Gossen Metrawatt SECU-cal 10 Operating Instructions

Operating Instructions
SECU-cal 10
Calibration Adapter for Testers per DIN VDE 0701-0702/ EN 62353 (VDE 751-1)
3-349-169-15
7/2.13
Function selector
1mA terminal
Reference terminals
Mains indicator lamp
Polarity display
for test instrument probe for measurement of R
PE
may not light up
Reference values for equivalent leakage current measurement I
EA
during
Tests with line voltage! – Leakage current measurement – Residual current measurement
with the help of an ammeter, see chapter 5 regarding connection
with max. 1 A test current
measurement!
Lateral imprint
showing the reference values
of test resistances R
PE
switch
Reference values for measurement of R
ISO
and equivalent leakage current I
EA
The indicated current values of 30 mA and 3 mA for switch positions 7 M and 70 M are merely guidance values. Usually, the capacitance of the connector cables is suffi­ciniently high to ensure that the current values actually mea­sured exceed considerably that level and can therefore not be used as reference values. This applies particularly for the test instruments of the SECUT­EST
2 GMC-I Messtechnik GmbH
®
... series.
Contents Page
1 Applications ..................................................................4
2 Safety Features and Precautions ..................................4
3 Connecting the Calibration Adapter to the Test
Instrument ....................................................................7
3.1 Tests without Line Voltage .......................................................... 7
3.2 Tests with Line Voltage ..............................................................7
4 Performing Tests without Line Voltage .........................8
4.1 Checking Display Values for Protective Conductor Resistance
Measurement ............................................................................ 8
4.2 Checking Display Values for Insulation Resistance Measurement ..8
4.3 Checking Display Values for Equivalent Leakage Current
Measurement ............................................................................ 9
4.4 Checking Display Values for Equivalent Device Leakage Current
Measurement per DIN EN 62353 (VDE 0751-1) ......................... 10
4.5 Checking Display Values for Equivalent Patient Leakage
Current Measurement per DIN EN 62353 (VDE 0751-1) ............. 10
5 Performing Tests with Line Voltage ............................11
5.1 Checking Display Values for Leakage Current Measurement ....... 11
5.2 Checking Display Values for Patient Leakage Current
Measurement .......................................................................... 12
6 Characteristic Values .................................................. 13
6.1 Nominal Range of Use .............................................................. 13
6.2 Ambient Conditions .................................................................. 13
6.3 Electrical Safety ....................................................................... 13
6.4 Mechanical Design ................................................................... 13
7 Maintenance ...............................................................14
7.1 Housing Maintenance ............................................................... 14
7.2 Device Return and Environmentally Compatible Disposal ............14
8 Repair and Replacement Parts Service
Calibration Center and Rental Instrument Service .....15
9 Product Support ..........................................................15
GMC-I Messtechnik GmbH 3

1 Applications

Attention!
!
The adapter is used for testing measuring accuracy of test instruments in accordance with DIN VDE 0701-0702 and DIN EN 62353 (VDE 0751-1). As a rule, these instruments must be tested once each year as set forth by accident pre­vention regulation BGV A3 (formerly BGV A2 and/or VBG4), as well as for certification in accordance with the ISO 9000 qual­ity standard.
All measuring ranges for tests required in accordance with DIN VDE 0701-0702 such as protective conductor resistance, insulation resistance, equivalent leakage current and residual and/or contact current must be tested.

2 Safety Features and Precautions

This instrument fulfills the requirements of the applicable Euro­pean and national EC guidelines. We confirm this with the CE marking. The relevant declaration of conformity can be obtained from GMC-I Messtechnik GmbH. The SECU-cal 10 calibration adapter is manufactured and tested in accordance with safety regulations IEC 61010-1 / EN 61010-1 / VDE 0411-1.
If used for its intended purpose, safety of the user and of the device is assured.
Read the operating instructions completely and carefully before using the adapter, and follow all instructions included therein. Observe the operating instructions included with the test instrument to be tested as well.
The calibration adapter may only be used for testing test instruments in accordance with DIN VDE 0404! Under no circumstances may the adapter be used in electrical systems!
4 GMC-I Messtechnik GmbH
Attention!
Before plugging the calibration adapter into a live
Attention!
Attention!
!!!
outlet (test socket or mains outlet) for the perfor­mance of tests with line voltage, set the rotary switch at the calibration adapter to the DI/I position.
Non-observance may result in:
Destruction of the adapter
Charging of the protective conductor with impermissible
fault currents of up to 8.25 mA
Tripping of 10 mA RCCBs
If the “NETZ/MAINS” and “POLARITÄT/POLARITY” lamps light up simultaneously during the test with mains voltage in rotary switch position DI/I, the polarity of the calibration adapter must be reversed before measurement is performed. Only the “NETZ/MAINS” lamp may remain illuminated. Non-obser­vance may result in erroneous measurements.
Only test instruments with a protective conductor test current of no greater than 1 A may be connected to the calibration adapter. Greater current values result in destruction of the adapter.
The device is not equipped with overcurrent or excessive temperature protection. Under no circumstances may the device’s load capacities be exceeded, because this may damage the device or reduce its level of accuracy.
GMC-I Messtechnik GmbH 5
Visual Inspection of Test Instruments
!
Perform a visual inspection of test instruments and their con­nector cables before connecting them to the calibration adapter. Damaged test instruments must first be repaired.
The calibration adapter may not be used:
with open housing
if visible damage is apparent
if the connection cable is damaged
if it no longer functions flawlessly
if the safety sockets are damaged
after excessive stress, i.e. if the load capacities specified in the technical data have been exceeded
after lengthy periods of storage under unfavorable condi­tions (e.g. humidity, dust, temperature)
The calibration adapter may only be repaired by the manufac­turer. Observance of technical measuring and safety require­ments cannot otherwise be assured.
Meanings of symbols on the device
The symbols on the device have the following meanings:
Indicates EC conformity
Warning concerning a source of danger (Attention, observe documentation!)
This device may not be disposed of with the trash. Further information regarding the WEEE mark can be accessed on the Internet at www.gossenmetrawatt.com by entering the search term ’WEEE’.
6 GMC-I Messtechnik GmbH
3 Connecting the Calibration Adapter to the Test
Instrument

3.1 Tests without Line Voltage

Make sure that the test socket is voltage-free
(e.g. set Mains VDE (Netz-VDE) switch to VDE position for SECUTEST
The earthing contact outlet of the calibration adapter may
only be plugged into the earthing contact outlet of the test
instrument, identified as test socket. Reversal of L and N at the outlet has no effect on measure­ment results.

3.2 Tests with Line Voltage

Before connecting the calibration adapter to the test
socket: Set the rotary switch to the DI/I position. The rotary switch must also be set to the DI/I position for auto-
matic test instruments, and for test panels with VDE MAINS selector switches before activating the test function (i.e. before starting the function test). If the “NETZ/MAINS” and “POLARITÄT/POLARITY” lamps light up simultaneously, mains plug polarity must be reversed for the test instrument or the calibration adapter before measure­ment is performed.
Only the “NETZ/MAINS” lamp may be illuminated during mea­surement.
®
11P, 15P or 21F workshop test panels).
GMC-I Messtechnik GmbH 7

4 Performing Tests without Line Voltage

Attention!
!
Connect the test instrument to mains power.Make sure that the test socket at the test instrument is
voltage-free.
Plug the earthing contact plug of the calibration adapter
into the test socket at the test instrument.

4.1 Checking Display Values for Protective Conductor Resistance Measurement

Use test instruments with test current of less than 1 A only. The reference resistors are destroyed at values of greater than 1 A.
Connect the probe from the test instrument to the “0.3 ”
or “1.0 ”socket at the calibration adapter.
Start “protective conductor measurement” at the test in-
strument.
The value displayed at the test instrument must lie within the test instrument’s specified operating error tolerance for the selected test type, plus calibration adapter error.
The reference values of the test resistances R on the front side of the calibration adapter and in the test report. Please use these values for assessing your DUT. Calibration adapter intrinsic error for protective conductor resistance measurement: 1%

4.2 Checking Display Values for Insulation Resistance Measurement

Start “insulation measurement” at the test instrument.Set the selector switch at the calibration adapter to either
70 or 0.1 M.
The value displayed at the test instrument must lie within the test instrument’s specified operating error tolerance for the selected test type, plus calibration adapter error.
Calibration adapter intrinsic error for various insulation resistance values
Value in M 70 0.1 7
Intrinsic error as % 10.5
are indicated
PE
8 GMC-I Messtechnik GmbH
4.3 Checking Display Values for Equivalent Leakage Current Mea-
Note
surement
Start “equivalent leakage current measurement” at the
test instrument.
Set the selector switch at the calibration adapter to either
3.5 or 7 mA.
The value displayed at the test instrument must lie within the test instrument’s specified operating error tolerance for the selected test type, plus calibration adapter error.
I
[mA] = 230 V
Display
Rxrepresents the resistors integrated into the SECU-cal 10. Any possi-
ble phase error is avoided through the use of ohmic resistors.
represents the internal resistance of the test instrument during equiv-
R
i
alent leakage current measurement. It amounts to typ. 2 k, accord­ing to standard.
Rx + R
i
Calibration adapter intrinsic error for equivalent leakage cur­rent measurement: 1 %
Before introduction of the DIN VDE 0701:2000-09 standard, the display value was calculated according to the following formula: I
[mA] = 230 x 1.06
Display
Rx + 2 k, i.e. the display value included an allowance of 6%. Veteran instruments therefore show an equivalent leakage current increased by 6 %. We recommend upgrading such instruments to the current version as soon as possible.
GMC-I Messtechnik GmbH 9

4.4 Checking Display Values for Equivalent Device Leakage Current Measurement per DIN EN 62353 (VDE 0751-1)

Start “equivalent leakage current measurement
(Ri = 1 k)” at the test instrument.
Connect the probe to the 0.3  socket.
In accordance with DIN EN 62353 (VDE 0751-1), the pro­tective conductor at the test socket is not connected. The probe functions as a return conductor.
The value displayed at the test instrument must lie within the test instrument’s specified operating error tolerance for the selected test type, plus calibration adapter error.
I
[mA] = 230 V
Display
Rxrepresents the resistors integrated into the SECU-cal 10. Any possi-
ble phase error is avoided through the use of ohmic resistors.
R
represents the internal resistance of the test instrument during equiv-
i
alent device leakage current measurement. It amounts to typ. 1 k, according to standard.
Rx + R
i
Nominal Values Switch Positions
70 M 3.29 A 3.29 A
7 M 32.8 A 32.8 A 2 M 115 A 115 A 1 M 0.23 mA 0.23 mA
0.5 M 0.46 mA 0.46 mA
0.1 M 2.25 mA 2.28 mA
3.5 mA 3.5 mA 3.55 mA 7 mA 7.0 mA 7.22 mA
Equivalent Leakage Current
Equivalent Device Leakage Current

4.5 Checking Display Values for Equivalent Patient Leakage Current Measurement per DIN EN 62353 (VDE 0751-1)

Plug the calibration adapter into the test socket at the test instrument. Plug the probe into sockets 4 and 5 at the test instrument. The probe handle is left open and may not be touched (hum). Connect one of the applied parts sockets (A through K) to the 1 mA socket at the SECU-cal 10. The included quick clip and a 2 mm cable can be used to this end. Start “equivalent patient leakage current measurement” at the
10 GMC-I Messtechnik GmbH
test instrument. The display value must lie within a range of
Attention!
Note
!
0.98 mA and 1.01 mA plus the measuring uncertainty of the test instrument. Other values cannot be tested with the SECU­cal 10. In order to test all of the sockets (A through K) for cor­rect functioning, connect each of these sockets with the 1 mA socket at the SECU-cal 10, one after the other. The display value must lie within the above specified range for each of the sockets.

5 Performing Tests with Line Voltage

Set the rotary switch at the calibration adapter to the DI/I position before plugging the calibration adapter into the test socket.
The following test with line voltage must be performed with an additional ammeter (calibrated multimeter, 1 mA respectively 10 mA AC range, e.g. METRAHIT X-TRA).

5.1 Checking Display Values for Leakage Current Measurement

The “POLARITÄT/POLARITY” lamp may not light up during measurement.
Start the requested leakage current measurement with
mains voltage at the test instrument.
Connect the “contact and probe current measurement”
socket at the test instrument to the “1 mA” safety socket at the calibration adapter via a series connected multime-
ter with the help of measurement cables. The values displayed at the multimeter and the test instrument must be in compliance by taking device tolerances into account.
Line voltage tolerances influence measurement results.
GMC-I Messtechnik GmbH 11

5.2 Checking Display Values for Patient Leakage Current Measurement

This test can be performed with all test instruments of the SECUTEST/SECULIFE series with sockets A through K for applied parts.
Please proceed as follows to calibrate the patient leakage cur­rent with SECU-cal 10:
Connect SECU-cal 10 with the test socket.Select function „I-leakage“ at the SECUTEST (via switch po-
sition or menu), and then patient leakage current.
Start measurement.Connect the applied part A to the 0.3 Ohm socket of
SECU-cal 10 via a series connected multimeter with the help of measurement cables.
Select „SFC Prot. Cond. interrupted“.Set the requested leakage current at the SECU-cal 10 test
instrument (e. g. 115 μA).
The values displayed at the multimeter and the test instrument must be in compliance by taking device tolerances into account.
12 GMC-I Messtechnik GmbH

6 Characteristic Values

6.1 Nominal Range of Use

Maximum Voltage For the measurement of
Maximum Current For the measurement of protective
AC frequency: 50 Hz 60 Hz, sinusoidal

6.2 Ambient Conditions

Operating Temperature 0 C  +40 C Storage Temperature –20 C  +60C

6.3 Electrical Safety

Safety Class I per IEC 61010-1 Operating Voltage 300 V Measuring Category I for insulation resistance
Pollution degree 2 EMC IEC/EN 61326

6.4 Mechanical Design

Protection housing: IP 40, terminals: IP 20 Extract from table on the meaning of IP codes
IP XY (1
Protection against
st
digit X)
foreign object entry
0 not protected 0 not protected
2 12.5 mm dia. 2
4 1.0 mm dia. 4 splashing water
Dimensions Weight approx. 0.26 kg
Insulation resistance: 600 V DC Equivalent leakage/equivalent device leakage current: 250 V AC Equivalent patient leakage current:
250 V AC
Leakage current: 250 V AC
conductor resistance:
1ADC/AC Insulation resistance: 2 mA Equivalent leakage current:
3.5 mA AC
II for all other measurements
IP XY (2nd digit Y)
Protection against the penetration of water
vertically falling drops
with enclosure tilted 15
125 mm x 66 mm x 41 mm (L x W x H)
eff
eff
GMC-I Messtechnik GmbH 13

7 Maintenance

Annual calibration, performed by the manufacturer, is recom­mended.

7.1 Housing Maintenance

No special maintenance is required for the housing. Keep out­side surfaces clean. Use a slightly dampened cloth for clean­ing. Avoid the use of solvents, cleansers and abrasives.

7.2 Device Return and Environmentally Compatible Disposal

The instrument is a category 9 product (monitoring and control instrument) in accordance with ElektroG (German Electrical and Electronic Device Law). This device is not subject to the RoHS directive.
We identify our electrical and electronic devices (as of August 2005) in accordance with WEEE 2002/96/EG and ElektroG with the symbol shown to the right per DIN EN 50419. These devices may not be disposed of with the trash. Please contact our service department regarding the return of old devices.
14 GMC-I Messtechnik GmbH
8 Repair and Replacement Parts Service
Calibration Center* and Rental Instrument Service
When you need service, please contact:
GMC-I Service GmbH
Service Center
Thomas-Mann-Strasse 20 90471 Nürnberg • Germany Phone +49 911 817718-0 Fax +49 911 817718-253 E-mail service@gossenmetrawatt.com www.gmci-service.com
This address is only valid in Germany. Please contact our representatives or subsidiaries for service in other countries.
* DAkkS Calibration Laboratory
for Electrical Quantities D-K-15080-01-01 accredited per DIN EN ISO/IEC 17025:2005
Accredited measured quantities: direct voltage, direct current values, DC resistance, alternating voltage, alternating current values, AC active power, AC apparent power, DC power, capacitance, frequency and temperature

9 Product Support

When you need support, please contact:
GMC-I Messtechnik GmbH
Product Support Hotline
Phone +49 911 8602-0 Fax +49 911 8602-709 E-Mail support@gossenmetrawatt.com
GMC-I Messtechnik GmbH 15
Edited in Germany • Subject to change without notice • A pdf version is available on the internet
GMC-I Messtechnik GmbH Südwestpark 15 90449 Nürnberg •
Germany
Phone +49 911 8602-111 Fax +49 911 8602-777 E-Mail
info@gossenmetrawatt.com
www.gossenmetrawatt.com
Loading...