The accessories available for your instrument are checked for
compliance with currently valid safety regulations at regular
intervals, and are expanded as required for new applications.
Currently up-to-date accessories which are suitable for your
measuring instrument are listed at the following web address
along with photo, order number, description and, depending
upon the scope of the respective accessory, data sheet and
operating instructions:
www.gossenmetrawatt.com
We calibrate and recalibrate all instruments supplied by GMC-I
Gossen-Metrawatt GmbH, as well as other manufacturers, at our
service center, for example after one year within the framework of
your test equipment monitoring program, as well as prior to use
etc.
Repair and Replacement Parts Service
DKD Calibration Laboratory* and Rental Instrument Service
If required please contact:
GMC-I Gossen-Metrawatt GmbH
Service Center
Thomas-Mann-Str. 20
90471 Nürnberg, Germany
Phone: +49 911 8602-0
Fax:+49 911 8602-253
e-mail: service@gossenmetrawatt.com
This address is only valid in Germany. Please contact our
representatives or subsidiaries for service in other countries.
* Calibration laboratory for measured electrical quantities,
DKD – K – 19701, accredited in accordance with
DIN EN ISO/IEC 17025:2005
Competent Partner
GMC-I Gossen-Metrawatt GmbH is certified in accordance with
DIN EN ISO 9001:2000.
Our calibration laboratory is accredited per DIN EN ISO/IEC
17025 by the Physikalisch-Technischen Bundesanstalt (German
Federal Institute of Physics and Metrology) and the Deutscher
Kalibrierdienst (German Calibration Service) under registration
number DKD–K–19701.
We offer a complete range of expertise in the field of metrology:
from test reports and factory calibration certificates, right on up to
DKD calibration certificates.
Our spectrum of offerings is rounded out with free test equipment
management.
As a full service calibration lab, we can calibrate instruments from
other manufacturers as well.
Accredited quantities: direct voltage, direct current value, direct current
resistance, alternating voltage, alternating current value, AC active
power, AC apparent power, DC power, capacitance, frequency,
temperature
4GMC-I Gossen-Metrawatt GmbH
Contact Persons
GMC-I Gossen-Metrawatt GmbH5
Table of Contents
ContentsPageContentsPage
1Safety Features and Precautions ....................................... 8
1.1Use for Intended Purpose ................................................................10
1.2Meanings of Danger Symbols ..........................................................10
1.3Meanings of Acoustic Warning Signals ............................................10
8 Technical Data ................................................................ 54
GMC-I Gossen-Metrawatt GmbH7
Safety Precautions
1Safety Features and Precautions
You have selected an instrument which provides you with a high
level of safety.
This instrument fulfills the requirements of applicable European
and national EC directives. This is confirmed by means of the
CE mark. A corresponding declaration of conformity can be
requested from GMC-I Gossen-Metrawatt GmbH.
The TRMS digital multimeter has been manufactured and tested
in accordance with the following safety regulations:
IEC 61010–1:2001 / DIN EN 61010–1/VDE 0411–1:2002.
When used for its intended purpose (see page 10), safety of the
operator, as well as that of the instrument, is assured. Their safety
is however not guaranteed, if the instrument is used improperly or
handled carelessly.
In order to maintain flawless technical safety conditions, and to assure
safe use, it is imperative that you read the operating instructions
thoroughly and carefully before placing your instrument into service,
and that you follow all instructions contained therein.
The multimeter is equipped with an automatic socket blocking
mechanism for your safety, and in order to safeguard your
instrument. This mechanism is linked to the rotary switch and only
allows access to those jacks which are actually required for the
selected function. It also prevents the user from turning the rotary
switch to impermissible functions after the measurement cables
have already been plugged in.
Measuring Categories and their Significance per IEC 61010-1
CATDefinition
Measurements in electrical circuits
I
which are not directly connected to the mains:
for example electrical systems in motor vehicles and aircraft, batteries etc.
Measurements in electrical circuits
II
which are electrically connected to the low-voltage mains:
with plugs, e.g. at home, in the office or laboratory etc.
Measurements in building installations:
III
stationary power consumers, distributor terminals, devices connected
permanently to the distributor
Measurements at power sources for low-voltage installations:
The measuring category and the maximum rated voltage which
are printed on the device apply to your measuring instrument, for
example 1000 V CAT III.
Observe the following safety precautions:
• The multimeter may not be used in potentially explosive
atmospheres.
• The multimeter may only be operated by persons who are ca-
pable of recognizing contact hazards and taking the appropriate
safety precautions. Contact hazards according to standard
exist anywhere, where voltages of greater than 33 V RMS or
70 V DC may occur. Avoid working alone when taking measurements which involve contact hazards. Be certain that a second person is present.
• Maximum allowable voltage between the voltage measuring
terminals, and between all connections and earth is 1000 V for
measuring category III, or 600 V for measuring category I V.
8GMC-I Gossen-Metrawatt GmbH
Safety Precautions
• Be prepared for the occurrence of unexpected voltages at
devices under test (e.g. defective devices). For example,
capacitors may be dangerously charged.
• Make certain that the measurement cables are in flawless
condition, e.g. no damage to insulation, no interruptions in
cables or plugs etc.
• No measurements may be made with this instrument in
electrical circuits with corona discharge (high-voltage).
• Special care is required when measurements are made in HF
electrical circuits. Dangerous pulsating voltages may be
present.
• Measurements under moist ambient conditions are not
permitted.
• Be absolutely certain that the measuring ranges are not
overloaded beyond their allowable capacities. Limit values are
included in chapter 8, “Technical Data”, in the table entitled
“Measuring Functions and Measuring Ranges” in the “Overload Capacity” column.
• The multimeter may only be operated with installed batteries or
rechargeable batteries. Dangerous currents and voltages are
otherwise not indicated, and the instrument may be damaged.
• The instrument may not be operated if the fuse cover or the
battery compartment lid has been removed, or if its housing is
open.
• The input for the current measuring range is equipped with a
fuse link (exception: METRAHIT⏐B
ASE).
Maximum permissible voltage for the measuring circuit (= rated
voltage of the fuse) is 1000 V AC/DC.
Use specified fuses only (see page 59)! The fuse must have a
breaking capacity of at least 30 kA.
Repair and Parts Replacement
When the instrument is opened, voltage conducting parts may be
exposed. The instrument must be disconnected from the measuring circuit before the performance of repairs or the replacement of
parts. If repair of a live open instrument is required, it may only be
carried out by trained personnel who are familiar with the dangers
involved.
Defects and Extraordinary Strains
If it may be assumed that the instrument can no longer be
operated safely, it must be removed from service and secured
against unintentional use.
Safe operation can no longer be relied upon:
• If the device demonstrates visible damage
• If the instrument no longer functions, or if malfunctioning
occurs
• After long periods of storage under unfavorable conditions,
e.g. humidity, dust or extreme temperature (see “Ambient Conditions” on page 60)
GMC-I Gossen-Metrawatt GmbH9
Safety Precautions
1.1 Use for Intended Purpose
• The respective multimeter is a portable device which can be
held in the hand during the performance of measurements.
• Only those types of measurements described in chapter 5 may
be performed with the measuring instrument.
• The measuring instrument, including measurement cables and
plug-on test probes, may only be utilized within the specified
measuring category (see page 59 and the table on page 8
regarding significance).
• Overload limits may not be exceeded. See technical data on
page 54 for overload values and overload limits.
• Measurements may only be performed under the specified
ambient conditions. See page 60 regarding operating temperature range and relative humidity.
• The measuring instrument may only be used in accordance
with the specified degree of protection (IP code) (see page 60).
1.2 Meanings of Danger Symbols
!
1.3 Meanings of Acoustic Warning Signals
Warning concerning a point of danger
(attention: observe documentation!)
Warning concerning dangerous voltage at the measurement
input: U > 55 V AC or U > 70 V DC
Voltage warning: > 1000 V (intermittent acoustic signal)
Current warning: > 10 A (intermittent acoustic signal)
Current warning: > 16 A (continuous acoustic signal)
* METRAHIT⏐BASE: Current measurement by means of clip-on current sensor with
voltage output only
→
chapter 4.1.2
2
→ chapter 3
3
→ chapter 5
4
→ chapter 6
5
→
6
→
chapter 1.2
7
→ chapter 3
8
→
chapter 4.1
9
*
chapter 3
1 Display (LCD) (see page 13 for significance of symbols)
2 MAN / AUTO shift key for manual/automatic measuring range selection
Increase parameter values
Operating mode menu:
3
ON / OFF | LIGHT
key for switching device and display illumination on and off
4 FUNC | ENTER multifunction key
Operating mode menu: Acknowledge entry (ENTER)
Increase measuring range or move decimal point to the right (MAN function)
5
6
Rotary switch
for measuring functions, (see page 15 for significance of symbols)
7 DKD calibration mark
8 Connector socket for ground
9 Connector socket for current measurement with automatic blocking
10 Connector socket for voltage, resistance, temperature, diode
capacitance measurement (
METRAHIT⏐TECH
11 DATA / MIN / MAX
ff
Key for freezing, comparing and deleting measured value, and for Min/Max
Decrease values
Operating mode menu: Select individual menu entries in direction of flow
12 MEASURE | SETUP
Key for switching back and forth between measuring and menu function
13 ZERO | ESC
Key for zero balancing
Operating mode menu:Exit current menu level and
14 Decrease measuring range or move decimal point to the left (MAN function)
15 Connector for power pack (METRAHIT⏐X-TRA⏐OUTDOOR only)
16 Infrared interface (METRAHIT⏐X-TRA⏐OUTDOOR only)
Select individual menu entries against direction of flow
METRA
HIT
⏐X-TRA⏐OUTDOOR
and
only) with automatic blocking
return to a higher level,
exit parameters configuration
without saving data
and
12GMC-I Gossen-Metrawatt GmbH
Symbols Used in the Digital Display
1 2910
3
4
24
23
22
2017
Battery level indicator
Battery full
Battery OK
Battery weak
Battery (almost) dead, U < 1.8 V
Interface indicator
METRAHIT⏐X-TRA⏐OUTDOOR:
Active data transmission ↓ to / ↑ from multimeter
IR interface in stand-by mode
(ready to receive starting commands)
2 ON: continuous operation (automatic shutdown deactivated)
3 MAN: manual measuring range selection is active
4 Digital display with decimal point and polarity display
5 max/min: Min/Max value storage
6 DATA: display memory, “freeze” measured value
7 STORE: memory mode active, with METRAHIT⏐X-
TRA⏐OUTDOOR only
8 ISO: no function here
9 Transformation factor (current clip factor),
with METRAHIT⏐B
ASE and METRAHIT⏐TECH only
10 IR: infrared interface display, with METRAHIT⏐X-TRA⏐OUTDOOR only
11 Scale for analog display
12 Pointer for analog display (bar graph – pointer)
depending upon setting in SET menu for the A.diSP parameter
Triangle appears: indicates overranging
13 Diode measurement selected
14 Selected type of current
15 TRMS measurement
16 Pt100(0): selected platinum resistance thermometer with
automatic recognition of Pt100/Pt1000,
with METRAHIT⏐X-
TRA⏐OUTDOOR only
17 Type K: temperature measurement with type K (NiCr-Ni) thermocouple
18 sec (seconds): unit of time
19 ΔREL: relative measurement with reference to offset
20 Unit of measure
21 ZERO: zero balancing active
22 Warning regarding dangerous voltage: U > 55 V AC or U > 70 V DC
23 Continuity test with acoustic signal is active
24 h (hours): unit of time
V~0/4 V~ AC TRMSAlternating voltage, TRMS AC, full bandwidth••••
Hz (V)1Hz ~ ACVoltage frequency, full bandwidth••••
V~ 2V Fil ~ AC TRMSAlternating voltage, TRMS AC, with low-pass (1 kHz)•••—
1kHz
Hz (V) 3Hz Fil ~ ACVoltage frequency, with low-pass (1 kHz)•••—
V~ 1 M
V~ 1V Fil ~ AC TRMSAlternating voltage, TRMS AC, up to 1 kHz, input 1 MΩ•••—
1kHz
Hz (V) 2Hz Fil ~ ACVoltage frequency, up to 1 kHz, input 1 MΩ•••—
Hz (V) 1 M
V 0/2 V DCDirect voltage••••
V1V DC AC TRMSPulsating voltage, TRMS ()••••
MHz0/2 MHz(High) frequency @ 5 V~•———
%1%Keying ratio @ 5 V~•———
Ω
Tem p TC0/2
Tem p RTD1°C Pt100/1000Temperature with resistance thermometer Pt100/Pt1000•———
A0/2 A DCDirect current value•••—
A1A DC AC TRMSPulsating current value, TRMS AC DC•••—
A~0/2 A~ AC TRMSAlternating current value, TRMS AC•••—
Hz (A)1Hz ~ ACCurrent frequency•••—
Hz ()1Hz ~ AC Current frequency—•—•
FUNC
DisplayMeasuring Function
1kHz
METRA
X-
TRA/
OUTDOOR
HIT
⏐
METRAHIT⏐
T
ECH
METRAHIT⏐
P
RO
METRAHIT⏐
B
ASE
Ω0/4 V~ AC TRMSAlternating voltage, TRMS AC, full bandwidth, input 1 MΩ•••—
1kHz
Ω 3Hz ~ ACVoltage frequency, full bandwidth, input 1 MΩ•••—
2
2
+=
V
ACDCVAC
— Ω(Direct current) resistance••••
0/2ΩContinuity testing Ω with acoustic signal••••
1
V DCDiode voltage••••
°C Typ-K
Temperature thermocouple Type K••••
— nFCapacitance••——
0/2 A DC Direct current value with AC DC clip-on current sensor 1 V:1/10/100/
A
1A DC AC TRMS
A
0/2 A~ AC TRMS
A
1000 A
Pulsating current value, TRMS, with AC DC clip-on current sensor, see above
Alternating current strength, TRMS, with clip-on current sensor, see above
Warning concerning a point of danger
(attention: observe documentation!)
Ground
CAT III / IV Measuring category III or IV device, see also
“Measuring Categories and their Significance per IEC
61010-1” on page 8
Continuous, doubled or reinforced insulation
The approval mark issued by the VDE test authority
applies to the following multimeters:
METRAHIT⏐X-
TRA⏐TECH⏐PRO⏐BASE
▲IR▼Position of the infrared interface, window on the top of
the instrument (METRAHIT⏐X-TRA⏐OUTDOOR only)
Position of the power pack connector socket,
see also chapter 3.1
5 V / 600 mA
(METRAHIT⏐X-TRA⏐OUTDOOR only)
Fuse for current measuring ranges, see chapter 9.3
(not METRAHIT⏐B
ASE)
The device may not be disposed of with the trash.
Further information regarding the WEEE mark can be
accessed on the Internet at www.gossenmetrawatt.com under the search term WEEE (see also
chapter 9.5).
Calibration seal (red seal):
B0730
DKD-K19701
01-04
Consecutive number
German Calibration Service – Calibration Laboratory
Registration number
Date of calibration (year – month)
see also “Recalibration Service” on page 64
Indicates EC conformity
GMC-I Gossen-Metrawatt GmbH15
Initial Start-Up – Setup
3Initial Start-Up
3.1 Inserting Batteries or Rechargeable Batteries
Be certain to refer to chapter 9.2 regarding correct battery
installation.
Momentary battery voltage can be queried in the Info menu (see
chapter 6.3).
!
Attention!
Disconnect the instrument from the measuring circuit before
opening the battery compartment lid in order to replace the
batteries.
Operation with the Power Pack
(accessory for METRAHIT⏐X-
Installed batteries are disconnected electronically if the NA X-TRA
power pack is used, and need not be removed from the
instrument.
If rechargeable batteries are used, they must be recharged
externally.
If the external power supply is switched off, the device is
automatically switched to battery operation without interruption.
3.2 Activation
Switching the Instrument On Manually
➭ Press the ON / OFF | LIGHT key until the display appears.
Power-up is acknowledged with a brief acoustic signal. As
long as the key is held depressed, all of the segments at the
liquid crystal display (LCD) are illuminated.
The LCD is depicted on page 13.
The instrument is ready for use as soon as the key is released.
TRA, not included, see chapter 10.3)
Display Illumination
After the instrument has been switched on, background illumination can be activated by briefly pressing the ON / OFF | LIGHT key. Illumination is switched back off by once again pressing the same
key, or automatically after approximately 1 minute.
Switching the Instrument On via PC (METRAHIT⏐X-
The multimeter is switched on after transmission of a data block
from the PC, assuming the “irStb ” has been set to “ir on ”
(see chapter 6.4).
However, we recommend using the power saving mode: “ir off ”.
Note
Electrical discharge and high frequency interference may
cause incorrect displays to appear, and may disable the
measuring sequence.
Disconnect the device from the measuring circuit. Switch the
instrument off and back on again in order to reset. If the
problem persists, briefly dislodge the battery from the
connector contacts (see also chapter 9.2).
3.3 Setting the Operating Parameters
Setting Time and Date
See the “t iME ” and “dAtE ” parameter in chapter 6.4.
Display Modes for the Analog Display
Selection can be made from two different display modes
(see “A.diSP ” parameter in chapter 6.4).
Display Modes for the Digital Display
Selection can be made from two different display modes
(see “0.diSP ” parameter in chapter 6.4).
TRA⏐OUTDOOR only)
16GMC-I Gossen-Metrawatt GmbH
3.4 Switching the Instrument Off
Switching the Instrument Off Manually
➭ Press the ON / OFF | LIGHT key until 0FF appears at the display.
Shutdown is acknowledged with a brief acoustic signal.
Automatic Shutdown
The instrument is switched off automatically if the measured value
remains unchanged for a long period of time (maximum measured
value fluctuation of approx. 0.8% of the measuring range per
minute, or 1 °C or 1 ° F per minute), and if none of the keys or the
rotary switch have been activated before a selected period of time
in minutes has elapsed (see “APoFF ” parameter on page 49.)
Shutdown is acknowledged with a brief acoustic signal.
Exceptions include:
Transmission and memory mode operation, continuous operation
and whenever a dangerous voltage is applied to the input (U > 55
V AC or U > 70 V DC).
Disabling Automatic Shutdown
The instrument can be set to continuous operation.
➭ Simultaneously press the and keys to
this end.
The “Continuous On” function is indicated at by means of the
ON display to the right of the battery symbol.
The “Continuous On” setting can only be cancelled by changing the
respective parameter, and not by switching the instrument off
(see “APoFF ” on page 49).
ON / OFF
LIGHT
FUNC
ENTER
Initial Start-Up – Setup
ON / OFF
LIGHT
ON/ OFF
LIGHT
ON/ OFF
LIGHT
ON/ OFF
LIGHT
Long
(1 s)
8.8.8.8.8
Auto-Range
000.00
Illumination On
000.00
Illumination Off
0ff
GMC-I Gossen-Metrawatt GmbH17
Control Functions
4Control Functions
4.1 Selecting Measuring Functions and Measuring Ranges
The rotary switch is linked to the automatic socket blocking
mechanism, which only allows access to two connector jacks for
each function. Be certain to remove the appropriate plug from its
respective jack before switching to and from the “A” functions.
The socket blocking mechanism prevents the user from inadvertently turning the selector switch to impermissible functions after
the measurement cables have been plugged in to the instrument.
4.1.1 Automatic Range Selection
The multimeter is equipped with auto-ranging for all measuring
functions, except for temperature measurement, diode and
continuity testing, and the MHz measuring function. Auto-ranging
is active as soon as the instrument is switched on. The instrument
automatically selects the measuring range which allows for
highest possible resolution of the applied quantity. When the
instrument is switched to frequency measurement, the previously
selected voltage measuring range remains active.
The Auto-Ranging Function
The multimeter is switched automatically to the next higher range
at ±(11999 d + 1 d → 01200 d), and to the next lower range at
±(01100 d - 1 d →10990 d).
Exception, capacitance measurement:
The multimeter is switched automatically to the next higher range
at ±(1199 d + 1 d → 0120 d), and to the next lower range at
±(0110 d - 1 d →1099 d).
4.1.2 Manual Measuring Range Selection
Auto-ranging can be deactivated and measuring ranges can be
selected manually in accordance with the following table by
pressing the MAN / AUTO button.
The desired measuring range can then be selected with the
scroll key.
or
The instrument is automatically returned to range selection when
the MAN / AUTO key is pressed, the rotary switch is activated or the
instrument is switched off and back on again.
Overview: Auto-Ranging and Manual Range Selection
FunctionDisplay
MAN / AUTO
Range switching sequence for:
V:100 mV*
Hz: 100 Hz
Ω: 100 Ω
10 MΩ ↔ 40 MΩ
A:METRAHIT⏐X-
↔
or
MAN / AUTO
* Via manual measuring range selection only
100 mA ↔ 1 A ↔ 10 A (16 A)
METRAHIT⏐T
10 mA
METRAHIT⏐P
A:METRAHIT⏐B
see chapter 5.7.3 ff
F:METRAHIT⏐X-
10 nF ↔ 100 nF ↔ 1 μF ↔ 10 μF ↔ 100 μF ↔ 1000 μF
Return to automatic measuring range selection—
Manual mode active:
utilized measuring range is fixed
↔
1 V↔ 10 V↔ 100 V↔ 1000 V
↔
1 kHz↔ 10 kHz↔ 100 kHz
↔
1 kΩ ↔ 10 kΩ ↔ 100 kΩ ↔1 MΩ
TRA⏐OUTDOOR:
ECH:
↔
100 mA↔ 1A↔ 10 A (16 A)
RO: 1A
↔
10 A (16 A)
ASE and METRAHIT⏐TECH:
TRA⏐OUTDOOR and METRAHIT⏐TECH:
↔
100 μA ↔ 1 mA ↔ 10 mA
MAN
MAN
18GMC-I Gossen-Metrawatt GmbH
Control Functions
4.1.3 Quick Measurements
Measurements performed using a suitable fixed measuring range
are executed more quickly than those which utilize automatic
range selection. Quick measurement is made possible with the
following two functions:
• Manual measuring range selection, i.e. selection of the measuring
range with the best resolution (see chapter 4.1.2).
or
•With the DATA function (see chapter 4.4). In this way, the
appropriate measuring range is selected automatically after the
first measurement and the second measurement is executed
more quickly.
The selected measuring range remains active for the subsequent
series of measurements with these two functions.
4.2 Zero Offset / Relative Measurements
Depending upon deviation from the zero point, zero balancing or
a reference value for relative measurements can be stored to
memory:
Deviation from zero
– With short-circuited measurement cables for V, Ω, A
– With open input for capacitance unit of measure F
0 to 200 digitsZERO ΔREL
> 200 to 5000 digitsΔREL
The applicable reference or correction value is deducted
individually for the respective measuring function as an offset from
all subsequent measurements, and remains in memory until
deleted, or until the multimeter is switched off.
Display
Zero balancing and reference value adjustment can be used for
auto-ranging, as well as for manual measuring range selection.
Zero Balancing
➭ Plug the measurement cables into the instrument and connect
the free ends to each other, except for capacitance
measurement in which case the ends of the cables are not
connected to each other.
➭ Briefly press the ZERO | ESC key.
The instrument acknowledges zero balancing with an acoustic
signal, and the “ZERO ΔREL” symbol appears at the LCD. The
value measured at the moment the key is pressed serves as a
reference value.
➭ Zero balancing can be cleared by once again pressing the
ZERO | ESC key.
Note
Setting the Reference Value
➭ Plug the measuring cables into the instrument and measure a
As a result of TRMS measurement, the multimeter displays a residual value of 1 to 30 digits with short-circuited
measurement cables as the zero point for V AC / I AC or
V(AC+DC) / I (AC+DC) measurements (non-linearity of
the TRMS converter). This has no influence on specified
accuracy above 2% of the measuring range (or 3% in the
mV range).
reference value (max. 5000 digits).
GMC-I Gossen-Metrawatt GmbH19
Control Functions
➭ Briefly press the ZERO | ESC key.
The instrument acknowledges storage of the reference value
with an acoustic signal, and the “ZERO ΔREL” or the “ΔREL”
symbol appears at the LCD. The value measured at the
moment the key is pressed serves as a reference value.
➭ The reference value can be cleared by once again pressing the
ZERO | ESC key.
Notes Regarding Relative Measurement
• Relative measurement effects the digital display only. The
analog display continues to read out the original measured
value.
• In the case of relative measurement, Ω / F or AC quantities
may also appear as negative values.
4.3 Display (LCD)
4.3.1 Digital Display
Measured Value, Unit of Measure, Type of Current, Polarity
4.3.2 Analog Display
Measured Value, Polarity
The analog display demonstrates the dynamic performance of a
moving-coil mechanism. This display is especially advantageous
for observing measured value fluctuation, and for balancing
procedures.
Two different display modes can be selected in the “SET ” menu
with the help of the “A.diSP ” parameter (see chapter 6.4):
• Bar graph
• Pointer: The current measured value is tracked in real-time.
The analog scale displays a negative range of 2 scale divisions for
the measurement of zero-frequency quantities, allowing for
precise observation of measured value fluctuation around zero. If
the measured value exceeds the negative range of 2 scale
divisions, polarity is reversed at the analog display.
Scaling of the analog scale is automatic. This is very helpful for
manual measuring range selection.
The measured value with decimal and plus or minus sign appears
at the digital display. The selected unit of measure and type of
current are displayed as well. A minus sign appears to the left of
the value during measurement of zero-frequency quantities, if the
plus pole of the measured quantity is applied to the “⊥” input.
Overranging
Overranging in the positive range is displayed by means of the
right triangle symbol.
The “0.diSP ” parameter can be used to determine whether lead-
ing zeros will be appear or be suppressed at the measured value
display (see chapter 6.4).
Overranging
Refresh Rate
In the bar graph and pointer modes, the analog display is
refreshed 40 times per second.
If the upper range limit of 12,000 digits is exceeded “0L” (overload)
appears at the display.
Exceptions: “0L” appears as of 1200 digits for capacitance
measurement, and as of 5100 digits for diode testing.
20GMC-I Gossen-Metrawatt GmbH
Control Functions
4.4 Measured Value Storage: DATA (auto-hold / compare)
An individual measured value can be automatically “frozen” with
the DATA function (auto-hold). This is useful, for example, when
contacting the measuring points with the test probes requires
your full attention. After the measuring signal has been applied
and the measured value has settled in in accordance with the
“condition” listed in the table below, the measured value is frozen
at the digital display and an acoustic signal is generated. The test
probes can now be removed from the measuring points, and the
measured value can be read from the digital display. If the
measuring signal falls below the value specified in the table, the
function is reactivated for storage of the next value.
Measured Value Comparison (DATA Compare)
If the currently frozen value deviates from the first saved value by
less than 100 digits, the acoustic signal is generated twice. If
deviation is greater than 100 digits, only a brief acoustic signal is
generated.
Note
The DATA function has no effect on the analog display, at
which the current measured value continues to appear.
However, when the digital display if “frozen”, the decimal
point is fixed as well (fixed measuring range, symbol:
MAN).
The selected measuring range should not be manually
changed as long as the DATA function is active.
The DATA function is deactivated by pressing and holding the
DATA/MIN/MAX key (for approx. 1 second), when the measuring
function is changed or when the instrument is switched off and
back on again.
V, A, Hz
F, MHz, %
of measuring range
100%
10%
Activated
DATA
Function
ActivateBriefblinksOnce
Save
(stabilized
measured
value)
Reactivate
Change to
Min/Max
ExitLong
1)
Reactivation results from falling short of specified measured value limits.
2)
Two acoustic signals are generated the first time a measured value is
saved as a reference value. For subsequent data hold, two acoustic
signals are only generated if the currently frozen value deviates from the
first saved value by less than 100 digits.
Key: MV = measured value, R = measuring range
Save
Press
DATA /
Min/Max
1)
BriefSee table in chapter 4.4.1
Reactivated
Measuring
Function
V, A, F, Hz,
MHz, %
Ω
V, A, F, Hz,
MHz, %
Ω
10,000 digits
t [s]
Save
ConditionResponse from Instrument
Display
MV
MV
DATA
Static
Blinks
Digital
Is displayed
Stored
Is cleared Is cleared
Measuring
Signal
> 10%
of R
≠ 0L
< 10%
of R
= 0L
Acoustic
Once
Twi ce
Twi ce
2)
GMC-I Gossen-Metrawatt GmbH21
Control Functions
Example
The voltage measuring range is set manually to 10 V.
The first measured value is 5 V, which is stored to memory
because it is greater than 10% of the measuring range (= 1 V),
and is thus reliably above the background noise level. As soon as
the measured values drops to less than 10% of the measuring
range, i.e. amounts to less than 1 V which corresponds to
removal of the test probes from the measuring point, the
instrument is ready to store a new value.
4.4.1 Saving Minimum and Maximum Values – MIN/MAX Function
Minimum and maximum measured values applied to the
measuring instrument’s input after the Min/Max function has been
activated can be “frozen” at the display. The most important use
of this function is the determination of minimum and maximum
values during long-term measured value observation.
The Min/Max function can be activated in all measuring functions.
The Min/Max function has no effect on the analog display, at
which the momentary measured value continues to appear.
Apply the measured quantity to the instrument and set the
measuring range with the MAN / AUTO key before activating the
Min/Max function.
The Min/Max function is deactivated by pressing and holding the
DATA/MIN/MAX key (for approx. 1 second), when the measuring
function is changed or when the instrument is switched off and
back on again.
Min/Max
Function
1
Activate
and save
2
Save and
display
3
Return to 1
StopLongAre deleted
Press
DATA /
Min/Max
2 x briefare saved
Brief
Brief
Brief
Min. and Max.
Measured Values
Storage continues in
background,
new min. and max.
values are displayed.
Same as 1,
stored values are not
deleted
Response from Instrument
Display
Measured Value
Digital
Momentary
measured
value
Saved min.
value
Saved max.
value
Same as 1
Momentary
measured
value
Acoustic
Max.
Signal
Min.
Max
and
min.
Min.1 x
Max.1 x
Same
as 1
de-
Is
leted
2 x
1 x
2 x
Note
As opposed to the DATA function, the Min/Max function
can also be used for temperature measurement.
22GMC-I Gossen-Metrawatt GmbH
Control Functions
4.5 Measurement Data Recording (METRAHIT⏐X-TRA⏐OUTDOOR only)
The METRAHIT⏐X-
TRA⏐OUTDOOR is capable of recording measure-
ment data using an adjustable sampling rate for long periods of
time in the form of measurement series. Data are stored to a battery backed memory module, and are retained even after the multimeter is switched off. The system acquires measured values relative to real-time.
Stored measured values can subsequently be read out with the
help of METRAwin 10 software. The only prerequisite is a PC which
is connected by means of an interface cable to the
USB X-TRA bi-directional interface adapter, which is in turn
plugged onto the METRAHIT⏐X-
24: Clear Memory
24: Clear Memory – appears after CLEAr
24: Querying Memory Occupancy
48: rAtE – set the sampling rate (METRAHIT⏐X-TRA⏐OUTDOOR only)
23: Starting Recording via Menu Functions
24: Ending Recording
The STORE Menu Function
➭ First set the sampling rate for memory mode operation (see rAtE
parameter in chapter 6.4) , and then start memory mode operation.
➭ First select the desired measuring function and an appropriate
measuring range.
➭ Check the battery charge level before starting long-term
measurement recordings (see chapter 6.3).
Connect the NA X-TRA power pack if required.
Starting Recording via Menu Functions
➭ Switch to the “SET ” mode by pressing MEASURE | SETUP and se-
lect the “StorE ” menu.
MEASURE
SETUP
1nFo
... StorE 000.0% StArt
FUNC
ENTER
FUNC
ENTER
➭ Memory mode operation is started by pressing FUNC | ENTER.
STORE appears underneath the analog display and indicates
that memory mode operation has been activated.
“StoP ” appears at the digital display.
➭ Press MEASURE | SETUP in order to return to the measuring
function.
GMC-I Gossen-Metrawatt GmbH23
Control Functions
During Recording
STORE is displayed underneath the analog display during memory
mode operation, and
000.3%
StoP
memory occupancy
can be controlled:
The following message appears as soon as memory is full:
“100.0%”.
In order to be able to observe measured values during recording,
switch to the measuring function by pressing MEASURE | SETUP.
The display is returned to the memory menu after once again
pressing MEASURE | SETUP.
A new memory block is created when another measuring function
is selected with the rotary switch or the FUNC | ENTER key. Data
storage then continues automatically.
Ending Recording
➭ “StoP ” appears at the display after pressing MEASURE | SETUP.
StoP start
FUNC
ENTER
➭ Acknowledge the “StoP ” display by pressing FUNC | ENTER.
STORE is cleared from the display, indicating that recording has
been ended.
➭ Press MEASURE | SETUP in order to return to the measuring
function.
➭ Memory mode operation can also be exited by switching the
multimeter off.
Querying Memory Occupancy
Memory occupancy can be queried during recording with the
help of the “1nFo ” menu (see also chapter 6.3).
Memory occupancy range: 000.1% to 099.9%.
MEASURE
SETUP
1nFo
FUNC
ENTER
bAtt:
...
0CCvP
%:
017.4
%
Memory occupancy can be queried before recording is started via
the “StorE” menu.
MEASURE
SETUP
... StorE017.4 % StArt
1nFo
FUNC
ENTER
Clear Memory
This function deletes all measured values from memory!
This function cannot be executed during memory mode
operation.
MEASURE
SETUP
CLEArempty
... StorE 017.4 % w StArt
1nFo
FUNC
ENTER
FUNC
ENTER
24GMC-I Gossen-Metrawatt GmbH
Control Functions
GMC-I Gossen-Metrawatt GmbH25
V/Hz, Ω, Temperature, and A/Hz Measurements
5Measurements
5.1 Voltage Measurement
Notes Regarding Voltage Measurement
• The multimeter may only be operated with installed batteries or
rechargeable batteries. Dangerous voltages are otherwise not
indicated, and the instrument may be damaged.
• The multimeter may only be operated by persons who are
capable of recognizing contact hazards and taking the appropriate safety precautions. Contact hazards exist anywhere, where
voltages of greater than 33 V RMS may occur.
The test probes may only be only gripped up to the finger
Do not touch the metallic test probes under any
guard.
circumstances.
• Avoid working alone when taking measurements which involve
contact hazards. Be certain that a second person is present.
• Maximum permissible voltage between the connector sockets,
(9 and 10) and ground (8) is 1000 V for measuring category II I,
and 600 V for measuring category IV.
• Be prepared for the occurrence of unexpected voltages at
devices under test (e.g. defective devices). For example,
capacitors may be dangerously charged.
• No measurements may be made with this instrument in
electrical circuits with corona discharge (high-voltage).
• Special care is required when measurements are made in HF
electrical circuits. Dangerous pulsating voltages may be
present.
• Be aware of the fact that dangerous voltage spikes are not displayed
during measurement with the low-pass filter.
We recommend measuring voltage without the low-pass filter first, in
order to be able to detect any dangerous voltages.
• Be absolutely certain that the measuring ranges are not
overloaded beyond their allowable capacities. Limit values are
included in chapter 8, “Technical Data”, in the table entitled
“Measuring Functions and Measuring Ranges” in the “Overload Capacity” column.
Scope of Functions, Voltage Measurement
Function
V AC / Hz TRMS
(Ri ≥ 9 MΩ)
V AC / LP filter 1 kHz
(Ri = 1 MΩ2)) TRMS
V AC+DC TRMS
(Ri ≥ 9 MΩ)
V DC (Ri ≥ 9 MΩ)
MHz at 5 V AC
Keying ratio as %
Freq. response, V AC
1)
A 1 kHz low-pass filter can be used in this case, in order to filter out high
frequency pulses of greater than 1 kHz, for example when performing
measurements at pulsed motor drives.
2)
Input resistance of approx. 1 MΩ. Erroneous displays resulting from
capacitive coupling during voltage measurement in power supply
systems are reduced to a minimum in this way.
METRA
HIT
⏐
X-TRA/
OUTDOOR
1)
20 kHz10 kHz10 kHz1 kHz
METRAHIT⏐
TECH
••••
•••—
••••
••••
•———
•———
METRAHIT⏐
PRO
METRA
BASE
HIT
⏐
26GMC-I Gossen-Metrawatt GmbH
5.1.1 Direct and Pulsating Voltage Measurement, V DC and V (DC+AC)
Note
Only to observe for METRAHIT⏐TECH: Set the CL iP parameter
to 0FF in the current clip setup menu. Otherwise all measured values are displayed in amperes, and are corrected by
the amount resulting from the selected transformation ratio
for an interconnected clip-on current sensor.
MEASURE
SETUP
FUNC
ENTER
➭ In accordance with the voltage to be measured, turn the rotary
switch to V or V.
➭ Connect the measurement cables as shown.
The “⊥” connector jack should be grounded.
Note
An intermittent acoustic signal warns the operator if the measured value exceeds the upper range limit in the 1000 V range.
Make sure that a current measuring range (“A”) has not been
activated when the multimeter is connected for voltage measurement! If the fuse’s blowing limits are exceeded as a result of operator error, both the operator and the instrument are in danger!
With the rotary switch in the V position, the multimeter is always
set to the 1 V measuring range immediately after it is switched on.
As soon a the MAN / AUTO key is pressed, and assuming the
measured value is less than 90 mV, the multimeter is switched to
the mV measuring range.
... SETrAtE ... CLIP
1nFo
1 / 10/100/1000 / 0ff
FUNC
ENTER
FUNC
ENTER
V/Hz, Ω, Temperature, and A/Hz Measurements
V
020.00
V
FUNC
ENTER
020.00
V
Ω
Tem p
Warnings regarding dangerous voltage:
> 55 V AC or > 70 V DC:
> 1000 V:
COM
– (+)
+ (–)
230.0
DC
V
V
DC
V
AC
V
Measuring Ranges:
V= : 100 mV…1000 V
V: 100 mV…1000 V
Max. 1000 V (< 10 kHz)
Max. 100 V (> 10 kHz)
Hz: 1Hz … 100 kHz
P
max
TRMS
= 3 x 106 V x Hz
for U > 100 V
GMC-I Gossen-Metrawatt GmbH27
V/Hz, Ω, Temperature, and A/Hz Measurements
5.1.2 Alternating Voltage Measurement with 1 MΩ Load Resistance
and Frequency Measurement with Selectable Low-Pass Filter
(METRAHIT⏐X-TRA⏐OUTDOOR⏐TECH⏐PRO only)
The measuring instrument includes a V
electricians with an input resistance of approximately 1 MΩ.
Erroneous displays resulting from capacitive coupling during
voltage measurement in power supply systems are reduced to a
minimum in this way.
Note
METRAHIT⏐TECH: see note in chapter 5.1.1.
➭ In accordance with the voltage to be measured, turn the rotary
switch to V~
➭ Connect the measurement cables as shown.
The “⊥” connector jack should be grounded.
Voltage Measurement
Note
An intermittent acoustic signal warns the operator if the
measured value exceeds the upper range limit in the 1000 V
range.
Make sure that a current measuring range (“A”) has not been
activated when the multimeter is connected for voltage measurement! If the fuse’s blowing limits are exceeded as a result of operator error, both the operator and the instrument are in danger!
➭ You can switch back and forth between voltage measurement
with and without low-pass filter.
➭ Press the FUNC | ENTER multifunction key repeatedly until unit of
measure V or V/Fil appears at the display.
or .
1MΩ
1 kHz
switch position for
1MΩ
Frequency Measurement
➭ Apply the measured quantity is the same way as for voltage
measurement.
➭ Manually select the measuring range for the voltage amplitude.
When the instrument is switched to frequency measurement,
the previously selected voltage measuring range remains
active.
➭ You can switch back and forth between frequency
measurement with and without low-pass filter.
Press the FUNC | ENTER multifunction key repeatedly until unit of
measure Hz or Hz/Fil appears at the display.
Lowest measurable frequencies and maximum allowable
voltages are included in chapter 8, “Technical Data”.
Measurement with Low-Pass Filter
!
Attention!
Be aware of the fact that dangerous voltage spikes are not
displayed during this type of measurement (see also “Voltage
Comparator”.
We recommend measuring voltage without the low-pass filter
first, in order to be able to detect any dangerous voltages.
A 1 kHz low-pass filter can be activated if required, in order to
filter out high frequency pulses of greater than 1 kHz, for example
when performing measurements at pulsed motor drives, i.e.
undesired voltages of greater than 1 kHz can be suppressed.
The active low-pass filter is indicated by the Fil display. The
multimeter is automatically switched to manual measuring range
selection.
28GMC-I Gossen-Metrawatt GmbH
V/Hz, Ω, Temperature, and A/Hz Measurements
V
Tem p
Ω
COM
~
Warnings regarding dangerous voltage:
> 55 V AC or > 70 V DC:
> 1000 V:
Voltage Comparator for Displaying Dangerous Voltage
The input signal or measuring signal is checked by a voltage
comparator for dangerous spikes, because these do not appear
at the display when the low-pass filter is used.
At voltages of greater than 55 V AC or 70 V DC, a danger symbol
appears at the display: .
230.0
Measuring Ranges:
V~: 100 mV…1000 V
Max. 1000 V (< 10 kHz)
Max. 100 V (> 10 kHz)
Hz: 1Hz … 100 kHz
P
= 3 x 106 V x Hz
max
for U > 100 V
1MΩ
V~
!
!
1 kHz
V~ with Filter
Hz with Filter
0225.6
0226.0
050.00
049.99
FUNC
ENTER
FUNC
ENTER
Hz
FUNC
ENTER
Hz
FUNC
ENTER
V
V
Fil
Fil
AC
TRMS
AC
TRMS
AC
AC
V~
V~
1 kHz
Hz
1 kHz
Hz
GMC-I Gossen-Metrawatt GmbH29
V/Hz, Ω, Temperature, and A/Hz Measurements
5.1.3 Alternating Voltage and Frequency Measurement V AC and Hz
with Selectable Low-Pass Filter
(METRAHIT⏐X-TRA⏐OUTDOOR⏐TECH⏐PRO only)
Note
METRAHIT⏐TECH: see note in chapter 5.1.1.
➭ In accordance with the voltage or frequency to be measured,
turn the rotary switch to V~ or Hz.
➭ Connect the measurement cables as shown.
The “⊥” connector jack should be grounded.
Voltage Measurement
Note
An intermittent acoustic signal warns the operator if the measured value exceeds the upper range limit in the 1000 V range.
Make sure that a current measuring range (“A”) has not been
activated when the multimeter is connected for voltage measurement! If the fuse’s blowing limits are exceeded as a result of operator error, both the operator and the instrument are in danger!
➭ You can switch back and forth between voltage measurement
with and without low-pass filter.
➭ Press the FUNC | ENTER multifunction key repeatedly until unit of
measure V or V/Fil appears at the display.
30GMC-I Gossen-Metrawatt GmbH
Frequency Measurement
➭ Apply the measured quantity is the same way as for voltage
measurement.
➭ Manually select the measuring range for the voltage amplitude.
When the instrument is switched to frequency measurement,
the previously selected voltage measuring range remains
active.
➭ You can switch back and forth between frequency
measurement with and without low-pass filter.
Press the FUNC | ENTER multifunction key repeatedly until unit of
measure Hz or Hz/Fil appears at the display.
Lowest measurable frequencies and maximum allowable
voltages are included in chapter 8, “Technical Data”.
Measurement with Low-Pass Filter
!
Attention!
Be aware of the fact that dangerous voltage spikes are not
displayed during this type of measurement (see also “Voltage
Comparator”.
We recommend measuring voltage without the low-pass filter
first, in order to be able to detect any dangerous voltages.
A 1 kHz low-pass filter can be activated if required, in order to
filter out high frequency pulses of greater than 1 kHz, for example
when performing measurements at pulsed motor drives, i.e.
undesired voltages of greater than 1 kHz can be suppressed.
The active low-pass filter is indicated by the Fil display. The
multimeter is automatically switched to manual measuring range
selection.
The specified measuring accuracy is not achieved when the filter
is activated and signals are greater than 100 Hz.
V/Hz, Ω, Temperature, and A/Hz Measurements
V
Tem p
Ω
COM
~
Warnings regarding dangerous voltage:
> 55 V AC or > 70 V DC:
> 1000 V:
Voltage Comparator for Displaying Dangerous Voltage
The input signal or measuring signal is checked by a voltage
comparator for dangerous spikes, because these do not appear
at the display when the low-pass filter is used.
At voltages of greater than 55 V AC or 70 V DC, a danger symbol
appears at the display:
230.0
Measuring Ranges:
V~: 100 mV…1000 V
Max. 1000 V (< 10 kHz)
Max. 100 V (> 10 kHz)
Hz: 1Hz … 100 kHz
P
= 3 x 106 V x Hz
max
for U > 100 V
V~
!
!
Hz
V~ with Filter
Hz with Filter
0230.6
050.03
0229.6
050.00
FUNC
ENTER
Hz
FUNC
ENTER
FUNC
ENTER
Hz
FUNC
ENTER
V
V
Fil
Fil
AC
TRMS
AC
AC
TRMS
AC
V~
Hz
V~
1 kHz
Hz
1 kHz
GMC-I Gossen-Metrawatt GmbH31
V/Hz, Ω, Temperature, and A/Hz Measurements
5.1.4 Transient Overvoltages
The multimeters are protected against transient overvoltages of
up to 8 kV with wave-front durations of 1.2 ms and halftimes of
50 μs in the voltage measuring range. If longer pulse durations
are expected, for example when conducting measurements at
transformers or motors, we recommend the use of our KS30
measuring adapter. It provides protection against transient
overvoltages of up to 6 kV with wave-front durations of 10, and
halftimes of 1000 μs.
Continuous load capacity is 1200 V
. Additional influence error
RMS
caused by the KS30 measuring adapter amounts to approximately -2%.
5.1.5 Voltage Measurements at Above 1000 V
Voltages of greater than 1000 V can be measured with a
high-voltage probe, e.g. the HV31) or the HV302) from GMC-I
Gossen-Metrawatt GmbH. It is absolutely essential to earth the
ground terminal in this case. Observe all applicable safety precautions!
Voltage Measurements at Above 1000 V
with the HV3 High-Voltage Probe
V
Ω
Tem p
Red
1)
HV3: 3 kV
2)
HV30: 30 kV, for (DC) voltages only
Black
COM
Black
x1000
x100
32GMC-I Gossen-Metrawatt GmbH
5.1.6 Frequency and Keying Ratio Measurements
(METRAHIT⏐X-
➭ Set the rotary switch to MHz or %.
➭ Connect the measurement cables as shown.
Make sure that a current measuring range (“A”) has not been
activated when the multimeter is connected for frequency or
keying ratio measurement!
!
Attention!
The applied signal voltage may not exceed 5 V.
Frequency Measurement, MHz
A 5 V signal with a frequency of up to 1 MHz is measured and
displayed using MHz as a unit of measure. Pulse frequency
demonstrates the reciprocal value of pulse period.
Keying Ratio Measurement, t
The ratio of pulse duration to pulse period is measured with
periodic square-wave signals and displayed as a percentage.
keying ratio (%) =
TRA⏐OUTDOOR only)
E/tP
pulse duration (tE)
pulse period (t
• 100
)
P
V/Hz, Ω, Temperature, and A/Hz Measurements
MHz
%
U
t
E
t
P
Pulse Time Quantities
f
P
t
E
t
P
t
P
t
E/tP
pulse frequency = 1/tP
pulse duration
pulse period
– tEinterpulse period
pulse or keying ratio
V
Ω
Tem p
MHz
1.0323
MHz
FUNC
ENTER
t
002.00
%
Measuring Ranges:
MHztE/t
100 Hz ... 1 kHz2 ... 98%
... 10 kHz5 ... 95%
... 100 kHz 10 ... 90%
COM
f
%
tE/t
P
P
P
Note
The applied frequency must remain constant during keying
ratio measurement.
GMC-I Gossen-Metrawatt GmbH33
!
Max. 5 V
V/Hz, Ω, Temperature, and A/Hz Measurements
5.2 Resistance Measurement, Ω
➭ Disconnect supply power from the electrical circuit of the
device to be measured, and discharge all high-voltage
capacitors.
➭ Make sure that the device under test is voltage-free.
Interference voltages distort measurement results!
Refer to chapter 5.1.1 regarding testing for the absence of
voltage with the help of the direct voltage measurement.
➭ Set the rotary switch to “Ω”.
➭ Connect the device under test as shown.
Ω
000.00
k
Ω
Measuring Ranges:
100 Ω … 40 MΩ
Ω
Note
Use short or shielded measurement cables in the case of
high-impedance resistance.
Improving Accuracy by means of Zero Balancing
Cable resistance and contact resistance can be eliminated in all
measuring ranges by means of zero balancing (see chapter 4.2).
34GMC-I Gossen-Metrawatt GmbH
0V !
!
R
x
V
Tem p
Ω
COM
R
x
5.3 Continuity Test
➭ Disconnect supply power from the electrical circuit of the
device to be measured, and discharge all high-voltage
capacitors.
➭ Make sure that the device under test is voltage-free.
Depending upon the selected limit value, the multimeter
generates a continuous acoustic signal in the case of continuity or
short-circuit, i.e. at a value of less than the selected limit value.
“0L” appears at the display in the case of an open connection.
The limit value can be adjusted in the “SETUP” menu (see also
chapter 6.4):
MEASURE
SETUP
FUNC
ENTER
(10 = default setting)
... SETrAtE ... bEEP
1nFo
1, 10, 20, 30, 40, 90Ω
FUNC
ENTER
FUNC
ENTER
V/Hz, Ω, Temperature, and A/Hz Measurements
000.8
Ω
FUNC
ENTER
DC
0.443
V
V
0V !
Ω
!
Tem p
COM
R < 1, 10, 20, 30, 40, 90 Ω
GMC-I Gossen-Metrawatt GmbH35
V/Hz, Ω, Temperature, and A/Hz Measurements
5.4 Diode Testing with a Constant Current of 1 mA
➭ Disconnect supply power from the electrical circuit of the
device to be measured, and discharge all high-voltage
capacitors.
➭ Make sure that the device under test is voltage-free.
Interference voltages distort measurement results!
Refer to chapter 5.1.1 regarding testing for the absence of
voltage with the help of the direct voltage measurement.
➭ Set the rotary switch to .
➭ Press the
➭ Connect the device under test as shown.
Conducting Direction and Short-Circuit
The instrument displays forward voltage in volts (display: 4
places). As long as voltage drop does not exceed the maximum
display value of 5.1 V, several series connected components or
reference diodes can be tested with a small reference voltage and
reference diodes.
FUNC
| ENTER
key.
0V !
000.8
Ω
FUNC
ENTER
DC
0.443
V
!
V
Tem p
Ω
COM
Reverse Direction and Interruption
The measuring instrument indicates overload 0L.
Measuring Range:
to 5.100 V
+
–
0.654
Note
Resistors and semiconductor paths connected in parallel to
the diode distort measurement results!
Forward DirectionReverse Direction
36GMC-I Gossen-Metrawatt GmbH
0.L
V/Hz, Ω, Temperature, and A/Hz Measurements
5.5 Temperature Measurement
Temperature measurement is performed with a type K
thermocouple (accessory, not included), which is connected to
the voltage input. Alternatively, a Pt100 or Pt1000 resistance
thermometer can be used with the METRAHIT⏐X-TRA⏐OUTDOOR .
Selecting the Unit of Measure for Temperature
MEASURE
SETUP
(°C = default setting)
5.5.1 Measurement with Thermocouples, Temp TC
➭ Set the rotary switch to “Temp
Note
METRAHIT⏐X-TRA⏐OUTDOOR only:
The last selected temperature measurement or the last
selected temperature sensor, i.e. type K or Pt100/Pt1000,
remains in memory and is accordingly displayed. Press the
FUNC | ENTER key in order to change to the other measuring
function if required.
➭ The reference temperature is measured at the internal
reference junction (see parameter “1tEMP ” in chapter 6.3
regarding querying).
... tEMP°C / °F
1nFo
FUNC
ENTER
TC
”.
FUNC
ENTER
Note
The internal reference temperature (temperature of the internal reference junction) is measured by a temperature sensor
inside of the instrument. This may be somewhat above room
temperature as a result of internal heat-up, or moving from
warmer to colder surroundings or vice versa.
➭ Connect the sensor to the two accessible jacks. The instru-
ment displays the measured temperature using the selected
unit of measure.
Tem p
TC
TC
0023.3
°Χ
FUNC
ENTER
Typ e K
V
Tem p
Tem p
RTD
Ω
COM
RTD
TE (TC)
0256.0
Measuring Range
TE (TC) K (NiCr-Ni) –250.0 … +1372.0 °C
°Χ
Pt1000
GMC-I Gossen-Metrawatt GmbH37
V/Hz, Ω, Temperature, and A/Hz Measurements
5.5.2 Measurement with Resistance Thermometers
(METRA
HIT⏐X-
TRA
⏐
O
UTDOOR
only)
➭ Set the rotary switch to “TempTC” or “Temp
The last selected temperature measurement or sensor, i.e. type K
or Pt100/Pt1000, remains in memory and is accordingly
displayed. Press the FUNC | ENTER key in order to change to the
other measuring function if required. The sensor type, i.e. Pt100
or Pt1000, is detected automatically and displayed.
There are two different ways to compensate for cable resistance:
Automatic Compensation
➭ Press the ZERO | ESC key.
“Short leads” appears at the display.
If you prefer to enter cable resistance directly, you can skip the
following entry prompt.
➭ Short circuit the measuring instrument’s connector cables.
“000.00 ” appears at the display. After pressing the FUNC | ENTER key, automatic compensation of cable resistance is
activated for all subsequent measurements. The short-circuit
can now be eliminated, and the device is ready for use.
Entering Cable Resistance
➭ Press the ZERO | ESC key once again in the automatic
compensation menu.
➭ Enter the known resistance of the connector cables with the
scroll keys: Select the digit to be changes with the keys,
and change the respectively selected digit with the
The default value is 0.1 Ω. Values can be selected within a
range of 0 to 50 Ω.
RTD
”.
keys.
➭
Upon pressing the
and the display is returned to the measuring function. Cable resistance
remains in memory even after the instrument has been switched off.
FUNC | ENTER
Temp
TC
key, the selected value is activated
TC
Tem p
RTD
Measuring Ranges
RTD Pt100 – 200.0 … +850.0 °C
RTD Pt1000 –150.0 … +850.0 °C
Short Leads
Automatic
Compensation
V
Tem p
Ω
COM
RTD
0023.3
0025.6
FUNC
ENTER
000.00
Ω
000.50
Ω
Enter Cable Resistance
FUNC
ENTER
R L
ZERO
R L
ESC
°Χ
°Χ
Typ e K
RTD
Pt1000
ZERO
ESC
FUNC
ENTER
38GMC-I Gossen-Metrawatt GmbH
5.6 Capacitance Measurement
(METRAHIT⏐X-
➭ Disconnect supply power from the electrical circuit of the
device to be measured, and discharge all high-voltage
capacitors.
➭ Make sure that the device under test is voltage-free.
Capacitors must always be discharged before measurement is
performed.
Interference voltages distort measurement results!
Refer to chapter 5.1.1 regarding testing for the absence of
voltage with the help of the direct voltage measurement.
➭ Set the rotary switch to “”.
➭ Connect the (discharged!) device under test to the sockets
with the measurement cables as shown.
Note
The “–” pole of polarized capacitors must be connected to
the “⊥” jack.
Resistors and semiconductor paths connected in parallel to
the capacitor distort measurement results!
TRA⏐OUTDOOR and METRAHIT⏐TECH only)
V/Hz, Ω, Temperature, and A/Hz Measurements
10.38
μ
F
!
V
0V !
+
Ω
Tem p
–
Measuring Ranges:
10 nF ... 1000 μF
COM
–
+
GMC-I Gossen-Metrawatt GmbH39
V/Hz, Ω, Temperature, and A/Hz Measurements
5.7 Current Measurement
Notes Regarding Current Measurement
• The multimeter may only be operated with installed batteries or
rechargeable batteries. Dangerous currents are otherwise not
indicated, and the instrument may be damaged.
• Set up the measuring circuit in a mechanically secure fashion,
and secure it against inadvertent breaks. Select conductor
cross-sections and lay out connections such that they do not
overheat.
• An intermittent acoustic signal warns of current greater than
10 A. An continuous acoustic signal warns of current greater
than 16 A.
• The input for the current measuring range is equipped with a
fuse link. Maximum permissible voltage for the measuring
circuit (= rated voltage of the fuse) is 1000 V AC/DC.
Use specified fuses only! The fuse must have a breaking capacity of at least 30 kA.
• If the fuse for the active current measuring range blows, “FUSE ”
appears at the digital display, and an acoustic signal is
generated at the same time.
• If a fuse should blow, eliminate the cause of overload before
placing the instrument back into service!
• Fuse replacement is described in chapter 9.3.
• Be absolutely certain that the measuring ranges are not
overloaded beyond their allowable capacities. Limit values are
included in chapter 8, “Technical Data”, in the table entitled
“Measuring Functions and Measuring Ranges” in the “Overload Capacity” column.
Scope of Functions, Current Measurement, Direct Connection
FunctionMETRAHIT⏐
A AC / Hz~100 μA
A AC+DC TRMS
A DC100 μA
1000 V fuse
TRA /
X-
O
UTDOOR
1/10/100 mA
1 A / 10 (16) A
100 μA
1/10/100 mA
1A / 10 (16) A
1/10/100 mA
1A / 10 (16) A
METRAHIT⏐
ECH
T
10/100 mA
1 A / 10 (16) A
10/100 mA
1A / 10 (16) A
10/100 mA
1A / 10 (16) A
•••—
METRAHIT⏐
RO
P
1 A / 10 (16) A
1 A / 10 (16) A
1 A / 10 (16) A
METRAHIT⏐
ASE
B
—
—
—
Scope of Functions, Current Measurement via Clip-On Current Sensor
FunctionMETRAHIT⏐
Transformation
Factor
A AC / Hz
A AC+DC
A DC
Hz (A AC)
X-
TRA /
OUTDOOR
... 30 kHz... 30 kHz... 30 kHz... 30 kHz
METRAHIT⏐
T
ECH
—•—•
—•—•
—•—•
—•—•
METRAHIT⏐
P
RO
METRAHIT⏐
B
ASE
40GMC-I Gossen-Metrawatt GmbH
5.7.1 Direct and Pulsating Current Measurement, Direct Connection,
A DC and A (DC+AC)
(METRAHIT⏐X-
➭ First disconnect supply power from the measuring circuit or
the power consumer (1), and discharge any capacitors.
➭ In accordance with the current to be measured, turn the rotary
switch to A or A.
➭ Select the current type appropriate for the measured quantity
by briefly pressing the FUNC | ENTER multifunction key. Each
time the key is pressed, the instrument is switched back and
forth between A DC and A (DC + AC)
by means of an acoustic signal. The current type is indicated at
the LCD by means of the DC or the (DC+AC)
➭ Safely connect the measuring instrument (without contact
resistance) in series to the power consumer (2) as shown.
➭ Switch supply power to the measuring circuit back on (3).
➭ Read the display. Make a note of the measured value if the
instrument is not being operated in the memory mode or the
transmission mode.
➭ Disconnect supply power from the measuring circuit or the
power consumer (1) once again, and discharge any
capacitors.
➭ Remove the test probes from the measuring point and return
the measuring circuit to its normal condition.
TRA⏐OUTDOOR⏐TECH⏐PRO only)
TRMS
, which is indicated
symbol.
TRMS
V/Hz, Ω, Temperature, and A/Hz Measurements
A
A
Current measurement
!
may only be performed
with installed batteries!
8.8.8.8.8
Measuring Ranges:
METRAHIT⏐PRO
1 A / 10 A (16 A max. 30 s)
METRAHIT⏐TECH
10 mA / 100 mA
1 A / 10 A (16 A max. 30 s)
METRAHIT⏐X-TRA/OUTDOOR
100 μA / 1 mA
10 mA / 100 mA
1 A / 10 A (16 A max. 30 s)
1
003.50
003.50
V
Ω
Tem p
3
A
FUNC
ENTER
A
COM mA A
2
DC
DC
AC
TRMS
I > 10 A
I > 16 A
A
A
!
~
R
x
GMC-I Gossen-Metrawatt GmbH41
~
R
x
Current
V/Hz, Ω, Temperature, and A/Hz Measurements
5.7.2 Alternating Current and Frequency Measurement,
Direct Connection, A AC and Hz
(METRAHIT⏐X-TRA⏐OUTDOOR⏐TECH⏐PRO only)
➭ First disconnect supply power from the measuring circuit or
the power consumer (1), and discharge any capacitors.
➭ In accordance with the current or frequency to be measured,
turn the rotary switch to A~ or Hz.
➭ Select the desired measured quantity by briefly pressing the
FUNC | ENTER multifunction key. Each time the key is pressed,
AC
switching is acknowledged with an acoustic signal.
➭ Safely connect the measuring instrument (without contact
resistance) in series to the power consumer as shown.
➭ Switch supply power to the measuring circuit back on (3).
➭ Read the display. Make a note of the measured value if the
instrument is not being operated in the memory mode or the
transmission mode.
➭ Disconnect supply power from the measuring circuit or the
power consumer (1) once again, and discharge any
capacitors.
➭ Remove the test probes from the measuring point and return
the measuring circuit to its normal condition.
and Hz are alternately selected, and
TRMS
A~
Hz
Current measurement
!
may only be performed
with installed batteries!
8.8.8.8.8
Measuring Ranges:
METRAHIT⏐PRO
1 A / 10 A (16 A max. 30 s)
METRAHIT⏐T
10 mA / 100 mA
1 A / 10 A (16 A max. 30 s)
METRAHIT⏐X-TRA/OUTDOOR
100 μA / 1 mA
10 mA / 100 mA
1 A / 10 A (16 A max. 30 s)
ECH
1
003.50
050.10
V
Ω
Tem p
3
A
FUNC
ENTER
Hz
COM mA A
2
AC
TRMS
AC
!
I > 10 A
I > 16 A
A~
Hz
~
R
x
42GMC-I Gossen-Metrawatt GmbH
~
R
Current
x
5.7.3 Direct and Pulsating Current Measurement with Clip-On Current Sen-
sor, A DC and A (DC+AC) (METRA
HIT⏐T
ECH
and METRA
HIT⏐B
ASE
only)
Transformer Output, Voltage/Current
When a clip-on current sensor is connected to the multimeter
(METRAHIT⏐T
ECH: V input, METRAHIT⏐BASE: V input), all current
displays appear with the correct value in accordance with the selected transformation factor.
The only prerequisite is that the current
sensor is equipped with at least one of the below listed transformation
factors, and that the factor has been previously selected in the following
Cl ip ≠
0FF
menu (
), see also chapter 6.4.
Current Clip Setup Menu
MEASURE
SETUP
FUNC
ENTER
1nFo
... SET rAtE ... CLIP
1 / 10/100/1000 / 0ff
FUNC
ENTER
FUNC
ENTER
V/Hz, Ω, Temperature, and A/Hz Measurements
METRAHIT⏐BASE
A
003.50
A
METRAHIT⏐TECH
A
FUNC
ENTER
V
003.50
V
A
1:1000
1:1000
DC
DC
AC
TRMS
A
A
Trans. Factors
CL IP
1:1
1 mV / 1 mA
1:10
1m V / 10 mA
1:100
1m V / 100 mA
1:1000
1 mV / 1 A
Measuring Ranges DMMClip Types
100 mV1 V10 V
100.00 mA1.0000 A10.000 AWZ12C
1.0000 A10.000 A100.00 AWZ12B, Z201A
10.000 A100.00 A1000.0 AZ202A
100.00 A1000.0 A(10000.0 A)
Z202A, Z203A,
WZ12C
METRAHIT⏐TECH
V
Tem p
Ω
COM
R
i
mA
Ri = 9 M
A
Ω
METRAHIT⏐BASE
V
Tem p
Ω
COM
Hz
R
i
Ri ~ 1 M
V
Ω
Maximum allowable operating voltage is equal to the current
transformer’s nominal voltage. When reading the measured value,
additional error resulting from the clip-on current sensor must also
be taken into consideration.
(default setting:
GMC-I Gossen-Metrawatt GmbH43
METRAHIT⏐TECH: OFF, METRAHIT⏐BASE: 1:1000)
~
Current
R
x
R
Current
x
~
V/Hz, Ω, Temperature, and A/Hz Measurements
5.7.4 Alternating Current Measurement with Clip-On Current Sensor,
A AC and Hz (METRAHIT⏐T
Transformer Output, Voltage/Current
When a clip-on current sensor is connected to the multimeter
(METRAHIT⏐T
ECH: V input, METRAHIT⏐BASE: V input), all current
displays appear with the correct value in accordance with the selected transformation factor.
sensor is equipped with at least one of the below listed transformation
factors, and that the factor has been previously selected in the following
Cl ip ≠
0FF
menu (
), see also chapter 6.4.
Current Clip Setup Menu
MEASURE
SETUP
FUNC
ENTER
... SETrAtE ... CLIP
1nFo
1 / 10/100/1000 / 0ff
ECH and METRAHIT⏐BASE only)
The only prerequisite is that the current
FUNC
ENTER
FUNC
ENTER
V~
METRAHIT⏐BASE
Hz
METRAHIT⏐TECH
Hz
Ri = 9 M
Ω
1MΩ
V~
A~
Ri = 1 M
1:1000
A~
003.50
AC
TRMS
A
FUNC
ENTER
1:1000
1kHz
Ω
050.10
Hz
Hz
AC
Trans. Factors
CL IP
1:1
1 mV / 1 mA
1:10
1m V / 10 mA
1:100
1m V / 100 mA
1:1000
1 mV / 1 A
Measuring Ranges DMMClip Types
100 mV1 V10 V
100.00 mA1.0000 A10.000 AWZ12C
1.0000 A10.000 A100.00 AWZ12B, Z201A
10.000 A100.00 A1000.0 AZ202A
100.00 A1000.0 A(10000.0 A)
Z202A, Z203A,
WZ12C
METRAHIT⏐TECH
V
Tem p
Ω
COM
R
i
A
mA
Ri = 1 MΩ/9 M
METRAHIT⏐BASE
V
Ω
Tem p
Ω
COM
Hz
R
i
Ri ~ 1 M
V
Ω
Maximum allowable operating voltage is equal to the current
transformer’s nominal voltage. When reading the measured value,
additional error resulting from the clip-on current sensor must also
be taken into consideration.
(default setting:
44GMC-I Gossen-Metrawatt GmbH
METRAHIT⏐TECH: OFF, METRAHIT⏐BASE: 1:1000)
~
Current
R
x
R
Current
x
~
5.7.5
Direct, Pulsating and Current Measurement
Transformer A DC, A (DC+AC), A AC and Hz (METRAHIT⏐T
Transformer Output, Current/Current
When a clip-on current transformer is connected to the multimeter (METRAHIT⏐T
with the correct value in accordance with the selected transformation factor. The only prerequisite is that the current transformer
is equipped with at least one of the below listed transformation
factors, and that the factor has been previously selected in the
following menu
Current Clip Setup Menu
MEASURE
SETUP
FUNC
ENTER
Trans. Factors
CL IP
1:1
1mA/1mA
1:10
1mA/10mA
1:100
1mA/100mA
1:1000
1 mA/1 A
ECH: mA/A input), all current displays appear
(Cl ip ≠ 0FF), see also chapter 6.4.
1nFo
... SET rAtE ... CLIP
1 / 10/100/1000 / 0ff
Measuring Ranges DMMClip Types
100 mA1 A10 A
100,00 mA1,0000 A10,000 A
1,0000 A10,000 A100,00 A
10,000 A100,00 A1000,0 A
100,00 A1000,0 A(10000,0 A)
FUNC
ENTER
with Clip-On Current
FUNC
ENTER
ECH only)
WZ12A, WZ12D,
WZ11A, Z3511,
Z3512, Z3514
V/Hz, Ω, Temperature, and A/Hz Measurements
A
A
1:1000
003.50
003.50
A
FUNC
ENTER
1:1000
A
DC
A
DC
A
AC
TRMS
METRAHIT⏐TECH
V
Tem p
Ω
COM
Hz
003.50
FUNC
ENTER
050.10
Hz
A
mA
R
i
Ri ≤ 1
Ω
A~
A~
AC
TRMS
A
Hz
AC
(default setting:
GMC-I Gossen-Metrawatt GmbH45
METRAHIT⏐TECH: OFF)
~
Current
R
x
Device and Measuring Parameters
6Device and Measuring Parameters
The instrument’s “SET ” mode (menu mode) makes it possible to
set operating and measuring parameters, query information and
activate the interface.
➭ The menu mode is accessed by pressing the MEASURE | SETUP
key, assuming that the instrument is switched on and set to
“Measure” (measuring mode operation).
“1nFo ” appears at the display.
➭ The main menus, i.e. the “SET ” and “tEMP ” menus, as well as
the “SEnd ” and “StorE ” menus included with the
METRAHIT⏐X-
returned to “1nFo” by activating the
direction).
➭ After selecting the desired main menu, sub-menus are
accessed by pressing the FUNC | ENTER key.
➭ The desired parameter is selected by repeatedly pressing the
or key.
➭ In order to check or change a parameter, acknowledge it with
the FUNC | ENTER key.
➭ The
keys can be used to position the cursor at the entry
position.
The desired value is selected with the help of the
➭ Changes can only be accepted with the FUNC | ENTER key.
➭ You can return to the sub-menu without making any changes
by pushing the ZERO | ESC key, and to the main menu by pressing the same key once again etc.
➭ You can switch to the measuring mode from any menu level by
pressing the FUNC | ENTER key.
TRA⏐OUTDOOR, are accessed, and the display is
keys (in any
keys.
After repeatedly pressing the MEASURE | SETUP key (without first
turning the multimeter off), you can return to the last selected
menu or parameter from the measuring mode.
Example: Setting Time
MEASURE
SETUP
FUNC
ENTER
... setrAte ... t iME
1nFo
10:24 (hh:mm)
FUNC
ENTER
FUNC
ENTER
Setting hours and minutes:
Change the setting, the entry position blinks.
ENTER
Advance to desired entry position.
Press and hold the key
FUNC
The new time setting is activated after
to change the setting rapidly.
acknowledgement.
46GMC-I Gossen-Metrawatt GmbH
6.1 Paths to the Various Parameters 6.2 List of All Parameters
Parameter X
TRATECHPROBASE
0.diSP••••48: 0.diSP – show/hide leading zeros
MEASURE
DC
020.00
Main Menus→
AC
TRMS
V
1nFo
FUNC
ENTER
Sub-Menus / Parameters ↓
QuerySetSetSetSet
bAtt:
VerSion
1tEMP:
dAtE:
t iME:
0CCvP: *
FUNC
ENTER
000.0
:
CLEAr
StArt/
StoP
Selection
FUNC
ENTER
SETUP
FUNC
ENTER
StArt
StoP
Selection
FUNC
ENTER
1nfo
*SETStorE *
FUNC
ENTER
rAtE *
0.diSP
.diSP
A
APoFF
bEEP
irStb
*
Addr *
dAtE
* METRAHIT⏐X-
** METRA HIT⏐B
TRA⏐OUTDOOR only
ASE only
tiME
CLiP **
Selection
FUNC
ENTER
tEMPSEnd
FUNC
ENTER
°C
°F
Selection
FUNC
ENTER
Addr
A.diSP••••49: A.diSP – analog display: select display mode
APoFF
bAtt
bEEP
CLEAr
CLIP
dAtE
EMpty
1nfo
irStb
1tEMP
0CCvP
rAtE
SEnd
SET
StArt
StoP
store
tEMP
t iME
vErSion
•———
••••
••••
••••
•———
—•—•
••••
•———
••••
•———
••••
•———
•———
•———
••••
•———
•———
•———
••••
••••
••••
Device and Measuring Parameters
Page: Header
53: Configuring Interface Parameters
49: APoFF – specified time for automatic shutdown
and continuous ON
48: bAtt – query battery voltage
49: bEEP – set the limit value for continuity testing
23: Measurement Data Recording (METRA HIT⏐X-
TRA⏐OUTDOOR only)
43:
Direct and Pulsating Current Measurement with
Clip-On Current Sensor, A DC and A (DC+AC)
(
METRAHIT⏐T
48: dAtE – query date, 50: dAtE – enter date
23: Measurement Data Recording (METRA HIT⏐X-
TRA⏐OUTDOOR only)
48: Querying Parameters – InFo Menu (as moving letters)
53: Configuring Interface Parameters
48: ItEMP – query reference temperature
23: Measurement Data Recording (METRAHIT⏐X-
DOOR
48: rAtE – set the sampling rate (METRAHIT⏐X-
DOOR
52: Activating the Interface
48: Entering Parameters – SETUP Menu
23: Measurement Data Recording (METRA HIT⏐X-
TRA⏐OUTDOOR only)
37: Temperature Measurement
48: tiME – query time, 50: tiME – set time
48: vErSion – query firmware version
only)
only)
ECH
and
METRAHIT⏐B
ASE
only)
TRA
TRA
⏐
OUT-
⏐
OUT-
GMC-I Gossen-Metrawatt GmbH47
Device and Measuring Parameters
6.3 Querying Parameters – InFo Menu (as moving letters)
bAtt – query battery voltage
MEASURE
vErSion – query firmware version
MEASURE
1nFobAtt
SETUP
1nFobAtt: vErSion: 2.09
SETUP
FUNC
ENTER
FUNC
ENTER
:
2.75 V.
ItEMP – query reference temperature
The temperature of the internal reference junction is measured with a
temperature sensor in close proximity to the input sockets.
MEASURE
dAtE – query date
MEASURE
D = day, M = month, Y = year
1nFobAtt: ...
SETUP
1nFobAtt: ...
SETUP
FUNC
ENTER
FUNC
ENTER
1tEMP: 24 °
dAtE: 31.12.
C
05 (DD.MM.YY)
Date and time must be reentered after replacing the batteries.
tiME – query time
MEASURE
1nFobAtt
SETUP
FUNC
ENTER
:
...
t iME: 13:46:
56
(hh:mm:ss)
h = hours, m = minutes, s = seconds
Date and time must be reentered after replacing the batteries.
rAtE – set the sampling rate (METRA HIT⏐X-TRA⏐OUTDOOR only)
The sampling rate specifies the time interval after which the
respective measured value is transmitted to the interface, or to
measured value memory.
Any one of the following sampling rates can be selected:
00:00.1, 00:00.2, 00:00.5, 00:01.0, 00:02.0, 00:05.0
[h:mm:ss.t] (h = hours, m = minutes, s = sec.,
This parameter determines whether or not leading zeros will
appear in the measured value display.
MEASURE
SETUP
1nFo
...
FUNC
SET rAtE ... 0.diSP
ENTER
FUNC
ENTER
0000.0 : with leading zeros (default setting)
0.0 : leading zeros suppressed
)
FUNC
ENTER
48GMC-I Gossen-Metrawatt GmbH
Device and Measuring Parameters
A.diSP – analog display: select display mode
One of two different display modes can be selected for the
analog display:
• bArG: bar graph
• Point: pointer
MEASURE
SETUP
FUNC
ENTER
... SETrAtE ... A.diSP
1nFo
bArG / Point
FUNC
ENTER
FUNC
ENTER
(bArG = default value)
APoFF – specified time for automatic shutdown and continuous ON
The instrument is switched off automatically if the measured value
remains unchanged for a long period of time, and if none of the
keys or the rotary switch have been activated before the specified
time “APoFF ” (entered in minutes) has elapsed.
If the on setting is selected, the multimeter is set to continuous
operation and
ON appears in the display to the right of the battery
symbol. In this case, the multimeter can only be switched off
manually. The “on ” setting can only be cancelled by changing the
respective parameter, and not by switching the instrument off.
MEASURE
SETUP
FUNC
ENTER
... SETrAtE ... APoFF
1nFo
10 ... 59 min on
FUNC
ENTER
FUNC
ENTER
bEEP – set the limit value for continuity testing
MEASURE
SETUP
FUNC
ENTER
1nFo
... SETrAtE ... bEEP
1, 10, 20, 30, 40, 90Ω
FUNC
ENTER
FUNC
ENTER
(10 Ω = default setting)
irStb – status of the infrared receiver in the stand-by mode
(METRAHIT⏐X-
TRA⏐OUTDOOR only)
See chapter 7.2 on page 53 regarding settings.
Addr – select device address (METRAHIT⏐X-TRA⏐OUTDOOR only
See chapter 7.2 on page 53.
(10 minutes = default setting)
GMC-I Gossen-Metrawatt GmbH49
Device and Measuring Parameters
dAtE – enter date
Entering the current date makes it possible to acquire measured
values in real-time.
MEASURE
SETUP
FUNC
ENTER
... SETrAtE ... dAtE
1nFo
31.12 (DD: day . MM: month)
2005 (YYYY: year)
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
Date and time must be reentered after replacing the batteries.
tiME – set time
Entering the correct time makes it possible to acquire measured
values in real-time.
MEASURE
SETUP
FUNC
ENTER
1nFo ... SETrAtE ... t iME
10:24(hh:mm)
FUNC
ENTER
FUNC
ENTER
Date and time must be reentered after replacing the batteries.
CLIP – set transformation factor (METRAHIT⏐BASE only)
See chapter 5.7.3 ff.
6.5 Default Settings
Previously entered changes can be undone, and the default
settings can be reactivated. This may be advisable under the
following circumstances:
• After the occurrence of software or hardware errors
• If you are under the impression that the multimeter does not
work correctly
➭ Disconnect the device from the measuring circuit.
➭ Remove the batteries temporarily (see also chapter 9.2).
The METRAHIT⏐X-TRA⏐OUTDOOR is equipped with an infrared interface for the transmission of measurement data to a PC. Measured
data are optically transferred through the instrument housing by
means of infrared light to an interface adapter (accessory), which
is
attached to the multimeter. The adapter’s USB interface allows for
the establishment of a connection to the PC via an interface cable.
Beyond this, commands and parameters can be transmitted from
the PC to the multimeter as well. The following functions can be
executed:
• Configuration and read-out of measuring parameters
• Measuring function and measuring range selection
• Start measurements
• Read out stored measured values
7.1 Activating the Interface
The interface is automatically activated for receiving operation
(multimeter receives data from the PC) as soon as the interface is
addressed by the PC, assuming that the “irStb ” parameter has
been set to “ir on ” (see chapter 7.2), or the instrument is already
switched on (the first command wakes up the multimeter, but
does not yet execute any further commands).
The “continuous transmission” operating mode is selected
manually as described below. In this operating mode, the
instrument continuously transmits measurement data to the PC
via the interface adapter, which can then be displayed with the
help of a terminal program.
Starting Continuous Transmission Operation with Menu Functions
MEASURE
SETUP
... SEndStArtsend
1nFo
FUNC
ENTER
FUNC
ENTER
The symbol blinks at the display in order to indicate interface
operation.
Stopping Continuous Transmission Operation with Menu Functions
MEASURE
SETUP
FUNC
ENTER
1nFo
... SEndsend
stop send
ENTER
FUNC
ENTER
FUNC
The symbol is cleared from the display.
Automatic Activation and Deactivation of Transmission Mode Operation
If the sampling rate is 10 seconds or longer, the display is
switched off automatically between samples in order to prolong
battery service life. The only exception is when the multimeter is
set to continuous operation.
As soon as an event occurs, the display is automatically switched
back on.
52GMC-I Gossen-Metrawatt GmbH
Interface Operation
7.2 Configuring Interface Parameters
irStb – status of the infrared receiver in the stand-by mode
There are two possible switching statuses for the infrared
interface when the multimeter is switched off:
ir on: IR appears at the display and the infrared interface is
active, i.e. signals such as making commands can be
received, and power is consumed even though the
multimeter is switched off.
ir oFF:IR does not appear at the display and the infrared
interface is switched off, signals cannot be received.
MEASURE
SETUP
FUNC
ENTER
... SET rAtE ... irStb
1nFo
ir on / ir oFF
FUNC
ENTER
FUNC
ENTER
(irstb = ir oFF = default setting)
Addr – Address
If several multimeters are connected to the PC via an interface
adapter, a separate address can be assigned to each instrument.
Address number 1 should be selected for the first instrument, 2
should be assigned to the second and so forth.
Residual value deviates within 1 ... 30 d from the zero point due to TRMS converter
when probe tips are short-circuited. See frequency influence on page 56.
5)
Switch-off time/Cool-down time > 30 min and TA ≤ 40 °C
4)
~
4)
~
±(1% rdg. + 30 d)
> 300 d
4)
4)
ECH)
±(1% rdg. + 30 d)
> 300 d
Overload Capacity
ValueTime
1)
1000 V
DC
AC
RMS
sine
0.2 A
10 A: ≤ 5 min
16 A: ≤ 30 s
0.2 A
10 A: ≤ 5 min
16 A: ≤ 30 s
Measuring input
10 A: 5 min
Measurement input
1000 V RMS Max. 10 s
1000 V
DC
AC
RMS
sine
continuous
continuous
continuous
Max. 10 s
2)
5)
5)
5)
5)
54GMC-I Gossen-Metrawatt GmbH
Technical Data
Meas.
Function
Measuring Range
Resolution at Upper
Range Limit
11,9991199~ / ValueTime
Input impedance
Discharge resistanceU
F
TRA
X-
OUTDOOR
TECH
Hz (V)
Hz (A)
Hz (A)
Hz (V)
Hz (A)
MHz
X-
TRA
OUTDOOR
%
X-
TRA
OUTDOOR
°C/°F
2)
At 0 ° ... + 40 °C
6)
Applies to measurements at film capacitors
7)
Lowest measurable frequency for sinusoidal measuring signals symmetrical to the zero point
8)
Overload capacity of the voltage measurement input:
power limiting: frequency x voltage, max. 3 x 10
10 nF10 pF10 MΩ0.7V1 + 6
100 nF100 pF1 MΩ0.7V1 + 6
1 μF1 nF100 kΩ0.7V1 + 6
10 μF10nF12kΩ0.7V1 + 6
100 μF100 nF3 kΩ0.7V5 + 6
1000 μF1μF3kΩ0.7V5 + 6
100.00 Hz0.01 Hz
1.0000 kHz0.1 Hz
10.000 kHz1 Hz
100.00 kHz10 Hz10 Hz
30.00 kHz10 Hz10 Hz
100 Hz ... 1 MHz
0.01 ...
100 Hz
2.0 ... 98%—0.01% 100 Hz ... 1 kHz1 Hz0.1 R per kHz> 2 V ... 5 V
5.0 ... 95%—0.01% ... 10 kHz1 Hz0.1 R per kHz> 2 V ... 5 V
10... 90%—0.01%... 100 kHz1 Hz0.1 R per kHz> 2 V ... 5 V
Pt100
– 200.0 ...
X-TRA
+850.0 °C
OUTD.
Pt1000
X-TRA
OUTD.
K
(NiCr-Ni)
– 150.0 ...
+850.0 °C
– 250.0 ...
+1372.0 °C
0.1 °C
6
V x Hz for U > 100 V
0 max
7)
f
min
1Hz
1 ... 100 Hz
Overload Capacity
1000 V
DC
AC
RMS
Sine
2)
Max. 10 s
Intrinsic Error under Reference Conditions
±(... % rdg. + ... d)
6)
with ZERO function active
6)
6)
6)
6)
6)
±(... % rdg. + ... d)
Hz (V) 8):
Hz(A)8):
0.05 + 3
10)
1000 V
Hz (A):
9
)
Max. 10 s
0.05 + 3> 2 V ... 5 V
1000 V Max. 10 s
±(... % rdg. + ... d)
11)
0.3 + 15
11)
0.3 + 15
11)
1% + 5 K
9)
Overload capacity of the current measurement input:
See current measuring ranges for maximum current values.
10)
Input sensitivity, sinusoidal signal, 10% to 100% of the measuring range
11)
Plus sensor deviation
1000 V
DC/AC
RMS
Sine
Max. 10 s
Key: R = meas. range, d = digit(s), rdg. = measured value (reading)
Power limiting: frequency x voltage max. 3 x 106 V x Hz for U > 100 V
3)
The accuracy specification for frequency response is valid within a dis-
1 V / 10 V
> 15 Hz ... 45 Hz
...
> 65 Hz ... 10 kHz
3 + 10—
> 65 Hz ... 1 kHz—3 + 10
play value range of 10% to 100% of the measuring range for both measuring modes with the TRMS converter in the AC and (AC+DC) ranges.
4)
METRAHIT⏐X-TRA⏐OUTDOOR: Frequency response up to 20 kHz,
METRAHIT⏐T
METRAHIT⏐P
METRAHIT⏐B
ECH: Frequency response up to 10 kHz,
RO: Frequency response up to 10 kHz,
ASE: Frequency response up to 1 kHz
56GMC-I Gossen-Metrawatt GmbH
Technical Data
Influencing
Quantity
Crest factor CF
5)
Except for sinusoidal waveshape
Influencing Qty.Sphere of Infl.Measured QuantityInfluence Error
Relative
humidity
Influencing
Quantity
Common Mode
Interference
Voltage
Series Mode
Interference
Voltage
Sphere of
Influence
1 ... 3
> 3 ... 5± 3 % rdg.
75%
3 days
instrument off
Sphere of Influence
Interference quantity max. 1000 VV> 120 dB
Interference quantity max. 1000 V
50 Hz ... 60 Hz, sine
Interference quantity: V,, respective
nominal value of the meas. range,
max. 1000 V , 50 Hz ... 60 Hz, sine
Interference quantity max. 1000 VV> 110 dB
Measured Quantity /
Measuring Range
V, A
V, A, Ω, Hz, °C1 x intrinsic error
Measured Quantity /
Measuring Range
1 V , 10 V> 80 dB
Influence Error
± 1 % rdg.
100 V> 70 dB
1000 V> 60 dB
V> 50dB
5)
Damping
Reference Conditions
Ambient temperature+23 °C ±2K
Relative humidity40 ... 75%
Measured qty. frequency45 ... 65 Hz
Measured qty. waveshapesine
Battery voltage3 V ±0.1 V
Response Time (after manual range selection)
Measured Quantity /
Measuring Range
V, V
AV, A
100 Ω ... 1 MΩ2s
10/40 MΩ5s
Continuity< 50 ms
°C (Pt100)max. 3 s
10 nF ... 100 μFmax. 2s
1 000 μFmax. 7s
>10 Hz1.5 s
Response Time
Digital Display
1.5 s
1.5 s
Measured Quantity Jump Function
of upper range limit value
of upper range limit value
of upper range limit value
Internal Clock
Time formatTT.MM.JJJJ hh:mm:ss
Resolution0.1 s
Accuracy±1 minute per month
Temperature
influence50 ppm/K
From 0 to 80%
From ∞ to 50%
From 0 to 50%
GMC-I Gossen-Metrawatt GmbH57
Technical Data
Data Interface (METRAHIT⏐X-TRA⏐OUTDOOR only)
TypeOptical via infrared light through the
Data transmission
(data transfer)Serial, bidirectional (not IrDa compatible)
ProtocolDevice specific
Baud Rate38,400 baud
Functions– Select/query measuring functions and
The USB X-TRA plug-in interface adapter (see accessories) is
used for adaptation to the PC’s USB port.
housing
parameters
–
Query/transmit momentary measurement
data
– Read out stored measurement data
Internal Measured Value Storage
(METRAHIT⏐X-
Memory capacity4 MBit / 540 kB for approx. 15,400
TRA⏐OUTDOOR only)
measured values with time stamp
Power Supply
Battery2 ea. 1.5 V mignon cell (2 ea. size AA),
Service lifeWith alkaline manganese: approx. 200
Battery testBattery capacity display with battery
Power OFF functionThe multimeter is switched off
Power pack connector socket
(METRAHIT⏐X-
TRA⏐OUTDOOR only)
alkaline manganese per IEC LR6
(2 ea. 1.2 V NiMH rechargeable battery
also possible)
hours
symbol in 4 segments: .
Querying of momentary battery voltage via
menu function.
automatically:
– If battery voltage drops to below approx.
1.8 V
– If none of the keys or the rotary switch
are activated for an adjustable duration
(10 to 59 min.), and the multimeter is not
in the continuous operating mode
If the NA X-TRA power pack (see
accessories) has been plugged into the
instrument, the batteries are disconnected
automatically.
Rechargeable batteries can only be
recharged externally.
58GMC-I Gossen-Metrawatt GmbH
Technical Data
Display
LCD panel (65 mm x 36 mm) with analog and digital display
including unit of measure, type of current and various special
functions
Background illumination
Background illumination is switched off approximately 1 minute
after it has been activated.
Analog
DisplayLCD scale with bar graph or pointer,
ScalingWith 4 division lines each
Polarity displayWith automatic switching
Overflow displayWith the symbol
Measuring rate40 per second and display refresh
Digital
Display / Char. Height
Number of places 4½ place11,999 steps
Overflow display“0L” is displayed for ≥12,000 digits
Polarity display“–” (minus sign) is displayed
Measuring rate10 measurements per second;
Refresh Rate2 times per sec., every 500 ms
depending upon A.diSP parameter setting
1 bar/pointer corresponds to 500 digits at
the digital display
(U and I)
7-segment characters / 15 mm
if plus pole is connected to “⊥”
40 per second with Min/Max function
except with capacitance, frequency and
keying ratio measuring functions
Acoustic Signals
For voltageIntermittent signal at above 1000 V
For currentIntermittent signal at above 10 A
Continuous signal at above 16 A
Fuse for METRAHIT⏐X-TRA⏐OUTDOOR⏐TECH⏐PRO
FuseFF (UR) 10 A/1000 V AC/DC,
10 x 38 mm,
switching capacity:
30 kA at 1000 V AC/DC,
protects the current input socket in the
100 μA to 10 A ranges
Electrical Safety
per IEC 61010-1:2001/VDE 0411-1:2002
Safety class
Measuring categoryIIIIV
Operating voltage1000 V600 V
Fouling factor
Tes t vo l t ag e
II
2
6.7 kV~
GMC-I Gossen-Metrawatt GmbH59
Technical Data
Electromagnetic Compatibility (EMC)
Interference
emissionEN 61 326: May 2004, class B
Interference
immunityEN 61326: May 2004, appendix E
IEC 61000-4-2: Dec. 2001
Feature B
8 kV atmospheric discharge
4 kV contact discharge
IEC 61000-4-3: Dec. 2001
Feature A
3V/m
Ambient Conditions
Operating
temp. range T
A
Storage
temp. range−25 °C ... +70 °C (without batteries)
Relative humidityMax.75%, no condensation allowed
Elevationto 2000 m
Deployment
−10 °C ... +50 °C
Indoors; outdoors only within specified
ambient conditions
Mechanical Design
HousingImpact resistant plastic (ABS)
Dimensions200 x 87 x 45 mm
WeightApprox. 0.35 kg with batteries
ProtectionHousing: IP 52
Table Excerpt Regarding Significance of the IP Code
Disconnect the instrument from the measuring circuit before
opening the battery compartment lid or fuse cover in order
to replace batteries or fuses!
9.1 Displays – Error Messages
MessageFunctionMeaning
FUSE
0L
9.2 Batteries
Current measurement Blown fuse
In all operating modes Battery voltage has fallen below 1.8 V
MeasurementIndicates overflow
Note
Removing the Batteries During Periods of Non-Use
The integrated quartz movement draws power from the
batteries even when the instrument is switched off. It is
advisable to remove the batteries during long periods of
non-use for this reason (e.g. vacation). This prevents
excessive depletion of the battery, which may result in
damage under unfavorable conditions.
Note
Battery Replacement for METRAHIT⏐X-TRA⏐OUTDOOR
Stored measurement data are lost when the batteries are
replaced. In order to prevent data loss, it is advisable to
backup your data to a PC with the help of METRA⏐V
IEW soft-
ware before replacing the batteries.
The selected operating parameters remain in memory,
although date and time must be reentered.
Battery
The current battery charge level can be queried in the “1nfo”
menu:
MEASURE
Make sure that no battery leakage has occurred before initial
start-up, as well as after long periods of storage. Continue to
inspect the batteries for leakage at short, regular intervals.
If battery leakage has occurred, carefully and completely clean
the electrolyte from the instrument with a damp cloth, and replace
the batteries before using the instrument.
If the “” symbol appears at the display, the batteries should
be replaced as soon as possible. You can continue working with
the instrument, but reduced measuring accuracy may result.
The instrument requires two 1.5 V batteries in accordance with
IEC R 6 or IEC LR 6, or two equivalent rechargeable NiCd
batteries.
1nFobAtt
SETUP
FUNC
ENTER
:
2.75 V.
62GMC-I Gossen-Metrawatt GmbH
Maintenance – Calibration
Replacing the Batteries
!
Attention!
Disconnect the instrument from the measuring circuit before
opening the battery compartment lid in order to replace the
batteries.
➭ Set the instrument face down onto the working surface.
➭ Turn the slotted screw on the lid with the battery symbols
counterclockwise.
➭ Lift off the lid and remove the batteries from the battery
compartment.
➭ Insert two new 1.5 V mignon batteries into the battery
compartment, making sure that the plus and minus poles
match up with the provided polarity symbols.
➭ When replacing the battery compartment lid, insert the side
with the guide hooks first.
Tighten the screw by turning it clockwise.
➭ Please dispose of depleted batteries in accordance with
environmental protection regulations!
9.3Fuse (METRA
METRA
Testing the Fuse
The fuse is tested automatically:
• When the instrument is switched on with the rotary switch in
the A position
• When the instrument is already on and the rotary switch is
turned to the A position
HIT⏐X-
HIT⏐PRO only)
TRA
⏐
O
UTDOOR
, METRA
HIT⏐T
ECH
and
• In the active current measuring range when voltage is applied
If the fuse is blown or has not been inserted, “FuSE” appears at
the digital display. The fuse interrupts the current measuring
ranges. All other measuring ranges remain functional.
DC
fuse
Replacing the Fuse
If a fuse should blow, eliminate the cause of overload before
placing the instrument back into service!
!
Attention!
Disconnect the instrument from the measuring circuit before
opening the fuse cover in order to replace the fuse!
➭ Set the instrument face down onto the working surface.
➭ Turn the slotted screw on the cover with the fuse symbol
counterclockwise.
➭ Lift off the cover and pry the fuse out using the flat side of the
fuse cover.
➭ Insert a new fuse. Make sure that the fuse is centered, i.e.
between the tabs at the sides.
➭ When replacing the fuse cover, insert the side with the guide
hooks first.
Tighten the screw by turning it clockwise.
➭ Dispose of the blown fuse with the trash.
A
Blown Fuse
GMC-I Gossen-Metrawatt GmbH63
Maintenance – Calibration
!
Attention!
Use specified fuses only!
If fuses with other blowing characteristics, other current
ratings or other breaking capacities are used, the operator
is placed in danger, and protective diodes, resistors and
other components may be damaged.
The use of repaired fuses or short-circuiting the fuse holder
is prohibited.
Note
Testing the Fuse with the Instrument Switched On
After inserting the fuse with the instrument switched on, the
instrument must be switched off briefly and then switched
back on, or briefly switched to a non-current measuring
range and then back to the “A” measuring range.
If contact is poor or the fuse is blown, FUSE appears at the
display.
9.4 Housing Maintenance
No special maintenance is required for the housing. Keep outside
surfaces clean. Use a slightly dampened cloth for cleaning. Avoid
the use of cleansers, abrasives or solvents.
9.5 Return and Environmentally Sound Disposal
The instrument is a category 9 product (monitoring and control
instrument) in accordance with ElektroG: German electrical and
electronic device law). This device is not subject to the RoHS
directive.
We identify our electrical and electronic devices (as of
August 2005) in accordance with WEEE 2002/96/EC
and ElektroG with the symbol shown at the right per
DIN EN 50419.
These devices may not be disposed of with the trash.
Please contact our service department regarding the return of old
devices (see page 4).
9.6 Recalibration Service
We calibrate and recalibrate all instruments supplied by GMC-I
Gossen-Metrawatt GmbH, as well as other manufacturers, at our
service center, for example after one year within the framework of
your test equipment monitoring program, as well as prior to use
etc. (see address on page 4).
64GMC-I Gossen-Metrawatt GmbH
9.7 Manufacturer’s Guarantee
All METRAHIT⏐ measuring and calibration instruments are
guaranteed for a period of 3 years after date of shipment. The
manufacturer’s guarantee covers materials and workmanship.
Damages resulting from use for any other than the intended
purpose, as well as any and all consequential damages, are
excluded.
Calibration is guaranteed for a period of 12 months.
Maintenance – Calibration
GMC-I Gossen-Metrawatt GmbH65
Accessories
10Accessories
10.1 General
The extensive accessories available for our measuring
instruments are checked for compliance with currently valid safety
regulations at regular intervals, and are expanded as required for
new applications. Currently up-to-date accessories which are
suitable for your measuring instrument are listed at the following
web address along with photo, order number, description and,
depending upon the scope of the respective accessory, data
sheet and operating instructions: www.gossenmetrawatt.de
(→ Measuring Technology - Portable → Digital Multimeters →
METRAHIT⏐ ... → Accessories).
10.2 Technical Data for Measurement Cables
(included: KS17-2 safety cable set)
Electrical Safety
Maximum rated voltage
Measuring category 1000 V CAT III, 600 V CAT IV
Maximum
rated current16 A
Ambient Conditions (EN 61010-031)
Temperature–20 °C ... + 50 °C
Relative humidity50 to 80%
Fouling factor2
10.3 NA X-TRA Power Pack (not included)
Use only the power pack from GMC-I Gossen-Metrawatt GmbH
in combination with your instrument. This assures operator safety
by means of an extremely well insulated cable, and safe electrical
isolation (nominal secondary ratings: 5 V / 600 mA). Installed
batteries are disconnected electronically if the power pack is
used, and need not be removed from the instrument.
66GMC-I Gossen-Metrawatt GmbH
10.4 Interface Accessories for METRA HIT⏐X-TRA⏐OUTDOOR (not
included)
USB X-TRA Bidirectional Interface Adapter
This adapter makes it possible to connect
METRAHIT⏐X-
TRA⏐OUTDOOR multimeters which are equipped with a
serial IR interface to the USB port at a PC. The adapter allows for
data transmission
between the multimeter and the PC.
METRAwin 10 PC Analysis Software
METRAwin 10 PC software is a multilingual, measurement data log-
ging program for recording, visualizing, evaluating and
documenting measured values from METRAHIT⏐ multimeters.
The following conditions must be fulfilled in order to allow for use
of METRAwin 10 :
Hardware
:
– IBM compatible Windows PC, Pentium processor with
200 MHz or better and at least 64 MB RAM
– SVGA monitor with at least 1024 x 768 pixels
– Hard disk with at least 40 MB available memory
–CD-ROM drive
– Microsoft compatible mouse
– Windows-supported printer
– 1 USB port for using USB X-TRA