Gossen Metrawatt METRISO XTRA Operating Instructions

Operating Instructions
METRISO XTRA
High-Precision Insulation, Low Resistance and Voltage Measurement Instrument
3-349-818-03
2/12.17
Red LED: limit value exceeded
Red LED:
– Interference voltage with device off
Setup Menu
Function Selector Switch
Slot for Carrying Strap
Slot for Carrying Strap
Green LED: limit value adhered to
– High test voltage at the measuring terminals
MEM: Key for memory
functions
ESC: Exit submenu /
activate instrument standby state
Control panel
Fixed Function Keys
LEDs section 4.3
CONTIN: Continuous
measurement of Rins and R (kΩ)
HELP: Access context
sensitive help
START: Start measurement
Softkeys
• Parameter selection
• Limit value specification
• Entry functions
• Memory functions
METRISO XTRA
during Riso/Rins
2 GMC-I Messtechnik GmbH
Battery Holder
Battery Compartment Lid
Battery Holder Contacts
Battery
Contact Spring
User Interface
Batteries, Fuses
Measuring
!
RS 232
+
COM
+
COM
SHIELD
SHIELD
Option KS-C (Z541F)
Option Z550A*
Charger Socket, Interfaces
These connections are located under a protective rubber flap.
a b
Replacement
Fuse
Socket for Z502R charger Caution!
Make sure that no batteries are inserted before connecting the charger.
The test instrument must remain off during the charging process.
Fuse
Inserting the Battery Holder (side view)
USB slave for PC connection
Port for connecting Barcode/RFID reader
GUARD
* Regarding Z550A:
Test Probe with measurement key remote control and LED for illuminating the measuring point: Do not use when measuring PI and DAR.
Compartment Lid
Battery
Compartment
Connections
Test Resistor 10 MΩ
GMC-I Messtechnik GmbH 3
Overview of Device Settings and Measuring Functions
Battery level
Bar graph display
Measurement
Upper range limit
Measuring
Parameter
Display panel
Save
Battery full
Battery OK
Battery weak
Battery (nearly)
Battery Level Indicator
BAT
BAT
BAT
BAT
depleted U < 8.5 V
Measuring Status
Measurement in progress: Bars moves from left
Measurement on pause: Bars are static.
to right.
||||||||||
||||||||||
Polarity reversal‚ L-PE / N-PE for documentation
setting
value
quantities
indicator
in progress
Memory Occupancy Display
MEM
Memory half full
MEM
Memory full > transfer data to PC!
These operating instructions describe a tester with software version SW-VERSION (SW1) 01.02.00.
!
Relative to Rotary Switch Setting
Switch setting,
see as of
RISO RINS
page 15
kΩ
page 20
RLO
page 21
V
page 14
SETUP
Picto­graph
Device settings, measuring functions
RINS Insulation resistance
U Voltage at the test probes
U
fix Fixed test voltage,
N
Uvariable Variable test voltage, limit value:
Uramp Variable test voltage,
PI Polarization index DAR Absorption index R Resistance measurement
RLO Low-resistance with polarity reversal RLO+, RLO–
R
OFFSET Offset resistance
U Voltage measurement
Tests: LCD pixel display
Battery test Ubat
limit values per VDE 0100
1 MOhm (for go/no-go display while saving the measured value)
ramp: triggering/breakdown voltage
Low-resistance, single-pole
LCD pixel display
All pixels off Green limit LED All pixels on Red limit LED Acoustic signal
Date/time CULT Language (D, GB)
Brightness
page 8
SET on: On-time for
Contrast Default settings
Device type Serial number Software version Hardware version Calibration date Adjustment date
LCD and tester
Scope of delivery:
1 Insulation and resistance measuring instrument 1 DAkkS calibration certificate 1 Set batteries (8 pieces in battery holder) 1Carrying strap 1 Alligator clip 1 KS17-4 cable set 1USB cable 1 Condensed operating instructions 1 Supplement Safety Information – Detailed operating instructions for download from our website
at www.gossenmetrawatt.com
4 GMC-I Messtechnik GmbH
Table of Contents Page Page
1 Applications ......................................................................6
1.1 Optional Z550A Remote Control .................................................. 6
1.2 Overview of Measuring Instrument Performance
Features ...................................................................................... 6
2 Safety Features and Precautions .....................................7
3 Initial Start-Up .................................................................. 7
3.1 Battery test ................................................................................. 7
3.2 Installing or Replacing Batteries ................................................. 7
3.3 Charging the Batteries in the Tester ........................................... 7
3.4 Device Settings – SETUP ........................................................... 8
4 General Operation ...........................................................11
4.1 Connecting the Instrument ....................................................... 11
4.2 Switching On, Monitoring and Switching Off ............................ 11
4.3 Optical Indicators ..................................................................... 11
4.4 Measurement Value Display and Memory ................................ 12
4.5 Help Function ............................................................................ 12
4.6 Setting Measuring Parameters
using Insulation Resistance Measurement as an Example ....... 12
4.7 Specifying Nominal Voltage for Uvar and Uramp ...................... 13
5 Measuring Direct Alternating Voltage ............................. 14
6 Measuring Insulation Resistance ...................................15
6.1 Measuring with Constant Test Voltage
and Nominal Value Selection via Rotary Switch Position ......... 16
6.2 Measurement with Constant Test Voltage
and Variably Adjustable Nominal Value .................................... 16
6.3 Measurement with Rising Test Voltage (ramp function)
and Variably Adjustable Final Value .......................................... 17
6.4 Insulation Resistance Measurement – Special Conditions ....... 17
6.4.1 Measurements with the Guard Cable ............................................ 17
6.5 Discharging the Device Under Test ........................................... 18
6.6 Evaluation of Measured Values ................................................. 18
6.7 Polarization Index Measurement .............................................. 18
6.7.1 Absorption Index (DAR) – DC Charging Test .................................. 19
11 Characteristic Values ...................................................... 29
12 Maintenance ...................................................................31
12.1 Firmware Revision and Calibration Information ....................... 31
12.2 Rechargeable Battery Operation and Charging ........................ 31
12.2.1 Charging Procedure with the Z502R Charger (accessory) .............. 31
12.3 Fuses ........................................................................................ 31
12.3.1 Melting Fuse ............................................................................... 31
12.3.2 Electronic Fuse ........................................................................... 31
12.4 Housing ..................................................................................... 32
12.5 Return and Environmentally Sound Disposal ............................ 32
13 Appendix .........................................................................32
13.1 Sample Connection Layouts for Insulation Resistance Measure-
ment ......................................................................................... 32
13.2 Error Messages ........................................................................ 34
13.3 Attaching the Test Probe Holder to the Carrying Strap ............. 36
13.4 Technical Data for Measurement Cables
(scope of delivery: KS17-4 safety cable set) ............................ 36
13.5 Optional Accessories (not included) ......................................... 37
14 Repair and Replacement Parts Service
Calibration Center and Rental Instrument
Service ............................................................................38
15 Recalibration ..................................................................38
16 Product Support .............................................................. 38
7 Test Resistor for Insulation Measurement for Checking the
Insulation Measuring Instrument .................................... 19
8 Measuring Resistance – kOhm Function ........................ 20
9 Measuring Low-Value Resistance of up to 10 Ohm
(protective conductor and equipotential bonding conductor) ... 21
10 Database ........................................................................24
10.1 Creating Distributor Structures, General ................................... 24
10.2 Transferring Distributor Structures ........................................... 24
10.3 Creating a Distributor Structure in the Test Instrument ........... 24
10.3.1 Creating Structures (example for electrical circuit) ......................... 25
10.3.2 Searching for Structural Elements ................................................26
10.4 Saving Data and Generating Reports ........................................ 27
10.4.1 Use of Barcode Scanners and RFID Readers ................................. 28
GMC-I Messtechnik GmbH 5

1 Applications

This instrument fulfills all requirements of applicable European and national EC directives. We confirm this with the CE mark. The rel­evant declaration of conformity can be obtained from GMC-I Messtechnik GmbH.
The METRISO XTRA insulation and resistance measuring instru­ment allows for quick and efficient testing of protective measures in accordance with DIN VDE 0100, ÖVE-EN 1 (Austria), SEV 1000 (Switzerland), and regulations specific to other countries as well.
The device is equipped with a microprocessor and complies with IEC/EN 61557 / VDE 0413 regulations:
Part 1: General requirements Part 2: Insulation resistance Part 4: Resistance of earth connection, protective conductors
and equipotential bonding
Part 10: Electrical safety in low-voltage systems up to 1000 V
AC and 1500 V DC – Equipment for testing, measuring or monitoring of protective measures
As well as requirements per VDE 0701-0702: Repair, modification and testing of electrical devices
The test instrument is especially well suited for:
•Systems setup
• Initial start-up
• Periodic testing
• Troubleshooting in electrical systems
The following measurements and tests can be performed with the insulation measuring instrument:
• Insulation resistance
• Low-resistance
• Voltage
The following can also be tested by using a shielded measure­ment cable:
• Floor covering electrostatic discharge capability

1.2 Overview of Measuring Instrument Performance Features

METRISO XTRA
Article No. M550S
Measurement
U = 50, 100, 250, 500, 1000 V
R
INS
Uvar = 50 ... 1000 V
R
INS
Uramp (U) = 100 ... 1000 V
R
INS
PI
DAR R 10 ... 10 kΩ
R
LO
U 0 ... 1000 V
Display Functions
Backlit display
Limit value LED (green/red) for: limit values per VDE 0100 in addition to acoustic signal
URamp limit value LED for:
indicating ramp characteristics
LED for dangerous contact voltage
(when switched off)
Battery level display
Special Functions
Discharge capacitive devices under test
Safety shutdown (UBatt < 8 V)
Data storage at the instrument
ETC software for data acquisition, data management and reporting
Features
Measuring categories: CAT II 1000 V / CAT II I 600 V / CAT IV 300 V Test resistance: 10 MΩ
Connections: charging socket, USB port (slave), RS 232 port
DAkkS calibration certificate
0.01 ... 10 Ω
R
INS RLO
R
INS

1.1 Optional Z550A Remote Control

The optional remote control with triggering key on the test probe and a key for illuminating the measuring point makes it possible to use the measuring instrument at difficult to access places.
For safety reasons, a predefined voltage is only applied to the test probes for as long as the triggering key is held depressed.
The start key on the test instrument should be used instead of remote control for triggering polarization and absorption index measurements, because these two measurements involve com­plete measuring cycles.
6 GMC-I Messtechnik GmbH

2 Safety Features and Precautions

Attention!
!
Attention!
!
Attention!
!
Attention!
!
!
XY123
2012-06
D-K
15080-01-01
Consecutive number
Registration number Date of calibration (year – month)
Deutsche Akkreditierungsstelle GmbH – calibration lab
The electronic measuring and test instrument is manufactured and tested in accordance with safety regulations IEC/EN 61010­1/VDE 0411-1 and EN 61 557. safety of the operator, as well as that of the instrument, is assured.
Read the operating instructions thoroughly and carefully before using your instrument. Follow all instructions contained therein.
The measuring and test instrument may not be placed into service:
• If the battery compartment lid has been removed
• If external damage is apparent
• If connector cable or measuring adapters are damaged
•If the instrument no longer functions flawlessly
• After extraordinary damage due to transport
• After a long period of storage under unfavorable conditions (e.g. humidity, dust or extreme temperature).
Opening of Equipment / Repair
The equipment may be opened only by authorized service per­sonnel to ensure the safe and correct operation of the equipment and to keep the warranty valid.
Even original spare parts may be installed only by authorized ser­vice personnel.
In case the equipment was opened by unauthorized personnel, no warranty regarding personal safety, measurement accuracy, conformity with applicable safety measures or any consequential damage is granted by the manufacturer.
Data Backup
We advise you to regularly transfer your stored data to a PC in order to prevent potential loss of data in the test instrument.
We assume no responsibility for any data loss. We recommend ETC software (Electric Testing Center) for back-
ing up, processing and managing data.
Meaning of Symbols on the Instrument
Warning concerning a point of danger (Attention, observe documentation!)
Protection class II device
When used for its intended purpose,
If battery voltage has fallen below the allowable lower limit, the pictograph shown at the right appears. The instrument does not function if the batteries have been depleted excessively, and no display appears.

3.2 Installing or Replacing Batteries

New batteries must be inserted for initial start-up, or if only one filled segment remains in the battery symbol.
Before opening the battery compartment (see page 5 for location), disconnect the instrument from the measuring circuit (mains) at all poles.
Eight 1.5 V size AA batteries in accordance with IEC LR 6 are required for operation of the insulation measuring instrument. Use new alkaline manganese batteries only.
Rechargeable NiCd or NiMH batteries may also be used. These can be charged externally or by connecting the Z502R charger to the test instrument. We recommend rechargeable NiMH batteries.
Always replace batteries in complete sets. Dispose of batteries in an environmentally sound fashion.
Loosen both slotted screws for the battery compartment lid
on the back, and remove the lid.
Remove the battery holder and insert eight 1.5 V size AA bat-
teries with correct polarity in accordance with the symbols.
Make sure that all of the batteries are inserted with correct polarity. If just one battery is inserted with reversed polar­ity, it will not be recognized by the instrument and may result in leakage from the batteries.
Push the battery holder into the battery compartment such
that the battery holder’s contacts touch the contact springs at the bottom of the battery compartment (see drawing on page 3). If the battery holder is not inserted as specified, the instrument cannot be supplied with power.
Replace the battery compartment lid and retighten the
screws.
Charging socket for extra-low direct voltage (Z502R charger)
Attention! Only rechargeable batteries may be inserted when the char­ger is connected.
EC mark of conformity
The device and included batteries may not be dis­posed of with the trash. Further information regarding the WEEE mark can be accessed on the Internet at www.gossenmetrawatt.com by entering the search term “WEEE”.
Calibration Seal (blue seal):
See also “Recalibration” on page 38.

3 Initial Start-Up

3.1 Battery test

Four different battery symbols, ranging from fully depleted to fully charged, continuously indicate the momentary charge level in the upper left-hand corner of the display
GMC-I Messtechnik GmbH 7
The instrument may only be placed into service if the battery compartment lid is securely fastened!

3.3 Charging the Batteries in the Tester

Use only the Z502R charger (available as an accessory) to charge batteries which have already been inserted into the test instrument.
Make sure that the following conditions have been fulfilled be­fore connecting the charger to the charging socket:
– Rechargeable batteries have been installed with
correct polarity (not standard batteries)
– The test instrument has been disconnected from the
measuring circuit at all poles
– The instrument must remain off during charging.
Refer to section 12.2.1 with regard to charging batteries which have been inserted into the tester.
If the batteries or the battery pack have not been used or recharged for a lengthy period of time (> 1 month), thus resulting in excessive depletion:
Observe the charging sequence (indicated by the LED at the charger) and initiate a second charging sequence if necessary (disconnect the charger from the mains and from the test instru­ment to this end, and then reconnect it).

3.4 Device Settings – SETUP

SETUP
LED and LCD test menu
Battery test menu
Brightness/contrast menu
Software revision level Calibration date
Display: date / time
Display: automatic shutdown
Display: automatic shutdown
of display illumination after 30 s.
of the tester after 120 s. Time, language, profiles
1
2
3
4
0b
0a
0
Return to main menu
Limit LED: test red
Limit LED: test green
Cell test
Inverse cell test
Hide all pixels
Show all pixels
Acoustic signal test
1
Return to main menu
Increase brightness
Decrease brightness
Increase contrast
Decrease contrast
Set time
Default settings
User interface language
3
3a 3b
3c
3d
Set date
On-time for display illumination / tester
0b
Return to submenu
0a
Display Illumination On-time
Brightness and Contrast Settings Time, On-Time and Default Settings
Menu Selections for Operating Parameters
LED Tests LCD and Acoustic Signal Tests
Test Instrument On-Time
Automatic shutdown, deactivated = OFF
8 GMC-I Messtechnik GmbH
SETUP
LED and LCD test menu
Battery test menu
Brightness/contrast menu
Software revision level Calibration date
Display: date / time
Display: automatic shutdown
Display: automatic shutdown
of display illumination after 15 s.
of the tester after 45 s.
Time, language, profiles
1
2
3
4
0b
0a
0
Return to main menu
Increase brightness
Decrease brightness
Increase contrast
Decrease contrast
Set time
Default settings
User interface language
3
3a 3b
3c
3d
Set date
On-time for display illumination / tester
Set time
Menu Selection for Operating Parameters
Adjusting Brightness and Contrast Set Time, Language, Profiles, Acoustic Signal
Set date
Select time/date
Increase
Increase
hours
Activate settings
minutes
3a
Increase seconds
Return to submenu
Decrease
Decrease
hours
minutes
Decrease
seconds
Select time/date
Increase
Increase
day
Activate settings
month
3b
Increase year
Return to submenu
Decrease
Decrease
day
month
Decrease
year
GMC-I Messtechnik GmbH 9
Significance of Individual Parameters
Note
Attention!
!
0a
0b23c
3d
4
Test Instrument On-Time
The period of time after which the test instrument is automatically shut off can be selected here. This selection has a considerable influence on the service life and the charging status of the batter­ies.
On-Time for LCD Illumination
The period of time after which LCD illumination is automatically shut off can be selected here. This selection has a considerable influence on the service life and the charging status of the batter­ies.
Submenu: Battery Level Query
Battery voltage U the battery to a load.
(Uo) can be queried here without subjecting
BAT
Data are lost when the language is changed, or if the instrument is reset to its default settings!
Back up your measurement data to a PC with the help of ETC software before press­ing the respective key. The prompt window shown at the right asks you to confirm deletion.
User Interface Language (CULTURE)
Select the desired language with the appropriate country
code and acknowledge your selection by pressing the key.
Measuring Sequence
If battery voltage drops to below 8.0 V during the course of a measuring sequence, this is only indicated by means of a pop-up window. Measured values are invalid. Measurement results cannot be saved to memory.
Press ESC in order to return to the main menu.
Default Settings (GOME SETTING)
The test instrument is returned to its original default settings when this key is activated.
Firmware Revision and Calibration Information (example)
Press any key in order to return to the main menu.
10 GMC-I Messtechnik GmbH
START/STOP SHIELD COM
Option Z550A
+
KS-C (Z541F)
+
C
O
M
SH
I
E
L
D
Limit
Green
Limit
Red
Uramp
Green
Uramp
Red
Red
!
Limit

4 General Operation

4.1 Connecting the Instrument

The test leads are connected to the “+” and “COM” jacks.
Special Case:
When inserting the 3-pole plug, make sure that the jack plug is inserted in the START/STOP position. Press and align the 3­pole plug such
Test Probe for Remote Triggering (Option Z550A)

4.3 Optical Indicators

LED Status
Function – Cause
Limit Value Indication
– Measured insulation resistance does not violate the limit
value.
– Measured low-resistance Rlo does not violate the limit
value.
Limit Value Indication
– Measured insulation resistance has fallen short of the
selected limit value.
– Measured low-resistance Rlo does has exceeded the
permissible limit value.
Indication of ramp characteristics
– Maximum ramp voltage (upper voltage limit) has been
reached (without breakdown).
that it is placed flush on the con­nection terminal. This is the only way to assure that the three contacts of the jack plug are properly con­nected with the command cables.
Special Case: Measuring High-Value Resistance with the KS-C Probe (Option)
When measuring electrostatic dis­charge capacity for floor cover­ings, the shielded cable should be connected to the COM and SHIELD jacks (KS-C accessory set, “cable set con­sisting of mea­surement cable and high-resis­tance measure­ment cable for measurements in the GΩ range”, see diagram). Be sure to observe color coding.
1
Function testing should be executed at regular intervals (see following section regarding testing the LEDs).
Testing the LED which Indicates Detection of Interference Voltage when Switched Off – OFF Switch Position
Apply a voltage of greater than 50 V (+ and COM jacks).Turn the rotary switch to the V position.Read the voltage value at the LCD.Turn the rotary switch to the OFF position.
Test results: If applied voltage is unchanged and the LED which indicates the detection of interference voltage lights up red, the
Indication of ramp characteristics
– Maximum ramp voltage (upper voltage limit) has not been
reached (due to breakdown). Breakdown voltage is displayed.
Interference voltage in switched-off condition lings of test voltage during insulation measurement
Dangerous voltage of greater than 50 V is present at the mea­surement inputs: – Initialization of insulation resistance and
low-resistance measurement is disabled.
– High test voltage is applied to the measuring terminals
(Riso/Rins, PI and DAR) during insulation measurement
Detection of interference voltage in the on-state
in the resistance measuring ranges after starting measurement The value of interference voltage is displayed in addition to signalling in the resistance measuring ranges for as long as it is present.
1
and signal-
LED is OK. In this case, the LED reliably indicates interference voltage even when the instrument is switched off. We recom-

4.2 Switching On, Monitoring and Switching Off

If supply voltage drops to a value of less than 8.5 V, the LOW BATT pop-up message appears: No more measurements can be started.
If battery voltage falls below the allowable limit value of 8.0 V, the
mend executing this test at regular intervals.
METRISO XTRA Measuring Functions, Measuring Ranges and Limit Values
Measuring Ranges
instrument cannot be switched on, or it is switched off. Measurements cannot be started in the resistance measuring
ranges in the event of interference voltage. The instrument only switches itself off automatically after comple-
tion of an (automatic) measuring sequence, and after the prede­termined on-time has expired (see page 8). On-time is reset to its original value as defined in the setup menu, as soon as any key or the rotary function switch is activated.
/ R
R
ISO
INS
/ R
R
ISO
INS
/ R
R
ISO
INS
R 10 ... 10 kΩ
R
LO
U 0 ... 1000 V
U = 50, 100, 250, 500, 1000 V
Uvar = 50 ... 1000 V
Uramp (U ) = 100 ... 1000 V
0.01 ... 10 Ω
If the instrument is switched off automatically with the rotary switch in any position other than OFF, it can be reactivated by pressing the ESC key. The instrument is also reactivated if the
Limit Values
rotary switch is activated and turned through the OFF position.
/ R
Limit R
The instrument can be switched off manually by turning the rotary switch to the OFF position.
GMC-I Messtechnik GmbH 11
ISO
PI limit Adjustable 1.0, 1.1, 1.5, 2.0, 3.0, 4.0
DAR limit Adjustable 1.25, 1.6
Limit R
LO
Fixed setting 50 kΩ @ U
INS
Fixed setting 100 kΩ @ U Fixed setting 500 kΩ @ U Fixed setting 1 MΩ @ U Fixed setting 1 MΩ @ U Fixed setting 1 MΩ @ U
Fixed setting 2 Ω
ISO/UINS
ISO/UINS
ISO/UINS
ISO/UINS
ISO/UINS
ISO/UINS
= 50 V
= 100 V
= 250 V
= 500 V
= 1000 V
= Uvar

4.4 Measurement Value Display and Memory

Note
HELP
1
2
3
2
The following appear at the display panel:
• Measurement values with abbreviations and units of measure
• Selected function
• Error messages Measurement values for automatic measuring sequences are
stored and displayed as digital values until the next measurement sequence is started, or until automatic shut-off occurs.

4.5 Help Function

A wiring diagram can be displayed for selected switch positions, as well as basic functions after they have been selected with the rotary selector switch:
Press the HELP key in order to query online help.Press the ESC key in order to exit online help.
If the upper range limit is exceeded, the upper limit value is dis­played and is preceded by the “>” symbol (greater than), which indicates measurement value overrun.
The depiction of LEDs in these operating instructions may vary from the LEDs on the actual instrument due to product improvements.

4.6 Setting Measuring Parameters using Insulation Resistance Measurement as an Example

1 Access the submenu for setting the desired parameter. 2 Select a parameter using the ↑ or ↓ scroll key. 3 The setting value is not permanently accepted for the respective measure-
ment until the key is pressed, after which the display is returned to the main menu. You can return to the main menu by pressing ESC instead of , without accepting the newly selected value.
12 GMC-I Messtechnik GmbH

4.7 Specifying Nominal Voltage for Uvar and Uramp

Note
Select value.
Select value.
Accept value.
Delete characters =
Save value (to list).
Select the EDIT menu.
change cursor position.
Nominal voltage can be freely selected within specified limits for the Uvar and Uramp measuring functions:
1 Access submenu for selecting the desired voltage:
Press the key with the icon.
2 Select the desired value with the left or right scroll key. The value is
accepted by pressing the key. The entire value is acknowledged by selecting and then pressing the key. The new nominal voltage appears at the main display.
Observe the predefined limits for the new setting value. New, freely selected limit values or nominal values included in the parameters list can be deleted/edited at the PC with the help of ETC software.
GMC-I Messtechnik GmbH 13

5 Measuring Direct Alternating Voltage

Note
V
L1 L2
N PE
L3
R
B
R
E
Max.
1.2 kV
U
+
COM
+
COMSHIELD
1
2
2
3
Direct voltage
Alternating voltage
Pulsating voltage
(DC + AC TRMS)
You can measure direct voltage, as well as sinusoidal alternating voltage with frequencies ranging from 45 to 65 Hz, with this test instrument.
Select Measuring Function
Select the V measuring function with the rotary switch.
Connection
Connect the measurement cables to the + and COM jacks.Contact the measuring point with both test probes.
Measurement
The measured value is displayed directly (without pressing the START key) – in analog format at the bar graph and in digital format at the matrix display.
After completing the measurement, switch the instrument off
by turning the rotary switch to the OFF position.
The ESC, START and CONTIN. keys have no function in this case.
Set the waveform parameter (voltage type).
Input impedance for the voltage measuring range is 10 MΩ.
14 GMC-I Messtechnik GmbH

6 Measuring Insulation Resistance

Note
Note
R
INS
U
var
U
ramp
HELP
1/2 2/2
L-PE
N-PE
1/4 2/4 3/4 4/4
L1-PE L2-PE L3-PE
N-PE
AUTO1 AUTO3
R
ISO/RINS
MEMORY
MEMORY R
ISO/RINS
Switch and save
Switch without saving
NEXT
Select Measuring Function
Select the R
depending on the measuring task, constant or variable test voltage (Uvar), or a ramp function (Uramp).
Connection
measuring function with the rotary switch and,
INS
Semiautomatic Measurement in Multipole Systems
Fast, semiautomatic polarity reversal is possible in all rotary switch positions for insulation resistance measurement. However, polar­ity selection is only relevant for documentation.
Fast Polarity Reversal
The polarity parameter is set to AUTO1 (single-phase mains volt­age) or AUTO3 (3-phase mains voltage).
There are two ways to quickly and conveniently switch amongst all polarity variants without switching to the parameter settings submenu:
• Press the NEXT key.
• Press the save key at the instrument twice after each mea­surement.
Connect the device under test to the + and COM jacks. Sample connection layouts for insulation resistance measurement
are included in section 13.1. Insulation resistance can only be measured at voltage-free
objects. If mains voltage or interference voltage is applied to the measurement inputs, measurement cannot be started.
Checking Measurement Cables Before Measurements
Before performing insulation measurement, the test probes on the measurement cables should be short-cir­cuited in order to assure that the instrument displays a value of less than 1 kΩ (see section 9). In this way, incor­rect connection can be avoided and broken measure­ment cables can be detected.
Continuous Measurement via the CONTIN. Key
The instrument’s batteries are rapidly depleted during the insulation resistance measurement. Stop continuous measurement with “constant test voltage” as soon as the display has settled in.
GMC-I Messtechnik GmbH 15
6.1 Measuring with Constant Test Voltage and Nominal Value
Note
Polarization index
Test voltage switch position:
50 V / 100 V / 250 V / 500 V / 1000 V
Absorption index
Insulation resistance
START
CONTIN
Variable test voltage
50 V ... 1000 V
(constant nominal value):
START
CONTIN
Selection via Rotary Switch Position
Set Parameters

6.2 Measurement with Constant Test Voltage and Variably Adjustable Nominal Value

A test voltage which deviates from nominal voltage, and is usually lower, can be selected for measurements at sensitive compo­nents, as well as systems with voltage limiting devices.
Set Parameters
The constant test voltage function offers two options:
After briefly pressing the START key, specified test voltage UN is read out and insulation resistance RINS is measured. As soon as the measured value is stable (settling time may be several seconds in the case of high cable capacitance values), mea­surement is ended and the last measured values for RINS and UINS are displayed. U is the voltage which is measured at the test probes during and after testing. This voltage drops to a value of less than 10 V (see section entitled “Discharging the Device Under Test”.
or
As soon as you press the CONTIN key, test voltage UN is applied and insulation resistance RINS is measured. Do not press the key again in order to stop measurement until the measured value has settled in (settling time may be several seconds in the case of high cable capacitance values). Voltage U, which is measured during testing, corresponds to voltage UINS. After once again pressing the CONTIN key, measurement is ended and the last measured values for RINS and UINS are displayed. U drops to a value of less than 10 V after measure­ment (see the section entitled “Discharging the Device Under Test”.
Start Measurement
Refer to section 4.7 regarding how to enter the variable voltage. The selected nominal voltage is displayed in the main menu.
Pole Selection Report Entry
The poles between which testing takes place can only be entered here for reporting purposes. The entry itself has no influence on the actual polarity of the test probes or pole selection.
Start Measurement
After pressing the START key, nominal voltage UN (50 to 1000 V) entered previously via the parameters menu is applied until the measured values settles in.
After pressing the CONTIN. key, the previously selected test voltage or nominal voltage U key is pressed once again.
(50 to 1000 V) is applied until the CONTIN.
N
U is the voltage which is measured at the test probes during and after testing. This voltage drops to a value of less than 10 V after measurement (see section 6.5, “Discharging the Device Under Test’ ’) .
Test voltage U
After pressing the START key, nominal voltage UN (50 ... 1000 V) selected with the rotary switch is applied until the measured val­ues settles in.
After pressing the CONTIN. key, the previously selected test voltage or nominal voltage U
key is pressed once again.
U is the voltage which is measured at the test probes during and
after testing. This voltage drops to a value of less than 10 V after measurement (see section 6.5, “Discharging the Device Under Test’’).
Test voltage U
16 GMC-I Messtechnik GmbH
(50 to 1000 V) is applied until the CONTIN.
N
is displayed for U
INS
.
is displayed for U
The instrument’s batteries are rapidly depleted during the insulation resistance measurement. Stop continuous measurement with “constant test voltage” as soon as the display has settled in.
INS
.
6.3 Measurement with Rising Test Voltage (ramp function)
Note
Attention!
!
Attention!
!
Note
Maximum test voltage
100 V ... 1000 V
(final ramp value)
Max. 1 mA (at 1 KΩ/V)
Current load:
START
+Measurement Cable
Measurement Cable
Guard Cable
Conductor
Insulation Material
Guard Rings
Contact Ring
(COM)
and Variably Adjustable Final Value
The “Uramp” rising test voltage function (ramp function) is used to detect weak points in the insulation, as well as to determine response voltage for voltage limiting components.
Set Parameters

6.4 Insulation Resistance Measurement – Special Conditions

Insulation resistance can only be measured at voltage­free objects.
If measured insulation resistance is less than the selected limit value, the limit LED lights up red.
Refer to section 4.7 regarding how to set the final ramp value. The selected nominal voltage value is displayed in the main menu.
Pole Selection Report Entry
The poles between which testing takes place can only be entered here for reporting purposes. The entry itself has no influence on the actual polarity of the test probes or pole selection.
Start Measurement
After briefly pressing the START key, test voltage is continuously increased until the specified nominal voltage U reached.
Insulation measurement with rising test voltage is ended:
• As soon as specified maximum test voltage U the measured value is stable
or
• After sparkover occurs at breakdown voltage
(100 ... 1000 V) is
N
is reached and
N
If an interference voltage of roughly – Interference Voltage” appears in a pop-up window, insulation resistance is not measured. In the case of interference voltage > 50 V, the “> 50 V” LED lights up.
In 3-phase systems, all conductors (L1, L2, L3 and N) must be tested against PE!
Do not touch the instrument’s terminal contacts during insulation resistance measurements!
If nothing has been connected to the terminal contacts, or if a resistive load component has been connected for measurement, your body would be exposed to a current of approximately 1 mA at a voltage of 1000 V. The noticeable shock may lead to injury (e.g. resulting from a startled reaction etc.).
15 V is present and “U
EXT

6.4.1 Measurements with the Guard Cable

The measurement of very high resistances necessitates extremely minimal measuring current and may be rendered problematic as a result of influences such as electromagnetic fields, humidity or surface pollution. An accurate test set-up is thus absolutely essential.
A guard cable must be used for measurements within a range of 100 GΩ (10 GΩ) 1TΩ, in order to prevent surface current from distorting measurement results. The guard rings prevent current at the surface of the insulation material from flowing from the +measurement cable to the –measurement cable, instead of through the insulation material itself.
Insert the plug from the guard cable into the appropriate jack
in the test instrument.
Plug the alligator clip onto the guard cable test probe.Connect the alligator clip to the guard ring between the two
measuring points at the insulation material under test.
Refer to section 6.1 to section 6.3 regarding the measuring
sequence.
As soon as the final ramp value is reached, the Uramp LED lights up green. If the final ramp value is not reached due to sparkover, the Uramp LED lights up red.
U is the voltage which is measured at the test probes during and after testing. This voltage drops to a value of less than 10 V after measurement (see section 6.5, “Discharging the Device Under Test’’).
Highest achieved test voltage U or any triggering or breakdown volt- age which occurs is displayed for U
GMC-I Messtechnik GmbH 17
.
INS
Measurement can be stopped at any time by pressing the START key or the CONTIN. key.
The following materials can be used as guard rings: alu­minum foil, copper foil or metallic hose clamps.

6.5 Discharging the Device Under Test

Attention!
!
Attention!
!
Note
PI
R
10min
R
1min
--------------=
Polarization Index
Test voltage switch position:
50 V / 100 V / 250 V / 500 V / 1000 V
Absorption index
Insulation resistance
START
If measurement is performed at a capacitive object such as a long cable, it becomes charged with up to approx. 1000 V!
Touching such objects is life endangering!
When an insulation resistance measurement has been performed on a capacitive object it is automatically discharged by the instru­ment after measurement has been completed. Contact with the device under test must be maintained to this end. The falling volt­age value can be observed at the U display.
Do not disconnect the DUT until less than 10 V is displayed for U!

6.6 Evaluation of Measured Values

Instrument measuring error must be taken into consideration in order to assure that the limit values set forth in DIN VDE regula­tions are not fallen short of. The required minimum display values for insulation resistance can be determined with the help of the table, “
Display Values in Consideration of Measuring Uncertainty page 30. These values take maximum device error into consider­ation (under nominal conditions of use). Intermediate values can be interpolated.
”, o n
Z550A Option
Use of the measurement cable with remote start/stop is not advisable for polarization index measurement PI or absorption index measurement DAR, because measure­ment is only conducted as long as the START key is pressed and held. sequence is nevertheless executed, start each measurement with the
Set Parameters
In order to assure that the entire measuring
START
key at the test instrument only.

6.7 Polarization Index Measurement

In the case of electrical machines which include components with windings (generator and motor windings), it’s advisable to con­duct polarization index testing. This procedure involves expanded testing of insulation resistance. The accumulation of moisture and contamination on windings can be detected as reduced insulation resistance.
DC measuring voltage from the METRISO XTRA is applied to the insulation for a duration of 10 minutes to this end. Measured val­ues are documented after one minute, and after ten minutes. If the insulation is good, the value measured after ten minutes is higher than the value measured after one minute. The relationship between the two measurement values is the polarization index.
The molecules within the insulation are aligned due to the applica­tion of measuring voltage over a long period of time, resulting in polarization. The polarization index indicates whether or not the molecules contained in the insulation can still be moved, thus allowing for polarization. This, in turn, is an indication of the condi­tion of the insulation. The more freely the charge carriers can be moved, the better is the condition of the insulation.
The following rules apply in general:
PI value < 1: Troubleshooting is required. PI value = 1 to 2: Maintenance is advisable. PI value = 2 to 3: DUT is OK!
PI values > 3: Error-free device under test
No immediate action is required. Preventive maintenance can be planned according to workload.
Start Measurement
Applications
Determination of moisture and contamination levels
The selected polariza­tion index menu is only displayed until the rotary
For PI measurement, test instrument on-time should be set to OFF (see section 3.4, “Device Settings – SETUP’’).
18 GMC-I Messtechnik GmbH
switch is activated or the parameter is changed.

6.7.1 Absorption Index (DAR) – DC Charging Test

DAR
R
60s
R
30s
---------=
Polarization Index
Test voltage switch position: 50 V / 100 V / 250 V / 500 V / 1000 V
Absorption index
Insulation resistance
START
+
COM
10M
Ω
+
COMSHIELD
R
ISO/RINS
10 MΩ
The absorption index test is part of the polarization index test. Insulation resistance measurements are placed in relationship to one another after 30 and 60 seconds.
Applications:
Set Parameters
Faster version of the polarization index test.
The selected polarization index menu is only displayed until the rotary switch is activated or the parameter is changed.
7 Test Resistor for Insulation Measurement for
Checking the Insulation Measuring Instrument
According to section 5.3.1.2 of VDE 0105-100 (EN 50110-1), the following applies: “These measuring instruments must be tested before, and if applicable after use.” The two outermost jacks on the connection panel must be con­nected to each other internally via a 10 MΩ test resistor to this end.
The sum of test resistor and cable resistance (for both cables), including test probes, amounts to 10 MΩ ±5%. This value allows for quick self-testing.
Start Measurement
Connect the measurement cables to the + and COM jacks.Insert the test probes into the above described jacks.Select the RiSO/iNS measuring function with the rotary
switch, as well as the desired test voltage, e.g. RiSO 100V.
Press the start key and view the measurement results.
GMC-I Messtechnik GmbH 19

8 Measuring Resistance – kOhm Function

Attention!
!
kΩ
+
C
O
M
R
x
START
CONTIN
Connector with
Shielded Cable,
Option KS-C (Z541F)
+
C
O
M
Shielded Cable
S
H
I
E
L
D
Resistances of greater than 10 Ω and less than 10 kΩ are mea­sured in this switch position. Refer to section 9 regarding resis­tances of less than 10 Ω .
Select Measuring Function
Select the kΩ measuring function with the rotary switch.
Connection
Connect the device under test to the + and COM jacks.
Special Case: Measuring High-Value Resistance with the KS-C Probe
Connection
Connect the device under test to the + and COM jacks with a
shielded cable via the optional KS-C probe.
Resistance can only be measured at voltage-free ob­jects. If mains voltage or interference voltage is applied to the measurement inputs, measurement cannot be started.
Start Measurement
Start an individual measurement by briefly pressing the START
key, or initiate continuous measurement by briefly pressing the CONTIN key.
20 GMC-I Messtechnik GmbH
9 Measuring Low-Value Resistance of up to
Attention!
!
Attention!
!
Note
Note
R
LO
HELP
+
COM
R
LO
+
COMSHIELD
R < 10
Ω
R
LO
R
x
0V !
!
Polarity: ± to PE
10 Ohm
(protective conductor and equipotential
bonding conductor)
According to the regulations, the measurement of low-value resis­tance at protective conductors, earth conductors or bonding con­ductors must be performed with (automatic) polarity reversal of the test voltage, or with current flow in one direction (+ pole to PE) and then the other (– pole an PE).
In the measuring function RLO, measurements are per­formed with currents about 200 mA. Please check before measuring whether your DUT or your circuit is designed for these high current values.
Low-value resistance can only be measured at voltage­free objects.
Select Measuring Function
Set Polarity Parameter (direction of current flow)
Select a polarity option or automatic polarity reversal.
Measuring ROFFSET
When using extension cables with resistance of up to 5.00 Ω, their resistance can be deducted automatically from the measure­ment results. Proceed as follows:
Select the Rlo measuring function with the rotary switch.
Connection
Short-circuit the end of the measurement extension cable with
the second test probe at the instrument.
Start measurement of offset resistance with OFFSET.
If the difference between RLO+ and RLO– is great than 10% with automatic polarity reversal, no offset is accepted. A pop-up window appears which must be acknowledged by any desired key. The respectively smaller value is otherwise stored to memory as an offset value. The maximum offset value is 5.00 Ω. Negative resis­tances may result due to the offset value.
The permissible value for R In the case of values of greater than 5.00 Ω, a pop-up window appears and the value is not accepted. The error message must be acknowledged and/or cleared by any desired key. The measured R measurement value for all subsequent R
R
OFFSET can be deleted at any time by pressing the OFFSET CLEAR
key.
OFFSET value will now be deducted from the actual
Only use this function when performing measurements with extension cables. When different extension cables are used, the above described procedure must always be repeated.
OFFSET is between 0.00 and 5.00 Ω.
measurements.
LO
Connect the device under test to the + and COM jacks.
GMC-I Messtechnik GmbH 21
Start Measurement
Attention!
!
Note
START
Measurement of + pole to PE
Measurement of – pole to PE
Automatic Polarity Reversal
After the measuring sequence has been started, the instrument performs measurement with automatic polarity reversal, first with current flow in one direction, and then in the other.
If the difference between RLO+ and RLO– is greater than 10% with automatic polarity reversal, RLO+ and RLO– values are dis­played instead of RLO. The respectively larger value, RLO+ or RLO–, appears at the top and is saved to the database as the RLO value.
Resistances which do not demonstrate a stable value until after a “settling in period” should not be measured with automatic polar­ity reversal. Measurement with automatic polarity reversal may lead to varying and/or inflated measurement values, and thus to an ambiguous reading.
Limit Value Indication
If the measured value is less than or equal to 2 Ω, the limit LED lights up green. If the measured value is greater than 2 Ω, the LED lights up red.
Evaluating Measurement Results
Differing results for measurements in both directions indicate volt­age at the DUT (e.g. thermovoltages or unit voltages).
Measurement results can be distorted by parallel connected impedances at load current circuits and by equalizing current, especially in systems which make use of “overcurrent protection devices” (previous neutralization) without an isolated protective conductor. Resistances which change during measurement (e.g. inductance), or a defective contact, can also cause distorted measurements (double display).
In order to assure unambiguous measurement results, causes of error must be located and eliminated.
In order to find the cause of measuring error, measure resistance in both current flow directions.
Final results
The measurement cannot be started until the test probes are in contact with the device under test. If a voltage of U > approx. 3 V is present at the device under test, a pop-up window appears which warns against interference voltage. This error message must be acknowledged and/or cleared by any desired key. Eliminate the interference voltage. If resistance is greater than 10 Ω, OL appears at the dis­play.
In the case of single-pole measurement, the respective value is saved to the database as RLO.
Measuring Low-Value Resistance
If an extension cable is used its resistance must be mea­sured and deducted from the measurement results.
Resistances which do not demonstrate a stable value until after a “settling in period” should not be measured with automatic polarity reversal, but rather one after the other with positive and negative polarity. Examples of resistances whose values may change during measurement include: – Incandescent lamp resistance, whose values change
due to warming caused by test current – Resistances with a large conductive component – Contact resistance
Polarity Selection Display Condition
+ pole to PE RLO+ None
– pole to PE RLO None
RLO If ΔRLO 10%
± pole to PE
22 GMC-I Messtechnik GmbH
RLO+ RLO
If ΔRLO > 10%
Evaluating of Measured Values
See table, “ on page 30.
Display Values in Consideration of Measuring Uncertainty
”,
Calculation of Cable Lengths for Common Copper Conductors
If the key is activated after performance of resistance measurement, the cable lengths corresponding to common con­ductor cross sections are displayed.
If results vary for the two different directions of current flow, cable length is not displayed. In this case, capacitive or inductive com­ponents are apparently present which would distort the calcula­tion.
This table only applies to cables made with commercially available copper conductors and cannot be used for other materials (e.g. aluminum)!
GMC-I Messtechnik GmbH 23

10 Database

10.1 Creating Distributor Structures, General

A complete distributor structure with customer, building and distrib­utor data can be created in the METRISO XTRA test instrument. This structure makes it possible to assign measurements to dis­tributors in various buildings and customer facilities.

10.2 Transferring Distributor Structures

The following data transfer operations are possible:
• Transfer a distributor structure from the PC to the test instru­ment.
• Transfer a distributor structure including measured values from the test instrument to the PC.
There are two possible procedures:
• On location or at the construction site: Create the distributor structure in the test instrument. A distrib­utor structure with up to 50,000 struc­tural elements can be created in the test instrument, which is saved to the instru­ment’s flash mem­ory.
or
• Create and save a distributor structure at a PC with the help of ETC report generating software (Electric Testing Center) (see con­densed operating instructions for ETC report generating soft­ware). The distributor structure is then transferred to the test instrument.
The test instrument and the PC must be con­nected with a USB cable in order to transfer dis­tributor structures and data.
The following image appears at the display during transfer of struc­tures and data.

10.3 Creating a Distributor Structure in the Test Instrument

Overview of the Meanings of Icons used to Create Structures
Icon Meaning
Main
Sub-
Level
Level
Memory menu, page 1 of 3
Cursor UP: scroll up
Cursor DOWN: scroll down
Note regarding ETC Report Generating Software
The following steps must be completed before using the soft­ware:
Install USB device drivers: (required for operation of the METRISO XTRA at a PC): GMC-I Driver Control software can be downloaded from Gossen Metrawatt's website at:
http://www.gossenmetrawatt.com
Products Software Software for Testers Utilities Driver Control
Install ETC report generating software: You can download the current ETC version free of charge from our homepage under section mygmc after registration or login:
http://www.gossenmetrawatt.com
Products Software Software for Testers
Protocol Software without Database →
ETC → myGMC → zum Login
ENTER: acknowledge selection + – change to sub-level
(open directory) or
+ change to main level
(close directory)
Display structure designation or ID number
Switch back and forth between structure designa­tion and ID number
Hide structure designation or ID number
Change display to menu selection
Memory menu, page 2 of 3
Add a structural element
Meaning of icons from top to bottom:
Customer, building, distributor, RCD and electrical cir­cuit (display of the icons depends on the selected
structural element). Selection: up/down scroll keys and In order to add a designation to the selected
structural element, refer to the edit menu below.
EDIT
For additional icons see edit menu below. Delete the selected structural element.
Show measurement data, if a measurement has been performed for this structural element.
24 GMC-I Messtechnik GmbH
Icon Meaning
Distributor
A check mark to the right of a structural element means that all measurements within the respective hierarchy have been passed.
x: At least one measurement has not been passed. No symbol: Measurement has not yet been performed.
Building
Customer
Same type of element as in the Windows Explorer:
+: Sub-object available, display by pressing ↵. –: Sub-objects are displayed, hide by pressing ↵.
Scroll up
Scroll down
Acknowledge selection /
Display object
Next page
change level
or ID number
Create object
Delete object
VΩA: show measurement data
Edit designation
Edit the selected structural element.
Memory menu, page 3 of 3
Search for ID number. > Enter complete ID number.
Search for text. > Enter full text (complete word).
Search for ID number or text.
Continue searching.
Edit menu
Cursor left: Select an alphanumeric character.
Cursor right: Select an alphanumeric character.
ENTER: Accept an individual character.
Acknowledge entry
Cursor left
Cursor right Delete characters.
Distributor Structure Symbology / Tree Structure

10.3.1 Creating Structures (example for electrical circuit)

After selection with the MEM key, all setting options for the cre­ation of a tree structure are made available on three menu pages (1/3, 2/3 and 3/3). The tree structure consists of structural ele­ments, referred to below as objects.
Select the position at which a new object will be added.
Switching amongst different types of alphanu­meric characters:
A Upper case letters
a Lower case letters
0Numbers
@ Special characters
Use the ↑↓ keys in order to select structural elements. Change to the sub-level with the key. Go to the next page with the >> key.
Create a new object.
GMC-I Messtechnik GmbH 25
Press the key in order to create a new object.
Select a new object from a list.
Scroll up
Scroll down
Acknowledge selection
Select character
Select character
Accept character
Delete characters
Character selection:
Save object designation
A, a, 0, @
Scroll up
Scroll down
Acknowledge selection /
Display object
Menu selection → page 3/3
change level
or ID number
Search for ID number
Search for text
Search for ID number or text
Select character
Select character
Accept character
Delete characters
Character selection:
Save object designation
Continue searching
Search for ID number
Search for text
Search for ID number or text
Search for ID number
Search for text
End search
Search for ID number or text
After selecting text search ...
Select the desired object from the list with the ↑↓ keys and acknowledge with the key.
Enter a designation (Example Distributor).
Enter a designation and then acknowledge it by entering a ✓.

10.3.2 Searching for Structural Elements

... and entering the desired text (only full matches are found – no wild cards, case sensitive) ...
... the first match is displayed. Further matches can be found by selecting the icon shown
at the right.
If no further matches are found, the message shown above is dis­played.
Mark the structural element from which the search will be started. Objects located underneath or next to this object will be included in the search.
Go to page 3/3 in the database menu.
26 GMC-I Messtechnik GmbH

10.4 Saving Data and Generating Reports

Note
Note
Preparing and Executing a Measurement Measurements can be performed and stored to memory for each
structural element. Proceed as follows, adhering to the prescribed sequence:
Select the desired measurement with the function selector
switch.
Start the measurement by pressing the START key. Upon completion of measurement, the “ Floppy Disk” softkey is
displayed. Briefly press the “Save Value” key.
The display is switched to the memory menu or the structural view.
Navigate to the desired memory location, i.e. to the desired
structural element / object, for which the measurement data will be saved.
If you would like to save a comment along with the
measurement, press the key shown at the right and enter a designation via the “EDIT” menu as described in section 10.3.1.
Complete data storage by pressing the “STORE” key.
Retrieving Saved Measured Values
Switch the display to the distributor structure by pressing the
MEM key and select the desired electrical circuit with the scroll keys.
Switch to page 2
by pressing the key shown here:
Display the measurement data
by pressing the key shown here:
One measurement with date and time, as well as any com­ment you might have entered, is displayed in each screen.
Example: RCD Measurement
Alternative Storage Procedure
The measured value can be saved to the last se-
lected object in the structural diagram by pressing and holding the “Save Value” key, without switching the display to the memory menu.
If you change the parameters in the measurement view, they are not saved for the structural element. A measure­ment with changed parameters can nevertheless be saved to the structural element, and any changed param­eters are documented in the report for each measure­ment.
An inverse displayed check mark in the header means that the respective measurement has been passed. An inverse displayed X means that the measurement has not been passed.
Scrolling amongst measurements
is possible with the keys shown here:
The measurement can be deleted with the key
shown here:
A prompt window asks you to confirm deletion.
With the help of the key shown at the right (MW: measured value / PA: parameter), the setting parame­ters can be displayed for this measurement.
Scrolling amongst measurements is possible with
the keys shown here:
GMC-I Messtechnik GmbH 27
Data Evaluation and Report Generation with ETC Software
Note
Note
All data, including the distributor structure, can be transferred to the PC and evaluated with the help of ETC software. Additional information can be entered here subsequently for the individual measurements. After pressing the appropriate key, a report including all measurements within a given distributor structure is generated, or the data are exported to an Excel spreadsheet.
The database is exited when the rotary selector switch is turned. Previously selected parameters in the database are not used for the measurement.

10.4.1 Use of Barcode Scanners and RFID Readers

Search for an Already Scanned Barcode
The search can be started from any switch setting and menu. Scan the object’s barcode.
The search is started based on the currently selected structural element down though lower hierarchical levels. The found bar­code is displayed inversely.
This value is accepted after pressing the ENTER key.
An already selected object cannot be found.
Continued Searching in General
Regardless of whether or not an object has been found, searching can be continued by pressing this key:
– Object found: Searching is continued underneath the previ-
ously selected object.
– No further object found: The entire database is searched at
all levels.
Reading In a Barcode for Editing
If the menu for alphanumeric entry is active, any value scanned by means of a barcode or RFID reader is accepted directly.
Using a Barcode Printer (accessory)
A barcode printer allows for the following applications:
• Read-out of ID numbers as encrypted barcodes; for quick and convenient acquisition for periodic testing
• Read-out of repeatedly occurring designations, such as test object types, encrypted as barcodes in a list, allowing them to be read in as required for comments.
28 GMC-I Messtechnik GmbH

11 Characteristic Values

METRISO XTRA
Meas.
Qty.
R
ISO
U
AC/DC
R
LO
R
1)
the indicated accuracy is only achieved with the optional accessory „shielded high-resistance measuring cable KS-C (article number Z541F)“
4)
up to 5 Ω
U
iSO
50 V
100 V
250 V / 500 V
Display range as of
01.0 Ω
Range Measuring Range
100 k 10 kΩ ... 99.9 kΩ 0.1 k
1 M 100 kΩ ... 999 kΩ 1 k
10 M 1.00 MΩ ... 9.99 MΩ 10 k
100 M 10.0 MΩ ... 99,9 MΩ 100 k
1000 V
1 G 100 MΩ ... 999 MΩ 1 M
10 G 1.00 GΩ ... 9.99 GΩ 10 M
100 G 10,0 GΩ ... 99.9 GΩ 100 M ±(8% rdg. + 3 d)
1 T 100 GΩ ... 999 GΩ 1 G ±(25% rdg. + 5 d) 1)±
100 V 10.0 V ... 99.9 V 0.1 V
1000 V 100 V ... 999 V 1 V
10 Ω 0.17 ... 9.99 Ω
100 Ω 10.0 ... 99.9 Ω 0.1 Ω
1 kΩ 100 ... 999 Ω 1 Ω
10 kΩ 1.00 ... 9.99 kΩ 10 Ω
Reso-
lution
0.01
Ω
Open-Circuit
Voltage
U
50 V /100 V:
1.25 U
250 V / 500 V / 1000 V:
1.1 U
4 V < U0 < 6 V
U0 max. 15 V
Test Current Intrinsic Uncertainty Measuring Uncertainty Overload Capacity
0max
ISO
IN = 1 mA
I
5 mA
K
ISO
——±(2.5% rdg. + 3 d) ±(5% rdg. + 3 d)
200 mA I
I ≤ 260 mA
1 mA I
I 1.3 mA
2)
does not conform to DIN EN 61557-2
3)
Display range up to 1.2 kV
±(5% rdg. + 3 d) ±(7% rdg. + 3 d)
1)
±(10% rdg. + 3 d)
(50% rdg. + 20 d)
±(2.5% rdg. + 3 d) ±(5% rdg. + 3 d)
4)
±(2.5% rdg. + 3 d) ±(5% rdg. + 3 d)
1)
1) 2)
1000 V AC/DC
TRMS
1000 V AC/DC
3)
TRMS
1000 V AC/DC
TRMS
1000 V AC/DC
TRMS
Breakdown Voltage (Uramp)
Parameter Range Intrinsic
Voltage range 100 1000 V Rise time 5 300 s — Measuring time 1 120 s /
uto / cont. measurement
a
Uncertainty
±(10% rdg. + 8 d)
——
Measuring Uncertainty
±(15% rdg. + 10 d)
Polarization Index (PI), Absorption Ratio (DAR)
t1 [min.] t2 [min.] Limit [min.]
PI 01:00 10:00 > 4.0 / > 3.0 / > 2.0 / > 1.5 / > 1.1 / > 1.0
DAR
00:30 01:00 > 1.60 / > 1.25
PI and DAR are calculated values. Insulation measurement specifications apply.
Reference Conditions
Reference temperature + 23 °C ± 3 K
Relative humidity 40 to 75% Measured quantity
frequency 45 to 65 Hz Measured quantity
waveshape Sine, deviation between TRMS and recti-
fied value < 1%
Battery voltage 9.5 V ±0.1 V Test resistance 10 MΩ±1%
Electrical Safety
Protection class II per IEC/EN 61010-1/VDE 0411-1 Pollution degree 2 Measuring category
CAT II 1000 V / CAT III 600 V / CAT IV 300 V Fuses Fuse link FF315mA/1000V, effective in all resis-
tance measuring ranges, 1 additional
replacement fuse in the battery compart-
ment Electronic fuse Protects low-resistance and resistance
measurements R
and R
LO
Electromagnetic Compatibility (EMC)
Interference emission EN 61326-1:2013, class B Interference immunity EN 61 326-1:2013
Power Supply
Batteries 8 ea. 1.5 V mignon cell (8 ea. size AA)
(alkaline manganese per IEC LR14) or 8 rechargeable NiMH batteries (must be recharged externally)
Z502R charger Broad-range charger with barrel connec-
tor, input: 100 to 240 V AC,
output: 16.5 V DC, 1 A (Mascot) Nominal range of use 8.5 to 12 V Battery test Battery capacity display with battery sym-
Battery saver circuit Automatic shutdown of display illumination
Service life For R
Safety shutdown If supply voltage is too low (U < 8 V), the
Recharging socket Installed rechargeable batteries can be
Charging time Approx. 2 hours *
* Maximum charging time with fully depleted rechargeable batteries.
A timer in the charger limits charging time to no more than 4 hours.
bol in 4 segments: .
Querying of momentary battery voltage via
menu function.
after 10 to 30 seconds (after the last time
the rotary switch is actuated) can be
selected in the setup menu (see page 8).
The test instrument is automatically
switched to the standby mode if the mea-
sured value remains unchanged for
approximately 15 minutes, and if none of
the controls are activated during this time.
The instrument is switched off automatically
if the measured value remains unchanged
for a long period of time and if none of the
keys or the rotary switch have been acti-
vated during on-time (specified in sec-
onds).
(1000 V / 1 MΩ) and RLO with
INS
25 s on-time and 1 subsequent measure-
ment each for a duration of 5 s
– With batteries (alkaline manganese):
700 measurements
– With rechargeable batteries (2000 mAh):
650 measurements
instrument is switched off, or cannot be
switched on.
recharged directly by connecting a charger
to the recharging socket:
charger Z502R
GMC-I Messtechnik GmbH 29
Displays
!
100
Ux/V
R
x
/kΩ
U
N
= 100 V
0
20
40
60
80
100
250
Ux/V
R
x
/kΩ
U
N
= 250V
0
50
100
150
200
250
0
100
200
300
400
500
500
U
N
= 500V
Ux/V
R
x
/kΩ
0
200
400
600
800
1000
1
U
N
= 1000V
Ux/V
R
x
/MΩ
Digital display Multiple display with dot matrix
Limit LED LED lights up red to indicate an exceeded
LED LED lights up red to indicate the presence
Uramp LED LED lights up green to indicate ramp
128 x 128 pixels, backlit (transflective), dimensions: 65 x 65 mm
limit value. LED lights up green to indicate adherence to the limit value.
of an external voltage (with the instrument switched off) or high test voltage during insulation mea­surement (Riso/Rins, PI and DAR) at the measuring terminals
sequence in progress. LED lights up red to indicate interrupted ramp sequence (e.g. in case of break­down).
Ambient Conditions
Accuracy temperature range 0 ... +40 °C
Operating temperature Storage temp. range –25 ... +70 °C (without batteries) Relative humidity Up to 75% (max. 85% during storage/
Elevation Max. 2000 m Calibration interval 1 year (recommended)
–10 ... +50 °C
transport), no condensation allowed
Table for determining maximum display values for low-value resistance in consideration of the instrument’s measuring uncertainty.
Limit Value
0.15 Ω 0.11 Ω
0.20 Ω 0.16 Ω 5.00 Ω 4.72 Ω
0.50 Ω 0.44 Ω 10.0 Ω 9.47 Ω
1.00 Ω 0.92 Ω 20.0 Ω 17.7 Ω
2.00 Ω 1.87 Ω 50.0 Ω 44.7 Ω
Maximum
Display Value
Limit Value
Maximum
Display Value
Voltage at Device Under Test During Insulation Resistance Measurement
Measuring voltage Ux at the device under test depending upon its resistance Rx at nominal voltages of 100, 250, 500 and 1000 V:
Mechanical Design
Dimensions 225 x 130 x 140 mm Weight Approx. 1.4 kg with batteries Protection Housing: IP 52, measurement cables and
connectors: IP 40 per DIN VDE 0470, part 1 / EN 60529, housing category 2
Excerpt from Table on the Meaning of IP Codes
(1
IP XY
st
digit X)
Protection Against Foreign
Object Entry
2 12.5 mm dia. 2 Dripping (at 15° angle) 3 2.5 mm dia. 3 Spraying water 4 1.0 mm dia. 4 Splashing water 5 Dust protected 5 Jet-water 6 Dust-proof 6 Powerful water jets
IP XY
(2nd digit Y)
Protection Against
Penetration by Water
Display Values in Consideration of Measuring Uncertainty
Table for determining minimum display values for insulation resistance in consideration of the instrument’s measuring uncertainty.
Limit Value
020 kΩ 025 kΩ 100 kΩ 111 kΩ 100 MΩ 111 MΩ 200 kΩ 219 kΩ 200 MΩ 219 MΩ 500 kΩ 541 kΩ 500 MΩ 541 MΩ
0.20 MΩ 0.25 MΩ
0.50 MΩ 0.57 MΩ
1.00 MΩ 1.11 MΩ 1.00 GΩ 1.11 GΩ
2.00 MΩ 2.19 MΩ 2.00 GΩ 2.19 GΩ
5.00 MΩ 5.41 MΩ 5.00 GΩ 5.41 GΩ
10.0 MΩ 11.1 MΩ 10.0 GΩ 11.1GΩ
20.0 MΩ 21.9 MΩ 20.0 GΩ 22.6 GΩ
50.0 MΩ 54.1 M
Minimum
Display Value
Ω 50.0 GΩ 55.9 GΩ
Limit Value
Minimum
Display Value
30 GMC-I Messtechnik GmbH

12 Maintenance

Note
Attention!
!
Attention!
!
Attention!
!
Attention!
!
Attention!
!
Attention!
!
BAT

12.1 Firmware Revision and Calibration Information

See section 3.4.

12.3 Fuses

If a fuse has blown due to overload, a corresponding error mes­sage appears at the display panel. The instrument’s voltage mea­suring ranges are nevertheless still functional.

12.2 Rechargeable Battery Operation and Charging

Check to make sure that no leakage has occurred at the rechargeable batteries in the Z502H battery pack at short, regular intervals, or after the instrument has been in storage for a lengthy period of time.
Prior to lengthy periods of rest (e. g. holiday), we recom­mend removing the (rechargeable) batteries. This helps to prevent excessive depletion or leakage of batteries, which, under unfavourable circumstances, may cause damage to the instrument.
If battery voltage has fallen below the allowable lower limit, the pictograph shown at the right appears. “Low Batt!!!” is also displayed along with a battery icon. The instrument does not function if the batteries have been depleted excessively, and no display appears.
Use only the Z502H battery pack.
Use only the Z502R charger (available as an accessory) to charge batteries which have already been inserted into the test instrument.
Make sure that the following conditions have been fulfilled be­fore connecting the charger to the charging socket:
– Rechargeable batteries have been installed with
correct polarity (not standard batteries).
– The test instrument has been disconnected from the
measuring circuit at all poles.
– The instrument must remain off during charging.
If the batteries or the battery pack have not been used or recharged for a lengthy period of time, thus resulting in excessive depletion:
Observe the charging sequence (indicated by the LED at the charger) and initiate a second charging sequence if necessary (disconnect the charger from the mains and from the test instru­ment to this end, and then reconnect it).

12.2.1 Charging Procedure with the Z502R Charger (accessory)

Insert the correct mains plug for your country into the charger.

12.3.1 Melting Fuse

This fuse is active in all resistance measuring ranges except for voltage measurement. A replacement fuse is included in the bat­tery compartment (FF315mA/1000V).
Checking the Fuse
If a resistance measuring range is selected with the rotary switch with a blown or defec­tive fuse in the instrument, and if measure­ment is started with the START or CONTIN. key, a pop-up window with the “blown fuse” icon appears. Prerequisite: The + and COM measurement jacks are not short circuited. This error message must be acknowledged and cleared by press­ing the ESC key.
Eliminate the cause of failure and replace the blown fuse.
Replacing the Fuse
Disconnect the instrument from the measuring circuit be­fore opening the battery compartment lid in order to re­place the fuse (refer to page 3 for location)! The rotary switch must be in the OFF position when the fuse is replaced.
Incorrect fuses may cause sever damage to the instru­ment. Only original fuses from GMC-I Messtechnik GmbH as­sure the required protection by means of suitable blow­ing characteristics. Short-circuiting of fuse terminals or the repair of fuses is prohibited! The instrument may be damaged if fuses with incorrect ampere ratings, breaking capacities or blowing charac­teristics are used!
Open the battery compartment lid by loosening the two
screws.
Remove the defective fuse and replace it with a new one. A
replacement fuse is included in the battery compartment.
Insert the new fuse.Replace the battery compartment lid and retighten the
screws.
Make sure that rechargeable batteries have been in­serted (not normal batteries). We recommend the use of rechargeable NiMH batteries (eneloop type).
Connect the charger to the test instrument with the barrel
connector, and then to the 230 V mains with the interchange­able plug.
Do not switch the test instrument on during charging. Monitor­ing of the charging process by the microprocessor might otherwise be disturbed, in which case the charging times specified in the technical data can no longer be assured.
Please refer to the operating instructions included with the
charger regarding the meanings of LED displays during the charging process.
Do not disconnect the charger from the test instrument until
the LED lights up green.
GMC-I Messtechnik GmbH 31
This fuse protects low-resistance (Rlo) and resistance measure­ments (kΩ) from overloading (electronic hardware circuit).
A pop-up window appears when the fuse blows.
This error message must be acknowledged and/or cleared by pressing any desired key.
Eliminate the cause of overloading.

12.3.2 Electronic Fuse

12.4 Housing

Pb Cd Hg
L1
L2
L3
N
PE
Between each active conductor and ground
L1
L2
L3
PEN
Between each active conductor and the PEN conductor
L1
L2
L3
N
PE
Between each active conductor (phase and neutral conductors) and ground
L1
L2
L3
PEN
Between each active conductor and the PEN conductor
No special maintenance is required for the housing. Keep outside surfaces clean. Use a slightly dampened cloth for cleaning. In par­ticular for the protective rubber surfaces, we recommend a moist, lint-free microfiber cloth. Avoid the use of cleansers, abrasives or solvents.

12.5 Return and Environmentally Sound Disposal

The instrument is a category 9 product (monitoring and control instrument) in accordance with ElektroG (German electrical and electronic device law). This device is subject to the RoHS direc­tive. Furthermore, we make reference to the fact that the current status in this regard can be accessed on the Internet at www.gossenmetrawatt.com by entering the search term WEEE.
In accordance with WEEE 2012/19EU and ElektroG, we identify our electrical and electronic devices with the symbol in accordance with DIN EN 50419 which is shown at the right. Devices identified with this symbol may not be disposed of with the trash. Please contact our service department regarding the return of old devices (see address in section 14).
If the batteries used in your instrument or accessory product are depleted, they must be disposed of properly in accordance with valid national regulations. Batteries may contain pollutants and heavy metals such as lead (Pb), cadmium (Cd) and mercury (Hg).
The symbol at the right indicates that batteries must not be disposed of with the trash, and must be brought to a designated collection point.

13 Appendix

13.1 Sample Connection Layouts for Insulation Resistance Measurement

Insulation Resistance Measurement per DIN VDE 0100, Part 600
32 GMC-I Messtechnik GmbH
Insulation Resistance Measurement in Different Types of Systems
TN–S
• With or without consumer
•N PE, separately
TN–C
• With or without consumer
• N – PE, together
TT
• With or without consumer
3-phase current (switched at all poles)
Attention:
– Open the overcurrent protective device – Disconnect the N conductor – Jumper the L and N conductors – Insulation measurement between L conductors and N to PE – Device switch can be open – if single-pole
In the distributor: individual and combined measurements
GMC-I Messtechnik GmbH 33

13.2 Error Messages

The following pop-up windows must be acknowledged and/or cleared by any desired key and the cause of the error must be eliminated.
Pop-Up Function Switch
Function / Meaning
Position
Interference voltage
All except U
R
LO
Remedy: Device under test must be disconnected from all sources of voltage.
OFFSET measurement does not make sense. Remedy: Check system. OFFSET measurement of RLO+ and RLO– is still possible.
R
INS
R
LO
/ kΩ / R
R
/ kΩ
LO
All
R
OFFSET
> 50 Ω:
OFFSET measurement does not make sense. Remedy: Check system.
Externally accessible fuse is blown. The voltage ranges remain functional even if the fuse has blown.
LO
Special case, R
blown fuse.
: Interference voltage during measurement may result in a
LO
Remedy: Replace fuse (replacement fuse in battery compartment).
Observe notes regarding fuse replacement in section 12.3.1.
The electronic fuse protects low-resistance (Rlo) and resistance measure­ments (kΩ) from overloading (electronic hardware circuit).
The voltage ranges remain functional even if the fuse has blown. Remedy: This error message must be acknowledged and cleared by
pressing the ESC key. Battery voltage is less than or equal to 8 V. Reliable measurement is no longer possible. Storage of measured values to memory is disabled. Remedy: Rechargeable batteries must be recharged, or replaced towards
the end of their service life.
34 GMC-I Messtechnik GmbH
Pop-Up Function Switch
Position
Database and Entry Operations
All
All Please enter a designation (alphanumeric).
All
All
Function / Meaning
The parameters you have selected do not make sense in combination with previously configured parameters. The selected parameter settings will not be saved.
Remedy: Enter other parameter settings.
Operation with a Barcode Scanner Error message appears when the “EDIT” entry field is opened and battery
voltage is less than 8 V. Output voltage is generally switched off during barcode scanner operation if U is less than 8 V in order to assure that remaining battery capacity is adequate for entering designations for devices under test and saving the measurement.
Remedy: Rechargeable batteries must be recharged, or replaced towards the end of their service life.
Operation with a Barcode Scanner Current flowing through the RS 232 port is too high. Remedy: The connected device is not suitable for this port.
All
All
All
All
All
SETUP
Operation with a Barcode Scanner Barcode not recognized, incorrect syntax
Data cannot be entered at this location within the structure. Remedy: Observe profile for preselected PC software (see SETUP menu).
Measured value cannot be saved at this location within the structure. Remedy: Make sure that you have selected the right profile for you PC
evaluation program in the SETUP menu (see section 3.4).
Memory is full.
Remedy: Save your measurement data to a PC and then clear memory at the test instrument by deleting the database or by importing an empty database.
Delete measurement or database.
This prompt window asks you to confirm deletion.
Data loss after changing language or profile, or after restoring default settings.
Back up your measurement data to a PC before pressing the respective key.
This prompt window asks you to confirm deletion.
GMC-I Messtechnik GmbH 35

13.3 Attaching the Test Probe Holder to the Carrying Strap 13.4 Technical Data for Measurement Cables

Attention!
!
1
Bottom Right
Detach the strap from the instrument: Turn out the slotted screw (M3) at the bottom.
Bottom Left
Strap
Clasp
Clasp
Front View
2
Slide the strap through the test probe holder.
Eyelet for Attachment to Tester
Side View
(METRISO INTRO, BASE, TECH, PRO, XTRA)
3
Front
Feed the strap through from the front of the test instrument and secure it with the slotted screw (M3).
Back
(scope of delivery: KS17-4 safety cable set)
Electrical Safety
Maximum rated voltage 600 V
Measuring category
Max. rated current: 1 A 1 A 16 A With safety cap attached Without safety cap ——
CAT IV CAT III CAT II
••
Ambient Conditions (EN 61010-031)
Temp erature – 20 °C ... + 50 °C Relative humidity Max. 80% Pollution degree
2
Using the KS17-4
Observe the instrument’s maximum values for electrical safety.
Measurements per DIN EN 61010-031 may only be per­formed in environments in accordance with measuring categories III and IV with the safety cap attached to the test probe at the end of the measurement cable.
In order to establish contact inside 4 mm jacks, the safety caps have to be removed by prying open the snap fastener with a pointed object (e.g. the other test probe).
1000 V
1000 V
36 GMC-I Messtechnik GmbH

13.5 Optional Accessories (not included)

Master Battery Pack (material no. Z502H)
8 LSD NiMH rechargeable batteries with reduced self­discharging (AA) (eneloop/Sanyo), 2000 mAh with sealed cells
Charger (material no. Z502R)
Broad-range charger for charging batteries in the mea­suring instrument Input: 100 to 240 V AC, output: 16.5 V DC, 0.6 A
IS0 Calibrator 1 (material no. M662A)
Calibration adapter for testing the accuracy of instru­ments used for measuring insulation resistance and low-resistance for test voltages of up to 1000 V (per VDE 0413, parts 1, 2, 4 and 10).
KS-C (material no. Z541F)
Cable set consisting of measurement cable and shielded high-resistance measurement cable for mea­surements in the GΩ range
1081 Probe (material no. GTZ3196000R0001)
Triangular probe for floor measurements per EN 1081, DIN VDE 0100-600 (standing surface insulation)
KS24 (material no. GTZ3201000R0001)
Cable set consisting of a 4 m long extension cable with a permanently attached test probe at one end and a contact protected socket at the other end, and 1 alli­gator clip which can be plugged onto the test probe.
Test Probe for Remote Triggering
(material no. Z550A) Optional plug-on measurement cable with a triggering key on the test probe and an additional key for illumi­nating the measuring point, including shielded, plug-in connector cable.
Barcode-Profiscanner-RS232
(material no. Z502F) Barcode scanner for RS 232 connection (laser sen­sor), variable barcode length, enhanced reading accu­racy, with coil cable
Barcodeleser B3261
(material no. GTZ3261000R0001) Barcode scanner for RS 232 connection (LED sensor)
SCANBASE RFID
(material no. Z751G) RFID read/write for RS 232 port (13.56 MHz)
GMC-I Messtechnik GmbH 37
14 Repair and Replacement Parts Service
Calibration Center* and Rental Instrument Service
If required please contact:
GMC-I Service GmbH Service-Center Thomas-Mann-Strasse 16-20 90471 Nürnberg • Germany Phone: +49 911 817718-0 Fax: +49 911 817718-253 E-mail service@gossenmetrawatt.com www.gmci-service.com
This address is only valid in Germany. Please contact our repre­sentatives or subsidiaries for service in other countries.
* DAkkS Calibration Laboratory for Electrical Quantities
D-K-15080-01-01 accredited per DIN EN ISO/IEC 17025
Accredited quantities: direct voltage, direct current value, direct current resistance, alternating voltage, alternating current value, AC active power, AC apparent power, DC power, capacitance, frequency and temperature
Competent Partner
GMC-I Messtechnik GmbH is certified in accordance with DINENISO9001.
Our DAkkS calibration laboratory is accredited by the Deutsche Akkreditierungsstelle GmbH (National accreditation body for the Federal Republic of Germany) in accordance with DIN EN ISO/IEC 17025 under registration number D-K-15080-01-01.
We offer a complete range of expertise in the field of metrology: from test reports and proprietary calibration certificates right on up to DAkkS calibration certificates.
Our spectrum of offerings is rounded out with free test equipment management.
An on-site DAkkS calibration station is an integral part of our service department. If errors are discovered during calibration, our spe­cialized personnel are capable of completing repairs using original replacement parts.
As a full service calibration laboratory, we can calibrate instru­ments from other manufacturers as well.

15 Recalibration

The measuring tasks performed with your instrument, and the stressing it’s subjected to, influence aging of its components and may result in deviation from the specified levels of accuracy.
In the case of strict measuring accuracy requirements, as well as in the event of use at construction sites with frequent stress due to transport and considerable temperature fluctuation, we recom­mend a relatively short calibration interval of once per year. If your instrument is used primarily in the laboratory and indoors without considerable climatic or mechanical stressing, a calibration inter­val of once every 2 to 3 years is sufficient as a rule.
During recalibration* at an accredited calibration laboratory (DIN EN ISO/IEC 17025), deviations from traceable standards demonstrated by your measuring instrument are documented. Ascertained deviations are used to correct displayed values during later use of the instrument.
We would be happy to perform DAkkS or factory calibration for you at our calibration laboratory. Further information is available at our website:
www.gossenmetrawatt.com ( Company DAkkS Calibration Center or FAQs Questions and Answers Regarding Calibra­tion).
Recalibration of your instrument at regular intervals is essential for the fulfillment of requirements according to quality management systems per DIN EN ISO 9001.
* Examination of the specification, as well as adjustment, are not included
in calibration. However, in the case of our own products, any required adjustment is performed and adherence to the specification is con­firmed.

16 Product Support

If required please contact:
GMC-I Messtechnik GmbH
Product Support Hotline
Phone: +49-911-8602-0 Fax: +49 911 8602-709 E-mail: support@gossenmetrawatt.com
Edited in Germany • Subject to change without notice • PDF version available on the Internet
GMC-I Messtechnik GmbH Südwestpark 15 90449 Nürnberg
Germany
Phone: +49 911 8602-111 Fax: +49 911 8602-777 E-mail: info@gossenmetrawatt.com www.gossenmetrawatt.com
Loading...