SECUTEST ST BASE(10) / PRO
and SECULIFE ST BASE(25)
Testers for Testing the Electrical Safety of Devices
per VDE 0701-0702, IEC 62353 and IEC 60974-4
Important
Read carefully before use.
Keep on file for future reference!
3-447-066-03
2/10.20
Please read the full operating instructions as well,
which are available as a PDF file at
www.gossenmetrawatt.com.
The condensed operating instructions do not replace
the full operating instructions!
Page 2
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
Note
MEM: Database
ESC
:Return
HELP: Help images
START
: Start/Stopp
– Single meas.
– Test seq.
Finger contact
PRINT
: Print via USB
Fixed Function Keys
Symbol display for devices
connected to the USB master
interface see below
Softkeys
2 USB master
– For keyboard
– for scanner
– for printer
– for USB stick
1 USB slave
– for PC
Test socket for
connecting DUTs
Attention!
Depending on the measuring
task, the test probe may be
charged with line v
oltage.
!
Rotary
Selector Switch
LC panel
Connection for
voltage supply
SECUTEST CLIP
(Z745H)
Sockets
2
nd
test probe
Special Symbol Displays:
– Measurement at IT line active
– OFFSET for RPE active
Kink protection grommet:
black: max. 16 A
green: max. 25 A
Mains on test socket
SFC: Normal condition
Mains on test socket
SFC: N interrupted
white/silver identified &
fused high current path
Fuse compartment
Voltage measurement
Rotary switch level: orange
Test sequences A1 ... A9
(test sequences per standard)
Rotary switch level: green
Single measurements
Test Pr o b e
Operating and Connection Overview
Features such as touchscreen, Bluetooth® etc. can be found in your instrument’s
data sheet.
2Gossen Metrawatt GmbH
Page 3
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
Note
1Safety Precautions
SECUTEST ST BASE(10), SECUTEST ST PRO and
SECULIFE ST BASE(25) test instruments are
manufactured and tested in accordance with
the following safety regulations:
IEC/EN 61010-1 / VDE 0411-1, DIN VDE
0404, IEC/EN 61577 / VDE 0413-2,-4 /
DIN
EN 61557-16 / VDE 0413-16
The “automatic test sequences” are semiautomatically controlled sequences whose
progress is interrupted by safety-relevant
stoppages and the associated warnings and
instructions. As a result, the level of protection provided to the user is greater than
demanded by sections 4.1.6 and 4.1.7 of
DIN EN 61557-16 with regard to “automatic
test sequences”.
The safety of the user, the test instrument
and the device under test (electrical equipment or electrical medical device) is only
assured when the instrument is used for its
intended purpose.
Read these condensed operating instructions and the
full operating instructions carefully and completely
before placing your test instrument into service (available at our homepage www.gossenmetrawatt.com for
download). Follow all instructions contained therein.
Make sure that the operating instructions are available
to all users of the instrument.
Tests may only be performed by a qualified
electrician, or under the supervision and
direction of a qualified electrician. The user
must be instructed by a qualified electrician
concerning performance and evaluation of
the test.
Suitable personal safety equipment is
required.
If you use any active or passive implanted
medical devices, please consult your doctor
or the manufacturer of these devices.
Manufacturers and importers of electrical
medical devices must provide documentation for the performance of maintenance by trained personnel.
Observe the following safety precautions:
•The instrument may only be connected to
electrical systems (TN, TT or IT) with a maximum of 240 V which complies with applicable safety regulations (e.g. IEC 60346, VDE
0100) and is protected with a fuse or circuit
breaker with a maximum rating of 16 A.
• Measurements within electrical systems
are prohibited.
• Be prepared for the occurrence of unexpected voltages at devices under test
(for example, capacitors can be dangerously charged).
• Make certain that the measurement
cables are in flawless condition, e.g. no
damage to insulation, no cracks in
cables or plugs etc.
• When using a test probe with coil cord
(SK2W): Grip the tip of the test probe
firmly, for example during insertion into a
jack socket. Tensioning at the coil cord
may otherwise cause the test probe to
snap back resulting in possible injury.
•
Measurement of insulation resistance and
equivalent leakage current
alternative measuring methods)
Testing is conducted with up to 500 V. Current limiting is utilized (I < 3.5 mA), but if
terminals L or N at the test socket are
touched, electrical shock may occur which
could result in consequential accidents.
• During operation with line voltage:
Exposed conductive parts may conduct
dangerous touch voltage during testing.
Do not touch under any circumstances!
Use a special cover in order to avoid
touch contact.
Mains power is disconnected at the
instrument if leakage current exceeds
approx. 10 mA (can be switched to 30
mA). However, this does not fulfill the
requirements specified for a PRCD. If the
“Proceed in case of limit violation” setting
is used, enhanced safeguarding against
touch contact and a 30 mA RCD must
be used, and personal protective equipment (PPE) must be worn (secure workstation).
(leakage current
4Gossen Metrawatt GmbH
Page 5
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
Attention!
!
Attention!
!
The function test may only be performed after the DUT has successfully
passed the safety test!
• Probe check for probe connector P1: Please
perform a probe check after each test.
If there is a fuse defect in test probe
P1 after having started the test, all
subsequent measurements which
are performed with this measurement path might be erroneously
rated as good!
Fuse replacement
The fuses may only be replaced when the
instrument is voltage-free, i.e. the instrument
must be disconnected from mains supply
power and may not be connected to a measuring circuit. The fuse type must comply
with the specifications in the technical data
or the labeling on the instrument.
Opening the instrument / repairs
The instrument may only be opened by
authorized, trained personnel in order to
ensure flawless operation and to assure that
the guarantee is not rendered null and void.
Even original replacement parts may only be
installed by authorized, trained personnel.
If it can be ascertained that the instrument
has been opened by unauthorized personnel, no guarantee claims can be honored by
the manufacturer with regard to personal
safety, measuring accuracy, compliance
with applicable safety measures or any consequential damages.
Any warranty claims will be forfeited when
the warranty seal has been damaged or
removed.
Switching power consumers (max. 16 A*)
Be absolutely sure to adhere to the
sequence specified below when switching
the live device under test. This prevents
excessive wear of the mains relays at the
test instrument.
Before measurement:
1) DUT: Turn the DUT off via its own switch.
2) Tes te r:
Switch line voltage to the test
socket.
3) DUT: Turn the DUT on via its own switch.
After measurement:
4) DUT: Turn the DUT off via its own switch.
5) Tes te r:
Deactivate line voltage to the test
socket.
* for currents > 16 A AC please use adapter
AT3-IIS32 (Z745X), for example
The test instrument may not be used:
•If external damage is apparent, for example
if parts which conduct dangerous touch
voltage are freely accessible, if the display
is broken or defective (in which case dangerous voltage or mains connection errors
might no longer be indicated)
•If the seal or sealing lacquer has been
removed as the result of repairs or manipulation carried out by a non-authorized/noncertified service provider.
•With damaged connection and/or measurement cables and patient ports, e.g.
interrupted insulation or kinked cable
•If the instrument no longer functions flawlessly
•After serious damage due to transport
In such cases, the instrument must be
removed from operation and secured
against unintentional use.
Gossen Metrawatt GmbH5
Page 6
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
!
Meanings of Symbols on the Instrument
The symbols on the instrument have the following meanings:
Warning regarding dangerous
electrical voltage
Warning concerning a point of
danger (attention: observe documentation!)
Indicates European Conformity
This device may not be disposed of
with the trash. Further information
regarding the WEEE mark can be
accessed on the Internet at
www.gossenmetrawatt.com by
entering the search term “WEEE”.
If the guarantee seal is damaged
or removed, all guarantee claims
are rendered null and void.
QR CODEQR CODE is a registered
trademark of
DENSO WAVE INCORPORATED
2Scope of Delivery
The scope of delivery varies depending on
which instrument variant has been ordered,
and is country-specific. Information concerning the scope of delivery can be found in
your order and the data sheet, in which all
order information is listed.
Example of a standard variant (standard
model):
1 SECUTEST ST BASE, SECUTEST ST PRO or
SECULIFE ST BASE(25) test instrument
1 Mains power cable
1 Test probe, 2 m, not coiled
1 USB cable, USB A to USB B,
length: 1.0 m
1 Plug-on alligator clip
1 KS17-ONE cable set for voltage measur-
ing inputs
(only with SECUTEST PRO or an instrument
with feature I01 and SECULIFE ST BASE(25))
1 Calibration certificate
1 Condensed operating instructions
– Set of complete operating instructions
available on the Internet for download
from
www.gossenmetrawatt.com
1 Card with registration key for
IZYTRONIQ BUSINESS Starter software
6Gossen Metrawatt GmbH
Page 7
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
Attention!
!
Attention!
!
Note
3Initial Start-Up
3.1Mains Connection
Nominal mains values: 100 to 240 V, 50 Hz to 400 Hz
➭ Connect the test instrument to the mains
cable via its inlet plug and insert the
mains plug into an electrical outlet. The
function selector switch can be set to
any position. If a mains outlet (earthing
contact outlet) is not available, or if only a
3-phase outlet is available, the adapter
socket can be used to connect the
phase conductor, the neutral conductor
and the protective conductor. The
adapter socket has three permanently
attached cables and is included with the
KS13 cable set (see wiring diagram in
the operating instructions).
If connection is not possible via an
earthing contact outlet: Shut down
mains power first. Then connect the
cables from the coupling socket to
the mains using pick-off clips in accordance with the diagram. Disconnection from mains power is only
possible with the mains plug.
Measurements in IT Systems
The setting IT system can be activated in
selector switch position SETUP (Setup 1/3)
under sub-menu All Measurement:
Parameter „Meas. at IT-mains“ = Yes: active
leakage current measurements (and/or all
measurements which include the PE at the
mains connection end) are blocked. Test
sequences which contain such kind of measurements are disabled as well.
3.2Detection of Mains Connection Errors
The device automatically recognizes mains
connection errors if the conditions in the following table have been fulfilled. The user is
informed of the type of error, and all measuring functions are disabled in the event of
danger.
Typ e of
Connection Error
Voltage at protective conductor PE
to finger contact
(START/STOP key)
Protective conduc-
tor PE & phase
conductor L
reversed and/or
neutral conductor
N interrupted
Line voltage
< 180 V / < 90 V
(depending on mains)
Tes t o n IT/ TN
system
1
10 A/25 A RPE measurements are only possible with line
voltages of 115/230 V and line frequencies of 50/60 Hz.
2
if the test person is highly insulated, the following error
message may appear: „Interference voltage at PE of
mains connection“
MessageCondition
Press START/
Display at the
instrument
Display at the
instrument
STOP button
U 25 V
Button PE:
< 1 M
Voltage at PE
> 100 V
U
L-N
U
L-N
Connection
N PE
> 20 k
In the event of mains connection errors as described in either of the first
two cases in the table above, immediately disconnect the test instrument
from the mains and arrange to have
the error eliminated!
Voltage at the
tive conductor PE
electrical system’s
may result in distorted
measurement values during testing for
the absence of voltage, or during leakage voltage measurements.
< 180 V
<90V
Measure-
All measure-
2
under certain
under certain
circumstances
ments
ments
disabled
Impossible
(no supply
power)
Possible
circum-
stances
Possible
protec-
1
Gossen Metrawatt GmbH7
Page 8
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
4
Overview of Functions of the
Tes te rs
Switch
setting,
Measuring functions,
test current/voltage
description as of
Single measurements,
rotary switch level: green
Measurements at voltage-free objects
R
PE
Page 15
R
INS
Page 16
R
Iptest current (200 mA)
R
U
protect. conductor resistance
PE
SECUTEST ST BASE(10)/PRO
SECULIFE ST BASE: 10 A
(Feature G01)
SECULIFE ST BASE25: 25 A
(Feature G02)
insulation resistance
INS
test voltage
INS
Measurements at DUTs with line voltage
I
PE
Page 18
I
T
Page 19
I
E
Page 20
I
A
Page 21
I
P
Page 22
U
Page 23
t
PRCD
Page 25
I
PE
I
PE~
I
PE=
U
U
I
T
I
T~)
I
T=
U
U
I
E
I
E~
I
E=
U
U
I
A
U
U
I
P
I
P~
I
P=
U
U
U
U
U
U
U
3
t
a
U
prot. conductor current, TRMS
AC component
DC component
test voltage
LPE
reference voltage (alternative)
Gen
touch voltage, TRMS
AC component
DC component
test voltage
LPE
reference voltage (alternative)
Gen
device leakage current, TRMS
AC component
DC component
test voltage
LPE
reference voltage (alternative)
Gen
leakage current from app. part
test voltage
LPE
reference voltage (alternative)
Gen
patient leakage current, TRMS
AC component
DC component
test voltage
LPE
probe voltage, TRMS
alt. voltage component
~
direct voltage component
=
RMS Voltage
AC voltage component
~
DC voltage component
=
time to trip for 10/30 mA PRCD
line voltage at the test socket
LN
2
Switch
setting,
Measuring functions,
test current/voltage
description as of
P
1
1)
Page 26
Function test at the test socket
Icurrent between L and N
Uvoltage between L and N
ffrequency
Pactive power
Sapparent power
PFpower factor
Probe measuring functions
EL1
Page 27
EXTRA
Page 28
Test sequences per standard
Page 32
Extension cord function test with EL1/VL2E/
AT3-IIIE adapter: continuity, short-circuit, re-
versed wires
Reserved for expansions in connection with
software updates
4
°C Temperature measurement 2 with Pt100/Pt1000
IZ Measurement of current at clamp with cur-
rent clamp sensor
rotary switch level: orange
Preconfigured (freely selectable) test sequences
A1
A2
A3
A4
A5
A6
A7
A8
A9
1
2
3
4
2
2
VDE 0701-0702, passive measurement type,
test socket
VDE 0701-0702, active measurement type,
test socket
VDE 0701-0702
EDP (active)
IEC 62353
IEC 62353
, parameters configuration for
(VDE 0751), passive measurement type
(VDE 0751), active measurement type
IEC 60974-4, connection type: test socket
IEC 60974-4
VDE 0701-0702,
cord
VDE 0701-0702
type, protection category – in each case automatic
10 A/25 A-RPE measurements are only possible with line
voltages of 115 V/230 V and line frequencies of 50 Hz/60 Hz.
Voltage measuring inputs only for
for devices with feature I01) and
Measurement of time to trip not possible in IT systems.
No checking for reversed polarity takes place when the
EL1 adapter is used.
, connection type: AT16-DI/AT32-DI
test
measurement type ext
(RPE, RISO), EL1/VL2E/
, connection type, measurement
SECUTEST ST PRO
SECULIFE ST BASE(25)
AT3-IIIE
ension
adapter
(or
8Gossen Metrawatt GmbH
Page 9
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
Note
Note
Changes in test sequences A1 ...
A9
are preserved even after switching off
the test instrument.
Features involving expansions (firmware) which can be enabled are available for an additional charge if they’re
not included in the scope of delivery.
Example: Feature KD01, “Z853S –
SECUTEST DB COMFORT”, as mentioned in this document.
Refer to the test instrument’s operating instructions for further information.
Gossen Metrawatt GmbH9
Page 10
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
5User Interface Symbols – Parameter and Softkey Symbols
Sym-
Setup
bols
Parameters and their significance
Page
Complete overviews of all symbols are included in the full operating instructions.
All measurements
1/3
have been standardized; Gnd fault sens.: continuous residual current monitoring (10/30 mA)
: Ref.voltage L-PE: voltage to which the measured values for leakage current
Automatic measurements: set parameters for test sequences: start and end view,
inclusive operation uncertainty (yes/no), auto measurement point (yes/no)
Database: deletion, statistics,
1/3
with inserted USB stick : save, restore database
System: set general device parameters;
1/3
2/3
2/3
2/3
2/3
3/3
date/time, volume, brightness,
Printer: printer selection for USB master interface
connected, disconnected
Te st er : select tester from list, add new tester
Culture: select language for operating instructions, keyboard and measuring
sequences by acknowledging the respective national flag; Reboot necessary!
Optionally connected external devices
USB stick, keyboard / barcode scanner, printer
System information: query software and hardware version, serial number, build
number, calibration data and storage occupancy
default settings, self-test
— Functions and their significance
Set classification parameters for the respective test sequence
(test sequences switch settings A1 ... A9)
Accept parameters, acknowledge message
Abort single measurement or test sequence
Evaluate measurement of visual inspection with OK or not OK (toggle key)
Continue test, next test step in the test sequence
Symbol left: Direct selection key measurement type (connection type...) or measuring method (direct ...) / Symbol right: Selection between two states (no submenu)
Start evaluation – record measured value. Each time this softkey is pressed, an
additional measured value is saved and the number is increased by one.
Symbol left: Repeat measured value recording
Symbol right: Repeat test step in the test sequence
Symbol left: Delete measured value
Symbol right: Skip individual tests in the test sequence
Display measured values from performed measurements and test sequences
Magnifying glass symbol: show (+) or hide (–) details regarding database objects
or selected measurements
Enter a new ID for a test object either before or after a test, and in case the ID
has not yet been entered to the structure
Save measurement data / save measurement data as (with display of directory
path / ID or new entry of an ID other than the preselected one)
10Gossen Metrawatt GmbH
Page 11
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
Note
6Internal Database
6.1Creation of Test Structures
A complete test structure with data regarding
customers and devices under test be created
in the test instrument. This structure makes it
possible to store the results of single measurements or test sequences of test objects
belonging to various customers. Manual single
measurements can be grouped together into a
so-called “manual sequence”.
Up to 50,000 data records can be stored in
the test instrument. The following applies in
this regard: 1 data record = 1 DUT or location node or customer or individual measurement.
A complete description of database creation is
included in the full operating instructions for
your test instrument.
6.2Export – Transmission and Storage of
Test Structures and Measurement Data
Structures set up in, and measurement data
saved to the test instrument can be imported
to the IZYTRONIQ report generating software
via a plugged-in USB stick (PRO and/or Feature KB01 only) or via the USB slave port.
Data can then be saved to the PC and
reports can be generated.
Data transfer to IZYTRONIQ should not
be started during single measurements or test sequences.
Furthermore, you can save the database to
a plugged-in USB stick for subsequent
restoring to the device memory.
6.3Import (PRO and/or Feature KB01 only)
The test structures created at the PC with
the help of the report generating software
can be loaded into the test instrument via an
USB stick or via the USB slave port.
Gossen Metrawatt GmbH11
Page 12
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
PRINT
ESC
HELP
MEM
Display Field
Key Field
Example Softkey Entries
Switch between
Scroll down
Transfer char. at cursor
Exit entry function
position to display field
without saving
Scroll up
key and display field
Delete characters
Toggling: uppercase,
Scroll left
Accept entry
Scroll right
7Data Input
7.1Keyboard or Softkey Entries
After selecting ID or any other object parameter, a keyboard is displayed which allows
for the entry of alphanumeric characters via
the fixed function keys and the softkeys.
Alternatively, entries can also be made with
the help of a USB keyboard or barcode
scanner which is connected to the instrument.
Procedure
(example: entering a DUT designation)
1 Switch the keyboard to uppercase, low-
ercase or special characters with the abc
key (Abc, ABC, Symb).
2 Select the desired alphanumeric charac-
ter or a line break with the scroll keys
(left, right, up and down). The selection
cursor can be accelerated by pressing
and holding the respective scroll key.
3 After pressing the key, the respec-
tive character appears in the display
field.
4 Repeat steps 1 through 3 until the com-
plete designation is shown in the display
field.
5 After pressing the green check-
mark, the selected character string
is saved.
7.2Entry via Touch Screen (
and/or Feature E01 only)
PRO
SECUTEST ST
The touch-screen allows for the convenient
entry of data and comments, parameters
and direct parameter selection.
12Gossen Metrawatt GmbH
Page 13
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
8Notes on Saving Single Measurements and Test Sequences
At the end of each test, test results can be
saved under an ID number which is unequivocally assigned to the respective test object.
Depending on the initial situation, i.e.
whether or not a test structure or database
is already available or an ID has already been
entered, the following different procedures
are used for saving:
Variant 1 – pre-selection of an existing ID
You’ve already set up a test structure in the
test instrument
IZYTRONIQ
Open the database view before starting the
measurement by pressing the MEM key.
Then select the test object or its ID within the
test structure by pressing the respective
scroll key. Exit the database view (MEM navigation) by pressing ESC and start the measurement. Press the Save as key at the
end of the measurement. The display is
switched to the SAVE view. The ID appears
with a green or orange background. Press
the Save key in order to complete the procedure.
Variant 2 – entry of an existing ID at the end of the
test
You’ve already set up a test structure in the
test instrument
IZYTRONIQ (
measurement without first opening the database. No test object was previously selected
in the database. Press the Save as key at
the end of the measurement. The following
message appears: “No DUT selected!”
Press the ID key. The softkey keyboard
appears.
If you enter an ID here which is already in the
database, the database view appears automatically (MEM navigation) and the test
object’s ID is displayed inversely. Acknowledge the entry by pressing the key. The
display is switched to the SAVE view. The ID
appears with a green or orange background.
Press the Save key once again in order to
complete the procedure.
or loaded one from the
.
or loaded one from the
SECUTEST PRO
only)
. You perform the
Variant 3 – entry of a new ID at the end of the test
You haven’t yet set up a test structure in the
test instrument, or the ID is not included in
the existing structure.
Press the Save as key at the end of the
measurement. The following message
appears: “No DUT selected!” Press the ID
key in order to enter the test object’s ID. The
softkey keyboard appears.
If you enter an ID here which is not yet
included in the database, a prompt appears
asking you if you want to enter a new object.
– : If you press , the display is switched
to the SAYVE view. The ID appears with a
green background. Press the key once
again in order to complete the procedure.
– : If you press , the database view
appears (MEM navigation). Go to the next
page (Process objects 2/3) by pressing ,
and then enter a new test object. Press
to this end. All possible object types
are displayed. Press “DUT”. The newly
entered ID appears in red to the right of
the ID parameter. Acknowledge the entry
by pressing the key. The display is
switched to the database view (MEM navi-
gation). The newly entered test object is
displayed inversely in the structure. Press
ESC in order to return to the SAVE view. The
ID appears with a green or orange background. Press the key once again in
order to complete the procedure.
– ESC: If you don’t want to save any mea-
sured values, press ESC twice in order to
go to the measuring view. If you press ESC
again, a prompt appears asking whether
or not you want to delete the measuring
points in order to continue with the measurement without saving.
Gossen Metrawatt GmbH13
Page 14
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
MEM
ESC
3
3
3
Select DUT
Start stop initiate saving
1234
Check end saving
9Single Measurements
Any measuring duration is possible. The
respective measurement is ended by pressing START/STOP. No limit values have been
entered for single measurements. As a
result, no evaluation of the measurement
results takes place.
Procedure for measuring with save function and
pre-selection of the (ME) device
If you’d like to save your single measurements to selected (ME) devices in a database (see section 6), the following procedure
is advisable.
1 Activate the database view (
igation) by pressing the
2 Select the (ME) device or its ID num-
ber for the following measurements
with the scroll keys.
3 Return to the measuring view by
pressing the ESC key or the START/
STOP key.
4 Start the test with the START/STOP
key before switching the DUT on
The measured value recording
symbol shown at the right
appears. Each time this key is
pressed, the currently displayed
value is saved to the clipboard
and the number shown in the symbol is
increased by one.
5 Switch off the DUT before finis-
hing the test with the START/STOP
key.
The Saveas symbol appears
(floppy disk icon with the number
of measured values saved to the
clipboard).
6 If you press the save symbol now,
the display is switched to the SAVE
view, where the pre-selected (ME)
device is highlighted.
7 After pressing the Save symbol
once again, acknowledgement of
successful storage appears. At the
same time, the display returns to the
measuring view.
MEM
MEM
key.
nav-
.
Procedure for measuring with save function and
subsequent entry of the (ME) device
As an alternative to the procedure described
above, you can start with step 4 and, after
measurement has been completed, assign
the results to a (ME) device or its ID number
which is included in the database: manually
by selecting ID and entering the ID via the
alphanumeric keyboard, or by scanning a
barcode.
14Gossen Metrawatt GmbH
Page 15
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
Select measuring function
1
R
PE
Select parameters
2
Set parameters
3
Connect DUT
4
PE(PD) – P1
P1
PE(mains) – P1
PE(mains)
P1
Start test
5
Acknowledge line voltage warning
6
!
With active: PE(TS)-P1 only
Save measured values to clipboard
6
Stop test
7
Save measurements under ID no.
8
9.1RPE – Protective Conductor Resistance for Protection Class I Devices
1
SECUTEST ST BASE(10) / SECUTEST ST PRO /
SECULIFE ST BASE
2)
SECULIFE ST BASE25
3)
The selected offset value is permanently stored to memory and
adapted for measurements in selector switch position
4
Connection also via EL1, VL2E, AT3 adapter, AT16DI/AT32DI
➭ Connect the DUT to the test socket.
➭ Contact all conductive parts which are
connected to the protective conductor
with test probe P1.
Special case: permanently installed DUT
Meas. ParameterMeaning
Mode (Measurement type)
admissible test current IP
passive: PE(TS) – P1
@ IP =
200 mA/10 A
active: PE(PD) – P1
@ IP = 200 mA/25 A
PE(mains) – P1
Permanently
connected devices
@ IP = 200 mA/10 A
PE(Mains) – P1 clip@ IP = 10 A, see section 9.16
P1 – P2
@ IP =
200 mA/10 A/25 A
Testing between the 2 protective
conductor terminals: at test sock-
4
et
and test probe P1
Like PE(TS) – P1
voltage to socket and with continuously increasing DC current (PRCDs)
Testing between ground terminal
at mains and test probe P1
Only test devices with feature H01:
2-pole measurement between test
probe 1 and 2, see section 9.15
but with mains
➭ Contact all conductive housing parts
with test probe P1.
IP(set)Ip
200 mA
10 A
25 A
1
2
Test cur.: 200 mA AC (+/–/± DC)
Test current: 10 A
Test current: 10 A
(Feature G01)
(Feature G02)
f – only at 200 mA ~ (AC)
50 to 200 Hz
Offset
> 0 to < 2
3
Test sequence (adjustable in increments)
Zero balancing for selected ref. point.
A1 ... A9
.
Gossen Metrawatt GmbH15
Page 16
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
Select measuring function
1
RINS
Select parameters
2
Set parameters
3
Connect DUT
4
LN(TS) – PE(TS)
LN(TS) – PE(TS)
P1
Start test
5
Save measured values to clipboard
6
Stop test
7
Save measurements under ID no.
8
9.2RINS – Insulation Resistance Measurement for Protection Class I Devices
Special case: inputs for application parts
U+/U– = increase/decrease UINS(set)
➭ Connect the DUT to the test socket.
➭
Meas. ParameterMeaning
Mode (Measurement type)
Contact the short-circuited inputs for the
application parts with test probe P1.
LN(TS) – PE(TS)
LN(TS) – P1See section 9.3
P1 – P2
PE(mains) – P1
Permanently
connected devices
PE(TS) – P1
LN(TS) – P1//PE(TS)
PC I: Testing between short-circuited LN mains terminals at test
socket and the DUT’s PE terminal
Only test devices with feature H01:
2-pole measurement between test
probe 1 and 2 instead of test socket
connection
Cable test: Testing between ground
terminal at mains and test probe P1
Testing between PE terminal at test
socket and test probe P1
Testing between short-circuited LN
mains terminals at test socket and test
probe P1, including PE at test socket
1
UINS(set)U+/U–
> 50 to < 500 VVariable test voltage can be en-
1
Connection also via EL1, VL2E, AT3-II I E, AT3-I IS, AT3-I I S32,
AT16DI/AT32DI or CEE adapter
16Gossen Metrawatt GmbH
tered with the numeric keypad
Page 17
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
Select measuring function
1
RINS
Select parameters
2
Set parameters
3
Connect DUT
4
LN(TS) – P1
P1
Start test
5
Save measured values to clipboard
6
Stop test
7
Save measurements under ID no.
8
9.3RINS – Insulation Resistance Measurement for Protection Class II Devices
➭ Connect the DUT to the test socket.
➭ Contact all conductive, exposed parts
with test probe P1.
U+ = increase UINS(set)
U– = decrease UINS(set)
Meas. ParameterMeaning
Mode (Measurement type)
LN(TS) – P1Testing between short-circuited LN
UINS(set)U+/U–
> 50 to < 500 VVariable test voltage can be en-
Gossen Metrawatt GmbH17
mains terminals at test socket and
test probe P1
(Connection via test socket, via
adapter VL2E, AT3-IIIE,
AT3-II S32 or AT16DI/AT32DI
tered with the numeric keypad
AT3-IIS,
)
Page 18
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
Select Measuring Function
1
IPE
Select parameters
2
Set parameters
3
Connect DUT
4
Direct / Diff / Alt
Start test
5
Acknowledge line voltage warning
6
!
Direct & differential & AT3 Adapter:
Save measured values to clipboard
7
Stop test
8
Save measurements under ID no.
9
2)
9.4IPE – Protective Conductor Current
Meas. ParameterMeaning
Mode (Measurement type)
only suitable for connection of DUT with test socket.
Not suitable for measurements at AT16DI or AT32DI
adapter.
Prior to all leakage current measurements, please
make sure that the measurement parameters „Ref.
voltage L-PE“ and „Testingfreq. Alt“ have been correctly set in the SETUP, see section 11.
➭ Connect the DUT (to test socket).
Direct
DifferentialDifferential current measurement
AlternativeEquivalent leakage current mea-
AT3 Adapter
Clip
Direct measuring method (via test
socket, AT16DI/AT32DI)
(via test socket)
suring method (via test socket
Only test devices with feature
Measurement with adapter AT3IIIE,
AT3-IIS or AT3-II S32
See section 9.17
Only test devices with feature
See section 9.16
I01:
I01:
➭Switch DUT on
1
)
➭Switch DUT off
Single Fault (SFC) – for direct measurement type only
Normal conditionNo single fault simulation active
N interruptedFault simulation - only phase and
protective conductor are connected with the DUT
2)
Polarity – for direct, differential measurement
and AT3 adapter only
normal Selection of polarity for line volt-
reverse
1
Connection also via VL2E, AT3 adapter, AT16DI/AT32DI
18Gossen Metrawatt GmbH
age at test socket
Page 19
9.5IT – Touch Current
Select measuring function
1
IT
Select parameters
2
Set parameters
3
Connect DUT
4
DirectP1/DiffP1/AltP1
P1
Start test
5
Acknowledge line voltage warning
6
!
Direct & differential:
Save measured values to clipboard
7
Stop test
8
Save measurements under ID no.
9
Meas. ParameterMeaning
Mode (Measurement type)
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
1
Connection also via
2
only suitable for connection of DUT with test socket.
Not suitable for measurements at AT16DI or AT32DI
adapter.
Prior to all leakage current measurements, please
make sure that the measurement parameters „Ref.
voltage L-PE“ and „Testingfreq. Alt“ have been correctly set in the SETUP, see section 11.
➭ Connect the DUT to the test socket.
➭ Contact additional, accessible, conduc-
tive parts which are not connected to the
protective conductor with test probe P1.
AT3-II IE, AT3-II S, AT3-II S32, AT16DI/AT32DI
Direct P1Direct measuring method
Differential P1Differential current measurement
Alternative P1
Permanent connection P1
Alternative P1–P2
Single Fault (SFC) – for direct measurement type only
Normal conditionNo single fault simulation active
N interruptedFault simulation - only phase and
PE interrupted
Polarity – for direct, differential measurement
and AT3 adapter only
normal Selection of polarity for line volt-
reverse
Gossen Metrawatt GmbH19
(via test socket 1)
(via test socket)
Equivalent leakage current measuring
method (via test socket 1 or VL2E)
Permanently connected DUT
Only test devices with feature H01:
Equivalent leakage current measurement method:
2-pole measurement between test
probe 1 and 2, see section 9.15
protective conductor are connected with the DUT
Fault simulation active - protect. conductor is disconnected from the DUT
for the duration of measurement
age at test socket
2
Page 20
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
Select measuring function
1
IE
Select parameters
2
Set parameters
3
Connect DUT
4
Direct / Alt
P1
Start test
5
Acknowledge line voltage warning
6
!
Direct & differential & AT3 Adapter:
Save measured values to clipboard
7
Stop test
8
Save measurements under ID no.
9
9.6IE – Device Leakage Current
Prior to all leakage current measurements, please
make sure that the measurement parameters „Ref.
voltage L-PE“ and „Testingfreq. Alt“ have been correctly set in the SETUP, see section 11.
➭ Connect the DUT to the test socket.
➭ Contact accessible, conductive parts
which are not connected to the protective conductor with test probe P1.
➭ For DUTs with application parts: addi-
tionally contact the short-circuited inputs
Meas. ParameterMeaning
Mode (Measurement type)
DirectDirect measuring method (via test
DifferentialDifferential current measurement
AlternativeEquivalent leakage current mea-
AT3 adapter
Clamp
Polarity 2) –
normal Selection of polarity for line volt-
reverse
1
Connection also via AT16DI/AT32DI (only differential
current methtod useful)
2)
Measurement must be performed for both mains polarities and the largest value is documented
socket1), optional probe contact
(via test socket)
suring method with probe contact
(via test socket, AT16DI/AT32DI)
Only test devices with feature
Measurement with adapter AT3IIIE,
AT3-IIS or AT3-II S32
See section 9.17
Only test devices with feature
see section 9.16
for direct, differential measurement
and AT3 adapter only
age at test socket
I01:
I01:
for the application parts with test probe
P1.
20Gossen Metrawatt GmbH
Page 21
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
Select measuring function
1
IA
Select parameters
2
Set parameters
3
Connect DUT
4
device under test
Direct P1/ PC 1
Start test
5
Acknowledge line voltage warning
6
!
Direct:
Save measured values to clipboard
6
Stop test
7
Save measurements under ID no.
9
9.7IA – Leakage Current from the Application Part
➭ Connect the DUT (to test socket).
➭ Contact the short-circuited inputs for the
application parts with test probe P1.
Meas. ParameterMeaning
Mode (Measurement type)
Direct P1Direct measuring method (via test
Alternative P1Equivalent leakage current mea-
Permanent connection P1Permanently connected DUT
Phase Angle – for Direct (P1) and Perm. c. only
0 ° or 180 ° Selectable phase angle of the in-
Polarity –
for direct, differential measurement
normal Selection of polarity for line
reverse
Prior to all leakage current measurements, please
make sure that the measurement parameters „Ref.
voltage L-PE“ and „Testingfreq. Alt“ have been correctly set in the SETUP, see section 11.
Gossen Metrawatt GmbH21
socket) with test probe P1
suring method (via test socket)
with test probe P1
ternal generator in relation to
mains phase angle
and AT3 adapter only
voltage at test socket
Page 22
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
Select measuring function
1
IP
Select parameters
2
Set parameters
3
Connect DUT
4
P1
Start test
5
Acknowledge line voltage warning
6
!
Direct
Save measured values to clipboard
7
Stop test
8
Save measurements under ID no.
9
9.8IP – Patient Leakage Current
Prior to all leakage current measurements, please
make sure that the measurement parameters „Ref.
voltage L-PE“ and „Testingfreq. Alt.“ have been correctly set in the SETUP, see section 11.
➭ Connect the DUT to the test socket.
➭ Contact the short-circuited inputs for the
application parts with test probe P1.
Meas. ParameterMeaning
Mode (Measurement type)
Direct P1Direct measuring method (via test
Permanent connection
P1
socket) with test probe P1
Permanently connected DUT
Single Fault (SFC) – for measrement type direct
P1 only
Normal conditionno single fault simulation active
N interruptedFault simulation active - only
PE interruptedFault simulation active - protective
Nom. voltage on
application part
Polarity –
normal Selection of polarity for line
reverse
1
Connection also via
22Gossen Metrawatt GmbH
phase and protective conductor
are connected with the DUT
conductor is disconnected from
the DUT for the duration of measurement
Fault simulation active - low voltage on application part
for direct, differential measurement
and AT3 adapter only
voltage at test socket
AT3-II I E, AT3-II S, AT3-I I S32, AT16DI/AT32DI
1)
Page 23
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
Select measuring function
1
U
Select parameters
2
Set parameters
3
Connect DUT
4
PE – P1 (with mains)
Probe Voltage
P1
PE – P1
Probe Voltage
P1
Start test
5
Acknowledge line voltage warning
6
!
Only for measurement
type:
(with mains)
Save measured values to clipboard
7
Stop test
8
Save measurements under ID no.
9
9.9U – Probe Voltage
Meas. ParameterMeaning
Mode (Measurement type)
PE – P1Measurement of voltages with
PE – P1 (with mains) Measurement of voltages with
Polarity
normal /
reverse
reference to PE, test socket remains voltage-free, e.g. for permanently connected DUTs
reference to PE, line voltage is applied to the test socket
Selection of polarity for line voltage at test socket
(with mains)“ only)
(with „PE – P1
➭ Connect the DUT to the test socket.
➭ Contact the ungrounded output for pro-
tective extra-low voltage with test probe
P1.
➭ Select line voltage polarity.
Special case: permanently installed DUT
➭ Contact all voltage conducting parts with
test probe P1.
Gossen Metrawatt GmbH23
Page 24
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
Select measuring function
1
U
Select parameters
2
Set parameters
3
Connect DUT
4
V – COM (with mains)
Measuring Voltage
V
COM
!
V – COM
Meas. Volt.
VCOM
Start test
5
Acknowledge line voltage warning
6
!
Only for measurement
type:
(with mains)
Save measured values to clipboard
7
Stop test
8
Save measurements under ID no.
9
9.10U – Measuring Voltage
SECUTEST ST PRO and SECULIFE ST BASE only
and/or Feature I01:
cables for the measurement of dangerous
voltages, please.
Special case: permanently installed DUT
➭ For testing mains power packs and charging
units: Connect the DUT mains terminal
with the test socket.
➭ Connect the DUT output, e.g. for mea-
suring safety extra low voltage (SELV),
with sockets V and COM.
Attention!
Only use the enclosed contact-protected KS17-ONE measurement
Meas. ParameterMeaning
Mode (Measurement type)
V – COMRMS value + AC + DC
V – COM (with mains) RMS value + AC + DC;
for permanently connected DUTs
for line voltage at test socket
Polarity
normal /
reverse
24Gossen Metrawatt GmbH
Selection of polarity for line voltage at test socket
(with mains)“ only)
(with „V – COM
Page 25
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
Select Measuring Function
1
tPRCD
2
Connect DUT
3
P1
Start test (test current 30 mA)
4
Acknowledge line voltage warning
5
!
Execute test
6
Testing is stopped automatically.
7
Save measurements under ID no.
8
9.11tPRCD – PRCD Time to Trip (portable residual current device)
➭ Activate the PRCD.
➭ Contact neutral conductor L at the
PRCD with test probe P1 (if required, ascertain by trial and error).
The PRCD is tripped.
Ascertained time to trip is displayed.
➭ Connect the PRCD to the test socket.
Gossen Metrawatt GmbH25
Page 26
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
Note
Select measuring function
1
P
Select parameters
2
Set parameters
3
Connect DUT
4
Starting test
5
Acknowledge line voltage warning
6
!
Save measured values to clipboard
7
Stop test
8
Save measurements under ID no.
9
9.12P – Functions Test
➭ Connect the DUT to the test socket.
Meas. ParameterMeaning
Polarity
normal Selection of polarity for line
reverse
The following connection types are possible:
• Test socket
• CEE adapter (only for connection via
single-phase CEE or „caravan socket“)
•AT3 adapter (AT3-I IIE, AT3-IIS, AT3-IIS32)
• AT16DI/AT32DI
It is possible to use the adapters listed
above for function testing (initial startup of DUT). Apparent/active power,
power factor and current consumption
can only be measured, however, if the
DUT is directly connected with the test
socket or via the CEE adapter (singlephase CEE socket only).
26Gossen Metrawatt GmbH
voltage at test socket
Page 27
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
This function allows for the evaluation of the
function of the active conductors L(1, 2, 3)
and N of an extension cable. The PE conductor
is not being tested in this process!
For the testing of R
tive single measurements.
XX—
XXX
XXX
PE
and R
see respec-
INS
Connection of EL1 Adapter
➭ Connect the EL1 adapter to the P1
probe sockets at the test instrument.
➭ Connect the plug at the end of the ex-
tension cord to the test socket.
➭ Connect the coupling socket at the end
of the extension cord to the plug at the
EL1 adapter.
Connection of Test Adapters VL2E and AT3-IIIE
➭ Connection examples are shown in sec-
tion 9.17.
Continuity test for L and N
See section 10, “Test Sequences in
Accordance with Standards” (switch
setting A8) with regard to testing extension cords per DIN VDE 0701-0702, for
which RPE and R
Gossen Metrawatt GmbH27
are measured.
INS
Page 28
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
Select measuring function
1
EXTRA
2
Select measuring function
1
EXTRA
Temp. oder IZ
2
9.14EXTRA – Special functions
SECUTEST ST BASE(10)
SECUTEST ST PRO
SECULIFE ST BASE(25)
(Feature I01) &
In this case, additional
measurement functions
are assigned to rotary
switch position EXTRA.
If a QR code is provided:
By scanning the code you can download the
current operating instructions from our
homepage www.gossenmetrawatt.com and
read them with a tablet PC.
28Gossen Metrawatt GmbH
➭ Select the desired measuring function.
Page 29
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
2Connect DUT
3
Start test
4
Save measured values to clipboard
5
Stop test
6
3Connect DUT
4
VCOM
Set parameters
5
Start test
6
Save measured values to clipboard
7
Stop test
8
Temp. – Temperature measurement
Temperature measurement works both with
a Pt100 and a Pt1000 temperature sensor
and automatically recognizes the utilized
sensor type.
IZ – Current clamp measurement
Meas. ParameterMeaning
Measurement type
Gossen Metrawatt GmbH29
V – COMA AC for permanently connected
V – COM (with mains) A AC; for line voltage at test
Polarity –
normal Selection of polarity for line
reverse
Adjust the clamp factor (cl. factor):
➭
DUTs
socket
for direct, differential measurement
and AT3 adapter only
voltage at test socket
– at the current clamp sensor
– at the test instrument
Page 30
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
P1P2
Switch position: RPE
Measurement type: P1-P2
P1P2
Switch position: RINS
Switch position: IT
Measurement type: P1-P2
VCOM
P1
Switch position: RPE
Measurement type: Clamp
V
COM
Switch position: IPE
Measurement type: Clamp
V
COM
Switch position: IE
Measurement type: Clamp
9.152-Pole Measurements with P1 and P2
Test Probes
SECUTEST ST PRO and SECULIFE ST BASE(25) only
and/or Feature H01
In case your DUT is not equipped with a
country-specific mains plug that fits into the
SECUTEST test socket or if it is a permanently installed DUT, the 2
combination with the 1
for 2-pole measurement (dual-lead measurement) of RPE, RINS and equivalent leakage current.
Measurements with test probe 1 against test
probe 2 (P1 – P2) are electrically isolated
from the mains. There is no voltage present
at the test probe.
Connection Example of RPE Measurement
Measuring of protective conductor resistance RPE at permanently installed safety
class I devices
nd
test probe, in
st
test probe, allows
9.16Measurement with Current Clamp Sensor at
Permanently Installed DUTs of Safety Class I
SECUTEST ST PRO and
SECUTEST
ST PRO
Parameter
Tra ns f.
Ratio
1 mV :1 mA
* for WZ12C only
Tra ns fo rm er
1 mV : 1 mA
Connection Example for RPE Measurement (WZ12C only)
Measurement of test current by enclosing
PE in the electric circuit. This type of measurement can only be selected if the test
current has been set at 10 A AC.
Connection Example for IPE Measurement (Direct)
SECULIFE ST BASE(25)
Clamp
Measuring
Ratio
(Switch*)
WZ12C
1 mA ...
Range
15 A
only
SECUTEST ST
PRO
Display
Range with
Clamp
0 mA ... 300 A
Connection Example of RINS or IT Measurements
Measuring of insulation resistance RINS or
touch current IT for safety class I devices
Measurement of protective conductor current by enclosing PE in the feeder.
Connection Example of IE Measurement
(Difference Measurement)
30Gossen Metrawatt GmbH
Measurement of device leakage current by
enclosing conductors
L and N
in the
feeder
.
Page 31
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
Attention!
!
EL1 adapter
P1
VCOM
P1
9.17Measurements with Test Adapter
Test with Adapter
EL1 VL2E AT3-
DUT terminals
Inlet plug 1P+N+PE 16 A
Schuko 1P+N+PE 16 A
CEE 1P+N+PE 16 A
CEE 3P+N+PE 16 A
CEE 3P+N+PE 32 A
5 x 4 mm sockets
—✔✔ — —
—✔ — — —
—✔✔ —✔
—✔✔✔ / —✔
—✔✔— / ✔✔
——— — ✔
Test instrument terminals
Schuko 1P+N+PE 16 A
Socket for test probe
Plug for V–COM
—— ✔✔ —
—✔✔ ——
1
—— ✔——
Active Testing
Protective conductor current IPE
– Direct method
– Differential current
—— ✔✔ —
——✔1✔—
Device leakage current IE
– Direct method
– Differential current
Touch current IT
—— ✔✔ —
——✔1✔—
—— ✔✔ —
Passive Testing
Protective conductor
resistance RPE
Insulation resistance
RINS
Protective conductor
current IPE (equiv. leakage current method)
Extension cables:
apart from RPE & RISO are conducted in switch position
single phase (3-pole)
3-phase (5-pole)
wire short circuit
wire interruption
wire reversal
1
Differential current method with
2
for IPE and IG: AT3-IIS or, alternatively, AT3-I I S32
✔✔ ✔ ✔ ✔
✔✔ ✔ ✔ ✔
—✔✔ ✔ ✔
the following additional measurements
✔✔ ✔ ——
—✔✔ ——
✔✔ ✔ ——
✔✔ ✔ ——
—✔✔ ——
AT1 6D I
2
IIIE
AT3 2D I
SECUTEST ST PRO
CEEAdapter
EL1
only
Connection Example with EL1
Connection Example with VL2E
Connection Example for Protective Conductor
Current Measurement IPE
1
(Direct Method) with
AT3-IIIE
Connection Example for Device Leakage Current
Measurement IG1 (Differential Current Method)
with AT3-IIIE
For information on the correct connection of test adapter and DUT as well as
particular aspects during the test sequence please refer to the operating instructions of the test adapters.
Gossen Metrawatt GmbH31
1
with SECUTEST ST PRO and/or Feature I01
or SECULIFE ST BASE(25) only
Page 32
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
Note
MEM
ESC
0
10Test Sequences in Accordance
with Standards
If the same sequence of single tests will be
run frequently (one after the other with subsequent report generation), for example as
specified in the standards, it’s advisable to
make use of test sequences. Limit values
have been entered for test sequences in
accordance with standards. And thus a go/
no-go evaluation takes place during measurement based on worst-case assessment.
If the momentary measured value is displayed in green, it lies within the limit values
specified in the standard. If the measured
value is red, is does not fulfill the requirements set forth in the standard. If the measured value is shown in amber, further entries
are required. Even if the DUT fails just a single test step, the test sequence is aborted
and testing in accordance with the selected
standard is failed.
With the help of the SECUTEST DB COMFORT feature (feature KD01 or
expansion Z853S), test sequences
can be modified such that they’re not
interrupted in the event of a limit value
violation. Amongst other things, this
is helpful for troubleshooting during
repair. Please refer to the operating
instructions for further information.
10.1General Procedure
1 Select the desired test sequence with
the rotary switch (A1 ... A9).
2 If no test object has been selected, enter
the ID number of the test object, for
example by means of a barcode scanner, after selecting ID.
3 As an alternative to step 2, activate
the database view with the MEM key:
4 Select the test object with the scroll
keys.
5 Return to the measuring view by
pressing the ESC key.
6 Start the test sequence with the
START/STOP key.
7 The measured value recording
symbol shown at the right appears.
Each time this key is pressed, the measuring or evaluation procedure is
restarted (see case B in section 10.2).
8 Proceed to the next measurement by
pressing the key shown to the right.
9 At the end of the test sequence,
you can generate a list of the
results of the individual test steps.
10 If you want to view details such as
the settings for the individual test
steps, select the desired measurement
with the cursor and press the + magnify-ing glass key.
11 The display is returned to the list of
test steps by pressing the – magnify-ing glass key.
12 Save the results of a successful test
sequence by pressing the Save key.
10.2Evaluation Procedure
The evaluation procedure can be started manually for some test steps within a given test
sequence, but all others are run automatically.
• Case A – automatic triggering of evaluation:
Evaluation (with a duration of, for example,
5 seconds) is started automatically as
soon as the measured value has stabilized. The worst value which occurs during
this duration is saved, and automatic
switching to the next test step ensues.
•
Case B – manual triggering of evaluation:
Evaluation is started after pressing
the measurement value recording symbol (display: 0). After a specified period
of time has elapsed, the worst value is
saved to the right of wc: (worst case),
and the number 1 is displayed in the
measurement value recording symbol
indicating that the first measured value
has been saved. Pressing the measured
value recording icon again restarts the
evaluation procedure. If the worst value
is worse that the value obtained for the
32Gossen Metrawatt GmbH
Page 33
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
Select test sequence
1
A2
Open database
2
MEM
Select Device
3
Switch to start view
4
ESC
ab
c
previous measurement, the new value is
used. However, if this value is better than
the previous worst value, the original
value remains in the display. Depending
on whether you want to delete the last
value saved to the clipboard or all values,
press the symbol shown below an
appropriate number of times.
Proceed to the next test by pressing
the symbol shown to the right.
10.3Sample Test Sequence
As-delivered Condition:
A1
VDE 0701-0702, passive meas. type, test socket
A2
VDE 0701-0702, active meas. type, test socket
A3
VDE 0701-0702, param. config. for EDP (active)
A4
IEC 62353 (VDE 0751), passive measement type
A5
IEC 62353 (VDE 0751), active measurement type
A6
IEC 60974-4, connection type: test socket
A7
IEC 60974-4
A8
VDE 0701-0702
A9
VDE 0701-0702,
type, protection category – in each case automatic
, connec. type: AT16-DI/AT32-DI adapter
, MT ExtCord, EL1/VL2E/AT3-IIIE adapter
connection type, measurement
Gossen Metrawatt GmbH33
Page 34
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
Sequence parameters
5
Set classification parameters
6
a
b
c
b
aca
b
Connect DUT
7
Check connection &start test sequence
8
Individual test steps can be configured with the sequence
parameters
, see detailed operating instructions.
A1 ... A9 switch setting
Meas. ParameterMeaning
StandardTest standard / extension cord
Protection class * PC1/PC2/PC3
Connection type * Test socket / permanent / adapter
Measurement type
(MT) *
Detected classifica-
tion
Auto-detection ofAny desired combinations for au-
* If the configurations of the classification parameters
are recognized automatically, they are marked with
an amber frame (in this case and ). They must
be entered manually if they’re not automatically
detected, or if they’re detected incorrectly.
Active or passive DUIT
(on test: on = passive, off = active)
No auto-detection:
tion parameters such as connection, protection class and measurement type must be entered
manually.
Always accept:
parameters activated under “Autodetection of” are detected automatically and accepted.
tomatic detection of:
– Connection
– Protection class (SK)
– Measurement type (MA)
all classifica-
all classification
➭ Connect the DUT to the test instrument
in accordance with the selected test sequence.
– Test socket
– Permanent connection
– Adapter
Switch position: A1 ... A7, A9
Connection depends on the type of DUT.
Switch position A8
For testing extension cords in accordance
with standards: connection to the test
socket via the following adapter:
– EL1:
–
for single-phase extension cords
VL2E/AT3-IIIE: for single- or three-phase
extension cords
The following checks are run automatically before the test sequence
is started:
• Probe test P1 (whether or not the probe
is connected and if fuse P1 is OK)
• Insulation test (whether or not the DUT is
set up in a well-insulated fashion)
• On test and short-circuit test.
In order to be able to detect a short-circuit at the DUT, testing is conducted
between L and N, as well as LN and PE.
If you’ve set the “
Detected classification
parameter for the respective test
sequence to “Always accept” and the
“
Auto-detection of
” parameter to “Connection and SK” (before triggering
following additional checks will be run before
the test sequence is started:
•Protection class detection for DUTs with
protective conductor
• Connection check: whether or not the
DUT is connected to the test socket. In
the case of protection class I: whether or
not the two protective conductor terminals are short-circuited.
34Gossen Metrawatt GmbH
Start
”
), the
Page 35
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
Note
Manual evaluation of visual inspection
9
Test step – start evaluation
10
Test step – automatic evaluation
11
Manual evaluation of functions test
12
Visual inspection passed
Visual inspection not passed
(sequence is aborted, test failed)
Resume test sequence
If the plug is disconnected from the test
socket during the test sequence, the
sequence is immediately disconnected.
Green measured value:
complies with standard
Record measuring point.
Delete last measuring point.
The measured value is ascertained automatically within a specified period of time. The
test sequence is then automatically
resumed.
Green measured value:
complies with standard
Functions test passed
Functions test not passed
(sequence is aborted, test failed)
Resume test sequence
➭ Remove DUT from service
Resume test sequence.
Gossen Metrawatt GmbH35
Page 36
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
End of sequence – display results
13
13
Display detailed results
14
Hide details
15
Confirm results
16
Save results under ID no.
17
Optional test step
(display of the memory screen depends on the
parameter pre-selection in the SETUP switch position:
Setup 1/3 > Auto. measurements > At end of
sequence > Memory screen. If set to events list,
is omitted.)
Optional test step
Optional test step
Switch to memory screen
(consideration of measuring error depends on
the parameter pre-selection for the SETUP switch
setting: Setup 1/3 > Auto. measurements >
Error considered. > Yes)
Optional test step
or with feature KD01 „Z853S – SECUTEST DB
COMFORT“:
Send measurement data to the PC,
via USB or Bluetooth® (Feature M01),
e. g. for saving to the IZYTRONIQ
report generating software (pushprint function), description see online
help for IZYTRONIQ
36Gossen Metrawatt GmbH
Save results
Page 37
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
11Measurement Parameters for
Individual Measurements and
Test Seq u en c es
Measurement parameters which apply for
both individual measurements and test
sequences, have to be entered in selector
switch position SETUP.
Setup 1/3 > All measurements
Meas. ParameterMeaning
Meas. at IT-mains
(Yes / No)
Ref.voltage L-PE
(110 V, 115 V, 220 V,
230 V, 240 V)
Testingfreq Alt
(48 Hz ... 400 Hz)
Yes : active leakage current mea-
surements (and/or all measurements which include the PE at the
mains connection end) are
blocked. Test sequences which
contain such kind of measurements are disabled as well.
The reference (line) voltage is the
voltage to which the measured
values for leakage current have
been standardized.
It is used in the case of leakage
current for mathematical adjustment of the current measuring
values to the predefined voltage.
Measurements with line voltage at the test socket:
The setting value has no influence
on the voltage with which the DUT
is supplied via the SECUTEST test
socket.
Leakage current measurements with measurement type
„Alternative“: The setpoint value
of the synthetical test voltage is
derived from the value specified
here.
Variable frequency setpoint value
for synthetical test voltage for all
leakage current measurements of
measurement type „Alternative“,
affecting the following measurements and/or rotary switch positions:
– Individual measurements
(green rotary switch level)
– Measurements in test se-
quences predefined ex factory
– Measurements in user-defined
test sequences
(only with database expansion
Z853R or feature KB01)
Gossen Metrawatt GmbH37
Page 38
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
12Repair and Replacement Parts Service
Calibration Center and Rental Instrument Service
If required please contact:
GMC-I Service GmbH
Service Center
Beuthener Straße 41
90471 Nürnberg, Germany
Phone: +49 911 817718-0
Fax:+49 911 817718-253
e-mail service@gossenmetrawatt.com
www.gmci-service.com
This address is only valid in Germany. Please
contact our representatives or subsidiaries
for service in other countries.
SECUTEST ST BASE(10) / PROund SECULIFE ST BASE(25)
14IZYTRONIQ Data Processing
Software
IZYTRONIQ is a database test software that
has been newly developed from scratch. It
enables the user to visualize and manage
the entire testing procedure for all our test
instruments and to document it in an auditproof manner. For the first time, it is thus
possible to combine the test and measurement data from a great variety of test instruments and multimeters in one test and generate one report report thereof. The intuitive
user guidance and modern design provide
for quick access to all functions.
The software is available in different sizes
and versions for trades, industry and vocational training purposes.
Depending on which instrument variant you
order, IZYTRONIQ is included in the scope
of delivery – for example with standard models and instrument sets (see section 2). If this
is not the case or if you would like to take
advantage of a variant with a larger scope of
functions, you can purchase IZYTRONIQ
separately. Detailed information can be
found on our website: