Goudsmit HGM09S Operating Instructions Manual

O
PERATING
INSTRUCTIONS
G
AUSSMETER
HGM09
S
Gaussmeter HGM09s Operating Instructions
Chapter 1
Page 2 / 75
© 2010
© 2010 © 2010
© 2010 Goudsmit Magnetic Systems
Goudsmit Magnetic SystemsGoudsmit Magnetic Systems
Goudsmit Magnetic Systems
All rights reserved. No part of these operating instructions may be reproduced or du­plicated without the author’s written consent.
We shall not be liable for the accuracy of these operating instructions nor for dam­ages which can result from the use of this manual. Since mistakes can never be avoided completely, despite all efforts, we would appreciate any given hint. We will be anxious to correct any faults known to us as soon as possible.
Edition Note
Data File HGM09_Manual_DE_1_00
Doc. Date 29.06.2010
Hardware Status VI
Software Status 290310
Current Documentation Status 1.00
Status of Documentation Concerns Page(s) No.
01/2010 Compilation
Gaussmeter HGM09s Operating Instructions
Chapter 1
Page 3 / 75
Tab
TabTab
Table of Contents
le of Contentsle of Contents
le of Contents
1 Safety Instructions 5
1.1 Safety Instructions for the Device 5
1.2 Safety Instructions for the Measuring Probes 6
1.3 Safety Symbols 6 2 Brief Introduction 7
2.1 Preparing a Measurement 7
2.2 Running a Measurement 7
2.3 Measuring Unit 8
2.4 Selecting the Measuring Range 8
2.5 Display 9
2.6 Status Display 9 3 Gaussmeter Function 10
3.1 The Hall Effect 10
3.1.1 Linear Properties of the Hall Probe 10
3.1.2 Non-linear Properties of the Hall Probe 11
3.2 Measurement Details 13
3.2.1 Sample Measurement with an NdFeB Magnet 13
3.2.2 Remanence and Hall Gaussmeter Measurement 14
3.2.3 Accuracy Based on Positioning and Direction 15
3.2.4 External Static Magnetic Fields 16
4 Control Elements and Connections 17
4.1 Front Side Overview 17
4.2 Ports Overview 17
4.3 Power Supply 18
4.4 Batteries 18
4.4.1 Charging the Batteries 19
4.5 Probe Connection 20
4.6 USB Interface 20
5 Operation 21
5.1 Keyboard 21
5.2 Display 22
5.3 Status Display 22
5.4 Switching on/off 24
5.5 Null 24
5.6 Measuring Range 25
5.6.1 Overview Measuring Ranges 25
5.7 Measuring Unit 26
5.8 DC/AC Field Measurements 26
5.8.1 DC Field Measurements 27
5.8.2 AC Field Measurements 28
5.9 Peak Value Measurement 30
5.9.1 Normal Peak Value Recording 30
5.9.2 Fast Peak Value Recording 31
5.10 Probe Data 33
6 Setup Menu 34
6.1 Settings 35
6.1.1 Operating Mode of the USB Interface 35
6.1.2 Selecting the Display Units 35
6.1.3 Operating Mode of the Peak Value Recording 36
6.1.4 DC/AC Field Measurement 36
6.1.5 Range Selection 37
6.1.6 Polarity Display (North/South) 37
6.1.7 Switching off the Device 37
6.1.8 Charging the Batteries 38
6.1.9 Brightness of Display Illumination 38
6.1.10 Display Contrast Adjustment 38
6.1.11 Version Remarks 39 7 Serial Interface 40
7.1 Introduction 40
7.2 Connecting the Gaussmeter to a Computer 41
7.2.1 Connector Plug 41
7.3 Direct Operation on the Computer 41
7.3.1 Example Normal Measuring Mode in Excel

42
7.3.2 Example Fast Peak Value Mode in Excel

42
7.3.3 Example Slow Peak Value Mode in Excel

42
7.4 Operation via Interface 43
7.4.1 Installation on the Computer 43
7.4.2 USB Interface Data Format 43
7.4.3 Character Set 43
7.4.4 Introduction to the SCPI Language 43
7.4.5 SCPI Data Types 45
7.4.6 The SCPI Status Model 48
7.5 Summary of SCPI Commands 51
7.5.1 Control Commands 51
7.5.2 Main Commands 51
7.5.3 Peak Value Functions 51
7.5.4 Probe Functions 52
7.5.5 Parameters 52
7.5.6 Device Functions 52
7.6 Explanation of the Individual SCPI Commands 53
7.6.1 Control Commands 53
7.6.2 Main Commands 57
7.6.3 Peak Value Function 60
7.6.4 Probe Functions 62
7.6.5 Parameters 63
7.6.6 Device Functions 67 8 Unit Conversion Table 69 9 Technical Data 70 10 Declaration of Conformity 72 11 Warranty and Copyright 73 12 Index 74
Gaussmeter HGM09s Operating Instructions
Chapter 1
Page 4 / 75
Table of Illustrations
Table of IllustrationsTable of Illustrations
Table of Illustrations
Display 9 Basic Assembly of a Hall Probe 10 Flux Line Characteristics of NdFeB Induction Disks 13 Field Strength Pattern of NdFeB Induction Disk 14
Front Side 17 Ports 17 Battery Box 18 Display 22 USB Connection 41 SCPI Status Model 49
Gaussmeter HGM09s Operating Instructions
Chapter 1 Safety Instructions
Page 5 / 75
1111 Safety Instructions
Safety InstructionsSafety Instructions
Safety Instructions
1.1
1.11.1
1.1 Safety Instructions
Safety InstructionsSafety Instructions
Safety Instructions for the Device
for the Device for the Device
for the Device
Use the measuring instrument only according to the user manual.
Do not replace any parts and do not make any modifications on the product with­out our explicit and written consent. Do not carry out any service measures at this device. For repair and maintenance please return the product to Goudsmit Mag­netic Systems or to your supplier, in order to make sure that all safety features re­main.
Handling malpractices may result in damage to the device and possibly in injury or death of persons.
The power supply is designed for a voltage range of 100 to 240VAC at 47 to 63Hz. Do not use the power supply on voltages beyond these areas.
Before use, check the power supply for cracks or missing plastic parts. Pay attention to the insulation. Do not use the power supply if it is damaged.
Do not dispose of the measuring device in normal household garbage. Please con­tact the manufacturer for the proper disposal of this instrument.
Only use magnetic field probes designed for this measuring device.
Observe the labeling of this measuring device before connecting a magnetic field probe to it.
Replace the rechargeable batteries only by batteries of the same type.
Make sure to insert the rechargeable batteries correctly into the measuring device and mind the correct polarity.
Do not dispose of the rechargeable batteries in the regular household garbage. Heed your national regulations for the disposal of old batteries.
Do not work in explosive environs or near inflammable gases or vapors with this device.
Environmental conditions The device is designed for the use in rooms with low condensation. See technical data.
Gaussmeter HGM09s Operating Instructions
Chapter 1 Safety Instructions
Page 6 / 75
1.2
1.21.2
1.2 Safety Instructions
Safety InstructionsSafety Instructions
Safety Instructions for the Measur
for the Measur for the Measur
for the Measuring Probes
ing Probesing Probes
ing Probes
The magnetic measurements should only be carried out in areas with a max. vol­tage of 60V DC, 30V AC RMS. The magnetic field probes are not electrically insu­lated. Please note that the probe holders and the housing might be electrically con­nected with the protective earth.
If you work in areas with voltages higher than 60V DC, 30V AC RMS or 42V peak values, act with particular caution because of electric shock hazard.
For measurements in high magnetic fields, please pay attention to the dangers that may arise by strong magnetic fields.
1.3
1.31.3
1.3 Safety
SafetySafety
Safety Symbols
Symbols Symbols
Symbols
Safety symbols can be found on various spots on the device.
Before using this connection or function read the corresponding instructions in the manual.
This symbol refers to information and references in the instruction manual which the user has to follow in order to avoid injuries to persons or damage to the device, or to obtain correct measuring results.
Gaussmeter HGM09s Operating Instructions
Chapter 2 Brief Introduction
Page 7 / 75
2222 Brief
BriefBrief
Brief Introduction
Introduction Introduction
Introduction
Measurements with the gaussmeter use the Hall effect as a measuring principle. A Hall probe is a symmetric semiconductor impressed by current. A magnetic field running vertically to this element generates an asymmetry on the chip and thereby creates an output voltage that, as a first approximation, is proportional to the prod­uct of magnetic field strength and the forced current. For higher magnetic field strengths this dependency is no more linear. This effect is automatically compen­sated in the device. The gaussmeter thus measures the magnetic flux density locally. The sensor only captures the component of the magnetic flux density which runs perpendicularly through it.
2.1
2.12.1
2.1 Preparing a Measurement
Preparing a MeasurementPreparing a Measurement
Preparing a Measurement
Make sure that the batteries are loaded.
Connect the device to a personal computer via the USB cable if required. Or
connect the included power supply via the USB socket.
Connect a measuring probe.
A suitable measuring probe can be plugged in via the Mini-DIN socket on the top of the device. Each measuring probe is calibrated individually. The calibra­tion data are stored in the probe memory. On inserting or changing a measur­ing probe these parameters are read in automatically.
Switch the device on and off by pressing the I/O
I/OI/O
I/O button longer (approx. 2 sec).
2.2
2.22.2
2.2 Running
RunningRunning
Running a Measurement
a Measurement a Measurement
a Measurement
After switching on the device, the current measuring value is shown continuously. The display additionally shows further information on the state of the device and the selected measuring range as well as the measuring mode.
The measuring range can be changed by pressing the RANGE
RANGERANGE
RANGE button. The characteristics of this button can be modified in the setup menu. By repeatedly pressing the RANGE
RANGERANGE
RANGE button both the measuring range and the display unit or
the DC/AC field measurement can be changed.
Gaussmeter HGM09s Operating Instructions
Chapter 2 Brief Introduction
Page 8 / 75
Insert the measuring probe into the measuring field after adjusting the re­quested measuring range and the requested unit. Especially for inhomogeneous magnetic fields, such as they occur on the surface and edges of magnets, keep in mind that the measured magnetic flux density depends very largely on the distance and the position. Further pay attention that the magnetic field compo­nent is measured in one direction only, so that a tilting of the measuring probe may lead to an error.
2.3
2.32.3
2.3 Measuring Unit
Measuring UnitMeasuring Unit
Measuring Unit
The gaussmeter shows measuring values in physical units of the SI-system as well as of the CGS system (Gauss system – particularly in use in North America).
The unit is either definitely preset in the setup menu or you can adjust it by pressing the RANGE
RANGERANGE
RANGE button. Each time you press the button the next unit is selected.
Please note that the same button selects the measuring range and the DC/AC field measurement, if necessary.
2.4
2.42.4
2.4 Selecting
Selecting Selecting
Selecting the Measuring Range
the Measuring Rangethe Measuring Range
the Measuring Range
By repeatedly pressing the RANGE button, you select the measuring range via four areas each. The maximum measuring value in this area is shown in the bottom left area of the display. You can also select an automatic range mode via the setup menu. In this operating mode, the measuring range is automatically aligned with the current measurement. Please note that no automatic range adjustment is possible in the peak mode.
If the measuring value exceeds the selected range limit, the display shows
-OL-
instead of the measuring value.
Gaussmeter HGM09s Operating Instructions
Chapter 2 Brief Introduction
Page 9 / 75
2.5
2.52.5
2.5 Display
Display Display
Display
A typical example display is demonstrated below.
Figure 1 Display
2.6
2.62.6
2.6 Stat
StatStat
Statuuuussss Display
DisplayDisplay
Display
In addition to showing the current measuring value, the gaussmeter display also shows the status information, a negative or positive peak value, if required, and the state of the USB interface as well as the charging status of the battery.
Unit
Measuring Value
Measuring Range
Operating Mode
Battery St
a
tus USB St
a
tus
Gaussmeter HGM09s Operating Instructions
Chapter 3 Gaussmeter Function
Page 10 / 75
3333 Gaussmeter
GaussmeterGaussmeter
Gaussmeter Function
Function Function
Function
3.1
3.13.1
3.1 The Hall Effect
The Hall EffectThe Hall Effect
The Hall Effect
3.1.1
3.1.13.1.1
3.1.1 Linear Properties of the Hall Probe
Linear Properties of the Hall ProbeLinear Properties of the Hall Probe
Linear Properties of the Hall Probe
The measurement is based on the diflection of charge carriers in a magnetic field in­side a conductor. For this reason the Lorentz force is the basis for measuring a mag­netic flux density. If you set a voltage between the beginning and the end of a flat electrical conductor, the carriers move with a speed of
enDrift
Ev
r
r
=
µ
, whereby
n
µ
represents the carrier mobility in the conductor. Due to their high mobility, the carri­ers are always electrons. Perpendicular to the current direction, a voltage can be measured that is ideally proportional to the magnetic flux density. Only the part of the flux density is effective which runs perpendicularly through the flat side of the con­ductor.
Figure 2 Basic Assembly of a Hall Probe
If you do not extract any current from the electrodes S1 and S2, but only measure the voltage, the following applies:
B
t
wIw
U
en
Hall
e
=
B
S2
C2
S1 C1
Gaussmeter HGM09s Operating Instructions
Chapter 3 Gaussmeter Function
Page 11 / 75
It follows that:
B
ten
I
U
e
Hall
=
1
With
This represents the idealized Hall effect.
In reality, the results deviate from this idealized effect.
Since there is a linear equation between the current and the measuring result, it follows that
BSB
ten
R
e
Hall
=
=
0
1
3.1.2
3.1.23.1.2
3.1.2 Non
NonNon
Non----linear
linear linear
linear Properties
PropertiesProperties
Properties of the Hall Probe
of the Hall Probe of the Hall Probe
of the Hall Probe
Contrary to the idealized description you find a non-linear performance:
(
)
offsetHALLHall
RBBSR ++=
2
0
1
α
For the used Hall probes the real description is true for flux densities of up to approx. 5000 mT.
3.1.2.1
3.1.2.13.1.2.1
3.1.2.1 Reasons for the Occurrence of R
Reasons for the Occurrence of RReasons for the Occurrence of R
Reasons for the Occurrence of R
offset
offsetoffset
offset
The largest deviation from the idealized Hall effect is the occurrence of an offset volt­age without a magnetic field. This effect is mainly caused by geometrical asymme­tries of the Hall element.
e
n
Carrier
e
Elementary charge of the electron (1.6022x10
-19
As)
w
Width of the path on which the electrons move
t
Effective force of the Hall element
B
Flux density in [Tesla]
Gaussmeter HGM09s Operating Instructions
Chapter 3 Gaussmeter Function
Page 12 / 75
3.1.2.2
3.1.2.23.1.2.2
3.1.2.2 Reasons for the Field Dependenc
Reasons for the Field DependencReasons for the Field Dependenc
Reasons for the Field Dependenceeee of the Sensitivity
of the Sensitivity of the Sensitivity
of the Sensitivity
There are several influences for the flux-density dependence of the sensitivity:
The carrier mobility depends on the flux density. This influence generally brings about a negative α
Hall
and is irrelevant for the used Hall sensors.
More important is the geometry of the used Hall sensors. The lamellar structure cre­ates a geometry-based field dependence of the sensitivity.
The non-homogeneous distribution of the current density in such a structure is the cause of this effect.
Already in field-free cases, the current distribution on the Hall element is complex. This entails a lowering of S0 and has an influence on the field dependence of the sensitivity.
A complex real-time correction of the handheld gaussmeter HGM09 compensates for the inherent non-linearities of the used Hall probes and guarantees a very stable zero point.
3.1.2.3
3.1.2.33.1.2.3
3.1.2.3 Field Dependenc
Field DependencField Dependenc
Field Dependenceeee of the Cross Current Resistance
of the Cross Current Resistance of the Cross Current Resistance
of the Cross Current Resistance
The complex current distribution is the cause of the resistance of a Hall probe. Cur­rent components which, just as the Hall voltage, run perpendicularly to the direction of the current feed, cause a diverted Hall effect. For the current source this effect re­sults in a flux density modulated resistor.
The device must have a sufficiently high dynamic for the measurement with fast mag­netic pulses, to be able to compensate this effect. The handheld gaussmeter HGM09 is optimized for this operating case.
3.1.2.4
3.1.2.43.1.2.4
3.1.2.4 Temperature Dependenc
Temperature DependencTemperature Dependenc
Temperature Dependenceeee of the Sensitivity
of the Sensitivity of the Sensitivity
of the Sensitivity
Due to the large band gap of the used Hall sensors, the temperature dependence of the probe sensitivity is low, it is approx. -0.06%/ºC.
3.1.2.5
3.1.2.53.1.2.5
3.1.2.5 Temperature Dependenc
Temperature DependencTemperature Dependenc
Temperature Dependenceeee of the Cross
of the Cross of the Cross
of the Cross Current Resistance
Current ResistanceCurrent Resistance
Current Resistance
The temperature dependence of the cross current resistance ranges at approx.
0.3%/ºC and is compensated automatically by the device.
Gaussmeter HGM09s Operating Instructions
Chapter 3 Gaussmeter Function
Page 13 / 75
3.2
3.23.2
3.2 Measurement
MeasurementMeasurement
Measurement D
D D
Deeeetails
tailstails
tails
The used Hall probes contain a very small active semiconductor area ranging at approx. 100µm. The local resolution of this measuring method is thus rather high. Also note that individual Hall probes measure one field component only.
3.2.1
3.2.13.2.1
3.2.1 Sample
Sample Sample
Sample Measurement
MeasurementMeasurement
Measurement with
withwith
with a
a a
annnn NdFeB Magnet
NdFeB Magnet NdFeB Magnet
NdFeB Magnet
Due to the high local resolution, the near-surface measurement with magnets may lead to misinterpretations because of the large field-strength gradients.
Figure 3 Flux Line Characteristics of NdFeB Induction Disks
Figure 3 shows an NdFeB magnet with a material remanence of 1400mT. In this exam­ple the magnetic disk has a thickness of 5mm with a diameter of 20mm. The magnet is supposed to be measured in 1mm distance from the surface.
Gaussmeter HGM09s Operating Instructions
Chapter 3 Gaussmeter Function
Page 14 / 75
Figure 4 Field Strength Pattern of NdFeB Induction Disk
The diagram in Figure 4 represents the measurement with a Hall probe which is moved in parallel to the surface of the magnet with a distance of 1mm distance to the measur­ing surface.
A minimum of the flux density emerges in the center of the magnet. Here approx. 230mT are measured. Due to the locally changing working points on the radius of the magnet, the flux density increases toward the outside. In the center the magnet carries the highest magnetic load in air and therefore provides the lowest flux density.
3.2.2
3.2.23.2.2
3.2.2 Remanence and Hall Gaussmeter Measurement
Remanence and Hall Gaussmeter MeasurementRemanence and Hall Gaussmeter Measurement
Remanence and Hall Gaussmeter Measurement
The remanence
B
r
is a measure for the aligned magnetic dipoles in the center of the
magnet.
B
r
is the theoretically maximal flux density that can be achieved if the magnet is
in magnetic idle. If it works against a magnetic resistance, it is always
B
<
B
r
.
On the surface of an individual magnet B <
B
r
/2 applies even more.
Which value is actually measured in the pole center depends on the geometry of the magnet.
Gaussmeter HGM09s Operating Instructions
Chapter 3 Gaussmeter Function
Page 15 / 75
As
B
r
as well as B are measured in the unit Tesla, the magnetic field measured on the
outside is often mixed up with the remanence.
Please note that a magnet
a magneta magnet
a magnet without back iron only shows a value clearly
shows a value clearlyshows a value clearly
shows a value clearly below the r
below the rbelow the r
below the re-
e-e-
e-
manence
manencemanence
manence on the surface. Due to the local and geometry-dependent measurement, the remanence of the workpiece cannot be checked reliably with a gaussmeter.
3.2.3
3.2.33.2.3
3.2.3 Accuracy
AccuracyAccuracy
Accuracy Based on
Based on Based on
Based on Positioning and Direction
Positioning and Direction Positioning and Direction
Positioning and Direction
Since the measuring value is dependent on the position, an accurate and repeatable measurement depends on the exact positioning of the probe during the measurement.
The measurement on the pole center of the magnet is most uncritical. When moving the probe on the pole surface of the magnet, the measuring value hardly changes at first. When changing the distance though, the measuring value varies considerably.
The smaller the
The smaller the The smaller the
The smaller the magnet to be measured, the
magnet to be measured, the magnet to be measured, the
magnet to be measured, the stronger eve
stronger evestronger eve
stronger even slight misalignments
n slight misalignments n slight misalignments
n slight misalignments
change the measuring value.
change the measuring value.change the measuring value.
change the measuring value. For quality-related research it is essential to ensure the positioning accuracy.
Since a Hall probe only records one field strength vector, the correct alignment relative to the magnet is important.
Please be particularly careful when measuring at the zero point at pole transitions. By slightly tilting the probe, you measure additional lateral field shares that seem to dis­place the zero passage.
In normal applications, a maximum flux density value is usually determined at a given position. The measuring probe is placed in position and varied in location and direction until the maximum is found. The device supports this measurement with the peak hold.
Gaussmeter HGM09s Operating Instructions
Chapter 3 Gaussmeter Function
Page 16 / 75
3.2.4
3.2.43.2.4
3.2.4 External Static Magnetic Fields
External Static Magnetic Fields External Static Magnetic Fields
External Static Magnetic Fields
Particularly in sensitive measuring ranges, an external static magnetic field, as e.g. the Earth’s field, can already become clearly noticeable. These external magnetic fields lead to a corruption of the measuring result.
To compensate external magnetic stray fields or asymmetries of the Hall probe, the de­vice can be reset.
For this purpose please hold the measuring probe into a field-free area, e.g. a zero Gauss chamber, or orient the measuring probe in a free field in east-west direction and press the NULL
NULLNULL
NULL button for approx. 3 seconds. The device then carries out a null balance
automatically.
The values are stored so that this balancing has to be carried out only in seldom cases.
If the magnetic field is too high during the automatic balancing, the correction is stopped with an error message.
Gaussmeter HGM09s Operating Instructions
Chapter 4 Control Elements and Connections
Page 17 / 75
4444 Control Elements and Connections
Control Elements and ConnectionsControl Elements and Connections
Control Elements and Connections
4.1
4.14.1
4.1 Front Side Overview
Front Side OverviewFront Side Overview
Front Side Overview
Figure 5 Front Side
4.2
4.24.2
4.2 Ports
PortsPorts
Ports Overview
Overview Overview
Overview
Figure 6 Ports
1 Display 2 RANGE
RANGERANGE
RANGE button Range selection 3 NULL button Reset 4 DATA
DATADATA
DATA button Data recording 5 IO
IOIO
IO button On/Off
1 PROBE Port for the probe 2 USB Port for computer or power supply
Gaussmeter HGM09s Operating Instructions
Chapter 4 Control Elements and Connections
Page 18 / 75
4.3
4.34.3
4.3 Power
PowerPower
Power Supply
Supply Supply
Supply
The gaussmeter can be operated with the included power supply. The power supply is connected via the USB port (type Mini-B) on the top of the device. The power supply is designed for a line voltage of 100 to 240V alternating current 50-60 hertz at a charging rate of max. 300mA. Use the power supply only when both the de­vice itself and the connector cable obviously do not show any damages.
4.4
4.44.4
4.4 Batteries
BatteriesBatteries
Batteries
The device is designed to operate with rechargeable batteries, type NiMH 2450mAh AA PK4. Alternatively the device can also operate with standard NiMH batteries of the same voltage and similar capacity. Only use the identical battery type if a change is necessary. Only change both batteries at the same time.
To insert or remove batteries, separate the device from the measuring probe and remove the cables from the USB plug. Switch off the device. Remove the protective cover and carefully open the battery box on the bottom side of the device. Mind the correct inserting position when exchanging the batteries. Mind the correct polarity. Normally old batteries must not be disposed of in the regular household garbage. Protect your environment and heed your national waste disposal regulations for old batteries.
Figure 7 Battery Box
The normal operating time with fully loaded batteries is approx. 10 hours. You can prolong this time by selecting a darker backlight. You can adjust the brightness in the parameter menu. Moreover, you can activate the automatic switch-off. The de­vice switches off automatically after 2 or 5 minutes, if no button is pressed and the measuring value does not change significantly.
Gaussmeter HGM09s Operating Instructions
Chapter 4 Control Elements and Connections
Page 19 / 75
4.4.1
4.4.14.4.1
4.4.1 Charging the Batteries
Charging the BatteriesCharging the Batteries
Charging the Batteries
The battery is charged via the power supply during operation or when connected to a personal computer. The top right edge of the display shows the battery charge condition.
Please note that the batteries are only charged when the device is switched on or when it remains connected to the power supply or personal computer after switch­ing off. The top right edge of the display shows the battery charging status.
Note
The device does not charge automatically when you feed it via the power supply and do not switch it on.
You can also avoid the charging by pressing the IO
IOIO
IO button in charging state once
again and hereby finally switch off the device.
In the setup menu you can manually deactivate the battery charging. This might be­come necessary if your personal computer cannot provide the required charging current. The interconnection of an active USB hub might help if such a problem oc­curs. Please note that this hub must however possess its own power supply.
Gaussmeter HGM09s Operating Instructions
Chapter 4 Control Elements and Connections
Page 20 / 75
4.5
4.54.5
4.5 Probe Connection
Probe ConnectionProbe Connection
Probe Connection
The magnetic measuring probe is plugged into the probe port (DIN Mini-DIN­socket) on the top of the device.
Note
Only use measuring probes which are approved by the manufacturer to operate with this device.
Note
The probe plugs must not be connected to the electric potential, the protective conductor or the plug shell. If you measure near current conducting parts, make sure that there is sufficient distance and a suf­ficiently good insulation.
4.6
4.64.6
4.6 USB Interface
USB InterfaceUSB Interface
USB Interface
The transmission of the measuring values and the entire gaussmeter control are possible via the installed standard USB interface.
The interface is specified according to the USB (Universal Serial Bus) 2.0. The interface is not insulated. Please note that the shielding of the measur­ing probes (e.g. the metallic handle) might be connected to the protective conductor of your personal computer via this USB interface.
The device supports two USB device classes. On the one hand the HID class. In this mode the data can be entered directly into any program. The gaussmeter acts similar to the keyboard of a personal computer.
On the other hand, the device can be operated in the USB device class CDC and thereby emulates a virtual serial interface on the personal com­puter or another processor. In contrast to the HID mode, the installation of a driver on the processor becomes necessary here.
Via the virtual serial interface all functions and parameters of the device can be set using the SCPI command, and the measuring values can be read out automatically.
Gaussmeter HGM09s Operating Instructions
Chapter 5 Operation
Page 21 / 75
5555 Operation
OperationOperation
Operation
5.1
5.15.1
5.1 Keyboard
KeyboardKeyboard
Keyboard
The required functions are selected and triggered via 4 buttons. The labeling of the button corresponds to the requested function, e.g. RANGE
RANGERANGE
RANGE changes the range of the
measuring value.
More complex parameter settings can be made via menu functions in the setup me­nu.
The button functions in detail:
With this button you change the measuring range, the measuring unit
and the measuring mode (DC or AC fields). The exact function of this but­ton depends on the settings in the setup menu. Further explanations, see there.
Via NULL the displayed peak value is set to zero in the peak value meas-
uring mode.
If you press the button for approx. 3 seconds, the magnetic offset values are compensated.
In the USB operating mode HID (keyboard) the current measuring values
are transmitted to the connected processor and can be entered into any program.
This button has no function in the other operating modes. In the USB op­erating mode CDC (virtual interface), pressing this button can be read out via an SCPI command.
To switch the device on and off, press the button for approx. 2 seconds.
If you press the button longer when switching on, the setup menu is called up.
Gaussmeter HGM09s Operating Instructions
Chapter 5 Operation
Page 22 / 75
5.2
5.25.2
5.2 Display
DisplayDisplay
Display
A typical display example is shown below.
Figure 8 Display
5.3
5.35.3
5.3 Stat
StatStat
Statuuuussss Display
Display Display
Display
Apart from the measuring value the display shows some status information.
Battery fully charged.
Residual capacity approx. 75%
Residual capacity approx. 50%
Residual capacity approx. 25%
Battery almost empty. Immediate charging required.
Battery status is determined.
Battery defective.
The top right area displays the battery charging status.
Battery charging.
Range
Unit
Measuring Value
Operating Mode
Battery St
a
tus USB Status
Gaussmeter HGM09s Operating Instructions
Chapter 5 Operation
Page 23 / 75
The device is connected to the computer via a virtual inter­face. The device is connected to the computer as a keyboard simulation. In this mode, the gaussmeter acts like a key­board. Pressing the
DATA
DATADATA
DATA
button causes the transmission of
the current measuring values to the selected program.
Interface
An internal error has occurred. Detailed error notes are given on quitting the setup menu.
Measuring Range
The bottom left area of the display shows the currently se­lected measuring range with the currently selected unit.
The equals sign to the right stands for DC field measure­ments.
The shift sign to the right stands for AC field measurements.
Operating Mode
Peak
in the bottom part shows the operating mode Fast­Peak-Registration, thus the fast recording of the maximum values of DC fields.
The display
Peak
with two measuring values in the penultimate line shows the two maximum and minimum measuring values determined during the slow peak value registration.
The current measuring value is 109.5mT, the smallest peak value was -112.2mT, the highest peak value was +109.5mT.
Gaussmeter HGM09s Operating Instructions
Chapter 5 Operation
Page 24 / 75
5.4
5.45.4
5.4 Switching on/off
Switching on/offSwitching on/off
Switching on/off
The device is switched on and off with the right bottom IO
IO IO
IO button. In each case the button must be pressed for approx. 1 second. This avoids an accidental switching on and off.
If the batteries are not yet fully charged during switch off, the charging
continues also after switching off. You can see this in the top right battery display. By pressing the ON
ONON
ON and OFF
OFF OFF
OFF button once again, the device is finally switched off.
During activated automatic switch-off, the device is turned off after 2 or 5 minutes without any activity.
5.5
5.55.5
5.5 Nu
NuNu
Null
ll ll
ll
During the peak value measurement the stored minimum and maximum values are reset by briefly pressing the NULL button.
To compensate external magnetic fields, press the NULL button for approx. 3 seconds. The device then carries out an automatic null
balance. During the null balance the text NULL SET is displayed. The null balance roughly takes 4 seconds. To avoid fatal errors, this process is interrupted by the error message OVERFLOW, if the adjacent external field is larger than 10% of the respective measuring range.
Note
Only carry out the null balance in a preferably field-free area.
Operation via the external interface:
:PEAK:NULL
Reset peak values
:NULL
Carry out null balance
Gaussmeter HGM09s Operating Instructions
Chapter 5 Operation
Page 25 / 75
5.6
5.65.6
5.6 Measuring Range
Measuring RangeMeasuring Range
Measuring Range
.
The gaussmeter has 4 measuring ranges. The range limit value is shown
at the bottom left of the display. The range limit values depend on the ranges themselves and on the selected measuring unit. The ranges can be switched manually or automatically. Regarding the manual switch, the measuring range is selected by multiple pressing of the RANGE
RANGERANGE
RANGE button. The automatic range selection mode is set via the setup menu, followed by the pa­rameter
Range off manually
.
With automatic range selection activated, the device switches to an insensitive range if the measuring value has exceeded 90% of the actual measuring range. A sensitive range is selected when the measuring value falls below 10% of the actual measuring range.
Operation via the external interface:
:RANG:SET {0|1|2|3}
Manual setting; 0 = most sensitive area
:RANG:AUTO
Automatic setting
5.6.1
5.6.15.6.1
5.6.1 Overview Measuring Ranges
Overview Measuring Ranges Overview Measuring Ranges
Overview Measuring Ranges
DC Field Measurement AC Field Measurement Fast Pulse Measurement
Unit Range Range Limit Resolution Range Limit Resolution Range Limit Resolution
Tesla 1 10 mT 1 µT 10 mT 10 µT 10 mT 10 µT
T 2 100 mT 10 µT 100 mT 100 µT 100 mT 100 µT
3 1000 mT 100 µT 1000 mT 1 mT 1000 mT 1 mT
4 4500 mT 1 mT 3000 mT 10 mT 4500 mT 10 mT
Ampere/m 1 10 kA/m 1 A/m 10 kA/m 10 A/m 10 kA/m 10 A/m
A/m 2 100 kA/m 10 A/m 100 kA/m 100 A/m 100 kA/m 100 A/m
3 1000 kA/m 100 A/m 1000 kA/m 1 kA/m 1000 kA/m 1 kA/m
4 3800 kA/m 1 kA/m 2500 kA/m 10 kA/m 3800 kA/m 10 kA/m
Gauss 1 100 G 10 mG 100 G 100 mG 100 G 100 mG
G 2 1000 G 100 mG 1000 G 1 G 1000 G 1 G
3 10 kG 1 G 10 kG 10 G 10 kG 10 G
4 45 kG 10 G 30 kG 100 G 45 kG 100 G
Oersted 1 100 Oe 10 mOe 100 Oe 100 mOe 100 Oe 100 mOe
Oe 2 1000 Oe 100 mOe 1000 Oe 1 Oe 1000 Oe 1 Oe
3 10 kOe 1 Oe 10 kOe 10 Oe 10 kOe 10 Oe
4 45 kOe 10 Oe 30 kOe 100 Oe 45 kOe 100 Oe
Gaussmeter HGM09s Operating Instructions
Chapter 5 Operation
Page 26 / 75
5.7
5.75.7
5.7 Measuring
Measuring Measuring
Measuring Unit
UnitUnit
Unit
You select the requested measuring unit by pressing the RANGE
RANGERANGE
RANGE button.
The function of the RANGE
RANGERANGE
RANGE button depends on the setting of the setup
menu under
Units
.
The measurement is carried out either in the unit Tesla, Gauss, kA/m or in Oersted. The display shows the selected measuring unit. The respec­tive range limit value changes according to the selected measuring unit. This is shown in the bottom left display area.
Also see the measuring ranges overview.
Operation via the external interface:
:UNIT {TESL|APM|GAUS|OE}
5.8
5.85.8
5.8 DC/AC Field Measurements
DC/AC Field MeasurementsDC/AC Field Measurements
DC/AC Field Measurements
With the gaussmeter you can carry out both DC field measurements (typical for permanent magnets) and AC field measurements (typical for transformers and similar components). You can moreover record fast single impulses.
Select the measuring mode via RANGE
RANGERANGE
RANGE or via the corresponding entry in
the setup menu.
Operation via the external interface:
:MODE {DC|AC}
Gaussmeter HGM09s Operating Instructions
Chapter 5 Operation
Page 27 / 75
5.8.1
5.8.15.8.1
5.8.1 DC Field Measurements
DC Field MeasurementsDC Field Measurements
DC Field Measurements
When measuring DC fields, the gathered measuring values are integrated via a time interval of 100 milliseconds. Influences of AC magnetic fields are thereby sup­pressed. This applies especially for interspersions of the AC network with a net fre­quency of 50 to 60 hertz.
The measuring values are emitted with positive or negative signs respectively.
Alternatively, the magnetic field polarity can also be displayed to (north/south). You can choose this by selecting the item POLE DISPLAY in the setup menu.
Operation via the external interface:
:MODE DC
5.8.1.1
5.8.1.15.8.1.1
5.8.1.1 DC Field Measurement Characteristics
DC Field Measurement CharacteristicsDC Field Measurement Characteristics
DC Field Measurement Characteristics
Field Strength Accuracy (1σ)
≤ 1.5 T ±0.5 %
> 1.5 T ±1.0 %
Gaussmeter HGM09s Operating Instructions
Chapter 5 Operation
Page 28 / 75
5.8.2
5.8.25.8.2
5.8.2 AC Field Measurements
AC Field MeasurementsAC Field Measurements
AC Field Measurements
In the operating mode AC field measurement the effective value (RMS) is calculated from the determined AC field shares. DC field shares are automatically suppressed in this type of measure­ment.
Note Please note that superimposed DC field shares must not be so large
that they exceed the measuring range. A small AC field together with a DC field may result in a display
overflow
.
Operation via the external interface:
:MODE AC
The accuracy of the AC field measurement depends on the AC field‘s frequency and wave shape.
5.8.2.1
5.8.2.15.8.2.1
5.8.2.1 AC Field Measurement Characteristics
AC Field Measurement CharacteristicsAC Field Measurement Characteristics
AC Field Measurement Characteristics
Maximum ratings for sinusoidal signal
Frequency B
eff
Accuracy (1σ)
≤2 kHz ≤1 T ≤±1.0 % ≤5 kHz ≤2 T ≤±2.0 %
The error of the AC field measurement is composed of the error of the DC field measurement and the error of frequency and shape factors.
Error DC Field Measurement
(1σ)
B ≤1.5 T ≤±0.5 %
B ≥1.5T ≤±1.0 %
Gaussmeter HGM09s Operating Instructions
Chapter 5 Operation
Page 29 / 75
Frequency Response Factor (Sinusoidal Field Pattern)
Frequency Factor
2 kHz 1.00 5 kHz 0.98 7 kHz 0.95
10 kHz 0.90
Shape Factor (Sinusoidal Field Pattern)
Field Strength
B
eff
Factor
700 mT 1.00 1000 mT 1.01 1500 mT 1.02 2000 mT 1.03
The frequency response factor and shape factor are multiplied with the basic accu­racy in dependence of the respective measuring parameters, and then added to the basic accuracy.
Example:
Measuring
B
eff
= 1000 mT at a frequency of 5 kHz.
The basic accuracy is 0.5%. The frequency response factor is 0.9%. The shape factor is 1.02.
The error from of these factors is 0.95 * 1.02 = 0.97 = -3%
For the measurement you have to calculate a total error of -3%±0.5%.
Gaussmeter HGM09s Operating Instructions
Chapter 5 Operation
Page 30 / 75
5.9
5.95.9
5.9 Peak Value Measurement
Peak Value MeasurementPeak Value Measurement
Peak Value Measurement
The device has 2 different operating modes for measuring peak values. They differ in speed, evaluation and resolution.
5.9.1
5.9.15.9.1
5.9.1 Normal Peak Value Recording
Normal Peak Value RecordingNormal Peak Value Recording
Normal Peak Value Recording
For the normal peak value recording of minimum and maximum values the displayed measuring value is analyzed continuously and the lowest or respectively highest value are determined from it.
In this operating mode approx. 10 measurements per second are carried out in high resolution. This enables recordings of slowly changing field sizes. A typical application is the determination of a maximum field strength value of a permanent magnet by manual positioning of the measuring probe on the surface.
Reset the value via the NULL button.
The automatic range selection cannot be used in this operating mode.
The accuracy corresponds to the DC field measurement.
Operation via the external interface:
:PEAK:MODE SLOW
Gaussmeter HGM09s Operating Instructions
Chapter 5 Operation
Page 31 / 75
5.9.2
5.9.25.9.2
5.9.2 Fast Peak Value Recording
Fast Peak Value RecordingFast Peak Value Recording
Fast Peak Value Recording
The fast recording of the maximum values of DC fields is required for short magnetic impulses, as they are generated in e.g. magnetizing systems.
In this operating mode you record peak values of the magnetic field starting from a duration of 10µs. The display shows the absolute highest amount.
Reset the value via the NULL button.
The automatic range selection cannot be used in this operating mode.
Note Please note that in sensitive measuring ranges magnetic AC fields
caused by e.g. transformers or electric lines might lead to a measuring value. Normally, these values would not be noticed during normal op­eration, because disturbances with a net frequency of 50 or 60 hertz are filtered out.
Operation via the external interface:
:PEAK:MODE FAST
5.9.2.1
5.9.2.15.9.2.1
5.9.2.1 Fast Peak Value Measurement Characteristics
Fast Peak Value Measurement CharacteristicsFast Peak Value Measurement Characteristics
Fast Peak Value Measurement Characteristics
Maximum ratings for the fast peak value measurement
Range
≤±1.0 %
≤±2.0 %
10 mT ≤70 Hz ≤100 Hz
100 mT ≤100 Hz ≤150 Hz
1 T ≤300 Hz ≤500 Hz
4.5 T ≤500 Hz (B <1.5T) ≤700 Hz
Gaussmeter HGM09s Operating Instructions
Chapter 5 Operation
Page 32 / 75
The error of the fast peak value measurement consists of the error of the DC field measurement and of a frequency factor that depends on the measuring range.
Error DC Field Measurement
(1σ)
B ≤1.5 T ≤±0.5 %
B ≥1.5T ≤±1.0 %
The range-depending frequency response factor is multiplied by the basic accuracy, depending on the respective signal frequencies and then added to the basic accu­racy.
Example:
Measurement in the range of 1000mT with a frequency of 1kHz.
The basic accuracy lies at 0.5%. The frequency response factor is 0.98 = -2%.
During this measurement, you have to calculate a total error of -2%±0.5%.
The frequency response factor as well as the shape factor are multiplied by the basic accuracy depending on the respective measuring parameters and then added to the basic accuracy.
Range-Depending Frequency Response Factor (Sinusoidal Field Pattern)
Measuring Range 4500 mT 1000 mT 100 mT 10 mT
Frequency 1.00 1.00 1.00 1.00
50 Hz 1.00 1.00 1.00 0.99
70 Hz 1.00 1.00 0.99 0.98 100 Hz 1.00 1.00 0.98 0.93 200 Hz 1.00 1.00 --- --­500 Hz 0.99 0.99 --- --­700 Hz 0.99 0.98 --- ---
1000 Hz 0.98 0.98 --- --­2000 Hz 0.92 --- --- ---
Gaussmeter HGM09s Operating Instructions
Chapter 5 Operation
Page 33 / 75
5.10
5.105.10
5.10 Probe Data
Probe DataProbe Data
Probe Data
For a correct measurement the gaussmeter must always know the necessary probe data. The connected probes contain a parameter memory which stores the probe pa­rameters, the serial number and the labeling. After changing the probe or switching on the device these data are read in automatically.
Gaussmeter HGM09s Operating Instructions
Chapter 6 Setup Menu
Page 34 / 75
6666 Setup Menu
Setup Menu Setup Menu
Setup Menu
You can change the device setup via the setup menu and hence adapt the device to the measurement task in an optimal way. By holding the IO
IOIO
IO
button after switching on, the setup menu appears on the display.
During operation you can also activate the setup menu by holding the RANGE
RANGERANGE
RANGE button for approx. 3 seconds.
The setup menu displays the selected menu item inversely. The individual menu item can be selected by the NULL and RANGE
RANGERANGE
RANGE buttons. On reaching the bottom or first line the menu entries are scrolled further down or up respectively.
You can change the selected menu item by pressing the DATA
DATADATA
DATA button. The indi­vidual possibilities are displayed one after the other. Quit the setup menu by clicking the IO
IOIO
IO button.
Gaussmeter HGM09s Operating Instructions
Chapter 6 Setup Menu
Page 35 / 75
6.1
6.16.1
6.1 Settings
SettingsSettings
Settings
The following setup possibilities can be adapted individually in order to be able to use the device for each application in an optimal way:
6.1.1
6.1.16.1.1
6.1.1 Operating Mode
Operating ModeOperating Mode
Operating Mode of the USB Interface
of the USB Interface of the USB Interface
of the USB Interface
OFF
OFFOFF
OFF
Switched off
The data connection is cut off. The device can however still be supplied with voltage via the USB connector.
COMP
COMPCOMP
COMPUUUUTER
TERTER
TER
Virtual serial con­nection
The interface is activated and the computer can operate the device and read out the measuring values via the SCPI protocol.
KEYB
KEYBKEYB
KEYBOOOOARD
ARDARD
ARD
Keyboard emulation The device acts similar to a key-
board of a connected computer. The data are transmitted after pressing the
DATA
DATADATA
DATA
button.
Operation via the external interface:
:PAR:USB {OFF|KEYB|COMP}
6.1.2
6.1.26.1.2
6.1.2 Selecting the Display Units
Selecting the Display UnitsSelecting the Display Units
Selecting the Display Units
GAUSS
GAUSSGAUSS
GAUSS
Display unit Gauss
KA/M
KA/MKA/M
KA/M
Display unit kA/m
TESLA
TESLATESLA
TESLA
Display unit Tesla
OERSTED
OERSTEDOERSTED
OERSTED
Display unit Oersted
SCROLL
SCROLLSCROLL
SCROLL
All units can be selected by multiple pressing of the
RANGE
RANGERANGE
RANGE
button.
Operation via the external interface:
:PAR:UNIT {ALL|TESL|GAUS|OE|APM}
Gaussmeter HGM09s Operating Instructions
Chapter 6 Setup Menu
Page 36 / 75
6.1.3
6.1.36.1.3
6.1.3 Operating Mode of
Operating Mode of Operating Mode of
Operating Mode of the
the the
the Peak Value Recording
Peak Value RecordingPeak Value Recording
Peak Value Recording
OFF
OFFOFF
OFF
Switched off The peak value recording is
switched off.
FAST
FASTFAST
FAST
Fast recording of max. values
The absolute higher peak value of the measuring value is recorded all 20µms. An automatic range switch is not possible. Only for DC field measurements.
SLOW
SLOWSLOW
SLOW
Slow recording of min. and max. values
The peak measuring values are re­corded and displayed all 100ms. An automatic range switch is not possible. Only for DC field meas­urements.
Operation via the external interface
:PAR:PEAK {OFF|SLOW|FAST}
6.1.4
6.1.46.1.4
6.1.4 DC/AC Field Measurement
DC/AC Field MeasurementDC/AC Field Measurement
DC/AC Field Measurement
BOTH
BOTHBOTH
BOTH
Manual selection via the
RANGE
RANGERANGE
RANGE
Button
AC~
AC~AC~
AC~
AC field measurements Measuring the AC field.
DC=
DC=DC=
DC=
DC field measurements Measuring the DC field. The peak
value measurement is only possible for DC field measurements.
Operation via the external interface
:PAR:ACDC {BOTH|DC|AC}
Gaussmeter HGM09s Operating Instructions
Chapter 6 Setup Menu
Page 37 / 75
6.1.5
6.1.56.1.5
6.1.5 Range Selection
Range SelectionRange Selection
Range Selection
Operation via the external interface
:PAR:RANG {MANU|AUTO}
6.1.6
6.1.66.1.6
6.1.6 Polarity Display (North/South)
Polarity Display (North/South)Polarity Display (North/South)
Polarity Display (North/South)
Operation via the external interface:
:PAR:POLD {OFF|ON}
6.1.7
6.1.76.1.7
6.1.7 Switching off th
Switching off thSwitching off th
Switching off the Device
e Devicee Device
e Device
Operation via the external interface:
:PAR:POFF {MANU|2MIN|5MIN}
MAN
MANMAN
MANUUUUALLY
ALLYALLY
ALLY
Manual range selection via the
RANGE
RANGERANGE
RANGE
button
AUTO
AUTOAUTO
AUTO
Automatic range switch at the range limits
The switch is made at > 90% and < 10% of the measuring range. For peak value meas­urements the automatic range switch is not possible.
OFF
OFFOFF
OFF
Switched off Only the sign is displayed.
ON
ONON
ON
Display of the north and south pole with the sign of the measuring value
For the exact definition of north and south pole, please consult the documentation of the used measuring probe. Only for DC field measurements.
MANUALLY
MANUALLYMANUALLY
MANUALLY
Manual switch off via the IO button
2 MIN
2 MIN2 MIN
2 MIN
Automatic switch off after 2 minutes without Activity
5 MIN
5 MIN5 MIN
5 MIN
Automatic switch off after 5 minutes without activity
The device is switched off when the buttons have not been pressed and the measuring value has not changed signifi­cantly.
Gaussmeter HGM09s Operating Instructions
Chapter 6 Setup Menu
Page 38 / 75
6.1.8
6.1.86.1.8
6.1.8 Charging the Batteries
Charging the BatteriesCharging the Batteries
Charging the Batteries
Operation via the external interface:
:PAR:CHAR {OFF|ON}
6.1.9
6.1.96.1.9
6.1.9 Brightness of Display Illumination
Brightness of Display IlluminationBrightness of Display Illumination
Brightness of Display Illumination
Op­eration via the external interface:
:PAR:LIGH {100|75|50|25|OFF}
6.1.10
6.1.106.1.10
6.1.10 Display Contrast A
Display Contrast ADisplay Contrast A
Display Contrast Adjustment
djustmentdjustment
djustment
Operation via the external interface:
:PAR:CON <0..20>
ON
ONON
ON
The batteries are charged via the power supply or the computer in operation
Charging is only possible when the device is switched on.
OFF
OFFOFF
OFF
The batteries are not charged
The charging is blocked.
25%
25%25%
25%..100%
..100%..100%
..100%
Adjustment of the brightness of the display illumination
OFF
OFFOFF
OFF
Switching off the display illu­mination
A minor brightness en­tails a longer battery life­time.
0%..100%
0%..100%0%..100%
0%..100%
Adjustment of the display contrast in 5% steps
Adjust the contrast corresponding to your lighting conditions.
Gaussmeter HGM09s Operating Instructions
Chapter 1
Page 39 / 75
6.1.11
6.1.116.1.11
6.1.11 Version Remarks
Version RemarksVersion Remarks
Version Remarks
After quitting the setup menu, two display pages emit some information regarding the device. If you want to look at the display longer, keep the IO
IOIO
IO button pressed.
The first page displays:
Information regarding the manu- facturer
Name of the device
Installed option
Serial number
Calibration information
The calibration information consists of the calibration date and the recommended date of the next recalibration.
The second page displays:
Information regarding the used measuring probe
o Type name
o Calibration date
Software status of the device
Hardware status of the device
In the event of a device failure, please see the third page for more detailed explana­tions.
Gaussmeter HGM09s Operating Instructions
Chapter 7 Serial Interface
Page 40 / 75
7777 Serial Interface
Serial InterfaceSerial Interface
Serial Interface
7.1
7.17.1
7.1 IIIIntrodu
ntroduntrodu
ntroduction
ctionction
ction
Via the installed serial interface all functions of the gaussmeter can be piloted by a controller (e.g. a personal computer). The interface can be used in two completely different ways.
On the one hand the gaussmeter can write the data directly into any PC application. In this KEYBOARD mode, the gaussmeter acts similar to a keyboard that is connected to a personal computer. The measuring data are transmitted automatically on press­ing the DATA
DATADATA
DATA button. A special installation of drivers is not necessary. This operating mode works for computers based on Windows as well as for Linux operating sys­tems or for Apple computers. The only requirement for the computer is that it has a connection for keyboards based on USB 2.0. This operating mode is e.g. suitable for filling out Excel sheets easily.
If you want to control the gaussmeter automatically and the connected computer is supposed to actively carry out the data exchange, a connection via a virtual serial in­terface is possible. In this operating mode a USB compliant CDC connection is estab­lished. The gaussmeter acts like an additional serial interface for a user program. However, in this case you must install a driver. For the standard Windows operating systems the driver is included in the delivery.
The data transmission itself, i.e. the protocol, is defined in close connection to the widely-used SCPI programming language for measuring devices.
Trademark of the respective trademark owner
Gaussmeter HGM09s Operating Instructions
Chapter 7 Serial Interface
Page 41 / 75
7.2
7.27.2
7.2 Connecting the Gaussmeter to a Computer
Connecting the Gaussmeter to a Computer Connecting the Gaussmeter to a Computer
Connecting the Gaussmeter to a Computer
7.2.1
7.2.17.2.1
7.2.1 Connector Plug
Connector PlugConnector Plug
Connector Plug
The USB port on the top of the gaussmeter is a Mini-B USB 2.0 socket.
The gaussmeter is connected to the computer via the supplied USB cable.
7.3
7.37.3
7.3 Direct Operation
Direct OperationDirect Operation
Direct Operation on the Computer
on the Computer on the Computer
on the Computer
The gaussmeter is connected to the computer via the USB interface. The device set­ting USB MODE is adjusted to KEYBOARD. The top right area of the display shows KEYB.
The installation of a driver depending on the operating system is generally not re­quired. The operating system solely has to support the HID (Human Interface Device) device class according to USB 2.0. This is the case for standard computers under Windows, Linux or Mac OS.
After starting a suitable program, e.g. Microsoft Excel, the measuring data can be transmitted to the respective data fields by simply pressing the DATA
DATADATA
DATA button. The decimal separator is selected automatically in dependence of the country version of the computer.
The transmitted number of digits corresponds to the number that is monitored on the device display. For normal measurements and fast peak value measurements a measuring value and a carriage return are transmitted. For slow peak value meas­urements the current value and both peak values are emitted separately by the tab character. These three values are then normally placed side by side.
Trademark of the respective trademark owner
Figure 9 USB Connection
USB port
Gaussmeter HGM09s Operating Instructions
Chapter 7 Serial Interface
Page 42 / 75
7.3.1
7.3.17.3.1
7.3.1 Example Normal Measuring Mode in Excel
Example Normal Measuring Mode in ExcelExample Normal Measuring Mode in Excel
Example Normal Measuring Mode in Excel
The marked values were transmitted by the gaussmeter.
Placement
starting in cell B3.
7.3.2
7.3.27.3.2
7.3.2 Example Fast Peak Value Mode in Excel
Example Fast Peak Value Mode in ExcelExample Fast Peak Value Mode in Excel
Example Fast Peak Value Mode in Excel
The marked values were transmitted by the gaussmeter.
Placement
starting in cell B3.
7.3.3
7.3.37.3.3
7.3.3 Example Slow Peak Value Mode in Excel
Example Slow Peak Value Mode in ExcelExample Slow Peak Value Mode in Excel
Example Slow Peak Value Mode in Excel
The marked values were transmitted by the
gaussmeter.
Placement
starting in cell B3.
Trademark of the respective trademark owner
Gaussmeter HGM09s Operating Instructions
Chapter 7 Serial Interface
Page 43 / 75
7.4
7.47.4
7.4 Operation via Interface
Operation via InterfaceOperation via Interface
Operation via Interface
7.4.1
7.4.17.4.1
7.4.1 Installation on the Computer
Installation on the ComputerInstallation on the Computer
Installation on the Computer
For the comprehensive operation on an external computer use the CDC device class (Communication Device Class) of the USB specification. This communication class defines a virtual serial interface on the connected computer. For the user program, the gaussmeter turns out to be an additional serial interface. The operation can be tested by simple terminal programs, as Hyperterm under Windows. Depending on the operating system, the installation of a driver on the computer might become nec­essary. You find more detailed notes on this in the driver installation documents.
7.4.2
7.4.27.4.2
7.4.2 USB Interface Data Format
USB Interface Data FormatUSB Interface Data Format
USB Interface Data Format
The data format of the virtual interface is predefined.
A possible adjustment or change of the parameters, as the transfer rate etc., has no effect. The data flow is stored in the USB protocol. The data are always transmitted with the maximum possible speed.
7.4.3
7.4.37.4.3
7.4.3 Character Set
Character SetCharacter Set
Character Set
The ASCII character set is used. The following control characters are used:
Character Octal Decimal Hex Function
<LF> 12 10 0A End of command line <CR> 15 13 0D Feed <ETX> 3 3 03 Abort
Other control characters can be used to achieve a clear format. They will be ignored.
7.4.4
7.4.47.4.4
7.4.4 Introduction to
Introduction to Introduction to
Introduction to the
the the
the SCPI Language
SCPI LanguageSCPI Language
SCPI Language
The programming language SCPI (Standard Commands for Programmable Devices) defines the way a measuring device (here the gaussmeter) can communicate with a controller. The SCPI language uses a hierarchical structure. The command tree con­sists of root directory commands that are placed on top and several levels below each root directory command. You have to specify the complete path to execute commands of the lower levels.
Trademark of the respective trademark owner
Gaussmeter HGM09s Operating Instructions
Chapter 7 Serial Interface
Page 44 / 75
7.4.4.1
7.4.4.17.4.4.1
7.4.4.1 Command Structure
Command StructureCommand Structure
Command Structure
The commands generally have a short and a long form. In the following descriptions, the short form is set in upper case. The attached long form is set in lower case. Only the characters of the short form are checked for syntactic correctness. Upper and lower case spelling is not distinguished. The same is true for the parameters. A pa­rameter is always emitted in long form and in upper case.
Example:
Command PROBe:POTEntialcoil:LENGth? Send PROBe:POTEntialcoil:LENGth? Send PROB:POTE:LENG? Send PROB:POTEntial:leng?
7.4.4.2
7.4.4.27.4.4.2
7.4.4.2 Path Separator „:“
Path Separator „:“Path Separator „:“
Path Separator „:“
If a colon is the first character of a command key word the next command code is a command of the root directory. If a colon is written between two command codes the colon entails a path to the next lower level of the current path of the command tree. Command codes must be separated from one another by a colon. You can omit the colon at the beginning of a command if the command is the first of a new program line.
7.4.4.3
7.4.4.37.4.4.3
7.4.4.3 Command Separator „;“
Command Separator „;“Command Separator „;“
Command Separator „;“
Several commands within the same command character string are separated by a semicolon. By means of a semicolon the indicated path is not changed. The two fol­lowing statements have the same meaning.
Example:
:IO:DIG:LOGI:IN POS;:IO:DIG:LOGI:OUT NEG :IO:DIG:LOGI:IN POS;OUT NEG
7.4.4.4
7.4.4.47.4.4.4
7.4.4.4 Parameter Separator „,“
Parameter Separator „,“Parameter Separator „,“
Parameter Separator „,“
If you require several parameters in one command, they have to be separated by a comma.
Example:
:PROB:SEAR:AREA 12,QMM
7.4.4.5
7.4.4.57.4.4.5
7.4.4.5 The Use of Blanks
The Use of BlanksThe Use of Blanks
The Use of Blanks
A parameter must be separated from a command key word by blanks (tab or space). Blanks are usually only ignored in parameter lists.
Gaussmeter HGM09s Operating Instructions
Chapter 7 Serial Interface
Page 45 / 75
7.4.4.6
7.4.4.67.4.4.6
7.4.4.6 Query
QueryQuery
Query Commands
Commands Commands
Commands
The controller can send out commands at any time, however a SCPI device (here the gaussmeter) will only answer, if it has expressly been instructed to do so. Only query commands (commands that end with a question mark) prompt the device to send a response. For queries, the device displays either measuring values or internal device settings.
Note
If you send two query commands without having read the reply to the first and thereupon try to read the first response, you might receive some data of the first reply, followed by the complete second re­sponse. Therefore do not send any query command without having read the reply beforehand. Commands and queries should not be sent via the same program line. This might result in an overflow of the raw data buffer, in the event that too many data are created.
7.4.4.7
7.4.4.77.4.4.7
7.4.4.7 System Commands
System CommandsSystem Commands
System Commands
Commands starting with an asterisk are called general commands. The commands with asterisk are used for controlling status operations in the gaussmeter.
7.4.5
7.4.57.4.5
7.4.5 SCPI
SCPISCPI
SCPI Data Types
Data Types Data Types
Data Types
The SCPI data language defines different data formats that are used in the program message and in the reply message. SCPI devices can normally accept commands and parameters in different formats. In particular the numerical parameters can be used very freely. Contrary to this general definition, the data formats are restricted in some places here.
7.4.5.1
7.4.5.17.4.5.1
7.4.5.1 Numerical Parameters
Numerical ParametersNumerical Parameters
Numerical Parameters
Commands that require numerical parameters, accept the generally used decimal notation of numerical characters including leading characters, decimal points and leading zeros. The scientific notation is supported. The technical unit is preset for most commands and will then not be transmitted. For some commands the unit is part of the command being an additional parameter.
Example:
:PROB:SEAR:AREA 12.345E-3,QM :PROB:SEAR:RES 123.5
Gaussmeter HGM09s Operating Instructions
Chapter 7 Serial Interface
Page 46 / 75
7.4.5.2
7.4.5.27.4.5.2
7.4.5.2 Discrete Parameters
Discrete ParametersDiscrete Parameters
Discrete Parameters
Discrete parameters are used in order to program setups that have a limited amount of values. You have a long and a short form for command key words. Upper and lower case can be mixed. Replies to queries are always only emitted in long form and in upper case.
Example:
:PEAK:MODE FAST
7.4.5.3
7.4.5.37.4.5.3
7.4.5.3 Boolean Parameters
Boolean ParametersBoolean Parameters
Boolean Parameters
Boolean parameters represent a single condition that is either true or untrue. As an untrue condition the gaussmeter accepts "OFF" or "0". As a true condition the gauss­meter accepts "ON" or "1". If a Boolean setup is interrogated, the device always emits "0" or "1".
Example:
:DISP:BARG ON :DISP:BARG? Answer:1<cr><lf>
7.4.5.4
7.4.5.47.4.5.4
7.4.5.4 Chara
CharaChara
Character String Parameter
cter String Parametercter String Parameter
cter String Parameter
Character string parameters can, in principle, contain a limited amount of ASCII­characters. A character string must begin and end with either an apostrophe (') or a quotation mark ("), with the same character at the beginning and at the end of the character string. The delimiter can be used within the character string by keying it in twice in a row without blank.
Example:
:PROB:IDEN "PART x" :PROB:IDEN 'PART x'
7.4.5.5
7.4.5.57.4.5.5
7.4.5.5 Entry
Entry Entry
Entry Ending
EndingEnding
Endingssss
Program messages that are sent to the gaussmeter must end with an
<LF>
character.
Likewise a
<CR>
character followed by an
<LF>
character is accepted. By terminating
a message, the current path is always set back to the root directory.
Gaussmeter HGM09s Operating Instructions
Chapter 7 Serial Interface
Page 47 / 75
7.4.5.6
7.4.5.67.4.5.6
7.4.5.6 Output Data
Output DataOutput Data
Output Data Format
Format Format
Format
Output data have the format that is shown in the following chart. Output data always terminate with a <CR> character followed by an
<LF>
character.
Types of Output Data
Output Data Format
Queries
<Parameter><cr><lf>
Text
"Character string" <cr><lf>
Numerical Value
+D.DDDDDDE+DD<cr><lf> (D = 0..9)
Gaussmeter HGM09s Operating Instructions
Chapter 7 Serial Interface
Page 48 / 75
7.4.6
7.4.67.4.6
7.4.6 The SCPI Status Model
The SCPI Status ModelThe SCPI Status Model
The SCPI Status Model
The status system records different device conditions in several register groups. The individual messages are grouped in the several registers. One bit of these registers is related to one message respectively.
The measuring event register stores the messages that are relevant for the measure­ment. The data error register stores the error messages and the standard event regis­ter the SCPI standard messages.
7.4.6.1
7.4.6.17.4.6.1
7.4.6.1 Event Register
Event RegisterEvent Register
Event Register
The event register can only be read. The signal bits are set by the device but are not deleted automatically.
Bits in an event register are deleted either by interrogation of this register (*ESR? or e.g. :STAT:QUES:EVEN?) or by the *CLS command. When interrogating an event register, the device emits a decimal value that corresponds to the sum of the binary place values of all bits that are set in this register.
7.4.6.2
7.4.6.27.4.6.2
7.4.6.2 Release Register
Release RegisterRelease Register
Release Register
The individual bits of the assigned event registers are masked via the release regis­ters. Only the released bits enter into the sum bit as an OR connection. The release registers are read- and writeable. The release registers are not deleted by a query. The *CLR command does not delete the release registers. The command :STATus:PRESet deletes the release register for the error data. When interrogating an event register, the device emits a decimal value that corresponds to the sum of the binary place values of all bits that are set in this register. For setting the bits in a re­lease register, a decimal value is transmitted that corresponds to the sum of the bi­nary place values of all bits set in this register.
7.4.6.3
7.4.6.37.4.6.3
7.4.6.3 Sum Register
Sum RegisterSum Register
Sum Register
In the sum register the results of the event registers are summarized after the mask­ing in the release register and assigned to individual bits. The deletion of the event in the event registers also sets back the respective sum bit in the status register. Via the release register of the sum register the individual sum bits can be re-merged to one whole message service request.
Gaussmeter HGM09s Operating Instructions
Chapter 7 Serial Interface
Page 49 / 75
7.4.6.4
7.4.6.47.4.6.4
7.4.6.4 Overview Status Model
Overview Status ModelOverview Status Model
Overview Status Model
Figure 10 SCPI Status Model
Gaussmeter HGM09s Operating Instructions
Chapter 7 Serial Interface
Page 50 / 75
7.4.6.5
7.4.6.57.4.6.5
7.4.6.5 Bit Definitions
Bit DefinitionsBit Definitions
Bit Definitions
Bit Definitions Sum Register
Bit
Decimal
Value
Definition
0 Measuring Event 1 One or several bits are set in the measuring event regis-
ter and activated in the release register.
3 Error Event 8 One or several bits are set in the error event register and
activated in the release register.
5 Standard Event 32 One or several bits are set in the standard event register
and activated in the release register.
6 Maintenance Request 64 One or several bits are set in the sum register and acti-
vated in the release register.
Bit Definitions Measuring Event Register
Bit
Decimal
Value
Definition
0 Overflow 1 An area overflow has emerged during the measurement.
1 Data available 2 A measurement is terminated. The data are available.
Bit Definitions Error Data Register
Bit
Decimal
Value
Definition
0 DATA Button 1 Is set when the DATA button has been pressed.
1 General
Calibration Error
2 Is set if the calibration data of the measuring probe could
not be read or if the internal calibration data are not con­sistent.
6 Internal
Calibration Error
64 Is set if the internal calibration data are not consistent.
7 Probe
Calibration Error
128 Is set if the calibration data of the measuring probe are
not consistent.
Bit Definitions Standard Event Register
Bit
Decimal
Value
Definition
0 OPC 1 Is set at the end of the SCPI command processing.
5 CME 32 An error has emerged upon the SCPI command process-
ing.
7 PON 128 Is set when the device is ready for use.
Gaussmeter HGM09s Operating Instructions
Chapter 7 Serial Interface
Page 51 / 75
7.5
7.57.5
7.5 Summary of SCPI Commands
Summary of SCPI CommandsSummary of SCPI Commands
Summary of SCPI Commands
The following spellings are used in the SCPI command syntax: optional key words or parameters are indicated in square brackets [ ]. Parameters within a command character string are indicated in braces { }. The parameter indicated in angle brackets < > must be replaced by a value.
7.5.1
7.5.17.5.1
7.5.1 Control Commands
Control CommandsControl Commands
Control Commands
*CLS
Reset the status registers.
*ESE?
Readout the event release register.
*ESR[?]
Readout and reset the standard event register.
*IDN?
Read the identification.
*OPC?
Display "1" for synchronization.
*OPC
Set the event bit "Operation terminated".
*RST
Reset the gaussmeter.
*SRE[?]
Readout and set the status byte release register.
*STB?
Readout the status byte sum register.
:STAT:PRES
Reset the error byte release register.
:STAT:QUES:ENAB
Readout and set the error byte release register.
:STAT:QUES:EVEN
Readout the error byte register.
:STAT:MEAS:ENAB
Readout and set the event release register.
:STAT:MEAS:EVEN
Readout the event register.
7.5.2
7.5.27.5.2
7.5.2 Main Commands
Main CommandsMain Commands
Main Commands
:MEAS?
Display the current measurement.
:READ?
Display the current measurement.
:UNIT[?]
Preset the physical unit.
:MODE[?] {DC|AC}
Preset the DC or AC field operation mode.
:RANG:SET {0|1|2|3}
Preset the measuring range.
:RANG:AUTO
Activate the automatic measuring range switch.
:RANG?
Interrogate the current measuring range.
:NULL
Null compensation of the measuring probe.
7.5.3
7.5.37.5.3
7.5.3 Peak Val
Peak ValPeak Val
Peak Value Function
ue Functionue Function
ue Functionssss
Gaussmeter HGM09s Operating Instructions
Chapter 7 Serial Interface
Page 52 / 75
:PEAK?
Interrogate the current peak value mode.
:PEAK:MODE {OFF|SLOW|FAST}
Select the peak value mode.
:PEAK:NULL
Reset the current peak values.
:PEAK:READ?
Display the stored peak value.
:PEAK:READ:MIN?
Display the stored minimum peak value.
:PEAK:READ:MAX?
Display the stored maximum peak value.
7.5.4
7.5.47.5.4
7.5.4 Probe Functions
Probe FunctionsProbe Functions
Probe Functions
:PROB:NAME?
Interrogate the probe designation.
:PROB:SN?
Interrogate the probe serial number.
:PROB:TYPE?
Interrogate the probe type.
7.5.5
7.5.57.5.5
7.5.5 Parameters
ParametersParameters
Parameters
:PAR:USB[?]
Select the operating mode of the USB interface.
:PAR:UNIT[?]
Select the magnetic unit.
:PAR:PEAK[?]
Select the peak value recording mode.
:PAR:ACDC[?]
Select the DC or AC field measurement.
:PAR:RANGe[?]
Switch on/off the automatic range selection.
:PAR:POLDetect[?]
Switch on/off the north/south pole display.
:PAR:POFF[?]
Set the automatic turn off time.
:PAR:CHARing[?]
Switch on/off the battery charging.
:PAR:LIGHt[?]
Set the brightness of the display illumination.
:PAR:CONTrast[?]
Set the display contrast.
:PAR:SAVE
Store the set parameters.
7.5.6
7.5.67.5.6
7.5.6 Device Functions
Device FunctionsDevice Functions
Device Functions
:SN:UNIT?
Readout the device serial number.
:SN:SW?
Readout the software version.
:SN:HW
Readout the hardware version.
:SN:CALI
Readout the calibration information.
Gaussmeter HGM09s Operating Instructions
Chapter 7 Serial Interface
Page 53 / 75
7.6
7.67.6
7.6 Explanation
ExplanationExplanation
Explanation of the Individual SCPI
of the Individual SCPI of the Individual SCPI
of the Individual SCPI Commands
Commands Commands
Commands
7.6.1
7.6.17.6.1
7.6.1 Control Commands
Control CommandsControl Commands
Control Commands
7.6.1.1
7.6.1.17.6.1.1
7.6.1.1 *CLS
*CLS*CLS
*CLS
Description: Resets the status register.
Mode: Command
Parameter: None
*RST
value: Not relevant
Example:
send *CLS<LF>
7.6.1.2
7.6.1.27.6.1.2
7.6.1.2 *ESE[?]
*ESE[?]*ESE[?]
*ESE[?]
Description: Reads out the event register. The gaussmeter displays a decimal
value that corresponds to the sum of the binary place values of all set bits in this register.
Mode: Command and query
Parameter: {<value>} (within the range 0..255)
*RST
value: Not relevant
Example:
send *ESE 22<LF> send *ESE?<LF> receive 22<CR><LF>
7.6.1.3
7.6.1.37.6.1.3
7.6.1.3 *ESR[?]
*ESR[?]*ESR[?]
*ESR[?]
Description: Reads out and resets the standard event register. The gaussmeter
displays a decimal value that corresponds to the sum of the binary place values of all set bits in this register.
Mode: Command and query
Parameter: {<value>} (in the range 0..255)
*RST
value: Not relevant
Example:
send *ESR?<LF>
receive 160<CR><LF>
Gaussmeter HGM09s Operating Instructions
Chapter 7 Serial Interface
Page 54 / 75
7.6.1.4
7.6.1.47.6.1.4
7.6.1.4 *IDN?
*IDN?*IDN?
*IDN?
Description: Reads the gaussmeter identification character string. The gaussmeter
displays the following identification text:
Mode: Query
Parameter: None
*RST
value: Not relevant
Example:
send *IDN?<LF> receive MAGSYS-MAGNET-SYSTEME,HGM09,0,150310,VI<CR><LF>
7.6.1.5
7.6.1.57.6.1.5
7.6.1.5 *OPC?
*OPC?*OPC?
*OPC?
Description: Issue of "1" to the output buffer after the command has been exe-
cuted. The command *OPC? can be placed at the end of a com­mand line in order to synchronize the controller with the command processing. The command itself has no function.
Mode: Query
Parameter: None
*RST
value: Not relevant
Example:
send *OPC?<LF> receive 1<CR><LF>
7.6.1.6
7.6.1.67.6.1.6
7.6.1.6 *OPC
*OPC*OPC
*OPC
Description: Sets the bit "Operation terminated" (bit 0) in the standard event regis-
ter after the command has been executed.
Mode: Command
Parameter: None
*RST
value: Not relevant
Example:
send *OPC<LF>
7.6.1.7
7.6.1.77.6.1.7
7.6.1.7 *RTS
*RTS*RTS
*RTS
Description: The gaussmeter is reset to the initial state. The internal parameters
are set to the preset value. The device carries out a complete reset.
Gaussmeter HGM09s Operating Instructions
Chapter 7 Serial Interface
Page 55 / 75
As the device interface is also reset, further commands on the inter­face might get lost.
Mode: Command
Parameter: None
*RST
value: Not relevant
Example:
send *RTS<LF>
7.6.1.8
7.6.1.87.6.1.8
7.6.1.8 *SRE[?]
*SRE[?]*SRE[?]
*SRE[?]
Description: Reads out and sets the status byte release register. The gaussmeter
displays a decimal value that corresponds to the sum of the binary place values of all set bits in this register.
Mode: Command and query
Parameter: {<value>} (within the range 0..255)
*RST
value: Not relevant
Example:
send *SRE<LF> receive 76<CR><LF>
7.6.1.9
7.6.1.97.6.1.9
7.6.1.9 *STB?
*STB?*STB?
*STB?
Description: Reads out the status byte sum register.
Mode: Query
Parameter: None
*RST
value: Not relevant
Example:
send *ESR 23<LF> send *STB?<LF> receive 0<CR><LF>
7.6.1.10
7.6.1.107.6.1.10
7.6.1.10 :STAT:PRESet
:STAT:PRESet:STAT:PRESet
:STAT:PRESet
Description: Resets the error byte release register.
Mode: Command
Parameter: None
*RST
value: Not relevant
Example:
send :STAT:PRES<LF>
Gaussmeter HGM09s Operating Instructions
Chapter 7 Serial Interface
Page 56 / 75
7.6.1.11
7.6.1.117.6.1.11
7.6.1.11 :STAT:QUES:ENABle[?]
:STAT:QUES:ENABle[?]:STAT:QUES:ENABle[?]
:STAT:QUES:ENABle[?]
Description: Reads out and sets the error byte release register. The gaussmeter
displays a decimal value that corresponds to the sum of the binary place values of all set bits in this register.
Mode: Command and query
Parameter: None
*RST
value: Not relevant
Example:
send :STAT:QUES:ENAB 64<LF> send :STAT:QUES:ENAB?<LF> receive 64<CR><LF>
7.6.1.12
7.6.1.127.6.1.12
7.6.1.12 :STAT:QUES:EVENt?
:STAT:QUES:EVENt?:STAT:QUES:EVENt?
:STAT:QUES:EVENt?
Description: Reads out the error byte register. The gaussmeter displays a decimal
value that corresponds to the sum of the binary place values of all set bits in this register.
Mode: Query
Parameter: None
*RST
value: Not relevant
Example:
send :STAT:QUES:EVEN?<LF> receive 0<CR><LF>
7.6.1.13
7.6.1.137.6.1.13
7.6.1.13 :STAT:MEAS:ENABle[?]
:STAT:MEAS:ENABle[?]:STAT:MEAS:ENABle[?]
:STAT:MEAS:ENABle[?]
Description: Reads out and sets the event release register. The gaussmeter dis-
plays a decimal value that corresponds to the sum of the binary place values of all set bits in this register.
Mode: Command and query
Parameter: None
*RST
value: Not relevant
Example:
send :STAT:MEAS:ENAB 123<LF> send :STAT:MEAS:ENAB?<LF> receive 123<CR><LF>
Gaussmeter HGM09s Operating Instructions
Chapter 7 Serial Interface
Page 57 / 75
7.6.1.14
7.6.1.147.6.1.14
7.6.1.14 :STAT:MEAS:EVENt?
:STAT:MEAS:EVENt?:STAT:MEAS:EVENt?
:STAT:MEAS:EVENt?
Description: Reads out the event register. The gaussmeter displays a decimal
value that corresponds to the sum of the binary place values of all set bits in this register.
Mode: Query
Parameter: None
*RST
value: Not relevant
Example:
send :STAT:MEAS:EVEN?<LF> receive 2<CR><LF>
7.6.2
7.6.27.6.2
7.6.2 Main Commands
Main CommandsMain Commands
Main Commands
7.6.2.1
7.6.2.17.6.2.1
7.6.2.1 :MEAS?
:MEAS?:MEAS?
:MEAS?
Description: Emits the current measurement (Like :READ?)
Mode: Query
Parameter: None
*RST
value: Not relevant
Example:
send :MEAS?<LF> receive 2.546313e-01<CR><LF>
7.6.2.2
7.6.2.27.6.2.2
7.6.2.2 :MODE[?]
:MODE[?]:MODE[?]
:MODE[?]
Description: The DC or AC field operation mode is preset.
Mode: Command and query
Parameter: {DC | AC}
DC DC operation mode
AC AC operation mode
*RST
value: DC
Example:
send :MODE AC<LF> send :MODE?<LF> receive DC<CR><LF>
Gaussmeter HGM09s Operating Instructions
Chapter 7 Serial Interface
Page 58 / 75
7.6.2.3
7.6.2.37.6.2.3
7.6.2.3 :NULL
:NULL:NULL
:NULL
Description: Null compensation of the measuring probe. This function should
only be carried out in sufficiently field-free areas.
Mode: Command
Parameter: None
*RST
value: Not relevant
Example:
send :NULL<LF>
7.6.2.4
7.6.2.47.6.2.4
7.6.2.4 :RANGe:SET
:RANGe:SET:RANGe:SET
:RANGe:SET
Description: The gaussmeter measuring range is preset.
Mode: Command
Parameter: { 0 | 1 | 2 | 3 }
0 Most sensitive range
3 Most insensitive range
*RST
value: Not relevant
Example:
send :RANG:SET 2<LF>
7.6.2.5
7.6.2.57.6.2.5
7.6.2.5 :RANGe:AUTO
:RANGe:AUTO:RANGe:AUTO
:RANGe:AUTO
Description: The automatic measuring range switch is activated.
Mode: Command
Parameter: None
*RST
value: Not activated
Example:
send :RANG:AUTO<LF>
Gaussmeter HGM09s Operating Instructions
Chapter 7 Serial Interface
Page 59 / 75
7.6.2.6
7.6.2.67.6.2.6
7.6.2.6 :RANGe
:RANGe:RANGe
:RANGe????
Description: Interrogates the current measuring range. 0 = most sensitive range.
Mode: Query
Parameter: None
*RST
value: Not relevant
Example:
send :RANG?<LF> receive 3<CR><LF>
7.6.2.7
7.6.2.77.6.2.7
7.6.2.7 :READ?
:READ?:READ?
:READ?
Description: Displays the current measurement.
Mode: Query
Parameter: None
*RST
value: Not relevant
Example:
send :READ?<LF> receive 2.546313e-01<CR><LF>
7.6.2.8
7.6.2.87.6.2.8
7.6.2.8 :UNIT[?]
:UNIT[?]:UNIT[?]
:UNIT[?]
Description: The physical unit of the measuring value is preset or interrogated.
Mode: Command
Parameter: {
TESL|APM|GAUS|OE|G|T
}
TESL T
Unit is Tesla
APM
Unit is A/m
GAUS G
Unit is Gauss
OE Unit is Oersted
*RST
value: Tesla
Example:
send :UNIT TESL<LF> send :UNIT?<LF> receive TESL<CR><LF>
Gaussmeter HGM09s Operating Instructions
Chapter 7 Serial Interface
Page 60 / 75
7.6.3
7.6.37.6.3
7.6.3 Peak Value Function
Peak Value FunctionPeak Value Function
Peak Value Function
7.6.3.1
7.6.3.17.6.3.1
7.6.3.1 :P
:P:P
:PEAK?
EAK?EAK?
EAK?
Description: Interrogates the current peak value mode.
Mode: Query
Parameter: None
*RST
value: Not relevant
Example:
receive OFF<CR><LF>
7.6.3.2
7.6.3.27.6.3.2
7.6.3.2 :PEAK:MODE[?]
:PEAK:MODE[?]:PEAK:MODE[?]
:PEAK:MODE[?]
Description: Selects and interrogates the peak value mode.
Mode: Command and query
Parameter: {OFF | SLOW | FAST}
OFF
No peak value recording
SLOW
Slow peak value recording
FAST
Fast peak value recording
*RST
value: OFF
Example:
send :PEAK:MODE SLOW<LF>
7.6.3.3
7.6.3.37.6.3.3
7.6.3.3 :PEAK:NULL
:PEAK:NULL:PEAK:NULL
:PEAK:NULL
Description: Resets the current peak values.
Mode: Command
Parameter: None
*RST
value: Not relevant
Example:
send :PEAK:MODE NULL<LF>
Gaussmeter HGM09s Operating Instructions
Chapter 7 Serial Interface
Page 61 / 75
7.6.3.4
7.6.3.47.6.3.4
7.6.3.4 :PEAK:READ?
:PEAK:READ?:PEAK:READ?
:PEAK:READ?
Description: The stored peak value is displayed. For SlowPeak the absolute larger
peak value with signs; for FastPeak the peak value; for normal measurement 0 is emitted.
Mode: Query
Parameter: None
*RST
value: Not relevant
Example:
send :PEAK:READ?<LF> receive -4.761955e-02<CR><LF>
7.6.3.5
7.6.3.57.6.3.5
7.6.3.5 :PEAK:READ:MIN?
:PEAK:READ:MIN?:PEAK:READ:MIN?
:PEAK:READ:MIN?
Description: The stored minimum peak value is displayed. For SlowPeak the
smaller peak value of max/min; for FastPeak the absolute larger peak value with signs; for normal measurement 0 is emitted.
Mode: Query
Parameter: None
*RST
value: Not relevant
Example:
send :PEAK:READ:MAX?<LF> receive 7.187624e-02<CR><LF>
7.6.3.6
7.6.3.67.6.3.6
7.6.3.6 :PEAK:READ:MAX?
:PEAK:READ:MAX?:PEAK:READ:MAX?
:PEAK:READ:MAX?
Description: The stored maximum peak value is displayed. For SlowPeak the lar-
ger peak value of max/min; for FastPeak the absolute larger peak value with signs; for normal measurement 0 is emitted.
Mode: Query
Parameter: None
*RST
value: Not relevant
Example:
send :PEAK:READ:MIN?<LF> receive -2.711216e-02<CR><LF>
Gaussmeter HGM09s Operating Instructions
Chapter 7 Serial Interface
Page 62 / 75
7.6.4
7.6.47.6.4
7.6.4 Probe Functions
Probe FunctionsProbe Functions
Probe Functions
7.6.4.1
7.6.4.17.6.4.1
7.6.4.1 :PROB:NAME?
:PROB:NAME?:PROB:NAME?
:PROB:NAME?
Description: Interrogates the probe name.
Mode: Query
Parameter: None
*RST
value: Not relevant
Example:
send :PROB:NAME?<LF> receive "HGM09 Probe T02.047.33.13 "<CR><LF>
7.6.4.2
7.6.4.27.6.4.2
7.6.4.2 :PROB:SN?
:PROB:SN?:PROB:SN?
:PROB:SN?
Description: Interrogates the probe serial number.
Mode: Query
Parameter: None
*RST
value: Not relevant
Example:
send :PROB:SN?<LF> receive "121109070"<CR><LF>
7.6.4.3
7.6.4.37.6.4.3
7.6.4.3 :PROB:TYPE?
:PROB:TYPE?:PROB:TYPE?
:PROB:TYPE?
Description: Interrogates the probe type.
Mode: Query
Parameter: None
*RST
value: Not relevant
Example:
send :PROB:TYPE?<LF> receive 0<CR><LF>
Gaussmeter HGM09s Operating Instructions
Chapter 7 Serial Interface
Page 63 / 75
7.6.5
7.6.57.6.5
7.6.5 Parameters
ParametersParameters
Parameters
7.6.5.1
7.6.5.17.6.5.1
7.6.5.1 :PAR:USB[?]
:PAR:USB[?]:PAR:USB[?]
:PAR:USB[?]
Description: Selects the USB interface operating mode. A change of this parame-
ter has an effect only after the next switching-on of the device. The changes must be stored by the command :PAR:SAVE, if necessary.
Mode: Command and query
Parameter: {OFF | KEYB | COMP}
OFF No data connection
KEYB Keyboard emulation
COMP Connection via interface
*RST
value: Not relevant
Example:
send :PAR:USB COMP<LF> send :PAR:USB?<LF> receive COMP<CR><LF>
7.6.5.2
7.6.5.27.6.5.2
7.6.5.2 :PAR:UNIT[?]
:PAR:UNIT[?]:PAR:UNIT[?]
:PAR:UNIT[?]
Description: Selects the magnetic unit. For ALL, the unit is also switched by means
of the RANGE
RANGERANGE
RANGE button. The changes must be stored by the command
:PAR:SAVE, if necessary.
Mode: Command and query
Parameter: {ALL | TESL | GAUS | OE | APM}
TESL T
Unit is Tesla
APM
Unit is A/m
GAUS G
Unit is Gauss
OE Unit is Oersted
*RST
value: Not relevant
Example:
send :PAR:UNIT ALL<LF> send :PAR:UNIT?<LF> receive ALL<CR><LF>
Gaussmeter HGM09s Operating Instructions
Chapter 7 Serial Interface
Page 64 / 75
7.6.5.3
7.6.5.37.6.5.3
7.6.5.3 :PAR:PEAK[?]
:PAR:PEAK[?]:PAR:PEAK[?]
:PAR:PEAK[?]
Description: Selects the peak value acquisition mode. The changes must be
stored by the command :PAR:SAVE, if necessary.
Mode: Command and query
Parameter: {OFF | SLOW | FAST}
OFF
No peak value recording
SLOW
Slow peak value recording
FAST
Fast peak value recording
*RST
value: Not relevant
Example:
send :PAR:PEAK SLOW<LF> send :PAR:PEAK?<LF> receive SLOW<CR><LF>
7.6.5.4
7.6.5.47.6.5.4
7.6.5.4 :PAR:ACDC[?]
:PAR:ACDC[?]:PAR:ACDC[?]
:PAR:ACDC[?]
Description: Selects the DC or AC field measurement. For BOTH, the DC/AC
field mode is also switched by the RANGE
RANGERANGE
RANGE button. The changes must
be stored by the command :PAR:SAVE, if necessary.
Mode: Command and query
Parameter: {BOTH | DC | AC}
BOTH Selection via the RANGE
RANGERANGE
RANGE button
DC DC field operating mode
AC AC field operating mode
*RST
value: Not relevant
Example:
send :PAR:ACDC DC<LF> send :PAR:ACDC?<LF> receive DC<CR><LF>
Gaussmeter HGM09s Operating Instructions
Chapter 7 Serial Interface
Page 65 / 75
7.6.5.5
7.6.5.57.6.5.5
7.6.5.5 :PAR:RANGe[?]
:PAR:RANGe[?]:PAR:RANGe[?]
:PAR:RANGe[?]
Description: Switches on/off the automatic range selection. The changes must be
stored by the command :PAR:SAVE, if necessary.
Mode: Command and query
Parameter: {MANU | AUTO}
MANU Selection via the RANGE
RANGERANGE
RANGE button
AUTO Automatic range selection
*RST
value: Not relevant
Example:
send :PAR:RANG MANU<LF> send :PAR:RANG?<LF> receive MANU<CR><LF>
7.6.5.6
7.6.5.67.6.5.6
7.6.5.6 :PAR:POLDetect[?]
:PAR:POLDetect[?]:PAR:POLDetect[?]
:PAR:POLDetect[?]
Description: Switches on/off the north/south pole display. The changes must be
stored by the command :PAR:SAVE, if necessary.
Mode: Command and query
Parameter: {OFF | ON}
OFF Switch off the pole display
ON Switch on the pole display
*RST
value: Not relevant
Example:
send :PAR:POLD ON<LF> send :PAR:POLD?<LF> receive OFF<CR><LF>
7.6.5.7
7.6.5.77.6.5.7
7.6.5.7 :PAR:POFF[?]
:PAR:POFF[?]:PAR:POFF[?]
:PAR:POFF[?]
Description: Sets the turn-off time or switches off the function respectively. The
device is switched off automatically after a determined period of in­activity. The changes must be stored by the command :PAR:SAVE, if necessary.
Mode: Command and query
Parameter: {MANU | 2MIN | 5MIN}
Gaussmeter HGM09s Operating Instructions
Chapter 7 Serial Interface
Page 66 / 75
MANU Switch off the automatic turn-off
2MIN Automatic turn-off after 2 minutes
5MIN Automatic turn-off after 5 minutes
*RST
value: Not relevant
Example:
send :PAR:POFF MANU<LF> send :PAR:POFF?<LF> receive MANU<CR><LF>
7.6.5.8
7.6.5.87.6.5.8
7.6.5.8 :PAR:CHARing[?]
:PAR:CHARing[?]:PAR:CHARing[?]
:PAR:CHARing[?]
Description: Switches on/off the battery charging. The changes must be stored by
the command :PAR:SAVE, if necessary.
Mode: Command and query
Parameter: {OFF | ON}
OFF Switch off the battery charging
ON Switch on the battery charging
*RST
value: Not relevant
Example:
send :PAR:CHAR OFF<LF> send :PAR:CHAR?<LF> receive OFF<CR><LF>
7.6.5.9
7.6.5.97.6.5.9
7.6.5.9 :PAR:LIG
:PAR:LIG:PAR:LIG
:PAR:LIGHt[?]
Ht[?]Ht[?]
Ht[?]
Description: Sets the brightness of the display illumination or switches off the il-
lumination respectively. The changes must be stored by the com­mand :PAR:SAVE, if necessary.
Mode: Command and query
Parameter: {100 | 75 | 50 | 25 | OFF}
25 .. 100 Brightness of the display illumination in %
OFF Switch off the display illumination
*RST
value: Not relevant
Gaussmeter HGM09s Operating Instructions
Chapter 7 Serial Interface
Page 67 / 75
Example:
send :PAR:LIGH 75<LF> send :PAR:LIGH?<LF> receive 100<CR><LF>
7.6.5.10
7.6.5.107.6.5.10
7.6.5.10 :PAR:CONTrast[?]
:PAR:CONTrast[?]:PAR:CONTrast[?]
:PAR:CONTrast[?]
Description: Sets the display contrast. The changes must be stored by the com-
mand :PAR:SAVE, if necessary. The value corresponds to 5% steps.
Mode: Command and query
Parameter: {<value>} (within the range 0..20)
0 .. 20 Display contrast in 5%
*RST
value: Not relevant
Example:
send :PAR:CONT 15<LF> send :PAR:CONT?<LF> receive 11<CR><LF>
7.6.5.11
7.6.5.117.6.5.11
7.6.5.11 :PAR:SAVE
:PAR:SAVE:PAR:SAVE
:PAR:SAVE
Description: Stores the set parameters.
Mode: Command
Parameter: None
*RST
value: Not relevant
Example:
send :PAR:SAVE<LF>
7.6.6
7.6.67.6.6
7.6.6 Device Functions
Device FunctionsDevice Functions
Device Functions
7.6.6.1
7.6.6.17.6.6.1
7.6.6.1 :SN:UNIT?
:SN:UNIT?:SN:UNIT?
:SN:UNIT?
Description: Emits the serial number of the device.
Mode: Query
Parameter: None
*RST
value: Not relevant
Example:
send :SN:UNIT?<LF> receive 010110078<CR><LF>
Gaussmeter HGM09s Operating Instructions
Chapter 7 Serial Interface
Page 68 / 75
7.6.6.2
7.6.6.27.6.6.2
7.6.6.2 :SN:SW?
:SN:SW?:SN:SW?
:SN:SW?
Description: Emits the software version.
Mode: Query
Parameter: None
*RST
value: Not relevant
Example:
send :SN:SW?<LF> receive 180310<CR><LF>
7.6.6.3
7.6.6.37.6.6.3
7.6.6.3 :SN:HW
:SN:HW:SN:HW
:SN:HW
Description: Reads out the hardware version.
Mode: Query
Parameter: None
*RST
value: Not relevant
Example:
send :SN:HW?<LF> receive VI<CR><LF>
7.6.6.4
7.6.6.47.6.6.4
7.6.6.4 :SN:CALI
:SN:CALI:SN:CALI
:SN:CALI
Description: Reads out the calibration information.
Mode: Query
Parameter: None
*RST
value: Not relevant
Example:
send :SN:CALI?<LF> receive 01JAN10 / 01JAN12<CR><LF>
Gaussmeter HGM09s Operating Instructions
Chapter 8 Unit Conversion Table
Page 69 / 75
8888 Unit
Unit Unit
Unit Conversion Table
Conversion Table Conversion Table
Conversion Table
This table shows the relationship between the displayed measuring values.
Size Unit Display Conversion
Mag. flux density B SI Tesla T
GT =4101
Flux density
C
GS
Gauss G
TG =4101
Mag. field strength H SI Ampere/meter A/m
OeAm
1000
4
1
1
π
=
Magnetic field strength
C
GS
Oersted Oe
1
4
1000
1−= AmOe
π
Gaussmeter HGM09s Operating Instructions
Chapter 9 Technical Data
Page 70 / 75
9999 Technical Data
Technical DataTechnical Data
Technical Data
General
GeneralGeneral
General
Power Supply Power supply unit 100..240 VAC, 50/60Hz, 0.3 A
max
USB interface Battery 2 x AA 1.2 V NiMH (rechargeable)
Power Consumption approx. 2.5 W
Environmental Conditions -10°C to 40°C;
< 80% relative humidity at 40°C non-condensing
Storage -40°C to 70°C
Housing Dimensions approx. 145 x 80 x 40 mm3
Weight approx. 250 g (incl. batteries, without probe)
Warranty 2 years
Accessories Power supply unit, data carrier with user manual, USB cable
Mathematical Functions Conversion of units and derived parameters
Linearization of the probe measurement
Data Protocol SCPI (standard commands for programmable measuring devices)
External Interfaces USB 2.0
Display Graphical, high-contrast LCD contrast adjustable via menu
Parameter Memory in
Probes
Storage of the calibration values
Measurement Prope
Measurement PropeMeasurement Prope
Measurement Properrrrties
tiesties
ties
Measuring Method Continuous acquisition of the magnetic field; conversion via 16 Bit A/D-
converter; analysis via 16 bit microprocessor system.
Display Updating Time Measuring values: approx. 100ms;
Display Resolution 3- to 4-digit, range-depending
Frequency Range DC/AC 0 Hz..5 kHz (effective value)
Measurement accuracy Accuracy
(1σ)
B Error ≤1.5 T ≤±0.5 %
DC Field Measurement
> 1.5 T ≤±1.0 %
Frequency B
eff
≤2 kHz ≤1 T ≤±1.0 %
AC Field Measurement
See Text
≤5 kHz ≤2 T ≤±2.0 %
Range
≤±1.0 %
≤±2.0 %
10 mT ≤70 Hz ≤100 Hz
100 mT ≤100 Hz ≤150 Hz
1 T ≤300 Hz ≤500 Hz
Peak Value Measurement
See Text
4.5 T ≤500 Hz (B <1.5T)
≤700 Hz
Gaussmeter HGM09s Operating Instructions
Chapter 9 Technical Data
Page 71 / 75
Units Tesla Gauss
Oersted Ampere/meter
4.5 T (1 mT)
45 kG (10 G)
45 kOe (10 Oe)
3800 kA/m (1 kA/m)
1 T (100 µT)
10 kG (1 G)
10 kOe (1 Oe)
1000 kA/m (100 A/m)
100 mT (10 µT)
1 kG (100 mG)
1 kOe (100 mOe)
100 kA/m (10 A/m)
Measuring Ranges
(Resolution)
DC Field Measurement
10 mT (1 µT)
100 G (10 mG)
100 Oe (10 mOe)
10 kA/m (1 A/m)
3.0 T (10 mT)
30 kG (100 G)
30 kOe (100 Oe)
2500 kA/m (10 kA/m)
1 T (1 mT)
10 kG (10 G)
10 kOe (10 Oe)
1000 kA/m (1 kA/m)
100 mT (100 µT)
1 kG (1 G)
1 kOe (1 Oe)
100 kA/m (100 A/m)
Measuring Ranges
(Resolution)
AC Field Measurement
10 mT (10 µT)
100 G (100 mG)
100 Oe (100 mOe)
10 kA/m (10 A/m)
4.5 T (10 mT)
45 kG (100 G)
45 kOe (100 Oe)
3800 kA/m (10 kA/m)
1 T (1 mT)
10 kG (10 G)
10 kOe (10 Oe)
1000 kA/m (1 kA/m)
100 mT (100 µT)
1 kG (1 G)
1 kOe (1 Oe)
100 kA/m (100 A/m)
Measuring Ranges
(Resolution)
Fast Pulse Measurement
10 mT (10 µT)
100 G (100 mG)
100 Oe (100 mOe)
10 kA/m (10 A/m)
Peak Hold t
signal
> 250 µs Standard transversal probe N (incl. in delivery): dimensions approx. 3.5 x 1.35 x 46 mm3 Axial probe: dimensions approx. 4 mm x 60 mm Transversal probe S: dimensions approx. 3.5 x 0.75 x 46 mm3
Probes
(Special Designs on Request)
All probes: Active range 0.15mm Handle bar approx. 11 mm x 100mm Cable length: 1.5m (special lengths available) Integrated parameter memory
Environmental Conditions
Operating Environment Specified accuracy for 0 °C up to 40 °C
Relative Operating Humidity
Up to 80 % relative humidity for temperatures up to 30 °C, linearly decreasing to 50 % relative humidity at 40 °C.
Storage Environment -20 °C up to 70 °C
Altitude 0 – 2000 meters as per IEC 61010-1 2nd Edition CAT III, 1000 V
Degree of Pollution Degree of pollution II
Notes: The technical data apply for a one-hour warm-up phase.
Gaussmeter HGM09s Operating Instructions
Chapter 10 Declaration of Conformity
Page 72 / 75
10
1010
10 Declaration of Conformity
Declaration of ConformityDeclaration of Conformity
Declaration of Conformity
EC Conformity Declaration in accordance with the EC Guid
EC Conformity Declaration in accordance with the EC GuidEC Conformity Declaration in accordance with the EC Guid
EC Conformity Declaration in accordance with the EC Guideeeelines
lineslines
lines
EG-Konformitätserklärung im Sinne der EG-Richtlinie Certificat de conformité en accord avec les directives Européennes
02-Dec-2009
Date /
Datum
M. Kopka Dipl.Ing. CEO
For further information, please contact your local Goudsmit Magnetic Systems sales office, agent or distributor, or Goudsmit Magnetic Systems, Petunialaan 19, 5582HA, Waalre, Netherlands. www.goudsmit-magnetics.nl
Für weitere Informationen kontaktieren Sie bitte Ihr örtliches Goudsmit Magnetic Systems Vertriebsbüro, Handelsvertreter oder Händler oder direkt
Goudsmit Magnetic Systems, Petunialaan 19, 5582HA, Waalre, Netherlands. www.goudsmit-magnetics.nl
Pour de plus amples informations, merci de prendre contact avec notre représentant local. Vous pouvez également nous contacter à l’adresse suivante:
Goudsmit Magnetic Systems, Petunialaan 19, 5582HA, Waalre, Netherlands. www.goudsmit-magnetics.nl
Manufact
ManufactManufact
Manufactuuuurer’s Name:
rer’s Name:rer’s Name:
rer’s Name:
Hersteller Name: Fabricant:
MAGSYS magnet systeme GmbH
Manufacturer’s Address:
Manufacturer’s Address:Manufacturer’s Address:
Manufacturer’s Address:
Hersteller Anschrift: Adresse:
Rohwedderstr. 7 44369 Dortmund – Germany
Declares under sole responsibility that the product as originally deliv
Declares under sole responsibility that the product as originally delivDeclares under sole responsibility that the product as originally deliv
Declares under sole responsibility that the product as originally deliveeeered
redred
red
Erklärt unter der Vorraussetzung, dass das Gerät dem Auslieferungszustand entsprechend Déclarons sous notre seule responsabilité que le produit
Product Name:
Product Name:Product Name:
Product Name:
Produkt Name: Nom du produit:
MAGSYS HAN
MAGSYS HANMAGSYS HAN
MAGSYS HANDDDDGAUSSMETER
GAUSSMETERGAUSSMETER
GAUSSMETER
Model Nu
Model NuModel Nu
Model Nummmmber:
ber:ber:
ber:
Model Nummer: Référence commerciale:
HGM09s
HGM09sHGM09s
HGM09s
Product Opt
Product OptProduct Opt
Product Optiiiions:
ons:ons:
ons:
Produkt Optionen: Options du produit
This declaration covers all options of the above pr
This declaration covers all options of the above prThis declaration covers all options of the above pr
This declaration covers all options of the above prooooduct
ductduct
duct
Diese Erklärung erfasst alle Optionen des Gerätes Cette déclaration couvre toutes les options du produit concerné
has been designed, constructed and manufactured in accordance with
has been designed, constructed and manufactured in accordance withhas been designed, constructed and manufactured in accordance with
has been designed, constructed and manufactured in accordance with the listed EC guidelines in their latest version
the listed EC guidelines in their latest versionthe listed EC guidelines in their latest version
the listed EC guidelines in their latest version
entwickelt, konstruiert und gefertigt wurde in Übereinstimmung mit den aufgeführten EG-Richtlinien in der jeweils letzten gültigen Fassung concerné par cette déclaration a été conçue et fabriqué selon les directives
Directive basse tension (2006/95/EC) Directive C.E.M. (2004/108/EC) and con
and conand con
and conforms with the following product standards:
forms with the following product standards:forms with the following product standards:
forms with the following product standards:
und den folgenden Vorschriften entspricht: et en conformité avec les normes suivantes:
EMC
EMCEMC
EMC
EMV C.E.M.
Standard
StandardStandard
Standard
Norm Norme
Limits
LimitsLimits
Limits
Grenzen Limites
IEC 61000-4-2:1995 / EN 61000-4-2:1995
4kV CD, 8kV AD
IEC 61000-4-3:1995 / EN 61000-4-2:1995
3 V/m, 80-1000 MHz
for the power supply
for the power supplyfor the power supply
for the power supply
für das Netzteil pour l’alimentation
EN 55022 Class B / EN61000-3-2,3 EN 55024 / EN 61000-4­2,3,4,5,6,8,11
Safety
SafetySafety
Safety
Sicherheit Sécurité
IEC 61010-1:2001 / EN 61010-1:2001
for the po
for the pofor the po
for the power supply
wer supplywer supply
wer supply
für das Netztel pour l’alimentation
TÜV EN 60950-1:2000 +A11 / CE
Gaussmeter HGM09s Operating Instructions
Chapter 11 Warranty and Copyright
Page 73 / 75
11
1111
11 Warranty and Copyright
Warranty and CopyrightWarranty and Copyright
Warranty and Copyright
Copyright
CopyrightCopyright
Copyright
© Copyright 2010 Goudsmit Magnetic Systems All rights reserved. No part of the manual or the device in-
cluding its control program (software) is allowed to be re­produced or duplicated without written authorization of the author.
Liability
LiabilityLiability
Liability
We do not take any liability for the correctness of this man­ual nor for any damages which might originate through the use of it. Since mistakes can never be avoided completely, despite all efforts, we would appreciate any given hint. We will be striving to fix any mistake as fast as possible.
Verification of Suitability
Verification of SuitabilityVerification of Suitability
Verification of Suitability
Goudsmit Magnetic Systems hereby certifies that this product has been inspected and tested before shipment in accor­dance with the stated technical data. Furthermore, Goudsmit Magnetic Systems declares that its calibration measurements are traceable to the Physikalisch-Technische Bundesanstalt, Braunschweig, or to the United States National Institute of Standards and Technology (former National Bureau of Stan­dards) and to other members of the International Standardi­zation Organization (ISO), provided that the calibration pos­sibilities of these institutes allow this.
Warranty
WarrantyWarranty
Warranty
The warranty is valid for two years starting from the date of shipment and comprises material and fabrication errors. Within the warranty period Goudsmit Magnetic Systems re­places or repairs products which have proved to be defec­tive.
Return under Warranty
Return under WarrantyReturn under Warranty
Return under Warranty
In case of warranty the device must be returned to a service point determined by Goudsmit Magnetic Systems. On return­ing the device under warranty, the buyer bears the costs for the shipment to Goudsmit Magnetic Systems, while Goudsmit Magnetic Systems bears the costs for returning the device to the buyer. However, the buyer bears all the costs for shipment, taxes and duties, if the device is returned to Goudsmit Magnetic Systems from abroad.
Extent of Warranty
Extent of WarrantyExtent of Warranty
Extent of Warranty
The warranty already mentioned does not apply for errors due to inadequate or insufficient maintenance on the part of the buyer, due to unauthorized modifications or maloperat­ion, to software or interfaces provided by the buyer, or to operations beyond normal ambient conditions. The buyer is
solely responsible for design and assembly of electric circuits in connection with this product. Goudsmit Magnetic Systems cannot be made liable neither for customers’ circuits nor for malfunctions of the product resulting from these. Moreover, Goudsmit Magnetic Systems will not accept warranty for damage which is traceable to customers’ circuits or to prod­ucts provided by the customer. Any further explicit or tacit warranty is excluded; Goudsmit Magnetic Systems especially rejects any tacit warranty concerning the qualification for a special purpose or to marketability.
Exclusive Measures
Exclusive Measures Exclusive Measures
Exclusive Measures
The above-mentioned measures are the only and exclusive measures on the side of the buyer. Goudsmit Magnetic Sys­tems is not liable for direct, indirect, particular damages or damages resulting from negligence or consequential dam­ages, no matter if they are founded on warranty, contract, offence or any further legal theory. This does not apply as far as Goudsmit Magnetic Systems is necessarily made liable by law.
Annotation
AnnotationAnnotation
Annotation
The information in this document is subject to alterations without prior announcement. Goudsmit Magnetic Systems does not take over warranty for this material including the marketability or its qualification for any determined purpose beyond. Goudsmit Magnetic Systems do not take over liabil­ity for errors in this manual or for accidental or consequen­tial errors within the context of shipment, performance or usage of this material. No portion of this document may re­produced, translated or transmitted, in any form or by any means, electronic, mechanical, photocopying, recording, or otherwise, without the express written permission of Goudsmit Magnetic Systems.
Safety
SafetySafety
Safety
Do not replace any parts and do not make alterations to the product without our explicit and written consent. Send the product to Goudsmit Magnetic Systems for repair and main­tenance to safeguard that all safety features be preserved. Handling malpractices may lead to damage at the device and possibly to injuries and death of persons.
Goudsmit Magnetic Systems Petunialaan 19 5582HA Waalre - Netherlands
Gaussmeter HGM09s Operating Instructions
Chapter 12 Index
Page 74 / 75
12
1212
12 IIIIndex
ndexndex
ndex
****
*CLS........................................... 53
*ESE .......................................... 53
*ESR .......................................... 53
*IDN ........................................... 54
*OPC.......................................... 54
*RTS .......................................... 54
*SRE .......................................... 55
*STB........................................... 55
::::
:MEAS........................................ 57
:MODE ....................................... 57
:NULL......................................... 58
:PAR
ACDC.................................... 64
CHAR.................................... 66
CONT.................................... 67
LIGH ..................................... 66
PEAK .................................... 64
POFF ..................................... 65
POLD .................................... 65
RANG ................................... 65
SAVE .................................... 67
UNIT ..................................... 63
USB.......................................63
:PEAK ........................................ 60
MODE...................................60
NULL .................................... 60
READ.................................... 61
MAX.................................61
MIN ..................................61
:PROB
NAME...................................62
SN ......................................... 62
TYPE..................................... 62
:RANG........................................ 59
AUTO.................................... 58
SET ....................................... 58
:READ ........................................ 59
:SN
CALI ..................................... 68
HW........................................ 68
SW......................................... 68
UNIT ..................................... 67
:STAT
MEAS
ENAB ............................... 56
EVEN ............................... 57
PRES ..................................... 55
QUES
ENAB ............................... 56
EVEN ............................... 56
:UNIT.......................................... 59
AAAA
AC Field Measurements ............. 28
Accuracy.....................................15
BBBB
Batteries .....................................18
Blanks.........................................44
Boolean Parameters ...................46
CCCC
Character Set..............................43
Character String Parameters ......46
Charging .....................................38
Charging Batteries ......................19
Command Separator...................44
Command Structure....................44
Conformity Declaration ...............72
Connector Plug ...........................41
Contents .......................................3
Contrast Adjustment ...................38
Control Commands ............... 51, 53
Cross Current Resistance ...........12
DDDD
Data Format................................43
DC Field Measurements .............27
DC/AC Field................................36
DC/AC Field Measurements .......26
Device Functions ..................52, 67
Direct Operation..........................41
Discrete Parameters ...................46
Display....................................9, 22
Display Brightness ......................38
Display Units...............................35
EEEE
Endings.......................................46
FFFF
Fast Peak Value Recording ........31
Field strength ..............................69
Flux density.................................69
Front Side ...................................17
GGGG
Gaussmeter Function..................10
HHHH
Hall Effect ...................................10
IIII
Illustrations ...................................4
Installation ..................................43
Interface......................................43
Introduction...................................7
KKKK
Keyboard ....................................21
LLLL
Linear Properties ........................ 10
MMMM
Main Commands .................. 51, 57
Measurement Details ................. 13
Measuring Range ................... 8, 25
Measuring Unit ....................... 8, 26
NNNN
Non-linear Properties.................. 11
Null............................................. 24
Numerical Parameters................ 45
OOOO
Operating Mode.......................... 35
Operation ................................... 21
Output Data................................ 47
PPPP
Parameter Separator.................. 44
Parameters........................... 52, 63
Path Separator ........................... 44
Peak Value................................. 36
Peak Value Functions .......... 51, 60
Peak Value Measurement .......... 30
Peak Value Recording................ 30
Polarity ....................................... 37
Ports Overview........................... 17
Power Supply ............................. 18
Preparing a Measurement ............ 7
Probe Connection....................... 20
Probe Data ................................. 33
Probe Functions ................... 52, 62
QQQQ
Query Commands ...................... 45
RRRR
Range Selection......................... 37
Remanence................................ 14
Running a Measurement .............. 7
SSSS
Safety Instructions.................... 5, 6
Safety Symbols ............................ 6
Sample Measurement ................ 13
SCPI............................... 45, 51, 53
SCPI Language.......................... 43
Sensitivity................................... 12
Serial Interface.......................... 40
Settings..................................... 35
Setup Menu................................ 34
Static Magnetic Fields ................ 16
Status Display ........................ 9, 22
Summary.................................... 51
Gaussmeter HGM09s Operating Instructions
Chapter 12 Index
Page 75 / 75
Switching off ............................... 37
Switching on/off .......................... 24
System Commands .................... 45
TTTT
Technical Data ........................... 70
UUUU
Unit ............................................ 69
USB ............................................ 41
USB Interface ............................ 20
VVVV
Version Remarks........................ 39
WWWW
Warranty..................................... 73
Loading...