GPI GM50 series Instruction Manual

Page 1
MS244G
0899 0001
Positive Displacement Flowmeters
GM50 series instruction manual
GM50 Pulse • GM50 Standard LCD • G540 Deluxe LCD • From serial No. CXXXX
Page 2
Thank you for purchasing a GPI GM Series Flow Meter. Please take a few minutes to read through this manual before installing and operating your meter. If you have any problems with the meter, refer to the maintenance and trouble shooting sections of this manual.
This manual contains connection and operating instructions for the M50 Series meters with pulse outputs and liquid crystal displays. For models with deluxe liquid crystal display an additonal instruction manual is
supplied. If you need further assistance, contact your local GPI representative or contact GPI by telephone or fax.
The GPI GM Series Flow Meter has incorporated the oval rotor principal into its design. This has proven to be a reliable and highly accurate method of measuring flow. Exceptional repeatability and high accuracy over a wide range of fluid viscosities and flow rates are features of the GM Series flow meter design. The low pressure drop and high pressure rating
means the GM Series flow meter is suitable for both gravity and pump (in line) applications.
The GPI GM Series flow meters are available in either aluminum or 316 stainless steel. Standard rotors are made from PPS (Polyphenylene Sulfide Resins) with optional 316 stainless steel rotors available for both stainless steel and aluminum models.
The M50 Series is available with either; * Standard Pulse * Standard LC Display and Pulse * Deluxe LC Display and Pulse
PLEASE READ THIS INFORMATION
CAREFULLY BEFORE USE!
Before use, confirm the fluid to be used is compatible with the meter (refer to the GPI fluid compatibility chart), or consult your local GPI representative for advice.
This meter will handle particle sizes up to 0.4mm/0.016”.
To prevent damage from dirt or foreign matter, GPI recommends a Y or Basket type 60 mesh strainer be installed as close as possible to the inlet side of the meter (if required contact GPI for further information).
Note: When a strainer is installed it should be regularly inspected and cleaned. Failure to keep the strainer clean will dramatically effect flow meter performance.
To prevent damage to the meter slowly fill the system with fluid (this will prevent damage caused by air purge).
Note: Failure to do this could damage the meter.
For pump applications, turn off the pump at the end of each day.
Maintenance can be carried out to the liquid crystal displays and pulse units without removing or isolating the meter from the line. When maintenance to any other part of the meter is required,
the meter must be isolated and the line pressure reduced.
The reed switch pulse unit can cause inaccurate rate counts when used with high speed counters. It is advised that a debounce circuit be used or alternatively use the hall effect sensor option.
PTB - Deluxe LC Display EEX1A IIC T6 (PTBnr EX-93.C4033X)
1
To the owner
Important Information
Page 3
1] GPI recommends that when setting
up pipework for meter installations a bypass line be included in the design. This provides the facility for a meter to be removed for maintenance without interrupting production. (See Fig.1)
2] Use thread sealant on all pipe
threads. For flanged versions 2” ANSI 150lb or DIN16, appropriate companion flanges and gaskets are required.
3] For pump applications ensure pipe
work has the appropriate working pressure rating to match the pressure output of the pump. The maximum working pressures are;
2” flanged stainless steel or aluminum series. See appropriate ANSI/DIN flange rules.
4] Install a wire mesh strainer (Y or
basket type 40 mesh as close as possible to the inlet side of the meter.
5] Ensure that the meter is installed so
that the flow of the liquid is in the
direction of the arrows embossed on the meter body.
6] The meter can be installed in any
orientation as long as the meter shafts are in a horizontal plane. (Refer to Fig.2 for correct installation) The register assembly may be orientated to suit the individual installation.
Note: Incorrect installation can cause premature wear of meter components.
7] Do not over tighten meter connections.
8] It is important that after initial
installation you fill the line slowly, high speed air purge could cause damage to the rotors.
9] Test the system for leaks.
10] Check the strainer for swarf or
foreign material, after the first 200 litres check periodically, particularly if the flow rate decreases.
2
Installation
When fluid passes through the meter, rotors turn. The magnets which are located in the rotors will pass across the pulser circuit board (containing either Reed switches or Hall Effect
sensors). A signal is received which is then sent by the Pulse Circuit Board (PCB) to the relevant LC display or receiving instrument.
Operation
1
2
Bypass Line
Flow Outlet
Flow Inlet
Strainer
Do Not Install Meter This Way
Page 4
Disassembly
Ensure that the fluid supply to the meter is disconnected, and the line pressure is released before disassembly, with the exception for repair or maintenance to the LC
Display or PCB where there is no necessity to isolate the meter from flow. Refer to the exploded parts diagram on pages 7 and 8 for item numbers.
1a] Units with Pulse Caps; Undo the
conduit connector, remove pulse cap (item 9) and remove the wires from the pulse terminal board (item 5).
1b] Standard LC Display; Mark the
display orientation with a marking pen, unscrew the four large screws (Item 30) on top of the LC Display. Carefully separate the LC Display from the plastic housing and disconnect the wires from the
Reed Switch Connections for PCB Terminals - refer Fig.3
Hall Effect Sensor Connections - refer Fig.4
3
Electrical Connections
Service Instructions
3
Contact rating 15VA
Maximum Voltage 150VDC
Configuration 1
2 x pulse outputs
Configuration 2
Link 2 & 3 for double
pulse output
LCD Versions
1 x Hall Effect Sensor 1 x Reed Switch
1 - Reed Switch 2 - Reed Switch 3 - HE Common ­4 - HE Signal 5 - HE Supply +
Pulse V
ersions Only
2 x Hall Effect Sensor
1 - HE Supply + 2 - HE 1 Signal 3 - HE Common ­4 - HE 2 Signal 5 - HE Ground
4
Hall Effect Voltage 4.5 to 24 VDC Current Draw Minimum 4.6mA Output NPN Open Collector 25mA
Page 5
This section of the instruction manual assists you in the operating of the GPI Standard LC Display. Please take a few minutes to acquaint yourself with the information that follows. Replace Battery
1] Remove the two large screws (Item
30) and the two small screws (Item
31) from the battery cover plate (Item 29).
2] Remove the battery cover plate
Standard LC Display
4
pulse terminal block. (See Standard LC Display instructions).
1c] Deluxe LC Display; Mark the
display orientation with a marking pen, remove the four retaining screws on the display face (Item
16). Lift off the display unit and remove the 9 pin connector at the back of the display unit. (Refer to separate Deluxe LC Display instruction manual).
2] Remove the mounting adaptor plate
and gasket (Item 14).
3] Loosen the eight cap head screws
(Item 7) that hold down the meter cap (Item 4), remove the screws, washers and lift off the cap.
4] Remove the o’ring (Item 2) from the o’ring
groove in the meter cap (Item 4).
5] Remove rotors (Item 3).
Reassembly
1] Before reassembling check the
condition of the rotors (replace if necessary).
2] Check that the smooth side of the
rotors (not the plug side) is facing you when inserting the rotors, the smooth side of the rotor is the
magnet side. There is no difference between rotor one or rotor two.
3] Replace the rotors (Item 3) onto the
shafts at 90oto each other (refer Fig.
5) and check their operation by turning either of the rotors. If the rotors are not in mesh correctly or do not move freely, remove one of the rotors and replace correctly at 90oto the other rotor. Re-check the operation of the rotors.
4] Replace the o’ring (Item 2) into
groove in the meter cap, if the o’ring has grown or is damaged in any way replace it with a new part.
5] Replace the meter cap making sure
that the locating pin in the body lines up with the hole in the meter cap. Insert the cap head screws (Item
7) and tighten in the sequence 1, 6, 2, 5, 3, 7, 4, 8.
6] The replacement of cables and
connectors are a reversal of the disassembly procedure, replace conduit fitting if required. When replacing the Standard LC Display or the Deluxe LC Display, confirm the orientation marks made on disassembly are aligned then screw the register into place.
7] Test the meter by turning the rotors
with a finger or by applying very low air pressure (no more than a good breath) to one end of the meter, before returning the meter to the line.
Pulse Circuit Board (PCB) Notes:
The PCB (Item 5) is fitted with (A) two reed switches; (B) hall effect sensors; or (C) one reed switch and one hall effect sensor. The PCB board is fastened to the meter cap (Item 4) by two screws and stand off’s. All care and caution should be taken when removing or handling the PCB as both the reed switch and hall effect sensor are fragile.
Individual reed switches or hall effect sensors are not available as individual replacement parts and are only available with the PCB (Item 5).
5
Rotors must be at 90oto each other.
Rotor #2
Rotor #1
6
Standard LC Display
Page 6
5
(Item 29) and gasket (Item 28).
3] Remove battery, and clean any
corrosion from the battery terminals (to protect terminals from corrosion or condensation coat the terminals with petroleum jelly).
4] Install the new battery.
5] Check gasket (Item 28) for damage
(replace if required). Position gasket (Item 28) and cover plate (Item 29) correctly, insert the screws and tighten.
Display Operation
All meter operations are reflected in the readout on the face of the meter. The readout contains three lines of information; refer to Fig. 7.
Note: The calibration is preset and is permanently stored in the meter’s computer.
The meter will turn on automatically when the liquid flows through the meter. The meter can also be turned on manually by pressing and releasing the display button.
T
urn Off
The meter turns off automatically two minutes after flow stops. When the meter is off, the readout is blank.
Batch T
otal
The resettable batch total indicates flow during a single use. This total can be reset by holding down the display button for three seconds until zeros appear. The batch total can be found by pressing and releasing the display button
until the bottom line displays “Total 2”.
Cumulative T
otal
The cumulative total is the total of all the liquid measured since the meter’s power supply was connected. This total cannot be manually zeroed, but will zero when the battery has been removed, goes dead, or when it reaches the maximum value of 999,999.
The cumulative total can be found by pressing and releasing the display button until the bottom line displays “Total 1 Locked”.
Flow Rate
The flow rate is the rate the fluid is flowing through the meter. The flow rate setting can be found by pressing and releasing the display button until the bottom line displays “flow rate”.
Propeller
Any time liquid flows through the meter, a small propeller displays.
Battery Replacement
Your meter is powered by a 9 volt lithium battery. (Replacement batteries; Lithium U9VL, Alkaline 522 or MN1604). If the meter’s readout should become dim or blank, the battery should be replaced.
When the battery either falls or is disconnected the batch and cumulative totals return to zero. The factory calibration is retained in the meter’s computer permanently and will not be affected by battery failure.
Auxiliary Facilities
Connection of an external 12 volt DC power supply facility;
1] Remove the 4 large screws (Item 30)
and remove the register from the coverplate.
2] Disconnect the internal battery.
3] Solder the external power leads to
the terminals marked ‘Ext 12V’ on the registers printed circuit board.
4] Re-assemble the register onto the
coverplate, replace the 4 screws (Item 30) and screw firmly together.
Connection of external pulse output
1] Remove the 4 large screws (Item 30)
and remove the register coverplate.
2] Connect external signal lead. (Refer
Fig. 8)
Pulse O/P and Ext Power are the connections located on the interface PCB, to use this output a 1K resistor will need to be soldered between the + voltage of the external power input and the + of the pulse O/P. Pulse signal represents the output cabling to the receiving instrument.
7
Top line indicates
“Preset Factory Calibration”
Bottom line indicates
“Totaliser” mode
Middle
line
indicates
“Flow”
Page 7
6
Meter Trouble Shooting
8
1KR (2 watt 1%)
Pulse O/P PCB Connection
Ext Power PCB
Connection 12~24
vdc 100mA
Pulse Signal to Instrument
TROUBLE SHOOTING GUIDE
TROUBLE
Fluid will not flow through meter
Reduced flow through the meter
Meter reading inaccurate
Meter not giving a pulse signal
LCD Register not working
CAUSE
a] Foreign matter blocking rotors b] Line strainer blocked
c] Damaged rotors d] Meter connections over tightened
e] Fluid is too viscous
a] Strainer is partially blocked b] Fluid is too viscous
a] Fluid flow rate is too high or too low b] Fluid is too viscous
c] Excess wear caused by incorrect
installation
a] Faulty hall effect sensor b] Faulty reed switch c] Magnets failed
a] Battery not connected properly b] Battery flat c] Faulty wiring connections
d] Faulty LC display e] Faulty connection from LC display
to Pulse PCB
REMEDY
a] Dismantle meter, clean rotors
(Strainer must be fitted in line)
b] Clean strainer c] Replace rotors (Strainer must be
fitted in line)
d] Re-adjust connections e] See specifications for maximum
viscosity
a] Clean strainer b] See specifications for maximum
viscosity
a] See “specifications” for minimum
and maximum flow rates
b] Consult GPI c] Check meter body and rotors.
Replace as required. Refer to installation instructions
a] Replace PCB Board b] Replace PCB Board c] Replace magnets
a] Check battery connections b] Replace battery c] Check wiring for loose or faulty
connections
d] Replace LC display e] Check wiring connections
Note: Internal 9 volt battery must be disconnected when using the display pulse output.
Page 8
7
Meter Parts Listing
1
23 4 56 7 8 910
1 1 MS283F Meter Body 2” ANSI 150lb Flange (Aluminium) 1 1 MS281D Meter Body 2” DIN16 Flange (Aluminium)
1 1 MS281F Meter Body 2” ANSI 150lb Flange (S/Steel) 1 1 MS283D Meter Body 2” DIN16 Flange (S/Steel)
21* BS252 “O” Ring (NBR) 21* BS252E “O” Ring (EPDM) 21* BS252TE “O” Ring (Teflon) 21* BS252V “O” Ring (Viton) 32* MS147S Rotors PPS (Polyphenylene Sulfide Resins) 32* MS147TS High Temperature Rotors (PPS) 32* MS147HS High Viscosity Rotors (PPS) 32* MS147HTS HighViscosity/High Temperature Rotors (PPS) 4 1 MS230 Meter Cap (Aluminium)
4 1 MS231 Meter Cap (Stainless Steel)
51* MS201-R PCB (Standard Reed Switch) 51* MS201-HE PCB (Hall Effect Sensor) 5 1 MS201-R/HE PCB ( 1 Reed Switch & 1 Hall Effect Sensor) 6 4 MS284S PCB Board Screws 76* MS243S Meter Cap Screws (Standard)
76* MS282S Meter Cap Screws (Stainless Steel)
81* MS300 Pulser Cap Gasket 9 1 MS160 Pulser Cap (Aluminium) 20mm Conduit Thread 9 1 MS160N Pulser Cap (Aluminum) 1/2” NPT Thread
9 1 MS170 Pulser Cap (Stainless Steel) 20mm Conduit Thread 9 1 MS170N Pulser Cap (Stainless Steel) 1/2” NPT Thread
10 4 MS115S Pulser Cap Screw (Stainless Steel) 11 1 MS37 Warning Lebel (Not Shown) 12 1 MS14 Explosion Proof Approval Label (Not Shown) 13 1 Customer to Specify Legend Plate (Not Shown) inc. Hammer Screws
Item
No.
No. Off.
Part or Set
(Order from this column only)
Part Description
Key: u Indicates recommended Spare Parts to stock Bold text indicates Stainless Steel Model Parts
Rec.
Parts
Page 9
8
MS280S Deluxe LCD Display (Complete)
14 1 MS279 Mounting Adaptor Plate 15 4 * MS117S Adaptor Screws 16 1 * MS69 LC Display Unit 17 4 * MS118S LC Display Mounting Screws 18 1 * MS127 Battery Retaining Screw 19 1 * MS126 Battery (Standard) CR2040 20 1 * MS68 Connector and Cable (Not Shown) 21 1 MS87 PTB Approval Label (Not Shown) 22 1 * MS128 Extended Life Battery 2/3 AA (Not Shown)
NOTE: Note Suitable for Intrinsic Use
33 1 * MS307 Gasket
MS275S Standard LC Display (Complete)
23 1 MS238 Mounting Adaptor Plate 24 4 * MS203S Adaptor Screws 25 1 * BS045 “O”Ring 26 1 * MS205 LC Display 27 1 * MS274 9 Volt Battery 28 1 * MS184S (Inc. MS184B) Battery Gasket 29 1 * MS184S (Inc. MS184A) Battery Cover 30 4 * MS206S Register Screws 31 4 * MS182S Battery Cover Plate Screws 32 1 Customer to Specify Battery Cover Label
Display Parts Listing
823242526282930
27
32 31
814
15
18 19
1716
33
Item
No.
No. Off.
Part or Set
(Order from this column only)
Part Description
Rec.
Parts
Page 10
9
Meter Specifications
Heating Jackets
Meter Type
Flow Ranges
(Litres per minute/US Gallons per minute)
Above 5 Centipoise
Below 5 Centipoise
Accuracy of Reading
Maximum Viscosity*
Maximum Operating Pressure
Maximum Operating Temperature
Pulse Type
Pulses Per Litre/US Gallon
Pulse
15 to 350/ 4 to 92 33 to 300/ 9 to 79 +/- 0.5% 1000 Centipoise As per flange rule 80°C/ 176°F
Dual Reed Switches or Hall Effect Sensor or combination HE Sensor/Reed Switch
6.68/ 13.3 or 25.28/ 50.34
Pulse with Standard
LC Display
15 to 350/ 4 to 92 33 to 300/ 9 to 79 +/- 0.5% 1000 Centipoise As per flange rule 80°C/ 176°F
Dual Reed Switches or Hall Effect Sensor or combination HE Sensor/Reed Switch
6.68/ 13.3 or 25.28/ 50.34
Pulse with Deluxe
LC Display
15 to 350/ 4 to 92 33 to 300/ 9 to 79 +/- 0.5% 1000 Centipoise As per flange rule 80°C/ 176°F
Dual Reed Switches or Hall Effect Sensor or combination HE Sensor/Reed Switch
6.68/ 13.3 or 25.28/ 50.34
1
2
3
Complete Assembly: GHJ500-1 - Aluminium BSP
GHJ500-2 - Aluminium NPT
Spare Parts Listing:
GHJ400-1: MS333 SHCS Screws
MS133B Jacket Body - Aluminium/BSP BS156 “O” Ring
GHJ400-2: MS333 SHCS Screws
MS133N Jacket Body - Aluminium/NPT BS156 “O” Ring
* Unless High Viscosity or High Temperature Rotors are fitted
Page 11
178mm
102mm
185mm
55
mm
10
Meter Dimensions
178mm
86mm
188mm
18mm
55mm
137mm
178
mm
55mm
GM50 Pulse Meter Dimensions
GM50 Pulse Meter with Standard LC Display
GM50 Pulse Meter with Deluxe LC Display
240mm
152mm
213mm
210mm
240mm
152mm
210mm
240
mm
127
mm
235mm
152mm
Page 12
Warranty
Great Plains Industries, Inc. Limited Warranty Policy
Great Plains Industries, Inc., 5252 East 36th Street North, Wichita, Kansas USA 67220-3205, hereby provides a limited one year warranty against defects in material and workmanship on all products manufactured by Great Plains Industries, Inc. This warranty shall extend to the purchaser of this product and to any person to whom such product is transferred during the warranty period.
The warranty period shall begin on the date of the original new equipment purchase. Warrantor’s obligation hereunder shall be limited to repairing defective workmanship or replacing or repairing any defective part or parts. This warranty shall not apply if:
a.)The product has been altered or modified outside the warrantor’s duly appointed
representative;
b.)The product has been subjected to neglect, misuse, abuse or damage or has been installed or operated other than in accordance with the manufacturer’s operating instructions.
To make a claim against this warranty, notice of claim must be given in writing to the company at its address below no later than 30 days after the expiration of the warranty period. Such notice shall identify the defect in the product. The company shall, within 14 days of receipt of such notice, notify the customer to either send the product, transportation prepaid, to the company at its office in Wichita, Kansas, or to a duly authorized service center. The company shall perform all obligations imposed on it by the terms of this warranty within 60 days of receipt of the defective product.
GREAT PLAINS INDUSTRIES, INC. EXCLUDES LIABILITY UNDER THIS WARRANTY FOR DIRECT, INDIRECT, INCIDENTALAND CONSEQUENTIAL DAMAGES INCURRED IN THE USE OR LOSS OF USE IF THE PRODUCT WARRANTED HEREUNDER.
The company herewith expressly disclaims any warranty of merchantability or fitness for any particular purpose other than for which it was designed.
This warranty gives you specific rights and you may also have other rights which vary from U.S. state to U.S. state.
NOTE: In compliance with MAGNUSON MOSS CONSUMER WARRANTY ACT - Part 702 (governs the resale availability of the warranty terms).
Loading...