2.2 RANGE ............................................................................................................................................5
2.5 POWER ............................................................................................................................................5
SECTION 3 – FEATURES
3.1 OPERATOR MENU ...........................................................................................................................5
3.2 DATA DISPLAY .................................................................................................................................5
3.3 LARGE OPERATION RANGE ...........................................................................................................5
APPENDIX A ................................................................................................................................... 33-35
Technical Support:
E-mail: dampers@greenheck.com or
Phone: 1-800-717-6540
3
1 – GENERAL INFORMATION
1.1 – DESCRIPTION
The Vari-Green® Thermal Air Flow Transmitter is used in conjunction with Vari-Green Probe(s). One Transmitter can
receive up to 32 individual flow sensors measuring a single ducted airflow or fan.
The Transmitter’s microprocessor averages flow and temperature signals from all sensors and outputs an analog
signal and local display of these averages. The Transmitter serves as the interface for all configuration and
diagnostic functions.
1.2 – THEORY OF OPERATION
The Vari-Green utilizes thermal dispersion technology
to measure airflow. As air moves across the surface of
an elevated temperature sensor, heat is transferred from
the sensor to the airflow in a mathematically defined
relationship between heat transfer rate and airflow
velocity.
Each Vari-Green sensing node utilizes a pair of precision
matched thermistors. One functions as a reference
sensor measuring the ambient air temperature at the
sensing node location, and the other thermistor is heated
to a preset temperature differential (∆T) above the
ambient air temperature. Airflow velocity is determined
using the measured ambient air temperature, the known
heat transfer characteristics of the heated thermistor,
and the power consumed to maintain the ∆T between
the two thermistors.
Using a combination of Computational Fluid Dynamics
(CFD) modeling and extensive full-scale wind tunnel
verification, the aerodynamically optimized cross section
of the sensing node aperture prevents error inducing
turbulence typical of sharp edged sensing tubes.
Using a combination of Computational Fluid Dynamics
(CFD) modeling and extensive full-scale wind tunnel
verification, the aerodynamically optimized cross section
of the sensing node aperture prevents error inducing
turbulence typical of sharp edged sensing tubes.
The aperture’s injection molded shape has been
engineered to minimize the angular flow effects
naturally present in any duct. The flared aperture
captures a representative sample of the ducted
airflow, while the contoured leading edges prevent the
creation of vortices. The center cross section of the
aperture functions in the same manner as a venturi. It
stabilizes and flattens the velocity profile at the point of
measurement ensuring the airflow maintains full contact
with the sensing thermistors. The flow conditioning
characteristic of the aperture compensates for turbulent
airflow with pitch and/or yaw angles up to ±30 degrees
and produces an accuracy of ±2% of reading for each
Vari-Green sensor.
4
2 – PERFORMANCE SPECIFICATIONS
2.1 – ACCURACY
Sensor Accuracy.
Velocity: ±2% of Reading
Temperature: ±0.1ºF
Overall System Accuracy.
Flow: ±2-3% of Actual Flow
Temperature: ±0.1ºF of Air Temperature
2.2 – RANGE
Velocity Calibration Range.
Low Range (0 to 5,000 FPM)
High Range (0 to 10,000 FPM)
Custom Range (consult Factory)
2.3 – PROBE ASSEMBLY
Mounting.
4 in. x 4 in. aluminum plate with four 0.25 in. holes for
fasteners and neoprene gasket.
Length.
8 in. up to 120 in.
Sensors Per Probe.
Maximum of 16
Connection.
Probe to Probe via RJ-45 cable for signal and power.
Probe to Transmitter via shielded cable with RJ-45
connection. 10 foot length standard with optional 50
foot and 100 foot lengths.
Temperature Limits.
–20ºF to 180ºF Storage
–20ºF to 140ºF Operating
Humidity Limits.
0 to 95% RH, non-condensing.
2.4 – TRANSMITTER
Display.
3 line, backlit graphical display. Flow on Line 1 is
double height and width for improved visibility. Lines
2 and 3 are standard size and are user selectable.
Output.
Dual analog outputs (0-5VDC, 0-10VDC, 4-20mA)
user selectable and configurable.
Configuration.
Via cover mounted membrane keypad
Filter.
Analog outputs filtered via user selectable digital low
pass filter.
Temperature Limits.
–20ºF to 180ºF Storage
–20ºF to 140ºF Operating
Humidity Limits.
0 to 95% RH, non-condensing
2.5 – POWER
Power Supply.
24VAC or 24VDC.
Power Consumption.
16 to 90VA dependent on number of sensor nodes (1 to
32 maximum)
Circuit Protection.
Power input is isolated, fused, and reverse polarity
protected. Earth Ground must be connected to the
terminal labeled "G".
3 – FEATURES
3.1 – OPERATOR MENU
The microprocessor program contains a Operator menu
system for selecting parameters.
Four pushbuttons UP, DN, ENT, and ESC allow the
operator access to the menu for setting configuration
modes and values. Refer to the Configuration
Programming (Section 5.6) instructions for operation
details.
Also contained in the Operator menu is an output
calibration feature. The calibration method is completely
digital and contains no analog potentiometer
adjustments which are susceptible to drift. Non-volatile
memory is used for storing all setup parameters and
calibration values, and will remain unchanged after set,
even when power to the unit is off.
3.2 – DATA DISPLAY
The Vari-Green Thermal Air Flow Transmitter utilizes a
high visibility backlit graphical LCD. 3/4" x 2-1/2" display
with three lines of process data, selectable for U.S. or
S.I. units. Also used for configuration interface.
3.3 – LARGE OPERATION RANGE
The Vari-Green Thermal Air Flow System can maintain
accurate and linear output of flows from 0 to 5,000 FPM
(Low Range) and 0 to 10,000 FPM (High Range). Consult
factory for custom range.
3.4 – ENCLOSURE
The Vari-Green Thermal Air Flow Transmitter is housed
in an aluminum NEMA 1 enclosure with hinged front
cover for access to power and signal terminals for field
connection.
3.5 – MULTIPLE OPERATING POWER
SELECTIONS
The Vari-Green Thermal Air Flow Transmitter can be
powered by either 24VAC or 24VDC.
5
4 – INSTALLATION
4.1 – RECEIVING AND INSPECTION
• Carefully remove theVari-Green Probe(s) or Station,
and Transmitter from the shipping container and
inspect for any damage. If any damage has occurred
in transit, contact freight carrier.
• Review the Factory Set-Up Information Sheet provided
separately and verify the W.O. # and Serial # match
those on the Vari-Green. Verify that the configuration
recorded on the Factory Set-Up Information Sheet
is correct for your application. If not, contact the
Greenheck dampers at 1-800-717-6540 for further
guidance.
Note: The Vari-Green has been configured to customer
provided parameters (see "Factory Setup Information
Sheet" provided separately), and requires no field
calibration/verification prior to installation.
4.2 – LOCATION
• The standard version of the Vari-Green Transmitter is
a NEMA 1 enclosure suitable for most clean indoor
locations. If additional protection is required, mount
the Transmitter in an enclosure with adequate NEMA
rating.
• The ambient temperature of the selected mounting
location must be between -20ºF to 140ºF.
Consideration should be given to units exposed to
direct sunlight.
• The selected mounting location should be rigid and
free of vibration.
4.3 – MOUNTING
Probes. See separate Installation Instructions for the
installation/mounting of Vari-Green Probes and Stations.
Transmitter. This section may be skipped if the
Transmitter was ordered factory mounted to the station.
– Tools Required: electric drill; #25 (0.1495”) bit;
screwdriver or nutdriver; and four #8-32 self-tapping
machine screws.
– The Vari-Green can be mounted in any position
provided it is secured using all four mounting holes.
– Reasonable consideration should be given to
clearances for electrical connections.
– Once a suitable location is found, use the unit as a
template to mark the centers of the four mounting
holes.
– Drill four pilot holes at the marked locations. With the
unit in position, install the four #8-32 screws.
®
Thermal Air Flow Transmitter
®
OFF ON
Figure 4.1
6
4.4 – PROCESS CONNECTIONS
See separate Installation Instructions provided with the Vari-Green Probes or Station for information on connecting
cables from the Probes or Station to the Transmitter.
If theVari-Green have been purchased with a factory mounted transmitter, all connector cabling has already been
installed.
4.5 – POWER/SIGNAL CONNECTIONS
Power wiring is done at terminal strip P8, and signal wiring is done at terminal strip P7. Both P7 and P8 are located
in the lower portion of the transmitter and are accessible by opening the front cover. See Figure 4.2 for location of
P7 and P8. Two conduit openings are provided in the bottom of the transmitter for the power and signal wiring.
Figure 4.2
WIRING.
It is recommended that power wiring be 14 awg to 18
awg, and signal wiring should be 14 awg to 22 awg.
14 awg is the maximum wire gauge that the terminals
can accommodate. No more than two wires should be
connected to any one terminal. To aid in the wiring of
both power and signal wires, P7 and P8 are removable
by pulling the terminal strip straight up and off the circuit
board. Once wiring has been completed, replace the
terminal strip by aligning with receptacle and pressing
firmly.
POWER 24 VAC/DC (Terminals L, N and G, P8).
Power required by the Vari-Green must be connected
to the two terminals labeled L(+) and N(–). Earth ground must be connected to the terminal labeled G. Power
Supply must be 20-28VAC or 20-40VDC.
OUTPUT #1 (Terminals 1(+) and 2(-), P7).
This output represents the measured airflow and is
sourced (powered) by the Vari-Green. Output can be
configured for 0-5VDC, 0-10VDC, or 4-20mADC. Review
Factory Set-Up Information Sheet for your configuration.
If output is to be 0-5VDC: Position both jumpers on
J3. See Figure 4.2.
If output is to be 0-10VDC: Position both jumpers on
J2. See Figure 4.2.
If output is to be 4-20mADC: Position both jumpers on
J4. See Figure 4.2.
In addition to positioning the jumpers for the desired
output type, the corresponding output type must be
selected in the Operator Menu (see Section 5.12).
7
4.5 – POWER/SIGNAL CONNECTIONS (con’t)
OUTPUT #2 (Terminals 3(+) and 4(-), P7).
This output can represent the measured flow (0 - Design
Full Range) or temperature (0 to 140ºF / -18 to 60ºC) and
is sourced (powered) by the Vari-Green. Output can be
configured for 0-5VDC, 0-10VDC, or 4-20mADC. Review
Factory Set-Up Information Sheet for your configuration.
If output is to be 0-5VDC: Position both jumpers on
J6. See Figure 4.2.
If output is to be 0-10VDC: Position both jumpers on
J5. See Figure 4.2.
If output is to be 4-20mADC: Position both jumpers on
J7. See Figure 4.2.
In addition to positioning the jumpers for the desired
output type, the corresponding output type must be
selected in the Operator Menu (see Section 5.12).
SNVT_temp_p105Temperature°C x 100Signed Long2 bytes
For BACnet, connections are as follows:
Terminal 5 – Net A or Net +
Terminal 6 – Net B or Net Terminal 7 – Network Ground
For BACnet objects, see Appendix A.
Set the S1 Termination switches on the BACnet module as follows:
Network Connection (Terminals 5, 6 & 7, P7).
This serial output is available with optional LonWorks or
BACnet (Version 1.5X only) communication.
For LonWorks, connections are as follows:
Terminal 5 – LonWorks B
Terminal 6 – LonWorks A
Terminal 7 – Ground
Two output variables are provided, flow volume and
temperature.
S1 Switch Position
Termination Type123
No TerminationOFFOFFOFF
End of LineONOFFOFF
Fail-safe BiasONONON
Note: The Fail-safe Bias termination includes End of Line termination.
4.6 – DISPLAY CONTRAST ADJUSTMENT
To compensate for different ambient lighting conditions and viewing angles, theVari-Green display's contrast can
be adjusted for optimum visibility. Contrast is adjusted using potentiometer R13. Turn R13 clockwise to increase
contrast (darken characters relative to background) or counterclockwise to decrease contrast.
8
5 – OPERATION
5.1 – INTRODUCTION
The Vari-Green has been configured at the factory to
customer specified parameters. Review this information
and verify that the Vari-Green set-up is correct for
your application. If any problems or discrepancies are
detected, contact Greenheck Dampers at 1-800-7176540 prior to proceeding.
5.2 – START-UP
1. After Installation has been verified in accordance with
Section 4, turn power switch located at the lower front
side to the ON position (see Figure 4.1).
2. Display will briefly indicate:
Followed by:
The number of active sensors will be represented by
the bar graph and listed below. There are 8 active
sensors in this example.
1 8 16 24 32
Sensors Found: 8
Total Sensors: 8
After 5 seconds, display will indicate:
The number of enabled sensors will be represented
by the bar graph and listed below. There are 8 active
sensors in this example.
1 8 16 24 32
After 20 seconds, the display will return to Normal
display mode, and the dot in the upper right of the
display indicating CPU Activated will blink rapidly (4
times per second). This is Normal operation mode
and no further user interface is required. If, however,
user would like to verify configuration or change user
selectable parameters, continue to Section 5.4 entitled
“CONFIGURATION”.
0 ACFM
VELO 0 AFPM
TEMP 68.0°F
Normal display mode
* Dot flashes 4 times a second, indicating that
CPU is active. If ALERT flashes instead of Dot, see
section for Diagnostics.
*
5.3 - NORMAL OPERATION
Under Normal operation, theVari-Green transmitter will
continuously display flow and velocity, or temperature.
5.4 - CONFIGURATION
The Vari-Green onboard microprocessor controls
Configuration: Operating parameter selection; input/
output activation and scaling, and display scaling.
The customer can verify configuration and change
certain parameters (within defined ranges) by
entering the Vari-Green Configuration mode. This is
accomplished using the four membrane pushbuttons
located on the Vari-Green cover. See Figure 4.1 for
location of pushbuttons.
Enabled Sensors: 8
Total Sensors: 8
After 5 seconds, display will indicate:
with the progressing from left to right.
The four pushbuttons used to interface with theVari-Green are as follows: UP DN ENT ESC
In addition to use during Configuration programming, pushbuttons can be used for certain functions when in the
Normal operation mode. The following list describes the pushbutton function when in the Normal operation mode
and in the User Setup (programming).
When in Normal Operation Mode.
ENT then ESC then UP then ENT Activates the Operator menu.
ENT then ESC then DN then ENT Activates the Supervisor menu.
ENT Activates Password selection menu.
ENT then UP then DN then ENT Displays ID, ESID, and WO#. Press UP and software version and maximum
velocity will be displayed. Press ESC and the display will go back to the
Normal operation mode.
When in User Setup.
UP or DN Use to scroll to the desired Main Menu item.
Use to scroll to the value or mode within a parameter.
ENTEnters user into specific selection sub-menu from Main Menu Selection.
Displays current setting of selected Parameter. Enters the selected value or
setting into memory.
ESC Use as an escape key to exit back to Main Menu selection to avoid scrolling
to Return to MAIN MENU
The following pushbutton combinations can be used to more quickly set user selected values (i.e. Process Min and
Max Values, etc.).
UP* + ESC Increases the second column digit.
DN* + ESC Decreases the second column digit.
UP* + DN Increases the third column digit. XXXXX
DN* + UP Decreases the third column digit.
UP* + ESC + DN Increases the fourth column digit.
DN* + ESC + UP Decreases the fourth column digit.
*Must be pressed and held before other button(s) are pressed.
Note: Pushbuttons are momentary type and should be quickly pressed and released to initiate desired change,
unless otherwise instructed to press and hold.
2nd Column
3rd Column
4th Column
5.6 – CONFIGURATION PROGRAMMING
1. With power ON and initialization complete (see Section 5.2), press ENT then ESC then UP then ENT to activate
the Operator menu. The display will indicate:
ENTER PASSWORD FOR
OPERATOR MENU
XXXX
Note: If Operator Menu password has been bypassed (see Section 5.7), the display will indicate as shown in Step 3
below.
2. Using UP and ENT, scroll to the numeric password for the Operator Menu. From the Factory, password is set for
1,111.
3. Once password is displayed, press ENT and display will indicate:
4. Press ENT and display will indicate:
By using UP and DN, the user can scroll through the following selections:
10
OPERATOR MENU
OPERATOR PASSWORD
Main Menu SelectionDescription
between VELOCITY,
between VELOCITY,
Operator Password SelectionAllows for the selection of password for the Operator Menu, and the
bypassing of the password
Flow ConfigurationAllows for the configuration of process variable such as: Duct size
and shape, process type, process maximum/units, and percent
lockdown
Display ConfigurationAllows the selection of parameter to be displayed on Lines 2 & 3,
and display filter
Output 2 Parameter SelectionAllows the selection Output 2 parameter
Low Pass Filter SelectionSelects amount of filtering applied to the output and to the display
Output Signal Type SelectionAllows the selection of signal type for Output 1 and 2
Transmitter Output CalibrationAllows for the zeroing and spanning of transmitter analog Outputs
1 and 2
K-factor ConfigurationTurns K-factor ON or OFF and allows for the calculation or selection
of exponent and gain values
Menu Inactivity Timeout SelectionAllows for the selection a time after which the unit returns to
normal operation if no activity in the Operator Menu
Network ConfigurationTurns the Network ON or OFF, and allows BACnet configuration
Flow Alarm Configuration
(available only with BACnet Network Option)
Temperature Alarm Configuration
(available only with BACnet Network Option)
Operator Menu ExitReturns display to Normal operation
Turns each alarm ON or OFF, select High and Low Setpoints,
Deadband selection and Alarm Delay
Turns each alarm ON or OFF, select High and Low Setpoints,
Deadband selection, and Alarm Delay
5. Once the desired selection is displayed, press ENT to enter its sub-menu.
Once the desired Selection is displayed, its sub-menu can be entered by pressing ENT.
At any time while in the Operator Menu, User can return to Normal operation mode by pressing ESC or scrolling to:
OPERATOR MENU
and pressing ENT.
EXIT
To better understand the process of Configuration Programming, arrows with pushbutton designation have been
included on the following flow chart of Display Configuration. This will aid in the navigation of the Configuration
Programming Process.
Operator Menu
Display Configuration
DN
UP
ENT
ESC
ESC
ENT or ESC
Display Filter
Selection
DN
Line 2 Parameter
Selection
DN
Line 3 Parameter
Selection
DN
Return to
MAIN MENU
UP
UP
UP
ENT
ENT or ESC
ENT
ENT or ESC
ENT
ENT or ESC
Line 2 Parameter
Line 2 Parameter
TEMPERATURE
Display Filter
0 (None)
VELOCITY
UP or DN will scroll
between 0 and 5
UP or DN wil scroll
TEMPERATURE, &
CUSTOM TEXT
UP or DN wil scroll
TEMPERATURE, &
CUSTOM TEXT
Output 2 Parameter
Selection
This example of navigation is similar for all Operator Menu selections.
11
Loading...
+ 24 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.