA with no options installed, a minimum current rating of 1.25A with
options installed.
Do not connect pins 3 or 4. The mating connector can be purchased from a local
electronics supplier. Recommended connectors are GSE PN: 26-20-3365 (24
AWG) or GSE PN: 26-20-3366 (22 AWG) / AMP PN: 640441-3 (24 AWG) or AMP
PN: 640440-3 (22 AWG). Connector cover GSE PN: 26-20-3389 or AMP PN:
643075-3.
Pin 1 Pin 2 Pin 3 Pin 4 Range
DC
+ -
Low
Batt
Shut
Down
12-36 VDC
Figure 19: Model 350/355 Internal Power Supply Model DC Connections (J12)
16
350/355 Technical Reference
Chapter 3: Option Installation
The capabilities of the Model 350 can be expanded with the use of one or more option kits. This chapter
provides installation procedures for these options.
Model 350/355 Options
The following options are available for the Model 350:
•Swivel Bracket Kit (die cast model) (GSE P/N 24350B-301C0)Allows the indicator to be securely mounted to any surface.
•Panel Mount Kit (die cast model)(GSE P/N 24350B-300C0) Allows the indicator to be easily mounted into existing cabinetry.
Allows the indicator to communicate with hazardous area indicator via fiber-optic cable.
SWIVEL BRACKET
The Model 350 die cast has an optional stainless steel swivel bracket for secure mounting to desks,
tabletops or walls. See the Mounting section beginning on page 9 for instructions on mounting the Model
350/355 using the swivel bracket.
Be sure that all parts shown in Figure 20 are accounted for.
2.Remove the two plastic Heyco plugs.
The plugs are located in the threaded bracket mounting holes on either side of the indicator. Be careful not
to scratch the finish of the indicator when removing these plugs.
Chapter 3
3. Place the two spacers against the bracket and then insert the thumb screws.
The spacer and bracket holes should align with the bracket mounting holes on either side of the enclosure.
PANEL MOUNT KIT
The Panel Mount Kit provides an easy way to mount the indicator into new or existing cabinetry. See Figure
21 for detailed instructions on panel mounting the Model 350 die cast. See also NTEP Panel Mount
Requirements on page 123 for additional panel mounting NTEP requirements.
Figure 21: Model 350 Panel Mount Installation (Die Cast Model)
18
350/355 Technical Reference
To install the Model 350 Panel Mount Kit:
1.Unpack the Panel Mount Kit.
Be sure that all parts shown in Figure 21 are accounted for.
2.Remove the two plastic Heyco plugs.
The plugs are located in the threaded bracket mounting holes on either side of the indicator. Be careful not
to scratch the finish of the indicator when removing these plugs.
3. Place the adhesive gasket around the panel cutout.
This step is optional. The gasket should be adhered onto the outside of the panel and centered around the
cutout.
4. Place the Model 350 through the panel cutout.
This step may require a second person to hold the indicator in place. The indicator should be centered
inside the cutout.
5. Using two small screws and star washers, attach the two panel mount brackets. The center holes of the
brackets should align with the threaded bracket mounting holes on either side of the indicator. The two
brackets should angle away from the indicator. Apply pressure on the mounting brackets in the directions
shown in Figure 21. This will ensure the indicator is drawn firmly against the panel.
6. Using four long screws and four lock-nuts, secure the indicator to the panel. Thread the lock-nuts onto the
screws so that they will not interfere with tightening the screws into the threaded bracket holes. Evenly
tighten the four screws until they are snug. Do not over-tighten. Now thread the lock-nuts down until they
are snug against the bracket.
7. Drill 1/16 (0.0625) diameter pilot hole on the front of the enclosure for the serial number label.
8. Add the serial number to the label.
9. Install the label by tapping the pins through the label into the front enclosure.
Analog Card Connections
The 16-bit electrically-isolated analog output module provides a highly accurate analog signal, proportional
to weight. This signal can be used for interfacing to PLCs, chart recorders, and other such devices.
See Analog Output Setup on page 67 for analog output software configuration details.
The Model 350/355 contains components which could be damaged by Electrostatic Discharge
(ESD) if serviced improperly. Use proper ESD precautions (wear a wrist strap connected to
ground, use grounded work stations, etc.) when opening the enclosure.
High voltages may exist within the enclosure! To prevent the risk of electrical shock, ALWAYS
unplug the Model 350/355 when opening the enclosure. Installation and servicing of the Model
350/355 should be performed by authorized and qualified service personnel only.
Never connect or disconnect option board cables while the indicator is powered. Doing so may
result in circuit board damage.
19
To install the Analog Output Module (350 die cast model):
1. Open the indicator. Remove the four screws from the back of the unit. It may
help to remove the swivel bracket, if installed.
2. Place the back of the indicator, open side down, on a firm surface. Using a
hammer and screwdriver, remove the appropriate knock-out.
It helps to place the screwdriver tip on the knockout pad, rather than in the
groove. The knock-outs do not require much force to remove. If only
installing one option, the left knock-out (as viewed from the rear) should be
removed. A small file may be used to remove any burrs.
3. Flip the back cover over and place the Analog Output Module, component
side down, over the four mounting holes. Install four 10mm screws to a
minimum of 8 in/lb of torque.
Chapter 3
Be sure the cable is already attached before installing the card. Also, be sure
the cable is attached to the left-most connector (as viewed from the
component side of the option board). The second connector is for ‘daisychaining’ another option card. The screws used to mount the option card are
self-tapping and will require added torque when first installed.
4. Attach the loose end of the cable to the serial I/O connector (J7) on the main
board or the open connector of a previously installed option card.
J7 is a 10-pin polarized connector. Be sure the cable is not twisted when installed. If this is a second option
card, route the cable to the open connector of the first option card.
5. Reinstall the back cover. Tighten the four screws to a minimum of 8 in/lb torque.
Be sure to avoid ‘pinching’ the cable between the housing halves. Affix all appropriate labels to the back of
the indicator.
To install the Analog Output Module without mounting bracket (Model 350/355 Stainless):
If only one option is being installed in a fiber-optic hub indicator use the following instructions. If two
options are being installed along with a fiber-optic option an option mounting plate is required. Refer
to page 27 for installation instructions.
1.Open the indicator.
Remove the eight screws from the back of the unit.
2. Locate the option mounting holes by lining up the option card with the four holes to the left of the power
supply section on the main board.
Figure 22: Option Board
Installation (Die Cast Model 350)
This is where the first option card should be installed.
20