Greisinger ALSCHU 480 Instruction manual

S39.0.01.6C-03 page 1 of 4
Instruction manual
ALSCHU 480
1 Specification
Power supply: 220/240V 50/60Hz (controlling device) Power consumption: approx. 1 Watt Control output: by outlet in the grounding contact adapter plug Switching voltage: max. 250 VAC Switching current: max. 10 A AC (resistive load) Breaking capacity: max. 2400 VA Dimensions: Controlling device: 110 x 65 x 45 mm (L x W x H)
EMC: The device corresponds to the essential protection ratings established in the
Regulations of the Council for the Approximation of Legislation for the mem­ber countries regarding electromagnetic compatibility (89/336 EWG).
2 Disposal instructions
The device may not be disposed in the regular domestic waste. Send the device directly to us (sufficiently stamped), if it should be disposed. We will dispose the device appropriate and environmental friendly.
3 Operating advices and service notes
1. Disconnect the power plug before opening the housing or before working at closed devices!
The ALSCHU 480 does not disconnect a connected device at the adapter plug power outlet from the mains! Therefore it is absolute essential for working at a device which is connected to the power outlet, to stake it off and thus to disconnect it from the supply voltage.
2. The mains switch of the alarm device turns only the device itself off, the adapter plug power outlet will be not disconnected!
3. For a durable and correct function you have to check the function in certain time lags, like every safety device. For this the device has to be set to alarm by raising the sensors not less than once a month.
4. The device has to be treated carefully and has to be used corresponding to the preceding specifica­tions (do not throw, do not let it dash against something, etc.).
5. The device may only be used in dry rooms.
GREISINGER electronic GmbH
D - 93128 Regenstauf, Hans-Sachs-Straße 26
Phone: +49 9402 / 9383-0, Fax: +49 9402 / 9383-33, www.greisinger.de
S39.0.01.6C-03 page 2 of 4
4 Installation and initiation (may only be effected through adequate skilled personnel!)
1. Turn the power switch of the alarm device (ALSCHU 480) to the off-position and turn the selector to setting "II" (to switch the power outlet from the shockproof adapter plug to currentless).
2. Hang up the alarm device via clamp to the favoured place (possibly choose an inconspicuous, but easily reachable arrangement; but pay attention to enough sonic radiation of the built in alarm trans­mitter!). The device may only be used in dry rooms.
3. Put the sensor(s) (water sensor, floating switch etc.) to the place which shall be protected and fix it as necessary.
4. Attach the sensor to the Ø2.5mm jack plug of the alarm device. At the use of more than one sensor the sensors are meant to be connected together via two way connectors (GAZ-1) or extension cables (VEKA 2, 5, 10) and then they must be attached to the alarm device.
5. The device which has to be switched in the case of alarm (e.g. lamp, siren, pump – attend the maxi­mum acceptable breaking capacity! At inductive loads use a RC-element) has to be plugged in the power outlet of the shockproof adapter plug. Pay attention that the on-off switch (if existing) of the connected device is at the position „ON“ after the initiation, otherwise the control by the ALSCHU 480 is not possible.
6. Plug the grounding contact adapter plug of the alarm device in the power outlet.
7. Switch the mains switch of the alarm device to "ON“. The operation indicator LED has to light now. Attention: if there is no sensor connected the alarm will be activated automatically (safety switching
against accidental dismantling of the sensor). Pay attention at using socket outlet adapters that a sensor is connected to all ends.
8. Put the selector switch of the device on position "I". The device which is attached to the power outlet of the shockproof adapter plug has to switch on. If this is not true you have to check the wiring and whether the power switch is set to on.
9. The favoured switch function can be chosen by the selector switch now:
I - in the case of alarm the power outlet in the shockproof adapter plug is currentless (the con-
nected device is turned off; break contact).
II - in the case of alarm the power outlet in the shockproof adapter plug is conducting (the con-
nected device is turned on; open contact).
10. Finally you have to carry out an operational check of the sensors. The alarm device has to be set to alarm, whereas the alarm has to resound. Depending on the chosen function ("I" or "II") the con­nected device has to react. At misoperation please check the wiring in a currentless status again.
Turn off the alarm device, wait minimum for 5sec. and then turn it on again. In the case of several connected sensors the test has to be repeated with every single sensor.
Your ALSCHU 480 is ready for operation now
5 Measures in case of alarm
- turn off the ALSCHU 480 and disconnect the power plug.
Attention: Please note that the alarm condition of the switch contact will be reset through the turn off of the alarm device! This means that a device which is connected to this outlet will be turned off and re­spectively on. If you do not want this put the selector switch I / II during the attendance on the comple­mentary position.
- determine the failure cause and clear the faults.
- turn on the alarm device again. Please note that between the unplug and plug of the device a time in-
terval of minimum 5sec. has to be passed. If necessary you have to put the selector switch I / II on the right operating position again.
S39.0.01.6C-03 page 3 of 4
6 Safety regulations
This device was designed and tested considering the safety regulations for electronic measuring devices. Faultless operation and reliability in operation of the measuring device can only be assured if the general safety precautions as well as the device-specific security advices in this users manual are considered at the usage of this device.
1. Faultless operation and reliability in operation of the measuring device can only be assured if the de­vice is used within the climatic conditions specified in the chapter „Specifications“.
2. Always disconnect the device from its power supply before opening it. Take care at the installation of the device and circuit points that the unit’s contacts are protected against direct contact.
3. Standard regulations for operating and safety for electrical, light and heavy current equipment have to be observed, with particular attention paid to the national safety regulations (e.g. VDE 0100).
4. When connecting the device to other devices (e.g. PC) the protective circuit has to be designed most thoroughly. Possibly internal connections in third-party devices (e.g. connection of ground with protec­tive earth) may lead to undesired voltage potentials.
5. If it is supposed that the device can not be prosecuted riskless the device has to be switched off and has to be assured against further usage through labelling.
The operator’s safety can be affected by the device if it for example:
- shows visible damage
- does not work anymore like prescribed
- was stored under inappropriate conditions for a longer time In case of doubt the device should strictly be send to the manufacturer for repairing /servicing it.
6. Warning: When running electric devices parts of them are unavoidable electrically live. Unless the
warnings are observed serious personal injuries or damage to property may result. Skilled personnel only should be allowed to work with this device. For trouble-free and safe op­eration of the device please ensure professional transport, installation and mounting as well as proper maintenance.
7. Warning: Do not use this product at safety systems or as an emergency stop device or in any other applications where failure of this product could result in personal injury or material damage. If this advice will be ignored it can result in serious injury or death of persons as well as material damage.
7 Accessories and spare parts
Socket outlet adapter:GAZ-1 socket outlet adapter for connecting two sensors Water sensor: GWF-1S additional water sensor Float switch: GSS-1 S plug-in float switch for electrically non-conductive media (normally
open / normally closed function can be selected by the customer) Level probe: GNS-1 S plug-in level probe with 2 stainless steel electrodes Extension cable: VEKA 2 Extension cable 2m
VEKA 5 Extension cable 5m VEKA 10 Extension cable 10m
S39.0.01.6C-03 page 4 of 4