Precision Meters - Sanitary Clamp
Use this meter in pre-process applications where
high accuracy is required without 3-A Approval.
Model GSCP
Tri-Clover® Clamp
GSCP shown here
with Local Display
A c cu r Ac y : ± 0 . 5%
Select Your Meter Size:
1/2 inch Meter with 3/4 or 1 inch Fitting
3/4 inch Meter with 1-1/2 inch Fitting
1 inch Meter with 1-1/2 inch Fitting
1-1/2 inch Meter with 1-1/2 inch Fitting
2 inch Meter with 2 inch Fitting
Sensor Options:
■ Low Drag Pickup
(1/2 in. turbines)
■ Standard Pickup
(3/4 to 2 in. turbines)
Electronics Options:
■ GG510 (Display with
Pulse Output)
■ GX510 (Display with
4-20 mA Output)
■ GA510 (4-20 mA Output)
■ SC510 (Scaled Pulse Output)
MODEL GSCP – SPECIFICATIONS
Design Type: Turbine
Housing Material:
Meter Sizes A
Meter ID: 1/2”
1/2”
3/4”
1”
1-1/2”
2”
Flow Range: 1/2” (050)
1/2” (051) 0.8 - 6 GPM (3.0 - 22 LPM)
3/4” (075) 1.6 - 16 GPM (6.0 - 60 LPM)
3/4” (075E) 2.3 - 23 GPM (8.7 - 87 LPM)
1” (100) 6.7 - 67 GPM (25.2 - 252 LPM)
1-1/2” (150) 17.7 - 177 GPM (67.0 - 670 LPM)
2” (200) 33 - 330 GPM (125.0 - 1250 LPM)
Accuracy (Linearity):
Repeatability:
Pressure Rating:
Operating T
ypical K-Factor: 1/2” (050)
T
1/2” (051)
3/4” (075)
3/4” (075E)
1” (100)
1-1/2” (150)
2” (200)
Wetted Materials:
Housing:
Sleeve Bearings:
Thrust Bearing:
Shaft:
Rotor:
Rotor Supports:
Recommended Strainer Size:
1/2”
3/4”
1”
1-1/2”
2”
Frequency Output:
1/2” (051)
3/4” (075)
3/4” (075E)
1” (100)
1-1/2” (150)
2” (200)
✝
GSCP-050 requires RF Pickup.
vailable (ID): 1/2” 3/4” 1” 1-1/2” 2”
1” Fitting
1-1/2” Fitting
1-1/2” Fitting
1-1/2” Fitting
2” Fitting
✝
0.6 - 6 GPM (2.2 - 22 LPM)
emperature Range:
1/2” (050)
316 Stainless Steel
3/4” Fitting
± 0.5%
± 0.1%
Limited by fitting size, clamp size & temp.
-100°F to +185°F (-74°C to +85°C)
✝
10,000
10,000
3,750
2,608
896
340
181
316 Stainless Steel
PTFE
440C Stainless Steel
316 Stainless Steel
CD4MCu Stainless Steel
316 Stainless Steel
40 mesh
40 mesh
40 mesh
18 mesh
14 mesh
100 - 1000 Hz
125 - 1000 Hz
100 - 1000 Hz
100 - 1000 Hz
100 - 1000 Hz
100 - 1000 Hz
100 - 1000 Hz
section 1: g series precision products
buy smart.buy value.buy
8
06/08 ML-1800-6