Graphtec GS-TH Operating Manual

GS-TH-UM-151
For GL100
Temperature / Humidity Sensor
USER’S MANUAL
Thank you very much for buying this GRAPHTEC product.
This product is a measurement sensor (hereafter “module”) that connects to the GL100-N/GL100-WL.
These directions describe preparations and cautions before measurement.
To ensure safety, please read the operation instructions, etc.
For details on the warnings and how to handle this module, please read the Quick Start Guide or USER'S MANUAL included on the CD-ROM (included in the GL100 packaging)
Confirmation of the exterior
After opening the package, please confirm that there are no problems (scratches and dirt) on the exterior before use.
Confirmation of the attached items.
User’s manual (this book): 1
If by any chance faults are found, please contact the store where you bought the item.
* Please note that items mentioned in this book may change without
prior notice.
MANUAL-TE604309041
1 Part Names
This section describes the name and function of each part.
1. Hook portion
2. Connector
3. Cable packing
GL100 main module
1. Hook portion ..................
2. Connector .....................
3. Cable packing ...............
Attaching just this module to the wall, etc. will damage the connector cable. Always connect to the GL100.
The temperature sensor measures the change in capacitance of the conductivity caused by moisture absorption. Therefore, dust, fumes and other organic compound may affect measurements. Usage in an environment with a large quantity of these substances floating about will cause large measurement deviations.
After connecting the GL100 to modules or sensors, please always check/set the time and date.
Used to mount to a wall. Used to connect to the connector on the GL100 module. This packing is used when connecting the connector.
Wall mounting hook
2 How To Measure 6 Combined Measurement
1. Power supply (Refer to Quick Start Guide or USER’S MANUAL.)
Connect this module while power is being supplied to the GL100 by a battery or USB cable.
2. Start-up and operation
(1) Screen display menu flow
After power-on, the GL100 is ready for operation by holding down [MENU] key. When the module is connected, “Module Type Recognition” screen is displayed. When the module is not connected, “Module Unconnected State” screen is
displayed.
Set the alarm information. Please set the number level.
A LAR M]                4/ 9 Al arm :O ff▽                
ALARM settings
Alarm Off
Level Off / Mode Level TEMP /HUM
Off H L
Value setting
(6) Temperature Unit Setting
On the OTHER-2 screen, you can switch the temperature setting between Celsius and Fahrenheit.
1. Combined temperature and humidity sensor and illumination / ultraviolet sensor measurement
Composite measurement can be done by using the branch adapter for GS (GS-DPA) and the illumination / ultraviolet sensor (GS-LXUV) (each sold separately).
Illumination /
Ultraviolet sensor
(GS-LXUV)
Temperature and humidity sensor (GS-TH)
Operate in accordance with the displayed instructions.
(5) ALARM setting
 S ENSO R ER ROR! !       
 P leas e co nnec t           the  sens or            QU IT k ey t o Po wer  Off
BA T LA N                 
”P le ase  w ait ”   
GL 100- TH               
 S leep ing! !              E NTER  key  to  sta rt  
QU IT k ey t o Po wer  Off
BA T LA N                 
 G L1 00- TH            I ni tia li zin g!!  
(2) Free-running screen
ST OP     ALM .        1 :2 8
TE MP :        +2 6. 7℃ HU M. :        +3 9. 8%
DE WP :      + 12 .1 8℃
BA T L AN  SD      S : 1 .0 s
Module unconnected state
<Operation> Connect the module.
Recognition of module types
Standby state
<Operation> Press [ENTER] key.
Module start-up
Hold down the [QUIT] key (approx. three seconds) to put the module into standby state.
When running on batteries, the module will automatically go into standby state after three minutes of no operation.
Press the [ENTER] key while in standby state to return to the free-running screen.
3. Setting
(1) Screen operation
Item selecting screen
Press the [MENU] key on the free-running screen to go to the setting screen. <How to set> Select the item with the directional keys ( ) and press the [ENTER] key.
[D ATA ]                 2/ 9 Sa mpl in g:1 s▽              Ca ptu re  MO DE :CO NT. ▽    Ca ptu re  DE ST :SD  Ca rd ▽
Fr ee  CA PA: 49 818 0     
Numerical entry screen
<How to set> Numbers can be inputted by increasing or decreasing the value with the and keys.
[D ate /T ime ]          7/ 9 Ty pe:  Y Y:M M: DD▽          Da te:  2 014 -0 1-0 4        Ti me:    03: 18 :26 ▽       
(2) AMP setting
The accumulated temperature can be set for the set temperature.
A MP]                   1/ 9
Mo de             RE F.T EM P ↑R EF. TE MP▽    ± 28. 00 ℃▽
(3) DATA setting
Set the Sampling and Capture Mode those will be recorded to the data recording media. The recorded data’s size will be displayed in the information for the SD card being recorded to. Please take note of it.
D ATA ]                 2/ 9 Sa mpl in g:1 s▽              Ca ptu re  MO DE :CO NT. ▽    Ca ptu re  DI ST :SD  Ca rd ▽
Fr ee  CA PA: 49 818 0     
(4) TRIGGER setting
Select the conditions for beginning data recording after measurement starts. Off :
Pressing the [START/STOP] key on this module will start/stop recording.
Start : The recording will start with the trigger source conditions after
pressing the [START/STOP] key. The recording will stop after pressing the [START/STOP key.
Stop : The recording will start after pressing the [START/STOP] key
and will be stopped with the trigger source conditions.
T RIG GE R]             3/ 9 TR IG  Se tti ng :Of f▽       TR IG  So urc e: Off ▽       
    ↑      1 s       2 s       3 s       6 s       1 0s         ↓    
AMP Input setting
Mode REF. TEMP
DATA recording condition setting
Sampling
Capture MODE Capture DIST
TRIGGER capture condition settings
TRIG setting TRIG Source
If the submenu shows then there are se lections in those directions.
REF. TEMP, REF. TEMP Value setting
500 ms, 1, 2, 5, 10, 20, 30 s, 1, 2, 5, 10, 20, 30, 60min CONT, 1 Hour, 24 Hour Memory, SD card
Off, Start, Stop Off Alarm Date Date, Time
3 Recording
(1) Recording
Press the [START/STOP] key to start measuring with the set conditions.
After pressing [START] key, when the module is in awaiting recording start, “ARMED” is displayed, and then when recording is started, "REC" is displayed.
RE C.     ALM .        1 :2 8
TE MP :        +2 6. 7℃ HU M. :        +3 9. 8%
DE WP :      + 12 .1 8℃
When battery replacement is required, “BAT” is displayed.
* You can switch to the Accumulation screen with the and keys when recording data.
The module’s status is shown with the lamp display.
BA T L AN  SD      S : 1 .0 s
LAN: displayed when the wireless LAN connection is enabled.
STATUS (Orange)
Accessing SD card Low battery Alarm active
POWER(Green)
Power supplying Wireless LAN connection possible status
When accessing an SD card, do not remove the SD card. The
data may not write properly or the SD card may be damaged.
When “low battery” is displayed, replace the battery or
connect the USB interface to supply power as soon as possible. Caution: Batteries cannot be replaced when recording data. Replace them after the recording has finished.
(2) Recording completion
Press the [START/STOP] key to stop measuring.
The screen display will change to the standby screen display.
Press [ENTER] key to change to the free-running screen display.
When alarm occurs, “ALM” is displayed.
Current time Note: The current time display can
be switched to the elapsed time with the [QUIT] key when recording.
“SD” is displayed during accessing the SD card.
Sampling interval
Access light Flash once every 5 seconds Flash once every 10 seconds
Flash once every 10 seconds Flash once every 5 seconds
GL 100 -T H                
 S lee pi ng! !                E NTE R  key  t o s tar t   
QU IT  ke y t o  Pow er  Of f
BA T L AN                    
4 How To Confirm The Data
Check the recorded data with the application software included with this module using the method below (for details, refer to the USER'S MANUAL).
(1) Connect the USB interface and check the online data (2) Insert the SD card into PC and check the data directly (3) Check the data directly from PC via wireless LAN
5 Specifications
Item
Measurement data
Measurement system Measurement temperature range Measured temperature accuracy
Measurement humidity range Measured humidity accuracy
Response time
Sampling interval
Alarm Waterproof treatment Temperature unit Usage environment
External dimensions [W×D×H] (approximate) Weight (approximate)
* We recommend replacing the module periodically.
Temperature, humidity, dew point temperature (calculated value), accumulated temperature *
The accumulated temperature is displayed only when recording.
C-MOS Sensor
-20℃ to 85℃
-20 ≤ TS < 0 ±0.8 (℃) 0 ≤ TS ≤ 60 ±0.5 (℃) 60 < TS ≤ 85 ±0.8 (℃)
0.0 to 100.0% RH
25℃ 0 ≤ RH < 10 ±10 (%) 10 ≤ RH < 20 ±8 (%) 20 ≤ RH ≤ 80 ±5 (%) 80 < RH ≤ 90 ±8 (%) 90 < RH ≤ 100 ±10 (%)
Temperature and humidity: 100 sec. (63% response, 25°C Air flow rate 1 m/s) * The sensor’s responsiveness is affected by the measured
temperature and air flow rate.
0.5, 1, 2, 5, 10, 20, 30 sec. 1, 2, 5, 10, 20, 30, 60 min. OFF / Level Sensor unit With waterproof filter Select from °C (Celsius) / °F (Fahrenheit) Within the measurement range of temperature and humidity However, set it to the environment if the Gl100 and this sensor are in the same environment. 15 × 45 × 10.2 mm (not including protruding parts)
14 g
Contents
Branch adapter for GS (GS-DPA)
GL100 main module
(1) Screen display menu flow
After connecting the power supply, connect this module and operate it in accordance with the content displayed on the screen.
* Refer to “2 How To Measure” above and the USER'S MANUAL for the
illumination / ultraviolet sensor.
(2) Free-running screen
ST OP     ALM .        1 :2 8
TE MP :      + 26 .1 6℃ HU M. :        +3 7. 3% IL LU M:          44 0 
                                  m 
UV -A :        0. 03 0 
BA T L AN  SD      S : 1 .0 s
              lx
                 W
Hold down the [QUIT] key (approx. three seconds) to put the module into standby state.
When running on batteries, the module will automatically go into standby state after three minutes of no operation.
Press the [ENTER] key while in standby state to return to the free-running screen.
2. Combined temperature and humidity sensor and CO2 sensor measurement
Composite measurement can be done by using the branch adapter for GS (GS-DPA) and the CO2 sensor (GS-CO2) (each sold separately).
If the CO2 sensor (GS-CO2) is included in the assembly, it cannot be powered with batteries.
CO2 Sensor
(GS-CO2)
Branch adapter for GS (GS-DPA)
* The Gl100’s power is
provided through a USB cable connection
GL100 main module
(1) Screen display menu flow
After connecting the power supply, connect this module and operate it in accordance with the content displayed on the screen.
* Refer to “2 How To Measure” above and the USER'S MANUAL for the
CO2 sensor.
(2) Free-running screen
ST OP     ALM .        1 :2 8
TE MP :      + 26 .1 6℃ HU M. :        +3 7. 3% DE WP :      + 10 .3 0℃
                                  PP
CO 2:             73 6 
                 m
BA T L AN  SD      S : 1 .0 s
< Extension cable > The module can be used approx. 1.5 m away from the GL100 by using an extension cable for GS (GS-EXC). However, you cannot connect and use multiple extension cables.
Hold down the [QUIT] key (approx. three seconds) to put the module into standby state.
Press the [ENTER] key while in standby state to return to the free-running screen.
CAUTION
It is not possible to be used by connecting two same sensors.