OXYGEN CONCENTRATOR
USER MANUAL
Model Numbers JB0160-010B, JB0160-010,
JB0160-010B-220, JB0160-010-220
and
JB0160-015B, JB0160-015, JB0160-015B-220,
JB0160-015-220 with O2 Sensor
Important: Do not operate this concentrator without
rst reading and understanding this manual!
Save this manual for future use.
CONTENTS
1 SAFETY ................................................................................... 3
Important Information .................................................................. 3
Before Installation......................................................................... 3
Placement ......................................................................................4
Fire Warning & Explosion ............................................................. 4
Maintenance .................................................................................. 5
Radio Frequency Interference ..................................................... 5
Additional Safety Warnings ..........................................................6
2 FEATURES .............................................................................. 8
Summary........................................................................................ 8
Characteristics .............................................................................. 8
3 HANDLING .............................................................................. 9
GF Health Products, Inc. Freight Policy ......................................9
Unpacking .....................................................................................9
Inspection .................................................................................... 10
Storage ........................................................................................10
4 OPERATION AND INSTALLATION ...................................... 11
Preparing the Concentrator ....................................................... 11
Turning the Concentrator On .....................................................12
Turning the Concentrator Off..................................................... 12
5 MAINTENANCE .................................................................... 13
Cleaning the Cabinet .................................................................. 13
Cleaning or Replacing the Filters (3 Types) .............................. 13
Cleaning the Optional Humidier Bottle ................................... 15
Replacing the Battery ................................................................. 15
6 TROUBLESHOOTING GUIDE .............................................. 16
7 BILL OF MATERIALS ...........................................................
Replacement Parts and Optional Accessories ........................ 18
18
8 SPECIFICATIONS ................................................................. 19
Normal Operation Conditions .................................................... 19
9 LIMITED WARRANTY ......................................................... 20
10 INDEX.................................................................................... 22
11 PRODUCT REGISTRATION ................................................
24
All specications and product congurations are subject to change without notication.
2 John Bunn O2 Premier Oxygen Concentrator
Federal Law (USA) restricts this device to sale by or on the order of a physician. This product should not be used unless the operator has been instructed by a qualified health care professional.
WARNING: Users who require continuous oxygenation must plan for alternate reserve sources of power and oxygen in the event of a failure or loss of power and oxygen. This device is to be used ONLY as an oxygen supplement device and is NOT intended to be a life­supporting or life-sustaining device.
1 SAFETY
The John Bunn O2 Premier oxygen concentrator is intend­ed for use as an oxygen supplement device in a home or care facility. Use the concentrator only as prescribed by a physician. This user manual serves as a reference for your oxygen concentrator.
Important Information
Please note the following special statements, used throughout this manual, and their signicance:
Note: Explanatory information.
Damage: Action could result in damage to equipment.
s
ATTENTION: Action could result in personal injury.
CAUTION: Action could result in electric shock.
WARNING: Action could result in fire and explosion.
Before Installation
Damage: During transport, always keep the concentrator
s
upright to prevent damage.
Damage: Only use stable and safe electrical power
s
sources.
3John Bunn O2 Premier Oxygen Concentrator
Damage: If the electrical power source becomes
s
unstable, discontinue use and find an alternate source.
Damage: The oxygen concentrator cabinet contains no
s
serviceable parts! The oxygen concentrator cabinet should ONLY be opened by an authorized equipment provider.
Placement
ATTENTION: The oxygen concentrator is intended to be rolled from room to room. Do not place the oxygen concentrator in surroundings where its airflow is obstructed.
ATTENTION: Be certain to place the unit so that all sides are at least 12 inches (30 cm) away from walls, draperies, furniture, or other obstructions. Do not place the unit in a confined area.
ATTENTION: The oxygen concentrator MUST be kept away from heat, fire and excessive water sources and conditions.
ATTENTION: Do not place items on top of the concentrator.
ATTENTION: NEVER block the concentrator’s air openings.
ATTENTION: Always place the concentrator on a hard surface. Never place the concentrator on a surface such as bed or couch, where the concentrator may tip or fall.
ATTENTION: Keep the concentrator openings free from lint, hair, etc.
Fire Warning & Explosion
WARNING: Keep the concentrator away from flammable and explosive areas.
4 John Bunn O2 Premier Oxygen Concentrator
WARNING: To reduce the risk of fire, special care must be taken when using oxygen concentrators. DO NOT SMOKE while using this device. Keep all matches, lighted cigarettes or other sources of ignition out of the room in which this concentrator is located. In public settings or when used in hospital or long-term care environments, prominently display NO SMOKING signs.
WARNING: Textiles and other materials that normally would not easily burn, can ignite and burn with great intensity in oxygen enriched air.
WARNING: A spontaneous and violent ignition may occur if oil, grease or greasy substances come in contact with oxygen under pressure. ALWAYS keep these substances away from the oxygen concentrator and all other oxygen equipment. DO NOT use any lubricants unless recommended by manufacturer.
WARNING: Failure to observe these warnings can result in severe fire, property damage and cause physical injury or DEATH.
Maintenance
CAUTION: The oxygen concentrator was specifically designed to minimize routine preventive maintenance. Only an authorized equipment provider or factory-trained personnel should perform preventive maintenance or adjustments to the oxygen concentrator.
Radio Frequency Interference
Most electronic equipment is inuenced by Radio Fre­quency Interference (RFI). ALWAYS exercise CAUTION with regard to the use of portable communications equipment in the area around such equipment.
5John Bunn O2 Premier Oxygen Concentrator
Additional Safety Warnings
WARNING: TO REDUCE THE RISK OF BURNS, ELECTROCUTION, FIRE OR INJURY TO PERSONS, avoid using while bathing. If continuous usage is required by the direction of a physician, the concentrator must be located in another room at least 9 feet (2.8m) from the bath.
WARNING: DO NOT reach for a concentrator that has fallen into water. UNPLUG IT IMMEDIATELY.
WARNING: In certain circumstances, oxygen therapy can be hazardous. Graham-Field recommends that you seek medical advice before using this product.
WARNING: Avoid creation of any spark near medical oxygen equipment. This includes sparks from static electricity created by any type of friction.
ATTENTION: DO NOT come in contact with the concentrator when you are wet.
ATTENTION: DO NOT place or store the concentrator where it can drop into water or other liquid.
ATTENTION: NEVER leave the concentrator unattended when plugged in.
ATTENTION: This device is to be used only in accordance with the prescription of a physician and this user manual. If at any time the patient or attendant concludes that the patient is receiving an insufficient amount of oxygen, contact the provider and/or physician immediately. NEVER make adjustments to the flow rate unless prescribed by a physician.
6 John Bunn O2 Premier Oxygen Concentrator
ATTENTION: ALWAYS supervise closely when this product is used near children or those who require close supervision.
ATTENTION: Only use this device as intended and described in this manual.
ATTENTION: DO NOT use parts, accessories or adapters other than those authorized by manufacturer.
ATTENTION: DO NOT connect the concentrator in parallel or series with other oxygen concentrators or oxygen therapy devices.
ATTENTION: JB0160-015B, JB0160-015, JB0160-015B-220 and JB0160-015-220 only: Discontinue concentrator use immediately if O2 sensor red warning light illuminates; oxygen concentration is too low for safe use. Contact your Graham-Field equipment provider immediately.
7John Bunn O2 Premier Oxygen Concentrator
2 FEATURES
Summary
The John Bunn O2 Premier oxygen concentrator is intend­ed for use as an oxygen supplement device in a home or care facility. Use the concentrator only as prescribed by a physician.
The concentrator separates oxygen from ambient air, providing the user a high concentration of oxygen. Oxygen concentrator usage is therapeutically equivalent to other types of oxygen delivery systems.
Characteristics
• Reliable, safe, complete plastic outer shell.
• Display screen shows total elapsed working hours.
• Pressure safety valve helps ensure operating pressure.
• Power loss alarm function.
• High and low pressure alarm function.
• Heat protection to ensure the safety of the compressor and concentrator.
• JB0160-015B, JB0160-015, JB0160-015B-220, and JB0160-015-220 only: O2 sensor (electro-chemical de­vice): Yellow fault light illuminates when O2 concen­tration is greater than 71% and less than 82%; system works. Red warning light illuminates when O2 concen­tration is less than 71%; system shuts off.
8 John Bunn O2 Premier Oxygen Concentrator
3 HANDLING
GF Health Products, Inc. Freight Policy For Your Protection, Read Carefully
The carrier accepted this merchandise “in good condition” and is responsible for safe delivery.
Before signing the freight bill, inspect the shipment care­fully for damage or missing pieces.
Apparent Loss or Damage
Should visual inspection show loss or damage, this MUST be noted on the freight bill and signed by the carrier’s agent. Failure to do so may result in the carrier failing to honor the claim. Please contact the carrier to obtain the paperwork necessary to le a claim.
Concealed Loss or Damage
If damage is discovered after delivery is made, a concealed damage claim must be entered with the freight carrier. When this occurs, make a written request to the carrier for inspection. This request for inspection must be made within 15 days of delivery. The carrier will provide all paperwork necessary to le a concealed damage or loss claim, since such damage or loss is the carrier’s responsi­bility.
Unpacking
Note: Unless the oxygen concentrator is to be used
immediately, retain containers and packing materials for storage until concentrator use is required.
1. Check for obvious damage to the carton or its contents. If damage is evident, please notify the carrier and your Graham-Field equipment provider.
2. Remove all loose packing from the carton.
3. Carefully remove all the components from the carton.
9John Bunn O2 Premier Oxygen Concentrator