High-Precision
Insulation, Low Resistance and Voltage Measurement Instrument
• Insulation measurement per EN 61557-2/VDE 0413, part 2
• Low-resistance measurement
• Intelligent filter: precise and measurement-dependent
activation for the measurement of very high resistances
• Digital and analog display, backlit
• Indication of dangerous contact voltage LED
• Acoustic signalling when limit value is exceeded
• Detection of interference voltage in switch position OFF*
• Overvoltage protection
Protects the instrument in the event of inadvertent connection
to mains power
– Fuse link for all resistance measuring ranges
– Electronic fuse for the protection of low resistance
and resistance measurement R
• Compact and rugged for service calls under harsh conditions
METRISO INTRO/TECH:
Voltage testing and measurement up to 1000 V
METRISO BASE/TECH:
One measuring point self-test with test resistance of 10 M
per IEC/HD 60364-6 / EN 50110
per EN 61557-4/VDE 0413, part 4
and R
LO
3-349-810-03
1/12.14
German
Accreditation Body
D-K-15080-01-01
Applications
METRISO INTRO/BASE/TECH insulation and resistance measuring instruments allow for quick and effective testing of protective
measures in accordance with DIN VDE 0100, ÖVE-EN 1 (Austria),
NIV/NIN SEV 1000
countries as well. The instruments are equipped with a microprocessor and comply with IEC/EN 61557 / VDE 0413 regulations:
Part 1: General requirements
Part 2: Insulation resistance measuring instruments
Part 4: Instruments for measuring resistance at earthing conduc-
tors, protective conductors and equipotential bonding
Part 10: Combined measuring equipment for testing, measuring
or monitoring protective measures
As well as requirements per VDE 0701-0702:
Repair, modification and testing of electrical devices
The insulation measuring instruments are suitable for the following
tasks:
• Measurement of insulation resistance at voltage-free
devices and systems, up to 1000 V depending upon variant
• Testing of the resistance of earthing conductors, protective
conductors and equipotential bonding
• Checking of test objects for absence of voltage
• Testing of electrostatic discharge capacity at floor coverings
(using shielded measurement cables) – EN 1081
CAT II I 600 V / CAT IV 300 V
Fuses
Fuse linkFF315mA/1000V, effective in all resis-
tance measuring ranges, 1 additional
replacement fuse in the battery compartment
Elektronic fusefor protecting low-resistance and resis-
tance measurement R
(not METRISO G500MM (M550K))
and R
LO
Ambient Conditions
Accuracy
temperature range0 ... +40 C
Operating
temperature–10 ... +50 C
Storage temp. range –25 ... +70 C (without batteries)
Relative humidityUp to 75% (max. 85% during storage/
ElevationMax. 2000 m
Calibration interval1 year (recommended)
transport), no condensation allowed
Displays
Digital DisplayWith additional bar graphs or pointer
depending on selection with a.disp
parameter, backlit (transflective);
leading zeros can be suppressed at the
digital display depending on selection with
0.disp parameter;
overranging indicated with 0L at display;
dimensions: 65 x 36 mm
Cable resistance
If measurement results for the two directions of current flow (polarity reversal) differ
by more than 10% (this corresponds to
typical measuring error for the instruments), both measured values are displayed next to each other with reduced
resolution.
Limit LEDLED lights up red to indicate an exceeded
limit value
LED lights up green to indicate adherence
to the limit value
LEDLED lights up red to indicate:
at LCDDetection of external voltage at the LCD
– the presence of an external voltage
bevore insulation testing
(U > 50 V)
with the device switched on or off
(device switched off not with M550N)
– the presence of the test/measuring voltage
during (insulation) measurement (U > 50 V)
the presence of a residual voltage
after insulation testing (U > 50 V)
with the device switched on or off
with the device switched on where U DC >
50 V and U AC > 40 V (50 Hz) for all measuring functions
Mechanical Design
Electromagnetic Compatibility (EMC)
Interference emission EN 61326-1:2006 class B
Interference immunity EN 61326-1:2006
GMC-I Messtechnik GmbH3
Dimensions225 x 130 x 140 mm
WeightApprox. 1.4 kg with batteries
ProtectionHousing: IP 52, measurement cables and
connectors: IP 40 per DIN VDE 0470, part
1 / EN 60529, housing category 2
Extract from table on the meaning of IP codes
(1
IP XY
st
Protection Against Foreign
digit X)
2 12.5 mm dia.2Dripping (at 15° angle)
3 2.5 mm dia.3Spraying water
4 1.0 mm dia.4Splashing water
5Dust protected5Jet-water
6Dust-proof6Powerful water jets
Object Entry
IP XY
(2nd digit Y)
Protection Against
Penetration by Water
Page 4
METRISO INTRO, BASE, TECH
High-Precision
Insulation, Low Resistance and Voltage Measurement Instrument
Power Supply
Batteries8 ea. 1.5 V mignon cell (8 ea. size AA)
Nominal range of use 8.5 ... 12 V
Battery testBattery capacity display with battery sym-
Battery saver circuitAutomatic shutdown of display illumination
Service lifeFor R
Safety shutdownIf supply voltage is too low, the instrument
(alkaline manganese per IEC LR14)
bol in 4 segments: .
Querying of momentary battery voltage via
menu function.
after 15 second s (after the last time the
rotary switch is actuated) can be set via
the blight parameter.
The test instrument is automatically switched
to the standby mode* when the measured
value remains unchanged and none of the
controls are activated during this time.
* Specified time “APoFF” (entered in minutes)
adjustable via SETUP menu (default setting
approx. 10 min).
(1000 V / 1 M) and RLO with 20
INS
seconds on-time and 1 measurement
each for a duration of 5 seconds:
– With batteries (alkaline manganese):
900 measurements
– With rechargeable batteries (2200 mAh):
850 measurements
is switched off, or cannot be switched on.
When the rotary switch is set to the OFF
position, the instrument is completely disconnected from the batteries (after
approximately 10 seconds).
Scope of delivery
1 Insulation and resistance measuring instrument
1 DAkkS calibration certificate (not METRISO INTRO)
1 Set batteries (not METRISO INTRO)
1Carrying strap
1 Alligator clip (not METRISO INTRO)
1 KS17-4 cable set
1 Condensed operating instructions
1 Supplement Safety Information
1 Detailed operating instructions for download from our website
at www.gossenmetrawatt.com
Accessories (not included)
ISO Kalibrator 1
Calibration adapter for the
rapid, efficient testing of
the accuracy of measuring instruments for insulation resistanced and lowimpedance resistances.
Cable Set KS-C
Cable set consisting of
measurement cable and
high resistance measuring
cable, for measurements
in the G- range.
Applicable Regulations and Standards
IEC 61010-1 / EN 61 010-1/
VDE 0411-1
DIN EN 61557 / VDE0413Part 1:2007-12 General requirements
EN 1081Testing of electrostatic discharge capacity for floor cover-
EN 60529
VDE 0470, part 1
DIN EN 61326-1
VDE 0843-20-1
Safety requirements for electrical equipment for measurement, control and laboratory use
– General requirements
Part 2:2008-02Insulation resistance measuring
instruments
Part 4:2007-12Instruments for measuring
resistance at earthing conductors,
protective conductors
and equipotential bonding
Part 10: 2001-12 Combined measuring equipment
for testing, measuring or
monitoring protective measures
ings in potentially explosive atmospheres
Test instruments and test procedures
Degrees of protection provided by enclosures (IP code)
Electrical equipment for measurement, control and labo-
ratory use – EMC requirements –
Part 1: General requirements
Cable Set KS24
Cable set KS 24 consists
of a 4 m long extension
cable with a permanently
mounted test probe at one
end and a contact protected socket at the opposite end, as well as an alligator clip for plugging onto
the test probe.
4GMC-I Messtechnik GmbH
Page 5
METRISO INTRO, BASE, TECH
Measurement key
Key for
Measuring point illumination
measuring point illumination
Safety cap
Safety collar
Test pro be
Connection example: module slot for INTRO / BASE / TECH / PRO / XTRA
Movable contact protection securely covers the magnetic contact by eleastic force
magnetic
strain relief
High-Precision
Insulation, Low Resistance and Voltage Measurement Instrument
Telearm 1
Reel TR25
Test Probe for Remote Triggering Z550A
Floor Probe
The 1081 floor probe can
be used for measuring the
resistance of insulating
floors in accordance with
DIN VDE 0100 Part 600
and EN 1081.
The test probe with integrated control module allows for remote
triggering in areas with difficult access or in situations which
require your full attention. Poorly lit measuring points can be iluminated with the integrated test probe lighting. The connection
cable is shielded from interfering influences.
Magnetic measuring contacts (patent) with magnetic strain relief (Z502U)
Drum with Measurement Cable TR50
50m measurement cable
coilded around a metal drum.
Connection to one end of the
cable is accomplished with a
jack which is integrated into
the drum. The other end is
equipped with a banana plug.
The drum axle with handle can
be removed for space saving
storage.
Cable resistance component
can be compensated for in
selector switch position R
LO
.
GMC-I Messtechnik GmbH5
Page 6
METRISO INTRO, BASE, TECH
High-Precision
Insulation, Low Resistance and Voltage Measurement Instrument
Operating Case METRISO G (Z550C)
Order Information
DescriptionTypeArticle number
Insulation measuring instrument for DIN VDE 0100, ÖVE-EN 1 (Austria) and NIV/NIN SEV
1000 (Switzerland), complies with IEC/EN 61 557/VDE 0413, parts 1, 2, 4 and 10
Test voltages from 50 V to 1000 V,
voltage measurement to 1000 V,
including low-resistance measurementMETRISO TECHM550P
METRISO TECH
for Remote Triggering (Z550A) and
Operating Case (Z550C)
Test voltages from 50 V to 500 V,
voltage measurement to 500 V,
including low-resistance measurementMETRISO BASEM550O
METRISO BASE
for Remote Triggering (Z550A) and
Operating Case (Z550C)
inclusive Test Probe
METRISO TECH-SetM551P
inclusive Test Probe
METRISO BASE-SetM551O
DescriptionTypeArticle number
Test voltages from 250 V to 1000 V,
voltage measurement to 1000 V,
including low-resistance measurementMETRISO INTROM550N
METRISO INTRO
for Remote Triggering (Z550A) and
Operating Case (Z550C)
Accessories (not included)
Calibration adapter for testing the accuracy of instruments used for measuring
insulation resistance and low-resistance
for test voltages of up to 1000 V (per
VDE 0413, parts 1, 2 and 4).
Cable set consisting of measurement cable and shielded high-resistance measurement cable for measurements in the G rangeKS-CZ541F
Alligator
and KS-C
Cable set consisting of a 4 m long
extension cable with a permanently
attached test probe at one end and a
contact protected socket at the other
end, and 2 alligator clips which can
be plugged onto the test probeKS24GTZ3201000R0001
Triangular probe for floor measurements per EN 1081, DIN VDE 0100600 (Standing-Surface Insulation)1081 probeGTZ3196000R0001
Telescoping rod for PE measurement Telearm 1GTZ3232000R0001
Reel with 25 m measurement cable TR25 reelGTZ3303000R0001
Drum with 50 m measurement cable TR50 drumGTY1040014E34
Test probe with START/STOP keyand
an additional key for illuminating the
measuring point, including shielded
cable and test probe holder for carrying belt
Magnetic Measuring contacts with
contact protection – Set with magnetic holder, measurement contacts
5,5 mm in diameter insulated,
CAT III 1.000 V / 4 A, temperature
between –10 °C and 60 °C, under
standard conditions and flat-head
screws holding force 1.200 g vertical to contact area;
measuring instrument connector:
angled multilam plug according (for
METRISO G series)
Operating case for
BASE / TECH / PRO / XTRA with
outer bag for measuring leads
inclusive Test Probe
clips (1 pair) for KS17-4
METRISO
INTRO /
METRISO INTRO-SetM551N
ISO calibrator 1 M662A
KY-95-3Z110J
Test Probe
for Remote Triggering
METRISO GZ550A
Set 1 – Magnetic Measuring TipsZ502U
Operating Case
METRISO GZ550C
For additional information regarding accessories please refer to
• www.gossenmetrawatt.com
Edited in Germany • Subject to change without notice • PDF version available on the Internet