1LCD
2 Battery compartment cover
3 mA, A connector jack for direct measurement of current to “max. 10 A”
4 A connector jack for current clamp measurement to “max. 30 V”
5 Connector jack for all measuring ranges except current measuring ranges
6“”connector jack for all measuring ranges
7 OFF/ON: ON/OFF switch
8 Resettable miniature circuit breaker (AUTO FUSE)
9 Rotary switch for function selection
10 Eyelet for carrying strap
11
MAN/AUTO
12 DATA and MIN/MAX: Key for measured value storage
: Key for manual and automatic measuring range selection
13 FUNC: Multifunction key
2 GMC-I Messtechnik GmbH
Digital Display Symbols
134
8
10
11131418
19
56
2
15
7
16
17
21
20
12
9
1 Continuous operation
2 Battery voltage display
3 Digital display with indication of decimal place and polarity
4 Manual measuring range selection
5 Display memory, “freeze measured value”
6 MIN-MAX storage
7 ZERO: zero balancing active
8 Selected current type, DC
9 USB Interface display
(when communication is active is displayed)
10 Unit of measure
11 Over-ranging
12 RPM measurement: Upm1/Upm2 (at 2/4 stroke engines)
13 Pointer for analog display
14 Analog display scale
15 Resistance thermometer: Pt100 / Pt1000
16 Thermocouple: type K
17 Current clamp measurement active
18 Negative analog display range exceeded
19 Transformer ratio (clip factor)
20 Acoustic signal activated (e.g. for continuity testing)
21 Diode measurement
Calibration Center* and Rental Instrument Service ...............................................28
19Product Support ....................................................................................................28
4 GMC-I Messtechnik GmbH
1Safety Features and Precautions
You have selected an instrument which provides you with a high level of safety.
This instrument fulfills the requirements of the applicable EU guidelines and
national regulations. We confirm this with the CE marking. The relevant declaration of conformity can be obtained from GMC-I Messtechnik GmbH.
The analog/digital multimeter has been manufactured and tested in accordance
with safety regulations IEC 61010–1:2010/DIN EN 61010–1:2011/VDE 0411–
1:2011. If used for its intended purpose, safety of the operator, as well as that of
the instrument, is assured. However, safety cannot be guaranteed if the instrument is used improperly or handled carelessly.
In order to maintain flawless technical safety condition, and to assure safe use, it is
imperative that you read the operating instructions thoroughly and carefully before placing
your instrument into service, and that you follow all instructions contained therein.
For your safety, as well as for protection of your multimeter, the instrument is
equipped with a miniature circuit breaker for the 10 A current measuring range.
Observe the following safety precautions
•The instrument may only be operated by persons who are capable of recognizing contact hazards and implementing appropriate safety precautions.
Contact hazards exist anywhere, where voltages of greater than 33 V RMS
may occur.
•Avoid working alone when taking measurements which involve contact
hazards. Be certain that a second person is present.
•Maximum allowable voltage between terminals (3), (4), (5), (6) and ground is equal to
300 V, category II.
•The current measuring range A is equipped with a magnetic circuit breaker.
Max. allowable voltage at the meas. circuit (= miniature circuit breaker nominal
voltage) is equal to 240 V~ (AC) and 50 V(DC) against earth in the “mA”
and “A” ranges.
•The instrument may only be used for current measurement in power systems if the
electrical circuit is protected with a fuse or a circuit breaker with a rating of up to
20 A, and if nom. voltage at the system does not exceed 240 V~ (AC) or 50 V (DC).
In order to conform to the CAT requirements, an additional slow-blowing fuse link
(T16A/500V) has been fitted in series with the miniature circuit breaker which can
only be replaced by service personnel once it is tripped.
•Be prepared for the occurrence of unexpected voltages at devices under test
(e.g. defective devices). For ex., capacitors may be dangerously charged.
•Make certain that the measurement cables are in flawless condition, e.g. no
damage to insulation, no interruptions in cables or plugs etc.
•No measurements may be made with this instrument in electrical circuits
with corona discharge (high-voltage).
•Special care is required when measurements are made in HF electrical
circuits. Dangerous pulsating voltages may be present.
•Measurements under moist ambient conditions are not permitted.
GMC-I Messtechnik GmbH 5
•Be absolutely certain that the measuring ranges are not overloaded beyond
!
Serial number
Registration number
Date of calibraion (year – month)
German Accrediation Body GmbH – Calibration lab
XY123
2012-08
D-K-
15080-01-01
their allowable capacities. Limit values can be found in the “Measuring
Ranges” table in chapter 15, “Characteristic Values”.
Meaning of symbols on the instrument
Warning concerning a point of danger
(Attention: observe documentation!)
Earth
Continuous doubled or reinforced insulation
CAT II Measurement category I I device
Indicates CE conformity
The device may not be disposed of with the trash.
Further information regarding the WEEE mark can be accessed on
the Internet at www.gossenmetrawatt.com by entering the search
term WEEE.
Calibration mark (blue seal):
Repairs, Parts Replacement and Balancing
When the instrument is opened, voltage conducting parts may be exposed. The
instrument must be disconnected from the measuring circuit before repair,
replacement of parts or balancing. If repair or balancing of a live, open instrument
is required, this may only be carried out by trained personnel who are familiar
with the dangers involved.
Defects and Extraordinary Strains
If it may be assumed that the instrument can no longer be operated safely, it
must be removed from service and secured against unintentional use.
Safe operation can no longer be relied upon:
•If the instrument demonstrates visible damage
•If the instrument no longer functions
•After long periods of storage under unfavorable conditions, e.g. humidity,
dust, excessive temperature (see “Ambient Conditions” on page 25).
6 GMC-I Messtechnik GmbH
2Initial Start-Up
Attention!
!
Note!
Note!
Installing Batteries
Disconnect the instrument from the measuring circuit at all poles before
opening the battery compartment!
➭ Fold the instrument closed.
➭ Insert a coin or a similar object into the slot between the housing and the
battery compartment cover and press down until the battery compartment
cover snaps open.
➭ Fold the instrument open as far as it will go and remove the battery
compartment cover.
➭ Insert two 1.5 V mignon batteries per IEC R6 or IEC LR6 into the battery
compartment, making sure they are poled in accordance with the symbols.
➭ Replace the battery compartment cover and press into position until it
audibly snaps into place.
Switching the Instrument On
➭ Set the toggle switch to the “ON” position.
Activation of the instrument is acknowledged with an acoustic signal.
If the instrument has been shut down automatically, press either the FUNC, DATA or MAN key to switch the multimeter back on, or set the toggle switch to the “OFF”
position and leave it there for at least 5 s before returning it to the “ON” position.
Electrical discharges and high frequency interference may lead to erroneous display values, and may disable the measuring sequence. Reset
the instrument by switching it off and then, after 5 s, back on again.
Switching the Instrument Off Manually
➭
Set the toggle switch to the “OFF” position and fold the instrument closed. The
battery is automatically disconnected when the multimeter is folded closed.
Automatic Shutdown (standby)
The instrument switches itself off automatically if the measured value remains
constant for a long period of time (maximum measured value fluctuation approx.
0.8% of the measuring range per minute, or 1 Celsius or 1 Fahrenheit per minute), and if no keys are activated for a period of approximately 10 minutes. Shutdown is acknowledged by a brief acoustic signal. Exception: continuous operation mode.
Electrical power is supplied to the processor even after the instrument
has been switched off automatically. A closed-circuit current of about
200 A remains. The instrument is only disconnected from the batteries
after it has been switched off manually with the toggle switch or with the
instrument folded.
GMC-I Messtechnik GmbH 7
Disabling Automatic Shutdown
The instrument can be switched to “CONTINUOUS ON”.
➭ Press the FUNC key while switching the instrument on with the toggle switch
until an acoustic signal is generated. The “CONTINUOUS ON” function
is indicated at the display with the symbol.
3Selecting Measuring Functions and Measuring Ranges
3.1Automatic Measuring Range Selection
The multimeter is equipped with auto-ranging for all measuring ranges except for
temperature measurement, diode and continuity testing. Auto-ranging is always
activated as soon as the instrument is switched on. Depending upon the measured quantity, the device automatically selects the measuring range which
allows for best possible resolution.
The instrument automatically switches to the next highest or next lowest
measuring range for the following measured quantities:
Measuring RangesResolution
V , V , A , mA ,
A, , 30 mF, Hz,
30 nF 300 F 3 ¾3 100280
Switching to the next
highest range occurs
at(... d + 1 d)
4 ¾31 0002 800
Switching to the next
lowest range occurs
at (... d –1 d)
1)
3.2Manual Measuring Range Selection – MAN/AUTO Key
Auto-ranging can be deactivated and ranges can be set manually in accordance
with the following table.
The manual mode is deactivated by pressing and holding the MAN/AUTO key for
about 1 s, by turning the rotary switch or by switching the instrument off and
back on again.
Key
MAN/AUTO
briefManual mode active: utilized measuring range is fixedMAN1 x
Switching sequence for:
V: 300 mV 3V 30 V 300 V 600 V 300 mV
A: 300 A 3mA 30 mA 300 mA 3A 10 A
brief
longReturn to automatic measuring range selection—2 x
If you wish to perform quicker measurements than those possible with the automatic measuring range selection function, make sure to establish the appropriate
measuring range:
•by manual measuring range selection, i. e. by selecting the measuring range
with the best resolution, see chapter 3.2.
or
•via DATA function, see chapter 5. After the first measurement, the proper
measuring range will be automatically determined so that measurements are
performed more rapidly from the second measured value onwards.
With both functions, the established measuring range is maintained for the subsequent series mode measurments.
4Display (LCD)
4.1Display Illumination
After the instrument has been switched on, background illumination can be activated by briefly pressing the DATA/MIN/MAX and MAN/AUTO keys simultane-
ously. Illumination is switched back off by once again pressing the same keys, or
automatically after approximately 1 minute.
4.2Digital Display
The measured value appears at the digital display with decimal point and sign (+
or -). The selected unit of measure and the type of current are displayed as well.
A minus sign is displayed to the left of the value for the measurement of zerofrequency quantities if the positive pole of the measured quantity is applied to the
jack.
If the measuring range upper limit is exceeded for the following measured
quantities, “OL” (overload) is displayed:
V (DC), I (DC), Hz, V~ (AC), I~ (AC), 30 mF: 30999 digits
300 F: 3099 digits
30 nF
The digital display is refreshed at different rates for the individual measured
quantities (see “Display Refresh Rates” on page 24.
4.3Analog Display
The analog display with simulated pointer and dynamic response equivalent to
that of a moving-coil mechanism, is refreshed 20 times per second. This display
is especially advantageous for the observation of measured value fluctuations,
and during balancing.
The analog display is equipped with a polarity indicator. The scale is extended 5
segments into the negative range for the measurement of zero-frequency
quantities, allowing for precise observation of measured value fluctuation around
the zero point. If the measured value exceeds the display range, a triangle is first
display at the left, and polarity at the display is reversed approximately 0.7
seconds later. If the measuring range is exceeded (> 30999 digits, in the F range
(except 30 mF) : > 3099) a triangle is displayed at the right.
Scaling at the analog display is automatic, which is quite helpful for manual
measuring range selection.
GMC-I Messtechnik GmbH 9
5Measured Value Storage – DATA / MIN-MAX Key
5.1DATA (hold / compare)
Measured values can be automatically frozen at the display with the DATA (hold)
function. This is especially useful when full attention is required for contacting the
measuring point. After the measured value has been applied to the multimeter
and the “condition” indicated in the following table has been fulfilled, the measured value is frozen at the digital display and an acoustic signal is generated.
The test probes can then be removed from the measuring point, and the measured value can be read from the digital display. If the measured value is less
than the limit value shown in the table, the instrument is reactivated for a new
measurement with the DATA hold function.
If the new measured value deviates from the previous value by less than 100 digits, the acoustic signal sounds twice (DATA compare).
Function
DATA
Activatebriefblinks1 x
Hold
Reactivate
Deactivatelo ng
1)
Reactivation occurs if the indicated measured value limits are fallen short of.
2)
Double acoustic signal after the first measured value has been stored.
Thereafter, double acoustic signal only occurs if the currently frozen value deviates from the
first stored value by less than 100 digits.
3)
Exception: 10% at 300
Key
DATA
1)
ConditionReaction at Instrument
, ,
, ,
Measured Value
Limits
(digits)
> 3.3% of MR
3)
OL
3)
> 3.3%
of range
< 3.3% of MR
3)
OL
< 3.3% 3)
of MR
Value, Digital
Measuring
Ranges
V, A, F, Hz, %
V, A, F, Hz, %
Display
Measured
displayed
frozen
measured
value
clearediscleared
DATA
is
displayed
blinks
is
Acoustic
Signal
1 x
is
2)
2 x
2 x
Key: MR = measuring range
The DATA function has no effect on the analog display, which continues to
indicate the current measured value. However, it must be observed that the
decimal place is fixed when the digital display is “frozen”.
The DATA function can be deactivated by pressing and holding the DATA key for
about 1 second, by turning the function selector switch or by switching the
instrument off and back on again.
10 GMC-I Messtechnik GmbH
5.2Storage of Minimum and Maximum Values “MIN/MAX” with Time Stamp
Minimum and maximum measured values which occur at the input of the measuring instrument after the MIN/MAX function has been activated can be saved
to memory. The most important application for this function is the determination
of minimum and maximum values during long-term observation of measured
quantities (same function as the slave pointer at an analog display).
The “MIN/MAX” function can be activated for all measuring ranges.
The MIN/MAX function has no effect on the analog display, which continues to
indicate the current measured value.
Apply the measured quantity to the instrument and select the measuring range
before activating the MIN/MAX function.
Measuring ranges can only be selected manually after the function has been
activated, and stored MIN/MAX values and time stamps are deleted in doing so.
The MIN/MAX function is deactivated by pressing and holding the DATA key for
about 1 second, by turning the function selector switch or by switching the
instrument off and back on again.
Function
MIN-MAX
1.
Activate and
Save and
Return to 1.
Deactivatelongare deletedis deletedis cleared2 x
GMC-I Messtechnik GmbH 11
Save
2.
Display
3.
2 x briefare saved
MIN and MAX
Key
Measured Values /
DATA
Time Stamps
brief
brief
Storage continues
in background,
briefMIN and hh:mm1 x
new MIN and MAX
briefstored MAX valueMAX1 x
values and time
stamps are displayed
brief
briefMAX and hh:mm1 x
brief
same as 1.,
stored values are not
deleted
Measured Value,
current measured
stored MIN valueMIN1 x
time elapsed to
stored MIN value
time elapsed to
stored MAX value
Reaction at Instrument
Display
Digital
value
same as 1.same as 1.1 x
MIN MAX
MIN and MAX
blink
MIN and h:mm:ss1 x
MAX and h:mm:ss1 x
Acoustic
Signal
2 x
6Voltage and Frequency Measurement
Note!
Attention!
!
– (+)
~
+ (–)
~
➭ Set the rotary switch to either V (TRMS) or V , depending upon the
voltage to be measured.
➭ You can switch back and forth between voltage and frequency
measurement in the V
➭ Connect the measurement cables as shown. The jack must be grounded.
An intermittent acoustic signal warns the operator if the measured value
exceeds the measuring range upper limit in the 600 V range.
Make certain that neither of the current measuring range is selected
(“A”) and that the measuring cables are connected to the right “V” und
“” jacks when the multimeter is connected for voltage measurement!
Both the operator and the instrument are in danger if the breaking
thresholds for the fuses are exceeded due to operator error!
switch position by pressing the FUNC key.
Zero Balancing in the
300 mV Measuring Range
➭ Select the 300 mV
measuring range.
➭ Connect the measure-
ment cables to the instrument and connect the
free ends to one another.
➭ Briefly press the FUNC key.
The instrument acknowledges
zero balancing with an
acoustic signal, and “000
(1 digit) and the ZERO
symbol appear at the LCD.
The voltage displayed at the moment the key was pressed is used as a reference
value (max. 20 00 digits which
corresponds to 20 mV).
This value is automatically subtracted from all subsequently measured values.
➭ Zero balancing can be cleared:
– by pressing and holding the FUNC key,
after which clearing is acknowledged by
– a twice repeated acoustic signal,
by switching the instrument off.
.00”
12 GMC-I Messtechnik GmbH
6.1Low-Voltage Measurement
– (+)~+ (–)
~
The instrument is equipped with a special 30 mV measuring range for measuring
voltage drop at fuses which is distinguished by high resolution (10 V) with a low
input resistance of 50 k.
➭ Set the rotary switch to “Temp RTD”.
➭ Select “V DC” measurement with probe by repeatedly pressing the FUNC
key until “mV DC” appears at the display.
➭ Connect
7Current Measurement
➭ First disconnect supply power from the measuring circuit or the power
➭ Depending upon the type of current, set the rotary switch to “A ” or “A ”.
➭
➭ Connect the measuring instrument securely to the power consumer in se-
Notes Concerning Current Measurement:
•The instrument may only be
•Set up the measuring circuit
•In the A measuring ranges, an intermittent acoustic signal generates a warn-
•All measuring ranges up to 10 A are protected by a resetable, 15 A/
•If a fuse or circuit breaker in the active current measuring range is tripped,
theprobetothe“” and “V” sockets.
consumer and discharge all capacitors, if any are present.
The symbol for the selected current type, (DC) or ~ (AC), appears at the LCD.
ries as shown in the diagram (without transition resistor).
used in power installations if
the electrical circuit is protected with a fuse or a circuit
breaker with a rating of up to
20 A, and if nominal voltage at
the system does not exceed
240 V~ (AC) or 50 V (DC).
in a mechanically secure
fashion such that it cannot
be inadvertently interrupted.
Use conductors with an
adequate cross section and
connectors of adequate
size in order to prevent
excessive warming.
ing if the measured value exceeds 10 A.
240 V AC/50 V DC auto-fuse. In order to conform to the CAT requirements,
an additional slow-blowing fuse link (T16A/500V) has been fitted in series
with the miniature circuit breaker which can only be replaced by service personnel once it is tripped.
FUSE appears at the digital display, and an acoustic signal is generated
simultaneously.
GMC-I Messtechnik GmbH 13
•Eliminate the cause of overload after the fuse or breaker is tripped before
Attention!
!
Note!
placing the instrument back into service!
If the secondary side of the current transformer with voltage output
remains open during operation, e.g. if cables are defective or have not
been connected, or due to a blown device fuse or incorrect
connection, dangerous voltages may occur at the terminals.
Motors with high starting current cause tripping of the circuit breaker,
except for measurement with clamp meters.
7.1Current Measurement with Current (Clamp) Sensors with Voltage Output
If a current sensor is connected to the multimeter, all current values are displayed
correctly in consideration of the transformer ratio. This presupposes that the current sensor is equipped with the required sensitivity, and that the
appropriate ratio is selected before measurement is performed.
➭ Turn the rotary switch to the position
➭ Select a current type,
or Upm2 (see below)
➭ Simultaneously press the FUNC and the MAN/AUTO keys. The currently
selected transformation ratio is displayed. The transformer ratio can be
changed by pressing the MAN or the DATA key, or the currently selected value
can be retained by pressing the FUNC key.
➭ Connect the current (clamp) sensor to the and the jacks.
A (DC) , A
, by pressing the FUNC key.
.
( ) (AC+DC), Hz, RPM Upm1
14 GMC-I Messtechnik GmbH
Please observe the specified operating conditions per IEC/EN 61010-2-32
M
I(A)
V/~
* for short-circuited measuring cables: residual value 1 ... 70 d at zero point due to TRMS converter
Current sensor
transformer ratio
max. measuring range
Measuring ranges
available in the multimeter
A A~ *
1/ 1 V / A
depending on the applied
current sensor
0 ... 300.00 mA/3.000 A/30.00 A
1 / 10 V / A0 ... 3.0000 A/30.000 A/300.00 A
1 / 100 V / A0 ... 30.000 A/300.00 A/3.0000 kA
1 / 1000V / A0 ... 300.00 A/3.0000 kA/30.000 kA
regarding measurement category, etc. for the applied current sensor.
Additional error caused by the current sensor must be taken into consideration.
7.2RPM Measurement at combustion engines
RPM is measured by acquiring pulses. The number of measurable pulses per
revolution varies depending upon the type of engine (2 or 4 stroke).
➭ Set the rotary switch to “”.
➭ Press the multifunction key (
(RPM measurement at 2 stroke engines: 1 pulse per revolution) or Upm2 (RPM
measurement at 4 stroke engines: 1 pulse for 2 revolutions)
The measured value then appears, for example “244.3
FUNC) repeatedly until unit of measure
is briefly displayed.
Upm
Upm1
”.
GMC-I Messtechnik GmbH 15
8Resistance Measurement
Note!
R
x
Voltage Drop
➭ Set the rotary switch to the
position. Overload is indicated
if no device under test has been
connected: “0.L M”. ➭ Before connecting the device under
test, make sure that it is voltage-
free. Interference voltages distort
measurement results! Perform a
voltage test first if required.
➭ Connect the device under test as
shown in the diagram.
Zero Balancing in the 30 , 300 and 3 kMeasuring Ranges
Resistance at cables, as well as contact resistances, can be eliminated for the
measurement of low-resistance values in the 30 , 300 and 3 k ranges by
means of zero balancing:
➭ Connect the measurement cables to the instrument and connect the free
ends to one another (short circuit the test probes).
➭ Briefly press the FUNC key.
The instrument acknowledges zero balancing with an acoustic signal, and
.00”, “ 000.00 ” or “0.0000 k”and the ZERO symbol appear at the
“00
LCD.
Resistance measured at the moment the key is pressed is used as a
reference value (max. 2000 digits). This value is automatically subtracted
from all subsequently measured values.
➭ Zero balancing can be cleared:
–by pressing and holding the FUNC key, after which clearing is acknowledged
by a twice repeated acoustic signal,
–by switching the instrument off.
9Continuity Testing
With the “acoustic signal” function activated, and exclusively in the 0 to 310
measuring range, the instrument generates a continuous acoustic tone for
measured resistance within a range of 0 to approximately 2 .➭ Turn the selector switch to the position. The and symbols appear at
the LCD.
➭ Connect the measurement cables to the device under test.
Continuity testing is very fast (< 50 ms) and is suitable for locating
connections with poor contact (e.g. due to vibration) in automotive
service applications.
16 GMC-I Messtechnik GmbH
10Diode Testing
Note!
Conducting
Direction
Blocking
Direction
(max. 5.1 V)
➭ Turn the selector switch to the position. Overload is indicated if no
device under test has been connected: “.0L V”.
➭ Make sure that the device under test is voltage-free. Interference voltages
distort measurement results! Perform a voltage test first if required.
➭ Connect the device under test as shown in the diagram.
Conducting Direction or Short-Circuiting
Conducting state voltage is displayed in volts at the measuring instrument. As
long as voltage drop does not exceed the maximum allowed display value of
5.1 V, several elements or reference diodes with minimal reference voltage can
be connected in series f or testing. If “.0L ” appears at the display, either the circuit
is interrupted or conducting state voltage is greater than 5.1 V.
Blocking Direction or Interruption
“.0L” is displayed at the instrument. If a value of less than 5.1 V is displayed, this
generally indicates that the diode’s blocking direction is defective.
Resistors and semiconductor paths which have been connected in
parallel to the diode distort measurement results!
GMC-I Messtechnik GmbH 17
11Capacitance Measurment
+
–
Duty Cycle (%) = ——————— • 100
Pulse Duration
Period
➭ Make sure that the device
under test is voltage-free.
Interference voltages distort
measurement results!
➭ Turn the rotary switch to the
“F” position.
➭ Connect the (discharged!)
device under test to the
and V jacks with the
measurement cables.
Zero Balancing in the 30 nF Measuring Range
Intrinsic capacitance at the measuring instrument and cables can be eliminated
for the measurement of small capacitance values in the 30 nF range by means of
zero balancing:
➭ Connect the measurement cables to the measuring instrument without a
device under test.
➭ Briefly press the FUNC key.
The instrument acknowledges zero balancing with an acoustic signal and
.00 ” and the ZERO symbol appear at the LCD. Capacitance measured
“00
at the moment the key is pressed is used as a reference value (max.
2000 digits). This value is automatically subtracted from all subsequently
measured values.
➭ Zero balancing can be cleared:
–by pressing and holding the FUNC key, after which clearing is acknowledged
by an acoustic signal,
–by switching the instrument off.
12Frequency Measurement – Duty Cycle Measurement
➭ Turn the rotary switch to the Hz position.
➭ Apply the measured quantity as described under voltage measurement.
➭ Smallest measurable frequencies and maximum allowable voltages are
listed in chapter 15, “Characteristic Values”.
The pulse-period ratio can be ascertained for square-wave signals with the duty
cycle measurement.
➭ Briefly press the multifunction key twice (FUNC). The instrument is switched
to duty cycle measurement. The duty cycle, i.e. the pulse duration of a sig-
nal as a percentage, is displayed at the LCD.
Note
The applied frequency must remain constant during duty cycle measurement.
18 GMC-I Messtechnik GmbH
13Temperature Measurement with Pt100 and Pt1000
Note!
Note!
Z3409
➭ Set the rotary switch to
“Temp RTD”.
➭ Connect the P
the and V jacks. The
instrument automatically
recognizes type of
connected sensor (Pt100
or Pt1000), and displays
measured temperature in
the selected unit of
measure.
Temperature Measurement
with Compensation for Sensor Cable Resistances from 0.1 to 50
Sensor cable resistances with values other than100 m can be compensated
for values of up to 50 as follows:
➭ Briefly press the FUNC and the MAN/AUTO keys simultaneously. The currently
selected cable resistance is displayed. This value can be increased with the
DATA key, or reduced with the MAN/AUTO key. Each time one of these keys is
briefly activated, the value is changed by 10 digits (0.1 ). The value can be
changed by means of rapid scrolling by pressing and holding the key.
➭ Switch back to temperature measurement by briefly acknowledging with the
FUNC key.
The new cable resistance value is retained, even after the instrument has been
switched off.
t sensor to
The characteristic cable resistance for the temperature sensors which
are available as accessories is automatically taken into consideration for
this measurement.
The default setting is Pt100
14Temperature Measurement with Type K Thermocouple
➭ Set the rotary switch to “Temp RTD”.
➭ Select “Temp TC K” measurement with thermocouple by repeatedly press-
ing the FUNC key until C or F appears at the display.
Select temperature unit of measure °C or °F by long pressing the FUNC key.
➭
➭ Connect the sensor to the “” and “V” sockets.
/Pt1000
with a cable resistance of 0.1 .
GMC-I Messtechnik GmbH 19
15Characteristic Values
Measuring
Function
V DC
Hz
C/F
1)
2)
3)
4)
Key: d = digit(s), rdg. = reading, MR = measuring range, MUL = measuring range upper limit
Measuring Range
30 mV10 V50k—
300 mV 10 V> 11 M11 M //<50pF
3 V 100 V11M11 M //<50pF
30 V 1 mV10 M10 M //<50pF
V
300 V 10 mV10 M10 M //<50pF
3)
600 V
300 A 10 nA160 mV
3 mA100 nA160 mV
30 mA1 A180 mV
A
300 mA10 A250 mV
3 A100 A360 mV
10 A1 mA920 mV
30 10 m1.3 Vmax. 250 A
300 10 m1.3 Vmax. 250 A
3k100 m1.3 Vmax. 150 A
30 k1 1.3 Vmax. 30 A
300 k10 1.3 Vmax. 3 A
3M100 1.3 Vmax. 0.36 A
30 M1k1.3 Vmax. 0.1 A
300 0.1 max. 8.4 VI
1)
5.1 V
30 nF10 pF10 M0.7 V
300 nF100 pF1 M0.7 V
3 F1 nF100 k0.7 V
F
30 F10nF11k0.7 V
300 F100 nF3 k0.7 V
300.00 Hz0.01 Hz1 Hz
3.0000 kHz0.1 Hz1 Hz
4)
30.000 kHz1 Hz1 Hz
300..00 kHz10 Hz1 Hz
1.0000 MHz100 Hz1 Hz
15...300 Hz: 2.0... 98.0%
%
... 10 kHz: 10.0... 90.0% 0.1 Hz0.1 Hz1 Hz
– 200.0 ... +850.0 CPt100
– 250.0 ... + 1372.0 CK / NiCr-Ni
Up to max. 5.1 V diode voltage, “OL” (overload) is displayed for higher values.
Smallest measurable frequency for sinusoidal measuring signal symmetric to zero point.
corresponds to 600 V CAT I
Input sensitivity, signal/sine: Hz (V): 10 to 100% MR except for mV: as of 30% MR;
H (I): 20 to 100% MR except for 3 A: as of 30% MR; Hz (clip): as of 30% MR
Resolution at MULInput Impedance
30 0003 000
100 mV10 M10 M //<50pF
1 mVmax. 8.4 VIk = 1 mA
0.1 Hz0.1 Hz1 Hz
0.1 C– 150.0 ... + 850.0 CPt1000
Approximate Voltage Drop at MUL
Open-Circuit Voltage
Discharge Resistance
2)
f
min
~
Meas. Cur. at MUL
= 1 mA
k
U
0 max
Power Limit
3 x 106 V x Hz
3 x 106 V x Hz... 3 kHz: 5.0... 95.0% 0.1 Hz0.1 Hz1 Hz
20 GMC-I Messtechnik GmbH
Measuring
Intrinsic Uncertainty under Reference Conditions
( % rdg. + d) ( % rdg. + d)
Range
30 mV 1+5—
300 mV 0.2 + 5
3 V0.2+30.5+30
30 V 0.2 + 30.5 + 30
300 V 0.2 + 30.5 + 30
600 V 0.2 + 30.5 + 30600 V CAT I
300 A0.5+51.5+30
3 mA0.5+51.5+30
30 mA0.5 + 51.5 + 30
300 mA0.5 + 51.5 + 30
3 A0.7+51.5+30
10 A0.7 + 51.5 + 30
30 1+5
300 0.2+5
3k0.2+5
30 k0.2+5
300 k0.2+5
3M0.2+5
30 M2+10
4) 7
4)
4)
2) 6)
~
1+30
2) 6
~
[to 1 k: (0.2 + 9 D)]
Overload Capacity
Valu eDura tion
600 V
(DC)
~ (AC)
eff sine
0.36 A
10 A
300 V
(DC)
~ (AC)
eff
Sinus
3
continuous
dauernd
max. 10 s
3+5
5.1 V0.5 + 3
30 nF1 + 6
300 nF1 + 6
3 F1 + 6
30 F1 + 6
300 F1 + 6
300.00 Hz
3.0000 kHz300 V
30 kHz300 V
300 kHz100 V
1000 kHz30
Pt 100– 200.0 ... + 850.0 C0.5% + 15
K / NiCr-Ni – 250.0 ... + 1372.0 C1% + 5 K
1)
2)
3)
4)
5)
6)
7)
Intrinsic error values valid as of 10 digits
0.1 + 5
(sinusoidal input
voltage> 2 ... 5)
0.1% MR 8 d
%
0.1% MR/kHz 8 d
At 0 ° to + 40 C
Values of less than 2 mV are suppressed in the 300 mV range
15 (20) 45 ... 65 Hz 10 kHz sine, see page 22 for influences.
After measurement with 10 A: at least 10 minute cool-down period
If “zero balancing” function is active, ZERO appears at display.
Plus sensor error
Specified intrinsic uncertainty valid for 3 to
with short-circuited test probes: residual value 1 to 30 d at zero point due to TRMS converter
4)
Max. Measuring Voltage
300 V
100% of the AC measuring ranges
5)
5)
5)
300 V
(DC)
~ (AC)
eff
sine
300 Vmax. 10 s
300 Vmax. 10 s0.1% MR/kHz 8 d
300 V (DC) /
~(AC)
eff sine
max. 10 s
max. 10 sPt1000– 150.0 ... +850.0 C0.5% + 15
1)
GMC-I Messtechnik GmbH 21
Influence Quantities and Influence Error
Influence
Quantity
Temperature
Sphere
of Influence
0 C
+21 C
and
+25 C
+40 C
Measured Quantity /
Measuring Range
300 A 300 mA
3 A / 10 A + ~1 + 10
300 300 k0.2 + 10
30 nF 30 F0.5 + 10
Influence Error
1)
( % + d) / 10 K
V0.2 + 10
~0.4 + 10
V
+ ~
0.5 + 10
3M0.2 + 10
30 M1 + 10
Hz / %0.5 + 10
C (Pt100)0.5 + 10
Influence
Quantity
Sphere
of Influence
(max. resolution)
Frequency
Intrinsic Uncertainty
( % rdg. + d)
2)
> 15 Hz ... 45 Hz3 + 30
> 65 Hz ... 1 kHz1.5 + 30
> 1 kHz ... 5 kHz2.5 + 30
> 5 kHz ... 10 kHz3 + 30
> 15 Hz ... 45 Hz3 + 30
> 65 Hz ... 10 kHz3 + 30
Frequency
Intrinsic Uncertainty 2)
( % rdg. + d)
> 15 Hz ... 45 Hz
> 65 Hz ... 1 kHz
3 + 30
> 15 Hz ... 45 Hz3 + 30
> 65 Hz ... 500 Hz1.5 + 30
> 500 Hz ... 1 kHz3 + 30
Frequency
V~
(AC)
Influence
Quantity
Frequency
I~
(AC)
3.0000 V
30.000 V
300.00 mV
300.00 V
600.0 V
Sphere of
Influence
(max. resolution)
300.00 A
3.0000 mA,
30.000 mA
10.000 A
300.00 mA
> 15 Hz ... 45 Hz3 + 30
3.0000 A
> 65 Hz ... 500 Hz1.5 + 30
> 500 Hz ... 1 kHz3 + 30
1)
With zero balancing
2)
Indicated error applies as of a display value of 10% of the measuring range.
22 GMC-I Messtechnik GmbH
Influence
1
2
3
4
5
CF
0
Voltage and Current
Measurement
Allowable crest factor CF for the
periodic quantity to be measured
depends upon the displayed value:
Digits
1000020 00030000
Quantity
Measured
Quantity
Waveshape
Sphere of
Influence
Crest
factor
CF
3)
Measured Quantity /
Measuring Range
1 2
>2 4 5% rdg.
>4 5 7% rdg.
V ~, A ~
Influence Error
1% rdg.
1)
Influence
Quantity
Relative
Humidity
Influence
Quantity
Common-
Mode
Interference
Volta ge
Series-Mode
Interference
Volta ge
1)
Except for sinusoidal waveshape
Reference Conditions
Ambient Temperature+23 C 3K
Relative Humidity40
Measured Qty. Frequency 45
Measured Qty. Waveshape sinusoidal
Battery Voltage3 V 0.1 V
GMC-I Messtechnik GmbH 23
Sphere of
Influence
75%
3 days
device off
interference quantity max. 300 VV> 90 dB
interference quantity max. 300 V ~
nom. value, respective measuring range,
max. 300 V ~, 50 Hz, 60 Hz sine
interference quantity max. 300 V
nom. value, respective measuring range
Measured Quantity /
Measuring Range
V, A
F, Hz, %
Sphere of Influence
50 Hz, 60 Hz sine
interference quantity V ~
75%
65 Hz
Influence Error
C
1 x intrinsic uncertainty
Measuring
Range
300 mV
30 V ~
300 V ~> 60 dB
V> 40 dB
V ~> 50 dB
Damping
> 60 dB
Response Time (after manual range selection)
Measured Quantity /
Measuring Range
V , V ~,
A, A~
30 3M2s
30 M5s
Continuity< 50 ms
C (Pt100/Pt1000)max. 3 s
30 nF 300 Fmax. 2 s
>10 Hzmax. 1.5 s
Digital Display
Response Time
1.5 s
1.5 s
Measured Quantity
Jump Function
from 0 to 80%
of the measuring range upper limit
from to 50%
of the measuring range upper limit
from 0 to 50%
of the measuring range upper limit
Display
LCD window (95 mm x 40 mm) with analog and digital display, including display
of unit of measure, current type and various special functions.
Ty pe
Background Illumination
Background illumination (by means of LEDs) is activated with two keys, and is
COG (chip on glass) for good legibility from various directions
switched off automatically after approximately 1 minute.
Analog:
DisplayLCD scale with pointer
Scale length80 mm for Vand A;
Scaling5
Polarity Displaywith automatic switching
Overload Displaytriangle appears
Sampling Rate20 measurements per second
Digital:
Display/Char. Height 7 segment characters / 20 mm
Number of Places4¾-place31000 steps
Overload Display“0L” appears
Polarity Display“–” is displayed if plus pole is connected to jack
Sampling Rate2 measurements per second
Display Refresh Rates
V (DC), V~ (AC), A, , ,
C (Pt100, Pt1000)twice per second
67 mm for all other ranges
0 30 with 35 scale graduations for ,
0 ... 30 with 30 scale graduations for all other ranges
Hzonce per second
Power Supply
Batteries2 ea. 1.5 V cells (4 x AA size)
alkaline-manganese per IEC LR6
24 GMC-I Messtechnik GmbH
Service Lifewith alkaline-manganese batteries: approx. 100 hours
Battery TestBattery capacity display with battery symbol in
Battery Saving Circuit
The instrument is switched off automatically
4 segments: “”
–if the measured value remains unchanged for a period of approximately 10
minutes, and if none of the control elements are operated during this period.
Automatic shutdown can be disabled.
–If battery voltage drops to below approx. 1.8 V
Fuses
300A to 10 A range
Electrical Safety
Safety ClassI I per IEC 61010-1:2010/VDE 0411-1:2011
Measurement Category
Operating Voltage300 V
Pollution degree2
Test Voltage2.3 kV~ per IEC 61010-1:2010/VDE 0411-1:2011
Electromagnetic Compatibility (EMC)
Interference Emission EN 61 326-1:2013 Class B
Interference Immunity EN 61326-1:2013
– Resettable 15A/240VAC/50VDC miniature circuit breaker,
– A fuse link is additionally connected in series to the
auto-fuse, the blowing or absence of which is detected
automatically (“FUSE” is displayed):
T16A/500VAC, 6,3 mm x 32 mm
Switching capacity 1.5 kA at 500 V AC and ohmic load
CAT II
EN 61326-2-1:2013
Ambient Conditions
Accuracy range0 C
Operating Temp.10 C
Storage Temperature 25 C
Relative Humiditymax. 75%, no condensation allowed
Elevationto 2000 m
Deploymentindoors, outdoors: only within the specified
Mechanical Design
Protectionhousing: IP 40, connector jacks: IP 20
Table Excerpt Regarding Significance of IP Codes
IP XY
st
(1
digit X)
Protection against
foreign object entry
+40C
+50C
+70C (without batteries)
ambient conditions
IP XY
(2nd digit Y)
Protection against the
penetration of water
2 12.5 mm dia.0not protected
4 1.0 mm dia.0not protected
Dimensions146 mm x 118 mm x 44 mm
Weightapprox. 450 gr. with batteries
GMC-I Messtechnik GmbH 25
16Maintenance
Attention!
!
Attention!
!
Disconnect the instrument from the measuring circuit before opening to
replace batteries or fuses!
16.1 Batteries
Make sure that no battery leakage has occurred before placing your instrument
into service for the first time, or after long periods of storage. Repeat this
inspection on a regular basis thereafter.
If battery leakage has occurred, clean the electrolyte from the device completely
and carefully with a dampened cloth and install new batteries before placing the
instrument back into service.
If the „“ symbol appears at the display, replace the batteries as soon as
possible. The instrument requires two 1.5 V batteries in accordance with IEC R 6
or IEC LR 6.
Replacing the Batteries
Disconnect the instrument from the measuring circuit at all poles before
opening the battery compartment!
➭ Fold the instrument closed.
➭ Insert a coin or a similar object into the slot between the housing and the
battery compartment cover and press down until the battery compartment
cover snaps open.
➭ Fold the instrument open as far as it will go and remove the battery
compartment cover.
➭ Insert two 1.5 V mignon batteries per IEC R6 or IEC LR6 into the battery
compartment, making sure they are poled in accordance with the symbols.
➭ Replace the battery compartment cover and press into position until it
audibly snaps into place.
➭ Please dispose of depleted batteries in an environmentally sound fashion!
16.2 Fuses
Eliminate the cause of overload after a fuse or breaker is tripped before placing
the instrument back into service!
10 A Range
A humming sound originating from the 15 A circuit breaker is normal when
measuring high value alternating current.
If the miniature 15 A circuit breaker for the active current measuring range is
tripped, FUSE appears at the digital display and an acoustic signal is generated
simultaneously. Make sure that the red pin is popped out in the tripped condition,
i.e. that no adhesion or fusion has taken place. Short circuit the and 15 A jacks
with the selector switch set to the continuity test position, and “0L” must appear
at the display.
26 GMC-I Messtechnik GmbH
Attention!
!
Inspect the measuring circuit and eliminate the cause of overload
Pb Cd Hg
before activating the reset button at the miniature circuit breaker.
The fuse link connected in series to the auto-fuse may only be replaced by
authorized service personnel.
16.3 Housing / Opening of Equipment / Repair
No special maintenance is required for the housing. Keep outside surfaces clean.
Use a slightly dampened cloth for cleaning. Avoid the use of cleansers, abrasives
or solvents.
The equipment may be opened only by authorized service personnel to ensure the safe
and correct operation of the equipment and to keep the warranty valid.
Even original spare parts may be installed only by authorized service personnel.
In case the equipment was opened by unauthorized personnel, no warranty
regarding personal safety, measurement accuracy, conformity with applicable
safety measures or any consequential damage is granted by the manufacturer.
16.4 Device Return and Environmentally Compatible Disposal
The instrument is a category 9 product (monitoring and control instrument) in
accordance with ElektroG (German Electrical and Electronic Device Law). This
device is subject to the RoHS directive. Furthermore, we make reference to the
fact that the current status in this regard can be accessed on the Internet at
www.gossenmetrawatt.com by entering the search term WEEE.
We identify our electrical and electronic devices in accordance with
WEEE 2012/19/EU and ElektroG with the symbol shown to the right
per DIN EN 50419.
These devices may not be disposed of with the trash. Please contact
our service department regarding the return of old devices.
If you use batteries or rechargeable batteries in your instrument or accessories
which no longer function properly, they must be duly disposed of in compliance
with the applicable national regulations.
Batteries or rechargeable batteries may contain harmful substances or heavy
metal such as lead (PB), cadmium (CD) or mercury (Hg).
They symbol shown to the right indicates that batteries or rechargeable
batteries may not be disposed of with the trash, but must be delivered
to collection points specially provided for this purpose.
17Multimeter Messages
MessageFunctionSignificance
FUSECurrent measurementBlown fuse or tripped miniature circuit breaker
OLIn all measuring modes Indicates ove rload
In all operating modesBattery voltage less than 2.3 V
This address is only valid in Germany.
Please contact our representatives or subsidiaries for service in other countries.
* DAkkS Calibration Laboratory for Electrical Quantities
D-K-15080-01-01 accredited per DIN EN ISO/IEC 17025
Accredited measured quantities: direct voltage, direct current values, DC
resistance, alternating voltage, alternating current values, AC active power, AC
apparent power, DC power, capacitance, frequency and temperature
If you order a DAkkS calibration certificate reprint for your instrument, please
provide us with the reference numbers indicated in the upper and lower most
fields of the calibration mark. We do not need the instrument’s serial number.
Edited in Germany • Subject to change without notice • A pdf version can be found on the internet