Adjusting: via keypress by editing offset and scale
Display: approx. 10 mm high, 4-digit LC-display
Min-/Max-Value Memory:min and max measured values are stored
Ambient conditions for electronics:
Nominal temperature:25°C
Operating condition:-10 to 50°C5 to 95 %RH (non-condensing) 850 ... 1100 hPa
Storage condition:-25 to 60°C5 to 95 %RH (non-condensing)700 ... 1100 hPa
(supply voltage)
< 180 s
63
BUS-
protocol
BUS
-devices
Uv = 12 - 30 V DC, max. 600 mA
or refer to type plate
(time to reach specification values)
2
2
/ °C (at 0-50°C)
2
BUS
, polarity free
Housing:ABS (IP20)
Dimensions:82 x 80 x 55 mm (without elbow-type plug and tube)
Mounting:With holes for wall mounting (in housing - accessible after cover has been removed)).
Mounting distance:50 x 70mm, max. shaft diameter of mounting screws is 4 mm.
Electrical connection:elbow-type plug conforming to DIN 43650 (IP65),
max. wire cross section: 1.5 mm², wire/cable diameter from 4.5 to 7 mm
EMC:The devices corresponds to the essential protection ratings established in the
regulations of the council for the approximation of legislation for the member countries
regarding electromagnetic compatibility (89/336/EWG).
In accordance with EN61326 +A1 +(appendix A, class B), additional errors: < 1% FS
When connecting long leads adequate measures against voltage surges have to be
taken.
Operating advice
CO2 gas is outweigh than air (rel. density = 1.52). The suggested mounting level is near the ground floor.
This device has been designed and tested in accordance with the safety regulations for electronic devices.
However, its trouble-free operation and reliability cannot be guaranteed unless the standard safety measures and special
safety advises given in this manual will be adhered to when using the device.
1. Trouble-free operation and reliability of the device can only be guaranteed if the device is not subjected to any other
climatic conditions than those stated under "Specification". If the device is transported from a cold to a warm
environment condensation may cause in a failure of the function. In such a case make sure the device temperature
has adjusted to the ambient temperature before trying a new start-up.
2. General instructions and safety regulations for electric, light and heavy current plants, including domestic safety
regulations (e.g. VDE), have to be observed.
3. If device is to be connected to other devices (e.g. via PC) the circuitry has to be designed most carefully. Internal
connection in third party devices (e.g. connection GND and earth) may result in not-permissible voltages impairing or
destroying the device or another device connected.
4. If there is a risk whatsoever involved in running it, the device has to be switched off immediately and to be marked
accordingly to avoid re-starting.
Operator safety may be a risk if:
-there is visible damage to the device
-the device is not working as specified
-the device has been stored under unsuitable conditions for a longer time
In case of doubt, please return device to manufacturer for repair or maintenance.
5. Warning:
Do not use these products as safety or emergency stop devices, or in any other application where failure of the
product could result in personal injury or material damage. Failure to comply with these instructions could result in
death or serious injury and material damage.
Disposal instructions
The device must not be disposed in the regular domestic waste.
Send the device directly to us (sufficiently stamped), if it should be disposed. We will dispose the device appropriate and
environmentally sound.
To mount the connection cable (3-, or 4-wire depending
on type of device) the elbow-type plug screw has to be
loosened and the coupling insert has to be removed by
means of a screw driver at the position indicated (arrow).
Pull out connection cable through PG glanding and
connect to the loose coupling insert as described in the
wiring diagram. Replace loose coupling insert onto the
pins at the transducer housing and turn cover cap with PG
glanding in the direction desired till it snaps on (4 different
starting positions at 90° intervals).
Re-tighten the screw at the angle plug.
T38.0.1X.6C-02aUser’s Manual EBG-CO2page 3 of 4
1.3 Assignment of elbow-type plug:
4-wire-connection (power supply,
3
1
EASY
2
.
.
.
.
The type current or voltage output is set by works and cannot be changed.
1= EASYBUS
2= EASYBUS
3= supply voltage +Uv
(4) = supply voltage -Uv
BUS
Uv
EASY
BUS
)
2 Operation
2.1 Display functions
2.1.1 Currently measured values: During normal operation the CO2 display value is displayed in ppm.
2.1.2 Min/Max Value Memory
watch min values (Lo):press ‘down‘(2) shortly oncedisplay changes between ‘Lo‘ and Min values
watch max values (Hi):press ‘up‘(3) shortly oncedisplay changes between ‘Hi‘ and Max values
restore current values:press ‘down‘(2) or ’up’(3) once againcurrent values are displayed
clear min values: press ‘down’(2) for 2 secondsMin values are cleared. The display shows shortly ‘CLr‘.
clear max values: press ‘up’(3) for 2 secondsMax values are cleared. The display shows shortly ‘CLr‘.
After 10 seconds the currently measured values will be displayed again.
2.1.3 Min/Max Alarm
Whenever the measured value is exceeding or undershooting the alarm-values that have been set, the alarm-value and
the meas. value will be displayed alternating.
AL.Lothe lower alarm boundary is been reached or is undershot
AL.Hithe upper alarm boundary is been reached or is exceeded
2.2 Error and system messages
DisplayDescriptionPossible fault causeRemedy
Err.1
Err.7
Err.9
Er.11
8.8.8.8
measuring range exceededWrong signal
System faultError in device
Sensor errorSensor defective
Calculation not possible
Segment test
Calculation variable missing
or invalid
The transducer performs a display test for 8 seconds after power up. After
that it will change to the display of the measuring.
Take care for fresh air
Disconnect from supply and reconnect.
If error remains: return to manufacturer
Check temperature
T38.0.1X.6C-02aUser’s Manual EBG-CO2page 4 of 4
3 Configuration
In the configuration the devices parameters can be changed.
The jumper has to be set, p.r.t. figure right-hand side. To set or remove
jumper, the housing cover has to be removed. Ex works the jumper is set.
To change parameters press “SET” (key 1) four seconds, then the
parameter selection is started with the first parameter (display shows
“dA.Lo”).
By pressing “SET” the desired parameter is selected, the editing of the
parameter values happens via keys 5 (key 3) or 6 (key 2).
Pressing “SET” again (after editing the parameter values) returns to the
parameter selection.
Pressing “SET” again after the last parameter finishes the configuration,
stores the changes and the instrument returns to the normal mode.
3.1 Description of configuration parameter:
a) 'AL.Hi': Upper boundary of alarm
At Al.Hi the boundary is set from which max. alarm will be given.
Selectable range: ‚AL.Lo‘ ... 2001
If the jumper is removed from the shown
contacts, the configuration is inaccessible,
values are protected.
Jumper
b) 'AL.Lo': Lower boundary of alarm
At Al.Lo the boundary is set from which min. alarm will be given.
Selectable range: -1 ... ‚AL.Hi‘
c) 'A.dEL': Alarm delay
The value at A.dEL declares the alarm delay in minutes.
e) 'OFFS': Offset of CO2-measuring
The offset of the measuring will be shifted by this value, the input is in ppm.
Selectable range: -200...+200 ppm or 'oFF': offset is deactivated (=0, ex works)
(calculation: see by scale)
e) 'SCAL': Scale correction of CO2-measuring
The scale of the measuring is changed by this value.
Selectable range: -10,00...+10,00 or 'oFF': scale is deactivated (=0, ex works)
The adjusting by offset and scale is intended to be used to compensate deviations of the CO
It is recommended to keep the scale correction deactivated (“oFF”). The display value is given by following formula:
CO2-display = measured value – offset
With a scale correction the formula changes:
CO2 display = (meas. value - offset) * ( 1 + scale correction/100)
-measuring.