6: Appendix A Specifications......................................................................................................44
Page 3www.greyline.com
1:General Description
1.1Introduction
This manualdescribes the operation of the Greyline Portaflow PT400 portable flowmeter. The flowmeter is
designed to work with clamp-on transducers to enable the flow of a liquid within a closed pipe to be measured
accurately without needing to insertanymechanicalparts through the pipe wall or protrude into the flow system.
Using ultrasonictransittime techniques, the Portaflow PT400 is controlled by a micro-processor system which
contains a wide range of data that enables it to beused with pipes with an outside diameter ranging from 13mm
up to 1000mm and constructed of almost any material. The instrument will also operate over a wide range of fluid
temperatures.
The PT400 series comprises two models which are identical in operation but designed to be used with a different
range of pipe diameters. The PT400 can be used with pipesin the range 13mm – 115mm and the PT500 with
pipes intherange 50mm – 1000mm.
Easy to operate, the Portaflow PT400 standard features are:
•Large, easy to read graphic display with backlighting
•Simple tofollowdual function keypad
•Simple ‘Quick Start’ set up procedure
•Continuous signal monitoring
•Pulse output
•4-20mA, 0-20mAor 0-16mA output
•Rechargeable battery
•Battery management
•Diagnostics
Volumetric flowrates are displayed in l/h, l/min, l/sec, gal/min, gal/h, USgals/min, USgals/h, Barrel/h, Barrel/day,
m³/s, m³/min, m³/h. Linear velocity is displayed in metres or feet persecond.
When operatinginthe‘FlowReading’ mode the total volumes, both positive and negative,are displayed up to a
maximum 12-digit number.
The flowmeter can be used to measure cleanliquids or oils that haveless than 3% byvolume of particulate
content. Cloudy liquids such as river water and effluent can be measured along with cleaner liquids such as
demineralised water.
Typical PortaflowPT400 applications include:
•River water
•Seawater
•Potable water
•Demineralised water
•Treated water
Page 4www.greyline.com
1.2Principles of Operation
When ultrasound is transmitted through a liquid the speed at which the sound travels through the liquid is
accelerated slightlyif it is transmitted in the same direction as the liquid flow and decelerated slightly if transmitted
against it. The difference intime taken by the sound to travel the same distance but in opposite directions is
therefore directly proportional to the flow velocity of the liquid.
The PortaflowPT400 system employs two ultrasonic transducers attached to the pipe carrying the liquid and
compares thetime taken to transmit an ultrasound signal in each direction. If the sound characteristics of the fluid
are known, the Portaflow microprocessorcanusethe results of the transit time calculations to compute the fluid
flow velocity. Once the flow velocity is known the volumetric flow can be easily calculated for a given pipe
diameter.
The Portaflowsystemcan be set up to operate in one of fourmodes determined mainly by the pipe diameter and
the typeof transducer set inuse. The diagram below illustrates the importance of applying the correct separation
distance between the transducers to obtain the strongest signal.
Fluid flow
Fluid flow
Fluid flow
Fluid flow
Fluid flow
Upstream
transducer
Separation
Distance
U
Separation
Distance
U
Separation
Distance
U
Separation
Distance
D
D
D
U
Reflex mode
This is the mode most commonly used.
The two transducers (U & D) are attached
to the pipe in line with each other and the
signals passing between themare reflected
by the opposite pipe wall.
The separation distance is calculated by
the instrument in response to entered data
concerning the pipe and fluid characteristics.
Reflex mode (double bounce)
In this mode the separation distance is
calculated to give a double bounce. This is
most likely to occur if the pipe diameter is
so small that the calculated reflex mode
separation distance would be impractical
for the transducers in use.
Reflex mode (triple bounce)
This illustration goes one step further to show
a triple bounce situation. This would normally
apply when working with very small pipes
relative to the transducer range in use.
Diagonal mode
This mode might be selected by the
instrument where relatively large pipes are
concerned. In this mode the transducers are
located on opposite sides of the pipe but the
separation distance is still critical in order
for the signals to be received correctly.
This mode might be used with the standard
‘A’ & ‘B’ transducer sets but for really large
pipe installation the optional transducer set ‘D’
might be recommended.
D
Figure 1.1 Operating modes
Page 5www.greyline.com
1.3Supplied Hardware
The Portaflowequipmentis supplied in a rugged plastic carrying case fitted with a foam insert to give added
protection fortransportation. The supplied components are shown in Figure 1.2.
TransducerCables
Chains (x2)
Ultrasonic
Couplant
GuideRails (x2)
Transducers
(Sensorsx2)
RuledSeparation bar
Test Block
4-20mA/
PulseOutput
Cable
Power Supply
(Batt. Charger)
Portaflow PT400
Instrument
Figure 1.2 Standard Portaflow equipment
Standard equipment
•Portaflow PT400 instrument with backlit graphic display
•Power supply - with UK, US, European adaptors. 110/240VAC
•4-20mA/Pulse Output cable
•2 lengths of chain each at 3.3 metres long
•Test block
•Transducer cables (x2) 2 metres long (one red and one blue)
•Transducer set 'A' (Transducers x2)
or
•Transducer set 'B' (Transducers x2)
•Set of guide rails for use with ‘A’ or ‘B’ transducers
•Ruled separation bar (2-piece)
•Manual
Page 6www.greyline.com
1.4Portaflow PT400 Instrument
The PortaflowPT400 is a microprocessorcontrolled instrument operated through a menu system using an inbuilt
LCD display and keypad. It can be used to display the instantaneous fluid flow rate or velocity, together with
totalised values.
The instrument can also provide a current or variable ‘pulse’ output proportional to the detected flow rate. These
outputs, which can be used with a range of external interface devices such as those found in BMS or site
monitoring systems, can be calibrated to suit a particular flow range.
1.4.1Connectors
4-20mA / Pulse Output
Transducer Cables
Reset pin-hole
LCD Display
Keypad
Battery Charger
Figure 1.3 Instrument details
Transducer connections
The transducers are connected to two colour-coded miniature coaxial sockets located on the top of the
instrument. Using the red/blue connector cables provided, the upstream transducer shouldalways be connected
to the RED socket and the downstream transducer to the BLUE one for a positive flow reading. It is safe to
connect or disconnect the cable while the instrument is switched on.
4-20mA and Pulse output connection
The 4-20mA/ ‘pulse’ output cable should be connected to the green7-pin connector on the top of the flowmeter,
as shown in Figure 1.3. A single cable that can be adapted for use for either of these output functions is included
in the Portaflow PT400 kit.The ‘tails’ on thefreeendof the cable must be terminated to suit the intended
application.
Red – 4-20mA positive
Black – 4-20mA negative
White – Pulse output
Green – Pulse return
Brown – Set Point (notinpresent use)
Blue – Set Point return (not in present use)
Thick Black – Cable screen
Battery charger connection
The suppliedbatterychargeris connected to the instrument by means of the grey 2-pin connector on the bottom
of theunit, as shown in Figure 1.3.
Page 7www.greyline.com
Note: The above connectors have different key-ways to prevent incorrect cable connection.
Numerical keypad
withdual function keys
ON/OFF
1.4.2Keypad
The instrument is configured and controlled via a 16-key tactile membrane keypad, as shown in Figure 1.4.
ON/OFF Key
The ON/OFF key is shown on the top left of the keypad. When turned ON an initialisation screen is displayed on
the LCDshowing theinstrument’sserialnumber and software revision. Once this appears, the instrument can be
started bypressingthe ENTER key once – the initialization screen is then replaced by a MAIN MENU which
provides accessto the remaining functions.
Menus and the menu selection keys
The PortaflowPT400 menus are arrangedhierarchally with the MAIN MENU being at the top level. Menu
navigation is achieved by three keys on the right hand side of the keypadwhich are used to scroll UP and DOWN
a menu list and SELECT a menu item. When scrolling through a menu an arrow-shaped cursor moves up and
down the left hand side of the screen to indicate the active menu choice which can then be selected by pressing
the ENTER (SELECT) key.
Scroll UP
ENTER
(SELECT)
Scroll DOWN
Scroll LEFTScroll RIGHT
Figure 1.4 Keypad
Some menushave more options than can be shownon the screen at the same time, in which case the
‘overflowed’ choices can be brought into view by continuing to scroll DOWN past the bottom visible item. Menus
generally ‘looparound’if you scroll beyond the first or last items.
If you select Exit it usually results in taking you back one level in the menu hierarchy, but in somecasesitmay go
directly to the ‘Flow Reading’ screen.
Some screens require you to move the cursor leftandrightalongthedisplayas well as up and down. This is
achieved using keys 5 (scroll LEFT) and 6 (scroll RIGHT).
Dual function numerical keypad
The block of keys shown in the centre of the keypadin Figure 1.4 are dual function keys. They can be used to
enter straight-forward numerical data, select the displayed flow units or provide quick access to frequently
required control menus.
Note: Some of the features accessed by these keys are restricted in the Portaflow 220 model range. An “Option
not available”message is displayed if you select a restricted function.
1.4.3Power supply and battery charging
Operating power is provided by an internal battery thatcanbe charged from theutility supply using the supplied
external charger.Whenyou first receive the unit you must put the battery on charge for a minimum of 6.5hrs
Page 8www.greyline.com
before use. A fully chargedbattery will power the instrument for up to 20hrs depending on the output utilisation
and backlight usage.
If the backlight has been enabled the display is illuminated for 10 seconds every time a key is pressed. If the
backlight is active continuously it would reduce the available battery operating time to 8hrs. Similarly, if the 420mA output is used constantly at 20mA, the battery life would reduce by 50%. It is therefore beneficial to turn off
the backlight and 4-20mA output facilities when they are not required.
When the instrument is operating in the ‘Flow Reading’ mode the percentage battery charge level is displayed
symbolically ontheLCDscreen. A warning message is triggered if the charge falls to approximately 30%, at
which point there is up to four hours of battery operation remaining, depending on usage. The battery can be
charged both while the instrument is in use or when switched off. The instrument’s internal data is stored in nonvolatile memory and will not be lost even if the battery discharges completely.
Key Point: The battery is not a user-changeable item. The instrument must be
returned to Greyline if the battery needs replacing.
Key Point: Only use the supplied charger, or special adaptor lead. Failure to comply
with this will invalidate your warranty.
1.5Transducers
Different transducersets are provided with the PF400A and PF400B Portaflow models and are not
interchangeable.
Transducer set 'A'
Supplied as standard on PF400A for use on pipes 0.51” to 4.525” (13mm to 115mm) outside diameter.
Transducer set 'B'
Supplied as standard on PF400B for use on pipes 1.97” to 39.37” (50mm to 1000mm) outside diameter.
Key Point: Always use the transducers that were supplied with the instrument.
Page 9www.greyline.com
2:Installation
Guide
20 x Diameter
10 x Diameter
Valid transducer location
The Portaflowequipment expectsa uniformflowprofile as a
2.1Transducer Positioning
In many applicationsan even flow velocity profile over
a full 360°is unattainable due, for example, to the
presenceof air turbulenceat the top of the flow and
possiblysludge in the bottom of the pipe. Experience
has shown that the most consistently accurateresults
are achievedwhen the transducer guide rails are
mounted at 45° with respect to the top of the pipe.
Air
rail
45°
Possible
sludge
Flow
distorted flow will produceunpredictablemeasurement
errors. Flowprofile distortions can result from upstream
disturbances such as bends, tees, valves, pumps and other
similar obstructions. To ensure a uniform profile the
transducers must be mounted far enough away from any
causeof distortionsuch thatitno longer has an effect.
Uniform Flow ProfileDistorted Flow Profile
Flow
Figure 2.1 Locating the transducers
To obtainthe most accurate results the condition of both the liquid and the pipe wall must be suitable to allow the
ultrasound transmission along its predetermined path. It is importantalso that the liquid flows uniformly within the
length ofpipe beingmonitoredand that the flow profile is not distorted by any upstream or downstream
obstructions. This is best achieved by ensuring there is a straight lengthof pipe upstream of the transducers of at
least 20timesthepipe diameter and 10 times the pipe diameter on the downstream side, as shown in Figure 2.1.
Flow measurements can be made on shorter lengths of straight pipe, down to 10 diameters upstream and 5
diameters downstream,butwhen the transducers are positioned this close to any obstruction the resulting errors
can be unpredictable.
Key Point: Do not expect to obtain accurate results if the transducers are positioned
close to any obstructions that distort the uniformity of the flow profile.
Page 10www.greyline.com
2.2Transducer Attachment
The transducers are fitted to adjustable guide rails which are secured to the pipe using wrap-around chains and
mechanically connectedtogether by a steel separation bar. The separation bar also acts as a ruler to allow the
distance between the transducers to be accurately set to the value determined by the Portaflow instrument.
When fitting the guide rails it is easiest to assemble them onto the separation bar and adjust to the required
separation distancebefore attaching them to the pipe.
2.2.1Preparation
1.Before you attach the transducers you should first ensure that the proposed location satisfies the
distance requirementsshownin Figure 2.1 otherwise the resulting accuracyoftheflowreadings may be
affected.
2.Prepare the pipe by degreasing it and removing any loose material or flaking paint in order to
obtain the best possible surface. A smooth contact between pipe surface and the face of the transducers is
an important factor in achieving a good ultrasound signal strength and therefore maximum accuracy.
2.2.2Attaching the guide rails
A
B
C
D
E
F
G
A: Tensioningthumb-wheel.
B: Tension bar.
C: Separationbar securing
screw.
Figure 2.2 Guide rail attachment
1.Slide the separation bar (D) into the front of the left hand guide rail, align the front edge of the
guide railwith ‘0’ on the ruler scale (E) and secure it in place by tightening the thumbscrew (C).
2.Slide the other end of the separation barintothefrontof the right hand guide rail, align the front
edge of the guide rail to the required separation distance (obtained from the Portaflow instrument) on the
ruler (F), then secure itin place by tightening the thumbscrew.
Page 11www.greyline.com
3On each guide rail, attach one end of a securing chain to a hook on the tensioning bar (B), wrap the chain
around thepipe (G) and then attach it to the hook on the other end of the tensioning bar whilst keeping the
chain as tight as possible.
4.Rotate the complete guide railassembly so that it is approximately 45° with respect to the top of
the pipe. Then tighten the chain by turning the tensioning thumb-wheel (A) on each guide block until the
assembly is securely attached to the pipe.
Note: If you are unable to get sufficient tension on the chain to hold the assembly in place, fully slacken the
tensioning thumb-wheel and shorten the effective length of the chain wrapped around the pipe by connecting
the tensioning bar to the next link in the chain, then re-tension.
2.2.3Mounting the transducers
C: Transducer coverplatesecuringscrew.
B:Transducer locatingslot/lug.
D: Transducer cableconnection.
A:Transducer cover plate.
G:Chain.
E:Ultrasonic
couplant
application.
Figure 2.3 Mounting the transducers
1.Slide the transducer cover plate (A) fully towards the outside of the guide assembly to allow
sufficient access to fit the transducer.
2.Clean the face of the transducer, removing alltracesof dirt and grease.
3.Apply a 1/8” (3mm) bead of ultrasonic couplant along the centre length of the transducer (E).
4.Fit the transducer into theguide block – ensuring the lugs on the sides of the transducer are
correctly located into the slots on the sides of the guide block (B).
5.Slide the transducer cover plate (A) over the top of the transducer and tighten the thumbscrew
(C) fingertight to secure the transducer in place. When securing the cover plate take care to leave sufficient
room around the transducer connector (D) to connect the cable.
6.Repeat the above steps forthe second transducer.
7.Connect the transducers to the Portaflow instrument usingthe coaxial cables provided. The RED
cable mustbe connected to theupstream transducer and the BLUE cable to the downstream transducer. If
you observe negative flow, swap the red andblue cables at the sensors.
Page 12www.greyline.com
3: Operating Procedures
Initial instrument setup
(Paragraph 3.1)
Battery charging, Set date/time, Language, Backlight
Connect and take basic flow readings
At a one-off location
(Paragraph 3.2)
QUICK STARTVIEW EDIT SITE DATA
Enter data
Attach sensors
FLOW READING
At a frequent location
(Paragraph 3.3)
Choose site / check data
Attach sensors
FLOW READING
Carry out any necessary calibration
(Paragraph 3.5)
How to adjust the Zero Flow Offset – Paragraph 3.5.2
How to adjust the Calibration Factor – Paragraph 3.5.3
How to adjust the Roughness Factor – Paragraph 3.5.4
How to adjust the Damping Factor – Paragraph 3.5.5
Set-up a monitoring or logging application
(Paragraph 3.6)
Manage site details
(Paragraph 3.4)
Set up a named site
Rename a site
Configure the interfaces
(Paragraph 3.7)
How to measure totalised flows – Paragraph 3.6.1
How to set up basic (manual) logging – Paragraph 3.6.2
How to set up auto-timed logging – Paragraph 3.6.3
How to set up logging ‘live’ to a PC– Paragraph 3.6.4
How to download saved data to a PC – Paragraph 3.6.5
How to print data – Paragraph 3.8
Before you use your Portaflow PT400 for the first time you should first charge the battery, then select the display
language andset-up the internal clock, as described below.
Charging the battery
1.Connect the external battery chargerto the charger socket at the bottom of the instrument then
switch ontheutilitysupply.
2.The instrument should indicate CHARGING
and an animated battery symbol indicates that the
battery is taking on charge.
3.Leave theinstrumentoncharge for 6.5 hours
before using it for the first time.
Selecting a language
The first time you switch on the instrument you may be asked to select a user language.
1.Switch on the instrument bypressing the ON/OFF button.
2.If necessary, select the required language using the UP/DOWN scroll keys then press the
ENTER key.
3.The selectedlanguage will be the default
when the instrument is next used. To change the
language againselectthe Change Language
option inthe SETUP INSTRUMENT screen (see
below)
4.The initialisationscreen will be displayed,
giving details of the instrument’s serial number
and software revision details.
CHARGINGDD-MM-YY HH:MM:SS
Serial # V 00.00.00.00.00.00.00. 0000
27Press to start
5.Press ENTER to start the instrument.
6.This is the MAIN MENU and is the starting
point for all the operations described in this
chapter.
MAIN MENUDD-MM-YY HH:MM:SS
Quick start
View/Edit Site Data
Data Logger
Setup RS232 /USB
Setup Instrument
Read flow
Page 14www.greyline.com
Setting the Date & Time
1.Select Setup Instrumentfrom the MAIN MENU.
The screen shown here should be displayed.
2.Select Set Date & Time and click ENTER.
3.A flashingcursorshould appear underthe first
date number. Enter the date sequence in dd-mmyy order then press ENTER.
4.Repeat this action to setthe time.
5.Select Exit then press ENTER to return to the
MAIN MENU.
Note: If you make a mistake when entering the data press the Delete key to move the cursor back to the number
you wish to change, then continue. If you enter an invalid number an ‘ERR:Invalid Date or Time!’ error message is
displayed on the second lineof the screen. If this occurs repeat the set date/time procedure.
3.1.2Enabling/disabling the backlight
When the backlight is enabledit will illuminate for 15 seconds every time a key is pressed so if it is not required it
is recommended that you disable it to prolong the battery discharge time.
2.Select Backlight from theSETUP
INSTRUMENT screenthenpress ENTER.
3.Select Enableor Disable as required.
4.Press ENTER to return to the SETUP
INSTRUMENT screen.
5.Select Exit then press ENTER to return to the
MAIN MENU.
3.2Using the Quick Start Menu
SETUP INSTRUMENTDD-MM-YY HH:MM:SS
Dim: mm
Set Date & Time:dd-mm-yy hh:mm:ss
Calibrate 4-20mA
Pulse output
Backlight:Disabled
Factory settings
Change Language
Exit
BACKLIGHTDD-MM-YY HH:MM:SS
Dim: mm
Enabled
Disabled
If you want to perform a ‘one-off’ flow reading at a particular pipe location the Quick Start menu provides the
quickest way to set up the Portaflow system and access the FLOW READING screen.
If the point at which you intend to take the measurement is likely to require regular monitoring it is best to set it up
as a ‘Site’ within the Portaflow PT400, which then stores the site parameters (See Managing Named Sites).
Page 15www.greyline.com
Before you can use the Portaflow system you need to obtain the following details (this information will be required
when setting up the Quick Start menu):
•The pipe outside diameter
•The pipe wall thickness andmaterial
•The pipe lining thickness andmaterial
•The type of fluid
•The fluidtemperature
Entering the site data
1.Select Quick Start from the MAIN MENU and press ENTER. You will then be presented with a
series of screens in which to enter the data mentioned above.
2.Select the dimension units (millimetres or
inches) used to measure the pipe, then press
ENTER.
3.Enter the pipe outside diameter dimension,
then press ENTER.
4.Enter the pipe wall thickness dimension, then
press ENTER.
DIMENSION UNITDD-MM-YY HH:MM:SS
Select thedimension units:
mm
Inches
OUTSIDE DIAMETERDD-MM-YY HH:MM:SS
Dimension:mm
Pipe outside diameter?58.0
PIPE WALL THICKNESSDD-MM-YY HH:MM:SS
Dimension:mm
Pipe outside diameter?58.0
Pipe wall thickness?4.0
5.If the pipe has a lining, enter the lining
thickness.
If nothingis entered the instrument automatically
assumes there is no lining.
6.Press ENTER to continue.
7.Select the pipe wall material from the list
provided, thenpressENTER.
If the material is notlisted select Other and enter
the propagation rate of the pipe wall material in
metres/sec. Contact Greyline if this is not known.
PIPE LINING THICKNESSDD-MM-YY HH:MM:SS
Dimension:mm
Pipe outside diameter?58.0
Pipe wall thickness?4.0
Pipe lining thickness?0.0
PIPE WALL MATERIALDD-MM-YY HH:MM:SS
Select pipe wall material
Mild Steel
S' less Steel 316
S' less Steel 303
Plastic
Page 16www.greyline.com
Cast Iron
Ductile Iron
Copper
Brass
Concrete
Glass
Other (m/s)
8.If a lining thickness value was entered earlier,
this screenis displayed to request that you enter
the lining material type. If no lining thickness was
entered this screen will be bypassed.
9.Select theliningmaterial from the list provided
then press ENTER.
If the material is notlisted select Other and enter
the propagation rate of the lining material in
metres/sec. Contact Greyline if this is not known.
10.Select the fluid type from the list provided and
press ENTER.
If the liquid is not listed select Other and enter a
propagation rate in metres/second.
PIPE LINING MATERIALDD-MM-YY HH:MM:SS
Select pipe lining material
Steel
Rubber
Glass
Epoxy
Concrete
Other (m/s)
FLUID TYPEDD-MM-YY HH:MM:SS
Select fluid type
Water
Glycol/water 50/50
Glycol/water 30/70
Lubricating oil
Diesel
Freon
Other (m/s)
11.If you need to alterthe fluid temperature from
that shown selecteither °C or °F with the cursor
and press the ENTER key.
12.Enter thenewtemperature value and press the
ENTER key.
13.The new temperature should now be indicated
in both °C and °F.
14.Select Continue.. and press ENTER.
15.The SENSOR SEPARATION screen now
displays a summary of the entered parameters and
informs you of the typeof sensor to be used, the
mode of operation and the distance to set up
between thesensors.
In this example it recommends type A-ST (A
standard) sensors operatingin the ‘Reflex’ mode
spaced at 32.2mm apart.
Take a note of these details
Pipe: 58.00 mm
Wall: 4.00
Sensors: A-STReflex
Temperature: 10.00°C 50.00°F
Set sensorseparationto32.2 mm
Press to continue, to select sens.
Page 17www.greyline.com
Note: Do not press ENTER until the transducers are fitted and connected to the instrument.
0.000
Attaching and connecting the transducers
16.Fit the designated sensors to the pipe using the appropriate guide rails as described in
Paragraph 2.2. Take great care to set the separation distance as accurately as possible.
17.Connect the redandblue coaxial cables between the sensors and the test instrument, ensuring
that the red connector on the instrument is connected to the ‘upstream’ sensor.
Taking a flow reading
18.Once thetransducers have been fitted and
connected press the Enter key.
19.This will take you from the SENSOR
SEPARATION screen to the FLOW READING
screen via a signal-checking screen (shown here).
20.Check thattheindicatedsignal strength on the
left of the screen is at least 2 bars (ideally 3 or 4). If
less than 2 bars are shown it indicates there could
be a problem with the transducer spacing,
alignment or connections; or itcould be due to an
application problem.
Flow monitoring
The FLOW READING screen is the one most used during normal monitoring operation. It shows the
instantaneous fluid flow together with totalised values (when enabled). In this mode you can select the flowrate
measurement unitsby pressing keys 7 (litres), 8 (Gallons) or 9 (m³), or change the display to show velocity by
pressing key 4.
FLOW READINGDD-MM-YY HH:MM:SS
Signal
+Total: 0.00litres
–Total: 0.00litres
3.3Using the System at a Regularly Monitored Location
Setting up the Portaflow system using the Quick Start method described in Paragraph 3.2 is easy andthe
recommended methodtousein a ‘one-off’ situation. But if you have a site location that you want to monitor on a
frequent basis it is better to set up a named ‘Site’ for that location so that you can recall it when needed and so
avoid theneed to re-enter the site details every time you want to install the equipment there.
Note: See Paragraph 3.4 for details of how to set-up and manage site details.
Dim: mm
l/min
Use this procedure to installthe equipment at a named site.
5.Scrolldown through the menu list and
enter/change the data that might have changed
since the last time the site was accessed.
6.When you are satisfied that the parameters are
correct select Save current site & read flow.
Pipe wall material: Mild Steel
Lining material: --------Sensor set: A-ST
Sensor mode: Reflex
Fluid type: Water
Save current site & read flow
Delete this Site
Download & save current site
Exit
7.If you need to alterthe fluid temperature from
that shown selecteither °C or °F with the cursor
and press the ENTER key.
8.Enter thenewtemperature value and press the
ENTER key.
9.The new temperature should now be indicated
in both °C and °F.
10.Select Continue.. and press ENTER.
11.The SENSOR SEPARATION screen now
displays a summary of the entered parameters and
informs you of the typeof sensor to be used, the
mode of operation and the distance to set up
between thesensors.
In this example it recommends type A-ST (A
standard) sensors operatingin the ‘Reflex’ mode
spaced 32.2mmapart.
Take a note of these details.
Note: If you press ENTER before the transducers are fitted and connected to the instrument, the instrument
will detect a low signalstrength and may indicate an ERROR condition.
FLUID TEMPERATUREDD-MM-YY HH:MM:SS
Enter Fluid Temperature
°C: 5.00
°F: 41.00
Continue..
SENSOR SEPARATIONDD-MM-YY HH:MM:SS
Site: MySite
Pipe: 58.00 mm
Wall: 4.00
Sensors: A-STReflex
Temperature: 10.00°C 50.00°F
Set sensorseparationto32.2 mm
Press to continue, to select sens.
Attaching and connecting the transducers
12.Fit the designated sensors to the pipe using the appropriate guide rails as described in
Paragraph 2.2. Take great care to set the separation distance as accurately as possible.
13.Connect the redandblue coaxial cables between the sensors and the instrument, ensuring that
the redconnectoron the instrument is connected to the ‘upstream’ sensor.
Taking a flow reading
14.Once thetransducers have been fitted and
connected press the ENTER key.
15.This will take you from the SENSOR
SEPARATION screen to the FLOW READING
screen via a signal-checking screen (shown here).
left of the screen is at least 2 bars (ideally 3 or 4). If
less than 2 bars are shown it indicates there could
be a problem with the transducer spacing,
alignment or connections; or itcould be due to an
application problem.
Flow monitoring
The FLOW READING screen is the one most used during normal monitoring operation. It shows the
instantaneous fluid flow together with totalised values (when enabled). In this mode you can select the flowrate
measurement unitsby pressing keys 7 (litres), 8 (Gallons) or 9 (m³), or change the display to show velocity by
pressing key 4.
3.4Managing Named Sites
If you want to monitor a particular site location frequently you can set up a named ‘Site’ tostore the installation
details, suchas pipe dimensions and material, required toset-up the Portaflow PT400 system. These can then be
recalled laterwhen revisiting that particularlocation.
The instrument can store up to 20 sites, the first site is reserved for QUICK START and cannot be renamed;
subsequent sitesare initially named EmptySite1 through to EmptySite19.
FLOW READINGDD-MM-YY HH:MM:S
SignalDim: mm
l/min
+Total: 0.00litres
–Total: 0.00litres
3.4.1Setting up a new site
1.Select View Edit Site Data from the MAIN
MENU.
2.Select Choose from list of sites.
3.Select one of the EmptySites from the
presented list (e.g. EmptySite 1 as shown).
4.Select Site name and press ENTER.
5.This opens the ALTER NAME screen.
6.Select Alter theSite Name and you will be
presented with a screen which allows you to enter
a newnameinmuch the same way as when
composing a mobile text message.
7.On completion press ENTER then select Exit.
This will take you back to the VIEW EDIT SITE
DATA screen.
ALTER NAMEDD-MM-YY HH:MM:SS
Site name:EmptySite1
Do you wish to:–
Alter the Site Name
Exit
Page 20www.greyline.com
8.Scroll down through the menu list and
enter/change the pipe parameters and other data
pertaining to the site.
Note thatthismenu allows you to choose a
Sensor Set, unlike the QUICK START menu
which recommended the sensor set to use. If you
enter an inappropriate sensor set in this menu you
will be presented with an error message later
when you go to the SENSOR SEPARATION
screen.
VIEW EDIT SITE DATADD-MM-YY HH:MM:SS
Dim: mm
Choose from list of sites
Site name:MyNewSite
Dimension units:mm
Pipe outside diameter:58.00
Pipe wall thickness:4.00
Pipe lining thickness:0.00
9.When all the data is correct you can either:
a)Select Save current site & read
flow to continue fitting the transducers and
opening the FLOW READING screen.
b)Select Delete this sitetodelete
the site name and values and restore itto
the original EmptySite name.
c)Select Download & save current
site to save the site details and download
them to a PC via RS232/USB provided the
RS232/USB linkis connected and correctly
set-up.
d)Select Exit to returnto the MAIN
MENU.
3.4.2Changing a site name
To changea site name use the same method described above for generating a newsite: but in this case access
a currentsitename to change rather than an EmptySite. If you change a site name while the site is logging the
logging will stop.
Save current site & read flow
Delete this Site
Download & save current site
Exit
Page 21www.greyline.com
3.5Instrument Calibration
The Portaflowis fully calibrated before it leaves the factory; however the following adjustments are provided to
allow you to further ‘fine tune’ your instrument to suit local conditions and application where necessary. Apart from
the zero flow offset adjustment, these are normally carried out only where the instrument is to be used in a
permanent or semi-permanent location.
3.5.1Adjusting the zero cut-off
This adjustmentallows you to set a minimum flow rate (m/s) below which the instrument will indicate ‘0’. The
default setting is 0.02 m/s but you may adjust this value if required.
1.With the instrument operating in FLOW READING mode, press the Options key to access the
FLOW READING OPTIONS menu shown.
2.Select ZeroCutoff(m/s) and press ENTER.
3.Enter thevaluefor the Zero Cutoff (e.g. 0.06
m/s) thenpress ENTER.
4.Scroll down to select Exit and press ENTER to
return to the FLOW READING screen.
3.5.2Adjusting the set zero flow offset
The Portaflowinstrumentoperatesbycomparing the time taken to send an ultrasonic signal between two
transducers in either direction. A Set zero flow offset adjustment is provided to compensatefor any inherent
differences between the two sensors, noise pick-up, internal pipe conditions etc. It can be used to ‘zero’ the flow
indication underno-flowconditions.
If you have adjusted the Zero Cutoff point to anywhere above ‘0’ you must reset it to
‘0’ before you can observe and adjust the Set zero flow offset, as its value is very
small. Once the Set zero flow offset has been calibrated you can then reapply the Zero
Cutoff if required.
1.Stop theliquid flow.
2.With the instrument in FLOW READING mode press theVelocity function key and observe the
reading (m/s). Any reading other than 0.000 indicates an offset errorandin practice this will typically be in the
range ±0.005m/s (possibly higher onsmaller diameter pipes). If a greater figure is shown it is worth
calibrating the offset to obtain a more accurate result. Continue as follows:
FLOW READING OPTIONDD-MM-YY HH:MM:SS
Data review
Zero Cutoff (m/s):0.00
Set zero flow (m/s):0.00
Damping (secs):10
Totaliser:Run
Reset +Total
3.Press the Options key to access the FLOW
READING OPTION screen shown.
4.Select Set zero flow (m/s) and press ENTER.
5.Press ENTER on the subsequent screen to
accept the change, which will return you to the
screen shown.
6.Scroll down to select Exit and press ENTER to
return to the FLOW READING screen.
FLOW READING OPTIONDD-MM-YY HH:MM:SS
Data review
Zero Cutoff (m/s):0.00
Set zero flow (m/s):0.00
Damping (secs):10
Totaliser:Run
Reset +Total
Page 22www.greyline.com
Key Point: In order to cancel any applied offset you must either read flow via Quick
Start or switch the Portaflow instrument OFF & ON. Any value that you trim-out
using the offset adjustment will be added/subtracted from the flow reading across
the whole range.
3.5.3Adjusting the calibration factor
Key Point: USE THIS FACILITY WITH CARE & ONLY WHERE NECESSARY
The Portaflow instrument is fully calibrated before leaving the factory and under
normal circumstances does not require further calibration when used on site.
This facility can be used to correct the flow indication where unavoidable errors occur due to the lack of a straight pipe or where the sensors are forced to be fitted
close to the pipe-end, valve, junction etc.
Any adjustment must be made using a reference flowmeter fitted in the system.
With the system running:
1.Stop (Stall) the Portaflow’s totaliser facility and zero it (Paragraph 3.6.1).
2.Run the Portaflow’s totaliser to measure the total flow over a 30-60 minute period, and note the
total flowindicated by the reference flow meter over the same period.
3.Calculate the % error between the Portaflow and reference meters. If the error is greater than
±1% calibrate the Portaflow as detailed below.
4.Press the Options key to access the FLOW
READING OPTION screen shown.
5.Scroll down and select Calibration factor.
6.Change the calibration factor according to the
error calculatedin step 3. For example, if the
Portaflow was reading 1% high then increase the
Calibration factor value by 0.010. Conversely, if the
reading is 1% low thendecrease the calibration
factor to 0.990.
7.Press Enter to apply the change.
8.Select Roughnessfactor or Exit as required.
3.5.4Adjusting the roughness factor
FLOW READING OPTIONDD-MM-YY HH:MM:SS
Data review
Zero Cutoff (m/s):0.00
Set zero flow (m/s):0.00
Damping (secs):10
Totaliser:Run
Reset +Total
Reset –Total
Calibration factor:1.000
Roughness factor:0.01
Diagnostics
Exit
The roughness factorcompensates for the condition of the internal pipe wall, as a rough surface will cause
turbulence and affects the flow profile of the liquid. In most situations it is not possible to inspect the pipe
internally and the true conditionisnotknown. In these circumstances experience has shown that the following
values canbe used:
Page 23www.greyline.com
Pipe MaterialRoughness Factor
Non ferrousmetal
Glass
Plastics
Light metal
Drawn steel pipes:
•Fine planed,polishedsurface
•Plane surface
•Rough planed surface
Welded steelpipes,new:
•Long usage,cleaned
•Lightly and evenly rusted
•Heavily encrusted
Cast iron pipes:
•Bitumen lining
•New, without lining
•Rusted / Encrusted
0.01
0.01
0.1
1.0
Page 24www.greyline.com
With the system running in FLOW READING mode:
1.Press the Options key to access the FLOW
READING OPTION screen shown.
2.Scroll down and select Roughness factor.
3.Change the roughness factor accordingto the
pipe materialand condition as described above.
4.Press Enter to apply the change.
3.5.5Adjusting the damping factor
By averaging-out the flow rate over several seconds, the Damping factor can be used to smooth out rapid
changes in flow rate to prevent wild fluctuations in the displayed flow value. It has a range of 1 to 199, with a
default setting of 10. With the system running in FLOW READNG mode:
1.Press the Options key to access the FLOW
READING OPTION screen shown.
2.Scroll down and select Damping (secs).
3.Change the value of the Damping factor as
required to remove any unwanted display
fluctuations. Increasing the value applies a greater
smoothing affect.
4.Press Enter to apply the change.
FLOW READING OPTIONDD-MM-YY HH:MM:SS
Data review
Zero Cutoff (m/s):0.00
Set zero flow (m/s):0.00
Damping (secs):10
Totaliser:Run
Reset +Total
Reset –Total
Calibration factor:1.000
Roughness factor:0.01
Diagnostics
Exit
FLOW READING OPTIONDD-MM-YY HH:MM:SS
Data review
Zero Cutoff (m/s):0.00
Set zero flow (m/s):0.00
Key Point: If the damping factor is set too high the value displayed may appear
stable but it may exhibit large step changes when the value is updated.
Page 25www.greyline.com
3.6Monitoring Functions
12.34
3.6.1How to measure totalised flows (manually)
The basic measurement indicated on the FLOW READING screen is the instantaneous flow rate, which in some
applications may vary over a period of time.Average flow rates are therefore often required in order to get a
better understanding of an application’s true performance. This is simply achieved by noting the total flow over a
specific period (for example 30-60 minutes) and then calculating the average flow rate over that period of time.
1.Press the Options key to access the FLOW
READING OPTION screen shown.
2.If the Totaliser is indicating Run, select it and
change it to Stall. Press ENTER.
3.Select Reset +Total and press ENTER.
4.Press ENTER on the subsequent screen to
accept the reset.
5.Press ENTER again to return to the FLOW
READING OPTIONS menu.
6.Select Reset –TotalandpressENTER.
7.Press ENTER on the subsequent screen to
accept the reset.
8.Press ENTER again to return to the FLOW
READING OPTIONS menu.
9.Note and record the current time.
10.Select Totaliserand change it to Run. Press
ENTER.
Note: thetotalisers begin to count up as soon as
Totaliser is put to Run.
FLOW READING OPTIONDD-MM-YY HH:MM:SS
Data review
Zero Cutoff (m/s):0.00
Set zero flow (m/s):0.00
Damping (secs):10
Totaliser:Stall
Reset +Total
FLOW READING OPTIONDD-MM-YY HH:MM:SS
Zero Cutoff (m/s):0.00
Set zero flow (m/s):0.00
Damping (secs):10
Totaliser:Run
Reset +Total
Reset –Total
11.Scroll down and select Exit to return to the
FLOW READING screen which will now indicate
the instantaneous flow together with the totalised
flow.
Note thatinsomeinstallation the measured flow
can be in either direction.Where this is the case
the upstreamflowis shown separately in the –
Total field.
FLOW READINGDD-MM-YY HH:MM:S
Dim:mm
Signal
l/min
+Total: 300.0litres
–Total: 0.00litres
Page 26www.greyline.com
Calculating the average flow
To calculatetheaverage flow wait for the allotted monitoring period to expire then divide the indicated total flow
by the time taken. This will give you the average flow in m/s, galls/hours or whatever units you select.
Note thatina bi-directional flow situation you must calculate the difference between the indicated positive and
negative flowtotalsbefore carrying out the average flow rate calculation.
How to stop the totaliser temporarily
If you want to stop the totaliser temporarily for operational reasons, set the Totaliser option to Stall in the FLOW
READING OPTIONS screen as described above. This will stop the totaliser operation without affecting its current
values.
Page 27www.greyline.com
3.7Configuring the Portaflow PT400 Interfaces
3.7.3 How to turn the 4-20mA output OFF/ON
1.With the instrument operating in the FLOW
READING mode, press the 4-20mA function key.
This will access the 4-20mA OUTPUT screen.
2.The ON/OFF status of the 4-20mA output is
shown on line 2 ofthedisplay.
3.To changetheON/OFF status select Output
Range and press ENTER.
3.7.3
4.Select Off, to turn OFF the 4-20mA Output or
select one of the output ranges to turn it ON.
5.Press ENTER to return to the 4-20mA
OUTPUT screen.
3.7.44-20mA signal calibration and ranging
4-20 mA OUTPUTDD-MM-YY HH:MM:SS
4-20 mA O/P is ONDim: mm
mA OutputReading:0.00
Output Range:4-20
Units:l/min
Flow at max. output:0.00
Flow at min. output:0.00
Output mA for error:22.00
Exit
4-20 mA OUTPUTDD-MM-YY HH:MM:SS
Dim: mm
Off
4-20mA
0-20mA
0-16mA
3.7.4
Key Point: The 4-20mA output has been calibrated in the factory and should not
require further adjustment. In the rare event that re-calibration is necessary, this
procedure should be carried out only by a trained engineer.
This procedure describes how to calibrate the 4-20mA output and ‘scale’ it to operate over a defined flow-rate
range.
Signal calibration
1.Select Setup Instrumentfrom the MAIN MENU,
to access the SETUP INSTRUMENT screen.
2.Select Calibrate4-20mA.
SETUP INSTRUMENTDD-MM-YY HH:MM:SS
Set Date & Time :dd-mm-yyhh:mm:ss
Calibrate 4-20mA
Pulse output
Backlight:Disabled
Factory settings
Change Language
Exit
Dim: mm
Page 28www.greyline.com
3.Connect a calibrated ammeter to the 4-20mA
output andadjusttheUP/DOWN Scroll keys
(Coarse) andLEFT/RIGHT Scroll keys 5 & 6 (fine)
until theoutput is exactly 4.0mA.
The DAC should indicate approximately 8000.
4.Press ENTER when done.
CALIBRATE 4mADD-MM-YY HH:MM:SS
Dim: mm
Adjust theoutputcurrent to 4mA
Use to set, 5/6 to trim
DAC Value:8000
Press when done
5.With the meter still connectedto the 4-20mA
output adjusttheScrollkeys to obtain an output of
exactly 20mA.
The DAC should indicate approximately 40000.
6.Press ENTER when done.
CALIBRATE 20mADD-MM-YY HH:MM:SS
Dim: mm
Adjust theoutputcurrent to 20mA
Use to set, 5/6 to trim
DAC Value:40000
Press when done
4-20mA Signal scaling
Note: The 4-20mA can be set to represent a particular flow range. It is also possible to enter a negative figure for
the minimum outputandthis would enable a reverse flow to be monitored.
7.With the instrument operating in the FLOW
READING mode, press the 4-20mA function key.
This will access the 4-20mA OUTPUT screen.
4-20 mA OUTPUTDD-MM-YY HH:MM:SS
4-20 mA O/P is ONDim: mm
mA OutputReading:0.00
Output Range:4-20
8.Select Flow at max. output and enter a value of
the flowrate that you want to associate with a
20mA output.
Units:l/min
Flow at max. output:0.00
Flow at min. output:0.00
Output mA for error:22.00
9.Select Flow at min. output and enter a value of
the flowrate that you want to associate with a 4mA
Exit
output.
This could be ‘0’.
10.Select Output mA for error and enter a value (max of about 23mA) that you want the 4-20mA output
to produceintheevent of an error (e.g. if the flow-rate is outside the set range).
11.Upon completion press ENTER to return to the FLOW READING screen.
3.7.5How to convert the measured current to flow rate
Assume the maximum flow rate is F
(l/min) and the minimum flow rate F
max
Page 29www.greyline.com
is ‘0’ (l/min), as shown.
min
F
max
Flow (l/min)
F
min
48121620
To calculatetheflowrate (l/min) for a measured current I(mA) then:
[0-16mA scale]
[0-20mA scale]
[4-20mA scale]
I (mA)
0-20mA
Flow rate
IF
-------------------------------------------F
–
maxFmin
20
min
Flow rate
+=
3.7.6How to turn the pulse output OFF/ON
1.With the instrument operating in the FLOW
READING mode, press the Pulse function key to
access thePULSE OUTPUT screen.
2.A Pulse output is ON message appears in the
second lineof the display.
3.Select Exit and press ENTER to return to the
FLOW READING screen.
3.7.7Pulse output signal calibration
1.With the instrument operating in the FLOW
READING mode, press the Pulse function key to
access thePULSE OUTPUT screen.
2.To changetheflowmeasurement units select
Flow units and press the required units function
key (7), (8), (9).
0-16mA
IF
-------------------------------------------F
–
maxFmin
16
+=
PULSE OUTPUTDD-MM-YY HH:MM:SS
Pulse output is ONDim: mm
Flow units:litres
Output:Off
Vol perpulse:10.00
Pulse width (ms):10
Exit
PULSE OUTPUTDD-MM-YY HH:MM:SS
Pulse output is ONDim: mm
3.Select Volper pulse and enter the required value. (In the example shown, a pulse is produced
every 10 litres of flow).
4.Select a Pulse width (inms) to suit theparticular application – e.g. electro-mechanical counter.
Refer to the manufacturer’s data sheet for the minimum pulse width.
5.Select Exit and press ENTER to return to the FLOW READING screen.
Page 30www.greyline.com
4:Maintenance & Repair
This instrumentdoes not contain any user-serviceable parts. The following notes are provided as a guide to
general equipment care
Do not disassemble this unit unless advised by Greyline.
Return the unit to an approved service agent or place of purchase
WARNING
1.Ensure the unit is switched off and disconnected from the mains, then wipe the exterior of the
instrument with a clean, damp cloth or papertowel. The use of a solvent may damage the surface.
2.The instrument contains a rechargable battery,dispose safely and inaccordance with the local
regulations in force in the country of operation.
3.Ensure allcables and connectors are kept clean and free from grease or contaminants.
Connectors may be cleaned with a general purpose cleaner if necessary.
4.Avoid theuseofexcessive grease/ultrasonic couplant on the sensors as this may impair the
performance of the equipment. Excessive grease/couplant can be removed from the sensors and guide rails
using anabsorbent paper towel and a general purpose solvent cleaner.
for further advice.
5.We recommend that the ultrasonic couplant is replaced on the sensors every 6 months,
especially onpipes where the application is too hot to touch. If the signal level drops below 30% this is also
an indication that the sensors need re-greasing.
6.Regularly checkall cables/parts for damage.Replacementparts are available from Greyline.
7.Ensure the person who services your instrument is qualified to do so. If in doubt, return the
instrument to Greyline with a detailed report on the nature of any problem.
8.Ensure thatsuitable precautions are taken when using any materials to clean the
instrument/sensors.
9.The instrument andsensors should be calibrated at least once every 12 months. Contact
Greyline or your local service agent for details.
10.When returning product to Greyline make sure it is clean and please notify Greyline if the
instrument has been in contactwith any hazardous substances.
11.If the instrument was supplied with dust or dirt caps make sure they are re-fitted when the
instrument is not in use.
Page 31www.greyline.com
5:Troubleshooting
5.1Overview
If you have a problem with your flow monitoring system it can be due to any of the following:
Faulty instrument
If you suspect the instrument is faulty you can check it out using a test block as described in Paragraph 5.4.. This
will establish that the instrument is functional and receiving a healthysignal from the connected transducers.
Incorrect setup
A low, or zero, signal could be caused byincorrect set-up such as:
•Incorrect sitedata entered into the instrument.
•Incorrect or non-matching ultrasonic transducers selected for use.
•Incorrectly fittedtransducers – lack of couplant applied, incorrect spacing, insecure
attachment.
•Poor connectionsbetween the probes and the instrument.
Application problem
If you are certain thatthe instrument is healthy and suitably set-up for the current site; and the probes are
properly assembledandfittedcorrectly, there could be an application problem concerned with the site.
Check suchconditions suchas:
Poor pipe outer surface quality
•Uneven surface preventing good surface contact with the transducer.
•Flaking paint (should be removed).
•Variable air gap in concrete-covered pipes affecting the ultrasonic signal quality.
Poor internal pipe construction
•Rough internal pipe walls affecting fluid flow (see roughness factor).
•Internal welds positioned in the transducer signal path affecting the signal quality.
•The ‘drippings’in galvanised-dipped pipes or other irregularities interferingwith the signal
path.
Incorrect probelocation
•Transducers locatedtoocloseto bends or valves,disturbing the flow profile.
•Transducers locatedtoocloseto insertion probes, disturbing the flow profile.
•For horizontal pipework transducers should not be positioned on the top of the pipe.
Poor fluid conditions within the pipe
•Fluid containsbubbles, high particle density or sludge.
•Air in the top of the pipe.
Low fluid flow within the pipe
•Pipe obstructions.
•Malfunctioning valve not opening fully (or closed inadvertently).
Liquid content problems
•Multiple liquidcontentsdo not comply accurately to expected sound speed criteria.
•Very hot pipe almost turns water to steam and therefore exhibits the wrong speed
characteristics –could be due to reduced pipe pressure.
•Flashover – liquid turns into a gas because of lower than required pressure.
Page 32www.greyline.com
5.2General Troubleshooting Procedure
Yes
Yes
Ensure sufficient acoustic couplant been applied to
If triple reflex modeis selected try double or single
START
Are any
status messages
Note any active messages.
Yes
Is the display blank?
Is the display
scrambled or
hung-up?
No
No
Recharge the battery.
If battery won't recharge then
replace the charger (if faulty) or
return the instrument for repair.
Ensure temperature is set correctly.Ensure transducer cables are connected correctly.
Ensure transducers are positioned in accordance
with the recommended distance from bends etc.
Ensure flow velocity >0.01m/s.
Problem probably due to non-constant fluid flow.
Ensure pipe data beenentered correctly.
Ensure fluid type been entered correctly.
Ensure correct transducer type been selected.
Ensure transducer separation distance correct.
Ensure transducers are aligned correctly.
Ensure pipe diameterwithin specifications of the
selected transducers.
Ensure the pipe is completely full.
Ensure pipe surface is not corroded, or protective
surface loose.
Check for particles in the fluid.
Refer to the message table in this
chapter to interpret themessage
and view suggested response.
the transducers?
reflex instead.
Check instrument using test block.
Figure 5.1 Troubleshooting chart
Page 33www.greyline.com
5.3Warning & Status Messages
FLOW RATE ERRORS
No flowsignal
Flow signal is poor
Zero cut-offerror!
Damping error!
Totaliser beyondmaximum!
Interpretation: This message appears when the transducers cannot send or
receive signals to each other.
Response: Firstly check that all cables are connected, transducers are on the pipe
correctly with sufficient couplant on the face.
This conditioncouldalso be due to a partially empty pipe, aerated liquid, particulate
content too high or when the condition of the pipe beingmeasuredispoor.
Interpretation: This warning appears when the signal is lower than 25%.
Response: This could be dueto an application problem, a poor quality pipe – see
also theconditions for No flow signal (above). Check for sufficient couplant.
Interpretation: You have entered anout-of-range value in the Zero cutoff field in
the Optionsmenu.
Response: Entera valid number.
Interpretation: You have entered anout-of-range value in the Damping field in the
Options menu.
Response: Entera valid number.
Interpretation: The totaliser has overflowed its maximum count. The counter will
roll-over and restart from zero but this message alerts you to the fact.
PULSE ERRORS
Pulse Rate > Max
Pulse volume error!
Pulse width error
Response: Reset the totaliser as described in Paragraph 3.6.1.
Interpretation: The flow rate exceeds the capability of the pulse output – i.e. too
many pulses per second are required than can be achieved.
Response: Narrowthe pulse width timeor increase the volume per pulse, as
described in Paragraph 3.7.7.
Interpretation: You have entered anout-of-range value in the Pulse volume error
field in the PULSE OUTPUT menu – see Paragraph 3.7.7.
Response: Entera valid number.
Interpretation: You have entered anout-of-range value in the Pulse width error
field in the PULSE OUTPUT menu – see Paragraph 3.7.7.
Response: Entera valid number.
Page 34www.greyline.com
4-20mA ERRORS
mA out> Max
Interpretation: The actual flow is higher than the maximum set on the mA range.
Response: Re-scale the 4-20mA output to be able to cope with the higher flow –
see Paragraph 3.7.4.
Calibration 20mA Error!NOTE: The 4-20mA output is calibrated before the instrument leaves the factory
and should not require further adjustment.
Interpretation: You have adjusted the DAC outside its accepted range when
calibrating the 20mA signal output.
Response: Re-calibratethe4-20mAoutput – see Paragraph3.7.4.
Calibration 4mAError!NOTE: The 4-20mA output is calibrated before the instrument leaves the factory
and should not require further adjustment.
Interpretation: You have adjusted the DAC outside its accepted range when
calibrating the 4mA signal output.
Response: Re-calibratethe4-20mAoutput – see Paragraph3.7.4.
BATTERY ERRORS
Battery Low
Interpretation: The battery has discharged to below 30% remaining. This leaves
the instrument with approximately 4 hours remaining, depending on power usage,
before it needs recharging.
Battery Exhausted
SET-UP ERRORS
Pipe OD out of range
Wall thickness out of range
Lining thickness out of range
Response: Rechargetheinternal battery at the earliest opportunity. Do not leave
the instrument for a prolonged period with a fully discharged battery.
Interpretation: The battery is approaching a fully discharged state and the
instrument is about to store the internal data and shut-down.
Response: Rechargethebattery.
Interpretation: You have entered anout-of-range value for the pipe outside
diameter dimension – i.e. larger or smaller than the unit or sensor can be used on.
Response: Entera valid number.
Interpretation: You have entered anout-of-range value for the pipe wall thickness
dimension – accepted range is 0.04” - 2.95” (1mm - 75mm).
Response: Entera valid number.
Interpretation: You have entered anout-of-range value for the lining thickness
dimension – acceptable range is 0” – 0.39” (0mm - 10mm).
Response: Entera valid number.
Page 35www.greyline.com
Temperature range
Interpretation: You have entered anout-of-range value for the fluid Temperature.
Accepted temperature range -4°F to 572°F (-20°C to +300°C).
Response: Entera valid number.
Invalid Date or Time
Sensors: INVALID
Mode: Err Typ
Interpretation: The entered Date or Time is invalid, or when setting up ‘timed’ data
logging the Stop time is set earlier than the Start time.
Response: Entera valid Date and Time.
Interpretation: The selected temperature is higher than the maximum allowed for
the sensor type.
Response: Selectalternative sensors or change the temperature.
Interpretation: The selected sensors are invalid and the mode cannot be verified.
Response: Selecta valid sensor typeand choose a modethat gives a non-zero
separation distance.
Page 36www.greyline.com
Test Block
A testblockis included with the Portaflow PT400 equipmentto
allow the transducers and inter-connecting cables to be
functionally checked.
1.Switch ON the instrument.
2.Select Quickstartandentertheparameters
shown in the table belowfor the appropriate transducer
type (A or B):
ParameterA SensorsB Sensors
Pipe outside diameter1.02” (26.0mm)2.09” (53.0mm)
Pipe wall thickness0.24” (6.0mm)0.28” (7.0mm)
Pipe lining thickness0.00.0
Pipe wall materialMild SteelMild Steel
Fluid typeWaterWater
Temp68°F (20°C)68°F (20°C)
Figure 5.2 Test block
3.When the above data is entered, the SENSOR SEPARATION screen will be displayed.
4.Use the UP/DOWN scroll keyto go to the SENSOR SELECTION menu.Select the appropriate
sensor (the default will be"A")and press ENTER to return to the SENSOR SEPARATION menu.
5.Select Sensormode and position the cursor at Diagonal and press ENTER to return to the
SENSOR SELECTION menu.
6.Select Exit and press ENTER to return to the SENSOR SEPARATION screen.
7.Check thattheparametersdisplayed are correct.
8.Apply acoustic couplant to thesensors and attach them to the test block with the connectors
positioned towards the centre of the test block as shown, and temporarily secure them in place using elastic
bands or tape.
9.Connect the sensors to the Portaflow PT400 instrumentusing the cables provided.
10.Press ENTER to go to the FLOW READING screen.
11.Select the Options key to go to the FLOW READING OPTION menu and set the Damping to at
least 10seconds.
12.Select Exit and press ENTER to return to the FLOW READING menu.
13.The flow reading value displayed is not important. The fact that a reading is obtained indicates
that the instrument is functioning. This value may fluctuate but this is normal.
14.The signal strength indicator at the left of the display should show 3–4 bars.
5.5Microprocessor Reset Facility
In therare event that thePortaflow PT400 instrument appears to totally hang-up, or displays total gibberish, you
can reset its microprocessor by carefully inserting a straightened paperclip into the pinhole located in the right-
Page 37www.greyline.com
hand side of the instrument to operate the internal reset switch. Hold the paperclip perpendicular to the instrument
while doing this.
Page 38www.greyline.com
5.6Diagnostics Display
This feature is designed for advanced users and is intended to provide information that will aid the user to
diagnose problems– e.g. no signal strength.
When operatinginthe FLOW READING mode you can access a diagnostics screen by pressing the Options
function key and then selecting Diagnostics from the FLOW READING OPTIONS screen. This will display the
operating values for the following parameters.
Calculated time (µs)
This is a value theinstrument predicts will be the time in µsecs that it should take for the acoustic wave to
propagate acrossa particular pipe size. This value is ascertained from the data entered by the user. i.e. Pipe size,
material, sensorsetetc.
Actual time (µs)
This is the value the instrument measures as the time taken for the acoustic wave to propagate across the pipe. It
is used to see if the signal is being taken from the burst, at the correct time to get the strongest signal. This value
is normally a few µs below the calculated µs value. If, however, this value is much greater than the calculated
time thenthere is a problem with the set-up.
Flow (m/s)
This displaysflowvelocityinm/secto 3 decimal places.
Signal strength
This is the averaged valueof Signal and should be a value between 800 and 1600 – where 800 is approximately
50%, and 1600 is approximately 100%.
UP/DN time difference
The differenceintransittimesbetween the upstream and downstream signals due to the fluid flow.
Propagation µs
Not currently in use.
Signal propagation
Not currently in use.
Fluid propagation rate
This is the sound speed of the fluidcalculated using the data entered by the user.
Fluid propagation rate
Calculated from the input dataconcerning the fluid type and temperature.
Sensor separation
The same value as displayedin the setup screen.
Page 39www.greyline.com
5.7 APPLICATIONS HOTLINE
For applications assistance,advice or information on any Greyline Instrument contact your Sales Representative, write to
Greyline orphone the Applications Hotline below:
United States:Tel: 315-788-9500Fax: 315-764-0419
Canada:Tel: 613-938-8956Fax: 613-938-4857
Toll Free:888-473-9546
Email:info@greyline.com
Web Site:www.greyline.com
Greyline Instruments Inc.
Canada:USA:
16456 Sixsmith Drive11451 Belcher Road South
Long Sault, Ont. K0C 1P0Largo, FL 33773
5.8 PRODUCT RETURN PROCEDURE
Instruments maybereturned to Greyline for service or warranty repair.
1. Obtain an RMA Number from Greyline -
Before shipping a product to the factory please contact Greyline by telephone, fax or email to obtain an RMA number
(Returned Merchandise Authorization). This ensures fast service and correct billing or credit.
When you contact Greyline please have the following information available:
1.Model number / Software Version
2.Serial number
3.Date of Purchase
4.Reason for return (description of fault or modification required)
5.Your name, company name, address and phone number
2. Clean the Sensor/Product -
Important: unclean products will not be serviced and will be returned to the sender at their expense.
1. Rinse sensor and cable toremove debris.
2. If the sensor has been exposed to sewage, immerse both sensor and cable in a solution of 1 part household
bleach (Javex, Clorox etc.) to 20 parts water for 5 minutes. Important: do not immerse open end of sensor cable.
3. Dry with paper towels and pack sensorandcablein a sealed plasticbag.
4. Wipe theoutside of the enclosure to remove dirt or deposits.
5. Returnto Greyline for service.
Page 40www.greyline.com
LIMITED WARRANTY
_____________________
Greyline Instruments warrants, to theoriginal purchaser,its products
to be free from defects in material and workmanship for a period of
one year from date of invoice. Greyline will replace or repair, free of
charge, any Greyline product if it has been proven to be defective
within the warranty period.This warranty does not cover any
expensesincurred intheremovalandre-installationoftheproduct.
If a product manufactured by Greyline shouldprove defective within
the first year, return it freight prepaid to Greyline Instruments along
withacopy ofyourinvoice.
Thiswarrantydoesnotcoverdamages duetoimproper installation or
handling, actsof nature, or unauthorized service.Modifications to or
tampering with any part shall void this warranty. This warranty does
not cover any equipment used in connection with the product or
consequentialdamages duetoadefectin the product.
Allimpliedwarrantiesare limited to thedurationof this warranty. This
is the complete warranty by Greyline and no other warranty is valid
against Greyline. Some states do not allow limitations on how long
an implied warranty lasts or limitation of incidental or consequential
damages, so the above limitations or exclusions may not apply to
you.
This warranty givesyou specific legal rights, andyou may also have
otherrights whichvaryfromstateto state.
GreylineInstrumentsInc.
Page 41www.greyline.com
Appendix A:Specifications
GENERAL
NEW! DSP Measurement
Technique:
Timing Resolution:50 pico-second, continuous signal level indication on display.
Improved! Flow Velocity
Range:
Turn Down Ratio:100:1
Accuracy:±0.5% to ±2% of flow reading for flow rate >0.2m/s and Pipe ID >75mm.
Repeatability:±0.5% of measured value or ±0.02m/s whichever is the greater.
Pipe Lining:Applicable pipe linings include Rubber, Glass, Concrete, Epoxy, Steel.
Pipe Lining Thickness:0mm – 10mm.
Pipe Wall Temperature
Range:
Min 13mm; Max 4000mm with D sensorset.
Standard sensor operating temperature is -4°F to 275°F (-20°C to +135°C).
Optional high temperature sensor operating temperature is -4°F to 392°F ( -20°C to +200°C).
TRANSDUCER SETS
Standard:Temperature Range -4°F to 275°F (-20°C to +135°C)
PF400A – Type 'A-ST' (2MHz)
PF400B – Type 'B-ST' (1MHz)
Page 42www.greyline.com
LANGUAGES
Standard Supported
Languages:
OUTPUTS
Analog Output:
Resolution:
Alarm Currents:
Isolation:
Maximum Load:
0.1% of full scale.
Any between 0–26mA.
1400V Opto-isolated.
620 Ohms.
Number Available: One open collector opto-isolated digital output.
Up to 400 pulses/sec (depending on pulse width).
400ms for 1 pulse/sec.
5ms for 100 pulses/sec.
150mA.
Battery:
Technology:5-cell NiMH.
Capacity:3.8AHr.
Operating time:Typically 20 hours continuous with backlight and 4-20mA output OFF.
Recharge Time:6.5 Hours.
Service Life:>400 charge/discharge cycles.
Power Supply/Charger:
Manufacturer:ECOPAC Model ECO-181WP12.
Input Voltage Range:90–264Vac.
Input Frequency
Range:
Output Voltage:12Vdc.
Max. Output Current:1.5A.
Approvals:UL, CUL, TUV, CB & CE.
47–63Hz.
MECHANICAL
Carrying case:
Rating:All components are contained in a hard-wearing polypropylene carrying case with a protective moulded foam
insert.
Page 43www.greyline.com
Enclosure:
Material:Flame retardant injection moulded ABS.
Dimensions:264mm x 168mm x 50mm.
Weight (Including
Battery):
Protection:IP54.
Keypad:
No. Keys:16.
Display:
Format:240 x 64 pixel graphic display, high contrast black-on-white, with backlight.
Viewing Angle:Min 30°, typically 40°.
2.5lbs (1.1 kg).
ENVIRONMENTAL
Operating Temperature:-4°F to 122°F (–20°C to +50°C).
Storage Temperature:-13°F to 149°F (–25°C to +65°C).
Operating Humidity:90% RH MAX at 122°F (+50°C).
APPROVALS
Safety:BS EN 61010.
EMC:BS EN 61326 - 1:2006, BS EN 61326-2-3:2006.
Battery Charger:EN61204 - 3.
SHIPPING INFORMATION
Box Dimensions:16.15” x 8” x 14” (410mm x 205mm x 355mm).
Weight:16.5lbs (7.5 kg).
Volumetric Weight:11lbs (5. kg).
Greyline Instruments reserves the right to alter specifications without notice.
Page 44www.greyline.com
PIPE CHARTS
Page 45www.greyline.com
Page 46www.greyline.com
Page 47www.greyline.com
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.