Graham Field 707A, 707AC, 707AX User Manual

USA, Corporate Headquarters:
Graham-Field Health Products
2935 Northeast Parkway
Atlanta, Georgia 30360
telephone: 800-347-5678, 770-447-1609
fax: 800-726-0601, 678-291-3232
www.grahameld.com
707A-INS-LAB-RevC08
Read this manual before operating the Labtron Digital Blood Pressure Monitor. Save this manual for future reference.
707A-INS-LAB-RevC08
Item 707A,
707A X, 707A C
Digital Blood Pressure
Monitor
w

User Manual
707A-INS-LAB-RevC08-book.indd 1 1/3/08 11:36:34 AM
Labtron Digital Blood Pressure Monitor 2
Labtron Digital Blood Pressure Monitor 19
GF, Graham-Field, and Labtron are registered trademarks of GF Health Products, Inc.
Velcro is a registered trademark of Velcro Industries B.V.
Packaging, warranties, products, and specications are subject to change without notice. GF Health Products, Inc. is not responsible for typographical errors.
CONTENTS
SAFETY GUIDELINES - PLEASE READ BEFORE USE ................................................................. 3
INTENDED USE ................................................................................................................................. 3
FEATURES ........................................................................................................................................
4
BATTERY INSTALLATION / REPLACEMENT ..................................................................................
5
BLOOD PRESSURE .......................................................................................................................... 5
PREPARING TO MEASURE BLOOD PRESSURE ................................................................... 6
MEASURING BLOOD PRESSURE ........................................................................................... 9
MAINTENANCE ...............................................................................................................................
13
TROUBLESHOOTING .....................................................................................................................
14
ERROR INDICATORS .............................................................................................................. 14
GENERAL TROUBLESHOOTING ........................................................................................... 15
SPECIFICATIONS * .........................................................................................................................
16
WARRANTY ..................................................................................................................................... 18
BLOOD PRESSURE MEASUREMENT LOG ..................................................................................
19
BLOOD PRESSURE MEASUREMENT LOG
Date Time
Pulse
Body
Condition
1/1 1/1 1/1
60
80
100
120
140
160
180
200
220
240
mmHg
128
134
123
84
90
76
70
73
69
7:00
13:30
20:00
707A-INS-LAB-RevC08-book.indd 2 1/3/08 11:36:34 AM
Labtron Digital Blood Pressure Monitor 18
Labtron Digital Blood Pressure Monitor 3
WARRANTY
GF Health Products, Inc. warrants the Labtron Digital Blood Pres­sure Monitor Model 707A/707A X/707A C for a period of one year for defects in workmanship and materials. During the warranty period, defective items will be repaired or replaced at manufacturer’s option at no charge.
SAFETY GUIDELINES - PLEASE READ BEFORE USE
Important! Read and understand these instructions before using the Labtron® Digital Blood Pressure Monitor. If you do not understand any part of these instructions, contact your medical professional or Graham-Field dealer for direction in the use of this product.
If components are damaged or missing, contact your Graham-Field dealer immediately. DO NOT use substitute parts.
GF Health Products, Inc. assumes no responsibility for any damage or injury caused by improper installation or use of this product.
INTENDED USE
The Labtron Digital Blood Pressure Monitor is intended to measure arterial blood pressure. This device is intended for use by adults, and not for use by children except under the supervision of an adult. This device is not intended to replace regular medical exami­nations. Review your procedure for using this monitor with your physician, who is the only person qualied to interpret blood pres­sure monitor results.
707A-INS-LAB-RevC08-book.indd 3 1/3/08 11:36:34 AM
Labtron Digital Blood Pressure Monitor 4
Labtron Digital Blood Pressure Monitor 17
FEATURES
Main features of the Digital Blood Pressure Monitor, referred to in this manual, are shown in the picture at right.
cuff
cuff
hose
LCD
bulb
ON/OFF button
exhaust valve
button
Digital Blood Pressure Monitor

Environmental conditions
Operation 50°F ~ 104°F (10°C ~ 40°C), 10% ~ 85% RH
Storage & transport
14°F ~ 140°F (-10°C ~ 60°C), 10% ~ 95% RH
Standard cited EN 1060-1 Noninvasive blood pressure measuring equipment
general requirements
EN 1060-3 Noninvasive blood pressure measuring equipment
supplementary requirements for electromechanical blood pressure measuring systems
EN 60601-1 Safety requirements for medical electrical equipment
EN 60601­1-2
Electromagnetic compatibility and safety for medical electrical equipment
EN 14971 Risk analysis for medical devices
* Specications are subject to change without notice
707A-INS-LAB-RevC08-book.indd 4 1/3/08 11:36:35 AM
Labtron Digital Blood Pressure Monitor 16
Labtron Digital Blood Pressure Monitor 5

Digital Blood Pressure Monitor Model
707A, adult-size cuff, ts 7.5 in. - 12.2 in. (19-31cm) arm 707A X, large adult-size cuff, ts 11.8 in. - 17.7 in. (30-45cm) arm 707A C, small adult-size cuff, ts 5.1 in. - 7.9 in. (13-29cm) arm
Measurement method Oscillometric
Display Digital LCD
Measurement range Pressure 20-280mmHg, Pulse 40-180bpm
Measurement accuracy Pressure ±3mmHg, Pulse ±5%
Ination system Manual ination with air pumping bulb
Deation system Automatic deation
Power 1.5V (AAA) alkaline batteries X 4
Automatic power-off Approximately 3 minutes after measurement
Reference method for clinical trials
Auscultatory measurement
Weight Approximately .75 lb (340g) without batteries
BATTERY INSTALLATION / REPLACEMENT
1. Slide the battery cover off in the direction of the arrow.
2. Install or replace the four AAA batteries, oriented as shown by battery compartment diagram, into the battery compartment.
3. Replace the battery cover. Ensure that the battery cover is securely fastened.
Note: Ensure that the batteries are correctly installed. Incorrect installation will prevent the device from operating.
BLOOD PRESSURE
Blood Pressure is a measure of the blood’s pressure in the circulatory system, which changes constantly during the course of the cardiac cycle. Blood pressure readings report two values. The higher reading (systolic pressure) shows the highest pressure
O P E N
707A-INS-LAB-RevC08-book.indd 5 1/3/08 11:36:35 AM
Labtron Digital Blood Pressure Monitor 6
Labtron Digital Blood Pressure Monitor 15
in the arteries occurring when the heart contracts. The lower reading (diastolic pressure) shows the lowest pressure in the arteries, which occurs right before the heart contracts. Blood pressure readings are written with the highest value rst, then the lowest value. Readings of 120/80 are considered to be normal, with high blood pressure being dened as a systolic pressure which is 140mmHg or more at rest and a diastolic pressure which is 90mmHg or more at rest. Only a patient’s physician is qualied to determine whether the readings obtained are normal for that person.
Preparing to measure blood pressure
1. Blood pressure uctuates continuously throughout the day. To consistently track your blood pressure, try to take your blood pressure at the same time each day.
2. Relax and remain still for 5 to 10 minutes before a blood pressure measurement, and between blood pressure measurements.
3. Refrain from eating, smoking, and drinking, especially alcoholic
General troubleshooting
Problem Recommended Action
Nothing appears in LCD
Ensure that batteries are correctly oriented in battery holder (polarity matches battery case placement indicators)
No measurement occurs
Ensure that cuff is positioned correctly Go over measurement procedure; measure again
Blood pressure measurements are too high or
too low
Ensure that user and cuff are positioned as instructed
Rest your arm on a table so that the cuff is level with your heart during
measurement Remain seated and still during entire measurement period Refrain from hand and body movements during measurement
Blood pressure measurements vary
Your blood pressure can uctuate considerably throughout the day. All of the following factors can inuence your blood pressure:
• Emotional state
• Daily activities / exercise
• Smoking
• Drinking alcoholic beverages
• Eating
• Taking certain medications
707A-INS-LAB-RevC08-book.indd 6 1/3/08 11:36:35 AM
Labtron Digital Blood Pressure Monitor 14
Labtron Digital Blood Pressure Monitor 7
TROUBLESHOOTING
Error indicators
Indication Cuff pressure reaches 330mmHg
Cause Twisted bladder inside cuff
Correction Ensure that bladder lies at within cuff
Indication "Err" displays, cuff pressure releases, then "0 pressure" displays
Cause Excessive arm movement or talking during measurement
Correction Turn monitor OFF then ON and measure again while remaining still
Indication Could not obtain pulse rate
Cause Cuff was too loosely applied, extremely weak pulse, or arrhythmia
Correction Reapply cuff and measure again; if problem persists after cuff is
correctly positioned, seek medical attention immediately
Batteries are exhausted. Replace with four new AAA alkaline batteries
beverages, before a measurement.
4. Remove any clothing on the upper arm to place the cuff directly on the skin. Constriction of the upper arm by clothing or a rolled-up sleeve, as shown at right, may cause an inaccurate reading; remove any constricting garments.
5. Sit down in a chair with back support and place both feet on the oor.
6. Extend the arm from which the blood pressure will be taken to the front or side. Rest your hand, palm-up, on a table as shown at right.
Note: Place the cuff at the same level as your heart to obtain accurate measurements. Your heart is located slightly to the right of your left armpit.
excessive
constriction
707A-INS-LAB-RevC08-book.indd 7 1/3/08 11:36:35 AM
Labtron Digital Blood Pressure Monitor 8
Labtron Digital Blood Pressure Monitor 13
7. Attach the cuff: Hold the area of the cuff where the hose exits the cuff rmly against the inside of the bare upper arm directly over the brachial artery, and the bottom of the cuff approximately one inch (2-3cm) above the elbow, as shown in the third pic­ture at right. Never place the cuff over clothing. Pull the opposite end of the cuff snugly around the arm and secure with the Velcro® fastener. The cuff should be snug, but not too tight. If one or two n­gers can t between the cuff and the arm, the cuff is properly secured. If you are unable to t the cuff as described above, a smaller or larger cuff should be used.
1 inch
MAINTENANCE
Recommended care and maintenance:
Do not drop or pull excessively on monitor components or subject them to strong impact.
Do not use monitor close to any device with a strong electrical eld such as television, microwave oven, X-Ray equipment, etc.
Never inate cuff above 300mmHg. Never inate cuff unless it is positioned on an arm.
Do not expose the monitor components to direct sunlight, high temperature, or high humidity.
Do not put any part of monitor in contact with sharp objects which could pierce the material and cause damage.
Do not dismantle or disassemble monitor components.
Cleaning Wipe off all components with a clean, dry cloth. Do not use solvents or other
petroleum-based cleaners. Do not press cuff with a hot iron.
Storage Always deate the cuff completely before storage.
Remove the batteries when the device is to be stored for extended periods of time. Note: Removing the batteries will erase all readings in memory.
707A-INS-LAB-RevC08-book.indd 8 1/3/08 11:36:36 AM
Labtron Digital Blood Pressure Monitor 12
Labtron Digital Blood Pressure Monitor 9
7. If performing measurement for the rst time, repeat the measure­ment two or more times to ensure that you and cuff are posi­tioned correctly and that results are accurate and consistent. Record your results; there is a log provided, with examples, at the end of this manual for this purpose. Always check with your phy­sician to ensure that readings are performed correctly. If you have concerns or questions about your results, always contact your physician. Only your physician is qualified to analyze blood
pressure.
Measuring blood pressure
Symbol key: LCD symbols you may see during measurement
Deating
Inating
Measurement error
Measuring
Low battery
1. Ensure that cuff is properly placed and that you are positioned as described in previous paragraphs.
2. Press the ON/OFF button. All LCD symbols will appear momen­tarily. If a downward arrow symbol displays, press the exhaust valve button (see earlier labeled picture) located between the ination bulb and the tube to release the residual pressure from the cuff.
707A-INS-LAB-RevC08-book.indd 9 1/3/08 11:36:36 AM
Labtron Digital Blood Pressure Monitor 10
Labtron Digital Blood Pressure Monitor 11
Note: The low battery symbol will also appear at this time while the
monitor checks the power level. This does not indicate that the batter­ies are low unless the low battery symbol remains on continuously.
3. After "0" displays, pressurize the cuff by squeezing the bulb until the pressure value displayed is 50mmHg above your usual systolic pressure. If you do not know your usual systolic pressure, squeeze the bulb just until 200mmHg displays.
4. The cuff will automatically deate after you stop pressurizing the cuff. If an upward arrow symbol displays, you must inate the cuff to a higher pressure than you did the rst time. If an upward arrow does not display, the monitor is taking your measurement. Remain calm and still to avoid causing any measurement error.
5. The systolic pressure, diastolic pressure, and pulse will display in the format shown at right. The device will completely deate when measurement is nished. To accelerate defla-
tion, press the exhaust valve button.
Systolic
Diastolic
Pulse rate
Note: To stop measurement in progress for whatever reason, press the exhaust valve button to release the pressure and turn off the device.
Note: If during measurement the power drops to an insufficient level, the device will end the measurement and display the low battery sym­bol. Replace the batteries and repeat the measurement.
Note: If the device cannot detect your pulse, it will end the measure­ment. Wait several minutes, ensure that you and the cuff are positioned properly, and try again.
6. The device will automatically shut off three minutes after com­pleting a reading, or you may turn it off manually by pressing the ON/OFF button.
707A-INS-LAB-RevC08-book.indd 10 1/3/08 11:36:36 AM