Gossen MetraWatt METRAVOLT 12D+L Operating Instructions Manual

Operating Instructions
METRAVOLT 12D+L
Voltage and Continuity Tester with Phase and Polarity Tester and Phase Sequence Indicator
3-349-592-02
4/4.12
Germany
Phone +49 911 8602-111 Fax +49 911 8602-777 E-Mail info@gossenmetrawatt.com www.gossenmetrawatt.com
2 GMC-I Messtechnik GmbH
1Test probes 2 Low voltage display (red LED):
as of 50 V AC / 120 V DC
3 LED (green): lights up for resistance
measurement from 0 to 1999 k (continuity is
indicated by means of a sound generator) 4 LEDs for phase sequence indication left/right 5 Polarity display (~ or –) 6 LCD panel for voltage and resistance 7 HOLD key: freezes the display value 8 /V and CAL key: shift key for resistance /
voltage measurement and zero balancing for
resistance measuring range 9 and test key : On/Off key (manual) and
function test
10 Handle bar
11 Connection cable
Meanings of symbols on the instrument:
Warning concerning a source of danger (attention: observe documentation!)
VDE testing authority approval mark
Indicates EC conformity
This device may not be disposed of with the trash. Further information regarding the WEEE mark can be accessed on the Internet at www.gossenmetrawatt.com by entering the search term ’WEEE’.
ED: On-time for voltage testing
1
3
2
5
6
8 9
10
4
7
11
!
GMC-I Messtechnik GmbH 3
Table of Contents Page
1 Applications ............................................... 3
2 Safety Precautions ..................................... 4
3 Initial Start-Up ............................................ 5
3.1 Battery ................................................................. 5
3.2 Testing for Correct Display and Function ................ 5
4 Measuring and Testing .............................. 6
4.1 General Instructions .............................................. 6
4.2 Testing Voltage and Polarity .................................. 6
4.2.1 Voltages of Less than 24 V .................................... 7
4.2.2 Voltages of 24 ... 1000 V AC / 1500 V DC ............. 7
4.2.3 Voltages of 1000 V AC ... 1200 V AC ..................... 7
4.2.4 Voltages greater than 1200 V AC/1500 V DC ......... 7
4.3 Testing of Phase and Phase Sequence ................... 7
4.3.1 Phase Test ........................................................... 8
4.3.2 Testing of Phase Sequence ................................... 8
4.4 Resistance and Continuity Testing .......................... 8
5 Characteristic Values ................................. 9
6 Maintenance – Recalibration ................... 10
6.1 Battery ............................................................... 10
6.2 Housing ............................................................. 11
6.3 Recalibration ...................................................... 11
6.4 Device Return and Environmentally Compatible
Disposal ............................................................. 12
7 Repair and Replacement Parts Service
Calibration Center
and Rental Instrument Service ................ 12
8 Product Support ....................................... 12
1 Applications
The METRAVOLT 12D+L is a 2-pole voltage tester with digital display, combined with an integrated continuity, phase and polarity tester, and a phase sequence indicator. It allows for safe detection and measurement of AC and DC voltages within a range of 24 to 1000 V at frequencies of up to 4 kHz. Resistance within a range of 0 to 1999 k can also be measured with the integrated continuity tester. Voltage and resistance values appear in digital format at the LCD. In addition to this, 3 LEDs indicate dangerous contact voltage and phase sequence, with another LED indicating resistance and a sound generator for continuity.
The METRAVOLT 12D+L can also be used in precipitation thanks to IP 65 protection.
4 GMC-I Messtechnik GmbH
2 Safety Precautions
You have selected an instrument which provides you with a high level of safety.
The METRAVOLT 12D+L voltage tester has been manufactured and tested in accordance with requirements set forth in DIN EN 61243-3 (VDE 0682 part 401) regarding 2-pole voltage testers. When used for its intended purpose, the safety of the operator, as well as that of the instrument, is assured.
In order to maintain flawless technical safety conditions, and to assure safe use, it is imperative that you read these operating instructions thoroughly and carefully before placing your instrument into service, and that you follow all instructions contained therein.
The instrument provides for excellent electrical safety by means of two series resistors immediately downstream from each of the two test probes as well as two completely autonomous test systems:
Low voltage test display which reliably indi­cates any existing voltage by means of one LED, even without batteries.
Digital display for exact test values.
The METRAVOLT 12D+L voltage tester has been approved by the VDE test authority for application of the VDE GS mark.
Observe the following safety precautions:
The instrument has a nominal voltage range of 24 to 1000 V AC/1500 V DC. The instrument may only be used in electrical systems whose nominal voltage lies within this range.
The faultless functioning of the display is only guaranteed in the temperature range from – 10 °C to + 55 °C.
Touch the instrument by the handles only during testing, and never touch the test probes.
The voltage tester is equipped with IP65 protection, and can be used in wet weather conditions (protected against jet water streams from all directions). However, powerful water jets, flooding and submersion or immersion must be avoided.
If one of the displays fails during the self-test described in chapter 3 – even if only partial failure occurs – or if no display appears indicating that the instrument is ready for use, the voltage tester may not be placed into operation.
The voltage tester may not be dismantled by un­authorized persons (see chapter 7 on page 12).
The voltage tester has a maximum allowable on-time of 2 minutes for voltage tests.
Voltage testers must be kept dry and clean. The housing can be cleaned with isopropyl alcohol or soapy water.