Gossen MetraWatt SECUTEST SII Operating Instructions Manual

Operating Instructions
SECUTEST®SII
Test Instrument for DIN VDE 0701, 0702 and 0751
3-349-223-03
8/11.05
SONDERS232 5 4
SECUTEST SI I
45678910
11
12
13
14 15
FUNCTION-TEST
230 V
MENU
OFF
VDE 0751
VDE 0701
0702
T. 240
T. 200
T. 1
VDE
N
L
L
N
SL
3
2
1
1
2
3
Feature F01
19
20
18
17
23
18
22
21
16
These operating instructions describe an instrument with firmware version 6xx.
Note
Contact problems with exposed conductive parts when using the standard probe with test tip
In order to assure good contact, surface coatings must be removed from devices under test with special tools at a suitable location so that the surface has a metallic shine. The tip of the test probe is not suitable for scratching away paint, because this may impair its coating and/or mechanical strength. The brush probe may be more suitable than the test probe in certain individual cases.
GOSSEN METRAWATT GMBH 3
Upper left figure
1 Jack for protective conductor at device under test (feature F01) 2 Jack for neutral conductor at device under test (feature F01) 3 Jack for phase conductor at device under test (feature F01) 4 Jack for connecting the probe 5 Jack for connecting the probe 6 Function selector switch
– VDE-: Automatic test sequence according to selected standard – OFF: Instrument is switched off (no disconnection from mains) – MENU: Setup: Device configuration
R
SL
: Individual tests
– FUNCTION TEST: Function test
7 scroll key for menu and parameter selection 8
scroll key for menu and parameter selection
9 LCD panel 10 Socket connector for RS 232 interface 11 Signal lamp for mains connection error 12
key for entry, for starting test sequences and for finger contact
13
help key (context sensitive)
14 Key next to the symbol for switching test voltage to
the test socket (only possible if symbol LED is blinking)
15 Signal lamp for the functions text
Lower Left figure
16 Push-buttons (left and right) for releasing the handle from its snap-in position 17 Standard outlet socket (test socket) for connecting the device under test 18 Push-buttons (left and right) for releasing the lid 19 Compartment for probe and accessories 20 Lid 21 Test probe 22 Carrying handle and tilt stand 23 Cover or printer module (feature E01)
Overview of Available Probe Types
1)
Accessory
Note
Use of probes other than those specified above:
Cables plugged into jacks 4 and 5 must be short-circuited in order to perform the probe test, either by plugging the cable ends together or by means of a conductive surface at the device under test (4-wire measurement). Remove any corrosion from the device under test.
Data Security
Measurement data, report data and user entries are stored to RAM at the SECUTEST
®
PSI printer module (feature E01), as long as the respective
battery supplies the required amount of voltage. Save your data to a PC on a regular basis in order to prevent any loss of
data at the printer module. We assume no liability for data loss.
We recommend the following PC programs for data processing and data management:
PS3 (transmission of measurement data to a PC, documentation, management, report generating and deadline follow-up)
SECU 601 (program for activating the data base function in the instrument (DBmed option), if not already available as a feature)
PC.doc-WORD (report and list generation
PC.doc-ACCESS (test data management)
i
Probe Type Application Special Features
Standard probe (test probe with coil-cable and alligator clip)
Text current: 200 mA
none
SK2
1)
Test current: 200 mA
Probe with plain cable (not coil-cable), 2meters long
SK5 option (feature KD01 or accessory)
Special probe in combination with “automatic recognition of measuring point change” function (see
chapter 13).
Brush probe
1)
Can be plugged onto all above listed probes and test probes
Leakage current, protective conduc­tor resistance
For contacting devices under test with rotating, vibrating, exposed conductive parts
!
Contents Page Contents Page
1 Applications ...................................................................................6
1.1 Table: Types of DUTs – Tests – Regulations ....................................................6
1.2 Table: Individual Measurements and Regulations ............................................6
1.3 Table: Leakage Current Types ........................................................................7
1.4 List of Possible Options for SECUTEST
®
SII Device Series .................................7
2 Safety Features and Precautions ...................................................8
3 Initial Start-Up ................................................................................9
3.1 Connection to the Mains (230 V, 50 Hz) ..........................................................9
3.2 Automatic Recognition of Mains Connection Errors ........................................10
4 General Notes ..............................................................................10
4.1 Online Instructions .......................................................................................10
4.1.1 Automatic Safety Class Selection ..................................................................10
4.1.2 Manual or Automatic Operating Sequences ................................................... 10
4.2 Online Help .................................................................................................10
4.3 Adjusting Contrast .......................................................................................11
4.4 Configuring Device Parameters, Setting Date and Time ..................................11
4.5 Configuring Measurement and Sequence Parameters .................................... 11
4.6 Setting Limit Values ..................................................................................... 11
4.7 Saving the Settings ......................................................................................11
5 Classification of Devices Under Test ............................................12
5.1 Safety Classes ............................................................................................12
5.2 Application Parts (electrical medical devices) .................................................12
6 Abbreviations ...............................................................................13
7 Connecting the Device Under Test ...............................................14
8 Configuring Device Parameters ...................................................15
9 Individual Measurements ............................................................16
9.1 Measuring Protective Conductor Resistance ................................................. 16
9.2 Insulation Resistance R
ISO
. .......................................................................... 17
9.3 Measuring Leakage Current ......................................................................... 19
9.3.1 Contact Current I
Contact
(probe current, housing leakage current) ...................19
9.3.2 Residual Current ΔI ..................................................................................... 19
9.3.3 Device Leakage Current I
DL
per DIN VDE 0751 .............................................19
9.4 Equivalent Leakage Current ......................................................................... 20
9.5 Probe Voltage U
probe
– Max. 300 V .............................................................. 21
9.6 Alternating / Direct Voltage U
AC/DC
– Max. 253 V (feature F01) ...................... 22
9.7 Resistance R (feature F01) ........................................................................... 22
9.8 Measurements with Accessories (feature F01) .............................................. 22
9.8.1 Alternating Current I
Z
with Clip-On Transformer (feature F01) ........................22
9.8.2 Temperature T with Pt100/1000 Sensor (feature F01) .................................. 23
10 Function Test ...............................................................................24
11 Measurements – Selector Switch Settings for Standards ...........26
11.1 Test Sequences .......................................................................................... 26
11.2 Setting Up Test Sequences .......................................................................... 27
11.3 Configuring Measuring Parameters ..............................................................27
11.4 Testing Devices in Accordance with DIN VDE 0701, Parts 1 and 200 ............. 28
11.5 Testing Devices in Accordance with DIN VDE 0701, Part 240 ........................ 30
11.6 Testing Devices in Accordance with DIN VDE 0702:2004 .............................. 32
11.7 Testing Extension Cables for VDE 0701, Part 1, and VDE 0702
(option: EL1 adapter) ................................................................................... 34
11.8 Testing Multiple Outlets for VDE 0702 (optional EL1 adapter) ......................... 35
11.9 Testing in Accordance with DIN VDE 0751 .................................................... 36
GOSSEN METRAWATT GMBH 5
Contents Page Contents Page
12 Storage in PSI Module (Feature E01)
and Database Operations (Feature KB01 or DBmed option) ........ 38
12.1 Storing Measurement Data in the PSI Module ................................................38
12.2 Database Operations ...................................................................................38
12.2.1 Setting Up Test Sequences with PC Software ................................................38
12.2.2 Storing Test Results to the SECUTEST
SII ......................................................38
13 Remote Control (Feature KD01 or SK5 Option) ............................ 38
14 Printing via PSI Module (Feature E01)
or with the Printer Adapter (DA-II accessory) ............................. 39
14.1 Print Out Test Data at the PSI Module ...........................................................39
14.2 Printing Test Results in Report Form .............................................................39
15 Direct Print-Out (feature KE01, SECU-dd option) ......................... 39
16 Characteristic Values .................................................................. 40
17 RS 232 Interface .......................................................................... 43
17.1 Transmission of Measurement Results to the SECUTEST®PSI ........................43
17.2 PC Connection ............................................................................................43
17.2.1 Software Evaluation of Measurement Results ................................................43
17.2.2 Instrument Control via Interface Commands ..................................................43
17.3 Interface Configuration and Protocol .............................................................43
18 Appendix ......................................................................................44
18.1 Evaluation of Measured Values for Individual Measurements
as well as for Calculated Quantities ..............................................................44
18.2 Index ..........................................................................................................45
19 Maintenance - Calibration ........................................................... 46
20 Repair and Replacement Parts Service
DKD Calibration Laboratory and Rental Instrument Service ........ 46
21 Product Support .......................................................................... 47
Condensed instructions for PS3 software for maintenance and equipment management are provided with the PS3 CD ROM.
They include important instructions regarding the following topics:
• PS3 demo with additional instructions
• Special features for reading measured values in from a data file
1 Applications
1.1 Table: Types of DUTs – Tests – Regulations
Attention!
!
The test instrument may not be used for measurements within electrical systems!
1.2 Table: Individual Measurements and Regulations
Key
Required test
Start-up and Modifi­cations
Testing after Repairs
Periodic Tes ti ng
Devices under test to be tested in accordance with the following regulations
DIN VDE 0751:2001
DIN VDE 0701 Part 1:2000
DIN VDE 0701 Part 240
DIN VDE 0751:2001
DIN VDE 0702:2004
DIN VDE 0751:2001
Laboratory instruments ••
Measuring and control instruments
••
Voltage generation devices
••
Electric tools ••
Electric heating devices ••
Electric motor devices ••
Lamps ••
Devices for entertain­ment electronics, information and commu­nications technology
••
Cable reels, extension and connection leads
••
Data processing and office equipment
••
Electrical medical de­vices, application parts
••
Individual Measurements per Regulation
Test current [A]
DIN VDE 0701,
part 1
DIN VDE 0701,
part 200
DIN VDE 0701,
part 240
DIN VDE 0702
DIN VDE 0751
Protective conductor resistance
0.2 ••••
Insulation resistance
••
Equivalent leakage current
••••
Equivalent (device) leakage current
Residual current
••
Contact current
••
Absence of voltage (exposed conductive parts)
Patient leakage current
Device leakage current
GOSSEN METRAWATT GMBH 7
1.3 Table: Leakage Current Types
Key
NC = Normal conditions PAT = Patient application part PE = Potential earthing
, mains protective conductor
DPE = Protective conductor of the device under test
1.4 List of Possible Options for SECUTEST®SII Device Series
Example for complete type designation (article number) of a SECUTEST
®
SII: M7030 E01 KD01 KE01 (only the designation of the basic device, M7030, and features other than 00 are specified)
1)
Each measured value is documented in this case, as opposed to the results of a test sequence for which the poorest value for each given test is displayed.
2)
Adapter set for international use (feature B00 - earthing contact socket - included)
3)
With configuration for hospital beds of Safety Class 2 (pre-adjusted)
4)
For multimeter functions, for measurements with accessories and for devices under test without plug, including 3 plug-on quick clips
DIN VDE 0701-1
DIN VDE 0702-1
DIN VDE 0751-1 (2001)
The following is measured:
Equivalent leakage current
Equivalent leakage current
PROBE (connected to protective conductor) to L & N
Equivalent device leakage current
PROBE (open protective conductor) to L & N
Contact current / mea­surement for absence of voltage
Contact current / mea­surement for absence of voltage
Probe to PE
Protective con­ductor to PE
Device leak­age current during opera­tion
Protective con­ductor inter­rupted, probe to PE
Protective conductor current with differential current measurement
Protective conductor current with differential current measurement
Device leak­age current during opera­tion, differential cur­rent method
See chapter
9.3.2
Features
00 01 02 03 04 05 06 07 08 09 10 11 99
Design AGM
Mains Connection for Respective Country
BD
F CH
2)
User Interface Language
C
DGB
Configuration (settings in the setup menu)
D
GM
3)
Printer Module SECUTEST
®
PSI
E
w/o with
Additional Test Sockets (cannot be retrofitted)
F
w/o
with
4)
Database DBmed Option (Z853H)
KB
w/o with
Remote Control SK5 Special Cable (Z745K)
KD
w/o with
Direct Printing After Each Measurement for Automatic Test Sequences
1)
Read-Out via
RS 232
, SECU-dd Option (Z853L)
KE
w/o with
DKD calibration certificate
L
w/o with
2 Safety Features and Precautions
The SECUTEST®SII test instrument has been manufactured and tested in accordance with the following safety regulations:
IEC 61010-1 / DIN EN 61010-1 / VDE 0411-1 and DIN VDE 0404
,
DIN VDE 0413 Part 2 and 4
.
When used for its intended purpose, the safety of the user, the test instrument and the device under test (electrical equipment or electrical medical device) is assured.
Read the operating instructions carefully and completely before placing your test instrument into service. Follow all instructions contained therein. Make sure that the operating instructions are available to all users of the instrument.
Tests may only be performed by a qualified electrician or under the supervision and direction of a qualified electrician. The user must be instructed by a qualified electrician in the execution and evaluation of tests.
Note
Manufacturers and importers of electrical medial devices must provide documentation for the performance of maintenance by trained personnel.
Observe the following safety precautions:
• The instrument may only be connected to electrical systems with a maximum of 230 V which are protected with a fuse or circuit breaker with a maximum rating of 16 A.
• Measurements within electrical systems are prohibited.
• Be prepared for the occurrence of unexpected voltages at devices under test (for example, capacitors can be dangerously charged).
• Make certain that the measurement cables are in flawless condition, e.g. no damage to insulation, no interruptions in cables or plugs etc.
• Grip the test probe firmly, for example during insertion into a jack socket. Tensioning at the coil cord may otherwise cause the test probe to snap back resulting in possible injury.
Measuring of insulation resistance (equivalent leakage current) The test is conducted with up to 500 V, with a current limit of I < 10 mA. However, when touching the terminals (L or N), you will get an electric shock which, in turn, may lead to accidents. Feature F01: When connecting the DUT via jacks 1 to 3, special care must be taken to avoid touching the open cables.
Leakage current measurement During leakage current measurement it is essential to ensure that the DUT is operated at mains voltage. Exposed conductive parts may be subject to hazardous contact voltage during the test and may not be touched under any cricumstances. (A mains shutdown takes place when the leakage current exceeds approx. 10 mA).
Attention!
!
The function test may only be performed after the DUT has successfully passed the safety test!
The measuring and test instrument may not be used:
• If it demonstrates visible damage
• With damaged connector cables, measuring cables or patient ports
• If it no longer functions properly
• After extraordinary stresses due to transport
In such cases, the instrument must be removed from operation and secured against unintentional use.
Meanings of Symbols on the Instrument
The symbols on the instrument have the following meanings:
Warning regarding dangerous electrical voltage
Warning concerning a point of danger (attention: observe documentation!)
Test socket
VDE testing authority approval mark
This device may not be disposed with the trash. For further details on the WEEE marking, please refer to our website www.gossenmetrawatt.de and enter search key ’WEEE’.
!
GOSSEN METRAWATT GMBH 9
3 Initial Start-Up
3.1 Connection to the Mains (230 V, 50 Hz)
Connect the mains plug at the test instrument to the mains power
outlet. The function selector switch can be set to any position. If a mains outlet (earthing contact outlet) is not available, or if only a 3-phase outlet is available, the adapter socket can be used to connect the phase conductor, the neutral conductor and the protective conductor. The adapter socket has three permanently attached cables and is included with the KS13 cable set
Attention!
!
If connection is not possible via an earthing contact outlet: Shut down mains power first. Then connect the cables from the coupling socket to the mains using pick-off clips in accordance with the diagram. Disconnection from mains power is only possible with the mains plug.
L1
N
green-yellow
green-yellow
PE
L1
L2
L3
N
PE
L1
L2
L3
N
green-yellow
U
L–N
= 230 V
Mains Connection
3.2 Automatic Recognition of Mains Connection Errors
The device automatically recognizes mains connection errors if the condi­tions in the following table have been fulfilled. The user is informed of the type of error, and all measuring functions are disabled in the event of danger.
1)
In SETUP – test sequence – IT system
Attention!
!
In either of the first two cases listed in the table above, immediately disconnect the test instrument from the mains and eliminate the error!
Note
Voltage at the electrical system’s protective conductor PE may result in distorted measurement values during testing for the absence of voltage, or during leakage voltage measurements.
4 General Notes
4.1 Online Instructions
The performance of measurements and tests is quick and easy with the SECUTEST
®
SII. Integrated online instructions inform the user regarding all required connections, necessary work steps, operator errors, measurement results and more in all measuring modes. Information and test results appear at the dot matrix LCD in plain text.
4.1.1 Automatic Safety Class Selection
Depending upon the type of mains plug or the connection configuration for the device under test, the test instrument recognizes the respective safety class and recommends its use for the measurement to be performed.
4.1.2 Manual or Automatic Operating Sequences
Depending upon selections made in the setup menu (selector switch in the Auto position), the next measurement is started automatically after the current measurement has been completed, or can only be started after manual acknowledgement. The integrated online instructions are adequate for most tests and measurements. However, the contents of these operating instructions should nevertheless be read and observed.
4.2 Online Help
Online help can be queried and displayed at the LCD for all measuring and test functions, and for almost all settings. Schematic diagrams which illustrate proper connection of devices under test to the SECUTEST
®
SII
can be displayed as well.
Press the following key in order to query online help:
Press the same key again in order to exit online help.
Note
Online help can be queried during measurement by pressing and holding the help key.
Typ e o f Mains
Connection Error
Message Condition Measurements
Voltage at protective
conductor PE to finger
contact
(key)
Text appears at
LCD
Press
key
U > 40 V
disabled
Protective conductor PE
and phase conductor L
reversed and/or
neutral conductor N
interrupted
lamp
lights up
Voltage at PE
>65V
impossible
(no supply power)
Contact voltage at
protective conductor PE
to neutral conductor N
or phase conductor L
Text appears at
LCD
U > 25 V
disabled, although
disabling can be
deactivated
1)
Mains voltage
too low lamp
lights up
U
L-N
< 90/180 V
possible
under certain
circumstances
!
!
GOSSEN METRAWATT GMBH 11
4.3 Adjusting Contrast
Set the selector switch to MENU.
Select the Setup menu, “back” is highlighted.
Activate contrast adjustment.
Press and hold the key.
Adjust contrast.
Return to the menu.
Store the contrast setting to permanent memory with the save function in the setup menu.
4.4 Configuring Device Parameters, Setting Date and Time
Device parameters and functions which are valid for all selector switch positions can be activated or deactivated with the selector switch in the
MENU position using the Setup parameter (see chapter 8 on page 15).
4.5 Configuring Measurement and Sequence Parameters
Measurement and sequence parameters, as well as functions, can be activated or deactivated in the Setup menu (selector switch in the VDE position) for the respective test regulation. Refer to chapter 11.3 on page 27 for the significance of the various parameters.
4.6 Setting Limit Values
Upon delivery, the limit values set forth (at the point in time of issue) in applicable national and international standards are stored to the test in­strument. Limit values for each of the respective standards can be queried and changed if required with the Setup > Limit Values menu (selector switch in the VDE... position), but changes can only be made which result in even stricter testing than is required by the respective standard.
Newly entered limit values become effective immediately. However, these are only stored to memory permanently after activating the Store function in the Setup menu for the respective standard.
If the limit values set forth in the standards for certain safety classes need to be restored despite individualized settings, the All Values per Standard menu function in the Limit Values submenu must be selected and acknowledged with the Enter key.
If the limit values set forth in the standards are changed, the instrument’s device software can be updated via the RS 232 interface.
4.7 Saving the Settings All of the settings and changes which have been entered to the Sequence,
Limit Values (selector switch in the VDE... position) and Zero Point (tempera- ture measurement) (selector switch in the MENU position) menus, as well as
the selected contrast setting are retained until the selector switch is turned, or the test instrument is disconnected from mains power. If settings and changes should be retained even after mains power has been interrupted, they must be saved in the Setup menu for the respective test regulation or selector switch position (acknowledge the “Save” function).
MENU
5 Classification of Devices Under Test
5.1 Safety Classes
Devices assigned to all of the following safety classes are equipped with basic insulation, and provide for protection against electrical shock by means of various additional precautions as well.
Safety Class I Devices
Exposed, conductive parts are connected to the protective conductor so that they are not charged with voltage if the basic insulation should fail.
Safety Class II Devices
These devices are equipped with double insulation or reinforced insulation.
Safety Class III Devices
These devices are powered with safety extra-low voltage (SELV). Beyond this, no voltages are generated which exceed SELV. These devices may not be connected to the mains.
Note: The DUT may only be connected to the test instrument at jacks 1 through 3 (Feature F01). Only a visual inspection, an insulation resistance measurement or a measurement of the supply voltage can be performed, see parameter “SC III U
V
” on page 27.
Parameter Classification (in the Sequence... menu)
The SECUTEST
®
SII always tests in accordance with the strictest limit val­ues of the respectively selected safety class. The test is failed if this limit value is exceeded. However, devices under test exist for which higher limit values are allowable.
If parameter classification is activated (= x), the user is asked if higher limit values are allowable for the DUT. If “Yes” is entered, reevaluation ensues and the DUT may pass the test.
Examples
If the insulation resistance test is failed, or if interference suppression capacitors have been replaced, equivalent leakage current measurement must be performed on DUTs with heating elements in accordance with DIN VDE 0702.
The test is failed by a DUT with 300 kW with deactivated classification function (= _), but it is passed with activated classification function (= x) if the question is answered accordingly.
Similar possibilities for passing the test, although it might otherwise be failed, are included in DIN VDE 0701, part 1.
If the DUT is connected via the jacks instead of the test socket, other limit values apply because higher power consumption is possible in this case (e.g. part 1 specifies a limit value of 1 mA per kW for equivalent leakage current).
See also table: “Maximum Allowable Limit Values for Equivalent
Leakage Current in mA” on page 20.
5.2 Application Parts (electrical medical devices)
Type B Application Parts (body)
Devices of this type are suitable for both internal and external patient applications, except for use in direct proximity to the heart.
These devices provide for adequate protection against shock especially as regards:
• Reliable leakage current
• Reliable protective conductor connection if utilized Devices of the following safety classes are allowable: I or II.
Type BF Application Parts (body float)
Same as type B, but with type F insulated application parts.
Type CF Application Parts (cardiac float)
Devices of this type are suitable for use directly at the heart. The application part may not be grounded.
Devices of the following safety classes are allowable: I or II.
GOSSEN METRAWATT GMBH 13
6 Abbreviations
AE Error condition: application part grounded B, BF, CF Classifications for application parts OE Operational earth
ΔI Differential current, residual current ΔI
max
Maximum residual current (during function test)
EDL
A1/A2
Equivalent device leakage current with designation A1/A2 (cross-reference in the standard)
EDL
PX±PE
Equivalent device leakage current for portable x-ray +PE: with additional PE conductor –PE: without additional PE conductor
EDL
SCII
Equivalent device leakage current for devices with
additional safety class 2 components HE Error condition: housing grounded I
L
, I , IHL, I
probe
Leakage current (differential, probe or contact current) I
Re
Residual current I
EL
, I
EDL
Equivalent device leakage current I
PE
Earth leakage current (current at protective conductor) IT system The IT system has no direct contact between active
conductors and grounded parts: bodies within the
electrical system are grounded. I
Lmax
Maximum load current (during function test) I
Z
Current at clip-on meter L Phase conductor PF Power factor MedGV German medical device ordinance MPG German medial product law MSELV Medical safety extra-low voltage N Neutral conductor NC Normal conditions P Active power PA Functional earth (equipotential bonding)
R Resistance (only with feature F01) R
ISO,
R-ISO Insulation resistance R-ISO APP-PE Insulation resistance: application part to PE R-ISO
INT. CARD.
Insulation resistance: intracardiac
(application at the heart) R-ISO NL-PE Insulation resistance: neutral/phase conductor to PE R
PE
, R-PE Protective conductor resistance
R-SL±Mains Protective conductor resistance limit value for
+mains: DUT with mains power cable
–mains: DUT without mains power cable
(protective conductor resistance limit value for mains
power cable only = 0.1 Ω) S Apparent power SELV Safety extra-low voltage SFC Single-fault condition PE Protective conductor U
AC/DC
AC/DC voltage (only with feature F01) U
REF
Reference voltage: voltage to which all leakage voltages
make reference (as a rule nominal line voltage) U-ISO Test voltage for insulation measurement U
L-N
Line voltage U
MEAS
Voltage with which testing was executed.
Displayed for all leakage current measurements. U
probe
Probe voltage tOn-time Temp Resistance temperature (only with feature F01) W Electrical energy ZVEH General Association of German Electricians
7 Connecting the Device Under Test
Connect the DUT in accordance with the schematic diagrams included
in the online help function.
Connection of the DUT to the test instrument depends upon:
• Type of connection: – With plug (“to test socket” parameter), applies to EL1 adapter as well – Without plug, single or multi-phase connection
(“to jacks” parameter), only with feature F01
– No connection to tester (“permanent connection” parameter)
• Whether or not an adapter is used: – Adapter to socket (customer specific adapter) – AT3-med to socket, adapter for devices
which are equipped with 5-pole, 16 A CEE plugs
– AT3-III to socket, adapter for devices
which are equipped 5-pole, 32 A CEE plugs, see AT3-III operating instructions for test sequence.
• The DUT’s safety class (I, II or I I I)
Note
The DUT must be switched on for all tests. Switches, relays, temperature regulators etc. must all be taken into consideration.
The test instrument automatically recognizes whether or not the DUT is connected to jacks 1 through 3. If the instrument is equipped with feature B00, 01 or 03, the instrument also recognizes whether or not a DUT has been connected to the test socket. As a default setting, the program sequence assumes that the plug from the DUT has been connected to the test socket.
Note
Safety Class II Devices with Safety Class I Mains Plugs
If the device under test is equipped with a safety class I plug although it complies with safety class II, safety class I is recog­nized by the test instrument. If this is the case, switch from safety class I to safety class II in the initial menu.
If the test instrument is unable to automatically recognize how the DUT has been connected, the recommended connection setup should be double checked and determined manually if necessary.
Position the
cursor at the second line in the start menu for the test
sequence.
A selection of possible connection setups can be displayed by activat-
ing the
key.
Select the desired connection setup with the
cursor and
acknowledge with the
key.
Omitting the Protective Conductor Test in the Case of Fully Insulated Devices
You are testing a fully insulated safety class I device (e. g. screen, sub­mersible pump, etc.), which is not equipped with an external protective conductor contact.
The decision as to the necessity of a protective conductor test in this case is to be taken by a qualified electrician who should also assume responsi­bility.
You can omit the protective conductor test by pressing the key
as
soon as the following instruction is shown: „Please connect the probe with the protective conductor of the DUT“.
Protective Conductor and Insulation Resistance Measurements for Permanently Installed Devices Under Test
Attention!
!
Deactivate the electrical system which supplies power to the device under test before connecting the test instrument!
Remove the mains fuses from the device under test and disconnect
the neutral conductor N inside the device under test.
Measuring Contact Current (absence of voltage)
Make sure that the contacted parts are not grounded.
GOSSEN METRAWATT GMBH 15
8 Configuring Device Parameters
General device parameters can be configured and saved using Setup with the selector switch in the MENU posi­tion.
Select Setup/Menu and acknowledge.
Select parameter and acknowledge, change setting and acknowledge. Settings x / – = function activated / deactivated
Limit values ...
Illumination Background illumination for the LCD. One of three
different conditions can be selected with the up and down scroll keys: x: continuously on, –: off Numbers 1 through 9: duration in minutes after
which illumination is automatically deactivated. Test time Duration of a single test (0 255 s) Reference voltage: Voltage to which leakage current makes reference
(as a rule nominal line voltage) Earth fault During the short-circuit test, testing is also performed
to determine whether or not a connection exists
between L/N and PE (short-circuit to exposed
conductive part). We assume that a short-circuit to
an exposed conductive part exists in the event of
leakage current greater than 15 mA from L/N to PE.
This value should be increased for some DUTs (in
particular high-current consumers), because greater
leakage currents are present.
Mains wait Line voltage is initially applied to the test socket.
However, testing does not begin until after the number of seconds selected in “Mains wait” has elapsed, e.g. in order to suppress measured values during the warm-up phase.
Modem... Modem function available upon request Test Sequence ...
Settings x / – = function activated / deactivated Single-fault
If the single-fault condition has been activated, the test is interrupted as a failure as soon as an error occurs.
Auto Class PSI Test results (passed or failed ) for the various selector
switch positions are automatically assigned to the 8 statistics channels.
inc. Service Error Measurement results are compensated by taking
service error into consideration (measuring error).
IT Network
Testing in IT systems can be performed by suppressing tests for U
PE-N
. The U
PE-N
test determines whether or not voltage is present at PE. (Leakage current measure­ment results may otherwise be distorted.)
Acst Sig, Seq An acoustic signal is generated for: incorrect con-
nection of the DUT, error in the electrical supply system and the next test step.
Acst Sig, Meas An acoustic signal is generated for: measured value
fluctuations and test current polarity reversal.
Auto Point This function can be enabled with the help of an
upgrade program, e.g. Z745K (see chapter 13 on page 38). An acoustic signal indicates whether or not the probe is connected to the protective conductor. The test sequence is run automatically. Rapid signal frequency: probe connected Slow signal frequency: next measuring point
Direct Printing See chapter 15 on page 39. Reports ... Reports which have been saved to memory can be
selected from a list with an ID number and displayed (see
chapter 14.2 on page 39
).
Service ... – Time and date settings
(if a printer module is used, the same time and date must also be entered to the PSI- menu)
– Service functions (password required)
MENU
Loading...
+ 33 hidden pages