All possible care has been taken in the preparation of this publication, but
Graseby Medical Limited accepts no liability for any inaccuracies that may be
found.
Graseby Medical reserves the right to make changes without notice both to this
publication and to the product which it describes.
No part of this publication may be reproduced, transmitted, transcribed, or stored
in a retrieval system or translated into any human or computer language in any
form or by any means without the prior permission of Graseby Medical Limited.
GRASEBY MEDICAL LIMITED
Colonial Way,
Watford,
Hertfordshire,
United Kingdom,
WD24 4LG
“Graseby”; “Smiths”; “omnifuse”; “omnistack”; and “Flo-Safer” are all
trademarks of the Smiths Group plc.
All other trademarks are acknowledged as the property of their respective owners.
Graseby Medical Ltd.
Using this manual
Read the first three chapters of this
manual for an overview of the Omnifuse
pump and to find out how to program the
pump for a basic infusion.
Warnings and Cautions
This gives a list of Warnings and
Cautions which you must be aware of
before using the pump. These are
repeated on the relevant pages in the
manual.
Table of Contents
Use the table of contents to see the
structure of the manual and the
headings in the order in which they
appear.
Chapter 1, Introduction
This introduces the Omnifuse pump. It
provides a high-level description of the
features of the pump.
Chapter 2, Basics
This chapter covers the external features
of the pump, and contains detailed
explanations of the techniques for using
the pump, for example, switching on and
loading or changing a syringe. It also
covers the care and maintenance of the
pump.
Chapter 3, Programming the pump
This chapter describes how to program
the Omnifuse pump for a Continuous
infusion. It explains how to use the pump
features designed for the clinical
environment, for example, setting the
occlusion alarm, viewing and resetting
totals, purging the line, using Sleep
mode, and giving a bolus.
Read the remaining chapters for details
on programming and running other
types of infusion and to find out how to
use the pump variants.
Chapter 4, Specialised infusions
This chapter explains how to program
using Mass Units, and describes how to
use the Preset Time, Preset Volume,
Intermittent and Circadian Rhythm
infusion modes on the Omnifuse.
Chapter 5, Protocols
This chapter explains how to program
the Omnifuse by selecting from a menu
of Drug Protocols.
Chapter 6, In-line occlusion sensing
This chapter explains how to use an
Omnifuse pump variant with an In-line
occlusion sensor.
Chapter 7, Lockable cover
This chapter explains how to use the
Omnifuse pump variant with a Lockable
cover.
Chapter 8, Troubleshooting
This chapter explains the warnings and
alarms that may be displayed by the
Omnifuse.
Specifications and Standards
This lists the specifications for the
Omnifuse pump and the standards with
which it complies.
Index
At the end of the manual is an index
which provides a list of key words, and
cross-references these to the relevant
pages in the manual.
Using this manual
Omnifuse Instruction Manual
Warnings/Cautions
Warnings/Cautions
Warnings
Graseby Medical Ltd.
Warnings tell you about dangerous
conditions that could lead to death or
serious injury to the user or patient, that
can occur if you do not obey all of the
instructions in this manual.
1. WARNING: To avoid over- or under- infusion,
always verify that the brand and size of the
loaded syringe are the same as the brand and
size displayed on the screen before starting an
infusion. Failure to do so may result in an
inaccurate delivery of medication, resulting in
patient injury or death.
2. WARNING: To avoid incorrect or inappropriate
configuration of the pump, the Configuration
must only be changed by qualified persons or
authorised personnel. Incorrect pump
configuration could lead to inappropriate
infusion resulting in patient injury or death.
3. WARNING: This equipment is not suitable for
use in the presence of flammable anaesthetic
mixtures with air, oxygen, or nitrous oxide.
The use of the device in presence of such
mixtures may lead to explosion or fire.
4. WARNING: To avoid possible malfunction of
the pump, do not
gamma rays or ionizing radiation, or to the RF
interference or strong electric/magnetic fields
emitted (for example) by diathermy equipment
or mobile telephones. If the pump is used in
the presence of, or in combination with
Magnetic Resonance Imaging (MRI) machine s
it must be protected from the magnetic field
emitted by such equipment. Malfunction of the
pump can cause incorrect infusion or loss of
infusion resulting in patient injury or death.
5. WARNING: Operation of the pump outside the
temperature limits defined in the specification
may result in erroneous operation. Ensure that
the temperature is with in the specified limits.
Failure to do so may result in patient injury or
death.
6. WARNING: In order to ensure that the
intended infusion is performed, data must be
entered correctly. Before confirming any
displayed data the user should ensure that it
is correct. Failure to do so may result in
compromised function of the product, patient
injury or death.
7. WARNING: Failure to follow the Mai nt e nan ce
Procedures described in Chapter 2 of the
OmnifuseService Manual may result in
compromised function of the product and lead
to patient injury or death.
expose the pump to X- rays,
8. WARNING: It is essential that clinical staff
remain within visual and audible range of the
pump so that they can respond promptly to
critical alarms. Failure to respond promptly to
an alarm may result in patient injury or death.
9. WARNING: The user should ensure that the
performance offered by the pump is fit for the
intended purpose. Failure to do so may result
in compromised function of the product,
patient injury or death.
10. WARNING: When the pump is carrying out an
infusion, to ensure that electrical safety is
maintained only items of equipment that
conform to EN60950 are to be connected to
the RS232 connector situated at the base of
the pump, otherwise patient safety may be
compromised.
11. WARNING: While Graseby Medical Limited
have taken all reasonable steps to ensure that
the pump operates correctly while under
remote control, it is the responsibility of the
person who designs and implements the
controlling device to ensure that the resulting
system (pump and controlling device) is fit for
its intended purpose. Failure to do so may
result in compromised function of the product,
patient injury or user injury.
12. WARNING: Do not use a faulty pump. If the
pump detects a fault, a continuous alarm will
sound and the screen will display a System
Fault message. If this happens, switch the
pump off, disconnect it from the mains and
take it to a suitably qualified engineer.
Incorrect performance of the pump can cause
complications resulting in patient injury or
death.
13. WARNING: Failure to use the mains lead
supplied with the pump will compromise the
pump’s ability to resist fluid ingress, resulting
in possible user or patient injury or death.
14. WARNING: Correct management of battery
charging is essential to ensure that the pump
can operate on batteries for the time specified.
Failure to do so may lead to impaired
functioning of the pump, resulting in patient
injury or death.
15. WARNING: The occlusion alarm level must be
checked before starting an infusion to ensure
that it is appropriate for that infusion. Failure
to do so may result in an unacceptably slow
time to occlusion alarm, resulting in patient
injury or death.
Omnifuse Instruction Manual
Graseby Medical Ltd.
16. WARNING: If an occlusion alarm occurs,
immediately clamp the line to eliminate the
possibility of a bolus being delivered to the
patient. Then inspect the fluid pathway for
kinks, clogged catheter, etc. in order to
remove the occlusion prior to restarting the
infusion. An unintentional bolus of medication
can result in patient injury or death.
17. WARNING: Use only the syringes and
administration sets listed in Specifications and Standards at the end of this manual .
Failure to do so may result in an inaccurate
delivery. Graseby Medical does not
guarantee performance of the pump if
syringes other than those listed are used.
Incorrect function or performance of the pump
can cause complications resulting in patient
injury or death.
18. WARNING: When using the In-line Occlusion
Sensing option, use only a Graseby Medical or
Graseby Medical-approved extension set (part
number 0130-0041). Graseby Medical does not
guarantee performance of the pump if
unapproved extension sets are used. Failure
to observe this warning may lead to
compromised performance of the pump,
resulting in patient injury or death.
19. WARNING: The volume of fluid contained in
the connecting tubing is a residual amount
and will not
made for this extra volume of fluid when
initially filling the syringe and purging the
system. Under-delivery of medication can
cause complications resulting in patient injury
or death.
20. WARNING: To avoid patient embolism, ensure
that the patient tubing is purged of all air
bubbles before administering any medication.
The pump provides a purge facility to assist
with this process. The presence of air within
the medication can result in complications
leading to patient injury or death.
21. WARNING: For safe operation of the pump, the
syringe must be correctly loaded. Ensure that
the syringe plunger is properly aligned before
closing the barrel clamp. Failure to do so may
result in inaccurate delivery, resulting in
patient injury or death.
22. WARNING: Ensure that your fingers are not in
the path of the pusher during syringe loading
or unloading. Failure to do so may result in
user injury.
be infused. Allowance must be
23. WARNING: To avoid syphoning of the syringe
contents (free flow), ensure that the syringe is
correctly loaded into the pump, that the
syringe plunger is properly engaged by the
pump’s pusher block and that the pump is
placed not more than 80cm above the infusion
site. Syphoning can result in over-infusion
leading to patient injury or death.
24. WARNING: To avoid over-infusion, do not
purge the infusion line when the
administration set is connected to the patient.
Over-infusion of medication can result in
patient injury or death.
25. WARNING: To avoid the pump becoming
detached from an IV pole always make sure
that the pump is securely fixed to the pole.
Always check the security and stability of the
assembly with the pump mounted.
If no IV pole is used make sure that the pump
is completely stable on a horizontal surface.
Failure to observe this warning may cause
damage to the Omnifuse pump and harm the
user or the patient. As a result, the user or
patient may suffer direct injury, or the
Omnifuse pump may fail to operate correctly,
leading to patient injury or death.
26. WARNING: Following a significant liquid spill
onto the pump, it should be wiped dry and
inspected by service personnel before being
returned to service. Failure to do so may result
in compromised functioning of the pump,
leading to patient or user injury or death.
27. WARNING: To avoid syphoning of the syringe
contents (free flow), ensure that the patient
line is clamped before loading or unloading
the syringe. Syphoning can result in overinfusion leading to patient injury or death.
28. WARNING: When the Omnifuse pump is fitted
with a Lockable cover, the cover and
associated security software should always be
used in accordance with local protocol. The
Lockable cover protects the syringe and its
contents from tampering; the security
software ensures that the infusion parameters
are not interfered with. Failure to use the
security software together with the Lockable
cover could result in an inaccurate delivery
leading to patient injury or death.
Warnings/Cautions
Omnifuse Instruction Manual
Warnings/Cautions
Cautions
Graseby Medical Ltd.
Cautions tell you about dangerous
conditions that may occur and cause
damage to the pump if you do not obey all
of the instructions in this manual.
1. CAUTION: Refer all service, repair and
calibrations only to qualified technical
personnel. Unauthorised modifications to the
pump must not
2. CAUTION: When turning the pump on, if
screens similar to those illustrated are not
displayed do not use the pump, and send the
pump to authorised service personnel.
3. CAUTION: Do not attempt to move the pump’s
pusher by hand. Always use the syringe Load
key
W
this caution may cause mechanical damage to
the pump.
4. CAUTION: Never carry the pump except by the
handle. Failure to do so may result in damage
to the case, or you may drop the pump and
cause it internal damage.
be carried out.
to move the pusher. Failure to observe
5. CAUTION: Do not use cleaning and
disinfecting agents other than the approved
ones specified here.
6. CAUTION: The pump must not
any liquids or exposed to strong organic
solvents. Wipe off spills immediately, and do
not allow fluid or residues to remain on the
pump. Additionally, the pump is not designed
to be autoclaved, steam-sterilised, EtOsterilised or subjected to temperatures in
excess of 45° C (113° F). Failure to observe
this caution may cause serious damage to the
pump.
7. CAUTION: Failure to use the mains lead
retainer means that the pump may be
accidentally disconnected from the AC mains
supply.
8. CAUTION: Users should bear in mind that the
syringe-ear clamp is for location only and may
not be powerful enough to hold the syringe in
place against the powerful negative backpressures that may be encountered in certain
clinical applications.
be immersed in
Omnifuse Instruction Manual
Graseby Medical Ltd.
Contents
Chapter 1 - Introduction
Introduction to the Omnifuse pump ...................................................... 1 - 1
Omnifuse pump features ................................................................................1 - 1
Trumpet curves ........................................................................................ S - 9
Index
ivOmnifuse Instruction Manual
Chapter 1
Introduction
Graseby Medical Ltd.Introduction to the Omnifuse pump
Introduction to the Omnifuse pump
The Omnifuse is part of the wide range of
Graseby pumps. Omnifuse syringe
pumps have been designed to meet the
growing needs of the clinical
environment. They are suitable for the
administration of drugs or other
parenteral fluids by or under the
supervision of healthcare professionals
such as physicians and nurses.
Omnifuse pump features
The features on the Omnifuse pump can
be tailored to suit individual hospital and
clinical requirements.
The pump offers a Continuous infusion
mode as well as other modes for
specialised infusions: Preset Volume,
Preset Time, Intermittent and Circadian
Rhythm.
The clear messages on the screen guide
you through an intuitive programming
and running sequence, with further
information available from on-screen
help explaining how to use a feature or
follow a prompt. The Omnifuse pump can
be run directly from the AC mains supply
or, when necessary, from its own
rechargeable backup batteries.
The pump has been designed to be simple
to use. Most actions are performed by
choosing options from the screen with
the Command wheel which you turn and
then press. There is a keypad for entry of
numerical data, for stopping and starting
the pump and for switching it on and off.
Carefully read the entire contents of
this manual before using the pump.
This Instruction Manual describes the
features of the standard Omnifuse pump,
explains how to use the pump and also
describes the pump variants.
The pump has a green Command wheel,
LCD surround and keypad.
Safety and security
Many safety features have been
incorporated into the Omnifuse pump:
• Self-test routines run each time the
pump is switched on to check it is
working correctly.
• Continual checks are also made
while the pump is running,
especially during an infusion.
• A wide range of audible and visual
alarms and warnings, including the
LCD screen flashing.
• Each infusion setting must be
reviewed and confirmed before an
infusion can start.
• Occlusion pressure is monitored as a
standard feature. The dry-side
pressure level is displayed while the
pump is running. The occlusion
alarm level setting is adjustable but
may be locked to restrict the upper
level if required.
• A history of 3000 of the pump’s most
recent events or actions, stamped
with their date and time can be
viewed on the pump screen, or
transferred to a PC.
Introduction
Omnifuse Instruction Manual 1 – 1
Introduction to the Omnifuse pumpGraseby Medical Ltd.
User interface
The screen and controls on the pump are
easy to use.
• All the information for programming
and running infusions is shown on
the LCD screen (Liquid Crystal
Display).
• Next to the screen is a multi-function
Command wheel which you turn and
Introduction
press to select items and enter
information.
• The pump has a numeric keypad for
entry of data such as bolus dose.
• An Infusion LED flashes to show
that the pump is infusing and
changes colour if the syringe nears
empty. A yellow LED is lit when the
pump is connected to the AC mains
supply.
• While you program the pump and
during an infusion, the screen shows
instructions on a message line. Help
and information icons lead to more
detailed information.
• Status icons show the battery level
and occlusion pressure level.
• A special feature lets you program
the pump ready for an infusion, then
leave it ‘asleep’ until it is needed.
• The pump totaliser shows totals for
the infusion and totals since last
reset. A shift totals feature is also
available.
Administration sets
Graseby medical have a wide range of
Flo-Safer administration sets that may
be used for specific infusion therapies.
These are listed in the Specification at
the end of this manual.
Pump configuration
The features required for a particular
clinical application must be configured
on the pump, by a qualified technician,
prior to use in the clinical area.
The technician uses Graseby Omnifuse
Technician PC software to enable
features on the Configuration menu,
then transfers this information to the
pump. Alternatively, the technician
accesses the Configuration menu directly
on the pump, using a special password.
Once the Omnifuse has been configured
for use, it can be programmed with
specific infusion settings according to the
clinical requirements of the patient and
medication.
In addition to the Configuration menu,
there is a Technician menu giving access
to information about the use of the pump.
This menu is normally accessed using PC
software but may be accessed directly on
the pump, using a special technician’s
password. For more details about
Configuration and Technician menus,
read the Omnifuse Technical User Manual.
Syringes
On the Omnifuse, the syringe can be
loaded as a one-handed operation.
Syringes of 2 to 60 ml capacity can be
used with the pump.
The syringe type and size are confirmed
as the first step in programming an
infusion. The syringe types that are
available on the pump’s menu depend on
which have been configured for use.
1 – 2Omnifuse Instruction Manual
Graseby Medical Ltd.Introduction to the Omnifuse pump
Bolus features
The Omnifuse pump may be configured
to offer one or two methods for
administering a bolus: a hands-on and/or
a pre-set bolus. The maximum bolus rate
can be set up to 800 ml/h although this
also depends on how the pump has been
set up and the size of syringe in use.
If no bolus is needed for the clinical area,
then the bolus feature can be turned off.
Mass Units
A wide range of mass units can be
configured for use on the Omnifuse. The
pump can be programmed to give a dose
of medication which is expressed as a
mass unit, rather than volume to be
infused over a period of time.
Protocols
The pump can offer the clinical user a
selection of protocols. Before the pump is
used, the hospital/community protocols
relevant to the clinical area must be
loaded into the pump. This is done using
the Omnifuse Drug Protocol
Management System.
Power
The Omnifuse is normally run from the
AC mains supply. However, it switches
automatically to internal rechargeable
batteries in the event of a power cut or if
it is disconnected from the mains.
On fully charged batteries, the pump
may be run for up to 10 hours depending
on the infusion rate.
External communications
The pump has an RS232 serial cable
connector with PC interface protocol.
This is used for configuring the pump
from a PC, is used to transfer the pump
hi sto ry to a PC and is also used to control
the pump externally.
Introduction
With Protocols in use, all the user in the
clinical area needs to do to set up the
pump for a patient is to load the syringe,
select the protocol and confirm the
parameters displayed on the screen.
Graphics
This feature allows you to view the
progress of the infusion in graphic form
as an infusion profile.
Omnifuse Instruction Manual 1 – 3
Introduction to the Omnifuse pumpGraseby Medical Ltd.
Omnifuse pump variants
The standard Omnifuse has no In-line
occlusion sensing capability or Lockable
cover. Three Omnifuse pump variants
are available for use in environments
where In-line Occlusion Sensing or a
Lockable cover are required.
The variants can be specified when the
pump is purchased, or the standard
pump can be upgraded to one of the
Introduction
variants, if it is returned to Graseby
Medical. The three variants are:
• Omnifuse with In-line Occlusion
Sensing
• Omnifuse with Lockable Cover
• Omnifuse with In-line Occlusion
Sensing and Lockable Cover.
In-line Occlusion Sensing
For clinical applications where occlusion
pressure needs to be monitored very
precisely, for example in a Neonatal unit,
the pump should be ordered with an inline occlusion sensor.
This is used with a special extension set,
which includes a pressure sensing disc,
to monitor occlusion pressure in the
infusion line while the pump is infusing.
Lockable Cover
The Lockable cover pump variant may be
used in clinical areas where it is
necessary to prevent the syringe from
being removed and to prevent any
change to the flow rate for the infusion.
The Lockable cover prevents physical
access to the syringe except by an
authorised person with a key. The cover
can be used with the Omnifuse’s Security
software enabled, which locks the
keypad once the infusion is started. This
prevents accidental changes to the
infusion rate, or access to the bolus
feature.
In-line Occlusion Sensing and
Lockable Cover
The Omnifuse is also available as a
pump fitted with both the In-line sensor
and the Lockable cover.
If required, an infusion can be started
without the in-line sensor, in which case
the pump reverts to the standard dryside occlusion sensing.
With In-line sensing in use, the Graph
hotspot offers a choice of infusion profile
graph or a pressure trend graph.
1 – 4Omnifuse Instruction Manual
Graseby Medical Ltd.PC software
PC software
With a PC linked to the pump by RS232
cable, you can transfer information
between the pump and PC.
Two software packages are available on
CD-ROM from Graseby Medical Ltd. for
use with the Omnifuse pump.
Omnifuse Drug Protocol
Management System
The Protocol support software package is
used to design and create Protocols and
download them for use on the Omnifuse.
Graseby Omnifuse Technician PC
Software
This software package is supplied with
the Service Manual. It is required for
technical maintenance and servicing an
Omnifuse pump, as well as pump
configuration. The Technician PC
software is used to:
• Perform service tests and calibration
on the pump.
• Set parameters and enable or disable
options that would otherwise be
changed using the pump’s own
Configuration menu.
• Download and examine the pump’s
history database.
• Keep a service database for pumps.
• Store configurations and settings
from the pump and restore them
later.
Introduction
Omnifuse Instruction Manual 1 – 5
Omnistack pump stacking systemGraseby Medical Ltd.
Omnistack pump stacking system
The Omnistack is a multi-pump stacking
system designed for mounting Omnifuse
syringe pumps. Up to four Omnifuse
pumps can be mounted on an Omnistack,
powered from a single mains inlet.
The Omnistack can be clamped to the
Graseby wheelbase, or to a suitable
pylon. The Omnistack must not be used
with a standard IV pole.
Introduction
For full details on how to use the
Omnistack, see the Omnistack Instruction Manual.
The part numbers for the Omnistack
stacking system, pole and wheelbase are
listed in the Accessories section of the
Specification at the back of this manual.
1 – 6Omnifuse Instruction Manual
Chapter 2
Basics
Parts of the Omnifuse pump 1
Switching the pump on and off 4
Omnifuse screen, icons and indicators 8
Loading and unloading a syringe 12
Care and maintenance 18
Page
Graseby Medical Ltd.Parts of the Omnifuse pump
Parts of the Omnifuse pump
This section shows the main external
features of the standard Omnifuse pump.
It has no In-line occlusion Sensor or
Lockable cover.
Rear of the pump
For information on in-line occlusion
sensing on the Omnifuse, see In-line occlusion sensing, Chapter 6 or for
details on the Lockable cover, see
Lockable cover, Chapter 7.
Basics
Omnifuse Instruction Manual 2 – 1
Parts of the Omnifuse pumpGraseby Medical Ltd.
Front of the pump
Basics
This illustration shows the main external features on the
front of the Omnifuse pump.
2 – 2Omnifuse Instruction Manual
Graseby Medical Ltd.Parts of the Omnifuse pump
The Command wheel, keypad and LEDs
Command wheel
You enter commands into the pump with
a multi-function Command wheel.
You can turn the wheel freely, either
clockwise or counter-clockwise.
If you press the wheel down, you will feel
a positive click.
Keypad
Next to the Command wheel is the
keypad where you will find the:
On/Off key
S
Start key (Green)
T
Stop key (Red).
U
The numeric keypad is for entering
numbers. It includes a Cancel key and
Decimal point key.
On the right of the keypad is:
Alarm Silence key (Red)
V
Syringe Load key (Blue).
W
• Turn the wheel in either direction to
highlight items on the screen such as
hotspots
• Press the wheel to select, accept or
confirm the parameters or action.
Basics
LEDs
The pump has three LEDs:
Infusion LED - This is green, or amber.
It is lit or flashes when an infusion is
running. For full details, see Pump indicators‚ page 2 - 10.
Mains LED - This is yellow. It is lit
when the pump is connected to the AC
mains supply and goes out if the pump is
operating on batteries.
Alarm LED - This flashes red when the
pump is sounding the alarm.
Omnifuse Instruction Manual 2 – 3
Switching the pump on and offGraseby Medical Ltd.
W
s
t
c
Switching the pump on and off
Before you switch the pump on, visually
check for damage to any part of the pump
or its connectors.
Plug it in to an AC mains supply if
possible. If necessary the pump can be
run on its internal batteries see Using the pump on batteries‚ page 2 - 6.
Switching on and power-up tests
To switch on, press On/Off S. The
initial screen shows:
The pump continues with its self test:
Basics
Service due message
The pump checks its service due date at
the end of the power-up tests. If a service
is due, you will see a message, for
example:
The pump can still be used. To continue
with your setup, simply press the
Command wheel. If however you wish
the pump to be serviced, switch off.
The pump will display the service due
message when you next switch it on.
Faulty pump
If the pump discovers a fault during the
The Pump ID and Location shown on this
screen are set up in the Technician menu
described in the Omnifuse Technical
User Manual.
CAUTION: When turning the pump on, if
screens similar to those illustrated are
not displayed do not use the pump, and
send the pump to authorised service
personnel.
2 – 4Omnifuse Instruction Manual
power-up tests, it displays a message. Do
not use the pump, but return it to a
qualified service engineer. For more
details, see Troubleshooting, Chapter 8.
ARNING: Do not use a faulty pump. If the
pump detects a fault, a continuous alarm will
ound and the screen will display a System
Fault message. If this happens, switch the
pump off, disconnect it from the mains and
ake it to a suitably qualified engineer.
Incorrect performance of the pump can
ause complications resulting in patient
injury or death.
Graseby Medical Ltd.Switching the pump on and off
Power-up tests completed
When the power-up tests are complete,
the screen shows the first of the syringe
loading instructions.
For example:
The pump is now ready for you to load
the syringe, as described on page 2 - 12.
Switching off
Before you can switch off, the pump must
be stopped. If an infusion is running:
• Press and hold the Stop
The screen displays:
U
key.
Basics
To switch off the pump:
• Press the On/Off key
down for two seconds.
As the pump switches off, the screen
displays:
Ensure that the pump is connected to the
AC mains to keep the batteries charged
and ready for use next time.
Omnifuse Instruction Manual 2 – 5
and hold it
S
Switching the pump on and offGraseby Medical Ltd.
r
Using the pump on batteries
In everyday use, the Omnifuse pump
should be connected to a suitable AC
mains supply, not used on batteries.
This way, the batteries are kept fully
charged, available for use in an
emergency.
• The yellow Mains LED is only lit
when the pump is connected to the
AC mains supply
• The battery icon P is shown on the
screen when the pump is running on
batteries
• The battery icon is displayed when
the pump is connected to the AC
mains supply, to show that the
batteries are being charged. When
Basics
the batteries are charged, the icon
disappears.
Pump with patient in transit
If a mains supply is unavailable - for
example, if the pump is infusing a
patient in transit - you can start or
continue an infusion with the pump
running on batteries.
Mains lead retainer
The mains lead is held in the pump with
a retainer. This is so that it cannot be
pulled out accidentally.
CAUTION: Failure to use the mains lead
retainer means that the pump may be
accidentally disconnected from the AC
mains supply.
WARNING: Failure to use the mains lead
supplied with the pump will compromise the
pump’s ability to resist fluid ingress,
resulting in possible user or patient injury o
death.
Mains supply failure
If the AC mains supply fails or if the
pump is disconnected from the mains, it
automatically switches to its internal
batteries.
If the pump is infusing, the infusion
continues without interruption. This
message is displayed:
• The yellow Mains LED goes out and
the battery icon P is shown on the
screen when the pump is running on
batteries.
The pump will display a warning
message approximately 30 minutes
before battery power runs out.
When the pump is reconnected to the
main supply, or mains power resumes,
the pump automatically switches back to
the AC mains supply.
Note: When the pump is being used
on battery, an independent battery
monitor is active. This provides an
additional check on the battery
condition.
If the pump’s alarm sounds but no
message is displayed on the screen,
the independent battery monitor has
detected a problem.
The alarm cannot be silenced and the
pump cannot be used. It must be
taken to a qualified technician for
repair.
2 – 6Omnifuse Instruction Manual
Graseby Medical Ltd.Switching the pump on and off
Recharging the batteries
While the pump is switched off, you
should leave it connected to the AC
mains supply so that the batteries are
always kept fully charged.
Recharging can take up to 10 hours if the
batteries are completely flat.
While the pump is running on the mains
supply, the battery icon P is displayed to
show that the batteries are being
charged. The battery icon disappears
once the batteries are fully charged.
Battery life
As a rough guide, an Omnifuse pump can
operate on battery power for up to 10
hours at a continuous infusion rate of
5 ml/h with the batteries starting from
fully charged. This period includes four
syringe load/unload cycles.
To ensure that the batteries perform as
specified they must always be fully
charged. The batteries must not be left
partially charged for extended periods,
for example when the pump is in store.
WARNING: Correct management of battery
charging is essential to ensure that the
pump can operate on batteries for the time
specified. Failure to do so may lead to
impaired functioning of the pump, resulting
in patient injury or death.
Omnifuse Instruction Manual 2 – 7
Basics
Omnifuse screen, icons and indicatorsGraseby Medical Ltd.
Omnifuse screen, icons and indicators
Moving around the screen
To move around on the screen you turn
the Command wheel in either direction.
The highlighting shows you where you
are.
As well as the main command on the
screen, you can see:
•The message line
• Hotspots
Basics
•Icons.
Message line
This gives you more details about the
command on the screen, or shows the
status of the infusion, or shows you a
warning or alarm message reminder.
Hotspots
A hotspot is an action word on the bottom
line of the screen. The hotspots you can
see are relevant to whatever the pump is
doing. The most common hotspot is
CONFIRM.
In many cases, the hotspot you need is
automatically highlighted on the screen,
so you can just press the Command
wheel.
If the relevant hotspot is not highlighted,
turn the Command wheel to highlight it,
then press the Command wheel.
When you press the wheel, the pump will
carry out the command. In this example,
if you press the Command wheel with
SKIP highlighted, the pump will skip the
syringe load process.
Icons on the pump screen
This example screen illustrates where
the Omnifuse screen icons appear:
Some icons are displayed on the screen
all the time, others only appear in special
circumstances.
The table on page 2 - 9 shows all the
icons that may appear.
2 – 8Omnifuse Instruction Manual
Graseby Medical Ltd.Omnifuse screen, icons and indicators
Screen icons
IconNameDescription
InfoDisplays settings and parameters for the
H
HelpDisplays a screen of help information relating
T
current infusion.
to the current state of the pump.
C Occlusion Level
(Dry-side)
Battery levelThis is displayed as a static icon when the
P
Hold (Graph)This appears when the pump is displaying the
I
Return (Graph)This is displayed when the pump is holding
F
Occlusion Level
(In-line)
The position of the vertical line shows the level
at which the occlusion alarm pressure has
been set, from Level 1 on the left to Level 5 on
the right. When the pump is running the icon
fills from left to right to show the current
pressure. You use this icon to set the alarm
pressure level, see How to set or change the occlusion alarm level‚ page 3 - 4.
This replaces the dry-side occlusion level icon
when the sensor is use on an Omnifuse with
In-line occlusion sensing, see In-line occlusion sensing, Chapter 6.
pump is running on batteries, or as a flashing
icon when the pump is connected to the AC
mains supply and the batteries are charging.
Graph screen. Select this icon to hold the
screen. While the screen is on hold, the icon
changes to the return icon shown below.
the Graph screen. Select this icon to return to
the running screen, or previous graph screen.
Basics
Zoom (Graph)This is displayed during long infusions, when
J
Omnifuse Instruction Manual 2 – 9
the graph display can be zoomed to show more
detail for a particular hour.
Omnifuse screen, icons and indicatorsGraseby Medical Ltd.
Pump indicators
There are three LEDs or indicators on
the front of the pump, positioned above
the numeric keypad.
Infusion indicator
The Infusion indicator shows whether or
not the pump is infusing, and changes
colour when the syringe is nearly empty.
See details in the table below.
Basics
Infusion LEDShows ...
FlashingMedication being
delivered.
SteadyInfusion suspended -
Stop key
pressed.
LED ColourShows ...
T
Mains indicator
The yellow Mains indicator is lit when
the pump is connected to the AC mains
supply.
Alarm indicator
The red Alarm indicator flashes when
the pump sounds the alarm.
was
GreenSyringe is full.
AmberSyringe is nearly
empty.
2 – 10Omnifuse Instruction Manual
Graseby Medical Ltd.Omnifuse screen, icons and indicators
Sounds on the Omnifuse
This table describes the sounds that you
may hear from the Omnifuse pump and
tells you what action is required if you
hear them.
SoundOccurs when...What you should do
BeepYou press a key or the
Command wheel.
Double beepThe pump requires
your attention.
AlarmThe pump has
discovered an operating
problem, for example
the syringe is empty or
there is an occlusion.
Continuous
System Alarm
The pump has detected
an internal error.
For more details on pump warnings and
alarms, see Troubleshooting, Chapter 8.
If the beep is too soft or too loud, call
a technician to change it for you via
the Key Beep Volume parameter in
the Configuration menu.
Read the message on the screen,
then carry out the suggested action.
Read the message on the screen,
then press the Alarm Silence key
and carry out the suggested
V
action.
Read the message on the screen.
Pressing Alarm Silence V will
have no effect. Switch off the pump
and hand it over to a qualified
technician for servicing.
Basics
Setting the alarm volume
The volume of the alarm on the
Omnifuse can be set to one of three
levels:
• 1 - Soft
•2 - Medium
•3-Loud.
Provided the pump is not infusing and
the cover is open, you can adjust the
volume to suit the operating
environment.
Omnifuse Instruction Manual 2 – 11
To adjust the alarm volume:
• Press and hold the Alarm Silence V
key and at the same time press the
1, 2 or 3 key on the keypad.
As you press a number, the pump
demonstrates the volume of the alarm at
your chosen level.
The last number you press sets the
volume that the pump will use for the
alarm, until set to a different volume.
Loading and unloading a syringeGraseby Medical Ltd.
Loading and unloading a syringe
The pump is designed to make loading a
syringe a one-handed operation. There
are three components that secure the
syringe:
• The syringe ear slot
• The barrel clamp
• The pusher.
The syringe is placed in the trough at the
front of the pump. The ear or flange of
the syringe should be placed into the
syringe ear slot. The barrel clamp is
lowered by hand onto the syringe.
The sections on the following pages
provide full details on loading/unloading
a syringe.
Basics
When to load the syringe
When you turn the pump on, you see the
screen below, unless a syringe is already
loaded:
Please note that Warnings relating to
syringe loading and unloading are listed
on page 2 - 14.
When to skip loading
If the syringe is not yet ready, or if
someone else is preparing the syringe
while you program the pump, you can
skip loading. The SKIP hotspot allows
you to go ahead with programming the
pump and load the syringe later on.
If a syringe is already loaded you will
only be asked to confirm the brand and
size.
Please read Warnings and Cautions concerning syringes, on page 2 - 14.
2 – 12Omnifuse Instruction Manual
Loading...
+ 100 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.