Note! For work with other type tubes (e.g. with different outer diameter, bottom shape or different
material [transparent plastic tubes]) it is necessary to perform calibration in the specified
tubes.
Perform calibration
3.4.1.
3.4.2. Switch ON the unit with the Power switch (Fig.2/ ) on the rear panel (position I).
, not more than 2 mm diameter
point is saved in the memory and the
next calibration value is displayed .
in the following sequence from the lower calibration value to the higher
values: 0.0, 0.5, 1.0, 2.0, 3.0, 4.0, 5.0, 6.0. The minimum requirement is to calibrate 2 points.
Attention! Ensure that the tube socket is empty!
If external power supply is used, connect it to the mains.
3.4.3. With a pin press Select button (fig.2/ ) on the backside of the unit (for Select and Install
buttons use thin pin ). The display indicates flashing figure "- - -"
the unit is ready to save calibration value of the first point - empty socket.
3.4.4. With a pin press Install button (fig.2/ ), “Empty socket”
(flashing figure "0.0”)
Note! Shake the tube with the corresponding standard, if necessary (it is recommended to
use vortex, e.g. Vortexer PV-1, for shaking).
3.4.5. Insert the tube with the calibration value corresponding standard into the socket (fig.3/ ) of
Densitometer.
3.4.6. With a pin press Install button. This point of calibration sequence is saved in the memory and
the next calibration value is displayed.
Note! It is recommended to calibrate 0 value and as many points as possible to obtain precise
results. The minimum requirement is to calibrate 2 points closest to working range (e.g.
0 and 6.0 for operation in 0 - 6.0 McF unit range).
3.4.7. Repeat steps 3.4.5.-3.4.6. until the calibration is complete (1-7 times, i.e. as many times as
many points the chosen calibration sequence has).
3.4.8. If a standard is not available, press Select button for the next calibration value to be displayed.
3.4.9. After installing the last standard value “6.0”, or skipping it (Select button), the unit enters
operation mode. The unit is ready for operation.
Note! If during calibration pressing of Install button does not cause switch to the next Standard
value, it means that the inserted in the densitometer socket Standard has lower turbidity
value than the previous Standard. Cause: turbidity of the inserted Standard does not
correspond to the necessary value (Standard has to be shaken or replaced).
3.4.10. At the end of calibration switch OFF the unit with the Power switch (position O). If external
power supply is used, disconnect the external power supply from the mains.
Fig.2 Rear panel
Power DC-12
Calibration
select
install
Page 9
DEN-1B
Operating instructions
Version 1.01 - June 2012