OCF 5.0 Open Channel Flow Monitor
CONNECTIONS:
POWER INPUT: The standard model requires AC power input between 100 to 240 VAC 50/60Hz. No
adjustments are necessary for voltages within this range.
Optional DC: 9-32 VDC. Connect to + and - terminals.
Optional Thermostat and Heater modules are available rated for 115 VAC or 230 VAC.
IMPORTANT NOTE: To comply with CSA/UL electrical safety standards, AC power input and relay
connection wires must have conduit entry to the instrument enclosure. Installation requires a switch,
overcurrent fuse or circuit breaker in the building (in close proximity to the equipment) that is marked
as the disconnect switch.
Risk of electric shock. Loosen cover screw to access connections. Only qualified personnel
!
should access connections.
Note: Use of instrumentation over 40°C ambient requires special field wiring.
Note: User replaceable fuse is 2 Amp 250V (T2AL250V).
Page 4
CONNECTIONS
OCF 5.0 Open Channel Flow Monitor
SENSOR
DATA LOGGER
4-20mA/0-5V
RLY2
RLY1
–
+
NO
C
NC
NO
C
NC
EXTRA RELAYS OPTION
NC
RLY3
AC
LN
AC
GROUND
NC
C
POWER
INPUT
C
NO
NO
RLY4
HEATER OPTION
NC
C
RLY5
NO
NC
C
RLY6
NO
GND
USB A
Plug
Page 5
OCF 5.0 Open Channel Flow Monitor
KEYPAD SYSTEM
The OCF 5.0 uses a menu system. Arrows show the four directions to leave a menu box. Press a key to
move to the next item in the direction shown. Move the cursor (underline) under numerals and increase
or decrease numerals with the and keys.
To store calibration values permanently (even through power interruptions), press .
Total50138 USG
Average 34.82 USG/m
Maximum 52.20 USG/m
Max Time11:08:00
Minimum 0.000 USG/m
Min Time9:15:00
RUN
The main display shows the units selected from the Units/Mode menu, Level,
Range, Flow, HRT, and Volume rate being measured, TOTALIZER and RELAY
states. The OCF 5.0 will start-up with this display and will return to this screen
after a timeout if keys are not pressed in other menus.
MESSAGE
³
Press from the RUN display to view status of the data logger and error/warning
messages provided by the instrument. The word
Message will appear on the
RUN display if error messages are being generated by the instrument. Refer to the
manual section Error/Warning Messages for a description. Press to return to the
main display.
STATUS
³
Press from the RUN display to view instrument status. Range will be displayed
in linear units.
Tot Displays the current totalizer reading.
EC Displays echo confidence.
Relays Energized relays will display as a white character on a black
background.
24 HR LOG
Press from the RUN display to view a formatted flow report from instruments
with a built-in data logger. Press to scroll down one day or repeatedly to scroll
to a specific date. Up to 365 days can be stored. Newest date will overwrite the
oldest. Press to return to the main display.
Page 8
OCF 5.0 Open Channel Flow Monitor
--Password----------
Password0000
PASSWORD
The
Password (a number from 0000 to 9999) prevents unauthorized access
to the
Calibration menu.
From the Run display press to get to
is 0000 and if it has not been changed press the to proceed to the
Selections
screen.
If a password is required, press to place the cursor under the first digit and or
to set the number, then to the second digit, etc. Press or to proceed to
the
Menu Selections screen.
A new password can be stored by going to
Password
.
Password. Factory default password
Menu
Special Functions/New
Page 9
OCF 5.0 Open Channel Flow Monitor
--Units/Mode-------ModeFlow
}
Linearft
VolumeUSG
Timemin
TemperatureC
--Units/Mode--------
ModeFlow
Lineari
}n
--Units/Mode--------
VolumeUS
}G
ft3
bbl
IMG
USMG
--Units/Mode-------TemperatureC
}
ft
mm
m3
IG
UNITS/MODE
From
Flow, HRT or Volume.
Mode press the and then the or to select Level, Range,
Range displays distance from the sensor to the target or liquid surface like a
tape measure. Range mode is useful to measure the exact distance
from the sensor to the zero level during calibration.
Level used to measure tank level in linear units, or "Head" in an open
channel for comparison with flow tables.
m
Volume displays tank inventory in engineering units.
HRT select Horizontal Round Tank mode to calculate and display volume
units in a horizontal round tank.
Flow calculates open channel flow through a flume or weir.
L
Volume, HRT or Flow gives the additional choice of volumetric units:
ft3 - cubic feet
USG - US gallons
USMG - US million gallons (FLOW only)
IG - Imperial gallons
F
IMG - Imperial million gallons (FLOW only)
m3 - cubic meters
L - liters
bbl - U.S. oil barrel
Press to store the selection, then to the next menu item and to enter.
Linear press and then or to select units of measurement.
Press to store the selection.
Time press and then or to select day, hr, min or
sec.
Temperature press and then or to select C or F (Centigrade or
Fahrenheit).
Press or to return to the Menu Selections screen.
Page 10
OCF 5.0 Open Channel Flow Monitor
--Calibration------Min Range1.000ft
}
Max Range12.000ft
Damping10%
LOE Time30sec
CALIBRATION
Press to
Calibration and to enter. Use or to position before
each menu item and to enter. When settings are completed press to store and
return to the Calibration menu.
Min Range Distance from the sensor face to highest expected level.
Max Range Distance fro m sensor to Zero level.
MaxVol For Volume and HRT mode only. Enter the maximum volume of
the vertical or horizontal tank.
Damping Minimum damping allows fast response to level changes.
Increasing damping slows the OCF 5.0's response to level changes
and is ideal to smooth the display and outputs in turbulent
conditions. Damping value is shown in percent (0-99%). Some
experimentation may be required to select the optimum damping
value. A value of 1% is recommended for most applications and for
fast level changes (up to ½ inch/sec - 13 mm/sec).
LOE Time Press and or to change the number of seconds without
receiving an echo before the OCF 5.0 displays ECHO LOSS, and
Control relays change state as calibrated under Relay Parameters.
Factory default is 30 seconds and is recommended for most
applications, Minimum is 1 second and maximum is 99 seconds.
Press from the
Units/Mode display to return to Menu Selections.
Page 11
OCF 5.0 Open Channel Flow Monitor
PAL MER BOWLUS
MAX LEVEL
V-NOTCH
ZERO LEVEL
SENSOR
SENSOR
MIN
RANGE
MAX LEVEL
MAX
RANGE
ZERO LEVEL
MIN
RANGE
MAX
RANGE
CALIBRATION
- for Open Channel Flow
1. Before starting the calibration determine:
a) MAX RANGE = ____________________ (Maximum range = distance from the Sensor to Zero flow point)
b) MAX LEVEL = ____________________
(Maximum level of flow through flume or weir)
c) MIN RANGE = __________________
(Distance from sensor to Max Level)
Minimum range = MAX RANGE - MAX LEVEL
(must be at least 8" / 20.3 cm depending on sensor model).
2. Check the maximum range with the sensor installed by:
a) When liquid is at zero level press to view the
Status menu. Use this range measured by the OCF 5.0 as the Max
Range
setting.
Range reading in the
or
b) Carefully measure distance from sensor to zero level with a tape measure,
and use this measurement as the
NOTE
: The OCF 5.0 will not detect targets beyond user entered Max Range.
Relay Press and or to select a relay (2 relays are standard, 4
additional are optional).
Function Press or to select Off, Temperature, Pulse
or
Flow.
Temperature Air temperature at the sensor location.
Press and and set the relay On and
Off for specific temperatures.
Pulse Press and set digits to the flow volume
increment required between relay pulses.
Use this feature for remote samplers,
chlorinators or totalizers. Minimum time
between pulses is 2.25 seconds and pulse
duration is 350 milliseconds.
Flow Press to Mode and to select
Pump, Low Alarm or Hi
Alarm
.
Note:
Function setting Flow will change according to selections made
under
Units /Mode eg Level, Range, Volume or HRT.
Press to return to
Menu Selections.
LOE Set relay state for ‘echo loss’ event (On, Off or Hold)
Alternate Pumps can be set to alternate when in pump mode.
Page 14
OCF 5.0 Open Channel Flow Monitor
--Data Logging------Log Site ID00
}
ModeLevel
Set Date Feb 18/2008
Set Time11:27:40
Interval10sec
LogStart
Mar 19/2009
12:28:41
99
Range
Flow
HRT
Volume
60min
30min
10min
5min
2min
1min
30sec
Stop
Delete
DATA LOGGING
Setup
Select Data Logging from Menu Selections.
Log Site ID
Enter a number from 00 to 99. The site ID will
become part of the downloaded file name to help
distinguish downloads from different instruments. Press
to store the setting.
Mode
Select Level, Range, Flow, HRT and
Volume. Flow (e.g. USGPM or l/sec). Press to
store the setting.
Set Date
Press or to scroll and select Month, Day and Year.
Press to store the setting.
Set Time
Press or to select the current time in Hours, Minutes
and Seconds. Press to store the setting.
Interval
Press or to select the logging interval. Flow rate
reading will be stored at each time interval. Press to
store the setting.
Note: Press to Log and or to Delete and to
delete the log file. Press and or to Start and to
restart the logger.
Log
, Start or Delete the log file. Delete old
Stop
file and start a new log to apply any changes that have
been made to the
Interval.
Log Site ID, Mode or
RETRIEVE LOG FILE
Plug a USB Flash Memory Drive (not supplied by Greyline) into the USB output
cable from the instrument. The instrument display will show the message
Downloading until the log file is transferred to the memory card and then display
Completed. The USB flash drive may be removed.
Download file names will appear in this format:
OCF__00A.LOG
MODELTAG DOWNLOAD
Page 15
OCF 5.0 Open Channel Flow Monitor
Tag is set according to the Log Site ID entered in the instrument Data Logging
menu.
Download letter will be A for the first download from an instrument. B for the
second, then C etc. At the letter Z a - character will appear indicating that the
maximum number of downloads for that instrument are on the USB flash drive.
Older files can be erased or moved from the flash memory drive or a new memory
drive can be used.
OPENING LOG FILES
Install Greyline Logger on your PC or laptop. Refer to the Help menu in the
program for detailed instructions.
Select File/Open/Instrument Log (.log) to open the log file from your USB flash
drive.
Page 16
OCF 5.0 Open Channel Flow Monitor
--Special FunctionsLanguageEnglish
}
Analog Out4-20mA
BacklightHigh
Reset TotalizerNO
Restore DefaultsNO
New Password0000
--Special Functions-
LanguageEnglish
BacklightHig
}h
Medium
Low
Key Hi/Lo
Key High
Key Med
Key Low
Off
SPECIAL FUNCTIONS
Language Select English, French or Spanish and
press .
Analog Out Select 4-20mA or 0-5V mode for the analog output.
Backlight Select High, Medium or Low for continuous
backlight.
Select
minute) after a keypress and then
Key Hi/Lo for high backlight (for 1
Lo backlight
until a key is pressed again.
Select
Key High, Med or Low for backlight
after a keypress and then backlight off until a key is
pressed again.
Reset Totalizer Press and select Yes to erase and restart the
totalizer at zero.
Restore Defaults Select Yes and press to erase all user settings
and return the instrument to factory default settings
New PasswordSelect any number from 0000 to 9999 and press .
Default setting of 0000 will allow direct access to the
calibration menus. Setting of any password greater
than 0000 will require the password to be entered to
access the calibration menus.
Press to return to
Menu Selections.
Page 17
OCF 5.0 Open Channel Flow Monitor
--Simulation-------TestActual
}
Flow250USG/m
4-20mA Flow4.00
Relays 1 2 3 4 5 6
SIMULATION
Exercises the 4-20mA (0-5V) output, digital display and control relays.
Test Select Maximum and press to simulate maximum Flow, Range or
Level and to output 20mA (5V) to the analog channel.
Select
Minimum and press to simulate minimum Flow, Range or
Level and to output 4mA (0V) to the analog channel.
To simulate and intermediate Flow, Range or Level set
Actual and then enter a value for the Flow, Range or
Level. The analog output and control relays will respond to the
simulated value.
Test to
Page 18
SENSOR MOUNTING METHODS
Notes:
1. Use the ¾" NPT "Isolation
Coupling" supplied and
only. Do not clamp sensor
tighten
body or stem.
2. Locate the sensor 1 ft (30 cm) from
the sidewall or obstruction for every
10 ft (3 m) depth.
3.mount in direct sunlight.
Do not
4. Extend sensor cable up to 500 ft
(150 m) with RG62AU coaxial only.
CROSS BAR MOUNT
CLAMP
DO NOT
CLAMP IN
THIS AREA
hand
3/4"
CONDUIT
ISOLATION
COUPLING
(SUPPLIED)
MUST BE
USED
FLANGE MOUNT
4" OR 6" BLIND FLANGE
TAPPED ¾" NPT
¾" NPT
NIPPLE
DO NOT
CLAMP IN
THIS AREA
OCF 5.0 Open Channel Flow Monitor
FLEXIBLE
CONDUIT
JUNCTION
BOX
(OPTION JB)
ISOLATION
COUPLING
(SUPPLIED)
MUST BE
USED
CONDUIT MOUNT
3/4"
CONDUIT
STANDPIPE LENGTH
AS SHORT AS POSSIBLE
STANDPIPE DIAMETER
AS LARGE AS POSSIBLE
TYPICAL STANDPIPE:
4” / 100 mm DIAMETER
12” / 300 mm LENGTH
NARROW DIAMETER
STANDPIPES (<4” / 100 mm)
MAY AFFECT ACCURACY
OF READING
STAND PIPE MOUNT
SMOOTH
GRIND OR FILE
PIPE EDGE
DO NOT
CLAMP IN
THIS AREA
3/4" NPT
DO NOT
CLAMP IN
THIS AREA
ISOLATION
COUPLING
(SUPPLIED)
MUST BE
USED
ANGLE MOUNT
NIPPLE
ISOLATION
COUPLING
(SUPPLIED)
MUST BE
USED
Page 19
OCF 5.0 Open Channel Flow Monitor
SENSOR MOUNTING/LOCATION
- Open Channel Flow Applications
Each sensor is equipped with a 3/4 inch isolation coupling which MUST be used in your installation. A
threaded nipple or length of conduit may be used to position the sensor at the desired height. The sensor
should be hand-tightened by turning the sensor stem only. DO NOT use a wrench and do not over
tighten.
IMPORTANT: Follow the flume manufacturer’s directions for transducer location. The transducer
should be centered above the flume approach section and mounted 8"/203 mm (minimum) above the
maximum liquid level (depending on sensor model).
The transducer should be protected from physical damage and the transducer cable should be routed in a
separate metal conduit.
Because the sensor is equipped with a temperature sensor, it should be shielded from direct sunlight.
Use the PVC “isolation coupling” supplied with the sensor and hand-tighten the sensor/coupling
assembly onto your mounting stand. Do not clamp directly to the sensor or to the isolation coupling.
12"
305
305 mm
12"
Page 20
OCF 5.0 Open Channel Flow Monitor
TYPICAL TRANSDUCER POSITIONING
FOR FLUMES AND WEIRS
Always refer to the flume or weir manufacturer’s instructions for correct measurement point upstream
from the flume or weir. Location of the sensor is critical for accurate flow measurement.
If manufacturer’s instructions are not available, the following guidelines are generally accepted.
1. PARSHALL FLUME:
Position the sensor at 2/3 Approach as
illustrated above. Sensor height must be
8" (203 mm) or more above the highest
water level.
2. PALMER BOWLUS FLUMES:
Position the sensor at ½ the flume
Diameter upstream from the throat of
the flume. Sensor height must be 8" (203
mm) or more above the highest water
D
level.
3. V-NOTCH WEIRS
3x MAX. HEAD
Position the sensor at 3 x maximum Head
upstream from the weir plate. Sensor height must
be 8" (20.3 cm) or more above the highest water
level.
SENSOR
POSITION
FLOW
WEIR
PLATE
STILLING WELLS
Stilling wells are recommended to reduce the effects of turbulence as water flows through the flume or weir. The
OCF 5.0 sensor is centered over the stilling well. Sensor height must be 8" (203 mm) or more above the highest
water level. The well must be kept clean of sediment and deposits on the side walls.
Page 21
OCF 5.0 Open Channel Flow Monitor
ZERO POSITIONING OF SENSOR
- Open Channel Flow Applications
Locate the sensor at the position upstream from the throat of the flume or weir plate as recommended by
the manufacturer. A technique for accurate sensor height adjustment is shown:
Page 22
OCF 5.0 Open Channel Flow Monitor
ENCLOSURE INSTALLATION
Locate the enclosure within 20 ft (6 m) of the sensor (500 ft -150 m optional). The enclosure can be wall
mounted with the four mounting screws (included) or panel mounted with Option PM Panel Mount kit
from Greyline Instruments.
Avoid mounting the enclosure in direct sunlight to protect the electronics from damage due to
overheating and condensate. In high humidity atmospheres, or where temperatures fall below freezing,
Option TH Enclosure Heater and Thermostat is recommended. Seal conduit entries to prevent moisture
from entering enclosure.
NEMA4X (IP66) WITH CLEAR COVER
1. Open hinged enclosure cover.
COVER
2. Insert #8 screws (supplied) through the four enclosure mounting holes
to secure the enclosure to the wall or mounting stand.
Additional conduit holes can be cut in the bottom of the enclosure when
ENCLOSURE
MOUNTING
HOLES
required. Use a hole saw or Greenlee-type hole cutter to cut the required
holes.
DO NOT make conduit/wiring entries into the top of the enclosure.
ENCLOSURE
Note: This non-metallic enclosure does not automatically provide
END VIEW
grounding between conduit connections. Grounding must be provided as part of the installation. Ground
in accordance with the requirements of the National Electrical Code. System grounding is provided by
connecting grounding wires from all conduit entries to the steel mounting plate or another point which
provides continuity.
CLEANING
Cleaning is not required as a part of normal maintenance.
Page 23
OCF 5.0 Open Channel Flow Monitor
PZxx Series Sensors
Troubleshooting
Resistance measured (between the shield and center wire) across the coaxial cable ends by mulitmeter
indicates ambient temperature.
Resistance vs. Temperature
Page 24
FUSE REPLACEMENT
1. Turn OFF power.
2. Loosen cover screw and open.
3. Remove power module.
4. Locate fuse on Power Board.
5. Replace fuse with 2 AMP/ 250V, 5 x 20mm fuse.
6. Reinstall power module into chassis.
OCF 5.0 Open Channel Flow Monitor
Fuse
POWER MODULE
Page 25
OCF 5.0 Open Channel Flow Monitor
APPLICATIONS HOTLINE
For applications assistance, advice or information on any Greyline Instrument contact your Sales
Representative, write to Greyline or phone the Applications Hotline below:
United States: Tel: 315-788-9500 Fax: 315-764-0419
Canada: Tel: 613-938-8956 Fax: 613-938-4857
Toll Free: 888-473-9546
Email: info@greyline.com
Web Site: www.greyline.com
Greyline Instruments Inc.
Canada USA:
16456 Sixsmith Drive 105 Water Street Long Sault, Ont. K0C 1P0 Massena, NY 13662
Page 26
OCF 5.0 Open Channel Flow Monitor
PRODUCT RETURN PROCEDURE
Instruments may be returned to Greyline for service or warranty repair.
1) Obtain an RMA Number from Greyline -
Before shipping a product to the factory please contact Greyline by telephone, fax or email to obtain an
RMA number (Returned Merchandise Authorization). This ensures fast service and correct billing or
credit.
When you contact Greyline please have the following information available:
1. Model number / Software Version
2. Serial number
3. Date of Purchase
4. Reason for return (description of fault or modification required)
5. Your name, company name, address and phone number
2) Clean the Sensor/Product -
Important: unclean products will not be serviced and will be returned to the sender at their expense.
1. Rinse sensor and cable to remove debris.
2. If the sensor has been exposed to sewage, immerse both sensor and cable in a solution of 1 part
household bleach (Javex, Clorox etc.) to 20 parts water for 5 minutes. Important: do not immerse open
end of sensor cable.
3. Dry with paper towels and pack sensor and cable in a sealed plastic bag.
4. Wipe the outside of the enclosure to remove dirt or deposits.
5. Return to Greyline for service.
3) Ship to Greyline -
After obtaining an RMA number please ship the product to the appropriate address below:
Canadian and International USA
Customers: Customers:
Greyline Instruments Inc. Greyline Instruments Inc.
16456 Sixsmith Drive 204 150th Avenue
Long Sault, Ont. K0C 1P0 Madeira Beach, FL 33708
RMA# RMA#
Page 27
LIMITED WARRANTY
_____________________________________
Greyline Instruments warrants, to the original purchaser, its
products to be free from defects in material and workmanship for a
period of one year from date of invoice. Greyline will replace or
repair, free of charge, any Greyline product if it has been proven to
be defective within the warranty period. This warranty does not
cover any expenses incurred in the removal and re-installation of
the product.
OCF 5.0 Open Channel Flow Monitor
If a product manufactured by Greyline should prove defective
within the first year, return it freight prepaid to Greyline
Instruments along with a copy of your invoice.
This warranty does not cover damages due to improper installation
or handling, acts of nature, or unauthorized service. Modifications
to or tampering with any part shall void this warranty. This
warranty does not cover any equipment used in connection with the
product or consequential damages due to a defect in the product.
All implied warranties are limited to the duration of this warranty.
This is the complete warranty by Greyline and no other warranty is
valid against Greyline. Some states do not allow limitations on how
long an implied warranty lasts or limitation of incidental or
consequential damages, so the above limitations or exclusions may
not apply to you.
This warranty gives you specific legal rights, and you may also
have other rights which vary from state to state.
Greyline Instruments Inc.
Page 28
OCF 5.0 Open Channel Flow Monitor
APPENDIX A - OPTIONS
EXTRA SENSOR CABLE
(OPTION XC)
Each Greyline OCF 5.0 includes 25 ft. (7.6m) RG62AU coaxial cable. Additional RG62AU coaxial
cable and Cable Junction Box (Option JB) may be ordered with the Flow Monitor, or the cable may be
spliced and extended up to 500 ft (152m) as required during installation. No adjustment is required when
the sensor cable is extended or shortened. Use only RG62AU (or RG62U) coaxial cable which is
available from Greyline Instruments or your local distributor. Nominal impedance of RG62AU cable is
93 ohms.
Extended sensor cable must be installed in metal conduit to prevent interference. Do not use BNC
coaxial connectors (TV cable type). Recommended installation with a metal junction box is illustrated
below:
SENSOR CABLE
25 ft (7.6 m)
RG62AU COAXIAL
Note: Optional Watertight steel NEMA4 Junction Boxes with terminal strips (Option JB) are available
from Greyline Instruments.
Page 29
SENSOR INTRINSIC SAFETY
(OPTION ISB)
When connected through Intrinsic Safety
Barriers, Greyline PZ** Series sensors are
certified for installation in a hazardous
location rated:
Class I, Groups C,D
Class II, Groups E,F,G
Class III
The Intrinsic Safety Barrier may be ordered
with the Greyline instrument and is supplied
mounted in the Greyline instrument
enclosure. Replacement barrier fuses (Part
No. ISB- 011239) may be purchased
separately. The instrument enclosure
containing the ISB Intrinsic Safety Barrier
must be installed in a non-hazardous
location.
OCF 5.0 Open Channel Flow Monitor
DRAWING NO: GN3SPEC-INST-03
1
2
-093-390-101
(9.6V, 27ohms)
STAHL 9001/02
3
4
SLT 5.0/OCF 5.0 / PSL 5.0
NOTE:
BARRIER-EQUIPPED UNITS ARE
FACTORY-WIRED WITH GROUND
THROUGH THE INSTRUMENT CHASSIS.
NON-HAZARDOUS LOCATION
HAZARDOUS LOCATION
CLASS I, GROUPS C,D
CLASS II, GROUPS E,F,G
CLASS III
SENSOR
Page 30
OCF 5.0 Open Channel Flow Monitor
ENCLOSURE HEATER AND THERMOSTAT - Option TH
Instruments can be factory-equipped with an Enclosure Heater and Thermostat or the module can be
customer-installed. The Thermostat is factory set to turn ON at 40°F (4.5°C) and OFF at 60°F (15.5°C).
Power consumption is 15 Watts.
TO AC POWER SUPPLY
ENCLOSURE SUNSCREEN - Option SCR
Do not mount instrument electronics in direct sunlight. Overheating will reduce the life of electronic
components and condensate may form during the heat/cool cycles and cause electrical shorts.
11" / 280 mm
Note:
Exposure to direct sunlight can cause
overheating and moisture condensation
which will reduce the operating life of
electronics.
Protect Instruments from direct sunlight
with this iridite finished aluminum sun
screen (Greyline Option SCR).
Seal conduit entries with caulking
compound to further reduce moisture
condensation.
11"
280 mm
5"
127 mm
Page 31
OCF 5.0 Open Channel Flow Monitor
POWER INPUT OPTION
9-32VDC
OCF 5.0 Level & Flow Monitors may be ordered factory-configured for 9-32VDC power input.
QUICK BENCH TEST:
Connect Sensor as shown below, then Power. When properly connected figures will show on the large
LCD display. Test operation of the OCF 5.0 by holding the sensor steadily and aiming at a flat, stable
target 12 to 28" (305 to 711 mm) away from the end of the sensor. Allow a few seconds for the OCF 5.0
to lock onto the target before displaying its distance. The OCF 5.0 will now display Range in ft or cm
(factory calibration).
CONNECTIONS:
POWER INPUT: Connect 9-32VDC/0.5 Amps to the + and - terminals. The Power Input GND must be
connected to the nearest Ground pole. A 1 amp fuse in line is recommended.
SENSOR
DATA LOGGER
EXTRA RELAYS OPTION
NC
C
NC
NO
C
NC
NO
C
NC
NO
C
NO
RLY4
RLY5
RLY6
GND
4-20mA/0-5V
RLY2
RLY1
RLY3
DC
+–
–
+
NO
C
NC
NO
C
NC
AC
GROUND
USB A
Plug
Page 32
OCF 5.0 Open Channel Flow Monitor
CONVERSION GUIDE
FROM TO MULTIPLY BY
US GALLONS CUBIC FEET 0.1337
US GALLONS IMPERIAL GALS 0.8327
US GALLONS LITRES 3.785
US GALLONS CUBIC METERS 0.003785
LITRES/SEC GPM 15.85
LITRES CUBIC METERS 0.001
BARRELS US GALLONS 42
BARRELS IMPERIAL GALS 34.9726
BARRELS LITRES 158.9886
INCHES MM 25.4
DEGREES F DEGREES C (°F-32) x 0.556
POUNDS KILOGRAMS 0.453
PSI BAR 0.0676
FOOT² METER² 0.0929
VOLUME CALCULATION FOR ROUND TANKS: 3.142 x R² x H
R = TANK RADIUS (½ TANK DIAMETER)
H = TANK HEIGHT
Page 33
OCF 5.0 Open Channel Flow Monitor
SPECIFICATIONS
Accuracy: ±0.25% of Range or 2
mm, whichever is
greater Repeatability
and Linearity: 0.1% F.S.
Displays:White, backlit matrix -
displays flow rate,
totalizer, relay states,
operating mode and
calibration menu
Calibration:built-in 5-key calibrator
with English, French or
Spanish language
Power Input: 100-240VAC, 50/60Hz,
30 Watts or
9-32VDC, 9 Watts max
Output:Isolated 4-20mA (1000
proportional pulse
Enclosure: watertight, dust tight NEMA4X (IP 66) polycarbonate with a clear
shatter-proof face
Environmental Conditions: Relative humidity up to 80% -23 to 60°C ambient temperature,
maximum 5000 m altitude, pollution degree 4, Installation Category II.
Sensitivity:adjustable. Damping: adjustable
Electrical Surge Protection: Sensor, 4-20mA output and AC power input ApproximateShipping Weight:10 lbs (4.5 kg)
7.4 / 188 mm"
6.46 / 164 mm"
OCF 5.0
Open Channel
Flow Monitor
CONDUIT ENTRY
LOCATION
10.94" / 278 mm
5.12 / 130 mm"
10 / 254 mm"
SIDE VIEW
Standard Sensor PZ15
Maximum Range:15 ft (4.57 m)
Minimum Range (Deadband):8" (203.2 mm)
Operating Frequency: 92 KHz
Beam Angle: 8°
Operating Temperature:-40° to 150° (-40° to 65°C)
Temperature Compensation: Automatic, continuous
Max. Operating Pressure: 20 psi (1.35 bar) Sensor Face: PVC
Sensor Body: PVC
Mounting: ¾" NPT
Cable Length:25 ft. (7.6 m) continuous
RG62AU coaxial. Optional
50 ft. (15 m) continuous
Max. Cable Length:500 ft. (152 m) RG62AU
coaxial (splice)
Hazardous Rating:CSA rated Intrinsically Saf e
Class I, Groups C,D, Class
II, Groups E,F,G with
Optional Intrinsic Safety
Barrier.
Page 34
~4.25"
(108 mm)
OVERALL
3.125"
79.4 mm
25 ft (7.6 m) RG62AU
COAXIAL CABLE
(50 ft 15 m OR
100 ft 30 m)OPTIONAL
3/4“ NPT
3/4" NPT
2.1" (53 mm)
ISOLATION
COUPLING
(SUPPLIED)
1.125"
(28.6 mm)
OCF 5.0 Open Channel Flow Monitor
Optional Sensor PZ32T
Maximum Range:32 ft. (10 m)
Deadband (blanking): Programmable, minimum 12” (305
mm)
Beam Angle:8° at 3 DB
Temperature Compensation: Automatic, continuous
Operating Frequency: 42 KHz Exposed Materials:PVC and Teflon
Operating Temperature:- 40° to 150°F (-40° to 65°C)
Operating Pressure:20 psi (1.35 Bar) maximum
Mounting:¾” NPT (PVC isolation coupling
supplied)
Sensor Cable:RG62AU coaxial, 25 ft. (7.6 m)
standard
Hazardous Rating:with optional Intrinsic Safety Barrier:
CSA, Class I,II,III, Div. I,II, Groups
C,D,E,F,G
Note: Max Range reduced to 25 ft
(7.6 m) with ISB option.
3/4"
NPT
ISOLATION
COUPLING
(SUPPLIED)
3/4"
NPT
SIDE
VIEW
END
VIEW
PVC
PVC
1-3/4 "
44.5mm
L
TEFLON
O
TEF
25' (7.6 m) RG62AU
COAXIAL CABLE
3-7/8 "
98mm
5"
(127 mm)
OVERALL
Optional Sensor PZ32TE
Maximum Range:32 ft. (10 m)
Deadband (blanking): Programmable, minimum 12” (305
mm)
Beam Angle:8° at 3 DB
Temperature Compensation: Automatic, continuous
Operating Frequency: 42 KHz Exposed Materials: Teflon Operating Temperature:-40° to 170°F (-40° to 76°C)
Operating Pressure:20 psi (1.35 Bar) maximum
Mounting:¾” NPT (Teflon isolation coupling
supplied)
Sensor Cable:RG62AU coaxial, 25 ft. (7.6 m)
standard
Hazardous Rating:with optional Intrinsic Safety Barrier:
CSA, Class I,II,III, Div. I,II, Groups
C,D,E,F,G,
Note: Max Range reduced to 25 ft
(7.6 m) with ISB option.
3/4"
NPT
TEFLON
ISOLATION
COUPLING
(SUPPLIED)
3/4"
NPT
SIDE
VIEW
END
VIEW
TEFLON
TEFLON
1-3/4 "
44.5mm
L
TEFLON
O
TEF
25' (7.6 m) RG62AU
COAXIAL CABLE
3-7/8 "
98mm
5"
(127 mm)
OVERALL
Page 35
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.