Introduction
The Telaire® 7001 is an easy-to-use handheld CO
2/Temperature monitor. The unit pro-
vides stable, highly accurate readings due
to our patented dual beam NDIR technology.
Equipped with a 0-4V output, the unit is perfect for long-term monitoring/recording.
The unit features a large, easy-to-read display with a push-button interface which allows for easy calibration, quick adjustments
for altitude correction, and simple toggling
between °C and °F temperature readings.
1
14
13
12
11
9
10
7
8
2
4
5
6
Features
1. Lightweight- Made of ABS Plastic. This
durable, lightweight plastic adds protection
while weighing less than one pound.
2. Display - The large display allows for CO
and T emperature readings at a glance. The
display also allows for adjustments to mode
settings using the push button interface.
3. Soft Touch - Attractive, gives comfort
and extra durability.
4. Up/Down Button - Used to increase or
decrease values while in the edit mode.
5. Mode Button - Toggles between menu
options.
6. Power - Turns the power on and off.
7. Enter - When toggling between menu op-
tions, the Enter button will select desired menu
option for editing (e.g. Altitude settings). After
changes have been made, the Enter key will
also confirm settings and return to the main menu.
8. Calibration Port - Used for gas connection during calibration.
9. Calibration Activation Switch -
Located in the battery compartment, this
switch is used for activating the calibration
process for CO
10. Voltage Output- The sensor is
equipped with an output for datalogging.
(see accessories).
11. Built-in Kickstand- For continuous
monitoring. The built-in kickstand simply
swings out from the base to support the monitor in an upright position.
12. Power Connection - For connecting the
supplied 6V power adapter.
13. Battery Compartment - 4 AA Batteries
are required for portable operation.
14. Battery Cover
2 or Temperature.
Display
Listed below are the modes of operation
visible on the display and is accessible
through the on-board buttons.
Warm-Up Mode - When the power button
is pressed, the monitor enters a one minute
warm-up indicated by the word WARM-UP
in the upper left corner.
Normal Operating Mode - After warm-up
the sensor will stabilize and display current
conditions.
Elevation Settings Mode - This feature allows the user to correct the monitor for elevation changes.
Calibration Settings Mode - Allows user
to set the CO
when performing a calibration.
Calibration in Prog Mode - Displays
when calibration is in progress.
Low Battery - Displays when the power
source is low.
3
Start-Up
Battery Operation - For portable use, the
monitor operates on 4 AA batteries. Expected battery life is as follows:
Normal (Non-Alkaline)..Up to 50 hours
Rechargeable............Up to 60 Hours
Industrial Alkaline ...Up to 70 Hours
Battery Installation - Remove the battery
cover (located on the back) by pressing the
2
pressure clip (located on the bottom near
the kickstand) and pull upward.
Follow the battery diagram imprinted on the
plastic on the back cover for proper installation. Replace battery cover when battery
installation is complete.
Low Battery - Depending on the battery, a
warning signal will flash (indicated by the
blinking Low Battery LCD) when there is
less than 30 minutes of battery life. At this
point the batteries should be replaced or the
AC adapter should be used as a substitute.
If operation continues, the unit will become
inoperable and only the Low Battery will be
blink on the LCD display.
AC Power - The sensor is shipped with a
6V DC 500mA AC/DC adapter.
T o use the AC adapter, connect the plug
into the back of the unit and plug the transformer into any standard wall outlet.
Use the supplied adapter. Using
the wrong adapter may cause
If power is lost during operation, battery operation will not function as a back up.
2 concentration or T emperature
NOTE
damage to the unit.
Operation
Power-Up
1. Press the Power Button.
2. A 2 second delay will occur before the
display becomes visible.
3. 10 seconds will elapse before current
CO
2 readings are displayed.
4. “Warm-up” will display for approximately one minute. During this one
minute warm-up, adjustments can not be
made to the sensor.
Elevation Correction - The sensor is
shipped with the elevation setting set at
“zero” or sea level. The sensor, like any
other gas measuring device is directly affected by altitude changes. If you are at an
altitude greater than 500 feet, an adjustment
should be made to assure sensor accuracy.
T o change the default setting using Elevation
Correction follow the steps below.
1. While in Normal Mode press the “Mode”
button once. The Elevation LCD will begin blinking.
2. Press Enter.
3. Press mode to toggle the elevation reading
between feet (ft) and meters (m)
4. Use the Up/Down button to adjust the altitude. The altitude can be adjusted in increments of 500 (feet) or 100 (meters). Once
the correct altitude is set, press Enter to
lock the setting and return to normal mode.
Display in Fahrenheit or Celsius - To
change the temperature to read in Celsius or
Fahrenheit, or to turn the temperature display off, use the Up or the Down button.
Stand-alone Monitoring - Once the batteries have been installed and the Elevation
correction has been made (as described in
the steps above), the sensor will begin to
accurately display current room conditions.
Kickstand - The sensor is equipped with a
kickstand located on the base of the unit.
For desktop monitoring, pull the kickstand
out to prop the unit on an angle.
Monitoring using an External
Datalogger - Voltage outputs for both CO
and temperature are available via an RJ-45
jack on the rear of the unit.
Calibration (CO2)
1. The calibration procedure will last approximately 5 minutes. Before performing
the calibration procedure, remove the battery cover (see “Battery Installation”) to
provide access to the CO
tivation switch. Connect the supplied AC
adapter to the back of the sensor. If you
do not have the AC adapter, new batteries should be used.
2. Power up the sensor and wait for the
Warm-up to end.
3. Next verify the Elevation correction has
been set. Refer to the steps in Elevation
Correction for procedure.
4. Once the Elevation is verified, press the
Mode button twice. The Calibration mode
will begin blinking.
5. Press Enter.
2 calibration ac-
6. Use the Up/Down Rocker button to adjust
the lower reading on the display to the
current ambient conditions. Pressing the
button once will change the readings in
increments of 10 ppm. To increase the
speed, press and hold the button.
NOTE
For best accuracy, a reference or
known concentration of CO
should be used when adjusting
the reading.
7. Next, on the backside of the unit locate
the push button switch (under the battery
cover, in the small round hole to the right
of the connector jack), use a small
pointed object to depress and hold the
switch for 5 seconds. The Calibration
light will begin to blink.
8. Press Enter.
9. Calibration In Progress will begin to
blink. At this point the unit will program itself based on the CO
put in Step 6. The calibration process will
take approximately 5 minutes.
When Calibration is complete, the display
will return to the steady Calibration
mode. Press Enter to return to the normal operation mode.
2 value that was in-
Temperature
Adjustment
Use this procedure to adjust the temperature
output when, for example, you wish to have
the temperature output match a reference sensor. The accuracy of a field adjustment is dependent upon the stability of the environment
is which the procedure is performed, and
upon the accuracy of the reference sensor.
1. Before performing the temperature adjustment, connect the supplied AC adapter to
the back of the sensor. If you do not have
the AC adapter, new batteries should be
used.
2
2. Power up the sensor and, using the
kickstand, place it on a flat surface in a
stable environment relatively free of
drafts or temperature changes. Wait 30
minutes for the unit to fully equilibrate with
the environment. Do not hold the unit in
your hand during this period.
3. Press the Mode button three times. The
word “TEMPERATURE” at the bottom of
the display will begin blinking.
4. Press Enter. Both the word “TEMPERATURE” and the numeric temperature display will begin blinking in unison.
5. Use the Up/Down Rocker button to adjust the temperature reading to match the
reference.
6. Press Enter. The temperature offset is
immediately adjusted, the blinking stops,
and the unit is now in normal operating
mode.
2
P/N 62292
31899
Specifications
Method
Dual Beam Absorption Infrared™
Sample Method
Diffusion or flow through (50 - 100 ml/min)
Warranty
18 months parts and labor
Performance
CO2 Channel
Measurement Range
0-4,000 ppm voltage output
0-10,000 ppm display
Sensitivity
± 1 ppm
Accuracy
±50 ppm or ±5% of reading, whichever is
greater
Repeatability
±20 ppm
Temperature Dependence
±0.1% of reading per °C or ±2 ppm per °C,
whichever is greater, referenced to 25°C
Pressure Dependence:
0.13% of reading per mm Hg
(Corrected via user input for elevation)
Annual Drift
± 20 ppm typical
Response Time
<60 seconds for 90% of step change
Warm-Up Time
<60 seconds at 22°C
Operating Conditions
32-122°F (0-50°C)
0-95% RH, non-condensing
Storage Temperatures
-40 to 140°F (-40 to 60°C)
Calibration Interval
12 months, offset adjustment using single
gas at 0-1000 ppm CO2. Full factory
calibration available
Temperature Channel
Temperature Range
Voltage output 32 to 104°F (0 to 40°C)
Display 32 to 122°F (0 to 50°C)
Display Resolution
0.1°F (0.1°C)
Display Options
°F, °C, or Off. Set with panel button.
Accuracy
±2°F (±1°C)
Response Time
20-30 minutes (case must equilibrate with
environment)
Calibration Interval
12 months, offset adjustment using
temperature standard at 50 to 86°F (10 to
30°C). Full factory calibration available
Output - Analog
CO2
0-4 VDC, 1mV/ppm (4,000 ppm max)
Temperature
0-4 VDC linear, 32-104°F (0-40°C)
Output Impedance
100 Ohms
Wiring Connection
Via RJ-45 to (accessory cable 2070 or
2071)
Output - Digital
Wiring Connection
Via RJ-45 to DB9 serial port cable
Display
LCD with independent CO2/ temperature
readings (panel buttons set elevation, °F/
°C, calibration functions)
Power Supply
Battery Type
Four AA batteries, not included
Battery Operation
48 hours (alkaline)
External
6 VDC from external AC/DC adapter,
included
Power Requirements
100 mA Peak, 20 mA average from 6V
Certification
FCC Class 15 Part B
Accessories
2070 Datalogging cable for analogoutput
2071 Datalogging cable Recordaire
2075 Calibration kit
1058 Recordaire® datalogger, 4-channel
1058GSW VG Graphing software
All specifications and instructions subject to
change without notice. Protected by United
States Patents - 5,060,508 and 5,163,332.
®
Telaire
®
7001
Carbon Dioxide and Temperature Monitor
Warranty
This Engelhard product has been examined
and tested for proper operation. Please operate this product only in accordance with
the instructions.
Engelhard warrants this product against defects in workmanship and materials for a period of 18 months from the date of purchase
by the original owner. If the product should
become defective within this warranty period
we will repair or exchange it. A return authorization number must be obtained from the
factory prior to returning equipment. Items received without a return authorization number
will be refused. Product to be serviced under
this warranty should be sent to Engelhard,
6489 Calle Real, Goleta, CA 93117. Shipment must be prepaid, properly packed and
insured.
Engelhard is not and will not be liable for
any consequential loss or damages that
may occur by reason of purchase and use
of this product. The responsibility of Engel-
hard, in any event, is strictly limited to the
replacement/repair of the product.
Engelhard seeks to present reliable information concerning the composition, properties
and use of its products, however; (1) All advice concerning selection and use of any
product is provided at no charge and with no
warranty. (2) No warranty is made hereby.
Products described herein are warranted to
conform to Engelhard’s specifications only at
the time of sale. All sales are subject to
Engelhard’s standard terms and conditions,
which are reproduced on the reverse side of
each invoice. All warranties of merchantability and fitness of purpose are disclaimed and
remedy for any breach of warranty is limited
to replacement of the defective product. (3)
Engelhard assumes no responsibility for any
patent liability arising from the use of any
product in a process, manner or formula not
designed by Engelhard.