GREISINGER G 1610 Operating Manual

Operating manual
G 1610
Dissolved oxygen measuring device
Permanently connected oxygen sensor
Waterproof

EN
Table of contents
2 / 30 B-H86.0.21.DB2-1.0
Table of contents
1 About this documentation ................................................................................................................. 4
1.1 Foreword............................................................................................................................................... 4
1.2 Purpose of the document...................................................................................................................... 4
1.3 Legal notices......................................................................................................................................... 4
1.4 Correctness of content.......................................................................................................................... 4
1.5 Layout of this document........................................................................................................................ 4
1.6 Further information ............................................................................................................................... 5
2 Safety ................................................................................................................................................... 6
2.1 Explanation of safety symbols .............................................................................................................. 6
2.2 Foreseeable misuse ............................................................................................................................. 6
2.3 Safety instructions ................................................................................................................................ 7
2.4 Intended use ......................................................................................................................................... 8
2.5 Qualified personnel............................................................................................................................... 8
3 Description .......................................................................................................................................... 9
3.1 Scope of delivery .................................................................................................................................. 9
3.2 Job description...................................................................................................................................... 9
4 The product at a glance ................................................................................................................... 10
4.1 The G 1610......................................................................................................................................... 10
4.2 Display elements ................................................................................................................................ 10
4.3 Operating elements ............................................................................................................................ 10
5 Bases for measurement ................................................................................................................... 12
5.1 Oxygen sensor.................................................................................................................................... 12
5.1.1 Explanation ......................................................................................................................................... 12
5.1.2 Design................................................................................................................................................. 13
5.1.3 Service life .......................................................................................................................................... 13
5.1.4 Operating position............................................................................................................................... 14
5.1.5 Measurement accuracy ...................................................................................................................... 14
5.1.6 Residue............................................................................................................................................... 14
5.2 Instructions for oxygen measurement................................................................................................. 14
5.2.1 Salinity correction ............................................................................................................................... 14
5.2.2 Environmental pressure, water depth and air pressure ...................................................................... 15
5.3 Commissioning, filling and maintenance of the sensor....................................................................... 15
6 Maintenance ...................................................................................................................................... 17
6.1 Operating and maintenance notices ................................................................................................... 17
6.2 Battery ................................................................................................................................................ 17
6.2.1 Battery indicator.................................................................................................................................. 17
6.2.2 Changing battery ................................................................................................................................ 17
6.3 Calibration and adjustment ................................................................................................................. 18
6.3.1 Automatic calibration in the air............................................................................................................ 18
6.3.2 Battery indicator.................................................................................................................................. 19
7 Operation........................................................................................................................................... 20
7.1 Commissioning ................................................................................................................................... 20
7.1.1 Explanation ......................................................................................................................................... 20
7.2 Configuration ...................................................................................................................................... 20
Table of contents
B-H86.0.21.DB2-1.0 3 / 30
7.2.1 Explanation ......................................................................................................................................... 20
7.2.2 Opening the configuration menu......................................................................................................... 20
7.2.3 Configuring parameters of the configuration menu............................................................................. 21
7.2.4 Adjustment of the measuring input ..................................................................................................... 22
7.2.5 Configuring parameters of the adjustment menu................................................................................ 23
8 Error and system messages............................................................................................................ 25
9 Disposal............................................................................................................................................. 26
10 Technical data................................................................................................................................... 27
11 Spare parts and accessories ........................................................................................................... 28
12 Service ............................................................................................................................................... 29
12.1 Manufacturer....................................................................................................................................... 29
12.2 Repairs ............................................................................................................................................... 29
12.3 Sales subsidiaries............................................................................................................................... 30
1 | About this documentation Dissolved oxygen measuring device
4 / 30 B-H86.0.21.DB2-1.0
1 About this documentation
1.1 Foreword
Read this document carefully and familiarise yourself with the operation of the product before you use it. Keep this document ready to hand and in the immediate vicinity of the product so that it is available to the personnel/user for reference at all times in case of doubt.
The product was developed according to the state of the art and fulfils the require­ments of the applicable European and national Directives. All corresponding docu­ments are available from the manufacturer.
Only technically qualified persons are permitted to carry out commissioning, operation, maintenance and decommissioning. The qualified personnel must have carefully read and understood the operating manual before beginning any work.
1.2 Purpose of the document
– This document describes the operation and maintenance of the product.
– Provides important information for working safely and efficiently with the product.
– In addition to the quick reference guide with all relevant legal and safety content in
hard copy, this document is a detailed reference option for the product.
1.3 Legal notices
The liability and warranty of the manufacturer for damages and consequential dam­ages are voided with misuse, disregarding this operating manual, disregarding safety notices, assignment of inadequately qualified technical personnel and arbitrary modi­fications of the product.
Only carry out the maintenance and service tasks on this product that are described in this documentation. In the process, adhere to the specified steps. For your own safety, only use original spare parts and accessories of the manufacturer. We assume no li­ability for the use of other products and resulting damage.
This document is entrusted to the recipient for personal use only. Any impermissible transfer, duplication, translation into other languages or excerpts from this operating manual are prohibited.
The manufacturer assumes no liability for print errors.
1.4 Correctness of content
The contents of this document were checked for corrected and are subject to a con­tinuous correction and updating process. This does not rule out potential errors. In the event that errors are discovered or in case of suggestions for improvement, please in­form us immediately via the indicated contact information in order to help us make this document even more user-friendly.
1.5 Layout of this document
Description
Each chapter is explained at the beginning in the description.
Dissolved oxygen measuring device About this documentation | 1
B-H86.0.21.DB2-1.0 5 / 30
Prerequisite
All mandatory prerequisites are then listed for each step.
Instruction
Tasks to be carried out by the personnel / user are represented as numbered instruc­tions. Adhere to the sequence of the specified instructions.
Representation
Shows an illustrative instruction or a configuration of the product.
Formula
Some instructions include a formula for a general understanding of a configuration, programming or a setting of the product.
Outcome of an action
Result, consequence or effect of an instruction.
Emphases
In order to simplify legibility and provide a clearer overview, various sections / informa­tion are emphasised.
1234 Display elements
Mechanical controls
Product functions
Product labels
Cross-reference [}p.4]
Foot notes
1.6 Further information
Software version of the product:
– V1.2 or later
– Version V0.9 or later
For the exact product name, refer to the type plate on the rear side of the product.
NOTE
For information about the software version, press and hold the ON button to switch on the product for longer than 5 seconds. The series is shown in the main display and the software version of the product is shown in the secondary display.
2 | Safety Dissolved oxygen measuring device
6 / 30 B-H86.0.21.DB2-1.0
2 Safety
2.1 Explanation of safety symbols
DANGER
This symbol warns of imminent danger which can result in death, severe bodily injury, or severe property damage in case of non-observance.
DANGER
This symbol indicates danger for living tissue as well as a variety of materials, which can be damaged or destroyed when coming into contact with this chemical. Caustic effect, protective equipment required!
CAUTION
This symbol warns of potential dangers or harmful situations which can cause damage to the device or to the environment in case of non-observance.
NOTE
This symbol indicates processes which can have a direct influence on operation or can trigger an unforeseen reaction in case of non-observance.
NOTE
This symbol instructs the use of eye protection which protects the eyes from harmful influences when working with powerful light, UV radiation, laser, chemicals, dust, splinters or weather influences.
NOTE
This symbol instructs the use of protective gloves which offer protection from mechan­ical, thermal, chemical, biological or electrical hazards.
2.2 Foreseeable misuse
The fault-free function and operational safety of the product can only be guaranteed if generally applicable safety precautions and the device-specific safety instructions for this document are observed.
If these notices are disregarded, personal injury or death, as well as property damage can occur.
Dissolved oxygen measuring device Safety | 2
B-H86.0.21.DB2-1.0 7 / 30
DANGER
Incorrect area of application!
In order to prevent erratic behaviour of the product, personal injury or property dam­age, the product must be used exclusively as described in the chapter Description [}p.9] in the operating manual.
– Do not use in safety / Emergency Stop devices!
– The product is not suitable for use in explosion-prone areas!
– The product must not be used for diagnostic or other medical purposes on pa-
tients!
– Not suitable for SIL!
2.3 Safety instructions
This product has been designed and tested according to the safety requirements for electronic measuring devices.
DANGER
Potassium hydroxide!
The electrode contains potassium hydroxide. This causes burns. All contact with the skin, clothing and eyes should be avoided. Nevertheless, should contact occur, take the following measures.
– Eyes: Flush with flowing water for at least 15 minutes, seek medical attention!
– Skin: Wash with large amounts of water for several minutes!
– Clothing: Remove immediately!
– If swallowed: Drink large amounts of water, do not induce vomiting and seek med-
ical attention!
CAUTION
Erratic behaviour!
On suspicion that the product can no longer be operated without danger, it must be decommissioned and prevented from recommissioning with appropriate labelling. The safety of the user can be impaired by the device if, for example, if it shows visible damage, it no longer works as specified or if it was stored for an extended period of time under unsuitable conditions.
– Visual inspection!
– In case of doubt, send the product to the manufacturer for repair or maintenance!
NOTE
If the product is stored at a temperature above 50 °C, or is not used for an extended period of time, the batteries must be removed. Leaks from the batteries are avoided as a result.
NOTE
This product does not belong in children's hands!
2 | Safety Dissolved oxygen measuring device
8 / 30 B-H86.0.21.DB2-1.0
2.4 Intended use
The product is designed for analysis of the oxygen concentration and/or oxygen satur­ation in freshwater and salt water. It is used, for example, for the monitoring of wells, sewer lines and aquaria. A minimum flow to the sensor of approx. 30 cm/sec is neces­sary for a correct measurement.
See Technical data [}p.27].
2.5 Qualified personnel
For commissioning, operation and maintenance, the relevant personnel must have ad­equate knowledge of the measuring process and use of the measurements, for which purpose this document makes a valuable contribution. The instructions in this docu­ment must be understood, observed and followed.
In order to ensure that no risks arise from the interpretation of the measurements in the concrete application, the user must have additional technical knowledge, because the user is liable in case of damage/danger due to misinterpretation as a result of in­adequate technical knowledge.
Dissolved oxygen measuring device Description | 3
B-H86.0.21.DB2-1.0 9 / 30
3 Description
3.1 Scope of delivery
Please check to ensure the completeness of the product after opening the package. You should find the following components:
– Quick reference guide
– Handheld measuring device, ready for operation, including batteries
– Permanently connected oxygen sensor
3.2 Job description
The product offers precision, speed and reliability in a compact, ergonomic housing. Additional impressive features include the dust-proof and waterproof design in accord­ance with IP 65/67 and the 3-line illuminated display, which offers overhead display at the push of a button. The product can be switched on, switched off and configured and the measurements and parameters can be adjusted and held with the operating ele­ments. The product with maintenance-friendly galvanised O2 sensor is an entry-level device suitable for everyday use. Concentrations in mg/l or ppm and saturation in per­centage can be read directly without using tables. Calibration with environmental air takes place at the push of a button. Use of a GSKA protective cap is recommended for field use in bodies of water in order to protect the membrane.
4 | The product at a glance Dissolved oxygen measuring device
10 / 30 B-H86.0.21.DB2-1.0
4 The product at a glance
4.1 The G 1610
G1610 GWO 5610
4.2 Display elements
Display
Battery indicator Evaluation of the battery status
Unit display Display of units, if applicable, with unstable symbol
or type of mode, min/max/hold
Main display Measurement of the current O2 value or value for
min/max/hold
Auxiliary display Corresponding temperature for the displayed O
2
value with unit.
Bar graph Progress for calibration and visualisation of the
electrode evaluation
NOTE
The unit display shows a rotating circle segment in the first position as long as the measurement is unstable, if the position is unoccupied by the unit display.
4.3 Operating elements
On / Off button
Press briefly Switch on the product
Activate / deactivate lighting
Long press Switch off the product
Reject changes in a menu
Up / Down button
Press briefly Display of the min/max value
Change value of the selected parameter
Long press Reset the min/max value of the current measure-
ment
Both simultaneously Rotate display, overhead display
Dissolved oxygen measuring device The product at a glance | 4
B-H86.0.21.DB2-1.0 11 / 30
Function key
Press briefly Freeze measurement
Return to measurement display
Call up next parameter
Long press, 2s
Start menu configuration, (ONF appears in the dis­play
Long press, 4s
Start automatic calibration, (AL appears in the dis­play
5 | Bases for measurement Dissolved oxygen measuring device
12 / 30 B-H86.0.21.DB2-1.0
5 Bases for measurement
5.1 Oxygen sensor
5.1.1 Explanation
The oxygen sensor is an active sensor. It consists of a platinum cathode, a lead anode and potassium hydroxide (KOH) as an electrolyte. If oxygen is present, it is reduced on the platinum cathode and the sensor delivers a signal. If no oxygen is present, no signal is delivered. The anode is consumed by the oxygen measurement. The sensor ages. Furthermore, the sensor loses water through the permeable membrane, in par­ticular, when it is stored in dry air. Therefore, it should be checked and maintained regularly and replaced as necessary.
DANGER
Potassium hydroxide!
The electrode contains potassium hydroxide. This causes burns. All contact with the skin, clothing and eyes should be avoided. Nevertheless, should contact occur, take the following measures.
– Eyes: Flush with flowing water for at least 15 minutes, seek medical attention!
– Skin: Wash with large amounts of water for several minutes!
– Clothing: Remove immediately!
– If swallowed: Drink large amounts of water, do not induce vomiting and seek med-
ical attention!
NOTE
Protective goggles must be worn for all of the following activities!
NOTE
Protective gloves must be worn for all of the following activities!
NOTE
Always store the oxygen sensor damp. It should always be stored in a storage bottle filled with water or in a container filled with water. After storage for an extended period, any potential deposit layers, such as algae, must be cleaned off of the membrane with a soft paper towel prior to measurement.
Dissolved oxygen measuring device Bases for measurement | 5
B-H86.0.21.DB2-1.0 13 / 30
5.1.2 Design
Storage bottle
Shaft
Membrane head
Refill opening
Platinum electrode
Platinum electrode
If oxygen is present, it is reduced on the platinum electrode and the sensor delivers a signal. Soiling on the platinum electrode or between the membrane and electrode can influence the measurement.
Storage bottle
The storage bottle is provided to keep the membrane moist. The service life of the sensor is extended as a result. Distilled or deionised water is in the storage bottle; do not add any other liquids!
Membrane head
The membrane head is covered with a thin plastic membrane. Faulty measurements will occur if the membrane is damaged or there are large air bubbles or even an air bubble ring on the membrane. This can also be the cause if a sensor can no longer be calibrated. The GWOK 02 membrane head is a spare part and can be re-ordered sep­arately.
Refill opening
Electrolyte must be filled or added for the initial commissioning of a sensor which is delivered dry, when performing maintenance or after use at high temperatures.
5.1.3 Service life
The sensor signal deteriorates relatively quickly at the end of the service life of the sensors. The electrode evaluation in %, therefore, can only be used as a guide value. A value of 70% does not mean that exactly 70% of the service life is still available, rather that the electrode signal has 70% of a comparison signal.
NOTE
The sensor evaluation is updated by the measuring device after a successfully per­formed calibration of the oxygen sensor.
The nominal service life can be reduced significantly due to use. Influential factors in­clude:
– Storage / operating temperature
– Contamination of the measured water
5 | Bases for measurement Dissolved oxygen measuring device
14 / 30 B-H86.0.21.DB2-1.0
– Mechanical stress of the sensor membrane
– Storage in dry air
– Continuous use in elevated carbon dioxide concentrations
5.1.4 Operating position
The oxygen sensor should be arranged vertically upwards with the connecting cable. A slight angle of inclination does not impair the measurement.
5.1.5 Measurement accuracy
The measurement accuracy can be impaired by:
– An inadequate flow below the necessary value of approx. 30cm/sec.
– The water temperature and sensor temperature must be the same. The most ac-
curate measurements are provided when the measuring temperature is calibrated.
5.1.6 Residue
Visible residue collecting as a reaction product in the interior of the membrane cap on the lead-oxide lead anode is red or brown from reaction with oxygen and lead carbon­ate and is white from reaction with carbon dioxide. These substances can collect on the membrane, but do not normally impair the measuring function and can be mostly removed in the course of maintenance of the sensor. Prior to screwing on the mem­brane cap, they should be removed as far as possible in order to prevent particles from being trapped between the membrane and the platinum cup. A rapid or excess­ive formation of lead carbonate after commissioning is an indication of air in the sensor. This is usually due to incomplete filling or a leak due to improper fitting of the cap / fill screw or membrane leak.
5.2 Instructions for oxygen measurement
The following must be observed when measuring dissolved oxygen:
– The storage bottle must be removed before the measurement.
– The sensor must have been calibrated.
– The sensor and the liquid to be measured must have the same temperature. The
temperatures of the two must match.
– The sensor must be immersed at least 3 cm into the liquid to be measured.
– A flow speed of at least approx. 30 cm/sec is necessary for exact measurements.
Either stir continuously or use an appropriate stirring device
– The measurement is sensitive to jarring! Therefore, make sure that the container
is not struck with the sensor while stirring, because this can significantly impair the measured value.
– The oxygen partial pressure, oxygen concentration in mg/l and oxygen saturation
in % are calculated from the sensor signal and temperature. In accordance with DIN38408-C22, the measurement is relative to water-vapour-saturated air.
5.2.1 Salinity correction
With increasing salinity SAL, which is the value for the salt content in the water, the sol­ubility of oxygen in water decreases, i.e. with the same oxygen partial pressure, fewer mg of oxygen are dissolved per litre of water. To determine this oxygen concentration,
therefore, the salinity of the medium must be entered first; see Configuring parameters of the configuration menu [}p.21]. A salinity correction is not necessary in freshwa-
Dissolved oxygen measuring device Bases for measurement | 5
B-H86.0.21.DB2-1.0 15 / 30
ter; the value is 0. A salinity of approx. 35 PSU is normal for sea water. The salinity correction is adjusted to aqueous media having a chemical composition corresponding to sea water. The International Oceanographic Tables (IOT) are used as a basis for the correction.
5.2.2 Environmental pressure, water depth and air pressure
The environmental pressure, water depth and air pressure play an important role at the place of measurement for the following points:
– Calculation of oxygen saturation in % SAT. Clean water can achieve 100 % satura-
tion in air. There must be no oxygen-depleting processes, such as biological de­caying processes, chemical effects or oxygen enhancing processes, such as ex­cessive ventilation or photosynthesis. This could result in an oversaturation above 100%.
– Calculation of the oxygen concentration in mg/l
– The valuation of calibration
Adjusting the pressure parameter on the product is recommended prior to calibration. In the scope of measuring accuracy, specifying the current air pressure in the region based on meteorological data or the standard pressure based on sea level is suffi­cient.
For example:
0 m above NN: 1013 hPa
300 m above NN: 978 hPa
600 m above NN: 943 hPa
1000m above NN: 899 hPa
5.3 Commissioning, filling and maintenance of the sensor
Description
The sensor is delivered dry. Therefore, the sensor is well-suited for storage. The sensor must be filled in good time before the measurement. A wait time of approx. 2 hours after filling should be planned in order to allow the sensor to stabilise. Commis­sioning with initial filling, filling and maintenance of the sensor are explained in the fol­lowing chapter.
Prerequisite
– Protective goggles
– Protective gloves
– A suitable screwdriver
– Pipette
– Household towel
– KOH electrolyte
– A spare GWOK 02 membrane head, if applicable
Instruction
1. Unscrew the membrane head.
2. Unscrew the seal screw from the refill opening.
NOTE
If the sensor was already filled and deposits have formed, it should be cleaned with KOH via the fill opening or removed. Loose residue can be rinsed out in the process. The platinum cup on the front of the membrane must be clean. Remove any soiling or electrolyte solution with a paper towel.
5 | Bases for measurement Dissolved oxygen measuring device
16 / 30 B-H86.0.21.DB2-1.0
3. Fill the pipette with the KOH electrolyte and initially fill the membrane head to ¾ full. Rinse off the excess electrolyte.
4. Slowly fill the sensor via the fill opening. In the process, tilt the sensor from side to side and tap the shaft to force out air bubbles. The sensor holds approximately 5 ml. If air bubbles no longer appear and the refill opening is filled up to the rim with KOH, screw the in the seal screw again.
5. Rinse off excess KOH and screw the sensor on with the membrane head. In the process, if there are any recognisable air bubbles under the membrane, KOH must be added.
6. After filling, the sensor should rest for 2 hours before a calibration is started.
Outcome of an action
The sensor has now been refilled. A sensor evaluation during the calibration should produce a result of 100%.
NOTE
Should it no longer be possible to calibrate the sensor or it only delivers unstable measurements, it must be maintained and/or the membrane head must be replaced
Dissolved oxygen measuring device Maintenance | 6
B-H86.0.21.DB2-1.0 17 / 30
6 Maintenance
6.1 Operating and maintenance notices
NOTE
If the product is stored at a temperature above 50 °C, or is not used for an extended period of time, the batteries must be removed. Leaks from the batteries are avoided as a result.
NOTE
The electrode should be stored in dry rooms at a temperature between 10 °C and 30 °C. If the storage temperature range is exceeded or undercut, the electrode can be destroyed. It should always be stored wet in distilled or deionised water.
6.2 Battery
6.2.1 Battery indicator
If the empty frame in the battery display blinks, the batteries are depleted and must be replaced. However, the device will still operate for a certain length of time.
If the BAT display text appears in the main display, the battery voltage is no longer ad­equate for operation of the product. Now the battery is fully depleted.
6.2.2 Changing battery
DANGER
Danger of explosion!
Using damaged or unsuitable batteries can generate heat, which can cause the batter­ies to crack and possibly explode!
– Only use high-quality and suitable alkaline batteries!
CAUTION
Damage!
If the batteries have different charge levels, leaks and thus damage to the product can occur.
– Use new, high-quality batteries!
– Do not use different types of batteries!
– Remove depleted batteries and dispose of them at a suitable collection point!
NOTE
Unnecessary screwing places the water-tightness of the product, among other things, at risk and should be avoided.
6 | Maintenance Dissolved oxygen measuring device
18 / 30 B-H86.0.21.DB2-1.0
NOTE
Read the following handling instructions before replacing batteries and follow them step by step. If disregarded, the product could be damaged or the protection from moisture could be diminished.
Description
Proceed as follows to replace the batteries.
Prerequisites
– The product is switched off.
Instruction
1. Unscrews the Phillips screws and remove the cover.
2. Carefully replace the two Mignon AA batteries. Ensure that the polarity is correct! It must be possible to insert the batteries in the correct position without using force.
3. The O-ring must be undamaged, clean and positioned at the intended depth. In or­der to facilitate assembly and avoid damage, a suitable grease can be applied.
4. Fit the cover on evenly. The O-ring must remain at the intended depth!
5. Tighten the Phillips screws.
Outcome of an action
The product is now ready for use again.
6.3 Calibration and adjustment
6.3.1 Automatic calibration in the air
Description
The following steps describe how to calibrate the product automatically.
Prerequisite
– The product is switched on.
NOTE
Calibration takes place in air saturated with water vapour. The GCAL 3610 calibration container or the storage bottle can be used for this purpose. The membrane of the electrode must be dry for the calibration. Prior to performing the calibration of the membrane, dab away water droplets from the membrane with a soft, dry cloth. The fol­lowing must be observed when using the storage bottle:
Only insert the electrode far enough into the storage bottle that the membrane does not come into contact with the water in the bottle.
Unscrew the lid of the storage bottle and only position it so that a small exchange of air and pressure equalisation can take place
Instruction
1. Place the electrode in the calibration container. If necessary, wait until the temper­ature has equalised and a stable value has been achieved.
2. Press the Function key for 4 seconds to open the Calibration menu. The display shows (AL.
3. Release the Function key.
4. The product determines the correct value automatically.
Outcome of an action
After successful completion of the calibration the assessment of the electrode condi­tion is displayed briefly in percent. An aged or contaminated electrode, incorrect ad­justment of the pressure, contamination of the platinum electrode or a damaged mem­brane can be the cause for a lower evaluation.
Dissolved oxygen measuring device Maintenance | 6
B-H86.0.21.DB2-1.0 19 / 30
If the calibration is not completed successfully an error message is displayed. (AL ERR. appears in the display See Error and system messages [}p.25]. Confirm the error
message pressing the Function key. The product restarts and the value of the last suc­cessful calibration is restored.
6.3.2 Battery indicator
If the empty frame in the battery display blinks, the batteries are depleted and must be replaced. However, the device will still operate for a certain length of time.
If the BAT display text appears in the main display, the battery voltage is no longer ad­equate for operation of the product. Now the battery is fully depleted.
7 | Operation Dissolved oxygen measuring device
20 / 30 B-H86.0.21.DB2-1.0
7 Operation
7.1 Commissioning
7.1.1 Explanation
Prerequisite
– Sufficiently full batteries are inserted in the product.
Instruction
– Press On/Off button.
Outcome of an action
Information about the configuration of the product appears in the display.
POFF
Automatic shut­off
Automatic shut-off activated. The product is switched off if no buttons have been pressed after the adjusted time
T.OF
Zero point correc­tion
If a zero point correction of the temperature sensor was made
T.SL
Gradient correc­tion
If a gradient correction of the temperature sensor was made
SAL
Salinity correction Links when the salinity correction is active
The product is now ready for measurement.
NOTE
The product must be calibrated to the electrode prior to starting the measurement. See Calibration and adjustment service.
7.2 Configuration
7.2.1 Explanation
The following steps describe how to adapt the product for your purposes.
NOTE
There are various configuration parameters available depending on the product ver­sion and configuration. They can differ depending on the product version and configur­ation.
7.2.2 Opening the configuration menu
Description
In order to configure the product, you must first open the Configuration menu. The menu is opened as shown in the illustration.
Prerequisite
– The product is switched on.
Instruction
1. Press the Function key for 2 seconds to open the Configuration menu.
2. (ONF appears in the display. Release the function key.
3. By briefly pressing the Function key, you can scroll through the parameters. Select the parameter you would like to configure.
4. When you have selected the desired parameter, change the parameter to the de­sired value with the Up button and the Down button.
Dissolved oxygen measuring device Operation | 7
B-H86.0.21.DB2-1.0 21 / 30
5. The changes are saved after running through the entire Configuration menu. STOR
appears in the display. The Configuration menu can be exited from any arbitrary parameter by pressing and holding the Function key for 2 seconds. The changes made up that point are saved.
Representation
Call up menu Next parameter Change value Save changes Discard
changes
2s Press: Single
step
Hold: Rapid change
2s 2s
Outcome of an action
The Configuration menu is closed after the last parameter.
NOTE
If the product is switched off without saving the configuration, the last save value is re­produced on the next start-up of the product.
7.2.3 Configuring parameters of the configuration menu
Description
The following representation shows the available parameters and various configura­tion options.
Prerequisite
– The Configuration menu is open. See Opening the configuration menu [}p.20].
Instruction
1. Select the desired parameter you would like to configure.
2. Adjust the desired configuration in the selected parameter with the Up button and Down button.
3. The available configuration options are listed for each parameter in the following representation.
Representation
Parameter Values Meaning
Input
INP
SAT %
Oxygen saturation in per cent
(ONC mg/l
Oxygen concentration in mg/l
(ONC ppm
Oxygen concentration in ppm
Pressure
SET.P
500 .. 4000
Environmental pressure in hPa, corresponding to mbar
Salinity correction
SAL
0 .. 70
Salinity in the measuring medium in PSU, corres­ponding to g/kg
7 | Operation Dissolved oxygen measuring device
22 / 30 B-H86.0.21.DB2-1.0
Shut-off time
POFF
OFF
No automatic shut-off
15 30 60 120 240
Automatic shut-off after a selected time in minutes, during which no buttons have been pressed
Backlighting
L,TE
OFF
Backlighting deactivated
15 30 60 120 240
Automatic shut-off of the backlighting after a selec­ted time in seconds, during which no buttons have been pressed
ON
No automatic shut-off of the backlighting
Temperature unit
UN,T
°(
Temperature display in °C
°F
Temperature display in °F
Factory settings
IN,T
NO
Use current configuration
YES Reset product to factory settings. IN,T DONE appears
in the display
Outcome of an action
The changed value is saved and the Configuration menu is closed. STOR appears in the display. If necessary, the product is restarted automatically in order to adopt the changed values.
NOTE
The configuration is closed if no button is pressed for 2 minutes. Any changes made up to that point are not saved. C.END appears in the display.
7.2.4 Adjustment of the measuring input
Description
The temperature input can be adjusted with the zero point correction and the gradient correction. If an adjustment is made, you change the pre-adjusted factory settings.
This is signalled with the T.OF or T.SL when the product is switched on. The standard settings of the zero point value and the gradient value is 0.00. It signals that no correc-
tion is made.
In order to adjust the product, you must first open the Adjustment menu. The menu is opened as shown in the illustration.
Prerequisites
– Sufficiently full batteries are inserted in the product.
– The product is switched off.
Instruction
1. Press and hold the Down button.
2. Press the On/Off button to switch on the product and open the Configuration menu. Release the Down button. The display shows the first parameter.
3. By briefly pressing the Function key, you can scroll through the parameters. Select the parameter you would like to configure.
Dissolved oxygen measuring device Operation | 7
B-H86.0.21.DB2-1.0 23 / 30
4. When you have selected the desired parameter, change the parameter to the de­sired value with the Up button and the Down button.
5. In order to save the new parameter value, press and hold the Function key for longer than 1 second.
Representation
Call up menu
Hold Release
Outcome of an action
The Configuration menu is closed after the last parameter.
NOTE
If the product is switched off without saving the configuration, the last save value is re­produced on the next start-up of the product.
7.2.5 Configuring parameters of the adjustment menu
Description
The following representation shows the available parameters and various configura­tion options.
Prerequisites
The Adjustment menu is open. See Adjustment of the measuring input [}p.22].
Instruction
1. Select the desired parameter you would like to configure.
2. Adjust the desired configuration in the selected parameter with the Up button and Down button.
3. The available configuration options are listed for each parameter in the following representation.
Representation
Parameter Values Meaning
Zero point correction
T.OF
0.00
No zero point correction
-5.00 .. 5.00
Zero point correction in °C. and/or at °F -9.00 ..
9.00
Gradient correction of the temperature
T.SL
0.00
No gradient correction
-5.00 .. 5.00
Gradient correction in %
Formula
Zero point correction: Displayed value = measured value – T.OF
Gradient correction °C: Display = (measured value – T.OF) * (1 + T.SL / 100)
Gradient correction °F: Display = (measured value – 32 °F – T.OF) * (1 + T.SL / 100) + 32 °F
Example calculation
– Zero point correction T.OF to 0.00 – Gradient correction T.SL to 0.00 – Display unit UNIT to °C
– Display in ice water -0.2 °C
7 | Operation Dissolved oxygen measuring device
24 / 30 B-H86.0.21.DB2-1.0
– Display in ice water setpoint T.OF = 0.0 °C
– Display in water bath 36.6 °C – Display in water bath setpoint T.SL = 37.0 °C – T.OF = display zero point correction – setpoint zero point – T.OF = -0.2 °C – 0.0 °C = -0.2 °C – T.SL = (setpoint gradient correction / (display gradient correction – T.OF) – 1) *100 – T.SL = (37.0 °C / (36.6 °C – (-0.2)) -1) *100 = 0.54
Outcome of an action
The changed value is saved and the Configuration menu is closed.
NOTE
If the product is switched off without saving the configuration, the last save value is re­produced on the next start-up of the product.
Dissolved oxygen measuring device Error and system messages | 8
B-H86.0.21.DB2-1.0 25 / 30
8 Error and system messages
Display Meaning Possible causes Remedy
No display, unclear char­acters or no response when but­tons are pressed
Battery depleted
System error
Product is defective
Battery depleted
Error in the product
Product is defective
Replace battery
Send in for repair
BAT
Battery depleted Battery depleted Replace battery
BAT LO
Battery depleted Battery depleted Replace battery
(AL ERR.2
Slope is too low
Incorrect oxygen ref­erence
Electrode contamin­ated or defective
Perform calibration in damp environmental air
Maintain the electrode
(AL ERR.3
Slope is too high
Incorrect oxygen ref­erence
Electrode contamin­ated or defective
Perform calibration in damp environmental air
Maintain the electrode
(AL ERR.4
Incorrect calibration temperature
Temperature too low or too high
Range of 5..40 °C
(AL ERR.5
Time exceeded dur­ing automatic calib­ration
Unstable electrode signal
Contaminated elec­trode
Temperature not equalised
Use calibration container
Maintain the electrode
Restart calibration
ERR.1
Measuring range ex­ceeded
Measurement too high
Electrode or product defect
Faulty calibration
The measurement is above the permissible range
Check electrode
Perform calibration
Send in for repair
ERR.2
Measuring range is undercut
Measurement too low
Electrode or product defect
Check electrode
Send in for repair
SYS ERR
System error Error in the product Switch product on/off
Replace batteries
Send in for repair
9 | Disposal Dissolved oxygen measuring device
26 / 30 B-H86.0.21.DB2-1.0
9 Disposal
Separation by material and recycling of device components and packaging must take place at the time of disposal. The valid legal regulations and directives applicable at the time must be observed.
NOTE
Fill in the return form available from the information base online at www.ghm-group.de and sent it in with the product.
NOTE
The device must not be disposed of with household waste. If the product is disposed of, please take it to a municipal collection point, where it will be transported to a dis­posal company in accordance with requirements of hazardous goods laws. Otherwise, return it to us, freight prepaid. We will then arrange for the proper and environment­ally-friendly disposal. Please dispose of empty batteries at the collection points inten­ded for this purpose.
Dissolved oxygen measuring device Technical data | 10
B-H86.0.21.DB2-1.0 27 / 30
10 Technical data
Measuring range O2 concentration O2 saturation Temperature
0.0 .. 20.0 mg/l
0.0 .. 20.0 ppm
0 .. 200 % 0 .. 50 °C
32 .. 122 °F
Accuracy (at nominal temper­ature)
±1.5 % of measured value
± 0.2 mg/l
±1.5 % of measured value
± 0.2 %
± 0.3 °C
Temperature compensation 0 .. 50 °C (or 32 .. 122 °F)
Nominal temperature 25°C
Measuring cycle
approx. 2 measurements per second
Connections Permanently connected oxygen sensor
Display 3-line segment LCD, additional symbols, illuminated (adjustable white, perman-
ent illumination)
Additional functions Min/Max/Hold
O2 calibration Automatic calibration in the air
Housing Break-proof ABS housing
Protection rat­ing
IP65 / IP67
Dimensions L*W*H [mm] and weight
108 * 54 * 28 mm without electrode
130 g, incl. battery, without electrode
190 g, incl. battery and electrode
Connections Permanently connected oxygen sensor
Operating conditions
Device -20 to 50 °C; 0 to 95 % r.h. (temporarily 100 % r.h.)
Electrode 0 .. 40 °C
Storage temperature 0 .. 40 °C
Current sup­ply
2*AA battery (included in the scope of delivery)
Current require­ment/
battery life
approx. 0.8 mA, approx. 2.7 mA with lighting
Service life > 3000 hours with alkaline batteries (without backlighting)
Battery indic­ator
4-stage battery status indicator,
Replacement indicator for depleted batteries: "BAT"
Auto-power-OFF function The device switches off automatically if this is activated
Directives and standards The devices conform to the following Directives of the Council for the harmon-
isation of legal regulations of the Member States:
2014/30/EU EMC Directive
2011/65/EU RoHS
Applied harmonised standards:
EN 61326-1:2013 Emission limits: Class B Immunity according to Table 2 Additional errors: < 0.5 % FS
EN 50581:2012
The device is intended for mobile use and/or stationary operation in the scope of the specified operating conditions without further limitations.
11 | Spare parts and accessories Dissolved oxygen measuring device
28 / 30 B-H86.0.21.DB2-1.0
11 Spare parts and accessories
A selection of spare parts and accessories for this product is listed below.
Article
Number Name Description
610049 Mignon batteryAAMignon AA spare battery
608012 GWOK 02 Spare membrane head
610049 GB AA AA battery
601414 GSKA 3600 PVC protective cap, submerged, for use in still bodies of
water
607267 GSKA 3610 Red brass protective cap, salt water resistant, also suit-
able for use with greater depths or with a flow
611371 GCAL 3610 Calibration bottle
611373 ST-G1000 Device protection bag with 1 round cut-out
603356 KOH 100 Spare electrolyte 100 ml
A complete list of all accessories and spare parts is available in our product catalogue or on our home page. We can also provide further information by phone.
Contact
Internet: www.greisinger.de
Tel: +49 94029383-52
Dissolved oxygen measuring device Service | 12
B-H86.0.21.DB2-1.0 29 / 30
12 Service
12.1 Manufacturer
If you have any questions, please do not hesitate to contact us:
Contact
GHM Messtechnik GmbH
GHM GROUP - Greisinger
Hans-Sachs-Str. 26
93128 Regenstauf | GERMANY
Phone: +49 94029383-52
info@greisinger.de | www.greisinger.de
WEEE reg. no. DE 93889386
12.2 Repairs
Defective products are repaired professionally and quickly in our service centre.
Open hours and contact
Monday to Thursday from 8:00 to 16:00
Friday from 8:00 to 13:00
GHM Messtechnik GmbH
Hans-Sachs-Str.26
Service Centre
93128 Regenstauf | GERMANY
Phone: +49 94029383-39
Fax: +49 94029383-33
service@greisinger.de
NOTE
Fill in the return form available from the information base online at www.ghm-group.de and sent it in with the product.
12 | Service Dissolved oxygen measuring device
30 / 30 B-H86.0.21.DB2-1.0
12.3 Sales subsidiaries
Austria
GHM Messtechnik GmbH
Office Austria
Breitenseer Str. 76/1/36
1140 Vienna | AUSTRIA
Phone +43 660 7335603
a.froestl@ghm-messtechnik.de
Brazil & Latin America
GHM Messtechnik do Brasil Ltda
Av. José de Souza Campos, 1073, cj 06
Campinas, SP
13025 320 | BRAZIL
Phone +55 19 3304 3408
Info@grupoghm.com.br
Czech Republic / Slovakia
GHM Greisinger s.r.o.
Ovci hajek 2 / 2153
158 00 Prague 5
Nove Butovice | CZECH REPUBLIC
Phone +420 251 613828
Fax +420 251 612607
info@greisinger.cz | www.greisinger.cz
Denmark
GHM Maaleteknik ApS
Maarslet Byvej 2
8320 Maarslet | DENMARK
Phone +45 646492- 00
Fax +45 646492- 01
info@ghm.dk | www.ghm.dk
France
GHM GROUP France SAS
Parc des Pivolles
9 Rue de Catalogne
69150 Décines-Charpieu (Lyon) | FRANCE
Phone +33 4 72 37 45 30
a.jouanilou@ghm-group.fr
India
GHM Messtechnik India Pvt Ltd.
209 | Udyog Bhavan | Sonowala Road
Gregaon ( E ) | Mumbai - 400 063
INDIA
Phone +91 22 40236235
info@ghmgroup.in | www.ghmgroup.in
Italy for Greisinger & Delta OHM
GHM GROUP – Delta OHM
Via Marconi 5
35030 Caselle di Selvazzano
Padova (PD) | ITALY
Phone +39 049 8977150
a.casati@ghm-messtechnik.de
Italy for Honsberg, Martens, Val.co
GHM GROUP – Val.co
Via Rovereto 9/11
20014 S. Ilario di Nerviano
Milano (MI) | ITALY
Phone +39 0331 53 59 20
alessandro.perego@valco.it
Netherlands
GHM Meettechniek BV
Zeeltweg 30
3755 KA Eemnes | NETHERLANDS
Phone +31 35 53805-40
Fax +31 35 53805-41
info@ghm-nl.com | www.ghm-nl.com
South Africa
GHM Messtechnik SA (Pty) Ltd
16 Olivier Street
Verwoerdpark, Alberton 1453
SOUTH AFRICA
Phone +27 74 4590040
j.grobler@ghm-sa.co.za
Loading...