GPI QME10, QMF05, QME05, QMF10, QME20 Owner's Manual

...
Page 1
SAVE THESE INSTRUCTIONS
Fixed and Portable Ultrasonic Flowmeter (UFM)
Owner’s Manual
TABLE OF CONTENTS
Key Aspects of QStar UFM..........................................3
Fixed
Approvals/CE...............................................................3
Measuring Principle ..................................................... 3
QStar Fixed UFM and Components ............................ 4
QStar Portable UFM and Components........................ 8
Operating ................................................................... 10
Getting Started .......................................................... 10
Preparing for Measurement ....................................... 13
Measuring with UFM.................................................. 15
Setup Parameters......................................................15
Measuring Windows .................................................. 31
Calibration ................................................................. 44
System Settings.........................................................44
Troubleshooting ......................................................... 46
Software Update ........................................................ 53
Fluid Properties ......................................................... 54
Specications ............................................................ 56
Portable
Page 2
QSTAR FIXED AND PORTABLE UFM OWNER’S MANUAL
Text identied with an exclamation mark contains important information that relates to the basic data
and operation of the device.
Text identied with the letter “i” contain supplementary and helpful information.
i
FIXED UFM
Text contained within this box applies specically
to the QStar FIXED UFM. If you have a PORTABLE
UFM, skip this section and go to the text with no
border, or the text in the PORTABLE UFM box.
FIXED UFM
PACKAGE INCLUDES:
Transmitter
Ultrasonic transducers
Spacer bar for the ultrasonic transducers
(for types F10/F21)
Stainless steel mounting belts
Getting Started (“Quick-start”) manual
USB drive with Owner’s Manual
Ultrasonic coupling grease
Other ultrasonic transducers for smaller or larger pipe dimensions, as well as clamp-on temperature sen­sors, are available on separate order. Contact GPI at www.GPImeters.net or toll-free (888) 996-3837.
PORTABLE UFM
Text contained within this box applies specically
to the QStar PORTABLE UFM. For a FIXED
UFM, skip this section and go to the text with
no border, or the text in the FIXED UFM box.
PORTABLE UFM
PACKAGE INCLUDES:
Hard-shell case
QStar Portable ow transmitter
Plug-in power adapter, plus IEC appliance power
cable
Transducer cables
Ultrasonic transducers
Spacer bar for the ultrasonic transducers
Cable for the 4 mA to 20 mA analog output (Mini
DIN, alligator clips)
Digital output cable for the relay/pulse output
(Mini DIN, alligator clips)
2
TEL: 888-722-3569 • 973-940-1684 • FAX: 1-973-940-1651 • www.GPImeters.com
USB cable
Stainless steel mounting chains (up to 16 in.)
Getting Started (“Quick-start”) manual
USB drive with operating instructions
Ultrasonic coupling grease
Measuring tape
Other ultrasonic transducers for smaller or larger pipe dimensions and clamp-on temperature sensors, are available on separate order. Contact GPI at www.GPImeters.net or toll-free (888) 996-3837.
Page 3
KEY ASPECTS OF QSTAR UFM:
APPROVALS/CE
Fixed or Portable system for measuring liquids in lled
piping systems.
Uses the ultrasonic transit-time differential method.
Heat measurement is included as standard application.
Clamp-on Fixed and Portable temperature sensors are optional.
Portable UFM can be operated in battery-powered
mode and on a power adapter for operation with 100% duty cycle. Fixed UFM can be operated on a power adapter.
Supports measurements on piping with diameters
from 1/2" to 240" (depending on the sensor used).
The uid to measure may have a temperature range
from -40° F to +300° F (depending on the transducer used).
You can save the measuring data to the internal
SD card, read the data via USB port and export this data using Microsoft® ofce software such as Excel (Portable UFM only).
The device is equipped with an electrically isolated
relay output and two 4mA to 20mA current outputs that can be operated in active and passive mode.
QStar UFM is compliant with the following European Directives and Standards
Test Specications
DIN EN 55011 B (11/2007)
DIN EN 61000-4-2 (09/2008)
DIN EN 61000-4-3 (06/2008)
DIN EN 61000-4-4 (07/2005)
DIN EN 61000-4-5 (06/2007)
DIN EN 61000-4-6 (10/2008)
DIN EN 61000-4-8 (12/2001)
DIN EN 61000-4-11 (02/2005)
Test Requirements
DIN EN 61000-6-1 (10/2007)
DIN EN 61000-6-3 (09/2007)
FIGURE 1: Measuring Principle
ULTRASONIC MEASURING PRINCIPAL
TEL: 888-722-3569 • 973-940-1684 • FAX: 1-973-940-1651 • www.GPImeters.com
3
Page 4
The UFM employs precise, ultrasonic transit-time dif-
cos221
)12(
TT
TT
π
4cos221
TT
ferential method. This method involves installation of two ultrasonic transducers on the surface of the piping and their interconnection with the electronic evaluation system. The ultrasonic transducers operate in alternating mode as transmitter and receiver with cyclic exchange of ultrasonic signals. Measurements cover the transit times of the upstream and downstream signals (t1, t2). The UFM measures the transit-time differential of the ultrasonic signals t1 and t2 that travel upstream and downstream. These signals are accelerated (t1) or retarded (t2). The difference that develops between both signal transit times
is proportional to ow velocity and is used in combination
with the piping geometry data for precise calculation of
the volumetric owrate.
FIXED UFM AND COMPONENTS
=
Lv
α
α
D
2
Calculation of ow velocity [m/s]
)12(
TT
LQ
=
Calculation of owrate [m3/s]
The ow transmitter uses a sophisticated cross-correlation
to detect signals. This ensures a reliable detection of
signals even in case of harsh circumstances like gas
and/or particle load.
FIGURE 2: UFM – with Mounted Ultrasonic Transducers (Bottom) and Flow Transmitters
The UFM consists of the ultrasonic transducers and the
ow transmitter that are mounted onto piping.
FLOW TRANSMITTER
The ow transmitter processes the signals and makes
the measurement results available to the user.
ULTRASONIC TRANSDUCERS
The ultrasonic transducers are mounted onto the pip­ing and transmit and receive the ultrasonic signals
that are used in the ow transmitter to calculate the volumetric owrate.
Ultrasonic transducers:
QMF-F10 (1 MHz) for pipe diameters 1.25" to 16" QMF-F21 (2 MHz) for pipe diameters 3/8" to 4"
Operating temperatures: -40° F to 300° F
FIGURE 4: Ultrasonic Transducers (F10/F21) typically used with Spacer Bar (not shown)
FIGURE 3: Flow Transmitter
4
TEL: 888-722-3569 • 973-940-1684 • FAX: 1-973-940-1651 • www.GPImeters.com
Ultrasonic transducer: QMF-F05 (0.5 MHz) for pipe diameters 8" to 240" Operating temperatures: -40° F to 180° F
(300° F optional on request)
FIGURE 5: Ultrasonic Transducers (Type F05)
Page 5
FIXED UFM AND COMPONENTS (Continued)
MOUNTING MATERIAL AND ACCESSORIES
Signal cables
Signal cables are a part of the ultrasonic transducers and cannot be separated from transducers.
Spacer bar for transducer mounting
For transducers F10 and F21 (Transducer QMF-F05 is mounted on pipes using textile tape rather than spacer bar).
FIGURE 6: Spacer bar
Metal Mounting Belt for Transducer Mounting
FIGURE 7: Mounting Belt (Stainless Steel)
QMF-PT100 Temperature sensors
The clamp-on temperature sensors collect tem­perature data in heating and cooling circuits. This data is then used to calculate heating and cooling
quantities.
FIGURE 9: Clamp-on temperature sensors, QMF-PT100 (optional)
FIGURE 10: QMF-PT100 (optional) Temperature Sensor Mounted with Metal Belt
FIGURE 8: Transducers Mounted with Spacer Bar and Mounting Belts
Coupling grease
Apply the ultrasonic coupling gel between the ultra-
sonic transducer and the piping in order to optimize
signal input.
TEL: 888-722-3569 • 973-940-1684 • FAX: 1-973-940-1651 • www.GPImeters.com
Interfaces of UFM
Open the cover to access the connecting terminals.
FIGURE 11: Front view of transmitter showing cover
Cover
Remove Screws
5
Page 6
FIXED UFM AND COMPONENTS (Continued)
CONNECTOR BOARD
FIGURE 12: Connecting Terminals
1 2 3 4 5 6 7
+ - - - - - + --
UP1 DN1 UP2 DN2 RS232/ 485
1 RS232/RS485 Interface boards
Digital Interface boards RS232 or RS485 are available as an option to provide digital com­munication via ASCI strings.
2 Input for temperature sensors
QMF-PT100 (3-wire)
The two temperature sensors (feed and return pipe) can be connected in order to measure
heat/thermal output. The ow transmitter always
includes the capability for heat/thermal output
measurement. Note that (if required by customer)
the QMF-PT100 inputs can also be used to reset the counters. If you use this reset function you cannot measure heat/thermal output at same time.
WW WW
QMF-
QMF-
PT100
PT100
Relay OUT2 OUT1 DO1 DO2 Reset Power
6 Hardware Reset
Used to reset unit (for hang-ups).
7 Power Supply
Location for connecting the supply voltage. QStar UFM is available as AC (90-240VAC) and DC (18-36VDC) version.
Always use the correct voltage for the UFM. Improper supply voltage might seriously
damage the ow transmitter. Check the type
of power supply on the name plate (printed
on right side of enclosure of ow transmitter.
CAUTION
+ - (DC)
3 Relay output (potential-free)
This output is potential-free NO (normally open) relay output. Use this output to establish an alarm (for example, when exceeding a certain
ow speed).
4 Analog output 4-20mA (active)
The 4-20mA outputs can be used to submit
measurement data like ow, thermal output and
velocity to the Programmable Logic Controller (PLC). These outputs are in active mode (supply
voltage provided internally by ow transmitter).
5 USB- Interface
Used for rmware updates. Standard USB (micro USB)- cable required to connect to a PC.
6
Wiring
The connection terminals are located under the lower cover. Remove the two screws and plastic cover to gain access to the connection terminals.
FIGURE 13: Accessing the Connection Terminals
Page 7
FIXED UFM AND COMPONENTS (Continued)
.5".015" - .030"
AC TERMINAL BLOCK WIRING
Use cables with 16-26 AWG wires. Wires should be stripped about .5 in. to allow proper contact
i
to terminals.
Put the stripped end of the related wire into the related hole. Wires will be held by spring. It might be helpful to use end sleeves or tin the wire ends.
FIGURE 14: Connection Terminals – Fasten Cables
Use a slotted screwdriver to unfasten wires, then press middle part between the two holes as shown to loosen the spring and remove the wires.
FIGURE 15: Connection Terminals - Unfasten Wires
Terminal Diagram
Terminals
UP1
DN1
UP2
DN2
RS485 Optional Interface board
QMF-
PT100
QMF-
PT100
Relay Relay output, external voltage required
OUT1
OUT2
DO1
DO2 Optional, Digital output 2
Designation
Connection of upstream transducer Red wire to be connected to +
Black wire to be connected to -
Connection of downstream transducer Red wire to be connected to +
Black wire to be connected to -
Connection of upstream transducer (path2) Red wire to be connected to +
Black wire to be connected to -
Connection of downstream transducer (path2) Red wire to be connected to +
Black wire to be connected to -
Connect temperature sensors (measuring colder temperature) to left QMF-PT100 terminal
WH = White wire from QMF-PT100 RD = Red wire from QMF-PT100
Connect temperature sensors (measuring warmer temperature) to right QMF-PT100 terminal
WH = White wire from QMF-PT100 RD = Red wire from QMF-PT100
Analog output1, 4…20mA, active, volt­age provided internally
Analog output2, 4…20mA, active, volt­age provided internally
Digital output (Transistor), passive
mode, external voltage required.
PE NL1
US EUROPE
BLACK
WHITE
GREEN
BROWN
BLUE
GREEN/YELLOW
Power
Supply
Either DC (18-36VDC) or AC (90-
240VAC). Check order conrmation or name plate on ow transmitter for
operating voltage.
CAUTION
• The 4-20mA outputs are set in active mode.
That means the required voltage is provided
by ow transmitter internally. DO NOT USE
additional external voltage.
• The digital output is set in passive mode and
requires external voltage to be operated.
• Relay is rated to max. 45V, 0.25mA. These
values must not be exceeded.
7
Page 8
PORTABLE UFM AND COMPONENTS
PORTABLE FLOW TRANSMITTER
Your UFM consists of the ultrasonic transducers and the ow transmitter that are mounted onto your piping.
The ow transmitter processes the signals and provides
the measurement results.
FIGURE 16: PORTABLE Flow Transmitter (Top) and Mounted Ultrasonic Transducers (Bottom)
MOUNTING MATERIAL AND ACCESSORIES
Signal cables
Spacer bar for transducer mounting
For QMP-F10 and QMP-F21 (QMP-F05 is mounted on pipes using textile tape rather than spacer bar).
FIGURE 17: Spacer bar
Chains for Transducer Mounting
FIGURE 18: Mounting Chain (Stainless Steel)
ULTRASONIC TRANSDUCERS
The ultrasonic transducers mount onto the piping to transmit and receive the ultrasonic signals used in the
ow transmitter to calculate the volumetric owrate.
Ultrasonic transducer QMP-F21
(2 MHz), RED housing, for pipe
diameters from 0.5 to 4.0 inches. Operating temperatures: -40° F to 300° F
Ultrasonic transducer QMP-F10
(1 MHz), BLUE housing. Pipe diam­eters: 1.5 to 16 inches. Operating temperatures: -40° F to 300° F
Ultrasonic transducer QMP-F05
(0.5 MHz), GREEN housing. For
pipe diameters from 8 to 240 inches. Operating temperatures:
-40° F to 176° F (300° F optional
on request)
FIGURE 19: Transducer QMP-F05 (500 kHZ) for large
pipes – Mounting with textile tape.
Coupling grease
Apply the ultrasonic coupling gel between the ultrasonic
transducer and the piping to optimize signal input.
8
Page 9
PORTABLE UFM AND COMPONENTS (Continued)
78
QMP-PT100 Temperature sensors (Optional)
The clamp-on temperature sensors collect tem­perature data in heating and cooling circuits.
Use this data to calculate heating and cooling quantities.
FIGURE 20: QMP-PT100 clamp-on temperature sensors
4-20mA analog output cable
FIGURE 21: Back view of Portable UFM connections
1234 56
Reset Down Up Relays/ T1/T2Analog OutUSBPower
Impulse OUT
1 Power Input
This jack is used to connect the plug-in power
adapter that is included with the UFM package.
2 USB Interface (Mini.USB Type B)
Enables access to the integrated SD memory
card from a PC. This card is used to store data logging information and measurement data (LOG
les). Windows XP or later versions detect the
internal SD Card as mass storage medium. No need to install additional drivers.
QMP-001
The analog output cables can be used to connect an external data logger or recorder to the ow transmit­ter for the transmission of measured values such as
owrates, or thermal output.
Cable for Relay/Pulse
The relay connecting cable can be used to trigger
alerts. For example, when exceeding a certain owrate.
Power adapter 100-240V, 47-63Hz, 1A
The power adapter is normally used to charge the
battery. Your UFM supports permanent operation by
using the power adapter.
3 Hardware Reset
Use a small screwdriver (or paper clip) to
press the reset button.
4/5 BNC Inputs for Ultrasonic
Transducers
Jacks for the ultrasonic transducers.
6 Relay/Pulse Output (4-Pol Mini DIN)
Electrically isolated output with NO (normally
open) contact. This internal NO contact is open unless an actuating signal is generated. Allows user to assign alarm or threshold limit functions to this output.
7 Input for Temperature Sensors
QMP-PT100 (6-Pole Mini DIN)
Receptacle for the optional temperature sen-
sors that enable the use of the internal heat measurement function of the UFM.
8 4-20mA Analog Output
(5-Pol Mini DIN)
These outputs can be assigned variables such
as the owrate and return a current that is
proportional to the value of the variables. The outputs operate in active (power provided by
ow transmitter) 2-wire mode.
9
Page 10
SAFETY INSTRUCTIONS
HOW TO NAVIGATE
The ow transmitter cannot be operated outside the
temperature range from -4° F to 140° F.
The ultrasonic transducers are sensitive to mechanical
stress such as impact and vibration. Always safeguard the transducers against strong vibration or impact to avoid damage or destruction.
The plug-in power supply is suitable for in-door use
only.
The plug-in power adapter or the power cable must
be replaced completely in the case of mechanical or electrical damage.
The ow transmitter is not approved for operation in hazardous locations. The standard ultrasonic trans­ducers are not approved for operation in hazardous
locations.
• The ultrasonic transducers may not be operated outside their specied uid temperatures.
OPERATING
Use the corresponding multifunctional buttons:
Arrow buttons for navigation
ENTER
NEXT
BACK
XYZ
Conrms entry
Conrms entries and opens the next window
Returns to the previous window
Increases the value
+
Reduces the value
Activates a certain function (depending on chosen menu)
No function
CONTROL BUTTONS
1. Turns on Portable transmitter. Fixed transmitter will automatically start when connected to voltage supply.
2. Switches the backlight On and Off.
3. Multifunctional buttons: Use this button to select the function that is displayed next to it on the screen.
GETTING STARTED
BASIC SETTINGS, MAIN MENU, NAVIGATION
Setting Language
1. Switch on the device. During the start sequence, press
the multifunctional button that is located next to the
“SETUP” eld.
10
Page 11
2. Conrm the “SETUP LANG.” button
2. Select “COMPL Setup” when window appears.
3. Use the arrows in the next window to select the dialog
language. Conrm entry with “Enter”. Exit the menu with “SETUP.”
CAUTION
The language setting selects the language used
in the menus. The language in the elds next to
the multifunctional button remains unchanged.
Navigation in Main Menu “Flow 1”
The “Flow 1” measuring window is automatically opened
with a delay of a few seconds after turning on the power.
The “Flow 1” measuring window provides an overview of all data that is necessary for ow and optional heat
measurements.
You are now in the main menu. Select all necessary
functions of the device in this menu.
To return to the measuring window, proceed as follows:
Select “ESC” -> “MEAS” in the next window.
To accelerate access to the main menu after
power on, select the start sequence “SETUP.”
i
Select “COMPL SETUP” in the next window.
Setting the Time and Date
After selecting the dialog language, the setup menu opens.
1. Scroll the “System Setup” menu command using the arrow keys.
1. Select “Setup”
11
Page 12
2. Select the “Time and Date” menu command
APPLIES TO PORTABLE QSTAR
The status bar is located in the uppermost row of the display.
3. Enter the time as: Hour (hh): Minute (mm): Second (ss). Enter the date as: Month (mm): Day (dd): Year (yy).
Time
Displays the current time (system time). The time stamp that is derived from the system time will be applied to the measurement data.
SD memory
Displays the free space on the internal SD memory card of the device (standard is 2 GB).
Backup battery
Provides information about the status of the recharge­able battery.
Load: The device is powered using the power
adapter while the battery is charged. The empty
battery needs a charging time of approximately ve
hours.
Full: The battery is in a charged state. The device
may be operated for approximately ve hours when the display backlight is switched off or for approxi-
mately three hours when it is switched on.
Percentage display
Displays the charging state of the battery.
The times specied applies to a new battery. The factual operating/load cycles may deviate from the specied
time values.
12
Page 13
PREPARING FOR MEASUREMENT
The following section elaborates on essential aspects that must be taken into account for successful owrate measurements.
STRAIGHT RUN REQUIREMENTS
The selection of the mounting location has a signicant impact on measurement quality. The charge and discharge areas listed in the following table should be taken into account.
FIGURE 22: Straight run requirements
13
Page 14
TRANSDUCER MOUNTING POSITIONS
4cos221
TT
Mounting Ultrasonic Transducers
CAUTION
The pipe must always be lled completely
at the mounting positions of the ultrasonic
transducers. It is not possible to take measurements on partially lled pipe.
The ultrasonic transducer can be operated in any mount-
ing position. However, conforming with the mounting
positions shown below is mandatory. The drawing shows the side view of the piping.
Mounting Ultrasonic Transducers On
Horizontal Pipe
FIGURE 23: Preferable mounting positions for ultrasonic transducers (1)
UFM uses the cross section of the pipe to calculate
the ow. The cross section is calculated from the inner
diameter setup (user setting). If there is sedimentation in the pipe, which decreases the real inner diameter, a small amount of uncertainty may result. The same hap-
pens when the inner diameter is not known/estimated.
D
2
πα
)12(
TT
LQ
=
To measure pipes with unknown wall thicknesses a wall thickness gauge is avaiable from GPI. Ask
i
GPI Customer Service Representative for more information or visit www.GPImeters.net.
Ultrasonic Transducers On Non-Planar
Surface
Never mount the transducers on non-planar surfaces such as welding seams or deformations. Always try to remove
thick and uneven protective paint coating from the piping
area where the ultrasonic transducers will be mounted.
On horizontal piping, mount the transducer with an offset of about +/-45% to the horizontal plane. Otherwise, there is a risk of the accumulation of bubbles in the upper sec-
tion and sedimentation in the lower section of the pipe.
FIGURE 24: Preferred mounting positions for ultrasonic transducers (2)
Mounting Ultrasonic Transducers
Following are the options for mounting the ultrasonic trans­ducers. The V-mode is standard for most applications.
V-Mode
FIGURE 25: Mounting of transducers in V-mode
Transducer 1 Transducer 2
In the V-mode, both ultrasonic transducers are mounted onto the same side of the pipe. This mode is the standard for small and medium pipe dimensions. The ultrasonic
signals are reected from the pipe wall.
14
Page 15
W-Mode
Transducer 1
SETUP PARAMETERS
FIGURE 26: Mounting of Transducers in W-Mode
Transducer 1 Transducer 2
The W-mode is a special method for mounting the ultra­sonic transducers. This method is usually employed on very small pipes.
Z-Mode
FIGURE 27: Mounting of Transducers in Z-Mode
Transducer 2
The Z-mode is a special method for mounting the ul­trasonic transducers. The signal is transmitted across a shorter distance with this installation method. Use for measuring large-scale piping systems, or where the sys-
tem is lled with heavily contaminated or gas-loaded uid.
Ths section denes the input of all data that is necessary for ow measurement.
“QUICK SETUP”: The Quick Setup guide offers step­by-step instructions on the essential tasks required to
setup parameters.
“COMPL SETUP”: The complete setup function enables access to all options and expert settings.
WHAT NEEDS TO BE SETUP
The pipe’s outer diameter or circumference.
The wall thickness of the pipe. The material and thick-
ness of the pipe lining, if such lining exists.
The pipe material
The uid
The type of ultrasonic transducers
The mounting mode for the ultrasonic transducers
Ultrasonic measurement is based on the signal transit time process. The ultrasonic signals
i
penetrate the piping and the uid. In order to calculate the signal transit time, each uid, pip-
ing material and existing lining will be assigned a sonic speed value, as well as the pipe diameter or circumference value. The tabular database
species the sonic speed values for the materials and uid. The sonic speed for materials not listed
in the tables must be entered manually. Tables that list additional sonic speed parameters for
different materials are available in the back of
these operating instructions.
MEASURING WITH UFM
5 Steps to Flow Measurement:
Select a suitable location for mounting the ultrasonic
transducers
Setup the UFM pipe parameters
Mount the ultrasonic transducers onto the piping
Perform a zero calibration
Start the ow measurement
15
Page 16
USING QUICK SETUP
START
Accessing the setup dialog:
After power on: Select “Setup” -> “Quick Setup” within the start sequence.
In the primary measuring window “Flow 1”: Select “Setup” -> “Quick “Setup.”
Enter the kinematic viscosity of the uid:
Specify whether to enter
the pipe circumference
or outer diameter.
1
2
Enter the outer diameter
OR
Enter the outside
circumference
i
Use a wall thickness meter if wall thickness is unknown, or consult published pipe standards.
16
Enter the pipe’s wall thickness
Page 17
USING QUICK SETUP (CONTINUED)
Choose pipe material
3
OR
4
Does the pipe have a lining YES/NO?
NO
OR5
YES. Enter the thickness of the lining.
Select the database, or user input if a material is not listed in the database.
17
Page 18
USING QUICK SETUP (CONTINUED)
OR
Choose lining material from
the database.
5
Enter speed of sound of
user-dened coating.
Select the uid:
OR
18
Enter the kinematic viscosity of the medium:
Page 19
USING QUICK SETUP (CONTINUED)
Enter the thermal capacity of the uid:
Enter the density of the uid:
6
7
8
END
Select a suitable ultrasonic transducer:
For information on suitable transducers for specific pipe dimensions, refer to section “QStar UFM and Components.”
Select a suitable mounting mode:
For information on suitable transducers for specific pipe dimensions, refer to section “QStar UFM and Components.”
Output of the distance between the
ultrasonic transducers:
i
i
The distance between transducers is specified in
i
inches and always measured between the faces of transducers 1 and 2. These dimensions are not the same for each mounting mode.
19
Page 20
MOUNTING DISTANCE
Distance
V-mode and W-mode
The distance between the ultrasonic transducers is always measured between their opposing surfaces in all mounting modes. After setting up the measuring point, the ow trans­mitter displays the distances that have to be set up using a measuring tape. When using a spacer bar in the V-mode, position the transducers conveniently using the spacer bar.
APPLIES TO FIXED UFM
FIGURE 29: V-Mode or W-Mode with Spacer Bar
FIGURE 28: Mounting of ultrasonic transducers V-mode
Mounting
Distance
Ultrasonic Transducer
V
t
1
FIGURE 31: Mounting of ultrasonic transducers (Z-mode)
t
2
FIGURE 30: V-Mode Mount with Fabric-Reinforced Tensioning Tapes for Large Pipe Diameters
Installation at Z-Mode
Transducer 1
Transducer 2
FIGURE 32: Mounting of ultrasonic transducers in Z-mode
using mounting belts
20
Page 21
APPLIES TO FIXED UFM
Grease coupling
U
ltrasonic Transducer
Cable
Sheath
FIGURE 33: Z-mode installation with fabric-reinforced tensioning tapes for large pipe diameters
ULTRASONIC TRANSDUCER MOUNTING
The ultrasonic transducer (F10 and F21) is made of plastic (PEEK) that has a beige color and is protected by means of a metal sheath. The sheath has two threaded bars on top that prepares the transducer for use of spacer bar.
The ultrasonic transducer type F05 consists only of the sensor sheath and the ultrasonic transducer.
FIGURE 34: View of the back side of the transducer.
Some coupling grease (approximately the size of a
peanut) has to be applied where the part touches the pipe wall after installation.
FIGURE 35: Lower side of ultrasonic transducer (touching pipe wall)
before mounting
Mounting in V-mode or W-mode
After setting up the parameters of the measuring point,
the ow transmitter displays the distance between
the transducers in inches (face-to-face, see Figure
36) and as number of grid holes for use of the spacer
bar (ultrasonic transducers type F10 and F21). Bar index number 4, for example, shows three free grid holes between the ultrasonic transducers. Install the
transducers on the spacer bar as shown in the gure.
Secure the transducers on the spacer bar using the
knurled screws .
FIGURE 36: Using spacer bar – bar index holes
numbering, Example number of holes = 4
21
Page 22
APPLIES TO FIXED UFM
Fix Transducers on Pipe Using Mounting Belts
The transducers are mounted by using the metal ten­sion belts. The length of the belts are related to the
maximum pipe size of the corresponding transducer (for example, when using –F10 (1 1/4" to 16") size of belt
is suited for max. 16"). If using the same transducer for smaller pipes, shorten the belts. The belts are fastened by a tightener (Figure 37) which can be tightened by using a screwdriver (Figure 38).
FIGURE 37: Tightener for Metal Tension Belts
FIGURE 38: Tighten metal belts with screwdriver
Z-mode Mounting of Ultrasonic
APPLIES TO PORTABLE UFM
FIGURE 39: V-mode or W-mode without spacer bar
FIGURE 40: V-mode or W-mode with spacer bar
FIGURE 41: V-Mode Mount with Fabric-Reinforced
Tensioning Tapes for Large Pipe Diameters
22
Page 23
APPLIES TO PORTABLE UFM
Distance
Z-Mode Installation
FIGURE 42: Z-mode mounting of ultrasonic transducers
Transducer 1
Transducer 2
FIGURE 43: Z-mode mounting of ultrasonic transducers
using mounting chains
ULTRASONIC TRANSDUCER MOUNTING
Basic structure of the ultrasonic transducer:
The ultrasonic transducer (F10 and F21) is made of plastic (PEEK) that has a beige color and is protected by means of a metal sheath. The transducer is secured
by means of a knurled screw that is passed through the
transducer support. This support can be shifted in axial
direction (arrow) with the help of the knurled screw (A).
The ultrasonic transducer type F05 consists only of the sensor sheath and the actual ultrasonic transducer.
Sheath
Ultrasonic Transducer
Layout of ultrasonic transducer
FIGURE 44: Z-mode installation with fabric-reinforced
tensioning tapes for large pipe diameters
Knurled screw (A)
Transducer face
Transducer support
Before mounting the ultrasonic transducer onto the piping, the beige transducer surface must be brought into the position underneath the bottom edge of the transducer support (screw transducer into support).
Apply acoustic gel on the sensing face of the ultrasonic transducer.
23
Page 24
APPLIES TO PORTABLE UFM
MountingDistance
Mounting in V-mode or W-mode
After setting up the parameters of the measuring point,
the ow transmitter displays the distance between the
transducers in inches (Figure 45) and as number of grid holes for use of the spacer bar (ultrasonic transducers type F10 and F21). Bar index number 5, for example,
is equivalent to the number of grid holes between the
ultrasonic transducers, plus the position at which the
knurled screw of the opposing transducer has to be
mounted. Install the transducers on the spacer bar as
shown in the gure. Secure the transducers on the spacer bar using the knurled screws (B).
FIGURE 45: Positioning the ultrasonic transducers by means of spacer bar index
Transducers
The mounting of transducers in Z-Mode might be use­ful for bigger pipes and/or applications with low signals
strengths (high particle/gas load). When using Z-mode,
the spacer bar cannot be used since the two transduc­ers are located on the opposite sides of the pipe. The
distance between the transducers (“mounting distance”) depends on application (mainly pipe size). In Figure 47
the principle is illustrated. The following sections show how to place the transducers correctly.
FIGURE 47: Transducers mounted in Z-Mode
Bar index number 5 example
Fix Transducers on Pipe Using Mounting Chains
FIGURE 46: Secure the ultrasonic transducers (types F10
and F21) by means of stainless steel tensioning chain.
Fix the ultrasonic transducers using the stainless steel chains (with or without spacer bar).
Attach the chains to the hooks on the transducers while keeping them under slight tension. Attach the ultrasonic transducers to the pipe by adjusting the knurled screw
(A) until the transducer is pressed slightly onto the pipe.
Use a plastic or paper template (not supplied) to mark the mounting positions. The example shows how to mark
the positions using a plastic template.
1. Wrap the plastic template once around the pipe at the
mounting position of the rst ultrasonic transducer
(transducers face has to be in line with the line to be drawn).
2. Using a felt tip pen, draw a line on the pipe along the template (corresponds with the pipe circumference)
24
TEL: 888-722-3569 • 973-940-1684 • FAX: 1-973-940-1651 • www.GPImeters.com
Page 25
FIGURES 48 AND 49: Attaching Plastic Template
Draw Line
FIGURE 51: Measure Required Distance (Given by Flow
Transmitter)
FIGURE 52: Set Up the Template to Mark the Mounting
Position of the Second Transducer
FIGURE 50: Pipe with First Mounting Line for First Transducer
3. After setting up parameters,the UFM displays the axial distance between the ultrasonic transducers (trans­ducer distance) is displayed. Measure the transducer distance based on the value displayed on the UFM,
starting from the rst line drawn to the position at which
the second line is to be drawn (mounting position for face of the second transducer).
4. Draw two cross hairs on the same axis, centered on the lines drawn with the help of the template.
FIGURE 53: Auxiliary marks
2
1
5. Mount the rst transducer. Its face is positioned on the axis of the rst line drawn. The transducer face (not the transducer) is centered onto the rst cross hair. Now calculate half of the pipe’s outer circumference.
25
Page 26
r = Radius of pipe including wall thickness
(“outer radius”)
Example: Radius (outer) = 250mm -> U =
2*3.1415*250mm/2 = 785.4mm
6. Position the zero line of the measuring tape onto the
center of the second cross hair drawn on the pipe (at
same level as rst transducer). Measure the previously calculated distance (half circumference). You should
now have located the precise position opposite to the
rst transducer. Draw a third line at this position.
FIGURE 54: Determining the Mounting Position for Transducer
EDIT PARAMETERS
Edit system parameters using the complete setup menu.
However, this method is less convenient, as it does not
offer a step-by-step guide to setting up the parameters. It is therefore recommended to use the main menu only in
situations that require editing of an individual parameter.
Editing the parameters includes access to piping, uid
and dialogs for selecting the ultrasonic transducer and mounting mode.
If you only want to change the mounting mode from V to
Z, simply select “Transducer parameters” and change it without having to run a complete “Quick Setup” session.
This document provides only a brief overview of how to setup the parameters and main menu in the form of structure diagrams that help identify the functions grouped in the respective menu.
7. Mount the second transducer. Its face is positioned
on the axis of the second line drawn. The transducer face is centered onto the third cross hair. The trans­ducers are now mounted precisely opposite to each
other and are prepared for measuring in Z-mode.
FIGURE 55: Right Mounting of Transducers in Z-mode
Navigate to the main menu:
In the primary measuring window “Flow 1”: Select “Setup”
-> “CMPL. SETUP”
Select the parameters to edit, for example, “Pipe Setup,” “Fluid Setup,” or “Transducer Setup.”
26
Page 27
Direct access to the pipe parameters:
PIPE SETUP
Outer Diameter Outer Circumference
OR
Enter outer diameter Enter outer circumference
Enter wall thickness
Pipe material
from database
Select from database: 1 Steel 2 Stainless steel...
OR
Does pipe have internal coating?
User-dened pipe material
Enter speed of sound
of your pipe material
YES NO
Enter thickness
of coating
Choose from
database
Select: 1 Rubber 2 ...
Material of coating
OR
OR
Enter
customized
coating
Enter speed of
sound of your
coating
END
27
Page 28
Setup Fluid Data
FLUID SETUP
Choose from database Set user-dened uid
OR
Enter outer diameter Enter outer circumference
Select from database: 1 Water 20 Degrees 2 ...
END
Enter speed of sound
of user-dened uid
Enter kinematic viscosity
Enter heat capacity
Enter density
Direct access to selection of ultrasonic transducer and mounting mode:
TRANSDUCER SETUP
Choose transducer
Choose mounting mode
END
28
Page 29
ZERO SETTING
Zero Calibration Using the Main Menu:
It is advisable to run a zero calibration before starting
measurements, if possible.
CAUTION
Prerequisite for error-free zero calibration is the complete setup of the device, proper installation of both ultrasonic transducers on the pipe, and their electrical interconnection with the ow transmitter. Zero-ow is required for error free calibration. Wait
two minutes after stopping process/shut-off pipe
to allow the ow to calm down.
Zero Calibration Starting in “Flow 1” Menu
1. Close the valves of the piping.
2. Navigate to the “Setup” window as follows, using one
of three options:
3. After power-on: Select “Setup” within the start se­quence
4. In the primary measuring window “Flow 1”: Select “Setup” and “Zero Setup”
5. Select “Set Zero”
In the main menu, select “damping/cutOff/Zero” -> “Zero calibration”
Stop ow of uid (close a valve)
Choose Set Zero
Deleting the zero value
In the main menu, select “Zero Setup” - “Zero” - “Delete zero”. This action deletes the zero calibration and resets
the meter to factory settings.
CAUTION
Once the zero offset has been set, it will remain in the system until it is deleted or a new zero offset has been set. We recommend to make new zero
setting at each new measurement when possible.
The following window opens on completion of zero
calibration:
The UFM displays the calculated correction value for
the signal transit-time on completion of zero calibration.
In the course of zero calibration with closed
pipe valves, the UFM calculates the transit-time
i
differential that may develop between the trans-
ducers and any residual ow. This calculated time (including zero) is automatically included for subse­quent calculations during ow measurement. This method enhances the precision of the ow measurements. If it is not possible to close the pipe valves, delete the zero value
that may have been set previously. If anything prevents
performing a zero calibration, take the corresponding
imperfection into account in your measurements. The
zero setpoint is retained in device memory until it is overwritten with a new zero setpoint. If it was possible to close the pipe valves, check the “Flow” column in one
of the three measuring windows to determine whether
or not the owrate is going down. Do not perform a zero calibration until a settled value is output to the ow display.
A stop valve is not available at all positions of the piping. The tolerances that develop during installation, including tolerances of the ultrasonic transducers and pipe data,
will lead to a certain zero offset error in the measuring equipment. Provided care was taken during installation, the ow velocity error should stay within the range from
0.00 ft./s to 0.09 ft./s. The zero offset error is reduced in proportion with increasing pipe size.
29
Page 30
HEAT MEASUREMENT
The integrated heat measurement function enables you
to determine the heat and cooling ow in your application
using QMF-PT100 or QMP-PT100 temperature sensors.
Introduction
The temperature sensor no.1 is installed in the warmer section, while temperature sensor no. 2 is installed in the cooler section of the circuit (The QMF-PT100 are
numbered on the cable). You can position the ultrasonic
transducers at the warmer or cooler section. GPI recom­mends installing the transducers in the cooler section,
as it is unlikely that they will be operated beyond their
permissible temperature limit in these sections.
QStar UFM displays the thermal output and the accu-
mulated heat quantity.
INSTALLING THE TEMPERATURE SENSOR
Temperature sensors can be mounted on piping using a metal belt (photo) or fabric-reinforced tape. Install the temperture sensor on the pipe as follows.
FIGURE 57: Mounted Temperature Sensor
FIGURE 56: Block Diagram of Heat Measurement
Thermal Output / Heat quantity
Volume Flow /
Ultrasonic Transducers
Quantity
Consumed Thermal Output / Heat quantity
Thermal Input / Heat Quantity
Feed temperature
Temperature Sensor T1
Return temperatur
Temperature Sensor T2
e
Q Flow Rate
QStar UFM shows heat (kWh) and thermal output (kW)
Calculating Thermal Output
The measuring method used is a relative measurement. This means that the measured
i
temperature values with absolute reference may deviate from this measuring equipment (for ex­ample, compared to submersion thermometers). It is important to set up a relation between both sensors. The ideal differential temperature be­tween the temperature sensors should amount
to zero degrees prior to installation on the piping.
Zero Setup of Temperature Sensors
From main menu “Flow 1”: Press button SETUP -> COMPL SETUP -> à CALIBRATION SETUP -> Select
“QMF-PT100 T2-T1”
The cross-sectional area of the pipe’s inner diameter [A] is multiplied by the ow velocity [v] and specic thermal capacity of the medium [c], as well as the differential temperature of both sensors, [T_hot- T_cold]. The product denes thermal output [Q] in W units.
Calculating heat (quantity)
The heat quantity is derived as a function of thermal
output over time.
30
The differential temperature between both temperature sensors should amount to approxi-
i
mately zero degrees prior to installation of the
transducers on the piping. Avoid touching the
transducers in the preliminary phases. To equal
both temperature sensors put them in a glass
of water for a couple of minutes. To check the
temperature of both sensors, use the values displayed in the “Heat quantity” measuring win­dow that is described in the following section.
Page 31
Select “READ OFFSET”. QStar UFM automatically
calculates the T2 to T1 offset. On completion of this calculation, the differential temperature T1 to T2 should
amount to approximately zero degrees. Use the “Reset Offset” command to reset the calculated differential temperature to zero.
Absolute thermal output measurements (absolute measurement)
Same procedure is applicable for second temperature sensor. In this case please choose PT100 T2 in calibra­tion menu.
MEASURING WINDOWS
THE MAIN DISPLAY “FLOW 1”
The device supports the alignment of the displayed temperature with a reference thermometer for each tem­perature sensor used. This functionality can be useful, for example, for heat measurements.
Example: The resistance thermometer installed in the pipe displays 348° F. However, the tempera-
i
ture sensors of your QStar UFM displays only 343° F. QStar UFM also supports manual adjustment of the offset. In this case, specify a setpoint of 348° F. The setpoint is an absolute value and not an offset.
Proceed as follows:
In the primary measuring window “Flow 1”: Select
“SETUP” -> “CMPL SETUP” -> CALIBRATION SETUP
-> “PT100 T1”
The “Flow 1” menu is the main menu that shows the current measured values like ow and thermal output.
Opening the central measuring window “Flow 1”
After power on: Wait approximately ve seconds for the
display of the start screen. The QStar UFM automatically
opens the central measuring window “Flow 1”.
Starting in the main menu: Select “ESC” > “MEAS.”
The “Flow 1” measuring window provides concise infor­mation that is important for ow and heat.
DATA IN MAIN MENU “FLOW 1”
PARAMETER DESIGNATION
Enter the absolute setpoint for T1. Caution: The setpoint is an absolute value and not an offset in terms of the tem-
perature displayed. You can select “RESET” to delete the
setpoint for the PT100 T1. In this case the temperature sensor will indicate the actual temperature.
Displays the current owrate
Totalizer = owrate counter. Dis­plays the actual owrate value.
Displays the actual thermal output
Displays the actual heat quantity
Returns the sonic speed of the
uid
31
Page 32
DATA IN MAIN MENU “FLOW 1”
PARAMETER DESIGNATION
Species the signal quality in [%]
Outputs the actual temperature of the connected QMF-PT100 temperature sensors. Signals the measurement status
Indicates the status of the measurement
Indicates deactivated data logging
Indicates ongoing data logging
Indicates USB-connection to PC
RESET
Resets the Totalizer (owrate counter) to zero
TOT
MEASURING WINDOW “HEAT”
Starting in the primary measuring window “Flow 1: Select ‘Flow 2’ -> Heat”
The “Heat” measuring window outputs concise informa-
tion that is important for heat measurements.
Change to the diagnostics windows. Only neces-
DIAG
sary for information or if you do not receive any measurement results
Change to measuring window “Flow 2” providing
FLOW 2
further information
Change to oscilloscope menu to watch signals
OSC
MEASURING WINDOW “FLOW 2”
How to open the central measuring window “Flow 2.”
Starting in the primary measuring window “Flow 1”: Select “Flow 2.”
Flow measuring window 2 provides concise information
that is important for ow measurements (without heat
measurement). The window also provides additional
ow velocity data.
DATA IN MENU “HEAT”
PARAMETER DESIGNATION
Displays the actual thermal output.
Displays the actual heat quantity.
Displays the differential temperature between the connected PT100 T1 and PT100 T2 temperature sensors.
FLOW 2
Change to measuring window “Flow 2.”
SELECTING THE PHYSICAL UNITS
How to access the “Units Setup” menu:
In the primary measuring window “Flow 1”: Select
“SETUP” -> “CMPL SETUP” -> Navigate to “Units Setup”
in the main menu.
Select the variable to change the physical unit:
ADDITIONAL CONTENT OF MENU “FLOW 2”
FUNCTION DESIGNATION
Displays the actual ow
velocity
FLOW 1
Change to the primary measuring window” “Flow 1”.
HEAT
Change to measuring window “Heat”
32
Page 33
SELECTING THE FLOW UNIT
To access the physical unit from submenu “Flow 1”
In the primary measuring window “Flow 1”: Select
“SETUP” -> “CMPL SETUP” -> Navigate to “Units Setup” in the main menu and then select “Flow.”
Select the parameter to change the unit:
SELECTING THE PHYSICAL UNIT FOR THERMAL OUTPUT
To access the physical unit submenu “Thermal output”
In the primary measuring window “Flow 1”: Select
“SETUP” ->”CMPL SETUP”. In the main menu, select “Units Setup”-> “Power.”
Select the physical unit:
SELECTING THE PHYSICAL UNIT FOR HEAT QUANTITY
Now select the desired unit.
SELECTING THE PHYSICAL UNIT FOR THE TOTALIZER
To access the physical unit submenu “Totalizer”
In the primary measuring window “Flow 1”: Select
“SETUP” -> “CMPL SETUP” -> Navigate to “Units Setup” in the main menu and then select “Total Flow”
Select the physical unit:
To access the physical unit submenu “Heat quantity”
In the primary measuring window “Flow 1”: Select
“SETUP” -> “CMPL SETUP” — In the main menu, select “Units Setup” -> “Heat.”
Select the physical unit:
33
Page 34
APPLIES TO FIXED QSTAR UFM
SETUP THE I/O
To access the I/O setup menu:
In the primary measuring window “Flow 1”: Select “SETUP” -> “CMPL SETUP”. In the “complete setup” menu, select “I/O Setup.”
SETUP THE 4 mA TO 20 mA CURRENT OUTPUTS
CAUTION
Any connection of an active 4 mA to 20 mA analog output of QStar UFM with an external device that also provides a voltage at its inputs will cause fatal damage to your QStar UFM and external device. Before you
interconnect both devices, always verify that
your external recording (PCS) system is set to passive state.
The outputs of QStar UFM are working in active mode in factory state. That means that the ow transmitter provides the required voltage to run the outputs.
Output passive
Passive 1
QStar UFM 24VDC
Passive 2
QStar UFM 24VDC
The mode (passive or active) of the outputs can be changed inside the QStar UFM. Always use the factory setting (active mode). If you need to change the mode, contact your GPI Customer Service Representative.
To access the “Analog output” menu:
In the primary measuring window “Flow 1”: Select
“SETUP” -> “CMPL SETUP”. In the main menu, select
-> “I/O Setup” -> Analog outputs”.
The inputs of the external device are connected directly to QStar UFM.
Outputs are activated (factory setting)
Active 1:
QStar UFM 24VDC
34
Select the analog output to be used:
Page 35
APPLIES TO FIXED QSTAR UFM (Continued)
Specify the variable to be output at the analog output:
Select the value to be assigned to the 4 mA output (Example Flow).
SETUP RELAY PARAMETERS
QStar UFM is equipped with a relay output.
This output can be assigned a function or a range.
Example of an external circuitry:
CAUTION
The relay has a NO (normally open) contact.
The contact is only closed when triggered by a setup parameter function. The contact will
retain its open state on power failure. You do
not have to observe the polarity of the wiring since it is potential free.
Select the value to be assigned to the 20mA output.
The analog outputs only support unidirectional ow.
QStar UFM can only provide positive ows to the outputs. The ow display must return positive
i
values. What happens if the value exceeds the
setup high or low limits? Example: You expect a owrate of 100m3/h and assigned this rate a value of 20 mA. However, the owrate may
reach 130 m³/h. This means that a value of 20
mA will also be output for owrates higher than
100 m3/h.
To access the “Relay” menu:
In the primary measuring window “Flow 1”: Select
“SETUP” -> “CMPL SETUP”. In the main menu, select
-> “I/O Setup” -> “Relay:
Choose value (lower limit) for activation of relay. Relay will be activated when measured value fall below the limit.
35
Page 36
APPLIES TO FIXED QSTAR UFM (Continued)
In addition to setup variables such as ow,
QStar UFM applies the corresponding unit
i
that is selected in the “Units Setup” setting
and appended to the respective variable that is displayed in a measuring window. Example:
If you selected the physical unit m³ for ow
variables, the values of the switching points are also the parameters set in cubic meter.
Select the upper limit value for the relay. Relay will be activated when measured value exceeds the maximum limit.
Example: A pump is operated in an applica­tion and the discharge volume of this pump
i
is measured. The pump has a maximum discharge volume of 6000 m3 /h and there is a risk of damage to the pump when the dis­charge volume drops to less than 150 m3/h. The objective is for QStar UFM to shut down the pump as soon as the value drops below a point at which damage to the pump can be expected. So in that case lower limit has to be set to 150m3/h, upper limit to 6000m3/h. Relay
will be activated if ow falls below 150m3/h or
exceeds 6000m3/h.
Connecting Digital Output
Example 1:
Connect the plus terminal of evaluation unit (for ex­ample, PCS) to the plus terminal of voltage supply. Connect the negative terminal of evaluation unit to the plus terminal of the DO. The negative terminal of the DO has to be connected to the negative terminal of external voltage supply.
QStar UFM
Example 2:
Connect a PNP transistor’s emitter to the positive
terminal of the external voltage supply. The basis of the transistor should be connected (via a resistance) to the positive terminal of the DO. The collector of the transistor should be connected to the positive terminal of evaluating unit (PCS). The negative terminal of the evaluating unit has to be connected to the negative terminals of the DO and the voltage supply.
QStar UFM
The physical unit cubic meter must have been
assigned to the ow variables in the physical
unit settings. The minimum activation point must have the parameters setup.
Setup the Digital Output (transistor)
QStar UFM provides one digital output (DO). The DO is a transistor output (open collector) and is typically
used as a counter for volume (m3) or heat (kWh). The
DO is operated in passive mode. That means a supply voltage (10…30VDC) has to be provided externally.
CAUTION
Be careful not to confuse the positive (+) and negative (-) voltage in order to avoid damage
to the ow transmitter.
36
To enter the menu for Digital outputs:
Starting from main menu “Flow 1”: Choose SETUP
-> COMPL SETUP -> I/O SETUP -> Digital Output.
Page 37
APPLIES TO FIXED QSTAR UFM (Continued)
Choose related data (heat or volume) to use with DO.
Choose the value (totalizer or heat) that represents
one pulse.
Example: To get a pulse at the DO each 10m3 enter 10 here. Units can be changed in units setup.
UPGRADE QSTAR UFM
Disconnect transmitter from power supply. Plug the RS232 board (QMF-RS232) into the two plugs in the region of the connection terminals (see QStar UFM and
Components: Interfaces of QStar UFM” to access
the connection terminals).
RS232 Board (QMF-RS232)
Now reconnect the power supply.
The unit in the digital output menu (for example, m3) is related to the chosen unit of the counter
i
(totalizer). Example: Choosing unit l (litres) for the totalizer in the units setup menu also
provides litres in the DO.
Choosing kWh as unit of the totalizer when
using it for heat measurement also provides
kWh in the DO.
SETUP THE SERIAL INTERFACE RS232
Using a separate RS232 interface board is an option. QStar UFM upgrades by simply plugging in the separate RS232 board. The RS232 sends ASCI strings including measurement data.
CONNECT RS232 BOARD TO A RECEIVER (PC OR PCS)
Requires a cable to be connected at the RS232 board
and –typically- to a SubD (others possible due to cus­tomers demands).
RS232 Interface board
TAG Designation
TX Transmit Data 3
Pin 9 Pol. Sub
D
RX Receive Data 2
GND Ground 5
37
Page 38
APPLIES TO FIXED QSTAR UFM (Continued)
Pin assignment of standard Sub-D 9 pols
ACTIVATE THE RS232 INTERFACE
To activate the RS232 go to menu “Flow 1”
From main menu “Flow 1”: Choose SETUP -> COMPL
SETUP -> SERIAL INTERFACE (6)
To activate RS232 interface press YES
CONFIGURE RECEIVER (for example, PC) TO RECEIVE
RS232 DATA
Choose interface where the RS232 is connected (for example, COM4).
Congure input according to Figure 58. Note: Bit rate cannot be changed (115200 bits/sec).
FIGURE 58: Congure Input for RS232 Communication
Then choose some parameters:
Start Date: Set date where RS232 communication should start.
Start Time: Set time where RS232 communication should start.
Duration: Set duration of RS232 communication.
Interval: Set interval (for example, 5 means that a data
package is sent each ve seconds).
After successful conguration you should receive data
as shown in Figure 59.
FIGURE 59: Receiving Data via RS232 Interface
To receive data use the Hyper Terminal provided
by Windows. If serial interface is not available
i
on the PC, use standard RS232 (Sub D-9) to USB converter.
If you do not set a start time and date the communication starts immediately after setting
i
duration and interval.
38
Page 39
APPLIES TO PORTABLE QSTAR UFM
SAVE, LOAD AND MANAGE DATA
Logging Data
The term data logging denotes the recording (saving) of measured value data on the internal SD Memory
Card. All measurement data like time and date, ow, velocity, totalizer and thermal output, heat quantity and
temperatures (when using temperature sensors). If temperature sensors should not be in use these values
are shown as “0” in the log les. The data is stored in a text le (*.txt) which allows easy and quick export into ofce software like Microsoft® Excel or similar.
Time-Controlled Data Logging
Your UFM supports time controlled data logging to the
internal SD memory card.
The time controlled data logging uses the in-
ternal system time which is set by user. Make
i
sure the system time is correct.
In the primary measuring window “Flow 1”: Select
“SETUP” -> “CMPL SETUP” -> “Data logger”
Navigate to the Data logger parameters eld and con­rm your entry with “NEXT”. Assign a le name (min. 4 characters). Use the arrow keys to select the letters, or delete a letter by means of “DEL”. Accept the letter with “ENTER”. Select “DONE” and press “ENTER” to
conclude data input
• Enter the start date May 25, 20YY
• Enter the start time 3:00 PM
• Enter the duration in 001:01:00:00 format
• Enter the interval in 00:00:60 format
The data logger does not average values
over the cycling period. The current values
i
are always stored. If you choose interval of 10
seconds the current values each 10 seconds
are stored (not the average over last 10 sec).
To delete or to rename existing log les please con­nect your UFM to your PC using the USB cable. Then
access the log les directly at the SD card.
Message “LOG ACTIVE” is being shown at the display
while data logging is going on. If there is no data log-
ging display shows “LOG INACTIVE.”
Continuous Data Logging:
Example: You want to log the data generated from May 25, 20YY, 3:00 PM to May 26, 20YY, 4:00 PM.
The data is logged at cyclic intervals of 60 seconds to the SD Card.
Start 15:00 Stop 16:00
25.05.20YY 26.05.20YY
27.05.20YY
39
Page 40
APPLIES TO PORTABLE QSTAR UFM (Continued)
Cancel active data logging
If you want to quit an ongoing data logging before the dened stop time please go to main window “Flow”: Choose “SETUP” -> COMPL SETUP -> DATA LOGGER
The following window appears:
To cancel data logging press YES. The stored data will
be still available on the SD card.
Save/load/edit parameters
The primary measuring window “Flow 1”: Select “SETUP” -> “CMPL SETUP” -> “Save/load parameters”.
Your UFM provides functions for saving and loading the parameter data. This saves time if you are taking
measurements at different locations.
To save the parameters, use the arrow keys to select the memory space and conrm your entry with “NEXT“:
Select “SAVE” to save the parameter data.
Select “LOAD” to load a parameter set from memory. Select “RENAME” to rename the stored le. Select Delete to delete the choosen parameter le.
Use the button to scroll through the details of the
stored parameter le.
The parameter les are stored into an internal
data logger and not in the SD card. The SD
i
card is reserved for data log les only. If you perform software reset stored parameter les
will be deleted.
READING DATA ON THE COMPUTER
Connect via USB port to a computer using the USB cable
supplied by GPI. Your UFM is automatically detected as mass storage medium (like a USB drive). Compat-
ible operating systems include: XP, WIN2000, Vista,
Windows 7, or MAC OS X.x. Your UFM is equipped
with a USB 2.0 interface.
When succesfully connected to a PC display will show
message USB ON in the main menu “Flow 1”.
Your current data setup is displayed:
Enter a le name. Use the arrow keys to select the let­ters, or delete a letter by means of “DEL”. Accept the letter with “ENTER”. Select “DONE” and press “ENTER”
to exit data input.
40
FIGURE 60: Display when connected to a PC
QStar`s SD card will be detected from Windows/Linux and can be accessed in order to rename/delete or
copy les.
Page 41
APPLIES TO PORTABLE QSTAR UFM (Continued)
Exporting data in Microsoft® Excel
Data logs are saved to a le with *.txt extension on the
SD Card. The data can be imported directly to a standard data processing software such as Microsoft® Excel.
Copy the desired le to your PC. This guide shows basics of exporting the les into Microsoft course working with the data logs is not restricted to
Excel and is possible with comparable software.
Free ofce software similar to Microsoft® Ofce can be found at www.openofce.org.
i
The log le will look like this: Each line represents the
data from one measuring point. The different values are separated by semicolon.
FIGURE 61: Log le opened by text editor
®
Excel. Of
SETUP THE 4 mA TO 20 mA CURRENT OUTPUTS
QStar UFM provides two 4 mA to 20 mA current outputs. These outputs can be assigned different measured values. The outputs are set actively by default. This means that your QStar UFM always provides a voltage at these outputs.
CAUTION
Any connection of an active 4 mA to 20 mA analog output of QStar UFM with an external device that also provides a voltage at its inputs will cause fatal damage to your QStar UFM and external
device. Before you interconnect both devices,
always verify that your external recording (PCS) system is set to passive state.
FIGURE 62: Data imported into Microsoft® Excel
SETTING THE PARAMETERS
How to access the I/O setup menu
In the primary measuring window “Flow 1”: Select “SETUP” -> “CMPL SETUP.” In the “complete setup” menu, select “I/O Setup.”
The outputs of your UFM are working in active mode in factory state. That means that the ow transmitter provides the required voltage to run the outputs.
The inputs of the external device are connected directly to your UFM.
Outputs are activated (factory setting).
Activ 1:
QStar UFM 24VDC
41
Page 42
APPLIES TO PORTABLE QSTAR UFM (Continued)
Current output 1 (I1A/B): Current ow is directed
from I1B to I1A.
Current output 2 (I2A/B): Direction of the current ow
is directed from I2B to I2A
Select the value to be assigned to the 20mA output. The analog outputs at your UFM only support unidi-
rectional ow.
How to access the “Analog output” menu:
In the primary measuring window “Flow 1”: Select “SETUP” -> “CMPL SETUP”. In the main menu, select
-> “I/O Setup” -> Analog outputs.
Select the analog output to be used:
Specify the variable to be output at the analog output:
Select the value to be assigned to the 4mA output (Example Flow)
QStar UFM can only provide positive ows to the outputs. The ow display must return posi-
i
tive values. What happens if the value exceeds
the setup high or low limits? Example: You expect a owrate of 100m3/h and assigned this rate a value of 20 mA. However, the owrate
may reach 130 m³. This means that a value of
20mA will also be output for owrates higher
than 100 m3/s.
When do you have an error situation?
i
If limits of the values previously assigned
to the 4mA and 20mA outputs were ex-
ceeded on expiration of a specic time
(burnout)
Signal loss
Example: You expect a maximum owrate of
100 m3 in your application and assigned this
rate the 20mA value. You also set the following
defaults for the error mode.
If an error is pending, the 4mA to 20mA output
enters the error mode and outputs 4mA on expi­ration of a delay time of 20 seconds. In practical
operation you may have a owrate of 130 m³
in the piping for a duration up to two minutes. In this case, QStar UFM would output 20mA for the duration of 20 seconds and then enter the error mode and output 4mA on expiration
of a timeout of 20 seconds. After the owrate
has dropped again to a value less than 100 m³ (on expiration of 100 seconds),the QStar UFM will automatically exit the error mode and
output the setup current for the ow.
42
Page 43
APPLIES TO PORTABLE QSTAR UFM (Continued)
SETUP THE RELAY
Your QStar UFM is equipped with a relay output.
This output can be assigned a function or a range.
Example of an external circuitry:
You link an alarm beacon with the output to signal that the owrate has undershot the lower threshold limit.
Color coding of the relay output cable:
In addition to setup variables such as ow,
QStar UFM applies the corresponding unit
i
that is selected in the “Units Setup” setting
and appended to the respective variable that
is displayed in a measuring window. Example:
If you selected the physical unit m³ for ow
variables, the values of the switching points
are also setup in cubic meter.
Select the upper limit value for the relay. Relay will be activated when measured value exceeds the maximum limit.
CAUTION
The relay has a NO (normally open) contact.
The contact is only closed when triggered by a setup function. The contact will retain its open state on power failure and with low rechargeable
battery. You do not have to observe the polarity
of the wiring since it is potential free.
To access the “Analog output” menu:
In the primary measuring window “Flow 1”: Select
“SETUP” -> “CMPL SETUP”. In the main menu, select
-> “I/O Setup” -> “Relay:
Choose value (lower limit) for activation of relay. Relay will be activated when measured value fall below the limit.
Example: A pump is operated in an applica-
tion and the discharge volume of this pump
i
is measured. The pump has a maximum
discharge volume of 6000 m3 /h and there is
a risk of damage to the pump when the dis-
charge volume drops to less than 150 m3/h.
The objective is for QStar UFM to shut down
the pump as soon as the value drops below
a point at which damage to the pump can be
expected. So in that case, lower limit has to be
set to 150m3/h, upper limit to 6000m3/h. Relay
will be activated if ow falls below 150m3/h or
exceeds 6000m3/h.
The physical unit cubic meter must have been
assigned to the ow variables in the physical
unit settings. The minimum activation point
must have been setup.
43
Page 44
CALIBRATION
SYSTEM SETTINGS
QStar UFM provides a calibrating function that can be
used to calibrate the ow. It may also be used to calibrate
the analog outputs and temperature sensors.
To access the “Calibration” menu:
In the primary measuring window “Flow 1”: Select
“SETUP” -> “CMPL SETUP” -> Navigate to “Calibration Setup” in the main menu”.
CALIBRATING FLOW
Select “Flow” from the calibration menu”.
To access the “System settings” menu:
In the primary measuring window “Flow 1”: Select
“SETUP” -> “CMPL SETUP” -> Navigate to “System Setup” in the main menu
Editing the Time and Date
To access the “Time and date” menu:
In the primary measuring window “Flow 1”: Select
“SETUP” -> “CMPL SETUP”. In the main menu, select “System Setup “ -> “Time and date
Enter an offset percentage that is relative to calibration reference.
To calibrate QStar UFM, we recommend you
take measurements at different ow velocities
i
then calculate the mean value of the results. GPI
calculates the mean value based on ve differ­ent ow velocities. The ow offset is retained in
device memory until it is overwritten with a new value word.
CALIBRATING THE TEMPERATURE SENSOR
For more information on temperature sensor calibration, refer to the section Measuring with QStar UFM: Heat measurement.
Enter the time with the format hh(hour):mm(minutes):ss (seconds). Enter the date with the format: mm(month). dd(day).yy(year).
Modifying the Display Backlight
To access the “Backlight” menu:
In the primary measuring window “Flow 1”: Select
“SETUP” -> “CMPL SETUP” — In the main menu, select “System Setup” -> “Backlight”
Adjust the display brightness by setting value (0% = no backlight, 100% maximum brightness).
44
Page 45
Change Language
MISCELLANEOUS
1. Switch on the UFM — Within the start sequence, press the multifunctional key next to the “SETUP” eld.
2. Conrm the “SETUP LANG.” button.
To access the “Miscellaneous” menu:
In the primary measuring window “Flow 1”: Select
“SETUP” -> “CMPL SETUP” — In the main menu, navi­gate to “Damping/CutOff/Zero” menu.
Flow Damping
In the primary measuring window “Flow 1”: Select
“SETUP” -> “CMPL SETUP”. In the main menu, navigate to “Damping/CutOff/Zero” -> “Damping Flow”
Specify an attenuation of signal output in this dialog. Enter a damping time. It is a T63 damping. That means after damping time the displayed value has reached 63% of real change.
3. Use the arrows in the next window to select the dialog
language. Conrm entry with “Enter” and exit the menu with “SETUP”.
CAUTION
The language setting changes the language used
in the menus. The language in the elds next to
the multifunctional button remain unchanged.
Example: Damping time 5 seconds, Flow change from 1 m3/h to 2 m3/h.
Display: Shows 1.63m3/h after 5 seconds, 2.00 m3/h after another 5 seconds (10 seconds in total).
Typical values are 5-30 seconds. The higher the damp-
ing the slower the measurement but the “smoother” the
measurement curve.
Flow Cut off
In the primary measuring window “Flow 1”: Select
“SETUP” -> “CMPL SETUP”. In the main menu, navigate to “System Setup” -> “Damping/Cut off/ Zero” -> “Cut off Flow”.
Only the ow velocities that are greater than the setup “Cut off” will be displayed. Lower velocities will be dis-
played as 0.
45
Page 46
Zero Setting
Impulse OUT
nUpRelays/ T1/T2 Analog Out
QMP-001
Software Reset
In the primary measuring window “Flow 1”: Select
“SETUP” -> “CMPL SETUP”. In the main menu, navigate to “System Setup” -> “ Damping/CutOff/Zero “ -> “Zero”.
System Setup-> “System information”
In the primary measuring window “Flow 1”: Select
“SETUP” -> “CMPL SETUP” — In the main menu, navigate to “System Setup” -> “Info Flow Module”. Check version
of software to see if an update is available.
RESETTING QSTAR UFM
Two different types of resets:
Hardware Reset
Use software reset in all other cases like getting an
implausible measurement or reading on the display. Data stored on SD card will not be deleted. Data stored
internally (parameter les) will be deleted.
There are different possibilities to make a software reset
Immediately after switching on, press “RESET SYS” at the starting screen.
From the main menu “Flow 1” choose SETUP
-> COMPL SETUP -> SYSTEM -> RESET SYS.
CAUTION
Reset the date and time after each software reset.
Software Reset
Hardware Reset
Hardware reset is only necessary when QStar UFM
hangs up and cannot be operated any longer by using the buttons. Data stored internally or on the SD card will not be deleted.
Fixed QStar UFM: The hardware reset can be activated by pressing the reset button located behind the cover.
Portable QStar UFM: The hardware can be reset by putting a small screwdriver or a paper clip into the hole
on the backside of the transmitter.
ResetDow
USBPower
TROUBLESHOOTING
Use the following check lists for troubleshooting and check all items listed. If an error prevents you from taking proper measurements after you completed the check, contact
GPI Customer Service Representative at: (888) 996-3837.
Make sure to have the following information available:
Pipe material
Pipe outer diameter
Wall thickness
Medium type and temperature
Type/length of the intake/discharge circuits
Type of transducer used
Use the “QStar UFM troubleshooting” information
for troubleshooting by means of step-by-step
i
elimination of items. This procedure helps when troubleshooting complex issues and in locating the fault systematically.
46
Page 47
TROUBLESHOOTING
A: NO MEASURE AT ALL
1. Are you absolutely certain that the piping is
completely lled?
2. Can you exclude gas load on the medium? Is a pump close by?
If there is gas load:
• Are the transducers installed at the 10 AM
or 2 PM position on the horizontal pipe?
• If there is gas load on the medium, is there
an option of mounting the transducers to a vertical pipe? If yes, mount the transducers to a vertical pipe.
3. What is the solid particle content of the uid?
4. Are you absolutely certain of the wall thick-
ness? Check the piping for the presence
of engraved information that may be of
relevance. Hot water piping is subject to the
formation of lime deposits that may propagate
the deposit of solid particles of uid with high content. Use a wall thickness meter. You can purchase a wall thickness meter from GPI. Check the wall thickness entry on the device.
5. Did you enter the outer diameter of the pipe, or its circumference? Check the correspond- ing values once again.
6. Have you selected the correct piping mate­rial? Are you certain the piping is not lined?
When taking measurements on a pipe made
of concrete you must presume that this piping does not have sound conducting properties.
Check the setup values once again.
7. Does the ultrasonic transducer match the application with regard to pipe dimensions and medium temperature? Is the temperature
of the uid within permitted limits? Did you
select a suitable ultrasonic transducers for the application with regard to the pipe dimen­sions? Once again, verify that you selected the correct ultrasonic transducer and that it is setup.
8. Which transducer mounting mode did you se­lect? Standard is the V-mode. If you selected the W-mode, you should rst try the V-mode.
9. Verify the proper transducer distance. Take the measurements at the transducer faces.
When using a spacer bar, check the number
of holes between the transducer mounting positions.
TROUBLESHOOTING
A: NO MEASURE AT ALL
10. Does the piping have a thick paint coating? If yes, try to remove the paint at the position where you are planning to install the ultrasonic
transducers using sand paper. Never take any
measurements on piping insulation, no matter what type of material is involved.
11. Did you apply an appropriate lm of acoustic coupling gel (Magnalube) to the transducer faces? Small transducers (types QMP-F10 and F21) need approximately 1.2 inches, whereas the large transducer (QMP-F05) requires ap­proximately 2.4 inches of coupling gel.
12. Do the transducers apply adequate pressure on the piping?
13. Are the transducers that are mounted without spacer bar precisely aligned along a common axis on the piping?
14. Are there sources of strong disturbance such as transformers,electrical drives, or sources of vibration in the immediate area of measure­ment?
15. Use the Z-mode for installation if all of your
efforts did not yield a satisfactory result. Setup
your device accordingly. You might also try to test your equipment on a different piping sec­tion that is close by in order to nd out whether or not you can perform a measurement. Check
all measuring leads for damage.
B: IMPRECISE MEASUREMENT RESULTS
1. Did you observe the corresponding upstream
and downstream distances? The quality of
measurements will deteriorate in propor­tion to shorter upstream and downstream distances.
2 . Did you perform a zero calibration after
having closed a stop valve on piping?
3 . Check the setup values with regard to the:
• Pipe outer diameter
• Wall thickness
• Piping material
• Transducer distance
47
Page 48
Measures to take if the piping is not lled
Amplitude
Amplitude
Amplitude
Amplitude
completely.
siphon
Sensor 1Sensor 2
If you cannot separate the piping, because it is plastic piping, for example, or if the application is in the planning or installation phase, it is recommended to use a siphon to
compensate for partially lled piping. The gradient of the siphon is calculated based on the expected ow velocity
and contamination load. Contact GPI for support if it is necessary to install a siphon for your application.
The time the signals needs from one transducer to the corresponding transducer (run time) depends on path
length (distance), speed of sound of materials, uid and ow velocity.
Depending on mounting mode the signals might look as
shown in Figure 64.
FIGURE 64: Signal Images
t1w
t1v
Pipe Signal
Pipe Signal
t1v
Signal
Signal
Mode-
Mode-
-
-
V
V
tp
tp
signal
signal
Send
Send
t1w
Signal
Signal
Mode-
Mode-
-
-
W
W
time
time
DIAGNOSTIC MENU
Oscilloscope/Auto-Window
Signal propagation
QStar UFM offers sophisticated diagnostic menus allow-
ing users to optimize challenging measurements in harsh
conditions. To use the diagnostic menu for troubleshoot-
ing, a background in signal propagation is provided.
Figure 63 illustrates which signals occur and how they progagate at the pipe.
FIGURE 63: Signal propagation
UP-Transd. V-Mode W-Mode
UP-Transd. V-Mode W-Mode
UP-Transd. V-Mode W-Mode
Signal
Desired
Desired Desired
-
Mode
Mode
Mode
-
-
-
Z
Z
Z
Signal
Signal
Signal
-
-
V
-
Mode
Signal
Signal
Signal
Z-Mode
Z-Mode
Z-Mode
Pipe Signal
Pipe Signal
Pipe Signal
W
-
W
W
Mode
-
-
Mode
Mode
Signal
Signal
Signal
-
-
V
V
-
-
Mode
Mode
-
At t=0 the UP-transducer starts signal transmission. After time T=tp you can see the pipe wall signal which reaches the corresponding transducer rst because of
shorter distance and (usually) faster propagation caused by higher speed of sound of pipe material.
After that, at time T=t1v, the V-mode signal reaches the transducer (single reection at opposite pipe wall). After
t1w (approximately 2x t1p) the W-mode signal is being detected. Usually, when having pipe materials highly conductible for ultrasonic signals (metals) and small pipes, the pipe wall signal has high amplitude which is similar to amplitude of V-mode signal. When having sound-absorbing materials (plastic, concrete) and/or
coating the pipe wall signal is usually weak indicated by
low amplitude. The V-mode signal is usually stronger than the W-mode signal.
The time between the different signals might be signi­cantly higher or lower depending on pipe sizes and speed
of sound of involved materials.
A different situation occurs when measuring in Z-mode.
FIGURE 65: Signals in Z-mode
The UP-transducer mounted at upstream position emits the send (or desired) signal. This signal propagates within
the pipe wall (“pipe signal”) as well as in the uid. The signal propagating in the uid is reected a couple of times at the pipe walls (Z-Mode signal, V-mode signal,
W-mode signal). The signal which represents the desired signal depends on mounting mode. For example, when mounted in V-mode the V-mode signal is the desired
signal. The other signals (Z-mode, W-mode, pipe wall)
may also disturb this signal. The corresponding transducer
is mounted in V-mode as standard but Z-mode and W-
modes are possible. This transducer receives the signals and returns another signal to UP transducer.
48
t1z
t1z
tp
tp
Send Signal
Send Signal
time
time
Beside the send signal, only the desired signal (Z-signal) is visible. The Z-signal reaches the receiving transducer after T=t1z.
Page 49
Oscilloscope Menu
Amplitude
Amplitude
With the oscilloscope menu you can analyze signals and
even manipulate them in order to handle challenging applications.
QStar UFM sets all necessary parameters for the signal processing. When manipulating signals
i
these settings will change. This might result in a loss of signals and/or failure in measurement. The changes done in the oscilloscope menu might remain valid even when leaving the menu. To get settings from QStar UFM the site should be setup again.
QStar UFM only works with the signals within the measur­ing window. The measuring window represents a certain
period of time which is analyzed by QStar UFM about signals. The measuring window can be moved back and
forth by using the arrow buttons.
FIGURE 66: Measuring Window
t1w
t1v
tp
tp
signal
signal
Send
Send
t1v
Signal
Signal
-
signal
signal
Pipe
Pipe
-
Mode
Mode
-
-
V
V
t1w
Signal
Signal
Mode-
Mode-
-
-
W
W
Time
Time
window displays (Figure 50). In this picture, window starts at 154 µSec. Use arrow buttons to change this starting time.
Signal Analysis Using Oscilloscope
The oscilloscope allows quick check of signal quality.
This helps to get accurate measurements. Good signal
to noise ratio and “sharpness” of signals are important
for best results.
Signal-to-noise ratio (SN)
SN indicates the difference of amplitudes between the desired signal and the noise. The higher the SN the bet­ter the signal processing. Figure 68 shows good signals with very good SN.
FIGURE 68: Oscilloscope Menu Showing Good Signals
Delay
Delay
Window
Window
Window
Zoom +
Zoom +
Zoom +
Zoom -Zoom -
Zoom -Zoom -
Zoom -Zoom -
There is almost no noise at all and a clear and distin-
guished signal. QStar UFM tries to set the measuring window so that only the desired signal (for example, V-mode) is visible.
FIGURE 67: Oscilloscope Menu Showing Desired Signals
Figure 69 shows a noisy signal with lower SN. Screen
shows noise around the signal which reaches 20-40%
of signal.
FIGURE 69: Oscilloscope Menu Showing Noisy Signal
Use ZOOM-/ZOOM+ buttons, to scale-up/scale-down
the measuring window (decrease/increase the time the window is open). Use arrow buttons to move window left or right (opens earlier/later).
QStar UFM only uses signals within the window. If you move the window, the signal will disappear and you might get wrong/no measurement.
QStar UFM is able to handle noisy signals thanks to its sophisticated signal processing. However,
i
if a high SN is encountered, the user should use caution in case of additional interferences.
Information about starting point (“delay”) of measuring
49
Page 50
If there are problems with your measurement SN can be improved by:
Figure 71 shows diffused signals. There is basically no phase shift. This might result in undetected signals.
Surface of pipe: Remove paint or rust
Coupling grease: Use sufcient coupling grease
(Magnalube).
Alignment of transducers: Make sure the trans­ducers are mounted as suggested by QStar UFM.
Make sure that center of transducers touches the pipe sufciently (might be a problem with very small pipes). You can check the oscilloscope
while mounting the transducers.
Make sure the transducers are not mounted on
welding seams.
Make sure to remove air from liquid.
Choose another mounting position (for example,
vertical pipe).
Make sure to provide sufcient straight run.
Make sure that signal cables are not disturbed
by power cables.
Pumps and motors (with frequency inverters)
generate electric noise and should be avoided.
Use Z-mode rather than V-mode to reduce path
length and therefore to increase signal strength.
Try V-mode rather than W-mode.
Use stronger transducers, for example, –F05
instead of –F10 or –F10 instead of –F21.
Sharpness of Signals
QStar UFM uses coded signals to make sure to detect signals in case of noise. Coded means QStar UFM makes phase shifts within a signal package.
Figure 70 shows a sharp signal. You can see approxi- mately 5 waves followed by a phase shift and other waves.
FIGURE 71: Diffuse Signals
QStar UFM is able to handle diffused signals until a certain limit. Nevertheless sharp signals
i
should be the goal.
If there are problems, the sharpness can be increased by:
• Choosing different signal sequence
Using other mounting mode
Using different transducer
Improving pipe surface: Removing paint and/
or rust
• Making sure to use sufcient amount of coupling
grease (Magnalube)
• Aligning of transducers: Make sure that transduc­ers are mounted as suggested and that the center
of transducers touches the pipe sufciently (might be problem at very small pipes). You can check
the oscilloscope while mounting the transducers.
• Making sure that transducers are not blocked by
welding seams
Choosing a different location for measurement
FIGURE 70: Sharpness of signals
50
To select a different signal sequence, enter the diagnostic
menu (Diag) and press button on the right of parameter
SendCode. Usually the signal sequences 3-Barker 5 or 4-Barker 7 are expected to bring best performance. You can try different sequences and check inuence in diag-
Page 51
nostic data (SigQ, see Troubleshooting: Oscilloscope
Amplitude
Amplitude
Amplitude
Menu of QStar UFM) as well as visually in oscilloscope.
Separating Signals (small pipes)
When measuring at small pipes (<1.9 in.) the distances between the received signals become smaller. In the worst case, signals might interfere as shown in Figure 72 Interfering signals.
FIGURE 72: Interfering Signals
FIGURE 73: Separated Signals
Signal
Signal
Mode-
Mode-
-
t1v
Pipe Signal
Pipe Signal
-
V
V
t1w
Signal
Signal
-
-
Mode
Mode
-
-
W
W
Send Signal
Send Signal
tp
time
time
tp
signal
Send
t1v
Zoom -
signal
Pipe
-
Mode
-
V
Window
Zoom +
Signal
-
t1w
Mode
-
W
Zoom
Signal
-
Time
-
---
In this case, there are not only the desired signal within the measuring window but also the interfering pipe wall signal and also a part of the W-mode signal.
QStar UFM uses sophisticated algorithms to separate desired signals from interfering signals.
i
These algorithms are mainly applied when having the auto window function activated. But especially when measuring at very small pipes interference might occur anyway. In that case the user can
separate signals (in order to lter out the desired
signal) manually in the oscilloscope mode.
Window
Window
Window
Zoom +
Zoom +
Zoom +
Zoom -Zoom -
Zoom -Zoom -
Zoom -Zoom -
Deactivate the auto-window in order to separate signals
manually. Always check the measured speed of sound
of the separated signal. When the (separated) desired signal is within the measuring window, the speed of sound (parameter vS in diagnostic menu) should be within +/- 20% of the expected speed of sound (for example, when measuring water at 68° F speed of sound should be within 1400-1500 ft/s). If vS exceeds the expected
values you most likely have separated the wrong signal
(example, pipe wall signal, then vS is showing speed of sound of pipe material).
Activate/Deactivate Auto-Window
When in main menu “Flow 1”, choose OSZ to enter the
oscilloscope menu.
Determine whether to use auto window or not. In case of problems with measurement we recommend to try
without auto window rst.
Choose “NO”
If interference does occur:
• Use W-Mode rather than V-mode (rst counter-
measure)
• Use Z-Mode rather than V-mode (when W-mode does not work)
• Deactivate Auto-Window and use ZOOM function/ shift function (arrow buttons) to kick out disturbing
signals of the measuring window.
• Use different signal sequence, for example, 3-Barker5 or 1Puls rather than 4-Barker 7
• Use transducer working at higher frequency
(F21 rather than F10) to get sharper signals.
When using F21 (2MHz) instead of F10 (1MHz)
the signals become half as wide (half of time)
which makes it easier to separate signals.
Display shows distinguished signals (UP and DOWN) in the oscilloscope.
Press DIAG and OSZ again to activate auto window. Answer with YES. Check if position of ultrasonic signals have been changed. Check if value for delay has been
changed. If yes, use auto window for your measurement.
Auto window is activated when you say YES and when
you leave oscilloscope mode. To deactivate auto window enter oscilloscope mode again and answer with NO).
51
Page 52
DIAGNOSTIC MENU
CAUTION
The diagnostic menu shows a lot of parameters that are mainly readable for experienced users. The diagnostic data are also suited for troubleshooting.
Diagnostic Window 1:
MENU 1: Diagnostic Data
Frequency: Shows the signal frequency which depends
on the selected transducer.
There are three different transducers using different
frequencies (500kHz, 1000kHz, 2000kHz).
Signal: Shows the type of signal code. Coded signals allow reliable detection of signals.
T1: Signal run time signal 1 (Upstream signal or down­stream signal depending on connection of cables).
T2: Signal run time of back signal 2 (Upstream signal or
downstream signal depending on connection of cables).
dT: Measured difference between T1 and T2 (minus
zero offset if applicable). dT is proportional to ow.
vF: Measured velocity
vS: Measured speed of sound of uid. Depends on
temperature.
SigQ: Number of valid signals (in percentage). QStar
UFM makes numerous measurements per second and automatically lters out unreliable signals. The number of
measurements in relation to valid measurements is SigQ.
The achievable number is related to the chosen quality (see Menu 2). The higher the quality the lower usually the SigQ. When setting quality to 0 there is basically no ltering of signals and all signals are used for measure­ment (with the risk of using bad signals.
K-Factor: Factor which compensates the inuence when measuring laminar ow (Re <8000). When
measuring in that area a (small) additional uncertainty might occur.
Reynolds: The number of Reynolds automatically affects the chosen compensation factor.
Send Code (=signal sequence): The used signal
sequence.
Diagnostic Menu 2:
MENU 2: Diagnostic Data
Theta: Angle between ultrasonic path and ow vector.
Path length: Length of acoustic path. Depends on pipe
size and on chosen installation mode of transducers.
(Delay): Represents the time when the measuring window starts.
Sensor distance: Distance between the two trans­ducers.
Bar index: Mounting positions when using spacer bar.
Gain: The required amplication (gain) of the signals
is automatically set continuously. 0 represents no
amplication (not applicable) while 255 represents the maximum amplication. The lower the gain the better
the signal transmission. Very high gain might indicate a disturbed signal (gas, particles, wrong mounting).
dT Corr: Time shift of signals created by zero setting.
Quality: The quality parameter represents the “thor-
oughness” of the internal lter. Filter means that QStar UFM checks each signal if plausible or not using the quality parameter. When choosing quality 0, there will be no ltering. That means that each signal is used for
measurement even the bad (and maybe wrong) ones.
Filtering of 100 would mean that ltering is very strict.
Both values ( 0/100) are not recommended. Typical values are 50-75. When not getting measurement it
might be useful to set quality lower (for example, set
to 20).
Diagnostic Menu 3:
52
Page 53
Menu 3: Diagnostic Data
T1/T2 (raw): Signal run times along complete signal path including pipe wall and transducer enclosure.
dt (raw): Measured time difference of T1/T2 (raw).
XDCR (DCR=Transducer): Signal run time within transducer.
Pipe delay: Signal run time within pipe wall.
qp: Compensation factor considering changes in
temperature (and therefore changes in speed of sound
of uid).
Coating delay: Delay in signal run time generated by coating (if applicable).
Lambda Corr.: Delay can be manually edited (not recommended, only for maintenance) by multiples of wave length of signal.
Delay Corr: Shows Lambda Corr in seconds.
SOFTWARE UPDATE
Check software version
Before making any software update, check version of
software.
Switch off transmitter and restart it. During “start” locate
the version at the bottom of start screen (see arrow). Note this value.
Consult GPI/website for latest revision levels. If you are using an earlier version, contact GPI for updates and procedures.
53
Page 54
FLUID PROPERTIES
SPEED OF WATER
° C V in/s
0 1403 1 1408 2 1413 3 1417 4 1422 5 1427 6 1431 7 1435 8 1439
9 1444 10 1448 11 1452 12 1455 13 1459 14 1463 15 1466 16 1470 17 1473 18 1476 19 1480 20 1483
° C V in/s
21 1486 22 1489 23 1492 24 1494 25 1497 26 1500 27 1502 28 1505 29 1507 30 1509 31 1512 32 1514 33 1516 34 1518 35 1520 36 1522 37 1524 38 1526 39 1527 40 1529 41 1531
° C V in/s
42 1532 43 1534 44 1535 45 1537 46 1538 47 1539 48 1541 49 1542 50 1543 51 1544 52 1545 53 1546 54 1547 55 1548 56 1549 57 1549 58 1550 59 1551 60 1551 61 1552 62 1552
° C V in/s
63 1553 64 1553 65 1554 66 1554 67 1554 68 1555 69 1555 70 1555 71 1555 72 1555 73 1555 74 1555 75 1555 76 1555 77 1555 78 1555 79 1555 80 1555 81 1555 82 1554 83 1554
° C V in/s
84 1554 85 1553 86 1553 87 1552 88 1552 89 1551 90 1551 91 1550 92 154958 93 1549 94 1548 95 1548 96 1547 97 1546 98 1545 99 1544
100 1543
° F V in/s
32 55226
33.8 55422
35.6 55613
37.4 55800
39.2 55983 41 56161
42.8 56335
44.6 56506
46.4 56672
48.2 56834 50 56992
51.8 57146
53.6 57297
55.4 57444
57.2 57587 59 57726
60.8 57862
62.6 57995
64.4 58124
66.2 58250 68 58372
° F V in/s
69.8 58492
71.6 58607
73.4 58720
75.2 58830 77 58937
78.8 59041
80.6 59142
82.4 59239
84.2 59335 86 59427
87.8 59516
89.6 59603
91.4 59687
93.2 59769 95 59847
96.8 59924
98.6 59997
100.4 60069
102.2 60137 104 60204
105.8 60268
° F V in/s
107.6 60330
109.4 60389
111. 2 60446 113 60501
114.8 60554
116.6 60604
118.4 60652
120.2 60698
122 60743
123.8 60785
125.6 60825
127.4 60863
129.2 60899
131 60933
132.8 60965
134.6 60995
136.4 61024
138.2 61050
140 61075
141.8 61098
143.6 61119
° F V in/s
145.4 61138
147.2 61156 149 61172
150.8 61185
152.6 61198
154.4 61209
156.2 61218 158 61225
159.8 61231
161.6 61235
163.4 61238
165.2 61239 167 61238
168.8 61236
170.6 61233
172.4 61228
174.2 61221 176 61213
177.8 61204
179.6 61193
181.4 61180
° F V in/s
183.2 61167 185 61152
186.8 61135
188.6 61117
190.4 61098
192.2 61077 194 61055
195.8 61031
197.6 6100709
199.4 60981
201.2 60954 203 60925
204.8 60896
206.6 60865
208.4 60832
210.2 60799 212 60764
54
Page 55
FLUID PROPERTIES (Continued)
DYNAMIC VISCOSITY
LIQUID ° F V in/s ° C pg/cm³ V m/s V (x10-6m2/s)
Acetone 68.00 46850 20 0.7905 1190 0.407
Aniline 68.00 65315 20 1.0216 1659 1762
Ether 68.00 39606 20 0.7135 1006 0.336
Ethylene glycol 68.00 65591 20 1.1131 1666 21.112
Chloroform 68.00 39409 20 1.4870 1001 0.383
Glycerin 68.00 75709 20 1.2613 1923 1188.5
Acetic acid 68.00 45630 20 1.0495 1159 1.162
Methyl acetate 68.00 46496 20 0.928 1181 0.411
Ethyl acetate 68.00 45827 20 0.900 1164 0.499
Heavy water 68.00 54646 20 1.1053 1388 1.129
Carbon tetrachloride 68.00 36929 20 1.5942 938 0.608
Mercury 68.00 57126 20 13.5955 1451 0.114
Nitrobenzene 68.00 57992 20 1.207 1473 1.665
Carbon disulde 68.00 45591 20 1.2634 1158 0.290
n-pentane 68.00 40630 20 0.6260 1032 0.366
n-hexane 68.00 42638 20 0.654 1083 0.489
Spindle oil 89.60 52126 32 0.905 1324 15.7
Gasoline 93.20 49213 34 0.803 1250 0.4 to 0.5
Water 56.30 57480 13,5 1. 1460 1.004(20°C)
SPEED OF SOUND OF PIPE MATERIALS
Material V m/s V in/s
Iron 3230 127165
Steel 3206 126220
Ductile cast iron
Cast iron 2460 96850
Stainless steel
Copper 2260 88976
Lead 2170 85433
Aluminum 3080 121260
Brass 2050 80709
VinyIchIoride 2640 103937
Acrylics 2644 104094
FRP 2505 98622
Mortar 2500 98425 Tar epoxy 2505 98622 Potyethylene 1900 74803 Teon 1240 48819
3000 118110
3206 126220
SPEED OF SOUND SUBJECT TO CHANGE OF TEMPERATURE IN WATER
LIQUID ° C ° F pg/cm³ V m/s V in/s
Acetone 20 68 0.7905 1190 46850 Aniline 20 68 1.0216 1659 65315 Alcohol 20 68 0.7893 1168 45984 Ether 20 68 0.7135 1006 39606 Ethylene glycol 20 68 1.1131 1666 65591 n-octane 20 68 0.7021 1192 46929 o-xylene 20 68 0.871 1360 53543 Chloroform 20 68 1.4870 1001 39409 Chlorobenzene 20 68 1.1042 1289 50748 Glycerin 20 68 1.2613 1923 75709 Acetic acid 20 68 1.0495 1159 45630 Methyl acetate 20 68 0.928 1181 46496 Ethyl acetate 20 68 0.900 1164 45827 Cyclohexane 20 68 0.779 1284 50551 Dithionic acid 20 68 1.033 1389 54685 Heavy water 20 68 1.1053 1388 54646
LIQUID ° C ° F pg/cm³ V m/s V in/s
Carbon tetrachloride 20 68 1.5942 938 36929 Mercury 20 68 13.5955 1451 57126 Nitrobenzene 20 68 1.207 1473 57992 Carbon disulde 20 68 1.2634 1158 45591 Chloroform 20 68 2.8904 931 36654 n-propyl alcohol 20 68 0.8045 1225 48228 n-pentane 20 68 0.6260 1032 40630 n-hexane 20 68 0.654 1083 42638 Light oil 25 77 0.81 1324 52126 Transformer oil 32.5 91 0.859 1425 56102 Spindle oil 32 90 0.905 1342 52835 Petroleum 34 93 0.825 1295 50984 Gasoline 34 93 0.803 1250 49213 Water 13.5 56 1. 1460 57480
Sea water (salinity: 3.5%)
16 61 1. 1510 59449
55
Page 56
SPECIFICATIONS QSTAR PORTABLE QSTAR FIXED
Operation: Intuitive via 8 main keys (Soft Keys), plain text display
Languages: English, Spanish and French
Units: Metric/US
Outputs:
Inputs: 2x PT100
Integrated Data Logger: 2 GB N/A
Data Logged: Measurement and totalizers N/A
Data Format: Can be exported into standard ofce programs N/A
Memory Cycle: Adjustable, 1 second to 24 hours N/A
Power Supply:
Protection Class: IP40 IP65, Ex
Housing: Aluminium, PVC PVC, wall-mounted
Dimensions: 10.4 x 7.5 x 2.7 in. 10.2 x 9.4 x 4.7 in.
Operating Temp: -4° F to 140° F
Transducer Temp: -40° F to 300° F
Weight: 3.3 lbs 2.9 lbs
Display: QVGA (320x240), black and white, adjustable backlighting
Carrying Case 20 x 16 x 16 N/A
2x 4-20 mA, 1x Relay, 1x MicroUSB
1x Pulse
Integrated rechargeable battery and 110V AC adapter 85-264VAC, 18-36VDC (opt.)
Battery Duration: Approximately 5 hours Power Consumption: 10 W
2x 4-20 mA, 1x Pulse, 1x MicroUSB
1x Relay, RS232 (opt.)
MEASUREMENT
Principle: Ultrasonic transit time difference with AFC technology
Values Meas: Flow, ow speed, heat ow
Totalizers: Heat quantity, volume
Meas. Range: +/- 98 ft/s
Signal Damping: 0 - 100 sec (adjustable)
Acoustic velocity, signal strength, SNR, signal quality,
Diagnostic Functions:
MODEL NO. DESCRIPTION
QME05 Ultrasonic Flowmeter (ENERGY-FIXED, .5 MHz) 8" to 240"
QME10 Ultrasonic Flowmeter (ENERGY-FIXED, 1 MHz) 1.5" to 16"
QME20 Ultrasonic Flowmeter (ENERGY-FIXED, 2 MHz) .5" to 4"
QMF05 Ultrasonic Flowmeter (FIXED, .5 MHz) 8" to 240"
QMF10 Ultrasonic Flowmeter (FIXED, 1 MHz) 1.5" to 16"
QMF20 Ultrasonic Flowmeter (FIXED, 2 MHz) .5" to 4"
amplitude, energy
Oscilloscope function allows graphical display and analysis of signals.
PIPE SIZES
(Above include temperature sensors)
MEASUREMENT ACCURACY
Inner Diameter Ø
.39 - .98 in.
.98-1.97 in.
1.97-11.81 in.
11.81-236.22 in.
Repeatability for majority of applications is <0.2%
MODEL NO. DESCRIPTION
QMP05 Ultrasonic Flowmeter (PORTABLE, .5 MHz) 8" to 240"
QMP10 Ultrasonic Flowmeter (PORTABLE, 1 MHz) 1.5" to 16"
QMP20 Ultrasonic Flowmeter (PORTABLE, 2 MHz) .5" to 4"
QMF-PT100 Temperature Sensor Kit, FIXED (16 FT)
QMP-PT100 Temperature Sensor Kit, PORTABLE (16 FT)
Pipe Wall Thickness Gauge
Range Deviation
6.56-98.42 ft/s 2.5% of reading 0-6.56 ft/s ± 0.16 ft/s
6.56-98.42 ft/s 1.5% of reading 0-6.56 ft/s ± 0.10 ft/s
6.56-98.42 ft/s 1% of reading 0-6.56 ft/s ± 0.07 ft/s
3.28-98.42 ft/s 1% of reading 0-3.28 ft/s ± 0.03 ft/s
PIPE SIZES
56
Page 57
Page 58
Page 59
Page 60
Limited Warranty Policy
Great Plains Industries, Inc. 5252 E. 36th Street North, Wichita, KS USA 67220-3205, hereby provides a limited warranty against
defects in material and workmanship on all products manufactured by Great Plains Industries, Inc. This product includes a 1 year warranty. Manufacturer’s sole obligation under the foregoing warranties will be limited to either, at Manufacturer’s option, replacing or
repairing defective Goods (subject to limitations hereinafter provided) or refunding the purchase price for such Goods theretofore paid
by the Buyer, and Buyer’s exclusive remedy for breach of any such warranties will be enforcement of such obligations of Manufacturer.
The warranty shall extend to the purchaser of this product and to any person to whom such product is transferred during the warranty period.
The warranty period shall begin on the date of manufacture or on the date of purchase with an original sales receipt. This warranty shall not apply if:
A. the product has been altered or modied outside the warrantor’s duly appointed representative;
B. the product has been subjected to neglect, misuse, abuse or damage or has been installed or operated other than in accor-
dance with the manufacturer’s operating instructions.
To make a claim against this warranty, contact the GPI Customer Service Department at 316-686-7361 or 888-996-3837. Or by mail at:
The company shall, notify the customer to either send the product, transportation prepaid, to the company at its ofce in Wichita, Kansas, or to a duly authorized service center. The company shall perform all obligations imposed on it by the terms of this warranty
within 60 days of receipt of the defective product.
GREAT PLAINS INDUSTRIES, INC., EXCLUDES LIABILITY UNDER THIS WARRANTY FOR DIRECT, INDIRECT, INCIDENTAL AND CONSEQUENTIAL DAMAGES INCURRED IN THE USE OR LOSS OF USE OF THE PRODUCT WARRANTED HEREUNDER.
The company herewith expressly disclaims any warranty of merchantability or tness for any particular purpose other than for which
it was designed.
This warranty gives you specic rights and you may also have other rights which vary from U.S. state to U.S. state.
Note: In compliance with MAGNUSON MOSS CONSUMER WARRANTY ACT – Part 702 (governs the resale availability of the war-
ranty terms).
Great Plains Industries, Inc.
5252 E. 36
Wichita, KS, USA 67220-3205
th
St. North
©2013 by GREAT PLAINS INDUSTRIES, INC., Wichita, KS Printed in U.S.A.
5252 East 36th Street North
Wichita, KS USA 67220-3205
316-686-7361 | FAX: 316-686-6746
888-996-3837 |
www.GPImeters.net
Loading...