Graham Field NEB-U-TYKE, JB0112-164 User Manual

Neb-u-Tyke™ RR Train
Nebulizer Compressor
Item JB0112-164
User Manual
Read this manual before operating the nebulizer. Save this manual for future reference.
Note: The most current version of this manual can be found online at www.grahameld.com.
Federal Law (USA) restricts this device to sale by or on the order of a physician. This device should not be used unless the operator has been instructed by a
qualied healthcare professional.
JB0112-164-INS-LAB-RevB09
This product has been tested and successfully approved with the following standards:
UL1431
EN 60601-1
CONTENTS
INTRODUCTION ...........................................................................................................3
INTENDED USE ...........................................................................................................3
SYMBOLS KEY ............................................................................................................3
IMPORTANT SAFETY PRECAUTIONS .......................................................................4
PRODUCT DESCRIPTION ...........................................................................................6
FEATURES AND COMPONENTS ................................................................................ 6
SET-UP ..........................................................................................................................7
OPERATION .................................................................................................................8
SAFETY OVERLOAD ................................................................................................... 9
MAINTENANCE ............................................................................................................9
CLEANING THE COMPRESSOR’S OUTER CASE .................................................... 9
CLEANING THE NEBULIZER CUP ...........................................................................10
REPLACING THE FILTER .......................................................................................... 10
STORAGE ....................................................................................................................11
TROUBLESHOOTING ................................................................................................11
SPECIFICATIONS .......................................................................................................12
LIMITED WARRANTY ................................................................................................13
COMPRESOR DE NEBULIZADOR MANUAL USUARIO (EN ESPAÑOL) ...............17
Graham-Field and John Bunn are registered trademarks, and Neb-u-Tyke is a trademark, of GF Health Products, Inc.
Packaging, warranties, products, and specications are subject to change with­out notice. GF Health Products, Inc. is not responsible for typographical errors.
John Bunn Neb-u-Tyke RR Train Nebulizer Compressor 2
INTRODUCTION
This user manual contains important information and safety precautions for the John Bunn Neb-u-Tyke RR Train Nebulizer Compressor.
Before using your Nebulizer Compressor, please read and understand this entire user manual. Take special note of all safety precautions marked DANGER and WARNING. Save this user manual for future reference.
Intended Use
The Neb-u-Tyke RR Train Compressor Nebulizer System is designed to provide a source of compressed air for aerosol therapy. It is used in conjunction with a jet (pneumatic) nebulizer to produce medicated aerosols for inhalation by pediatric and adult patients with respiratory symptoms.
Indications for therapy include asthma, chronic bronchitis, infection of the upper respiratory tract, chronic obstruc­tive pulmonary disease, and other respiratory disorders in accordance with a physician’s prescription. Do not use this product for any other purpose than the preceding. This device can be used with adult or pediatric patients under physician’s prescription.
Symbols Key
O Off
I On
~ Alternating current
IPX0 Enclosed equipment without protection against ingress of water
Attention (read the instructions)
II Class II (not applicable for UL1431)
BF: This product is in a rdance with the degree of protection against electric shock for type BF equipment. (not applicable for UL1431)
Disposal of Electrical & Electronic Equipment (WEEE): This product should not be treated as household waste, but submitted to an appropriate collection point for the recycling of electrical and electronic equipment. For more detailed information about the recycling of this product, please contact your local city office, household waste disposal service, or the Graham-Field authorized distributor from whom you purchased this product.
John Bunn Neb-u-Tyke RR Train Nebulizer Compressor 3
IMPORTANT SAFETY PRECAUTIONS
When using electrical products, especially when children are present, always follow basic safety precautions, includ­ing the following:
DANGER: Indicates a potential hazard situation or unsafe practice that, if not avoided, will result in death or serious personal injury.
WARNING: Indicates a potential hazard situation or unsafe practice that, if not avoided, could result in death or seri­ous personal injury.
s
Caution: Indicates a potential hazard or unsafe prac­tice that, if not avoided, could result in minor personal injury or product/property damage.
Note: Provides application recommendations or other useful information to ensure that you get the most from your prod­uct.
READ ALL INSTRUCTIONS BEFORE USING
DANGER: To reduce the risk of electrocution:
1. Always unplug this product immediately after use.
2. Do not use this product while bathing, showering, wash­ing dishes, or close to water sources of any kind.
3. Do not place or store device where it can fall or be pulled into a tub or sink.
4. Do not place device in or drop into water or other liquid.
5. Do not reach for a device that has fallen into water. Un­plug it immediately.
WARNING: To reduce the risk of burns, electrocution, re
or personal injury:
1. Never leave this device unattended when plugged in.
2. This product contains small parts that may present a choking hazard to small children, and the tubing may present a strangulation hazard. Always use close supervi-
sion when this device is used by or near children or those who require close supervision.
John Bunn Neb-u-Tyke RR Train Nebulizer Compressor 4
3. Children under the age of 3, or any patient who is unable to use a mouthpiece properly under supervision, should use a mask.
4. Use this device only as intended and described in this manual. Do not use attachments or accessories not rec­ommended by GF Health Products, Inc.
5. Never operate this product if: a. it has a damaged cord or plug, b. it is not working properly or it has been dropped or
damaged,
c. it has been dropped into water.
Return the product to your Graham-Field authorized dis-
tributor for examination and repair.
6. Keep the cord away from heated surfaces.
7. Never block the air openings of this product or place it on a soft surface, such as a bed or couch, where the air openings may be blocked. Keep the air openings free of lint, hair and the like.
8. Never drop or insert any object into any openings or hose.
9. Never use this device while drowsy or sleeping.
10. Do not use this device outdoors or operate where aerosol (spray) products are being used.
11. Do not use this device in the presence of ammable anes­thesia or where oxygen or nitrous oxide is being adminis­tered. This device has no AP or APG protection.
12. Connect this device to a properly grounded outlet only.
13. This device has a polarized plug (one blade is wider than
the other). As a safety feature, this plug will t into a polarized outlet only one way. If the plug does not t fully into the outlet, reverse the plug. If it still does not t, con­tact a qualied electrician. Do not modify the plug in any
way. Do not attempt to defeat this safety feature.
14. Unplug the device before lling or cleaning the nebulizer.
15. This device is oil-free. Do not lubricate.
16. Do not remove the jet nebulizer while the device is operating.
John Bunn Neb-u-Tyke RR Train Nebulizer Compressor 5
17. Do not leave medication inside the jet nebulizer when it is not in use.
18. Use the nebulizer intermittently; the maximum operating period is 30 minutes. The nebulizer must stand unused for 30 minutes between uses.
19. The mouthpiece and masks are for use by one person only; in order to avoid infection, do not share them.
PRODUCT DESCRIPTION
Features and Components
Please note nebulizer features and components, labeled in the illustration below.
ON/OFF switch
filter cap
air outlet connector for tubing
tubing
adult mask
Neb-u-Tyke RR Train
T adapter
pediatric mask
reservoir tube
nebulizer cup
mouthpiece
Nebulizer Compressor
Note: RR Train replacement parts are available at www.grahamfield.com or from your Graham-Field authorized distributor.
John Bunn Neb-u-Tyke RR Train Nebulizer Compressor 6
SET-UP
Note: Before initial operation, clean the nebulizer cup as­sembly following the instructions given in MAINTENANCE/ Cleaning the Nebulizer Cup.
WARNING: Ensure that the nebulizer compressor is un­plugged during set-up.
1. Before each use, inspect the compressor and nebulizer
cup assembly for damage or wear; replace as needed.
2. Place the compressor on a table or other at stable sur-
face (not on the oor), in an area that is clean and dust­free. Be sure you can easily reach the controls when seated.
3. Open the nebulizer medication cup: Twist the cap coun-
ter-clockwise until it stops, then lift cap off.
4. Pour the prescribed medication into the nebulizer medi-
cation cup. Do not ll with more than 5 ml per treat­ment.
cap
5. Install cap on cup. Turn clockwise to
tighten.
6. If using the mouthpiece for treatment,
proceed to step 7. If using a mask, install the bottom part of the mask directly onto the top of the nebulizer medication cup cap as shown at right. Proceed to step 9.
7. Insert mouthpiece into large end of
T adapter.
John Bunn Neb-u-Tyke RR Train Nebulizer Compressor 7
cup
install cap
on cup
mask
cap
install mask
on cap
T adapter
mouthpiece
assemble
mouthpiece
and T adapter
Note: The reservoir tube, shown in the illustration earlier in this manual, acts as a holding tank for the mist, reducing the amount lost during treatment. Its use is optional. If you wish to use the reservoir tube, install it now on the small end of the T adapter.
8. Install T adapter onto top of nebulizer
medication cup cap.
9. Connect the tubing to the adapter on the
bottom of the nebulizer medication cup.
10. Connect the tubing's other end to the
install
T adapter
on cap
T adapter
cup
compressor air outlet connector.
WARNING: Before use, check that all parts are securely connected. Use of an improperly assembled nebulizer could diminish or prevent adequate delivery of medication and the effectiveness of the treatment.
OPERATION
s
Caution: The Neb-u-Tyke RR Train Nebulizer Compres­sor is designed for intermittent use only. Do not op­erate it continuously for more than 30 minutes for a single use without turning it off and allowing a cooling period of least 30 minutes.
1. Ensure that the ON/OFF switch is in the OFF position.
2. Plug the power cord into a properly grounded electrical
wall outlet.
3. Move the ON/OFF switch to the ON position to start the
compressor.
4. Place the nebulizer mouthpiece between teeth and close
mouth (or install the mask on face).
5. As aerosol starts owing, inhale deeply and slowly
through mouthpiece (or mask).
WARNING: Never use a nebulizer with a clogged nozzle. If the nozzle is clogged, the aerosol mist will be reduced, altered or stopped, which will diminish or eliminate the ef­fectiveness of the treatment. If clogging occurs, replace the nebulizer.
John Bunn Neb-u-Tyke RR Train Nebulizer Compressor 8
Note If you need to interrupt treatment, move the ON/OFF switch to the OFF position.
6. When the treatment is complete, move the ON/OFF
switch to the OFF position to turn off the compressor.
7. Unplug the compressor from power outlet.
8. Disconnect nebulizer from tubing.
9. Clean nebulizer as instructed in MAINTENANCE/Clean-
ing the Nebulizer Cup.
Safety Overload
Note: Do not exceed 30 minutes of continuous operation.
1. The motor of this device has a built-in thermal overload
protector. Should the motor overheat, the protector will automatically shut off the motor. Should this occur, move the ON/OFF switch to the OFF position and allow the motor to cool down for approximately 30 minutes before turning it on again.
2. If the overload protector shuts off the motor frequently,
you may have an unstable voltage situation.
3. If the device shuts down and cannot restart, it may need
to be replaced. Call your Graham-Field authorized dis­tributor immediately.
MAINTENANCE
Cleaning the Compressor’s Outer Case
WARNING: Electrical shock hazard. Do not remove the outer case of the device. All disassembly and maintenance
of the device must be performed by a qualied service technician. Refer service to qualied service personnel.
WARNING: This device does not require oil. Do not attempt to lubricate any internal parts.
WARNING: Unplug device before cleaning. Do not sub­merge in water for cleaning.
Wipe the compressor's outer case with a clean, damp cloth weekly to keep it dust-free. Do not use any powdered type cleaners or soap.
John Bunn Neb-u-Tyke RR Train Nebulizer Compressor 9
Cleaning the Nebulizer Cup
All parts of the disposable nebulizer, except tubing, should be thoroughly cleaned after each use. If your physician species a specic cleaning procedure, follow that proce­dure.
WARNING: To prevent possible risk of infection from con­taminated medication, clean the nebulizer after each treat­ment.
1. After each use, disassemble the nebulizer cup, T adapt-
er and mouthpiece completely, wash all parts in warm soapy water, and rinse thoroughly. Allow to air dry. Store in a clean, dry location.
2. Once daily, after washing as above, soak nebulizer com-
ponents in a fresh solution consisting of one part warm water and one part white vinegar for 30 minutes fol­lowed by a warm water rinse. Allow to air dry. Store in a clean, dry location.
Replacing the Filter
Note: Change the filter once every 30 days or when the filter changes color. Additional filters can be purchased at www.grahamfield.com or from your Graham-Field authorized distributor.
1. Ensure that ON/OFF switch is in the OFF position. Open
the lter cap located next to the ON/OFF switch.
2. Remove the dirty lter with a small, pointed object. Dis-
card the lter.
3. Replace with a clean lter.
4. Replace the lter cap.
s
Caution: Do not use cotton or other materials as a filter.
s
Caution: Do not wash or clean the filter.
s
Caution: Do not reuse the filter. Do not operate com­pressor without a filter. Compressor damage can result.
s
Caution: The disposable nebulizer kit is for single pa­tient use only.
John Bunn Neb-u-Tyke RR Train Nebulizer Compressor 10