Gould Advance DSO400u User Manual

Page 1
Gould 400 Series Digital Storage Oscilloscopes Operators Manual
Page 2
Gould 400 Series Digital Storage Oscilloscopes Operators Manual
Copyright@1990 Gould lnc., Instruments Division (Gould Electronics Ltd.)
Page 3
400 Operators Manual
A
Contents
Introduction
Getting Started
Advanced Features
The Menus
Performance Checking
Waveform Processing
1
2
3
4
5
Battery Unit Operat io n
Alphabetical Summary
Appendix 1 Error Messages
Appendix 2 Specification
Appendix 3 Remote Operation
ppendix 4 Front & Back
Pictures
I
Service Facilities
6
7
A1
A2
A3
Page 4
Contents
400 Operators Manual
Introduction
Getting Started
1.
1.1 Safety and Power Requirements
1.1.1
International Safety Warning
1.1.2
Grounding
1.1.3
Live Parts
1.1.4
Ventilation and Dust
1.1.5
Operating Temperatures
1.1.6
Power & Frequency
Requirements
1.1.7
1.2 Using the Buttons
1.3 Start-Up Display
1.4 Obtaining a Trace
1.5 Horizontal Adjustments
1.6 Vertical Adjustments
Fuse Requirements
1.4.1
AUTO SETUP
1.4.2
Channel Selection (Off/On/lnv)
1.4.3
Coupling (AC/DC/Gnd)
1.5.1
TIME/DIVISION
1.5.2
Aliases
1.5.3
Position
1.5.4
Magnification
1.6.1
VOLTS/DIVISION
1.6.2
Position
1.6.3
Variable/Uncalibrated
1.6.4
Add
1.7 Manually Obtaining a Trace
1.8 Operating Hints
1.8.1
1.8.2
Advanced Features
2.
2.1 Trigger Control
2.2 Captu e Facilities
2.3 Display Modes
2.4 Cursor Measurements
1.8.3
1.8.4
1.8.5
2.1.1
2.1.2
2.1.3
2.1.4 Slope
2.1.5
2.1.6
r
2.2.1
2.2.2
2.2.3
2.4.1
2.4.2
2.4.3
Intensities too low Trace off the top or bottom
Trace not being acquired Trace unstable Trace has flat top or bottom
Selecting Source and Coupling Level Trigger Point (T)
Trigger Mode (Norm/Auto) Trigger Delay
Trace Hold S/Shot and Run
Acquisition Status-AFTS
Cursor and Datum Selection
Making Measurements
of screen
(+/-)
The Cursor& Datum Lines
Plot
2.5
2.5.1
2.5.2
2.5.3
2.5.4
2.5.5
2.5.6
2.5.7
2.5.8
Internal Colour Plotter Pens
Pen Changing Paper Selection Paper Loading Internal plotting External plots Plot positioning and scaling
Data Transfer
2.6
2.6.1
2.6.2
2.6.3
Syntax Receiving Data Error Status
Real Time Clock
2.7
3.
The Menus
Additional Buttons
3.1
3.1.1
3.1.2
3.1.3
3.1.4
The Numeric Buttons Menu Traces Control Post Store
Control Master Menu
3.2
Status Menu
3.3
3.3.1
3.3.2
3.3.3
3.3.4
3.3.5
3.3.6
Mode Max/Min V/Div Probe set Timebase Trigger
Display and Trigger Menu
3.4
3.4.1
3.4.2
3.4.3
3.4.4
3.4.5
Probe ratio Max/Min Dot Join Averaging
Trig Pos'n
Display Intensity
3.5
Reference Trace
3.6
RS423 Interface Menu
3.7
3.7.1
RS423 Plot Connections
3.7.2
RS423 Data Connections
Special Functions Menu
3.8
Post Store Master Menu
3.9
3.10 Save Trace Menu Recall Memory Menu
3.11
3.12 Plot Menu
Performance checking
4.
Risetime
4.1
Bandwidth
4.2
Trigger sensitivity
4.3
Trigger bandwidth
4.4
Timebase calibration
4.5
Vertical Calibration
4.6
Max-Min (Alias Detector)
4.7
Page 5
400 Series Operators Manual
Contents
Waveform Processing Functions
5.
5.1 Cursor Measurements
5.1.1 Voltage and Time
5.1.2 Peak-Peak
5.1.3 Max-Min
5.1.4 Risetime (falltime)
5.1.5 Overshoot (preshoot)
5.1.6 Pulse Width
5.1.7 Frequency, Period, Duty Cycle
5.1.8 RMS
5.1.9 Area
5.2 Trace Manipulation
5.2.1 Filter Frequency
5.2.2 Filter Current Trace
5.2.3 Invert Current Trace
5.2.4 Integrate Signal
5.2.5 Integrate Current Trace
5.3 Trace arithmetic
5.4 Persistence/Limits Testing
5.4.1 Limits
5.4.2 Persistence
Battery Unit Operation
6.
Alphabetical Summary of controls
7.
Appendix 1: Error Messages
Appendix 2: Specification
Appendix 3: Remote DC operation of the DSO
Appendix 4: Front and Back Pictures
Service Facilities
Figure List
1.2a Single Function Buttons
1.2b The Toggles
1.2c
Pressure Sensitive Buttons
1.3
Start-Up Display
1.4
Obtaining a trace
1.4.1 An AUTO SETUP Display
1.5
Horizontal Controls
1.5.2 Alias Generation
1.6
Vertical Controls
2.1
Trigger Controls
2.1.6 Trace Capture with Delay
2.2
Capture Controls
2.4
The Cursor and Datum Line Controls
2.4.2 The Cursor and Datum Lines
2.5.5 Paper Loading
2.5.8a Relative Plot Positions
2.5.8b Plot Dimensions 3
3.1
3.2
3.3
3.4
3.5
3.6
3.7
3.7.1 400to Gould 6120 or HP7475 Connections
3.7.2a 400 to IBM-PC/XT Connections
3.7.2b 400 to IBM-PC/AT Connections
3.8
3.9
3.10 Save Trace Menu
3.11 Recall Trace Menu
3.12 Plot Menu
5.1a Cursor Measurements Menu
5.1b Example of Calculations on trace
5.1.1 Voltage and Time Measurement
5.1.2 Peak to Peak Measurement
5.1.3 Max-Min Measurement
5.1.4 Risetime Measurement
5.1.5 Overshoot Measurement
5.1.6 Pulse Width Measurement
5.1.7 Freq. Period and Duty Cycle Measurement
5.1.9 Area Measurement
5.2
5.4 6
6.2 A2 A3.2 DC Power Connections A4a
A4b
Menu Overview Numeric Buttons Control Master Menu
A Status Menu A Display and Trigger Menu Display Intensity Menu
Reference Trace Menu An RS423 Interface Menu
Special Functions Menu Post Storage Master Menu
Trace Manipulation Menu Persistence/Limits Testing Menu Battery Unit Interconnections Battery Unit Front Panel Indicators
400 Series Dimensions
Rear View RS423 Connections
Page 6
400 Operators Manual
Introducti
N
on
Introduction
The Gould 400 series instruments are Digital Storage Oscilloscopes (DSOs). They include all the features expected of advanced modern oscilloscop es designed for the professional engineer, whilst retaining the user­friendliness essential for those learning to use such instruments for the first time.
The 400 range consists of three basic models; the 400 the 420 and the 450. All versions can be fitted with a battery unit which provides the instrument with a fully automatic built in Nickel Cadmium battery and charger which allows uninterrupted operation of the DSO in the even t of an AC supply failure and complete operation independent of an AC supply.
Obtaining a trace is especially simple - just connect the signal and press the Auto Setup button - the 400 does the rest. Having obtained a trace, readily accessible datum lines and a cursor make it easy to take automatic timing and voltage measurements directly from the display. On the 420 and 450 models, the waveform processing function increases the power of the cursor measurements in terms of both capture and post storage analysis and measurement functions.
The innovative use of ergonomically designed pressure sensitive push button controls provides a combination of precision and flexibility for ease of operation.
More advanced features of the 400 series include a com­prehensive range of menu-controlled functions. For example, the Display and Trigger menu operates features such as the trigger delay and pre-trigger display functions. The pre-trigger display function allows the signal prior to the trigger point to be captured and displayed.
Three complete traces can be stored for future use and recalled to the display via the Save Trace and Recall Trace menus respectively. With the built in battery back up facility, these will be retained even when the instrument is switched off.
In addition to the above features, the 420 and 450 instru­ments have a built in 4 colour plotter and a battery backed Real Time Clock. This provides a simple and convenient method of obtaining permanent hard copy plots of the screen display. The plots will contain the date and time of acquisition together with the date and time of plotting.
The 450 has all the features of the 420 and has a signal bandwidth of 50MHz and an extra timebase range of 50ns/div.
umbers circled in the text refer to the controls shown
on the front panel picture in Appendix 3.
Page 7
Getting Started 1.1-1.1.6
400 Series Operators Manual
1.1 Safety and Power Requirements
1.1.1 International Safety Warning (as required for I.E.C. 348 Cat I)
This instrument has been designed and tested in accordance with IEC publication 348, and has been supplied in a safe condition. This manual contains information and warnings which must be observed to keep the instrument in a safe condition. The instrument should not be switched on if it is damaged and it should not be used under wet conditions.
For the correct and safe use of this instrument it is essential that both operating and service personnel follow generally accepted safety procedures in addition to the safety precautions specified in this manual.
Whenever it is likely that safety-protection has been impaired, the instrument must be made inoperative and be secured against any unintended operation. Qualified maintenance or repair personnel should be informed. Safety protection is likely to be impaired if, for example the instrument shows visible damage or fails to perform the intended measurements correctly.
1.1.2 Grounding
THE INSTRUMENT MUST BE GROUNDED.
AC
The instrument must be operated with a protective ground connected via the yellow/green conductor of the supply cable. This is connected to the instrument before the line and neutral connections when the supply plug is inserted into the socket on the back of the instrument. If the final connection to the supply is made elsewhere, ensure that the ground connection is made before line and neutral.
DC
If the unit is disconnected from the AC supply, and powered from an isolated DC source or the internal battery unit, the unit will not be grounded. Independent provision must be made to maintain the case at a safe potential, by grounding the safety ground terminal on the DSO rear panel. The safety ground terminal is connected to the instrument case, to the ground of the input signals (outer of the BNC connectors), and also internally to the negative side of the DC supply.
WARNING: Any interruption of the protective ground
conductor inside or outside the instrument is likely to make the instrument dangerous. Intentional interruption is prohibited.
Signal connections to the instrument should be connected after the ground connection is made and disconnected before the ground connection is removed, i.e. the supply lead must be connected whenever signal leads are connected.
1.1.3 Live Parts
The instrument should not be operated with covers removed. The covers protect the user from live parts and
they should be removed only by suitably qualified personnel for maintenance and repair purposes.
WARNING: Removing the covers may expose voltages in
excess of 8000V at the side of the display tube; these may be present for up to one minute after the instrument has been disconnected from the power source.
1.1.4 Ventilation and Dust
The instrument relies on forced air cooling via a fan and ventilation slots. Adequate ventilation can usually be achieved by leaving a 75mm (3" gap) around the instrument.
The instrument should not be operated in dusty environments. If the CRT filter requires cleaning it can be easily removed by
pressing in its right hand edge as shown by the moulded arrow.
1.1.5 Operating Temperatures
The instrument is designed to be operated in an environment having an ambient temperature of between 0 and 50 degrees C, (0 to 45 degrees if the battery unit is fitted) and to operate with full accuracy between 15 and 35 degrees C.
Note: Direct sunlight, radiators and other heat sources
should be taken into account when assessing the ambient temperature.
The instrument may occasionally be subjected to temperatures between 0
and -10ºC without degradation of its safety.
1.1.6 Power and Frequency Requirement s
The instrument uses less than 85V A (200V A if battery unit fitted) and operates from line voltages of 90V to 130V, and 190V to 265V, at 45 to 400Hz. Under the extreme conditions of 90V and 45Hz, the instrument will still operate correctly even if there is a half cycle dropout in the line supply.
The instrument may be powered from a direct current supply in the range 12V to 33V. The unit cannot be damaged by applying power to both inputs simultaneously.
Before connecting the instrument to the supply, ensure that the rear panel AC supply voltage selectors are set to the appropriate voltage.
Access to the voltage selector can only be made if the AC supply connector is removed, and is by hinging open the panel connector from its top edge. The selector barrel can then be moved to the required setting. This same operation provides access to the AC supply fuse, which must be changed to suit the supply voltage as shown in table 1.1. 7.
The AC power connection is via a standard IEC, CEE 22 connector and the DC power input is via the supplied 0.25" spade connector (Gould part No. 457839.) See appendix 4 for polarity details.
Page 8
400 Operators Manual
Figure 1.2a Single Function Buttons
Getting Started
Figure 1.2b The Toggles
Figure 1.2c The Pressure Sensitive Buttons
Page 9
Getting Started 1.1.7-1.4.1
pply
gg
yp
P
y
400 Series Operators Manual
CAUTION: The negative DC terminal is connected to the
safety ground terminal and the instrument case. Care should be taken to avoid ground loops when, for example, the instrument is used in a vehicle and powered by that vehicle's supply.
1.1.7 Fuse Requirements
The fuse arrangement shown in table 1.1. 7 must be followed, and additionally in the UK, a 3A fuse (5A if the battery unit is fitted) should be fitted in the line supply plug .
Su
Slow Blow Gould Su
Voltage Fuse Rating Part No Manufacturer/Type No
IEC (ULlCSA) DSO
230V 0.5A (0.6A) 457452 BeswickrrDC488,
115V lA (1.2A) 457454 BeswickrrDC488,
12VDC 5A (6A) 457979 HRC type
BATTERY UNIT
230V lA (1.2A) 457454 BeswickrrDC488,
115V 2A (2.5A) 457455 BeswickrrDC488,
Table
1.1.7
Fuse ratings
ested t
es.
Littlefuse/239, Schurter/FSP
Littlefuse/239
Schurter/FSP
Beswick/S505,
Wickman 19181
Littlefuse/239,
Schurter/FSP
Littlefuse/239,
Schurter/FS
1.3 Start-Up Display
When switched on with the POWER button will go through its automatic self-calibration sequence and then display information similar to that shown in Figure
Figure
1.3
Start-Up Displa
The trace is visible across the centre of the screen. At the top will be the sensitivity of the two input channels and the timebase speed. It any input is inactive, information for that channel will not be displayed.
Users with no previous experience of a DSO will find that the 400 responds like a conventional Real Time Oscilloscope while it is repetitively triggered, but in the absence of further valid triggers, it retains the last trace for continuous display.
(1)
.
the instrument
1.3.
Note: For different DC voltages, the DC fuse rating multiplied
by the voltage used should equal60W. e.g. at 30V a 2A
is required. For additional safety under extreme
fuse fault conditions, the DC fuse (HRC) type. This fuse should have a 1500A break capacity.
1.2 Using the Buttons
There are three types of buttons. The first type usually have only one function: when pressed, that one function will be
(
activated e.g. AUTO SETUP
The second type are toggles: each press of the button either switches something on or off e.g. Add next item in a sequence e.g. Off/On/lnv
1.2b. The third type are Pressure Sensitive: the effect caused depends
upon how hard the button is pressed. For example, a gentle
press of a Pos'n (8) button will move the trace slowly. If the
button is pressed harder, the trace will move more quickly. See Figure 1.2c.
24)
is
a high rupture current
see appendix 3
.
See Figure 1.2a.
(10) , or selects the
(13) . See Figure
1.4 Obtaining a Trace
1.4.1 AUTO SETUP
To display an input signal, connect it via either the CHI socket
See Figure 1.4
(9) or the CH2 socket (12) and press AUTO SETUP (24) .
Assuming you've applied a regular signal, say a 2kHz sine­wave with an amplitude of perhaps 5V peak to peak, it will almost immediately be represented on the display. An example is shown in Figure 1.4.1.
It the display is very dim or has completely disappeared see section 1.8.1.
AUTO SETUP will attempt to arrange the display so that two
to five complete cycles appear, with the amplitude set so that the height of the trace is between two and five screen divisions. Also, it selects Auto trigger to ensure that the screen is frequently updated and a trace will be visible.
If signals are connected to both channels, the highest amplitude takes priority.
Page 10
A
400 Series Operators Manual
Getting Started 1 .4.2-1 .5.1
Figure
1.4
Obtaining a Trace
1.4.3 Coupling (AC/DC/Gnd)
These buttons control the type of coupling between the input signal and the instrument. DC is the most generally applicable, and AUTO SETUP will normally set this control to DC, where possible. The input impedance is I M ohm in parallel with a capacitance of 28pF.
(14)
AC This is used to remove any DC component from input
signals. Suitable input signals (i.e. the bandwidth) are from 4Hz to 20MHz.
Gnd The input signal is internally disconnected from the inputs and
the amplifier grounded. A OV reference signal is displayed.
DC The input signal is directly coupled to the instrument so all
frequency components of the input signal will be displayed. The bandwidth will be from DC to 20MHz.
1.5 Horizontal Adjustments
Figure
1.4.1
n AUTO SETUP Display
1.4.2
Channel Selection (Off/On/lnv)
A channel may be switched on or off with its Off/On/Inv button. If the channel is on, its trace can be displayed in either normal or inverted mode.
Off The channel is deactivated. On The trace is a true representation of the input signal. Inv The input signal is inverted before being displayed. If
there is any DC component in the signal this will also be inverted and could cause the trace to disappear from the screen. Such an unwanted DC component can be removed by selecting AC coupling. Any vertical shift applied to the trace is not inverted. The trigger point remains at the same point on the waveform regardless of inversion.
(13)
Figure
1.5
Horizontal Controls
1.5.1
TIME/DIVISION
TIME/DIV
With a timebase of say 200/Ls, each horizontal screen division represents 200/Ls worth of signal. The timebase is shown near the top of the display - e.g. TB=200/Ls.
Assuming a 2kHz signal is applied as mentioned earlier, when the timebase is set to 500ms/div, an interesting phenomenon may appear: an 'alias'.
(33) These buttons control the sweep rate of the trace.
The timebase can be varied from 100ns/div to 50s/div in a I, 2, 5 sequence of values. The button marked 'ns' decreases the time/div, the button marked 'sec' increases the time/div.
Page 11
Getting Started 1 .5.2-1 .6.3
400 Series Operators Manual
1.5.2 Aliases
An alias is a false image. The instrument is a digital oscilloscope and so takes frequent samples of the input signal in order to update the trace. Thus, if the signal frequency is higher than the sample frequency, one sample will be taken from a particular point on the waveform and the next sample may be taken from a point slightly further along on a subsequent cycle. It will then display the wave as being much slower than it really is. See Figure 1.5.2. This effect is only likely to occur if there are more than 100 cycles of waveform across the screen.
Figure
1.5.2
Alias Generation
There is a 'Max/Min' glitch detection feature which can be used to detect aliases. With dot join on, this usually results in a filled-in wave envelope being seen on the screen, indicating a high
A.
frequency. See Section 3
are 20mV to 50V per division at the probe tip. See Section
3.4.1
.
If the Uncal light is on, then these buttons vary the
1.6.3.
sensitivity continuously. See Section
1.5.3 Position
Position (26) These buttons move all traces to the right or
left.
The position of the cursor (Section trace so it will move with the applied shift. With x-magnified
1.5.4),
traces (Section displayed on the screen. To bring it back into view use the CURSOR < > buttons
the cursor can be off the part of the trace
(5).
2.4)
is fixed in relation to the
1.5.4 Magnification
Mag (32) Switches horizontal magnification on or off When
switched on, a xl0 expansion is applied to any displayed trace, which will expand around the centre of screen. The timebase setting is adjusted to reflect the expansion.
The instrument displays 50 dots (samples) per screen division, each plotted dot value being obtained from the 512 byte acquisition memory. At an expansion of x 10. there are five dots per division.
1.6 Vertical Adjustments
Each channel has its own set of vertical controls. See Figure
1.6.1 VOL TS/DIVISION
V/DIV
(11) Adjusts the sensitivity of the instrument over discrete
calibrated ranges from 2m V to 5V per screen division in 1,2,5 steps. With a x10 probe the ranges
1.6.
1.6.2 Position
(8) These move their respective traces up and down
Pos'n
the display.
If Trace Hold is on (Section 2.2) or a S/Shot capture has been made, any part of the trace which was captured off-screen vertically will be shown by a horizontal line.
1.6.3 Variable/Uncalibrated
Var (15) When this is set to 'Uncal', the coarse setting of the
attenuator remains unchanged, but a variable attenuation is
004.
applied to the input signal in the range of 1 to with an initial setting of IV, the actual sensitivity of the channel could be set by this control to anywhere between I V and 2.5V per division. The V/DIV buttons are used to vary the uncalibrated sensitivity.
Thus,
Example screen display:
CH1 =5V Channel is set to a sensitivity of 5 Volts per screen
division.
CH2>20mV Channel 2 is uncalibrated and the attenuator is set to a
sensitivity greater than 20m V per screen division.
Page 12
400 Series Operators Manual Getting Started 1.6.4 - 1.8.5
1.6.4 Add
Add (10) Displays the sum, or if one channel is inverted the
difference, of the input signals. The original traces disappear and the resultant trace is displayed as a Channel 2 trace.
1.7
Manually Obtaining a Trace
This subsection describes how to obtain a trace without using AUTO SETUP.
You may need to consult later sections of the manual as not all of the operating features have been discussed so far.
1. When switched on, the instrument will go through its
power-up sequence, checking the internal calibration.
2. Decide to which channel you are going to apply the
signal (CH1 or CH2).
3. Make sure the chosen channel is active by setting the
Off/On/lnv
4. Set the AC/DC/Gnd button
nel to Gnd.
5. If necessary, turn that channel's Uncal light out by
pressing Var once.
6. Make sure that horizontal mag is not selected. (32)
7. Adjust the TIME/DIV setting
of
5μs.
8. Set the display Mode (31) to Refr.
9. Set the TRIGGER Norm/Auto button
10. Use the CHl/CH2/Ext/Line button (17) to choose the
source for trigger signals.
11. Set the trigger coupling with the hf rej/AC/DC button.
(13) for that channel to On.
(14) for the chosen chan-
(15)
(33) to give a timebase
(20) to Auto.
(18)
12. If necessary adjust the position of the trace using the
Position and Pos’n buttons.
13. Ensure that Trace Hold (25) is not selected.
14. Select Run.
15. Apply the signal through a BNC connector to the chosen
CH input socket. (9) or (12)
16. Set the AC/DC/Gnd button
either DC or AC, as appropriate.
17. Adjust the gain of the chosen channel using the V/ DIV
buttons by pressing the Var button once.
18. Adjust the timebase setting using the TIME/DIV but-
tons.
19. If the display is unstable, adjust the trigger Level. (22)
1.8
Operating Hints
The following list gives some of the more commonly met problems in operating digital oscilloscopes, how to correct them, and a brief explanation of what was wrong.
(27)
(11) . For intermediate settings, set Uncal on
(33)
(26)
&
(8)
(14) for the channel to
(15)
You may need to consult later sections of the manual as not all of the operating features have been discussed so far.
1.8.1 Problem: Traces and Alphanumerics very dim or completely disappeared.
Intensities too low
- Press the control button menu, then press numeric button 3 to select the intensity menu. Further presses of numeric button 3 will increase the brightness of the alphanumerics on the display. The traces can now be intensified using numeric button I.
Note:- The display may not become visible until the numeric
3
button
1.8.2 Problem: Trace off the top or bottom of the screen. Too
much vertical shift
Input has large DC offset
1.8.3 Problem: Trace not being acquired.
Instrument in single capture mode
Trigger level incorrect
Trigger source on the wrong input
Trigger coupling on an unsuitable setting
Trace Hold on
Timebase on very slow acquisition
1.8.4 Problem: Trace is unstable even when triggered. Alias
Noisy input
Trigger on Auto
1.8.5 Problem: Trace has a very flat top or bottom.
Trace captured when off screen vertically and Pos’n shift has been used
has been pressed several times.
- correct with that channel's Pos’n buttons.
- AC couple input signal.
- Correct with Pos’n buttons.
- Use a less sensitive range.
- Press Run.
- Select Auto and DC trigger, then adjust the level
control until the trigger level indicator bars are lined up
with the centre of the trace.
- Change trigger source. (CH1, CH2, Ext, Line) (17)
- Change trigger coupling. (hf rej, AC, DC) (18)
- Release Trace Hold.
-Adjust TIME/DIV.
- Check for alias using Max/Min (glitch detection), and select a faster timebase range.
- Select hf rej trigger coupling. (18)
- Adjust trigger Level.
- With low frequency inputs (below 20Hz), Auto trigger
will initiate triggers overriding the input triggers. Select
Norm trigger.
- Use less sensitive V/DIV range when acquiring trace. (11)
- Re-position trace prior to capture.
(27)
(20)
(7) to get the control master
(8)
(8)
(25)
(33)
(22)
Page 13
A
dvanced Features 2.1-2.1.5
p
2 Advanced Features
2.1 Trigger Control
Trigger facilities discussed here are controllable directly from the front panel. For the more advanced menu-controlled facilities see Section 3.4.
The power-down trigger setting will be retained on future
ower-up.
400 Series Operators Manual
Table
2. 1. 1
Useful Frequency Ranges
Types
Coupling
hfrej 10Hz to 15kHz AC 4Hz to 20M Hz
DC DCt020MHz
2.1.2 Level
The trigger level is the threshold at which the instrument will respond to potential triggers; the trace actually has to pass through the level indicated for a trigger to be valid. See Section
2.1.3.
The level is indicated on the display by two bars, one on each side. of the screen, and is adjusted by the level buttons. For an internal trigger the range is approximately divisions and on external approximately signal is AC coupled, the level bars will be offset from the actual trigger position on the screen by any DC offset present.
2.1.3 Trigger Point (T)
Input
of
Coupling
+/
- 3V. If the trigger
+/-
(22)
10
Figure
2. 1
Trigger Controls
2.1.1 Selecting Source and Coupling
The lowest button in the TRIGGER section of the front panel selects the source of the trigger.
CH1/CH2/Ext/line
of trigger signals. When Ext is selected, the source is the
'EXT TRIG' socket (16) in the lower right corner of the
front panel.
Selecting line is meaningful only if the instrument is powered from the mains. Triggering is then from an internal pulse having a fixed phase relation to the mains voltage waveform. To change this phase relationship, use the trigger delay buttons. See Section 2.1.6.
(17) Steps through the possible sources
The trigger point is indicated on the display by a 'T' near the bottom of the screen underneath the trigger. An arrow next to the T indicates that the trigger point is off the screen.
2.1.4 Slope
A trigger is generated when the selected source signal passes through the chosen trigger level. This transition may be either on a rising or a falling edge. The rising edge is considered to be a positive slope and the falling edge a negative slope.
+/-
(19)This button selects positive
slope triggers.
2.1.5 Trigger Mode (Norm/Auto)
The trigger system operates in either Normal or Auto mode.
In Normal mode, display captures can only occur when a valid trigger input has been received.
In Auto mode, if no valid trigger has been received for some time the instrument will generate its own trigger and initiate a capture. This ensures that the screen is constantly updated irrespective of the input signal. However, if valid input triggers are received at a rate of 20Hz or more, the instrument will start all captures with these triggers and not generate its own.
Auto/Norm (20)This button selects which trigger mode the
instrument is operating in.
Trig'd
receiving valid trigger signals.
(+/-)
(+)
or negative (-)
(21) This LED lights up when the instrument is
hf rej/
AC/DC
(18) Steps through the available trigger cou-
pling options; hf rej is a 15kHz low-pass filter ('high frequency reject'). All the couplings can be used with any source except Line, with which the input coupling is not selectable.
Page 14
400 Series Operators Manual Advanced Features 2.1.6-2.2.1
2.1.6
Trigger Delay
The amount of data which is acquired for display before or after the trigger is determined by the pre or post trigger delay selected, see figure 2.1.6. The value of this delay is shown in the bottom right hand of the trace display.
Pretrig This allows a section of the trace that occurred
before the trigger point to be included in the display. The amount of pre-trigger visible can be set with the front panel PreTrig/Delay buttons. set anywhere between 0% and 100% of screen width, in steps as small as 0.4%.
Trig Delay With this set, the oscilloscope will acquire a trace
following both the specified trigger and a specified delay. The amount of delay is shown beside the 7 key. Delay is controlled by the PreTrig/Delay keys. The size of the steps by which the delay can be incremented or decremented is controlled by the TIME/DIV keys, i.e., step size is 0.04 x the time/ div setting. This can be demonstrated by varying both the timebase and delay settings, while observing the last two lines of the Display and Trigger Menu (See section 3.4). Once set, the delay time remains the same regardless of the selected timebase range.
(23) Pre-trigger can be
(23)
2.2 Capture Facilities
The capture facilities allow the traces to be frozen.
Figure
2.2
The Capture Controls
2.2.1
Trace Hold
There are two ways to freeze the display: a single-shot capture whereby a full trace is acquired then frozen (Section
Figure 3.4a Trace Capture with Delay
Page 15
Advanced Features 2.2.2-2.4.2
s
400 Series Operators Manual
2.2.2), or by pressing one of the buttons labelled Trace Hold.
Trace Hold:
(25) Freezes the relevant trace or traces the moment
they are pressed. A second press releases the trace( s).
2.2.2 Single Shot and Run
These buttons are used to set the instrument into either a capture­and-freeze mode (S/Shot), or a free running continuous capture and re-capture mode (Run).
S/Shot
(30) Arms the instrument for a single-shot triggered
acquisition. The Arm'd light will be illuminated to show that the button has been pressed.
Arm'd
(29) Illuminates after the S/Shot button has been pressed;
it will stay lit until either a valid trigger has been received or until the Run button is pressed.
Stor'd
Run
(28) Illuminates on completion of a single-shot
acquisition. This is after the instrument has been Arm'd, triggered and a complete trace acquired. The light will stay on until the instrument is re-armed by pressing S/Shot again or Run is pressed.
(27) This button puts the instrument in continuous capture
mode: the instrument will automatically rearm itself after each capture is completed so that the display will be updated with each triggered sweep.
2.2.3 Acquisition Status - AFTS
These characters appear on the screen and can be used with the Arm'd and Stor'd lights to determine the 'status of the acquisition in progress.
A Stands for 'Armed'. F Indicates that the acquisition store has been filled with
enough pre-trigger information to be ready for a fresh acquisition.
T Stands for 'Triggered': the instrument has received or
generated a valid trigger; acquisition has begun.
S 'Stored': acquisition is complete.
2.3 Display Modes
socket input to control the Y (vertical) component. It provides an X- Y display of the data captured by the time base and trigger systems. The timebase should be set slow enough to capture the whole of the signal of interest. This mode is not available with magnified traces (Mag).
2.4 Cursor Measurements
The instrument allows you to take direct measurements from the screen display automatically, using inbuilt datum lines and cursor.
Figure
2.4
The Cursor and Datum Line Control
2.4.1 Cursor and Datum Selection
The cursor and datum lines are switched on or off using the Select Trace button.
Select Trace
2.4.2 The Cursor and Datum Lines
There are three lines, as indicated in Figure 2.4.2. Movement of them is aCH1eved using the DATUM and CURSOR buttons, as described below.
(4)
(4) Successive presses of this button places the
cursor and datum lines on to the displayed traces in turn. i.e. channel, channel 2, reference, off, channel etc. If any of the traces are not in use the lines skip to the next valid selection.
The following three display modes are directly selectable using the
Mode button.
Refreshed The instrument will imitate the style of a conventional
Roll This mode is like a chart recorder: the display scrolls from
(31)
realtime 'scope: the display is plotted from left to right as it is acquired.
right to left until a trace has been acquired. The scrolling effect is most noticeable on the slower timebase ranges. The rolling display is frozen by the trigger in S/Shot mode and is unaffected by trigger in Run.
X-V This mode allows the CH1 socket input to control the X
(horizontal) component of the trace, and the CH2
Figure
2.4.2
The Cursor and Datum lines
Page 16
400 Series Operators Manual
A
r
d
dvanced Features 2.4.3-2.5.5
Dat Selecting this option on the Dat/Pos/Mag button (3)
allows the position of the datum lines to be controlled
using the DATUM buttons. (2)
DATUM
(5) Moves the cursor to the right and left along the trace.
Pos/Mag These selections on the Dat/pos/mag button
Pos This option on the Dat/Pos/Mag button (3)
Mag The vertical magnitude of the selected trace can be varied
2.4.3 Making Measurements
The instrument displays, at the bottom of the screen, the time and voltage difference between the intersection of the horizontal and vertical datum lines and the cursor.
The choice of points between which to make the measurements can influence the accuracy: while the measurements made will always be within the design limits, placement of the cursor at the desired position is easier at those points where the slope of the waveform is at its steepest. For example, on a standard sine wave, the most accurate measurements of wavelength are likely to be made if they are taken between two OV crossing points.
(2)
Used for moving the two datum lines. CURSOR
(3)
allow the DATUM buttons to control post storage position and magnitude. They can only be used with stored waveforms. Pressing menu button 1 (opposite the restore message) will cancel the effect of pos and mag on the selected trace i.e. the one with the cursor
X and Y shift to be applied to the trace on which the cursor is placed, using the POST STORAGE/ DATUM
keys. (2)
from X4 to XO.062, using the POST STOR- AGE/DATUM keys. displayed near the top centre of the CRT.
No horizontal magnification is available using this control, but x10 magnification may be applied to all stored traces using the horizontal mag button. See Section 1.5.4).
(2)
The magnification factor is
on
it.
allows both
The normal pen sequence is:
Pen No. Colou
1 Black 2 Blue 3 Green
4 Re
The pen holder is colour coded with this colour sequence. When pens are fitted in this order, trace 1, the border and all the alphanumerics are plotted in black, trace 2 in blue, the reference trace in green and the graticule in red. The cursor and datum lines and their measurements are plotted in the same colour as the trace to which they relate. The cursor colour also applies to the HPGL output on both the 420 and 450.
2.5.3 Pen Changing
Underneath the plotter lid at the front of the plotter there are two small blue buttons. The right hand button when viewed from the front of the OSO is the Pen change button, and the left hand one is the paper advance button.
CAUTION: When changing a pen, extreme care should
be taken not to damage any of the plotter mechanism.
NEVER
move the pen carriage by hand.
To change a pen, ensure that the instrument is switched on and
then proceed as follows:
1. Press the pen change button once. This rotates the pen
carriage and moves the carriage to the right hand side of the plotter.
2. Gently press the white grooved eject lever towards the
front of the instrument. The top pen will be ejected from the pen carriage and can be removed from the plotter.
3. Insert the tip of the new pen into the carriage guide hole.
Gently snap the other end of the pen into place.
4. When the pen has been changed, press the left hand
button - the paper advance button - and the carriage will move back to the left hand side ready for the next plot.
2.5
Plot
2.5.1 Internal Colour Plotter Operation
When the internal colour plotter is fitted to the 400 OSO, plots of the instrument's display can be sent to it by selecting internal on the Plot Menu, see section 3.12.
The internal plotter is automatically selected on power up if it is fitted. If required the plot output can be directed via the RS423 port to an external plotter by selecting RS423 on the Plot Menu, see section 3.12.
2.5.2 Pens
Only use Gould pens - part number 04101175 for a pack of 4; 1 of each colour.
2.5.4 Paper selection
Only use Gould paper - Part number 04101165 for a pack of 8
rolls. Other paper types may damage the mechanism.
2.5.5 Paper Loading
CAUTION: When loading paper, extreme care should be taken
not to force any of the plotter mechanism. paper through by hand, always use the paper advance button to prevent damage to the plotter mechanism.
1. With scissors, cut the end of the paper square.
2. Open the plotter cover by depressing the catch and
lifting the cover until it is fully open.
3. Remove the shaft from the paper cradle and if the plotter
has been used previously remove the plastic or cardboard tube from the shaft.
NEVER
pull
Page 17
A
dvanced Features 2.5.6-2.5.8 400 Series Operators Manual
4. Insert the shaft through the paper roll and fit the roll into
the paper cradle.
5. Feed the paper from the top of the roll as shown in
figure 2.5.5 below and push it through the slot in the rear of the printer mechanism.
6. Press the paper advance button - the left hand blue
button at the front of the printer assembly - until the paper reappears at the front.
7. Feed enough paper through to pass through the top
cover, ensure that it is running around the mechanism smoothly and that it is straight.
7. Close the plotter cover and ensure that it latches shut.
Figure
2.5.5.
Paper Loading
2.5.6
Internal Plotting
When a plot is initiated, the Hold All lamp lights and the message "PLOTTING - PRESS PLOT TO ABORT" appears on the screen. After about 9 seconds the hold All lamp goes out indicating that the data has been transferred to the plot buffer and that the instrument can be used again. The plotting message will remain on the screen until the plot is complete when it will disappear.
In Auto Plot Mode, the instrument is re-armed after the plot data has been transferred to the plot buffer, but the front panel controls will remain in-active during plots.
If the plot button is pressed while the plot is in progress, the plot will be aborted. Any changes to the instrument status during a plot will have no effect on the plot as the data has already been transferred and stored in the plot buffer.
2.5.7 External plots
A plot of the display can be sent to an external plotter by
pressing the plot button @ . The parameters for the plotter
and the plot format can be set as described in section 3.7 and
3.12.
The display is held while a plot is in progress.
Plots can be aborted by a second press of the plot button. If the plot is aborted, the pen will be left in a random position and the external plotter may need to be reset to restore the pen to its start position.
2.5.8
Plot positioning and scaling
The output plot from the 400 has the relative positions shown in figure 2.5.8a. These may be shifted and scaled to position the plot exactly over a pre-defined grid using the following procedure with reference to figure 2.5.8b.
The sides of the pre-defined grid are X and Y, the exact positions of PI and P2 are found by calculating the dimensions A, B, C and D:
Measure the vertical dimension of the required grid, note this as Y.
Measure the horizontal dimension of the required grid, note this as X.
U sing these values of X and Y, A, B, C and D can be calculated as follows:
Measure and mark the exact positions of PI and P2 from the grid edges using figure 2.5.8b and the values obtained above.
The PI and P2 points can now be set on the plotter, refer to the plotter manual for details.
Page 18
400 Series Operators Manual
s
Advanced Features
Figure 2.5.8a Relative Plot Positions
Figure 2.5.8b Plot Dimension
Page 19
A
dvanced Features 2.6-2.7
400 Series Operators Manual
2.6 Data Transfer
Trace data for either CH1, CH2, or the reference trace can be transferred to and from the 400 oscilloscope and a computer.
There are four I/O commands which are used for the data transfer, they are: ST1, STI, ST3, and ST? The first three commands are used to transfer data and the fourth one to interrogate the 400 for its error status.
2.6.1 Syntax
To receive the data for trace one, the computer sends "STl". The 400 then responds with:
STl =20m V, 10us, # HdataCRLF
Where #H is:
#H for hexadecimal number base, #0 for octal number base,
#B for binary number base, or nothing for decimal number base.
Data is: the trace information as described below. CRLF is: carriage return, line feed.
If the 400 receives ST3 and the reference trace is unused, the 400 will respond with "ST3=UNUSED"
Note: When data is being sent to the 400 #D may be used for decimal number base.
For decimal, octal and hexadecimal, the "data" field consists of 501 data points separated by commas.
For binary transfer, the "data" field is as follows:
"lldddd ... ddddcc"
Where:
II
is a 2-byte count of the data that follows including
the 2-byte checksum. This will have the value 01F7 (hex) or 503 (decimal). dddd ... dddd is 501 bytes of trace data. cc is a 2-byte checksum of the preceding data including the 2-byte length.
The most significant byte is sent first for both the length and checksum fields.
Note that there are no comma separators between data bytes during binary transfer.
2.6.2 Receiving Data
The computer may send strings to the 400 at any time in the same syntax as above. The 400 converts the data into a binary form in an internal memory, and when all the data has been collected, transfers the entire memory to the destination trace store.
For decimal, octal, and hexadecimal the 400 will cease receiving as soon as a CR or LF is received, and then only transfer the data received to the required store.
For binary transfer, the 400 will cease receiving when the amount of data indicated by the length field has been received.
Unless the selected trace for received data is held or acquisition is not armed, the selected memory data could be overwritten almost immediately by a fresh acquisition.
Note: The number base for received data is set by the #X field of the incoming string and not by the menu selection. See section 3.7
The strings for the horizontal and vertical ranges must be exactly as follows, anything else will result in error status 102 being generated, as will sending 50ns to a 400 or 420.
50s 20s 10s 5s
0.2s
0.ls 50ms 20ms10ms 5ms 2ms
1ms 0.5ms 0.2ms 0.1ms 50us 20us 10us 5us 2us 1μs 0.5μs 0.2μs 0.1μs 50ns 5V 2V IV O.SY 0.2V 0.1V 50mV
20mV 10mV 5mV 2mV
2s Is 0.5s
2.6.3 Error Status
If the 400 receives "ST?", it will respond with a decimal error status string as follows:
o
No error
96 Invalid command 102 Syntax error 103 Number out of range 104 Length error - Binary transfer only 105 Checksum error - Binary transfer only
If an error condition is detected, the 400 will not make the transfer of data between its internal memory and the destination.
2.7 Real Time Clock (420 and 450 only)
A real time clock is fitted to the 420 and 450 instruments. This enables the current date and time together with the date and time of acquisition to appear on any screen plot, and to be transferred to and from a computer if the data transfer facility is used. See section 2.6. When trace data is transferred to the DSO the date and time are checked to ensure that they are valid and syntax error 102 is returned if they are invalid.
The clock will be kept running for at least ten years, even with the power disconnected, by its own internal battery back-up.
The date and time are set via the special functions menu. See section 3.8.
Page 20
400 Series Operators Manual
3. The Menus
The menu structure is summarised in Figure 3. The numbers 1 to 9 refer to the numeric buttons used to select the particular menus and menu items (Section 3.1.1). All front panel controls remain "live" when menus are displayed, allowing the control status to be changed at any time.
The Menus 3
Figure 3 Menu Overview
Page 21
The Menus 3.1-3.2
400 Series Operators Manual
3.1 Additional Buttons
3.1.1 The Numeric Buttons
The buttons 1 to 7 (36) are used in conjunction with the menu
system to provide a large number of extra functions not otherwise available directly from the front panel. Pressing these buttons when the menus are displayed operates the menu functions described in Sections 3.2 to 3.12. Buttons have extra functions described below.
0, 8
and 9
Figure
3. 1
The Numeric Buttons
3.1.2 0: Menu/Traces
Menu/Traces: (35)Toggles the display between the traces and
the last used menu.
3.1.3 9: Control
Control: (7)
3.1.4 8: Post Store
Post Store: (6) Brings the Post Storage Master Menu onto the
The current display, whatever it may be, is
replaced with the Control Master Menu.
display.
3.2 Control Master Menu
Each entry on the Control Master Menu is also a menu, cov­ering one set of functions. The text is lined up with the numeric buttons 1-7 on the side of the display. To obtain any secondary menu simply press the relevant button.
Figure 3.2 Control Master Menu
The options available on the Control Master Menu are as follows:
STATUS Displays information about the present setup of the
instrument. See section 3.3.
DISPLAY AND TRIGGER Controls probe sensitivity,
averaging, interpolation, the glitch detect function, and trigger delay. See section 3.4.
DISPLAY INTENSITY The brightness of the
alphanumerics, graticule, cursor and traces can all be adjusted independently. See section 3.5.
REFERENCE TRACE Either one of the currently displayed
traces can be copied to the reference trace. See section
3.6.
RS423 INTERFACE Setting the RS423 output protocol.
See section 3.7.
SPECIAL FUNCTIONS Enables the user to disable the auto-
calibration, calibrate either channel, or force a full calibration. See section 3.8.
Page 22
400 Series Operators Manual
The Menus 3.3
3.3 Status Menu
This menu allows the various horizontal, vertical and trigger settings of the instrument to be viewed. A typical display is shown in Figure 3.3.
Figure
3.3
A Status Menu
3.3.1 Mode:
There are three possible display modes:
Refresh, Roll and X- Y. They are described in Section
2.3.
3.3.4 Probe Set: The probe gain settings are shown as x l ,
x10 or x100. These may be set using the Display and Trigger Menu. See Section 3.4.
3.3.5 Timebase: The sweep rate of the timebase is shown in s,
ms,
/-LS
or ns per division.
3.3.6 Trigger This section of the display shows the selected
trigger options. The choices of source are CH1, CH2, Ext and Line. The available couplings are AC, DC and high frequency reject. These are discussed in Section
2.1.
The trigger level is shown in terms of the number of screen divisions above the bottom of the graticule. The trigger slope is shown as either +ve (rising edge) or -ve (falling edge).
Following this is the trigger delay setting, given in s, ms,
/-LS
or ns as appropriate. For pre-trigger operation, the figure is given as a percentage; 0% places the trigger point at the left-hand edge of the screen and 100% on the right-hand edge.
3.3.2 Max/Min: The glitch detector is selected using the Display and Trigger Menu. See Section 3.4.
3.3.3 V/Div:
The input sensitivity of the channels is shown in volts per division. The range is 2mV to 5V per division. If Add mode is selected, a + sign is shown between the channels.
Along with the input sensitivity of the channels, other information is also shown. The symbols used are as follows:
-
Trace inverted
Uncalibrated
>
AC coupled
~
=
Calibrated
Trace off top of display
Trace off bottom of display
Page 23
The Menus 3.4 400 Series Operators Manual
)
f
d
atest value at
3.4
Display and Trigger Menu
This menu controls probe ratios, maximin detection, inter­polation, and the trigger delay function.
3.4.1
Probe Ratio:
can be set for each of the two input channels with successive presses of buttons 1 and 2. The ratios available are xl, x10 and x100. Thereafter, probes of the chosen ratios may be attached to the appropriate inputs; the new sensitivities at the probe tip will be displayed correctly.
3.4.2
MaxiMin: Using the number 3 key, the maximin or glitch
detection function can be switched on or off.
The Max-Min function is designed to detect narrow glitches which may occur between samples. It operates on the signal before it is placed in the acquisition store and can detect very narrow glitches, down to Any glitches detected will at least be displayed as a spike. They will be detected whether they are positive (max) or negative (min). The function works in the timebase ranges of 100 microseconds per division down to 50s per division.
An independent probe attenuation ratio
2f.Ls
wide.
Figure
3.4
Display and Trigger Menu with
Trigger Delay Selected.
3.4.3
Dot Join
On some X-magnified or high speed traces, gaps can be visible between the sample points; if Off is selected, the points alone will be displayed. Selecting On causes the data to be automatically joined by straight lines.
3.4.4
Averaging
ratio of repetitive signals by averaging them. This function, which converts the display into a weighted average of previous acquisitions is toggled On and Off by button 5. Button 6 is used to determine the averaging factor to be used in the algorithm
Button 4 selects the dot-joining function.
It is possible to improve the signal to noise
Where: n =The data point (1 to 501
A =New value at n
an
=
averaging factor (2 to 256)
=
L
=
e.g. In continuous mode, if 8 acquisitions are selected, any single acquisition that deviates from an otherwise consistent level, contributes 1/8 of its deviation to the display. In S/Shot acquisitions contribute to the resultant held display.
only the selected number of
Old value at n
n
3.4.5
Trig Pos’n:
pretrigger percentage or the post trigger delay time. The value can be altered as described in section 2.1.6. Figure
2.1.6 shows the display relative to the trigger position for both pre and delay trigger values.
This displays the current value of either the
Page 24
400 Series Operators Manual
The Menus 3.5-3.6
3.5 Display Intensity
This menu allows independent control over the brightness of various display features.
TRACES/CURSOR: The brightness of the cursor, datum lines
and all traces will be adjusted. The number 1 key increases the intensity and the number 2 key decreases it.
Figure
3.5
Display Intensity Menu
ALPHA: The brightness of all the alphanumerics will be
adjusted. The number 3 key increases the intensity and the number 4 key decreases it.
GRAT: The brightness of the graticule (or grid) will be
adjusted. The number 5 key increases the intensity and the number 6 key decreases it.
3.6 Reference Trace
The menu provides the option of temporarily storing a trace for comparison with the two Channels. It will be lost when it is replaced with an alternative trace.
Figure
3.6
Reference Trace Menu
To copy a currently displayed and held trace into the reference trace, simply press either button 2 or button 3 depending on whether the held trace is in Channel or Channel 2 respectively. A copy of the trace will then be automatically taken. If the trace is not held prior to pressing the button, a fresh acquisition will be made before the copy is taken.
To display a previously stored reference trace, use the button to toggle the selection to On.
Page 25
The Menus 3.7
s
400 Series Operators Manual
3.7 RS423 Interface Menu
This menu allows the parameters of the RS423 output to the remote plotter or other data handling device to be set.
Figure
3.7 An RS423 Interface Menu
Data is always output with no parity, one start bit, eight data bits and one stop bit.
The data that is output is the complete waveforms, including the reference trace, plus information about the timebase and volts per division. The data is output in a form suitable for HPGL plotters, such as the Gould 6120. See section 3.12.
FF represents the top of the screen. The maximum number of characters per trace
transfer is 1522
Binary:
Each data point is sent as a single byte.
00000000 represents the bottom of the screen, 10000000 represents the middle of the screen, and 11111111 represents the top of the screen.
The maximum number of characters per trace
transfer is 525
BLOCK LENGTH: Successive press of button 4 changes the
block length from 0 to 80 in steps of 10 i.e. the sequence 0, 10,20,30,40,50,60,70,80, 0 ... A block length of 0 selects infinite length blocks.
During data output from the 400, if the next value to be sent would exceed the block length, then a CRLF is sent instead of the comma. Block length has no effect on binary transfers, all bytes of data are always sent.
3.7.1 RS423 Plot Connections
The RS423 port is configured at the factory as shown in table
3.7.1. The instrument can be connected to a Gould colorwriter 6120 or a HP7475 plotter using the Gould plotter cable part number 0409-1232. The connections of this cable are shown in figure 3.7.1.
SPEED: This is selected with button 1. The data rate may be
75,150,300,600, 1200,2400,4800 or 9600 baud.
HANDSHAKE: Key 2 is used to select from handshaking Off,
or
XON-XOFF,
NUMBER BASE: Button 3 sequentially selects from decimal,
hexadecimal, octal, and binary as the number base to be used in any data transmission from the 400. See section
2.6
The value range for the data points for each number base is as follows:
Decimal: Each data point is sent as a signed decimal number. ­128 represents the bottom of the screen,
o
represents the middle of the screen, and 128 represents the top of the screen. The maximum number of characters per trace
transfer is 2522
Octal: Each data point is sent as three unsigned octal digits. 000 represents the bottom of the screen, 200 represents the middle of the screen, and 377 represents the top of the screen.
The maximum number of characters per trace
transfer is 2023
Hexadecimal: Each data point is sent as two unsigned hex digits. 00 represents the bottom of the screen, 80 represents the middle of the screen, and
CTS-RTS.
Pin No. Name Description
1 2 RXD Receive Data
3
4 CTS Clear to send 5 RTS Request to send 7 9
Table 3.7. 1 Standard RS423 Pin Connections
Figure 3.7.1 400
3.7.2 RS423 Data Connections
To connect the 400 to a computer use the connections shown in figure 3.7 .2a or figure 3.7 .2b. A prewired cable for each of these arrangements is available; part number 0410-9248 for the PC/XT and part number 0410-1247 for the
OV
TXD Transmit Data
OV OV
to
Gould 6120
Protective Ground
Signal Ground Ground
or
HP7475 Connection
AT.
Page 26
400 Series Operators Manual
The Menus 3.8
Figure
Figure
3.
7.2a 400
3.
7.2b 400
to
IBM-PC/XT Connections
to
IBM-A T Connections
3.8 Special Functions
This menu controls the selection of auto calibration and can force the instrument to perform a full calibration.
Figure
3.8
Special Functions Menu
AUTO-CAL: successive presses of button I enables and
disables the instrument's automatic re-calibration procedure.
When Auto-cal is enabled, the change of any vertical or timebase range will, after a 2 second pause, initiate a calibration of the new operating conditions. If the operating conditions are not changed, a recalibration is initiated at 15 minute intervals, alternating between channels 1 and 2. Each calibration takes about one second.
If the occasional interruption of the normal function of the instrument is unacceptable to the user the Auto-cal facility can be disabled as described above. Recalibration is then recommended 30 minutes after switch on from cold, and additionally if there is a significant change in the ambient operating temperature.
CALIBRATE CURRENT SETUP: Pressing button 2 forces the
instrument to recalibrate for the vertical range and mode and the timebase range that it is currently set at. The actual calibration occurs approximately 2 seconds are pressing the button. This feature does not operate when the instrument is in Roll mode.
FULL CALIBRATION: Button 3 forces the instrument to
perform a complete internal recalibration ignoring existing information. All internal memories are cleared including the instrument status data and any stored traces. To obtain the best results from this calibration, the instrument must have been operating for a minimum
1/
hour. This calibration should only be required if
of
2
there have been repairs to the signal processing sections of the unit or if the internal battery maintaining the status memories has been totally discharged or disconnected.
Page 27
The Menus 3.9
400 Series Operators Manual
REAL TIME
Note: No channel calibrations will be performed if the instrument is armed, in a data capture cycle, in continuous roll mode or if Auto-cal is disabled, they will simply be postponed until these conditions end. Full calibration is always available.
If the unit is unable to complete any of the above self calib­rations. an error message will appear. See appendix 1. These messages should only occur if there is a major fault requiring the unit to be serviced.
When performing a full calibration connect a to the Ext Trig input. This ensures that the trigger input is correctly calibrated for a
CLOCK (420 and 450 only):
clock is set using buttons Menu. The entire date and time is displayed, with one section highlighted. The highlighted element is selected using button 5 and can then be increased or decreased using button 4 or 6 as appropriate.
Once the correct date and time has been set, pressing button 7 enters the data. This makes it easy to set a time that is a few seconds in advance of real time and then press button 7 at a time signal. If the menu is exited without pressing button 7, then the date and time will remain unchanged.
The clock is kept running even with the power discon­nected for at least ten years by its own internal battery back-up.
son
source signal.
The real time
son
termination
3.9
Post Storage Master Menu
The Post Storage Master Menu provides access to three further menus which control the operation of the three battery-backed trace memories, and the plotter.
Figure
3.9
Post Storage Master Menu
Lines 4 to 7 only appear on a 420 or 450 instrument and are the waveform processing functions. Their operation is fully described in section 5.
Page 28
400 Series Operators Manual
The Menus 3.10-3.11
3.10 Save Trace Menu
This menu allows you to store up to three traces, including the reference trace, to the battery-backed memories.
Figure
3.
10 Save Trace Menu
Because these memories are battery-backed, the traces stored in them will not be lost even when the instrument is powered down.
SAVE TRACE: Using the 1, 2 or 3 keys selects either Trace
Trace 2 or the Reference Trace respectively to be saved. Once the trace has been selected, a fresh option will appear, described below.
SAVE TRACE TO MEMORY: Any of the three battery backed
memories may be selected.
1,
3.11 Recall Memory Menu
Traces previously saved using the Save Trace menu can be recalled using this menu.
Figure
3. 11
Recall Memory Menu
RECALL MEMORY One of the three memories may be recalled
by pressing the button 1,2 or 3. Having chosen a memory, you will be asked into which display trace you would like to place it.
RECALL MEMORY TO TRACE Pressing button 5, 6 or 7 will
cause the trace data from the previously selected memory to be displayed on trace 1, trace 2, or the reference trace.
NOTE:
is not armed, the selected memory data could be over-
written soon after it
Unless the selected trace is held, or acquisition
is displayed.
Page 29
The Menus 3.12
3.12 Plot menu
This menu allows the user to set the format of the plot sent to the internal colour plotter or to the external plotter via the RS423 interface. Refer to Section 3.7 for RS423 connections.
400 Series Operators Manual
Figure
3.12
Plot Menu
PLOT MODE: This can be set to either Single or Auto with
successive presses of the number 1 button. In Single plot mode one press of the Plot button wil1 produce one plot output. In Auto mode, a press of the plot button will initiate a plot. The sequence thereafter continues automatically. At the end of the plot the trigger circuit is rearmed for the next acquisition, and at the end of an acquisition, the plot is started. An external plotter must have an auto feed facility to function correctly in this mode. If it does not, then the feed command from the 400 may confuse the plotter.
PLOT WITH GRATICULE: The user has the option of
including the screen grid in the plots with On or excluding it with Off.
GRAT LINE TYPE: A solid or broken graticule can be
selected for the plot, by successive presses of button 3.
PLOT WITH CURSORS: If On is selected, then any plots
made will include the cursor and datum lines. Otherwise, they will be omitted.
PLOT OUTPUT: This line only appears if the instrument is
fitted with an internal colour plotter. The internal plotter is automatical1y selected on power up if it is fitted. If required the plot output can be directed, via the RS423 port, to an external plotter by selecting RS423 with button 5.
ADVANCE PAPER: Pressing button 6 when the internal
plotter is selected causes the paper to be advanced approximately 15mm per press. This line only appears if the internal plotter is selected on line 5.
Page 30
400 Series Operators Manual
Performance Checking 4.1-4.5
4. Performance Checking
The aim of this section is to allow the user of a 400 oscillos-
cope to verify the major analogue performance parameters.
Should any of these fail to be within specification then it is
recommended that the instrument be re- calibrated. A skilled
technician with the aid of the service manual should be able to
perform this task; alternatively return the instrument to your
Gould distributor.
4.1 Risetime
Equipment required:
Fast edge pulse generator <Ins: Bradley oscilloscope calibrator type 192 or Tektronix PG506 or similar; 50 Ohm precision coaxial cable; 50 Ohm precision terminator.
Connect the pulse generator to a channel input through the coax cable and the terminator, which should be at the 400 end of the cable.
1. Select 100mY per division on the input channel and set
the timebase to
2. Set the generator to give 600mY peak to peak at a
repetition rate of 1 MHz and adjust the trigger level to give a stable trace.
3. Select the cursor and change the time base to 500ns per
division.
4. Apply x10 trace magnification using the Mag button to
give 50ns per division, and adjust the horizontal Position buttons to centre the fast edge.
5. Position the cursor and datum lines on the edge such that
the cursor is 60m Y below the high level and the datum lines cross 60m Y above the low level.
The rise time can now be read from the cursor line on the bottom row of text. This figure should not be greater than 18ns.
Suss per division.
4.2 Bandwidth
Equipment required:
Levelled signal generator: Tektronix SG503 or similar; 50 Ohm coaxial cable; 50 Ohm terminator.
Connect the signal generator to a channel input through the coax cable and the terminator, which should be at the 400 end of the cable.
1. Select 5mY per division on the input channel and set the
50jLs
timebase to
2. Set the amplitude of the signal generator to give six
divisions at 50kHz or similar reference frequency.
3. Change the timebase range to 500ns per division.
per division.
4. Increase the frequency on the signal generator until the peak to peak amplitude reduces to 4.2 divisions.
The measured frequency should be greater than 20MHz.
4.3 Trigger Sensitivity
Equipment required:
Signal generator: Tektronix SG503 or similar; 50 Ohm coaxial cable;
50 Ohm terminator.
Connect the signal generator to a channel input through the coax cable and the terminator, which should be at the 400 end of the cable.
1. Select 100mY per division on the input channel and set
50jLs
the timebase to
2. Select Auto trigger.
3. Set the signal generator to 50kHz or similar reference
frequency and adjust the amplitude to give 0.3 of a division peak to peak.
4. Select AC Coupling and Norm on the trigger con-
trols.
It should be possible to find a suitable trigger level to obtain a stable triggered picture.
per division.
4.4 Trigger Bandwidth
Equipment required:
Levelled signal generator: Tektronix SG503 or similar; 50 Ohm coaxial cable; 50 Ohm terminator.
Connect the signal generator to a channel input through the coax cable and the terminator, which should be at the 400 end of the cable.
1. Select 100m Y per division on the input channel and
500ns per division on the timebase.
2. Set the signal generator to 20MHz and adjust its
amplitude to give one and one half divisions on the screen.
It should be possible to adjust the trigger level to give a stable triggered trace.
4.5 Timebase Calibration
Equipment required:
Time calibrator: Bradley oscilloscope calibrator type 192 or Tektronix TG501 or similar;
50 Ohm coaxial cable; 50 Ohm terminator.
Connect the calibrator to a channel input through the coax cable and the terminator, which should be at the 400 end of the cable.
Page 31
Performance Checking 4.6-4.7
400 Series Operators Manual
1. Select a suitable timebase range to view the signal and
set the channel attenuator to give between two and five vertical divisions of signal.
2. Select DC Coupling and Norm on the trigger con-
trols.
3. Adjust the trigger level to give a stable trace.
4. Change the timebase to 500ns per division.
5. Set the calibrator to produce markers every 500ns.
6. Select the cursors.
With the cursor and time datum on identical positions on any two markers, the time difference should read in multiples of 500ns to within 1 %.
Note: Failure of this specification point is indicative of a
major system fault and the instrument should be ser­viced immediately.
4.6 Vertical Calibration
Equipment required:
Oscilloscope calibrator: Bradley type 192 or similar; Coaxial cable.
Connect the vertical calibration output of the oscilloscope calibrator to the 400 through the coax cable.
500ILs
1. Set the timebase to
per division for the input channel.
2. Set the calibrator to give 25mV peak to peak, i.e. 5
screen divisions.
3. Adjust the trigger level to give a stable picture.
4. Switch the cursor on.
per division and select 5m V
The peak to peak measurement should be 25m V to within 3% and the trace should be 5 divisions high to These measurements should be repeated on all the attenuator ranges, each time setting the calibrator to 5 screen divisions for the selected range.
+/-
0.1 divisions.
4.7 Max-Min (Alias Detector)
Equipment required:
Signal generator: Tektronix SG503 or similar; 50 Ohm coaxial cable; 50 Ohm terminator.
Connect the signal generator to an input channel through the coax and the terminator, which should be at the 400 end of the cable.
1. Set the timebase to 500ns per division and select 100m
V per division on the input channel.
2. Set the generator to give approximately 5 divisions at 1
MHz and adjust the trigger level to give a stable trace.
3. Select
4. Carefully adjust the frequency of the signal generator to
5. Select the Max-Min function from the Display and
If Max-Min is functioning correctly and dot join is off then two roughly horizontal lines will be displayed. These will be at the peak levels of the input signal.
20ILs
per division on the timebase. You should notice that the screen picture will change considerably with small changes in input frequency.
produce a sine wave of approximately 2 to 5 cycles. This is an alias.
Trigger Menu.
Page 32
400 Series Operators Manual
Waveform Processing 5.1-5.1.1
5. WAVEFORM PROCESSING FUNCTIONS (420 and 450 only)
All the waveform processing functions are accessed through the Post-Storage Master Menu. Selections 1, 2 and 3 on this menu perform as described in section 3.9 to 3.12.
Selections 4 to 7 provide the waveform processing functions.
When performing any waveform processing that modifies a displayed trace, the new data could be overwritten almost immediately by a fresh acquisition unless the selected trace is held or further acquisitions are prevented e.g. the instrument is in single shot mode and the trace stored or no triggers are provided.
5.1 Cursor Measurements
The 400 DSO has 3 user controlled lines for use with this section of facilities. Their operation is fully described in section
2.4. The cursor is moved along the selected trace using the
◄► cursor buttons (5) the two datum lines are moved using the datum buttons (2) the time datum line with ◄► and the voltage datum with and ▼.
Pressing button 4 on the Post Storage Master Menu changes the display to the Cursor Measurements Menu - figure 5.1a. The cursor is used to determine the trace of interest and the cursor and time datum lines define the portion of the trace to be measured. Most trace measurements are with respect to the voltage datum line.
If any of the calculations are set to either Pulse Width or Risetime, lines 5,6 and 7 are displayed.
The calculation results are displayed in the three lines at the bottom of the trace display; figure 5.1b shows an example. Cursor measurements on timebase speeds of 200 and 100ns/div, i.e. when ETS - equivalent time sampling - is in operation, are only valid on traces that the cursor is on and that are triggered.
Figure
5. 1
b. Example
If all the cursor measurements are set to off, the cursor data line on the display will show the message; "TRI CURSOR
MEASUREMENTS OFF", and if the data is invalid for the
measurement selected, the cursor line will display the message "TRI INVALID DATA". The trace number in both the above messages will change according to the selected trace.
of
Calculations on Trace
Figure 5.1a. Cursor Measurements Menu
Three calculations are possible using buttons 2, 3 and 4. Each calculation is selected from a roll around list containing the following options. Voltage and Time, Peak-Peak, Max-Min, Risetime (falltime), Overshoot (preshoot), Pulse Width, Frequency, Period, Duty Cycle, RMS, Area and off. A calculation can only be selected once e.g. if calculation 1 is set to its default of Voltage and time, then Voltage and time will not appear in the possible selections for calculations 2 or 3.
5.1.1 Voltage and Time
The display shows the voltage between the voltage datum line and the point where the cursor line crosses the trace and the time between the time datum line and the cursor line.
Figure
5.1.1.
Voltage and time Measurement
Page 33
Waveform Processing 5.1.2-5.1.4
5.1.2
Peak-Peak
The display will show the peak to peak voltage difference between the most positive and negative points of the trace between the cursor and the time datum.
400 Series Operators Manual
5.1.4
Risetime (falltime)
Position the cursor and the time datum at the 0% and 100% points of the waveform. The software will calculate the rise or fall time between the low and high percentage points of the waveform.
Figure
5.1.2.
Peak
to
Peak Measurement
5.1.3
Max-Min
The display shows the voltages of the maximum and minimum points of the waveform between the cursor and time datum measured with respect to the voltage datum.
Figure
5.1.4.
Risetime Measurement
The default percentage points are 10% and 90%. Using buttons 5, 6 and 7 these values can be changed. Button 5 selects the point to be changed, either Risetime High or Risetime Low. When the appropriate point is selected, it can be changed by pressing button 6 or 7 as required. Low can be set from 0 to 49% in 1 % steps and high from 51 to 100% also in 1 % steps.
Percentages set are retained even on power down, but are reset to their default vales if a full calibration is performed.
If Pulse Width is also selected as one of the calculations, pressing button 5 steps around 4 entries; Risetime Low, Risetime high, Pulse Width Left and Pulse Width Right.
Figure
5.1.3.
Max-Min Measurement
Page 34
400 Series Operators Manual
5.1.5
Overshoot (preshoot)
Position the cursor and the time datum at the 0% and 100% points of the waveform. The software calculates the overshoot as a percentage of the voltage difference between the 0% and 100% points.
Figure
5.1.5
Overshoot Measurement
Waveform processing 5.1.5-5.1.8
The default percentage points are 50% and 50%. Using buttons 5, 6 and 7 these values can be changed. Button 5 selects the point to be changed, either Pulse Width Left or Pulse Width Right. When the appropriate point is selected, it can be changed by pressing button 6 or 7 as required. Both of them can be set from 5% to 95% in 1 % steps.
Percentages set are retained even on power down, but are reset to their default vales if a full calibration is performed.
If Risetime is also selected as one of the calculations, pressing button 5 steps around 4 entries; Pulse Width Left, Pulse Width Right, Risetime Low and Risetime High.
5.1.
7 Frequency, Period, Duty Cycle
The three parameters frequency, period and duty cycle of a waveform are displayed with this selection. The voltage datum defines the zero crossing voltage. The cursor and time datum must enclose at least three zero crossings.
If the voltage datum does not cross the trace, the mean of the waveform will be taken as the zero crossing line.
The frequency is the inverse of the period calculated above.
The duty cycle is the ratio of the mark or space at the time
datum to whole pulse period expressed as a percentage.
5.1.6
Pulse Width
The software measures the time between the 50% points of a pulse. The time datum and cursor should be positioned on either side of the pulse to be measured. Both positive or negative pulses can be measured.
The first pulse between the time datum and the cursor is measured and if the leading edge of the pulse is noisy, the noise may be taken as a pulse and be measured rather than the pulse of interest.
Figure
5.1.7.
Freq. Period and Duty Cycle Measurement
5.1.8
RMS
This selection calculates and displays both the RMS voltage with respect to the voltage datum and the AC-RMS voltage with respect to the mean of the trace. Both voltages are cal­culated on the waveform bounded by the cursor and time datum.
Figure
5.1.6.
Pulse Width Measurement
Page 35
Waveform Processing 5.1.9-5.2.4
400 Series Operators Manual
5.1.9 Area
This selection calculates the area bounded by the trace and the cursor and datum lines.
Figure
5.1.9.
Area Measurement
The result is given in volt-seconds or similar units.
If the trace falls below the voltage datum then the area will be negative. Positive and negative areas are added algebraically to obtain the displayed result.
5.2 Trace Manipulation
Pressing button 5 on the Post Storage Master Menu changes the display to the Trace Manipulation Menu - figure 5.2. The trace to be manipulated is selected with the cursor and the required function with button 2, 3 or 5.
If no trace has the cursor on it when button 2, 3 or 5 is pressed, the warning message "NO displayed.
Trace manipulations are with respect to the voltage datum line.
After any trace manipulation has been performed pressing button 7 labelled Restore, will return the display to the original stored waveform.
TRACE SELECTED" will be
Figure
5.2.
Trace Manipulation Menu
5.2.1 Filter Frequency
The trace can be filtered by a single low pass filter and the 3dB point displayed on the trace display.
The frequency of the filter applied to the waveform when button 2 is pressed is selected using button 1.
One of six filter stages can be chosen from the roll-around list. The actual frequencies available depend on the current timebase as shown in table 5.2.1.
5.2.2 Filter Current Trace
Pressing button 2 causes the software to perform the filtering.
5.2.3 Invert Current Trace
If button 3 is pressed, the selected trace will be inverted about the voltage datum. The original trace can be 'Restored' using button 1.
5.2.4 Integrate Scaling
To allow an integrated waveform to fit on the screen a scaling factor can be applied to the integrate function. Pressing button 4 steps around the factors xl, xO.5, xO.2 and xO.1. The selected factor is highlighted.
Page 36
400 Series Operators Manual
F2F3F
F
Waveform Processing 5.2.5-5.3
Timebase
50s 111mHz 4504mHz 21.3mHz 1O.3mHz 5.05mHz 2.51mHz 20s
10s 556mHz 227mHz 106m Hz 51.5mHz 25.3mHz 12.5mHz
500ms l1.1Hz 4055Hz 2.13Hz 1.03Hz 505mHz 251mHz 200ms 100ms 55.6Hz 22.7Hz 1O.6Hz 5.15Hz 2053Hz 1.25Hz
50ms 20ms 10ms
5ms 1. 11KHz 455Hz 213Hz 103Hz 50.5Hz 25.1Hz 2ms 1ms
500J.Ls 200J.Ls
100J.Ls
50J.Ls 20J.Ls
10J.Ls
5J.Ls 2J.Ls 1J.Ls
500ns 11.1MHz 4.55MHz 2.13MHz 1.03MHz 505KHz 251KHz 200ns 100ns
(450 only) 50ns
F1
4 F5
278m Hz 114mHz 53.2mHz 25.8mHz 12.6mHz 6.27mHz
1. 11Hz 455mHz 213mHz 103mHz 50.5mHz 25.1mHz
5s
2.78Hz 1.14Hz 532mHz 258mHz 126mHz 62.7mHz
2s
5056Hz 2.27Hz 1.06Hz 515mHz 253mHz 125mHz
Is
27.8Hz 1104Hz 5032Hz 2058Hz 1.26Hz 627mHz
111Hz 45.5Hz 2103Hz 1003Hz 5.05Hz 2.51Hz 278Hz 114Hz 53.2Hz 25.8Hz 12.6Hz 6.27Hz 556Hz 227Hz 106Hz 5105Hz 2503Hz 12.5Hz
2.78KHz 1.14KHz 532Hz 258Hz 126Hz 62.7Hz
5.56KHz 2.27KHz 1.06KHz 515Hz 253Hz 125Hz
11.1KHz 4.55KHz 2.13KHz l.03KHz 505Hz 251Hz
27.8KHz 1104KHz 5032KHz 2.58KHz 1.26KHz 627Hz
55.6KHz 22.7KHz 10.6KHz 5.15KHz 2.53KHz 1.25KHz
111KHz 45.5KHz 2103KHz 1003KHz 5.05KHz 2051KHz 278KHz 114KHz 53.2KHz 25.8KHz 12.6KHz 6.27KHz
556KHz 227KHz 106KHz 5105KHz 2503KHz 12.5KHz
l.l1MHz 455KHz 213KHz 103KHz 5005KHz 25.1KHz
2.78MHz 1.14MHz 532KHz 258KHz 126KHz 62.7KHz
5.56MHz 2.27MHz 1.06MHz 515KHz 253KHz 125KHz
27.8MHz l104MHz 5.32MHz 2.58MHz 1.26MHz 627KHz
55.6MHz 22.7MHz 1O.6MHz 5.15MHz 2.53MHz 1.25MHz 111MHz 45.5MHz 21.3MHz 1O.3MHz 5.05MHz 2.51MHz
6
Table
5.2. 1.
Filter Frequencies
5.2.5
Integrate Current Trace
When button 5 is pressed, the display will show the integrated waveform of the selected trace. The voltage datum is taken as zero for integration and the cursor reads out the value in Vs ­volt-seconds - or similar units.
5.3
Trace Arithmetic
Pressing button 6 on the Post Storage Master Menu selects the Trace Arithmetic Menu, figure 5.
Using buttons 1,2 and 3 the traces involved and the calculation required can be chosen with the selected option being highlighted.
When multiply is selected the scale line next to button 5 appears so that a scale factor can be applied to the calculation result to allow it to fit on the screen.
The result of the arithmetic will be stored in the highlighted trace selected with button 4.
Pressing button 7 - Execute - causes the calculation to happen.
Figure
5.3.
Trace Arithmetic Menu
Page 37
Waveform Processing 5.4-5.4.2
400 Series Operators Manual
5.4 Persistence/Limits Testing
Button 7 on the Post Storage Master Menu selects the Per­sistence/Limits Menu - figure 5.4.
Figure
5.4.
Persistence/Limits Testing Menu
5.4.1
Limits
The upper and lower limits to be tested against are set by displaying the desired limit trace on the screen and then pressing either button 1 or 2 as required. When a limit has been
successfully set, the message LIMIT SET is displayed. Limit
traces may be sent to or read from the 400 with an external computer, using the mnemonics STU for the upper limit and STL for the lower limit, using the data transfer facility described in section 2.6.
The band between the upper and lower limits will be cross­hatched and the limits testing will be performed on all the data points of both the upper and lower limit traces within the area bounded by the cursor and time datum lines.
Limit testing may be turned On or Off or set to Stop-on-fail by pressing button 3. When turned On, each successive acquisition is tested and a warning message displayed when a trace is outside the limits. This warning and trace will be overwritten by the next acquisition.
If the limits mode is set to Stop-on-fail, acquisitions are halted when a trace that is outside the limits is captured,
allowing it to be examined. This screen picture can be plotted, and with the internal colour plotter, the upper and lower limits will be in the same colour as the reference trace to distinguish them from the failed trace.
During limits testing the trigger level and position markers are not displayed.
Test Limits and Display Limits cannot be turned on if the DSO is in X- Y mode or if X-Mag is turned on. Similarly, XMag or X - Y mode cannot be selected if either Test Limits or Display Limits is on. If any of these situations occur one of the following messages will be displayed as appropriate.
NO LIMITS DISPLAY IN XY
NO XY WHEN LIMITS DISPLAYED NO LIMITS DISPLAY IN X-MAG NO LIMITS TESTING WITH X-MAG NO X-MAG WHEN LIMITS DISPLAYED NO X-MAG WITH LIMITS TESTING
Also, Display Limits cannot be turned on if persistence is on.
Trying to select this mode results in the message; NOT WHEN PERSISTENCE.ON.
A record of all failing traces can be automatically plotted by setting the limits testing mode to stop on fail, and selecting Auto Plot from the plotter menu. Any acquired trace that then falls outside the limits will be plotted. If the plot button is not pressed to initiate a continuous sequence, only the first failing trace will be acquired and plotted.
5.4.2
Persistence
The 400 is not capable of variable persistence but either the number of acquisitions accumulated before the display is cleared or the absolute time between display clearances can be set from the Persistence/Limits Menu.
The choice of Persistence for time or number of acquisitions is made using button 5 with the actual time of number of cycles set by button 6 or 7 as appropriate.
When using the persistence mode, the trigger level and position markers are not displayed.
If Display Limits is on, then persistence cannot be selected and
the message NOT WHEN LIMITS DISPLAYED is shown if
button 5 is pressed.
Page 38
400 Series Operators Manual
Battery Unit Operation 6.1
6.0 Battery Unit Operation
The 400-107 battery unit option for the 400 series DSOs provides the instruments with a fully automatic built in Nickel Cadmium battery and charger which allows uninterrupted operation of the DSOs in the event of an AC supply failure and complete operation independent of an AC supply.
The battery is fully protected against damage from overcharge or over-discharge and automatically receives a fast recharge followed by a continuous trickle charge when AC power is restored or re-applied to the unit.
A DSO with the battery option operates as an integral unit. All the normal functions of the DSO with the exception of line trigger are available and function according to the main operators manual. Full operation of the DSO is also available while the battery is recharging.
In the following text, battery unit refers to the option and DSO refers to the main oscilloscope. The circled numbers in this section refer to the numbered controls in figure 6.2
An AC supply is applied to the battery unit. This charges the batteries and is linked through the battery unit to power the DSO. The battery unit then also provides a DC supply to the DSO for use in the absence of an AC supply.
The battery unit is electrically connected to the DSO by two rear panel cables; one for the AC supply and one for the DC supply.
CAUTION: Before use ensure that the connections between
the battery unit and the
DSO are correct, see figure 6.0.
6.1 AC Supply Operation
To operate either the DSO or the battery unit the master on/off
(37)
switch
If the instrument has been unused for a period of a month or more at 20°C or for a week or more at 50°C, individual cells within the batteries may have lost different amounts of charge. 24 hours operation on charge is required prior to discharge.
With an AC supply applied to the battery unit, and the master power switch section 6.3.
When the DSO power is also switched on, the DSO operates from the AC supply. Normal operation of the oscilloscope is possible, and the batteries remain on charge.
on the front of the battery unit must be turned on.
(37)
turned on, the batteries are on charge. See
Figure
6.
Battery Unit Interconnections
Page 39
Battery Unit Operation 6.2-6.5
6.2
DSO Battery Operation
400 Series Operators Manual
Limited battery capacity is available to operate the DSO if the AC supply is removed for any reason during fast charge. After full discharge, the battery capacity will be approximately half the elapsed recharge period.
If the unit is used outside the temperature range of + 15°C to + 35°C, the normal operation may be limited as discussed in section 6.5.
6.4
Alarm Conditions
Figure
6.2
Battery Unit Front panel indicators
If no AC supply is available, the DSO can be powered from the battery unit simply by turning both the Master power switch
(37)
and the DSO power switch use on an AC supply that is interrupted, the operation of the DSO is uninterrupted as it changes to DC operation.
Whenever the batteries are supplying power to the DSO the
(41)
Using battery lamp
When the batteries approach full discharge their output is automatically cut off to prevent damage due to deep discharge. In this state no lamps on the battery unit will be illuminated.
Similarly, there will be no lamps illuminated if there is no AC supply and the DSO is switched off, whether or not the battery unit is switched on.
6.3
Battery Recharge
To recharge the batteries, the master on/off switch the front of the battery unit must be on and AC power supplied to the rear socket.
After any period of battery operation the battery unit will power up in fast charge mode when the AC supply is restored or re-applied. After a time period dependant on the discharge time, which allows at least 80% of the charge withdrawn to be replaced, the unit reverts to standby charge mode. This replenishes approximately 5% of the total battery capacity per hour, and is maintained continuously to hold the batteries in the fully charged state.
The batteries will be charged by the AC supply whether or not the DSO itself is turned on.
When AC is applied, either indicator will be illuminated to show the charge mode.
(39)
Indicator indicator
No harm can be caused to the batteries by leaving them on standby charge for indefinite periods, although if they have been left in this mode for a long time, one cycle of full dis­charge and fast recharge may be required to obtain maximum capacity.
shows that the battery is on fast charge, and
(40)
shows that the battery is on standby charge.
is illuminated.
(1)
on. If the DSO is in
(39)
or indicator
(37)
(40)
on
The alarm lamp power from the batteries while an AC supply is present. i.e. when the DSO should operate from the AC supply. This condition only occurs if the interconnecting AC cable is not connected or the DSO's AC supply fuse is open.
Once this condition is detected, the battery supply to the DSO is automatically switched off.
When the fault has been remedied, the alarm condition can be cleared by turning the master power switch least 10 seconds and then on again.
The alarm lamp battery temperature trip operates - see section 6.5 - or if a severe fault blows the internal battery protection fuse. Replacement of this fuse requires removal of the instrument covers and it should only be attempted by suitably qualified personnel. See section 1.1.3.
6.5
Extreme Temperature Operation
(38)
will flash if the DSO attempts to draw
(37)
(38)
will be illuminated constantly if the
off for at
and Storage
The DSO may be powered from the battery unit at any ambient temperature up to +45°C. However, the internal battery temperature detector may inhibit full fast charge at temperatures above 35°C. In this condition, both the standby charge lamp constantly.
Charging is interrupted until the battery temperature reduces naturally to below the trip temperature, when the alarm lamp will be extinguished. Standby charge will then be resumed. At least 24 hours standby charge may then be required to regain full charge.
Although the battery unit may be stored at temperatures up to 50°C, and operated at temperatures up to 45°C, it is recommended that the period of such usage is limited. Pro­longed exposure to high temperatures will adversely affect the battery life.
The battery unit may be used or recharged at temperatures down to O°C but repeated recharge below 15°C may adversely affect the battery life.
The batteries will not be damaged by storage in the discharged state, however, it is recommended that they are fully charged before being unused for long periods. 24 hours operation on charge is recommended before use, after the unit has been unused for more than a month at 20°C or a week at 50°C.
(40)
and the alarm lamp
(38)
will be on
Page 40
400 Series Operators Manual
6.6 Independent Operation of DSO
In any circumstances if the battery facility is not required the AC supply link from the battery unit to the DSO can be disconnected and an AC supply fed directly into the DSO. In this mode, the master power switch DC cable disconnected from the DSO.
This mode of operation allows the use of the DSO on an AC supply above the 45°C limit for battery operation, but below the 50°C limit for the DSO alone.
Similarly, if the DC cable from the battery unit is removed from the DSO, an external DC supply can be used to power the DSO. For example, a high capacity battery could operate the DSO remote from an AC supply for long periods. The batteries cannot be re-charged from an external DC supply.
(37)
must be off or the
Battery Unit Operation 6.6
Page 41
A
lphabetical Summary
400 Series Operators Manual
7. Alphabetical Summary of Controls
AC As part of AC/DC/Gnd, AC removes DC components
from signals of between 4Hz and 20MHz. As part of hf rej/ AC/DC it sets AC coupling for the trigger signal,
which may be between 10Hz and 20MHz.
AC/DC/Gnd
Add
Arm'd
AUTO SETUP
Auto/Norm
CH1/CH2/Ext/Line CH1(X)
being for the connection of signals up to +/ -400Y peak to Channel. It is used for the X component of an X- Y mode trace. It can also be a trigger source if selected using
CH1/CH2/Ext/Line.
CH2(Y)
CURSOR
Control
Dat
(14)
Controls the type of coupling between the
instrument and the input signal. AC is used to remove
DC components from signals of between 4Hz and
20MHz. DC couples the input signal directly to the
instrument, so all frequency components of the signal up
to 20MHz will be displayed. Gnd internally disconnects
the inputs from the instrument; a 0V reference signal is displayed instead.
(10)
Displays the sum (or difference if one channel is inverted) of the channels. The original traces disappear and the resultant trace is displayed as a new Channel 2 trace.
(29)
Illuminates after S/Shot is pressed; it will stay lit
until either a valid trigger has been received or the Run button has been pressed.
(24)
Will attempt to arrange the display so
that two to five complete cycles appear, with the amplitude set so that the height of the trace is between
two and five screen divisions. Also, it selects Auto
trigger to ensure that the screen is frequently updated and a trace will be visible. If the frequency of the input signal is less than 20Hz, this function may not operate correctly.
(20)
Selects the trigger mode. Norm means that
valid triggers must be received to initiate captures. Auto triggering is the same as Norm except that if no valid
trigger has been received for some time, an artificial trigger will be generated: the instrument will generate its own triggers if no valid trigger has been received for
0.05s (i.e. 20Hz is the lower limit).
(17)
Steps through trigger sources.
(9)
One of two channel signal input sockets, this one
(12)
As CH1, but CH2 is used for the
of an X- Y trace, where CH1 is the X component.
(7)
These two buttons control the movement of the
cursor.
(2) Selects the control master menu.
Part of Dat/Pos/Mag
AGE/DATUM and CURSOR keys to be used to con-
trol the movement of the cursor and datum lines.
(3)
. Allows the POST STOR·
Y
component
DC As part of AC/DC/GND, DC couples the input signal
directly to the instrument so that all frequency com-
ponents from DC to 20MHz are displayed. As part of hf rej/AC/DC it couples triggers from DC to 20M Hz to
the instrument.
Ext
External trigger source (connected via EXT TRIG
socket).
EXT TRIG
Gnd
hf rej/AC/DC
Socket
(16)
for the connection of external triggers
of up to +/-250Y peak.
Part of AC/DC/Gnd, Gnd disconnects the input signal
from the instrument and displays a 0V reference signal instead.
(18)
Selects trigger coupling. The hf rej (high
frequency reject) option is a 15kHz low-pass filter. Any coupling may be used with any trigger source except
Line where the coupling is not selectable.
Inv Part of Off/On/Inv; this option is used to invert the
display of a channel's trace - i.e. to negate it so that it is displayed upside down.
Level
(22)
Adjusts the trigger level, which is indicated by
two trigger bars on the display.
Line
Mains frequency triggers. Input coupling may not be
selected with this option.
Mag
(32)
Used to control the horizontal magnification of the
traces; when selected, a xl0 expansion around the centre of the screen is applied to the traces. The timebase setting is altered to reflect the expansion.
Mag
part of Dat/Pos/Mag
nitude of a stored waveform to be varied using the
POST STORAGE/DATUM keys.
Menu/Traces
Mode
Norm
ns
(33)
Off/On/lnv
channel is on, its trace can be displayed Inverted (upside down).
Plot
(34)
(35)
menu and the trace display.
(31)
Selects one of three display modes. Refreshed mode is the normal oscilloscope mode whereby the trace is plotted from left to right. Roll mode is like a chart recorder: the display scrolls from right to left as the trace
is acquired. X-Y mode takes the CH1 (X) input as the x (horizontal) component and the CH2(Y) input as the y
(vertical) component.
Part of Auto/Norm. The instrument will only initiate captures when a valid trigger is received when in Norm mode.
Decreases the horizontal TIME/DIV.
(13)
Switches the channel On or Off. If the
Produces a plot of the current display on a suitable
plotter. If a plot is in progress, pressing this button will abort it.
Pos Part of Dat/Pos/Mag
the cursor is placed to be shifted using the POST STORAGEIDATUM keys.
Pos'n
(8)
Moves that channel's trace vertically.
Position Post Store POWER
(26)
Moves all traces horizontally.
(6)
Selects the post storage master menu.
(1)
Used to switch the instrument on or off.
(3)
this allows the vertical mag-
Swaps the display between the last used
(3)
this allows the trace on which
Page 42
A
400 Series Operators Manual
lphabetical Summary
Pre-Trigger
Refr Refreshed mode is the normal oscilloscope mode whereby
Roll This display mode operates like a chart recorder: the
Run
(33)
Sec Select Trace S/Shot
Stor'd
TIME/DIV
(23)
Used in conjunction with the menus to
control the degree of pre-trigger.
the trace is plotted from left to right.
display scrolls from right to left as the trace is acquired.
(27)
Puts the instrument in continuous capture mode; the instrument automatically re-arms itself after each acquisition.
Increases the horizontal TIME/DIV.
(4)
Places the cursor onto a displayed trace.
(30)
Arms the instrument for a single-shot
triggered acquisition sweep (capture). The Arm'd light
will be illuminated to show that this button has been pressed.
(28)
Illuminates on completion of a single-shot
acquisition. It will stay lit until the instrument is re-
armed or until Run is pressed.
(33)
Controls the sweep rate of the trace. The timebase can be varied from 100ns/div to 50s/div in a 1,2,5 sequence of values.
Trace Hold
Trig'd
Uncal Controlled by the Var button. When lit, the coarse
Var
V/DIV
X-V mode takes the
+/-
(25)
Freezes the relevant trace or traces until
pressed again.
(21)
This lights up when the instrument is receiving
valid triggers.
setting of the attenuator remains unchanged, but a variable gain is applied to the input signal. This control applies attenuation in the range of 1 to 0.4. Thus, with an initial setting of 1 V, the actual sensitivity of the channel may be set anywhere between 1 V and 2.5V per division.
The V/DIV buttons are used to vary the sensitivity.
(15)
Toggles between Uncal
(11)
Adjusts the sensitivity of the instrument over
discrete calibrated ranges from 2mV/div to 5V/div in 1, 2, 5 steps. With a x10 probe the ranges are 20m V/ div to 50V/div at the probe tip. Input voltages must not exceed
+/-400V peak. If the Uncal light is on, then these
buttons control the sensitivity over an uncalibrated continuous range of values.
CH1(X)
component and the component.
(19)
Selects triggering on positive or negative slopes.
&
Cal.
input as the x (horizontal)
CH2(Y)
input as the y (vertical)
Page 43
A
ppendix 1 Error Messages
400 Series Operators Manual
Appendix 1: Error Messages
When something is wrong, an error message will be displayed in inverse text near the bottom of the display. Also in some cases a message appears to confirm that an action has occurred. Brief explanations of when these messages occur are given below in alphabetical order.
1. CALIBRATION DISABLED
If Calibrate channel or 2 is pressed when Auto-cal is disabled.
2. LEVEL FIXED WHEN SOURCE=LINE If the trigger level controls are touched when line trigger
is selected.
3. NO ADD IN XY MODE
If trying to select ADD when in X- Y mode.
4. NO ADD WHEN HELD If the add button is pressed when either or both CH1 or
CH2 are held.
5. NO BINARY WITH X-ON X-OFF If X-ON X-OFF is selected and binary number base is
also selected.
6. NO CURSORS IN XY If either the Datum or Cursor position keys or the select
trace button is pressed when in X-Y mode.
7. NO HORIZONTAL MAG If the post storage <l or l> keys are pressed when the
cursor and X-MAG are on.
8. NO LIMITS DISPLAY IN XY
If limit display is turned on while the instrument is in X­Y mode.
9. NO LIMITS DISPLAY IN X-MAG
If limit display is turned on while the instrument is in X­MAG mode.
10. NO LIMITS TESTING ON REF TRACE If limit testing is turned on while the cursor is on the
reference trace.
11. NO LIMITS TESTING WITH X-MAG If limit testing is turned on while the instrument is in X-
MAG mode.
12. NO POST STORAGE IN XY If the cursor is selected and then the unit is switched to
X-Y mode and then a post storage key is pressed.
13. NO TRACE SELECTED If the cursor is off and either the restore, cursor position
or post storage movement keys are pressed.
14. NO X-MAG IN ACTIVE ROLL a) If the X-MAG key is pressed while the instrument is
in ROLL mode on timebase ranges slower than 20ms/div, with the traces not stored or either channel not held.
b) If in ROLL mode on timebase ranges faster than 50ms/div with continuous acquisitions and X-MAG on, and then the timebase range is changed to be 50ms/div or slower.
15. NO X-MAG IN XY MODE If X-MAG is pressed when in X-Y mode.
16. NO X-MAG WHEN LIMITS DISPLAYED If X-MAG is pressed while display limits is on.
17. NO X-MAG WITH LIMITS TESTING If X-MAG is pressed while the instrument is in limits testing mode.
18. NO XY WHEN LIMITS DISPLAYED If XY is pressed while display limits is on.
19. NO XY WHEN CHANNELS ADDED If trying to select X-Y mode when channel ADD is
20. NO XY WITH X-MAG If trying to select X- Y mode when X-MAG is on.
21. NOT WHEN LIMITS DISPLAYED If persistence is turned on while display limits is on.
22. NOT WHEN PERSISTENCE ON
If display limits is turned on while the instrument is in persistence mode.
23. PLOTTING - PRESS PLOT AGAIN TO ABORT Appears whilst plot is in progress.
24. TR1 CURSOR MEASUREMENTS OFF
If all three cursor measurement options are set to off, this message appears.
25. TR1 INVALID DATA
If the data is invalid for the measurement selected.
26. TRACE COPIED
To confirm that a trace has been copied to the reference trace.
(Note also that if a trace is copied to the reference trace, the reference trace is automatically turned on as well).
27. TRACE MEMORY UNUSED
a) If trying to turn the reference trace on when it has not been previously copied/saved.
b) If an attempt is made to recall a backup memory when that memory has not been previously saved.
28. TRACE NOT YET STORED
If the cursor is selected and a post-storage key pressed but the trace is not stored or held.
29. TRACE RECALLED
To confirm that a trace has been recalled.
30. TRACE SAVED
To confirm that a trace has been saved.
31. USE CH1 POS'N KEYS IN XY
If either horizontal position key is pressed when in X-Y mode.
32. WAITING FOR TRIGGER
If S/SHOT has been selected but no trigger has been detected within 2s.
on.
Page 44
400 Series Operators Manual
A
ppendix 1 Error Messages
33. X-MAG RESTORED WHEN STORED/HELD
If the traces are stored or held in ROLL mode. Then if X-MAG is selected and either S/SHOT or RUN is
pressed, or either channel unheld.
i.e. X-Mag has been turned off while the display is in
active roll mode, but it will be restored when the display is once again stored or held.
34. O.IILS TO IOOILS/DIV, NO MAX-MIN If max/min is selected when timebase is faster than 100μs/DIV.
Page 45
Appendix 2 Specification
A
400 Series Operators Manual
Appendix 2: Specification
DISPLAY CRT
Graticule
Intensity
VERTICAL SYSTEM
Two identical channels, CH1 and CH2. Inputs via BNC connectors
Sensitivity Accuracy
Variable Sensitivity
Input Impedance Input Coupling
Bandwidth
400and420: 450:
Input protection Expansion
HORIZONTAL SYSTEM
Sweep rate
400 and 420: 450:
Transient capture Repetitive sampling:
400 and 420: 450:
Sample rate accuracy: Expansion:
TRIGGER DELAY
Trigger delay range: Trigger delay accuracy: Pre-trigger: Resolution:
TRIGGER SYSTEM
Variable level control with Auto/Normal facility,
resolution of less than 0.1 div.
:
5 inch diagonal screen Raster scan, vertically scanned
:
Electronically generated 8 x 10 divisions with
0.2 sub divisions
:
Separate controls for Traces, Graticule and
Alphanumerics
:
2mV/div to 5V/div in 1-2-5 sequence
:
+/- 2.5% of reading +/- 1 digitising level
(1/30 of a division)
:
>2.5:1 range allowing continuous
adjustment of sensitivity between ranges
:
lMnl28pF
:
DC-GND-AC
:
DC: 0-20MHz (-3dB) AC: 4Hz-20MHz (-3dB) DC: 0 ­50MHz (-3dB) AC: 4Hz ­50MHz (-3dB)
:
400V DC or peak AC at 10kHz or less
:
Post storage x 0.062 to x 4.00
(Timebase)
:
27 ranges in 1-2-5 sequence
28 ranges in 1-2-5 sequence
:
500ns/div to 50s/div.
200 and 100ns/div
200,100 and 50ns/div
+/- 0.01 % of sample time
x
10 with linear dot interpolation
20ns to 5000s
+/-
0 to 100% of sweep in 0.4% steps
2% of time/div, 20ns min
0.01
%,
+/-
1
ns
uto/Normal Mode:
when insufficient signal (20Hz - 20M Hz) is present or when the selected level is outside the range of the
input signal.
Source: Coupling: Slope:
Sensitivity 400 and 420:
Internal DC Coupled <0.3 div DC to 2M Hz
External DC Coupled <150mV DC to 2MHz
Sensitivity 450:
Internal DC Coupled <0.3 div DC to 2MHz
External DC Coupled < 150m V DC to 2MHz
Range:
External input impedance: External input protection: Trigger jitter:
DISPLAY MODES
Refreshed:
Roll:
Dot join:
X-Y: X-Y display is 8
Single trace: Dual trace:
CH1, CH2, External or Line
DC, AC, high frequency reject filter
+ve or -ve
AC Coupled <0.3 div 10Hz to 2MHz
AC Coupled <150mV 10Hz to 2MHz
Internal +/- 10 divisions External +/- 3V
(unexpanded), +/- 2ns
0.2fLS/Div and 0.lfLs/Div, +/- 2ns
Stored data and display updated by triggered
sweep
Stored data and display updated continuously for
timebases 50ms/div to 50sec/div. Trigger stops the updating process
Refresh and roll operate as repetitive single shot for timebase ranges faster than 50ms/div
Dots are joined by vertical raster lines. Linear dot interpolation is provided when the trace is X­expanded
are updated by triggered sweep. There is no dot joining, used as the X (8 bit resolution 25 steps/div) and CH2 as the Y (7 bit resolution 15 levels/div) deflection
x
CH1 or CH2
CH1 and CH2
In Auto the timebase free runs
<1.5 div DC to 20MHz
AC Coupled <0.3 div 10Hz to 2 MHz
<1.5 div 4Hz to 20MHz
<600mV DC to 20MHz
AC Coupled <150mV 10Hz to 2MHz
<600m V 4Hz to 20MHz
<1.5 div DC to 50MHz
<1.5 div 4Hz to 50MHz
<600mV DC to 50MHz
<600mV 4Hz to 50MHz
100K!1/10pF
250V DC or peak AC
50s/Div to 0.5fLs/Div, +/- 2% of time/Div
x
8 divisions. Stored data and display
10 expansion or cursor in this mode. CH1 is
Page 46
A
400 Series Operators Manual
ppendix 2 Specification
Add: CH1 and CH2 can be added to give the algebraic sum of
the two channels. Addition is post storage
Invert: Both channels may be independently inverted Single
Freezes store at the end of a single triggered sweep
Shot: Display trace hold:(all) Freezes the display immediately
Channel 1 Trace hold: Freezes channel 1 display
immediately
Channel 2 Trace hold: Freezes channel 2 display
immediately
Reference trace: One reference trace can be displayed in
addition to the two input channels. This can display a waveform memory of a trace copied from CH1 or CH2
ACQUISITION SYSTEM Maximum sample rate:
ously on each channel
Vertical resolution: 8 Bits (1 in 256) 30 levels per division Record length: 501 points per channel
ACQUISITION MODES
Normal mode:
(Repetitive capture is only on timebase ranges 100 or 200ns/div, which gives an equivalent sample rate of 2ns/sample on the 100ns/div range.)
X-Y Mode: Bandwidth 20MHz (-3dB) 50MHz on 450.
Phase difference <4° @20MHz. Acquisition rate dependent on the timebase range.
Averaging: Averages can be set from 2 to 256 in binary
sequence, selected from the menu system. Averaging operates continuously or, using single shot for the set number of acquisitions. (weighted average)
Peak detection: Minimum pulse width Zus, 100% probability
of capture. Operates on timebase range 100p..s/ div or slower
MEMORY
Waveform memory:
waveform data storage. These memories are Nonvolatile
Set-Up: The control set-up is retained on memory during
power down
Retention time: The memory support is trickle charged and
will retain information for 1 month after power down
ON SCREEN MEASUREMENTS and ALPHANUMERIC DISPLAY
Datum Lines:
time and voltage datums
Horizontal and vertical full screen amplitude
100 megasamples/sec simultane-
Transient and repetitive signal capture.
3 reference memories are selectable for
Cursor: The measurement cursor can be assigned to a trace
and measurements made in time and voltage with respect to the datums
Cursor measurement display:
displayed on screen
Accuracy: Voltage
(1/30 of a division) Time
Resolution: Voltage 0.4% of F.S.D.
Time 0.2% of F.S.D.
Trigger Indication: On screen trigger level and trigger point
indication
Alphanumerics: Display on screen to indicate vertical sen-
sitivity and input coupling for each channel, timebase speed and pre-trigger or trigger delay. Arrow for off screen indication of trigger point and traces
MENU SELECTION
Control master menu:
Status, display and trigger facilities, display intensity, reference trace control, RS423 interface, and special functions
Menu/traces: Alternately switches between trace display and
the last menu selection
Post storage master menu: Selects menus for fast access
to save/recall trace and plot output parameters
AUTO SET UP
Automatically sets the front panel controls to display any applied repetitive input signal for frequencies greater than 20Hz. Trigger and timebase priority is CH1, CH2 and Line
RS423 INTERFACE
Serial interface port for bi-directional waveform data and associated range parameter transfer
Baud rate: 75, 150,300,600, 1200,2400,4800,9600 Data bits: 8
Parity: None
Start-stop:
DIGITAL PLOTTER OUTPUT
The instrument can directly output to suitable HPGL format plotters via the RS423 interface port
Plot mode: Manual or auto selection to output a stored trace Annotation: Range and scaling annotation, graticule , cursors
and cursor readout can all be included in the plot output
Colors: Color pens automatically selected when
available
+/-
2.5% of reading,
+/-
0.01%, of reading
Fixed, one start bit, one stop bit
6
Voltage and 6 Time
+/-
1 digitising level
+/-
1 digit
Selects menus for prime functions:-
Page 47
Appendix 2 Specification
400 Series Operators Manual
Internal Color Plotter (420 and 450 only)
Paper type: Gould part No. 04101165 (8 roll pack). Paper length:
Pen type: Gould part No. 04101175 (4 color pack). Pen life: Minimum of250 metres line length approximately
55 metres, equivalent to approximately 300
plots.
175 plots.
WAVEFORM PROCESSING FUNCTIONS (420 and 450 only)
Cursor Measurements
Voltage and Time
The cursor measures the voltage and time relative to the voltage and time datum lines respectively.
Peak-Peak
Calculates peak-to-peak voltage of the waveform bracketed between the time datum and cursor.
Max-Min
Displays maximum and minimum voltage excursions of a waveform relative to the voltage datum position. The cursor and datum bracket the waveform of interest.
Risetime (falltime)
Calculates the rise or fall time between the 10% and 90% points of a signal; The 0% and 100% points are set by the cursor and time datum.
Overshoot (preshoot)
Calculates the over or pre shoot of a signal as a percent of the 100% point; The 0% and 100% points are set by the cursor and time datum.
Pulse Width
Calculates the time between 50% points of a pulse, with the pulse being bracketed between the time datum and the cursor.
Frequency, Period, Duty Cycle
Calculates the average frequency, period and duty cycle of a waveform. Uses either the voltage datum as the zero crossing point or the mean of the waveform. The cursor and time datum set the limits of the area of interest.
RMS
Calculates the root mean square (RMS) voltage of a waveform bracketed between the cursor and time datum. The voltages are calculated with respect to both the voltage datum (DC-RMS) and the mean of the waveform (AC-RMS).
Area
Calculates the area under a waveform bracketed bet­ween the cursor and the voltage and time datum lines.
Trace Manipulation
Filter
6 selectable stages of low pass filter for each timebase range.
Invert
Inverts the trace about the voltage datum.
Integrate
Calculates the indefinite integral and displays the resultant waveform. A scaling factor of xl, x0.5 , x0.2 or x0.1 can be applied to the integrated waveform.
Trace Arithmetic Add
Adds any two traces and displays the result in a selected trace.
Subtract
Subtracts any two traces and displays the result in a selected trace.
Multiply
Multiplies any two traces and displays the result in a selected trace. A scaling factor of xl, x0.5 , x0.2 or x0.1 can be applied to the multiplied waveform.
Limits testing
Tests a waveform with respect to two definable limit waveforms and will display a TEST FAILED message if the acquired signal fall outside the limits. Acquisition can be continuous or stop-on-fail.
Persistence
Either the time, or number of acquisitions between screen clearances can be set to give a persisted trace display. Times of 0.5, 2, 5, 15, 30 or 90 seconds or multiple acquisitions of 10, 20, 50, 100, 200 or 500 traces can be selected. Both options can be set to infinity to give a permanent persistence display.
REAL TIME CLOCK (420 and 450 only)
24 hour time and date set via a menu. Printed out on plots to give a permanent record of acquisition time. The time and date is maintained for approximately ten years with the instrument's power disconnected.
BATTERY UNIT 400-107
Operation of DSO from fully charged battery:
1.5 hours minimum.
90% recharge time from full discharge:
Battery life to 70% capacity:
200 complete discharge cycles at 20°C 100050% discharge cycles at 20°C.
AC supply range: 90 to 130V or 180 to 265V, 45 to 400Hz. Indication provided for:
Battery in use (discharging) Battery on fast charge Battery on standby charge Alarm conditions
5 hours maximum
Page 48
400 Series Operators Manual
A
ppendix 2 Specification
ENVIRONMENTAL
Temperature:
Operating:0°C to +50°C. (0°C to +45°C with battery unit) Storage: -l0°C to +70°C. (-l00C to +50°C with battery unit)
Humidity:
Non-operating IEC 68-2-Db cycling 25°C to 45 °C with 95% RH, 6 cycles (144 hours)
Vibration:
5-500Hz at 1g rms for 15 minutes
IEC 68-2-6 Test Fc. 15 cycles of I minute duration 10 to 55Hz at 0.6mm peak to peak displacement in each of the three major axis (4g at 55Hz)
Safety EMC:
EMI to BS6527, VDE 0871 Class A ESD
10kV to IEC 801-2, BS6667-2
MISCELLANEOUS
Calibrator:
KHz calibration signal on front panel
Ground:
POWER REQUIREMENTS AC Voltage:
Frequency:
option.
DC Voltage: Power:
Full Specification: + 15°C to + 35°C.
Operating IEC 68-2-Ca at 45°C with 95% RH.
MIL spec 8100. Random frequency vibrations of
:
Designed for IEC 348 Cat I standard
1
V peak to peak +/- 1%, frequency approx 1
Front panel ground reference
90 to l30V or 190 to 265V
45 to70 Hz. 400Hz operation is available as an
12 to 33V
85V
A170W
approx
WEI HTG 400:
420&450: 400 with battery unit: 400 and 450 with battery unit:
DIMENSIONS
See Figure A2.
ACCESSORIES SUPPLIED
Operating manual Line cord DC power connector
OPTIONAL ACCESSORIES
Rack mounting kit PN 04090490 Carrying case (soft padded) PN 04101221
Carrying case (hard, foam lined) Front facia cover Accessory Pouch PN 04101223 Service manual Probe DC to 100MHz switched x 1 and x I 0 PBI2 Probe DC to 250MHz 1200V x 100 Probe DC to 7MHz 15kV x 1000
External TV Trigger Unit
PN 04101220
PN 04101224
5.5Kg (12Ib) approx.
6.5Kg (14.2Ib) approx. 1l.7Kg (25.7Ib) approx.
12.7Kg (27.9Ib) approx.
PN 04101222
PBI7
PB27
Page 49
Appendix 2 Specification
400 Series Operators Manual
Figure
A2
400 Family Dimensions
Page 50
400 Series Operators Manual
2
Appendix 3 Remote Operation
Appendix 3: Remote DC Operation of the 400 series DSOs
As well as the possibility of having an internal battery unit ­option 107 see section 6- the 400 series DSOs can be operated from external DC voltages between 11.5 and 33V measured at the DSO. This has the advantage of uninterruptible operation which an oscilloscope powered from an inverter type of DC supply does not have. Inverter type DC supplies are powered by batteries and generate AC power in the absence of an incoming AC supply. In order for them to start working they first have to detect the loss of the AC supply which can take a few seconds.
Any applications which require continuous operation of a DSO regardless of AC supply fluctuations or drop-outs, as well as those requiring a remote power source will benefit from the versatility of the 400 series' DC power feature. Some examples of these applications are uninterruptible power supply development and troubleshooting, and troubleshooting of supply interference caused by large electric maCH1nes.
The DC operation power dissipation of the 400 is about 60W and of the 420 and 450 about 62W (67W when printing.) This dissipation is a constant power load to the source.
This means that less current is drawn at higher voltages and conversely more current at lower voltages. The user should ensure that there is sufficient source voltage to overcome cable and connector voltage losses that may occur because of the relatively high current involved. See table A3.1.
Figure A3.2 below shows the 400's internal arrangement of the power connections. It can be seen that current will be drawn from the AC supply as long as the external DC input voltage is lower than the AC input voltage minus the two diode voltage drops in the bridge rectifier.
This simple diode O-Ring of the two power sources means that switCH1ng from one source to another is very quick there is no interruption in the operation of the instrument while one source takes over from the other.
The diode in series with the external DC source protects the instrument against reverse polarity voltages and allows both the AC and DC supplies to be connected at the same with the source supplying the higher voltage being the one from which current is drawn.
Table of Annealed Copper wire sizes vs. length for
O.5V drop at 5 Amp current
Diameter Cross Sectional Area
(mm)
0.8 0.5 2.9
0.98
0.75 4.35
1.25 7.25
1.26
1.78 2.5 14.5
4.0 23.2
2.26
6.0 34.8
2.76
8.0 46.4
3.19
3.57 10.0 58.0
TableA3.1
External DC from battery unit option or separate DC source
FigureA3.
(mm)
Length
M
AC from internal transformer
Page 51
A
ppendix 4 Front & Back Pictures
s
400 Series Operators Manual
Fig. A4a Rear View
Fig. A4b RS423 Connection
Page 52
400 Series Operators Manual
Appendix 4 Front & Back Pictures
Page 53
Fig. A4c Front Panel Control
Page 54
Service Facilities
400 Operators Manual
Service Facilities
Gould and its distributors and agents maintain comprehensive after sales facilities and, whether or not it is still under guarantee, the instrument should be returned to the local Gould service centre or distributor through whom it was supplied for servicing if this is necessary. The type and serial number of the instrument should always be quoted, together with full details of any fault and service required.
Equipment returned for servicing must be adequately packed, preferably in the box in which the product was supplied and shipped, with transportation charges prepaid. We accept no responsibility for instruments arriving damaged.
Our Sales and Service Department and those of approved distributors and agents are ready to assist you at all times.
The Gould Service Department and those of approved dis­tributors and agents can provide maintenance and repair information by telephone or letter, if required.
PRINCIPAL GOULD SERVICE OFFICES
AUSTRIA
Gould Electronics GmbH, Niederlassung Wien, Mauerbachstrasse 24, A-1140 Wien. Telephone (222) 972506. Telex 01-31380. Fax (222) 97250638
FRANCE
Gould Electronique, 57 Rue Saint Sauveur, Ballainvilliers, 91160 Longjumeau. Telephone (6) 934-1067. Telex 600824 GOULD. Fax (6) 934-2073
Adresse postale: B.P. 115,91162 Longjumeau Cedex.
GERMANY Gould Instruments, Waldstrasse 66,6057 Dietzenbach Telephone (06074) 4908-0. Fax (06074) 4908-48.
WL490
UNITED KINGDOM Gould Electronics Ltd., Instrument Systems,
Roebuck Road, Hainault, liford, Essex IG6 3UE. Telephone 081-5001000. Telex 263785. Fax 081-5010116.
U.S.A
Gould Inc., Recording System Division,
3631 Perkins Avenue, Cleveland, Ohio 44114. Telephone (216) 361-3315. Telex 1961123 GLD RS UT. Fax (216) 881-4256.
And other Gould Agents and Distributors worldwide.
Manual Part Number 457400
Page 55
Notes
Page 56
Page 57
t
400 Operators Manual Amendment Sheet
November 1990
400 System Rese Should the system memory become corrupted, the instrument could lock out the front panel operation or begin to behave erratically. Since the internal system variables are held in non-volatile memory, turning the set off an on again may not clear them. In this unlikely event a system reset may be necessary.
A system reset should only be used as a last resort as all the internal memories will be cleared erasing all reference traces, calibration details, and set-up details.
To perform a system reset, switch the instrument off. Press and hold the Hold All and the Plot buttons. Keep these buttons pressed and turn the instrument on using the Power button. The instrument will then start up in its reset condition. This can be confirmed by noting that on power up the instrument performs a complete calibration.
Amendment to 400 series Operators Manual Part Number
457403
Page 1 of 1 Part No. 460150