Immediately after receipt, unpack the KONSTANTER and all
included accessories, and inspect for damage and completeness.
Unpacking
Great care must be exercised when removing the electronic
device from the package.
☞ Pull the KONSTANTER from its package.
☞ In doing so, do not grasp rotary knobs, terminals or jacks in
order to avoid damage.
☞ Do not allow the KONSTANTER to fall out of the packaging.
Controls, displays, terminals and internal components might
otherwise be damaged.
☞ Do not, under any circumstances, touch the KONSTANTER’s
electrical terminals before grounding yourself to the housing
in order to neutralize potential differences. Sensitive
electronic circuitry might otherwise be damaged as a result
of electrostatic discharge.
☞ The KONSTANTER is delivered in recyclable packaging, which
provides for adequate protection during transport as
substantiated by testing. If the instrument is repacked at a
later point in time, the same packaging or its equivalent must
be used.
Visual Inspection
☞
Compare the order number or type designation included on
the packaging and/or the serial plate with the particulars
shown in the shipping documents.
☞ Make sure that all accessory components have been
included
(see chapter 1.3, “Options and Accessories”).
☞ Inspect the packaging, as well as mechanical instrument
and accessory components for possible transport damage.
Complaints
If damage is discovered, immediately file a claim with the
freight forwarder (save the packaging!). If other defects are
detected, or in the event that service is required, inform your
local representative, or contact us directly at the address
included in the last page of this handbook.
IIWarnings and Safety Precautions
The KONSTANTER has been manufactured and tested in
accordance with the electrical safety regulations listed under
“Technical Data” as a safety class I device, and has been shipped
from the factory in flawless technical safety condition. In order to
maintain this condition and to assure safe operation, users must
observe all notes and warnings included in these operating
instructions. These are identified with the following headings:
ATTENTION!
A note concerning operation, practical advice or other information
which must be adhered to in order to prevent damage to the
KONSTANTER, and to assure correct operation.
WARNING!
An operating procedure, practical advice or other information
which must be adhered in order to assure safe operation of the
KONSTANTER, and to prevent personal injury.
The most important warnings are summarized below.
Reference is made to these warnings at appropriate points
within the operating instructions.
WARNING I – Protective Grounding
The KONSTANTER may only be placed into operation after the
protective conductor has been connected. Interruption of the
protective conductor, either inside or outside of the
KONSTANTER, or disconnection of the protective conductor
terminal may transform it into a source of danger. Intentional
interruption is prohibited.
The device is connected to the mains by means of a 3 conductor
cable with mains plug. The plug may only be inserted into a
suitable outlet with earthing contact. This safety precaution must
not be defeated though the use of an extension cable without
protective conductor.
WARNING II – Impaired Safety
If it can be assumed that safe operation is no longer possible, the
KONSTANTER must be removed from service and secured against
inadvertent use. It must be assumed that safe operation is no longer
possible:
☞ If the KONSTANTER demonstrates visible damage
☞ If the KONSTANTER no longer functions
☞ After lengthy periods of storage under conditions which deviate
from specified storage conditions
☞ After extraordinary stressing due to transport
WARNING III – Opening Housing Panels
Voltage conducting parts may be exposed when housing panels
are opened, as long as the KONSTANTER is connected to supply
power.
Any contact with these exposed conductive parts is life
endangering.
For this reason, housing panels may only be opened and/or
removed by trained personnel who are familiar with the dangers
involved.
WARNING IV – Repair by Trained Personnel
Voltage conducting parts may be exposed when housing panels
are opened, as long as the KONSTANTER is connected to supply
power.
Maintenance and repair work, as well as internal balancing, may
only be performed by trained personnel who are familiar with the
dangers involved.
The KONSTANTER must be disconnected from all external
voltage sources before work of this type is performed, in as far as
this is possible. A five minute waiting period must be observed
after disconnection in order to allow internal capacitors to
discharge to safe voltage levels.
WARNING V – Fuse Replacement
Only specified fuse types with the specified nominal current rating
may be used to replace blown fuses (see “Technical Data” and
specifications on the serial plate).
Tampering with fuses or fuse holders is prohibited (“repairing”
fuses, short-circuiting fuse holders etc.).
Significance of Symbols
Indicates EC conformity
Observe EGB directives
Warning concerning a source of danger
(attention: observe documentation!)
4GMC-I Messtechnik GmbH
1Technical Description
1.1 Features and Range of Applications
Depending upon where they are used and prevailing local
conditions, electrical and electronic devices may be subject to
significant supply power fluctuation in the absence of stabilizing
and back-up systems.
Automotive electrical system voltage characteristics during starter
motor operation offer a typical example.
R&D, production and test departments must be able to assure
that electrical equipment is capable of executing all required
functions under these types of conditions.
Type SSP 120, SSP 240 and SSP 320 KONSTANTER power
supplies from GMC-I Messtechnik GmbH support users in fulfilling
this objective.
In particular in automated test systems for routine testing,
SSP KONSTANTERs are capable of high throughput rates. Short response times assure accurate emulation of rapidly changing
voltage and current characteristics.
Power consumer characteristics relative to dynamic supply power
can thus be tested and simulated very easily.
The integrated calibration procedure and the included calibration
certificate are especially important for use in production and testing
departments which are certified in accordance with ISO 9000.
SSP KONSTANTERs (single output system power supplies are
programmable, voltage and current regulated DC power supplies
with 120, 240 or 320 W of output power.
They are microprocessor controlled, and are furnished with an
addressable RS 232C serial interface as standard equipment. Up
to 30 KONSTANTERs can thus be controlled from a single PC
port.
An IEEE 488 interface can be integrated as an option. Complete
control of all device functions, as well as querying of measured
values, setting parameters and device statuses is possible via
both interfaces.
Manual adjustment of voltage and current is accomplished by
means of two rotary pulse encoders with selectable adjusting
sensitivity.
Two 4-place multifunction displays are used for precision read-out
of measured values (V, A and W), as well as for menu-driven
configuration of numerous additional functions such as limiting
setting ranges, overvoltage protection, overcurrent shutdown
delay and programmable, digital signal inputs and outputs.
Up to 243 settings can be saved to memory and recalled either
individually or sequentially, for example in order to generate
specific voltage or current characteristics. And innovative circuit
technology allows for nearly load independent response times of
less than 1 ms.
Output parameters can thus be superimposed with AC signal
right up into the kHz range via the analog interface which is
furnished as standard equipment.
The measuring function includes a Min-Max value memory, limit
monitoring signals and a HOLD function.
1.2 Functions
Adjustable functions:
❑ Voltage and current setpoint values
❑ Voltage and current limit values (soft-limits)
❑ Activate / deactivate the output
❑ Overvoltage protection trigger value
❑ Overcurrent response (limiting with or without shutdown)
❑ Delay time for overcurrent shutdown
❑ Start-up performance (power-on status)
❑ Reset device settings
❑ Save / recall device settings
❑ Save / recall setting sequences
❑ Sequence definition
❑ Trigger input function selection
❑ Function selection for digital control inputs
❑ Operating parameters for the measuring function (Min-Max
value memory, limit monitoring, display resolution)
❑ Momentary measured voltage, current and power values
❑ Minimum and maximum measured voltage and current
values
❑ Current device settings (individual or complete)
❑ Current operating status (control mode, overtemperature,
busy)
❑ Occurred events (mains or phase failure, overtemperature,
overvoltage, overload, programming error)
❑ Memory contents
❑ Device ID
❑ Calibration date
Protective Functions and Additional Functions
❑ Sensing terminals protected against polarity reversal and
automatic switching to auto-sensing
❑ Control panel disabling
❑ Output overvoltage protection
❑ Output protected against reverse polarity
❑ Overtemperature protection
❑ Backup battery for device settings memory
❑ Master-slave series operation
❑ Inrush current limiting
❑ Temperature controlled fans (with 240 and 320 W models)
GMC-I Messtechnik GmbH5
1.3 Options and Accessories
1.4 Functional Principle
Options
(See order information on last page.)
Devices can be equipped with an IEEE 488 plug-in interface
module for integration of SSP KONSTANTERs into IEC bus
controlled systems.
The interface is in compliance with requirements set forth in the
IEEE488.2 standard (IEC 625-2), and is frequently designated
as a GPIB (general purpose interface bus) or an HP-IB (Hewlett
Packard interface bus).
Programming of all device functions, as well as querying
measured and configured parameters, is possible via the
interface. Several specific interface functions are also provided.
The interface option is supplied as a separate plug-in module for
easy, subsequent installation to the KONSTANTER.
Included accessories:
The following accessories are included with the
SSP KONSTANTER:
❑ These operating instructions
❑ 1 mains power cable with earthing contact plug
(approx. 1½ meters)
Additionally available accessories:
(See order information on last page.)
❑ 19" rack adapter, 1 x 32 N
required for installation of one series SSP 32 N or SLP 32 N
KONSTANTER to a 19" rack
❑ 19" rack adapter, 2 x 32 N
required for installation of two series SSP 32 N or SLP 32 N
KONSTANTERs to a 19" rack
❑ Mains jumper cable, 0.4 meters
The cable is equipped with a 10 A inlet plug and a 10 A inlet
socket. It is used for looping mains power through when
several KONSTANTERs are mechanically combined into a
multi-channel unit. In this way, the complete unit requires
only one mains power cable.
❑ RS 232 bus cable, 2 meters
For connecting one KONSTANTER to an RS 232 interface.
(extension cable, 9-pin socket / 9-pin plug connector)
Power Supply
Required DC supply power is generated for each respective
circuit from mains power which has been fed to the power pack
via an interference suppression filter, a wire fuse, the mains switch
and inrush current limiting.
The Central Processing Unit (CPU)
Overall control of the SSP KONSTANTER is accomplished by
means of the CPU on PCB A. It uses an 80C32 8-bit
microcontroller with 64 kilobytes of program memory and 32
kilobytes of battery-backed CMOS RAM.
An 11 MHz pulse generator establishes the clock frequency for
the processor, and creates a time reference for the serial
interface.
A watchdog circuit monitors processor activity and disables
access battery-backed RAM in the event of supply power failure.
Operation
Generally speaking, all of the functions of the SSP KONSTANTER
can be controlled with the adjusting elements at the front panel,
via the RS 232 interface or via the optional IEEE 488 plug-in
interface module.
Displays and Control Panel
Both 4-place 7-segment displays and the individual LEDs are
controlled statically by the processor via the registers. Front panel
key operations are processed by means of a group interrupt and
subsequent direct querying. The rotary pulse encoders control
the increment and decrement counters depending upon the
direction of rotation. Each time an adjusting element is activated,
an interrupt occurs at the CPU which then causes an appropriate
response.
Interface Option
The SSP KONSTANTER with installed interface module allows for
control via the IEEE 488 bus in addition to serial operation.
Remote Control
Device messages received by the interface are forwarded to the
CPU where they are first saved to RAM. After receiving an end-ofmessage character, data are checked for correct syntax,
plausibility and limit values. Valid commands are subsequently
executed.
6GMC-I Messtechnik GmbH
Setup Procedure
Setup data are processed and forwarded to the respective I/O
controller of the respective function unit. Each setting value for
output voltage, output current or overvoltage protection triggering
is converted to a proportional control voltage by a 12 bit DAC,
and is fed to the respective controller or comparator as a setpoint
or a reference quantity.
Actual output voltage is ascertained by a voltage monitor, whose
automatic sensor switching inputs are connected either to the
output terminals or the sensing terminals.
Actual output current is acquired as a voltage drop at a shunt
situated in the negative output conductor, and is amplified by the
current monitor to a scaled signal.
In order to achieve rapid downward adjustment of output voltage
even with minimal output load, the device is equipped with a
limited sink function (limited to approximately 15 W continuous
power). The sink function is implemented by means of BET
technology (bidirectional energy transfer). This technology assures
charging and discharging of the output capacitor within an equally
short period of time, even at no-load operation. The sink function
is activated as soon as, and for as long as output voltage exceeds
the current setpoint value (also in the event of energy recovery
from a parallel connected voltage source).
The source and sink functions are enabled when the output On /
Off status is set to ON, and the source function is disabled when
the status is set to Off, in which case the sink function is
deactivated after approximately 300 ms (highly resistive output).
Measuring Procedure
Monitor-amplifier output signals, which are proportional to actual
output voltage and current, are fed to an analog multiplexer (MUX)
which switches one of the two signals to the input of the analogdigital converter (ADC) depending upon the desired measured
quantity. The converter, controlled directly by the CPU, functions in
accordance with the dual-slope principle and traces measured
value acquirement back to a time measurement (meter reading).
The actually measured decimal value is calculated based upon this
data, and is stored to RAM. Depending upon the circumstances,
the measured value display is refreshed, an extreme value
comparison is performed for the Min-Max function or the measured
value is made available at the computer interfaces’ data output
buffer.
Monitoring Functions
– Control Mode Recognition and Overload
An electrically isolated digital signal is derived from the output
signals of the voltage and current regulators, which indicates
the currently active control mode (constant voltage or constant
current mode), as long as overload operation is not active.
“Overload” indicates that power limiting has been triggered as a
result of selected parameter settings and prevailing load. These
operating conditions are evaluated by the CPU (e.g. for OCP
function), are indicated with LEDs and are used to generate
status and event registers for computer control.
– Overvoltage Monitoring
After bypassing the voltage monitor, device output voltage is
additionally compared to an adjustable limit value within a range
of 3 V to 120% nominal voltage by a comparator, and the
output is deactivated if the limit value is exceeded and an OVP
message is generated (“CV” LED blinks, status and events
registers).
– Overcurrent Monitoring
Overcurrent monitoring can be activated and deactivated.
Response time is defined with the delay parameter. If, with
activated OCP function, the device is in the current regulating
mode for at least the duration of the selected delay time (“CC”
LED is illuminated), the power output is switched off and an
OCP message is generated (“CC” LED blinks, status and
events registers).
–Temperature Monitoring
Temperature is acquired by means of PTC resistors at
representative points (circuit breaker heat sink), and is
converted into a proportional electrical signal. As of an initial
threshold temperature of approximately 70C, fan voltage is
increased in a linear fashion as temperature increases. The
device can be operated at any output load up to the maximum
specified operating temperature after the fan has been
activated. If ventilation is impeded, or at excessive ambient
temperatures, temperature at the sensor may reach the upper
threshold of approximately 90C. If this is the case, an
overtemperature message is entered to the status and event
registers. The output is deactivated by the CPU after 5
seconds, and OTP triggering is indicated by the blinking
“Pmax” LED. After sufficient cooling has occurred, the status
message is cleared and a ready for operation message is
entered to the event register. If the POWER-ON function is set
to “Recall”, the output is reactivated automatically. If the
POWER-ON function is set to “Standby” or “Reset”, the output
remains deactivated and can only be reactivated by pressing
the OUTPUT key, or by means of a command from the control
computer. After actual temperature has fallen to below the
lower threshold value, fan speed is reduced automatically.
GMC-I Messtechnik GmbH7
1.5 Technical Data
U/V
U
nom
I / A
I
nom
0
0.5 I
nom
P
nom
Short-Term
Working Range
Voltage Setting
Current Setting
0.5
U
nom
Range
Range
380.5
17
1488
221.5
Dimensions in millimeter
1.5.1 General Data
Power Supply
ConnectionInput:10 A IEC inlet plug
Output:10 A IEC inlet socket,
Line voltage230 V ~ +10 -15%, 47 to 63 Hz
Power consumptionSee chapter 1.4.3
Inrush CurrentMax. 50 A
Mains fuse1 ea. T 4 A / 250 V (6.3 x 32 mm, UL)
internal: 1 ea. T 5 A 250 V (5 x 20 mm)
Output
Connection
OutputFront panel, 2 ea. 4 mm safety jacks
Rear , 6-pole plug-in screw terminal block
SensorRear panel, included in 6-pole plug-in
screw terminal block
Analog interfaceRear, 11-pole plug-in screw terminals
Regulator typePrimary switched-mode regulator with
BET technology
Operating modesAdjustable constant voltage / constant
current source with automatic sharp
transition
Output isolationFloating output with “safe electrical
separation” from the mains input and
computer interfaces
Max. allowable potential, output–ground:
120 V, capacitance, output–ground
(housing) 60 nF
Output operating range
without switch,
without fuse
S
Analog Interface
Connection11-pole plug-in screw terminal block
Reference potential Output minus pole, floating TRG input
Connector pin assignments:
Pin DesignationFunction
1SIG1 OUT
2SIG2 OUT
3TRG IN –
4TRG IN +
5 +15 V
6AGND
7U
–Analog, inverting voltage control input
set
+Analog voltage control input
8U
set
+Analog current control input
9I
set
10 U-MONOutput voltage measuring output
11 I-MONOutput current measuring output
Digital, programmable, open-collector outputs (max.
30 V / 20 mA)
Auxiliary voltage, +15 V / max. 50 mA
reference point, connected to -output
(0 ... –5 V correspond to 0 ... U
(0 ... +5 V correspond to 0 ... U
(0 ... +5 V correspond to 0 ... I
(0 ... 10 V correspond to 0 ... U
(0 ... 10 V correspond to 0 ... I
, Ri = 10 k)
nom
, Ri = 10 k)
nom
, Ri = 10 k)
nom
, Ri = 9.8 k)
nom
, Ri = 9.4 k)
nom
Addressable V.24 – RS 232C Interface
Input9-pin subminiature socket connector
Output9-pin subminiature plug connector
Operating modehalf-duplex, asynchronous, XON / XOFF
Transmission speed adjustable from 50 to 19,200 bits per second
Device addressselectable from 0 to 30 or UNL (unlist)
Max. Setting Rate approx. 15 settings per second
Max. Sampling Rateapprox. 7 measurements per second
IEC 625 – IEEE 488 Interface (optional)
Connection24-pin socket connector
per IEC 625.1, IEEE488.1
InterfaceSH1 SOURCE HANDSHAKE
functionshandshake source function
AH1 ACCEPTOR HANDSHAKE
handshake sink function
T6 TALKER
talker function with serial polling
and automatic unlisting,
no secondary address and no
Short-term range:
If lengthy operation in the short-term
range occurs, overtemperature protection may be triggered resulting in output
shutdown (see also short-term power in
chapter 1.5.3).
talk-only operation
L4 LISTENER
listener function with automatic
unlisting, no secondary
address and no listen-only operation
SR1 SERVICE REQUEST
service request function
RL1 REMOTE / LOCAL
remote to local switching
function with disabling
DC1 DEVICE CLEAR
reset function including selected
device clear
PP1 PARALLEL POLL
parallel polling function with remote
configuring
DT1 DEVICE TRIGGER
trigger function
C0 No controller function
E1/2 Open collector driver
Codes / formatsPer IEEE 488.2
Device addressSelectable from 0 to 30 or UNL (unlist)
Figure 1.5Dimensional Drawing (benchtop device)
Max. setting rateApprox. 40 settings per second
Max. sampling rate Approx. 15 measurements per second
8GMC-I Messtechnik GmbH
Electrical Safety
Safety classI
Overvoltage
categoryII for mains input
I for output and interfaces
Fouling factor2
Earth leakage
current2.5 mA, typ.
IEC 61010-1:1990 + A1:1992 / DIN EN 61010-1: 1993 /
VDE 0411-1:1994
Unless otherwise specified, all entries are maximum values and apply within an operating
temperature range of 0 to 50C, within the nominal power range and the nominal line
voltage range of 230 V
10 % after a warm-up period of 30 minutes.
Percentages make reference to the respective setting value or measured value.
SSP 240-20SSP 240-40SSP 240-80SSP 320-32
32 N 20 RU 20 P32 N 40 RU 12 P32 N 80 RU 6 P32 N 32 RU 18 P
0 ... 20 V0 ... 40 V0 ... 80 V0 ... 32 V
0 ... 20 A0 ... 12 A0 ... 6 A0 ... 18A
max. 240 Wmax. 240 Wmax. 240 Wmax. 320 W
max. 320 Wmax. 360 Wmax. 360 Wmax. 430 W
-0.5 A / K-0.3 A / K-0.15 A / K-0.5 A / K
5 mV / 10 mV, 5 mV
5 mA / 10 mA, 5 mA
0.15% +40 mV
0.5% +70 mA
25 mV
30 mA
5 mV
8 mA
15 mV
eff
50 mA
eff
0.5 mA
eff
40 mV
600 s
10 mV
3.33 mA / 10 mA, 3.33 mA
0.15% +45 mV
0.5% +45 mA
18 mV
30 mA
5 mV
8 mA
15 mV
eff
25 mA
eff
0.5 mA
eff
80 mV
300 s
20 mV
2 mA
0.15% + 80 mV
0.5% +25 mA
18 mV
15 mA
5 mV
5 mA
15 mV
eff
20 mA
eff
0.5 mA
eff
160 mV
200 s
50 mA
450 mV450 mV800 mV450 mV
40 mV
1 ms / 1 ms
1 ms / 1 ms
200 mA
< 5 ms / < 5 ms
< 5 ms / < 5 ms
2 mV / 10 mV, 2 mV
2 mA, 10 mA, 2 mA
0.15% +40 mV
0.5% +70 mA
0.65% +1.4 W
80 mV
1 ms / 1 ms
1 ms / 1 ms
120 mA
< 5 ms / < 5 ms
< 5 ms / < 5 ms
10 mV, 4 mV
2 mA / 10 mA, 1.2 mA
0.15% +40 mV
0.5% +25 mA
0.65% +1 W
160 mV
4 ms / 4 ms
4 ms / 4 ms
60 mA
< 10 ms / < 10 ms
< 10 ms / < 10 ms
10 mV, 8 mV
0.6 mA, 1 mA
0.15% + 80 mV
0.5% +15 mA
0.65% +1.2 W
10 mV
5 mA / 10 mA, 5 mA
0.15% +50 mV
0.5% +70 mA
30 mV
40 mA
10 mV
20 mA
30 mV
eff
(Ua > 10%U
eff
0.5 mA
eff
64 mV
500 s
64 mV
1 ms / 1 ms
1 ms / 1 ms
180 mA
< 5 ms / < 5 ms
< 5 ms / < 5 ms
10 mV, 4 mV
2 mA, 10 mA, 2 mA
0.1 W, 0.1 W
0.15% +40 mV
0.5% +70 mA
0.65% +1.4 W
nom
)
0 ... 25 V
0.1 V
2% +0.2 V
0 ... 50 V
0.2 V
2% +0.4 V
0 ... 100 V
0.4 V
2% + 0.8 V
20 A12 A6 A18 A
40 V80 V100 V64 V
230 V~ +10 / 15%
47 ... 63 Hz
510 VA, 350 W
45 VA, 15 W
620 VA
230 V~ +10 / 15%
47 ... 63 Hz
500 VA, 340 W
45 VA, 15 W
690 VA
230 V~ +10 / 15%
47 ... 63 Hz
500 VA, 340 W
45 VA, 15 W
690 VA
> 68%> 70%> 70%> 69%
200 kHz200 kHz200 kHz200 kHz
K330AK331AK332AK334A
1)
Specified values are increased by a factor of approximately 1.2 within the functional range for line voltage, namely -10% to -15%.
0 ... 40 V
0.2 V
2% +0.4 V
230 V~ +10 / 15%
47 ... 63 Hz
650 VA, 460 W
50 VA, 15 W
770 VA
GMC-I Messtechnik GmbH11
2Initial Start-Up
Warning!
Attention!
!
ANALOG INTERFACEOUTPUT
OUTIN RS 232
230V 5
FUSE T4
➁
➂
➁
Interface
Module
Ribbon
Cable
➁
➀
➀
➁
19“ Dummy Plate
19“ Abutment
➃
19'' Connector Plate
19'' Abutment
➃
➃
2.1 Preparing for Operation
2.1.1 Installing the IEEE 488 Interface Module
2.1.2 Installation to 19'' Device Racks
The SSP KONSTANTER housing allows for use as a benchtop
instrument, as well as for installation to a 19'' rack. One
KONSTANTER and a cover plate, or two devices next to each
other can be installed to the rack.
The benchtop instrument can be quickly retrofitted for rack
mounting.
The KONSTANTER must be disconnected from the
mains before installing the interface module. Switch the
KONSTANTER off, and then pull the mains plug from the
outlet. Wait at least 5 minutes before opening the
KONSTANTER, in order to assure that the capacitors
have been discharged to a safe voltage level.
Capacitors may otherwise be charged with life
endangering voltage, and coming into contact with them
may result in severe injury.
The interface module may be damaged by electrostatic
discharge. Observe guidelines for handling electrostatic
sensitive devices. Do not touch electrical contacts or PCB
components.
➀ Disconnect the KONSTANTER from the mains, and pull the
mains plug from the outlet. Wait for at least 5 minutes in
order to assure that the capacitors have been adequately
discharged.
➁ Loosen the 4 screws in the housing cover and lift the cover
off.
➂ Unscrew the cover plate at the left-hand side of the rear
housing panel. Save the screws for use in the next step..
Retrofit for Rack Mounting a Single Device
Use the accessory adapter: 19'' adapter 1 x 32 N.
This accessory components includes a 19'' abutment and a 19''
blanking plate.
➀ Loosen the 4 screws at the KONSTANTER front panel.
➁ Pull out the filler strips from the left and right sides.
Figure 2.1.2 aRack Installation of a Single KONSTANTER
➂ Replace the filler strips with the 19'' abutment on one side,
and with the 19'' blanking plate on the other side, and
secure them with the four screws.
➃ Unscrew the feet from the bottom of the housing. Remove
the rubber inserts from the feet to this end. The screws are
then exposed.
➄ Now install the KONSTANTER to the rack. Save all loose
parts for possible future use.
Figure 2.1.1 a Opening the KONSTANTER Housing
the rack. The guide rails, as well as the front panel mounting
screws used to secure the KONSTANTER, are rack-specific and
The KONSTANTER must be attached to guide rails at one side of
➃ The interface module:
must be procured from your rack supplier.
With the ribbon cable facing forward, push the interface module
through the housing rear panel and into the device interior.
Securely screw the sheet metal cover on the interface
module to the housing rear panel from the outside. Use the
two screws referred to in step➂.
➄ Insert the plug connector from the interface module into the
connector strip on the primary printed circuit board. Do not
twist the ribbon cable.
Retrofit for Rack Mounting Two KONSTANTERs
Use the accessory adapter: 19'' adapter 2 x 32 N.
This accessory component includes two 19'' abutments and one
19'' connector plate.
➀ Loosen the 8 screws at the KONSTANTER front panels.
➁ Pull out the filler strips from the left and right sides of each
device.
Figure 2.1.2 bRack Installation for Two KONSTANTERs
➂ Replace the filler strips with the 19'' abutments at the far left
and far right, and with the 19'' connector plate in the middle,
and secure them with the eight screws.
Figure 2.1.1 bInstalling the Interface Module
➅ Return the housing cover and secure it with the four screws.
The KONSTANTER can now be connected as usual.
12GMC-I Messtechnik GmbH
Screw both housings to the through-holes and threads in
the plug protector at the rear.
➃ Unscrew the feet from the bottom of the housing. Remove
Attention!
!
Warning!
the rubber inserts from the feet to this end. The screws are
then exposed.
➄ If you would like to electrically connect the two
KONSTANTERs, use the “mains jumper cable” and “RS 232
bus cable” accessories.
➅ Install the two devices to the rack. Save all loose parts for
possible future use.
The KONSTANTERs must be attached to guide rails at both sides
of the rack. The guide rails, as well as the front panel mounting
screws used to secure the KONSTANTERs, are rack-specific and
must be procured from your rack supplier.
2.1.3 Combining Benchtop Devices
Up to 3 benchtop KONSTANTERs can be stacked in combination
(see also chapter 5 for electrical connection options via the analog
interface).
➀ Unscrew the feet from the bottom of the housing. Remove
the rubber inserts from the feet to this end. The collar screws
are then exposed.
Four oblong slots now become visible at the bottom of the
housing.
➁ Turn the 4 collar screws from the feet into the 4 threaded
holes on the top of the other device housing. Save the 4
retaining washers and feet for possible future use.
➂ Set the KONSTANTER without feet onto the top of the other
KONSTANTER. The screws from the bottom KONSTANTER
must protrude through the oblong slots in the base of the
other KONSTANTER. Carefully push the top KONSTANTER
back until the screws snap into place.
➃ Screw the two KONSTANTERs together via the through-
holes and threads in the plug protector at the back. The top
KONSTANTER is thus secured against shifting.
➄ If you would like to electrically connect the two
KONSTANTERs, use the “mains jumper cable” and “RS 232
bus cable” accessories.
2.1.4 Connection to the Mains
Observe WARNING I!
Before switching the KONSTANTER on, is must be
assured that available mains power complies with the
supply power values specified at the mains connection
on the back of the device.
The KONSTANTER requires 230 V ~ supply power. Connect the
mains inlet plug at the rear panel to a mains outlet with earthing
contact using the included power cable.
Rated power consumption is specified on the serial plate at the
bottom of the KONSTANTER.
A mains outlet is included above the mains inlet plug for looping
mains power through to an additional KONSTANTER.
This mains outlet can neither be switched on and off, nor is it protected
with a fuse.
If mains power is looped through, make sure that overall
current consumption does not exceed 10 A at the
incoming supply lines!
Suitable “mains jumper cables” are available as
accessories (see order information on last page).
2.1.5 Connecting Power Consumers
The output leads are connected either at the front panel with
4 mm safety plugs to the “” and “” safety jacks, or at the rear
panel to the “” and “” outputs at the 6-pole screw terminal.
If loads are connected to both the front and the rear panel (not
allowed in the event of parallel connection due to danger of
overload!), constant voltage regulation applies to the terminals at
the rear panel. There are two terminals each for “” and “” for
load connection at the rear panel.
In the case of load current of greater than 10 A, these terminals must
be parallel connected due to the specified contact rating.
Make sure that the utilized cables have an adequate cross-section,
and that polarity is not reversed. It is advisable to twist the output
leads and to identify polarity at both ends.
The yellow-green safety jack at the front panel is connected to
PE, and can be used to connect ground cables or cable shield, or
can be used as an earth connection point for one of the output
poles.
2.1.6 Connection to Computer Interfaces
If the KONSTANTER is used within computer controlled systems,
one of the two connections described below must be established
via the appropriate interface.
Comments
The KONSTANTER cannot be remote controlled via both
interfaces simultaneously. The interface which first initiates
action after mains power has been switched on is activated,
and the other remains inactive.
In order to assure that existing bus activity is not interfered
with, all affected KONSTANTERs should be switched off while
establishing the bus connection.
Both interfaces are equipped with a common ground (GND),
and are electrically isolated from the output in accordance with
specified electrical safety regulations.
Interface configuration is described in chapter 4.7.4.
RS 232C Interface
Most controllers include two serial ports, which are commonly
designated COM1 and COM2, and which are equipped with 25 or
9-pin subminiature plug connectors.
A suitable cable with a length of approximately 2 meters is
available as an accessory for connecting the KONSTANTER and
the controller. Cables with other lengths are commercially
available. Appropriate adapters are available as well, in the event
that your controller is equipped with a 25-pin plug connector.
If you would like to fabricate the connector cable yourself, you’ll
need a 3-conductor shielded cable in order to establish
connection as shown in Figure 2.1.6 a.
GMC-I Messtechnik GmbH13
Figure 2.1.6 aPin Assignments for 9-Pin Plug and
TxD_IN 2
3
5
RxD_IN
GND_IN
2 RxD
3 TxD
5 GND
SSP
9-pin Plug9-pin Socket Connector
RxD_OUT 2
3
5
TxD_OUT
GND_OUT
2 TxD_IN
3 RxD_IN
5 GND_IN
9-pin Plug9-pin Socket Connector
SSP
OUT
IN
RS 232
RS 232
PC/Controller
SSP
IN
RS 232
COM1/COM2
IN OUT
IN OUT
IN OUT
IN OUT
Accessory:
RS 232 Bus Cable, 0.4 meters
Accessory: RS 232 Bus Cable, 2 meters (for example)
Socket Connectors
The serial interface furnished with this KONSTANTER series is
addressable. Up to 30 KONSTANTERs can be addressed via this
interface.
Data Queries
If a device (slave) is requested to return data to the controller, it
may not be addressed in the interim. In other words, the controller
must wait until all requested data have been fully received. No
data may be transmitted to any other devices during this period of
time.
In the case of multiple device serial operation, good timing must be
assured. Data collision is thus avoided, which may result in
interruption of data transfer and deletion of output buffer
contents.
If the slave does not respond within a specified period of time
(TIMEOUT), the controller can try to synchronize the device with a
universal command, or execute a reset and request the required
data once again.
IEEE 488 Interface
Up to 15 IEC bus controlled devices (including the controller)
can be interconnected to create a system.
These devices are connected to the bus with suitable,
commercially available cables with 24-pin plug connectors.
If your IEC bus system is equipped with the previously
common 25-pin subminiature plug connectors, you’ll need a
suitable adapter cable.
In order to assure reliable data transfer, the cable length
between any two devices should not exceed 2 meters, and
overall length should not exceed 15 meters.
Double shielded connector cable is recommended if the
KONSTANTERs are operated in proximity to strong sources of
interference or their power cables.
2.2 Switching the Instrument On
After the described preparations have been completed, the
device can be switched on.
Press the mains switch [1] at the front panel until it snaps into
place in order to turn the device on.
Power-Up Test
The microprocessor included in the device then starts a power-up
Figure 2.1.6 b Interconnection via the Serial Interface
If RS 232 communication is to be expanded to several
KONSTANTERs, additional arrangements must be made
regarding data protocols.
Addressing
A separate address is assigned to each interconnected
KONSTANTER. If the KONSTANTER receives its own address, it
is switched to the “addressed” status after which it accepts all
subsequent commands and data, until it is unlisted upon receipt
of a different device address.
In compliance with IEEE 488, an address range of 0 to 31 is used,
although address 31 is reserved as a general unlist command
(UNL). The serial interface’s device address is thus identical to the
address of the optional IEEE 488 interface (if included).
Universal Commands
In addition to the command
Addr xxDevice address,
other general commands can be used which are accepted
without previous addressing.
*TRG Trigger command, synchronization
DCLDevice clear
IFCInterface clear
Explanations of these commands and their syntax are included in
chapter 6.
14GMC-I Messtechnik GmbH
test. The following operations are performed during the test
routine (duration approximately 8 seconds):
– Reset all functional units (except battery-backed setup
memory)
– ROM test
–RAM test
– Initialize computer interfaces if installed
– Ascertain device type
– Check the A-D converter timer
– Recall last settings if required
The “LOCKED/SEQ” LED [12] blinks while this routine is running,
and all other LEDs and all digital display segments light up
(display test). If the device has been equipped with the optional
“IEEE 488 computer interface”, the selected IEC bus device
address then appears briefly at the display (e.g. “Addr 12”).
After successful completion of the self-test, the READY lamp is
continuously illuminated and the display is switched to measured
value indication for voltage (Uout) and current (Iout).
If this status is not achieved despite correctly selected device
address (0 to 30), even after repeatedly switching the device on
and off with abbreviated self-test, the device is probably
defective. If this is the case, contact your local representative.
Abbreviated Power-Up Test
In order to shorten power-up time, or if problems occur with the
normal power-up test, an abbreviated power-up test can be
used:
With the device switched off, press and hold the <ENTER>
key.
Turn the mains switch on.
Release the <ENTER> key after approximately 1 second.
If this procedure is used, only essential initialization steps are run
during power-up.
After initial power-up, the device has the following basic
configuration:
Interface functionsStandard “pon” status
Device functions
– Output statusInactive
– Voltage setpoint0 V
– Current setpoint0 A
– Voltage setting limitNominal output voltage
– Current setting limit Nominal output current
– OVP trigger value25 V (for 20 V models)
50 V (for 40 V models)
100 V (for 80 V models)
– Current limiting modeLimiting without shutdown
– Shutdown delay0 ms
– TRIGGER input Inactive
– Min-Max measured value memoryOff
– Power ON mode Reset configuration
– Manual operationEnabled
– Memory contents Deleted
The desired settings can be selected starting with this basic
configuration.
After a warm-up period of approximately 30 minutes, the
instrument operates at maximum accuracy.
When the device is powered up again at a later point in time, active
device configuration depends upon the last setting selected for
the POWER_ON function (—> page 69):
– Default settings or
– Last used device settings or
– Last used device settings with inactive output
Power-Up with RESET
In order to assure that the connected power consumer is not
endangered by any previous device settings, the device can be
initialized with the “POWER_ON RST” function by pressing and
holding the <CE/LOCAL> key during the power-up routine.
In order to switch the device off, activate the mains switch one again.
The device is then disconnected from mains power and the
output is deactivated. The last device configuration, as well as any
settings which have been saved to battery-back configuration
memory, are retained.
Caution!
Avoid switching the device on and off in a rapid, repeated fashion. This
temporarily impairs the effectiveness of the inrush current limiting
function, and may result in a blown fuse.
GMC-I Messtechnik GmbH15
3Controls, Display Elements and Terminals
12356 89 12
171615141311107418
Note: Numbers in brackets make reference to the figures included below.
[1]Mains Switch <POWER>
For turning the KONSTANTER on and off
After switching the KONSTANTER on, a self-test is performed
with a duration of approximately 8 seconds. After successful
completion of the self-test, the KONSTANTER briefly displays
its interface address and the version number of the installed
firmware, one after the other. The POWER-ON function then
configures the device with predetermined settings and the
device is ready for use.
When the KONSTANTER is switched off, it is disconnected from
mains power and the output is immediately deactivated. The last
device configuration, as well as any settings which have been
saved to battery-backed setup memory, are retained.
[2]Output On-Off Key <OUTPUT>
The power output can be activated and deactivated by
pressing the <OUTPUT> key. The red LED above the <OUTPUT>
key is illuminated as long as the output is active.
No significant output voltage overshooting occurs when the
output is activated and deactivated.
Activation sequence:
If the power output is activated, current and voltage are initially
set to “0” for transition from the “highly resistive” state.
Setpoints do not become active until after this transition has
been completed.
Deactivation sequence:
Current is set to 0 A and voltage to 0 V for approximately
350 ms. A sink is thus activated which discharges the output
capacitors to the greatest possible extent. The sink is then
deactivated and the output becomes highly resistive. However,
the output terminals are not electrically isolated.
[3]Control Mode, Protective Function and Status Displays
No LEDs light up: Output has not been activated
LEDs lights up:
Indicates the output’s operating status (control mode) if it has
been activated:
Green CV LEDConstant voltage regulation (Uout = Uset)
Green CC LED Constant current regulation (Iout = Iset)
Blinking LED:
Indicates the reason for automatic deactivation of the output
Green CV LEDOVP (overvoltage protection)
Overvoltage protection has been triggered,
because output voltage has exceeded the
selected trigger value OVP / OVSET.
Possible causes:
☞ Voltage setpoint USET has been set too high manually,
memory recall, programming error or Uset control signal to
the analog interface
☞ Voltage transients caused, for example, by switching
inductive power consumers (perhaps too little difference
between selected USET and OVSET values)
☞ Unipolar power recovery from the connected consumer
(e.g. DC motor)
16GMC-I Messtechnik GmbH
☞ During auto-sensing: Sensing lead polarity is reversed, or an
output lead is/was interrupted or was not taken into consideration when adjusting OVSET, so that the voltage at
the output terminals which is relevant for the OVP function is
increased by the amount to be compensated for at both leads,
and is higher than USET voltage as controlled by the
sensors at the load side (too little difference between
selected USET and OVSET values).
☞ Parallel connected voltage sources
Green CC LED OCP (overcurrent protection)
Overcurrent protection is active.
If the OCP ON function has been activated,
the output was operated for a duration of
t > DELAY in the current limiting mode
(current regulation). The output has been
deactivated.
Possible causes:
☞ Current setpoint ISET has been set too low manually, memory
recall, programming error or Iset control signal to the
analog interface
☞ Current transients caused by switching inductive
consumers (DELAY time may be set too low)
Yellow Pmax LED OTP (overtemperature protection)
Electronic power limiting has been
(overload! Pout > Pnom).
Possible causes:
triggered
☞ Impaired cooling, e.g. air inlet or exhaust vents are
obstructed.
☞ Excessive ambient temperature
The KONSTANTER is capable of continuously supplying
nominal power at ambient temperatures of up to 50 °C
(measured at the air inlet vents). Approximately 120 to
130% nominal power can be drawn intermittently
(triggering point for electronic power limiting). Continuous
operation at these levels may cause triggering of the
overtemperature protection function.
☞ The fan has failed.
☞ A device error or defect has occurred.
After the cause of triggering has been eliminated, the output can be
reactivated.
[4]Output
Selected constant voltage or constant current is available from
the safety jacks at the front panel.
blueNegative output pole
redPositive output pole
yellow-greenEarth connection point for the output or
shield terminal if shielded output leads are
used. The ground terminal is connected to
the housing and the earthing contact at the
mains connection.
The power consumer can also be connected to the OUTPUT
interface [22] at the rear panel.
[5]Left-Hand Display with [7]
[6]Display Parameter Indicators (LEDs)
As a default setting, the measured output voltage value Uout
appears in volts at the left-hand display.
The display can be switched to the momentary measured voltage value Uset by slightly turning the voltage adjusting knob [5], or by
pressing the <ENTER> key.
The LEDs allocated to the display indicate which value is being
displayed (green LED = measured value, yellow LED = setting
value).
Uout / V (gr) = Measured output voltage value in V
Uset / V (ye) = Output voltage setpoint in V (blinking
indicates that the displayed value has not
yet been set)
If no adjustment is made for a period of 10 seconds during the
display of a setting value, the display returns to Uout
automatically.
Additional KONSTANTER functions can be selected with the
<SELECT>, <FUNCTION>, <RCL> and <SAVE> keys. The function
code or the parameter name appears at the left-hand display
in this case.
[7]Rotary Knob for Voltage Adjustment <Uset>
Voltage is adjusted in the usual fashion with the rotary knob.
However, the adjusting element is not a potentiometer, but
rather a rotary pulse encoder which generates 24 pulses per
revolution, whose setting resolution (step size per pulse) can
be set to either coarse, medium or fine with the <SELECT>
function. This allows for convenient, precise adjustment on the
one hand, and also assures that no change occurs to the
selected value when switching between remote control and
manual operation.
When the voltage adjusting knob is turned, the left display is
first switched to the Uset display and the current voltage
setpoint appears. After approximately 0.4 seconds, one of the
decimal places starts blinking at the display in order to indicate
selected adjusting sensitivity. From this point on, turning the
rotary knob changes the display value, and thus the setpoint
value, at the selected decimal place.
Clockwise rotation increases the value, and counterclockwise
rotation decreases the value.
If no adjustment is made for a period of 10 seconds, the
display automatically returns to the measured voltage value
Uout. The display can be immediately switched to the Uout
value by pressing the <CE/LOCAL> key.
☞ Detailed explanations are included in chapter 4.6.
[8]Right-Hand Display with [10]
[9]Display Parameter Indicators (LEDs)
As a default setting, the measured output current value Iout
appears in amperes at the right-hand display.
The display can be switched to the momentary current setpoint Iset by slightly turning the current adjusting knob [8], or by
pressing the <ENTER> key twice.
The LEDs allocated to the display indicate which values are
being displayed (green LED = measured value, yellow LED =
setting value).
Iout/A (gr) =Measured current value in A
Iset / A (ye) =Current setpoint in A (blinking indicates that
the displayed value has not yet been set)
If no adjustment is made for a period of 10 seconds during the
display of a setting value, the display returns to Iout
automatically.
GMC-I Messtechnik GmbH17
Additional KONSTANTER functions can be selected with the
Attention!
!
<SELECT>, <FUNCTION>, <RCL> and <SAVE> keys. The
respective measured value or setting parameter appears at the
right-hand display in this case.
[10] Rotary Knob for Adjusting Current <Iset/A>
The same applies to this rotary knob, with reference to output
current, as is also the case with the voltage adjusting knob [5].
[11] Parameter Adjusting Keys and
The <> (increment) and <> (decrement) keys are used to
select or adjust all KONSTANTER functions and their
parameters.
☞ Detailed explanations are included in chapters 4.6, 4.7
The respectively illuminated LED indicates the current
operating state of the computer interface:
REMOTE on:The KONSTANTER is being remote
controlled, front panel controls are
disabled.
ADDR/DATA on:The KONSTANTER has been addressed
and is receiving or transmitting data
(applies to IEC bus operation only).
SRQ/STS on: The KONSTANTER is transmitting a service
request.
LOCKED/SEQ on:Indicates that manual controls are disabled:
Front panel controls are disabled and
protected against unauthorized or
inadvertent change.
This display only applies to disabling of the
front panel controls by means of manual
operation or a control signal applied to the
TRIGGER input (for T_MODE LLO). It does not
indicate disabling of manual switching to
local control by means of the IEC bus LOCAL LOCKOUT command.
blinking:Sequence mode display in
disabled and enabled state:
– Slow blinking: sequence in hold status
– Fast blinking: sequence in run status
Sequence mode display
LLO is displayed when any key or rotary
knob is activated (
and disabling:
l
ocal lockout).
[13] Function Selection Key <SELECT>
Selection of displayable measured values
Resolution of display values and setting values
☞ Detailed explanations are included in chapters 4.6, 4.7
and 4.8.
[14] Recall Key <RCL>
Recall of stored KONSTANTER settings, value pairs (Uset and
Iset) and any associated dwell time setting for possible
readjustment.
☞ Detailed explanations are included in chapter 4.10.
[15] <ENTER> Key
The <ENTER> key is used for several functions:
Execute selected functions
Acknowledge selected text parameters
Go to the next lowest function menu level
Switch the display from Uout / Iout to Uset / Iset with active
cursor
Setpoint changes are acknowledged with the <ENTER> key
in this mode.
Move cursor back and forth between Uset and Iset in the
Uset / Iset display
Activate the recalled memory location
☞ Detailed explanations are included in chapters 4.6, 4.7
and 4.12.
[16] <CE/LOCAL> Key
This key has several functions:
Abort an operation
Switching from remote to local control
Disable front panel controls
In combination with special functions
☞ Detailed explanations are included in chapter 4.13.
[17] Save Key <SAVE>
Saves device settings, value pairs (Uset and Iset) and any
associated dwell time. In the manual mode, this key also
fulfills the function of the STORE computer command.
Disabling of front panel controls in order to prevent
inadvertent or unauthorized changes to settings is only
possible in combination with the <CE/LOCAL> key.
☞ Detailed explanations are included in chapter 4.9.
[18] Function Selection Key <FUNCTION>
No special key is assigned to device functions which are
normally seldom adjusted or used. These functions are
controlled in a menu-driven fashion.
☞ A description of the menu and the procedure for setting all
functions and parameters is included in chapter 4.7.
[19] RS 232 Interface
☞ Detailed explanations regarding remote control of
KONSTANTER functions via the RS 232 serial interface are
included in chapter 2.1.3.
☞ Interface-specific commands are included in chapter 6.4.
[20] IEEE 488 Bus Connection
Connection for remote control of KONSTANTER functions:
☞ Detailed explanations are included in chapter 2.1.1.
☞ Interface-specific commands are included in chapter 6.4.
The electrical contacts of all interfaces are connected to
components which may be damaged by electrostatic
discharge. Neutralize potential differences between
yourself and the device by grasping the housing before
touching these contacts!
18GMC-I Messtechnik GmbH
[21] Analog Interface
Attention!
!
Attention!
!
Attention!
!
Warning!
GOSSEN METRAWATT
IEC 625/IEEE 488 INTERFACE F. SSP-KONSTANTER BEST.NR. K380A
+
-
+
+
-
+
-
+
SENSE
-
+
SENSE
I-MON
U-MON
Iset
Uset
Uset
AGND
15V
TRG IN
TRG IN
SIG2 OUT
SIG1 OUT
ANALOG INTERFACEOUTPUT
OUT
IN
RS 232
230V 50-60Hz
FUSE T4/250V
1920
2122
232425
The analog interface facilitates the following functions:
Remote adjustment of output voltage and current with
analog control voltages ranging from 0 to 5 V, or 5 to 0 V
☞ chapter 5.8 / chapter 5.4
External measurement or recording of output voltage and
current based on monitor signals, 0 to 10 V
☞ chapter 5.4 / chapter 5.5
Supply external controllers with +15 V auxiliary power
Linking of several devices for master-slave operation
☞ chapter 5.9 / chapter 5.10
Varying the internal output resistance value
☞ chapter 5.8
Control of a selected device function via the floating trigger
input
☞ chapter 5.7
The electrical contacts of this interface are connected to
components which may be damaged by electrostatic
discharge. Ground yourself by grasping the housing
before touching these contacts!
All of the control cables connected to the analog
interface should be shielded. Ground the shield directly
to the device using the shortest possible connection
lead. One of the threaded holes for the plug protector at
the rear panel can be used for this connection. Use a
suitable screw and washer to secure the connection and
assure good contact.
[22] OUTPUT Interface
The output interface offers two options:
Take constant voltage or constant current from the terminal
strip at the rear panel of the KONSTANTER.
Connect sensing leads for compensation of voltage drops in
the output leads
☞ chapter 5.3
[23] Exhaust Vents
The exhaust vents are required for regulating the temperature
inside the device. Warm air generated during operation of the
device is discharged via these vents with the help of a
temperature controlled fan.
The exhaust vents may not be closed or obstructed,
because heat accumulation inside the device may result
in malfunctioning, failure or damage to the
KONSTANTER.
[24] Mains Input
Mains input with looped through mains outlet for inlet plug. The
looped through mains outlets make it possible to directly
connect up to three devices using two short power cables with
inlet plugs. In this way, only one mains power cable is required
to operate all three devices.
[25] Mains Fuse
Fusing for the 230 V mains power input
All devices:T 4.0 A / 250 V (6.3 x 32 mm)
Only fuses of the type and nominal current rating
specified here may be used when replacing blown fuses.
Tampering with fuses or fuse holders is absolutely
prohibited (“repairing” fuses, short-circuiting fuse holders
etc.).
GMC-I Messtechnik GmbH19
4Manual Operation and Device Functions
Uset / V
Uout
t
(1)
(2)
(3)
4.1 Menu Structure
After the power-up sequence has been completed, the device is
switched to the basic operating mode by means of which the
device’s basic functions can be executed, such as:
• Select desired output voltage Uset
• Select desired output current Iset
• Adjust allowable working range with soft-limits Ulim and Ilim
• Adjust overvoltage protection and overcurrent shutdown
Additional setup menus can be accessed with the function key.
These include:
• SEt (setup) extended Setup functions
•AnIF (analog interface) Analog interface settings
• SEq (sequence function) Sequence function settings
After selecting the desired setup menu with the function key, the
respective menu level appears at the display.
The function and arrow keys are used to scroll through the
respective menu levels and select the desired settings.
The <CE/LOCAL> key can be used to shift back up one level at a
time, until the basic operating menu once again appears.
KONSTANTER
Initializing the Procedure
☞ Slightly turn (1) the Uset knob [7].
!The display is switched from Uout (measured voltage value)
to Uset (voltage setpoint). The decimal place for the selected
setting resolution blinks.
!The green Uout/V LED [6] goes out and the yellow Uset/V LED
[6] lights up. Continuous illumination of the Uset/V LED
indicates that values are activated immediately during
adjustment.
Selecting Resolution
!3 step widths are possible: 0.01 V, 0.1 V or 1 V.
!The blinking decimal place [5] indicates which step width will
be used for setpoint adjustment.
☞ Repeatedly press the <SELECT> key [13] until the desired
decimal place blinks at the display.
Executing the Procedure
Values become immediately active during adjustment.
☞ Adjustment (2) with the Uset rotary knob [7]:
Clockwise rotation Increases the value
Counterclockwise rotation Decreases the value
All intermediate values are run through semi-linearly and are
read out to the output (assuming it is active).
4.2 Setting Output Voltage and Output Current
There are two ways to set output voltage and output current:
• Direct selection
Changes become immediately effective when this method is utilized,
assuming the output is active.
• Pre-selected settings
Output voltage or current is preset with the scroll keys to the
desired value. The selected value is activated at the output
after acknowledging with the <ENTER> key.
4.2.1 Direct Selection (rotary knobs and scroll keys)
The operating concept allows for direct selection of output
voltage and/or output current with the rotary knobs, with
immediate activation of the new values at the output.
In the default configuration, current output values appear at the
display with voltage at the left and current at the right-hand side.
This is indicated by means of the two LEDs above the display. If
the output is active (indicated by illumination of the red LED above
the <OUTPUT> key), the LEDs in the diagram indicate the control
state. Depending upon the selected output quantities and the
load situation, either output voltage or output current is regulated.
CV (constant voltage) indicates voltage regulation, and CC
(constant current) indicates current regulation.
If the working point is not within the allowable control range, the
power LED lights up.
The display can be switched to a representation of the
corresponding setpoint by slightly turning one of the rotary knobs
(Uset or Iset). This change is indicated by the respective LED
above the display. The blinking decimal place indicates the
resolution with which adjustment will take place. Resolution can
be adjusted with the <SELECT> key.
The setting can then be changed to the desired value with the
rotary knob.
As soon as the selected setpoints have been activated, they can
also be adjusted with the scroll keys, in which case resolution can
also be pre-selected.
Figure 4.2.1 aContinuous Adjustment of Uset
☞ Adjust (3) with the <> key or the <> key [11]
<> (increment) Increases the value
<> (decrement) Decreases the value
!Each time the key is pressed, output voltage is changed
by an amount which corresponds to the value selected
with the resolution setting function.
!Pressing and holding the respective key results in rapid
scrolling, regardless of the step width.
Attention! Uset may not be set to a value which exceeds Ulim!
Uset – Output Voltage
• Adjustment with the Uset rotary knob [7] is not enabled until 0.4
seconds after the digital display [5] has been switched to the
respective function. This delay time prevents inadvertent changes to Uset during selection of the Uset display. The rotary
knob must be adjusted after this time period has elapsed in
order to change the Uset setting.
• After delay time has elapsed, output voltage is adjusted directly
as the knob is rotated, if the output is active. The currently
active value always appears at the display [5].
• If no adjustment is made for a period of 10 seconds, the
display automatically returns to the measured voltage value
Uout. The display can be immediately switched to the Uout
value by pressing the <CE/LOCAL> key [16].
• There are two ways to adjust Uset manually:
– Adjustment with immediate activation of the new value:
Adjustment of the setpoint has an immediate effect on
load output quantities.
– Setpoint adjustment:
Adjustment of the setpoint does not effect load output
quantities until after enabling.
20GMC-I Messtechnik GmbH
Iset – Output Current
Iset / A
Iout
t
(1)
(2)
(3)
Uset / V
ENTERENTER
ENTER
Uout Iout
Select Uset
Uout Unchanged!
CE/LOCAL
Uout Iout
Iout Unchanged!
Activate
Select Iset
Uset
The procedure for selecting output current Iset is identical to the
procedure for selecting output voltage Uset (4.1.2).
However, the following controls and displays must be substituted:
• Iset rotary knob (chapter 3 [10]
• Right-hand display (chapter 3 [8])
• Yellow Iset/A LED (chapter 3 [9])
• Setting resolution:
4 step widths are possible: 0.001 A / 0.01 A / 0.1 A / 1 A
Attention: Iset may not be set to a value which exceeds Ilim!
!Each time the key is pressed, output voltage is changed
by an amount which corresponds to the value selected
with the resolution setting function.
!Pressing and holding the respective key results in rapid
scrolling, regardless of the step width.
☞ Acknowledge the selected value with the <ENTER> key
[15]. The new value is now activated at the output [4], and
the yellow Uset/V LED [6] is continuously illuminated.
☞ Repeatedly press the <ENTER> key [15] in order to switch
back and forth between Uset and Iset.
4.2.2 Pre-selected Setting (ENTER, scroll keys)
Figure 4.2.2 a Adjusting Uset with a Specified Fixed Value
If the application requires that switching to a new setpoint takes
place in a single jump by pressing a key (without semi-continuous
adjustment), this can be accomplished as follows.
Proceed to the basic setting menu with the <CE/LOCAL> key, i.e.
to the display of momentary output values. Press the <ENTER>
key in order to switch to the pre-selection setting mode. After
switching to this mode, the currently valid setpoint values appear
at both displays (indicated by the LEDs above the display).
As a default function, voltage adjustment is always activated first,
which is indicated by a blinking decimal place at the
corresponding display. Repeatedly press the <ENTER> key in
order to switch back and forth between current adjustment and
voltage adjustment.
The new setpoint can be selected with the scroll keys (and only
with the scroll keys), after selecting voltage or current setpoint
adjustment and the desired resolution. The SET LED blinks in
order to indicate that a change has been made but not yet
activated. The new setpoint is activated after acknowledging with
the <ENTER> key.
Pre-Selecting a Setpoint
☞
Press the <ENTER> key [15].
!The display [5] is switched from Uout (measured voltage value)
to Uset (voltage setpoint). The decimal place which corresponds
to the selected resolution setting blinks.
!The green Uout/V LED goes out, and the yellow Uset/V LED [6]
lights up (as long as the Uset/V LEDis continuously illuminated,
direct adjustment with immediate activation of the new value
can be executed with the rotary knob [7]).
Setting resolution:
!3 step widths are possible: 0.01 V, 0.1 V or 1 V.☞ Repeatedly press the <SELECT> key [13] until the desired
decimal place (step width) blinks at the display.
Output Voltage:
☞ Press either the <> key or the <> key [11]!
<> (increment) Increases the value
<> (decrement) Decreases the value
!The yellow Uset/V LED[5] starts blinking.
!A blinking Uset/V LED indicates that new values are displayed
but not activated during adjustment. The old Uset value
remains active.
☞ Press the <> key or the <> key [11] until the desired
value appears at the display [5].
GMC-I Messtechnik GmbH21
• Output voltage jumps from the old value to the new value in a
single step (no intermediate values).
Attention: Uset may not be set to a value which exceeds Ulim!
4.3Switching the Power Output On and Off
The <OUTPUT> key (chapter 3 [2]) functions independently of the
current operating state of the KONSTANTER.
Details regarding control mode, protective function and status
displays which provide information concerning the output’s
operating mode are included in chapter 3 [3].
Closed Loop Control Mode
The power output can be activated or deactivated by pressing the
<OUTPUT>key [2].
OUTPUT OFF OUTPUT ON
REMOTE status (remote control)
(chapter 3[12], chapter 4.13)
The <OUTPUT> key [2] is disabled and has no function.
LOCAL LOCKED status (front panel controls disabled)
(chapter 3[12], chapter 4.13)
The <OUTPUT> key [2] is disabled and has no function.
trG out (T_MODE OUT) and External Trigger Active
The output can be activated an deactivated with the trigger
signal (trG parameter OUT selected).
If the power output has been disabled by applying an external trigger signal to the analog interface, it cannot be switched on,
neither by means of a command nor with the keys. The
corresponding command is not executed and bit 4 in event
register B is set (output-on error). “Err 25” also appears briefly
at the display as a warning in the event of manual operation.
The status of the power output
by the red LED above the <OUTPUT> key [2]:
LED on = output activated
LED off = output deactivated
In the case of OUTPUT OFF, the control mode displays are also
switched off
(chapter 3[3]).
If the output has be deactivated as a result of overtemperature protection, the yellow “240 W” LED[3] in the control mode
display blinks. The output cannot be activated until the
KONSTANTER has returned to its normal operating
temperature.
(chapter 3[4/22]) is indicated
If the Pon RCL function is active, the output is automatically
reactivated after the device has returned to its normal
operating temperature.
Functions which may influence the status of the output
include:
• Manual and computer-aided settings for Iset may not exceed
Ilim.
• Ilim cannot be set to a value which is less than current
setpoint Iset. Iset must be reduced far enough to allow for the
new Ilim setting.
FunctionsMeaningManual
Operation
OVP (OVSEt)Overvoltage
protection
OCPOvercurrent
Pon
(POWER_ON)
SEqSequencechapter 4.7.4page 70
trG (T_MODE)page 28page 74
4.4
Limiting the Allowable Working Range: Ulim, Ilim
Allowable setting limits for voltage and current can be restricted in
order to assure ideal matching to the working ranges of the
connected power consumer. The Ulim and Ilim setting functions
are provided to this end. Setting options can be selected with the
respective <SELECT> key. Setting resolution is selected with the
<RESOL> key. The setting itself is entered with the scroll keys.
New settings become immediately active.
Attention!
These settings represent so-called soft limits. This means that
values can be selected both manually and via the computer
interface which lie within these limit values, and that a
corresponding error message is otherwise generated.
Attention!
The actual output quantity is the sum of the digitally selected
setpoint value and the setpoint value specified via the analog
interface. This makes it possible to select values which exceed
the specified soft limit.
protection
page 25page 69
page 26page 68
page 26page 69
Remote
Operation
Settings
See description on page 25.
4.5 Description of OVP and OCP Protection Functions
Protection for the connected power consumer and the
KONSTANTER by means of the following functions:
OVP – overvoltage protection
Functions
• Protection for the connected power consumer
• If voltage at the output terminals exceeds the selected OVP
value, the power output is deactivated.
• Triggering of overvoltage protection causes immediate (< 200
s) deactivation of the output (OUTPUT OFF). The HF power
transmitter is disabled, and the electronic sink for discharging
the output capacitors for approximately 350 ms is activated.
In addition, bit 4 (OVPA) is set in event register A. Bit 4 remains
set in status register A for as long as the trigger value is
exceeded.
•The CV LED blinks as soon as overvoltage protection OVP is
triggered (see also chapter 3 [3]), and the red OUTPUT LED
goes out.
• As soon as the shutdown condition no longer exists, the
power output can be reactivated by pressing the <OUTPUT> key, by transmitting a trigger signal to the analog interface or by
means of computer control (“OUTPUTON”
lights up).
Settings
See description on page 25.
➅ red OUTPUT LED
Ulim – Setting the Upper Voltage Limit Value
Functions
• Upper setting limit (soft limit) for Uset
• Prevents inadvertent violation of the maximum voltage value
when adjusting Uset.
• Protection for the connected power consumer
• Ulim has higher priority than Uset.
• Manual and computer-aided settings for Uset may not exceed
Ulim.
• Ulim cannot be set to a value which is less than a previously
selected Uset value. Uset must be reduced far enough to allow
for the new Ulim setting.
Settings
See description on page 25.
Ilim – Setting the Upper Current Limit Value
Functions
• Upper setting limit (soft limit) for Iset
• Prevents inadvertent violation of the maximum current value
when adjusting Iset.
• Protection for the connected power consumer
• Ilim has higher priority than Iset.
OCP – Overcurrent Protection
Functions
• Protects the power consumer from continuous overcurrent.
• Deactivates the power output when load current Iset has been
reached, and the output is switched to the current regulating mode.
• Current can nevertheless be allowed to exceed Iset for
specified, short periods of time by specifying a delay time (see
below), for example:
!Starting current for electric motors
!In-rush current for capacitive power consumers
!For testing the breaking performance of circuit breakers,
motor protecting switches, fuses etc.
!For determining the short-term load capacity of contacts
and cables, as well as electrical and electronic
components
!In order to maintain short response times when
programming voltage increases
•The CC LED blinks as soon as overvoltage protection OCP is
triggered (see also chapter 3 [3]), and the red OUTPUT LED
goes out.
• The power output can be reactivated at any time by pressing
the <OUTPUT> key, by transmitting a trigger signal to the analog
interface or by means of computer control (OUTPUTONOUTPUT LED lights up).
➅ red
Settings
See description on page 26.
22GMC-I Messtechnik GmbH
dLY – Output Off Delay for OCP
Functions
• Delay time prior to deactivation of the power output after
current regulation has been triggered (Iout = Iset)
• Only enabled with activated OCP function (OCP ON)
• If output current Iout drops below Iset before DELAY time
elapses, the shutdown sequence is aborted.
• If current regulation is triggered again, the routine is started
over (at 00.00).
• The default setting after RESET (*RST) is 00.00.
Settings
See description on page 26.
4.7Operating Menu via the FUNCTION Key
The FUNCTION menu consists of the following functions for
configuring the KONSTANTER’s parameters:
Function groupFunctionParameter (numeric / text)
■ Setup
SEtUlimNP: xx.xx
IlimNP: xx.xx
OVPNP: xxx.x
OCPTP: oFF / on
dLYNP: xx.xx
PonTP: rSt / SbY / rcL
-
UI
rndTP: 0 / -1 / -2
TP: oFF / on / rSt
4.6 Display of Momentary Output Values Uout, Iout
and Pout
Uout – Momentary Measured Voltage Value
• Appears at the left-hand display [5].
•The green Uout/V LED [6] lights up and indicates display of the
momentary measured voltage value.
• The measured voltage value is automatically displayed again,
approximately 10 seconds after the last setting has been
made.
• The momentary measured voltage value is displayed
immediately if the user exits the device functions setting mode
by pressing the <CE/LOCAL> key [16].
Iout – Momentary Measured Current Value
• Appears at the right-hand display [8].
• The green Iout/A LED [9] lights up and indicates display of the
momentary measured current value.
• The measured current value is automatically displayed again,
approximately 10 seconds after the last setting has been
made.
• The momentary measured current value is displayed
immediately if the user exits the device functions setting mode
by pressing the <CE/LOCAL> key [16]
Pout – Display Momentary Output Power
Calculated internally based upon momentary measured values
Uout and Iout.
Settings
See description in chapter 4.8.1.
UI_– Display Measured Values in
Functions
• Reads out stored values for Umin, Umax, Imin and Imax at the
display, or via the computer interface.
• Stored Min-Max values can be read out regardless of the
status selected for the MINMAX function.
Settings
See description in chapter 4.8.1.
U/I Min-Max Memory
■ Analog Interface
AnIFtrGTP: oFF / out / rcL / SEq / LLO / UI
SiG1TP: oFF / on / out / Mode / SEq /
SSEt / U-Lo / U-Hi / I-Lo / I-Hi
SiG2TP: oFF / on / out / Mode / SEq /
SSEt / U-Lo / U-Hi / I-Lo / I-Hi
■ Sequence Function
SEq*tSEtNP: xx.xx
SSEtNP: oFF / on
tdEFNP: xx.xx
StrtNP: xxx (11... 252)
StoPNP: xxx (12... 253)
rEPNP: xxx (cont. or 1 ... 255)
SEq**TP: Go, Strt, StoP, hold, StEP, cont.
* Sequence function: appears at left-hand display, right-hand display is blank.
** Parameter selection for sequence control: appears at left-hand display, selected parameter
appears at right-hand display.
■ Interface Configuration
buSAddrNP: 0 / 1 / ... / 13 /... / 30 / (UNL)
bAUdNP: 50 / 75 / 150 / 200 / ... / 4800 /
9600 / 19200
dbitNP: 7 / 8
PbitTP: nonE / ZEro / EVEn / odd / onE
SbitNP: 1 / 2
■ Adjustment (calibration, chapter 7)
CALdAtETP: mm.yylast/new adjustment date
CALTP: Strt
MEASUoFF
UoFFNP: x.xxxentry of ext. measured value U
MEASUFS
UFSNP: xx.xxentry of ext. measured value U
MEASIoFF
IoFFNP: x.xxxentry of ext. measured value I
MEASIFS
IFSNP: xx.xxentry of ext. measured value I
CALEnd
NP: Numeric Parameters
• Can be selected as a specified value or as a value within an
interval with corresponding resolution.
•Adjust resolution (decimal place) by pressing the <SELECT> key.
• Select with the <> key or the <> key.
• Displayed value = selected value.
-
TP: Text Parameters
• Select parameter with the <> key or the <> key.
• Acknowledge the selected value by pressing the <ENTER> key.
GMC-I Messtechnik GmbH23
Settings
FUNCTION
FUNCTION
ENTER
FUNCTION
+
Jump to
last edited
function group
Jump to
last edited
function
e.g.
e.g.
e.g.
*
CE/LOCAL
FUNCTION
FUNCTION
FUNCTION
+
FUNCTION
+
FUNCTION
+
FUNCTION
+
☞
Press the <FUNCTION> key [14] at the front of the
KONSTANTER.
If no settings have yet been changed via the FUNCTION menu
after switching the device on, the SEt function group is
accessed upon pressing the <FUNCTION> key.
If settings have already been changed in the function menu,
the last edited function group or function is accessed
automatically.
• Accessing the last edited function:
☞ If you want to configure a function included in the current
function group, select the function group by pressing the
<FUNCTION> key. Detailed information is included in the
respective section of this chapter.
☞ If you want to configure a function from another function
group, first return to the function groups menu by pressing
the <CE/LOCAL> key.
• Forward scrolling through function groups in this menu:
☞ Repeatedly press the <FUNCTION> key or
☞ Press and hold the <FUNCTION> key and repeatedly press
the <> key at the same time.
• Reverse scrolling through function groups in this menu:
☞ Press and hold the <FUNCTION> key and repeatedly press
the <> key at the same time.
• Accessing functions in the selected function group:
☞ Press the <ENTER> key.
Figure 4.7 aAccessing the Functions Submenu
☞ If the function groups menu is currently open, you can
access the following function groups:
Figure 4.7 bSwitching Amongst Individual Function Groups
Note
The order in which the commands are explained in the following sections
corresponds to the actual sequence for manual operation of the
SSP KONSTANTER. This should make it easy to configure device functions
and parameters without leafing back and forth through the instructions.
☞ If the desired function group does not appear immediately
at the left-hand display, repeatedly press the <FUNCTION>
key until the name of the desired function group appears
at the left-hand display (SEt in this example).
24GMC-I Messtechnik GmbH
4.7.1 SET – “Setup” Function Group
FUNCTION
+
FUNCTION
+
FUNCTION
+
FUNCTIONPress on ce
ENTER
Press repeatedly if necessary
Jump to last
edited setup function
Cursor position
SELECT
Resolution of
numeric parameter
Select a value for
the upper voltage limit
(numeric parameter)
CE/LOCAL
CE/LOCAL
CE/LOCAL
CE/LOCAL
Uout
Iout
*
*
*
Ulim – Upper Voltage Setting Limit
Functions
See functions description on page 22.
Settings
The procedure is described in principle in chapter 4.7.
!
The Ulim display appears in the
window for the setup function,
along with a related numeric
parameter [V].
Ilim – Upper Current Setting Limit
Functions
See functions description on page 22.
Settings
The procedure is described in principle in chapter 4.7.
!The Ilim display appears in
the window for the setup
function, along with a
related numeric parameter
[A].
OVP – Overvoltage Protection Trigger Value
Functions
See functions description on page 22.
Settings
The procedure is described in principle in chapter 4.7.
!The OVP display appears in
the window for the setup
function, along with a
related numeric parameter
[V].
☞ The OVP trigger value should be set at least 1 V higher than the
desired USET output voltage in order to prevent undesired
triggering of the OVP function resulting from overshooting due
to sudden output discharging (minimum values: chapter 1.5.3,
“Electrical Data”)!
☞
The
OVP
trigger value makes reference to the prevailing voltage value
between the output terminals of the SSP. This voltage is increased
USET
by the
parameter during sensing mode operation (remote
sensing) by an amount equal to voltage drop at the output leads. For
OVP
and
USET
this reason, the above defined difference between
must be correspondingly increased during sensing mode operation.
☞ The selected parameter becomes immediately active!
Note
• Overvoltage protection response time is less than 200 μs.
Output voltage generated by the device may exceed OVSET for
the duration of this response time. Maximum overshooting
can be approximately calculated as follows:
Uout = ISET [A] x 200 [μs] / Cout [μF]
ISET = selected current setpoint
Cout = capacitance of the output capacitor
Subsequent discharging time for the output capacitor
Figure 4.7.1 a Path to Ulim Settings
GMC-I Messtechnik GmbH25
depends upon load, and corresponds to the specified values
for response time at Unom 1 V included in chapter 1.5.3.
• Possible causes for triggering overvoltage protection are listed
in chapter 3 [3], “Control Modes”.
OCP – Overcurrent Protection
*
Select the text parameter
from an options loop
From the SEt function group
ENTER
CE/LOCALFUNCTION
*
*
Select the text parameter
from an options loop
From the SEt function group
ENTER
CE/LOCALFUNCTION
*
Pon – Specifying a Power-On Status
Functions
See functions description on page 22.
Setting Parameter
☞
OFF: OCP function is inactive
Continuous current limiting (current regulation)
☞ ON: OCP function activated
The output is deactivated as soon as current limiting has been
active for the specified DELAY time.
☞ The selected text parameter does not become active until the
<ENTER> key is pressed!
Settings
The procedure is described in principle in chapter 4.7.
!The OCP display and a related
text parameter appear in the
setup function window.
Functions
• Use automatic KONSTANTER configuration after power on.
Setting Parameters:
☞
rSt: (RESET) Factory default settings are utilized
☞ SbY: (STANDBY) Same settings as prior to shutdown, power
output remains inactive (OUTPUT OFF)
Disabled panel controls are re-enabled after mains ON.
From firmware version 4.003 onwards:
Disabled panel controls remain blocked even after mains ON.
☞ rcl (RECALL) Same settings as prior to shutdown.
Disabled panel controls remain blocked even after mains ON.
Settings
The procedure is described in principle in chapter 4.7.
!The Pon display and a related text parameter appear in the
setup function window.
Figure 4.7.1 b Path to Selection of the OCP Text Parameter
Note
Load current for the output capacitor is also acquired by the
current regulator, and is limited to a value of Iset Ioad. If the OCP
function has been activated and ISET and DELAY have been set to
low values, the output may even be deactivated if output voltage
is increased. For this reason, DELAY must at first be set somewhat
higher than the resulting output voltage response time.
dLY – Output Off Delay for OCP
Functions
See functions description on page 23.
Settings
The procedure is described in principle in chapter 4.7.
!
The
DELAY
the window for the setup
function, along with a related
numeric parameter [t].
display appears in
Figure 4.7.1 cPath to Selection of the Pon Text Parameter
Note
• The status of the Pon function is not saved as a device setting
to SETUP memory with the <SAVE> key.
• The following settings are recommended, depending upon
how the KONSTANTERs is used:
! Pon rSt:Use in computer controlled systems
! Pon SbY:Common laboratory use
! Pon rcL:Applications which must continue in an
unchanged fashion after power failures.
26GMC-I Messtechnik GmbH
UI_– U/I Min-Max Measured Value Memory (MINMAX)
Select the text parameter
from an options loop
From the SEt function group
*
ENTER
CE/LOCALFUNCTION
Select the text parameter
from an options loop
From the SEt function group
*
ENTER
CE/LOCALFUNCTION
rnd – Rounding Off the Displayed Measured Value
Functions
• Saveminimum and maximum measured voltage and current
values.
• Reads out stored values for Umin, Umax, Imin and Imax at the
display, or via the computer interface.
• The MINMAX function can be temporarily set to OFF, for
example before deactivating the output when changing
devices under test. Stored values are then no longer updated,
nor are they automatically reset.
• Stored Min-Max values can be read out regardless of the
status selected for the MINMAX function.
• Stored Min-Max values can always be reset with the UI
_RST
command.
• Stored Min-Max values can also be reset with the *RST
command, or with the key combination <CE/LOCAL> + <ENTER> (= RESET).
• The status of the MINMAX function is also saved to SETUP
memory by pressing the SAVE key.
•If the Pon function is set to SbY or rcl, the MINMAX function is
activated upon power-up, but previously measured Min-Max
values are lost when the KONSTANTER is switched off.
•The MINMAX function can also be controlled via the trigger input
at the analog interface.
Setting Parameter
☞
OFF: Storage of Min-Max values is deactivated
☞ ON: Storage of Min-Max values is activated
☞ RST: contents in the Min-Max memory are reset or replaced
with the momentary measured value:
Umin = UoutUmax = Uout
Imin = IoutImax= Iout
Functions
• Only effective for Uout and Iout measuring function
• Number of decimal places which appear at the display
• Stabilizes the display when fluctuation is expected at the last
decimal place of the measured value
Setting Parameter
☞
0: no rounding
☞ -1: Rounded off one decimal place
☞ -2: Rounded off two decimal places
Settings
The procedure is described in principle in chapter 4.7.
!The rnd display and a related text
parameter appear in the setup function
window.
Settings
The procedure is described in principle in chapter 4.7.
!The UI_ display and a related text
parameter appear in the setup
function window.
☞ Default setting after RESET (*RST): OFF!
Figure 4.7.1 dPath to Selection of the Pon Text Parameter
Figure 4.7.1 e Path to Selection of the rnd Text Parameter
☞ Default setting after RESET (*RST): unchanged
GMC-I Messtechnik GmbH27
4.7.2 AnIF – “Analog Interface” Function Group
Select the text parameter
from an options loop
From the AniF function group
*
ENTER
CE/LOCALFUNCTION
The analog interface allows for remote control of the
KONSTANTER.
An isolated digital control input (TRG IN+ / TRG IN-) and digital signal
outputs (SIG1 / SIG2) are available in addition to analog setting
options.
Other devices can be controlled in this way (e.g. counters, alarms,
SSP KONSTANTERs and many more). A master-slave system can
be set up in combination with a second or several
SSP KONSTANTERs (see also chapter 5).
Observe specified maximum current and voltage values for signal
outputs.
SEQUENCE: Sequential execution of specified values for Iset,
Uset and dwell time (see also chapter 4.7.3).
The SEQUENCE function group must be set to SEQUENCE ON
for this function.
TRIGGER Signal Influence on SEQUENCE Status
LOW _
HIGH _ LOWEnds sequence execution by jumping to
HIGH
The SEQUENCE function is started beginning
with the start address,
the stop address.
SEQUENCE GO
☞ LLO: (LOCAL LOCKED) Operation via front panel controls is
disabled.
trG – Function Selection for Trigger Input
Functions
• Floating optocoupler input (TRG IN) at the analog interface
• The trigger input is controlled with a digital signal
(low: 0 V, high: 4 ... 26 V).
• Response time is approximately 1 to 15 ms.
Setting Parameter
☞
OFF: Trigger input function is deactivated, signals at the trigger
input are ignored.
☞ OUT: Trigger input influences the OUTPUT: output on / off
TRIGGER Signal Influence on OUTPUT Status
LOWStatus depends upon manual setting or
LOW _
HIGHOUTPUT in the OFF state,
HIGH
HIGHOUTPUT remains OFF or
_ LOWOUTPUT is activated,
programming command
OUTPUT is switched OFF
cannot be switched on – neither manually
nor with a programming command
except in the event of OTP or OVP
☞ RCL: Memory recall from defined start to defined stop address
(step-by-step sequence control)
TRIGGER Signal Influence on step-by-step control
Edge
Low High
Start trigger signal
TRIGGER Signal Influence on Front Panel Controls
LOWAll front panel controls are functional.
HIGHAll front panel controls are disabled,
except for the mains switch,
no activation with the LOCAL key.
☞ UI_
MIN-MAX function: Storage of Min-Max values for U and I is
controlled by the trigger,
Min-Max function active, (UI _
TRIGGER Signal Influence on Min-Max Status
LOWThe Min-Max function is active.
LOW _
HIGHThe Min-Max function is inactive.
HIGH
HIGHThe Min-Max function is deactivated.
Levels in the in Min-Max memories
remain unchanged.
_ LOWValues in the Min-Max memories are
replaced with momentary output values.
The Min-Max function is activated.
ON)
☞ The status of the trG function is not saved as a device setting
with the SAVE key.
Settings
☞
Procedure for configuring trG ➁ see also
chapter 4.7
!The trG display and a related text
parameter appear in the AnIF function
window.
☞ Default setting after RESET (*RST):
unchanged
HIGH The trigger signal is a high pulse with
a duration of 11 to 800 ms.
A high pulse with a duration > 1.0 s
resets the address counter to the start
address at any point in time, and
execution begins with the next trigger
signal.
Edge
High Low
The high low edge of the (short) trigger
signal results in step-by-step control of
the currently selected sequence,
regardless of the specified time and
number of repetitions.
Recall of the memory’s contents begins
with the START address. Each trigger
signal increasers the address by 1, until
the STOP address is reached. The next
pulse once again causes execution of the
contents of the START address.
Figure 4.7.2 aPath to Selection of the trG Text Parameter
☞ SEQ: Controlling Execution of the SEQUENCE Function
28GMC-I Messtechnik GmbH