Gradian UAM Quick user guide

UAM VENTILATOR
QUICK SET UP GUIDE
1
Setting Up the Ventilator
preparing the UaM and the Ventilator
1. Ensure that equipment such as the ventilator and the patient monitor are securely attached to the top shelf of the UAM
2. Ensure that all cables and sample lines are correctly attached – refer to the manual of each individual manufacturer.
3. Connect these essential items:
a. Patient Circuit with Y piece. b. Ventilator to UAM using the ventilator drive hose. c. Ventilator patient flow sensor between the patient Y piece and the
bacteria filter. d. Ventilator patient flow sensor sample lines at rear of ventilator e. Patient bacterial filter to prevent flow sensor contamination f. Theatre anaesthetic gas scavenging system
Ventilator ConneCtionS
1. Flow sensor connection – The patient inspired and expired
gas is measured with a variable orifice flow sensor. a. Distal to patient – The clear tube is connected furthest from the patient. b. Proximal to patient – blue tube. The blue tube is connected closest to
the patient.
note: A breathing circuit filter must be used to prevent contamination.
2. Bellows connection – 17 mm male taper that is attached to the UAM
anaesthesia machine using a 15 mm x 1.5 m black corrugated breathing hose.
3. Mains power inlet socket-IEC power cable.
3
4
2
1
2
perforMing the pre-USe CheCkS
1. Check the ventilator for labeling to indicate if the machine has any faults or
needs to be serviced
2. Check for visible signs of damage.
3. Check the correct connection of the patient circuit and any auxiliary
equipment such as patient monitoring equipment, gas scavenger etc.
4. Perform a patient circuit leak check: a. With the patient circuit connected check that all connections are secure
and all tubes used are in good condition.
b. With the UAM anaesthesia machine in manual bellows mode block
the patient Y piece and ensure that the manual bellows on the UAM anaesthesia machine can generate 30 cmH2O or more pressure when the bellows is pushed down without the bellows descending due to a leak.
iSolating the hand BellowS
Push the lever to the rear to isolate the hand bellows and pull it forward when the hand bellows is in use.
The Manual Bellows Isolator lever isolates the Hand bellows from the breathing circuit when the ventilator is being used.
MAINS INDICATOR
Yellow: Mains applied
tUrning the Ventilator on
To turn the ventilator on press the
Green: Ventilator ON
ON/OFF push button
On/O push button for 1 second.
Start Up Self teSt
If the patient circuit was connected to a patient or a test lung, the ventilator will alert you to remove the Y piece and press to repeat tests.
When the ventilator turns on it will perform a series of self-tests and the bellows will move up and down. The screen will be blank at this time.
note: During these tests, the breathing circuit must be disconnected from the patient.
3
patient SeleCt SCreen
Once the self-tests are over the patient select screen will appear.
The patient select screen allows the user to choose patient ventilator settings by weight – 3kg, 10kg, 25kg, 50kg, 70kg and 100kg or by selecting one of 6 user-defined settings.
Connect, pulse oximeter probe, NIBP cu and ECG leads for patient monitoring.
Patient select screen with body weight setting for 50Kg selected. Note that the selected key is now black and that the ventilator settings for this weight are displayed on the left hand side
Starting the Ventilator
Observe that no alarms are indicated in the alarms dialogue box – this indicates that there are no system errors, including mains power, battery failures. The alarms are disabled until the ventilator is switched to the “start” screen allowing the ventilator to be turned on without nuisance alarms.
Once the user has chosen the initial settings the actual settings can be adjusted as required prior to starting to ventilate. If the settings are to the users satisfaction, pressing the green “Start Case” button shown on the lower left hand side of the display will start the ventilator.
Patient settings are displayed in the “SET” column on the right hand side vertical column. These settings can be adjusted before or during a case. All functions require “OK” to be pressed to confirm the setting change. If the setting change is made and the OK button is not pressed the screen will time out and stay at the original setting.
4
Loading...
+ 9 hidden pages