Graphtec GL820 Specifications

Comprehensive Data Logger
20 channel Portable Logger
8
PulseHumidityTemp. LogicVoltage
20CH
Fast
Sampling
10ms
Wide LCD
5.7inch
20 insulated channels, each with multifunction input capacity
Contains an insulated input system which ensures signals do not cause any distortions among other channels. Suitable for recording voltage, temperature, humidity, pulse, and logic signals.
20mV to 50V wide range
Supported 9 type of thermocouples: R, S, B, K, E, T, J, N, W
Measured the humidity easily using the optional sensor
Supports 4 channels pulse or logic digital I/O using optional cable B-513.
Maximum 10ms Sampling 5.7 inch TFT colour LCD
Provides faster sampling rates for voltage
measurements.
Achieves up to 10ms sampling interval when
limiting the number of channels in use.
Utilises an industry leading 5.7-inch TFT color LCD monitor Providing enhanced ability to read data in waveform and digital forms to measure and analyze data in real time.
Built-in 2GB Flash memory for reliable long term measurement
The 2GB Flash Memory enables a securelong term data measurement
usage without applying an external storage device.
Supports USB 2.0 memory sticks for external storage.
The GL820 saves measured data directly on to the USB
memory sticks. The device can be replaced during
measurement without loosing any data.
Ring memory function
The most recent data is saved when internal
(VGA: 640 x 480 dots) .
memory or external memory is configured in ring memory mode.
Built-in
memory
2GB
Ring
memory
WWW.GRAPHTECAMERICA.COM/INSTRUMENTS
GL820 main unit specifications
Item Number of analog input channels
External input output
Sampling interval Time scale Trigger
function
Alarm function
Pulse input function
Input Output
Action Source
Combination Condition
Detecting method Condition
Alarm output Accumulating count
*8
mode
*8
*8
*8
Instant count mode
Rotation count (RPM) mode
Max. input pulse rate
Calculation function
Search function
Interface to PC Storage device
Data saving function
Ring capturing mode
USB memory device emulation Engineering scale function
Display
Operating environment
Between channels Statistical
Captured data Others
Size Formats
Power source
Power consumption
Weight
Software specifications
Item Supported OS
Functions GL820 settings control Controlled units Captured data Displayed information Display modes
Warning functions File format conversions Report functions
Description Windows XP / Vista / 7 (32 bits and 64 bits edition) Control GL820, Real-time data capture, Replay data, Data format conversion Input settings, Memory settings, Alarm settings, Trigger settings
Up to 10 units or 500 channels Transfers data in real-time (in binary or CSV format), saved data in GL820 or the USB memory Analog waveforms, Logic waveforms, Pulse waveforms, Digital values Y-T waveforms, Digital values, Report, X-Y graph (specified period of data, data reply only) Sends E-mail to the specified address when the alarms occur Converts the specified period data or all data to the CSV format (thinning function is available) Creates the daily or monthly report automatically (can also export directly to Excel)
Description
20 ch, Expandable up to 200 ch by unit of 20 ch Trigger or Sampling input 1 ch, Logic or Pulse input 4 ch
Alarm output 4 ch 10 ms to 1 h (in 10ms to 50ms, voltage only and limited channel), External 1 sec to 24 hour /division Start or stop capturing data by the triggerr Start: Off, Input signal, Alarm, External Stop: Off, Input signal, Alarm, External OR or AND condition at the level of signal or edge of signal Analog: Rising, Falling, Window-in, Window-out Pulse: Rising, Falling, Window-in, Window-out
Logic: Rising, Falling Level or edge of signal Analog: Rising, Falling, Window-in, Window-out Pulse: Rising, Falling, Window-in, Window-out Logic: Rising, Falling
4 channels, Output type: Open collector (pulled-up to 5 V by resistor 10 kΩ) Accumulating the number of pulses from the start of measurement
Range: 50, 500, 5 k, 50 k, 500 k, 5 M, 50 M, 500 M counts/F.S.
Counting the number of pulses per sampling interval Range: 50, 500, 5 k, 50 k, 500 k, 5 M, 50 M, 500 M counts/F.S.
Counting the number of pulses per second and then it is converted to RPM Range: 50 rpm, 500 rpm, 5 krpm, 50 krpm, 500 krpm, 5 Mrpm, 50 Mrpm, 500 Mrpm /F.S.
50 k pulses/sec or 50k counts per sampling interval (16 bits counter is used) Addition, Subtraction, Multiplication and Division for analog input Select two calculations from Average, Peak, Max., Min., RMS Search for analog signal levels, values of logic or pulse or alarm point
in captured data
Ethernet (10 BASE-T/100 BASE-TX), USB (Full speed) Built-in Flash memory (2 giga-bytes), USB memory device Direct saving of data into built-in Flash memory or USB memory device Setting conditions, Screen copy Function: ON/OFF, Number of capturing point: 1000 to 2000000 (size of the
capture data will be limited to 1/3 of available memory when in Ring Mode)
*8
, Clock, Week or Time
*8
, Clock, Week or Time
*9
USB Memory emulation mode (Transfer or delete the file in built-in memory)
Set based on the reference point of the scaled output and input signal for each channel (Voltage measurement: four points are necessary to scale the output, Temperature measurement: two points are necessary to scale the output).
5.7 inch TFT color LCD (VGA: 640 x 480 dots) Waveform + Digital, Waveform only, Calculation + Digital, Expanded digital 0 to 45 °C, 5 to 85 %RH
(When operating with battery pack 0 to 40 °C, charging battery 15 to 35 °C)
AC adapter (100 to 240 V, 50/60 Hz), DC power (8.5 to 24 V DC, max. 26.4 V)
32 VA or lower (when operating with AC adapter, displaying LCD, charging battery pack)
approx. 232 x 152 x 50 mm approx. 900 g (Excluding AC adapter and battery pack)
*10
, Battery pack
*10
Analog input specifications
Item Type of input terminal Input method Measurement
range
Voltage Temperature
Humidity
Filter Measurement
accuracy
Voltage
*11
Tempe-
Thermocouple
rature
R/S
B
K
E
T
J
N
W
RTD Pt100 JPt100
A/D Converter Maximum
input voltage
Withstand voltage
*8
: Logic alarm cable (B-513) option is required.
Input signal of External sampling, Logic, Pulse; Maximum voltage: 24 V, Threshold: approx. 2.5 V, Hysteresis: approx. 0.5 V
*9 : Size of the USB memory device is unlimited. Maximum file size is limited to 2GB.
*10: DC drive cable (B-514) or battery pack (B-517) option is required.
*11: Subject to the following conditions;
• Room Temperature is 23°C ±5°C.
• When 30 minutes or more have elapsed after power was turned on.
• Filter is set to 10.
• Sampling rate is set to 1s with 20 channels.
• GND terminal is connected to ground.
Logic alarm cable (B-513)
20ch extension terminal set (B-538)
*12
: Operating environment: -25°C to 80°C
Pt1000
Between + / - terminal Between channels Between channel / GND
Between channels 350 V p-p (1 minute) Between channel(-)/ GND
DC drive cable (B-514)
Description Screw terminal (M3 screw) Scans by the photo-MOS-relay, all channels isolated, balanced input
20, 50, 100, 200, 500 mV, 1, 2, 5, 10, 20, 50 V, and 1-5 V/F.S. Thermocouple: K, J, E, T, R, S, B, N, and W (WRe5-26) Resistance Temperature Detectors (RTDs): Pt100, JPt100(JIS), Pt1000(IEC751) 0 to 100%
(using humidity sensor (B-530 optional), power is supplied to only one sensor)
Off, 2, 5, 10, 20, 40 (moving average in selected number)
0.1 % of F.S. Measurement range
0 °C TS 100 °C
100 °C < TS 300 °C R: 300 °C < TS 1600 °C S: 300 °C < TS 1760 °C
400 °C TS 600 °C
600 °C < TS 1820 °C
-200 °C TS -100 °C
-100 °C < TS 1370 °C
-200 °C TS -100 °C
-100 °C < TS 800 °C
-200 °C TS -100 °C
-100 °C < TS 400 °C
-200 °C TS -100 °C
-100 °C < TS 100 °C 100 °C < TS 1100 °C
0 °C TS 1300 °C 0 °C TS 2000 °C
Reference Junction Compensation (R.J.C.): ±0.5 °C
Measurement range
-200 °C to 850 °C (FS = 1050 °C)
-200 °C to 500 °C (FS = 700 °C)
-200 °C to 500 °C (FS = 700 °C) ΣΔ type, 16 bits (effective resolution: 1/40000 of measuring full range) 60 V p-p 60 V p-p
60 V p-p
Measurement accuracy
± 5.2 °C ± 3.0 °C ± (0.05 % of reading + 2.0 °C) ± (0.05 % of reading + 2.0 °C)
± 3.5 °C ± (0.05 % of reading + 2.0 °C)
± (0.05 % of reading + 2.0 °C) ± (0.05 % of reading + 1.0 °C) ± (0.05 % of reading + 2.0 °C) ± (0.05 % of reading + 1.0 °C)
± (0.1 % of reading + 1.5 °C) ± (0.1 % of reading + 0.5 °C)
± 2.7 °C ± 1.7 °C ± (0.05 % of reading + 1.0 °C) ± (0.1 % of reading + 1.0 °C) ± (0.1 % of reading + 1.5 °C)
Measurement accuracy ±1.0 °C ±0.8 °C
±0.8 °C
350 V p-p (1 minute)
Battery pack (B-517)
Humidity sensor
12
(B-530) *
Extension terminal base kit (B-537)
Standard accessories
Item
AC adapter CD-ROM
Quick Start Guide
Options and accessories
Item
Logic alarm cable DC drive cable Battery pack Humidity sensor
Extension terminal base kit 20 ch extension terminal set
Description 100 to 240 V AC, 50 / 60 Hz (with specified type of power cord) User’s manual (PDF format), Application software
Model number B-513 B-514 B-517
*12
B-530
B-537 B-538
17462 Armstrong Avenue
Irvine, CA 92614
Remarks 2 m long (no clip on end of cable) 2 m long (no clip on end of cable)
1 piece (7.4 V 2200 mAh, 17Wh)
3 m long (with power plug) Terminal base, cable Terminal base, terminal unit (20 ch), fixing plate