GREISINGER GDH 01 AN, GDH 07 AN, GDH 13 AN, GDH 14 AN, GDH 12 AN Operating Manual

H18.0.0x.6C-01 page 1 of 3
Operating Manual for
GDH __ AN
Specification: GDH 01 AN GDH 07 AN GDH 12 AN GDH 13 AN GDH 14 AN
manometer manometer vakuum-/barometer manometer manometer
Measuring range: 0 ... 1999 Pa rel. 0.0 ... 199.9 mbar rel. 0 ... 1300 mbar abs. 0 ... 1999 mbar rel. 0.00 ... 10.00 bar rel.
(0.00 ... 19.99 mbar rel.)
or new calibration of sensor)
Resolution: 1 Pa (0.01 mbar) 0.1 mbar 1 mbar 1 mbar 0.01 bar
Accuracy (device): 1 Pa ±1 digit 0.1 mbar ±1 digit 1 mbar ±1 digit 1 mbar ±1 digit 0.01 bar ±1 digit
(at nominal temperature 25°C)
Drift (device): 0.01 % / K 0.01 % / K 0.01 % / K 0.01 % / K 0.01 % / K
Sensor: (relative pressure) piezoresistive relative pressure sensor, externally mounted in plastic case, 2 connection terminals
for plastic tube 6 x 1 mm (4 mm i.d.), approx. 1 m of 4-wire PVC connecting cable with 4-pin mini DIN plug.
Sensor: (absolute pressure) piezoresistive absolute pressure sensor, externally mounted in plastic case, connection terminal for
plastic tube 6 x 1 mm (4 mm i.d.), approx. 1 m of 4-wire PVC connecting cable with 4-pin mini DIN plug.
Application area: Sensor suitable for non-corrosive and non-ionising gases and liquids (do not use for water - air
cushion or hydrophobic filter)
Sensor accuracy:
(typ.)
hysteresis and linearity: ± 0.5% full scale ± 0.2% full scale ± 0.2% full scale ± 0.2% full scale ± 0.2% full scale temperature influence 0 - 50°C: ± 0.4% full scale ± 0.4% full scale ± 0.4% full scale ± 0.4% full scale ± 0.4% full scale
with option „double accuracy”: ±0.1%f.s. / ±0.2% f.s. ±0.1% f.s. / ±0.2% f.s. ±0.1% f.s. / ±0.2% f.s. ±0.1% f.s. / ±0.2% f.s.
Nominal temperature: 25°C
Operating temperature: device: 0 to 50°C
sensor: -40 to 85°C (temperature in sensor is compensated from 0 to 70° C)
Relative humidity: 0 to 80 %RH (non-condensing)
Storage temperature: -10 to 70°C
Display: 3½ digit LCD-display , approx. 13 mm high
Analog output:0 - 1 V DC equiv. 0 to 2000 digit
connection over 3.5 mm dia jack connector (plug included in scope of supply)
Power supply: 9V battery, type JEC 6F22 (in scope of supply)
Power consumption: approx. 5 mA
Mains connection: 2.5 mm socket for external 10 - 12V supply voltage. (suitable power pack: GNG10)
Battery will be automatically disconnected as soon as power pack is connected.
Low battery warning: "BAT" displayed automatically in case of low battery
Dimensions (device): approx. 150 x 86 x 30 mm (H x W x D), impact resistant ABS plastic housing
with integrated pop-up clip for table-top or suspended use.
Dimensions (sensor): approx. 67.5 x 26 x 15 mm (H x W x D), with suspension eye
Weight: approx. 340g (device ready for operation)
EMC: The device corresponds to the essential protection ratings established in the Regulations of the
Council for the Approximation of Legislation for the member countries regarding electromagnetic compatibility (89/336/EWG). Additional error: <1%
GREISINGER electronic GmbH
H18.0.0x.6C-01 page 2 of 3
Safety Requirements:
This device has been designed and tested in accordance with the safety regulations for electronic devices.
However, its trouble-free operation and reliability cannot be guaranteed unless the standard safety measures and special safety advises given in this manual will be adhered to when using the device.
1. Mains operation:
When using a power supply device please note that operating voltage has to be 9 to 12 V DC. Do not apply overvoltage!! Cheap 12V-power supply devices often have excessive no-load voltage. We, therefore, re­commend using regulated voltage power supply devices. Trouble-free operation is guaranteed by our power supply GNG10.
Prior to connecting the power supply to the mains make sure that the operating voltage stated at the power supply is identical to the mains voltage.
2. If device is to be connected to other devices (e.g. via serial interface) the circuitry has to be designed most carefully. Internal connection in third party devices (e.g. connection GND and earth) may result in not-permissible voltages impairing or destroying the device or another device connected. Warning: If device is operated with a defective mains power supply (short circuit from mains voltage to output voltage) this may result in hazardous voltages at the device (e.g. sensor socket at interface)
3. Trouble-free operation and reliability of the device can only be guaranteed if the device is not subjected to any other climatic conditions than those stated under "Specification". If the device is transported from a cold to a warm environment condensation may cause in a failure of the function. In such a case make sure the device temperature has adjusted to the ambient temperature before trying a new start-up.
4. If there is a risk whatsoever involved in running it, the device has to be switched off immediately and to be marked accordingly to avoid re-starting.
Operator safety may be a risk if:
there is visible damage to the device
the device is not working as specified
the device has been stored under unsuitable conditions for a longer time
In case of doubt, please return device to manufacturer for repair or maintenance.
5. Warning: Do not use this product as safety or emergency stop devices, or in any other application where failure of the product could result in personal injury or material damage.
Failure to comply with these instructions could result in death or serious injury and material damage.
Disposal instructions:
The device must not be disposed in the regular domestic waste. Send the device directly to us (sufficiently stamped), if it should be disposed. We will dispose the device appropriate and environmentally sound.
Loading...