Gossen MetraWatt SECUTEST SII Operating Instructions Manual

Page 1
Operating Instructions
SECUTEST®SII
Test Instrument for DIN VDE 0701, 0702 and 0751
3-349-223-03
8/11.05
Page 2
SONDERS232 5 4
SECUTEST SI I
45678910
11
12
13
14 15
FUNCTION-TEST
230 V
MENU
OFF
VDE 0751
VDE 0701
0702
T. 240
T. 200
T. 1
VDE
N
L
L
N
SL
3
2
1
1
2
3
Feature F01
19
20
18
17
23
18
22
21
16
These operating instructions describe an instrument with firmware version 6xx.
Note
Contact problems with exposed conductive parts when using the standard probe with test tip
In order to assure good contact, surface coatings must be removed from devices under test with special tools at a suitable location so that the surface has a metallic shine. The tip of the test probe is not suitable for scratching away paint, because this may impair its coating and/or mechanical strength. The brush probe may be more suitable than the test probe in certain individual cases.
Page 3
GOSSEN METRAWATT GMBH 3
Upper left figure
1 Jack for protective conductor at device under test (feature F01) 2 Jack for neutral conductor at device under test (feature F01) 3 Jack for phase conductor at device under test (feature F01) 4 Jack for connecting the probe 5 Jack for connecting the probe 6 Function selector switch
– VDE-: Automatic test sequence according to selected standard – OFF: Instrument is switched off (no disconnection from mains) – MENU: Setup: Device configuration
R
SL
: Individual tests
– FUNCTION TEST: Function test
7 scroll key for menu and parameter selection 8
scroll key for menu and parameter selection
9 LCD panel 10 Socket connector for RS 232 interface 11 Signal lamp for mains connection error 12
key for entry, for starting test sequences and for finger contact
13
help key (context sensitive)
14 Key next to the symbol for switching test voltage to
the test socket (only possible if symbol LED is blinking)
15 Signal lamp for the functions text
Lower Left figure
16 Push-buttons (left and right) for releasing the handle from its snap-in position 17 Standard outlet socket (test socket) for connecting the device under test 18 Push-buttons (left and right) for releasing the lid 19 Compartment for probe and accessories 20 Lid 21 Test probe 22 Carrying handle and tilt stand 23 Cover or printer module (feature E01)
Overview of Available Probe Types
1)
Accessory
Note
Use of probes other than those specified above:
Cables plugged into jacks 4 and 5 must be short-circuited in order to perform the probe test, either by plugging the cable ends together or by means of a conductive surface at the device under test (4-wire measurement). Remove any corrosion from the device under test.
Data Security
Measurement data, report data and user entries are stored to RAM at the SECUTEST
®
PSI printer module (feature E01), as long as the respective
battery supplies the required amount of voltage. Save your data to a PC on a regular basis in order to prevent any loss of
data at the printer module. We assume no liability for data loss.
We recommend the following PC programs for data processing and data management:
PS3 (transmission of measurement data to a PC, documentation, management, report generating and deadline follow-up)
SECU 601 (program for activating the data base function in the instrument (DBmed option), if not already available as a feature)
PC.doc-WORD (report and list generation
PC.doc-ACCESS (test data management)
i
Probe Type Application Special Features
Standard probe (test probe with coil-cable and alligator clip)
Text current: 200 mA
none
SK2
1)
Test current: 200 mA
Probe with plain cable (not coil-cable), 2meters long
SK5 option (feature KD01 or accessory)
Special probe in combination with “automatic recognition of measuring point change” function (see
chapter 13).
Brush probe
1)
Can be plugged onto all above listed probes and test probes
Leakage current, protective conduc­tor resistance
For contacting devices under test with rotating, vibrating, exposed conductive parts
!
Page 4
Contents Page Contents Page
1 Applications ...................................................................................6
1.1 Table: Types of DUTs – Tests – Regulations ....................................................6
1.2 Table: Individual Measurements and Regulations ............................................6
1.3 Table: Leakage Current Types ........................................................................7
1.4 List of Possible Options for SECUTEST
®
SII Device Series .................................7
2 Safety Features and Precautions ...................................................8
3 Initial Start-Up ................................................................................9
3.1 Connection to the Mains (230 V, 50 Hz) ..........................................................9
3.2 Automatic Recognition of Mains Connection Errors ........................................10
4 General Notes ..............................................................................10
4.1 Online Instructions .......................................................................................10
4.1.1 Automatic Safety Class Selection ..................................................................10
4.1.2 Manual or Automatic Operating Sequences ................................................... 10
4.2 Online Help .................................................................................................10
4.3 Adjusting Contrast .......................................................................................11
4.4 Configuring Device Parameters, Setting Date and Time ..................................11
4.5 Configuring Measurement and Sequence Parameters .................................... 11
4.6 Setting Limit Values ..................................................................................... 11
4.7 Saving the Settings ......................................................................................11
5 Classification of Devices Under Test ............................................12
5.1 Safety Classes ............................................................................................12
5.2 Application Parts (electrical medical devices) .................................................12
6 Abbreviations ...............................................................................13
7 Connecting the Device Under Test ...............................................14
8 Configuring Device Parameters ...................................................15
9 Individual Measurements ............................................................16
9.1 Measuring Protective Conductor Resistance ................................................. 16
9.2 Insulation Resistance R
ISO
. .......................................................................... 17
9.3 Measuring Leakage Current ......................................................................... 19
9.3.1 Contact Current I
Contact
(probe current, housing leakage current) ...................19
9.3.2 Residual Current ΔI ..................................................................................... 19
9.3.3 Device Leakage Current I
DL
per DIN VDE 0751 .............................................19
9.4 Equivalent Leakage Current ......................................................................... 20
9.5 Probe Voltage U
probe
– Max. 300 V .............................................................. 21
9.6 Alternating / Direct Voltage U
AC/DC
– Max. 253 V (feature F01) ...................... 22
9.7 Resistance R (feature F01) ........................................................................... 22
9.8 Measurements with Accessories (feature F01) .............................................. 22
9.8.1 Alternating Current I
Z
with Clip-On Transformer (feature F01) ........................22
9.8.2 Temperature T with Pt100/1000 Sensor (feature F01) .................................. 23
10 Function Test ...............................................................................24
11 Measurements – Selector Switch Settings for Standards ...........26
11.1 Test Sequences .......................................................................................... 26
11.2 Setting Up Test Sequences .......................................................................... 27
11.3 Configuring Measuring Parameters ..............................................................27
11.4 Testing Devices in Accordance with DIN VDE 0701, Parts 1 and 200 ............. 28
11.5 Testing Devices in Accordance with DIN VDE 0701, Part 240 ........................ 30
11.6 Testing Devices in Accordance with DIN VDE 0702:2004 .............................. 32
11.7 Testing Extension Cables for VDE 0701, Part 1, and VDE 0702
(option: EL1 adapter) ................................................................................... 34
11.8 Testing Multiple Outlets for VDE 0702 (optional EL1 adapter) ......................... 35
11.9 Testing in Accordance with DIN VDE 0751 .................................................... 36
Page 5
GOSSEN METRAWATT GMBH 5
Contents Page Contents Page
12 Storage in PSI Module (Feature E01)
and Database Operations (Feature KB01 or DBmed option) ........ 38
12.1 Storing Measurement Data in the PSI Module ................................................38
12.2 Database Operations ...................................................................................38
12.2.1 Setting Up Test Sequences with PC Software ................................................38
12.2.2 Storing Test Results to the SECUTEST
SII ......................................................38
13 Remote Control (Feature KD01 or SK5 Option) ............................ 38
14 Printing via PSI Module (Feature E01)
or with the Printer Adapter (DA-II accessory) ............................. 39
14.1 Print Out Test Data at the PSI Module ...........................................................39
14.2 Printing Test Results in Report Form .............................................................39
15 Direct Print-Out (feature KE01, SECU-dd option) ......................... 39
16 Characteristic Values .................................................................. 40
17 RS 232 Interface .......................................................................... 43
17.1 Transmission of Measurement Results to the SECUTEST®PSI ........................43
17.2 PC Connection ............................................................................................43
17.2.1 Software Evaluation of Measurement Results ................................................43
17.2.2 Instrument Control via Interface Commands ..................................................43
17.3 Interface Configuration and Protocol .............................................................43
18 Appendix ......................................................................................44
18.1 Evaluation of Measured Values for Individual Measurements
as well as for Calculated Quantities ..............................................................44
18.2 Index ..........................................................................................................45
19 Maintenance - Calibration ........................................................... 46
20 Repair and Replacement Parts Service
DKD Calibration Laboratory and Rental Instrument Service ........ 46
21 Product Support .......................................................................... 47
Condensed instructions for PS3 software for maintenance and equipment management are provided with the PS3 CD ROM.
They include important instructions regarding the following topics:
• PS3 demo with additional instructions
• Special features for reading measured values in from a data file
Page 6
1 Applications
1.1 Table: Types of DUTs – Tests – Regulations
Attention!
!
The test instrument may not be used for measurements within electrical systems!
1.2 Table: Individual Measurements and Regulations
Key
Required test
Start-up and Modifi­cations
Testing after Repairs
Periodic Tes ti ng
Devices under test to be tested in accordance with the following regulations
DIN VDE 0751:2001
DIN VDE 0701 Part 1:2000
DIN VDE 0701 Part 240
DIN VDE 0751:2001
DIN VDE 0702:2004
DIN VDE 0751:2001
Laboratory instruments ••
Measuring and control instruments
••
Voltage generation devices
••
Electric tools ••
Electric heating devices ••
Electric motor devices ••
Lamps ••
Devices for entertain­ment electronics, information and commu­nications technology
••
Cable reels, extension and connection leads
••
Data processing and office equipment
••
Electrical medical de­vices, application parts
••
Individual Measurements per Regulation
Test current [A]
DIN VDE 0701,
part 1
DIN VDE 0701,
part 200
DIN VDE 0701,
part 240
DIN VDE 0702
DIN VDE 0751
Protective conductor resistance
0.2 ••••
Insulation resistance
••
Equivalent leakage current
••••
Equivalent (device) leakage current
Residual current
••
Contact current
••
Absence of voltage (exposed conductive parts)
Patient leakage current
Device leakage current
Page 7
GOSSEN METRAWATT GMBH 7
1.3 Table: Leakage Current Types
Key
NC = Normal conditions PAT = Patient application part PE = Potential earthing
, mains protective conductor
DPE = Protective conductor of the device under test
1.4 List of Possible Options for SECUTEST®SII Device Series
Example for complete type designation (article number) of a SECUTEST
®
SII: M7030 E01 KD01 KE01 (only the designation of the basic device, M7030, and features other than 00 are specified)
1)
Each measured value is documented in this case, as opposed to the results of a test sequence for which the poorest value for each given test is displayed.
2)
Adapter set for international use (feature B00 - earthing contact socket - included)
3)
With configuration for hospital beds of Safety Class 2 (pre-adjusted)
4)
For multimeter functions, for measurements with accessories and for devices under test without plug, including 3 plug-on quick clips
DIN VDE 0701-1
DIN VDE 0702-1
DIN VDE 0751-1 (2001)
The following is measured:
Equivalent leakage current
Equivalent leakage current
PROBE (connected to protective conductor) to L & N
Equivalent device leakage current
PROBE (open protective conductor) to L & N
Contact current / mea­surement for absence of voltage
Contact current / mea­surement for absence of voltage
Probe to PE
Protective con­ductor to PE
Device leak­age current during opera­tion
Protective con­ductor inter­rupted, probe to PE
Protective conductor current with differential current measurement
Protective conductor current with differential current measurement
Device leak­age current during opera­tion, differential cur­rent method
See chapter
9.3.2
Features
00 01 02 03 04 05 06 07 08 09 10 11 99
Design AGM
Mains Connection for Respective Country
BD
F CH
2)
User Interface Language
C
DGB
Configuration (settings in the setup menu)
D
GM
3)
Printer Module SECUTEST
®
PSI
E
w/o with
Additional Test Sockets (cannot be retrofitted)
F
w/o
with
4)
Database DBmed Option (Z853H)
KB
w/o with
Remote Control SK5 Special Cable (Z745K)
KD
w/o with
Direct Printing After Each Measurement for Automatic Test Sequences
1)
Read-Out via
RS 232
, SECU-dd Option (Z853L)
KE
w/o with
DKD calibration certificate
L
w/o with
Page 8
2 Safety Features and Precautions
The SECUTEST®SII test instrument has been manufactured and tested in accordance with the following safety regulations:
IEC 61010-1 / DIN EN 61010-1 / VDE 0411-1 and DIN VDE 0404
,
DIN VDE 0413 Part 2 and 4
.
When used for its intended purpose, the safety of the user, the test instrument and the device under test (electrical equipment or electrical medical device) is assured.
Read the operating instructions carefully and completely before placing your test instrument into service. Follow all instructions contained therein. Make sure that the operating instructions are available to all users of the instrument.
Tests may only be performed by a qualified electrician or under the supervision and direction of a qualified electrician. The user must be instructed by a qualified electrician in the execution and evaluation of tests.
Note
Manufacturers and importers of electrical medial devices must provide documentation for the performance of maintenance by trained personnel.
Observe the following safety precautions:
• The instrument may only be connected to electrical systems with a maximum of 230 V which are protected with a fuse or circuit breaker with a maximum rating of 16 A.
• Measurements within electrical systems are prohibited.
• Be prepared for the occurrence of unexpected voltages at devices under test (for example, capacitors can be dangerously charged).
• Make certain that the measurement cables are in flawless condition, e.g. no damage to insulation, no interruptions in cables or plugs etc.
• Grip the test probe firmly, for example during insertion into a jack socket. Tensioning at the coil cord may otherwise cause the test probe to snap back resulting in possible injury.
Measuring of insulation resistance (equivalent leakage current) The test is conducted with up to 500 V, with a current limit of I < 10 mA. However, when touching the terminals (L or N), you will get an electric shock which, in turn, may lead to accidents. Feature F01: When connecting the DUT via jacks 1 to 3, special care must be taken to avoid touching the open cables.
Leakage current measurement During leakage current measurement it is essential to ensure that the DUT is operated at mains voltage. Exposed conductive parts may be subject to hazardous contact voltage during the test and may not be touched under any cricumstances. (A mains shutdown takes place when the leakage current exceeds approx. 10 mA).
Attention!
!
The function test may only be performed after the DUT has successfully passed the safety test!
The measuring and test instrument may not be used:
• If it demonstrates visible damage
• With damaged connector cables, measuring cables or patient ports
• If it no longer functions properly
• After extraordinary stresses due to transport
In such cases, the instrument must be removed from operation and secured against unintentional use.
Meanings of Symbols on the Instrument
The symbols on the instrument have the following meanings:
Warning regarding dangerous electrical voltage
Warning concerning a point of danger (attention: observe documentation!)
Test socket
VDE testing authority approval mark
This device may not be disposed with the trash. For further details on the WEEE marking, please refer to our website www.gossenmetrawatt.de and enter search key ’WEEE’.
!
Page 9
GOSSEN METRAWATT GMBH 9
3 Initial Start-Up
3.1 Connection to the Mains (230 V, 50 Hz)
Connect the mains plug at the test instrument to the mains power
outlet. The function selector switch can be set to any position. If a mains outlet (earthing contact outlet) is not available, or if only a 3-phase outlet is available, the adapter socket can be used to connect the phase conductor, the neutral conductor and the protective conductor. The adapter socket has three permanently attached cables and is included with the KS13 cable set
Attention!
!
If connection is not possible via an earthing contact outlet: Shut down mains power first. Then connect the cables from the coupling socket to the mains using pick-off clips in accordance with the diagram. Disconnection from mains power is only possible with the mains plug.
L1
N
green-yellow
green-yellow
PE
L1
L2
L3
N
PE
L1
L2
L3
N
green-yellow
U
L–N
= 230 V
Mains Connection
Page 10
3.2 Automatic Recognition of Mains Connection Errors
The device automatically recognizes mains connection errors if the condi­tions in the following table have been fulfilled. The user is informed of the type of error, and all measuring functions are disabled in the event of danger.
1)
In SETUP – test sequence – IT system
Attention!
!
In either of the first two cases listed in the table above, immediately disconnect the test instrument from the mains and eliminate the error!
Note
Voltage at the electrical system’s protective conductor PE may result in distorted measurement values during testing for the absence of voltage, or during leakage voltage measurements.
4 General Notes
4.1 Online Instructions
The performance of measurements and tests is quick and easy with the SECUTEST
®
SII. Integrated online instructions inform the user regarding all required connections, necessary work steps, operator errors, measurement results and more in all measuring modes. Information and test results appear at the dot matrix LCD in plain text.
4.1.1 Automatic Safety Class Selection
Depending upon the type of mains plug or the connection configuration for the device under test, the test instrument recognizes the respective safety class and recommends its use for the measurement to be performed.
4.1.2 Manual or Automatic Operating Sequences
Depending upon selections made in the setup menu (selector switch in the Auto position), the next measurement is started automatically after the current measurement has been completed, or can only be started after manual acknowledgement. The integrated online instructions are adequate for most tests and measurements. However, the contents of these operating instructions should nevertheless be read and observed.
4.2 Online Help
Online help can be queried and displayed at the LCD for all measuring and test functions, and for almost all settings. Schematic diagrams which illustrate proper connection of devices under test to the SECUTEST
®
SII
can be displayed as well.
Press the following key in order to query online help:
Press the same key again in order to exit online help.
Note
Online help can be queried during measurement by pressing and holding the help key.
Typ e o f Mains
Connection Error
Message Condition Measurements
Voltage at protective
conductor PE to finger
contact
(key)
Text appears at
LCD
Press
key
U > 40 V
disabled
Protective conductor PE
and phase conductor L
reversed and/or
neutral conductor N
interrupted
lamp
lights up
Voltage at PE
>65V
impossible
(no supply power)
Contact voltage at
protective conductor PE
to neutral conductor N
or phase conductor L
Text appears at
LCD
U > 25 V
disabled, although
disabling can be
deactivated
1)
Mains voltage
too low lamp
lights up
U
L-N
< 90/180 V
possible
under certain
circumstances
!
!
Page 11
GOSSEN METRAWATT GMBH 11
4.3 Adjusting Contrast
Set the selector switch to MENU.
Select the Setup menu, “back” is highlighted.
Activate contrast adjustment.
Press and hold the key.
Adjust contrast.
Return to the menu.
Store the contrast setting to permanent memory with the save function in the setup menu.
4.4 Configuring Device Parameters, Setting Date and Time
Device parameters and functions which are valid for all selector switch positions can be activated or deactivated with the selector switch in the
MENU position using the Setup parameter (see chapter 8 on page 15).
4.5 Configuring Measurement and Sequence Parameters
Measurement and sequence parameters, as well as functions, can be activated or deactivated in the Setup menu (selector switch in the VDE position) for the respective test regulation. Refer to chapter 11.3 on page 27 for the significance of the various parameters.
4.6 Setting Limit Values
Upon delivery, the limit values set forth (at the point in time of issue) in applicable national and international standards are stored to the test in­strument. Limit values for each of the respective standards can be queried and changed if required with the Setup > Limit Values menu (selector switch in the VDE... position), but changes can only be made which result in even stricter testing than is required by the respective standard.
Newly entered limit values become effective immediately. However, these are only stored to memory permanently after activating the Store function in the Setup menu for the respective standard.
If the limit values set forth in the standards for certain safety classes need to be restored despite individualized settings, the All Values per Standard menu function in the Limit Values submenu must be selected and acknowledged with the Enter key.
If the limit values set forth in the standards are changed, the instrument’s device software can be updated via the RS 232 interface.
4.7 Saving the Settings All of the settings and changes which have been entered to the Sequence,
Limit Values (selector switch in the VDE... position) and Zero Point (tempera- ture measurement) (selector switch in the MENU position) menus, as well as
the selected contrast setting are retained until the selector switch is turned, or the test instrument is disconnected from mains power. If settings and changes should be retained even after mains power has been interrupted, they must be saved in the Setup menu for the respective test regulation or selector switch position (acknowledge the “Save” function).
MENU
Page 12
5 Classification of Devices Under Test
5.1 Safety Classes
Devices assigned to all of the following safety classes are equipped with basic insulation, and provide for protection against electrical shock by means of various additional precautions as well.
Safety Class I Devices
Exposed, conductive parts are connected to the protective conductor so that they are not charged with voltage if the basic insulation should fail.
Safety Class II Devices
These devices are equipped with double insulation or reinforced insulation.
Safety Class III Devices
These devices are powered with safety extra-low voltage (SELV). Beyond this, no voltages are generated which exceed SELV. These devices may not be connected to the mains.
Note: The DUT may only be connected to the test instrument at jacks 1 through 3 (Feature F01). Only a visual inspection, an insulation resistance measurement or a measurement of the supply voltage can be performed, see parameter “SC III U
V
” on page 27.
Parameter Classification (in the Sequence... menu)
The SECUTEST
®
SII always tests in accordance with the strictest limit val­ues of the respectively selected safety class. The test is failed if this limit value is exceeded. However, devices under test exist for which higher limit values are allowable.
If parameter classification is activated (= x), the user is asked if higher limit values are allowable for the DUT. If “Yes” is entered, reevaluation ensues and the DUT may pass the test.
Examples
If the insulation resistance test is failed, or if interference suppression capacitors have been replaced, equivalent leakage current measurement must be performed on DUTs with heating elements in accordance with DIN VDE 0702.
The test is failed by a DUT with 300 kW with deactivated classification function (= _), but it is passed with activated classification function (= x) if the question is answered accordingly.
Similar possibilities for passing the test, although it might otherwise be failed, are included in DIN VDE 0701, part 1.
If the DUT is connected via the jacks instead of the test socket, other limit values apply because higher power consumption is possible in this case (e.g. part 1 specifies a limit value of 1 mA per kW for equivalent leakage current).
See also table: “Maximum Allowable Limit Values for Equivalent
Leakage Current in mA” on page 20.
5.2 Application Parts (electrical medical devices)
Type B Application Parts (body)
Devices of this type are suitable for both internal and external patient applications, except for use in direct proximity to the heart.
These devices provide for adequate protection against shock especially as regards:
• Reliable leakage current
• Reliable protective conductor connection if utilized Devices of the following safety classes are allowable: I or II.
Type BF Application Parts (body float)
Same as type B, but with type F insulated application parts.
Type CF Application Parts (cardiac float)
Devices of this type are suitable for use directly at the heart. The application part may not be grounded.
Devices of the following safety classes are allowable: I or II.
Page 13
GOSSEN METRAWATT GMBH 13
6 Abbreviations
AE Error condition: application part grounded B, BF, CF Classifications for application parts OE Operational earth
ΔI Differential current, residual current ΔI
max
Maximum residual current (during function test)
EDL
A1/A2
Equivalent device leakage current with designation A1/A2 (cross-reference in the standard)
EDL
PX±PE
Equivalent device leakage current for portable x-ray +PE: with additional PE conductor –PE: without additional PE conductor
EDL
SCII
Equivalent device leakage current for devices with
additional safety class 2 components HE Error condition: housing grounded I
L
, I , IHL, I
probe
Leakage current (differential, probe or contact current) I
Re
Residual current I
EL
, I
EDL
Equivalent device leakage current I
PE
Earth leakage current (current at protective conductor) IT system The IT system has no direct contact between active
conductors and grounded parts: bodies within the
electrical system are grounded. I
Lmax
Maximum load current (during function test) I
Z
Current at clip-on meter L Phase conductor PF Power factor MedGV German medical device ordinance MPG German medial product law MSELV Medical safety extra-low voltage N Neutral conductor NC Normal conditions P Active power PA Functional earth (equipotential bonding)
R Resistance (only with feature F01) R
ISO,
R-ISO Insulation resistance R-ISO APP-PE Insulation resistance: application part to PE R-ISO
INT. CARD.
Insulation resistance: intracardiac
(application at the heart) R-ISO NL-PE Insulation resistance: neutral/phase conductor to PE R
PE
, R-PE Protective conductor resistance
R-SL±Mains Protective conductor resistance limit value for
+mains: DUT with mains power cable
–mains: DUT without mains power cable
(protective conductor resistance limit value for mains
power cable only = 0.1 Ω) S Apparent power SELV Safety extra-low voltage SFC Single-fault condition PE Protective conductor U
AC/DC
AC/DC voltage (only with feature F01) U
REF
Reference voltage: voltage to which all leakage voltages
make reference (as a rule nominal line voltage) U-ISO Test voltage for insulation measurement U
L-N
Line voltage U
MEAS
Voltage with which testing was executed.
Displayed for all leakage current measurements. U
probe
Probe voltage tOn-time Temp Resistance temperature (only with feature F01) W Electrical energy ZVEH General Association of German Electricians
Page 14
7 Connecting the Device Under Test
Connect the DUT in accordance with the schematic diagrams included
in the online help function.
Connection of the DUT to the test instrument depends upon:
• Type of connection: – With plug (“to test socket” parameter), applies to EL1 adapter as well – Without plug, single or multi-phase connection
(“to jacks” parameter), only with feature F01
– No connection to tester (“permanent connection” parameter)
• Whether or not an adapter is used: – Adapter to socket (customer specific adapter) – AT3-med to socket, adapter for devices
which are equipped with 5-pole, 16 A CEE plugs
– AT3-III to socket, adapter for devices
which are equipped 5-pole, 32 A CEE plugs, see AT3-III operating instructions for test sequence.
• The DUT’s safety class (I, II or I I I)
Note
The DUT must be switched on for all tests. Switches, relays, temperature regulators etc. must all be taken into consideration.
The test instrument automatically recognizes whether or not the DUT is connected to jacks 1 through 3. If the instrument is equipped with feature B00, 01 or 03, the instrument also recognizes whether or not a DUT has been connected to the test socket. As a default setting, the program sequence assumes that the plug from the DUT has been connected to the test socket.
Note
Safety Class II Devices with Safety Class I Mains Plugs
If the device under test is equipped with a safety class I plug although it complies with safety class II, safety class I is recog­nized by the test instrument. If this is the case, switch from safety class I to safety class II in the initial menu.
If the test instrument is unable to automatically recognize how the DUT has been connected, the recommended connection setup should be double checked and determined manually if necessary.
Position the
cursor at the second line in the start menu for the test
sequence.
A selection of possible connection setups can be displayed by activat-
ing the
key.
Select the desired connection setup with the
cursor and
acknowledge with the
key.
Omitting the Protective Conductor Test in the Case of Fully Insulated Devices
You are testing a fully insulated safety class I device (e. g. screen, sub­mersible pump, etc.), which is not equipped with an external protective conductor contact.
The decision as to the necessity of a protective conductor test in this case is to be taken by a qualified electrician who should also assume responsi­bility.
You can omit the protective conductor test by pressing the key
as
soon as the following instruction is shown: „Please connect the probe with the protective conductor of the DUT“.
Protective Conductor and Insulation Resistance Measurements for Permanently Installed Devices Under Test
Attention!
!
Deactivate the electrical system which supplies power to the device under test before connecting the test instrument!
Remove the mains fuses from the device under test and disconnect
the neutral conductor N inside the device under test.
Measuring Contact Current (absence of voltage)
Make sure that the contacted parts are not grounded.
Page 15
GOSSEN METRAWATT GMBH 15
8 Configuring Device Parameters
General device parameters can be configured and saved using Setup with the selector switch in the MENU posi­tion.
Select Setup/Menu and acknowledge.
Select parameter and acknowledge, change setting and acknowledge. Settings x / – = function activated / deactivated
Limit values ...
Illumination Background illumination for the LCD. One of three
different conditions can be selected with the up and down scroll keys: x: continuously on, –: off Numbers 1 through 9: duration in minutes after
which illumination is automatically deactivated. Test time Duration of a single test (0 255 s) Reference voltage: Voltage to which leakage current makes reference
(as a rule nominal line voltage) Earth fault During the short-circuit test, testing is also performed
to determine whether or not a connection exists
between L/N and PE (short-circuit to exposed
conductive part). We assume that a short-circuit to
an exposed conductive part exists in the event of
leakage current greater than 15 mA from L/N to PE.
This value should be increased for some DUTs (in
particular high-current consumers), because greater
leakage currents are present.
Mains wait Line voltage is initially applied to the test socket.
However, testing does not begin until after the number of seconds selected in “Mains wait” has elapsed, e.g. in order to suppress measured values during the warm-up phase.
Modem... Modem function available upon request Test Sequence ...
Settings x / – = function activated / deactivated Single-fault
If the single-fault condition has been activated, the test is interrupted as a failure as soon as an error occurs.
Auto Class PSI Test results (passed or failed ) for the various selector
switch positions are automatically assigned to the 8 statistics channels.
inc. Service Error Measurement results are compensated by taking
service error into consideration (measuring error).
IT Network
Testing in IT systems can be performed by suppressing tests for U
PE-N
. The U
PE-N
test determines whether or not voltage is present at PE. (Leakage current measure­ment results may otherwise be distorted.)
Acst Sig, Seq An acoustic signal is generated for: incorrect con-
nection of the DUT, error in the electrical supply system and the next test step.
Acst Sig, Meas An acoustic signal is generated for: measured value
fluctuations and test current polarity reversal.
Auto Point This function can be enabled with the help of an
upgrade program, e.g. Z745K (see chapter 13 on page 38). An acoustic signal indicates whether or not the probe is connected to the protective conductor. The test sequence is run automatically. Rapid signal frequency: probe connected Slow signal frequency: next measuring point
Direct Printing See chapter 15 on page 39. Reports ... Reports which have been saved to memory can be
selected from a list with an ID number and displayed (see
chapter 14.2 on page 39
).
Service ... – Time and date settings
(if a printer module is used, the same time and date must also be entered to the PSI- menu)
– Service functions (password required)
MENU
Page 16
9 Individual Measurements
9.1 Measuring Protective Conductor Resistance
Definition
Protective conductor resistance is the sum of the following resistances:
• Connector cable or device connector cable resistance
• Contact resistance at plug and terminal connections
• Extension cable resistance if utilized
Resistance is measured:
• Between each exposed conductive part of the housing and the earthing contacts at the mains and the device plug (if a removable mains connector cable is used), or the protective conductor terminal for permanently installed devices.
• As a 4-pole measurement
• Between the earthing contacts at the mains plug and the earthing contacts at the device plug for device connector cables
• Between the earthing contacts at the mains plug and the earthing contacts at the coupling socket for extension cables
Connecting Safety Class I Devices to the Test Socket
When the DUT is connected, resistance is measured between the protec­tive conductor terminal at the test socket or at the PE jack, and the probe connection at the DUT (contact with conductive parts of the housing).
In order to measure protective conductor resistance, contact a
conductive part of the housing with the probe, which is connected to the protective conductor.
During measurement, the connector cable must only be moved to the extent it is accessible during repair, modification or testing. If a change in resistance occurs during the manual test step of the conti­nuity test, it must be assumed that the protective conductor is damaged, or that one of the connector contacts is no longer in flawless condition.
Testing Extension Cables
See test sequence in chapter 11.7 on page 34.
Note
“DUT connection: SC I/II” is not displayed when the test is performed individually, but rather only during the automatic test sequence.
MENU
Ω
Page 17
GOSSEN METRAWATT GMBH 17
Combined Test – Differential Protective Conductor Resistance Zero balancing is also possible for protective conductor measurement. With
zero balancing, all subsequent measurements are adjusted with an offset such that 0 Ω is displayed for a selected reference point which is connected to the protective conductor. When test points are contacted with the probe which are electrically connected to this reference point, dif­ferential resistance ΔR
PE
between the reference point and the contacted
test point is displayed. The mains release key must be activated during measurement in order
to perform zero balancing. The acquired value can either be applied (the value remains in memory until the instrument is disconnected from mains power), permanently saved or deleted.
Maximum Allowable Limit Values for Protective Conductor Resistance for Connector Cables with Lengths of up to 5 m
1)
This value may not exceed 1 Ω for permanently connected data processing systems (DIN VDE 0701, part 240).
2)
Total protective conductor resistance: max. 1 Ω
9.2 Insulation Resistance R
ISO
.
Definition
Safety Class I
Insulation resistance is measured between short­circuited mains terminals and the protective conductor.
Safety Classes II and III
Insulation resistance is measured between short­circuited mains terminals and external conductive parts which can be contacted with the probe.
Exception for Permanently Installed Safety Class I Devices
Attention!
!
Deactivate the electrical system which supplies power to the device under test before connecting the test instrument!
Remove the mains fuses from the device under test and disconnect
the neutral conductor N inside the device under test.
Connect the probe to phase conductor L at the device under test in
order to measure insulation resistance.
Test Standard Test Current
Open-
Circuit Voltage
R
PE
Housing –
Device Plug
R
PE
Housing –
Mains Plug
VDE 0701 Part 1:2000 VDE 0702:2004 VDE 0751:2001
> 200 mA 4 V < U
L
< 24 V
0.3 Ω
1)
+ 0.1 Ω
2)
for each
additional 7.5 m
VDE 0751:2001 0.2 Ω
Ω
f
o
r
S
C
I
I
Page 18
Sequence
Note
All switches at the device under test must be set to the on posi­tion during measurement of insulation resistance, including tem­perature controlled switches and temperature regulators as well. Measurement must be performed in all program stages for devices equipped with program controllers.
R-ISO
Start measurement.
Nominal voltage is 500 V DC in this case. Nominal voltage can be adjusted within a range of 50 V to 550 V DC.
Note
When insulation measurement is first started from the menu, nominal voltage is always set to 500 V. Open-circuit voltage is always greater than nominal voltage.
Minimum Allowable Limit Values for Insulation Resistance
* with activated heating elements
(if heating power > 3 kW and R
ISO
< 0.3 MΩ: leakage current measurement
required)
Notes Concerning Test Standard 0702
If the measured value is less than 0.3 MΩ for safety class I devices which include a heating element, equivalent leakage current must be measured as described in chapter 9.4 on page 20, and this test must be passed. The same applies to mains powered safety class II electronic devices if the required value of 2 MΩ is not complied with. All exposed, conductive parts of safety class II and III devices, as well as of battery-powered devices must be scanned with the probe and insula­tion resistance and/or leakage current must be measured.
Batteries must be disconnected from their terminals during testing of battery powered devices.
Test Standard Test Voltage
R
ISO
SCI SCII SC III Heat
VDE 0701 Part 1:2000
500 V
1MΩ 2MΩ 0.25 MΩ 0.3 MΩ
VDE 0702: 2004
1MΩ 2MΩ 0.25 MΩ 0.3 MΩ
VDE 0751: 2001
2MΩ 7MΩ
70 MΩ 70 MΩ
Page 19
GOSSEN METRAWATT GMBH 19
9.3 Measuring Leakage Current
Attention!
!
During leakage current measurement it is essential to ensure that the DUT is operated at mains voltage. Exposed conductive parts may be subject to hazardous contact voltage during the test and may not be touched under any cricumstances. (A mains shutdown takes place when the leakage current exceeds approx. 10 mA).
Select the I
xx
measurement and start.
Each time line voltage is applied to the test socket, L and N are reversed, if this function has been se­lected in the leakage cur­rent menu (see chapter
9.3 on page 19). Frequency response is
taken into consideration in accordance with the dia­gram to the right when leakage current is measured.
9.3.1 Contact Current I
Contact
(probe current, housing leakage current)
Current which flows from housing parts which are not connected to the protective conductor via an external conductive connection to earth or another part of the housing. Flow of current via the protective conductor is excluded in this case.
The current’s AC component is measured. The DC component can also be measured by means of indi­vidual measurement (but not with a test sequence).
9.3.2 Residual Current ΔI
Sum of instantaneous current values which flow via the L and N conduc­tors at the device mains connection (also known as differential current). Residual current is practically identical to fault current in the event of an error. Fault current: Current which is caused by an insulation defect, and which flows via the defective point.
Attention: Residual current also includes contact current.
9.3.3 Device Leakage Current I
DL
per DIN VDE 0751
Device leakage current is determined by means of differential current measurement.
Maximum Allowable Limit Values for Leakage Current in mA
* for devices with a heating power > 3.5 kW
Note 1: Devices which do not have any exposed conductive parts which are
connected to the protective conductor, and which correspond with I
DL
and, if applicable, IPL, e.g. data processing equipment with
shielded power pack Note 2: Permanently connected devices with protective conductor Note 3: Portable x-ray devices with mineral insulation
Key
I
Contact
Housing leakage current (probe or contact current) ΔI Residual current I
DL
Device leakage current
10 10210310410510
6
+20
0
–20
–40
–60
Frequency (f) in Hz
Relative Magnitude (dB):
20 log
U(f)
U(f=10)
μΑ
2
L
N
1
1 Test socket (w/o protective conductor contact) 2Device housing
Tes t Standard
I
PE
I
Contact
ΔII
DL
NC SFC NC SFC
VDE 0701
-1: 2000
SC I: 3.5 1 mA/kW
*
0.5
SCI: 3.5
1 mA/kW
*
SC II: 0.5
VDE 0701
-240
0.25
VDE 0702: 2004
SC I: 3.5 1 mA/kW
*
0.5 3.5
VDE 0751: 2001
General 0.5
Notes 1 & 3 2.5
Note 2 5.0
SC II 0.1
Page 20
9.4 Equivalent Leakage Current
General
Measurement of equivalent leakage current is required for:
• DIN VDE 0701 and 0702, after passing the insulation test
Equivalent Device Leakage Current I
EDL
(DIN VDE 0751)
Measurement of equivalent device leakage current is required for:
• Electrical medial devices in accordance with DIN VDE 0751, part 1
Tes t S e t u p
A high-impedance power supply is connected between the short­circuited mains terminals and all exposed metal parts of the housing (which are connected to one another).
Measurement
Current which flows over the insulation at the device under test is measured.
Maximum Allowable Limit Values for Equivalent Leakage Current in mA
IELEquivalent leakage current I
EHL
Equivalent housing leakage current
PE Protective conductor
1)
For devices with heating power of greater than 3.5 kW
2)
This limit value is not taken into consideration in the DIN VDE 0751 standard.
Connection
Refer to the schematic diagrams included with the online help for connection instructions.
Connection Exception for Permanently Installed Safety Class I Devices
Current is measured between the probe, with which the L and N conduc­tors must be contacted, and the protective conductor terminal PE at the device under test for permanently installed safety class I devices under test.
Attention!
!
Disconnect mains power before connecting the test instrument!
Remove the mains fuses from the device under test and disconnect
the neutral conductor N inside the device under test.
Connect the probe to phase conductor L and neutral conductor N at
the device under test in order to measure equivalent leakage current.
Connection Exception for Multi-phase Devices
Equivalent leakage current measurement is not appropriate for multi­phase devices.
~
DUT
Probe
mA
SL
L N
Probe
DUT
~
mA
L N
PE
Test Standard I
EL
I
EHL
VDE 0701-1: 2000
SC I: 3.5
1 mA/kW
1)
SC II: 0.5
VDE 0702: 2004
3.5
1 mA/kW
1)
VDE 0751: 2001
SC II 0.2
2)
SC I (PE or parts connected to PE)
1
Permanently connected devices with PE 10
Portable x-ray devices with additional PE 5
Portable x-ray devices without additional PE 2
Devices with mineral insulation 5
Page 21
GOSSEN METRAWATT GMBH 21
Sequence
Current, which would flow during leakage current measurement con­ducted in accordance with device regulations and with nominal line volt­age, is displayed during this type of equivalent leakage current measure­ment. Leakage current measurement in accordance with the respective device regulations is usually not possible, because the device would have to be set up in an electrically isolated fashion, or connected to an earth isolated power supply to this end.
Equivalent Leakage Current I
EL
DIN VDE 070x / 2 K
Select the I-EL measurement and start.
Equivalent leakage current is measured between short-circuited N and L, and the protective conductor PE.
Measuring circuit resistance is equal to 2 kΩ for VDE 0701/0702 for simulation of mean body resistance of a human being.
Equivalent Device Leakage Current I
EDL
for VDE 0751 / 1 K
Select the I-EDL measurement and start. Equivalent device leakage current is measured between short-circuited N
and L, and the probe. Measuring circuit resistance is equal to 1 kΩ for VDE 0751 for the
simulation of mean patient resistance.
9.5 Probe Voltage U
probe
– Max. 300 V
Voltage is measured between the mains PE terminal at the test instrument and the probe. In this case the probe can also be used as a phase finder.
Start the U
probe
measurement.
Page 22
9.6 Alternating / Direct Voltage U
AC/DC
– Max. 253 V (feature F01)
Direct, alternating and pulsating voltages of up to 253 V can be measured between the N (2) and L (3) connector jacks.
Select the U
AC/DC
measurement and start.
9.7 Resistance R (feature F01)
Resistance of up to 150 kΩ can be measured between the PE (1) and N (2) connector jacks.
Select the R measurement and start.
9.8 Measurements with Accessories (feature F01)
9.8.1 Alternating Current I
Z
with Clip-On Transformer (feature F01)
Connection
Alternating current can be measured in two measuring ranges (1 mA 10 A ~, 1 A 100 A~) with a clip-on current-voltage transformer connected to the N (2) and L (3) jacks (e.g. the WZ12C).
Start the IZ measurement.
Select measuring range.
Page 23
GOSSEN METRAWATT GMBH 23
9.8.2 Temperature T with Pt100/1000 Sensor (feature F01) Connection
Temperature can be measured within a range of –200 °C to + 850 °C with a Pt100 or Pt1000 sensor (default setting) connected to the PE (1) and N (2) jacks.
Select the Tem p measurement and start. Select Pt100 or Pt1000 by means of
“select measuring range” with the key. The temperature unit of measure can be selected in the “TEMPERATURE” setup menu. Selection can be made amongst °C (Celsius), °F (Fahrenheit) or Kelvin. Zero balancing is also accessible via the “TEMPERATURE” setup menu.
Zero Balancing
Sensor cable resistance can be compensated with this function:
Short circuit the ends of the sensor leads and determine resistance as
shown below.
Zero Balancing
The determined value can be stored directly (
key) or changed first. The
data entry menu is opened with the key.
Change the measured value
manually with the help of the
and
keys.
Press the
key in order to
acknowledge the selected value, and to display other menu functions at the bottom of the window.
Save the selected value by activating the “store value” key
before
exiting the balancing function with the
key.
The “delete value” command can only be accessed via the “change value” menu. The “no zero balancing” setting is saved at the same time by activating the
key.
Page 24
10 Function Test
The device under test can be subjected to a function test with line voltage via the integrated test socket.
In addition to testing with the selector switch in the function test position, or with the Function parameter in the MENU position (feature F01 only), a function test can also be performed immediately after safety testing has been passed in accordance with the selected standard (not possible for safety class III devices).
Attention!
!
The function test may only be performed after the DUT has successfully passed the safety test.
Note
Each time line voltage is applied to the test socket, phase conductor L and neutral conductor N are automatically reversed, if the “mains polarity reversal” function has been activated (= x) in the I leakage selector switch position.
Note
The function test is only possible if the device under test has been connected to the test socket (21).
Measurements
The function test includes the following measurements:
– Voltage U-LN between the L and N conductors – Residual current ΔI
(corresponds to fault current between L and N)
– Load current I
L
– Active power P – Apparent power S (calculated) – Power factor PF (cos ϕ calculated, display > 10 W) – Electrical energy W – On-time t for U
L–N
at the socket (21)
The following values are also displayed for all selector switch positions except MENU, after the function test has been completed:
– Maximum residual current ΔI
max
– Maximum load current I
Lmax
– Maximum active power P
max
Power factor is calculated from active power and apparent power. Power factor corresponds to cos ϕ for sinusoidal quantities (line voltage and load current).
Attention!
!
Starting the Function Test
For reasons of safety, the device under test must be switched off before the function test is started. This precaution prevents inadvertent start-up of a device under test which may represent a hazard during operation, e.g. a circular saw or a disc grinder.
Ending the Function Test
After completion of the function test, devices under test must be turned off with their own switch – especially devices with relatively high inductivity.
Function Test
Page 25
GOSSEN METRAWATT GMBH 25
Short-Circuit Test
1 Test for shorts between conductors
N and L.
2 Test to determine whether or not
the N or L conductors are short­circuited to the protective conductor.
Note
A short-circuit at the device under test is recognized automati­cally by the test instrument. A message appears at the display (9), and the function test is disabled.
If the lamp blinks (15), line voltage can be switched to the test socket with the key (14), and the measurement can be started. If the lamp (15) is lit continuously, line voltage is present at the test socket.
The test socket can be rendered voltage-free with the key (14), or the function test can be ended with the
key (12).
Page 26
11 Measurements – Selector Switch Settings for Standards
If measurements need to be performed in accordance with given stan­dards which require specific tests, and if results need to be documented with a test report, an automatic test sequence is advisable instead of indi­vidual measurements.
Connect the SECUTEST
®
SII to mains power.
A mains connection test is initialized (see chapter 3.2 on page 10).
Connect the DUT to the test socket at the SECUTEST
®
SII (see chap­ter 7 on page 14). The test instrument initializes connection type recognition.
Set the selector switch
to the appropriate standard.
If the test instrument is equipped with features B00, 01 or 03, and if the DUT has been connected to the test socket, safety class recognition is performed. Otherwise, the safety class must be specified manually. Move the cursor up into the third line in the initial window with the key, and acknowledge with the key. A safety class can now be selected with the
and keys and acknowledged with the key.
The test sequence can be configured in the Setup… menu, limit values
can be changed if necessary and database options can be selected.
The test sequence is started by selecting Start test and acknowledging
with the
key (see following chapter, “Test Sequences”.
Tests which have already been included in chapters 9 through 14 are not described here again. The only exception is measurement of extension cables.
11.1 Test Sequences
Test sequences for the various standards are always run in the same order, assuming that the device under test has been properly connected and the mains connection test has been passed. The test sequence can be run step by step with manual activation of each subsequent step if this function has been specified, or automatically. Step by step manual operation can be selected if “Manual Sequence” has been activated under Sequence... in the setup menu in the initial window.
• Visual inspection: If “Visual Inspection” has been activated under Sequence
If a part is recognized as defective by the user, it must be identified as such by selecting it with the cursor and acknowledging with the key.
• Protective conductor measurement (for SC I devices under test only)
Note
If a protective conductor connection is not possible, the mea­surement can be skipped with
(when the message „Connect
probe to protective conductor“ is shown on the display).
• Evaluation of protective conductor testing
• Insulation resistance measurement
DIN VDE 0751: Only if ISO-R is activated under Sequence
in the setup menu
DIN VDE 0702: Only if ISO-R is activated in the initial window
• Evaluation of insulation test
• Leakage current measurement
• Evaluation of each individual leakage current measurement
• Evaluation of the overall test
• Perform function test if required:
The function test can be performed each time a safety test has been successfully completed. The blinking signal lamp indicates that the function test should be started. Beyond this, the function test can also be started from the Function Test selector switch position. See chapter 10 on page 24 regarding performance of the function test.
• Display test results
(the worst measured values for the test sequence)
• Save test results and print if required
Page 27
GOSSEN METRAWATT GMBH 27
11.2 Setting Up Test Sequences
All possible sequence settings for all of the regulations are listed below.
Select the Setup... menu from the initial program window and acknowledge
Repair tests, periodic testing
Save All of the settings in the setup menu, i.e.
configuration of measuring parameters and current limit values, can be saved with this command. These values remain active even after setting the selector switch to a different position, and after disconnection
from mains power. Sequence See page 27. Limit values See chapter 4.6 on page 11. Database Start with ID No.
x: Before each measurement is started, an entry prompt appears requesting entry of an ID no. An individual number can be entered (max. 20 characters) with the keypad at the PSI module (optional), read in with a barcode scanner (optional) or directly selected from a list. If an incorrect entry is made: Only complete lines can be deleted, and deletion is only possible with the
key at the instrument.
ID no. = test sequence (feature KB01 or DBmed option)
See chapter 12 on page 38.
11.3 Configuring Measuring Parameters
Depending upon the test regulation, various measuring parameters can be configured for the test sequence (settings: x / – = function activated / deactivated). All possible parameters for all of the regulations are listed below. The Sequence ... menu is accessed via the setup parameter in the initial program window for the respective regulation.
Select the Sequence ... menu and acknowledge.
Select a parameter, acknowledge, change and acknowledge the change.
General Parameters
Visual inspection This menu appears at the very beginning of the test
sequence.
Manual sequence Each test step must be acknowledged with the
key (see test sequence in chapter 8 on page 15 regarding test duration for automatic sequence)
Auto-store After testing is finished, test data are automatically
stored to the SECUTEST
®
SII (feature KB01 or
Dbmed option), or the SECUTEST
®
PSI (accessory).
Polarity reversal L an N are reversed each time line voltage is applied
to the test socket.
Classification Questions regarding classification appear if limit
values are exceeded (see chapter 5 on page 12).
SC III U
V
Supply voltage is measured instead of insulation resistance for active devices under test.
R-ISO LN-PE Insulation resistance measurement is performed
between phase/neutral conductors and the protective conductor.
Additional Parameters for DIN VDE 0702/0751
Auto test method
The instrument recognizes whether or not the device under test can be switched on: Leakage or residual current, or insulation resistance and earth leakage current, are measured accordingly
Adapter for socket Limit values are activated for permanently connected
devices. A device under test which is normally permanently installed can be connected to the test socket via an adapter. No voltage may be applied to the
test socket when this test method is used.
Page 28
11.4 Testing Devices in Accordance with DIN VDE 0701, Parts 1 and 200
The following measurements can be performed in accordance with the above mentioned standard:
• Protective conductor measurement R
PE
(permanent connection or
with plug)
– Part 1/200: Test current: ±200 mA DC
• Insulation resistance measurement R
ISO
Part 1
The following safety class I through III appliances and electrical equipment can be tested in this selector switch position, for example:
• Devices with electric motors
• Electrical heating devices
• Electrical tools
• Light fixtures SC I extension cables can also be tested (see chapter 11.7 on page 34).
Part 200
Mains powered, safety class I through III electronic devices, e.g. consumer electronics, can be tested in this selector switch position:
• Stereo equipment, televisions The following maximum values for equivalent leakage current apply for
mains powered electronic devices per DIN VDE 0701, part 200:
– Devices with single-phase power supply:1 mA – Devices with multi-phase power supply: 0.5 mA
Check connection parameters and start test.
To Socket This is the default setting. Refer to chapter 7 on page
14 for other types of connection.
Class If the test instrument is equipped with features B00,
01 or 03, and if the DUT has been connected to the test socket, safety class recognition is performed (SC I or SC II). Otherwise, the safety class must be specified manually.
Ext. Cable x: Extension cables or connector cables which are
longer than 5 m can be tested with the help of the EL1 adapter (optional), either separately or in combination with a device.
ID No. See parameters database in chapter 11.2 on page
27.
Setup See chapter 11.2 on page 27 regarding setup of the
measuring sequence.
Note
Extension cables can only be tested in the VDE 0701 part 1 and VDE 0702 selector switch positions if the EL1 accessory adapter is used (see chapter 11.7 on page 34).
Page 29
GOSSEN METRAWATT GMBH 29
Test Sequence per VDE 0701
Select connection, select test regulation VDE 0701 part 1 or 200, classify the DUT (SC I, II or III), part 1: extension cable X/– (with/without)
Visual inspection
OK?
SC I: RPE
OK?
I
L
OK?
Short-circuit
at DUT?
Tes t OK?
Yes /No
No
Yes
Display results, save/print report
Yes
No
Yes
No
No
Tes t OK?
Framed with dashed line: The test is only run if it has been activated in the initial program window, or in the Setup menu under Sequence ...
Switch line voltage to test socket, start function test
Start
function test
Safety Class I only: Contact all exposed, conductive parts with the
probe. The test can be repeated as often as desired for various protective conductor parts with the manual sequence*
* If it is not clear whether or not all
exposed, conductive parts are connected to one another or to the protective conductor, testing can be performed in the manual mode.
Switch DUT on
RISO+I
EL
OK?
or
Yes
No
No
Yes
Page 30
11.5 Testing Devices in Accordance with DIN VDE 0701, Part 240
Testing safety class I and II data processing equipment and office machines as individual devices and in combination with one another.
The following measurements can be performed in accordance with the above mentioned standard:
• Protective conductor R
PE
(permanent connection or with plug)
Test current: DC ±200 mA
• Housing leakage current I
• According to DIN VDE 0701, part 240, the device protective conduc­tor must be tested after maintenance, repair or modification of data processing equipment and office machines, and exposed, conductive parts must be tested for the absence of voltage. This applies to:
• Safety class I devices for all exposed, conductive parts which are accessible to the user, and which are not connected to the protective conductor
• Safety class II devices (totally insulated devices) for all exposed, conductive parts which are accessible to the user
with the mains plug poled in both directions
Setting Up the Test Sequence
See chapter 11.4 regarding the test sequence.
Special Parameters
Combined Testing
Safety class I and II devices can be tested individually or in combination. All protective conductor connections are tested first for interconnected safety class I devices, and then – as is also the case for interconnected safety class II devices – all exposed, conductive parts.
Connecting the Device Under Test
Connect the test instrument and the DUT as described below:
– Connect both devices to separate mains outlets.
The outlets to which the test instrument and the safety class I DUT are connected must share a common protective conductor!
– Or connect the test instrument to the mains and the DUT to the
test socket at the test instrument.
Data Processing / Office Machines
Permanently connected or with plug
To test socket at the instrument
The requirement for testing with the mains plug poled in both directions can be fulfilled
by connecting the DUT to the test socket at the instrument, and by activating mains polarity reversal under “Sequence” in the setup menu. Each time the key (14) is activated, phase conductor L and neutral conductor N are reversed at the test socket.
Attention!
!
Testing with mains polarity reversal or with the mains plug poled in both directions results in interruption of supply power to the affected data processing equipment or office machine. This test may thus only be conducted with the consent of the operator of the data processing equipment or office machine. If the DUT is defective, the electrical system’s RCCB may be tripped during testing which would also result in interruption of supply power to the affected equipment or office machine. The manufacturer of the test instrument assumes no liability for loss of data or other damage which results from use of the test instrument.
Page 31
GOSSEN METRAWATT GMBH 31
Test Sequence per VDE 0701, Part 240
Select connection, select test regulation VDE 0701, part 240, classify the DUT (SC I, II or III), combination testing X/– (yes/no)
Visual inspection
OK?
SC I: RPE
OK?
Short-circuit
at DUT?
Yes /No
No
Yes
Display results (for combination testing: additional display of differential resistance), save/print report
Yes
No
No
No
Tes t OK?
Framed with dashed line: The test is only run if it has been activated in the initial program window, or in the Setup menu under
Sequence ...
Switch line voltage to test socket, start function test
Start
function test
IL
OK?
Switch line voltage to test socket, start measurement
Reverse mains polarity, perform measurement again
Yes
Contact all exposed, conductive parts with the probe. The test can be repeated as often as desired for various protective conductor parts with the manual sequence*.
* If it is not clear whether or not all
exposed, conductive parts are connected to one another or to the protective conductor, testing can be performed in the manual mode.
Switch DUT on
Page 32
11.6 Testing Devices in Accordance with DIN VDE 0702:2004
The following measurements can be performed in accordance with the above mentioned standard:
• Protective conductor R
PE
(permanent connection or with plug)
Test current: DC ±200 mA
• Insulation measurement R
ISO
(can be deactivated, e.g. if danger exists that voltage sensitive components in data processing systems might be damaged) plus equivalent leakage
current
or
• Contact current for safety class I I or
• Differential current
Check connection parameters and start test.
To Socket This is the default setting. Refer to chapter 7 on page
14 for other types of connection.
Class If the test instrument is equipped with features B00,
01 or 03, and if the DUT has been connected to the test socket, safety class recognition is performed (SC I or SC II). Otherwise, the safety class must be specified manually.
Ext. Cable x: Extension cables or connector cables which are
longer than 5 m can be tested with the help of the EL1 adapter (optional), either separately or in combination with a device.
ISO-R x: Insulation resistance measurement is performed.
ID No. See parameters database in chapter 11.2 on page
27.
Setup See chapter 11.2 on page 27 regarding setup of the
measuring sequence.
Page 33
GOSSEN METRAWATT GMBH 33
Test Sequence per VDE 0702:2004
Select connection, select test regulation VDE 0702, classify DUT (SC I, II or III), extension cable X/– (with/without)
Visual inspection
OK?
SC I: RPE
OK?
Short-circuit
at DUT?
Yes /No
No
Display results, save/print report
Yes
No
Yes
No
Tes t OK?
Framed with dashed line: The test is only run if it has been activat­ed in the initial program window, or in the Setup menu under Sequence ...
Switch line voltage to test socket, start function test
Contact all exposed, conductive parts with the probe
Start
function test
Switch DUT on
I
L
*
OK?
R
ISO
OK?
Yes
No
R
ISO
and IL OK: Yes
I
L
*
OK?
R
ISO
and IL OK?
Yes /No
* Residual current,
contact current (direct) or equivalent leakage current
If there are reservations against an insulation measurement
No
Page 34
11.7 Testing Extension Cables for VDE 0701, Part 1, and VDE 0702 (option: EL1 adapter)
Extension Cables up to 5 m Long
Protective conductor resistance between the earthing contact at the mains plug and all exposed metal parts may not exceed 0.3 Ω for safety class I devices. This value may not exceed 1 Ω for permanently connected data processing systems (DIN VDE 0701, part 240).
Extension and Connector Cables Longer than 5 m
Per DIN VDE 0702, an additional cable resistance of 0.1 Ω, however, not more than 1 Ω, is allowable as of a length of 5 m for each additional
7.5 m.
Resistance testing for cables longer than 5 m is thus advisable (see also limit values on page 17).
Note
The EL1 accessory adapter is required for testing for short­circuiting and interruption of single-phase extension cables.
Connecting the Extension Cable or the Multiple Outlet
Note
The keys in the adapter’s handle have no function.
Performing the Test
Connect the extension cable to the EL1 adapter as shown in the figure above.
Select the “Ext. Cable” test from the initial menu with the cursor and acknowledge with the
key: Ext. Cable x WITH EL1.
Select “Start Test” with the key.
Start the measurement sequence with the key.
First perform and acknowledge visual inspection of the extension cable.
Enter the length of the cable with the and keys. Acknowledge with the
key.
PROBE 5 4
(21)
EL1
SECUTEST
Device Under Test = Extension Cable
Page 35
GOSSEN METRAWATT GMBH 35
11.8 Testing Multiple Outlets for VDE 0702 (optional EL1 adapter)
up to version 5.10:
Do not select “WITH EL1” in the initial menu. The following must appear in the line: Ext. Cable – WITH EL1. “Manual sequence” must be selected.
from version 5.11 onwards:
Select “X WITH EL1” in the initial menu. The following must appear in the line: Ext. Cable X WITH EL1. “Manual sequence” must be selected.
A visual inspection must always be performed. It may thus be
necessary to unreel the cable from its drum or reel.
Protective conductor resistance measurement: Contact the first outlet
with the EL1 adapter. Each time you are ready to contact the next outlet, press the
key to repeat the test.
After measurement has been completed for all outlets, reinsert the test
probe in order to continue with the performance of subsequent tests as described in chapter 11.6.
Page 36
11.9 Testing in Accordance with DIN VDE 0751
The following measurements can be performed in accordance with the above mentioned standard:
• Protective conductor measurement R
PE
, test current: 200 mA DC
• Insulation measurement R
ISO
(can be additionally activated)
– R-ISO LN-PE
(Insulation resistance LN to protective conductor)
• Equivalent device leakage current I
EDL
• Device leakage current (direct measurement or by means of differential
current)
• Patient leakage current with probe
(Patient leakage current is not possible with SFC „Mains at Application Part“)
• Equivalent patient leakage current with probe
Leakage current is converted to reference voltage (see limit values, chapter 8 on page 15). Reference voltage must be matched to the supply voltage range.
Check connection parameters and start test.
To Socket This is the default setting. Refer to chapter 7 on page 14 for
other types of connection.
Class If the test instrument is equipped with features B00, 01 or 03,
and if the DUT has been connected to the test socket, safety class recognition is performed (SC I or SC II). In all other cases, or if it is not clear whether or not all exposed, conductive parts are connected to one another or to the protective conductor, the safety class can be selected manually.
Type Select the type of device under test from a list (device type).
If “Old devices” is selected, limit values specified in DIN VDE 0702 are used.
Ap. parts... (BF): The type of application parts can be changed manually:
Select the “Ap. Parts” line with the
or key, acknowledge
with the
key and change with the or key.
ID No. See parameters database in chapter 11.2 on page 27. Setup Refer to chapter 11.2 on page 27 regarding setup of the
measuring sequence.
Page 37
GOSSEN METRAWATT GMBH 37
Test Sequence per VDE 0751
Select connection, select test regulation VDE 0751, classify the DUT (SC I or II), application part? (type B/BF/CF)
Visual inspection
OK?
SC I: R
PE
OK?
R
ISO
OK?
Short-circuit
at DUT?
Yes /No
No
Display results, save/print report
Yes
No
Tes t OK?
Framed with dashed line: The test is only run
If it has been activated in the initial
program window, or in the Setup menu under Sequence ...
or – If at all possible
Switch line voltage to test socket, start function test
Start
function test
I
EDL
ΔI
Contact all exposed, conductive parts with the probe. The test can be repeated as often as desired for various protective conductor parts with the manual sequence*.
* If it is not clear whether or not all
exposed, conductive parts are connected to one another or to the protective conductor, testing can be performed in the manual mode.
Contact all exposed, conductive parts with the probe.
Yes
Switch DUT on
I
DL
No
ΔI or IDL: only if the device is switched electronically
OK?
No
NoNo
Yes
Yes
I
EPL
I
PL
OK?
NoNo
Yes
Application parts B, BF, CF
Page 38
12 Storage in PSI Module (Feature E01)
and Database Operations (Feature KB01 or DBmed option)
12.1 Storing Measurement Data in the PSI Module
Upon completion of a measurement – „Test passed/Test failed“ is shown on the display – you can save the measurement data to the memory of the PSI module.
Press the STORE key at the PSI module to this end.
A text entry field is shown.
You may now enter your comments on the measurement and/or an
identification number.
Press the STORE key once more to save the measurement data and
your comments. The message: „Data are stored“ is displayed.
A detailed description is given in the operating instructions of your PSI module under chapter „How to display, print and store protcol“.
12.2 Database Operations
12.2.1 Setting Up Test Sequences with PC Software
This function must be activated with the help of upgrade software, e.g. the Z853H program.
Up to 99 test sequences can be uploaded for each of the selector switch positions from a PC via the serial port at the SECUTEST
®
SII test instru­ment. After testing has been performed, the test results are saved to the same database, if a PSI module has not been connected. The maximum possible sum of test sequences and test results is 127.
Test results are stored to the PSI module if one has been connected. The database can be used in different ways:
(see also chapter 11.2 on page 27)
1. Stored test results can be displayed
(select the ID no. and then activate the “from database” function). The desired test sequence can be selected with the scroll keys.
2. If “ID no. = test sequence” has been activated (X), the first two
characters of the ID number determine the test sequence.
Example: ID number = 037890sk3r
Test sequence number 03 is executed. Test sequence 03 is the test sequence which appears at the third position from the top of the list.
If this number does not exist, the default sequence is executed.
3. Executed test sequences are stored to memory at the instrument as test results. The test results can also be displayed at a later point in time with the “test sequence settings” submenu in the “general setup” menu.
Note
The database can only be generated or deleted with the help of a PC with a terminal program, or appropriate application software.
12.2.2 Storing Test Results to the SECUTEST
®
SII
This function must be activated with the help of upgrade software, e.g. the Z853H program.
If no PSI module has been connected, up to 99 reports can be stored to the test instrument. The reports can be viewed as required at the instru­ment and can be printed out, for example with the help of DA-II software or a terminal program.
The reports are sorted by time and date and are displayed with the ID number. If no ID number was assigned, date and time are automatically saved instead.
13 Remote Control (Feature KD01 or SK5 Option)
This function must be activated with the help of upgrade software, e.g. the Z745K program.
The protective conductor measurement is expanded to include the function: “automatic recognition of measuring point change”. During protective conductor measurement, the instrument recognizes whether or not the probe is in contact with the protective conductor, and indicates these two possible conditions by means of different acoustic signals.
This function is helpful if several protective conductor connections need to be tested. The function can be activated with the “auto measuring point” function in the “test sequence setup” menu (see “changes to the switch position menu” above).
Page 39
GOSSEN METRAWATT GMBH 39
14 Printing via PSI Module (Feature E01)
or with the Printer Adapter (DA-II accessory)
14.1 Print Out Test Data at the PSI Module
Upon completion of a measurement – „Test passed/Test failed“ is shown on the display – you can print out the measurement data via your PSI module.
Press the PRINT key at the PSI module to this end.
The print menue is displayed.
Select „Protocol“ with your cursor and acknowledge with
.
Press the PRINT key once more to start the print-out.
A detailed description is given in the operating instructions of your PSI module.
14.2 Printing Test Results in Report Form
Test instruments can be connected to commercially available printers with the DA-II printer adapter (Z745M), even if they are not equipped with a parallel port, thus allowing for on-site print-out of test reports etc.
The only prerequisite is that the printer module is set up for direct report generation.
Note
Only printers disposing of the ASCII type font are suitable for use.
Connecting a Centronics Printer
Connect the test instrument to a Cen­tronics printer via the RS 232 port with the help of the DA-II adapter. The PSI module may not be connected.
You can shift to the Report menu from any of the displayed test results (1
st
page) with the help of the key.
Measurement results for the current test can be stored to memory, the re­sults of the current test can be printed out to the corresponding report form, previously stored test results can be queried (scroll: DBmed option, see chapter 12) and all saved measurement results can be printed out from this menu.
Print-out is executed directly to the Centronics printer via printer adapter DA-II.
The matching report form is automati­cally selected for the standard selected with the selector switch.
15 Direct Print-Out (feature KE01, SECU-dd option)
After completion of each test (individual test or at the end of a test sequence), test results are read out directly via the RS 232 interface. If the SE­CUTEST
®
PSI has been connected, the results are printed directly to paper.
Page 40
16 Characteristic Values
1)
The device leakage current step of the test sequence is executed by means of
differential current measurement for testing per DIN VDE 0751.
2)
As of 25 mA: shutdown by residual current measurement within 100 ms
3)
Measured value P and calculated value S are compared and the smaller of the two is displayed.
4)
Measuring circuit becomes highly resistive, indication at display
5)
For DIN VDE 0701/0702: 2 kΩ, for DIN VDE 0751: 1 kΩ
6)
This measuring range with DIN VDE 0751 only
Func-
tion
Measured Quantity Measuring Range /
Nominal Range of
Use
Resolu-
tion
Nominal
Voltage
U
N
Open-Cir-
cuit Volt-
age U
0
Nominal
Current
I
N
Short­Circuit Cur. I
K
Internal
Resis-
tance R
I
Reference
Resist.
R
REF
Measuring
Error
Intrinsic Error Overload Capacity
Value Duration
Tests per DIN VDE 0701 / 0702 / 0751
Device Protective
Conductor
Resistance R
PE
0.000 2.100 Ω 1mΩ
4.5 9V DC
>200 mA
DC
——
±(5% rdg. + 10 d)
>10d
±(2.5% rdg. + 5d)
> 10 d
253 V cont.
2.11 31.00 Ω 10 mΩ
0.050 1.500 MΩ 1kΩ 50 500
VDC
1.0 • U
N
1.5 • U
N
>1mA <10mA
±(5% rdg. +10 d)
±(2.5% rdg. +5 d)
> 10 d
253 V cont.
Insulation Resistance
R
ISO
1.01 10.00 MΩ 10 kΩ
10.1 310.0 MΩ 100 kΩ±(10% rdg. +10 d) ±(10% rdg. +10d)
Equivalent Leakage
Current
IEL or. I
DL
0.00 21.00 mA 10 μA
230 V~
– 20/
+10 %
< 3.5 mA > 72 kΩ 1/2 kΩ
5
±(5% rdg. +10 d)
±(2.5 % rdg. +5 d)
> 10 d
253 V cont.
20.1 120.0 mA 100 μA
Equivalent Patient Leakage Current
I
EPL
0.0 ... 310.0 μA100nA
230 V~
– 20/
+10 %
< 3.5 mA > 72 kΩ
1kΩ
±10 Ω
±(5% rdg. +10 d)
±(2.5% rdg. +5 d)
> 10 d
253 V cont.0.300 ... 2.100 mA 1 μA
2.00 ... 11.00 mA 10 μA
Contact Current or
Housing Leakage
Current I
Probe
or I
DL
0 310 μA
6
0.300 3.500 mA
0.1 μA 1 μA
————1/2kΩ
5
±(5% rdg. +10 d)
±(2.5% rdg. +5 d)
> 10 d
253 V cont.
Patient Leakage
Current I
PL
AC/DC Compo-
nents Measured
Separately
0.0 ... 310.0 μA100nA
————1kΩ ±(5% rdg. +10 d)
±(2.5% rdg. +5 d)
> 10 d
253 V
cont.
2 4
0.300 ... 3.100 mA 1 μA
3.10 ... > 15.00 mA 10 μA
Differential Current ΔI
between L and N
1
0.000 … 3.100 mA
~
3.00 31.00 mA~
2
1 μA
10 μA
— — ————
±(10% rdg. +10 d)
> 10 d
±(5% rdg. +5 d)
> 10 d
22
Function Test
Line Voltage U
L–N
207.0 ... 253.0 V~ 0.1 V ±(2.5% rdg. +5 d) 253 V cont.
Load Current I
V
0 ... 16.00 A
RMS
10 mA ±(2.5% rdg. +5 d) 20 A 10 min.
Active Power P 0 ... 3700 W
3
1W
±(5% rdg. +10 d)
>20d
253 V cont.
20 A 10 min.
Apparent Power S 0 ... 4000 VA 1 VA Calculated value U
L–N
• I
V
±(5% rdg. +10 d)
>20s
Power Factor LF,
sinusoidal: cos
ϕ
0.00 ... 1.00 0.01 Calculated value P / S, display > 10 W ±(10% rdg. +5 d)
Differential Current
ΔI between L and N
per DIN VDE 0702
0.00 ... 31.00 mA~ 10 μA— —
±(10% rdg. +10 d)
> 10 d
±(5% rdg. +5 d)
22
U
Probe
Probe voltage
0 ... 253.0 V
, and
0.1 V
±(2.5% rdg. +5 d)
>10d
253 V cont.
Page 41
GOSSEN METRAWATT GMBH 41
Multimeter Measurements for Devices with Feature F01
Reference Ranges
Line Voltage 230 V ±0.2 % Line Frequency 50 Hz ±0.1 % Waveshape Sine (deviation between effective and rectified
value < 0.5%) Ambient Temperature +23 °C ±2K Relative Humidity 40 60% Load Resistance Linear
Nominal Ranges of Use
Line Voltage 207 V 253 V Line Frequency 50 Hz Line Voltage
Wave sh ape Sine Temperature
0 °C +50 °C
Ambient Conditions
Storage Temperature – 20 °C ... + 60 °C Operating Temperature – 10 °C ... + 50 °C Accuracy Range 0 °C ... + 50 °C Relative Humidity Max. 75%, no condensation allowed Elevation Max. 2000 m Deployment Indoors, except within specified ambient
conditions
Influencing Quantities and Influence Error
Func­tion
Measured Quantity Measuring Range /
Nominal Range of Use
Reso­lution
Open­Circuit Voltage U
0
Short­Circuit Current I
K
Internal Resis­tance R
I
Measuring Error
Intrinsic Error Overload Capacity
Value Duration
U
AC/DC
Voltage
0 ... 253.0 V
,
and
0.1 V
±(2.5 % rdg. +5 d)
>10d
253 V cont.
Low Voltage, SC II I
±(5% rdg. +10 d)
R Resistance 0 ... 150.0 k
Ω 100 Ω <20V– 1.1mA ±(1% rdg. +3 d) 253 V cont.
I
Clip
Current with
Clip-On Current-
Voltage Transformer
WZ12C
0.000 ... 10.00 A 1mA 1.5MΩ
±(3% rdg. +10 d)
>10d without clip-on device
253 V cont.
0 ... 100 A
1A 1.5MΩ 253 V cont.
Tem p.
Tem pe ra tu re
with Pt100/Pt1000
Sensor
– 200 ... – 50
°C1°C
<20V– 1.1mA
±(2% rdg. +1 °C) 10 V cont.
– 50.1 ... + 300.0
°C0.1°C ±(1% rdg. +1 °C) 10 V cont.
+300 ... +850
°C1°C ±(2% rdg. +1 °C) 10 V cont.
Influencing Quantity / Sphere of Influence
Designation per DIN VDE 0404
Influence Error ± % of Measured Value
Position Change E1
Change in Test Setup Supply Power E2 2.5
Temperature Fluctuation
E3
Specified influence error applies per 10 K change in temperature:
0 21
°C and 25 … 40 °C
1 in case of PE measurement
0.5 for all other measuring ranges
Current at Device Under Test E4 2.5
Low-Frequency Magnetic Fields E5 2.5
Impedance at Device Under Test E6 2.5
Capacitance, Insulation Measurement E7 2.5
Waveshape of Measured Current
E8
49 51 Hz 2 for capacitive load (for equivalent
leakage current)
45 100 Hz 1 (for contact current)
2.5 for all other measuring ranges
Page 42
Power Supply
Line Voltage 207 V 253 V Line Frequency 50/60 Hz Power Consumption Approx. 30 VA
for function test Continuous max. 3600 VA,
power is conducted through the instrument only, switching capacity ≤ 16 A
Electrical Safety
Safety Class I per IEC 61010-1/EN 61 010-1/VDE 0411-1 Nominal Voltage 230 V Test Voltage 3.7 kV 50 Hz Measuring Category II Fouling Factor 2 Safety Shutdown Where differential current at the
DUT > 25 mA Shutdown time: < 100 ms Probe current: > 10 mA, < 1 ms
Electromagnetic Compatibility
Generic Standard DIN EN 61326:2002
Mechanical Design
Display Multiple dot matrix display, 128 x 128 pixels Dimensions LxWxH: 292 mm x 138 mm x 243 mm Weight Standard instrument: approx. 4.0 kg Protection Housing: IP 40
Terminals: IP 20 per DIN VDE 0470 part 1 / EN 60529
Extract from table on the significance of IP codes
RS 232 Interface
Type RS 232C, serial, per DIN 19241 Configuration 9600, N, 8, 1 Connection 9-pin subminiature socket connector
Interference Emission Class
EN 55022 B
Interference Immunity Test value Feature
EN 61000-4-2 Contact/atmos. - 4 kV/8 kV A
EN 61000-4-3 10 V/m C
EN 61000-4-4 Mains connection - 2 kV B
EN 61000-4-5 Mains connection - 1 kV A
EN 61000-4-6 Mains connection - 3 V A
EN 61000-4-11 0.5 period / 100% A
IP XY
(1
st
digit X)
Protection against
foreign object entry
IP XY
(2nd digit Y)
Protection against the
penetration of water
0 not protected 0 not protected 1 50.0 mm dia. 1 vertically falling drops
2 12.5 mm dia. 2
vertically falling drops with
enclosure tilted 15° 3 2.5 mm dia. 3 spraying water 4 1.0 mm dia. 4 splashing water
Page 43
GOSSEN METRAWATT GMBH 43
17 RS 232 Interface
The RS 232 port is used to connect the SECUTEST®PSI module (acces­sory), which can be installed to the lid of the SECUTEST
®
SII, to a PC or a
barcode scanner.
17.1 Transmission of Measurement Results to the SECUTEST
®
PSI
Test results – except for results from individual measurements (selector switch in MENU position) and the function test – can be transmitted from the SECUTEST
®
SII to the SECUTEST®PSI module, where they can be stored and printed out in the form of measuring, test and statistical reports at any time.
17.2 PC Connection
Connection to an IBM compatible PC is also possible. The PC is connected to the interface at the test instrument, or to the interface port at a previously installed SECUTEST
®
PSI module.
17.2.1 Software Evaluation of Measurement Results
Convenient software programs such as PC.doc-WORD, ACCESS or PS3 allow for easy generation of measuring and test reports, as well as archiving of measured data.
17.2.2 Instrument Control via Interface Commands
All key functions included with the SECUTEST
®
SII can be simulated with the help of interface protocols, and the following parameters can be queried:
• Type of measurement and measuring range
• Test setup
• Measurement sequence progress
• Detailed measurement results
17.3 Interface Configuration and Protocol
The interface included with the SECUTEST
®
SII is in compliance with the
RS 232 standard. Technical Data: Baud Rate 9600 baud, permanently set
Character Length 8 bits Parity none Stop Bits 1 Data Protocol per DIN 19244, X_ON / X_OFF protocol
Connector Pin Assignments for 9-Pin Subminiature Socket Connector:
1: External in + (for internal use only) 2: TXD (transmitter output) 3: RXD (receiver input) 4: External in + 5: GROUND 6: +5 V (500 mA output, for barcode scanner only) 7: Ext. in – 8: Control output 9: +9 V (1.5 A output, for PSI module only)
9
876
54
3
21
Page 44
18 Appendix
18.1 Evaluation of Measured Values for Individual Measurements as well as for Calculated Quantities
In order to assure that the limit values for the individual measurements are always observed, device measuring error must be taken into consideration. The table in the appendix allows for calculation of the required minimum display value for each respective measurement which must appear at the instrument in consideration of measuring error (under nominal conditions of use), in order to assure that the required limit value is not fallen short of (DIN VDE 0413, part 1). Intermediate values can be interpolated.
Measuring Error for Test Sequences
The test instrument takes respective measuring error into consideration during automatic test sequences, and corrected results are entered into the test report, as long as this function has been activated in the setup switch position under “include service error”.
Tables for the calculation of minimum display values for insulation resistance and maximum display values for protective conductor resistance, equivalent leakage current, probe current and residual current in consideration of device measuring error:
1)
Depending upon resolution
R
ISO
MΩ RPE Ω
Limit Value
Minimum
Display Value
Limit Value
Maximum
Display Value
0.100 0.115 0.100 0.085
0.250 0.273 0.200 0.180
0.500 0.535 0.300 0.275
1.000 1.060 0.400 0.370
2.000 2.200 0.500 0.465
5.000 5.350 0.600 0.560
7.000 7.450 0.700 0.655
10.00 10.60 or 12.5
1
0.800 0.750
20.00 23.00 0.900 0.845
75.00 83.50 1.000 0.940
1.100 1.035
I
EL
mA I
probe
mA IRI mA
Limit Value
Maximum
Display Value
Limit Value
Maximum
Display Value
Limit Value
Maximum
Display Value
1.00 0.85 0.100 0.085 0.25 0.12
3.50 3.23 0.250 0.227 0.50 0.35
7.00 6.55 0.500 0.465 1.00 0.80
10.00 9.40 1.000 0.940 2.00 1.70
15.00 14.15 2.000 1.890 3.50 3.05
20.00 18.90 3.500 3.315 5.00 4.40
7.00 6.20
10.00 8.90
15.00 13.40
20.00 17.90
25.00 22.40
Page 45
GOSSEN METRAWATT GMBH 45
18.2 Index
A
Absence of Voltage ................................................ 6, 7
Acoustic Signal, Measurement ................................. 15
Acoustic Signal, Sequence ....................................... 15
Adapter for Socket ................................................... 27
Alternating / Direct Voltage UAC/DC ......................... 22
Auto Class PSI ........................................................ 15
Auto Point ............................................................... 15
Auto Test Method .................................................... 27
Auto-Store .............................................................. 27
C
Classification ..................................................... 12, 27
Configuring Device Parameters .......................... 11, 15
Configuring Measurement Parameters ...................... 11
Connecting the Device Under Test ............................ 14
Contact Current ......................................... 6, 7, 14, 19
Contrast Adjustment ................................................ 11
D
DBmed Option ......................................................... 38
Device Leakage Current ................................... 6, 7, 19
Differential Current Method ........................................ 7
Direct Printing ......................................................... 15
E
Earth Fault .............................................................. 15
Equivalent Device Leakage Current ............. 6, 7, 20, 21
Equivalent Leakage Current .............................. 6, 7, 21
Extension Cables ..................................................... 34
F
Features ................................................................... 7
Finger Contact ........................................................ 10
First Measured Values ............................................. 28
Frequency Response ............................................... 19
Function Test .......................................................... 24
I
Illumination ..............................................................15
Incl. Service Error ....................................................15
Individual Measurements .......................................... 16
Insulation Resistance ...........................................6, 17
Insulation Resistance Limit Values ............................18
Interface ..................................................................43
IT Network ...............................................................15
L
Limit Values ............................................................15
M
Mains Connection Errors ..........................................10
Mains Power Outlet ....................................................9
Mains Wait ..............................................................15
Manual Sequence ....................................................27
Measurements with Accessories ............................... 22
Measuring Error ....................................................... 44
Modem ...................................................................15
O
Online Help .............................................................10
P
Patient Leakage Current ............................................. 6
Periodic Testing ......................................................... 6
Polarity Reversal ...................................................... 27
Printer Adapter ........................................................ 39
Probe Voltage Uprobe ..............................................21
Protective Conductor Resistance ................................. 6
R
Reference Voltage ..............................................15, 36
Reports ...................................................................15
Residual Current ..................................................6, 19
Residual Current Method ............................................ 7
Resistance R ...........................................................22
R-ISO LN-PE ............................................................27
S
Safety Class I Devices .............................................. 12
Safety Class II Devices ............................................. 12
Safety Class III Devices ............................................ 12
Saving Settings to Memory ...................................... 11
SC I I I UV ............................................................... 27
Service ................................................................... 15
Setting Date and Time ............................................. 15
Setting Limit Values ................................................. 11
Short-Circuit Test .................................................... 25
Single Fault ............................................................. 15
SK5 Option ............................................................. 38
T
Test Current .............................................................. 6
Test Sequence ........................................................ 15
Test Time ............................................................... 15
Testing after Repairs ................................................. 6
Type B Application Parts .......................................... 12
Type BF Application Parts ........................................ 12
Type CF Application Parts ........................................ 12
V
Visual Inspection ..................................................... 27
Z
Zero Balancing .................................................. 17, 23
Page 46
19 Maintenance - Calibration
Housing Maintenance
No special maintenance is required for the housing. Keep outside surfaces clean. Use a slightly dampened cloth for cleaning. Avoid the use of cleansers, abrasives or solvents.
Calibration
According to the new DIN VDE 0702:2004-06 standard the following ap­plies as from 1/6/2004:
„Measuring instruments used for periodic testing must be inspected and calibrated at regular intervals.“
Please contact us to fix an appointment for the inspection and calibration of your test instruments: GOSSEN METRAWATT Calibration Center, Phone 0911-8602-256, address see below.
Device Return and Environmentally Compatible Disposal
The instrument is a category 9 product (monitoring and control instru­ment) in accordance with ElektroG (German Electrical and Electronic De- vice Law). This device is not subject to the RoHS directive.
We identify our electrical and electronic devices (as of August 2005) in accordance with WEEE 2002/96/EG and ElektroG with the symbol shown to the right per DIN EN 50419 .
These devices may not be disposed with the trash. Please contact our service department regarding the return of old devices.
20 Repair and Replacement Parts Service
DKD Calibration Laboratory* and Rental Instrument Service
If required please contact:
GOSSEN METRAWATT GMBH Service Center Thomas-Mann-Str. 20 90471 Nürnberg, Germany Phone: +49-(0)-911-8602-0 Fax: +49-(0)-911-8602-253 E-mail: service@gossenmetrawatt.com
This address is only valid in Germany. Please contact our representatives or subsidiaries for service in other countries.
* Calibration Lab for Measured Electrical Quantities: DKD – K – 19701
Accredited per DIN EN ISO/IEC 17025
Accredited quantities: direct voltage, direct current value, direct current resistance, alternating voltage, alternating current value, AC active power, AC apparent power, DC power, capacitance, frequency
Page 47
GOSSEN METRAWATT GMBH 47
Competent Partner
GOSSEN METRAWATT GMBH is certified in accordance with DIN EN ISO 9001:2000.
Our DKD calibration lab is accredited by the Physikalisch-Technischen Bundesanstalt (German Federal Institute of Physics and Metrology) and the Deutscher Kalibrierdienst (German Calibration Service) in accordance with DIN EN ISO/IEC 17025 under registration number DKD–K–19701.
We offer a complete range of expertise in the field of metrology: from test
reports and factory calibration certificates, right on up to DKD calibration certificates. Our spectrum of offerings is rounded out with free test equip­ment management.
Our service department includes an
on-site DKD calibration bench
. If errors are discovered during calibration, our specialized personnel are capable of completing repairs using original replacement parts. As a full service cali­bration lab, we can calibrate instruments from other manufacturers as well.
21 Product Support
If required please contact:
GOSSEN METRAWATT GMBHProduct Support Hotline Phone: +49-(0)-911-8602-112 Fax: +49-(0)-911-8602-709 E-mail support@gossenmetrawatt.com
Page 48
Edited in Germany Subject to change without notice PDF version available from the Internet
GOSSEN METRAWATT GMBH Thomas-Mann-Str. 16-20 90471 Nürnberg •
Germany
Phone: +49-(0)-911-8602-0 Fax: +49-(0)-911-8602-669 E-mail: info@gossenmetrawatt.com www.gossenmetrawatt.com
Loading...