GPI TM050-P, TM100-P, TM150-P, TM200, TM150 User Manual

...
ENGLISH
IMPORTANT NOTICE
Use TM Series meters with water and other chemicals compatible with wetted compo­nents. Do not use to meter fuel or incompatible chemicals. TM Series meters are available with either a computer for local electronic display, or a conditioned signal output module to provide a digital signal to customer interfacing equipment. TM Series meters with computer display mea­sure in gallons or litres. Refer to the Calibration Section for details.
TM Series Electronic Water Meters
User Manual
Rev. B 920786-02
04/13
TABLE OF CONTENTS
English ...........................................................1
Español ..........................................................8
Deutsch .......................................................15
Italiano .........................................................22
Français .......................................................29
SAVE THESE INSTRUCTIONS
These meters are not legal for trade applications. TM Series meters are very sensitive to electric
noise if operated within 1 to 2 inches of some electric motors or other sources of electronic noise.
INSTALLATION
Connections
Install your meter in-line either horizontally or vertically or at the end of the hose adjacent to the nozzle. Installation to metal connections is not recommended. Install as follows:
1. Plan to install turbine with a minimum straight pipe length as follows:
Upstream from the turbine, allow a mini-
mum straight pipe length of 10 times the internal diameter of the turbine.
Downstream from the turbine, allow a
minimum straight pipe length of 5 times the internal diameter of the turbine.
2. For Spigot (Pipe) End use only primer and solvents approved for PVC gluing.
For NPT and BSP Fittings wrap all connec-
tions with 3 to 4 wraps of thread tape (optional to use pipe thread sealant). Make sure the tape does not intrude into the flow path.
TM Meter with Computer Display TM Digital Pulse Meter
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
2
3. Attach meter with arrow pointed in the direc­tion of fluid flow.
4. For NPT and BSP Fittings - Hand tighten the meter at the housing ends. Do not use a wrench or similar tool to tighten. This can damage the housing.
OPEN COLLECTOR SIGNAL OUTPUT
Wiring Diagram 1
+9 to 35 volt DC
Open Collector
Signal
3
2
1
4
Red
White
Black
Common (Ground)
Terminal Block
Customer
Interfacing
Equipment
The terminal block is identified as follows:
Pin #1 6 volt square wave (not used) Pin #2 +9 to 35 volt DC Input Pin #3 Common Ground Pin #4 Open Collector signal Output
Resistor
To protect against leakage, seal all pipe threads with an appropriate sealing com­pound. Make sure the sealing compound does not intrude into the flow path.
NOTE: If connecting to new male pipe threads,
burrs and curls can adversely affect accu­racy. Correct the problem prior to turbine installation.
NOTE: Do not over tighten the flange bolts. This
may cause the gasket to be compressed into the flow stream and may decrease the accuracy of the meter.
Compatibility of this product’s material and the process fluid and/or environment should be considered prior to putting into service.
WARNING
Product should never be operated outside its published specifications for temperature or pressure. See specifications for your model.
WARNING
Make sure flow and pressure have been eliminated from process pipe prior to installing or removing product.
WARNING
Always use appropriate thread sealant or flange gaskets when connecting product to
process piping.
WARNING
Installation near high electromagnetic fields and high current fields is not recommended and may result in inaccurate readings.
CAUTION
CAUTION
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
3
Conditioned Signal Output Module Wiring
This conditioned signal output module can be wired to provide an open collector signal output or 6-volt square wave output.
Open Collector Signal Output
To achieve an open collector signal output, reference Wiring Diagram 1. The terminal block is located on the back side of the module. The module is factory assembled for open collec­tor signal output. Please provide the (820 ohm minimum) resistor.
Ten feet (3 m) of cable is provided with the module. Trim it to desired length or extend it as necessary. Distances up to 5,000 feet (1,524 m) can be achieved for open collector signal output.
Square Wave Output
To achieve square wave output, reference Wir­ing Diagram 2 and use an Electronic Digital Meter Battery Kit (sold separately) for battery power. The terminal block and battery location are located on the back side of the module. Access as follows:
1. Remove the four Phillips-head screws from the front of the module and lift the module from the turbine.
2. To change terminal block connections, loosen the appropriate screws. Reconnect the wires in the proper positions and tighten the screws.
3. Install the batteries. Make sure the positive post is in the correct position.
4. Position the module on the turbine housing. To avoid moisture damage, make sure the seal is fully seated. Tighten the four screws on the front of the module.
Ten feet (3 m) of cable is provided with the module. Trim the cable to desired length or extend it as necessary.
Verify Meter Accuracy
Before using, check the meter’s accuracy and verify calibration.
1. Make sure there is no air in the system by starting the flow until it runs steadily. Then, stop or divert the flow using a valve or nozzle.
SQUARE WAVE OUTPUT
Wiring Diagram 2
3
2
1
4
White
Black
Common (Ground)
Terminal Block
Customer
Interfacing
Equipment
6 volt Square Wave Signal
The terminal block is identified as follows:
Pin #1 6 volt square wave Pin #2 +9 to 35 volt DC Input (not used) Pin #3 Common Ground Pin #4 Open Collector signal Output (not used)
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
4
2. Meter an exact known volume into an ac­curate container. For best results, meter with one continuous full stream.
3. Check the volume against the display or recording equipment. If the amount metered is accurate, further calibration is not neces­sary. If not, refer to the Calibration Section for further instructions.
OPERATION
Computer Display – Batch and Cumulative Totals
The computer maintains two totals. The Cumula­tive Total provides continuous measurement and cannot be manually reset. The Batch Total can be reset to measure flow during a single use. The Cumulative Total is labeled TOTAL 1, Batch Total is labeled TOTAL 2 BATCH.
When the Cumulative Total reaches a display reading of 999,999 the computer will highlight an X10 icon. This indicates to the operator that a zero must be added to the 6 digits shown. When the next rollover occurs, the computer will highlight an X100 icon. This indicates to the operator that two zeros must be added to the 6 digits shown.
Press the DISPLAY button briefly to switch between the TOTAL 1, TOTAL 2 BATCH and FLOWRATE. Press DISPLAY briefly to display the TOTAL 2 BATCH. Hold the DISPLAY button for 3 seconds to reset the Batch Total to zero.
When fluid is flowing through the meter, a small propeller icon is highlighted.
NOTE: Totalization counts total units without
differentiating between gallons, litres or field calibrated units.
Flowrate Feature
To use this feature, press and release DISPLAY button until FLOWRATE icon appears. The factory set time base will be highlighted to the right of FLOWRATE (M = minutes, H = hours, D = days). When FLOWRATE is invoked, the display will be indicating rate of flow.
Activate the Meter
Computer is on continuously and always ready to perform. The computer is powered by field replaceable batteries. When display becomes
dim, faded or the low battery message appears (see below), the batteries need to be replaced. Reference the Maintenance Section for details.
Factory and Field Calibration
All calibration information is visible to the user as icons on the top line of the display, above the numeric digits.
All units are configured with a “factory” calibra­tion. Both gallons and litres are available (“GL” or “LT” will be displayed). While holding the CALIBRATE button, briefly press DISPLAY to toggle between gallons and litres. This factory calibration (indicated with FAC) is permanently programmed into the computer and is not user adjustable.
NOTE: Your computer may have other units of
measure programmed into it. If so, holding the CALIBRATE button and momentarily pressing the DISPLAY button will toggle through all factory set units. Other possible units are: IGL (imperial gallon), QT (quart), CF (cubic feet), CM (cubic meter), BL (42 gal. barrel), CC (cubic centimeter) or OZ (ounce).
Switching between different units will not corrupt the Total’s contents. For example, in GL mode, the computer totalizes 10.00 gallons, if the user switches to LT mode, the display will read 37.85 litres (the same volume, different unit).
The “field” calibration may be set by the user, and can be changed or modified at any time using the calibration procedure described in the Calibration Section. Totals or flowrate derived from the field calibration are invoked when the FAC icon is no longer visible on the top line of the display.
CALIBRATION
Verify Accuracy Before Beginning Field Calibration
For the most accurate results, dispense at a flowrate which best simulates your actual oper­ating conditions. Avoid “dribbling” more fluid or repeatedly starting and stopping the flow. This can result in less accurate calibrations.
Make sure you meet the meter’s minimum flowrate requirements:
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
5
TM Series Meters
½ inch meter 1 GPM (3.8 LPM) ¾ inch meter 2 GPM (7.5 LPM) 1 inch meter 5 GPM (18.8 LPM) 1 ½ inch meter 10 GPM (37.5 LPM) 2 inch meter 20 GPM (75 LPM)
The use of a uniformly dependable, accurate calibration container is recommended for the most accurate results. Due to high flowrate, it is strongly recommended that calibration be completed with a combination of volume and weight using fine resolution scales.
For best results, the meter should be installed and purged of air before field calibration.
Field Calibration with Computer Display
Field Calibration and Factory Calibration are defined in the Operation Section. Factory calibration settings are programmed into each computer during manufacturing, using water at 70° F (21° C). Readings using the Factory Calibration (FAC) may not be accurate in some situations, for example, under extreme tem­perature conditions, non-standard plumbing configurations or with fluids other than water.
Field Calibration Procedures (Correction Factor Method)
1. To calibrate, press and hold the CALIBRATE and DISPLAY buttons for about 3 seconds until you see FLdCAL. Release both buttons and you will see CF - 00.0. You are now in the field calibration mode and values from
-99.9% to +99.9% can be entered.
2. The
+
/– position appears either as an “under­score” character for plus, or as a “hyphen” character for minus. The DISPLAY button selects the position and the CALIBRATE button toggles this character.
3. The DISPLAY button can then be pushed to select the numeric positions. Press the CALIBRATE button to scroll from 0 to 9. Enter the percentage of change you want the display to correct. When satisfied with the value, press both CALIBRATE and DIS­PLAY buttons simultaneously. CALEnd will be displayed and unit will go back to normal operation, less the FAC (factory calibration) icon.
4. All enabled units-of-measure remain visible and selectable – the entered correction will be applied to all enabled units.
5. To return to factory calibration (FAC), press and hold both CALIBRATE and DISPLAY buttons for about 3 seconds until FAcCAL is displayed. Then release buttons. Unit should return to normal operation and FAC icon is visible.
Calibration with Conditioned Signal Output Module
The K-factor of your meter appears on the calibration report as the number of pulses per gallon. The factor is determined during produc­tion using water at 70° F (21° C). This K-factor may be used for “single point” calibration and provide acceptable accuracy. However, readings may not be accurate when using this calibra­tion method in some situations. For example, extreme temperature conditions, non-standard plumbing configurations or with fluids other than water.
MAINTENANCE
Proper handling and care will extend the life and service of the meter.
Turbine Rotor
The meter is virtually maintenance-free. How­ever, it is important the rotor moves freely. Keep the meter clean and free of contaminants.
If the rotor does not turn freely, apply a pene­trating lubricant on the rotor, shaft and bearings. Remove any debris or deposits from the rotor using a soft brush or small probe. Be careful not to damage the turbine rotor or supports.
Blowing compressed air through the tur­bine assembly could damage the rotor.
CAUTION
CAUTION
CAUTION
Do not allow liquids to dry inside the turbine.
Handle the rotor carefully. Small scratches or nicks can affect accuracy.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
6
Battery Replacement
The computer display is powered by two 3-volt lithium batteries which may be replaced while the meter is installed. When batteries are removed or lose power, the batch and cumula tive totals and the field and factory calibrations are re tained.
(Battery) – Avoid mechanical or electrical abuse. Batteries may explode or cause burns, if disassembled, crushed or exposed to fire or temperatures in excess of 212°F (100°C). Do not short circuit or install with incorrect polarity. DO NOT INCINERATE.
Batteries should ONLY be replaced with P/N 113520-1 Kit (Includes two each P/N 902004-2 Batteries). Do not mix old with new. Do not use other brands or technologies.
Open battery cells should be disposed of in accordance with local regulations. Lithium batteries are best disposed of as a non-hazardous waste when fully or mostly discharged. EPA does not list or exempt Lithium as a hazardous waste. If waste lithium batteries are still fully charged or only partially discharged, they can be considered a reactive hazardous waste because of unconsumed lithium remaining in the battery. Such batteries may qualify as “Universal Waste” in many jurisdictions within the U.S. and thus can be shipped for disposal or recycling in accordance with Universal Waste requirements.
CAUTION
WARNING
If the display becomes dim, blank or the low battery message appears (see below), replace the batteries as follows:
1. Remove the four Phillips-head screws from the face of the meter and lift the faceplate from the turbine.
2. Remove the old batteries and clean any corrosion from the terminals.
3. Install new batteries. Make sure the positive post is in the correct position.
4. When the batteries are replaced, the face­plate will power ON. Check the display to ensure normal functions have resumed before assembling again.
5. Reseat batteries, if necessary, and posi­tion the faceplate on the turbine housing. To avoid moisture damage, make sure the seal is fully seated. Tighten the four screws on the faceplate.
SPECIFICATIONS
Inlet and Outlet:
Spigot (Pipe) End Models: TM050/TM050-P ½ inch Schd. 80, Spigot (Pipe) TM075/TM075-P ¾ inch Schd. 80, Spigot (Pipe) TM100/TM100-P 1 inch Schd. 80, Spigot (Pipe) TM150/TM150-P 1 ½ inch Schd. 80, Spigot (Pipe) TM200/TM200-P 2 inch Schd. 80, Spigot (Pipe) NPT Models: TM050-N/TM050-N-P ½ inch NPT TM075-N/TM075-N-P ¾ inch NPT TM100-N/TM100-N-P 1 inch NPT TM150-N/TM150-N-P 1 ½ inch NPT TM200-N/TM200-N-P 2 inch NPT
BSP Models: TM100-B/TM100-B-P 1 in. BSP TM150-B/TM150-B-P 1 ½ in. BSP TM200-B/TM200-B-P 2 in. BSP
Design Type: Turbine Wetted Components:
Housing: PVC Journal Bearings: Ceramic Shaft: Tungsten Carbide Rotor and Supports: PVDF Retaining Washer: Stainless Steel
Max. Working Pressure: 225 PSIG @ 73° F BSP: 150 PSIG @ 73° F
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
7
U.S. Measurement
Unit of Measure: Gallon Flow Range:
½ inch 1 - 10 GPM ¾ inch 2 - 20 GPM 1 inch 5 - 50 GPM 1 ½ inch 10 - 100 GPM 2 inch 20 - 200 GPM
Accuracy with Computer: ± 3.0% of read-
ing (Accuracy can be improved with field calibration)
Operating Temperature: +32° to +140° F
(Do not allow fluid to freeze inside meter.)
Storage Temperature: –40° to +158° F Product Weight:
*
Spigot (Pipe) NPT BSP
½ in. .38 lbs. .55 lbs. N/A ¾ in. .43 lbs. .67 lbs. N/A 1 in. .49 lbs. .84 lbs. .84 lbs. 1 ½ in. .66 lbs. 1.38 lbs. 1.47 lbs. 2 in. .78 lbs. 1.78 lbs. 1.93 lbs.
Dimensions - Inches (W x H x L):
**
Without Fitting With NPT Fitting
½" 2.1 x 2.5 x 4.3 2.1 x 2.7 x 6.0 ¾" 2.1 x 2.7 x 4.4 2.1 x 2.9 x 6.1 1" 2.1 x 2.9 x 4.5 2.1 x 3.1 x 6.5 1 ½" 2.1 x 3.6 x 5.4 2.3 x 3.8 x 7.6 2" 2.4 x 4.1 x 5.5 3.5 x 4.4 x 7.9
With BSP Fitting
½" N/A ¾" N/A 1" 2.1 x 3.1 x 6.7 1 ½" 2.2 x 3.7 x 7.6 2" 2.7 x 4.2 x 7.8
* Weight with computer display. Conditioned signal
output module adds .30 lbs.
** Dimensions with computer display. Conditioned
signal output module adds 1.1 inch to height.
Metric Measurement
Unit of Measure: Litre Flow Range:
½ inch 3.8 - 38 LPM ¾ inch 7.6 - 76 LPM 1 inch 19 - 190 LPM 1 ½ inch 38 - 380 LPM 2 inch 76 - 760 LPM
Accuracy with Computer: ± 3.0% of read-
ing (Accuracy can be improved with field calibration)
Operating Temperature: 0° to +60° C
(Do not allow fluid to freeze inside meter.)
Storage Temperature: –40° to +70° C Product Weight:
*
Spigot (Pipe) NPT BSP
½ inch .172 kg .249 kg N/A ¾ inch .195 kg .304 kg N/A 1 inch .222 kg .381 kg .381 kg 1 ½ inch .299 kg .626 kg .666 kg 2 inch .354 kg .807 kg .875 kg
Dimensions - cm (W x H x L):**
Without Fitting With NPT Fitting
½" 5.3 x 6.4 x 10.9 5.3 x 6.9 x 15.2 ¾" 5.3 x 6.9 x 11.2 5.3 x 7.3 x 15.5 1" 5.3 x 7.4 x 10.9 5.3 x 7.9 x 16.5 1 ½" 5.3 x 9.1 x 13.7 5.8 x 9.7 x 19.3 2" 6.1 x 10.4 x 14.0 8.9 x 11.2 x 20.0
With BSP Fitting
½" N/A ¾" N/A 1" 5.3 x 7.9 x 17.0 1 ½" 5.6 x 9.4 x 19.3 2" 6.9 x 10.7 x 19.8
* Weight with computer display. Conditioned signal
output module adds .14 kg.
** Dimensions with computer display. Conditioned
signal output module adds 2.8 cm to height.
PARTS
The following replacement parts and acces­sories are available for the TM Series meters:
Part No. Description
113435-1 Conditioned Signal Output Module 113520-1 Battery Replacement Kit 116000-1 Calibration Container, Large (5 gallon) 125508-03 ½ inch, Turbine Assy Kit 125508-04 ½ inch NPT, PVC Turbine Assy Kit 125510-03 ¾ inch, Turbine Assy Kit 125510-04 ¾ inch NPT, PVC Turbine Assy Kit 125512-03 1 inch, Turbine Assy Kit 125512-04 1 inch NPT, PVC Turbine Assy Kit 125514-03 1 ½ inch, Turbine Assy Kit 125514-04 1 ½ inch NPT, PVC Turbine Assy Kit 125516-03 2 inch, Turbine Assy Kit 125516-04 2 inch NPT, PVC Turbine Assy Kit 901002-52 Seal 125512-05 1 inch BSP, PVC Turbine Assy Kit 125514-05 1 ½ inch BSP, PVC Turbine Assy Kit 125516-05 2 inch BSP, PVC Turbine Assy Kit
Computer Kits:
125509-03 ½ inch, Computer Assy Kit 125511-03 ¾ inch, Computer Assy Kit 125513-03 1 inch, Computer Assy Kit 125515-03 1 ½ inch, Computer Assy Kit 125517-03 2 inch, Computer Assy Kit
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
Loading...