Grafco 8375, 8376 Operating Instructions Manual

DRY HEAT STERILIZERS
Catalog Nos. 8375/8376
(Both Models are Equipped with Automatic Timer Switch, Thermostat and Thermometer)
OPERATING INSTRUCTIONS Timer Switch
The timer switch serves as the ON/OFF switch for the sterilizer. To activate the timer and turn the sterilizer on, turn the knob clockwise (right) past the number “10”. You can then set the timer for the desired time as indicated on the calibrated scale dial.
To set the timer for continuous operation or “warm up”, turn the knob counter-clockwise (left) to the “HOLD” position. CAUTION: the sterilizer will remain on until it has been manually turned off or set for a timed sterilization cycle.
Thermostat & Thermometer
Before using the sterilizer for the FIRST time, it is necessary to establish a thermostat setting that will maintain the selected temperature.
For 320ºF (160ºC) set the number “7” on the thermostat dial to match and line up with the marking dot on the side of the sterilizer. Set the timer to “HOLD” position.
After approximately 30 minutes for the Model 200 or 45 minutes for the Model 300, check the thermometer temperature. The thermometer temperature should be near 320ºF (160ºC). The smaller the number of the dial setting, the lower the temperature setting. Temperatures may vary during the heat-up period.
Pilot Lights
The sterilizer has two (2) pilot lights. The left light will be on as long as the timer switch is in the “ON” position. The right light is on only while the heating elements are operating and heating the sterilization chamber.
Sterilization Procedures
Prepare all instruments for sterilization by using an Ultrasonic Cleaner, as recommended by the Centers for Disease Control (CDC). Thoroughly rinse the instruments and pat dry with a paper towel. The American Dental Association (ADA) recommends bagging the instruments prior to sterilizing. Sterilization bags specific to dry heat sterilization must be used. Instruments should be placed in a single layer and distributed evenly on the sterilizer’s trays. Instruments should not touch each other.
DO NOT OVERLOAD THE STERILIZER.
Maximum Loading
Model 200 Sterilizer (Cat. No. 8375)–1.3 pounds per tray (includes the weight of the tray) Model 300 Sterilizer (Cat. No. 8376)–1.4 pounds per tray (includes the weight of the tray)
Spore Test Strips The Centers for Disease Control (CDC) recommends the sterilizer be tested weekly. For Dry Heat Sterilizers, use Bacillus Atrophaeus spore test strips.
Power Source
Always check for correct voltage before operating sterilizer. Standard units operate on 115 VAC electrical power. Other voltages are available upon request. Voltage rating is printed on nameplate (located near power cord).
Time and Temperature Chart*
*Time cycle begins after sterilizer reaches operating temperature.
Note: Registration card accompanying your Dry Heat Sterilizer must be completed and returned within two weeks.
Distributed By: GF Health Products, Inc. Atlanta, GA 30360 Made in USA www.grahamfield.com
Revised 11/08