Analog Insulation, Low Resistance and Voltage Measurement Instrument
3-349-815-03
1/9.14
Overview of Included Features Scope of Delivery
ab
Replacement Fuse
Fuse Link
Battery Compartment Lid
METRISOPRO
Article numberM550R
Measurements:
U = 50, 100, 250, 500, 1000 V
R
INS
R
LO
U10 ... 1000 V
Display Functions:
Scale illumination
Limit value LED (green/red) for:
additional acoustic signal,
limit values per VDE 0100
Caution LED
for dangerous contact voltage
– When switched off
– For active test voltage
FUSE LED
For blown fuse
Battery level display
Special Functions:
Discharge capacitive devices under test
Safety shutdown (UBatt < 8 V)
0.17 ... 4
R
INS RLO
3
3
3
3
3
3
3
3
3
1 Insulation and resistance measuring instrument
1 Factory calibration certificate
1 Set of batteries (8 pieces in battery holder)
1Carrying strap
1 Alligator clip
1 KS17-4 cable set
1 Condensed operating instructions
1 Supplement Safety Information
– Detailed operating instructions for download from our website
at www.gossenmetrawatt.com
Battery Compartment Lid and Location of the Fuses
(housing bottom)
Features:
Measuring category
CAT II 1000 V / CAT III 600 V / CAT IV 300 V
Test resistor: 10 M
Factory calibration certificate
2GMC-I Messtechnik GmbH
3
3
3
Inserting the Battery Holder (side view) Connections (housing top)
Battery Holder
Battery Compartment Lid
Battery Holder
Contacts
Battery
Contact
Spring
User Interface
Compartment
Shielded
Cable
Te st
Resistor
Tes t
Resistor
Test Pro b e w i th
Control Keys
1
Connector with
Shielded Cable,
Option KS-C (Z541F)
(G500/G10
G500/G1000)
Measuring Connections
+
COM
+
COM
10 M
GUARD
SHIELD
Guard connection
see section 5.4
GMC-I Messtechnik GmbH3
1
Z550A accessory as option: test probe with measurement key and an additional
key for illuminating the measuring point, including shielded, plug-in connector
cable
User Interface
Red LED: limit value violated
Red LED:
– Interference voltage with device off
Function
Selector Switch
Guide for
Carrying Strap
Guide for
Carrying Strap
Green LED: limit value complied with
– Blinking indicates standby mode
Control Panel Keys
(see description on page 5)
Display Panel
(see description on page 5)
Refer to page 10 for a description of the LEDs.
– Test voltage active
4GMC-I Messtechnik GmbH
abcd
e
1
2
4
5
63
➀
➁
➃
➅
Control Panel Keys – Adjusting Screw
a START: Starts insulation resistance measurement depending
on rotary selector switch setting (see measuring ranges
➀ and
Analog Display Symbols
➁), or low-resistance measurement with automatic polarity
reversal (see low-resistance measuring range
Reactivates the instrument from the standby mode.
b CONTIN./ZERO*
Starts long-term measurement for insulation and low resistance (low resistance: polarity in one direction only).
Each measurement takes up to 3 minutes; the instrument is
then switched to standby. Reactivates the instrument from the
standby mode.
c Bat. TEST: Starts the battery test (see battery level display
and section 3.1).
d 10k-1T/100k-100M
Switches between measuring ranges for insulation resistance
measurement (see measuring ranges
tion 5.2).
Reactivates the instrument from the standby mode.
➀
and ➁, as well as sec-
Mechanical Zero Point – Adjusting Screw (e)
If necessary, use the adjusting screw (e) to set the pointer to the
mechanical zero point (measuring range lower limit for all measuring ranges, except battery level indicator) with the switch in the
OFF position.
Electrical Zero Point – ZERO Function (Roffset)
ZERO function (Roffset): for subtracting cable resistance from the measurement results with the help of the CONTIN. and 10k-1T/
100k-100M keys (see section 7.3).
*the supplementary labelling „ZERO“ depends on the product series
9Technical Data ........................................................................19
6GMC-I Messtechnik GmbH
1Applications
These instruments fulfills all requirements of applicable European
and national EC directives. We confirm this with the CE mark. The
relevant declaration of conformity can be obtained from GMC-I
Messtechnik GmbH.
The METRISO PRO insulation and resistance measuring instrument allows for quick and efficient testing of protective measures
in accordance with DIN VDE 0100, ÖVE-EN 1 (Austria), SEV 1000
(Switzerland), and regulations specific to other countries as well.
The device is equipped with a microprocessor and complies with
IEC/EN 61557 / VDE 0413 regulations:
Part 1:General requirements
Part 2:Insulation resistance measuring instruments
Part 4:Instruments for measuring resistance at
ductors, protective conductors and equipotential bonding
Part 10: Combined measuring equipment for testing, measuring
or monitoring protective measures
as well as requirements per VDE 0701-0702:
Repair, modification and testing of electrical devices
earthing con-
The test instrument is especially well suited for:
•Systems setup
• Initial start-up
• Periodic testing
• Troubleshooting in electrical systems
The following measurements and tests can be performed with the
insulation measuring instruments:
• Insulation Resistance
• Low-resistance
• Voltage
The following can also be tested by using a shielded measurement cable:
The electronic measuring and test instrument is manufactured
and tested in accordance with safety regulations IEC/EN 610101/VDE 0411-1 and EN 61557. When used for its intended purpose, safety of the operator, as well as that of the instrument, is
assured.
Read the operating instructions thoroughly and carefully before
using your instrument. Follow all instructions contained therein.
The measuring and test instrument may not be placed into service:
• If the battery compartment lid has been removed
• If external damage is apparent
• If connector cable or measuring adapters are damaged
• If the instrument no longer functions flawlessly
• After extraordinary damage due to transport
• After a long period of storage under unfavorable conditions
(e.g. humidity, dust or extreme temperature)
Opening of Equipment / Repair
The equipment may be opened only by authorized service personnel to ensure the safe and correct operation of the equipment
and to keep the warranty valid.
Even original spare parts may be installed only by authorized service personnel.
In case the equipment was opened by unauthorized personnel,
no warranty regarding personal safety, measurement accuracy,
conformity with applicable safety measures or any consequential
damage is granted by the manufacturer.
Meanings of Symbols on the Instrument
Warning concerning a point of danger
(attention, observe documentation!)
Protection class II device
CAT II / III Device assigned to measuring category
CAT II 1000 V / CAT III 600 V / CAT IV 300 V
EC mark of conformity
The device and included batteries may not be disposed of with the trash. Further information regarding
the WEEE mark can be accessed on the Internet at
www.gossenmetrawatt.com by entering the search
term “WEEE”.
8GMC-I Messtechnik GmbH
3Initial Start-Up
Note
Attention!
!
Attention!
!
Attention!
!
3.1Battery Test
A battery test should be conducted after inserting the batteries, or
if none of the measuring range LEDs lights up during insulation
resistance measurement. Press and hold the Bat. TEST key to this
end. Pointer deflection within the scale for the battery test indicates the condition of the batteries with an average load amounting to a test voltage of 1000 V. The rotary selector switch position
is not taken into consideration. The left-hand scale limit represents minimum required supply voltage, and the right-hand limit
represent maximum available supply power.
If the pointer is only deflected into the minimum supply
power range, several measurements can still be executed
with test voltages of less than 1000 V, because the battery test is conducted with a load which amounts to a test
voltage of 1000 V.
The instrument does not function if the batteries have been
depleted excessively, and alarm indication does not work either.
3.2Installing or Replacing Batteries
New batteries must be inserted for initial start-up, or if only minimal supply power is indicated.
Eight 1.5 V size AA batteries in accordance with IEC LR 6 are
required for operation of the insulation measuring instrument. Use
new alkaline manganese batteries only.
Rechargeable NiCd or NiMH batteries may also be used. These
can only be recharged externally. We recommend rechargeable
NiMH batteries.
Always replace batteries in complete sets.
Dispose of batteries in an environmentally sound fashion.
Ð Loosen both slotted screws for the battery compartment lid
on the back, and remove the lid.
Ð Remove the battery holder and insert eight 1.5 V size AA bat-
teries with correct polarity in accordance with the symbols.
Make sure that all of the batteries are inserted with correct polarity. If just one battery is inserted with reversed polarity, it will not be recognized by the instrument and may
result in leakage from the batteries.
Ð
Push the battery holder into the battery compartment such that
the battery holder’s contacts touch the contact springs at the
bottom of the battery compartment (see drawing on page 3).
If the battery holder is not inserted as specified, the instrument
cannot be supplied with power.
Ð Replace the battery compartment lid and retighten the
screws.
Before opening the battery compartment (see page 5 for
location), disconnect the instrument from the measuring
circuit (mains) at all poles.
GMC-I Messtechnik GmbH9
The instrument may only be placed into service if the
battery compartment lid is securely fastened!
4General Operation
KS-C
(Z541F)
+
C
O
M
S
H
I
E
L
D
G
U
A
R
D
Limit
Green
Limit
Red
!
Red
The test leads are connected to the “+” and “COM” jacks.
When measuring electrostatic
discharge capacity for floor
coverings, the shielded cable
should also be connected to
the COM and SHIELD jacks (KS-C
accessory set, “cable set consisting of measurement cable
and high-resistance measurement cable for measurements
in the G range” (see diagram). Be sure to observe color coding.
4.1Switching On, Monitoring and Switching Off
The instrument is switched on as soon as the rotary switch is
turned to any position other than OFF.
If battery voltage falls below the allowable limit value (U < 8 V), the
instrument cannot be switched on, or it is immediately switched
off.
Measurement cannot be started in the resistance measuring
ranges in the event of interference voltage.
The instrument is switched to the standby mode after 3 minutes,
if neither the rotary switch nor any keys have been activated during this time.
A brief acoustic signal is generated before the instrument
switches itself off automatically with the rotary switch in a position
other than OFF. Moreover, 6 brief acoustic signals indicate every
10 minutes that the device is still on standby. The device should
be switched off by setting the rotary selector switch to OFF in
order to save (rechargeable) batteries.
The instrument can be activated once again by pressing the START
key.
The instrument can be switched off manually by turning the rotary
switch to the OFF position.
For purposes of transport and maintenance, we recommend turning the rotary selector switch to the OFF position in order to avoid
switching the test instrument on inadvertently.
Start-Up Test
When the instrument is switched on, i.e. as soon as the rotary
switch is turned to any position other than OFF, all LEDs light up
briefly: limit, caution, blown fuse, insulation resistance measuring
ranges, voltage measurement and scale illumination.
Optical Indicators
LEDStatus Function – Cause
Limit value indication
–Measured insulation resistance does not violate the limit
value.
– Measured low-resistance Rlo does not violate the limit
value.
Limit value indication
–Measured insulation resistance has fallen short of the limit
value.
–Measured low-resistance Rlo has exceeded the permissible
limit value.
Detection of interference voltage in the off state
If dangerous voltage of greater than 50 V is present at the measure-
ment inputs:
–Initialization of the insulation resistance and
low-resistance measurements is disabled.
–Presence of test voltage is indicated.
1
Function testing should be executed at regular intervals (see following section regarding testing the LEDs).
1
10GMC-I Messtechnik GmbH
Testing the LED which Indicates Detection of Interference Voltage
when Switched Off – OFF Switch Position
Ð
Apply a voltage of greater than 50 V (+ and COM jacks).
Ð Turn the rotary switch to the V position.
Ð Read the voltage value at the analog display.
Ð Turn the rotary switch to the OFF position.
Test results: If applied voltage is unchanged and the LED which
indicates the detection of interference voltage lights up red, the
LED is OK. In this case, the LED reliably indicates interference
voltage even when the instrument is switched off. We recommend executing this test at regular intervals.
Limit Values for Insulation and Low-Resistance Measurements
Limit R
/ R
Limit R
ISO
LO
INS
100 k @ U
500 k @ U
2
50 k @ U
1M @ U
1M @ U
ISO/UINS
ISO/UINS
ISO/UINS
ISO/UINS
ISO/UINS
= 50 V
= 100 V
= 250 V
= 500 V
= 1000 V
Limit LEDs for Insulation Resistance and Low-Resistance
Measuring Ranges
The LimitLED lights up green if the measured insulation resistance
value does not violate the limit value. The Limit LED lights up red, if
the measured insulation resistance value violates the limit value. If
the LimitLED does not light up at all, this means that the selected
test voltage value has not been reached. A battery test is advisable in this case.
4.2Analog Display
Insulation Resistance Measuring Ranges
The logarithmic representation of the upper resistance scale
allows for quick recognition of the order of magnitude.
In order to achieve the required accuracy during protective measures testing, the 10k-1T/100k-100M key makes it possible to
switch to the expanded lower measuring range of 100 k
100 M.
The two orange LEDs,
➀ and ➁, indicate which of the two resis-
tance measuring ranges is active at the moment.
Low-Resistance Measuring Range
LED ➃ lights up orange to indicate that low-resistance measurement is active. LED
➃ blinks when the measuring range is
exceeded.
Voltage Measuring Ranges
The two lower scales are for voltage measurement and the battery test (see section 3.1). LED
➃ lights up orange in order to indi-
cate that voltage measurement is active.
Measuring Range Overflow in the Resistance Measuring Ranges
The LED assigned to the respective measuring range, namely ➀,
➁ or ➃ blinks orange in the event of measuring range overflow.
Scale Illumination
Scale illumination is active for about 15 seconds after switching
the instrument on, after a measurement is started and after
changing measuring ranges.
GMC-I Messtechnik GmbH11
5Insulation Resistance Measurement
Note
Note
Attention!
!
!
– Riso/Rins Function
5.1Connection
Checking the Measurement Cables
Before performing insulation measurement, the test
probes on the measurement cables should be short-circuited in order to assure that the instrument displays a
value very close to 0 (see section 7).
This makes it possible to detect interrupted measurement
cables, which simulate high insulation resistance.
Ð Connect the device under test to the + and COM jacks.
Sample connection layouts for insulation resistance measurement
are included in section 13.1.
Insulation resistance can only be measured at voltagefree objects. The measurement cannot be started until
the test probes are in contact with the device under test.
If line voltage or interference voltage of greater than 50 V
is applied to the measurement inputs, this is indicated by
means of a double acoustic signal. Insulation measurement cannot be started as long as interference voltage is
present.
5.2Executing the Measurement
Note: Condensation must be ruled out when performing measure-
ments at close to the freezing point.
Ð Select the measuring function, as well as the desired test volt-
age depending on the device under test, with the rotary
switch, e.g. R
iSO
100 V.
Do not touch the instrument’s terminal contacts or the
conductive ends of the two test probes during insulation
resistance measurement!
If nothing has been connected to the terminal contacts, or if a
resistive load component has been connected for measurement,
your body would be exposed to a maximum short-circuit current.
The resulting electrical shock is not life endangering. However, the
noticeable shock may lead to injury (e.g. resulting from a startled
reaction etc.).
Ð Start an individual measurement by briefly pressing the START
key, or initiate continuous measurement by briefly pressing
the CONTIN./ZERO key.
Ð Contact the measuring point with both test probes.
A double acoustic signal indicates that measurement has been
started.
The Caution! LED remains illuminated for as long as test voltage
is applied to the test probes.
LED
➀ lights up to indicate that the upper measuring range of
10 k 1T has been complied with (default setting).
Ð Wait until the pointer settles in. Depending on the device
under test, this may take anywhere from several seconds to
half a minute if, for example, large capacitors (long cables)
have to be charged.
Ð Read the measured value from the upper voltage scale
If the Limit LED lights up green, the insulation measurement is
valid and the limit value has not been violated.
If the limit value is fallen short of, the Limit LED lights up red.
➀.
12GMC-I Messtechnik GmbH
Note
If the Limit
Note
Note
+
COM
+
COM
SHIELD
GUARD
L1
L2
N
PE
L3
R
B
R
E
R
INS
R
x
0V !
!
R
INS
R
INS
1
2
➀
➁
!
selected test voltage value has not been reached. A battery
test is advisable in this case (see section 3.1 on page 9).
Ð Switching to the Measuring Range with Higher Resolution
In order to perform measurement with the enhanced accuracy
required for protective measures testing, select the measuring
range with higher resolution, namely 100 k 100 M, with
the help of the 10k–1T/100k–100M key immediately after the
measured value is displayed. LED
measuring range, lights up.
LED
does not light up at all, this means that the
➁, which is assigned to this
Ð Read the measured value from the lower voltage scale
If the Limit LED lights up green, the insulation measurement is
valid and the limit value has not been violated. If the limit value is
fallen short of, the Limit LED lights up red.
After test voltage is switched off – Caution! LED goes out
(U > 50 V) – the pointer remains frozen in place for about 3 sec-
GMC-I Messtechnik GmbH13
onds. When the Limit LED and the LED for measuring range
both go out, the measurement has been completed and the
pointer returns to its neutral position.
Three-Phase Systems
All conductors (L1, L2, L3 and N) must be tested against
PE!
The instrument’s batteries are exposed to excessive
stress during insulation resistance measurement. For this
reason, it’s advisable to perform individual rather than
continuous measurements.
➁.
➁
Special Case: Capacitive Devices Under Test
Caution!
Note
!
!
+Measurement Cable
–Measurement Cable
Guard Cable
Conductor
Insulation
Material
Guard Rings
Contact Ring
(COM)
A guard cable must be used for measurements within a range of
100 G (10 G) 1T, in order to prevent surface current from
distorting measurement results. The guard rings prevent current
at the surface of the insulation material from flowing from the
If measurement is performed at a capacitive object such
as a long cable, it becomes charged with up to approximately 1000 V (test voltage)! Touching such objects is life
endangering!
+measurement cable to the –measurement cable, instead of
through the insulation material itself.
Ð Insert the plug from the guard cable into the appropriate jack in
the test instrument.
Ð Plug the alligator clip onto the guard cable test probe.
5.3Ending the Measurement – Safe Discharging
Individual measurement: Measurement is ended automatically as
soon as the measured value has settled in.
Continuous measurement is ended by once again briefly pressing
Ð Connect the alligator clip to the guard ring between the two
measuring points at the insulation material under test.
Ð Refer to section 5.2 on page 12 regarding the measuring se-
quence.
the CONTIN./ZERO key.
When an insulation resistance measurement has been performed
on a capacitive object it’s automatically discharged by the instrument. Contact between the object and the instrument must nevertheless not be interrupted.
Do not disconnect until:
•The Caution! LED (U > 50 V) is no longer illuminated.
As an alternative,
the device under test can be discharged in a
controlled manner (due to touch hazard) by switching to V and
contacting the device under test until the Caution! LED is no
longer illuminated and the measuring movement indicates 0 V.
Do not reverse polarity while discharging, because internal surge
protection will otherwise be triggered.
5.4Measurements with the Guard Cable
The measurement of very high resistances necessitates extremely
minimal measuring current and may be rendered problematic as a
result of influences such as electromagnetic fields, humidity or
The following materials can be used as guard rings: aluminum foil, copper foil or metallic hose clamps.
surface pollution. An accurate test set-up is thus absolutely
essential.
14GMC-I Messtechnik GmbH
6Measuring Direct, Alternating and
Attention!
!
Note
!
+
COM
+
COM
SHIELD
GUARD
L1
L2
N
PE
L3
R
B
R
E
Max.
1.2 kV
U
V
4
Pulsating Voltage – V Function
You can measure direct voltage, as well as sinusoidal alternating voltage with frequencies ranging from 45 to 65 Hz with this test instrument
Pointer deflection at the measuring instrument is always positive
for the measurement of direct voltage, regardless of the polarity of
the connections. Alternating voltage is always indicated as an
RMS value.
Voltage measurement is used to test for the absence of voltage
before performing the insulation resistance measurement, as well
as for automatic discharging of capacitive devices under test. The
falling voltage value can be observed at the display.
Ð Select the V measuring function with the rotary switch, LED ➃
lights up permanently.
Ð Connect the measurement cables to the + and COM jacks.
Ð Contact the measuring point with both test probes.
A double acoustic signal is generated and the Caution!
LED lights up red in order to indicate the presence of
voltage of greater than 50 V.
Ð Read the measured value from the V scale. The measured
value is displayed directly – without pressing the START key.
Ð After completing the measurement, switch the instrument off
by turning the rotary switch to the OFF position.
No more than the maximum permissible voltage of
1000 V may be applied. Input impedance for the voltage
measuring range is roughly 10 M.
The START and CONTIN./ZERO keys have no function in this case.
.
GMC-I Messtechnik GmbH15
7
Attention!
!
Note
+
COM
+
COM
10 M
GUARD
SHIELD
R
LO
R < 5
R
LO
R
x
0V !
!
R
LO
4
➃
Measuring Low-Resistance of up to 5
– RLO Function
According to the regulations, the measurement of low-resistance at
protective conductors, earth conductors or equipotential bonding
must be performed with (automatic) polarity reversal of the test voltage, or with current flow in one direction, and then in the other.
In the measuring function RLO, measurements are performed with currents higher than 200 mA.
Please check before measuring whether your DUT or
your circuit is designed for these high current values.
Connection
Low-resistance can only be measured at voltage-free
objects.
Ð Connect the device under test to the + and COM jacks.
Ð Select the Rlo measuring function with the rotary switch.
Attention: The measurement cannot be started until the test
probes are in contact with the device under test.
If line voltage or interference voltage of greater than 50 V is
applied to the measurement inputs, this is indicated by means of
a double acoustic signal. Low-resistance measurement cannot be
started as long as interference voltage is present. The display is
not returned to low-resistance measurement until voltage is less
than approximately 8 V. If resistance is greater than 5 , LED ➃
blinks.
Measurement Types
You can choose one of two different types of measurement:
• Measuring sequence with automatic polarity reversal (reversal
of current flow direction) via the START key
• Manual measurement with positive polarityvia the CONTIN./
ZERO key
16GMC-I Messtechnik GmbH
7.1Measurement with Automatic Polarity Reversal
Limit
< Limit R
LO
= OK 3
Ð Start measurement in both current flow directions by briefly
pressing the START key, the LED ➃ lights up during measurement.
After the measuring sequence has been started, the instrument
performs measurement with automatic polarity reversal, first with
current flow in one direction (), and then in the other ().
Measurement is executed in the background. During both measurements, the pointer briefly moves back towards 0 , indicating
to the user that measurement is being performed with both polarities in the background.
The respectively larger (worse) measured value is displayed when
measurement has been completed (after the relay has audibly
switched two times).
The Limit LED lights up in order to indicate that the measured
value is valid. When the Limit LED and ➃ go out, the measurement has been completed and the pointer returns to its neutral
position.
Limit Value Indication
If the measured value is less than or equal to 2 , the Limit LED
lights up green. If the measured value is greater than 2 , the Limit LED lights up red.
Resistances which do not demonstrate a stable value until after a
“settling in period” should not be measured with automatic polarity reversal. In this case, measurement with automatic polarity
reversal may lead to varying and/or inflated measurement values,
and thus to an ambiguous reading.
7.2Measurement with Manual Polarity Reversal
In order to determine whether or not test results are independent
of current flow direction, measurement can be performed manually in one direction of current flow (positive polarity). Measurement in the other direction of current flow (negative polarity) is
then conducted by switching the test probes.
Ð Start continuous measurement by briefly pressing the CONTIN./
ZERO key, the LED ➃ lights up during measurement.
The red or green Limit LED indicates whether or not the limit value
has been violated (see section 7.1).
Measurement takes up to 3 minutes, after which the instrument is
switched to the standby mode. If you want to stop measurement
prematurely, press the CONTIN./ZERO key once again, after which
the relay switches audibly. Measurement is ended after roughly 5
seconds.
Differing results indicate voltage at the device under test (e.g.
thermovoltage or unit voltages).
Measurement results can be distorted by parallel connected
impedances in load current circuits and by equalizing current,
especially in systems which make use of “overcurrent protection
devices” (previous neutralization) without an isolated protective
conductor. Resistances which change during measurement (e.g.
inductance), or a poor contact, can also cause distorted measurements.
Examples of resistances whose values may change during measurement include:
–Incandescent lamp resistance with changing values caused
by warming due to measuring current
– Resistances with a large inductive component
In order to assure unambiguous measurement results, causes of
error must be located and eliminated.
GMC-I Messtechnik GmbH17
7.3Taking Measurement Cables and Extension Cables into
+
COM
+
COM
SHIELD
GUARD
10M
R
ISO/RINS
10 M
Account (up to 5 ) – ZERO Function (Roffset)
Ohmic measurement cable resistance can be subtracted from the
measurement results automatically. Proceed as follows:
Ð Select function Rlo with the rotary selector switch.
Ð Short circuit the test probes at the ends of the measurement
cables.
Ð Press and hold the CONTIN./ZERO key.
Ð Additionally press the measuring range selection key
(10k–1T/100k–100M) until the pointer indicates “0”.
LED ZERO/FUSEÐ Release both keys.
During the course of subsequent low-resistance measurements,
cable resistance R
as long as the rotary selector switch is not turned to the OFF position. Perform low-resistance measurement as described above in
the preceding pages. The stored value is retained, even if the
instrument has meanwhile been in standby operation provided
that the rotary selector switch position has not been changed.
➅ lights up red permanently:
is taken into consideration, i.e. subtracted,
Offset
8Test Resistor for Insulation Measurement for
Checking the Insulation Measuring Instrument
According to section 5.3.1.2 of VDE 0105-100 (EN 50110-1), the
following applies: “These measuring instruments must be tested
before, and if applicable after use.”
The two outermost
be connected to each other internally via a 10 M test resistor to
this end.
The sum of test resistor and cable resistance (for both cables),
including test probes, amounts to 10 M ±5%. This value allows
for quick self-testing.
18GMC-I Messtechnik GmbH
metallic
jacks on the connection panel must
Ð Connect the measurement cables to the + and COM jacks.
Ð
Insert the test probes into the above described metallic jacks.
Ð Select the R
as well as the desired test voltage, e.g. R
Ð Press the start key and view the measurement results.
The pointer should remain frozen at 10 M for about 5 seconds.
When the Limit LED and the LED for measuring range
out, the measurement has been completed and the pointer
returns to its neutral position.
measuring function with the rotary switch,
iso/ins
iSO
100 V.
➀ both go
9Technical Data
2
134
5
1
2
4
5
3
Meas.
Scale /
Qty.
Standard
VDE 0413
R
INS
R
LO
U
AC/DC
U
BAT
1
Accuracy specified here is only achieved with the following optional accessory:
“shielded high-resistance measurement cable KS-C (article number Z541F)”.
2
Relative to scale length:
Scale 1: 83.13 mm
Scale 2: 75.05 mm
Scales 3 and 4: 67.02 mm
3
Within the identified range on the respective scale (nominal range of use)
4)
with ZERO balancing
5)
0.17 ... 2 : 20%
Measuring
Range
100 k ...
100 M
10 k ... 1 T
0 ... 5 0.17 ... 4
Nominal Range
of Use
100 k ...
10 M
100 k ...
100 G
4)
Nominal / Open-
Circuit Voltage
50 V /100 V:
1.25 U
ISO
250 V /500 V /
1000 V:
1.1 U
ISO
4 V < U0 < 6 V
0 ... 1000 V10 ... 1000 V———2.5%—
8 ... 12 V8.5 ... 12 V————
Nominal
Current I
N
Short-
Circuit Cur-
rent I
Intrinsic Uncertainty
K
under Reference
Conditions
2.5%
1 mA 5 mA
2.5%
Test current IN 200 mA2.5%
Measuring
2
Uncertainty
Overload
3
Capacity
30% of
measured value
1
5)
10%
measured value
of
1000 V AC/DC
TRMS
1000 V AC/DC
TRMS
1000 V AC/DC
TRMS
GMC-I Messtechnik GmbH19
Displays
!
Analog display
Measuring movement Moving-coil mechanism with core magnet
Scale length83.13 mm (longest scale)
Limit LEDLED lights up red to indicate an exceeded
limit value. LED lights up green to indicate
compliance with the limit value.
LEDLED lights up red to indicate the presence
of interference voltage (when instrument is
switched off), test voltage during insulation
measurement or residual voltage after
insulation testing (U > 50 V)
Analog display LEDs Description see page 5
Reference Conditions
Reference temperature
Relative humidity40 ... 75%
Measured quantity
frequency
Measured quantity
waveformSine, deviation between TRMS and recti-
Battery voltage9.5 V 0.1 V
Test resistor10 M1%
Normal position
of use30°
+23 C 3K
45 Hz ... 65 Hz
fied value < 1%
Electrical Safety
Standard
VDE regulationVDE 0411, part 1, 1994-03
Protection classII
Pollution degree2
Measuring category
Fuses
Fuse link
Electronic fusefor the protection of low-resistance mea-
CAT II 1000 V / CAT III 600 V / CAT IV 300 V
FF315mA/1000V, effective in all resistance
measuring ranges, 1 additional replacement fuse in the battery compartment
surement R
LO
Electromagnetic Compatibility (EMC)
Interference emission EN 61326-1:2006 class B
Interference immunity EN 61326-1:2006
Power Supply
Batteries8 ea. 1.5 V mignon cell (8 ea. size AA)
Nominal range of use
Battery test
Battery saver circuit
Service lifeFor R
Safety shutdown
(alkaline manganese per IEC LR14) or 8
rechargeable NiMH batteries (must be
recharged externally)
8.5 ... 12 V
Battery capacity display via the Bat. TEST key
Standby function
(1000 V / 1 M) and RLO with 20
INS
s on-time and 1 measurement each for a
duration of 5 s
– With batteries (alkaline manganese):
900 measurements
– With rechargeable batteries (2000 mAh):
850 measurements
If supply voltage is too low, the instrument
is switched off, or cannot be switched on.
When the rotary switch is set to the OFF
position, the instrument is completely disconnected from the batteries (after
approximately 10 seconds).
20GMC-I Messtechnik GmbH
Ambient Conditions
Accuracy temp. range
Operating temperature
Storage temp. range –25 ... +70 C (without batteries)
Relative humidityUp to 75% (max. 85% during storage/
ElevationMax. 2000 m
Calibration interval1 year (recommended)
0 ... +40 C
–10 ... +50 C
transport), no condensation allowed
Mechanical Design
Dimensions225 x 130 x 140 mm
WeightApprox. 1.5 kg with batteries
ProtectionHousing: IP 52, measurement cables and
connectors: IP 40 per DIN VDE 0470,
part 1 / EN 60529, housing category 2
Excerpt from Table on the Meaning of IP Codes
IP XY
st
(1
digit X)
2 12.5 mm dia.2Dripping (at 15 angle)
3 2.5 mm dia.3Spraying water
4 1.0 mm dia.4Splashing water
5Dust protected5Jet-water
6Dust-proof6Powerful water jets
Protection Against
Foreign Object Entry
IP XY
(2nd digit Y)
Protection Against
Penetration by Water
GMC-I Messtechnik GmbH21
Display Values in Consideration of Measuring Uncertainty
100
Ux/V
R
x
/k
U
N
= 100 V
0
20
40
60
80
100
250
Ux/V
R
x
/k
U
N
= 250 V
0
50
100
150
200
250
Table for determining minimum display values for insulation resistance in consideration of the instrument’s measuring uncertainty:
Voltage at Device Under Test During Insulation Resistance
Measurement
Measuring voltage Ux at the device under test depending upon its
resistance R
and 1000 V:
at nominal voltage UN = 50 V, 100 V, 250 V, 500 V
x
Table for determining maximum display values for low-value resistance in consideration of the instrument’s measuring uncertainty.
Limit Value
0.15 0.11
0.20
0.16 5.00 4.72
0.50 0.44 10.0 9.47
1.00 0.92 20.0 17.7
2.00 1.87 50.0 44.7
22GMC-I Messtechnik GmbH
Maximum
Display Value
Limit Value
Maximum
Display Value
10List of Abbreviations and their Meanings
0
100
200
300
400
500
500
UN = 500 V
Ux/V
R
x
/k
0
200
400
600
800
1000
1
UN = 1000 V
Ux/V
R
x
/M
Voltage
U
ISO
U AC/DC
U
BAT
Current
I
N
I
K
Resistance
LIMITLimit value for insulation resistance or low-resistance
Offset Correction value for measurement cable resistance
R
ISO
R
LO
Test voltage or nominal voltage
Measured voltage (sinusoidal alternating voltage)
Operating voltage
Nominal current (insulation resistance measurement)
Test current (low-resistance measurement)
Short-circuit current (insulation resistance measurement)
When only one filled segment remains in the battery symbol,
install a new set of batteries or charge the rechargeable batteries.
Check to make sure that no leakage has occurred at batteries or
rechargeable batteries at short, regular intervals, or after the
instrument has been in storage for a lengthy period of time.
We recommend removing the batteries during lengthy
periods of non-use (e.g. vacation). This prevents excessive battery depletion or leakage, which may result in
damage to the instrument under unfavorable conditions.
If leakage has occurred, the electrolyte must be carefully and
completely removed with a damp cloth before installing new batteries.
Replacing the Batteries
See section 3.2 and section 11.3.
11.2Fuses
If a mechanical or electronic fuse has blown due to faulty operation or overload depending on the measuring function, the ZERO/FUSE LED blinks. The instrument’s voltage measuring ranges are
nevertheless still functional.
11.2.1 Fuse Link – ZERO/FUSE LED
This fuse is active in all resistance measuring ranges except for
voltage measurement. A replacement fuse is included in the battery compartment (FF315mA/1000V).
Disconnect the instrument from the measuring circuit before opening the battery compartment lid in order to replace the fuse (refer to page 5 for location)!
Checking the Fuse
If a resistance measuring range is selected with the rotary switch
and measurement is started with a blown or defective fuse in the
instrument, the ZERO/FUSE LED blinks. Prerequisite: The + and
COM measurement jacks are not short circuited.
After eliminating the cause of error and replacing the defective
fuse, the
again been switched on from rotary switch position OFF.
ZERO/FUSE LED goes out after the instrument has once
Incorrect fuses may cause severe damage to the instrument.
Only original fuses from GMC-I Messtechnik GmbH assure the required protection by means of suitable blowing characteristics.
Short-circuiting of fuse terminals or the repair of fuses is
prohibited!
The instrument may be damaged if fuses with incorrect
ampere ratings, breaking capacities or blowing characteristics are used!
24GMC-I Messtechnik GmbH
Replacing the Fuse
Pb Cd Hg
Ð Open the battery compartment lid by loosening the two
screws.
Ð Remove the blown fuse and insert a new one. A replacement
fuse is included in the battery compartment.
Ð Insert the new fuse.
Ð Replace the battery compartment lid and retighten the
screws.
11.2.2 Electronic Fuse – ZERO/FUSE LED
This fuse only protects low-resistance measurement (Rlo) from
overloading (electronic hardware circuit).
As the electronic fuse blows faster than the mechanical fuse in the
event of faulty operation of the instrument with live components,
not only the instrument, but also the fuse link is protected in the
low-resistance measuring range.
This means that, if interference voltage occurs during low-resistance measurement, the electronic fuse is blown. The measurement is aborted and the ZERO/FUSE LED blinks.
After eliminating the cause of error, the ZERO/FUSE LED goes out
after the instrument has once again been switched on from rotary
switch position OFF.
11.3Housing
No special maintenance is required for the housing. Keep outside
surfaces clean. Use a slightly dampened cloth or a plastic cleaner
for cleaning. Avoid the use of cleansers, abrasives or solvents.
Return and Environmentally Sound Disposal
The instrument is a category 9 product (monitoring and control
instrument) in accordance with ElektroG (German electrical and
electronic device law). This device is subject to the RoHS directive. Furthermore, we make reference to the fact that the current
status in this regard can be accessed on the Internet at
www.gossenmetrawatt.com by entering the search term WEEE.
We identify our electrical and electronic devices in
accordance with WEEE 2012/19EU and ElektroG using
the symbol shown at the right per DIN EN 50419.
These devices may not be disposed of with the trash.
Please contact our service department regarding the return of old
devices (see address in section 14).
If the (rechargeable) batteries used in your instrument are depleted,
they must be disposed of properly in accordance with valid
national regulations.
Batteries may contain pollutants and heavy metals such as lead
(Pb), cadmium (Cd) and mercury (Hg).
The symbol to the right indicates that batteries must not
be disposed of with the trash, and must be brought to a
designated collection point.
GMC-I Messtechnik GmbH25
12Recalibration
L1
L2
L3
N
PE
Between each active conductor and ground
L1
L2
L3
PEN
Between each active conductor and the PEN conductor
13Appendix
The measuring tasks performed with your instrument, and the
stressing it’s subjected to, influence aging of its components and
may result in deviation from the specified levels of accuracy.
In the case of strict measuring accuracy requirements, as well as
in the event of use at construction sites with frequent stress due
to transport and considerable temperature fluctuation, we recommend a relatively short calibration interval of once per year. If your
instrument is used primarily in the laboratory and indoors without
considerable climatic or mechanical stressing, a calibration interval of once every 2 to 3 years is sufficient as a rule.
During recalibration at an accredited calibration laboratory (DIN
EN ISO/IEC 17025), deviations from traceable standards demonstrated by your measuring instrument are documented. Ascertained deviations are used to correct display values during later
use of the instrument.
We would be happy to perform DAkkS or factory calibration for
you at our calibration laboratory. Further information is available at
our website:
www.gossenmetrawatt.com ( Company DAkkS Calibration
Center or FAQs Questions and Answers Regarding Calibration).
Recalibration of your instrument at regular intervals is essential for
the fulfillment of requirements according to quality management
systems per DIN EN ISO 9001.
* Examination of the specification, as well as adjustment, are not included in calibra-
tion. However, in the case of our own products, any required adjustment is performed and adherence to the specification is confirmed.
13.1Sample Connection Layouts for Insulation Resistance
Measurement
Insulation Resistance Measurement per DIN VDE 0100, Part 600
26GMC-I Messtechnik GmbH
L1
L2
L3
N
PE
Between each active conductor (phase and neutral
conductors) and ground
L1
L2
L3
PEN
Between each active conductor and the PEN conductor
TN–S
• With or without consumer
•N – PE, separately
TN–C
• With or without consumer
• N – PE, together
Insulation Resistance Measurement in Different Types of Systems
GMC-I Messtechnik GmbH27
TT
• With or without consumer
3-phase current (switched at all poles)
Caution:
– Open the overcurrent protective device.
– Disconnect the N conductor.
– Jumper the L and N conductors.
– Insulation measurement between L conductors and N to
PE.
In the distributor: individual and combined measurements
28GMC-I Messtechnik GmbH
13.2Attaching the Test Probe Holder to the Carrying Strap
1
Bottom Right
Detach the strap from the instrument:
Turn out the slotted screw (M3) at the
bottom.
Bottom Left
Strap
Clasp
Clasp
Front View
2
Slide the strap through the test probe
holder.
Eyelet for Attachment to Tester
Side View
(METRISO INTRO, BASE, TECH, PRO, XTRA)
GMC-I Messtechnik GmbH29
13.3Technical Data for Measurement Cables
Attention!
!
3
Front
Feed the strap through from the front
of the test instrument and secure it
with the slotted screw (M3).
Back
(scope of delivery: KS17-4 safety cable set)
Electrical Safety
Maximum rated voltage600 V
Measuring category
Max. rated current: 1 A1 A16 A
With safety cap attached
Without safety cap——
CAT IVCAT IIICAT II
1000 V
Ambient Conditions (EN 61010-031)
Tem pe rat ur e– 20 C ... + 50 C
Relative humidityMax. 80%
Pollution degree
2
Using the KS17-4
1000 V
—
Measurements per DIN EN 61010-031 may only be performed in environments in accordance with measuring
categories III and IV with the safety cap attached to the
test probe at the end of the measurement cable.
In order to establish contact inside 4 mm jacks, the safety caps
have to be removed by prying open the snap fastener with a
pointed object (e.g. the other test probe).
30GMC-I Messtechnik GmbH
Observe the instrument’s maximum values for electrical safety.
13.4Optional Accessories (not included)
ISO Calibrator 1 (material no. M662A)
Calibration adapter for testing the accuracy of instruments used for measuring insulation resistance and
low-resistance for test voltages of up to 1000 V (per
VDE 0413, parts 1, 2, 4 and 10).
KS-C(material no. Z541F)
Cable set consisting of measurement cable and
shielded high-resistance measurement cable
surements in the G-
KS24(material no. GTZ3201000R0001)
Cable set consisting of a 4 m long extension cable with
a permanently attached test probe at one end and a
contact protected socket at the other end, and 2 alligator clips which can be plugged onto the test probe
1081 Probe (material no. GTZ3196000R0001)
Triangular probe for floor measurements
per EN 1081,
lation)
Z550A(material no. Z550A)
Optional plug-on measurement cable with measurement key on the test probe and an additional key for
illuminating the measuring point, including shielded,
plug-in connector cable and test probe holder for
attachment to the carrying strap.
This address is only valid in Germany. Please contact our representatives or subsidiaries for service in other countries.
* DAkkS calibration laboratory for electrical quantities, registration no.
D-K-15080-01-01, accredited per DIN EN ISO/IEC 17025:2005
Accredited quantities: direct voltage, direct current value, direct current resistance,
alternating voltage, alternating current value, AC active power, AC apparent power, DC
power, capacitance, frequency and temperature
GMC-I Messtechnik GmbH31
Competent Partner
GMC-I Messtechnik GmbH is certified in accordance with
DIN EN ISO 9001:2008.
Our DAkkS calibration laboratory is accredited by the Deutsche
Akkreditierungsstelle GmbH (national accreditation body for the
Federal Republic of Germany) under registration number D-K15080-01-01 in accordance with DIN EN ISO/IEC 17025:2005.
We offer a complete range of expertise in the field of metrology:
from test reports and factory calibration certificates right on up to
DAkkS calibration certificates.
Our spectrum of offerings is rounded out with free test equipment management.
An on-site DAkkS calibration station is an integral part of our service
department. If errors are discovered during calibration, our specialized personnel are capable of completing repairs using original
replacement parts.
As a full service calibration laboratory, we can calibrate instruments from other manufacturers as well.
Edited in Germany Subject to change without notice PDF version available on the Internet