Immediately after receipt, unpack the KONSTANTER and all
included accessories, and inspect for damage and
completeness.
Unpacking
•Other than the usual care exercised in handling electronic
equipment, no additional precautions are required when
unpacking the instrument.
•The KONSTANTER is delivered in recyclable packaging,
which provides for adequate protection during transport as
substantiated by testing. If the instrument is repacked at a
later point in time, the same packaging, or its equivalent, must
be used.
Visual Inspection
•Compare the order number or type designation included on
the packaging and/or the serial plate with the particulars
shown in the shipping documents.
•Make sure that all accessory components have been included
( Kap. 1.3, “Options and Accessories”).
•Inspect the packaging, as well as mechanical instrument and
accessory components for possible transport damage.
Complaints
If damage is discovered, immediately file a claim with the freight
forwarder (save the packaging!). If other defects are detected, or
in the event that service is required, inform your local
representative, or contact us directly at the address included in
the last page of this handbook.
IIWarnings and Safety Precautions
The KONSTANTER has been manufactured and tested in
accordance with the electrical safety regulations listed in the
technical data as a safety class I device, and has been shipped
from the factory in flawless technical safety condition. In order to
maintain this condition and to assure safe operation, users must
observe all notes and warnings included in these operating
instructions. These are identified with the following headings:
Caution!
A note concerning operation, practical advice or other information which
must be adhered to in order to prevent damage to the KONSTANTER, and
to assure correct operation.
WARNING!
An operating procedure, practical advice or other information which must
be adhered in order to assure safe operation of the KONSTANTER, and to
prevent personal injury.
The most important warnings are summarized below. Reference is made
to these warnings at appropriate points within the operating instructions.
WARNING I – Protective Grounding
The KONSTANTER may only be placed into operation after the
protective conductor has been connected. Interruption of the
protective conductor, either inside or outside of the
KONSTANTER, or disconnection of the protective conductor
terminal may transform the KONSTANTER
potential danger. Intentional interruption is prohibited.
62N: The device is connected to the mains by means of a 3
conductor cable with mains plug. The plug may only be inserted
into a suitable outlet with earthing contact. This safety precaution
must not be defeated though the use of an extension cable
without protective conductor.
64N: These devices are permanently connected to the mains with
a 5-pole terminal block for 3-phase current (5 conductor).
WARNING II – Impaired Safety
If it can be assumed that safe operation is no longer possible, the
KONSTANTER
against inadvertent use. It must be assumed that safe operation is
no longer possible:
•If the KONSTANTER
•If the KONSTANTER
•After lengthy periods of storage under conditions which
deviate from specified storage conditions
•After extraordinary stressing due to transport
WARNING III – Opening Housing Panels
Voltage conducting parts may be exposed when housing panels
are opened, as long as the KONSTANTER
power.
Any contact with these exposed conductive parts is life
endangering.
For this reason, housing panels may only be opened and/or
removed by trained personnel who are familiar with the dangers
involved.
WARNING IV – Repair by Trained Personnel
Voltage conducting parts may be exposed when housing panels
are opened, as long as the KONSTANTER
power.
Maintenance and repair work, as well as internal balancing, may
only be performed by trained personnel who are familiar with the
dangers involved.
The KONSTANTER
sources before work of this type is performed, in as far as this is
possible. A 5 minute waiting period must be observed after
disconnection in order to allow internal capacitors to discharge to
safe voltage levels.
WARNING V – Fuse Replacement
Only specified fuse types with the specified nominal current rating
may be used to replace blown fuses (see Technical Data and
specifications on the serial plate).
Tampering with fuses or fuse holders is prohibited (“repairing”
fuses, short-circuiting fuse holders etc.).
must be removed from service and secured
demonstrates visible damageno longer functions
must be disconnected from all external power
into a source of
is connected to supply
is connected to supply
Significance of Symbols
Indicates EC conformity
Observe EGB directives
Warning concerning a source of danger
(attention: observe documentation!)
This device may not be disposed of with the trash. Further
information regarding the WEEE mark can be accessed
on the Internet at www.gossenmetrawatt.com by entering
the search term’WEEE’.
4GMC-I Messtechnik GmbH
1Technical Description
1.1 Features and Range of Applications
SSP KONSTANTERs (single-output system power supplies) are
manual and remote controllable DC power supplies for laboratory
and system use. Despite high output power, the power supplies
are small and lightweight.
The floating output features “safety electrical separation” from the
mains input as well as optional computer interfaces, and is
classified as a safety extra-low voltage circuit (SELV) in
accordance with VDE/IEC. Wide ranging nominal power values
are available from the voltage and current controlled output.
The devices are generally equipped with a control panel and
display, as well as an analog interface.
An additional interface can be plugged into the device for
integration into computer controlled systems. Two interface
variants are available to this end, as described in chapter 1.3.
Manual adjustment of voltage and current is accomplished by
means of two rotary knobs with selectable adjusting sensitivity.
Numerous additional functions can be activated with various keys
(see table below).
Two 4-place digital LED displays read out measured values and
settings. LEDs indicate the current operating mode, selected
display parameters and the status of device and interface
functions.
The analog interface can be used to set output voltage and
current with the help of external control voltages, and also serves
to link several devices for master-slave operation. Deactivation of
the output, control panel disabling and individual or sequential
retrieval of saved settings can be controlled via the floating
optocoupler input.
1.2 Functions
Adjustable functions:
–Voltage and current setpoint values
–Voltage and current limit values (soft-limits)
–Activate / deactivate the output
–Overvoltage protection trigger value
–Overcurrent response (limiting with or without shutdown)
–Delay time for overcurrent protection
–Starting performance (power-on status)
–Reset device settings
–Save device settings
–Recall device settings individually or sequentially
–Control panel disabling
–Function selection for trigger input
–Round off measured value for display
–Service request masks (SRQ masks)
–Activate/deactivate digital displays
–Self-triggering when mains power is switched on
Retrievable information:
–Momentary measured voltage and current values
–Minimum and maximum measured voltage and current values
–Momentary output power
–Current device settings (individual or complete)
–Current device status (control mode, overtemperature, busy)
–Occurred events (mains or phase failure, overtemperature,
overvoltage, overload, programming error)
–Device ID
1
Additional Functions
–Sensor terminals protected against polarity reversal and
automatic switching to auto-sensing
–Overtemperature protection
–Output protected against reverse polarity
1
1
1
1
–Backup battery for device settings memory
–Recognition of mains or phase failure
–Inrush current limiting
1)
Via computer interfaces only
1.3 Options and Accessories
Devices can be equipped with a plug-in interface module (2
variants) for integration of SSP KONSTANTERs into computer
controlled systems.
Variant 1 includes one digital computer interface:
–One V.24 / RS 232C bit-serial interface
This interface is provided in minimized format only
(conductors: RxD, TxD, GND).
Variant 2 includes two digital computer interfaces:
–One IEC 625 bit-parallel, byte-serial interface
This interface fulfills all requirements of the IEEE 488 standard
and is commonly designated as a GPIB (general purpose
interface bus) or an HP-IB (Hewlett Packard interface bus).
–One V.24 / RS 232C bit-serial interface
This interface is provided in minimized format only
(conductors: RxD, TxD, GND).
Programming of all device functions, as well as querying of
measured and configured parameters is possible via both
interfaces. The IEC bus interface provides several additional,
specific interface functions.
The interface option is supplied as a separate plug-in module and
can be easily plugged into the device at the rear.
Included accessories:
–1 set operating instructions
–1 mounting kit for rack installation
–1 mains power cable with earthing contact plug
(for 62 N only)
Additionally available accessories:
–RS 232 bus cable
–IEEE – IEEE bus cable
–Device driver for LabView software
–Device driver for LabWindows / CVI software
–Device driver for HPVEE / VXI PnP software
1
GMC-I Messtechnik GmbH5
1.4 Functional Principle
62 N: D
64 N: I+K+M
Figure 1.4 shows a schematic diagram of the SSPKONSTANTER. The delineations indicate distribution of the circuit
to the individual printed circuit boards, as well as subdivision of
the mains power circuit, the control and interface circuit, and the
output circuit.
Figure 1.4Schematic Diagram of the SSP KONSTANTER
Printed Circuit Board Functions Overview
PCB A:Central processing unit
PCB B:IEEE 488 – RS 232C interface (optional)
PCB C:Control panel and display
PCB D:
Required DC supply power is generated for each respective
circuit from mains power which has been fed to the power pack
via an interference suppression filter, a wire fuse, the mains switch
and inrush current limiting. Series 64 N SSP KONSTANTERs are
supplied with power from a 3-phase mains system (relative to the
neutral conductor).
Central Processing Unit (CPU)
Overall control of the SSP-KONSTANTER is accomplished by
means of the CPU on PCB A. It uses an 80C32 8-bit
microcontroller with 64 kilobytes of program memory and 32
kilobytes of battery-backed CMOS RAM.
An 11 MHz pulse generator establishes the clock frequency for
the processor, and creates a time reference for the measuring
function and the serial interface.
A watchdog circuit monitors processor activity and disables
access to battery-backed RAM in the event of supply power
failure.
Operation
The SSP-KONSTANTER can be operated with the controls at the
front panel, or by means of the optional IEEE 488 and RS 232C
plug-in interface module.
Displays and Control Panel
The two 4-place, 7-segment displays and the keys at the front
panel are managed by a controller module in multiplex mode. The
individual LEDs are driven statically via a register, and the rotary
pulse encoders control increment-decrement counters relative to
direction of rotation. Each time an adjusting element is activated,
an interrupt occurs at the CPU which then causes an appropriate
response.
Interface Option
If the SSP-KONSTANTER is equipped with a plug-in interface
module, the device can also be controlled either via the IEEE 488
bus or the RS 232C serial interface.
Remote Control
Device messages received by the interface are forwarded to the
CPU where they are first saved to RAM. After receiving an end-ofmessage character, data are checked for correct syntax,
plausibility and limit values. Valid commands are subsequently
executed.
Setup Procedure
Setup data are processed and forwarded to the respective
function unit via I/O control and an optocoupler assuring electrical
isolation. Each setting value for output voltage, output current or
overvoltage protection triggering is converted to a proportional
control voltage by a 12 bit DAC, and is fed to the respective
controller or comparator as a setpoint or a reference quantity.
Actual output voltage is ascertained by a voltage monitor, whose
automatic sensor switching inputs are connected either to the
output terminals or the sensor terminals.
Actual output current is acquired as a voltage drop at a shunt
located in the negative output conductor, and is amplified by the
current monitor to a scaled signal.
In order to achieve rapid downward adjustment of output voltage
even with minimal output load, the device is equipped with a
limited sink function (limited to approximately 25 W per 1000 W
output power) for discharging the output capacitor. This function
is activated as long as output voltage exceeds the current
setpoint value (also in the event of energy recovery from a parallel
connected voltage source).
The source and sink function is enabled when the output On / Off
controls are set to ON, and the source function is disabled when
controls are set to Off and the sink setpoint is set to Unominal
after approximately 300 ms (high impedance for Uout <
Unominal).
Measuring Procedure
Monitor amplifier output signals, which are proportional to actual
output voltage and current, are fed to an analog multiplexer (MUX)
which switches one of the two signals to the input of the analogdigital converter (ADC) depending upon the desired measured
quantity. The ADC functions in accor
with the synchronous
dance
voltage-frequency conversion principle, and makes a squarewave signal available at its output whose frequency is
proportional to the measured quantity fed to the input. An
optocoupler is used to assure electrically isolated signal
transmission to a binary counter whose 40 ms gating time is
derived from the quartz controlled pulse frequency of the CPU by
means of an additional counter. After gating time has elapsed, the
CPU acquires the counter value and calculates a measured value
which is saved to RAM as a decimal number. Depending upon
the circumstances, the measured value display is refreshed, an
extreme value comparison is performed for the Min-Max function
or the measured value is made available at the computer
interface’s data output buffer.
Monitoring Functions
– Control Mode Recognition and Overload
An electrically isolated digital signal is derived from the output
signals of the voltage and current regulators, which indicates
the currently active control mode (constant voltage or
constant current mode), as long as overload protection is not
active. “Overload” indicates that power limiting has been
triggered as a result of selected parameter settings and
prevailing load. These operating conditions are evaluated by
the CPU (e.g. for OCP function), are indicated with LEDs and
are used to generate status and event registers for computer
control.
– Overvoltage Monitoring
If the voltage monitor is bypassed, device output voltage is
additionally compared to an adjustable limit value within a
range of 3 V to 120% nominal voltage by a comparator, and
the output is deactivated if the limit value is exceeded and an
OVP message is generated (LED display, status and events
register).
– Temperature Monitoring
Temperature is converted to a proportional electrical signal by
PTC resistors at representative points (conductor bar, diode
stack, choke and additional switching transistors (for series 62
N, 500 W)) and are fed to a two-step threshold trigger. The
lower threshold value corresponds to a temperature of
approximately 75° C, and the upper threshold value is
approximately 85° C. As soon as the lower threshold is
exceeded at any of the temperature sensors, the fan(s) is/are
switched to high speed by the fan controller. The device can
be operated at any output load up to the maximum specified
operating temperature after this ventilation has been
activated. If ventilation is impeded, or at excessive ambient
temperatures, temperature at the sensor may reach the upper
threshold. If this is the case, an overtemperature message is
entered to the status and events register. After 5 seconds, the
OTP LED is activated and the output is shut down by the
CPU. After sufficient cooling has occurred, the status
message is cleared and a ready for operation message is
entered to the event register. If the POWER-ON function is set
to “Recall”, the output is reactivated automatically. If the
POWER-ON function is set to “Standby” or “Reset”, the
output remains deactivated and can only be reactivated by
pressing the OUTPUT key, or by means of a command from
the control computer. After actual temperature has fallen to
below the lower threshold value, fan speed is reduced
automatically.
GMC-I Messtechnik GmbH7
1.5 Technical Data
U / V
U
nominal
0,5
P
nominal
P
typ
P
dyn
(< 5 ms)
0.5 I
nominal
I
nom
0
I / A
Voltage
Setting Range
Current
Setting Range
P
typ
1.3 x P
nominal
U
nominal
1.5.1 General Data
Output
Regulator typePrimary switched-mode regulator
Operating modesAdjustable constant voltage / constant
current source with automatic sharp
transition
Output insulationFloating output with “safe electrical
separation” from the mains input and
computer interfaces
Max. allowable potential, output to ground:
120 V
Capacitance, output to ground (housing)
Series 62 N: 500 W / 1000 W: typ. 90 nF
Series 64 N: 2000 W / 3000 W: typ. 180 nF
Output operating range
Power Supply
Line voltage62 N:230 V ~ +10 / –15%,
47 … 63 Hz
64 N:3 x 400 / 230 V ~ +10 / -15%,
47 … 63 Hz
Inrush CurrentMax. 50 A
S
Fuse62 N:1 ea. M 15 A / 250 V
(6.3 x 32 mm), UL
64 N:3 ea. M 15 A / 250 V
(6.3 x 32 mm), UL
Electrical Safety
Safety classI
Overvoltage
categoryII for mains input
I for output and interfaces
Fouling factor2
Earth leakage
current62 N: <3 mA
64 N: <1 mA
eff
eff
Electrical IsolationRated Voltage Test Voltage
Output – mains280 V
eff
4 kV ~ (type test)
Output – bus /
ground120 V
S
1.5 kV ~
Mains – bus /
ground230 V
eff
2.2 kV –
Bus – groundNo electrical isolation
Analog Interface
Functions– Auto-sensing mode
– Programmable trigger input
– Voltage control input (0 ... 5 V)
– Current control input (0 ... 5 V)
– Voltage monitor output (0 ... 10 V)
– Current monitor output (0 ... 10 V)
– Master-slave parallel operation
– Master-slave series operation
EN 61000-4-3: 1996 + A1: 1998, feature A
EN 61000-4-4: 1995, feature C
EN 61000-4-5: 1995, feature B
EN 61000-4-6: 1996, feature B
EN 61000-4-11: 1994, feature A
Ambient Conditions
Climatic categoryKYG per DIN 40 040
Temperature range Operation:0 to 40° C
Storage:–20 to +70° C
Relative humidityOperation: 75%,
no condensation allowed
Storage: 65%
CoolingWith integrated fan
(2-step temperature control)
Inlet vent:side panel
Outlet vent: rear panel
Operating noiseNoise pressure level at a distance of 30 cm
with fan set to low / high
62 N:64 N:
front18 / 28 dBA20 / 30 dBA
rear23 / 35 dBA26 / 38 dBA
left and right 20 / 30 dBA26 / 36 dBA
8GMC-I Messtechnik GmbH
1.5.2 Mechanical Data
TxD_IN 2
3
5
RxD_IN
GND_IN
2 RxD
3 TxD
5 GND
SSP
9-pin subminiature9-pin subminiature
PC / Controller
socket connectorplug connector
IN
RS 232
COM1/COM2
Twist with
opposing wires
Remote enable
Data bus
Shield connected
to ground
Attention
Service request
Interface clear
No data accepted
Data valid
End or identify
Data bus
Not ready for data
Module Type
Benchtop device, suitable for rack mounting
Dimensions
(W x H x D)See also dimensional drawings.
62 N: 19" x 2 std. height units x 500 mm
64 N: 19" x 4 std. height units x 500 mm
Weight62 N:500 W: approx. 12 kg
1000 W : approx. 13 kg
64 N:2000 W: approx. 22 kg
3000 W : approx. 28 kg
RS 232C interface (optional) approx. 0.1 kg
IEEE 488 – RS 232C interface (optional)
approx. 0.14 kg
Terminals (rear panel)
Main input62 N:10 A IEC inlet plug with
earthing contact (L + N + PE)
64 N:5-pole screw terminal block
for cable connection,
0.75 ... 2.5 mm2 (3 L + N + PE)
OutputBus bars with drill holes for M8 screws and 4
mm diameter drill holes
Analog interface14-pin plug connector with screw clamps
The device must be switched off when installing the interface module.
The interface module may be damaged by electrostatic discharge.
Observe guidelines for handling electrostatic sensitive devices. Do not
touch electrical contacts or PCB components.
1. Unscrew the cover plate at the left-hand side of the rear
housing panel.
2. Carefully insert the interface module into the open slot and
press it onto the plug connector.
3. Fasten the interface module with the screws taken from the
cover plate.
2.1.2 Installation to 19'' Device Racks
The SSP-KONSTANTER housing allows for use as a benchtop
instrument, as well as for installation to a 19'' rack.
The benchtop instrument can be quickly converted to a rack
mount device:
1. Unscrew the handles at the front.
2. Pull out the filler strips at the sides and replace them with the
included rack-mount fastening tabs.
3. Replace the front handles (if you prefer to leave the handles
out, turn M4 screws with a maximum length of 8 mm in to the
open threaded holes.)
4. Unscrew the feet from the bottom of the housing.
5. Save all loose parts for possible future use.
Attention!
The device must be attached to guide rails at both sides of the rack. The
guide rails, as well as the front panel mounting screws, are rack-specific
and must be procured from your rack supplier.
2.1.3 Connection to the Mains
Observe WARNING I!
Caution!
Before switching the SSP KONSTANTER on, it must be assured that
available mains power complies with the supply power values specified
at the mains connection on the back of the device.
– Series 62 N (500 W, 1000 W) :
These devices require 230 V supply power and are connected
to a mains outlet with earthing contact with the included power
cable via the mains connector plug [35] at the rear panel.
– Series 64 N (2000 W, 3000 W):
WARNING!
These devices may only be connected to mains supply power
by a qualified electrician.
These devices require 3-phase 120/400 V supply power with
neutral and phase conductors (3 L + N + PE).
A 5-conductor power cable with a minimum wire cross-section
of 1.5 square mm is required for connection to mains supply
power, and is connected to the terminal block [35] at the rear
panel:
L1:Phase
L2: Phase
L3: Phase
N: Neutral conductor
PE: Protective conductor
The cable must be secured with the cable clamp [38] for
reliable strain relief.
2.1.4 Connecting Power Consumers
The output leads are connected to the output terminal bars [33] at
the rear panel by means of ring-type cable lugs. The terminal bars
are equipped with drill holes for M8 screws to this end. 4 mm drill
holes are included as well, which can be used for connecting
measurement cables, ground cables or cable shields.
Connection:
Remove the safety cap.
Connect the output leads to the terminal bars with suitable
screws and washers.
Make sure that the utilized cables have an adequate cross-
section, and that polarity is not reversed. It is advisable to twist
the output leads and to identify polarity at both ends.
Avoid exerting of force at the terminal bars.
Arrange the leads such that they can be fed through the
opening in the safety cap.
Snap the safety cap back into place.
In order to be able to take advantage of highly constant output
voltage at the consumer even if long leads are used, sensing
leads can be used to compensate for voltage drops within the
output leads ( chapter 5.2).
2.1.5 Connection to Computer Interfaces
If the device is used within computer controlled systems, one of
the two connections described below must be established via the
optional interface.
Comment
The device cannot be remote controlled via both interfaces
simultaneously. The interface which first initiates action after
mains power has been switched on is activated, and the other
remains inactive.
In order to assure that existing bus activity is not interfered with,
all affected devices should be switched off while establishing the
bus connection.
Both interfaces are equipped with a common ground
are electrically isolated from the output in accordance with
specified electrical safety regulations.
Connection
Configure the interface as described in chapter 4.7.4 before
connecting.
a) IEC Bus
Up to 15 IEC bus controlled devices (including controllers) can
be interconnected to create a system.
These devices are connected to the bus with suitable,
commercially available cables with 24-pin plug connectors.
If your IEC bus system is equipped with the previously
common 25-pin subminiature plug connectors, you will need a
suitable adapter cable.
Both cable types are available as accessories (see last page).
In order to assure reliable data transmission, cable length
between devices should not exceed 2 m, and overall length
should not exceed 15 m.
Double shielded connector cable is recommended if bus
devices are operated in proximity to strong sources of
interference or their power cables.
(GND), and
14GMC-I Messtechnik GmbH
b) RS 232C Interface
Only two devices can be connected to each other with this
serial interface, namely a controller and the device to be
controlled.
If you intend to control several devices with a single controller,
the controller must be equipped with suitable interfaces. Most
controllers include two serial ports which are commonly
designated COM1 and COM2, and which are equipped with
25 or 9-pin subminiature plug connectors.
Suitable cable is available in various lengths from commercial
outlets for connecting the SSP-KONSTANTER to the
controller. Appropriate adapters are available as well, in the
event that your controller is equipped with a 9-pin plug
connector.
If you intend to fabricate the connector cable yourself, you will
need a 3-conductor shielded cable in order to establish the
connection as shown in Figure 1.5.2.
2.2 Switching the Instrument On
After the described preparations have been completed, the
device can be switched on.
Press the mains switch [4] at the front panel until it snaps into
place in order to turn the device on.
Power-Up Test
After switching the device on, the POWER lamp [5] lights up and
the fan is started. The microprocessor included in the device then
starts a power-up test. The following operations are performed
during the test routine (duration approximately 8 seconds):
– Reset all functional units (except battery-backed configurations
memory)
– ROM test
–RAM test
– Initialize computer interfaces if installed
– Ascertain device type
– Check the ADC timer
– Recall last settings if required
The READY lamp [22] blinks while this routine is running, and all
other LEDs and all digital display segments light up (display test).
If the device has been equipped with the “IEEE 488 – RS 232
computer interface” option, the selected IEC bus device address
then appears briefly at the display (example: “Addr 12”).
After successful completion of the self-test, the READY lamp is
continuously illuminated and the display is switched to measured
value indication for voltage (Uout) and current (Iout).
After initial power-up, the device has the following basic
configuration:
Interface functionsStandard “pon” status
Device functions
– Output statusInactive
– Voltage setpoint0 V
– Current setpoint0 A
– Voltage setting limitNominal output voltage
– Current setting limit Nominal output current
– OVP trigger value62.5 V (for 52 V models)
100 V (for 80 V models)
– Current limiting modeLimiting without shutdown
– Shutdown delay0 ms
– TRIGGER input Inactive
– Min-Max measured value memoryOff
– Power ON mode Reset configuration
– Manual operationEnabled
– Memory contents Deleted
The desired settings can be selected starting with this basic
configuration.
After a warm-up period of approximately 30 minutes, the
instrument operates at maximum accuracy.
When the device is powered up again at a later point in time, active
device configuration depends upon the last setting selected for
the POWER_ON function (—> page 70):
– Default settings or
– Last used device settings or
– Last used device settings and inactive output
Power-Up with RESET
In order to assure that the connected power consumer is not
endangered by any previous device settings, the device can be
initialized with the “POWER_ON RST” function by pressing and
holding the <CE/LOCAL> key during the power-up routine.
In order to switch the device off, activate the mains switch once
again. The device is then disconnected from mains power and the
output is deactivated. The last device configuration, as well as
any settings which have been saved to battery-backed
configurations memory, are retained.
Caution!
apid, repeated fashion. This
Avoid switching the device on and off i
temporarily impairs the effectiveness of the inrush current limiting
function, and may result in a blown fuse.
n a r
If this status is not achieved despite a correctly selected device
address (0 to 30), even after repeatedly switching the device on
and off with abbreviated self-test, the device is probably
defective. If this is the case, contact your local representative.
Abbreviated Power-Up Test
In order to shorten power-up time, or if problems occur with the
normal power-up test, an abbreviated power-up test can be
used:
With the device switched off, press and hold the <ENTER>
key.
Turn the mains switch on.
Release the <ENTER> key after approximately 1 second.
If this procedure is used, only essential initialization steps are run
during power-up.
GMC-I Messtechnik GmbH15
3Controls, Display Elements and Terminals
Slot for installing optional
interfaces
Shown with optional
IEEE 488 – RS 232C interface
12 3 45
6
7
910
1112
1314
15 16
17
18 1920212223
24 25
26313736
35
3433
32
2631 323536 37
3834332730
8
Series 64 N
19" Rack
Series 62 N
Rear Panel
Series 62 N
19" Rack
Series 64 N
Rear Panel
1
16GMC-I Messtechnik GmbH
Note: Numbers in brackets make reference to figures on page 19.
[1]Rack mounting tabs (included accessories)
for mounting to a 19" rack
Assembly:
– Unscrew the handles at the front.
– Pull out the filler strips at the sides and replace them with
the included rack-mount fastening tabs.
– Replace the front handles.
[2]Front handles
For carrying the device or pulling it from the 19" rack
The handles can be removed if desired (turn M4 screws with
a maximum length of 8 mm into the empty threaded holes).
[3]Device feet
For use as a benchtop instrument
The device can be tilted back slightly by folding the front feet
out.
As a rule, the feet must be removed in order to install the
device to a 19" racks.
[4]Mains switch
For turning the device on and off
After switching the device on, a self-test is performed with a
duration of approximately 8 seconds. After successful
completion of the self-test, the “POWER-ON” function
configures the device with predetermined settings and the
device is ready for use.
When the device is switched off, it is disconnected from
mains power and the output is immediately deactivated. The
last device configuration, as well as any settings which have
been saved to battery-backed configurations memory, are
retained.
[5]POWER indicator lamp
Indicates that the device is switched on (power on).
[6]Control mode display
The illuminated LED indicates the current operating status
(control mode) of the output, as long as it is active:
– Green “CV” LED lights up: constant voltage mode (Uout =
Uset),
– Green “CC” LED lights up: constant current mode (Iout =
Iset)
– Yellow “Pmax” LED lights up: electronic power limiting is
active (Pout > Pnominal)
– No LEDs illuminated: output is inactive
[7]Output On-Off key <OUTPUT>
The power output can be activated and deactivated by
pressing the <OUTPUT> key. The respective LED is
illuminated as long as the output is active.
No significant output voltage overshooting occurs when the
output is activated and deactivated.
The output capacitor is rapidly discharged by a sink when
the output is deactivated. After approximately 350 ms (500
ms for 80 V models), the output becomes highly resistive,
but it is not isolated from the output terminals.
[8]Rotary knob for adjusting voltage
Voltage is adjusted in the usual fashion with the rotary knob.
However, the adjusting element is not a potentiometer, but
rather a rotary pulse encoder which generates 24 pulses per
revolution, and whose step size per pulse (setting resolution)
can be adjusted to either coarse, medium or fine with the
<RESOL> key [19]. This allows for convenient, precise
adjustment on the one hand, and also assures that no
change occurs to the selected value when switching
between remote control and manual operation.
When the voltage adjusting knob is turned, the left display is
first switched to the Uset display and the current voltage
setpoint appears. After approximately 0.4 seconds, one of
the decimal places starts blinking at the display in order to
indicate the selected adjusting sensitivity. From this point on,
turning the rotary knob changes the display value, and thus
the setpoint value, at the selected decimal place. Clockwise
rotation increases the value, and counterclockwise rotation
decreases the value. If no adjustment is made for a period of
10 seconds, the display automatically returns to measured
voltage value Uout. The display can be immediately
switched to the Uout value by pressing the <ENTER> or the
<CE/LOCAL> key.
[9]Left-hand display with
[10]
Display par
ameter indicators and
[11] Display selector key <SELECT>
As a default setting, measured output voltage value Uout
appears in volts at the left-hand display. The display can be
switched to any of the following values by (repeatedly)
activating the respective <SELECT> key:
– Uset= Output voltage setpoint in volts
– Ulim= Upper setting range limit for Uset
– OVset = Trigger value for output overvoltage protection
in volts
– Pout= Momentary measured output power in watts
(calculated as Uout x Iout)
The LEDs assigned to the display indicate the type of
displayed values (green LEDs = measured values, yellow
LEDs = setting values).
If no adjustment is made for a period of 10 seconds during
the display of a setting value, the display returns to Uout.
Measured power Pout can be indicated at the left, as well as
at the right-hand display. This makes it possible to evaluate
output power relative to output voltage or output current.
Additional device functions can be selected with the
<FUNCTION>, <SAVE> and <RCL> keys. The function
code appears at the left-hand display in this case.
“Err” appears as an identifier for system messages.
[12] Rotary knob for adjusting current <Iset>
The same applies to this rotary knob with reference to
output current, as is also the case with the voltage adjusting
knob [8].
[13] Right-hand display with
[14] Display parameter indicators and
[15] Display selector key <SELECT>
As a default setting, measured output current value Iout
appears in amperes at the right-hand display. The display
can be switched to any of the following values by
(repeatedly) activating the respective <SELECT> key:
– Iset= Output current setpoint in amperes
– Ilim= Upper setting range limit for Iset
– DELAY = Shutdown delay time for the OCP function in
seconds
– Pout= Currently measured output power in watts
(calculated as Uout x Iout)
The LEDs assigned to the display indicate the type of
displayed values (green LEDs = measured values, yellow
LEDs = setting values).
If no adjustment is made for a period of 10 seconds during
the display of a setting value, the display returns to Iout.
Additional device functions can be selected with the
<FUNCTION>, <SAVE> and <RCL> keys. The respective
setting parameter or measured value appears at the righthand display in this case.
“Err” appears as an identifier for system messages (see
chapter 8.6).
[16] Parameter adjusting keys <> and <>
In order to prevent operator errors to the greatest possible
extent, the two rotary knobs, [8] and [12], are used
exclusively for adjusting output voltage and output current.
These two keys are used for the selection and adjustment of
parameters for all other adjustable device functions
according to the following method:
a) Adjusting numeric parameters
After selecting the function to be adjusted with the
<SELECT>, <SAVE> or <RCL> key, the current numeric
GMC-I Messtechnik GmbH17
value for the selected parameter appears at the display.
This value can be increased <> or decreased <> within
predefined limits with the parameter adjusting keys.
Pressing the key briefly results in a single step, and if the
key is pressed and held the value is advanced through a
series of consecutive steps. The setpoint value is
changed simultaneously along with the display value.
b) Selecting text parameters (setting alternatives)
After selecting the function to be adjusted with the
<FUNCTION> key, the current parameter status appears
at the display in text format. Any of the alternative settings
can be selected by repeatedly pressing the <> or the
<> key. The parameter blinks at first to indicate that the
displayed alternative has not yet become effective. The
selected parameter value does not become effective until
it is acknowledged with the <ENTER> key. If
acknowledgement does not ensue, the device function is
exited and the respective setting remains unchanged.
c) Browsing through the SEQUENCE register
chapter 4.11 RCL and chapter 4.8 SELECT
[17] Protective functions status display
These indicators provide information concerning the setting
status (yellow LED) or triggering (red LED) of protective
functions. Illumination of the respective LEDs has the
following significance:
OVP (overvoltage protection)
Overvoltage protection has been triggered, because output
voltage has exceeded the selected trigger value (OVSET).
The output is deactivated.
Causes:
– Voltage setpoint USET has been set too high manually, or
as a result of memory recall, programming error or Uset
control signal to the analog interface
– Voltage transients caused, for example, by switching
inductive power consumers (perhaps too little difference
between selected USET and OVSET values)
– During auto-sensing: Sensing lead polarity is reversed, or
an output lead is/was interrupted or was not taken into
consideration when adjusting OVSET, so that the voltage
at the output terminals which is relevant for the OVP
function is increased by the amount to be compensated
for at both leads, and is higher than USET voltage as
controlled by the sensors at the load side (too little
difference between selected USET and OVSET values).
– Unipolar power recovery from the connected power
consumer (e.g. DC motor)
– A device error or defect has occurred.
After the cause of triggering has been eliminated, the
output can be reactivated with the OUTPUT ON
command.
As long as the latter has no occurred, the output can be
reactivated after an adequate cool-down period. If the
POWER-ON function has been set to “RCL”, the output is
reactivated automatically.
urrent protection)
OCP (over
c
Overcurrent protection has been triggered because the
output has been operated in the current limiting mode
(current control) for a duration greater than the DELAY value
with activated OCP ON function.
Causes: page 26 and page 68
The output can be reactivated with the OUTPUT ON
command.
OCP ON
Overcurrent shutdown is enabled.
page 26 and page 68
LOCAL LOCKED
The front panel controls are disabled, and are thus protected
against unauthorized or inadvertent adjustment.
This display only applies to disabling of the front panel
controls by means of manual adjustment or a control signal
applied to the TRIGGER input (for T_MODE TRG). It does
not indicate disabling of manual switching to local control by
means of the IEC bus LOCAL LOCKOUT command.
[18] Function selector key <FUNCTION>
No special key is assigned to device functions which are
normally seldom adjusted or used. These functions are set in
a menu-driven fashion by means of the following procedure:
1. Select the desired device function
with the <FUNCTION> key:
– Scroll forward: <FUNCTION> + <> or
<FUNCTION> alone
– Scroll back:<FUNCTION> + <>
Left display:Device function code
Right display:Currently selected parameter setting
or saved measured value
2. Select the desired function parameter
by (repeatedly) activating the <> or the <> key ([16]).
Left display:Unchanged
Right display:For adjusting the selected function
parameter (blinking indicates that
adjustment has not yet been executed)
OTP (overtemperature protection)
Overtemperature protection has been triggered because the
device has overheated. The output is deactivated.
Causes:
– Impaired cooling, e.g. air inlet or exhaust vents are
obstructed.
– Excessive ambient temperature The device is capable of
continuously supplying nominal power at ambient
temperatures of up to 50° C (measured at the air inlet
vents). Approximately 120 to 130% nominal power can be
drawn intermittently (triggering point for electronic power
limiting). Continuous operation at these levels may cause
triggering of the overtemperature protection function.
– One or more fans have failed.
– Output “fluctuates”. In the case of complex loads, control
fluctuations may occur ( chapter 5.2) which result in
increased power loss and thus causes overheating.
– A device error or defect has occurred.
18GMC-I Messtechnik GmbH
3. Acknowledge and execute the selected setting
by pressing the <ENTER> key.
Both displays return to default values Uout and Iout.
If the rotary knobs or other function keys are activated, the
function menu is exited and settings remain unchanged.
Exceptions:
<OUTPUT> is autonomously active.
<RESOL> remains active for numeric parameters, and is
otherwise inactive.
Function parameters which do not blink in their entirety are
accepted without acknowledging with <ENTER>.
The last open menu appears when the functions menu is
reopened.
[19] Resolution selection key <RESOL>
This key can be used to vary step size (setting resolution) for
device functions with numeric parameter settings, whose
displayed values can be increased or decreased with the
rotary knobs, or the <> and <> keys.
The decimal place to be increased or decreased blinks.
Blinking can be positioned at any one of the three lowest
(right-most) decimal places by repeatedly pressing the
<RESOL> key, thus allowing for the selection of fine,
medium or coarse setting resolution.
A different resolution can be selected for the left and righthand displays, and settings remain valid until a new
resolution is selected, or until the device configuration is
reset. Manual resolution adjustment after reset (*RST): fine
(right-most decimal place)
[20] Save key <SAVE> and
[21] Recall key <RCL>
The memory function is controlled with these two keys (
chapter 4.11 and page 62).
[22] <ENTER> key
The parameter value selected for a given device function
setting is acknowledged and executed by pressing this key.
[23] <CE/LOCAL> key
This key has several functions:
1. Abort an operation
If the <CE/LOCAL> key is activated while a device function
is displayed for adjustment, the display returns to its default
value and no change is made to the selected device setting.
2. Switch from remote to local control
If the device is being remote controlled via one of the
computer interfaces (REMOTE LED illuminated), all of the
control panel elements are disabled except for the mains
switch and the <CE/LOCAL> key. The device can be
returned to manual operation and the control panel elements
can once again be enabled by pressing the <CE/LOCAL>
key ( REMOTE LED off), without causing any changes to
current device settings.
If the device is being controlled via the IEC bus, the
<CE/LOCAL> key can be disabled with the LOCAL
LOCKOUT command, making key-operated return to
manual mode operation impossible.
3. RST – reset device settings (RESET)
The reset command is triggered by pressing and holding the
<CE/LOCAL> key and simultaneously activating the
<ENTER> key. This command returns most device functions
to their predefined default settings. Default settings are
described in chapter 4.16 and on page 62.
4. Disabling front panel controls
By pressing and holding the <CE/LOCAL> key and
simultaneously activating the <RCL> key, all control panel
elements are disabled except for the mains switch and the
<CE/LOCAL> key, and the LOCAL LOCKED LED lights up.
Disabling front panel controls prevents unauthorized or
inadvertent adjustment of device settings.
In order to reactivate the controls, the <CE/LOCAL> key
must be pressed and held for at least 4 seconds ( LOCAL
LOCKED LED off).
The <CE/LOCAL> key can also be disabled by applying a
signal to the TRIGGER input at the
analog interface
selecting the appropriate setting for the T_MODE function,
thus preventing manual reactivation of the front panel
controls.
[24] READY indicator
Indication of ready for operation:
– LED on: The device is ready for operation and the controls
are enabled.
– Blinking LED: The device is performing a self-test and
cannot be operated at the moment, or the SEQUENCE
mode is active ( page 36 and page 71).
– LED off: The device is not ready for operation.
[25] Interface status displays: REMOTE, ADDR and SRQ
Indication of computer interface operating status:
– REMOTE LED on: Device is being remote controlled, front
panel controls are disabled.
– ADDR LED on: The device has been addressed and is
receiving or transmitting data (applies to IEC bus operation
only).
– SRQ LED on: The device is transmitting a service request
(applies to IEC bus operation only).
[26] Device serial plate
For identification of the device
Contains particulars regarding the manufacturer, device
type, type designation, order number, serial number,
hardware revision level and power consumption.
[27] IEC 625 bus interface
For remote control of device functions via the IEC 625 bus
(= IEEE 488 bus) ( chapter 2.1.5).
Caution!
The electrical contacts of this interface are connected to
components which may be damaged by electrostatic discharge.
Ground yourself by grasping the housing before touching these
contacts!
[28] —
[29] —
[30] RS 232C interface
For controlling device functions via the RS 232C serial port
( chapter 2.1.5).
Caution!
The electrical contacts of this interface are connected to
components which may be damaged by electrostatic discharge.
Ground yourself by grasping the housing before touching these
contacts!
[31] Ground terminal
The output or cable shields can be grounded here if shielded
output cables or control cables for the analog interface are
used.
The ground terminal is connected to the housing and the
earthing contact at the mains connection.
and
GMC-I Messtechnik GmbH19
[32] Analog interface
The analog interface facilitates the following functions:
– Remote adjustment of output voltage and current with
analog control voltages ranging from 0 to 5 V ( chapter
5.3 / chapter 5.4)
– External measurement or recording of output voltage and
current based on monitor signals
0 to 10 V ( chapter 5.5 / chapter 5.6)
– Connection of sensing leads for the compensation of
voltage drops within the output leads ( chapter 5.2)
– Linking of several devices for master-slave operation
( chapter 5.8 / chapter 5.9)
– Varying internal output resistance ( chapter 5.10)
– Control of a selected device function via the floating
TRIGGER input ( chapter 5.7)
Caution!
The electrical contacts of this interface are connected to
components which may be damaged by electrostatic discharge.
Ground yourself by grasping the housing before touching these
contacts!
[33] Power output +/–
Terminals for connecting the power consumer
This is a floating output and can be grounded with the
positive or the negative pole. A detailed functional
description of the output and possible connection options is
included in chapter 5.
The outputs of all device models are classified as “safety
low-voltage circuits” (SELV) by the specified electrical safety
regulations (voltage in the event of error: < 120 V–).
However, contact protection measures are required for the
outputs of the 80 V models.
The output should always be deactivated before connecting
consumers in order to prevent sparking and induced voltage
transients.
[34] Air vents
Exhaust vents for the integrated fan
In order to assure adequate device cooling, exhaust vents
may not be obstructed.
The fan(s) is/are equipped with a 2-step temperature
controller.
[35] Mains connection
Connection for mains supply power.
[36] Line fuse(s)
Fusing for mains supply power
[37] Fuse ratings
WARNING!
Only fuses of the type and nominal current rating specified here
may be used when replacing blown fuses.
Tampering with fuses or fuse holders is prohibited (“repairing”
fuses, short-circuiting fuse holders etc.).
[38] Power cable strain relief
(for series 64 N only)
20GMC-I Messtechnik GmbH
4Manual Operation and Device Functions
Uout / V
Uset / V
Ulim / V
OVset / V
Pout / W
SELECT
UsetIset
Iout / A
Iset / A
Ilim / A
DELAY / s
Pout / W
SELECT
/
/
/
OUTPUT
Uset / V
Ulim
t
(1)
(2)
(3)
Important menu functions can be selected directly with the
<SELECT> keys.
4.1 Menu Structure
After the power-up sequence has been completed, the device is
switched to the basic operating mode – indicated by the
illuminated READY LED – by means of which the device’s basic
functions can be executed, such as:
• Select desired output voltage Uset
• Select desired output current Iset
• Adjust allowable working range with soft-limits Ulim and Ilim
• Adjust overvoltage and overcurrent protection
Additional setup menus can be accessed with the function key.
These include:
• SEt (setup) Extended setup functions
•AnIF (analog interface) Analog interface settings
• SEq (sequence function) Sequence function settings
After selecting the desired setup menu with the function key, the
respective menu level appears at the display.
The function and arrow keys are used to scroll through the
respective menu levels and select the desired settings.
The CE/Local key can be used to shift back up one level at a
time, until the basic operating menu once again appears.
4.2Setting Output Voltage Uset and Output Current Iset
There are two ways to set output voltage and output current:
• Direct selection
• Pre-selected settings
Changes become immediately effective when this method is utilized,
assuming the output is active.
Output voltage or current is preset with the arrow keys to the
desired value. The selected value is activated at the output
after acknowledging with the ENTER key.
Uout
Uset/V
Ulim/
OVset
Pout
4.2.1 Direct Selection (rotary knobs and arrow keys)
The operating concept allows for direct selection of output
voltage and/or output current with the rotary knobs, with
immediate activation of the new values at the output.
In the default configuration, momentary output values appear at
the display with voltage at the left and current at the right-hand
side. This is indicated by means of the two LEDs to the right of
the display. If the output is active (indicated by illumination of the
red LED above the OUTPUT key), the LEDs in the diagram
indicate the control mode. Depending upon the selected output
quantities and the load situation, either output voltage or output
current is regulated. CV (constant voltage) stands for voltage
regulation, and CC (constant current) stands for current
regulation.
If the working point is not within the allowable control range, the
power LED lights up.
The display can be switched to a representation of the
Iout/A
Iset/A
Ilim/A
Delay/s
Pout/
corresponding setpoint by slightly turning one of the rotary knobs
(Uset or Iset). This change is indicated by the respective LED to
the right of the display. The blinking decimal place indicates the
resolution with which adjustment will take place. Resolution can
be changed with the RESOL key.
The setting can then be changed to the desired value with the
rotary knob.
As soon as the selected setpoints have been activated, they can
also be adjusted with the arrow keys, in which case resolution
can also be pre-selected.
The setting mode can also be accessed with the SELECT key.
The LEDs next to the display indicate the selected function.
Initializing the Procedure
☞ Slightly turn (1) the Uset knob (see Figure 4.2.1 a).
!The display is switched from Uout (measured voltage value) to
Uset (voltage setpoint). The decimal place for the selected setting resolution blinks.
!
The
green Uout/V LED
goes out, and the
yellow Uset/V LED
lights up.
Selecting a Resolution
!3 step widths are possible: 0.01 V, 0.1 V or 1 V.
!
The
blinking decimal place
indicates which
step width
will be used for
setpoint adjustment.
☞ Repeatedly press the <RESOL> key [19] until the desired
decimal place blinks at the display.
Executing the Procedure
!Values become immediately active during adjustment.
☞ Adjustment (2) with the Uset rotary knob (see Figure 4.2.1 a):
Clockwise rotation Increases the value
Counterclockwise rotation Decreases the value
All intermediate values are run through semi-linearly, and are
read out to the output (assuming it is active).
Figure 4.2.1 aContinuous Adjustment of Uset
☞ Adjust (3) with the <> or the <> key:
<> (increment) Increases the value
<> (decrement) Decreases the value
!Each time the key is pressed, output voltage is changed by an
amount which corresponds to the value selected with the
resolution setting function.
!Pressing and holding the respective key results in rapid
scrolling, regardless of the step width.
Attention! Uset may not be set to a value which exceeds Ulim!
GMC-I Messtechnik GmbH21
Uset – Selecting a Setpoint for Output Voltage
Iset / A
Ilim
t
(1)
(2)
(3)
Uset / V
ENTERENTER
ENTER
Uout Iout
Select Uset
Uout unchanged!
CE/LOCAL
Uout Iout
Iout unchanged!
Activate
Select Iset
Uset
Function
• Adjustment with the Uset rotary knob [8] is not enabled until 0.4
seconds after the digital display [9] has been switched to the
respective function. This delay time prevents inadvertent changes to Uset during selection of the Uset display. The rotary
knob must be adjusted after this time period has elapsed in
order to change the Uset setting.
• After delay time has elapsed, output voltage is adjusted directly
as the knob is rotated if the output is active. The currently
active setpoint appears continuously at the display [9].
• If no adjustment is made for a period of 10 seconds, the
display automatically returns to measured voltage value Uout.
The display can be immediately switched to the Uout value by
pressing the <CE/LOCAL> key.
• There are two ways to adjust Uset manually:
– Adjustment with immediate activation of the new value:
Adjustment of the setpoint has an immediate effect on
load output quantities.
– Setpoint adjustment:
Adjustment of the setpoint does not effect load output
quantities until after activation.
Iset – Selecting a Setpoint for Output Current
The procedure for selecting output current Iset is identical to the
procedure for selecting output voltage Uset (page 22).
However, the following controls and displays must be substituted:
• Iset rotary knob (chapter 3 [12])
• Right-hand display (chapter 3 [13])
• Yellow Iset/A LED (chapter 3 [14])
• Setting resolution:
3 step widths are possible: 0.01 A, 0.1 A or 1 A.
Attention: Iset may not be set to a value which exceeds Ilim!
Pre-selecting a Setpoint
☞ Press the <ENTER> key [15].
!The display [5] is switched from Uout (measured voltage value)
to Uset (voltage setpoint). The decimal place which corresponds
to the selected resolution setting blinks.
!The green Uout/V LED goes out, and the yellow Uset/V LED [6]
lights up (as long as the Uset/V LEDis continuously illuminated,
direct adjustment with immediate activation of the new value
can be executed with the rotary knob [7]).
• Setting resolution:
!3 step widths are possible: 0.01 V, 0.1 V or 1 V.
☞ Repeatedly press the <SELECT> key [13] until the desired
decimal place (step width) blinks at the display.
• Output Voltage:
☞ Press either the <> key or the <> key [11]!
<> (increment) Increases the value
<> (decrement) Decreases the value
!The yellow Uset/V LED [5] blinks. This indicates that new
values are displayed but not activated during adjustment. The old Uset value remains active.
☞ Press the <> key or the <> key [11] until the desired
value appears at the display [5].
!Each time the key is pressed, the pre-selected setpoint
value is changed by an amount which corresponds to the
value selected with the resolution setting function.
!Pressing and holding the respective key results in rapid
scrolling, regardless of the step width.
☞ Acknowledge the selected value with the <ENTER> key
[15]. The new value is now activated at the output [4], and
the yellow Uset/V LED [6] is continuously illuminated.
☞ Repeatedly press the <ENTER> key [15] in order to switch
back and forth between Uset and Iset.
Figure 4.2.2 a Adjusting Uset with a Specified Fixed Value
Figure 4.2.1 bContinuous Adjustment of Uset
4.2.2 Pre-selected Setting (ENTER, arrow keys)
• Output voltage jumps from the old value to the new value in a
single step (no intermediate values).
Attention:Uset may not be set to a value which exceeds Ulim!
If the application requires that switching to a new setpoint takes
place in a single jump by pressing a key (without semi-continuous
adjustment), this can be accomplished as follows.
Proceed to the basic setting menu using the CE/LOCAL key, i.e.
to the display of momentary output values. Press the ENTER key
in order to switch to the pre-selection setting mode. After
switching to this mode, the currently valid setpoint values appear
at both displays (indicated by the LED to the right of the display).
As a default function, voltage adjustment is always activated first,
which is indicated by a blinking decimal place at the
corresponding display. Repeatedly press the ENTER key in order
to switch back and forth between current adjustment and voltage
adjustment.
The new setpoint can be selected with the arrow keys (and only
with the arrow keys), after selecting voltage or current setpoint
adjustment and the desired resolution. The SET LED blinks in
order to indicate that a change has been made but not yet
activated. The new setpoint is activated after acknowledging with
the ENTER key.
22GMC-I Messtechnik GmbH
4.3Switching the Power Output On and Off
4.4Limiting the Allowable Working Range: Ulim, Ilim
The <OUTPUT> key [7] (chapter 3 ) functions independently of the
current operating state of the KONSTANTER.
Further details regarding how the device reacts when the
<OUTPUT> key [7] is activated are included in chapter 3.
Closed Loop Control Mode
The power output can be activated or deactivated by pressing the
<OUTPUT>key [7].
OUTPUT OFF OUTPUT ON
REMOTE status (remote control)
(chapter 3[25], page 69)
The <OUTPUT> key is disabled and has no function.
LOCAL LOCKED status (front panel controls disabled)
(chapter 3[17], chapter 4.12)
The <OUTPUT> key is disabled and has no function.
trG out (T_MODE OUT) and External Trigger Active
The output can be activated and deactivated with the trigger
signal (trG parameter OUT selected).
If the power output has been disabled by applying an external trigger signal to the analog interface, it cannot be switched on,
neither by means of a command nor with the keys. The
corresponding command is not executed, and bit 4 in event
register B is set (output-on error). “Err 25” also appears briefly
at the display as a warning in the event of manual operation.
The status of the power output
the red LED above the <OUTPUT> key:
LED on = output activated
LED off = output deactivated
In the case of OUTPUT OFF, the control mode displays are also
switched off (chapter 3[6]).
If the output has be deactivated as a result of overtemperature protection, the red OTP LED[17] in the control mode display
lights up. The output cannot be activated until the
KONSTANTER has returned to its normal operating
temperature.
If the Pon RCL function is active, the output is automatically
reactivated after the device has returned to its normal
operating temperature.
Functions which may influence the status of the output
include:
FunctionsMeaningManual
OVP (OVSEt)Overvoltage
protection
OCPOvercurrent
protection
Pon (POWER_ON)page 27page 70
SEqSequencechapter
trG (T_MODE)page 29page 74
(chapter 3[33]) is indicated by
Remote
Operation
page 24page 69
page 26page 68
4.7.3
Operation
page 71
Allowable setting ranges for voltage and current can be limited in
order to assure ideal matching to the working ranges of the
connected power consumer. The Ulim and Ilim setting functions
are provided to this end.
Setting options can be selected with the respective <SELECT>
key. Setting resolution is selected with the <RESOL> key. The
setting itself is entered with the arrow keys. New settings become
immediately active.
Attention!
These settings represent so-called soft limits. This means that
values which lie within these limits can be selected both manually
and via the computer interface, and that a corresponding error
message is otherwise generated.
Attention!
The actual output quantity is the sum of the digitally selected
setpoint value and the setpoint value specified via the analog
interface. This makes it possible to select values which exceed
the specified soft limit.
Ulim – Setting the Upper Voltage Limit Value
Function
• Upper setting limit (soft limit) for Uset
• Prevents inadvertent violation of the maximum voltage value
when adjusting Uset.
• Protection for the connected power consumer
• Ulim has higher priority than Uset.
• Manual and computer-aided settings for Uset may not exceed
Ulim.
• Ulim cannot be set to a value which is less than a previously
selected Uset value. Uset must be reduced far enough to allow
for the new Ulim setting.
Settings
The setting procedure is described in chapter 4.8.1.
Ilim – Setting the Upper Current Limit Value
Function
• Upper setting limit (soft limit) for Iset
• Prevents inadvertent violation of the maximum current value
when adjusting Iset.
• Protection for the connected power consumer
• Ilim has higher priority than Iset.
• Manual and computer-aided settings for Iset may not exceed
Ilim.
• Ilim cannot be set to a value which is less than current
setpoint Iset. Iset must be reduced far enough to allow for the
new Ilim setting.
Settings
The setting procedure is described in chapter 4.8.1.
GMC-I Messtechnik GmbH23
4.5 Description of OVP and OCP Protection Functions
Protection for the connected power consumer and the
KONSTANTER by means of the following functions:
OVP – overvoltage protection
Function
•
Protection
•If voltage at the output terminals exceeds the selected
value, the power output is deactivated.
•Triggering of overvoltage protection causes immediate (< 200
transmitter is disabled, and the electronic sink for discharging
the output capacitors over a period of approximately 350 ms is
activated. In addition, bit 4 (OVPA) is set in event register A. Bit 4
remains set in status register A for as long as the trigger value is
exceeded.
•The
triggered (see also chapter 3 [17]), and the
out.
•As soon as the shutdown condition no longer exists, the power
output can be reactivated by pressing the
transmitting a
computer control
Settings
The setting procedure is described in chapter 4.8.1.
Note
•The
that the desired
undesired triggering of the
overshooting due to sudden output discharging (minimum
values: chapter 1.5.3)!
•The
value between the output terminals of the SSP. This voltage is
increased by the
(remote sensing) by an amount equal to voltage drop at the
output leads. For this reason, the above defined difference
between
during sensing mode operation.
•The triggering threshold for overvoltage protection is identical to
the displayed value, and is always active!
•Overvoltage protection response time is less than 200 µs.
Output voltage generated by the device may exceed
the duration of this response time. Maximum overshooting can
be approximately calculated as follows:
Uout = ISET [A] x 200 [µs] / Cout [µF]
ISET = selected current setpoint
Cout = capacitance of the output capacitor
Subsequent discharging time for the output capacitor depends
upon load, and corresponds to the specified values for response
time at Unom —>1 Volt included in chapter 1.5.3.
•Possible causes for triggering overvoltage protection are listed in
chapter 3 [17].
OCP – Overcurrent Protection
Functions
•
Protects
•Deactivates the power output when load current
reached, and the output is switched to the
mode
•Current can nevertheless be allowed to exceed Iset for specified,
short periods of time by specifying a delay time (see below), for
example:
!
!
!
!
for the connected power consumer
OVSET
s) deactivation of the output (
OVP
LED lights up as soon as overvoltage protection is
trigger signal
(“
OUTPUT ON” ➩ red
OVP
trigger value (OVSET) should be set at least 1 V higher
USET
output voltage in order to prevent
OVP
trigger value makes reference to the prevailing voltage
USET
parameter during sensing mode operation
OVSET
and
USET
the power consumer from continuous overcurrent.
.
Starting current for electric motors
In-rush current for capacitive power consumers
For testing the breaking performance of circuit breakers,
motor protecting switches, fuses etc.
For determining the short-term load capacity of contacts and
cables, as well as electrical and electronic components
OUTPUT OFF
to the analog interface or by means of
OVP
function resulting from
must be correspondingly increased
). The HF power
red OUTPUT
<OUTPUT> key,
OUTPUT
LED lights up).
current regulating
LED goes
OVSET
Iset
has been
by
for
!
In order to maintain short response times when
programming voltage increases
•The red
chapter 3 [17]), and the
•The power output can be reactivated at any time by pressing the
<OUTPUT> key,
interface
OUTPUT
OCP
LED lights up as soon as
red OUTPUT
by transmitting a
or by means of
LED lights up).
computer control (OUTPUT ON ➩ red
OCP
is triggered (see also
LED goes out.
trigger signal
to the analog
Settings
See description on page 26.
DELAY – Output Off Delay for OCP
Functions
•Delay time prior to deactivation of the power output after
triggering of current regulating (
•Only enabled with activated
•If output current
the shutdown sequence is aborted.
•If current regulation is triggered again, the routine is started once
again (at 00.00).
•The default setting after
Iout
drops below
Iout = Iset)
OCP
function (
Iset
RESET (*RST
OCP ON
before
) is 00.00.
)
DELAY
time elapses,
Settings
The setting procedure is described in chapter 4.8.1.
4.6 Display of Momentary Output Values Uout, Iout and
Pout
Uout – Display Momentary Measured Voltage Value
•Appears at the
•The
•The measured voltage value is automatically displayed again
•The momentary measured voltage value is displayed
green Uout/V
momentary measured voltage value.
approximately 10 seconds after the last setting has been made.
immediately if the user exits the device functions setting mode by
pressing the
Iout – Display Momentary Measured Current Value
•Appears at the
•The green
momentary measured current value.
•The measured current value is automatically displayed again
approximately 10 seconds after the last setting has been made.
•The momentary measured current value is displayed
immediately if the user exits the device functions setting mode by
pressing the
Pout – Display Momentary Output Power
Repeatedly press the
until the respective Pout LED lights up. Momentary output power,
based upon measured Uout and Iout values, appears at the
corresponding display ([9] or [13]).
UI
_
– Display Measured Values in U/I Min-Max Memory
Functions
•Reads out stored values for
display, or via the computer interface.
•Stored Min-Max values can be read out regardless of the status
selected for the
Settings
See description on page 28.
left-hand display
LED [10] lights up and indicates display of the
<CE/LOCAL>
right-hand display
Iout/A
LED [14] lights up and indicates display of the
<CE/LOCAL>
<SELECT>
MINMAX
[9]
key [23].
[13]
key [23]
function.
key [11] or the
Umin, Umax, Imin
<SELECT>
and
Imax
key [15]
at the
24GMC-I Messtechnik GmbH
4.7 Operating Menu via the FUNCTION Key
FUNCTION
FUNCTION
ENTER
FUNCTION
+
Jump to
last edited
function group
Jump to
last edited
function
e.g.
e.g.
e.g.
*
CE/LOCAL
FUNCTION
FUNCTION
FUNCTION
+
FUNCTION
+
FUNCTION
+
The FUNCTION menu consists of the following functions for
configuring the KONSTANTER’s parameters:
* Sequence function: appears at left-hand display, right-hand display is blank.
** Parameter selection for sequence control: appears at left-hand display, selected parameter
appears at right-hand display.
TP: oFF / on / rSt
-
Figure 4.7 aAccessing the Functions Submenu
☞ If the function group menu is currently open, you can access
the following function groups:
■ Interface Configuration
bUSAddrNP: 0 / 1 / ... / 13 /... / 30 / (UNL)
bAUdNP: 50 / 75 / 150 / 200 / ... / 4800 /
9600 / 19200
dbitNP: 7 / 8
PbitTP: nonE / ZEro / EVEn / odd / onE
SbitNP: 1 / 2
NP: Numeric Parameters
• Can be selected as a specified value or as a value within an
interval with corresponding resolution.
•Adjust resolution (decimal place) by pressing the <RESOL> key.
• Select the value with the <> key or the <> key.
• Displayed value = selected value.
TP: Text Parameters
• Select parameters with the <> key or the <> key.
• Acknowledge your selection by pressing the <ENTER> key.
Settings
☞ Press the <FUNCTION> key at the front of the KONSTANTER.
If no settings have yet been changed via the FUNCTION menu
after switching the device on, the SEt function group is accessed by
pressing the <FUNCTION> key.
If settings have already been changed in the function menu, the last edited function group or function is accessed automatically.
• Accessing the last edited function:
☞ If you want to configure a function included in the current
function group, select the function group by pressing the
<FUNCTION> key. Detailed information is included in the
respective section of this chapter.
If you want to configure a function from another function group, first
☞
return to the
function group menu
• Exit the function menu completely:
☞ Press the <CE/LOCAL> key once again.
!Device is returned to Uout / Iout display.
by pressing the
<CE/LOCAL>
key.
☞ If the desired function group does not appear immediately at
the left-hand display, repeatedly press the <FUNCTION> key
until the name of the desired function group appears at the
left-hand display (SEt in this example).
•
Forward scrolling
through function groups in this menu:
☞ Repeatedly press the <FUNCTION> key or
☞ Press and hold the <FUNCTION>key and repeatedly press the
<> key at the same time.
• Reverse scrolling through function groups in this menu:
☞ Press and hold the <FUNCTION> key and repeatedly press the
<> key at the same time.
•
Accessing functions in the selected function group:
☞ Press the <ENTER> key.
Figure 4.7 bSwitching Amongst Individual Function Groups
Note
The order in which the commands are explained in the following sections
corresponds to the actual sequence for manual operation of the SSP
KONSTANTER. This makes it easy to configure device functions and parameters
without leafing back and forth through the instructions!
GMC-I Messtechnik GmbH25
4.7.1 SET – “Setup” Function Group
FUNCTION
+
FUNCTION
+
FUNCTION
+
FUNCTIONPress once
ENTER
Press repeatedly if necessary
Jump to last
edited setup function
Start-up performance after power on
CE/LOCAL
CE/LOCAL
CE/LOCAL
CE/LOCAL
Uout
Iout
Display of measured
values
*
Select the text parameter
from an options loop
From the SEt function group
ENTER
*
Activate
the setting
CE/LOCAL
FUNCTIONSAVE
RCL
Uset/IsetFUNCTION
MENU
SAVE
MENU
RCL
MENU
FUNCTION
MENU
Abort
without change
OCP – Activate Overcurrent Protection
Functions
See functions description on page 24.
Setting Parameters
☞ OFF (default setting after RESET (*RST))
OCP function inactive
Continuous current limiting (current regulation)
☞ ON
OCP function activated
The output is deactivated as soon as current limiting has been
active for the specified DELAY time.
Settings
☞ The procedure for configuring OCP is described in principle in
chapter 4.7 and in Figure 4.7.1 a.
!However, the OCP display and a
related text parameter appear in
the setup function window.
☞ The selected text parameter does
not become active until the
<ENTER> key is pressed!
Figure 4.7.1 bPath to Selection of the OCP Text Parameter
Note
Load current for the output capacitor is also acquired by the
current regulator, and is limited to a value of Iset load. If the OCP
function has been activated and ISET and DELAY have been set to
low values, the output may even be deactivated if output voltage
is increased. For this reason, DELAY must at first be set somewhat
higher than the resulting output voltage response time.
Figure 4.7.1 a Path to Pon Settings
and other setting options in the Set menu
26GMC-I Messtechnik GmbH
Pon – Output Switching Status, Response After Power On
Select the text parameter
from an options loop
From the SEt function group
ENTER
*
CE/LOCAL
FUNCTIONSAVE
RCL
Uset/IsetFUNCTION
MENU
SAVE
MENU
RCL
MENU
FUNCTION
MENU
Activate
the setting
Abort
without change
Select the text parameter
from an options loop
From the SEt function group
*
ENTER
CE/LOCAL
FUNCTIONSAVE
RCL
Uset/IsetFUNCTION
MENU
SAVE
MENU
RCL
MENU
FUNCTION
MENU
Activate
the setting
Abort
without change
UI_ – Configure U/I Min-Max Measured Value Memory (MINMAX)
Functions
• Automatically determines KONSTANTER configuration after
power on.
Setting Parameters:
☞ rSt (RESET): Factory default settings are utilized.
☞ SbY (STANDBY): Same settings as prior to shutdown, power
output remains inactive (OUTPUT OFF).
Disabled front panel controls are re-enabled after power on.
As from firmware version 3.006:
Disabled front panel controls remain disabled after power on.
☞ rcl (RECALL): Same settings as prior to shutdown. Disabled
front panel controls remain disabled after power on.
Settings
☞ The procedure for configuring Pon is described in principle in
chapter 4.7 and in Figure 4.7.1 a.
!However, the Pon display and a related text parameter appear
in the setup function window.
Functions
• Saveminimum and maximum measured voltage and current
values.
• Reads out stored values for Umin, Umax, Imin and Imax at the
display, or via the computer interface.
• The MINMAX function can be temporarily set to OFF, for
example before deactivating the output when changing
devices under test. Stored values are then no longer updated,
nor are they automatically reset.
• Stored Min-Max values can be read out regardless of the
status selected for the MINMAX function.
• Stored Min-Max values can always be reset with the
_ RST command.
UI
• Stored Min-Max values can also be reset with the *RST
command, or with the key combination <CE/LOCAL> + <ENTER> (= RESET).
• The status of the MINMAX function is also saved to SETUP
memory by pressing the SAVE key.
•If the Pon function is set to SbY or rcl, the MINMAX function is
activated upon power-up, but previously measured Min-Max
values are lost when the KONSTANTER is switched off.
•The MINMAX function can also be controlled via the trigger input
at the analog interface (see page 29).
Setting Parameters
☞ OFF: Storage of Min-Max values is deactivated.
☞ ON: Storage of Min-Max values is activated.
☞ RST: Contents in the Min-Max memory are reset or replaced
with momentary measured values:
Umin = UoutUmax = Uout
Imin = IoutImax= Iout
Figure 4.7.1 cPath to Selection of the Pon Text Parameter
Note
• The status of the Pon function is not saved as a device setting
to SETUP memory with the <SAVE> key.
• The following settings are recommended, depending upon
how the KONSTANTER is used:
!Pon rSt: use in computer controlled systems
!Pon SbY:common laboratory use
!Pon rcL:applications which must continue in an unchanged
fashion after power failures.
Settings
☞ The procedure for configuring UI_ is described in principle in
chapter 4.7 and in Figure 4.7.1 a.
!However, the UI_ display and a
related text parameter appear in the
setup function window.
Default setting after
☞
RESET
(*
RST
):
OFF!
GMC-I Messtechnik GmbH27
Figure 4.7.1 d Path to Selection of the Pon Text Parameter
UI_– Display Measured Values in
CE/LOCAL
Uout
Iout
FUNCTIONSAVERCL
Abort
Abort
Abort
Abort
FUNCTION
+
FUNCTION
ENTER
FUNCTION
FUNCTION
FUNCTION
FUNCTION
Uset/IsetFUNCTION
MENU
SAVE
MENU
RCL
MENU
FUNCTION
MENU
Select the text parameter
from an options loop
From the SEt function group
*
ENTER
CE/LOCAL
FUNCTIONSAVE
RCL
Uset/IsetFUNCTION
MENU
SAVE
MENU
RCL
MENU
FUNCTION
MENU
Activate
the setting
Abort
without change
U/I Min-Max Memory
rnd – Rounding Off the Displayed Measured Value
Functions
• Reads out stored values for Umin, Umax, Imin and Imax at the
display, or via the computer interface.
• Stored Min-Max values can be read out regardless of the
status selected for the MINMAX function.
Settings
☞ Select from the Set menu with the <FUNCTION> key.
☞ After U
_, U , I_ or I is displayed for the first time, browsing
amongst these values is possible with the <> and <> keys.
!U_, U , I_ or I appears in the setup function display, and the
corresponding measured value appears at the left-hand
display.
☞ Default setting after RESET (*RST): Measured value memory is
cleared!
Applications
The ROUND function can be used where fluctuating measured
values are to be expected due to the application (e.g. in
manufacturing), which may cause unnecessary concern.
Functions
The ROUND function defines how many decimal places will be
displayed for measured Uout and Iout values.
Setting Parameters
☞0: no rounding
☞ –1: rounded off by one decimal place
☞ –2: rounded off by two decimal places
Settings
☞ The procedure for configuring Rnd is described in principle in
chapter 4.7 and in Figure 4.7.1 a.
!However, the rnd display and a
related text parameter appear in the
setup function window.
Figure 4.7.1 ePath to Selection of the rnd Text Parameter
☞ Default setting after RESET (*RST): unchanged
28GMC-I Messtechnik GmbH
Loading...
+ 64 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.