Graham Field US0208PL, US0208, US0458 User Manual

Page 1
MODEL US0208 / US0208PL SEMI-ELECTRIC BED
MODEL US0458 / US0458PL FULL-ELECTRIC / LOW BED
USER MANUAL
READ THIS MANUAL BEFORE OPERATING THE PATRIOT™ HOMECARE BED.
SAVE THIS MANUAL FOR FUTURE USE.
CAN BE FOUND ONLINE AT www.grahameld.com
US0208-INS-LAB-RevD11
Page 2
CONTENTS
1 SAFETY ............................................................................................................................................................................................... 3
WARNING / CAUTION / NOTICE SUMMARY ............................................................................................................................. 3
2 INTRODUCTION, PATRIOT HOMECARE BED USER MANUAL ...................................................................................................... 4
THERMAL MOTOR PROTECTION .............................................................................................................................................. 4
GRAVITY-DOWN FEATURE FOR HEAD AND KNEE ELEVATION ............................................................................................ 4
3 ASSEMBLY INSTRUCTIONS .............................................................................................................................................................. 5
HEAD AND FOOT ASSEMBLY .................................................................................................................................................... 5
MOTOR ASSEMBLY .................................................................................................................................................................... 6
REMOVE MOTOR SLIDE CAPS INSTALL MOTORS
INSTALL MOTOR SLIDE CAPS ........................................................................................................................................... 7
MOTOR ASSEMBLY TROUBLESHOOTING ........................................................................................................................ 7
FULL-ELECTRIC BED ONLY (US0458 / US0458PL): CASTER POSITION .............................................................................. 8
STANDARD (CONVENTIONAL) BED HEIGHT LOW BED HEIGHT
BED END ASSEMBLY ................................................................................................................................................................. 9
ALL BED MODELS: ASSEMBLE BED HEAD END ............................................................................................................. 9
ALL MODELS: ASSEMBLE BED FOOT END ..................................................................................................................... 9
ALL MODELS: INSTALL BED END LOCKING HARDWARE FULL-ELECTRIC BED ONLY (US0458 / US0458PL): ASSEMBLE HI/LO MOTOR SHAFT TO FOOT END GEAR BOX
HI/LO ROD ASSEMBLY / BED LENGTH ................................................................................................................................... 13
SEMI-ELECTRIC BED ONLY (US0208 / US0208PL): ASSEMBLE HI/LO ROD ............................................................... 13
FULL-ELECTRIC BED ONLY (US0458 / US0458PL): ASSEMBLE HI/LO ROD ............................................................... 13
ALL MODELS: ATTACH HI/LO ROD TO GEARBOX OR MOTOR SHAFT PINS ............................................................. 14
4 BED OPERATION .............................................................................................................................................................................. 15
PENDANT HAND CONTROL OPERATION .............................................................................................................................. 15
SEMI-ELECTRIC BED ONLY (US0208 / US0208PL) ......................................................................................................... 15
FULL-ELECTRIC BED ONLY (US0458 / US0458PL) ........................................................................................................ 15
CRANK OPERATION: SEMI-ELECTRIC BED ONLY (US0208 / US0208PL) ........................................................................... 16
TRENDELENBURG POSITION, ALL MODELS ........................................................................................................................ 16
REVERSE TRENDELENBURG POSITION, ALL MODELS ...................................................................................................... 16
5 SPECIFICATIONS, ALL MODELS .................................................................................................................................................... 17
6 ACCESSORIES: BED RAILS ............................................................................................................................................................ 18
BED RAIL SAFETY .................................................................................................................................................................... 18
LUMEX PATRIOT LIBERTY BED RAILS ................................................................................................................................... 18
GF6500B-1: ASSIST BAR .................................................................................................................................................. 18
GF6580B-1: QUARTER RAIL ............................................................................................................................................. 19
GF6590B-1: HALF NO-GAP HEAD RAIL .......................................................................................................................... 19
GF6570B-1: FULL LENGTH RAIL ...................................................................................................................................... 19
LUMEX UNIVERSAL BED RAILS ............................................................................................................................................. 20
GF6570A-1: FULL LENGTH SIDE RAIL ............................................................................................................................ 20
LX5088: NO-GAP BED RAIL .............................................................................................................................................. 20
GF6580A-1: LOW UNIVERSAL HALF RAIL ...................................................................................................................... 21
7 SERVICE PARTS, ALL MODELS ..................................................................................................................................................... 22
BED COMPONENTS .................................................................................................................................................................. 22
BED END COMPONENTS BED SERVICE PARTS
8 MAINTENANCE AND SAFETY CHECKS ......................................................................................................................................... 25
9 WARRANTY ....................................................................................................................................................................................... 26
10 INDEX ................................................................................................................................................................................................ 27
............................................................................................................................................................... 6
............................................................................................................................................................... 8
......................................................................................................................................................... 23
............................................................................................................................................................... 24
.......................................................................................................................................... 6
.................................................................................................................... 8
............................................................................................ 10
12
GF Health Products, Inc. is not responsible for typographical errors.
Packaging, warranties, products and
specications are subject to change without notice.
Graham-Field and Lumex are registered trademarks, and Patriot is a trademark, of GF Health Products, Inc.
© 2011, GF Health Products, Inc.
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 2
Page 3
1 SAFETY
The safety statements presented in this chapter refer to the basic safety information that the op­erator of the Patriot Homecare Bed shall pay attention to and abide by. There are additional safety statements in other chapters or sections, which may be the same as or similar to the following, or specic to the operations. Please note the following special statements, used throughout this manual, and their signicance:
WARNING: Indicates a potential hazard situation or unsafe practice that, if not avoided,
could result in death or serious personal injury.
CAUTION: Indicates a potential hazard situation or unsafe practice that, if not avoided,
could result in moderate or minor personal injury.
s NOTICE: Indicates a potential hazard or unsafe practice that, if not avoided, could result in
product or property damage.
Info: Provides application recommendations or other useful information to ensure that you get the most from your product.
WARNING / CAUTION / NOTICE SUMMARY
WARNING: Keep hands and feet clear of all moving parts. WARNING: Do not allow small children on or near bed during operation. WARNING: Do not allow this device to be operated by small children. WARNING: When operating the hi/lo, knee, or back function of the bed, ALWAYS ensure that the
individual conned to the bed is positioned properly within the connes of the bed. DO NOT let any extremities protrude over the side or between the bed rails when performing any functions.
WARNING: Ensure that body weight is evenly distributed over the surface of the bed. Do not lie, sit
or lean in such a way that your entire body weight is placed only on the raised head or foot sections of the bed. This includes when assisting the user in repositioning or transferring in or out of bed.
WARNING: In order not to invalidate the warranty or create a hazardous condition resulting
in serious personal injury, GF Health Products, Inc. recommends the use of only Lumex replacement parts.
WARNING: Before operating, ensure that the bed frame rivets are properly seated in the bed
end hooks; otherwise, personal injury or damage could result.
WARNING: Keep all moving parts free of obstructions (i.e. blankets/sheets, heating blan-
kets/pads, tubing, wiring, and other types of products).
WARNING: Ensure that Bed End Locking Hardware is installed in order to prevent personal
injury when moving the bed or using a bed-mount trapeze.
WARNING: WARNING/CAUTION labels applied to the bed outline hazards or unsafe prac-
tices that could result in personal injury and/or property damage; do not remove them.
WARNING: Ensure that the power cord DOES NOT get tangled around the bed, bed rails or legs
during normal operation of the bed.
WARNING: Do not leave the bed unattended when plugged in.
s NOTICE: The pendant hand control cord must be routed and secured properly to ensure that
the cord does NOT become entangled and eventually severed during use.
s NOTICE: When using nasal or masked type administering equipment, oxygen or air tubing
MUST be routed and secured properly to ensure that tubing does NOT become entangled and eventually severed during normal operation of bed.
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 3
Page 4
2 INTRODUCTION, PATRIOT HOMECARE BED USER MANUAL
This manual contains assembly and maintenance instructions for your Model US0208 / US0208PL Semi-Electric and US0458 / US0458PL Full-Electric / Low Patriot bed. Read the entire manual care­fully before using your bed, and refer to it during use if you have questions.
THERMAL MOTOR PROTECTION
This product is powered by a quiet, efcient DC motor system. When the bed is plugged into a power outlet, the electrical transformer is always energized in a stand-by condition. The Okimat actuator casing may feel warm to the touch even when the bed is not being used; this is a normal condition.
Thermal protection activation (protection against overheating caused by excessive or extended periods of operation), which causes the bed to shut down operation when overheated, is a safety feature of this product.
Causes of thermal protection activation include multiple or repeated adjustment of long duration, or the use of multiple functions at one time. To ensure trouble-free operation, always allow a slight pause between multiple adjustments, and avoid operating more than one function at a time. If thermal protection activation occurs, the bed pendant hand control will not function. The thermal protection will reset after a period of time dependent on the severity of the overheating. To short­en the reset time, temporarily unplug the power cord from the power outlet.
GRAVITY-DOWN FEATURE FOR HEAD AND KNEE ELEVATION
Patriot beds are equipped with a gravity-down safety feature. The motors push the head and knee grid decks up, but do not pull the knee and head down. As the motor plunger retracts, the head and knee lower freely, minimizing bed damage and/or personal injury from items caught under the grid deck.
Info: When using telescoping deck-mount bed rails, the head and knee may not lower without weight on the bed.
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 4
Page 5
3 ASSEMBLY INSTRUCTIONS
HEAD AND FOOT ASSEMBLY
hook
foot
section
hook
Assembly, head and foot sections
Note: Grid deck not shown for clarity
rivet
rivet
motor cam
head
section
1. Place the head and foot sections on their sides at approximately right angles to each other with motor cams up, as shown above.
2. Slide the head and foot sections together until the foot section’s hook catches the head sec­tion’s rivet on both sides. Ensure both rivets are completely seated before continuing.
head wing
linkage plate
head
foot
straighten
Straighten sections
Note: Grid deck not shown for clarity
3. With the rivets fully seated in the hooks as described in step 2, rotate and straighten the bed as shown above. Position head wing as shown above.
s NOTICE: While balancing the frame, do not elevate the knee too far or the linkage may over-
extend and lock. If the linkage locks, manually push the large linkage plate toward the foot of the bed.
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 5
Page 6
MOTOR ASSEMBLY
WARNING: Assemble motor on bed as follows. While it is possible to
install the motor backwards, avoid it! Incorrect installation could cause the bed to move in unexpected ways. Note that the motor ends are labeled and embossed with head and foot symbols (shown at right). Ensure that bed head and motor head, and bed foot and motor foot, are matched.
Remove motor slide caps
slide cap
motor
Remove motor slide caps
1. Remove the slide caps from both ends of the motor as shown above. Plungers should be recessed in motor cavities; if not, plug the power cord into any grounded outlet and operate “down” func­tion on the pendant hand control, then unplug the power cord.
Head
Foot
Info: Motor comes from factory with pendant hand control plugged into motor.
2. Install 9V backup battery (not provided) and slide into storage area.
Install Motors
cam
head
foot
tap here
Install motor
3. Hold the bed securely, positioned with motor cams up, as shown above.
4. Position the motor over both cams.
5. Tap either end of the motor housing (shown below) with enough force to snap the motor as­sembly into position on the cam.
6. Repeat step 5 for other end of motor.
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 6
Page 7
Install motor slide caps
pressure
7. Position the caps as shown at right so that the grooves in the slide cap are aligned with the grooves in the motor.
slide cap
8. Apply straight downward pressure to each end of the cap while sliding into the closed position.
9. Repeat step 8 for the other cap.
10. Ensure that all four corners of each slide cap are engaged.
motor
Install motor slide caps
WARNING: Ensure that the motor is assem-
bled on bed as instructed above before use.
Motor assembly troubleshooting
Problems associated with improper motor installation are listed below, with appropriate x.
Problem Probable cause Fix
Head operates with knee control button and knee operates with head control button
Knee lifting linkages bind against grid deck, preventing knee deck from elevating.
Motor is installed backwards Remove motor and reverse it (see previ-
ous page)
Motor is installed backwards, enabling knee to elevate too high and bind linkage
Remove motor and reverse it (see previ­ous page)
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 7
Page 8
FULL-ELECTRIC BED ONLY (US0458 / US0458PL): CASTER POSITION
Any full-electric Patriot bed can be used as either a standard-height or low-height bed by moving the casters.
fig 1: casters mounted at
standard height
fig 2: casters mounted at low
height, bed in lowest position
fig 3: casters mounted at low
height, bed in elevated position
Standard (conventional) bed height
To achieve standard bed height, insert casters into outside legs, as shown in g. 1 above.
Low bed height
To achieve low bed height, insert casters into inner short legs, as shown in gs. 2 and 3 above.
When in the lowest position, the casters are engaged for bed movement, as shown in g. 2 above.
When elevated, the bed is off the casters and cannot be moved, as shown in g. 3 above.
Due to the clearances involved, only minor travel and movement are recommended in the low po­sition. The casters can be moved to the outer legs to provide clearance if the bed must be moved or rolled.
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 8
Page 9
BED END ASSEMBLY
All bed models: Assemble bed head end
1. Insert one locking and one non-locking caster into bed end. Locking casters must be installed diagonally from each other. See previous page for full-electric caster position options.
2. Standing directly in front of the bed end, position it as close to the bed as possible.
3. With one hand, reach over end and lift up the bed.
bed head end
frame (head)
Attach bed head end
4. Tilt the bed end slightly back toward you as shown above.
rivet
lock
Engage lower rivet
Lower bed onto lock
5. Engage lower rivet slightly with corner lock and tilt end toward bed as shown at above left.
6. Lower bed onto corner lock as shown at above right.
All models: Assemble bed foot end
After bed head end has been attached, perform steps 1-6 in All models: Assemble bed head end above to install the foot end.
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 9
Page 10
All models: Install bed end locking hardware
1. After bed head end has been attached, install Bed End Locking Hardware to act as a latch for each corner.
fender
washer
nylon
cord
Bed End Locking Hardware, part number 690-1000-100, includes four lanyards, each as-
clevis pin
with hole
hairpin clip
sembled from a clevis pin, hairpin clip, fender washer, and nylon cord, as shown at right. Bed
Lanyard
end locking hardware, while not necessary for normal bed operation, will prevent the bed ends from accidentally rolling off when mov­ing or pushing the bed, or while installing or using bed-mount trapezes.
WARNING: GF Health Products, Inc. recommends the use of Bed End Locking Hardware at
all times.
bed end
frame bracket
bed end bracket (corner lock)
a b c
lanyard — hairpin clip removed
2. Refer to accompanying illustrations while installing Bed End Locking Hardware. Remove the hairpin clip from the end of the lanyard assembly, as shown in a, and set it aside.
3. Align the cord end with the frame bracket top hole closest to the bed end. Insert the cord end through the hole so that an equal amount of cord extends outside either side of the frame bracket hole, as shown in b.
4. The lanyard assembly will now hang loosely with the clevis pin and fender washer end on the outside, as shown in c.
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 10
Page 11
d fe
5. From the inside, reattach the hairpin clip to the cord’s looped end as shown in d. This prevents the cord from slipping back through the frame bracket hole.
6. From the outside, insert lanyard clevis pin with washer into the small hole directly below the bed end bracket top hook so that the washer rests against the bracket, as shown in e.
7. From the inside, insert the hairpin clip, now attached to the cord, into the clevis pin hole to secure, as shown in f.
8. Follow previous steps 1-7 for other bed end corners, ensuring that all four corners are securely fastened and locked in place.
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 11
Page 12
Full-electric bed only (US0458 / US0458PL): Assemble hi/lo motor shaft to foot end gear box
FOOT
END
notches in hi/lo motor shaft coupler
slide over
gear box shaft
pin
gear box
hi/lo motor shaft
Attach hi/lo motor shaft
to foot end gear box
hi/lo motor
You must rst attach the coupler end of the motor shaft to the foot-end gear box (see above).
1. Note the position of the hi/lo motor plug on the Okimat actuator next to the pendant hand control plug — plug the hi/lo motor into the actuator, routing the hi/lo motor cord under the
fixed base frame cross-angle as shown below to ensure clearance with the hi/lo drive shaft.
2. Plug the power cord into any grounded outlet.
3. Pull the motor shaft toward the foot end. Using the hi/lo function on your pendant hand con­trol, slowly turn the hi/lo motor shaft until the coupler matches up with the gear box shaft pin as shown above.
4. Slide the coupler until it rests completely on the pin. You may need to manipulate the foot end slightly from side to side and/or lift the hi/lo motor itself for proper alignment.
5. Unplug the power cord.
HEAD
FOOT
Route cord under cross-angle
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 12
Page 13
HI/LO ROD ASSEMBLY / BED LENGTH
The hi/lo rod assembly consists of an inner and outer tube with a spring button lock to keep them together.
1. Assemble outer and inner tubes:
Semi-electric bed only (US0208 / US0208PL): Assemble hi/lo rod
Hi/lo rod assembly
semi-electric bed
US0208 / US0208PL
spring button
inner tube
outer tube
inserted spring button position
Depress spring button and insert inner tube into outer tube until third hole is reached as shown above. Release spring button to lock it into hole.
Full-electric bed only (US0458 / US0458PL): Assemble hi/lo rod
Hi/lo rod assembly
full-electric bed
spring button
inner tube
outer tube
US0458 / US0458PL
inserted spring button position
Depress spring button and insert inner tube into outer tube until rst hole of second hole set is reached as shown above. Release spring button to lock it into hole.
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 13
Page 14
All models: Attach hi/lo rod to gearbox or motor shaft pins
Both ends of the hi/lo rod assembly have a slot to mount on the foot end motor shaft pins and bed end gearbox shaft pins. Follow steps 2-3 to mount the hi/lo rod assembly on the gearboxes.
foot end
foot end gearbox
Fig 1: Semi-Electric Bed:
Attach hi/lo rod assembly
to foot end gearbox
notch at end of
over pin at end of m
FOOT END
end of hi/lo motor shaft
Fig 2: Full-Electric Bed:
Attach hi/lo rod assembly
to foot end hi/lo motor shaft
hi/lo rod rests
otor shaft
end of
rod (notch)
hi/lo
head end
head end gearbox
Fig 3: All Bed Models:
Attach hi/lo rod assembly
to head end gearbox
2. Connect the hi/lo rod assembly to the bed.
Semi-electric bed: Attach either end of hi/lo rod assembly to the foot end gearbox as shown in Fig 1 above.
Full-electric bed: Attach either end of hi/lo rod assembly to the foot end hi/lo motor shaft as shown in Fig 2 above.
3. Attach other end of hi/lo rod assembly to head end gearbox as shown in Fig 3 above.
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 14
Page 15
4 BED OPERATION
WARNING: Ensure that bed is secure and assembled as instructed on previous pages
before operating.
Plug the power cord into any grounded outlet.
WARNING: Do not leave the bed unattended when plugged in.
Info: In the event of a power failure, use a standard 9V battery to lower the bed’s head and foot sections (the 9V battery connection is attached to the motor). When used with the battery, the motor will only lower, not lift, the head and/or foot sections.
PENDANT HAND CONTROL OPERATION
Semi-electric bed only (US0208 / US0208PL)
The semi-electric pendant hand con­trol is shown at right. Use the pendant hand control to raise and lower bed head and foot. Each button has a differ­ent function, as noted in the picture.
Full-electric bed only (US0458 / US0458PL)
The full-electric pendant hand con­trol is shown at right. Use the pendant hand control to raise and lower bed, head and foot. Each button has a differ­ent function, as noted in the picture.
to raise head
to lower head
and foot at the
same time
to lower head
to raise head
to raise foot
to raise bed
to raise foot
to raise head and foot at the same time
to lower foot
Semi-electric
pendant hand control
to lower head
to lower foot
to lower bed
Full-electric
pendant hand control
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 15
Page 16
CRANK OPERATION: SEMI-ELECTRIC BED ONLY (US0208 / US0208PL)
Use the crank at the center foot of the semi­electric bed (shown at right) to change bed
foot
height (raise or lower entire deck).
crank
Crank operation
TRENDELENBURG POSITION, ALL MODELS
(HEAD END AT LOWER ELEVATION THAN FOOT END)
1. Lower bed to its lowest height. Disengage the hi/lo rod by compressing the spring button and removing the hi/lo rod from the motor or bed end.
2. Raise the foot bed end using the pendant hand control or crank (semi-electric bed only). The head end will remain in the low position.
REVERSE TRENDELENBURG POSITION, ALL MODELS
(HEAD END AT HIGHER ELEVATION THAN FOOT END)
1. Raise bed to its highest height. Disengage the hi/lo rod by compressing the spring button and removing the hi/lo rod from the motor or bed end.
2. Lower the foot bed end using the pendant hand control or crank (semi-electric bed only). The head end will remain in the high position.
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 16
Page 17
5 SPECIFICATIONS, ALL MODELS *
* Specifications are subject to change without notice.
Deck
High position
height
Low position
Bed length
Info: To extend grid length (80 in.) to 84 in., install optional kit 690-0084-000 (all beds).
Deck width
US0208 / US0208PL:
24.0 inches
US0208 / US0208PL:
15.0 inches
All models: 87.0 inches
Deck height: High Position
US0458 / US0458PL with cast­ers in low position: 20.0 inches
Deck height: Low Position
US0458 / US0458PL with cast­ers in low position: 9.5 inches
Bed Length
Deck Width
US0458 / US0458PL with casters in standard position:
23.5 inches
US0458 / US0458PL with casters in standard position: 13 inches
All models: 36.0 inches
Weight *
Normal operating conditions 450 lb, evenly distributed including mattress, accessories, rails, etc.
* Maximum patient weight 350 lb
Construction
Frame 10 ga (.135) high strength, low alloy steel (roll formed) Sleeping surface 11 ga (.120) cold rolled steel
Finish
5 stage bath 1. S21F acid bath (cleans grease and oil from metal).
2. Fresh water rinse.
3. QuiliPhos 20 phosphate coat (inhibits rust and helps maintain electrical charge during powder application).
4. Fresh water rinse.
5. QuiliRinse 404 (seals metal for powder coating).
All parts are then coated with a colored epoxy powder and baked 35 minutes at 375 degrees F. The powder coating nish is approximately 2.5 mils thick. The coating is smooth, uniform nish without runs, wrinkles, or grit.
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 17
Page 18
6 ACCESSORIES: BED RAILS
Patriot Homecare Bed decks are equipped with patent-pending slots specically for bed rail instal­lation, providing a xed place to mount GF bed rails. Two styles of Lumex bed rails are described in Section 6:
1) The Lumex Patriot Liberty bed rail system, designed specically for Patriot Beds, and
2) Lumex Universal bed rails
Please read the following bed rail safety information.
BED RAIL SAFETY
WARNING: If bed rails are not properly installed and adjusted, personal injury and/or dam-
age to the bed rail could result.
For information on Bed Rail safety, please see the following brochure: “A Guide to Bed Safety: Bed Rails in Hospitals, Nursing Homes and Home Health Care: The Facts” on the FDA website, at
www.fda.gov/MedicalDevices/ProductsandMedicalProcedures/GeneralHospitalDevicesandSupplies/HospitalBeds/
WARNING: GF Health Products, Inc. assumes no responsibility for any damage or injury
caused by improper installation of bed rails or unauthorized bed rail use. Only Lumex bed rails are recommended for use with Patriot beds. Please consult the instructions that ac­company your Lumex bed rails for complete rail assembly and installation instructions.
LUMEX PATRIOT LIBERTY BED RAILS
Lumex Patriot Liberty Bed Rails are designed specically for use with Lumex Patriot Beds. Devel­oped using the FDA Hospital Bed System Dimensional Guidance to Reduce Entrapment and tested to the Hospital Bed Safety Workgroup and FDA Dimensional Test Methods for Bed
System Zones 1-4 and International Standard IEC60601-2-38 Requirements for the Safety of Electrically Operated Hospital Beds, Lumex Patriot Liberty Bed Rails offer added protection and
reduce the risk of entrapment. For installation of these FDA compliant-bed rails, follow the installa­tion instructions that accompany them. Patriot Liberty Bed Rail descriptions follow.
GF6500B-1: Assist Bar
The Assist Bar, shown at right, is a secure handhold and support for active patients who need assistance getting in and out of bed on their own. This bar is bolted directly to either side of the bed frame to support the split frame when a patient enters or exits the bed. It can also assist patients ca­pable of grasping the bar and maneuvering themselves with rotating and turning in bed.
GF6500B-1
Assist Bar
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 18
Page 19
GF6580B-1: Quarter Rail
GF6580B-1
Quarter Rail
head end slot
pivot point
slot
pivot
point
foot end
head end slots
head end
slotted head deck angle
foot end pivot point
foot end pivot point slot
foot end slot
foot end
head end
foot deck angle
foot end slots
The Quarter Rail, shown above, is approximately 18 inches in length when mounted on the Patriot bed. It can be used as a head assist, head rail, or foot rail.
Mounting slot locations utilized when the Quarter Rail is used as a head rail are shown in “head end slots” above (opposite slots are directly across bed).
Mounting slot locations utilized when the Quarter Rail is used as a foot rail are shown in “foot end slots” above (opposite slots are directly across bed).
GF6590B-1: Half No-Gap Head Rail
head end slot
pivot point slot
pivot point
head end
slotted head deck angle
GF6590B-1
Half No-Gap Head Rail
foot end
mounting slots
The Half No-Gap Head Rail, shown above, provides full head coverage of 31.5 inches and extends to eliminate the gap between the head board and the end of the rail. The GF6590B-1 mounted on the bed’s head and the GF6580B-1 mounted on the bed’s foot combine to create the Patriot split rail system.
Half No-Gap Head Rail mounting slot locations are shown in “mounting slots” above (opposite slots are directly across bed).
GF6570B-1: Full Length Rail
head end slot
head end
GF6570B-1
Full-Length Rail
pivot point
foot end slot
mounting slots
foot end
head end
The Full Length rail, shown above, is a nominal, telescoping full length rail that provides maximum length rail perimeter coverage.
Full Length Rail mounting slot locations are shown in “mounting slots” above (opposite side slots are directly across bed).
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 19
Page 20
LUMEX UNIVERSAL BED RAILS
If you install Lumex universal bed rails, follow the installation instructions that accompany them; the following instructions are provided to describe the differences between the installation of bed rail crossbraces on Patriot grid beds and spring beds.
WARNING: Do not use the bed rails as handles for getting into or out of bed; this could
result in personal injury and damage to the bed rails.
GF6570A-1: Full Length Side Rail
GF6570A-1
Full-Length Side Rail
HEADBOARD
FOOTBOARD
A
Spring Buttons
G1
Spring Buttons
G4
Position spring-loaded crossbraces #A into slots G1 and G4 on both sides of bed with spring but­tons facing inward as shown above.
LX5088: No-Gap Bed Rail
LX5088
No Gap Side Rail
HEADBOARD
B
G2
Spring Buttons
Spring Buttons
G3
A
FOOTBOARD
B
Position spring-loaded crossbraces #B into slots G2 and G3 on both sides of bed with spring but­tons facing inward as shown above.
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 20
Page 21
GF6580A-1: Low Universal Half Rail
Cable tie
A
B
GF6580A-1
Low Universal Half Rail
Secure crossbrace to head deck angle with cable tie
SUGGESTED RAIL TO HEADBOARD MINIMUM OF 9.25"
HEADBOARD
Spring Buttons
Spring Buttons
G2
Follow same procedure for mounting half rails on foot-end using G3 slots.
G3
FOOTBOARD
Head end: Position spring-loaded crossbrace #B into slots G2 on both sides of bed. Position cross­brace #A so that the stabilizer rests underneath the head deck angle. Secure the crossbrace to the head deck angle with a cable tie. Ensure spring buttons face inward as shown above.
Info: Suggested rail-to-headboard minimum is 9.25 inches.
Foot end: Position spring-loaded crossbrace #B into slots G3 on both sides of bed. Position cross­brace #A so that the stabilizer rests underneath the foot deck angle. Secure the crossbrace to the foot deck angle with a cable tie. Ensure spring buttons face inward as shown at head end above.
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 21
Page 22
7 SERVICE PARTS, ALL MODELS
Service parts for all bed models are shown below and on the following two pages. See pictures and tables for appropriate service parts.
Info: To order service parts, contact our Customer Service department at 800-365-2338 (fax 920-929-8213).
BED COMPONENTS
HEAD
FOOT
Part number Description Semi-electric Full-electric
US0208 US0208PL US0458 US0458PL
690-7100-901 Head section, includes hi/lo rod X X X X 690-7105-902 Foot section, semi-electric, includes motor X X 690-7104-902 Foot section, full-electric, includes motors X X 690-7001-000 End set, semi-electric, walnut panels X 690-7601-000 End set, semi-electric, plastic panels X 690-7004-000 End set, full-electric, walnut panels X 690-7604-000 End set, full-electric, plastic panels X
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 22
Page 23
BED END COMPONENTS
HEAD
6
FOOT
5
1
8
2
3
7
4
Item Part number Description Semi-electric Full-electric
US0208 US0208PL US0458 US0458PL
1 690-7001-901 Head bed end, semi-electric, walnut panels X
690-7601-901 Head bed end, semi-electric, plastic panels X 690-7004-901 Head bed end, full-electric, walnut panels X 690-7604-901 Head bed end, full-electric, plastic panels X
2 690-7001-902 Foot bed end, semi-electric, walnut panels X
690-7601-902 Foot bed end, semi-electric, plastic panels X 690-7004-902 Foot bed end, full-electric, walnut panels X 690-7604-902 Foot bed end, full-electric, plastic panels X
3 999-0301-904 Caster, set of 4: 2 locking & 2 non-locking X X X X
100-6330-014 Caster, non-locking, each X X X X
100-6330-015 Caster, locking, each X X X X 4 100-4200-012 Caster socket, each X X X X 5 554-2001-015 End panel, head/foot, semi-electric, walnut
panels
690-5001-025 End panel, head/foot, semi-electric, plastic
panels
690-5004-015 End panel, head/foot, full-electric, walnut
panels
690-5004-025 End panel, head/foot, full-electric, plastic
panels
6 554-2001-059 Capping, bed end, head/foot, semi-electric X X
690-5004-059 Capping, bed end, head/foot, full-electric X X 7 600-2001-912 Foot elbow, full-electric X X
600-2001-916 Foot elbow with crank, semi-electric X X
999-0353-000 Foot crank only, semi-electric X X 8 600-2001-910 Elbow, head end, semi-electric & full-electric X X X X 9 690-1000-100 End locking hardware, set of 4 (not shown) X X X X
X
X
X
X
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 23
Page 24
BED SERVICE PARTS
FOOT
HEAD
1
4
2
3
Item Part number Description Semi-electric, H7920SE Full-electric, H7920FE
US0208 / US0208PL US0458 / US0458PL
1 690-2001-913
690-2001-411 Pendant hand control, semi-electric X
690-3001-943 Motor and pendant hand control, full-electric X
690-2001-414 Pendant hand control, full-electric X 2 690-7001-944 Motor and bracket, hi/lo X
690-3001-945 Motor, hi/lo, full-electric X 3 554-2018-901 Hi/lo rod assembly X X 4 690-2001-412 Motor slide covers X X 5 690-1000-100 End locking hardware, set of 4 (not shown) X X
Motor and pendant hand control, semi-electric
X
Info: US0208 and US0208PL are identified on the foot section component label as Patriot Semi-Electric Bed Model H7920SE.
Info: US0458 and US0458PL are identified on the foot section component label as Patriot Full-Electric Bed Model H7920FE.
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 24
Page 25
8 MAINTENANCE AND SAFETY CHECKS
Perform the following as needed or between patient placements, whichever happens sooner:
Electronics and Motors
Bed frames and sleeping surface
Cleaning
Lubrication and mechanical
Check all controls to ensure that all functions work properly.
Check all cables for damaged or frayed wires. Power cord
Check to ensure that all plugs are fully inserted or attached. Check to ensure that all wires are routed and attached properly so not to interfere with any moving parts. Inspect motor casings for excessive wear and cracks Inspect motor slide caps for excessive wear and deformation
Visually check all welds Head section
Check joints between sleeping surface sections for loose fasteners. Inspect frames and sleeping surface for bending and excessive wear The metal parts of the bed are covered with a powder coating. Clean all coated parts with mild detergent
and warm water. Periodically raise head and foot sections of the bed and remove dust from frame. Also, periodically remove mattress and clean mattress deck.
Lubricate all caster roller and swivel bearings with light machine oil. Check all bolts and tighten as needed.
Foot control Head control Hi/Lo (if applicable)
Pendant hand control cord
Foot section Main Frame
WARNING: The bed’s electronic system contains no serviceable parts. Do not open the
motors, cables, or pendant hand control. Only an authorized equipment provider or factory­trained personnel may perform service or adjustments on these components.
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 25
Page 26
9 WARRANTY
GF Health Products, Inc. warrants the
Patriot US0208 / US0208PL Semi-Electric Homecare Bed
Patriot US0458 / US0458PL Full-Electric / Low Bed
as follows:
Two-year limited warranty for defects in workmanship and materials of mechanical components, frame, and electronics.
During the warranty period, defective items will be repaired or replaced at manufacturer’s option. The warranty does not include any labor charges incurred in replacement part(s) installation or any associated freight or shipping charges to the manufacturer.
The warranties contained herein contain all the representations and warranties with respect to the subject matter of this document, and supersede all prior negotiations, agreements and under­standings with respect thereto. The recipient of this document hereby acknowledges and repre­sents that it has not relied on any representation, assertion, guarantee, warranty, collateral con­tract or other assurance, except those set out in this document.
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 26
Page 27
10 INDEX
A
Assembly instructions 5
B
Bed end assembly 9
Bed end locking hardware, install 10
Bed extension kit, 84” 17
Bed height, low, full-electric 8
Bed height, standard, full-electric 8
Bed length adjustment 13
Bed operation 15
Bed rails 18
Bed rail safety 18
C
Caster position (full-electric bed only) 8
CAUTION statement, signicance 3
Crank operation (semi-electric bed only) 16
D
Deck height (high position) 17
Deck height (low position) 17
F
Full-electric 13
H
Head and foot assembly instructions 5
Hi/lo rod assembly 13
I
Info statement, signicance 3
Introduction 3
L
Lumex Patriot Liberty bed rails 18
Lumex universal bed rails 20
M
Maintenance, bed frame and sleeping surface 25
Maintenance, cleaning 25
Maintenance, electronics 25
Maintenance, lubrication and mechanical 25
Maintenance and safety checks 25
Motor assembly 6
Motor assembly troubleshooting 7
Motor slide caps, install 7
Motor slide caps, remove 6
N
NOTICE statement, signicance 3
P
Pendant hand control operation, full-electric 15
Pendant hand control operation, semi-electric 15
Power failure; battery operation 15
S
Service parts, all models 22
Service parts, bed components 22
Service parts, bed end components 23
Specication, construction 17
Specication, weight 17
Specications, dimensions 17
T
Thermal motor protection 4
Trendelenburg position 16
Trendelenburg position, reverse 16
W
WARNING statement, signicance 3
Warranty 26
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 27
Page 28
U.S.A. Corporate Headquarters:
GF Health Products, Inc. (“Graham-Field”)
2935 Northeast Parkway
Atlanta, Georgia 30360
telephone: 770-368-4700
www.grahamfield.com
Loading...