Graham Field 760000-INS-LAB User Manual

Page 1
760000
ALTERNATING PRESSURE / LOW AIR LOSS MATTRESS SYSTEM
USER MANUAL
Important: Do not operate the Mattress System
without rst reading and understanding this manual! Save this manual for future use.
760000-INS-LAB-RevA08
Page 2
Contents
INTRODUCTION ..................................................................................... 3
INTENDED USE OF THIS DEVICE ........................................................ 3
IMPORTANT SAFETY PRECAUTIONS .................................................. 3
DANGER ................................................................................................. 4
WARNING ............................................................................................... 4
CAUTION ................................................................................................ 5
EQUIPMENT SYMBOLS ......................................................................... 5
PRODUCT DESCRIPTION ......................................................................6
PUMP AND MATTRESS SYSTEM ......................................................... 6
PUMP: FRONT ........................................................................................ 6
PUMP: REAR .......................................................................................... 7
PUMP: FRONT PANEL ........................................................................... 7
FRONT PANEL FIRST ROW (CYCLE TIME ADJUSTMENT) ......... 7
FRONT PANEL SECOND ROW (PRESSURE ADJUSTMENT) ...... 7
FRONT PANEL THIRD ROW (THERAPY MODE ADJUSTMENT) . 7
FRONT PANEL FOURTH ROW ....................................................... 8
INSTALLATION ........................................................................................ 9
UNPACKING ........................................................................................... 9
PUMP AND MATTRESS INSTALLATION .............................................. 9
OPERATION ...........................................................................................10
PRESSURE ADJUSTMENT ................................................................. 11
QUICK DISCONNECT .......................................................................... 11
TRANSPORT CAP ................................................................................ 12
CPR FUNCTION ................................................................................... 12
MAINTENANCE ..................................................................................... 12
CLEANING ............................................................................................ 12
GENERAL MAINTENANCE .................................................................. 13
FUSE REPLACEMENT .................................................................. 13
AIR FILTER REPLACEMENT ........................................................ 13
STORAGE ............................................................................................. 14
SPECIFICATIONS .................................................................................. 14
WARRANTY ........................................................................................... 15
GF, Graham-Field, Lumex and AltaDyne are registered trademarks of GF Health Products, Inc.
GF Health Products, Inc. is not responsible for typographical errors. Packaging, warranties, specifications and products are subject to change without notice.
2 760000-INS-LAB-RevA08
Page 3
IntroduCtIon
Use this manual for set-up and operation of the Alternating Pressure / Low Air Loss Mattress System. Read all instruc­tions before using the mattress system. Save this manual for future reference.
Intended use of thIs devICe
The intended use of the Alternating Pressure / Low Air Loss Mattress System is:
• The treatment of pressure ulcers while optimizing user comfort.
• Pain management as prescribed by a physician.
• As described above, in either a homecare or long-term care setting.
Important safety preCautIons
The safety statements presented in this chapter refer to the basic safety information that should be observed by those using the Alternating Pressure / Low Air Loss Mat­tress System. There are additional safety statements in other chapters or sections, which may be the same as or similar to the following, or specic to the operations.
DANGER: Indicates an imminent hazard situation that, if not avoided, will result in death or serious injury.
WARNING: Indicates a potential hazard situation or unsafe practice that, if not avoided, could result in death or serious injury.
s
Caution: Indicates a potential hazard or unsafe prac­tice that, if not avoided, could result in minor person­al injury or product/property damage.
Note: Provides application recommendations or other useful information to ensure that you get the most from your product.
760000-INS-LAB-RevA08 3
Page 4
danger
To reduce the risk of electrocution:
DANGER: Always unplug this product immediately after using.
DANGER: Do not use this product while bathing.
DANGER: Do not place or store this product where it can fall or be pulled into a tub or sink.
DANGER: Do not place this product in or drop into water or other liquid.
DANGER: Do not reach for a product that has fallen into water. Unplug it immediately.
WarnIng
To reduce the risk of burns, electrocution, re, or
personal injury:
WARNING: Do not leave this product unattended when plugged in.
WARNING: Always use close supervision when this product is used by, on, or near children or those who require close supervision.
WARNING: Use this product only as intended and described in this manual. Do not use attachments or accessories not recommended by Graham-Field.
WARNING: Never operate this product if
a) it has a damaged cord or plug b) it is not working properly c) it has been dropped or damaged d) it has been dropped into water.
Return the product to your Graham-Field equipment provider.
WARNING: Keep the cord away from heated surfaces.
4 760000-INS-LAB-RevA08
Page 5
WARNING: Never block the air openings of this product or place it on a soft surface, such as a bed or couch, where the openings may be blocked. Keep the air opening free of lint, hair, and other similar particles.
WARNING: Never drop or insert any object into any opening or hose.
WARNING: Do not use this product in the presence
of ammable anesthesia, or where oxygen or nitrous oxide is being administered.
WARNING: Connect this product only to a properly grounded power outlet.
WARNING: Maximum patient weight capacity for this
product is 400 lb (189 kg).
CautIon
s
Caution: The pump can be used only with the accom­panying mattress. Do not use it for any other purpose.
equIpment symbols
Attention! Consult accompanying documents (this manual)
Indicates that this product is according to the degree of protection against electric shock for type BF equipment
Functional earth (for UL only)
Class II (for UL only)
Disposal of Electrical & Electronic Equipment (WEEE): Do not treat this product as household waste. For more detailed information with regard to returning and recycling
this product, please consult your local city ofce, household
waste disposal service, or Graham-Field Health Products equipment provider.
760000-INS-LAB-RevA08 5
Page 6
For U.S. and Canada only
pump and mattress system
CPR valves
mattress-pump connectors
mattress
foot end symbol
pump
quick connector
E228589 53DG
Medical Equipment: Air Pump
with respect to electrical shock, re and
mechanical hazards only in accordance with UL60601-1 and CAN/CSA C22.2 No. 601.1
produCt desCrIptIon
pump and mattress system
pump: front
6 760000-INS-LAB-RevA08
Page 7
pump: rear
main power
switch
power cord
receptacle
fuse
hanger
filter
cycle time adjust button; cycle time indicators at right of button
decrease pressure button;
pressure level indicators
at right of button
panel lock button; panel lock
indicator at right of button
therapy adjust button;
therapy mode indicators
at right of button
power ON/OFF button; ON/OFF
indicator at right of button
low pressure indicator (upper) power failure indicator (lower)
alarm mute button
service indicator
increase pressure button
pump: front panel
Front panel rst row (cycle time adjustment)
There are four cycle time settings. Press the cycle time adjust button to cycle between the settings. The selected cycle time setting's indicator will illuminate.
Front panel second row (pressure adjustment)
There are ten pressure (rmness) settings. Press the de­crease pressure button at the left end of the row to de­crease pressure (softer), and press the increase pressure button at the right end of the row to increase pressure (rmer). The selected pressure level setting's indicator will illuminate.
Front panel third row (therapy mode adjustment)
There are four therapy mode settings: Autofirm (to ll mattress quickly, or to fully inate mattress during patient ingress/egress for better support), Alternating Pressure (every other tube lls, then deates), Static (all tubes remain full), and Seat Inflation (when the head is raised to
760000-INS-LAB-RevA08 7
Page 8
increase pressure under the buttocks). Press the therapy adjust button at the left end of the row to cycle between the settings. The selected therapy mode setting's indicator will illuminate.
Front panel fourth row
Panel lock button: The panel lock protects the panel set­tings from accidental change. Press to lock; press for three seconds to unlock. When the panel remains untouched for three minutes, it locks automatically. When the panel is locked, the panel lock indicator will illuminate.
Power button (ON/OFF): The power button turns power to the mattress system ON or OFF. Press to turn ON; press to turn OFF. When power is on, the power (ON/OFF) indicator will illuminate. When power is turned OFF, mattress system will slowly deate. Note: Main power switch on side of pump
must be ON in order for the panel power button to function.
Alarm mute button: The alarm mute button temporarily suspends the LED indicator and buzzer when either the low-pressure alarm or power failure alarm is activated. Should the problem not be resolved within ve minutes, the alarm will resume.
Low pressure indicator: The low pressure indicator illumi­nates, and the alarm sounds, when pressure level is lower than the selected pressure. If the low pressure indicator persistently illuminates inappropriately:
1. Check to ensure all connections are properly and se
-
curely connected per installation instructions.
2. Check for any leakage (tubes or connecting hoses). If necessary, contact your Graham-Field equipment pro­vider to replace any damaged tubes or hoses.
3. If problem persists, contact your Graham-Field equip­ment provider.
Power failure indicator: The power failure indicator illu­minates, and the alarm sounds, when there is a power failure. Depress the alarm mute button to disable both alarm and LED.
8 760000-INS-LAB-RevA08
Page 9
Service indicator: Mechanical failure, service required;
hanger
contact your Graham-eld equipment provider.
InstallatIon
unpaCkIng
1. Before unpacking, check for obvious damage to the car­ton or its contents. If damage is evident, please notify the carrier and your Graham-Field equipment provider.
2. Remove all loose packing from the carton.
3. Carefully remove all components from the carton.
4. Inspect all components. If damage is evident, please notify the carrier and your Graham-Field equipment provider.
pump and mattress InstallatIon
1. Remove all sheets, blankets, and pillows from the bed.
2. Place the mattress on top of the bed frame; ensure that the foot end symbol faces up at foot end as shown at right. Secure mat­tress to bed with straps.
3. Fold open the wire hangers on the back of the pump as shown at right.
4. Either hang the pump on the bed's foot end, as shown at right, or place the pump on a at sur­face easily accessible to the care­giver and/or doctor.
760000-INS-LAB-RevA08 9
Page 10
s
main
power
switch
power button
Caution: Place the device in a position where the care­giver and/or doctor can access it easily.
5. Connect the mattress-pump air hose quick connector; ensure that the connector has clicked securely into place before continuing.
s
Caution: Ensure that the air hoses are not kinked or tucked under the mattress.
WARNING: Locate all cords so that they will not be stepped on, tripped over, or otherwise subjected to damage or stress.
WARNING: Do not use a household extension cord if
the electrical cord does not reach the power outlet.
Use of an improper extension cord could result in re and electric shock. If an extension cord must
be used, use a three conductor cord with ground, properly wired, in good electrical condition, and keep it as short as possible.
WARNING: Ensure that the local power voltage is appropriate for the pump unit.
6. Plug the power cord into a properly grounded electrical outlet.
operatIon
1. Turn the main power switch to the ON position.
2. Press the front panel power button. The power indica­tor will illuminate.
10 760000-INS-LAB-RevA08
Page 11
Note: The mattress system will automatically enter Autofirm mode for the quickest inflation at set-up; initial inflation takes approximately twenty minutes.
3. The pump will begin delivering air into the mattress.
4. When the selected pressure level is reached, the pump will enter alternating mode. The mattress can then be adjusted to the desired pressure.
5. Replace all sheets, blankets, and pillows on the bed.
pressure adjustment
Generally, a lighter patient will need a lower (softer) set­ting while a heavier patient will need a higher (rmer) setting, but pressure adjustment must ultimately be based on the patient's weight distribution. A short, stocky person may be the same weight as a tall, thin person, but their settings will not be the same. The mattress should be rm enough to support the patient, yet soft enough so the patient will sink down to eliminate pressure points. Initially inate the mattress to maximum pressure. Then, when patient lies on the mattress, test the pressure by placing four ngers below the mattress under the patient's but­tock area; your ngers should just t. If the t is too loose, decrease the setting. If your ngers do not t well, increase the setting two notches rmer. Wait ten minutes and then check again. You should only have to adjust this setting one time per patient. Please consult your physician for an appropriate setting.
quICk dIsConneCt
Pull the power plug from the wall connector to disconnect the device quickly.
760000-INS-LAB-RevA08 11
Page 12
transport Cap
quick connector
transport cap
In case of power failure or trans­port: Disconnect the quick connec­tor and cap the mattress with the transport cap to slow deation.
Cpr funCtIon
If CPR must be performed on the patient while the mattress is in use, to immediately deate the mat­tress, either:
1. Pull the CPR valves from the mattress in the location of the arrow shown at right, or
2. Disconnect the quick connector from the pump in the location of the arrow shown at right.
maIntenanCe
CleanIng
Perform the following cleaning procedures before use.
Pump: Wipe the pump with a clean, damp cloth and mild detergent, and keep it away from dust. Air dry.
s
Caution: Do not use corrosive or powdered cleansers to clean the pump.
s
Caution: Do not immerse or soak the pump.
s
Caution: Do not clean with phenolic products.
Mattress: Wipe the mattress unit with a clean, damp cloth
and mild detergent. The mattress may also be cleaned us­ing a 10% solution of sodium hypochlorite (bleach) diluted in water. Air dry all parts thoroughly before use.
s
Caution: Do not clean with phenolic products.
s
Caution: After cleaning, air dry the mattress without direct exposure to sunlight.
12 760000-INS-LAB-RevA08
Page 13
general maIntenanCe
fuse
filter
1. Check main power cord for abrasion or excessive wear.
2. Check mattress cover for signs of wear or damage. En­sure mattress cover and tubes are connected correctly.
3. Check airow from the air hose connector. The airow should alternate between each connector every half­cycle time if it’s in alternating mode.
4. Check the air hoses for any kink or break. For replace­ment, please contact your GF equipment provider.
Fuse replacement
tool needed: small screwdriver
1. If you suspect a blown fuse, discon­nect the plug from the wall outlet immediately.
2. Use a small screwdriver to remove the cover of the fuse holder (fuse locations are shown in picture at right).
3. Insert a new fuse of the correct rating (T1A/250V, VDE approved).
4. Replace the fuse holder cover. Ensure that fuse holder cover is securely installed.
Air lter replacement
tool needed: small screwdriver
1. Use a small screwdriver to remove the air lter plate located at rear of pump (lter location shown in picture at right).
2. Remove the lter. The lter is reusable, if not torn, and can be washed gently with a mild detergent and water. Dry the lter before use.
3. Replace the lter and cover. Ensure that lter cover is securely installed. Replace the air lter regularly if it is gray, torn, or the environment is dirty.
760000-INS-LAB-RevA08 13
Page 14
storage
1. Lay the mattress out at and upside down.
2. Roll from the head end toward the foot end.
3. Stretch the foot-end strap around the rolled mattress to prevent unrolling.
4. Store mattress and pump in a dust-free environment with no exposure to direct sunlight.
s
Caution: Do not fold, crease or stack mattress.
speCIfICatIons
Pump Specication
Power Supply 120V System: AC 120V 60 Hz
Note: see rating label on product
Fuse Rating T1A, 250V
Cycle time 10, 15, 20 or 25 minutes
Air input >8 LPM
Dimensions (L x W x H) 11.4" x 7.9" x 4.6" (29.1 x 20 x 11.7 cm)
Weight 5 lb (2.2 kg)
Environment Temperature Operation: 50°F to 104°F (10°C to 40°C)
Storage: 5°F to 122°F ( -15°C to 50°C) Shipping: 5°F to 158°F ( -15°C to 70°C)
Humidity Operation: 10% to 90% non-condensing
Storage: 10% to 90% non-condensing Shipping: 10% to 90% non-condensing
Classication Class II, Type BF, IPX0
Applied Part: Air Mattress Not suitable for use in the presence of a
ammable anesthetic mixture (No AP or
APG protection)
14 760000-INS-LAB-RevA08
Page 15
Mattress Specication
Model 8" mattress
Dimensions (L x W x H) 78.7" x 35.4" x 8" (200 x 90 x 20.3 cm)
Weight 16.5 lb (7.5 kg)
Warranty
GF Health Products, Inc. warrants the Lumex AltaDyne Plus 760000 Alternating Pressure / Low Air Loss Mattress Sys­tem for a period of twelve months for defects in workman­ship and materials. During the warranty period, defective items will be repaired or replaced at GF Health Products' option at no charge.
760000-INS-LAB-RevA08 15
Page 16
USA, Corporate Headquarters:
Graham-Field Health Products
2935 Northeast Parkway
Atlanta, Georgia 30360
telephone: 800-347-5678, 770-447-1609
fax: 800-726-0601, 678-291-3232
www.grahamfield.com
760000-INS-LAB-RevA08
© 2008 GF Health Products, Inc.
Loading...