Gossen MetraWatt SECULIFE ST, SECULIFE STHV Operating Instructions Manual

Page 1
Operating Instructions
SECULIFEST and SECULIFEST HV
Test Instruments for Portable Appliance Testing according to Health and Safety Policy and in Accordance with the Medical Product Law as well as for Routine Testing
3-349-448-03
16/11.11
Page 2
45678910
!
24
23
22
21
25
22
26
27
20
19
Probe connection
Insert the double plug of the probe into sockets 4 and 5 such that the plug with the white ring contacts socket 5 (vertical bar).
Note: Contact problems with exposed conductive parts when using the standard probe with test tip
In order to assure good contact, surface coatings must be removed from devices under test with special tools at a suit­able location so that the surface has a metallic shine. The tip of the test probe is not suitable for scratching away paint, be­cause this may impair its coating and/or mechanical strength. The brush probe (Z745G) may be more suitable than the test probe in certain individual cases.
Measurements at jacks 1 – 2 – 3
Always start with the measurement before contacting the measuring point. Between jacks 1 and 2, a maxi-
11
12
13
RS232
IOIOI
PROBE
SONDE
Setup
3
2
1
HV
Function Test
V
Aux
Applied Parts AC E G I
BDFHK
!
Iso
I leakage
/
PE
Auto
L
3
max. 253 V
N
2
max. 10 V
SL
1
mum of 10 V may be applied. Between jacks 2 and 3,
14 15 1816 17
up to 253 V may be applied.
Attention: Jacks (2) and (3) are short-circuited during all measurements at the test socket! (Exception: see chapter 12.2)
Standard equipment
1
Test instrument with 10 + 2 connectors for application parts 1 Probe cable with test probe 1 Clip-on alligator clip for test probes 3 Clip-on quick-clamp terminals 10 conductor patient connection cable 2 mm 10 Clip-on alligator clip 2 mm 1 Calibration certificate per DKD 1 Operating instructions 1Carrying strap
Up-to-date PC software (free start-up programs or demo software for data management, report and list generation) is available on our website for download.
These operating instructions describe an instrument with firmware version 7.24.
2 GMC-I Messtechnik GmbH
Page 3
1 Jack for protective conductor at device under test
Note
i
!
2 Jack for neutral conductor at device under test 3 Jack for phase conductor at device under test 4 Jack for connecting the probe 5 Jack for connecting the probe 6 Function selector switch
– Function Test: Function test – Auto: Automatic test sequence according to selected standard – PE: Protective conductor test – Iso/HV: Insulation test / high-voltage test – I leakage: Leakage current measurement – V : Multimeter functions – Aux: Auxiliary multimeter functions
– Setup: Device configuration 7 scroll key for menu and parameter selection 8 scroll key for menu and parameter selection 9 LCD window
10 Socket connector interface RS232 for (P)SI module SECUTEST PSI/SI+, storage
adapter
SECUSTORE
, barcode or RFID scanner 11 Signal lamp for mains connection error 12 key for entry and for starting test sequences and finger contact 13 help key (context sensitive) 14 Key next to the symbol for switching test voltage to
the test socket (only possible if symbol LED is blinking) 15 Signal lamp for the functions test 16 Functional earth (equipotential bonding) 17 Operational earth 18 Connector jacks for application parts 19 Push-buttons (left and right) for releasing the handle from its snap-in position 20 Earthing contact socket for service purposes (Feature B01),
e.g. for connecting a notebook or an A4 format printer, Terminal Data see page 63 21 Standard outlet socket (test socket) for connecting the device under test 22 Push-buttons (left and right) for releasing the lid 23 Lid 24 Compartment for probe and accessories 25 Cover or (P)SI module (accessory
SECUTEST PS I or SECUTEST SI+)
26 Carrying handle and tilt stand 27 Test probe (accessory probe with coil-cable SK2W (Z745N))
Overview of Available Probe Types
Probe Type Application Special Features
Standard probe (test probe with coil-cable and alligator clip)
1)
SK2
1)
SK2W
Feature KD01 with probe SK5 Restriction with
Brush probe Can be plugged onto all above listed probes and test probes
1)
1)
Accessory
Max. test current: 25 A Probe with cable
Max. test current: 25 A Probe with cable (no coil-cable),
Max. test current: 25 A Probe with cable (coil-cable),
Feature G01 (I short-circuit current < 25 A
Leakage current, protective conductor resistance
>25A):
K
(no coil-cable)
2 meters long
2 meters long
Special probe in combination with “automatic recognition of measur­ing point change” function (see chapter 17)
For contacting devices under test with rotating, vibrating, exposed conductive parts
when using other probes than those specified above
The cables plugged into the sockets (4 and 5) must be short circuited for testing with the probe, i.e. by plugging the ends of the cable together, or via a conductive surface at the device under test (4-wire measurement). Remove any corrosion from the device under test.
Data Security
Measurement data, report data and user entries are stored to RAM at the (P)SI module (accessory), as long as the respective battery supplies the required amount of voltage.
Save your data to a PC on a regular basis in order to prevent any loss of data at the (P)SI module. We assume no liability for data loss. We recommend the following PC programs for data processing and data management:
PS3 (transmission of measurement data to a PC, documentation,
management, report generating and deadline follow-up)
PC.doc-WORD/EXCEL (report and list generation
PC.doc-ACCESS (test data management)
ELEKTROmanager/PROTOKOLLmanager for SECUTEST...
patManager (Report and list generation)
GMC-I Messtechnik GmbH 3
Page 4
Contents Page Contents Page
1 Applications ...................................................................................6
8 Configuring Device Parameters ...................................................18
1.1 Table: Types of DUTs – Tests – Regulations .............................................. 6
1.2 Table: Individual Measurements and Regulations .....................................7
1.3 Table: Leakage Current Types ...................................................................7
1.4 List of Possible Options and Standard Types .............................................. 8
9 Measuring Protective Conductor Resistance ...............................19
9.1 Maximum Allowable Limit Values for Protective Conductor Resistance
for Connector Cables with a Length of up to 5 m ..................................... 20
2 Safety Features and Precautions ...................................................9
2.1 Notes Regarding the High-Voltage Test (only Feature F02 or SECULIFE ST
HV) .......................................................10
3 Initial Start-Up ..............................................................................11
3.1 Connection to the Mains (115 V / 230 V, 50 Hz / 60 Hz) ............................11
3.2 Automatic Recognition of Mains Connection Errors ..................................12
4 General Notes ..............................................................................12
4.1 Online Instructions ....................................................................................12
4.1.1 Changing the User Interface Language .........................................................12
4.1.2 Automatic Safety Class Selection ..................................................................13
4.1.3 Manual or Automatic Operating Sequences ...................................................13
4.2 Online Help ................................................................................................13
4.3 Adjusting Contrast .....................................................................................13
4.4 Configuring Device Parameters, Setting Date and Time ...........................14
4.5 Configuring Measurement and Sequence Parameters .............................14
4.6 Setting Limit Values ..................................................................................14
4.7 Saving the Settings ...................................................................................14
5 Classification of Devices Under Test ............................................15
5.1 Safety Classes ...........................................................................................15
5.2 Application Parts (electrical medical devices) ..........................................15
6 Abbreviations ...............................................................................16
7 Connecting the Device Under Test ...............................................17
10 Insulation Measurement .............................................................. 20
10.1 Insulation Resistance R
10.2 Equivalent Leakage Current ..................................................................... 22
10.3 High-Voltage Test (Feature F02 or SECULIFE ST
. ....................................................................... 20
INS
HV) ................................. 24
11 Leakage Current Measurement ...................................................26
11.1 Earth Leakage Current IPE (Feature KA01) ................................................ 26
11.2 Housing Leakage Current I
11.3 Patient Leakage Current I
11.4 Patient Auxiliary Current IPA (Feature KA01) ............................................. 27
11.5 Residual Current I
RC .......................................................................................................27
(probe current, contact current) ................. 26
HL
PL ......................................................................................... 27
11.6 Device Leakage Current ILC per IEC 62353 (VDE 0751-1) ......................... 27
12 Multimeter Functions ...................................................................30
12.1 Probe Voltage U
12.2 Alternating / Direct Voltage U
12.3 Resistance R ............................................................................................. 31
– Max. 300 V ........................................................... 30
probe
– Max. 253 V .....................................30
AC/DC
13 Measurements with Accessories .................................................32
13.1 Alternating Current IC with Clip-On Converter ..........................................32
13.2 Protective Conductor Resistance R
13.3 Temperature T with Pt100/1000 Sensor ................................................... 33
via Clip-On Meter ..........................32
PE
14 Function Test ...............................................................................34
4 GMC-I Messtechnik GmbH
Page 5
Contents Page Contents Page
15 Measurements in Accordance with National and
International Standards with Selector Switch in Auto Position ... 36
15.1 Test Sequences .........................................................................................36
15.2 Setting Up Test Sequences .......................................................................37
15.3 Configuring Measuring Parameters ..........................................................39
15.4 Testing Devices in Accordance with DIN VDE 0701, Part 1 .......................40
15.5 Testing Devices in Accordance with DIN VDE 0701, part 240 ...................42
15.6 Testing Devices in Accordance with DIN VDE 0701-0702:2008 ................44
15.7 Testing Extension Cables for VDE 0701-0702 (VDE 0701 Part 1)
(optional EL1 adapter) ...............................................................................46
15.8 Testing Multiple Outlets for VDE 0701-0702 (optional EL1 adapter) .........47
15.9 Testing in Accordance with DIN EN 60 950 ...............................................48
15.10 Testing Devices in Accordance with EN 61010 ........................................50
15.11 Testing Devices in Accordance with EN 60335 ........................................52
15.12 Testing in Accordance with IEC 62353/VDE 0751 .....................................54
15.13 Testing in Accordance with EN 60601 (Feature KA01) .............................56
16 Storing in (P)SI Module (Accessory) and Database Operations
(Feature KB01) ............................................................................. 58
16.1 Storing Measurement Data in the (P)SI Module ........................................58
16.2 Database Operations .................................................................................58
16.2.1 Storing Test Results to the Test Instrument ...................................................58
16.2.2 Uploading Report Templates into Test Instrument, Reading Out From the Test
Instrument, Editing at the PC and Re-Saving to the Test Instrument ................58
16.2.3 Reading Out and Saving Test Results / Test Data from the (P)SI Module ..........58
20 RS 232 Interface .......................................................................... 64
20.1 Transmission of Measurement Results to the (P)SI module .....................64
20.2 PC Connection ...........................................................................................64
20.2.1 Software Evaluation of Measurement Results ................................................64
20.2.2 Instrument Control via Interface Commands ..................................................64
20.3 Interface Definition and Protocol ..............................................................64
21 Appendix ..................................................................................... 65
21.1 Evaluation of Measured Values for Individual Measurements
as well as for Calculated Quantities .........................................................65
21.2 Evaluation of Measured Values during Equivalent Leakage Current
Measurement (Automatic Test Sequence According to Standard) ........... 65
21.3 Index ......................................................................................................... 66
22 Maintenance - Recalibration ....................................................... 68
22.1 Housing Maintenance ...............................................................................68
22.2 Recalibration ............................................................................................. 68
22.3 Safety Checks ...........................................................................................68
22.4 Device Return and Environmentally Compatible Disposal ........................69
23 Repair and Replacement Parts Service
Calibration Center and Rental Instrument Service ...................... 69
24 Product Support .......................................................................... 70
17 Recognition of Probe to Protective Conductor
(Feature KD01) ............................................................................. 58
18 Storing Test Results and Printing in Report Form ....................... 59
19 Characteristic Values ...................................................................60
GMC-I Messtechnik GmbH 5
Page 6

1 Applications

Attention!
!

1.1 Table: Types of DUTs – Tests – Regulations

Start-up and Modifications
Devices under test to be tested in accordance with the following regulations
EC 62353 :2007
I
Laboratory instruments Measuring and control in­struments
Voltage generation devices  Electric tools  Electric heating devices  Electric motor devices  Lamps 
Devices for entertainment electronics, information and communications technology
Cable reel, extension and connection leads
Data processing and office equipment
Electrical medical devices, application parts
Testing after Repairs
-0702:2008
DIN VDE 0701
DIN EN 62353:2008
(VDE 0751-1)





Periodic Tes t in g
-0702:2008 1
EC 62353 :2007
I
DIN VDE 070
DIN EN 62353:2008
Routine Testing
EC 62353 :2007
I
DIN EN 62353:2008
DIN EN 60950/50 116
DINEN61010
DIN EN 60335/50 106
IEC 60601/DIN EN 60601
The test instrument may not be used for measurements within electrical systems!
Applicable Standards
German National European International
DIN EN 61010 EN 61010 IEC 61010
DIN EN 60601 EN 60601 IEC 60601
DIN EN 60335-1 EN 60335-1 IEC 60335-1
DIN EN 60950 EN 60950 IEC 60950
IEC 62353 :2007
DIN EN 62353:2008
EN 62353 IEC 62353
(VDE 0751-1)
6 GMC-I Messtechnik GmbH
Page 7

1.2 Table: Individual Measurements and Regulations

Individual Measurements per Regulation
Tes t Cur re nt [ A]
DIN VDE 0701-0702
DIN VDE 0701
DIN VDE 0701
DIN VDE 0701
DINEN60950
DINEN61010
DINEN60335
IEC 62353
0.2
Protective Conductor Resistance
Insulation Resistance
Equivalent Leakage Current
High-Voltage Test
Equivalent (Device) Leakage Current
Equivalent Patient Leakage Current
Residual Current
Contact Current
Absence of Voltage (exposed conductive parts)
Housing Leakage Current
Earth Leakage Current
Patient Leakage Current
Total Patient Leakage Current
Patient Auxiliary Current
Device Leakage Current
Single Fault Conditions N
PE
Mains at Application Part

10
25
   
 AC AC
  

 



Key
Standards printed in grey half-tone will be superseded by the new DIN VDE 0701-0702:2008 standard.
Required test
IEC 601/EN 60 601 2nd
IEC 601/EN 60 601 3rd



1.3 Table: Leakage Current Types

DIN VDE 0701-0702
Equivalent Leakage Current
Contact Current/ Measurement for Absence of Volta ge
Protective Conductor Current with Residual Current Measurement
IEC 62353 (VDE 0751-1)
Equivalent Device Leakage Current
Equivalent Patient Leakage Current
Housing Leakage
Patient Leakage Current NC
Device Leakage Current during Operation, Direct Measurement
Device Leakage Current during Operation, with residual Current Measurement
DIN EN 60601-1 The following is
IEL interrupted from N
Current NC
Patient Leakage Current NC
Patient Auxiliary Current NC
Earth Leakage Current NC
Key
NC = normal condition PAP = patient application part PE = Potential earthing PC = Protective conductor of the DUT
, system protective conductor
measured:
PROBE (connected to protective conductor) to L & N
PROBE (connected to protective conductor) to L & N
L & N & PE to Patient Jacks
Probe to PE
Patient Jack to PE
Patient Jack to Patient Jack
Protective Conductor to PE
Protective Conductor Interrupted, Probe + PAP to PE
See chapter 11.5
GMC-I Messtechnik GmbH 7
Page 8

1.4 List of Possible Options and Standard Types

Features 00 01 02 03 04 05 06 07 08 09 10 11 XX
Mains Connection for Respective Country
User Interface Language C DUKF I ECZENL High Voltage Testing
HV-DC F without
AC Test Current 50/60 Hz for Protective Conductor Measurement
Test Sequence for IEC 60601
Data Memory for up to 125 Tests
Recognition of Probe to Protective Conductor
Direct Printing after each Measurement for Automatic Test Sequences Read-Out via RS232
1)
Each measured value is documented in this case, as opposed to the results of a test sequence for which only the poorest value for each given test is displayed. (via the PSI module, the SECUSTORE storage adapter or a PC)
5)
1)
B D
G 10 A 25 A
KA without with
KB without with
KD without with
KE without with
D + ser-
vice socket
4)
UK
max.
6.126 kV DC ( 4KVAC)
F/CZE DK
2)
Adapter set for international use (comes with Feature B01)
4)
if adapter (feature B11) is applied: HV-DC max. 1.5 kV DC
5)
without function test values and without comments on DUT
4)
China/AUS
4)
CH
adapter set 2)
4)
Designation Type
Test instrument with test current 200 mA DC and 10 A AC Sequences for IEC 61010, IEC 60 335, IEC 60950 and IEC 60601, data memory for up to 125 tests
SECULIFE ST M693A
Same instrument as M693A, however, suited to international use with adapter set for mains connection in the respective user country and English user interface language SECULIFE ST M693B
8 GMC-I Messtechnik GmbH
Article Number
Designation Type
Same instrument as M693A, however, with test current 200 mA DC and 25 A AC, additionally with high-voltage test max.
6.126 kV DC ( 4 KV AC) SECULIFE ST Same instrument as M693C, however,
suited to international use with adapter set (application of adapter: high-voltage test max. 1.5 kV DC) for mains connection in the respective user country and English user interface language SECULIFE ST
Article Number
HV M693C
HV M693D
Page 9

2 Safety Features and Precautions

Note
Attention!
!
This instrument fulfills the requirements of the applicable European and na­tional EC guidelines. We confirm this with the CE marking. The relevant decla­ration of conformity can be obtained from GMC-I Messtechnik GmbH.
The test instrument has been manufactured and tested in accordance with the following safety regulations:
IEC 61010-1 / DIN EN 61010-1 / VDE 0411-1, DIN VDE 0404, DIN VDE 0413 Part 2 and 4 and DIN VDE 0104 (Feature F02 or
Only when used for its intended purpose, the safety of the user, the test instrument and the device under test (electrical equipment or electrical medical devices) is assured.
Read the operating instructions carefully and completely before placing your test instrument into service, and follow all instructions contained therein. Make sure that the operating instructions are available to all users of the instrument.
Tests may only be performed by a qualified electrician or under the supervision of a qualified electrician. The user must be instructed by a qualified electrician concerning performance and evaluation of the test.
Manufacturers and importers of electrical medical devices must provide documentation for the performance of maintenance by trained personnel.
Observe the following safety precautions:
• The instrument may only be connected to electrical systems with a maximum of 230 V +10% which are protected with a fuse or circuit breaker with a maximum rating of 16 A.
• Measurements within electrical systems are prohibited.
• Be prepared for the occurrence of unexpected voltages at devices under test (for example, capacitors can be dangerously charged).
• Make certain that the measurement cables are in flawless condition, e.g. no damage to insulation, no interruptions in cables or plugs etc.
• When using a probe with coil cable (SK2W): Grip the test probe firmly, for example after insertion into a jack socket. Tensioning at the coil cord may otherwise cause the test probe to snap back resulting in possible injury.
Measurement of Insulation Resistance and Equivalent Leakage Current This test is performed with a maximum voltage of 500 V, with a current
SECULIFE STHV)
limit having been set (I < 3.5 mA). However, contacting the terminals (3 or 2) causes an electric shock which, in turn, may result in accidents.
Leakage Current Measurement
During leakage current measurement it is imperative to ensure that the de­vice under test is operated at line voltage. Exposed conductive parts may be charged with hazardous contact voltage during the test and may con­sequently not be touched under any circumstances. (There is a power shutdown as soon as the leakage current is higher than approx. 10 mA).
The function test may only be performed after the DUT has successfully passed the safety test!
Switching loads on and off
When switching the DUT on or off under load it is imperative that you ad­here to the order indicated below. This helps to prevent excessive wear and tear of the mains relays at the test instrument.
Before the measurement:
1) DUT: Switch the DUT off at the proprietary switch.
2) SECULIFE ST or SECULIFE ST
3) DUT: Switch on the DUT at the proprietary switch.
After the measurement:
4) DUT: Switch the DUT off at the proprietary switch.
5) SECULIFE ST or SECULIFE ST
The measuring and test instrument may not be used:
• If it demonstrates visible damage
• With damaged connector cables, measuring cables or patient ports
• If it no longer functions properly
• After extraordinary stresses due to transport In such cases, the instrument must be removed from operation and
secured against unintentional use.
HV:
Apply line voltage to the test socket .
HV:
Disconnect the test socket from the line .
GMC-I Messtechnik GmbH 9
Page 10
Meanings of Symbols on the Instrument
Attention!
!
!
The symbols on the instrument have the following meanings:
Warning regarding dangerous electrical voltage
Warning concerning a point of danger (Attention: observe documentation!)
Test socket
Mark of approval from VDE test authority
Indicates EC conformity
This device may not be disposed of with the trash. For further details on the WEEE marking, please refer to our website www.gossenmetrawatt.com and enter search term ’WEEE’.
2.1 Notes Regarding the High-Voltage Test (only Feature F02 or SECULIFE ST
The KS13 cable set, or similar cable sets, may not be used for the high-voltage test. The high-voltage test may only be performed directly via the test socket!
Do not hold the device under test in your hand during testing,
especially when testing safety class II devices. Make sure that the device under test does not make contact with any equipment or persons during testing.
Liability Exclusion In the event of sparkover, PCs operated in proximity to the test instrument may
“crash” resulting in data loss. All data and programs should be suitably backed up before high-voltage testing is performed, and computers should be shut down if necessary. A crash may occur even if no RS 232 connection has been established.
The manufacturer of the test instrument assumes no liability for direct or consequential damage to computers, peripherals or data which occurs during high-voltage testing.
The manufacturer assumes no liability for defects at the device under test which result from high-voltage testing. As a rule, defects can only occur at devices under test which are not in compliance with applicable standards, which were previously damaged or which have been improperly repaired, because high-voltage testing is required for type and routine testing by IEC 61010-1/EN 61010-1 / VDE 0411, part 1, as well as EN 60 335, EN 60601 and EN 60950.
HV)
10 GMC-I Messtechnik GmbH
Page 11

3 Initial Start-Up

Attention!
!
L1
N
green-yellow
green-yellow
PE
L1
L2
L3
N
PE
L1
L2
L3
N
green-yellow
U
L–N
= 115 V / 230 V
Connection to the Mains

3.1 Connection to the Mains (115 V / 230 V, 50 Hz / 60 Hz)

õ Connect the mains plug at the test instrument to the mains power out-
let. The function selector switch can be set to any position. If a mains outlet (earthing contact outlet) is not available, or if only a 3-phase outlet is available, the adapter socket can be used to connect the phase conductor, the neutral conductor and the protective con­ductor. The adapter socket has three permanently attached cables and is included with the KS13 cable set.
If connection is not possible via an earthing contact outlet: Shut down mains power first. Then connect the cables from the coupling socket to the mains using pick-off clips in accordance with the diagram. Disconnection from mains power is only possible with the mains plug.
GMC-I Messtechnik GmbH 11
Page 12

3.2 Automatic Recognition of Mains Connection Errors

Attention!
!
Note
Attention!
!
!
!
The test instrument automatically recognizes mains connection errors, if the conditions in the following table have been fulfilled. The user is informed of the type of error, and all measuring functions are disabled in the event of danger.
Typ e of Ma ins
Connection Error
Voltage at protective
conductor PE to finger
contact
(key)
Protective conductor PE
and phase conductor L
reversed and/or
neutral conductor N
interrupted
Contact voltage at
protective conductor PE
to neutral conductor N
or phase conductor L
Mains voltage
too low lamp
1)
In SETUP – test sequence – IT system
In either of the first two cases listed in the table above, immediately disconnect the test instrument from the mains and eliminate the error!
Message Condition Measurements
Text appears at
LCD
lamp
lights up
Text appears at
LCD
lights up
Press
key
U 40 V
Voltage at PE
>65V
U 25 V
U
< 90/180 V
L-N
disabled
impossible
(no supply power)
disabled, although
disabling can be
deactivated
possible
under certain
circumstances

4 General Notes

4.1 Online Instructions

Integrated online instructions inform the operator regarding all required connections, necessary work steps, operator errors, measurement results and more in all measuring modes. Information and test results appear at the dot matrix LCD in clear text.

4.1.1 Changing the User Interface Language

If you require a different language for the user interface of the test instru­ment, you can load it into your test instrument by means of the update and options installation program „SECU-Up“. The program is available for download from our website www.gossenmetrawatt.com (Products > Software > Software for Testers > SECU-Up).
Upon installation on your PC and starting the program, you proceed by selecting the „Update“ menu and choosing a language from the following list: Deutsch, English, Français, Italiano, .
Only one language at a time can be installed on the test instrument, the one previously installed is overwritten in the process.
1)
During data transmission, the test instrument and PC may not be disconnected from the mains power supply under any cir­cumstances. No other programs under WINDOWS may be activated during the update.
Voltage at the electrical system’s protective conductor PE may result in distorted measurement values during testing for the absence of voltage, or during leakage voltage measurements.
12 GMC-I Messtechnik GmbH
Page 13

4.1.2 Automatic Safety Class Selection

Note
Auto
Depending upon the type of mains plug or the connection configuration for the device under test, the test instrument recognizes the respective safety class and recommends its use for the measurement to be per­formed.

4.1.3 Manual or Automatic Operating Sequences

Depending upon selections made in the setup menu (selector switch in the Auto position), the next measurement is started automatically after the current measurement has been completed, or can only be started after manual acknowledgement. The integrated online instructions are ade­quate for most tests and measurements. However, the contents of these operating instructions should nevertheless be read and observed.

4.2 Online Help

Online help can be queried and displayed at the LCD for all measuring and test functions, and for almost all settings. Schematic diagrams which illustrate proper connection of devices under test to the test instrument can be displayed as well.
õ Press the following key in order to query online help:

4.3 Adjusting Contrast

Set the selector switch to Auto.
Select the “Setup” menu, “return” is highlighted.
Activate contrast adjustment.
Press and hold the ENTER key.
õ Press the same key again in order to exit online help.
Adjust contrast.
Online help can be queried during measurement by pressing and holding the help key.
Return to the menu.
Store the contrast setting to permanent memory with the save function in the setup menu.
GMC-I Messtechnik GmbH 13
Page 14

4.4 Configuring Device Parameters, Setting Date and Time

Device parameters and functions which are valid for all selector switch positions can be activated or deactivated with the selector switch in the
Setup position (see chapter 8 on page 18).

4.5 Configuring Measurement and Sequence Parameters

Measurement and sequence parameters, as well as functions, can be activated or deactivated in the setup menu (selector switch in the Auto position) for the respective test regulation. Refer to chapter 15.3 on page 39 for the significance of the various parameters.

4.6 Setting Limit Values

Upon delivery, the limit values set forth (at the point in time of issue) in ap­plicable national and international standards are stored to the test instru­ment. Limit values for each of the respective standards can be queried and changed if required with the setup menu (selector switch in the Auto position), but changes can only be made which result in even stricter test­ing than is required by the respective standard.
Newly entered limit values become effective immediately. However, these are only stored to memory permanently after activating the store function in the setup menu.
If the limit values set forth in the standards for certain safety classes need to be restored despite individualized settings, the menu function all values per standard in the limit values sub-menu must be selected and acknowl­edged with the key.
If the limit values set forth in the standards are changed, the instrument’s device software can be updated via the RS 232 interface.

4.7 Saving the Settings

All of the settings and changes which have been entered to the configure, limit values (selector switch in the Auto position) and zero point (temperature measurement) (selector switch in the Aux position) menus, as well as the se-
lected contrast setting are retained until the selector switch is turned, or the test instrument is disconnected from mains power. If settings and changes should be retained even after mains power has been interrupted, they must be saved in the setup menu for the respective test regulation or selector switch position.
14 GMC-I Messtechnik GmbH
Page 15

5 Classification of Devices Under Test

5.1 Safety Classes

Devices assigned to all of the following safety classes are equipped with basic insulation, and provide for protection against electrical shock by means of various additional precautions as well.
Safety Class I Devices
Exposed, conductive parts are connected to the protective conductor so that they are not charged with voltage if the basic insulation should fail.

5.2 Application Parts (electrical medical devices)

Type B Application Parts (body)
Devices of this type are suitable for both internal and external patient applications, except for use in direct proximity to the heart.
These devices provide for adequate protection against shock especially as regards:
• Reliable leakage current
• Reliable protective conductor connection if utilized Devices of the following safety classes are allowable:
I, II, III or devices with internal electrical power supply.
Safety Class II Devices
These devices are equipped with double insulation or reinforced insula­tion.
Safety Class III Devices
and Devices with Internal Power Supply
These devices are powered with Safety Extra Low Voltage (SELV). Beyond this, no voltages are generated which exceed SELV. These devices may not be connected to the mains. They may only be connected to the test instrument at jacks 1 through 3.
Type BF Application Parts (body float)
Same as type B, but with type F insulated application parts.
Type CF Application Parts (cardiac float)
Devices of this type are suitable for use directly at the heart. The applica­tion part may not be grounded.
Devices of the following safety classes are allowable: I, II or devices with internal electrical power supply.
Note: The DUT may only be connected to jacks 1 to 3 at the test instru­ment. It is only possible to perform a visual inspection, a measurement of the insulation resistance or the supply voltage, see parameter “SC II I U on page 39.
V
Classification Parameter (in the Sequence ... menu)
The test instrument always performs testing in accordance with the strict­est limit values of the respectively selected safety class. The test is failed if this limit value is exceeded.
However, higher limit values are allowed for certain devices under test. If the classification parameter has been activated (= x), the user is asked if
higher limit values are allowable for these devices. If the user responds with “Yes”, the DUT is reevaluated and the test may be passed.
Devices with Internal Power Supply
Devices with internal power supply are tested like permanently connected Safety Class II or II I devices.
GMC-I Messtechnik GmbH 15
Page 16

6 Abbreviations

AE Error condition: application part grounded AP Apparent power (during function test) B, BF, CF Classifications for application parts DEFI Defibrillator
I Residual current, fault current (during function test)I
max
RC
wc
HE Error condition: housing grounded I
CLIP
I-EHC
SCII
I-EHL
A1/A2
I-EHL
FRPE
IEL, I-EL Equivalent leakage current
, I-EDL
I
EDL
I
, I-EPL Equivalent patient leakage current
EPL
, IHL, I-HL Leakage current (differential, probe or contact current)
I
L
(Ia) (Maximum) load current (during function test)
I
L(max)
I
LC
I
PA
I
PE
I
PL
I
PMAP
IRC, (IRe) Residual current (current at protective conductor during
, I-HL Contact current (housing leakage current)
I
C, IHL
IT-system The IT system makes no direct contact between active
L Phase conductor connection of DUT MedGV German medical device ordinance
Maximum residual current (during function test) Residual current worst case
Current at clip-on meter Equivalent device leakage current for devices with
additional safety class II components Equivalent device leakage current with note A1/A2
(cross-reference within the standard) Equivalent device leakage current for portable x-ray devices
+PE: with additional protective conductor –PE: without additional protective conductor
Equivalent device leakage current (current at protective conductor)
Device leakage current Patient auxiliary current Earth leakage current (current at protective conductor) Patient leakage current Mains to application part (patient leakage current measurement)
automatic test sequence)
conductors and grounded parts: bodies within the elec­trical system are grounded.
MSELV Medical safety extra-low voltage N Neutral conductor connection of DUT NC Normal condition OE Operational earth P Active power (during function test) PA Functional earth (equipotential bonding) PE Protective conductor connection of DUT PF Power factor (during function test) RResistance
, R-INS Insulation resistance
R
INS
R-INS AP-PE Insulation resistance: application part to PE R-INS
INT. KARD.
Insulation resistance: intercardiac (application in proximity to the heart)
R-INS NL-PE Insulation resistance: neutral/phase conductor to PE
, R-PE Protective conductor resistance
R
PE
R-PEmains Protective conductor resistance limit value for
+mains: device under test with mains cable –mains: device under test without mains cable (protective conductor resistance limit value for mains cable only = 0.1 
SELV Safety extra-low voltage SFC Single-fault condition t On-time (during function test) T, Temp Te mper a tur e U
AC/DC
U
REF
, U-HV High-voltage
U
HV
, U-INS Test voltage for insulation measurement
U
INS
, U-LN Line voltage
U
LN
U
MEAS
U
Probe
AC/DC voltage Reference voltage: voltage to which leakage current is
related (as a rule nominal line voltage)
Voltage at which testing was executed. Displayed for all leakage current measurements.
Probe voltage W Electrical energy (during function test) ZVEH General Association of German Electricians
MLV manufacturer’s limit value MPG German medical product law
16 GMC-I Messtechnik GmbH
Page 17

7 Connecting the Device Under Test

Note
Note
Attention!
!
Attention!
!
Attention!
!
õ Connect the DUT in accordance with the schematic diagrams included
in the online help function.
Connection of the DUT to the test instrument depends upon:
The type of device under test:
electrical equipment or not, with or without application parts
The type of connector included with the DUT:
– With plug (“to test socket” parameter), applies to EL1 adapter as well – Without plug, single or multi-phase connection (“to jacks” parameter) – No connection to tester (“permanent connection” parameter),
see also chapter 3.1
Whether or not an adapter is used:
– Adapter to socket (customer specific adapter) – AT3-II S to socket, adapter for devices which are equipped with
5-pole, 16 A CEE plugs
– AT3-III E to socket, adapter for devices which are equipped with
5-pole, 32 A CEE plugs, see AT3-III operating instructions for test sequence.
• The DUT’s safety class (I, II or III)
The DUT must be switched on for all tests. Switches, relays, temperature regulators etc. must all be taken into consideration.
The test instrument automatically recognizes whether or not the DUT is connected to jacks 1 through 3. The instrument also recognizes whether or not the DUT has been connected to the test socket. As a default set­ting, the program sequence assumes that the plug from the DUT has been connected to the test socket.
õ Position the cursor at the third line in the start menu for the test se-
quence.
õ A selection of possible connection setups can be displayed by
activating the key.
õ Select the desired connection setup with the cursor and
acknowledge with the key.
For omitting the protective conducter test in the case of fully insulated devices see page 65.
Protective Conductor and Insulation Resistance Measurements for Permanently Installed Devices Under Test
Deactivate the electrical system which supplies power to the
device under test before connecting the test instrument!
õ Remove the mains fuses from the device under test and disconnect
the neutral conductor N inside the device under test.
Measuring Contact Current (absence of voltage)
Make sure that the contacted parts are not grounded.
High-Voltage Test (Feature F02 or SECULIFE ST
The KS13 cable set, or similar cable sets, may not be used
for the high-voltage test. The high-voltage test may only be
performed directly via the test socket!
HV)
Safety Class II Devices with Safety Class I Mains Plugs
If the device under test is equipped with a safety class I plug although it complies with safety class II, safety class I is recog­nized by the test instrument. If this is the case, switch from safety class I to safety class II in the initial menu.
If the test instrument is unable to automatically recognize how the DUT has been connected, the recommended connection setup should be double checked and determined manually if necessary.
GMC-I Messtechnik GmbH 17
õ Connect the device under test to the test socket.
õ Safety class II only: Connect the probe to jacks 4 and 5.
Make sure that the application parts are not connected during
the high-voltage test!
Page 18

8 Configuring Device Parameters

Setup
duration selected in seconds under “Mains pause” has elapsed.
Automode x: for fully automatic test sequences the messeges
are mainly suppressed
Test Sequence ...
General device parameters can be configured and saved with the selector switch in the Setup position.
Settings x / – = function activated / deactivated Single-fault
If the single-fault condition has been activated, the test is interrupted as a failure as soon as an error occurs.
Auto Class PSI Test results (passed or failed) for the various selector
switch positions are automatically assigned to the 8 statistics channels.
inc. Service Error Measurement results are compensated by taking
Select a menu and acknowledge.
IT Network Testing in IT systems can be performed by suppress-
service error into consideration (measuring error).
ing tests for U whether or not voltage is present at PE.
PE-N
. The U
test determines
PE-N
(Leakage current measurement results may other-
Select parameter and acknowledge, change setting and acknowledge.
Limit values ...
Settings x / – = function activated / deactivated Illumination Background illumination for the LCD. One of three
different conditions can be selected with the up and down scroll keys: x: continuously on, –: off
Acst Sig, Seq
Acst Sig, Meas Acoustic signal is generated for: Measured value
Auto meas. point Prerequisite: Feature KD01. An acoustic signal indi-
numbers 1 through 9: duration in minutes after which illumination is automatically deactivated.
Test time Duration of a single test (0 255 s) Reference voltage: Voltage to which leakage current makes reference
(as a rule nominal line voltage)
Earth fault: During the short-circuit test, testing is also performed
to determine whether or not a connection exists
Direct Print-Out Reports Reports which have been saved to memory can be
between L/N and PE (short-circuit to exposed con­ductive part). We assume that a short-circuit to an exposed conductive part exists in the event of leak­age current greater than 15 mA from L/N to PE. This value should be increased for some DUTs (in particu-
Select template a report template can be selected for print-out from
Service – Time and date settings
lar high-current consumers), because greater leak­age currents are present.
Mains wait Line voltage is first applied to the test socket.
18 GMC-I Messtechnik GmbH
However, testing does not begin until after the
wise be distorted.) Acoustic signal is generated for: Incorrect connection of the DUT, error in the electrical supply system, next test step.
fluctuations, test current polarity reversal
cates whether or not the probe is connected to the protective conductor. The test sequence is run auto­matically. Rapid signal frequency: probe connected to PE Slow signal frequency: next measuring point Prerequisite: Feature KE01, see chapter 18 on page 59.
selected from a list with an ID number and displayed (see chapter 18 on page 59).
of 5 different templates.
(if a (P)SI module is used, the same time and date must also be entered to the (P)SI menu)
– Service functions (password required)
Page 19

9 Measuring Protective Conductor Resistance

Note
PE
Connecting Safety Class I Devices to the Test Socket
When the DUT is connected, resistance is measured between the protec­tive conductor terminal at the test socket or at the PE jack and the probe connection at the DUT (contact with conductive parts of the housing).
õ In order to measure protective conductor resistance, contact a
conductive part of the housing with the probe, which is connected to the protective conductor.
During measurement, the connector cable must only be moved in as far as it is accessible during repair, modification or testing. If a change in resistance occurs during the manual test step of the conti­nuity test, it must be assumed that the protective conductor is damaged, or that one of the connector contacts is no longer in flawless condition.
Definition
Protective conductor resistance
Testing Extension Cables
See test sequence in chapter 15.7 on page 46.
is the sum of the following resistances:
• Connector cable or device connector cable resistance
• Contact resistance at plug and terminal connections
“Connection of the DUT: SC I/II” is not displayed when the test is performed individually, but rather only during the automatic test sequence.
• Extension cable resistance if utilized
Resistance is measured:
• Between each conductive part of the housing and the earthing contacts at the mains and the de-
vice plug (if a removable mains connector cable is used), or the protec-
Selecting Test Current and Polarity Test current amperage (200 mA DC or 10 A AC (Feature G00 or SECULIFE ST) or
200 mA DC, 25 A AC (Feature G01 or SECULIFE ST
HV)) as well as polarity can
both be changed by pressing the or the key.
tive conductor terminal for permanently installed devices.
• as 4-pole measurement
• Between the earthing contacts at the mains plug and the earthing contacts at the device plug for device connector cables
• Between the earthing contacts at the mains plug and the earthing contacts at the coupling socket for extension cables
Testing with 10 A Test Current (Feature G00 or SECULIFE ST) or 25 A (Feature G01 or SECULIFE ST
HV)
The test has a maximum duration of 30 s (fixed value) if 10 A or 25 A test current is used. After this time period has elapsed, the last measured value is frozen and “data hold, measurement stopped” appears at the dis­play. If the test instrument becomes excessively warm, testing cannot be repeated until after a waiting period of 1 minute. When testing with 10 A or
GMC-I Messtechnik GmbH 19
25 A, the last measurement can be repeated if the test results in failure.
Page 20
Combined Testing – Differential Protective Conductor Resistance
Iso / HV
f
o
r
S
C
I
I
Zero balancing is also possible for protective conductor measurement. With
zero balancing, all subsequent measurements are adjusted with an offset such that 0 is displayed for a selected reference point which is con­nected to the protective conductor. When test points are contacted with the probe which are electrically connected to this reference point, differen­tial resistance R point is displayed. The mains release key must be activated during measurement in order to perform zero balancing. Press key
between the reference point and the contacted test
PE
„Store
value“ in order to save the reference and/or correction value. The meas­sage „Zero point corrected“ concerning the reference value is displayed during all future measurements.
Attention: It is absolutely essential to delete the reference value after the reference value has been stored and the test has been performed as it is taken into account in all future tests. To delete the value, the procedure is the same as for storage, press key
„Delete value“.

9.1 Maximum Allowable Limit Values for Protective Conductor Resistance for Connector Cables with a Length of up to 5 m

R
PE
Housing –
Device Plug
0.2 0.3
0.2 0.2
3)
Open-Circuit
Voltage
< 24 V
4 V < U
L
Tes t Sta nd ard Te st C urre nt
VDE 0701­0702:2008
IEC 62353 (VDE 0751-1) EN 61010
EN 60335 EN 60950 EN 60601 0.1 0.2 
1)
This value may not exceed 1 for permanently connected data processing systems (DIN VDE 0701-702, DIN VDE 0701, part 240).
2)
Permanently connected cable
3)
Feature G00 = 10 A (SECULIFE ST) / G01 = 25 A (SECULIFE ST HV)
> 200 mA
10 A
~/25 A
to test socket
only
R
PE
Housing –
Mains Plug
1)
0.3 
2)
+ 0.1
additional 7.5 m
2)

10 Insulation Measurement

10.1 Insulation Resistance R
Definition
Safety Class I
Insulation resistance is measured between short­circuited mains terminals and the protective conductor.
Safety Classes II and I II
Insulation resistance is measured between short­circuited mains terminals and external conductive parts which can be contacted with the probe.
INS
.
20 GMC-I Messtechnik GmbH
Page 21
Exception for Permanently Installed Safety Class I Devices
Attention!
!
Attention!
!
Note
Note
Deactivate the electrical system which supplies power to the device under test before connecting the test instrument!
õ Remove the mains fuses from the device under test and disconnect
the neutral conductor N inside the device under test.
õ Connect the probe to phase conductor L at the device under test in
order to measure insulation resistance.
Sequence
Measurement of Insulation Resistance (equivalent leakage current)
This test is performed with a maximum voltage of 500 V, with a current limit having been set (I < 3.5 mA). However, contacting the terminals (3 or 2) causes an electric shock which, in turn, may result in accidents.
All switches at the device under test must be set to the ON position during measurement of insulation resistance. This also applies to temperature controlled switches and temperature regulators. Measurement must be performed in all program stages for devices equipped with program controllers.
R-INS
Start measurement.
Nominal voltage is 500 V DC in this case.
Nominal voltage can be adjusted within a range of 50 V to 550 V DC.
When insulation is first started from the menu, nominal voltage is always set to 500 V. Open-circuit voltage is always greater than nominal voltage.
Minimum Allowable Limit Values for Insulation Resistance
R
Tes t St anda rd Tes t Vo lta ge
VDE 0701­0702:2008
IEC 62353 (VDE 0751-1)
500 V
SCI SCII SCIII Heat
1M 2M 0,25 M 0.3 M 2M 7M
INS
70 M 70 M
* for Safety Class I devices with activated heating elements
Notes
All exposed, conductive parts of safety class II and III devices, as well as of battery-powered devices must be scanned with the probe and insula­tion resistance and/or leakage current must be measured.
Batteries must be disconnected from their terminals during testing of bat­tery powered devices.
*
GMC-I Messtechnik GmbH 21
Page 22

10.2 Equivalent Leakage Current

~
DUT
Probe
mA
PE
L N
Probe
PAT
DUT
~
mA
L N
PE
mA
Probe
PAT
DUT
L N
~
PE
Equivalent Patient Leakage Current I
(IEC 62353 (VDE 0751-1))
EPL
Test S etu p
General
The equivalent leakage current measurement is a method for measuring protective conductor current (DIN VDE 0701-0702) and/or device leakage current (IEC 62353 (VDE 0751)).
A high-impedance power supply is connected between the short-cir­cuited mains terminals and all exposed metal parts of the housing (which are connected to one another).
Measurement of Equivalent Leakage Current I
per DIN VDE 0701-0702
EL
A high-impedance power supply is connected between each pa­tient port and all exposed metal parts of the housing (which are connected to one another). The mains terminals are short-cir­cuited, and connected to the same point on the housing.
Measurement
The current which flows over the insulation at the device under test is measured separately for each appli­cation part.
Measurement is always performed with an AC power supply with current limiting. Varying line voltages are taken into account.
Maximum Allowable Limit Values for Equivalent Leakage Current in mA
Test Standard I
VDE 0701­0702:2008
Measurement of Equivalent Device Leakage Current I (VDE 0751-1))
(per IEC 62353
EDL
IEC 62353 (VDE 0751-1)
The patient ports are linked with each other and are connected to the same point on the housing.
22 GMC-I Messtechnik GmbH
IELEquivalent leakage current I
Equivalent device leakage current
EDL
I
Equivalent patient leakage current
EPL
PE Protective conductor
1)
for devices with heating power of  3.5 kW
2)
with and without line voltage applied to the application part
EL
SC I: 3.5
1 mA/kW
SC II: 0.5
1)
SC I (parts within or connected to PE)
Permanently connected devices with PE 10
Portable x-ray devices with additional PE 5
Portable x-ray devices without additional PE 2
I
EDL
SC II 0.2 Type BF 5
Devices with mineral insulation 5
1 Type CF 0.05
I
EPL
2)
2)
Page 23
Connection
Attention!
!
Refer to the schematic diagrams included with the online help for connection instructions.
Equivalent Leakage Current I
DIN VDE 0701-0702 / 2 K
EL
Connection Exception for Permanently Installed Safety Class I Devices
Current is measured between the probe, with which the L and N conduc­tors must be contacted, and the protective conductor terminal PE at the device under test for permanently installed safety class I devices under test.
Disconnect mains power before connecting the test instrument!
õ Remove the mains fuses from the device under test and disconnect
the neutral conductor N inside the device under test.
õ Connect the probe to phase conductor L and neutral conductor N at
the device under test in order to measure equivalent leakage current.
Sequence
Current is displayed during this type of equivalent leakage current measurement which would flow during leakage current measurement conducted in accordance with device regulations with nominal voltage. Leakage current measurement in accordance with the respective device regulations is usually not possible, because the device would have to be set up in an electrically isolated fashion, or connected to an earth isolated power supply to this end.
For the evaluation of measured values during equivalent leakage current measurement please refer to chapter 21.2.
Select the I-EL measurement and start.
Equivalent leakage current is measured between short-circuited N and L, and the protective conductor PE.
Measuring circuit resistance is equal to 2 k for VDE 0701-0702 for the simulation of the mean body resistance of a human being.
Equivalent Device Leakage Current I
for IEC 62353 (VDE 0751-1) / 1 K
EDL
Select the I-EDL measurement and start Equivalent device leakage current is measured between short-circuited N
and L, and the probe. Measuring circuit resistance is equal to 1 k for IEC 62353/VDE 0751 for
the simulation of mean patient resistance.
Equivalent Patient Leakage Current I
(IEC 62353 (VDE 0751-1))
EPL
Select the I-EPL measurement and start
Equivalent patient leakage current is measured between short-circuited N and L and the respective application part. Jacks A through K (application parts) are connected separately for each application part.
Groups of cables or sensors can be assigned to application parts in the test sequence start menu in accordance with IEC 62353 (VDE 0751-1) or EN 60601.
GMC-I Messtechnik GmbH 23
Page 24

10.3 High-Voltage Test (Feature F02 or SECULIFE ST HV)

Caution: High-Voltage!
Attention!
!
Only safety class I and II devices can be tested, which can be connected to the test socket.
The high-voltage test is performed with direct voltage. In order to comply with requirements for alternating voltage, testing is performed with 1.5­fold direct voltage. This multiplier is applied automatically during testing. A selected nominal voltage of 3.5 kV thus results in a DC output voltage of
5.25 kV.
Output voltage is measured for the entire duration of the test, and its min­imum value is determined. The minimum voltage value is indicated as the test result. If this value is less than the pre-selected test voltage, the test is failed.
The conversion factor must be taken into consideration for testing and calibration of the SECULIFE ST
HV.
The device is designed such that special measures in accordance with DIN VDE 0104 (high-voltage test) need not be observed.
This is accomplished by means of the following characteristics:
1. Continuous short-circuit current is less than 3 mA (DC).
2. Discharge energy (at 5.25 kV) is less than 350 mJ.
In order to comply with high-voltage test regulations despite the minimal continuous short-circuit current, the charging capacitors are connected to the test socket (L, N) with relatively low-value protective resistors. This results in a peak short-circuit current value of approximately 5 A (at 5 kV), which causes plainly audible and visible sparking.
High-voltage testing can only be performed via the test socket. The pro­tective conductor at the socket is grounded during the high-voltage test.
To uc h neither the test socket nor the device under test during voltage testing! A high-voltage of up to 5.5 kV is present at the test socket output! A current may flow over your body, and although it does not reach life endangering levels, the resulting shock is plainly discernible. Previous testing of the protective conductor is absolutely essential for safety class I devices because if the protective conductor is in-
terrupted, the high-voltage test does not charge all dielectric ma­terial and the scope of testing is thus insufficient.
Connection
õ Insert the mains plug from the device under test into the test socket.
õ Safety class II: Connect the probe to jacks 4 and 5.
Make sure that the application parts (18) are not connected dur­ing high-voltage testing!
Individual Test
õ Set the rotary selector switch to the Iso/HV position.
õ Select the U-HV menu with the key and acknowledge with the
key.
õ If the device under test has not yet been switched on, a message to
this effect appears at the display. Nominal voltage is then selected.
õ Select the desired nominal voltage
from the display with the key. The instrument multiplies this nomi­nal voltage by a factor of 1.5, which represents the actual test voltage (see chapter 10.3). Note: For SC I, the test voltage can be set to maximum 1.5 kV.
õ Test voltage is applied to the test
socket, and in turn the device under test, for as long as the
key is
held depressed. Activation of high­voltage is indicated with an acoustic signal. Currently measured minimum output voltage U divided by 1.5), test voltage U
, and remaining test time are displayed.
DC
(measured value
HV AC
24 GMC-I Messtechnik GmbH
Page 25
õ Safety class II: Contact all exposed, conductive parts, simultaneously
Attention!
!
Note
if possible, in order to avoid unnecessarily long test durations or repeats. Avoid contacting individual parts sequentially.
õ After the key is released, the equivalent AC voltage is displayed which
occurred during testing. This voltage is indicated as the test result. If this value is less than the pre-selected nominal voltage, the test is failed.
õ If you want to repeat the test, press the key. The test sequence is
started over again with selection of nominal voltage.
If sparkover should occur, the test is immediately interrupted and the voltage measured at the moment sparkover occurred is display as U If the device is disconnected from the test socket during testing,
HV AC
.
even though this is prohibited, the test is immediately inter­rupted. The following message is displayed: “Caution, device under test is still charged!”
Testing as Part of a Test Sequence
õ Select the high-voltage menu under the setup function for the respec-
tive test regulation in order to set parameters for the high-voltage test.
õ Enter the desired AC nominal voltages for safety classes I and II. The
instrument multiplies the respective AC nominal voltage by a factor of
1.5, which represents the actual DC test voltage (see chapter 10.3). Test voltage is determined based upon the selected, or the automati­cally recognized safety class.
õ Save the setup values to memory.
õ Start the high-voltage test by pressing the key, if “manual
sequence” has been selected in the setup menu.
õ Safety class II only: Contact the device under test with the probe.
Activation of high-voltage is indicated with an acoustic signal. The selected nominal voltage of 3.5 kV results in a DC output voltage of
maximum 5.25 kV. The high-voltage test is ended auto-
matically after the test duration has elapsed.
: Equivalent test voltage in DC
U
HV DC
U
: Measured DC value divided
HV AC
by 1.5
If the value U lected nominal voltage U failed.
is less than pre-se-
HV AC
AC
, the test is
In the event of sparkover, the voltage measured at the moment sparkover occurred is dis­played in the test results as minimum value U reason the test was failed.
, along with the
HV AC
GMC-I Messtechnik GmbH 25
Page 26

11 Leakage Current Measurement

Attention!
!
I leakage
10 10210310410510
6
+20
0
–20
–40
–60
Frequency (f) in Hz
Relative Magnitude (dB): 20 log
U(f)
U(f=10)

2
3
L
N
1 Test socket (w/o protective conductor contact) 2 Patient connector cable
(insulated application part)
3 Device housing
1
11.1 Earth Leakage Current I
(Feature KA01)
PE
Current which flows from the power pack over the insulation to the protective conductor, and thus to earth.
The protective conductor is disabled during this measurement.
Select the Ixx measurement and start. Each time line voltage is applied to the
test socket, L and N are reversed, if this function has been selected in the leak­age current menu (see chapter 11 on page 26).
Attention: Leakage Current Measurement
During leakage current measurement it is imperative to ensure that the de­vice under test is operated at line voltage. Exposed conductive parts may be charged with hazardous contact voltage during the test and may con­sequently not be touched under any circumstances. (There is a power shutdown as soon as the leakage current is higher than approx. 10 mA).
Frequency response is taken into consideration in accordance with the digram to the right when leakage current is mea­sured.
11.2 Housing Leakage Current I
Current which flows from housing parts which are not connected to the protective conductor via an external conductive connection to earth or another part of the housing. Flow of current via the protective conductor is excluded in this case.
The AC component is mea­sured. The DC component can also be measured if individual measurement is performed (in­stead of a test sequence).
EN 60601/VDE 0751: The following sequence is programmed for measuring and documenting several exposed conductive parts: If the acoustic signal chain changes from long to short intervals, it means that measurement is completed and the next measuring point can be se­lected and scanned (key ).
(probe current, contact current)
HL
If each measured value is to be recorded (printed), it can be done by pressing the key after each measurement (on condition that „Direct print-out“ is activated, see chapter 18).
26 GMC-I Messtechnik GmbH
Page 27

11.3 Patient Leakage Current IPL

Note
Note
Attention!
!
2
3 (N) L (L) N
1 Test socket w/o
protective conduc-
tor contact 2 Application part 3 Device housing 4 Measuring circuit

4
IEC 62353/
I
LC
1
VDE 0751
Current which flows from an ap­plication part via the patient to earth. This current may be caused by an unintentional interference voltage at the patient, and may flow via the patient and an insu­lated, floating type F application part to earth. Useful patient cur­rent is excluded in both cases.
AC and DC components are measured.
Patient leakage current must also be measured if application parts are available. The displayed test voltage must be documented.
11.5 Residual Current I
RC
Sum of instantaneous current values which flow via the L and N conduc­tors at the device mains connection (also known as differential current). Residual current is practically identical to fault current in the event of an er­ror. Fault current: Current which is caused by an insulation defect, and which flows via the defective point.
11.6 Device Leakage Current I
per IEC 62353 (VDE 0751-1)
LC
Device leakage current is the sum of all leakage currents from the housing, all accessible conductive parts and all applica­tion parts to PE. Measurement must be performed for both mains polarities and the largest value is documented.
11.4 Patient Auxiliary Current I
Current which flows within the patient between the electrodes of the application part. Use for in­tended purpose is assumed. Fur­thermore, the current should not cause any physiological effects. For example, this is the case for input current from amplifiers, or current used for impedance plethysmography.
AC and DC components are measured.
GMC-I Messtechnik GmbH 27
(Feature KA01)
PA
The highest device leakage current value and line voltage must be documented.
The protective conductor is interrupted during this measurement.
Page 28
Key for Tables
IPEEarth leakage current in the operating state (alternative: measurement of IEL) I
Housing leakage current (probe or contact current)
HL
I
Residual current
RC
I
Device leakage current
LC
I
Patient leakage current
PL
I
Patient auxiliary current
PA
Maximum Allowable Limit Values for Leakage Current in mA
I
Test Standard
VDE 0701­0702:2008
SC I: 3.5 1 mA/kW
I
PE
NC SFC NC SFC
*
0.5
I
HL
RC
SC I: 3.5
1 mA/kW
*
SC II:
0.5
IEC 62353 (VDE 0751-1)
General 0.5 1
EN 60601
Notes 1 & 3 2.5 5
0.1 0.5
Note 2 5.0 10
* for devices with a heating power > 3.5 kW Note 1: Devices without any accessible conductive parts which are connected to
the protective conductor, and which comply with I e.g. electronic data processing devices with shielded power pack
and conceivably IPL,
HL
Note 2: Permanently connected devices with protective conductor Note 3: Portable x-ray devices and devices with mineral insulation
I
LC
General 0.5
Notes 1 & 3 2.5
Note 2 5.0
SC II 0.1
I
Tes t Standard
IEC 62353/ VDE 0751
EN 60601
Type B Type BF Type CF Type B Type BF Type CF
NCSFCNCSFCNCSFCNCSFCNCSFCNCSFC
Direct
0.01
Current
Alt.
0.1 0.1 5 * 0.01
Current
Direct
0.01 0.05 0.01 0.05 0.01 0.05 0.01 0.5 0.01 0.05 0.01 0.05
Current
Alt.
0.1 0.5 0.1
Current
PL
0.01 0.01
0.5
0.01
5 *
* Only with line voltage at the application part
I
PA
0.05 *
0.05
0.1 0.5 0.1 0.5 0.01 0.05
0.05
*
28 GMC-I Messtechnik GmbH
Page 29
This page has been deliberately left blank.
GMC-I Messtechnik GmbH 29
Page 30

12 Multimeter Functions

Attention!
!
V
12.2 Alternating / Direct Voltage U
– Max. 253 V
AC/DC
Direct, alternating and pulsating volt­ages of up to 253 V can be measured between the 2 and 3 connector jacks.
Furthermore, it is possible to switch be­tween minimum, maximum and mo­mentary measured value via key . This is particularly useful in combination with the SECULOAD test adapter for weld- ing equipment (article number Z745V).
Two procedures are to be distinguished as from firmware version 7.24:
12.1 Probe Voltage U
Voltage is measured between the mains PE terminal at the test instrument and the probe. In this case the probe can also be used as a phase finder. For IEC 61010: A selection can be made with the up scroll key as to whether testing will be conducted under normal conditions, or with interrupted protective conductor.
For conducting the measurement the DUT must be put into service via key (14).
– Max. 300 V
probe
1. DUT not connected to test socket (permanent connection)
õ Select switch position V
and then measurement U
AC/DC
.
õ Connect the measurement cables to jacks 2 and 3.
õ Scan the measuring point with the test probes.
õ Read the measured values.
õ Remove the test probes from the measuring point and pull off the
measurement cables from jacks 2 and 3.
õ Return to the Multimeter menu with ENTER.
Start the U
measurement.
probe
30 GMC-I Messtechnik GmbH
Page 31
2. DUT on test socket (new! as from firmware version 7.24)
Attention!
!
It is imperative that you observe the specified sequence of test steps:
No device or component may initially be connected to jacks 1 through 3! (Jacks 2 and 3 are short-circuited during all measurements at the test socket; Exception: as soon as the express command to connect the measurement cable appears in the display, short-circuit is cancelled, see below)
õ Remove all cables that may be plugged into jacks 1 through 3.
õ Connect the DUT to the test socket. õ Select switch position V
and then measurement U
õ Switch the DUT on (a short-circuit test is conducted).
õ Put the DUT into service by switching line voltage to the test socket
with key (14).
õ Please do not fail to observe the following:
Do not connect the measurement cables until the following command appears in the display: „Connect the measurement cable to jacks 2 and 3 for voltage measurement“.
õ Scan the measuring point with the test probes.
õ Read the measured values.
õ Remove the test probes from the measuring point and pull off the
measurement cables from jacks 2 and 3.
õ Return to the Multimeter menu with ENTER.
Measurement of Safety Extra Low Voltage (see Procedure No. 2)
Line voltage can be switched to the DUT via the test socket with key 14, e.g. in order to measure protective extra low voltage at the DUT output.
AC/DC
.

12.3 Resistance R

Resistance of up to 150 k can be measured between the 1 and 2 con­nector jacks.
Select the R measurement and start.
The voltage measured at the DUT output must be a protective extra low voltage electrically isolated from the mains, otherwise an overcurrent protection device in the building installation may trip.
GMC-I Messtechnik GmbH 31
Page 32

13 Measurements with Accessories

Aux
13.1 Alternating Current I
with Clip-On Converter
CLIP
Connection
Alternating current can be measured in two measuring ranges (1 mA 10 A ~, 1 A 100 A~) with a clip-on current-voltage converter connected to the 2 and 3 jacks (e.g. the WZ12C).
Start the I
measurement.
CLIP
Select measuring range.
13.2 Protective Conductor Resistance R
via Clip-On Meter
PE
Connection
Protective conductor resistance can be determined with the WZ12C clip-on current transformer.
25 A AC test current (Feature G01 or
SECULIFE ST
HV only):
Use the Z864A shunt in addition to measuring range matching.
P: potential lead for 4-wire measurement
The potential lead must be connected to the outgoing protective conductor at the distributor.
Cable resistance is measured from the device under test to the test instru­ment without potential lead P. This value may deviate significantly from actual protective conductor resistance, because the lead is measured along with the test instrument installation. Resistance from the probe terminal to contact P at the protective conductor is measured with the potential lead.
32 GMC-I Messtechnik GmbH
Page 33
Select RPE measurement and start.

13.3 Temperature T with Pt100/1000 Sensor Connection

Temperature can be measured within a range of – 200 C to +850 C with a Pt100 or Pt1000 sensor (default setting) connected to the PE (1) and N (2) jacks.
Select the Temp measurement and start. Select Pt100 or Pt1000 by means of
“select measuring range” with the key. The temperature unit of measure can be selected in the “TEMPERATURE” setup menu. Selection can be made amongst C (Celsius), F (Fahrenheit) or Kelvin. Zero balancing is also accessible via the “TEMPERATURE” setup menu.
Zero Balancing
Sensor cable resistance can be compensated for with this function:
õ Short circuit the ends of the sensor leads and determine resistance as
shown below.
Zero Balancing
The determined value can be stored di­rectly ( key), or changed first. The data entry menu is opened with the key.
õ Change the measured value
manually with the help of the
and keys.
õ Press the key in order to acknowledge the selected value, and to
display other menu functions at the bottom of the window.
Save the selected value by activating the “store value” key before exiting the balancing function with the key.
The “delete value” command can only be accessed via the “change value” menu. The “no zero balancing” setting is saved at the same time by activating the key.
GMC-I Messtechnik GmbH 33
Page 34

14 Function Test

Attention!
!
Note
Note
Attention!
!
Function Tes t
The device under test can be subjected to a function test with line voltage via the integrated test socket.
In addition to testing with the selector switch in the function test position, a function test can also be performed immediately after safety testing has been passed in accordance with the selected standard (not possible for safety class III devices).
The function test may only be performed after the DUT has successfully passed the safety test.
Each time line voltage is applied to the test socket, phase con­ductor L and neutral conductor N are reversed, assuming that “polarity reversal = X” is selected in the I leakage switch setting.
The function test is only possible if the device under test has been connected to the test socket (21).
Measurements
The function test includes the following measurements:
–Voltage U – Residual current I
between the L and N conductors
LN
(corresponds to fault current between L and N) – Load current I – Active power P
a
– Apparent power AP (calculated) – Power factor PF (cos calculated, display > 10 W) – Electrical energy W – On-time t for U
at socket (21)
LN
The following values are also displayed in all test sequences in switch po­sition AUTO after the function test has been completed:
– Maximum residual current I – Maximum load current I – Maximum active power P
max
Lmax
max
Power factor is calculated from active power and apparent power. Power factor corresponds to cos for sinusoidal quantities (line voltage and load current).
Starting the Function Test
For reasons of safety, the device under test must be switched off before the function test is started. This precaution prevents inadvertent start-up of a device under test which may represent a hazard during operation, e.g. a circular saw or a disc grinder.
Ending the Function Test
After completion of the function test, devices under test must be turned off with their own switch – especially devices with relatively high inductivity.
34 GMC-I Messtechnik GmbH
Page 35
Short-Circuit Test
Note
1 Test for shorts between N and L. 2 Test to determine whether or not
the N or L conductors are short­circuited to the protective conductor.
A short-circuit at the device under test is recognized automatically by the test instrument. A message appears at the display (9), and the function test is disabled.
If the lamp blinks (15), line voltage can be switched to the test socket with the key (14), and the measurement can be started. If the lamp (15) is lit continuously, line voltage is present at the test socket.
The test socket can be rendered voltage-free with the key (14), or the
function test can be ended with the key (12).
GMC-I Messtechnik GmbH 35
Page 36
15 Measurements in Accordance with National and
Note
Note
International Standards with Selector Switch in Auto Position
If measurements need to be performed in accordance with given stan­dards which require specific tests, and if results need to be stored and documented with a test report, an automatic test sequence is advisable instead of individual measurements.
for tests per EN 60950, EN 61 010 and EN 60335
The DUT must be connected to the test socket for the automatic test sequence! The DUT must comply with safety class I or II.
õ Connect the test instrument to mains power.
A mains connection test is initialized (see chapter 3.2 on page 12).
õ
Connect the DUT to the test socket at the test instrument (see chapter 7 on page 17). The test instrument initializes
õ Set the selector switch to the Auto position. If the DUT has been con-
nected to the test socket, safety class recognition is performed. Other­wise, the safety class must be specified manually. In the initial program window, move the cursor up with the knowledge with the key. A safety class can now be selected with the
and keys and acknowledged with the key.
õ In order to select the regulation for which testing is to be performed,
move the cursor up within the initial program window with the into the top line and acknowledge with the key. The desired stan­dard can now be selected with the with the key. If you always want to test in accordance with a spe­cific standard, this can be stored as the default standard in the setup menu. Otherwise, the selected standard only remains active until the test instrument is disconnected from mains power (factory default set­ting: VDE 0701-0702).
õ The test sequence can be configured in the Setup menu, limit values
can be changed if necessary and database options can be configured.
õ The test sequence is started by selecting Start test and by acknowledg-
ing with the key (see “Test Sequences”).
Tests which have already been included in chapters 9 through 14 are not described here again. The only exception is measurement of extension cables.
and keys and acknowledged
connection type recognition
key into the third line and ac-
key

15.1 Test Sequences

Test sequences for the various standards are always run in the same order, assuming that the device under test has been properly connected and the mains connection test has been passed. The test sequence can be run step by step with manual activation of each subsequent step if this function has been specified, or automatically. Step by step manual operation can be selected in the initial program window in the setup menu under “Sequence” ... “manual sequence”.
Safety class: SC I or SC II is automatically recognized by the test instru-
ment. If the safety class is not correctly recognized, you can change it manually. This manual correction is not stored however.
Visual inspection: If “visual inspection” has been activated
If a part is recognized as defective by the user, it must be identified as such by selecting it with the cursor and acknowledging with the key.
.
Protective conductor resistance measurement
This test cannot be skipped. Exception: This measurement can be skipped by pressing sage „Connect probe to protective conductor“ is displayed) if a protective conductor connection is not possible.
Evaluation of protective conductor testing
High-voltage test for DIN VDE 0701 Part 1, Appendix E, EN 60950,
EN 61010, EN 60335, EN 60 601, if selected in the setup menu.
Insulation resistance measurement
IEC 62353: only if selected in the setup menu of the
EN 60601: only if selected in the initial program window
DIN VDE 0701-0702 only if R-INS is activated in the initial program
Evaluation of insulation test
Leakage current measurement (various single-fault conditions are run de-
pending on classification)
Evaluation of each individual leakage current measurement, see also
chapter 21.2
• Evaluation of the overall test
in the initial program window under “Sequence”.
(for SC I devices under test only)
(when mes-
initial program window under “Sequence ...”
under „Test Conditions“
window
36 GMC-I Messtechnik GmbH
Page 37
•Perform function test if required:
The function test can be performed each time a safety test has been successfully completed. The blinking signal lamp indicates that the function test should be started. Beyond this, the function test can also be started from the Function Test selector switch position. See chapter 14 on page 34 regarding performance of the function test.
•Display test results
(the worst measured values for the respective test sequence)
• Save test results and print if required

15.2 Setting Up Test Sequences

The order of a test sequence cannot be changed! However, it is possible to change limit values and measuring parameters, and test steps can be added or skipped.
Principally, two test sequences are distinguished which are described on the following page:
• Sequence 1 (repair tests and periodic tests)
• Sequence 2 (type tests)
Settings for the Test Sequence
Settings for the test sequence are not stored under a type designation, but apply generally for the preset sequence of the respective test stan­dard.
Type-related Settings (Type*) (for tests per EN 60950, EN 61 010 and EN 60 335 only)
The type-related settings only include limit values. These limit values can be stored for each type under an individual type designation. Measure­ments the limit values of which have been deleted are skipped.
A type can be stored as default type. This type is indicated if the rotary switch is set to Auto.
If no default type has been stored, the following appears for type in the Auto rotary switch position: GENERAL and the general limit values of the se- lected test standard apply.
Overview of adjustable measuring parameters (see chapter 15.3 for their meaning)
Adjustable measuring parameters for test sequences according to standard
To test socket  Adapter for socket  Test standard (safety) class  Ext. Cable WITH EL1 ————— Test standard Combined Testing —————— Test standard R-INS LN-PE —— ————— Test standard AP-Type ——————  Test standard Limit values Test standard (Type) Typ e    Typ e
R-INS LN-PE ——————— Test standard R-INS AP-PE ——————— Test standard HV test ——————— Test standard Patient Auxiliary Current
Visual inspection  Test standard Manual sequence  Test standard Autostore  Test standard Mains polarity reversal  Test standard Classification  Test standard SK III U
V
R-INS LN-PE   Test standard HV test ——— — — Test standard HV test duration ——— — — Test standard Auto (test) method ——— Test standard R-PE AC > 10 A  Test standard R-PE with clip ——————— Test standard No I HL for SC I ——————— Test standard Mains wait  Test standard First measured values R-INS AP-PE —————— Test standard
DIN VDE 0701
Part 1
DIN VDE 0701
Part 240
DIN VDE 0701-0702
DINEN60950
DINEN61010
DINEN60335
Measuring parameters of the initial program window
Measuring parameters in the Test Cond. menu (test conditions)
——————— Test standard
Measuring parameters in the Setup menu
 Test standard
—————— Test standard
See chapter 8 for device parameters in the test sequence
IEC 62353
IEC 601/EN 60601
can be stored
under
GMC-I Messtechnik GmbH 37
Page 38
All of the possible sequence settings for all of the regulations are listed below.
Select the Setup... menu from the initial program window and acknowledge.
Sequence 1 (repair tests, periodic testing)
store All of the settings in the setup menu, i.e. the configu-
ration of the measuring parameters and the current limit values can be saved with this command for the selected test standard. These values remain active even after setting the selector switch to a different position, and after disconnection from mains power.
Sequence See page 39. Limit Values See chapter 4.6 on page 14. Database Start with ID no.
x: Before each measurement is started, an entry prompt appears requesting entry of an ID no. An individual number can be entered (max. 20 characters) with the keypad at the (P)SI module (optional), read in with a barcode scanner (optional) or directly selected from a list. If an incorrect entry is made: Only complete lines can be deleted, and deletion is only possible with the key at the instrument. ID no. = test sequence (Feature KB01 or
SECULIFE ST HV
).
See chapter 16 on page 58.
Additional Parameters
R-PE with clip x: Protective conductor resistance can be
determined with the help of the WZ12C current clip.
High-voltage x: A high-voltage test is conducted
(prerequisite: Feature F02 or SECULIFE ST
HV)
Sequence 2 (type tests)
Type: ........ Sequences 1 and 2 differ primarily as regards the
type parameter. The user can set up any desired type of device under test with sequence 2, for which the same limit values and measuring parameters always apply. Up to 125 different types can be defined. DUT types are predefined for sequence 1 (EN 60601). Designations made up of alphanumeric characters (max. 10 characters) can be entered at the keypad of a (P)SI module, or at a PC with a terminal program. The test sequence, including all limit values, is saved together with the selected type. We recommend using the characters which identify the respective standard as the initial characters in the designation. If a type is queried from the initial program window which does not belong to the currently selected stan­dard, these characters identify the test sequence as belonging to another standard.
Limit Values See chapter 4.6 on page 14. Sequence See page 39. Store as Default All of the settings in the setup menu, i.e. the configu-
ration of the measuring parameters and the current limit values can be saved with this command. These values remain active even after setting the selector switch to a different position, and after dis­connection from mains power.
Delete The type which has been selected in the initial pro-
gram window can be deleted.
38 GMC-I Messtechnik GmbH
Page 39

15.3 Configuring Measuring Parameters

Depending upon the test regulation, various measuring parameters can be configured for the test sequence (settings: x / – = function activated / deactivated). All of the possible parameters for all of the regulations are listed below. The Sequence... menu is accessed via the setup parameter in the initial program window for the respective regulation.
Select the Sequence... menu and acknowledge.
Select a parameter, acknowledge, change and acknowledge the change.
Visual inspection This menu appears at the very beginning of the test
sequence.
Manual sequence Each test step must be acknowledged with the
key (see test sequence in chapter 8 on page 18 regarding test duration for automatic sequence).
Autostore After testing is completed, test data are automatically
stored to the test instrument(Feature KB01), or in the (P)SI module (accessory).
Mains polarity reversal L an N are reversed each time line voltage is applied
to the test socket.
Classification Questions regarding classification appear if limit
values are exceeded, see chapter 5 on page 15.
SC III U
V
Supply voltage is measured instead of insulation resistance for active devices under test.
R-INS LN-PE Insulation resistance measurement is performed
between phase and neutral conductors, and the protective conductor.
R-INS AP-PE Insulation resistance measurement is performed from
an application part to the protective conductor.
HV test A high-voltage test is performed
(prerequisite: Feature F02 or SECULIFE ST
HV)
HV test duration Duration of a high-voltage test (5 ... 60 s)
Auto (test) method The instrument recognizes whether or not the device
under test can be switched on: Leakage or residual current, or insulation resistance and earth leakage current are measured accordingly.
R-PE AC > 10 A
Protective conductor measurement with 200 mA DC 10 A AC oder 25 A AC
(feature G00 or
(
feature
G01 or
SECULIFE ST)
SECULIFE ST HV)
,
R-PE with clip Protective conductor resistance can be determined
with the help of the WZ12C current clip
No I
for SCI Housing leakage current is not tested for SCI.
HL
Mains wait A waiting period between switching mains power on
and starting the test can be entered here, e.g. in order to suppress the display of recorded values which occur during warm-up of the device under
test. Patient auxiliary current Measurement is performed with patient auxiliary
current Adapter for socket Limit values are activated for permanently connected
devices. A device under test which is normally per-
manently installed can be connected to the test
socket via an adapter. No voltage may be applied to the
test socket when this test method is used.
First measured values A menu for the entry of the first measured value
appears during the test sequence.
GMC-I Messtechnik GmbH 39
Page 40

15.4 Testing Devices in Accordance with DIN VDE 0701, Part 1

Note
You can either change over to testing in accordance with the new VDE 0701-0702 standard or perform the following measurements in ac­cordance with the above-mentioned standard:
• Protective conductor measurement R (permanent connection or with plug)
PE
– Part 1: Test current: 200 mA DC – Appendices: Test current
Test current: 10 A AC (Feature G00 or
: 25 A AC (Feature G01 or
SECULIFE ST HV)
SECULIFE ST)
• High-voltage test as supplement (Appendix E) (Prerequisite: Feature F02 or SECULIFE ST
• Insulation resistance measurement R
HV)
INS
Part 1
The following safety class I through III appliances and electric equipment can be tested with this standard, for example:
• Devices with electric motors
• Electric heating devices
• Electric tools
• Light fixtures
• Stereo equipment, television sets
SC I extension cables can also be tested (see chapter 15.7 on page 46).
Appendices (previously part 260)
Appendix E: Electric tools
To Socket This is the default setting. Refer to chapter 7 on page
17 for other types of connection.
Class If the DUT has been connected to the test socket,
safety class recognition is performed (SC I or SC II). Otherwise, the safety class must be specified manu­ally.
Ext. Cable
WITH EL1
x: Extension cables or connector cables which are longer than 5 m can be tested with the help of the EL1 adapter (optional), either separately or in combination with a device, see chapter 15.7.
ID No.
See parameters database in chapter 15.2 on page 37.
Setup See chapter 15.2 on page 37 regarding setup of the
measuring sequence.
Check connection parameters and start test.
Extension cables can only be tested with the VDE 0701-0702
(VDE 0701 part 1) parameter settings, if the EL1 accessory
adapter is used (see chapter 15.7 on page 46).
40 GMC-I Messtechnik GmbH
Page 41
Test Sequence per VDE 0701
Select connection, select test regulation
VDE 0701 part 1
, classify DUT (SC I, II or III),
part 1: extension cable X/– (with/without), Appendix E: HV test X/– (with/without)
Visual inspection
O.K.?
SC I: RPE
OK?
I
L
OK?
Short-circuit
at DUT?
Te st OK?
Yes /N o
No
Yes
Display results, save/print report
Yes
No
Yes
No
No
Tes t
O.K.?
Framed with dashed line: The test is only run if it has been activated in the initial program window, or in the Setup menu under Sequence...
Switch line voltage to test socket, start function test
Start
function test
Safety Class I only: Contact all exposed, conductive parts with the
probe. The test can be repeated as often as desired for various pro­tective conductor parts with the manual sequence*.
* If it is not clear whether or not all ex-
posed, conductive parts are connect­ed to one another or to the protective conductor, testing can be performed in the manual mode.
Switch DUT on
RINS+I
EL
OK?
or
Yes
No
No
Yes
Yes
Test HV? HV OK?
Apply high voltage
Yes
No
Yes /N o
GMC-I Messtechnik GmbH 41
Page 42

15.5 Testing Devices in Accordance with DIN VDE 0701, part 240

Attention!
!
You can either change over to testing in accordance with the new VDE 0701-0702 standard or perform the following measurements in ac­cordance with the above-mentioned standard:
Testing safety class I and II data processing equipment and office machines as individual devices and in combination with one another.
The following measurements can be performed in accordance with the above mentioned standard:
• Protective conductor R Test current: DC ±200 mA
• Housing leakage current I
• According to DIN VDE 0701, part 240, the device protective conduc­tor must be tested after maintenance, repair or modification of data processing equipment and office machines, and exposed, conductive parts must be tested for the absence of voltage. This applies to:
• Safety class I devices for all exposed, conductive parts which are accessi­ble to the user, and which are not connected to the protective conductor
• Safety class II devices (totally insulated devices) for all exposed, conductive parts which are accessible to the user
with the mains plug poled in both directions.
Setting Up the Test Sequence
See chapter 15.4 regarding the test sequence.
Special Parameters
Combined Testing
Safety class I and II devices can be tested individually or in combination. All protective conductor connections are tested first for interconnected safety class I devices, and then – as is also the case for interconnected safety class II devices – all exposed, conductive parts.
Connecting the Device Under Test
õ Connect the test instrument and the DUT as described below:
– Connect both devices to separate mains outlets.
The outlets to which the test instrument and the safety class I DUT are connected must share a common protective conductor!
– Or connect the test instrument to the mains and the DUT to the test
socket at the test instrument.
(permanent connection or with plug)
PE
Data Processing / Office Machines
Permanently connected or with plug
To test socket at the instrument
The requirement for testing with the mains plug poled in both directions can be fulfilled
by connecting the DUT to the test socket at the instrument, and by activat­ing mains polarity reversal in the setup menu under “Sequence”. Each time the key (14) is activated, phase conductor L and neutral conductor N are reversed at the test socket.
Testing with mains polarity reversal or with the mains plug poled in both directions results in interruption of supply power to the affected data processing equipment or office machine. This test may thus only be conducted with the consent of the operator of the data processing equipment or office machine. If the DUT is defective, the electrical system’s RCCB may be tripped during testing which would also result in interruption of supply power to the affected equipment or office machine. The manufacturer of the test instrument assumes no liability for the loss of data or other damage which result from use of the test instrument.
42 GMC-I Messtechnik GmbH
Page 43
Test Sequence per VDE 0701, part 240
Select connection, select test regulation VDE 0701, part 240 classify DUT (SC I, II or II I), combination testing X/– (yes/no)
Visual inspection
OK?
SC I: RPE
OK?
Short-circuit
at DUT?
Yes /N o
No
Display results (for combination testing: additional display of differential resistance), save / print report
Yes
No
No
Tes t OK?
Framed with dashed line: The test is only run if it has been activated in the initial program window, or in the Setup menu under
Sequence....
Switch line voltage to test socket, start function test
Start
function test
Switch line voltage to test socket, start measurement
Reverse mains polarity, perform measurement again
Contact all exposed, conductive parts with the probe. The test can be repeated as often as desired for various protective conductor parts with the manual sequence*.
* If it is not clear whether or not all
exposed, conductive parts are connected to one another or to the protective conductor, testing can be performed in the manual mode.
Switch DUT on
Yes
IL
OK?
Yes
No
GMC-I Messtechnik GmbH 43
Page 44

15.6 Testing Devices in Accordance with DIN VDE 0701-0702:2008

The following measurements can be performed in accordance with the above mentioned standard:
• Protective conductor R Test current: DC 200 mA Test current: 10 A AC Test current: 25 A AC
• Insulation measurement R
exists that voltage sensitive components in data processing systems
(permanent connection or with plug)
PE
(Feature G00 or (Feature G01 or
INS
SECULIFE ST)
SECULIFE ST HV)
(can be deactivated, e.g. if danger
might be damaged) plus equivalent leakage current
or
• Contact current for safety class II
or
• Residual current
Check connection parameters and start test.
To Socket This is the default setting. Refer to chapter 7 on page
17 for other types of connection.
Class If the DUT has been connected to the test socket,
safety class recognition is performed (SC I or SC II). Otherwise, the safety class must be specified manu­ally.
Ext. Cable
WITH EL1
x: Extension cables or connector cables which are longer than 5 m can be tested with the help of the EL1 adapter (optional), either separately or in combi­nation with a device, see chapter 15.7.
R-INS LN-SL x: Insulation resistance measurement is performed.
ID No. See parameters database in chapter 15.2 on page
37.
Setup See chapter 15.2 on page 37 regarding setup of the
measuring sequence.
44 GMC-I Messtechnik GmbH
Page 45
Test Sequence per DIN VDE 0701-0702:2008
Select connection, select test standard VDE 0701-0702, classify DUT (SC I, II or I I I), extension cable X/– (with/without)
Visual inspection
OK?
SC I: R
PE
OK?
Short-circuit
at DUT?
Yes /N o
No
Display results; save/print report
Yes
No
No
Te st OK?
Framed with dashed line: The test is only run if it has been activated in the initial program window of in the Setup menu under Sequence ...
Switch line voltage to test socket
Contact all exposed, conductive parts with the probe
Start
function test
Switch DUT on
R
INS
OK?
Yes
I and/or IC with MLV O.K.: Yes
I
L
*
OK?
Yes
* • Equivalent leakage current IEA
• Residual current I and
contact current (direct) I
B
with MLV
OK?
Enter manufacturer’s limit values
Yes
start function test
No (I
EL
not OK)
No
No (I or I
C
not OK)
dev.
heating power?
No
No
Test without insulation measurement,
if there are reservations against an insulation measurement
Protective conductor: not available, not connected, has no contact
I
L
O.K.: Yes
GMC-I Messtechnik GmbH 45
Page 46
15.7 Testing Extension Cables for VDE 0701-0702 (VDE 0701 Part 1)
Note
Note
PROBE 5 4
(21)
EL1
SECUTEST
Device Under Test = Extension Cable
Tes t Soc ke t
(optional EL1 adapter)
Extension Cables up to 5 m Long
Protective conductor resistance between the earthing contact at the mains plug and all exposed metal parts may not exceed 0.3 for protec­tion class I devices. This value may not exceed 1 for permanently con-
nected data processing systems (DIN VDE 0701, part 240).
Extension and Connector Cables Longer than 5 m
An additional cable resistance of 0.1 , however, not more than 1 , is al­lowable as of a length of 5 m for each additional 7.5 m.
Resistance testing for cables longer than 5 m is thus advisable (see also limit values on page 20.
The EL1 accessory adapter is required for testing for short­circuiting and interruption of single-phase extension cables.
Connecting the Extension Cable or the Multiple Outlet
Performing the Test
õ Connect the extension cable to the EL1 adapter as shown in the figure
above.
õ Select the “
Ext. Cable
and acknowledge with the key:
WITH EL1
” test from the initial menu with the cursor
x
.
õ Select “Start Testing” with the key.
õ Start the test sequence with the key.
õ
First perform and acknowledge visual inspection of the extension cable.
õ Enter the length of the cable with the
and keys. Acknowledge with
the key.
For conductors with a rated current of > 16 A, the cross section has to be taken into account during measurement (only 3-phase adapter application, not EL 1).
The keys in the adapter’s handle have no function.
46 GMC-I Messtechnik GmbH
Page 47

15.8 Testing Multiple Outlets for VDE 0701-0702 (optional EL1 adapter)

õ Select “
Ext. Cable
pear in the line: “
WITH EL1
Ext. Cable
” in the initial menu. The following must ap-
WITH EL1
x“. “Manual sequence” must be se-
lected.
õ A visual inspection must always be performed. It may thus be
necessary to unreel the cable from its drum or reel.
õ Protective conductor resistance measurement: Contact the first outlet
with the EL1 adapter. Each time you are ready to contact the next outlet, press the key to repeat the test.
GMC-I Messtechnik GmbH 47
Page 48

15.9 Testing in Accordance with DIN EN 60 950

Tests for safety class I and II data processing equipment and office machines. The following measurements can be performed in accordance with the
above mentioned standard:
• Protective conductor R Test current: : 25 A AC
Testing of dielectric strength by applying DC high voltage (1.5-fold value)
PE
(Feature G01 or
SECULIFE ST HV)
(Prerequisite: Feature F02 or SECULIFE ST HV)
• Device leakage current I
According to DIN EN 60950, data processing equipment and office ma­chines must be subjected to a device protective conductor test and a high-voltage test before they are brought into circulation. This applies to:
• Safety class I devices for all exposed, conductive parts which are accessible to the user, and which are not connected to the protective conductor
• Safety class II devices (totally insulated devices) for all exposed, conductive parts which are accessible to the user
• Housing leakage current
with the mains plug poled in both directions
Connecting the Device Under Test
õ Connect the test instrument to the mains, and the device under test to
the test socket at the test instrument.
Check connection parameters and start test.
To Socket This is the default setting. Refer to chapter 7 on page
17 for other types of connection.
Class If the DUT has been connected to the test socket,
safety class recognition is performed (SC I or SC II). Otherwise, the safety class must be specified manu­ally.
Type A specific DUT type can be selected from a list of
devices under test, if these have been entered in the setup menu.
ID No. An individual serial number (up to 10 characters) can
be entered with the keypad at the (P)SI module (accessory), or can be read in with a barcode scan­ner (accessory). If an incorrect entry is made: Only complete lines can be deleted, and deletion is only possible with the key at the instrument.
Setup Refer to chapter 15.2 on page 37 regarding setup of
the measuring sequence.
48 GMC-I Messtechnik GmbH
Page 49
Test Sequence per EN 60 950
Select connection, select test regulation EN 60950 classify DUT (SC I, II or I II), setup/sequence: R-INS / HV test X/– (with/without)
Visual inspection
OK?
SC I: R
PE
OK?
Short-circuit
at DUT?
Te st
OK?
No
Yes
Display results, save / print report
Yes
No
Yes
No
No
Te st OK?
Framed with dashed line: The test is only run if it has been activated in the Setup menu under
Sequence...
.
Switch line voltage to test socket, start function test
Start
function test
Safety Class I only: Contact all exposed, conductive parts with the
probe. The test can be repeated as often as desired for various protective conductor parts with the manual sequence*.
* If it is not clear whether or not all
exposed, conductive parts are connected to one another or to the protective conductor, testing can be performed in the manual mode.
Switch DUT on
Activate high-voltage
Test R
INS
?
Yes
R-INS OK?
Test HV?
Yes
HV OK?
I
L
OK?
No
No
Switch line voltage to test socket, start measurement
Reverse mains polarity, perform measurement again
Yes /N o
Yes /N o
GMC-I Messtechnik GmbH 49
Page 50

15.10 Testing Devices in Accordance with EN 61010

The following measurements can be performed in accordance with the above mentioned standard. Connection must be made via the test socket:
• Protective conductor R Test current: 10 A AC Test current: 25 A AC
PE
(Feature G00 or (Feature G01 or
SECULIFE ST)
SECULIFE ST HV)
• Testing of dielectric strength by applying high voltage with direct volt­age (1.5-fold value) (Prerequisite: Feature F02 or SECULIFE ST
• Insulation measurement R
(can be deactivated)
INS
• Housing leakage current under normal condition I condition I
with interrupted protective conductor
HLSF
HV)
and single-fault
HLNC
Check connection parameters and start test.
To Socket This is the default setting. Refer to chapter 7 on page
17 for other types of connection.
Class If the DUT has been connected to the test socket,
safety class recognition is performed (SC I or SC II). Otherwise, the safety class must be specified manu­ally.
Type A specific DUT type can be selected from a list of
devices under test, if these have been entered in the setup menu.
ID No. An individual serial number (up to 10 characters) can
be entered with the keypad at the (P)SI module (accessory), or can be read in with a barcode scan­ner (accessory). If an incorrect entry is made: Only complete line can be deleted, and deletion is only possible with the key at the instrument.
Setup Refer to chapter 15.2 on page 37 regarding setup of
the measuring sequence.
50 GMC-I Messtechnik GmbH
Page 51
Test Sequence per EN 61 010
Select connection, select test standard EN 61010, classify DUT (SC I, II or I II), setup/sequence: R-INS-/HV test X/– (with/without)
Visula inspection
OK?
SC I: RPE
OK?
Short-circuit
at DUT?
Te st
OK?
No
Yes
Display results, save / print report
Yes
No
Yes
No
No
Te st OK?
Framed with dashed line: The test is only run if it has been activated in the Setup menu under
Sequence...
.
Switch line voltage to test socket, start function test
Start
function test
* If it is not clear whether or not all
exposed, conductive parts are connected to one another or to the protective conductor, testing can be performed in the manual mode.
Switch DUT on
Apply high voltage
Test R-I NS ?
Yes
R-INS OK?
Test HV ?
Yes
HV OK?
IHL
NC/SFC
No
No
Yes /N o
Yes /N o
Safety Class I only: Contact all exposed, conductive parts with the probe. The test can be repeated as often as desired for various protective conductor parts with the manual sequence*.
Select connection, select test standard EN 61010, classify DUT (SC I, II or I II), setup/sequence: R-INS-/HV test X/– (with/without)
Visula inspection
OK?
SC I: R
PE
OK?
Short-circuit
at DUT?
Te st
OK?
No
Yes
Yes
No
Yes
No
No
Te st OK?
Framed with dashed line: The test is only run if it has been activated in the Setup menu under
Sequence...
.
Switch line voltage to test socket, start function test
Start
function test
Switch DUT on
Apply high voltage
Test R
INS
?
Yes
R
INS
OK?
Test HV ?
Yes
HV OK?
I
HL
NC/SFC
No
No
Yes /N o
Yes /N o
Safety Class I only: Contact all exposed, conductive parts with the probe. The test can be repeated as often as desired for various protective conductor parts with the manual sequence*.
GMC-I Messtechnik GmbH 51
Page 52

15.11 Testing Devices in Accordance with EN 60335

The following tests can be performed in accordance with VDE 0700, part 500, (as part of DIN EN 50106:1998) in compliance with special rules for routine testing of devices for which EN 60335-1 and EN 60967 apply:
• Testing of the protective conductor connection by means of resistance measurement R Test current: 10 A AC Test current: 25 A AC
Testing for dielectric strength by applying DC high-voltage (1.5-fold value)
(permanent connection or with plug)
PE
(Feature G00 or (Feature G01 or
SECULIFE ST)
SECULIFE ST HV)
(Prerequisite: Feature F02 or SECULIFE ST HV)
•Function test
The following tests can be performed in accordance with EN 60335-1:1994:
• Testing for dielectric strength by applying high-voltage (Prerequisite: Feature F02 or SECULIFE ST
HV)
• Testing for Equivalent Leakage Current
Additional testing options:
• Insulation resistance measurement R
INS
• Residual current
Check connection parameters and start test.
If desired, the test sequence can be adapted to the respective device
under test in the following menu in Setup... under Sequence....
To Socket This is the default setting. Refer to chapter 7 on page
17 for other types of connection.
Class If the DUT has been connected to the test socket,
safety class recognition is performed (SC I or SC II). Otherwise, the safety class must be specified manu­ally.
Type A specific DUT type can be selected from a list of
devices under test, if these have been entered in the setup menu.
ID No. An individual serial number (up to 10 characters) can
be entered with the keypad at the (P)SI module (accessory), or can be read in with a barcode scan­ner (accessory). If an incorrect entry is made: Only complete line can be deleted, and deletion is only possible with the key at the instrument.
Setup Refer to chapter 15.2 on page 37 regarding setup of
the measuring sequence.
52 GMC-I Messtechnik GmbH
Page 53
Test Sequence per EN 60 335
Select connection, select test regulation EN 60335 classify DUT (SC I, II or I II), setup/sequence: R-INS / HV test X/– (with/without)
Visual inspection
OK?
SC I: R
PE
OK?
Short-circuit
at DUT?
No
Yes
Display results, save / print report
Yes
No
Yes
No
No
Te st OK?
Framed with dashed line: Test is only run if it has been activated in the Setup menu under
Sequence....
Switch line voltage to test socket, start function test
Start
function test
Safety Class I only: Contact all exposed, conductive parts with the
probe. The test can be repeated as often as desired for various protective conductor parts with the manual sequence*.
* If it is not clear whether or not all
exposed, conductive parts are connected to one another or to the protective conductor, testing can be performed in the manual mode.
Switch DUT on
Activate high-voltage
Test R
INS
?
Yes
R
INS
OK?
Test HV?
Yes
HV OK?
No
No
Yes /N o
Yes /N o
Tes t
O.K.?
Tes t IEL?
NC/SFC
Test IRC?
IEL OK?
NC/SFC
IRC OK?
Yes
Yes
Yes /N o
Yes /N o
No
No
GMC-I Messtechnik GmbH 53
Page 54

15.12 Testing in Accordance with IEC 62353/VDE 0751

The following measurements can be performed in accordance with the above mentioned standard:
• Protective conductor measurement R Test current: 200 mA DC, Test current: 10 A AC Test current: 25 A AC
(Feature G00 or (Feature G01 or
PE
SECULIFE ST)
SECULIFE ST HV)
• Insulation measurement (can be additionally activated) – R-INS LN-PE
(insulation resistance, LN to protective conductor)
– R-INS AP-PE (insulation resistance, app. part to protective conductor)
• Equivalent device leakage current I
• Equivalent patient leakage current I
EDL
EPL
• Device leakage current (direct or residual current)
• Contact current
• Patient leakage current (direct or mains at application part )
Leakage current is converted to reference voltage (see limit values chapter 8 on page 18). Reference voltage must be adapted to the supply voltage range.
Check connection parameters and start test.
To Socket This is the default setting. Refer to chapter 7 on page 17 for
other types of connection.
Class If the DUT has been connected to the test socket, safety
class recognition is performed (SC I or SC II). In all other cases, or if it is not clear whether or not all exposed, conductive parts are connected to one another or to the protective conductor, the safety class can be selected manually.
Type Select the device type (DUT) from the list.
If “old devices” is selected, limit values from DIN VDE 0701­0702 are used.
AP-Type:... (BF): Application parts are recognized automatically, and can
be manually changed as well:
Select the App. prt. ... line with the
edge with the
key and change with the or key.
or key, acknowl-
02: The number of configured groups is displayed here.
After selecting the App. prt. ... line with the or key,
acknowledging twice with the key and then pressing the
or key, the “Configure application parts” menu appears
(see chapter 15.13 on page 56).
ID No. See parameters database in chapter 15.2 on page 37. Setup Refer to chapter 15.2 on page 37 regarding setup of the
measuring sequence.
54 GMC-I Messtechnik GmbH
Page 55
Test Sequence per IEC 62 353/VDE 0751
Select connection, select test standard IEC 62353/VDE 0751, classify DUT (SC I, I I or III), application part? (type B/BF/CF)
Visual inspection
OK?
SC I: R
PE
OK?
R
INS
OK?
Short circuit
at DUT?
Yes /N o
No
Display results; save / print reports
Yes
No
Te st OK?
Framed with dashed line: The test is only run, – if it has been activated in the initial
program window, or in the Setup menu under Sequence
or – if possible at all
Switch line voltage to test socket, start function test
Start
function test
I
EDL
I
RC
Contact all exposed, conductive parts with the probe. The test can be repeated as often as desired for various protective conductor parts with the manual sequence.*
* If it is not clear whether or not all
exposed, conductive parts are con­nected to one another or to the protective conductor, testing can be performed in the manual mode.
Measurement 1: Contact all exposed, conductive parts with the probe
Yes
Switch DUT on
I
LC
I
PL
OK?
Connect application parts.
I
EPL
, only if the DUT can be switched on
I
PA
, for type B only
Mains to application part for type F
No
IRC or ILC, only if the DUT is switched electronically
OK?
No
NoNoN
Yes
Yes
Yes
I
EPL
Measurement 2
: Connect application parts
If the limit values of leakage currents are exceeded in the following sequence, the first measured values or the manufacturer’s values can be entered instead for those leakage currents which did not pass.
GMC-I Messtechnik GmbH 55
Page 56

15.13 Testing in Accordance with EN 60601 (Feature KA01)

The following leakage and auxiliary currents can be measured in accor­dance with the above mentioned standard in the operating mode, as well as under normal and single-fault conditions:
• Protective conductor measurement R test current: 10 A AC Test current: 25 A AC
• Earth leakage current I
• Housing leakage current I
(Feature G00 or
(Feature G01 or
PE
HL
,
PE
SECULIFE ST)
SECULIFE ST HV)
• Patient leakage current IPL (with nominal voltage at application part)
• Patient auxiliary current I
PA
Leakage current is converted to reference voltage (see limit values chapter 8 on page 18). Reference voltage must be adapted to the supply voltage range.
Check connection parameters and start test.
To Test Socket This is the default setting. For other connection types, please
refer to chapter 7 on page 17.
Class If the DUT has been connected to the test socket, safety
class recognition is performed (SC I or SC II). In all other cases, or if it is not clear whether or not all exposed, conduc­tive parts are connected to one another or to the protective conductor, the safety class can be selected manually.
Type The DUT type can be selected from a list.
If the limit values of the 3rd issue are to be taken into account, select a DUT type ending with ... 3rd under this parameter.
Cond. Various test conditions can be activated here, including
insulation resistance measurement.
App Prt See below and on page 54. ID-No. See parameters database in chapter 15.2 on page 37. Setup Refer to chapter 15.2 on page 37 regarding setup of the
measuring sequence.
Configure Application Parts
Entry is made here indicating whether or not the application parts are to be tested. Jacks A through K (for connection of cables or probes) can also be assigned to groups (application parts) for mutual testing.
Selecting Preset Test Combinations
õ Use the cursor to select test combinations with groups of 1, 2, 5 or 10
application parts and acknowledge with the key. The groups are assigned automatically after the application parts have been selected.
Selecting User Defined Test Combinations
õ Select the respective application part from the JACK column with
the cursor and acknowledge with the key. You can set up a user defined group for each application part which includes 1 to 10 application parts with the cursor keys in the GR column (GRoup). Acknowledge your settings with the key.
As long as at least one group number has been entered, the application parts test is preset in the “To Socket” window. The application part type with the strictest limit value takes precedence in the initial window. All groups are set to this type. Various types can be assigned to the groups with “direct print” (adjustable option in the setup menu). If testing is not to be performed with application parts, the group assign­ment must be cancelled with the “clear” function.
The TYPE is filled in automatically if the safety class has already been entered to the “To Socket” window.
Flowchart, see next page.
56 GMC-I Messtechnik GmbH
Page 57
Visual inspection
OK?
SC I: RPE
OK?
Short-circuit
at DUT?
Tes t
OK?
Yes /N o
No
Display results, save / print report
Yes
No
Yes
No
No
Framed with dashed line: The test is only run if it has been activat­ed in the initial program window, or in the Setup menu under Sequence...
.
Switch line voltage to test socket, start function test
Start
function test
Run all
tests?
Contact all exposed, conductive parts with the probe. The test can be repeated as often as desired for various protective conductor parts with the manual sequence
1)
1)
If it is not clear whether or not all exposed, conductive parts are connected to one another or to the protective conductor, testing can be performed in the manual mode.
No
Set test conditions:
1. Potential conductor interrupted
2. Operational earth interrupted
3. Connect housing to protective conductor
4. Connect application parts to PE
Earth leakage current, housing leakage current2), patient leakage current
3)
, patient auxiliary current
4)
Mains poles
reversed?
No
Reverse L and N
Set single-fault conditions:
1. N interrupted
5)
2. Normal condition
3. Protective conductor interrupted
4. Mains to application part
5)
Run
all SFC?
Yes
Yes
No
2)
no N interrupted
3)
Test is only performed if at least one type BF or CF application part is connected.
4)
Test is only performed if at least two type BF or CF application parts are connected
5)
Test method: equiv­alent leakage current
R-INS
OK?
Yes
Yes /N o
Switch DUT on
Activate high-voltage
Yes
HV OK?
No
Tes t HV?
Additional insulation resistance measurement between application parts an PE
Select connection, select test regulation EN 60601 classify DUT (SC I, II or I II), test conditions (amongst others R
INS
measurement), application part? (type B/BF/CF)
Prüfung
O.K.?
GMC-I Messtechnik GmbH 57
Page 58
16 Storing in (P)SI Module (Accessory) and Database Operations
(Feature KB01)

16.1 Storing Measurement Data in the (P)SI Module

Upon completion of a measurement – message „Test passed/failed“ is displayed – you can file the measurement data in the memory of the (P)SI module.
õ Press the STORE key at the (P)SI module.
A text entry field is displayed.
õ You may now enter your comments on the measurement
and/or an ident. no.
õ Press the STORE key once more in order to save the measurement data
and your comments. The message: „Data are being stored“ is shown on the display.
A detailed description is included in the operating instructions for your (P)SI module in the chapter „Display, print and store protocol“.

16.2 Database Operations

16.2.1 Storing Test Results to the Test Instrument

If no (P)SI module has been connected, up to 125 reports can be stored to the test instrument (without function test values and without comments on DUT). The reports can be viewed as required at the instrument and can be printed out, for example with the help of a terminal program, see chap­ter 18.
The reports are sorted by time and date and are displayed with the ID number. If no ID number was assigned, date and time are automatically saved instead.
As an alternative, consecutive numbering can be selected.
16.2.2 Uploading Report Templates into Test Instrument, Reading Out From the
Test Instrument, Editing at the PC and Re-Saving to the Test Instrument
The „Test result“ menu allows for storage of up to 4 report templates in the test instrument. Report templates are uploaded from a file at a PC to the test instrument (“Load file”). Templates can also be uploaded from the test instrument to a PC (“Templates from Secutest”), and then edited and re­saved.
After completion of a test (“Passed” or “Failed” appears at the instru­ment’s LCD), the test results are read out to one of these report templates
in the form of report data (depending upon the selector switch position or test regulation) via the RS 232 interface. The report menu in the test in­strument is activated with the and keys to this end. In order to read out the report data at a PC, it is necessary to connect the SECUSTORE stor- age adapter with the RS232 interface. The report can, for example, be displayed with the PC analysis program WinProfi (as from version 3.06) or ETC (as from version 1.22).

16.2.3Reading Out and Saving Test Results / Test Data from the (P)SI Module

Test results saved to the test instrument can be displayed, edited, printed (only PSI module) or saved after reading them into a PC using the desired report template.
Data can be processed immediately after completion of the test, or from the database (Feature KB01). Data can be saved with or without a report template (e.g. for further processing with PS3).
Test results saved to the (P)SI module can also be read out, printed (only PSI module), saved or processed by means of a report template.
Easy method of printing out reports: Activate the „test result“ menu in the update and options installation pro-
gram. When the test result is displayed in the test instrument select the print function (press key „Cursor up“, then set the cursor on „print“ and press ENTER).
17 Recognition of Probe to Protective Conductor
(Feature KD01)
The protective conductor measurement is expanded to include the function: “automatic recognition of measuring point change”. During protective conductor measurement, the instrument recognizes whether or not the probe is in contact with the protective conductor, and indicates these two possible conditions by means of acoustic signals.
This function is helpful if several protective conductor connections need to be tested. The function can be activated in the “test sequence setup” menu with “auto measuring point” (see “changes to the switch position menu” above).
58 GMC-I Messtechnik GmbH
Page 59

18 Storing Test Results and Printing in Report Form

(read out)
You can shift to the Protocol menu from any of the displayed test results (1 page) with the help of the key.
Storing to the Test Instrument
Measurement results for the current test can be stored to the test instru­ment, the results of the current test can be printed out to the corresponding re­port form, previously stored test results can be queried (scroll: Feature KB01, see chapter 16) and all stored mea­surement results can be read out from this menu.
The report template automatically complies with the standard of the se­lected switch position, provided that parameter „Select template“ has been deactivated.
If parameter „Select template“ is activated, it is possible to select one of 5 different report templates. Templates 1 to 4 can be modified via the up­date and options installation program SECU-Up, see chapter 16.2.2.
st
Direct Print-Out (Feature KE01 in Combination with a PSI Module or SECUSTORE Storage Adapter)
After completion of each test (individual test or at the end of a test sequence), test results are read out directly via the RS232 interface.
When the PSI module is connected (accessory, not part of the standard equipment), the result is directly printed on paper. When the SECUSTORE storage adapter is connected, the test result is stored to memory.
Storing to the SECUSTORE Storage Adapter (Accessory)
Connect the SECUSTORE storage adapter with the test instrument via the RS232 interface. The (P)SI module may not be connected.
GMC-I Messtechnik GmbH 59
Page 60

19 Characteristic Values

Please refer to chapter 1.2 on page 7 to check as to which of the following measurements is required for the relevant regulation
Open-
Nominal Current
U
0
DC
...
N
>1mA <10mA
N
– 20/
Measured Quantity
Device Protective Conductor Resis­tance R
PE
Insulation Resistance R
INS
Equivalent Leakage Current I
EL
Meesuring Range/ Nominal Range of
Use
0.000 ... 2.100 1m
2.11 ... 31.00 10 m
Resolu-
tion
Nominal
Voltage
U
N
Circuit
Voltage
4.5 ... 9 V
0.000 ... 2.100 1m —<6VAC—
0.050 ... 1.500 M 1k
10.1 ... 310.0 M 100 k
0.00 ... 21.00 mA 10 A
20.1 ... 120.0 mA 100
A
50 ...
500 V DC
1.0 • U
1.5 • U
230 V
+10 %
Contact Current (Absence of Voltage) I
probe
Residual Current I between L and N
Equivalent Device and/or Patient Leakage Current I
EDL and/or IEPL
Leakage Current I
All Leakage Current
7
I
L
1
As of 25 mA: shutdown by residual current measurement within 100 ms
2
Exception earth leakage current: only 0.000 3.100 mA
3
Measuring circuit is highly resistive, indication at display
4
Measurement with AC test current (Feature G00 or G01) is not possible at the sockets (1 through 3). Feature G01 or SECULIFE ST SC5 special cable is used.
5
Test duration max. 40 s, protection against overheating: measurement cannot be restarted until a waiting period of 1 minute has elapsed.
6
Calculated value
7
AC and DC are measured for patient leakage current and patient auxiliary current.
0 ... 3.500 mA 1 A— — — — 2k
0.000 ... 3.100 mA
RC
3.00 ... 31.00 mA
A
1
10
————— —
A
0.0 ... 310.0 A0.1A
0.000 ... 2.100 mA 1 A
2.101 ... 21.00 mA 10
20.1 ... 120.0 mA 100
20.0 ... 310.0 A 100 nA
L
0.210 ... 3.600 mA 1 A
3.10 ... > 15.00 mA
9
10 A
HV: > 25 A; Short-circuit current is less than 25 A if the
A
A
110 % of
highest line
voltage
230 V
– 20/
+10 %
———1k
6
60 GMC-I Messtechnik GmbH
Short­Circuit
I
Current
N
>200mA
>10A
AC
—<3.5mA>72k
Internal
Resis-
tance
I
K
DC
4
>5 s
R
I
——
no protection
2k
—<3.5mA>72k
8
The data are only valid for the values displayed at the test instrument. Data which are transmitted via the RS232 interface may deviate.
9
only applies to earth leakage current or mains at application part
Key: of rdg. = of reading (measured value), D = digit
Reference
Resistance
R
REF
1k
50
Measuring
Uncertainty
(5% rdg.+10 digits)
> 10 d
(5% rdg.+10 digits
(10% rdg.+10 digits)(10 % rdg.+10digits)
(5% rdg.+10 digits)
(5% rdg.+10 digits)
(10% rdg.+10 digits)
> 10 digits
(5% rdg.+10 digits)
(5% rdg.+10 digits)
8
Intrinsic
Uncertainty
(2.5% rdg.+ 5 digits)
> 10 digits
(2.5 % rdg.+5 digits)
)
> 10 digits
(2.5 % rdg.+5 digits)
> 10 digits
(2.5 % rdg.+5 digits)
> 10 digits
(5 % rdg.+5 digits)
> 10 digits
(2.5 % rdg.+5 digits)
> 10 digits
(2.5 % rdg.+5 digit)
> 10 digit
Overload Capacity
8
Value Time
253 V cont.
253 V cont.1.01 ... 10.00 M 10 k
253 V cont.
253 V
253 V
253 V
cont.
11
cont.
1 3
cont.
1 3
5
2
Page 61
Func-
Measured Quantity Measuring Range /
tion
Nominal Voltage U
Load C urrent I
Active Power P 0 ... 3700 W
Apparent Power AP 0 ... 4000 VA 1 VA Calculated Value U
Functions Test
Power Factor PF,
sinusoidal: cos
Residual Current
between L and N
U
AC/DC
U
Probe
R Resistance 0 ... 150.0 k 100 <20V– 1.1mA
I
CLIP
Tem p
10
Measured value P and calculated value S are compared, and the smaller value is displayed.
Volta ge
Probe Voltage
Current via
Clip-On Current-
Voltage Converter
WZ12C
Tem p er at u re
with Pt100 / Pt1000
Sensor
Nominal Range of Use
103,5 V  126,5 V
LN
207.0 ... 253.0 V
0 ... 16.00 A
a
0.00 ... 1.00 0.01 Calculated Value P / AP, Display > 10 W (10% rdg.+5 digits)
I
0.00 ... 31.00 mA 10 A——
0 ... 253.0 V
, and
0 ... 253.0 V
, and
0.000 ... 10.00 A 1mA 1.5M
0 ... 100 A
– 200 ... – 50
– 50.1 ... + 300.0
+300 ... +850
Reference Ranges
Line Voltage 115 / 230 V 0.2% Line Frequency 50/60 Hz 0.1% Waveshape sine (deviation between effective and
Resolu­tion
0.1 V (2.5%rdg.+5 digits) 253 V cont.
10 mA (2.5%rdg.+5 digits) 20 A 10 min
RMS
10
1W
0.1 V (5% rdg.+10 d)
0.1 V
1A 1.5M 253 V cont.
Open­Circuit Voltage U
0
Short­Circuit Current I
K
Internal Resis­tance R
I
• I
LN
V
C1C
C0.1C (1% rdg.+1 C) 10 V cont.
<20V– 1.1mA
C1C (2% rdg.+1 C) 10 V cont.
Nominal Ranges of Use
Line Voltage 103.5 V 126.5 V or 207 V 253 V Line Frequency 50 Hz or 60 Hz Line Voltage Waveshape sine
rectified value < 0.5%)
Temperature 0 C + 50 C
Ambient Temperature + 23 C 2K Atmospheric Humidity 50% relative 5% Load Resistors linear
Measuring Uncertainty
(10% rdg.+10 d)
> 10 digits
8
Intrinsic Uncertainty
(5% rdg.+10 digits)
>20digits
(5% rdg.+10 digits)
>20digits
(5% rdg.+5 digits)(2.5%rdg.+5 digits)
>10digits
(2.5%rdg.+5 digits)
>10digits
8
Overload Capacity
Value Duration
253 V cont.
20 A 10 min
11
253 V cont.
253 V cont.
(1% rdg.+3 digits) 253 V cont.(3% rdg.+10 digits)
>10digits without clip
253 V cont.
(2% rdg.+1 C) 10 V cont.
GMC-I Messtechnik GmbH 61
Page 62
Influencing Quantities and Influence Error
Influencing Quantity / Sphere of Influence
Position Change E1
Change in Test Setup Supply Power E2 2.5
Temperature Fluctuation
0 21
C and 25 40 C
Current at Device Under Test E4 2.5
Low-Frequency Magnetic Fields E5 2.5
Impedance at Device Under Test E6 2.5
Capacitance, Insulation Measurement E7 2.5
Waveshape of Measured Current
49 51 Hz 2 for capacitive load (for equivalent
45 100 Hz 1 (for contact current)
Designation per DIN VDE 0404
E3
E8
Influence Error % of Measured Value
Specified influence error applies per 10 K change in temperature:
1 in case of PE measurement
0.5 for all other measuring ranges
leakage current)
2.5 for all other measuring ranges
Ambient Conditions
Storage Temperature – 20 C ... + 60 C Operating Temperature – 10 C ... + 50 C Accuracy Range 0 C ... + 50 C Relative Humidity max. 75%, no condensation allowed Elevation max. 2000 m Deployment indoors, outdoors: only under specified ambient
conditions
Power Supply
Line Voltage 103.5 V 126.5 V or 207 V 253 V Line Frequency 50 Hz or 60 Hz Power Consumption approx. 30 VA
for 10 A test current approx. 95 VA, test duration max. 70 s for 25 A test current approx. 180 VA, test duration max. 70 s for function test continuous max. 3600 VA,
power is conducted through the instrument only, switching capacity  16 A
RS 232 Data Interface
Type RS 232C, serial, per DIN 19241 Format 9600, N, 8, 1 Connector 9-pin subminiature socket connector
Electrical Safety
Safety Class I per IEC 61010-1/EN 61 010-1/VDE 0411-1 Nominal Voltage 115/230 V Test Voltage 3.7 kV, 50 Hz Measuring Category 250 V CAT II (is not valid for the jacks 1, 2 and 3) Contamination Level 2 Safety Shutdown for residual current at
device under test > 25 mA, disconnecting time < 100 ms probe current > 10 mA, < 1 ms
Electromagnetic Compatibility
Product standard DIN EN 61326-1
Interference emission
EN 55011 B
Interference immunity
EN 61000-4-2 Contact/Atmos. – 4 kV/8 kV A EN 61000-4-3 3 V/m or 1 V/m A EN 61000-4-4 1 kV B EN 61000-4-5 1 kV bzw. 2 kV A EN 61000-4-6 3 V/m A
EN 61000-4-11
Test Value Evaluation Criteria
0.5/1/25 Periods A 250 Periods C
Class
62 GMC-I Messtechnik GmbH
Page 63
Mechanical Design
Display multiple dot matrix display, 128 x 128 pixels Dimensions test instruments without high-voltage module:
LxWxH: 292 mm x 138 mm x 243 mm test instruments with high-voltage module: LxWxH: 292 mm x 138 mm x 300 mm
Weight standard test instrument: approx. 4.5 kg
instrument with HV test: approx. 5.24 kg instrument with 25 A PE test: approx. 5.5 kg instr. with 25 A PE & HV test: approx. 5.9 kg
Protection housing: IP 40
terminals: IP 20 per DIN VDE 0470, part 1/EN 60529
Extract from table on the meaning of IP codes
IP XY
st
digit X)
(1
0 not protected 0 not protected 1 50.0 mm
2 12.5 mm 3 2.5 mm
4 1.0 mm
Protection against
foreign object entry
 
 
IP XY
(2nd digit Y)
1 vertically falling drops
2
3spraying water 4 splashing water
Protection against the
penetration of water
vertically falling drops with
enclosure tilted 15
High-Voltage Test (Prerequisite: Feature F02 or SECULIFE ST
HV)
Transducer
Nominal Voltage, AC
Open-Circuit Voltage, DC Intrinsic Error, Uo Uo 1.5%
Nominal Current per DIN VDE 0104 < 3.5 mA DC
Short-Circuit Current discharge current > 5 A at 6 kV
Resistance to Interference Voltage
UN~ adjustable in 10 V steps
Uo
in 100 V steps
0.5 0.99 kV 1 4.0 kV ((U
· 1.5) · 1.011) + 60 V
N~
none
Test Duration as long as START key is pressed (max. 60 sec.)
Measuring
Measuring Range Display Range Intrinsic Error, Uo 0 Uomax 0.000 > 10.00 kV DC 1.5% rdg. + 2 digits
Maximum Test Voltage
SC I* DUTs 1.5 kV SC II DUTs 4 kV
Service socket (20) – Terminal data
* devices with protective conductor terminal
(Prerequisite: Feature B01) Line Voltage 103.5 V ... 126.5 V or 207 V ... 253 V Line Frequency 50 Hz or 60 Hz Current output protected by building installation (16 A; fed through to the mains plug)
GMC-I Messtechnik GmbH 63
Page 64

20 RS 232 Interface

Note
1: External in + (for internal use only) 2: TXD (transmitter output) 3: RXD (receiver input) 4: External in + 5: GROUND 6: +5 V
(500 mA output, for barcode scanner only) 7: Ext. in – 8: Control output 9: +9 V (for (P)SI module only)
9
876
54
3
21
Jack RS232 is provided for the connection of the following instruments:
• (P)SI module (accessory), which can be inserted into the lid of the tester
•PC
• Barcode scanner of the following type: B3261 with RS232 interface (article number: GTZ3261000R0001) Z720A with RS232 interface (article number: Z720A)
or RFID scanner of the following type: Z751G with RS232 interface (article number: Z751G)

20.1 Transmission of Measurement Results to the (P)SI module

Test results – except for results from individual measurements and the function test – can be transmitted from the test instrument to the (P)SI module, where they can be stored and printed out in the form of measur­ing, test and statistics reports at any time.

20.2 PC Connection

Connection to an IBM compatible PC is also possible. The PC is connected to the interface at the test instrument, or to the interface port at a previously installed (P)SI module.

20.2.1Software Evaluation of Measurement Results

Convenient software programs such as PC.doc-WORD/EXCEL, PC.doc­ACCESS or PS3 allow for easy preparation of measuring and test reports, as well as archiving of measured data.

20.2.2Instrument Control via Interface Commands

All key functions included with the test instrument can be simulated with the help of interface protocols, and the following parameters can be queried:
• Type of measurement and measuring range
• Test setup
• Measurement sequence progress
• Detailed measuring results
64 GMC-I Messtechnik GmbH

20.3 Interface Definition and Protocol

The interface included with the test instrument is in compliance with the RS 232 standard.
Technical Data: Baud Rate 9600 baud, permanently set
Character Length 8 bits Parity none Stop Bits 1 Data Protocol per DIN 19244
X_ON / X_OFF protocol
Connector Pin Assignments, 9-Pin Subminiature Socket Connector:
A detailed description of the interface protocole is available from our Product Support, refer to chapter 24 for contact details.
Page 65

21 Appendix

21.1 Evaluation of Measured Values for Individual Measurements as well as for Calculated Quantities

In order to assure that the limit values for the individual measurements are always observed, device measuring error must be taken into consideration. The table in the appendix allows for calculation of the required minimum display value for each respective measurement which must appear at the instrument in consideration of measuring error (under nominal conditions of use), in order to assure that the required limit value is not fallen short of (DIN VDE 0413, part 1). Intermediate values can be interpolated.
Measuring Error for Test Sequences
The test instrument takes respective measuring error into consideration during automatic test sequences, and corrected results are entered into the test report, as long as this function has been activated in the setup menu under “include service error”.
Omitting the Protective Conductor Test in the Case of Fully Insulated Devices
You are testing a fully insulated safety class I device (e. g. screen, sub­mersible pump, etc.), which is not equipped with an external protective conductor contact.
The decision as to the necessity of a protective conductor test in this case is to be taken by a qualified electrician who should also assume responsibility.
You can omit the protective conductor test by pressing the key as soon as the following instruction is shown: „Please connect the probe with the protective conductor of the DUT“.
Tables for the calculation of minimum display values for insulation resistance and maximum display values for protective conductor resis­tance, equivalent leakage current, probe current and residual current in consideration of device measuring error:
R
M RPE
INS
Limit Value
0.100 0.115 0.100 0.085
0.250 0.273 0.200 0.180
0.500 0.535 0.300 0.275
1.000 1.060 0.400 0.370
2.000 2.200 0.500 0.465
Minimum
Display Value
Limit Value
Maximum
Display Value
5.000 5.350 0.600 0.560
7.000 7.450 0.700 0.655
10.00 10.60 or 12.5
20.00 23.00 0.900 0.845
75.00 83.50 1.000 0.940
1)
Depending upon resolution
mA I
I
EL
Limit Value
1.00 0.85 0.100 0.085 0.25 0.12
3.50 3.23 0.250 0.227 0.50 0.35
7.00 6.55 0.500 0.465 1.00 0.80
10.00 9.40 1.000 0.940 2.00 1.70
15.00 14.15 2.000 1.890 3.50 3.05
20.00 18.90 3.500 3.315 5.00 4.40
Maximum
Display Value
Limit Value
1)
probe
0.800 0.750
1.100 1.035
mA I mA
Maximum
Display Value
Limit Value
Maximum
Display Value
7.00 6.20
10.00 8.90
15.00 13.40
20.00 17.90
25.00 22.40

21.2 Evaluation of Measured Values during Equivalent Leakage Current Measurement (Automatic Test Sequence According to Standard)

During equivalent leakage current measurement, L and N are intercon­nected whereupon a test voltage of 230 V is applied between LN and PE and the leakage current is measured. Thus the most adverse case (N in­terrupted) is being tested.
As a rule, this results in at least double the value of the direct leakage cur­rent measurement (since all discharge capacitors are placed in parallel).
If frequency converters are used in addition to that, it is no longer possible to compare the measured values between direct leakage current and equivalent leakage current method. In this case, we recommend conduct­ing individual measurements in accordance with the residual current method.
GMC-I Messtechnik GmbH 65
Page 66

21.3 Index

A
Absence of Voltage ...................................... 7
Acst Sig, Meas ........................................... 18
Acst Sig, Seq ............................................. 18
Alternating / Direct Voltage UAC/DC ........... 30
Ambient Conditions .................................... 62
App prt ....................................................... 56
Auto (Test) Method ..................................... 39
Auto Class PSI ........................................... 18
Auto R-PE .................................................. 18
Automode .................................................. 18
Autostore ................................................... 39
C
Calibration .................................................. 68
Classification ........................................ 15, 39
Combined Testing ...................................... 42
Configuring Device Parameters ............. 14, 18
Configuring Measurement Parameters ........ 14
Connecting the Device Under Test ............. 17
Contact Current ...................................... 7, 17
Contact Problems ........................................ 2
Contrast Adjustment ................................... 13
cross section .............................................. 46
D
Data Security ................................................ 3
Device Leakage Current ............................... 7
Devices with Internal Power Supply ............ 15
Differential Current Measurement .................. 7
Direct Print-Out .......................................... 18
E
Earth Fault .................................................. 18
Earth Leakage Current ............................ 7, 26
Electrical Safety .......................................... 62
Electromagnetic Compatibility ..................... 62
Equivalent (Device) Leakage Current ............. 7
Equivalent Device Leakage Current .. 7, 22, 23
Equivalent Leakage Current .......................... 7
Equivalent Patient Leakage Current .. 7, 22, 23
Extension Cables ........................................ 46
F
Finger Contact ............................................ 12
Firmwareversion ........................................... 2
First Measured Values ................................ 39
Frequency Response .................................. 26
Function Test ............................................. 34
H
High-Voltage Test ............................. 7, 17, 24
Housing Leakage Current ....................... 7, 26
HV Test ...................................................... 39
HV Test Duration ........................................ 39
I
ILC ............................................................. 27
Illumination ................................................. 18
Incl. Service Error ....................................... 18
Influencing Quantities and Influence Error ... 62
Insulation Resistance .............................. 7, 20
Insulation Resistance Limit Values .............. 21
Interface ..................................................... 64
IT Network .................................................. 18
L
Limit Values...
(menu - selector switch Setup position) ...... 18
M
Mains Connection Errors ............................ 12
Mains Polarity Reversal ............................... 39
Mains Power Outlet .................................... 11
Mains Wait ........................................... 18, 39
Manual Sequence ...................................... 39
Measurements with Accessories ................ 32
Measuring Error .......................................... 65
Measuring Protective Conductor Resistance 19
Mechanical Design ..................................... 63
Multimeter Functions .................................. 30
N
No I-HL for SC I .......................................... 39
Nominal Ranges of Use .............................. 61
O
Online Help ................................................. 13
Options
EL1 adapter ....................................... 46
List of Possible Options and Standard
Types ................................................... 8
P
Patient Auxiliary Current .................... 7, 27, 39
Patient Leakage Current ......................... 7, 27
Periodic Testing ............................................ 6
Power Supply ............................................. 62
Probe Voltage Uprobe ................................ 30
GMC-I Messtechnik GmbH 66
Page 67
Protective Conductor Resistance .................. 7
R
Reference Ranges ...................................... 61
Reference Voltage ..........................18, 54, 56
Reports...
(menu - selector switch Setup position) ......18
Residual Current ..................................... 7, 27
Residual Current Method .............................. 7
Resistance R .............................................. 31
R-INS AWT-SL ........................................... 39
R-INS LN-SL .............................................. 39
R-PE AC > 10 A .........................................39
R-PE with clip ....................................... 38, 39
T
Test Conditions .......................................... 56
Test Current ................................................. 7
Test Sequence... (menu - selector switch Setup
position) ..................................................... 18
Test Time ................................................... 18
Testing after Repairs ..................................... 6
total patient leakage current ......................... 7
Type B Application Parts ............................ 15
Type BF Application Parts .......................... 15
Type CF Application Parts .......................... 15
V
Visual Inspection ......................................... 39
S
Safety Class I Devices ................................ 15
Safety Class II Devices ................................ 15
Safety Class III Devices ............................... 15
Safety Extra Low Voltage ...................... 15, 31
Saving Settings to Memory ......................... 14
SC I I I UV ................................................... 39
SECULOAD ................................................ 30
Select template .......................................... 18
Service socket
connection ........................................... 3
Terminal data ..................................... 63
Service...
(menu - selector switch Setup position) ......18
Setting Limit Values ....................................14
Setting of Time and Date ............................ 18
Short-Circuit Test ....................................... 35
Single Fault ................................................. 18
Single Fault Conditions ................................. 7
Socket Adapter .......................................... 39
Switching loads on and on ........................... 9
GMC-I Messtechnik GmbH 67
W
Welding equipment ..................................... 30
Z
Zero Balancing ..................................... 20, 33
Page 68

22 Maintenance - Recalibration

22.1 Housing Maintenance

No special maintenance is required for the housing. Keep outside surfaces clean. Use a slightly dampened cloth for cleaning. Avoid the use of cleansers, abrasives or solvents.

22.2 Recalibration

The respective measuring task and the stress to which your measuring in­strument is subjected affect the ageing of the components and may result in deviations from the guaranteed accuracy.
If high measuring accuracy is required and the instrument is frequently used in field applications, combined with transport stress and great tem­perature fluctuations, we recommend a relatively short calibration interval of 1 year. If your measuring instrument is mainly used in the laboratory and indoors without being exposed to any major climatic or mechanical stress, a calibration interval of 2-3 years is usually sufficient.
During recalibration* in an accredited calibration laboratory (DIN EN ISO/IEC 17025) the deviations of your instrument in relation to traceable standards are measured and documented. The deviations de­termined in the process are used for correction of the readings during subsequent application.
We are pleased to perform DKD or factory calibrations for you in our cali­bration laboratory. Please visit our website at www.gossenmetrawatt.com ( Services DKD Calibration Center or FAQs Calibration ques­tions and answers).
By having your measuring instrument calibrated regularly, you fulfill the re­quirements of a quality management system per DIN EN ISO 9001.
Standards DIN VDE 0701-0702 and IEC 63353 (VDE 0751) stipulate that only measuring instruments which are regularly tested and calibrated may be used for testing.

22.3 Safety Checks

Check the safety of your test instrument on a regular basis. We recom­mend the same test intervals as for recalibration.
The SECUTEST... is designed as a totally insulated instrument in accor­dance with standards IEC 61010 and VDE 0404. The protective conduc­tor is only used for measurement purposes and is therefore not accessible in idle state. A protective conductor test at the test socket can be per­formed as follows:
Ð Connect the SECUTEST... to a multiple distribution box.
Ð Conduct a contact current measurement for permanently installed
DUTs (the test socket may not be connected to any load).
Ð Measure the protective conductor resistance between the neighbour-
ing socket of the multiple distribution box and the test socket.
Ð The measured value may not exceed 0.3 .
Insulation resistance between LN and PE in the SECUTEST... equals ap­prox. 150 kfor metrological reasons. This must be taken into account when performing the safety checks, i.e. the protective conductor current measurement must produce a value below 3.5 mA, instead of insulation resistance measurement (if the equiv­alent leakage current measurement method has been chosen, the value must remain below 7 mA).
Apart from this, the SECUTEST... features 3 exposed conductive parts for which contact current measurement must produce a value below 0.5 mA:
• RS232 interface
• metallized start key
• protective conductor clip at the test socket.
* Verification of specifications or adjustment services are not part of the calibration.
For products from our factory, however, any necessary adjustment is frequently performed and the observance of the relevant specification is confirmed.
68 GMC-I Messtechnik GmbH
Page 69

22.4 Device Return and Environmentally Compatible Disposal

The test instrument is a category 9 product (monitoring and control instru­ment) in accordance with ElektroG (German Electrical and Electronic De­vice Law). This device is not subject to the RoHS directive.
We identify our electrical and electronic devices (as of August
2005) in accordance with WEEE 2002/96/EG and ElektroG with the symbol shown to the right per DIN EN 50419 . These devices may not be disposed with the trash.
Please contact our service department regarding the return of old de­vices, address see chapter 23.
23 Repair and Replacement Parts Service
Calibration Center* and Rental Instrument Service
If service is required please contact:
GMC-I Service GmbH
Service Center
Thomas-Mann-Strasse 20 90471 Nürnberg · Germany Phone +49 911 817718-0 Fax +49 911 817718-253 E-Mail service@gossenmetrawatt.com www.gmci-service.com
This address is only valid in Germany. Please contact our representatives or subsidiaries for service in other countries.
* Calibration Laboratory for Electrical Quantities
DKD – K – 19701 accredited per DIN EN ISO/IEC 17025:2005
Accredited measured quantities: direct voltage, direct current values, DC resistance, alternating voltage, alternating current values, AC active power, AC apparent power, DC power, capacitance, frequency and temperature
Competent Partner
GMC-I Messtechnik GmbH is certified in accordance with DIN EN ISO 9001:2008.
Our DKD calibration laboratory is accredited by the Deutscher Kalibrierdi­enst (German Calibration Service) in accordance with DIN EN ISO/ IEC 17025:2005 by under registration number DKD–K–19701.
We offer a complete range of expertise in the field of metrology: from test
reports and proprietary calibration certificates right on up to DKD calibration certificates. Our spectrum of offerings is rounded out with free test equipment manage­ment.
An on-site DKD calibration station is an integral part of our service depart­ment. If errors are discovered during calibration, our specialized personnel are capable of completing repairs using original replacement parts. As a full service calibration laboratory, we can calibrate instruments from other manufacturers as well.
GMC-I Messtechnik GmbH 69
Page 70

24 Product Support

If support is required please contact:
GMC-I Messtechnik GmbH
Product Support Hotline
Phone +49 911 8602-0 Fax +49 911 8602-709 E-Mail support@gossenmetrawatt.com
70 GMC-I Messtechnik GmbH
Page 71
GMC-I Messtechnik GmbH 71
Page 72
Edited in Germany Subject to change without notice A pdf version is available on the internet
Phone +49 911 8602-111 GMC-I Messtechnik GmbH Südwestpark 15 90449 Nürnberg
Germany
Fax +49 911 8602-777
info@seculife.eu
E-Mail
www.
seculife.eu
Loading...