Gradian iM8B_manual Equipment Packet: Oxygen Concentrator

Page 1
EDAN INSTRUMENTS, INC.
Patient Monitor
Multi-Parameter
Manual Ver: V1.1 Release Date: Apr. 2008 Part Numbe: MS1R-100464 -V1.1
Page 2
EC Conformity Information
Manufacturer: EDAN Instruments , Inc.
Addr: 3/F-B, Nanshan Medical Equipm ents Park, Nanhai R d 1019#, shekou ,
Nanshan Sh enzhen, 518067 P.R. China
European Representative: Shangh ai International Holding C orp. Gmb H (Europe)
Addr: Eiffestrasse 8 0 D-20537 Hamburg German y
Product: Patient Moni tor
Model: M8B
Classificatio n (MDD, Annex IX): IIb
We herewith declare that the above mentioned product(s) meet the transposition into
national law, the provisions of Council Directive 93/42/ EEC of 14 June 19 93
concerning medical devices-as amended b y Directive 98/79/EC on in vitro
diagnostic medical devices.
All supporting d ocumentation is ret ained at the premises of the manufacturer.
DIRE CTI VES
General Applicable Directives:
Medical Device Directive: COUNCIL DIRECTIVE 93/42/EEC of 1 4 June 1993
concerning medical devices (MDD 93/ 42/EEC).
Standards applied: EN6060 1-1:1990+A1:19 93+A2:1995, IEC60 601-1-2:2001,
EN60601-2-27:1994, IEC606 01-2-30:1999, EN60601-2-34:2000,
IEC60601-2-49:200 1, IS O99 19:2005, EN12470-4:2001, EN106 0-1:1995,
EN1060-3:19 97, ANSI/ AAMI SP-10:2002, ISO9001:2000, ISO13485:2003, EN
ISO14971:2000, IEC 6 0601-1:1988, BS EN60601-1-4:1997, ISO 21647:2004,
ANSI/AAM I SP-1 0-200 2, EN1041:1998, IEC /TR 60878-2003, EN980:2003, EN
60417-2-20 00, EN475:1995, EN540:1993, ISO1 000:1998, EN ISO780:1999 .
Notified Body: TÜV SÜD Product Service GmbH, R idlerstr 65, D-80339 MÜnchen,
German y.
Identification number
Page 3
Copyright
Copyright b y EDAN Instruments, Inc. 2008. All rights reserved.
Statement
EDAN Instruments, Inc. (hereinafter called EDAN) makes no w arranty o f any kind
with regard to t his material, incl u ding, but not limited to the implied warranties of
merchantability and fitness for a particular purpos e. EDAN owns all rights to this
unpublished work and intends to maintain t his work as confidential. EDAN may als o
seek to maintain this work as an unpublished copyright. This publication is to be
used solely for the purposes of reference, operation, maint enance, or repair of this
compan y equipment. No part of this can be disseminat ed for other purpos es.
In the event of inadvertent or deliberate publication, EDAN intend to enforce its
rights to this work under copyright laws as a published work. Those h aving access to
this work may not copy, u se, or disclose the information in t his work unless
expressl y authorized by thi s company to do so.
All information contained in this p ublication is believed to be correct. EDAN shall
not be liable for errors contained herein nor for incid ental or consequential damages
in connection with the furnishing, performance, or use of this material. This
publication may refer to information and protected by copyrights or patents and does
not convey any license under the patent rights of this company, nor the rights of
others. T his company does not assume any liability arising out of any infringements
of patents or other rights of third parties.
Content of this manual is subject to changes without prior notice.
Responsibility of the Manufacturer
This company is responsible for safety, reliabilit y and perform ance o f t his
equipment only in the conditio n that:
All install ation, expansion, change, modification and repai r of this equipment are
conducted by EDAN quali fied personnel; and,
Applied electrical appliance is in compliance with relevant National Standards;
I
Page 4
and,
The monitor is o perat ed under strict observance of this manual.
NOTE:
This equ ipment is not intended for family usage.
WARNING
This monitor is not a de vice for treatment purpose.
Using This Label Guide
This guide is desi gned to give key concepts on safety precautions.
WARNING
A WARNING label advises agai nst certain a cti ons or situations that could
result in personal injury or death.
CAUTION
A CAUTION l abel advises against actions or situations that could damage
equipment, produce inaccurate d ata, or invalidat e a proced ure.
NOTE: A NOTE provides useful information regarding a function or procedure.
Warranty
Workma nship & Materials
EDAN guarantees new equipment other than accessories to be free from defects in
workmanship and materials for a period of one year (six m onths for multi-site probes
and SpO2 sensor) from d ate of shipment under normal u se and service. This
compan y's obligation u nder this warranty is limited to repairing, at this companys
option, an y part which upon this company's ex aminati on proves defective.
Exempti ons
This company's obligation or liability under t his warranty does not incl ude any
transportation or other charges or liability for direct, indirect or consequential
damages or delay resulting from the improper use or application of the product or the
II
Page 5
substitution upon i t of parts or accessories not approved by this company or repaired
by an yone other than this compan y authorized representative.
This warranty shall not extend to any instrument which has been subjected to misuse,
negligence or accident; any instrument from which this compan y's ori gin al serial
number tag or product identification markings have been altered or removed, or any
product of an y other manufacturer.
Safety, Reliability and Performance
This company is not responsibl e for the e ffects o n safet y, reliabilit y and performance
of the Multi-parameter Patient Monitor if:
Assem bl y operations, exten sio ns, re-adjusts, modifications or repairs are carried
out by persons other than th ose authorized by this compan y.
The Multi-parameter Patient Monitor is not used in accordance wi th the
instructions.
Return Policy
Return Procedure
In the event that it becomes necessary to return a unit to EDAN, the following
procedure should be followed:
1) Obtain return aut horization. Contact EDAN Service Department and obtain a
Customer Service Authorization number. The Company number must appear on
the outside of the shipping con tainer. Return shipm ents will not be accepted if
the Company number is not clearl y visible. Please provide the model number,
serial number, and a brief description of the reason for return.
2) Freight policy. The customer is respon sible for freight charges when equipment
is shipp ed to EDAN for service (this includes customs charges).
III
Page 6
Revision History
This user ma nua l shall be revised whenever cha n ges in regulatory requirements di c tate.
Dat e ECO Version Revision history
21st, Jan. 2 008 V1.0
8th, Apr. 2008 ECO-MPM-8005 V1.1
1st edition
Add CE information; add
3-channel recording.
IV
Page 7
Contents
1 Introduction ....................................................................................................................1
1.1 General Information ................................................................................................ 1
1.2 Screen Display........................................................................................................ 2
1.3 Button Functions..................................................................................................... 5
1.4 Interfaces................................................................................................................ 7
1.5 Built-in Chargeable Battery ..................................................................................... 9
2 Installation of Monitor.................................................................................................. 11
2.1 Open the Package and Check .............................................................................. 11
2.2 Install Wall Mount for M8B (Optional).................................................................... 11
2.3 Connect the Power Cables ................................................................................... 13
2.4 Power on the Monitor............................................................................................ 13
2.5 Connect Patient Sensors ...................................................................................... 14
2.6 Check the Recorder.............................................................................................. 14
3 System Menu ................................................................................................................ 15
3.1 Patient Information Setup .....................................................................................15
3.2 Default Setup ........................................................................................................ 17
3.3 Mark Event............................................................................................................ 18
3.4 Face Select........................................................................................................... 19
3.5 Time Setup............................................................................................................ 19
3.6 Recorder Setup..................................................................................................... 20
3.7 Module Setup........................................................................................................ 22
3.8 Tracing Waveforms Selection ............................................................................... 22
3.9 Monitor Version..................................................................................................... 23
3.10 Alarm Volume...................................................................................................... 23
3.11 Key Volume......................................................................................................... 24
3.12 Drug Calculation ................................................................................................. 25
3.13 Waveform Demonstration ................................................................................... 25
3.14 Maintenance ....................................................................................................... 25
V
Page 8
4 Face Select.................................................................................................................... 32
4.1 Select Operating Screen....................................................................................... 32
4.2 Standard Screen ................................................................................................... 32
4.3 Trend Screen ........................................................................................................ 33
4.4 oxyCRG Screen.................................................................................................... 34
4.5 Viewbed Screen.................................................................................................... 35
4.6 Large Font Screen ................................................................................................ 36
5 Alarm ............................................................................................................................. 39
5.1 Alarm Modes......................................................................................................... 39
5.1.1 Alarm Level.................................................................................................. 39
5.1.2 Alarm Modes................................................................................................ 39
5.1.3 Alarm Setup .................................................................................................41
5.2 Alarm Cause ......................................................................................................... 42
5.3 SILENCE ..............................................................................................................43
5.4 Parameter Alarm................................................................................................... 43
5.5 When an Alarm Occurs......................................................................................... 44
6 Freeze............................................................................................................................ 45
6.1 General ................................................................................................................. 45
6.2 Enter/Exit Freeze Status .......................................................................................45
6.3 FROZEN Menu ..................................................................................................... 46
6.4 Reviewing Frozen Waveform................................................................................ 46
7 Recording (Optional).................................................................................................... 47
7.1 General Information on Recording........................................................................ 47
7.2 Recording Type..................................................................................................... 47
7.3 Recording Startup ................................................................................................. 49
7.4 Recorder Operations and Status Messages ......................................................... 50
8 Trend and Event ........................................................................................................... 52
8.1 Trend Graph.......................................................................................................... 52
8.2 Trend Table ........................................................................................................... 53
8.3 NIBP Recall .......................................................................................................... 55
VI
Page 9
8.4 Alarm Event Recall ............................................................................................... 56
9 Drug Calculation and Titration Table (Optional) ........................................................ 58
9.1 Drug Calculation ................................................................................................... 58
9.2 Titration Table ....................................................................................................... 60
10 Safety Guidance ......................................................................................................... 62
11 Maintenance/Cleaning................................................................................................ 70
11.1 System Check ..................................................................................................... 70
11.2 General Cleaning ................................................................................................ 70
11.3 Cleaning Agents.................................................................................................. 71
11.4 Sterilization.......................................................................................................... 72
11.5 Disinfection ......................................................................................................... 72
11.6 Replacement of Fuse .......................................................................................... 72
12 ECG/RESP Monitoring ............................................................................................... 73
12.1 What Is ECG Monitoring ..................................................................................... 73
12.2 Precautions during ECG Monitoring ................................................................... 74
12.3 Monitoring Procedure.......................................................................................... 74
12.3.1 Preparation ................................................................................................ 74
12.3.2 Installing ECG Lead ................................................................................... 75
12.4 ECG Screen Hot Keys ........................................................................................ 80
12.5 ECG Menu .......................................................................................................... 81
12.6 ECG Alarm Information....................................................................................... 87
12.6.1 Alarm Message .......................................................................................... 87
12.7 ST Segment Monitoring (Optional)...................................................................... 90
12.8 Arr. Monitoring (Optional) .................................................................................... 93
12.9 Measuring RESP ................................................................................................ 99
12.9.1 How to Measure RESP? ............................................................................ 99
12.9.2 Setting up RESP Measurement ................................................................. 99
12.9.3 Installing Electrode for RESP Measurement .............................................. 99
12.9.4 RESP Menu ............................................................................................. 100
12.9.5 RESP Alarm Message.............................................................................. 102
VII
Page 10
12.10 Maintenance and Cleaning ............................................................................. 102
13 SpO2 Monitoring ....................................................................................................... 104
13.1 What is SpO2 Monitoring................................................................................... 104
13.2 Precautions during SpO2/Pulse Monitoring....................................................... 105
13.3 Monitoring Procedure........................................................................................ 106
13.4 Limitations for Measurement............................................................................. 107
13.5 SpO2 Menu ....................................................................................................... 107
13.6 Alarm Description.............................................................................................. 109
13.7 Maintenance and Cleaning ................................................................................111
14 NIBP Monitoring ....................................................................................................... 112
14.1 Introduction ....................................................................................................... 112
14.2 NIBP Monitoring................................................................................................ 113
14.2.1 NIBP Measuring....................................................................................... 113
14.2.2 NIBP Parameter Setup and Adjustment ................................................... 116
14.3 NIBP SETUP menu........................................................................................... 117
14.4 NIBP Alarm Message and Prompt Message..................................................... 120
14.5 Maintenance and Cleaning ............................................................................... 123
15 TEMP Monitoring...................................................................................................... 125
15.1 TEMP Monitoring .............................................................................................. 125
15.2 TEMP SETUP Menu......................................................................................... 126
15.3 TEMP Alarm Message ...................................................................................... 127
15.4 Care and Cleaning............................................................................................ 128
16 IBP Monitoring (Optional) ........................................................................................ 130
16.1 Introduction ....................................................................................................... 130
16.2 Precautions during IBP Monitoring.................................................................... 130
16.3 Monitoring Procedure........................................................................................ 131
16.4 IBP Menu .......................................................................................................... 132
16.5 Alarm Information.............................................................................................. 140
16.6 Maintenance and Cleaning ............................................................................... 141
17 Accessories and Ordering Information .................................................................. 143
VIII
Page 11
18 Service Information.................................................................................................. 146
Appendix.................................................................................................................... 147
A1.1 Classification .................................................................................................... 147
A1.2 Specifications ................................................................................................... 147
A1.2.1 Size and Weight ...................................................................................... 147
A1.2.2 Environment ............................................................................................ 147
A1.2.3 Display..................................................................................................... 148
A1.2.4 Battery ..................................................................................................... 148
A1.2.5 Recorder (Optional)................................................................................. 148
A1.2.6 Recall ...................................................................................................... 148
A1.2.7 ECG......................................................................................................... 148
A1.2.8 RESPIRATION ........................................................................................ 150
A1.2.9 NIBP ........................................................................................................ 151
A1.2.10 SpO2...................................................................................................... 152
A1.2.11 TEMPERATURE .................................................................................... 152
A1.2.12 IBP......................................................................................................... 153
Appendix ................................................................................................................... 154
A2.1 Electromagnetic Emissions-for all EQUIPMENT and SYSTEMS ..................... 154
A2.2 Electromagnetic Immunity - for all EQUIPMENT and SYSTEMS ..................... 154
A2.3 Electromagnetic Immunity-for EQUIPMENT and SYSTEMS that are not
LIFE-SUPPORTING.................................................................................................. 155
A2.4 Recommended Separation Distances .............................................................. 156
IX
Page 12
M8B Multi-Parameter Patient Monitor User Manual
1 Introduction
1.1 General Information
M8B Multi-parameter patient monitor monitors parameters such as ECG, RESP, SpO2, NIBP,
dual-TEMP and dual-IBP (optional). It is adaptable to adult, pediatric, and neonatal usage. The user
can select different parameter configuration according to different requirements.
It centralizes the function of parameter measurement module, display, record and output to compose
a compact, portable device. Its built-in replaceable battery provides convenience for patient
movement. On the high-resolution display screen, 7 waveforms and all the monitoring parameters
can be displayed clearly. The POWER switch is on the left of the front panel (Figure 1-1). The POWER indicator lights when the monitor is powered on (Figure 1-1). The CHARGE indicator shows the charging status (Figure 1-1). The ALARM indicator flashes when the alarm is triggered (Figure 1-1). The
sockets of various sensors are on the left panel. Other sockets and power plug-in are on the rear
panel. The recorder is on the right panel.
M8B Multi-parameter patient monitor is a user-friendly device with operations conducted by a few
buttons and rotary knob on the front panel (Figure 1-1⑤⑥). Refer to Button Functions.
Figure 1-1 Button Functions of M8B
- 1 -
Page 13
M8B Multi-Parameter Patient Monitor User Manual
Figure 1-2 M8B Multi-parameter Patient Monitor
M8B Multi-parameter patient monitor can monitor:
ECG: Heart Rate (HR),
2-channel ECG waveform,
Arrhythmia and ST-segment analysis (optional);
RESP: Respiration Rate (RR),
Respiration Waveform;
SpO2: Oxygen Saturation (SpO2), Pulse Rate (PR),
SpO2 Plethysmogram;
NIBP: Systolic Pressure (NS), Diastolic Pressure (ND), Mean Pressure (NM);
TEMP: Channel-1 Temperature (T1), Channel-2 Temperature (T2),
Temperature Difference between two channels (TD);
IBP: Channel-1 SYS, DIA, MAP,
Channel-2 SYS, DIA, MAP,
Dual-IBP waveforms;
M8B Multi-parameter patient monitor provides extensive functions as visual & audible alarm,
storage for trend data, NIBP measurements, alarm events, and drug dose calculation and so on.
1.2 Screen Display
M8B Multi-parameter patient monitor is equipped with a high-resolution multicolor TFT LCD
screen. The patient parameters, waveforms, alarm messages, bed number, time, monitor status and
other data can be reflected from the screen.
The screen is divided into three areas: 1 Information Area; 2 Waveform Area; 3 Parameter Area.
- 2 -
Page 14
M8B Multi-Parameter Patient Monitor User Manual
Figure 1-3 Main Interface of M8B
Information Area ( )
The Information Area is at the top and bottom of the screen, displaying operating state of the
monitor and status of the patient.
The information area contains following data:
Bed number of the monitored patient
ADU Type of patient. Three options: Adult, Pediatric, Neonate Name Name of the monitored patient. When the user input patients name, this
name will be displayed on the right side of the Patient type. If the user
does not input patient name, this position will be vacant.
12-05-2007 Current date
"11:33:23" Current time
Indicates the status of mains power supply
means the mains power supply is on,
means the mains power supply is off.
Indicates the battery and its capacity in the monitor
means there is a battery in the monitor, and the green part
indicates the battery’s charge.
means there is no battery is equipped in the monitor.
Indicates the audio alarm is turned off.
- 3 -
Page 15
M8B Multi-Parameter Patient Monitor User Manual
Indicates the audio alarm is paused.
Indicates the alarm for parameters is turned off.
Other information of the Information Area comes up only with respective monitoring status. They
are:
Signs indicating the operating status of the monitor and the sensors are displayed at the right
side of patient name.
Alarm message is displayed at the right most area. FREEZE appears when the waveforms are frozen.
Waveform Area ()
Seven waveforms can be displayed at the same time. The sequence of waveforms can be adjusted.
Under the maximum configuration, the system can display 2 ECG waveforms, SpO2 waveform,
Respiration waveform (can be from ECG module), 2 IBP waveforms.
In the TRACE SETUPmenu, all the waveforms are listed. The user can select the waveform to be displayed, and adjust the display position. Refer to Waveform Sequence Selection” for details.
The name of the waveform is displayed on the upper left part of the waveform. The name of ECG is
user-selectable. Gain and filter way of this channel are displayed as well. A 1mv scale is marked on
the right of ECG waveform. The IBP waveform scale can also be selected according to the actual
requirement. Its range is described in the part: IBP Monitoring. In the IBP waveform area, the
waveform scale is displayed. The three dotted lines for each IBP waveform from up to down
represent respectively the upper limit scale, reference scale and lower limit scale. The values of
these three scales can be set. The specific method is given in the part: IBP Monitoring.
When a certain menu is displayed, some waveforms become invisible. Main display is restored
when you exit the menu.
The user may set up the rate to refresh the waveform. The method to adjust the refreshing rate of
each waveform is discussed in the setup description of each parameter.
Parameter Area()
Parameter area is on the right of Waveform area, and parameters are displayed corresponding to
waveforms basically. They are:
ECG:
Heart Rate (Unit: bpm) ST-segment analysis of Channel 1 & 2 – ST1, ST2 (Unit: mV) PVCs (Premature Ventricular Contraction) events (Unit: event/min)
Oxygen Saturation (SpO2):
SpO2 (Unit: %)
- 4 -
Page 16
M8B Multi-Parameter Patient Monitor User Manual
NIBP:
(From left to right) Systolic pressure, Mean pressure, Diastolic pressure (Unit: mmHg or
kPa)
TEMP:
Temperature of channel1 and channel2 and their temperature difference: T1, T2, TD (Unit:
°C or °F)
RESP:
Respiration Rate (Unit: breath/min)
IBP:
The blood pressure of channel 1 and 2. From left to right, there are Systolic pressure, Mean
pressure and Diastolic pressureunit: mmHg or kPa)
Alarm Indicator and Alarm Status
Under normal status, the alarm indicator does not light.
When alarming, the alarm indicator lights or flashes. The color of light represents the alarm level. Refer to Alarm Functions for details.
Refer to relative content of parameter for Alarm information and prompt.
Charge Indicator and Charge Status
The light can indicate the status of charging. When the battery is charged, the light color will turn to
orange.
1.3 Button Functions
All the operations to M8B Multi-parameter patient monitor can be finished by several buttons and a
knob. They are:
Figure 1-4 Buttons
- 5 -
Page 17
M8B Multi-Parameter Patient Monitor User Manual
Menu Press to call up the SYSTEM MENU. Refer to Chapter System
Menu and Chapter Trend and Event for details.
Start Press to fill gas into cuff and start blood-measuring. During the
measuring process, press the button to stop measuring and
deflating.
Record
Silence
Press to start a real time recording. The recording time is set in
RT REC TIME of RECORD submenu.
When the SYSTEM MENU MAINTAIN USER MAINTAIN ALARM SETUP is set ON Press this button
to enter the status of ALARM SILENCE. All the alarm audio
will be closed. At the same time, Alarm Pause ×× s and
will be displayed in the interface. When repress it or the pause
time is over, the system will stop pausing, and resume the normal monitoring status, and Alarm Pause ×× s and icon
will vanish.
Press this button and hold for more than 3 seconds can turn off
the audio alarm. displays on interface. Press or hold the
button again can resume the alarm.
Freeze
Main
Rotary Knob
NOTE:
Whether an alarm will be reset depends on the status of
the alarm cause. But by pressing SILENCE button
(suspend alarm) can permanently shut off audio sound of
the Lead Off or Sensor Off alarms. So the user can exit
the Alarm Silence Status by Technical Alarm.
When in normal mode, press this button to freeze all the
waveforms on the screen. When in FREEZE mode, press to
restore the waveform refreshing.
Press to return to the main interface.
The user may use the rotary knob to select the menu item and
modify the setup. It can be rotated clockwise or anticlockwise
and pressed like other buttons. The user may use the knob to
realize the operations on the screen and in the system menu
and parameter menu.
- 6 -
Page 18
M8B Multi-Parameter Patient Monitor User Manual
Method to Use the Knob to Operate on the Screen:
The rectangular mark on the screen that moves with the rotation of the knob is called cursor.
Operation can be performed at any position at which the cursor can stay.
When the cursor is in the waveform area, the user may immediately modify the current setup. When
the cursor is in the parameter area, the user may open the setup menu of the corresponding
parameter module so as to set up the menu items of the module.
Operating method:
Move the cursor to the item where the operation is wanted.
Press the knob.
One of the following four situations may appear:
1 The cursor with background color may become into the frame without background
color, which implies that the content in the frame can change with the rotation of the
knob.
2 Menu or measuring window may appear on the screen, or the original menu is replaced
by the new menu.
3 A check mark appears at the position, indicating that the item is confirmed. 4 The system immediately executes a certain function.
1.4 Interfaces
For the convenience of operator, interfaces of different function are in different sites of the monitor.
Right Side of the Monitor
At the right side of the monitor, there is the bracket of hydrophobic bottle from the recorder’s paper inlet cover ().
Figure 1-5 Right panel
- 7 -
Page 19
M8B Multi-Parameter Patient Monitor User Manual
② ③ ④ ⑤ ⑥ ⑦
Left Side of the Monitor
Connectors for cables and sensors are as shown in Figure 1-6.
TEMP1 probe connector TEMP2 probe connector IBP2 transducer connector IBP1 transducer connector ECG cable connector
NIBP cuff connector
SpO2 sensor connector
IBP2
S
T1
T2
CO
O2
p
CO2
BP1
I
ECG
BP
I
N
Figure1-6 Left Panel
This symbol means BE CAREFUL". Refer to the manual.
Indicates that the instrument is IEC60601-1 Type CF equipment. The unit displaying this
symbol contains an F-Type isolated (floating) patient applied part providing a high degree of
protection against shock, and is suitable for use during defibrillation.
Rear Panel
- 8 -
Page 20
M8B Multi-Parameter Patient Monitor User Manual
Figure 1-7 Rear panel of M8B
Sockets on the rear panel are shown in Figure 1-7.
Network Interface (reserved): Standard RJ45 Socket, for connecting to MFM-CMS of EDAN. Equipotential grounding terminal for connection with the hospitals grounding system. Fuse box, used to put fuse in. Power supply socket: AC100-240 V, 50/60 Hz.
1.5 Built-in Chargeable Battery
M8B Multi-parameter Patient Monitor is equipped with a built-in chargeable battery. When switch
on AC power supply, the battery will be charged automatically until full electric energy. There is a sign in the lower left corner of screen to show the charging status, and the green part is
the electric energy of battery. When the monitor is not equipped with battery, the battery status will
be showed as the sign , which means no battery.
One battery can support the working of monitor. Under the cable connectors is the cover of battery
slot. See figure1-8 Battery Slot cover.
The charging time for battery resume to 90% quantity of electric charge needs 6 hours.
- 9 -
Page 21
M8B Multi-Parameter Patient Monitor User Manual
OPEN
Figure 1-8 Battery Slot Cover
Replace Battery
When the lifespan of battery is over, or foul odor and leakage has been detected, please contact with
manufacturer or local distributor for replacement of battery.
WARNING
Do not unplug the battery when monitoring. The unexpected power off can not impact the device normal working, for it has battery for standby.
WARNING
Stop using the battery if abnormal heat, odor, discoloration, deformation or abnormal
condition is detected during use, charge, or storage. Keep it away from the monitor.
WARNING
Make sure the device is used in the appointed range of voltage, the effect of power supply can be not noticeable.
- 10 -
Page 22
M8B Multi-Parameter Patient Monitor User Manual
2 Installation of Monitor
Open the package and check
Install wall mount for M8B
Connect the power cables
Power on the monitor
Connect patient sensors
Check the recorder
NOTE:
To ensure that the monitor works properly, please read Chapter Safety Guidance, and
follow the steps before using the monitor.
2.1 Open the Package and Check
Open the package and take out the monitor and accessories carefully. Keep the package for possible
future transportation or storage. Check the components according to the packing list.
Check for any mechanical damage.
Check all the cables, modules and accessories.
If there is any problem, contact the distributor immediately.
2.2 Install Wall Mount for M8B (Optional)
Figure 2-1
- 11 -
Page 23
M8B Multi-Parameter Patient Monitor User Manual
Figure 2-2
Figure 2-3
Installation Steps:
1. Outline along the part 1 edge against the wall with the desired location.
2. Drill 4 holes of 6 mm diameter in the wall, and knock the plastic nuts in the holes completely.
3. Knock 4 screws in the plastic nuts, when tighten them, secure the channel (part1) at the desired
location of the wall (as shown in Figure 2-1).
4. Slide the arm (part2) into channel (part1) as Figure2-2 shown to the desired location.
5. Tighten the manual screw and secure the arm (part2) at the desired position within the channel
(part1).
6. Face the side with duplexing piece of the device holding plate to the underside of the M8B, fix
it with M4×25mm screw (as shown in Figure 2-3).
7. Pull out the fixing screw from the fixing hole.
8. At the same time, put the arm (part2) into the M8B with the device holding plate, and secure the
fixing screw totally through the fixing hole.
9. The M8B can be adjusted maximum 15 degrees by rotating the black knob.
- 12 -
Page 24
M8B Multi-Parameter Patient Monitor User Manual
2.3 Connect the Power Cables
Connection procedure of the AC power line:
Make sure the AC power supply complies with following specification: 100~240 VAC, 50/60
Hz.
Apply the power line provided with the monitor. Plug the power line to INPUT interface of the
monitor (Socket in Figure 1-7). Connect the other end of the power line to a grounded
3-phase power output.
NOTE:
Connect the power line to the jack special for hospital usage.
Connect to the ground line if necessary. Refer to Chapter Safety Guidance for details.
NOTE:
When the battery configuration is provided, after the device is transported or stored, the
battery must be charged. Powering on without connecting AC power supply may cause the
device out of work. Switch on AC power supply can charge the battery no matter if the
monitor is powered on.
2.4 Power on the Monitor
Power on, LOGO information will be displayed on the screen.
NOTE:
Check all the functions that may be used to monitor and make sure that the M8B is in good
status.
NOTE:
If chargeable battery is provided, charge it after using the device every time to ensure the
electric power is enough.
WARNING
If any sign of damage is detected, or the monitor displays some error messages, do not use
it on any patient. Contact biomedical engineer in the hospital or Customer Service Center
immediately.
NOTE:
The interval between twice presses of POWER should be more than 1 minute.
- 13 -
Page 25
M8B Multi-Parameter Patient Monitor User Manual
NOTE:
After continuous 360-hour runtime, please restart the M8B to ensure the monitors steady
performance and long lifespan.
2.5 Connect Patient Sensors
Connect all the necessary patient sensors between the monitor and the patient.
NOTE:
For information on correct connection, refer to related chapters.
2.6 Check the Recorder
If your M8B is equipped with a recorder, open the recorder door to check if paper is properly
installed in the slot. If no paper is present, refer to Chapter Recording for details.
- 14 -
Page 26
M8B Multi-Parameter Patient Monitor User Manual
3 System Menu
M8B Multi-parameter patient monitor features flexible configurations. You can configure various
aspects of the monitor, including the parameters to be monitored, sweeping speed of the waveforms,
audio signal volume, and output content. Press the MENU button on the front panel to call up SYSTEM MENU menu. You can perform
following operations in this menu.
Figure 3-1 System Memu
Review of trend graphs/tables and NIBP measurements will be described in Chapter Trend and
Event.
3.1 Patient Information Setup
NOTE:
To clear current patient data, refer to New Patient for details.
Pick "PATIENT SETUP" in SYSTEM MENU to call up the following menu.
- 15 -
Page 27
M8B Multi-Parameter Patient Monitor User Manual
Figure 3-2 Patient Setup
You can setup the following patient information:
DEPT. Department in which the patient receives treatment.
PAT NO Patient No.
BED NO Patient bed number (Range: 1~254)
DOCTOR Name of the doctor.
NAME Patient name (Valid characters: A~Z, 0~9; Max. length: 12 characters)
SEX Patient gender (Available options: "F" for Female, "M" for Male)
PAT TYPE Patient type (Available options: ADU, PED, and NEO)
ADMIT Hospitalization starting date (format: year\month\day)
BIRTH Patient date of birth (format: year\month\day)
HEIGHT(cm/inch) Patient height (Increase/decrease by 0.5 cm or 0.5 inch per switch)
WEIGHT(kg/lb) Patient weight (Increase/decrease by 0.5 kg or 0.5 lb per switch)
BLOOD Patient blood type (Pick A, B, O, AB, or N. "N" represents unknown blood
type)
NEW PATIENT Admission of new patient
Also in this menu, the user may select NEW PATIENT item to access CONFIRM TO UPDATE
- 16 -
Page 28
M8B Multi-Parameter Patient Monitor User Manual
PATIENT dialog box as shown below, in which the user decide whether to monitor a new patient.
Figure 3-3 Confirm to Update Patient
Pick YES to delete all information of the patient being currently monitored and exit the menu.
Pick NO to give up updating the patient and the system will keep the information of the current
patient and exit the menu.
NOTE:
Selecting YES will delete all information about the currently monitored patient.
3.2 Default Setup
NOTE:
Select any item in this sub-menu to cancel the current setup and use the selected default
setup.
Figure 3-4 Default Menu
In this sub-menu, you can select both the factory default and the user-defined default. Also in this
- 17 -
Page 29
M8B Multi-Parameter Patient Monitor User Manual
sub-menu, you can save the current configuration as the user-defined default configuration. At this
time, the system will automatically save all the setups in the parameter menu, ECG lead, gain and
filter way as the user-defined default configuration according to the patient type. And the dialog box
as shown below will pop up.
Figure 3-5 Confirm Save Default Config
Pick YES to save the current patient type config as user default configuration.
Pick NO to give up the operation.
3.3 Mark Event
There are four types of events that you can define. Select MARK EVENT item in SYSTEM SETUP to call up the following menu:
Figure 3-6 Mark Event
To mark the event: Use the rotary knob to select one from event A, B, C and D. There is a @
signal for the one selected. To cancel your selection, repress the knob at selected item. Press EXIT
to return to the previous menu.
The point of using event function:
To differentiate the patient events that have impact on parameter monitoring, such as dose taking,
- 18 -
Page 30
M8B Multi-Parameter Patient Monitor User Manual
injection, therapy status, etc.
Event will be displayed on the Trend Graph, Trend Table to assist to analyse patient parameter of
the time event happened.
3.4 Face Select
Select FACE SELECT item in SYSTEM SETUP menu to access FACE SELECT dialog box
as shown below, in which five selections are available: STANDARD SCREEN, TREND SCREEN,
oxyCRG SCREEN, VIEWBED SCREEN and LARGE FONT SCREEN. Only one selection can be
chosen for each time.
Figure 3-7 Face Select
Enter LARGE FONT SCREEN can select three modes, see as follows:
Figure 3-8 Large Font Face Select
3.5 Time Setup
Select TIME SETUP item in SYSTEM SETUP menu to access the sub-menu of TIME SETUP as shown below. System time is in format of year, month, day, hour, minute and second. Pick the item you wish to modify and turn the knob. Select EXIT item to return to the previous
- 19 -
Page 31
menu.
M8B Multi-Parameter Patient Monitor User Manual
Figure 3-9 Time Setup
3.6 Recorder Setup
Select "RECORD" in SYSTEM SETUP menu to call up the following menu:
Figure 3-10 Record
In the sub-menu, the user may select the REC WAVE1, REC WAVE2 or REC WAVE3 items,
maximum 3 waveforms can be printed out.
The output waveforms can be selected for the following items:
- 20 -
Page 32
M8B Multi-Parameter Patient Monitor User Manual
ECG1, ECG2
SpO2
RESP
IBP1, IBP2
ECG1 waveform and ECG2 waveform. (There will be 7 ECG
waveforms on the screen when Full-Lead display). If no ECG
waveform is currently displayed on the screen, this item cannot be
picked.
SpO2 Plethysmogram. (If no SpO2 waveform is currently displayed on
the screen, this item cannot be picked. But under ECG Full-Lead
display mode, this item can be picked, although no SpO2 waveform is
currently displayed on the screen. )
RESP waveform. (If no RESP waveform is currently displayed on the
screen, this item cannot be picked. But under ECG Full-Lead display
mode, this item can be picked, although no RESP waveform is
currently displayed on the screen.)
IBP1 waveform and IBP2 waveform. (If no IBP waveform is currently
displayed on the screen, this item cannot be picked. But under ECG
Full-Lead display mode, this item can be picked, although no IBP
waveform is currently displayed on the screen.)
OFF No display for this waveform.
RT REC TIME represents real-time recording time, for which two selections are available:
CONTINUAL and 8S. CONTINUAL means once pressing the Record button on the front panel, the recorder will continuously print out the waveform or parameter until the Record
button is pressed again.
TIMING REC TIME represents time interval between two times of timing recording. 10
selections are available: OFF, 10MIN, 20MIN, 30MIN, 40MIN, 50MIN, 1HOUR, 2HOURS, 3HOURS and 4HOURS. It means that the system will trigger the recording operation
according to the selected time interval. The recording time is fixed at 8 seconds.
NOTE:
REC TIME has the priority compared with TIMING REC TIME.
REC RATE: 25.0 mm/s or 50.0 mm/s..
CLEAR REC TASK can be used by the user to stop recorder from printing out too many tasks.
NOTE:
The recorder is an optional part.
NOTE:
If two same waveforms are selected, one of them is switched to a different waveform
- 21 -
Page 33
M8B Multi-Parameter Patient Monitor User Manual
automatically.
NOTE:
When ECG waveforms are selected for printing, with gain of ×1, ×0.5 or ×0.25, 3-channel waveform can be printed out; however, with gain of ×2, only 2-channel waveform can be printed out to avoid overlapping of waveforms, and the third waveform will be omitted. The 3-channel waveform can be printed only in real-time recording, while it is not available in other recording modes, such as alarm review recording, alarm triggered recording.
3.7 Module Setup
Select MODULE SETUP item in SYSTEM SETUP menu to call up the following menu:
Figure 3-11 Module Setup
You can choose the parameter you want to monitor from this menu, so that you can enhance the
display efficiency, and avoid interference from other messages.
3.8 Tracing Waveforms Selection
Select TRACE SETUP item in SYSTEM SETUP menu to call up the following menu:
- 22 -
Page 34
M8B Multi-Parameter Patient Monitor User Manual
Figure 3-12 Trace Setup
You can define the traces displayed on the screen in this menu. The waveforms available for selection are those whose modules have been selected in MODULE SETUP menu.
3.9 Monitor Version
Pick VERSION to show the software version information of this monitor.
3.10 Alarm Volume
The system provides five levels of alarm volume and an alarm silence function. The system will
give audio alarm prompt (including no alarm sound) based on the selection.
The user may select different level of volume as per clinical requirement. The method is listed
below: Press ALARM SETUP item in SYSTEM SETUP menu to call up ALARM SETUP
sub-menu as shown below, in which the user may set up the alarm volume and other alarm
information. Refer to Chapter Alarm for details.
- 23 -
Page 35
M8B Multi-Parameter Patient Monitor User Manual
Figure 3-13 Alarm Setup
Pick ALARM VOL to set the volume of alarm sounds by turning the knob. The selections are 1~5. You can also set alarm parameters in MAINTAIN USER MAINTAIN ALARM SETUP.
Refer to Chapter5 Alarm for details.
WARNING
If the alarm volume is off, when there is alarm, the M8B can not give audio alarm prompt, so
the user should use this function cautiously.
3.11 Key Volume
Select SELECTION item in SYSTEM SETUP menu to call up SELECTION sub-menu as shown below. Select KEY VOL item and set the volume. The selections are OFF, LOW, MED, HIGH.
Figure 3-14 Selection
- 24 -
Page 36
M8B Multi-Parameter Patient Monitor User Manual
3.12 Drug Calculation
The M8B provide drug calculation and titration table display functions for fifteen different drugs.
For details, please refer to the Chapter: Drug Calculation and Titration Table.
3.13 Waveform Demonstration
Select DEMO item in SYSTEM MENU to call up INPUT DEMO KEY. After entering the
password, the system enters the Demonstration Waveform status.
The purpose of waveform demonstration is only to demonstrate the machine performance, and for
training purpose. In clinical application, this function is not recommended because the DEMO will
mislead the hospital workers to treat the waveform and parameter as actual data of the patient,
which may result in delay of treatment or mistreatment.
3.14 Maintenance
Select MAINTAIN item in SYSTEM MENU to call up ENTER MAINTAIN PASSWORD
dialog box as shown below, in which you can enter password and then customize maintenance
settings. User can not execute factory maintenance function, which is only available for the service
engineers of EDAN or local representative authorized by EDAN.
Figure 3-15 Enter Maintain Password
User Maintain
Input the password into the ENTER MAINTAIN PASSWORD box and press CONFIRM, the USER MAINTAIN” menu will pop up, in which you can set up following items.
- 25 -
Page 37
M8B Multi-Parameter Patient Monitor User Manual
Figure 3-16 User Maintain
LANGUAGE: You can set one of the eight interface languages, which are “CHINESE”,
“ENGLISH”, “RUSSIAN”, “ITALIAN”, “SPANISH”, “FRENCH”, “POLISH” and “GERMAN”.
NOTE:
Please restart the monitor after changing the language.
LEAD NAMING: You can select AHA or EURO. To know the difference between these
two styles, refer to Chapter ECG/RESP Monitoring.
LOCAL NET NO: Physical No. of monitor.
ALM SETUP>>: You can set up parameters of alarm. For more details refer to Chapter Alarm:
- 26 -
Page 38
M8B Multi-Parameter Patient Monitor User Manual
Figure 3-17 Alarm Setup
OTHER SETUP>>: you can setup some parameters, such as module type. See as follows:
Figure 3-18 Other Setup
NURSE CALL: turn on or off nurse call. When a new alarm of physiological parameter
occurs, it gives 3-second NURSE CALL alarm; if the system alarm or the audio alarm is
turned off, the NURSE CALL is unavailable. It is connected to RJ45 socket, the same port
as connected to Ethernet. NURSE CALL occupies the 7th and 8th pins of RJ45. When alarm
occurs, the 7th and 8th pins are in short circuit, otherwise they are disconnected.
SERVER IP: Setup server IP.
SERVER PORT: Setup server port.
CO2 CHECK>>: For calibrating CO2. Unavailable for M8B.
GAS CALIBRATION>>: For calibrationing gas for AG monitoring. Unavailable for M8B.
- 27 -
Page 39
M8B Multi-Parameter Patient Monitor User Manual
Factory Maintain
Factory maintenance function is only available for the service engineers of EDAN or representative
authorized by EDAN.
1. Enter factory maintain through password.
Figure 3-19 Factory Maintain
VGA SIZE: set this item to 10.4 WIDTH_SCREEN.
MODULE SELECT:
- 28 -
Page 40
M8B Multi-Parameter Patient Monitor User Manual
Figure3-20 Select Module
NOTE:
For M8B, the AG, CO2 functions are unavailable.
2. Enter representative maintain through password.
Figure 3-21 Representative Maintain
- 29 -
Page 41
SELECT MODULE:
M8B Multi-Parameter Patient Monitor User Manual
Figure3-22 Select Module
For changing password:
Figure 3-23 Change Password Interface
Prompts for changing password:
- 30 -
Page 42
M8B Multi-Parameter Patient Monitor User Manual
Figure 3-24 Save Failure
Figure 3-25 Save Successfully
- 31 -
Page 43
M8B Multi-Parameter Patient Monitor User Manual
4 Face Select
The M8B has five different operating screens, which are “Standard Screen”, “Trend Screen”, “oxyCRG Screen”, “VIEWBED SCREEN”and “Large Font Screen”. When required, you can select
different operating screen for necessary information. Lets probe into these four operating screens
one by one.
4.1 Select Operating Screen
Select FACE SELECT item in SYSTEM SETUP menu to access FACE SELECT dialog box
as shown below, in which five selections are available: STANDARD SCREEN, TREND SCREEN,
oxyCRG SCREEN, VIEWBED SCREEN and LARGE FONT SCREEN. Only one selection can be
chosen for each time.
Figure 4-1 Face Select
4.2 Standard Screen
In the FACE SELECT menu, select the STANDARD SCREEN option to enter the Standard
Screen. The Standard Screen displays to us the parameters in the Parameter area and the waveforms
being monitored. This screen is the basic operating screen of the monitor.
- 32 -
Page 44
M8B Multi-Parameter Patient Monitor User Manual
Figure 4-2 Standard Screen
4.3 Trend Screen
Enter TREND SCREEN In the FACE SELECT menu, select theTREND SCREENoption to enter the Trend Screen.
Figure 4-3 Trend Screen
- 33 -
Page 45
M8B Multi-Parameter Patient Monitor User Manual
Position of trend graph
Trend graph is located to the right of the corresponding waveform in the Waveform area. Its color is
the same as that of the corresponding parameter.
Trend length
Dynamic trend length is 2 hours. On the trend graph, the scale of the right end of the X-axis is 0
hour while the left end is 2-hour.
Select trend parameter
If M8B are located at the same position on the trend graph, by selecting the corresponding hot key
of a parameter on the trend graph, you can have the trend graph of this parameter displayed on the
screen. For example, in ECG trend graph, you can select hot keys such as HR, ST or PVCs, then the
system will display their corresponding trend graphs respectively.
Close trend screen In the FACE SELECT menu, select options of other operating screens to close the Trend Screen.
4.4 oxyCRG Screen
Enter oxyCRG SCREEN In the FACE SELECT menu, select the“oxyCRG SCREEN”option to enter the oxyCRG Screen.
Trend graph of oxyCRG screen
Figure 4-4 oxyCRG Screen
- 34 -
Page 46
M8B Multi-Parameter Patient Monitor User Manual
Located at the lower part of the screen, oxyCRG screen consists of three trends: HR Trend, SpO2
Trend and RR Trend or Compressed Resp. Waveform.
Select oxyCRG trend length
There are two hot keys at the bottom part of the oxyCRG Screen, which are 4MIN/2MIN/1MIN,
RR/RESP WAVE.
By using hot keys for trend time, you may select to display trend graphs of three different lengths,
i.e., 1 min, 2 min and 4 min.
Select RR trend or Compressed Resp. Waveform
By using the hot keys for RR/RESP WAVE, you may select either RR trend graph or compressed Resp. Wave. They occupy the same position. Therefore, if you select RR, the position displays the dynamic trend of RR. If you select RESP WAVE, the position displays the compressed Resp.
Wave.
Close oxyCRG
In the FACE SELECT menu, select other options of operating screen to close the oxyCRG Screen.
4.5 Viewbed Screen
If another monitor is connected on the same LAN of this monitor, you can use this monitor to view any measured waveform and information about all measured parameters from another monitor.
Enter Viewbed Screen
Select the VIEWBED SCREEN option in the FACE SELECT menu. Viewbed Screen window occupies the space of the bottom four waveforms.
Figure 4-5 Viewbed Screen
Hot key of Viewbed
There are two hot keys in the Viewbed Screen: Select Bed Number and Select Waveform.
- 35 -
Page 47
M8B Multi-Parameter Patient Monitor User Manual
The hot key of Select Bed Number displays the bed numbers and patient names of other monitors currently connected on the LAN. You can select a monitor to be monitored according to the patient name and bed number. If at this time no other monitors are connected on the same LAN of this monitor, the hot key of Bed Number will therefore display N/A. After you use this hot key to select a monitor to be viewed, the system will toggle to the display of the selected monitor for your view. The selected waveform is one of those listed in the hot key of Select Waveform. The hot key of Select Waveform is used to select a waveform generated by the monitor being viewed. If the hot key of Select Waveform displays N/A, it indicates that the bedside monitor being viewed has no waveforms. You can use this hot key to select and therefore view different waveforms of the monitor being viewed.
Alarm indicator of Viewbed
On the upper right side of the Viewbed Screen, there is an Alarm Indicator used to tell the alarm status of the monitor being viewed. The activity of this alarm indicator is identical with that of the alarm lamp on the panel of the monitor being viewed. That is to say, if the monitor being viewed occurs medium/low level alarm, this alarm indicator illuminates yellow; if it occurs high level alarm, this alarm indicator illuminates red. If the monitor being viewed has no alarm or the alarm is screened, the icon for this alarm indicator will not be displayed.
Parameter area of Viewbed Screen
Under the hot key of Select Bed Number is the Parameter area, in which parameters of all monitors being viewed are displayed.
Waveform area of Viewbed Screen
Under the hot key of Select Waveform is the Waveform area. The Sweep manner (refreshing or scrolling) of the waveform is identical with that of this monitor. The feature description of the displayed waveform is given above the waveform. Sweep speed is also identical with that set up for the same waveform on this monitor.
Close Viewbed Screen
In the FACE SELECT menu, select options of other operating screens to close the Viewbed Screen.
Rules for automatically selecting monitor to be viewed and waveform
When you turn on the monitor or enter Viewbed Screen, the system will automatically select a networked bedside monitor and a waveform of this monitor for you to view. If the monitor being currently viewed is disconnected, the viewed monitor will automatically close, clear displays of all alarms, parameters and waveforms. However in this situation, the Viewbed Screen still displays. If you want to view another monitor, you must select again through using hot keys. If a measure module of the viewed monitor is plugged out or closed, its corresponding waveform will disappear and the waveform in the Waveform area will not be refreshed. Instead this Waveform area will display empty. At this time, if you want to view other waveforms of this monitor, you need to select again.
4.6 Large Font Screen
- 36 -
Page 48
M8B Multi-Parameter Patient Monitor User Manual
Large Font Screen is a kind of operating screen, just like Standard Screen, Trend Screen and other
operating screens. It is used by customer to meet different display requirements in monitoring.
Enter Large Font Screen
Choose LARGE FONT SCREEN in FACE SELECT menu to enter LARGE FONT FACE SELECT. There are three modes, see as follows:
Figure 4-6 Large Font Face Select
Three display modes
1. ECG+SpO2+NIBP display mode:
Figure 4-7 ECG+SpO2+NIBP display mode
- 37 -
Page 49
M8B Multi-Parameter Patient Monitor User Manual
2. SpO2+NIBP display mode:
3. SpO2 display mode:
Figure 4-8 SpO2+NIBP display mode
Figure 4-9 SpO2 display mode
Exit Large Font Screen
In the LARGE FONT FACE SELECT menu, choose EXIT to return to FACE SELECT screen.
- 38 -
Page 50
M8B Multi-Parameter Patient Monitor User Manual
5 Alarm
This chapter gives general information about the alarm and measures to be taken accordingly.
Alarm setup and prompt messages are provided in respective parameter setup sections.
5.1 Alarm Modes
5.1.1 Alarm Level
Each alarm, either technical or physiological, has its own level. For alarm of higher level, when it occurs, the system will give prompt in various ways. Some alarms level can be set by the user via
software. Others can not be changed once defined by the system. Alarms in M8B are divided into
three levels which are high, medium and low.
High-level alarm indicates the patient’s life is in danger or the monitor under using has serious
problem in technical respect. It is the most serious alarm.
Medium-level alarm means serious warning.
Low-level alarm is a general warning.
Alarms are classified into three categories, which are physiological alarm, technical alarm and general alarm. Physiological alarm refer to those alarms triggered by patient’s physiological
situation which could be considered dangerous to his or her life, such as heart rate (HR) exceeding
alarm limit (parameter alarms). Technical alarm refer to system failure which can make certain
monitoring process technically impossible or make monitoring result unbelievable. Technical alarm
is also called System Error Message. General alarm belongs to those situations that can not be
categorized into these two cases but still need to pay some attention.
M8B Multi-parameter patient monitor has preset the alarm level for the parameters. You can also
modify the alarm level using the method described in this chapter.
Alarm level of the System Error Message (technical alarm) is preset in the system.
All technical alarm level and general alarm level, some of the physiological alarm levels are preset
in the system and can not be changed by user.
5.1.2 Alarm Modes
When alarm occurs, M8B Multi-parameter patient monitor may raise the users attention in at least
three ways, which are audio prompt, visual prompt and description. Audio and visual prompt is
given by TFT display device, the speaker on the display device and the alarm indicator. Description
is displayed on the screen. Physiological alarm is displayed in the Physiological Alarm area. Most
- 39 -
Page 51
M8B Multi-Parameter Patient Monitor User Manual
of technical alarms are displayed in the Technical Alarm area. Technical alarms related to NIBP
measurement are displayed in the NIBP Technical Alarm area at the bottom of NIBP parameter area.
NOTE:
The Physiological Alarm area is on the upper right part of the screen. The Technical Alarm
area is to the left side of the Physiological Alarm area.
NOTE:
If M8B is connected to the external alarm prompt system (e.g. the alarm speaker and
indicator connected onto the rear panel of the monitor), when alarm occurs, the external
alarm prompt system responds in the same way as the the monitor.
NOTE:
The concrete presentation of each alarm prompt is related to the alarm level.
How to indicate that the measured parameter has exceeded its alarm limits:
When physiological alarm of the monitored parameter exceeds the alarm limit, besides using the
above-mentioned three ways to give the alarm prompt, the M8B gives alarm by making the
monitored parameter flash in the frequency of 1Hz.
Screen Display
When the measured parameter exceeds its alarm limits and triggers a physiological alarm, the corresponding parameter value will flash. * signal appears on the screen indicating the occurrence of alarm. Red *** indicates high-level alarm, yellow ** indicates medium-level alarm, and yellow * indicates low-level alarm. Technical alarm will not prompts * signal.
Lamp light
The high/medium/low-level alarms are indicated by the system in following different visual ways:
Alarm Level Visual Prompt
High Alarm indicator flashes in red with high frequency.
Medium Alarm indicator flashes in orange with low frequency.
Low Alarm indicator lights on in orange.
Alarm Sound
The high/medium/low-level alarms are indicated by the system in following different audio ways:
- 40 -
Page 52
M8B Multi-Parameter Patient Monitor User Manual
Alarm level Audio prompt
High
Mode is DO-DO-DO------DO-DO, DO-DO-DO------DO-DO, which
is triggered once every 10 seconds.
Medium Mode is DO-DO-DO, which is triggered once every 25 seconds.
Low Mode is DO-, which is triggered once every 30 seconds.
NOTE:
When alarms of different levels occur at the same time, the M8B prompts the one of the
highest level.
5.1.3 Alarm Setup
Setup alarm in the ALARM SETUP menu
Press the ALARM SETUP button on the SYSTEM SETUP menu to call up ALARM SETUP menu (default menu) as shown below. In the ALM SEL item, the user may set up the information about common alarm setup and the alarm setup of each parameter.
Figure 5-1 Alarm Setup
COMMON ALM SETUP
Select COMMON ALM SETUP selection in ALM SEL item. This operation may call up the dialog box as the default one.
ALARM VOL: have five selections: 1, 2, 3, 4 and 5. ALM REC TIME: which has three selections: 8S, 16S, 32S.
Alarm setup of each parameter
- 41 -
Page 53
M8B Multi-Parameter Patient Monitor User Manual
In the ALARM SETUP menu, select the ALM SEL item to set up the alarm information for following parameters. They are HR, ST, PVCs, SpO2, NIBP, IBP (1, 2), RESP, TEMP, CO2, CO and AG. For example: Method to set up HR alarm information: Step 1: Select the HR ALM SETUP option in the ALM SEL item. Then the menu only displays HR setup items. Step 2: You can set up five items in this menu, which are HR ALM (on/off of the alarm switch), ALM LEV (alarm level), ALM REC (alarm recording switch), ALM HI (higher limit of HR alarm), ALM LO (lower limit of HR alarm). You can move the cursor onto the item to be setup by using the knob and press the knob to make the setup. The method for setting the alarm information of other parameters is the same as HR.
Setup alarm in the User Maintain menu
You can also setup alarm parameters in SYSTEM MENU MAINTAIN USER MAINTAIN ALARM SETUP. See as follows:
Figure 5-2 Alarm Setup in User Maintain
ALM PAUSE TIME: set up the duration of Alarm Pause status. Three options are available, 1
minute, 2 minutes and 3 minutes.
ALARM SILENCE: when it set to ON, hold the Silence button on front panel for 3 seconds,
the alarm will be silenced. During the silence, the monitor gives Low alarm for alarm silenced
state per 3 minutes. Press this button again to turn on the alarm system.
SENSOR OFF ALARM: turn on or off the sensor off alarm. When this item is set ON, press
Silence button on front panel can pauze the audio alarm. And press again to resume the audio
alarm; at the alarm paused state, it will give alarm if sensor off alarm occurs.
5.2 Alarm Cause
- 42 -
Page 54
M8B Multi-Parameter Patient Monitor User Manual
Alarm occurs when:
1. Physiological alarm is evoked;
2. Alarm for error of the system (technical alarm) is evoked;
3. General alert occurs.
A. Conditions that activate the parameter alarms:
When the measurement value exceeds the alarm limit and the alarm is set ON. Alarm will not activate if the alarm is set OFF.
B. Conditions that activate the system alarms (technical alarm):
Upon the system error, the monitor prompts alarm immediately.
C. General alert
In some circumstances, alerts will behave as physiological alarm but in normal sense, we do not
regard them as real patient health related items.
5.3 SILENCE
SILENCE
Enter SYSTEM MENU MAINTAIN USER MAINTAIN ALARM SETUP, if the ALARM
SILENCE is set to ON, press Silence button to turn off the audio alarm or paused it.
1. Audio alarm paused icon
When the ALARM SILENCE is ON, press Silence button on front panel, then the audio alarm is
paused. And the paused time can be set in ALARM SETUP menu, see figure 5-2. The audio
alarm paused icon displays in Information area. Press Silence button again can resume the audio
alarm.
2. Audio alarm off icon
Press the Silence button for more than 3 seconds, the audio alarm is turned off. Then press
Silence button again or hold it for few seconds can turn on the audio alarm. At the audio alarm
off state, the monitor gives a Low alarm beep per 3 minutes to prompt that the alarm is turned
off.
NOTE:
Whether an alarm will be reset depends on the status of the alarm cause.
5.4 Parameter Alarm
The setup for parameter alarms is in their menus. In the menu for a specific parameter, you can
- 43 -
Page 55
M8B Multi-Parameter Patient Monitor User Manual
check and set the alarm limit, alarm status. The setup is isolated from each other.
When a parameter alarm is off, a symbol displays near the parameter. If the alarms are
turned off individually, they must be turned on individually.
For the parameters whose alarm is set to ON, the alarm will be triggered when at least one of them
exceeds alarm limit. The following actions take place:
1. Alarm message displays on the screen as described in alarm mode;
2. The monitor beeps in its corresponding alarm class and volume;
3. Alarm lamp flashes.
5.5 When an Alarm Occurs
NOTE:
When an alarm occurs, you should always check the patient's condition first.
The alarm message appears at the top of the screen on the right side. It is needed to identify the
alarm and act appropriately, according to the cause of the alarm.
1. Check the patient's condition.
2. Identify the cause of the alarm.
3. Identify which parameter is alarming or which alarm is happening.
4. When cause of alarm has been over, check that the alarm is working properly.
You will find the alarm messages for the individual parameter in their appropriate parameter
chapters of this manual.
- 44 -
Page 56
M8B Multi-Parameter Patient Monitor User Manual
6 Freeze
General
Freeze & Unfreeze
Review Frozen Waveforms
6.1 General
When monitoring a patient, you may freeze the waveforms of interest so as to view them carefully.
Generally you can review maximally 120 seconds of a frozen waveform. The Freeze function of
this monitor has following features:
Freeze status can be activated on any operating screen;
At the same time of entering the Freeze status, the system exits all other operating menus.
Besides, the system freezes all waveforms in the Waveform area of the Basic Screen, and also
freezes Full Lead ECG waveforms and extra waveforms in the Full Lead ECG interface(if
have). Nevertheless the Parameter area refreshes normally.
The frozen waveforms can be reviewed.
6.2 Enter/Exit Freeze Status
Enter Freeze Status
In the Non-Freeze status, press the Freeze button on the control panel of the monitor to let the
system exit the Menu being currently displayed (if available), then enter the Freeze status and display the popup FROZEN menu. In the Freeze status, all other waveforms are frozen. In other
words, the system will no longer refresh all other waveforms.
Exit Freeze Status
In the Freeze status, executing any of the following operations will command the system to exit the
Freeze status:
Select the EXIT option on the FROZEN menu; Press the Freeze button on the control panel again;
Press the non-immediate-to-execute button (such as a button once pressed, a menu will pop up
for you to further select an option) on the front panel and system buttons of Menu and Main;
Execute any operation that may trigger the adjustment of the screen or display of a new menu.
After exiting the Freeze status, the system will discharge the Freeze status, clear screen waveforms
and resume display real-time waveforms. In the Screen Refresh mode, the system will sweep the
waveforms from the left to right in the Waveform Area.
- 45 -
Page 57
M8B Multi-Parameter Patient Monitor User Manual
6.3 FROZEN Menu
Press the FREEZE button on the control panel, the FROZEN menu will appear on the bottom part
of the screen. At the same time, the system enters the Freeze status.
Figure 6-1 Frozen
RECALL: used to review frozen waveforms.
EXIT: after pressed, the system closes the FROZEN menu and exits the Freeze status.
NOTE:
Pressing the Freeze button repeatedly in short time may result in discontinuous
waveforms on the screen.
6.4 Reviewing Frozen Waveform
By moving the waveform, you may review a waveform of 120 seconds before the moment when it
is frozen. For a waveform less than 120 seconds, the remaining part is displayed as a straight line. Use the rotary snob on the control panel to move the cursor to the RECALL option on the
FROZEN menu. Press the knob. By turning the knob left or right, frozen waveforms on the screen
will move left or right correspondingly. There is an arrow indicating upward under the right side of the last waveform. There is also a time scale beside the arrow. -0S is used to mark the moment
when waveforms are frozen. With waveforms moving right, this time mark will in turn change into
-1S, -2S, -3SThese time marks are applied to all waveforms on the screen.
- 46 -
Page 58
M8B Multi-Parameter Patient Monitor User Manual
7 Recording (Optional)
General information on recording
Instructions for configuring and recording
Recording messages
7.1 General Information on Recording
A thermal dot matrices recorder with 48mm wide printout paper is used for M8B Multi-parameter
Patient Monitor.
Performance of the Recorder
Waveform record is printed out at the rate of 25 mm/s or 50 mm/s.
It can record up to 3 waveforms.
English printout.
User-selectable real-time recording time and waveform.
Auto recording interval is set by the user, the waveform is in accordance with the real time
recording.
7.2 Recording Type
M8B Multi-parameter patient monitor provides several stripe recording types:
Continuous real-time recording
8 second real-time recording
Auto 8 second recording
Alarm recording
Frozen waveform recording
Trend graph, trend table recording
Arrhythmia review recording
Drug calculation titration recording
NIBP review recording
Alarm review recording
oxyCRG recording
NOTE:
When ECG waveforms are selected for printing, with gain of ×1, ×0.5 or ×0.25, 3-channel waveform can be printed out; however, with gain of ×2, only 2-channel waveform can be printed out to avoid overlapping of waveforms, and the third waveform will be omitted.
The 3-channel waveform can be printed only in real-time recording, while it is not available
- 47 -
Page 59
M8B Multi-Parameter Patient Monitor User Manual
in other recording modes, such as alarm review recording, alarm triggered recording.
Real-time Recording
Real-time recording starts as you press the Record button on the recorder.
The waveforms for continuous real-time recording and continuous 8 second recording are
automatically set by the monitor (usually the first three waveforms displayed on the screen). You
can also configure it through the menu. Refer to related section for details.
In RECORD menu, the user can choose three waveforms to be printed out. The User can setup one
or two waveforms off. Thus, the real time record will print out one or two waveforms. If three
waveforms are off, the real time record will print out measure parameters only.
NOTE:
When the system is executing a recording task, the system can start executing next alarm
recording task only when the current one is finished.
Auto Recording
The monitor starts the recorder for 8 seconds according to interval time set in the TIMING REC TIME of the RECORD menu. Refer to Chapter Recorder Setup for details.
Alarm Recording
Parameter Alarm
The monitor records waveforms 4, 8, or 16 seconds prior to and after the alarm (totally 8, 16 or 32
seconds) (which can be selected in System Menu). All parameter values during the alarm will also
be recorded.
When parameter alarm occurs, two recorded waveforms can be printed out.
In order to avoid repeated printout of alarm waveforms: If more than two parameter alarms are switched on and triggered simultaneously, the recorder
will print out those of the highest level. If of the same alarm level, the latest alarm will be
printed out.
If an alarm occurs during the alarm of another parameter, it will be printed out after the current
recording is finished.
If many alarms occur at the same time, some of waveforms will be stored for printout in turn.
ST Segment Alarm
The M8B records 2-channel ECG waveforms 4, 8, or 16 seconds prior to and after the alarm (totally
8, 16 or 32 seconds) (which can be selected in the ECG SETUP menu). All parameter values during
the alarm will also be recorded.
- 48 -
Page 60
M8B Multi-Parameter Patient Monitor User Manual
Arrhythgmia Alarm
The M8B records 2-channel ECG waveforms 4, 8, or 16 seconds prior to and after the alarm (totally
8, 16 or 32 seconds). All measurement results during the alarm will also be recorded.
Titration Table
The M8B can print out the message in the current TITRATION window.
Notes on Recording
Recording types:
Real time Report
Periodic Report
Para Alarm Report
Titration Table
Patient bed number, name, sex, height, weight, date of birth, admission date
Parameter name and value
Recording time
Waveform name
Waveform scale (for ECG waveform)
ECG lead, scale, filter mode, (if having ECG waveforms, it will be printed out within the first
second or when changing the lead, gain and filter mode during real-time recording.)
IBP scale (the first second of IBP waveform)
Date and time.
7.3 Recording Startup
You can start the recording in the following ways:
Continuous real-time
recording
8 second real-time
Press the Record button to start/stop the recording.
Press the Record button to start recording. It will automatically
recording
stop in 8 seconds.
Auto recording Record the three waveforms selected in RECORD menu
according to the setup time interval in RECORD menu. It will
automatically stop in 8 seconds.
Alarm recording When alarm recording is set ON, it automatically starts when
alarm occurs.
- 49 -
Page 61
M8B Multi-Parameter Patient Monitor User Manual
Trend graph recording Access the TREND GRAPH menu, then press the Record
button to start recording.
Trend table recording Access the TREND TABLE menu, then press the Record
button to start recording.
Arrhythmia review
recording
Enter the ECG SETUP menu via hot key, select ARR ANALYSISARR RECALL, then press the Record button
to start recording.
Alarm review recording Access the ALARM RECALL menu, then press the Record
button to start recording.
NIBP review recording Access the NIBP RECALL menu, then press the Record
button to start recording.
Titration table recording Access the DRUG CALC menu from the SYSTEM MENU
menu. Pick the TITRATION button in the menu to access the TITRATION window. Pick the REC button to print out the
titration currently displayed in the window.
NOTE:
You can press the Record button on the control panel to stop the current recording
process.
Access the RECORD menu from the SYSTEM SETUP menu. Then pick the CLEAR REC TASK button to stop all recording tasks.
7.4 Recorder Operations and Status Messages
Record Paper Requirement
Only standard thermosensitive record paper can be used, otherwise the recorder may not function,
the recording quality may be poor, and the thermosensitive printhead may be damaged.
Proper Operation
When the recorder is working, the record paper goes out steadily. Do not pull the paper outward
with force, otherwise the recorder may be damaged.
Do not operate the recorder without record paper.
Paper Out
When "RECORDER OUT OF PAPER" alarm is displayed, the recorder can not start. Please insert
- 50 -
Page 62
M8B Multi-Parameter Patient Monitor User Manual
record paper properly.
Inserting Paper
Place fingers under the two sides flange of the recorder casing, pull outwards directly to release
the casing.
Insert a new roll of paper into the paper cassette, printing side facing upwards.
Ensure proper position and tidy margin.
Pull about 2 cm of the paper out, and close the recorder casing.
NOTE:
Be careful when inserting paper. Avoid damaging the thermo-sensitive print head. Unless
when inserting paper or shooting troubles, do not leave the recorder catch open.
Removing Paper Jam
When the recorder functions or sounds improperly, you should open the recorder casing to check if
there is a paper jam. Removing the paper jam in the following way:
Cut the record paper from the feeding edge.
Open the recorder casing.
Re-insert the paper.
- 51 -
Page 63
M8B Multi-Parameter Patient Monitor User Manual
8 Trend and Event
M8B Multi-parameter patient monitor provides 96-hour trend data of all parameters, storage of 500
NIBP measurement results and 60 alarm events. This chapter gives detailed instruction for review
of all data.
8.1 Trend Graph
■ The latest 1-hour trend is displayed every 1 or 5 seconds;
■ The latest 96-hour trend is displayed every 1, 5 or 10 minutes;
Pick "TREND GRAPH" in the SYSTEM MENU to call up the following menu:
Figure 8-1 Trend Graph Menu
In the trend graph, the Y-axis stands for the measured value and X-axis time. " " is the cursor of
the trend graph, the parameter value of the position pointed by the cursor is displayed below the
trend graph and the corresponding time is displayed above the trend graph. Other trends except
NIBP trend are displayed as continuous curves. In NIBP trend graph, "▼" indicates systolic value,
"▲" indicates diastolic value, and "*" indicates mean value.
To select trend graph of a specific parameter
Pick PARA SELECT item and select a requested parameter name by turning the knob.
- 52 -
Page 64
M8B Multi-Parameter Patient Monitor User Manual
To select 1-hour or 96- hour trend graph
Pick RESOLUTION item, choose 1 or 5 sec for 1-hour trend graph and 1, 5 or 10 min for 96-hour
trend graph.
To view other trend curves
When " " appears on the right part of the screen, pick "L-RIGHT", turn the knob clockwise to
view later trend curves. When " " appears on the left part of the screen, pick the same item, turn
the knob counterclockwise to view earlier trend curve.
To change the display scale
Pick the ZOOM button to adjust the Y-axis scale and thus change the trend curve in proportion.
The value beyond maximum value will be represented by the maximum value.
To obtain trend data of a specific time
The time to which the cursor points will change as the knob is turned. Parameter at this time is
displayed below the X-axis. When " " appears on the right part of the screen, the trend graph
pages down for later trend curve as the cursor moves here. When " " appears on the left part of
the screen, the trend graph pages up for earlier trend curve as the cursor moves here.
Mark Event
If an event is marked A, B, C, or D, then the corresponding event type will display on the axis time
of the trend graphsuch as , , or .
Operation Example
To view the NIBP trend graph of the last 1 hour:
Pick the Menu button on the lower right of the screen.
Pick TREND GRAPH item in the SYSTEM MENU.
Select parameter: pick the PARA item and turn the knob until NIBP appears.
Select 1S or 5S in the RESOLUTION item.
Pick the L-Right button and turn the knob to view changes of the trend graph time and trend
curve.
Stop at requested trend time section for careful review. Pick the ZOOM button to adjust the
display scale if necessary.
For measurement result of a specific time, pick CURSOR to move the cursor to the point,
corresponding time and value will display on above and below respectively.
Pick EXIT to return to trend graph display.
8.2 Trend Table
The latest 96-trend table data can be displayed at every 1, 5, 10, 30, or 60 minutes.
- 53 -
Page 65
M8B Multi-Parameter Patient Monitor User Manual
Pick TREND TABLE in the SYSTEM MENU to call up the following menu:
Figure 8-2 Trend Table
Time in response to each group of trend data is displayed at the leftmost list with date in bracket.
Marked event corresponds to marking time. Trend data of each parameter is divided into 8 groups.
HR, PVC
ST1, ST2
RR
T1, T2, TD
SpO2, PR
NIBP NS/NM/ND
IBP1, IBP2
To choose trend table of different resolution
Pick the RESOLUTION item and turn the knob to change its content so as to change the time
interval of trend data.
To view other trend data
When " " appears on the upper part of the screen, pick UP-DOWN button and turn the knob
counterclockwise to view later trend data. When " " appears on the lower part of the screen, pick
the same item and turn the knob clockwise to view earlier trend data.
- 54 -
Page 66
M8B Multi-Parameter Patient Monitor User Manual
To obtain trend data of different parameter
Pick L-RIGHT to select one from the 8 groups of parameters. A " " by the rightmost item
indicates following page available. A " " by the leftmost item indicates previous page available.
Mark Event
If an event is marked A, B, C, or D, the corresponding event type will display on the axis time of
the trend table.
Operation Example
To view a NIBP trend table:
Pick the Menu button on the lower right of the screen to access SYSTEM MENU.
Pick TREND TABLE.
Pick L-RIGHT and switch to NIBP by turning the knob.
Pick RESOLUTION to select requested time interval.
Pick UP-DOWN and turn the knob to view NIBP trend data of different time.
Pick EXIT to return to SYSTEM MENU.
8.3 NIBP Recall
M8B Multi-parameter patient monitor can review the latest 500 NIBP measurement data.
Pick NIBP RECALL in the SYSTEM MENU to invoke the result and time of the latest 15
measurements, as shown in the figure below.
Figure 8-3 NIBP Recall
- 55 -
Page 67
M8B Multi-Parameter Patient Monitor User Manual
Data is listed chronologically from the latest to the earliest. 15 measurements can be displayed in
one screen. Pick UP-DOWN to view other trend curve up to 500 results. When you press the RECORDbutton, the recorder will pint out the metrical data of current window.
8.4 Alarm Event Recall
M8B Multi-parameter patient monitor can display the latest 60 alarm events. Select ALARM RECALL in the SYSTEM MENU to access ALARM RECALL
CONDITION menu as shown below.
Figure 8-4 ALARM RECALL CONDITION
In this menu, the user may select the conditions for alarm review, including:
1. Start and End time of review:
The user may select the start time of review in the item of START.
Then the user may select the end time of review. Two selections are available: current time and
the user-defined time.
For user-defined end time, the user can use the knob to select.
2. ALARM RECALL EVENT
In the pull-down list of ALARM RECALL EVENT, the user can select the parameter whose
alarm events he wants to review. The selections include ALL (alarm events of all parameters),
ECG, REST, SpO2, NIBP, IBP, TEMP, HR_H>180 (the value of HR is higher than the upper
alarm limit), HR_L<60 (the value of HR is below the lower alarm limit), SpO2<90%, IBP_H>200mmHg, IBP_L<40mmHg, RR_H>40, RR_L<10, TEMP_H>40ºC, TEMP_L<34ºC. After setting up all the review conditions, press the ALARM RECALL button to access ALARM RECALL window.
- 56 -
Page 68
M8B Multi-Parameter Patient Monitor User Manual
ALARM RECALL
The ALARM RECALL window is as shown below, in which following data are displayed:
Time span (Format: month-day-year hour: minute-month-day-year hour: minute). Event type. Serial number (Format: NO. ×× of ××). The value at the time of alarm. NIBP result is with time. Two 8/16/32-second waveforms.
Figure 8-5 ALARM RECALL Menu
To view all waveforms during the alarming process
Pick L-RIGHT and turn the knob to view all 8/16/32-second waveforms stored.
To view other alarm events
Events of up to 60 are listed chronologically from the latese to the earliest. Pick UP-DOWN button
and turn the knob to view later or earlier events.
Recording
Pick REC to print our all data and waveform this event.
- 57 -
Page 69
M8B Multi-Parameter Patient Monitor User Manual
9 Drug Calculation and Titration Table (Optional)
M8B Multi-parameter patient monitor provides Drug calculation and titration table display
functions for fifteen drugs and outputs the content of titration table on the recorder.
9.1 Drug Calculation
The drug calculations that can be performed by the system are AMINOPHYLLINE,
DOBUTAMINE, DOPAMINE, EPINEPHRINE, HEPARIN, ISUPREL, LIDOCAINE, NIPRIDE,
NITROGLYCERIN and PITOCIN. Besides DRUG A, DRUG B, DRUG C, DRUG D and DRUG E
are also provided to flexibly replace any of the drugs. Select DRUG CALC in SYSTEM MENU, the following DRUG CALC display appears:
Figure 9-1 DRUG CALC
The following formulas are applied for dose calculation:
Concentrat = Amount / Volume
INF Rate = DOSE / Concentrat
Duration = Amount / Dose Dose = Rate × Concentrat DRIP Rate = INF Rate / 60 × DROP Size
Operating Method:
In the Drug Calculation window, the operator should first select the name of the drug to be
calculated, and then confirm the patient weight. Afterwards, the operator should also enter other
known values.
- 58 -
Page 70
M8B Multi-Parameter Patient Monitor User Manual
Turn the knob to select the value of the item to be calculated. Turn the knob to change the value.
When it is the required value, press the knob to view the calculation result. Each item has its calculation range. If the result exceeds the range, it displays ---.--.
NOTE:
For the drug calculation, the prerequisite is that the operator must enter the patient weight
and drug name at first. The system first gives a group of random initial values, which can
not be used by the operator as the calculation reference. Instead, he should enter a new
group of values at the doctors instruction.
NOTE:
Each drug has its fixed unit or unit series. Operator must select the proper unit at the
doctors instruction. If the result exceeds the system-defined range, it will display ---.
NOTE:
After entering a value, a conspicuous prompt will appear in the menu warning the operator
to confirm the correctness of the entered value. The correct value is the guarantee for the
reliability and safety of the calculated results.
NOTE:
For each entered value, the system will always give a dialog box asking for the user s
confirmation. You must be careful when answering each box. The calculated result is
reliable only after the entered value is confirmed to be correct.
Select the Drug Name:
Turn the knob to pick the DRUG NAME item in DRUG CALC menu. The user may select the drug
name in the pull-down list, including AMINOPHYLLINE, DOBUTAMINE, DOPAMINE,
EPINEPHRINE, HEPARIN, SUPREL, LIDOCAINE, NIPRIDE, NITROGLYCERIN, PITOCIN,
Drug A, Drug B, Drug C, Drug D and Drug E. Calculation for only one type can be generated each
time.
NOTE:
A, B, C, D and E are only code for drugs instead of their real names. The units for these five
drugs are fixed. The operator may select the appropriate units according to the convention
of using these drugs. The rules for expressing the units are:
“mg series units are fixedly used for drug A, B and C: g, mg, mcg.
- 59 -
Page 71
M8B Multi-Parameter Patient Monitor User Manual
unit series units are fixedly used for drug D: unit, k unit, m unit.
mEq is fixedly used for drug E.
Patient Weight:
After accessing the DRUG CALC window, the operator should enter the patient weight into the first
or the second item. The entered weight will be used as the independent data only for the calculation
of drug concentration.
NOTE:
This drug calculation function acts only as a calculator. That means the patients weight in
Drug Calculation menu and Patient Information menu are independent on each other.
When the Weight in Drug Calculation is changed, the Weight in Patient Information will
not change. In this way, we can say, the Drug Calculation menu is independent on other
menus in the system. Any change of it will not affect other information about the patient
being currently monitored.
9.2 Titration Table
Access Titration Table:
Select TITRATION item in DRUG CALC menu to enter titration table display.
Titration table display for drug is as following:
Figure 9-2 TITRATION
- 60 -
Page 72
M8B Multi-Parameter Patient Monitor User Manual
Method to operate the titration table:
1. In the TITRATION table, turn the knob to pick BASIC item. Press and turn the knob to select
either INF RATE or DOSE or DRIP RATE.
2. Then turn the knob to pick STEP item. Press and turn the knob to select step. 1 ~ 10 are
available for selection with the increment being 1.
3. Turn the knob to pick DOSE TYPE item. Press and turn the knob to select the unit in the
pull-down list.
4. Use UP-DOWN item in the table to view the data in previous or following pages.
5. Turn the knob to pick REC item. After pressing the knob, the recorder prints out the data
displayed in tne current titration table.
6. Turn the knob to pick EXIT to return to DRUG CALC menu.
Total amount, dose, volume, INF rate, drip rate and patient weight and drug name are displayed on
the top of the titration table. Meaning of each English identifier is:
AMOUNT: drug amount
VOLUME: liquid volume
DOSE/min: drug dose
INF RATE: flow rate
DRIP RATE: drop rate
WEIGHT: patient weight
- 61 -
Page 73
M8B Multi-Parameter Patient Monitor User Manual
10 Safety Guidance
The M8B Multi-parameter Patient Monitor is designed to comply with the International National
Safety requirements for medical electrical equipment IEC60601-1, EN60601-2-27 and
EN60601-2-30. This device has floating inputs and is protected against the effects of defibrillation
and electrosurgery. If the correct electrodes are used and applied in accordance with the
manufacturer instructions (see Chapter ECG/RESP Monitoring), the system can restore screen
display within 10 seconds after defibrillation.
This symbol indicates that the instrument is IEC 60601-1 Type CF equipment. The unit displaying
this symbol contains an F-Type isolated (floating) patient applied part providing a high degree of
protection against shock, and is suitable for use during defibrillation.
Environment
Follow the instructions below to ensure a completely safe electrical installation. The environment
where the M8B Multi-parameter patient monitor will be used should be reasonably free from
vibration, dust, corrosive or explosive gases, extremes of temperature, humidity, and so on. For a
cabinet mounted installation, allow sufficient room at the front for operation and sufficient room at
the rear for servicing with the cabinet access door open.
M8B Multi-parameter patient monitor operates within specifications at ambient temperatures between 5 and 40. Ambient temperatures that exceed these limits could affect the accuracy of
the instrument and cause damage to the modules and circuits. Allow at least 2 inches (5cms)
clearance around the instrument for proper air circulation.
Power Source Requirements
Refer to APPENDIX .
Grounding the M8B
To protect the patient and hospital personnel, the cabinet of the M8B Multi-parameter patient
monitor must be grounded. Accordingly, the M8B is equipped with a detachable 3-wire cable which
grounds the instrument to the power line ground (protective earth) when plugged into an appropriate
3-wire receptacle. If a 3-wire receptacle is not available, consult the hospital electrician.
Connect the grounding wire to the equipotential grounding terminal on the main system. If it is not
evident from the instrument specifications whether a particular instrument combination is hazardous
or not, for example due to summation of leakage currents, the user should consult the manufacturers
concerned or else an expert in the field, to ensure that the necessary safety of all instruments
concerned will not be impaired by the proposed combination.
- 62 -
Page 74
M8B Multi-Parameter Patient Monitor User Manual
Equipotential Grounding
Protection class 1 instruments are already included in the protective grounding (protective earth)
system of the room by way of grounding contacts in the power plug. For internal examinations on
the heart or the brain, the M8B must have a separate connection to the equipotential grounding
system. One end of the equipotential grounding cable (potential equalization conductor) is
connected to the equipotential grounding terminal on the instrument rear panel and the other end to
one point of the equipotential grounding system. The equipotential grounding system assumes the
safety function of the protective grounding conductor if ever there is a break in the protective
grounding system. Examinations in or on the heart (or brain) should only be carried out in
medically used rooms incorporating an equipotential grounding system. Check each time before use
that the instrument is in perfect working order. The cable connecting the patient to the instrument
must be free of electrolyte.
WARNING
If the protective grounding (protective earth) system is doubtful, the M8B must be supplied
by inner power only.
Condensation
Make sure that during operation, the instrument is free of condensation. Condensation can form
when equipment is moved from one building to another, thus being exposed to moisture and
differences in temperature.
Safety Precautions
WARNING and CAUTION messages must be observed. To avoid the possibility of injury, observe
the following precautions during the operation of the instrument.
WARNING
The M8B is provided for the use of qualified physicians or personnel professionally trained.
And they should be familiar with the contents of this user manual before operation.
WARNING
Only qualified service engineers can install this equipment. And only service engineers
authorized by EDAN can open the shell.
WARNING
EXPLOSION HAZARD-Do not use M8B in a flammable atmosphere where concentrations
- 63 -
Page 75
M8B Multi-Parameter Patient Monitor User Manual
of flammable anesthetics or other materials may occur.
WARNING
SHOCK HAZARD-The power receptacle must be a three-wire grounded outlet. A hospital
grade outlet is required. Never adapt the three-prong plug from the monitor to fit a two-slot
outlet.
WARNING
SHOCK HAZARD-Do not attempt to connect or disconnect a power cord with wet hands.
Make certain that your hands are clean and dry before touching a power cord.
WARNING
Accessory equipment connected to the analog and digital interfaces must be certified
complying with the respective IEC standards (e.g. IEC 950 for data processing equipment
and IEC 60601-1 for medical equipment). Furthermore all configurations shall comply with
the valid version of the system standard IEC 60601-1-1. Everybody who connects
additional equipment to the signal input connector or signal output connector configures a
medical system, and is therefore responsible that the system complies with the
requirements of the valid version of the system standard IEC 60601-1-1. If in doubt, consult
our technical service department or your local distributor.
WARNING
Before using the rechargeable lithium-ion battery (hereinafter called battery), be sure to
read the user manual and safety precautions thoroughly.
WARNING
Do not place battery in the M8B with the (+) and (-) in the wrong way around.
WARNING
Do not connect the positive (+) and negative (-) terminals with metal objects, and do not put
the battery together with metal object, which can result in short circuit.
WARNING
Do not unplug the battery when monitoring.
- 64 -
Page 76
M8B Multi-Parameter Patient Monitor User Manual
WARNING
Do not heat or throw battery into fire.
WARNING
Do not use, leave battery close to fire or other places where temperature may be above 60.
WARNING
Do not immerse, throw, and wet battery in water/seawater.
WARNING
Do not destroy the battery: Do not pierce battery with a sharp object such as a needle; do
not hit with a hammer, step on or throw or drop to cause strong shock; do not disassemble
or modify the battery.
WARNING
Use the battery only in the M8B. Do not connect battery directly to an electric outlet or
cigarette lighter charger.
WARNING
Do not solder the leading wire and the battery terminal directly.
WARNING
If liquid leaking from the battery gets into your eyes, do not rub your eyes. Wash them well
with clean water and go to see a doctor immediately.
WARNING
If liquid of the battery leaks onto your skin or clothes, wash well with fresh water
immediately.
WARNING
Keep away from fire immediately when leakage or foul odor is detected.
- 65 -
Page 77
M8B Multi-Parameter Patient Monitor User Manual
WARNING
Stop using the battery if abnormal heat, odor, discoloration, deformation or abnormal
condition is detected during use, charge, or storage. Keep it away from the monitor.
WARNING
Do not use a battery with serious scar or deformation.
WARNING
Only patient cable and other accessories supplied by EDAN can be used. Or else, the
performance and electric shock protection can not be guaranteed.
WARNING
The user should check the monitor and accessories before use.
WARNING
Be sure that all electrodes have been connected to the patient correctly before operation.
WARNING
When connecting the cables and electrodes, make sure no conductive part is in contact
with the ground. Verify that all ECG electrodes, including neutral electrodes, are securely
attached to the patient.
WARNING
Do not touch the patient, bed or instrument during defibrillation.
WARNING
Please set the alarm according to the individual status of patient to avoid delaying treatment.
Ensure there will be alarm audio prompt when alarming.
WARNING
Devices connecting with monitor should be equipotential.
- 66 -
Page 78
M8B Multi-Parameter Patient Monitor User Manual
WARNING
When the M8B and electrosurgical device are used together, the user (physician or nurse)
should guarantee the safety of patient.
WARNING
The M8B will be damaged if the water level in the hydropohobic bottle reachs the bottom of
drainpipe.
WARNING
Please replace the hydrophobic bottle if it is nearly full.
WARNING
Please sterilize and disinfect timely to prevent the cross infection between patients.
CAUTION
Electromagnetic Interference - Ensure that the environment in which the M8B is installed
is not subject to any sources of strong electromagnetic interference, such as radio
transmitters, mobile telephones, etc.
CAUTION
The device is designed for continuous and is ordinary (i.e. no drip or splash-proof).
CAUTION
Keep the environment clean. Avoid vibration. Keep it far from corrosive medicine, dust area,
high-temperature and humid environment.
CAUTION
Do not immerse transducers in liquid. When using solutions, use sterile wipes to avoid
pouring fluids directly on the transducer.
CAUTION
Do not use autoclave or gas sterilizer to the M8B, recorder or any accessories.
- 67 -
Page 79
M8B Multi-Parameter Patient Monitor User Manual
CAUTION
The M8B and reusable accessories could be sent back to the manufacturer for recycling or
proper disposal after their useful lives.
CAUTION
Remove a battery whose life cycle has expired from the monitor immediately.
CAUTION
Avoid liquid splash and excessive temperature. The temperature must be kept between 5 and 40 while working. And it should be kept between -20 and 55 during
transportation and storage.
CAUTION
Before use, the equipment, patient cable and electrodes etc. should be checked.
Replacement should be taken if there is any evident defectiveness or aging symptom which
may impair the safety or performance.
Explanation of Symbols in the Monitor
This symbol means BE CAREFUL. Refer to the user manual.
This symbol indicates that the equipment is IEC 60601-1 Type CF equipment. The unit
displaying this symbol contains an F-Type isolated (floating) patient applied part providing
a high degree of protection against shock, and is suitable for use during defibrillation.
This symbol consisting of two parts, see below.
It indicates that the equipment should be sent to the special agencies according to local
regulation for separate collection after its useful life.
It indicates that the equipment is put on the market after 13th August 2005.
- 68 -
Page 80
M8B Multi-Parameter Patient Monitor User Manual
Serial number
CE Mark
Authorised representative in the European community
Manufacture date
Manufacturer
Equipotential grounding system
Standby. It designates that the switch or switch position which one part of the device has
been switched on, while the device is at the status of stand-by.
- 69 -
Page 81
M8B Multi-Parameter Patient Monitor User Manual
11 Maintenance/Cleaning
11.1 System Check
Before using M8B, do the following:
Check if there is any mechanical damage;
Check if all the outer cables, inserted modules and accessories are in good condition;
Check all the functions of the M8B to make sure that the monitor is in good condition.
If you find any damage on the monitor, stop using the monitor on patient, and contact the
biomedical engineer of the hospital or Customer Service immediately.
The overall check of the monitor, including the safety check, should be performed only by qualified
personnel once every 6 to 12 months, and each time after fixing up.
All the checks that need to open the M8B should be performed by qualified customer service
technician. The safety and maintenance check can be conducted by persons from this company. You can obtain the material about the customer service contract from the local company’s office.
WARNING
If the hospital or agency that is responding to using the M8B does not follow a satisfactory
maintenance schedule, the M8B may become invalid, and the human health may be
endangered.
NOTE:
To prolong the life of chargeable battery, charging it once every month at least is
recommended, and it must be done after the electric energy is run out.
WARNING
Replace batteries according to the instruction of our servicing engineer.
11.2 General Cleaning
WARNING
Before cleaning the monitor or the sensor, make sure that the equipment is switched off and
disconnected from the power line.
M8B Multi-parameter patient monitor must be kept dust-free.
- 70 -
Page 82
M8B Multi-Parameter Patient Monitor User Manual
Regular cleaning of the monitor shell and the screen is strongly recommended. Use only
non-caustic detergents such as soap and water to clean the monitor shell.
CAUTION
Please pay special attention to the following items:
1. Avoid using ammonia-based or acetone-based cleaners such as acetone.
2. Most cleaning agents must be diluted before use. Follow the manufacturer's
directions carefully to avoid damaging the monitor.
3. Do not use the grinding material, such as steel wool etc.
4. Do not let the cleaning agent enter into the chassis of the system.
5. Do not leave the cleaning agents at any part of the equipment.
11.3 Cleaning Agents
Examples of disinfectants that can be used on the instrument casing are listed below:
Diluted Ammonia Water
Diluted Sodium Hypochlorite (Bleaching agent).
NOTE:
The diluted sodium hypochlorite from 500 ppm (1:100 diluted bleaching agent) to 5000 ppm
(1:10 bleaching agents) is very effective. The concentration of the diluted sodium
hypochlorite depends on how many organisms (blood, mucus) on the surface of the chassis
to be cleaned.
Diluted Formaldehyde 35% ~ 37%
Hydrogen Peroxide 3%
Alcohol
Isopropanol
NOTE:
M8B and sensor surface can be cleaned with hospital -grade ethanol and dried in air or with
crisp and clean cloth.
NOTE:
This company has no responsibility for the effectiveness of controlling infectious disease
using these chemical agents. Please contact infectious disease experts in your hospital for
details.
- 71 -
Page 83
M8B Multi-Parameter Patient Monitor User Manual
11.4 Sterilization
To avoid extended damage to the equipment, sterilization is only recommended when stipulated as
necessary in the Hospital Maintenance Schedule. Sterilization facilities should be cleaned first.
Recommended sterilization material: Ethylate, and Acetaldehyde.
Appropriate sterilization materials for ECG lead, blood pressure cuff are introduced in relative
chapters respectively.
WARNING
Please sterilize and disinfect timely to prevent the cross infection between patients.
CAUTION
Follow the manufacturers instruction to dilute the solution, or adopt the lowest possible
density.
Do not let liquid enter the monitor.
No part of this monitor can be subjected to immersion in liquid.
Do not pour liquid onto the monitor during sterilization.
Use a moistened cloth to wipe up any agent remained on the monitor.
11.5 Disinfection
To avoid extended damage to the equipment, disinfection is only recommended when stipulated as
necessary in the Hospital Maintenance Schedule. Disinfection facilities should be cleaned first.
Appropriate disinfection materials for ECG lead, SpO2 sensor, blood pressure cuff, TEMP probe,
IBP sensor are introduced in relative chapters respectively.
CAUTION
Do not use EtO gas or formaldehyde to disinfect the monitor.
11.6 Replacement of Fuse
Unscrew the fuse cap anticlockwise, replace the fuse (protector tube) and screw down the fuse cap clockwise. Fuse size: Ф5x20, Rated value: T1.6AL/250V.
NOTE:
Switch off the power switch of the M8B before examining the fuse.
- 72 -
Page 84
M8B Multi-Parameter Patient Monitor User Manual
12 ECG/RESP Monitoring
12.1 What Is ECG Monitoring
Monitoring the ECG produces a continuous waveform of the patient's cardiac electric activity to
enable an accurate assessment of his current physiological state. Only proper connection of the
ECG cables can ensure satisfactory measurement. On the Normal Display, The monitor provides
display of 2-channel ECG waveforms.
The patient cable consists of 2 parts
The cable that connects to the monitor,
The lead set that connects to the patient.
Using a 3-lead set, one waveform is displayed on the screen. Using a 5-lead set, the ECG can
derive up to two waveforms from two different leads. For requested lead, you may choose from
the left side of ECG waveform.
The monitor displays the Heart Rate (HR), ST segment and Arrhythmia analysis.
Lead off detecting, detect all the electrods, indicate the broken off leads.
Anti-electrotome function: if the monitor works with high-frequency electrotome, it will not be
deadlock or restarting.
Every ECG channel has Pacing impulse rejection and Bandpass filter circuit.
Defibrillation protection (needs 1K resistance ECG cables in series) and hardware clampe
function.
A 20 seconds monitor stabilization period shall be allowed before testing. The active noise suppression is less than 0.1µA, has Tall T-wave rejection capability.
Response time of heart rate meter to change in heart rate is less than 10s.
The type of averaging done to compute the minute heart rate, its updating interval is 1s.
Alarm for tachycardia with different plus is given within 10s.
NOTE:
In the default settings of the M8B, the ECG waveforms are the first two waveforms from top
in the Waveform Area.
NOTE
The monitor can only be used on one patient at the same time.
NOTE
The defibrillator cables should be used in the ECG monitoring, it can avoid burned by high frequency.
- 73 -
Page 85
M8B Multi-Parameter Patient Monitor User Manual
12.2 Precautions during ECG Monitoring
WARNING
Do not come into contact with the patient, table, or the monitor during defibrillation.
WARNING
Use only the original ECG cable for monitoring.
WARNING
When connecting the cables and electrodes, make sure no conductive part is in contact
with the ground. Verify that all ECG electrodes, including neutral electrodes, are securely
attached to the patient but not the conductive part or ground.
NOTE:
Interference from a non-grounded instrument near the patient and ESU interference can
cause inaccuracy of the waveform.
EN60601-1-2 (protection against radiation is 3v/m) specifies that the electrical field density
exceeding 1v/m may cause measurement error in various frequencies. It is accordingly
suggested that do not use equipment generating electrical radiation near ECG/RESP
monitoring devices.
NOTE:
The simultaneous use of cardiac pacemaker or other patient-connected equipments does
not cause safety hazard.
NOTE:
If the pacemaker signals are higher than the claimed range, the heart rate may be calculated incorrectly.
12.3 Monitoring Procedure
12.3.1 Preparation
1. Prepare the patient's skin prior to placing the electrodes.
The skin is a poor conductor of electricity, therefore preparation of the patient's skin is
- 74 -
Page 86
M8B Multi-Parameter Patient Monitor User Manual
important to facilitate good electrode contact to skin.
Shave hair from sites, if necessary.
Wash sites thoroughly with soap and water. (Never use ether or pure alcohol, because this
increases skin impedance).
Rub the skin briskly to increase capillary blood flow in the tissues and remove skin scurf
and grease.
2. Attach clip or snap to electrodes prior to placement.
3. Put the electrodes on the patient. Before attaching, apply some conductive jelly on the
electrodes if the electrodes are not electrolyte self-supplied.
4. Connect the electrode lead to the patient's cable.
5. Make sure the monitor is ready with power supply.
WARNING
Placed the electrode carefully and make sure contact good.
WARNING
Check everyday whether there is skin irritation resulted from the ECG electrodes. If so,
replace electrodes every 24 hours or change their sites.
WARNING
Check if the lead connection is correct before monitoring. Unplug the ECG cable from the
socket, the screen will display the error message ECG LEAD OFF and the audible alarm
is activated.
NOTE:
For protecting environment, the used electrodes must be recycled or disposed of properly.
12.3.2 Installing ECG Lead
Placing the Electrodes for ECG Monitoring
Electrode placement for 3-lead set (Figure 12-1)
Red (R) electrode - Be placed near the right shoulder, directly below the clavicle.
Yellow (L) electrode - Be placed near the left shoulder, directly below the clavicle.
Green (F) electrode - Be placed on the left hypogastrium.
NOTE: The following table gives the corresponding lead names used in Europe and America respectively. (Lead names are represented by R, L, F in Europe, whose
corresponding lead names in America are RA, LA and LL.)
- 75 -
Page 87
M8B Multi-Parameter Patient Monitor User Manual
America Europe
Lead names Color Lead names Color
RA White R Red
LA Black L Yellow
LL Red F Green
Figure 12-1 Electrode Placement for 3-lead Set
Electrode placement for 5-lead set (Figure 12-2)
Red (R) electrode - Be placed near the right shoulder, directly below the clavicle.
Yellow (L) electrode - Be placed near the left shoulder, directly below the clavicle.
Black (N) electrode - Be placed on the right hypogastrium.
Green (F) electrode - Be placed on the left hypogastrium.
White (C) electrode - Be placed on the chest as illustrated in the Figure 12-3.
NOTE:
The following table gives the corresponding lead names used in Europe and America
respectively. (Lead names are represented by R, L, N, F and C respectively in Europe,
whose corresponding lead names in America are RA, LA, RL, LL and V.)
America Euro
Lead names Color Lead names Color
RA White R Red
LA Black L Yellow
LL Red F Green
- 76 -
Page 88
M8B Multi-Parameter Patient Monitor User Manual
RL Green N Black
V Brown C White
Lewis
R
Red
C
White
N
Black
Figure 12-2 Electrode Placement for 5-lead Set
L
Yellow
F
Green
NOTE:
To ensure patient safety, all leads must be attached to the patient.
For 5-lead set, attach the C (V)-electrode to one of the indicated positions as below (Figure 12-3):
V1 On the 4th intercostal space at the right sterna margin.
V2 On the 4th intercostal space at the left sterna margin.
V3 Midway between V2 and V4 electrodes.
V4 On the 5th intercostal space at the left clavicular line.
V5 On the left anterior axillary line, horizontal with V4 electrode.
V6 On the left middle axillary line, horizontal with V4 electrode.
V3R-V7R On the right side of the chest in positions corresponding to those on the left.
VE Over the xiphoid position.
V7 On the 5th intercostal space at the left posterior axillary line of back.
V7R On the 5th intercostal space at the right posterior axillary line of back.
- 77 -
Page 89
M8B Multi-Parameter Patient Monitor User Manual
Figure 12-3 C-electrode Placement for 5-lead Set
Recommended ECG Lead Placement for Surgical Patients
WARNING
When using Electrosurgery (ES) equipment, leads should be placed in a position in equal
distance from Electrosurgery electrotome and the ES grounding plate to avoid cautery.
Electrosurgery equipment wire and ECG cable must not be tangled up.
Monitoring ECG leads are mainly used for monitoring the patient’s vital signs. When using the
M8B with other Electrosurgery equipments, we advice to use the counteracting defibrillation ECG
lead.
The placement of the ECG leads will depend on the type of surgery that is being performed. For
example, with open chest surgery the electrodes may be placed laterally on the chest or on the back.
In the operating room, artifacts may affect the ECG waveform due to the use of ES (Electrosurgery)
equipment. To help reduce this you can place the electrodes on the right and left shoulders, the right
and left sides near the abdomen, and the chest lead on the left side at mid-chest. Avoid placing the
electrodes on the upper arm, otherwise the ECG waveform will be too small.
WARNING
When using the M8B with the defibrillator or other high frequency equipments,
counteracting defibrillation ECG lead should be used to avoid cautery.
- 78 -
Page 90
M8B Multi-Parameter Patient Monitor User Manual
WARNING
When using Electrosurgery (ES) equipment, do not place an electrode near the grounding
plate of the Electrosurgery device, otherwise there will be a great deal of interference with
the ECG signal.
Using 5-lead ECG Set
You can set the leads on ECG CH1 and ECG CH2 according your needs. The lead label is displayed
on the upper left part of the waveform. You can set them corresponding to any two from I, II, III,
AVR, AVL, AVF and V. If you set both to the same value, one of them will be adjusted to another
option automatically. (Figure 12- 4)
F
Figure 12-4 ECG Lead
NOTE:
If an ECG waveform is not accurate, while the electrodes are tightly attached, try to change
the lead.
NOTE:
Interference from a non-grounded instrument near the patient and ESU interference can
cause inaccuracy of the waveform.
Normal QRS complex should be:
Tall and narrow with no notches.
With tall R-wave completely above or below the baseline.
With pacemakr spike no higher than R-wave height.
With T-wave less than one-third of the R-wave height.
- 79 -
Page 91
M8B Multi-Parameter Patient Monitor User Manual
① ② ③
With P-wave much smaller than the T-wave.
Figure 12-5 Standard ECG Waveform
12.4 ECG Screen Hot Keys
Figure 12-6 Hot Key for ECG
Leads of channel 1:
1The selectable leads are I, II, III, aVR, aVL, aVF, V. 2When the ECG is 5-lead, the selectable leads are: I, II, III, aVR, aVL, aVF, V; when ECG is 3-lead, the selectable leads are: IIIIII. 3Leads on the ECG wave must not have the same name. Otherwise, the system will
automatically change the ECG waveform name that has the same name as the waveform being
currently adjusted to another name.
Waveform gain of channel 1: used to adjust the size of ECG waveforms
Signal amplification and collection of Channel 1 ECG (12 bits, 500Hz), it can collect plus value
- 80 -
Page 92
M8B Multi-Parameter Patient Monitor User Manual
for each channel setting as ×0.25, ×0.5, ×1, ×2, or AUTO mode. Under "AUTO" mode, the
monitor chooses an appropriate level automatically. A 1mv scale displays on each ECG
channel's right side. The height of 1mV bar is directly proportional to the waveform amplitude.
NOTE:
When the input signals are too large, the peak of the waveform may be not able to be
displayed. In this case the user may manually change the setup method of ECG waveform
according to the actual waveform so as to avoid the occurrence of the unfavorable
phenomena.
Filter method: used for displaying clearer and more detailed waveform
There are three filter modes for selection. DIAGNOSTIC, MONITOR and SURGERY modes
may reduce perturbance and interference from Electrosurgery equipment. The filter method is
the item applicable for both channels, which is always displayed at the waveform place of the
channel 1 ECG waveform.
NOTE:
Only in Diagnosis mode, the system can provide non-processed real signals. In M8B or
Sugery mode, ECG waveforms may have distortion of different extent. In either of the latter
two modes, the system can only show the basic ECG and the results of ST analysis may
also be greatly affected. In Surgery mode, results of ARR analysis may be somewhat
affected. Therefore, it is suggested that in the environment having relative small
interference, youd better monitor a patient in Diagnosis mode.
Leads of channel 2: refer to for detailed information.Waveform gain of channel 2: refer to for detailed information.
NOTE:
Pacemaking signal detected is marked by a " | " above the ECG waveform.
12.5 ECG Menu
Pick the ECG hot key on the screen, and the following menu will pop up.
- 81 -
Page 93
ECG Alarm Setting
M8B Multi-Parameter Patient Monitor User Manual
Figure 12-7 ECG Setup
HR ALM: pick "ON" to enable prompt message and data record during the ECG alarm; pick
"OFF" to disable the alarm function, and there will be a beside ECG.
WARNING
In order to avoid endangering the patients life, the user should use this function cautiously.
ALM LEV: selectable from HIGH, MED, LOW. Level HIGH represents the most serious
case.
ALM REC: pick "ON" to enable report printing upon ECG alarm.
ALM HI: used to set up the upper limit of ECG alarm.
ALM LO: used to set up the lower limit of ECG alarm.
ECG alarm is activated when the heart beat exceeds set ALM HI value or falls below ALM LO
value.
The adjusting range of ECG alarm limits is: 15~350.
ECG alarm limits for default:
ALM HI ALM LO Step
HR ADU 120 50 1
HR PED 160 75 1
HR NEO 200 100 1
The adjusting range of ECG alarm limits
Max.ALM HI Min. ALM LO Step
HR ADU 300 15 1
- 82 -
Page 94
M8B Multi-Parameter Patient Monitor User Manual
HR PED 350 15 1
HR NEO 350 15 1
NOTE:
Please set the alarm limits according to clinical condition of individual patient. The upper
limit shall not exceed 20 beats per minute higher than the patient's heart rate.
WARNING
Response time for heart rate meter to change in heart rate calculation is less than 10s. The updating interval of averaging type done to compute the heart rate per minute is 1s.
HR FROM
ECG, SpO2, AUTO and BOTH may detect heart rate. AUTO distinguishes heart rate source
according to the quality of signal. When the quality of ECG signal and SpO2 signal is the same,
ECG takes priority over SpO2. By picking ECG, the monitor prompts HR and activates HR beep.
By picking SpO2, the monitor prompts PULSE and activates pulse beep. BOTH mode displays HR
and PR simultaneously, when this item is picked, PR parameter is displayed to the right side of
SpO2. As for the sound of HR or PR in BOTH mode, HR is given the priority, i.e., if HR is available,
whose sound will be sent out, but if HR is not available, then the sound will be for PR.
HR CHANNEL
"CH1" to count the heart rate by CH 1 waveform
"CH2" to count the heart rate by CH 2 waveform
LEAD TYPE: used to select either 5 LEADS or 3 LEADS.
SWEEP
Available options for ECG SWEEP are 12.5, 25.0, and 50.0 mm/s.
ST ANALYSIS
Pick this item to access ST ANALYSIS menu, the detailed information about the menu is to be
discussed in the following section.
ARR ANALYSIS
Pick this item to access ARR ANALYSIS menu, the detailed information about the menu is to
be discussed in the following section.
OTHER SETUP
Pick this item to access ECG SETUP menu as shown below:
- 83 -
Page 95
M8B Multi-Parameter Patient Monitor User Manual
Figure 12-8 ECG Setup
In the sub-menu, following functions are available:
ECG DISPLAY
Select NORMAL DISPLAY to display 2 ECG waveforms for 5-lead (for 3-lead, only 1 ECG
waveform is displayed.). Select MULTI-LEADS DISPLAY, the waveform area on the screen only displays 7 ECG waveformsoccupying the area of 7 waveforms. Select HALF-SCAN
MULTI-LEADS, those 7 ECG waveforms are displayed on the screen, occupying the area of 4
waveforms.
Figure 12-9 MULTI-LEADS DISPLAY
- 84 -
Page 96
M8B Multi-Parameter Patient Monitor User Manual
Figure 12-10 HALF-SCAN MULTI-LEADS display
NOTE:
If 3 LEADS is selected in the ECG SETUP menu, only NORMAL DISPLAY can be selected
for ECG DISPLAY item in the sub-menu.
BEAT VOL
Six selections are available: 0, 1, 2, 3, 4, 5. 5 indicates maximum volume. 0 indicates no
sound.
PACE
"ON" detected signal will be marked by a " | " above the ECG waveform
"OFF" for non-pacemaking patient.
NOTE:
If monitoring a patient with the pacemaker, set PACE to On. If monitoring a patient without
pacemaker, set PACE to Off.
If PACE is on, the system will not perform some types of ARR analysis. For detailed
information, please refer to the section: ARR ALARM.
PACE FROM: to set ON or OFF PACE FROM which means all the ECG waveforms occupy 2
channels waveform area. It functions only when the ECG DISPLAY is set to NORMAL
DISPLAY.
ECG CAL
Pick this item to start ECG calibrating process. Picking this item again or change lead name on
the screen can end calibrating process.
It can use Standardizing voltage to set the display width for 1mV signal, such as X1 is for
- 85 -
Page 97
M8B Multi-Parameter Patient Monitor User Manual
10mm, X2 is for 20mm.
ADJUST WAVE POS
Pick this item to call up the ADJUST WAVE POS dialog box and adjust the position of ECG
waveforms on the screen. The user may use CH NAME item to select the channel to be
adjusted, UP-DOWN to adjust the position of the selected channel on the screen, BACK TO
DEFAULT to let the waveform go back to the default position on the screen.
Figure 12-11 ADJUST WAVE POS Menu
DEFAULT
Pick the DEFAULT item to call up the ECG DEFAULT CONFIG dialog box, in which you can
select the FACTORY DEFAULT CONFIG or the USER DEFAULT CONFIG item. After
selecting any of the items and exiting the dialog box, the system will pop up a dialog box
asking for your confirmation.
Figure 12-12 ECG Default Configure
WARNING
For pacemaker patient, the pacing impulse analysis function must be switched on,
otherwise, the pacing impulse may be counted as normal QRS complex, which results in
failure of detection.
- 86 -
Page 98
M8B Multi-Parameter Patient Monitor User Manual
NOTE:
When Pacer Switch is On, the Arrhythmia events related to PVCs will not be monitored. At
the same time, the ST analysis will not be performed either.
If the monitor can do ST segment monitoring and Arrhythmia monitoring, please refer to section
12.7 and 12.8.
12.6 ECG Alarm Information
12.6.1 Alarm Message
Alarms occurring in the process of ECG measurement contain two types: physiological alarm and
technical alarm. For the audio and visual features during the appearance of these alarms in the
process of ECG measurement, please refer to the related description in Chapter Alarm. In the
screen, physiological alarm messages are displayed in the Physiological Alarm area. Technical
alarms messages are displayed in the Technical Alarm area.This section does not describe the
content about Arr. and ST analysis.
Tables below describe respectively the possible various alarms those may occur during the
measurement.
Physiological alarms:
Message Cause Alarm level
ECG SIGNAL WEAK
Can not detect the signal in designated time period.
High
HR TOO HIGH HR measuring value is above the upper alarm limit User-selectable
HR TOO LOW HR measuring value is below the lower alarm limit User-selectable
Technical alarms: (take American Standard for example)
Message Cause
Alarm level
Remedy
ECG LEAD OFF
ECG C LEAD OFF
More than one ECG electrodes fall off the skin or ECG cables fall off the monitor.
ECG electrode C falls off the skin or ECG cables fall off the monitor.
Low
Low
- 87 -
Make sure that all electrodes, leads and patient cables are properly connected.
Page 99
M8B Multi-Parameter Patient Monitor User Manual
ECG V LEAD OFF
ECGLEAD OFF
ECGLEAD OFF
ECG F LEAD OFF
ECG electrode V falls
off the skin or ECG
cables fall off.
ECG electrode falls
off the skin or ECG
cables fall off the
monitor.
ECG electrode falls
off the skin or ECG
cables fall off the
monitor.
ECG electrode F falls
off the skin or ECG
cables fall off the
monitor.
Low
Low
Low
Low
ECG L LEAD OFF
ECG R LEAD OFF
ECG LL LEAD OFF
ECG LA LEAD OFF
ECG electrode L falls
off the skin or ECG
cables fall off the
monitor.
ECG electrode R falls
off the skin or ECG
cables fall off the
monitor.
ECG electrode LL falls off the skin or ECG cables fall off the monitor.
ECG electrode LA falls off the skin or ECG cables fall off the monitor.
Low
Make sure that all electrodes, leads and patient cables are properly connected.
Low
Low
Low
ECG RA LEAD OFF
ECG electrode RA falls off the skin or ECG cables fall off the monitor.
Low
- 88 -
Page 100
M8B Multi-Parameter Patient Monitor User Manual
ECG EXCEED
ECG INIT ERR
ECG INIT ERR 1
ECG INIT ERR 2
ECG INIT ERR 3
ECG INIT ERR 4
ECG INIT ERR 5
ECG measuring value is beyond measuring range.
ECG module failure
ECG module failure
High
High
High
Check lead connection and patient condition
Stop using measuring function of ECG module, notify biomedical engineer or Manufacturer’s service staff.
Stop using measuring function of ECG module, notify biomedical engineer or Manufacturers service staff.
ECG INIT ERR 6
ECG INIT ERR 7
ECG INIT ERR 8
ECG COMM STOP
ECG COMM ERR
HR ALM LMT ERR
ECG NOISE
ECG module failure or communication failure
High
Functional safety failure High
ECG measuring signal is greatly interfered.
Low
Stop using measuring function of ECG module, notify biomedical engineer or Manufacturer’s service staff.
Stop using HR alarm function, notify biomedical engineer or Manufacturers service staff.
Check lead connection and patient condition
HR EXCEED
HR measuring value is beyond measuring range.
High
- 89 -
Check lead connection and patient condition
Loading...