Grainger 4802 Operating And Service Manual

SCALE-TRONIX
4802
PEDIATRIC SCALE
SECOND GENERATION WEIGHING SYSTEM
OPERATING AND SERVICE MANUAL
(SERIAL NUMBERS 21000 AND UP)
NOW WITH DISPOSABLE DURACELL BATTERIES
ECO-FRIENDLY DESIGN
SPECIFICATIONS: 4802-PEDIATRIC SCALE
GENERAL
SCALE-TRONIX® 4802 DIGITAL PEDIATRIC SCALE: Completely self-contained scale; accurately
weighs infants and toddlers. The scale is contained in a portable, heavy-duty
stainless steel enclosure, with integral electronic digital readout and a sturdy transparent acrylic
weighing cradle with length gauge. The patient’s
weight is automatically locked-in and displayed during the weighing process, and simultaneously stored within the scale’s memory for later recall. This feature allows the infant to be removed from the scale prior to charting the weight. A sealed membrane pushbutton assembly provides the operator with a choice of pounds/ounces or kilograms for the weight display. Other front panel pushbuttons provide on/off/zero, reweigh and recall of the previous patient’s weight.
The scale can be operated directly from AC line power, or when portability is desired, from internal batteries. Features include automatic shut-off when operating from battery to conserve battery life; continuous operation when powered from AC line; and automatic switching from batteries to line power. Sturdy handles provide for easy transport and cradle protection.
WEIGHT DISPLAY Large, bright, 0.43” light-emitting diode display.
Easily read in all lighting conditions. Selectable weight display in grams, kilograms, pounds, ounces or pounds/ounces. Weight units can be restricted and locked out by hospital. Indicator light to show selected units. Previous weight held in memory for later recall.
WEIGHING CAPACITY 10-20,000 GRAMS (20KG), 44 lbs.
ACCURACY
5 grams
OPTIONS
Diaper option provides 1 gram resolution for weights less than 1 KG.
Manual part No. 070060 (G), revised July 2014.
UNIVERSAL POWER SUPPLY
Medical grade UL listed external power supply. 100-240 VAC 50/60 Hz. Captive but replaceable line cord.
Six (6) “DURACELL” D size alkaline batteries. With “LOW BATTERY” warning indicators.
WEIGHING CRADLE
Removable molded acrylic cradle. Smooth weighing surface provides easy cleaning and sanitary maintenance. Length measurement in centimeters and inches or centimeters only. OPTIONAL: Extra-long or four-sided cradles.
DIMENSIONS Overall: 27-1/2” x 16 x 7 inches WEIGHT (uncrated) 25 pounds
ACCESSORIES
Matching stainless steel cart for mobile use. Disposable cradle liners.
WARRANTY
One (1) year, parts and labor.
TESTING CERTIFICATION STANDARDS
Meets UL544 standard. Meets Canadian Standards CSA C22.2 No. 125. Constructed to International Electrotechnical Commission IEC-601-1.
FACTORY AND SERVICE
P.O. Box 15 Wheaton, IL 60189 1-800-879-2001 1-630-653-3377
MARKETING AND SALES
200 East Post Road White Plains, NY 10601 1-800-873-2001 1-914-948-8117
Internet: www.scale-tronix.com
PEDIATRIC WEIGHING SYSTEM
MODEL 4802
SECTION I GENERAL DESCRIPTION…………………………………….1 SECTION II ASSEMBLY – SET UP ………………………………………...3 SECTION III OPERATION OF SCALE ………………………………………7 SECTION IV TECHNICAL DESCRIPTION …………………………………12 SECTION V PARTS LIST …………………………………………………….20 SECTION VI WARRANTY……………………………………………………..24
This User’s Manual is intended to provide all necessary information and data for the proper
operation and service of the model 4802 PEDIATRIC SCALE with serial numbers 21000 and above. This manual should be studied before installing or placing the scale in service and kept
in a safe, handy place for future reference. Refer to the scale’s serial and model numbers in all
correspondence. Feel free to call for service and operational assistance.
C
B
D
E
F
G
A
E
ILLUSTRATION 1
A. Weighing cradle with length gauge. B. Heavy duty Stainless Steel Cabinet. C. Front panel with membrane pushbuttons D. Digital display E. Carrying handles F. Rear handle and cord holder G. Line cord, power supply, and plug (on rear panel).
SECTION I
GENERAL DESCRIPTION
1.0 GENERAL Congratulations! You have just purchased the only
truly modern, full-featured Pediatric scale. Introduced in 1979, The Scale-Tronix 4800 was the
preferred pediatric scale for hospitals due to its accuracy, reliability and rugged stainless steel
construction. Years later, we updated the scale’s
electronics to make it even more user friendly and called the 4802. We added the 2 to denote the second generation of our great scale. Currently, most of these scales are still in everyday use. Scale-Tronix has not stopped innovating and trying to incorporate all of the great features and durability of our previous models to meet the challenges of our modern world. We now introduce the new improved 4802 with the significant improvement of using disposable long life alkaline “D” size flashlight batteries.
The 4802 can be operated from AC line power and left on continuously for fast convenient weights; or it can be transported on a cart, operation from its self-contained “D” batteries for true portable
operation. The scale’s internal microcomputer will
automatically keep the scale “on” when plugged in
and time out after a period of inactivity when
operating from the batteries. Successive weighing’s
will automatically reset the time and keep the scale turned on as long as required.
We have replaced the rechargeable battery with standard heavy duty disposable DURACELL alkaline D size flashlight batteries which provides several thousand weighing’s (DURACELL only). A flashing low battery annunciator provides adequate warning to change batteries soon.
The clear acrylic weighing cradle readily unplugs from the scale for easy clean-up or exchange. A variety of cradle sizes and configurations are available to suit your weighing needs, including four-sided and oversized cradles. A length gauge, marked in both inches and centimeters or centimeters only, is conveniently provided on top of the cradle for
measuring the infant’s length during the weighing
procedure. Disposable paper cradle liners are available to reduce clean-up and prevent cross-contamination.
The scale’s enclosure is constructed of heavy gauge stainless steel for durability and minimum maintenance. Carrying handles are provided on both sides of the enclosure for safe, convenient movement. In addition to serving as a handle, the rear handle serves as a line cord holder, and along with the side handles, serve as a bumper to prevent damage or interference with the scale if it is pushed against a wall or other surface. Users appreciate these helpful safeguards and features.
A stainless steel weighing cart is available that makes the model 4802 PEDIATRIC SCALE a self-contained mobile weighing station. The cart also features a dispenser for disposable scale liners.
The scale employ the latest in microcomputer control combined with user friendly software. This provides the operator with an accurate, rugged and easy-to-use scale. Advanced Digital Signal Processing (DSP) provides accurate data at all times.
Audible response by a series of beeps provides excellent operator feedback for ease of operation. The beeper can be disabled if required.
An internal non-volatile memory stores the last
patient’s weight for later recall and charting. User
programmable features allow the operator to select automatic shut-off times, weight resolution, weight units, etc.
A Data Port provides serial data output to a printer, personal computer or EMR data collection system. Consult the factory for further details.
1.10 MECHANICAL FEATURES:
A summary of the mechanical features follows:
A. Heavy gauge stainless steel enclosure for
durability and low maintenance. Angled front panel to maximize readability.
B. Side handles for carrying convenience and
cradle protection.
C. Rear handle for carrying, cord storage and
scale protection.
1
D. Easily detachable acrylic cradle with length
gauge. Variety of cradle sizes/configurations available.
E. Optional matching cart turns scale into a
mobile weighting station.
1.20 ELECTRONIC FEATURES:
A summary of the electronic features follows:
A. Microcomputer equipped electronics with
digital signal processing.
B. Sealed membrane panel with touch
pushbuttons.
C. Large, bright, light-emitting diode display with
function annunciators.
D. Pushbutton control of turn-on, turn-off.
E. Pushbutton selection of weight units.
Selection of weight units retained in memory for subsequent use. Pounds or kilograms can be locked out to suit requirements.
F. Automatic weight lock-in and display.
Reweigh function allows confirmation of weight without removing patient from scale. Weight stored in non-volatile memory for subsequent recall and display.
G. Pushbutton zero to remove weight of cradle
and blankets, etc.
H. Operation from AC line or internal batteries;
automatically selected. Automatic shut-off to conserve battery life. “LOW BATTERY” front panel annunciator.
I. User programmable “set-up” feature allows
selection of weight units, resolution, shut-off time, beeper, etc. Pounds or kilograms can be locked out to suit hospital requirements.
J. Universal to 100 to 240 volt power supply
with captive but readily replaceable AC line cord with IEC receptacle.
2
SECTION II
ASSEMBLY – SET UP
The SCALE-TRONIX® model 4802 PEDIATRIC SCALE is usually shipped with the scale and cradle in
a single carton. If the oversized cradle option was ordered, this cradle will be shipped in a separate carton. Please save the carton(s) and shipping materials for possible future use if the scale is ever to be returned to the factory for service or upgrades. Carefully inspect the carton(s) for shipping damage before unpacking. Contact the shipping company and SCALE-TRONIX® immediately if shipping damage is found. All claims must immediately be made directly with SCALE-TRONIX®.
2.10 ASSEMBLY INSTRUCTIONS (Refer to illustration 1)
The SCALE-TRONIX® model 4802 PEDIATRIC SCALE is fully assembled at the factory and tested.
The scale is shipped with the weighing cradle removed. This insures protection of the delicate weighing transducers.
Carefully remove the acrylic weighing cradle from the shipping carton. Remove the scale from the carton and place it on a counter top or stable cart. Position the cradle over the scale so the four metal mounting posts line up with the holes in the top of the scale. Note the length gauge and rotate the cradle if
necessary so the length gauge is “right reading”.
Apply a gentle downward pressure to “plug-in” all four posts so the cradle rests firmly on the scale.
2.20 CHECK-OUT
1. Turn the scale on by briefly pressing the “ON / (OFF) / ZERO” pushbutton on the
front panel. A single “beep” will be heard.
The scale should display “888888” and illuminate the “KILOS’, “POUNDS”,
“OUNCES”, “PRIOR WEIGHT” and “LOW BATTERY” annunciators as a test of their
operation. Observe these digits and annunciators to insure they light. The scale will then briefly display its model number, “4802”. During this short time period the scale also performs some internal checks of its operation.
2. After a short period of settling, the scale will
then automatically establish its “zero” value,
removing the weight of the cradle and any miscellaneous items left in the cradle. The display will show “0.000 KILOS” or “0 POUNDS O.O OUNCES” depending on its previous mode. A series of two beeps will be heard signaling the setting of zero.
3. Place an object to be weighed on the cradle. A one or two liter bag of IV solution will do. A series of moving dashes (“-----“) will be noted on the display indicating motion of the cradle.
4. The scale will calculate and display the value of weight on the cradle and signal it by sounding a series of three beeps. The weight will be displayed in either pounds/ounces or kilograms depending on the currently selected mode. To obtain the weight in the other units press the ‘KILOS / POUNDS” pushbutton on the front panel. Note the appropriate annunciator will illuminate as the weight is displayed. Keep in mind that if the hospital has a requirement of locking out pounds (or kilograms) that this switch will not be functional.
5. Place an additional small object on the cradle. Press the ‘REWEIGH” pushbutton on the front panel. The scale will calculate the new weight and re-display it as it sounds a series of three beeps.
6. If your scale was purchased with the optional “DIAPER WEIGHING CAPABILITY perform steps 6, 7, and 8; otherwise skip to step 9.
Obtain a stack of about 25 standard sized (2”
x 3-1/2”) business cards to use as test weights. Insure that the scale is properly placed on a steady surface and that they are not any fans or overhead vents blowing on the cradle (with measurements as sensitive as these, small breezes can affect the weight readings!) Remove all items from the cradle and press the “ON / (OFF) / ZERO” pushbutton on the front panel to obtain a zero reading. If necessary, set the weight mode to
“KILOS” by pressing the “KILOS / POUNDS” pushbutton.
7. Count out fifteen (15) of the business cards and place them on the cradle. Note the
reading on the scale’s display when it
3
stabilizes. Normally this will produce a reading of about 15 grams for typical small business cards. (Since the cards can vary in weight, this is only an approximation.)
8. Now toss an additional business card on the cradle and press the REWEIGH” pushbutton. This will normally produce an increase in the weight reading of 1 gram. Repeat this process with additional cards, pressing the “REWEIGH” pushbutton each time and noting that the reading increases in steps of 1 gram. (Note: the exact weight of the cards may cause the reading to vary by 0, 1 or 2 grams.)
9. When powered from an AC line the scale will remain on continuously. Press and hold the “ON / (OFF) / ZERO” pushbutton on the front panel until the scale turns off. Remove all objects from the cradle. Unplug the scale from the AC line. Turn the scale back on by briefly pressing the “ON / (OFF) / ZERO” on the front panel.
10. After the self-test and zero process, the scale will return to displaying “0.000 KILOS” or “0 POUNDS 0.0 OUNCES” depending on its previous mode. Press the “RECALL PRIOR WEIGHT” pushbutton on the front panel. The weight value from the previous weighing, which was automatically stored in the scales memory, will be displayed along with a flashing PRIOR WEIGHT” annunciator. The weight may be displayed in either kilograms or pounds/ounces by pressing the “KILOS/POUNDS” pushbutton (unless locked out).
11. Leave the scale on and determine that it shuts off after a period of about 90 seconds. (Pressing any of the pushbuttons will reset the time and start the timing period over.) If the timing period is different from the specified 90 seconds see section 2.40 “OPTION SET-UP”.
12. If the “LOW BATTERY” annunciator flashes during the process, it indicates the batteries need replacing. Replace with size D “DURACELL” alkaline batteries only.
13. If the optional cart has been purchased follow the assembly instructions included with it. The model 4802 PEDIATRIC SCALE may then be mounted on it.
14. This completes the assembly and check-out procedure.
2.30 CALIBRATION The SCALE-TRONIX® model 4802 PEDIATRIC
SCALE is fully tested and calibrated before leaving
the factory. No further calibration is normally required. The calibration procedure is outlined in section 4.40.
2.40 OPTION CUSTOM “SET-UP” The SCALE-TRONIX® model 4802 PEDIATRIC
SCALE incorporates a unique feature that lets you
“customize” the scale to best suit your needs. This
“set-up” mode allows the characteristics of the scale
to match your particular requirements. Items that can be selected include the automatic shut-off time, weighting units, beeper, etc.
2.41 ENTERING THE “SET-UP” MODE
To enter the “set-up” mode, start with the scale turned off. Note the “Print/Send/Data” button located directly to the right of the “RECALL PRIOR WEIGHT” pushbutton on the front panel of the enclosure. Press and hold this “Print/Send/Data” pushbutton on the front panel while turning the power on with the “ON / (OFF) / ZERO” pushbutton. Once
the scale turns on, release the “Print/Send/Data”
pushbutton, then press and release it 5 more times. The scale will display “SEt-UP” and produce a series of four long beeps. “Set-up” mode has now been entered.
The “RECALL PRIOR WEIGHT” pushbutton is used to select the particular option value. Once this value is selected, the scale can be advanced to the next option by again pressing the “Print/Send/Data” pushbutton or it can be shut-off by pressing and holding the “ON / (OFF) / ZERO” pushbutton.
2.42 SETTING THE OPTIONS
First enter the “set-up” mode as explained above.
1. Press the “Print/Send/Data” pushbutton. The scale will display “SOFt” (software version). Press the “Print/Send/Data” pushbutton again. The software version will be display (i.e. “U 1.00.4”). This is provided as a reference aid in determining which features are contained within the software.
2. Press the “Print/Send/Data” pushbutton.
The scale will display “dAtE” (date). Again
4
press the “Print/Send/Data” pushbutton. The release date of the software will be displayed. A display of “07.15.97” would correspond to July 15, 1997. This is provided as a reference aid in determining which features are contained within the software.
3. Press the “Print/Send/Data” pushbutton. The scale will display “SCALE”. Press the “Print/Send/Data” pushbutton again; the model number of the scale will be displayed “4802”. This is provided as a reference aid to help with customer trouble-shooting and is set at the factory.
4. Press the “Print/Send/Data” pushbutton. The scale will display “AutOFF” (automatic shut-off time). Again press the “Print/Send/Data” pushbutton. The currently selected number of seconds before the scale shuts-off will be displayed. This is normally set at the factory to be 90 seconds. It can be altered in 15 second increments from 15 seconds to 180 seconds (3 minutes). An additional “Cont” (continuous) option is provided for some special applications; this prevents the scale from automatically shutting-off; the scale can then only be turned off by pressing and holding the “ON / (OFF) / ZERO” pushbutton. Use of the continuous mode of operation is not recommended since it will discharge the battery if the operator forgets to deliberately turn the scale “off”. To alter the length of the automatic shut-off time press the “RECALL PRIOR WEIGHT” pushbutton. The automatic shut-off time will advance through the available options. Stop at the desired time interval.
5. Press the “Print/Send/Data” pushbutton. The scale will display “AC con” (AC power continuous operation). This feature is used in the model 4802 PEDIATRIC SCALE to cause the scale to remain “on” as long as AC power is provided (automatic shut-off will not occur. Pressing the “Print/Send/Data” pushbutton again will show this option selected as “On” or “OFF”. To alter the selection, press the “RECALL PRIOR WEIGHT” pushbutton.
6. Press “Print/Send/Data” pushbutton. The scale will display “UnitS” (weight units). This is the start of the various weight units (kilograms, pounds, ounces, pounds/ounces)
that can be selected. Use the “RECALL PRIOR WEIGHT” pushbutton to set each weight unit “On” or “OFF”. NOTE: If all the units are set to “OFF” the scale will automatically default to “KILOS”.
7. Press the “Print/Send/Data” pushbutton again. The “KILOS” annunciator will light and the display will show “On” or “OFF”. Setting this to “On” will allow display of the weight in kilograms with a display of “0.000”. Normal setting as shipped from the factory is “On”.
8. Press the “Print/Send/Data” pushbutton again. The “POUNDS” and “OUNCES” annunciators will light and the display will show “On” or “OFF”. Setting this to “On” will allow display of “0” (pounds) 0.0
(ounces)”. Normal setting as shipped from
the factory is “On”.
9. Press the “Print/Send/Data” pushbutton again. The “POUNDS” annunciator will light and the display will show “On” or “OFF”. Setting this to “On” will allow display of the weight in pounds with a display of “0.000”. Normal setting as shipped from the factory is "OFF”.
10. Press the “Print/Send/Data” pushbutton again. The “OUNCES” annunciator will light and the display will show “On” or “OFF”. Setting this to “On” will allow display of the weight in ounces with a display of “0.00”. Normal setting as shipped from the factory is “OFF”.
11. Press the “Print/Send/Data” pushbutton. The scale will display “bEEPEr” (audio beeper). Press the “Print/Send/Data” pushbutton again. “On” or “OFF” will be displayed. This may be changed by pressing the “RECALL PRIOR WEIGHT” pushbutton. Setting this option to “OFF” will eliminate the audible signal that occurs when a front panel pushbutton is pressed or a weight is locked in Normal setting as shipped from the factory is “On”.
12. Press the “Print/Send/Data” pushbutton. If equipped at the factory with the data port the scale will display “rS-232”, indicating the start of the serial data port options. Separate detailed instructions for this feature are available from the factory. Call us at
5
Loading...