GENERAL DESCRIPTION
OF THE UNIVERSAL ANAESTHESIA MACHINE®
2
DEFINITION OF STATEMENTS2
SPECIFICATIONS3-4
COMPONENTS OF THE UAM®5-9
CHECKLISTS10
Prior To Every Operating Session 10
Prior To Every Patient Case 11
Following Every Patient Case 11
Following Every Operating Session 11
USING THE CONTROL SCREEN11
Calibrating The Oxygen Monitor 11
Calibrating The Apnoea Alarm Sensor 12
Setting The Oxygen Alarms 12
Setting The Apnoea Alarm On Or Off 12
FILLING THE VAPORIZER13
USING THE UAM 13
Balancing Gas Supply And Patient Demand 13-14
Using The Bellows 15
Using The Oxygen Concentrator 15
Using An Attached Oxygen Cylinder 16
Using A Pipeline Connector 16
Using The Emergency Oxygen Inlet 16
Using Nitrous Oxide 16
Using An Ayre’s T-Piece (ATP) 17
Using The UAM To Provide Oxygen For An Awake Patient 18
PREVENTATIVE MAINTENANCE19
Cleaning The Halothane Vaporiser 19
Replacing The Oxygen Sensor 19
Replacing The Oxygen Concentrator Filter 20
Cleaning The Breathing Circuit 20-22
APPENDIX ONE: BREATHING CIRCUIT DIAGRAM23
HOW TO GET HELP
EMAIL:
service@gradianhealth.org
WEB: www.gradianhealth.org
PHONE: +1.212.537.0340
ADDRESS: Gradian Health Systems, 160 Fifth Avenue, 7th Floor, New York, NY 10010, USA
GENERAL DESCRIPTION OF THE UNIVERSAL ANAESTHESIA MACHINE®
The UAM® combines elements from both continuous flow and draw-over designs to
provide a workstation that delivers controlled anaesthesia in all hospital settings. Oxygen
can be provided from a variety of sources, and if none of those sources is available the
machine will automatically draw in room air without the need for the user to change
settings. The vaporizer provides calibrated delivery of anaesthetic agent and the manual
bellows ensures that safe assisted respiration can continue without requiring a high
pressure gas source.
The system has been designed to accept standard anaesthesia accessories such as the
breathing circuit, oxygen sensor, and standard connectors for the oxygen and nitrous
oxide supplies. The UAM is also designed for simple field service. Most components can be
removed with standard tools and a replacement part installed by local technicians without
removing the workstation from the hospital.
The UAM carries the CE Mark, having passed through EU regulatory inspections. It
is manufactured in an ISO-certified factory in England. It is used in a wide variety of
hospitals, from internationally recognized academic centers to resource-constrained
district hospitals.
DEFINITION OF STATEMENTS
The following terminology and statements are important for the operator to understand
before consulting the user’s manual for the UAM:
WARNING: Indicate a possibility of injury to the operator or others.
CAUTION: Indicate a possibility of damage to the equipment.
NOTE: Indicate points of interest for proper operation of the equipment.
Gradian Health Systems, UAM, the Gradian logo, Universal Anaesthesia Machine are trademarks (™)
of Gradian Health Systems, LLC, registered ® in the U.S.
Printed in the United States of America. P/N: 1973-511 Issue A, January 2013 CE 0120
The UAM is manufactured by OES Medical Ltd., Abingdon, UK.
2
SPECIFICATIONS
PHYSICAL AND ELECTRICAL
g Workstation: 146cm x 53cm x 69cm, 130kg
g Top Shelf: 46cm x 31cm, 88cm above floor, 35kg weight limit
g Drawer: 10cm x 29cm x 32cm, 35kg weight limit
g Materials: Aluminum frame, vertical dovemount side rails on both sides,
nylon internal tubing
g Mains power supply: 240V, 50-60Hz
g Oxygen monitor power: CE-marked, fused and medical grade
g Isolator protects concentrator, monitor and sockets
g 3.5” TFT touch screen for oxygen display and alarm setting
g Membrane switch for oxygen and alarm settings
g Front lockable castors
OXYGEN SUPPLY AND MONITORING
g Oxygen concentrator flow rate: 0.1 to 10.0 liters per minute
g Maximum oxygen concentration: 95%
g Alternative sources: cylinder yoke, pipeline or portable
g Automatic room air intake when patient minute volume exceeds fresh gas flow
g Accuracy of glass rotameters: +/- 2.5% when using 100% O2
g Integrated inspiratory oxygen monitoring uses MOX-3 fuel cell
g Calibration for room air (21%) and 100% oxygen
g Pressure sensitive apnoea or high flow alarm
g Adjustable minimum and maximum oxygen alarm settings
g Up to ten hour battery backup, trickle recharge from mains power
g Membrane keypad and touch screen
OPTIONAL USE OF NITROUS OXIDE
g Sources: pipeline or cylinder
g Hypoxic cut-off: A solenoid automatically shuts off nitrous oxide delivery if O2 level
falls below a minimum of 25%
g Differentiated flow control knob
g Independent glass rotameters accurate to +/- 2.5%
VENTILATION
g Unique bellows for adult and pediatric use made from durable,
long-lasting silicone rubber
g 1600cc capacity
g Inspiratory pressure relief valve set at 55cm water
g Aluminum bellows block
g Silicone balloon inflating valve at the expiratory port
g 15mm female/22mm male ISO standard taper connection for breathing circuit
g 30mm male ISO standard taper connection for AGSS scavenging equipment
3
VAPORIZERS
g Stainless steel and plated brass construction
g Separate units for isoflurane and halothane
g Agent delivery range: 0.5% to 4%
g 120ml capacity
REAR PANEL CONNECTORS
g Mains isolator switch, 2 UK style power sockets
g Oxygen and nitrous pipeline NIST connectors and cylinder yokes
g Multi-diameter external oxygen hose connector
4
COMPONENTS OF THE UAM
Patient Vital Signs Monitor
Oxygen monitor and
control screen
Oxygen and Nitrous Oxide
Flow Meters
Backbar with air inlet,
pressure relief valve and
2 liter reservoir bag
Low resistance vaporizer
Oxygen concentrator
on/off switch
Oxygen cylinder and
pipeline pressure gauges
Nitrous Oxide cylinder
and pipeline gauges
SEE INSET PAGE 6
Bellows
Pressure relief valve
and gauge
Fenton Balloon
inflating valve
Y-piece connectors
Water trap
Oxygen concentrator
status screen
Integrated oxygen
concentrator
Castors (front lockable)
5
Oxygen monitor
and control screen
Oxygen and Nitrous
Oxide Flow Meters
Backbar
a. Air inlet valve
b. Pressure relief valve
c. Reservoir bag
Low resistance vaporizer
Oxygen concentrator
on/off switch
Oxygen cylinder and pipeline
pressure gauges
Nitrous Oxide cylinder
and pipeline gauges
6
Bellows
Pressure relief gauge
Oxygen sensor
1st one-way valve
2nd one-way valve
Pressure relief valve
Y-piece connectors
Water trap
7