Gossen Metrawatt METRISO XTRA Operating Instructions

Operating Instructions
METRISO XTRA
High-Precision Insulation, Low Resistance and Voltage Measurement Instrument
3-349-818-03
2/12.17
Red LED: limit value exceeded
Red LED:
– Interference voltage with device off
Setup Menu
Function Selector Switch
Slot for Carrying Strap
Slot for Carrying Strap
Green LED: limit value adhered to
– High test voltage at the measuring terminals
MEM: Key for memory
functions
ESC: Exit submenu /
activate instrument standby state
Control panel
Fixed Function Keys
LEDs section 4.3
CONTIN: Continuous
measurement of Rins and R (kΩ)
HELP: Access context
sensitive help
START: Start measurement
Softkeys
• Parameter selection
• Limit value specification
• Entry functions
• Memory functions
METRISO XTRA
during Riso/Rins
2 GMC-I Messtechnik GmbH
Battery Holder
Battery Compartment Lid
Battery Holder Contacts
Battery
Contact Spring
User Interface
Batteries, Fuses
Measuring
!
RS 232
+
COM
+
COM
SHIELD
SHIELD
Option KS-C (Z541F)
Option Z550A*
Charger Socket, Interfaces
These connections are located under a protective rubber flap.
a b
Replacement
Fuse
Socket for Z502R charger Caution!
Make sure that no batteries are inserted before connecting the charger.
The test instrument must remain off during the charging process.
Fuse
Inserting the Battery Holder (side view)
USB slave for PC connection
Port for connecting Barcode/RFID reader
GUARD
* Regarding Z550A:
Test Probe with measurement key remote control and LED for illuminating the measuring point: Do not use when measuring PI and DAR.
Compartment Lid
Battery
Compartment
Connections
Test Resistor 10 MΩ
GMC-I Messtechnik GmbH 3
Overview of Device Settings and Measuring Functions
Battery level
Bar graph display
Measurement
Upper range limit
Measuring
Parameter
Display panel
Save
Battery full
Battery OK
Battery weak
Battery (nearly)
Battery Level Indicator
BAT
BAT
BAT
BAT
depleted U < 8.5 V
Measuring Status
Measurement in progress: Bars moves from left
Measurement on pause: Bars are static.
to right.
||||||||||
||||||||||
Polarity reversal‚ L-PE / N-PE for documentation
setting
value
quantities
indicator
in progress
Memory Occupancy Display
MEM
Memory half full
MEM
Memory full > transfer data to PC!
These operating instructions describe a tester with software version SW-VERSION (SW1) 01.02.00.
!
Relative to Rotary Switch Setting
Switch setting,
see as of
RISO RINS
page 15
kΩ
page 20
RLO
page 21
V
page 14
SETUP
Picto­graph
Device settings, measuring functions
RINS Insulation resistance
U Voltage at the test probes
U
fix Fixed test voltage,
N
Uvariable Variable test voltage, limit value:
Uramp Variable test voltage,
PI Polarization index DAR Absorption index R Resistance measurement
RLO Low-resistance with polarity reversal RLO+, RLO–
R
OFFSET Offset resistance
U Voltage measurement
Tests: LCD pixel display
Battery test Ubat
limit values per VDE 0100
1 MOhm (for go/no-go display while saving the measured value)
ramp: triggering/breakdown voltage
Low-resistance, single-pole
LCD pixel display
All pixels off Green limit LED All pixels on Red limit LED Acoustic signal
Date/time CULT Language (D, GB)
Brightness
page 8
SET on: On-time for
Contrast Default settings
Device type Serial number Software version Hardware version Calibration date Adjustment date
LCD and tester
Scope of delivery:
1 Insulation and resistance measuring instrument 1 DAkkS calibration certificate 1 Set batteries (8 pieces in battery holder) 1Carrying strap 1 Alligator clip 1 KS17-4 cable set 1USB cable 1 Condensed operating instructions 1 Supplement Safety Information – Detailed operating instructions for download from our website
at www.gossenmetrawatt.com
4 GMC-I Messtechnik GmbH
Table of Contents Page Page
1 Applications ......................................................................6
1.1 Optional Z550A Remote Control .................................................. 6
1.2 Overview of Measuring Instrument Performance
Features ...................................................................................... 6
2 Safety Features and Precautions .....................................7
3 Initial Start-Up .................................................................. 7
3.1 Battery test ................................................................................. 7
3.2 Installing or Replacing Batteries ................................................. 7
3.3 Charging the Batteries in the Tester ........................................... 7
3.4 Device Settings – SETUP ........................................................... 8
4 General Operation ...........................................................11
4.1 Connecting the Instrument ....................................................... 11
4.2 Switching On, Monitoring and Switching Off ............................ 11
4.3 Optical Indicators ..................................................................... 11
4.4 Measurement Value Display and Memory ................................ 12
4.5 Help Function ............................................................................ 12
4.6 Setting Measuring Parameters
using Insulation Resistance Measurement as an Example ....... 12
4.7 Specifying Nominal Voltage for Uvar and Uramp ...................... 13
5 Measuring Direct Alternating Voltage ............................. 14
6 Measuring Insulation Resistance ...................................15
6.1 Measuring with Constant Test Voltage
and Nominal Value Selection via Rotary Switch Position ......... 16
6.2 Measurement with Constant Test Voltage
and Variably Adjustable Nominal Value .................................... 16
6.3 Measurement with Rising Test Voltage (ramp function)
and Variably Adjustable Final Value .......................................... 17
6.4 Insulation Resistance Measurement – Special Conditions ....... 17
6.4.1 Measurements with the Guard Cable ............................................ 17
6.5 Discharging the Device Under Test ........................................... 18
6.6 Evaluation of Measured Values ................................................. 18
6.7 Polarization Index Measurement .............................................. 18
6.7.1 Absorption Index (DAR) – DC Charging Test .................................. 19
11 Characteristic Values ...................................................... 29
12 Maintenance ...................................................................31
12.1 Firmware Revision and Calibration Information ....................... 31
12.2 Rechargeable Battery Operation and Charging ........................ 31
12.2.1 Charging Procedure with the Z502R Charger (accessory) .............. 31
12.3 Fuses ........................................................................................ 31
12.3.1 Melting Fuse ............................................................................... 31
12.3.2 Electronic Fuse ........................................................................... 31
12.4 Housing ..................................................................................... 32
12.5 Return and Environmentally Sound Disposal ............................ 32
13 Appendix .........................................................................32
13.1 Sample Connection Layouts for Insulation Resistance Measure-
ment ......................................................................................... 32
13.2 Error Messages ........................................................................ 34
13.3 Attaching the Test Probe Holder to the Carrying Strap ............. 36
13.4 Technical Data for Measurement Cables
(scope of delivery: KS17-4 safety cable set) ............................ 36
13.5 Optional Accessories (not included) ......................................... 37
14 Repair and Replacement Parts Service
Calibration Center and Rental Instrument
Service ............................................................................38
15 Recalibration ..................................................................38
16 Product Support .............................................................. 38
7 Test Resistor for Insulation Measurement for Checking the
Insulation Measuring Instrument .................................... 19
8 Measuring Resistance – kOhm Function ........................ 20
9 Measuring Low-Value Resistance of up to 10 Ohm
(protective conductor and equipotential bonding conductor) ... 21
10 Database ........................................................................24
10.1 Creating Distributor Structures, General ................................... 24
10.2 Transferring Distributor Structures ........................................... 24
10.3 Creating a Distributor Structure in the Test Instrument ........... 24
10.3.1 Creating Structures (example for electrical circuit) ......................... 25
10.3.2 Searching for Structural Elements ................................................26
10.4 Saving Data and Generating Reports ........................................ 27
10.4.1 Use of Barcode Scanners and RFID Readers ................................. 28
GMC-I Messtechnik GmbH 5

1 Applications

This instrument fulfills all requirements of applicable European and national EC directives. We confirm this with the CE mark. The rel­evant declaration of conformity can be obtained from GMC-I Messtechnik GmbH.
The METRISO XTRA insulation and resistance measuring instru­ment allows for quick and efficient testing of protective measures in accordance with DIN VDE 0100, ÖVE-EN 1 (Austria), SEV 1000 (Switzerland), and regulations specific to other countries as well.
The device is equipped with a microprocessor and complies with IEC/EN 61557 / VDE 0413 regulations:
Part 1: General requirements Part 2: Insulation resistance Part 4: Resistance of earth connection, protective conductors
and equipotential bonding
Part 10: Electrical safety in low-voltage systems up to 1000 V
AC and 1500 V DC – Equipment for testing, measuring or monitoring of protective measures
As well as requirements per VDE 0701-0702: Repair, modification and testing of electrical devices
The test instrument is especially well suited for:
•Systems setup
• Initial start-up
• Periodic testing
• Troubleshooting in electrical systems
The following measurements and tests can be performed with the insulation measuring instrument:
• Insulation resistance
• Low-resistance
• Voltage
The following can also be tested by using a shielded measure­ment cable:
• Floor covering electrostatic discharge capability

1.2 Overview of Measuring Instrument Performance Features

METRISO XTRA
Article No. M550S
Measurement
U = 50, 100, 250, 500, 1000 V
R
INS
Uvar = 50 ... 1000 V
R
INS
Uramp (U) = 100 ... 1000 V
R
INS
PI
DAR R 10 ... 10 kΩ
R
LO
U 0 ... 1000 V
Display Functions
Backlit display
Limit value LED (green/red) for: limit values per VDE 0100 in addition to acoustic signal
URamp limit value LED for:
indicating ramp characteristics
LED for dangerous contact voltage
(when switched off)
Battery level display
Special Functions
Discharge capacitive devices under test
Safety shutdown (UBatt < 8 V)
Data storage at the instrument
ETC software for data acquisition, data management and reporting
Features
Measuring categories: CAT II 1000 V / CAT II I 600 V / CAT IV 300 V Test resistance: 10 MΩ
Connections: charging socket, USB port (slave), RS 232 port
DAkkS calibration certificate
0.01 ... 10 Ω
R
INS RLO
R
INS

1.1 Optional Z550A Remote Control

The optional remote control with triggering key on the test probe and a key for illuminating the measuring point makes it possible to use the measuring instrument at difficult to access places.
For safety reasons, a predefined voltage is only applied to the test probes for as long as the triggering key is held depressed.
The start key on the test instrument should be used instead of remote control for triggering polarization and absorption index measurements, because these two measurements involve com­plete measuring cycles.
6 GMC-I Messtechnik GmbH

2 Safety Features and Precautions

Attention!
!
Attention!
!
Attention!
!
Attention!
!
!
XY123
2012-06
D-K
15080-01-01
Consecutive number
Registration number Date of calibration (year – month)
Deutsche Akkreditierungsstelle GmbH – calibration lab
The electronic measuring and test instrument is manufactured and tested in accordance with safety regulations IEC/EN 61010­1/VDE 0411-1 and EN 61 557. safety of the operator, as well as that of the instrument, is assured.
Read the operating instructions thoroughly and carefully before using your instrument. Follow all instructions contained therein.
The measuring and test instrument may not be placed into service:
• If the battery compartment lid has been removed
• If external damage is apparent
• If connector cable or measuring adapters are damaged
•If the instrument no longer functions flawlessly
• After extraordinary damage due to transport
• After a long period of storage under unfavorable conditions (e.g. humidity, dust or extreme temperature).
Opening of Equipment / Repair
The equipment may be opened only by authorized service per­sonnel to ensure the safe and correct operation of the equipment and to keep the warranty valid.
Even original spare parts may be installed only by authorized ser­vice personnel.
In case the equipment was opened by unauthorized personnel, no warranty regarding personal safety, measurement accuracy, conformity with applicable safety measures or any consequential damage is granted by the manufacturer.
Data Backup
We advise you to regularly transfer your stored data to a PC in order to prevent potential loss of data in the test instrument.
We assume no responsibility for any data loss. We recommend ETC software (Electric Testing Center) for back-
ing up, processing and managing data.
Meaning of Symbols on the Instrument
Warning concerning a point of danger (Attention, observe documentation!)
Protection class II device
When used for its intended purpose,
If battery voltage has fallen below the allowable lower limit, the pictograph shown at the right appears. The instrument does not function if the batteries have been depleted excessively, and no display appears.

3.2 Installing or Replacing Batteries

New batteries must be inserted for initial start-up, or if only one filled segment remains in the battery symbol.
Before opening the battery compartment (see page 5 for location), disconnect the instrument from the measuring circuit (mains) at all poles.
Eight 1.5 V size AA batteries in accordance with IEC LR 6 are required for operation of the insulation measuring instrument. Use new alkaline manganese batteries only.
Rechargeable NiCd or NiMH batteries may also be used. These can be charged externally or by connecting the Z502R charger to the test instrument. We recommend rechargeable NiMH batteries.
Always replace batteries in complete sets. Dispose of batteries in an environmentally sound fashion.
Loosen both slotted screws for the battery compartment lid
on the back, and remove the lid.
Remove the battery holder and insert eight 1.5 V size AA bat-
teries with correct polarity in accordance with the symbols.
Make sure that all of the batteries are inserted with correct polarity. If just one battery is inserted with reversed polar­ity, it will not be recognized by the instrument and may result in leakage from the batteries.
Push the battery holder into the battery compartment such
that the battery holder’s contacts touch the contact springs at the bottom of the battery compartment (see drawing on page 3). If the battery holder is not inserted as specified, the instrument cannot be supplied with power.
Replace the battery compartment lid and retighten the
screws.
Charging socket for extra-low direct voltage (Z502R charger)
Attention! Only rechargeable batteries may be inserted when the char­ger is connected.
EC mark of conformity
The device and included batteries may not be dis­posed of with the trash. Further information regarding the WEEE mark can be accessed on the Internet at www.gossenmetrawatt.com by entering the search term “WEEE”.
Calibration Seal (blue seal):
See also “Recalibration” on page 38.

3 Initial Start-Up

3.1 Battery test

Four different battery symbols, ranging from fully depleted to fully charged, continuously indicate the momentary charge level in the upper left-hand corner of the display
GMC-I Messtechnik GmbH 7
The instrument may only be placed into service if the battery compartment lid is securely fastened!

3.3 Charging the Batteries in the Tester

Use only the Z502R charger (available as an accessory) to charge batteries which have already been inserted into the test instrument.
Make sure that the following conditions have been fulfilled be­fore connecting the charger to the charging socket:
– Rechargeable batteries have been installed with
correct polarity (not standard batteries)
– The test instrument has been disconnected from the
measuring circuit at all poles
– The instrument must remain off during charging.
Refer to section 12.2.1 with regard to charging batteries which have been inserted into the tester.
If the batteries or the battery pack have not been used or recharged for a lengthy period of time (> 1 month), thus resulting in excessive depletion:
Observe the charging sequence (indicated by the LED at the charger) and initiate a second charging sequence if necessary (disconnect the charger from the mains and from the test instru­ment to this end, and then reconnect it).

3.4 Device Settings – SETUP

SETUP
LED and LCD test menu
Battery test menu
Brightness/contrast menu
Software revision level Calibration date
Display: date / time
Display: automatic shutdown
Display: automatic shutdown
of display illumination after 30 s.
of the tester after 120 s. Time, language, profiles
1
2
3
4
0b
0a
0
Return to main menu
Limit LED: test red
Limit LED: test green
Cell test
Inverse cell test
Hide all pixels
Show all pixels
Acoustic signal test
1
Return to main menu
Increase brightness
Decrease brightness
Increase contrast
Decrease contrast
Set time
Default settings
User interface language
3
3a 3b
3c
3d
Set date
On-time for display illumination / tester
0b
Return to submenu
0a
Display Illumination On-time
Brightness and Contrast Settings Time, On-Time and Default Settings
Menu Selections for Operating Parameters
LED Tests LCD and Acoustic Signal Tests
Test Instrument On-Time
Automatic shutdown, deactivated = OFF
8 GMC-I Messtechnik GmbH
SETUP
LED and LCD test menu
Battery test menu
Brightness/contrast menu
Software revision level Calibration date
Display: date / time
Display: automatic shutdown
Display: automatic shutdown
of display illumination after 15 s.
of the tester after 45 s.
Time, language, profiles
1
2
3
4
0b
0a
0
Return to main menu
Increase brightness
Decrease brightness
Increase contrast
Decrease contrast
Set time
Default settings
User interface language
3
3a 3b
3c
3d
Set date
On-time for display illumination / tester
Set time
Menu Selection for Operating Parameters
Adjusting Brightness and Contrast Set Time, Language, Profiles, Acoustic Signal
Set date
Select time/date
Increase
Increase
hours
Activate settings
minutes
3a
Increase seconds
Return to submenu
Decrease
Decrease
hours
minutes
Decrease
seconds
Select time/date
Increase
Increase
day
Activate settings
month
3b
Increase year
Return to submenu
Decrease
Decrease
day
month
Decrease
year
GMC-I Messtechnik GmbH 9
Significance of Individual Parameters
Note
Attention!
!
0a
0b23c
3d
4
Test Instrument On-Time
The period of time after which the test instrument is automatically shut off can be selected here. This selection has a considerable influence on the service life and the charging status of the batter­ies.
On-Time for LCD Illumination
The period of time after which LCD illumination is automatically shut off can be selected here. This selection has a considerable influence on the service life and the charging status of the batter­ies.
Submenu: Battery Level Query
Battery voltage U the battery to a load.
(Uo) can be queried here without subjecting
BAT
Data are lost when the language is changed, or if the instrument is reset to its default settings!
Back up your measurement data to a PC with the help of ETC software before press­ing the respective key. The prompt window shown at the right asks you to confirm deletion.
User Interface Language (CULTURE)
Select the desired language with the appropriate country
code and acknowledge your selection by pressing the key.
Measuring Sequence
If battery voltage drops to below 8.0 V during the course of a measuring sequence, this is only indicated by means of a pop-up window. Measured values are invalid. Measurement results cannot be saved to memory.
Press ESC in order to return to the main menu.
Default Settings (GOME SETTING)
The test instrument is returned to its original default settings when this key is activated.
Firmware Revision and Calibration Information (example)
Press any key in order to return to the main menu.
10 GMC-I Messtechnik GmbH
START/STOP SHIELD COM
Option Z550A
+
KS-C (Z541F)
+
C
O
M
SH
I
E
L
D
Limit
Green
Limit
Red
Uramp
Green
Uramp
Red
Red
!
Limit

4 General Operation

4.1 Connecting the Instrument

The test leads are connected to the “+” and “COM” jacks.
Special Case:
When inserting the 3-pole plug, make sure that the jack plug is inserted in the START/STOP position. Press and align the 3­pole plug such
Test Probe for Remote Triggering (Option Z550A)

4.3 Optical Indicators

LED Status
Function – Cause
Limit Value Indication
– Measured insulation resistance does not violate the limit
value.
– Measured low-resistance Rlo does not violate the limit
value.
Limit Value Indication
– Measured insulation resistance has fallen short of the
selected limit value.
– Measured low-resistance Rlo does has exceeded the
permissible limit value.
Indication of ramp characteristics
– Maximum ramp voltage (upper voltage limit) has been
reached (without breakdown).
that it is placed flush on the con­nection terminal. This is the only way to assure that the three contacts of the jack plug are properly con­nected with the command cables.
Special Case: Measuring High-Value Resistance with the KS-C Probe (Option)
When measuring electrostatic dis­charge capacity for floor cover­ings, the shielded cable should be connected to the COM and SHIELD jacks (KS-C accessory set, “cable set con­sisting of mea­surement cable and high-resis­tance measure­ment cable for measurements in the GΩ range”, see diagram). Be sure to observe color coding.
1
Function testing should be executed at regular intervals (see following section regarding testing the LEDs).
Testing the LED which Indicates Detection of Interference Voltage when Switched Off – OFF Switch Position
Apply a voltage of greater than 50 V (+ and COM jacks).Turn the rotary switch to the V position.Read the voltage value at the LCD.Turn the rotary switch to the OFF position.
Test results: If applied voltage is unchanged and the LED which indicates the detection of interference voltage lights up red, the
Indication of ramp characteristics
– Maximum ramp voltage (upper voltage limit) has not been
reached (due to breakdown). Breakdown voltage is displayed.
Interference voltage in switched-off condition lings of test voltage during insulation measurement
Dangerous voltage of greater than 50 V is present at the mea­surement inputs: – Initialization of insulation resistance and
low-resistance measurement is disabled.
– High test voltage is applied to the measuring terminals
(Riso/Rins, PI and DAR) during insulation measurement
Detection of interference voltage in the on-state
in the resistance measuring ranges after starting measurement The value of interference voltage is displayed in addition to signalling in the resistance measuring ranges for as long as it is present.
1
and signal-
LED is OK. In this case, the LED reliably indicates interference voltage even when the instrument is switched off. We recom-

4.2 Switching On, Monitoring and Switching Off

If supply voltage drops to a value of less than 8.5 V, the LOW BATT pop-up message appears: No more measurements can be started.
If battery voltage falls below the allowable limit value of 8.0 V, the
mend executing this test at regular intervals.
METRISO XTRA Measuring Functions, Measuring Ranges and Limit Values
Measuring Ranges
instrument cannot be switched on, or it is switched off. Measurements cannot be started in the resistance measuring
ranges in the event of interference voltage. The instrument only switches itself off automatically after comple-
tion of an (automatic) measuring sequence, and after the prede­termined on-time has expired (see page 8). On-time is reset to its original value as defined in the setup menu, as soon as any key or the rotary function switch is activated.
/ R
R
ISO
INS
/ R
R
ISO
INS
/ R
R
ISO
INS
R 10 ... 10 kΩ
R
LO
U 0 ... 1000 V
U = 50, 100, 250, 500, 1000 V
Uvar = 50 ... 1000 V
Uramp (U ) = 100 ... 1000 V
0.01 ... 10 Ω
If the instrument is switched off automatically with the rotary switch in any position other than OFF, it can be reactivated by pressing the ESC key. The instrument is also reactivated if the
Limit Values
rotary switch is activated and turned through the OFF position.
/ R
Limit R
The instrument can be switched off manually by turning the rotary switch to the OFF position.
GMC-I Messtechnik GmbH 11
ISO
PI limit Adjustable 1.0, 1.1, 1.5, 2.0, 3.0, 4.0
DAR limit Adjustable 1.25, 1.6
Limit R
LO
Fixed setting 50 kΩ @ U
INS
Fixed setting 100 kΩ @ U Fixed setting 500 kΩ @ U Fixed setting 1 MΩ @ U Fixed setting 1 MΩ @ U Fixed setting 1 MΩ @ U
Fixed setting 2 Ω
ISO/UINS
ISO/UINS
ISO/UINS
ISO/UINS
ISO/UINS
ISO/UINS
= 50 V
= 100 V
= 250 V
= 500 V
= 1000 V
= Uvar

4.4 Measurement Value Display and Memory

Note
HELP
1
2
3
2
The following appear at the display panel:
• Measurement values with abbreviations and units of measure
• Selected function
• Error messages Measurement values for automatic measuring sequences are
stored and displayed as digital values until the next measurement sequence is started, or until automatic shut-off occurs.

4.5 Help Function

A wiring diagram can be displayed for selected switch positions, as well as basic functions after they have been selected with the rotary selector switch:
Press the HELP key in order to query online help.Press the ESC key in order to exit online help.
If the upper range limit is exceeded, the upper limit value is dis­played and is preceded by the “>” symbol (greater than), which indicates measurement value overrun.
The depiction of LEDs in these operating instructions may vary from the LEDs on the actual instrument due to product improvements.

4.6 Setting Measuring Parameters using Insulation Resistance Measurement as an Example

1 Access the submenu for setting the desired parameter. 2 Select a parameter using the ↑ or ↓ scroll key. 3 The setting value is not permanently accepted for the respective measure-
ment until the key is pressed, after which the display is returned to the main menu. You can return to the main menu by pressing ESC instead of , without accepting the newly selected value.
12 GMC-I Messtechnik GmbH
Loading...
+ 26 hidden pages