GS Corpuls3 Safety tests

Document name and location:
S:\Reparaturen\STK_Protokolle\STK_c³_with_test_instruction _EN.docx
Software version 2.1.0 Creation date:
2013-05-07
Release date:
2013-05-15
Originator name:
Carsten Fuchs
Release name:
Udo Franke
corpuls
³
Safety Technical Test, Test instruction and test report 1/16
Warning
Implementation of Safety Technical Tests implies the occurrence of high voltages which may result in severe injuries or death. Only authorized qualified personnel are permitted to carry out Safety Technical Tests.
Attention
Safety Technical Tests should only be carried out by authorized personnel with calibrated measuring equipment and simulators.
Description
Manufacturer or supplier
Model
Part no.
Safety analyzer to perform protec­tive earthing-, device leakage current-, isolation resistance measurement and high voltage test
EKU Elektronik GmbH, D-56291 Leiningen MTP 6000 (or comparable)
Defibrillator/Pacer analyzer
Fluke Biomedical Division, Everett, WA 98203, U.S. Impulse 7000 (or comparable)
Multi parameter/ECG simulator
Fluke Biomedical Division PS420 (or comparable)
Digital multimeter
Fluke Corporation Series 80V (or comparable)
Digital fine manometer
GREISINGER electronic GmbH, D-93128 Regenstauf GMH 3160-07 (or comparable)
Mains charger with magnetic plug, 12 volts DC / 9 amp.
GS Elektromedizinische Geräte G. Stemple GmbH, D-86916 Kaufering
04501.120
Test cable set for Safety Technical
Test Patient Box corpuls³
W0029.14100
Test cable set for Safety Technical
Test Defibrillator Unit corpuls³
W0029.24100
Test cable for Safety Technical
Test Display Unit corpuls³ with LAN
W0029.34100
Paddle holder for leakage current
and H/V testing, corpuls³
W0029.031263
Paddle for leakage current and high
voltage testing corpuls³
(2 pcs. required)
W0029.04326
Test cable Safety Technical Test Charger with magnetic plug corpuls³
W0029.04501
Safety Technical Test
For legal requirements and periods of Safety Technical Tests see chapter 9.4.1 of Operators Instruction.
Implementation of Safety Technical Tests is only accepted when carried out according to both, this test instruction and the Operators Instruction, and at least documented according to this test report.
Test equipment, adapters and facilities
Document name and location:
S:\Reparaturen\STK_Protokolle\STK_c³_with_test_instruction _EN.docx
Software version 2.1.0 Creation date:
2013-05-07
Release date:
2013-05-15
Originator name:
Carsten Fuchs
Release name:
Udo Franke
corpuls
³
Safety Technical Test, Test instruction and test report 2/16
Adapter for leakage current test Masimo Rainbow® Module
W4212.00002
T-connector cable
04325.09
1
Check of external mechanical intactness of the device
1.1
Check of mechanical intactness. The device must neither have form cracks nor fractures. In case of damage, present the device to an authorized national or local GS representative for repair. Swivel the complete device about 30°. The mechanisms must neither be too rough-running nor automatically tilting.
1.2
Check covers and fasteners for completeness.
1.3
Check of shock paddles intactness. The paddles must be free from form cracks in their synthetic material; the helix cable, the electrode plates and the safety electrodes must be free of any damage; rubber caps of the buttons must be available and undamaged, labels must be available. The connector must easily and surely be lockable with the therapy cable. The electrode surfaces should be free of gel leftovers and other contamination.
1.4
Check of attached cables and sensors. Check cleading of cables and sensors for insulating damage; if necessary remove paste-on labels installed by the operator. Continued use of broken cables and sensors is forbidden.
1.5
Check printer assembly, lid and locking mechanism. The printer lid should easily release and lock.
1.6
Check of labels, warning and safety instruction plates for completeness and readability. Check of type plates for readability. Note the serial numbers of each single component to be tested in the test report sheet.
1.7
Check interlocking mechanisms between the single components. Therefore unlock and relock the components. The interlocking mechanism must be free of damage.
1.8
Check the magnetic connector contact panels of the of the single components for damage and pollution. Clean the panels if necessary.
1.9
Shaking test. Shake each of the single components to detect loose parts inside.
Warning
Do not open the device! Inner parts may be under high voltage. Disregarding may cause severe injuries or death! In case of defect contact your authorized national or local GS representative for inspection and potential repair.
Test instruction for Safety Technical Test
Document name and location:
S:\Reparaturen\STK_Protokolle\STK_c³_with_test_instruction _EN.docx
Software version 2.1.0 Creation date:
2013-05-07
Release date:
2013-05-15
Originator name:
Carsten Fuchs
Release name:
Udo Franke
corpuls
³
Safety Technical Test, Test instruction and test report 3/16
2
Protection against electrical shock hazard
Attention
During following measurements, all single components must be separated from electrical power supply (power supply pack). Take the batteries out of the single components!
Attention
Performance of both the leakage current measuring of application parts (item 2.6) and the high voltage test (item 2.7) requires the application of the specified GS­specific adapters and facilities. Using other adapters and facilities may result in device damage and termination of warranty rights!
2.1
Check of protective earthing of the external power supply pack (IEC 62353:2007, item 5.3.2.2, image 1, protection class I). Connect the mains plug to the safety tester and measure with the probe at the minus pole of the corpuls magnet connector using the test cable P/N W0029.04501 to activate the switch inside the male connector. Testing current must be not less than 200 mA and must not exceed a limit of 10 A. This step must not be done for power supplies with part number 04500.1202 and 04501.1201.
2.2
Measuring of device leakage current of external power supply pack (IEC 62353:2007, item 5.3.3.2.3, image 4, protection class I) (direct measurement).
2.3
Measuring of insulation resistance of the external power supply pack (IEC 62353:2007, item 5.3.4, image 9). Testing voltage 500 V (DC).
2.4
Measuring of line resistance between potential equalization plug and minus pins of power contact field of the Defibrillator Unit (by using a Hirschmann clip or an alligator clip). Remove cover if required. Testing current must be not less than 200 mA and must not exceed a limit of 10 A. The potential equalization plug may be covert by a plastic cover.
2.5
Measuring of line resistance between safety electrodes of connected shock paddles and potential equalization plug by using a resistance measuring gauge. Testing current must not exceed a limit of 1 A.
2.6
Measuring of leakage current of application parts (IEC 62353:2007, item
5.3.3.3.1, image 6, protection class I). Application parts at which is not currently measured, must be grounded (to the potential equalization plug of the Defibrillator Unit or to minus contact of the corpuls magnet connector of the Patient Box). Measure between:
2.6.1
High voltage output of therapy cable/therapy socket with connected shock paddles and potential equalization plug (connect the shock paddles to the therapy cable/therapy socket and put on the adapter P/N w0029.031263).
2.6.2
High voltage output of therapy cable/therapy socket with connected T-connector cable (P/N 04325.09) and potential equalization plug. The measuring is only required if internal shock electrodes (shock spoons) are applied.
2.6.3
High voltage output of test signal connector and potential equalization plug.
Document name and location:
S:\Reparaturen\STK_Protokolle\STK_c³_with_test_instruction _EN.docx
Software version 2.1.0 Creation date:
2013-05-07
Release date:
2013-05-15
Originator name:
Carsten Fuchs
Release name:
Udo Franke
corpuls
³
Safety Technical Test, Test instruction and test report 4/16
2.6.4
Short-wired test contact pins of the paddle trays and potential equalization plug (if the test contact pins were replaced by accessory bags Part no. 04322.50 and 04322.51, the original condition must be restored for the time of measurement).
2.6.5
ECG inputs (4-pole and 6-pole connected) and minus contact of the corpuls magnet connector of the Patient Box.
2.6.6
SpO
2
input and minus contact of the corpuls magnet connector of
the Patient Box.
2.6.7
CO
2
input and minus contact of the corpuls magnet connector of the
Patient Box.
2.6.8
Input Temp1 and minus contact of the corpuls magnet connector of the Patient Box.
2.6.9
Input Temp2 and minus contact of the corpuls magnet connector of the Patient Box.
2.6.10
Input IBP1 and minus contact of the corpuls magnet connector of the Patient Box.
2.6.11
Input IBP2 and minus contact of the corpuls magnet connector of the Patient Box.
2.6.12
CPR input and minus contact of the corpuls magnet connector of the Patient Box.
2.6.13
LAN connector and minus contact of the corpuls magnet connector of the Display Unit.
2.7
High voltage test of application parts (IEC 60601-2-4:2002, item 20, image
107). Application parts at which is not currently measured, must be grounded (to the potential equalization plug of the Defibrillator Unit or to minus contact of the corpuls magnet connector of the Patient Box). Set the current limiter of the high voltage generator to 500 µA. No high voltage flashover must appear; no release of the current limiter. Apply 5 kV (DC) for 30 sec each between:
2.7.1
High voltage output of therapy cable/therapy socket with connected shock paddles and potential equalization plug (connect the shock paddles to the therapy cable and put on the adapter P/N w0029.031263).
2.7.2
High voltage output of test signal connector and potential equalization plug.
2.7.3
Short-wired test contact pins of the paddle trays and potential equalization plug (if the test contact pins were replaced by accessory bags Part no. 04322.50 and 04322.51, the original condition must be restored for the time of measurement).
2.7.4
ECG inputs (4-pole and 6-pole connected) and minus contact of the corpuls magnet connector of the Patient Box.
2.7.5
SpO
2
input and minus contact of the corpuls magnet connector of
the Patient Box.
2.7.6
CO
2
input and minus contact of the corpuls magnet connector of the
Patient Box.
2.7.7
Input Temp1 and minus contact of the corpuls magnet connector of the Patient Box.
2.7.8
Input Temp2 and minus contact of the corpuls magnet connector of the Patient Box.
Document name and location:
S:\Reparaturen\STK_Protokolle\STK_c³_with_test_instruction _EN.docx
Software version 2.1.0 Creation date:
2013-05-07
Release date:
2013-05-15
Originator name:
Carsten Fuchs
Release name:
Udo Franke
corpuls
³
Safety Technical Test, Test instruction and test report 5/16
2.7.9
Input IBP1 and minus contact of the corpuls magnet connector of the Patient Box.
2.7.10
Input IBP2 and minus contact of the corpuls magnet connector of the Patient Box.
2.7.11
CPR input and minus contact of the corpuls magnet connector of the Patient Box.
2.7.12
LAN connector and minus contact of the corpuls magnet connector of the Display Unit.
3
General functions (charging, network connection)
For all following checks reinstall the batteries to the single components. The single components will automatically switch on. Shutdown the single components by operating the PWR on/off buttons and reassemble them to a compact device.
3.1
Switch on the compact device by operating the PWR button of the Display Unit. Wait for boot-up (approx. 20 sec) then check network status.
3.2
Separate Defibrillator Unit, and then separate Patient Box from Display Unit (modular mode). Check network status.
3.3
Check network status on the LCD of Patient Box.
3.4
Connect power supply pack to corpuls magnet connector of Defibrillator Unit. Check charging control lamp of Defibrillator Unit.
3.5
Connect power supply pack to corpuls magnet connector of Patient Box. Check charging control lamp of Patient Box.
3.6
Reassemble Patient Box with Display Unit (semi-modular mode). Check charging control lamp of Display Unit.
3.7
Connect power supply pack to corpuls magnet connector of the Display Unit. Check charging control lamp of Display Unit.
3.8
Reassemble all single components to a compact device; connect power supply pack to corpuls magnet connector of Defibrillator Unit. Check charging control lamp of Display Unit.
4
Device function Display Unit
For following tests operate the device in modular mode.
4.1
Check acoustics and keyboard of the Display Unit. Push every button of the keyboard. The speaker must answer with a tone.
4.2
Check printer functionality. Select menu item “Test page“.
4.3
Check alarm and printer stop in case of lack of paper. Remove paper reel and start printer operation.
4.4
Check visual alarm indication. In case of an alarm, the alarm lamp in the jogdial must light up.
4.5
Check GSM function. Install SIM and establish a connection to a fax recipient. Transmit a diagnosis ECG to the fax recipient.
4.6
Check the Health Insurance Card Reader. Insert an applicable card into the card reader slot; patient’s data must appear on the screen.
4.7
Check the LAN function. Compare the MAC address of the device with the papers of the device. Connect the device to a DHCP server. Dial the assigned IP address of the device using the ping networking utility.
Loading...
+ 11 hidden pages